Jewell Instruments LCF-196 Specifications

Page 1
LCF-196 Biaxial Series Inclinometer
Making Sense out of Motion…
Biaxial Inclinometer is a
±14.5 to ±90 device, with
lower ranges available with customization.
The Jewell LCF 196 Series is a
Features & Benefits
• Very low hysteresis
• Excellent turn on repeatability
• High Accuracy Sensor
• 500g Shock Capability
• IP65 Sealed Housing
• Less > 0.02% Non-linearity
• Bias Temp Sens >50µg/°C
• Small diameter
Applications Pin Out (Options: C-connector, P-Pin)
• Geophysical Measurement
• Earth Movement Monitoring
• Oil & Gas Well Logging
• Dam Monitoring
• Heavy Construction, Grading
• Ship & Barge Leveling
• Deviation Surveys
• Continuous Casting
• Weapons Platform Leveling
• Data Logging
Jewell Instruments LLC , 850 Perimeter Road, Manchester, NH 03103
[email protected] • www.jewellinstruments.com • Tel (800) 227-5955
Page 2
LCF-196 Biaxial Series Inclinometer
Making Sense out of Motion…
Performance Specifications
STATIC/DYNAMIC
Input Range, ° ±14.5 ±30.0 ±90.0
Full Range Output (FRO) VDC ±1%¹
±5.0 ±5.0 ±5.0
Non Linearity (%FRO Max.)²
0.02 0.02 0.10
Scale Factor, Volts/g, Nominal
20.0 10.0 5.0
Scale Factor Temp. Sensitivity (SFTS), PPM /°C Maximum
100 100 100
Natural Frequency, Hz Nominal³
30.0 30.0 30.0
Bandwidth (-3dB), Hz Nominal
30.00 30.00 30.00
Input Axis Misalignment, ° Maximum
1.00 1.00 1.00
Bias, Volts, Maximum
0.040 0.020 0.020
0° Output Temp Sensitivity, Volts/°C, Maximum
0.001 0.0005 0.0003
Resolution and Threshold
ELECTRICAL
Input Voltage Range, (VDC)⁴
±12 to ±19
Input Current, mA, Nominal
15
Output Impedance, Ohms, Nominal
100
Noise, Vrms, Maximum
0.002 0.001 0.001
ENVIRONMENTAL
Operating Temp Range Survival Temp Range Vibration grms
10
Shock
500g, 0.001 sec
Seal
MIL-STD 202, Method 112
Weight
11.0 oz.
Notes:
1- Full Range is defined as "from negative full input angle to positive full input angle."
2 - Nonlinearity is specified as deviation of output referenced to theoretical sine function value, independent of misalignment.
3 - Output phase angle = -90°
4 - Unit Power connections can easily be adapted for operations from single-ended, floating power supplies of 24 to 34 Volts DC.
Custom Capabilities
Available in lower ranges with customization.
Single Axis lower cost until available.
LCF-196-14.5 475196-001
LCF-196-30 475196-002
LCF-196-90 475196-003
Jewell Instruments LLC , 850 Perimeter Road, Manchester, NH 03103
[email protected] • www.jewellinstruments.com • Tel (800) 227-5955
3 µradians
-40°C to +80°C
-60°C to +90°C
How to Order