Limitation of Liability.......................................................................................................1-6
Use of Third Party Computer Products ..........................................................................1-6
Iris Diagnostics Contact Information ..............................................................................1-7
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™(2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20051-1
1. Introduction
Intended Use
This Service Manual is intended for the
generation Automated Urine Microscopy Analyzer – HVL (High Volume
Laboratory).
There are two HVL configurations:
• The
• The
These instruments have a new fluidic design, the capability of connecting
Load and Unload stations, and improved error reporting.
The
vitro diagnostic use device composed of the
Microscopy Analyzer (
the AUTION MAX™ AX-4280 Automated Urine Chemistry Analyzer (AX-
4280), and a workstation. The
complete routine urinalysis profile, including urine test strip chemistry
panel, specific gravity, color, clarity, and microscopic analysis, providing
quantitative or qualitative counts of formed elements, such as cells, casts,
crystals, and organisms.
Optionally, the
as a stand-alone device for microscopic analysis, and the results from the
iQ
This manual is intended to be used by Iris Diagnostics trained personnel.
®
iQ
®
200 can be combined with other urine chemistry results.
®
iQ
200 Sprint and the
®
iQ
200 Sprint can process 101 samples per hour
iQ
200Sprint™ Automated Urinalysis System (
nd
200 2
generation can process 60 samples per hour
iQ
®
iQ
200), connected physically and electronically to
®
iQ
200 System is used to automate the
®
iQ
200 Automated Urine Microscopy Analyzer can be used
®
iQ
200 2
®
iQ
200 System) is an in-
®
200Automated Urine
nd
How to use this Manual
References are made to other sections within the Manual. In order to
avoid searching for a specific document or section, electronic hyperlinks
have been created to let you jump to other locations and to give you
immediate access to related information.
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™(2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20051-2
1. Introduction
To follow a link:
1. Select the hand tool
2. Links are indicated in the following manner:
• The text is colored blue and underlined. Example: Iris Diagnostics
Contact Information.
3. Position the pointer over the linked area on the page until the pointer
changes to the hand with a pointing finger . Then click the link.
4. To return to the
previous section, click
on the Go to Previous
View button.
5. The Bookmark pane,
located on the left side
of the screen, can be
used as a “linked” table
of contents. Click on
the + or - sign located
next to the title to
expand or collapse the
selection for a
bookmark. Clicking on
a bookmark will take you automatically to the starting page of the
selected section.
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™(2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20051-3
1. Introduction
Precautions and Warnings
The Manual includes information and warnings that must be observed by
the operator to ensure safe operation of the system. Important messages
are highlighted with borders and special icons identifying the type of
message enclosed.
There are four types of messages.
Notes
NOTE: Highlights important facts, gives helpful information and tips, and clarifies procedures.
Cautions
CAUTION: Electrical caution! Unplug before handling.
CAUTION: If any liquid is spilled onto the Sampler, wipe the area clean
before testing. Leaving the spill may cause crystals to form and block the
movement of the sample racks.
Warnings
WARNING: Identifies potentially hazardous situations that could result in
serious injury to laboratory personnel.
Biological Warning
WARNING: Wear protective gloves to prevent exposure to pathogens.
Discard contaminated materials according to applicable regulations.
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™(2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20051-4
1. Introduction
Warranty
Iris Diagnostics, a subsidiary of Iris International, (Iris) warrants that the
products manufactured by it, its divisions or subsidiaries and sold
hereunder shall be free from defects in material and/or workmanship,
under normal use and service, for the period expiring twelve (12) months
from the completion of installation under standard procedure by Iris
Diagnostics or an authorized Iris Diagnostics distributor, or upon
Purchaser’s signature on Iris Warranty/Acceptance form, or eighteen (18)
months from shipment, whichever occurs first. Iris Diagnostics makes no
warranty whatsoever regarding products manufactured by persons other
than Iris Diagnostics, its Divisions or Subsidiaries and Purchaser’s sole
source of warranty therefore, if any, is the original manufacturer’s
warranty.
