Invacare Vicair iShear User guide [de]

Page 1
shear genius.
USER MANUAL
iShear
ishear.com
Page 2
Preface
Thank you for purchasing the iShear. This manual contains information necessary for the user to
become familiar with the iShear. We kindly ask you to carefully read this information before
(see Chapter 2 Safety).
This document is product-specic and therefore only applies to the product as indicated in
Paragraph 1.1 Product identication. In this manual, the iShear will be referred to as ‘the iShear’
or ‘the product’.
Refer to the table of contents at the start of this manual to locate information relevant to you.
Keep a downloaded version of this manual on the device with the iShear app and/or other
devices in an accessible place near the product.
For more information, please contact:
Vicair B.V.
Bruynvisweg 5
1531 AX WORMER
The Netherlands
www.ishear.com
+31 (0)75 642 9999
© 2017 iShear
All rights reserved. No parts of this work may be reproduced in any form or by any means –
graphic, electronic, or mechanical, including photocopying, recording, taping, or information
storage and retrieval systems – without the written permission of the publisher.
Products that are referred to in this document may be either trademarks and/or registered
trademarks of the respective owners.
While every precaution has been taken in the preparation of this document, the publisher and
the author assume no responsibility for errors or omissions, or for damages resulting from the
use of information contained in this document or from the use of programs and source code
that may accompany it. In no event shall the publisher and the author be liable for any loss of
prot or any other commercial damage caused or alleged to have been caused directly or
indirectly by this document.
Wormer, The Netherlands
Revision 05/2017
user manual | 3
Page 3
Table of contents
Preface 3
1 Introduction to this manual 5
1.1 Product identication 5
1.2 Target groups 5
1.3 Glossary of terms 6
1.4 Symbols used in this manual 7
1.5 Disclaimer 8
2 Safety instructions 9
5 Using the iShear 24
5.1 Connecting the iShear with
the iShear app 24
5.2 Placing the iShear 25
5.3 Measuring with the iShear 28
5.4 Removing the iShear after
measuring 35
5.5 Working with saved measurements 35
6 Maintenance, cleaning and storage 43
6.1 Maintenance and cleaning 43
6.2 Disinfecting 44
6.3 Storage 44
3 Product description 11
3.1 TSF 11
3.2 iShear 13
3.3 iShear app 15
3.4 Intended use and reasonably
foreseeable misuse 18
3.5 Technical Specications and
system requirements 19
4 Before using the iShear 21
4.1 Preparing the iShear 21
4.2 Downloading the iShear app 21
4.3 Signing up and Login 22
4.4 Conguring your device 23
7 Troubleshooting 46
8 Disposing of the iShear 47
9 Warranty 48
| iShear4
Page 4
1 Introduction to this manual
1.1 Product identication
The iShear is identied by the identication sticker on the product (see Figure 1).
Figure 1 - Location of the identication sticker with serial number
Table 1 - Product Identication
Product / Model
Serial number (SN)
Manufacturer
1.2 Target groups
iShear
6 digits. For instance 100003. See Figure 1
Vicair B.V.
This document applies to dierent target groups, see Table 2.
Table 2 - Target groups
Target group Denition
Owner The owner of the iShear
User
Client The person sitting in the wheelchair
The operator of the iShear and iShear app. No special skills or training necessary.
user manual | 5
Page 5
1.3 Glossary of terms
Terms that are used in this document are dened in Table 3.
Table 3 – Glossary of terms
(iShear) App
The iShear application for Apple iOS and Android devices, see Paragraph 3.3
Bluetooth
Client
File
Force Unit
iShear
Lateral aps
Log
Log session
Menu symbol
Seat plane
Sensor bar
Share
Sync
TSF
A Bluetooth Low Energy (BLE / Bluetooth Smart) connection
between the iShear and the device with the iShear app is necessary.
