Jewell Instruments LSO Data sheet

LSOC/LSOP "L" 4-20 mA Series Inclinometer
Making Sense out of Motion…
Input Ranges From
±1.0°
High Resolution and Low
Non-Linearity
The Jewell
LSOC/LSOP "L" Series
4-20 mA
_
Output Flexure Suspension
Servo Fluid Damped Inclinometer is
designed for applications where high
levels of shock, vibration and electrical
noise are present and/or long cable
runs are required.
Features & Benefits
hysteresis of less than 0.0005% of full
range output.
stand shocks in excess of 1500g and
vibration of 20 grms.
small as 0.000006"/ft.
balanced sensor technology.
Low white noise spectral density
of better than 0.15
µV/Sq Root HZ
Applications
electric arc furnance and idle control
mining, tunneling, and overturn detection
Extremely high resolution and low
Extremely robust designed to with-
Responds to changes of slope as
High accuracy closed-loop force
Outline Diagram
Pin Out (Options C-connector, P-pin)
Steel processing: continous casting,
Heavy construction: paving, grading,
Structrural monitoring: walls of dams,
support, columns, bridges, and others
Railway: automated train controls, rail
leveling and grinding, and rail bed analysis
Military Applications: Measuring of
angular tilt where high levels of shock and vibration are present
Jewell Instruments LLC, 850 Perimeter Road, Manchester, NH 03103
sales@jewellinstruments.com www.jewellinstruments.com Tel (800) 227-5955
LSOC/LSOP "L" Series Inclinometer
Making Sense out of Motion…
Performance Specifications
STATIC/DYNAMIC
Input Range (°) (Note 1) ± 1.0 ± 3.0 ± 14.5 ± 30.0 ± 90.0
ELECTRICAL
Custom Capabilities
Internal temperature sensor and thermal modeling for the highest levels of accuracy over
a wide temperature range
Available in LSR Series package configuration for applications requiring a more compact solution
Factory set zero biasing for non-horizontal measurements
Solder terminals and flying leads in place of military connector
Custom input ranges from ±0.5° to ±90° available
Custom output impedances available
How to Order
Full Range Output (FRO), mADC ± 1.0% 4 to 20 4 to 20 4 to 20 4 to 20 4 to 20
Non Linearity (%FRO, Max.) (Note 2)
Scale Factor, Ma/G (Nominal) 458 152.8 31.3 16.0 8.0
Scale Factor Temp Sens (PPM/°C, Max.) 400 300 100 100 100
Natural Frequency, Hz (Nominal) (Note 3) 0.3 2.0 15 20 30
Bandwidth (-3dB), Hz (Nominal) 0.3 2.0 15 20 30
Transverse Axis Misalignment, ° Max. 0.15 0.25 0.50 1.00 1.00
Output at Zero Tilt, Ma 12 ± 0.6 12 ± 0.6 12 ± 0.3 12 ± 0.3 12 ± 0.3
Zero Tilt Temp Sens, mA/°C Max. 0.015 0.01 0.0032 0.0032 0.0016
Resolution and Threshold, μ rad Max. 1 1 1 1 1
Input Voltage, VDC 20 to 30 (24 Nominal) Operating Temp Range -18 to +71°C
Input Current (mA, Max.): 40 Survival Temp Range -40 to +71°C
Noise, mA rms Max. 0.01 Vibration 20 grms
Notes: 1 - Full range is defined as "from negative full input
angle to positive full input angle." The inclinometer output is proportional to the sine of the tilt angle.
2 - Referenced to theoretical sine value independent of misalignment.
3 - Output phase angle = -90°.
*Specifications subject to change without notice on account of continued product development
0.05 0.05 0.10 0.10 0.05
ENVIRONMENTAL
Shock 1500g, 0.5 msec, ½ sine
Seal MIL-STD 202, Method 112
LSOC-1L 02550278-506 LSOC-3L 02550278-507 LSOC-14.5L 02550278-508 LSOC-30L 02550278-509 LSOC-90L 02550278-510
Jewell Instruments LLC, 850 Perimeter Road, Manchester, NH 03103
sales@jewellinstruments.com www.jewellinstruments.com Tel (800) 227-5955