No warranty extended by Iris Diagnostics shall apply to any products
which have been modified, altered, or repaired by persons others than
those authorized or approved by Iris or to products sold as “used.”
Iris Diagnostics’ obligation under this warranty is limited SOLELY to the
repair or replacement, at Iris Diagnostics’ option, of defective parts,
F.O.B. warehouse or local Iris office, or as otherwise specified by Iris.
Repairs or replacement deliveries shall not interrupt or prolong the term of
this warranty. Iris Diagnostics warranty does not apply to consumable
materials, except as specially stated in writing, not to products or parts
thereof manufactured by Purchaser.
This limited warranty is made on condition that immediate written notice
of any defect be given to Iris Diagnostics and that Iris inspection reveals
that the Purchaser’s claim is valid under the terms of this warranty.
Iris Diagnostics makes no warranty other than the one set forth herein or
that which may be provided in a separate warranty covering the
applicable product category. Such limited warranty is in lieu of all other
warranties, expressed or implied, including but not limited to any
expressed or implied warranty of merchantability or fitness for particular
purposes and such constitutes the only warranty made with respect to the
products.
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™(2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20051-5
1. Introduction
Limitation of Liability
Iris Diagnostics shall not be liable for any loss of use, revenue or
anticipated profits, or for any consequential or incidental damages
resulting from the sale or use of the products.
Use of Third Party Computer Products
Iris Diagnostics does not recommend that the workstation provided as a
®
iQ
functional part of the
®
iQ
200System be employed for performing any software or hardware-
based applications other than those specifically furnished to operate and
support the Iris instrument system, or those recommended and offered by
Iris Diagnostics specifically as accessories or enhancements for the Iris
instrument system. No other third party application software should be
installed in these microcomputers in addition to those provided or
recommended by Iris Diagnostics, without the expressed approval of Iris
Diagnostics Technical Services, in order to avoid potential performance
and reliability problems which can result from incompatibility factors,
errors in use of such software, or software-based “viruses.”
Installation of such third party software, or non-approved electronic cards
or other devices, without advance Iris Diagnostics approval may affect the
terms of or void any Iris Diagnostics warranty otherwise in effect, covering
Iris Diagnostics supplied software and hardware on the microcomputers
and the overall performance and reliability of the entire Iris Diagnostics
instrument system.
200Automated Urine Microscopy Analyzer or the
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™(2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20051-6
1. Introduction
Iris Diagnostics Contact Information
Iris Diagnostics
A Division of Iris International, Inc.
9172 Eton Avenue
Chatsworth, CA 91311
USA
Telephone:
From U.S. locations (800) PRO-IRIS (776-4747)
From outside the U.S. +1-818-709-1244
Fax: +1-(818) 700-9661
e-mail: sales@proiris.com
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™(2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20051-7
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-1
2. Installation Procedures
Installation Requirements
Space Requirements
®
iQ
The
0.9m) height countertop of approximately 36 inches wide by 24 inches
deep (91cm by 61cm), with vertical clearance of at least 24 inches
(61cm).
With optional Load and Unload stations, the width requirement is 60
inches (147 cm.)
Workstation
The Workstation is composed of a desktop computer, video monitor,
keyboard and mouse requiring a suitable counter or desktop, providing
comfortable access and a good viewing angle. It is suggested that, if
possible, the WorkStation be placed in a work area, which can be semidarkened to increase screen display visibility and reduce glare.
The data connections are made between the
(and the AX-4280 and the WorkStation, if applicable) via the special
cables provided.
200 is a tabletop unit requiring free space on a standard (36 inch:
®
iQ
200 and the Workstation
Location
For most laboratories, the
area which can be used for preparation of urine tubes and sample racks.
All data transmissions from the
routed to the Workstation, which manages and controls all
communications to the Laboratory Information System (LIS) via serial
connections.