Person sitting in the wheelchair
A saved single measurement of TSF
The unit in which the TSF is measured: Kilograms (Kg), Newton (N) or Pounds (Lbs)
The physical product, see Paragraph 3.2 iShear
The left and right sides of the iShear mat
A saved over-time measurement of TSF
Continuous measuring over a set time period
The horizontal plane between wheelchair seat base and the wheelchair cushion
The aluminium bar of the iShear
Share File(s) or Log(s)
Synchronise saved Logs between iShear and iShear app
Total shear force in the seat plane measured by the iShear (anterior – posterior force parallel to the seat plane)
| iShear6
Page 6
1.4 Symbols used in this manual
The symbols below will be used on the product and throughout this manual to focus the
attention of the user on important information
Table 4 – Used symbols
Manufacturer
Date of manufacture
Serial number (on sensor bar – see Figure 1)
Lot number (on mat label – for example 27.01.2017)
Caution: Consult instructions for use for important cautionary information such as warnings and precautions
CE
Protected against tools, wires etc. greater than 2.5 millimeters. Protected against dripping water
Keep dry
IP31
Do not dispose of with other waste products; contact the manufacturer for instructions
Waste Electrical and Electronic Equipment recycling Directive: Battery
Model or Type: iSHEAR
FCC ID : ZAT26M1 IC : 451H-26M1
Danger! The signal word that indicates a hazard with a high level of risk which, if not avoided, will result in death or serious injury.
Warning! The signal word that indicates a hazard with a medium level of risk which, if not avoided, could result in death or serious injury and/ or cause serious damage to the product.
Caution! The signal word that indicates a hazard with a low level of risk which, if not avoided, could result in minor or moderate injury and/or cause damage to the product.
Notice The signal word that indicates information considered important, but not hazard-related (e.g. messages relating to property damage).
user manual | 7
Page 7
1.5 Disclaimer
The product owner is responsible for ensuring that only the latest revision of this document is
used, as supplied by Vicair. The latest revision of this manual can be found in the iShear app and
at iShear.com/support. Make sure to install app-updates when provided. The user of the iShear
should always have a downloaded oine version of the manual available for reference.
Furthermore, the product owner is to ensure that the user is aware and in possession of this
document and is familiar with its contents.
| iShear8
Page 8
2 Safety instructions
This chapter informs the user of the iShear about safety. It is important to read and adhere to
these instructions. Safety instructions are also mentioned in the chapters to which they apply.
• Improper use and/or care of the iShear may lead to improper measurements, malfunction,
failure or damage to the iShear.
• Do not use the iShear with clients weighing less than 45 kg / 99 lbs or more than 120 kg /
264lbs.
• Do not use the iShear in temperatures below -10° C / 14 F or higher than 40° C / 104 F.
• Do not pivot on the iShear.
• Do not pull at the sides and/or the sensor bar of the iShear when the client is seated on the
cushion.
• Do not remove the iShear when the client is seated on the cushion.
• Do not pull at the sensor bar and/or use excessive force when removing the iShear.
• Do not soak the iShear in uids, keep away from water, (bodily) uids and humidity.
• Do not continue to use the iShear when it is soiled with urine or soaked with any other kind
of uid. Contact Vicair.
• With incontinent clients, ensure incontinence material is being used when using the iShear.
• Before each session check the iShear for any visible damage to the sensor bar and/or mat
(holes, tears, torn seams, dents etc.). In case of any visible damage, do not use the iShear.
Contact Vicair.
• Do not remove the mat from the sensor strips of the iShear.
• Do not place the client directly on top of the iShear.
• Before starting a session with a new client, the iShear must be disinfected to avoid cross
contamination.
• An iShear session may only be performed with the iShear placed underneath a seat cushion.
• Do not pinch or wrinkle the iShear.
• Do not fold the iShear in any other way than in the way it is originally placed in the storage
case.
• Do not place any objects on top of a folded iShear.
• Do not use excessive force when placing the iShear in the chair.
• Be careful when using secondary positioning systems to transfer the client into the chair, as
they may damage the iShear or alter its measurements.
• Check after transferring the client if the iShear and cushion are still in correct position.
• Always store the iShear in the original iShear carrying case after each session.
user manual | 9
Page 9
• If the iShear is not in use for an extended period, remove the batteries from the battery
compartment of the sensor bar.
• Do not use undiluted bleach or Hydrogen Peroxide to clean the iShear.
