iStat Portable Clinical Analyzer User manual

Portable Clinical Analyzer
& Test Cartridges
Manufacturer: i-STAT Corporation
104 Windsor Center Drive East Windsor, NJ 08520 • USA USA Tech Suport: (800) 366-8020 Fax: (609) 443-9310
Art: 714199-00D Rev. Date: 06/29/04
Emergo Europe P.O. Box 18510 2502 EM The Hague The Netherlands Tel: (31)70 345 8570 Fax: (31)70 346 7299
User Guide

System Components

Infrared
Light-
Emitting
Diode
Display
Screen
Soft
Keys
Key Pad
Cartridge Port
Power-on
Indicator
IR
Link
Cradle
Analyzer
Calibrant
(beneath
IR Link/Printer Cradle
Connected to Printer
Power Switch
Paper Advance
Solution
label)
Snap
Cover
AC Adapter
Port
Cartridge
Contact Pads
Sensors
Sample Well
IR Receiver
Cradle
IR Link
Status Light
Contact Pads
Plug to Central Data Station
Electronic Simulator
For in vitro diagnostic use.
See System Manual for instructions.
Intended Use: The i-STAT® Portable Clinical Analyzer is intended for use with i-STAT cartridges for the in vitro quantification of various analytes in whole blood.
© Copyright 2004. i-STAT Corporation. All Rights Reserved.

Quality Check Messages and Codes

Blood Collection

MESSAGE ACTION EXPLANATION
SIM Use Electronic Simulator Simulator must be used once every 24
BAT
SFT Update software Software will expire 15 days after SFT is
Date Invalid Check clock on Status Page. The software checks for an unrealistic date.
Dead Batteries Replace Batteries The analyzer does not have enough power
TEMPERATURE OUT OF RANGE
CARTRIDGE ERROR Use another cartridge
CARTRIDGE TYPE NOT RECOGNIZED
CARTRIDGE PREBURST Use another cartridge Calibrant released too soon. Take care not
SAMPLE POSITIONED SHORT OF FILL MARK
SAMPLE POSITIONED BEYOND FILL MARK
INSUFFICIENT SAMPLE Use another cartridge Cartridge is underfilled or bubbles trapped
UNABLE TO POSITION SAMPLE
CARTRIDGE NOT INSERTED PROPERLY
ANALYZER ERROR
ANALYZER ERROR
SYSTEM INTERRUPTED Use another cartridge Conditions such as batteries removed be-
Invalid or Expired CLEW Update software Software expired or corrupt. Verify that the
Replace batteries soon Approximately 50 more cartridges can
Check Status Page. Use another cartridge.
Use another cartridge The analyzer could not identify the type of
Use another cartridge Sample did not reach fill mark. Tilt
Use another cartridge Cartridge is overfilled. Fill new cartridge
Use another cartridge Inadequate seal; sample overflowed well;
Reinsert cartridge Push cartridge straight through door until it
Use Electronic Simulator If the analyzer passes the simulator check,
See System Manual See System Manual
hours, or as needed for regulatory compli­ance.
be tested.
first displayed.
Check Status Page. If date is correct, call Technical Services.
to complete the testing cycle.
The analyzer is too warm or too cool, or the room is too warm or too cool.
The analyzer detected a problem with the calibrant solution.
cartridge inserted. Check software version. Check to see if cartridges are expired.
to press over cartridge label area.
cartridge to aid sample flow into sample chamber.
to fill mark. Press on snap closure, and not over sample well when closing.
in sample.
sample clotted; snap closure left open.
will go no further.
continue to use. If not, call i-STAT Technical Services.
fore analyzer deactivated when cartridge is still in analyzer.
analyzer’s date is correct.
Acceptable Samples for Cartridges
Arterial: Plain syringe, heparinized syringe labeled for analytes to be
tested and filled to capacity, or syringe with minimum volume of heparin to prevent clotting (10 U/mL of blood). For ionized calcium, use balanced heparin syringes. Mix heparinized syringes by rolling between palms for at least 5 seconds in 2 directions, then invert the syringe repeatedly for at least 5 seconds. Test for lactate immediately. Test for other analytes within 10 minutes.
Avoid drawing air into syringes for blood gas and ionized calcium
tests.
If not tested immediately, remix and discard 2 drops of blood before
filling cartridge.
Do not use iced samples.
Venous: Collection tube with lithium or
sodium heparin filled to capacity and mixed by gentle inversion at least 10 times. Test within 10 minutes.
Do not leave tourniquet on for more
than 2 minutes.
Do not draw above an I.V.
Skin puncture: Lithium heparin capillary
tubes for testing all analytes but ionized calcium. For all analytes including ionized calcium, use plain or balanced heparin capillary tubes. Test immediately.
Allow alcohol to dry over puncture site before collecting sample.
Do not “milk” finger or heel while collecting sample.
Coagulation Tests:
The ACT test can be performed using venous or arterial samples,
while the PT/INR test can be performed using capillary or venous samples.
Use plain plastic syringes or plastic evacuated tubes with no
anticoagulant, activators, or serum separators.
Test sample immediately upon draw.
Some experts recommend drawing and discarding a sample of at
least 1 ml prior to drawing samples for coagulation testing.
If a second measurement is needed, draw a fresh sample.
Loading...
+ 5 hidden pages