Requirements
The
nor frequency parameters are customer configurable. Uninterruptible
power supplies are recommended and available from Iris Diagnostics for
the
system operation during short power outages and brownouts. This allows
for an orderly shutdown of instruments without the loss of data.
Installation
The
Diagnostics.
®
iQ
200 is placed near an open benchtop work
®
iQ
200 (and the Ax-4280, if used) are
®
iQ
200 uses 90 to 240V AC, 50-60Hz input source. Neither voltage
®
iQ
200 and the Workstation (and the AX-4280 if applicable) to maintain
®
iQ
200 will be installed by a factory-trained representative from Iris
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-2
2. Installation Procedures
Shipping Cartons
®
iQ
The
Instrument Carton
Item number Description Qty
800-3320
800-3322
700-3103 Printer
250-3082 Printer Cable 1
700-3120
210-3004
700-3101
700-3022
700-3008
700-3007
700-3781
700-3701
700-3702
250-3501
700-3011
200 is shipped in three cartons.
iQ
Computer Version 3.0
Flat Panel Monitor
Keyboard and Mouse
Sample Racks (1 Set of 9) 1
Calibrator Rack
Control/Focus Rack
iQ Lamina Cap & Filter
Drain Tubing
Lamina Tubing
Serial Cable, PC to Microscopy Module
Operators Manual CD
®
200 60/hour or
®
iQ
200 Sprint
TM
101/hour
1
1
1
1
1
1
1
1
1
1
1
1
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-3
2. Installation Procedures
Starter Kit A Carton
Item number Description Qty
475-0047
475-0021
475-0003
660-0036
800-3211
®
iQ
200 Lamina
Iris Diluent
Iris System Cleanser
Culture Tube, 16x100 250 pack
®
iQ
200 Dilution Labels Package
Starter Kit B Carton (needs refrigeration)
Item number Description Qty
800-3104
800-3103
®
iQ
200 Control/Focus Set (four bottles)
®
iQ
200 Calibrator (four bottles)
1
1
1
1
1
1
1
Optional
Item number Description Qty
800-3810
800-3514
800-3515
800-3501
Load and Unload Set
Load Station (only)
Unload Station (only)
Ax-4280 Bridge Connection Kit
1
1
1
1
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-4
2. Installation Procedures
Installation
To ensure safety, we advise you to read these installation instructions
very carefully and take all proper precautions.
Precautions
The entire unit weighs approximately 140 lbs (63.5k). The Load and
Unload Stations weigh 14 lbs (6.35) each. Choose a place to set up the
unit before completing its assembly.
If the unit must be moved, separate the Sampler from the analyzer before
moving. If these two units come apart while being carried, it may result in
injury or severe damage.
Always keep a distance of at least 2 inches (5 cm) between the rear of
the unit and the wall. If this distance is not maintained, the connecting
tubes and cables may overheat and / or be subjected to mechanical
stress.
Do NOT use power frequencies or voltage other than those specified in
this document. Connection to an inappropriate power source may
damage the instrument.
Make certain that the power supply for the
that provides power to no other instruments or appliances. If power is not
clean and steady, a UPS and/or power conditioner is recommended.
Do NOT disassemble or modify the unit. Doing so may cause injury
and/or instrument malfunction.
Place the unit on a stable and level surface free of vibration. Failure to do
so may cause injury or malfunction of the unit.
Do NOT place the unit where it may be affected by chemicals, corrosive gases
or electronic noise. Doing so may cause injury or malfunction of the unit.
Do NOT place the unit where it may be affected by water, direct sunlight
or draft. This may yield incorrect results, and the unit may be damaged.
Select a room to set up the unit where the temperature can be controlled
between 50
80%.
o
F (10oC) and 86oF (30oC), and humidity in a range of 20% to
®
iQ
200 is from a dedicated line
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-5
2. Installation Procedures
Unpacking the Instrument
Carefully remove the instrument from the shipping container. Remove all
packing material from inside the instrument surrounding the SPA, the
microscope and the waste well.