• Cleaning products or disinfectants must be thoroughly wiped o the iShear mat after
cleaning/disinfecting. Allow the iShear mat to dry thoroughly before use or storage.
• Make sure that the lateral parts of the cushion are in between the lateral aps of the iShear.
• Make sure that the lateral aps of the iShear are not covering the top of the cushion.
• Make sure there is no loss of seat depth when placing the iShear in the wheelchair.
• Measurement data and patient data are not in any way collected or stored by Vicair.
• Within the app you have the option to store patient information. It is the user’s responsibility
to safeguard this (sensitive) patient information.
• Measured values can be dierent with dierent client weight.
• The iShear is not a product selection tool.
• Friction in the back system could create a dierent TSF value.
• The values measured by the iShear do not provide an indication of the risk of tissue damage.
• Because of internal friction measured values between -2 kg and 2 kg are less accurate.
• A (re-)calibration of the sensors may only be performed after specic instructions from Vicair.
• For safe disposal of the iShear: please contact Vicair for instructions.
| iShear10
Page 10
3 Product description
The iShear is a medical device that measures the sliding force in the seat plane of a client sitting
in a wheelchair. The iShear app is the user interface of the product.
In this manual ‘sliding force’ will be referred to as ‘TSF’, as explained in 3.1.1.
By using the iShear and iShear app you can measure the TSF in dierent wheelchair set-ups and
create a situation in which TSF is as low as possible. This decreases the negative eects of TSF on
the client.
3.1 TSF
3.1.1 Explanation of TSF
The total shear force in the seat plane measured by the iShear is the anterior-posterior force
parallel to the seat plane. It is measured between the cushion and the seat base interface. From
now on we will refer to this as TSF. This paragraph further explains this denition.
In Figure 2 the client is sitting upright in the wheelchair. The forces shown in gure 2 are the
vertical forces, perpendicular to the seat plane. The weight of the client pushes down on to the
cushion (gravity – green arrows) and the cushion ‘pushes back’ with a force in the opposite
direction (normal force – red arrows).
In Figure 3, the client is leaning against the back support. This situation introduces horizontal
forces, parallel to the seat plane. Besides the vertical forces, the client exerts a horizontal force
into the back support and the back support ‘pushes back’ with a horizontal force in the opposite
direction (red arrow in back support).
In addition, there are horizontal forces in the seat plane. Because the client is pushing back in
the back support his/her body consequently pushes forward in the seat plane (green arrow in
seat plane). This horizontal force in the seat plane is the ‘sliding force’ and what is referred to as
the TSF in this document. The red arrow in the seat plane is the friction force.
user manual | 11
Page 11
Figure 2 - Situation without sliding forces Figure 3 - Situation with sliding forces
3.1.2 Wheelchair settings that inuence TSF
In a clinical setting, the TSF can be inuenced by setting up the wheelchair and its components.
TSF variations can be expected with variations in (e.g.):
• Seat angle
• Back angle
• Tilt angles
• Foot supports
• Back support interventions
• Shell back supports
• Tension adjustable back supports
• Custom molded back supports
• Non-adjustable back supports
• Propulsion (hand and foot propulsion)
• Secondary support systems
| iShear12
Page 12
3.2 iShear
The iShear consists of the following components (see Figure 4, 5 and 6):
• Sensor bar (see Figure 5)
The Sensor bar is made of aluminium. It holds:
– On/O button
– On/O indicator light
– Battery compartment
– Button to unlock battery compartment
– Bluetooth connection device
– Log session indicator
– U-shaped metal prole
– Product identication sticker (see Figure 1)
• Sensor strips (see Figure 6)
• 4 AAA batteries
• Mat
The iShear mat has a zero friction middle layer and a high friction top layer. The left and right
sides of the mat are referred to as ‘lateral aps’ (see Figure 4).
The iShear is equipped with a carrying case (see Figure 7 )
Figure 4 - Components of the iShear
Sensor Bar
Mat
Lateral ap
user manual | 13
Page 13
On/o button
Log session indicator light
(ashing)
On/o indicator
light (ashing)
Button to unlock battery compartment
Battery compartment for four AAA
batteries
Figure 5 - Sensor bar components
Figure 6 - Sensor bar with sensor strips and U-shaped metal prole for stabilisation
Figure 7 - Carrying case
| iShear14
Page 14
3.3 iShear app
Measurements of the iShear are registered by the iShear app. It is the user interface of the iShear
and features the following functions:
• Single measurements of TSF
• Over-time measurement of TSF = ‘Log session’
• View, share and compare measurements
For system requirements, refer to Paragraph 3.5 Technical Specications and system
requirements.
3.3.1 Login screen
The Login screen is the screen that appears when you open the app without being logged in.
Figure 8 - Login screen
Fill in existing account details,
using username and password
Tap to login
Tap to request a password reset for
an existing account
Sign up to create an account
user manual | 15
Page 15
3.3.2 Connect screen
The Connect screen is the screen that appears when you open the app when you are logged in.
Tap to go to Menu
Tap to scan for the iShear
Scanned iShear
Logged in username
Figure 9 - Connect screen
3.3.3 Measurement screen
The Measurement screen is the screen from which you take measurements.
Tap to go to Menu
TSF indicator
Reset button
Figure 10 - Measurement screen
| iShear16
iShear battery indicator
Measurement force unit
Left-right balance indicator
Save le (single measurement)
Go to over-time measurement
(Log session)
Page 16
3.3.4 Menu screen
The Menu screen can be entered either by swiping to the right or tapping the Menu button.
Tap to leave Menu screen
Figure 11 - Menu screen
Functions (see Figure 11):
• Force Unit: set the force unit to Kilograms (Kg), Newton (N) or Pounds (Lbs)
• Update speed: set the update interval for the measurement
• Saved Files: show saved single measurements
• Saved Logs: show saved Log sessions (over-time measurements)
• Support: go to Support screen
• Connected to: shows currently connected iShear
• Disconnect: disconnects the iShear from the app
• Login by: connected username
• Logout: Log out your user account
user manual | 17
Page 17
3.3.5 Support screen
• A (re-)calibration of the sensors may only be performed after specic
instructions from Vicair.
The Support screen (Figure 12) can be entered via the Menu screen. There you can nd the serial
number of the connected iShear, its rmware version and the app version.
Go to iShear manual
Go to calibration options
Go to rmware updates
Back to Measurement screen
Figure 12 1- Support screen
3.4 Intended use and reasonably foreseeable misuse
• Do not use the iShear in temperatures below -10° C / 14 F or higher
than 40° C / 104 F.
• Do not use the iShear with clients weighing less than 45 kg / 99 lbs or
more than 120 kg / 264 lbs.
• Improper use and/or care of the iShear may lead to improper
measurements, malfunction, failure or damage to the iShear.
• Measurement data and patient data are not in any way collected or
stored by Vicair.
• Within the app you have the option to store patient information. It is
the user’s responsibility to safeguard this (sensitive) patient
information.
• The iShear is not a product selection tool
• The values measured by the iShear do not provide an indication of
the risk of tissue damage.
• A (re-)calibration of the sensors may only be performed after specic
instructions from Vicair.
| iShear18
Page 18
The iShear is intended for measuring the total shear force in the seat plane of a client sitting in a
wheelchair. It is intended to be used in combination with the iShear App, using a Bluetooth
connection (Bluetooth Low Energy/BLE/Bluetooth Smart, see 3.5 Table 5).
Misuse can occur when the iShear is used beyond its capacity, capability or operational
boundaries (see Paragraph 3.5 Technical Specications and system requirements). Misuse can
also occur if one does not keep to the safety precautions (see Chapter 2 Safety) or use the iShear
for any other purposes than dened in this chapter.
Clinical applications include:
• Evaluate total setup of the wheelchair in combination with pressure mapping
• Measure the inuence of seat and back angles on TSF
• Evaluate dierent seat and back solutions on TSF
• Evaluate eect of leg positioning on TSF
• Determine rotational forces in the seat plane
• Evaluation of a pelvic angle tilt/rotation on the TSF
• Measure TSF over time
• Evaluate the inuence of secondary positioning aids on TSF
• Educational purposes (end users, care givers)
The iShear helps you to set up a wheelchair. The values measured do not provide an indication of
the risk of tissue damage.