Connecting the Sampler
In order to protect the Analyzer and Sampler from damage during
transport, each section is secured with tape. Follow the procedure below
before installing the unit.
1. Remove the left side panel of
the analyzer and the EMI
shielding panel.
2. Remove the four screws
securing the STM SIC and
diaphragm pumps.
3. Connect the grey ribbon cable
from the sampler to J01 on
PCBA 101-5014 STM SIC.
4. Connect the multi-colored
wiring cable to J02 on PCBA
101-5014 STM SIC.
5. Connect the black specimen tube detector cable to J9 on PCBA 1015014 STM SIC.
6. Connect the ground cables to the chassis of the microscopy module.
7. Insert the hooks of the sampler into the slots of the analyzer. Be
careful not to catch the cables.
8. Reattach the STM SIC and diaphragm pumps to the chassis.
Ground
Adjust the Height of the Analyzer
• Using a 7/16” wrench, loosen the
locking nuts located under the
analyzer.
• Adjust the height of the Sampler by
rotating the feet. The space between
the bench top and the analyzer must
be ¾ inch.
• After adjusting the height, tighten the
locking nuts.
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-6
¾ inch
2. Installation Procedures
Adjust Sampler
Two adjustment feet are located under the Sampler. They are accessed
from the top of the Sampler. Each foot must be in contact with the
benchtop.
• Using a flat head
screwdriver, remove the two
rubber caps located on the
front corners of the
Sampler. The top of the
feet (screws) can be seen.
• Using a Phillips screwdriver
(+), turn each of the two
screws clockwise until the
adjustment feet touch the
bench top.
• Reinsert the rubber caps.
CAUTION: Make sure the height of the Sampler is adjusted correctly.
Otherwise, the Sampler and the Probe may be damaged.
Installing the Lamina Container
• Install the Lamina cap with
filter on a fresh Lamina
container.
• Connect one end of the
Lamina tubing to the top
fitting on the right side of
the analyzer. Connect the
other end to the fitting of
the Lamina cap.
Connecting the Drain Tubing
• Connect the drain tubing to the bottom fitting on the right side of
the analyzer.
• The Drain Tubing can be directed to the designated waste area
according to local regulations.
• If direct drainage is not an option, slide the Drain Tubing inside a
waste container (approximately 2 in. (5cm)) Make sure the waste
container is lower than the microscopy module.
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-7
2. Installation Procedures
Computer Connections
1. Connect the video monitor cable to the video monitor and the video port.
2. Connect the keyboard cable to the keyboard port.
3. Connect the mouse cable to the mouse port.
4. Connect the power cord into the power inlet, and then plug into an
electrical outlet.
5. Connect the printer cable to the printer and the printer port.
6. Connect the 100-pin camera data cable to the frame grabber port.
7. Connect one end of the RS232 cable to the serial port on the
computer; connect the other end to the serial port 1 (top port) on the
®
back of the
iQ
200.
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-8
2. Installation Procedures
iQ200 Rear Panel SIC Connection Configurations
WARNING: Before making any connections to Load or Unload ports on
the Rear Panel SIC of the iQ200, ensure the jumpers are in the correct
configuration depending on the system options.
AX-4280 Connection to the iQ200
The second port labeled
“Load” should have all
three jumpersremoved
from the port. CAUTION! If
the jumpers are not
removed, a fuse will blow
on the backplane of the
card cage (F10) inside the
Microscopy Module.
Load Station Connection to the
iQ200
The second port labeled
“Load” should have all
three jumpersinstalled in
the port. CAUTION! If the
jumpers are not in place,
the connection will not
work and no power will be
supplied to the Load
Station.
Unload Station Connection to
the iQ200
The third port labeled
“Unload” should have all
three jumpersinstalled in
the port. CAUTION! If the
jumpers are not in place,
the connection will not
work and no power will be
supplied to the unload
Station. Currently, the only
configuration for this port is
all three jumpers in place.