3.5 Technical Specications and system requirements
Table 5 – Technical Specications and system requirements
• Weight: 1,7 kg / 3.75 lbs
• Dimensions (HxWxD) 27 x 690 x 615 mm / 1.06 x 27.17 x 24.21 inches
• Material sensor bar: aluminium
• Material top layer mat: PU coated tricot and nylon fabric
• Batteries: 4 AAA 1,5V LR03 batteries
• Supported operating systems: iOS 8.0 or newer, compatible with iPhone, iPad and iPod Touch Android 4.3 or newer
• Connection requirements: Bluetooth Low Energy (BLE, also known as Bluetooth Smart). This Low Energy Bluetooth version is in general available on devices since 2011 (rst BLE version was Bluetooth 4.0).
user manual | 19
Page 19
• Wheelchair requirements: The iShear can be used in a seat plane of 35-50 cm / 133⁄4-20 inch width (X) and 35-50 cm / 133⁄4-20 inch depth (Y). (see Figure 12).The back-support canes must be at least 30 cm / 11.8 inch apart from each other
• Allowed user weight on the iShear: Minimum: 45 kg / 99 lbs, Maximum: 120 kg / 264 lbs
• Ambient temperature Minimum: -10 °C / 14 F, Maximum: 40 °C / 104 F
• Relative humidity Operating: 30 to 75% RH (non-condensing) Storage: 10 to 100% RH (non-condensing)
Y
X
Figure 12-2 - Wheelchair requirements
| iShear20
Page 20
4 Before using the iShear
4.1 Preparing the iShear
• Before each session check the iShear for any visible damage to the
sensor bar and/or mat (holes, tears, torn seams, dents etc.). In case of
any visible damage, do not use the iShear. Contact Vicair.
• Do not fold the iShear in any other way than in the way it is originally
placed in the storage case.
• Do not place any objects on top of a folded iShear.
• Do not pinch or wrinkle the iShear.
• Do not remove the mat from the sensor strips of the iShear.
To make the iShear ready for use:
1. Take the iShear from its carrying case.
2. Unlock the battery compartment (see 3.2 Figure 5), take it out of the sensor bar and place 4
AAA batteries in the compartment.
3. Re-insert the battery compartment into the sensor bar.
4.2 Downloading the iShear app
The iShear app is currently only available for Apple iOS devices running on iOS 8.0 or newer and
Android devices running on Android 4.3 or newer.
When you are going to use your own personal device (phone/tablet) with the iShear download
the iShear app on that device.
When your company or institution uses one company-owned device (phone/tablet) that stays
with the iShear, download the app on that shared device.
To download the iShear app:
1. Go to the app store / play store and search for ‘iShear’.
2. Tap ‘Download’ and install the app.
user manual | 21
Page 21
4.3 Signing up and Login
1. Tap ‘Sign Up’ in the Login screen. (see Figure 13)
2. When you use your own personal device (phone/tablet) to download the iShear app on, you
can choose to sign up with an individual account using your own email address. When your
company or institution uses one company-owned device (phone/tablet) that stays with the
iShear, download the app on that shared device. In this case it might be useful to create a
shared company account with the company name as user name and a general company
email address. (see Figure 14)
3. After lling in a username, email address and your password (the same password twice),
check the ‘I agree to the terms and conditions’ box.
4. Tap ‘Sign Up’.
5. You will receive a Sign Up conrmation via the email address that you have signed up with.
This is to validate the used email address.
6. Conrm your email address through the link in the Sign Up conrmation email.
7. You can now go to the app and log in. The app oers to remember your account
(see gure 15). If you prefer this then you can tap “Yes”. Otherwise tap “No”.
Figure 13 - Sign up
| iShear22
you@company.com
Your username
I agree to the terms and conditions.
info@company.com Group username
Figure 14 - Create an individual or group account
Page 22
Figure 15 - Succesful Login- Remember account
If – at any time – it is necessary to login, you can do this by using the registered username and
password.
4.4 Conguring your device
Congure your Apple or Android device so that the iShear app will work properly. Consult the
user manual of your device for the following congurations:
• Bluetooth
Turn on the Bluetooth function. Bluetooth version 4 (Bluetooth Low Energy – BLE) is required to
be able to connect. The iShear can now connect with your device.