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-9
2. Installation Procedures
iQ200 / AX-4280 Bridge Installation Instructions
“iQ200-AX4280 Sample bridge kit” PN 800-3501.
1. Remove left side cover from the AX-4280
sampler.
2. Unplug connector # S1-10 (white label) from
the circuit board and replace with connector#
S1-10 (Side Sampler) stored in bottom of
sampler.
3. At the bottom of the sampler, remove the
screw (as indicated on the picture on the right)
securing carry arm stop post. Remove the
post from the sampler.
4. Extend the lever from the sampler. Attach the
rack push arm to the lever with two (2, M3x5)
flat head and one (1, M3x6 pan head) screws.
5. Break the tab on the side cover to allow the
rack to transfer from the Load Station to the
sampler.
6. Replace left side cover.
7. Back out the screw located inside the right
frame of the test strip disposal compartment to
hold the spring loaded return plate.
8. Install the new waste box provided with the
connection kit.
9. Install an alignment bracket under each
analyzer aligning the pins with the two holes
located under the front of each instrument.
Align the brackets and join with the connection
bracket. Secure with four (4, M3x10 pan head)
screws.
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-10
2. Installation Procedures
10. Install the bridge faceplate between the two
samplers. Center and secure with the
thumbscrews located underneath the
faceplate.
11. Secure the bridge between the two samplers
with two (2, M3x6 flat head) screws. Make
sure the guide spring points to the left. Check
for proper alignment and adjust the analyzer
level and/or position to provide a smooth
transition.
12. Make the cable connections between the
analyzers, and the computer according to the layout diagram.
NOTE: Do not use the grey cable provided with the Arkray kit, the pin out
for this cable will not work due to incorrect pin positioning.
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-11
REMOVE jumpers on
COM2 “Load” port
before connecting
directly to AX-4280
2. Installation Procedures
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-12
2. Installation Procedures
13. Power up the AX-4280 while holding the STOP button to access the
Service mode.
a. Press -4280-.
b. Press 1: “User set up” then press 2: “User detail”.
c. Press ENTER until “Sampler” is displayed.
d. Press the “-” button until [LOWER] is displayed and then press
ENTER.
e. Save these settings when prompted.
f. Press ESC until the first menu is displayed.
g. Press 2: “Adjust”. Press 2: “Pulse & LED”. Press 1: “Pulse
adjust”.
h. Press the “-“ button until [RACK] is displayed, then press ENTER.
i. Press the “-“ button until [Carry send] is displayed. Place a rack
on the left side of the AX-4280 sampler in the area where the rack
can be transferred to the iQ200 from the AX-4280 and press
ENTER. Verify that the carry arm moves the rack to the iQ
sampler and is positioned on the iQ 200 sampler fully.
j. If the distance needs to be adjusted, change the pulse number on
the AX-4280 to the proper position. A larger number carries the
rack further. Use the keypad to change the value.
Press 7 to increase the pulse number by 1
Press 4 to decrease the pulse number by 1
Press 9 to increase the pulse number by 10
Press 6 to decrease the pulse number by 10
k. Press MENU until the third menu is displayed.
l. Press 2: System Option. Press 1: Edit rack ID.
m. Enter the following rack ID information-
Rack numbers 1-11 Standard ID
Rack number 12 Start
Rack numbers 13-24 Standard ID
Rack numbers 25-27 Standard ID
Rack number 28 STAT/CTRL
Rack numbers 29-30 PASS
Rack number 31 NONE
14. Save these settings when prompted.
15. Return to Standby mode.
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-13
2. Installation Procedures
16. Load the AX Connectivity CD PN 700-3106 on the Analysis
Processor.
17. From the Main Menu screen of the iQ, click on the Settings button.
18. Click on Interfaces.
19. Verify that the “Enable direct connection” checkbox with the AX-
4280 is selected.
Iris Diagnostics, a Division of Iris International, Inc.
®
iQ
200Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/20052-14
Loading...
+ 185 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.