• Free hard disk space
Make sure there is enough space available in order to install the iShear app and save
measurements.
user manual | 23
Page 23
5 Using the iShear
5.1 Connecting the iShear with the iShear app
To establish a Bluetooth connection between the iShear and the iShear app:
1. Press the on/o button on the sensor bar (see 3.2 – Figure 5)
A ashing blue light indicates that the iShear is turned on.
2. Open the iShear app on your device and log in with your username and password.
3. Tap ‘Scan Done Press to rescan’ (see Figure 16)
4. Tap the iShear you wish to connect to (see Figure 16). When the connection is successful, the
Measurement screen (see Figure 10, chapter 3.3.3) will open automatically.
Figure 16 - Connect the iShear
| iShear24
Page 24
5.2 Placing the iShear
• Do not use the iShear with clients weighing less than 45 kg / 99 lbs
• Do not use the iShear in temperatures below -10° C / 14 F or higher
• Before each session check the iShear for any visible damage to the
• With incontinent clients, ensure incontinence material is being used
• Do not pull at the sides and/or the sensor bar of the iShear when the
• Before starting a session with a new client, the iShear must be
• Cleaning products or disinfectants must be thoroughly wiped o the
• Do not use excessive force when placing the iShear in the chair.
• Do not pinch or wrinkle the iShear.
• Do not place any objects on top of a folded iShear.
• Do not place the client directly on top of the iShear.
• An iShear session may only be performed with the iShear placed
• Be careful when using secondary positioning systems to transfer the
• Check after transferring the client if the iShear and cushion are still in
• Make sure there is no loss of seat depth when placing the iShear in
• Make sure that the lateral parts of the cushion are in between the
• Make sure that the lateral aps of the iShear are not covering the top
or more than 120 kg / 264 lbs.
than 40° C / 104 F.
sensor bar and/or mat (holes, tears, torn seams, dents etc.). In case of
any visible damage, do not use the iShear. Contact Vicair.
when using the iShear.
client is seated on the cushion.
disinfected to avoid cross contamination.
iShear mat after cleaning/disinfecting. Allow the iShear mat to dry
thoroughly before use or storage.
underneath a seat cushion.
client into the chair, as they may damage the iShear or alter its
measurements.
correct position.
the wheelchair.
lateral aps of the iShear.
of the cushion.
user manual | 25
Page 25
To place the iShear in the wheelchair. These steps correspond to image I-V of the Quick Guide
provided with the iShear:
1. Transfer the client out of the wheelchair.
2. Remove the seat cushion from the wheelchair.
3. Place the iShear on the wheelchair seat base with the sensor bar placed horizontally between
the back support canes. The lateral aps should be folded upwards against the side guards/
armrests of the wheelchair. The space between the sensor bar and the wheelchair canes
should be the same on both sides of the sensor bar (see Figure 17).
y
x=y
x
Figure 17
| iShear26
Page 26
4. Place the seat cushion back in the wheelchair, on top of the iShear. Make sure the lateral
aps are positioned between the sides of the wheelchair cushion and the side guards of the
wheelchair. The rear end of the wheelchair seat should be equal to the rear end of the iShear
mat (see Figure 18).
Figure 18
5. Transfer the client back into the wheelchair.
user manual | 27
Page 27
5.3 Measuring with the iShear
• Do not pull at the sides and/or the sensor bar of the iShear when the
client is seated on the cushion.
• Do not pivot on the iShear.
• Do not remove the iShear when the client is seated on the cushion.
• Do not soak the iShear in uids, keep away from water, (bodily) uids
and humidity.
• Do not continue to use the iShear when it is soiled with urine or
soaked with any other kind of uid. Contact Vicair.
• Improper use and/or care of the iShear may lead to improper
measurements, malfunction, failure or damage to the iShear.
• An iShear session may only be performed with the iShear placed
underneath a seat cushion.
• Do not pinch or wrinkle the iShear.
• Measurement data and patient data are not in any way collected or
stored by Vicair.
• Within the app you have the option to store patient information. It is
the user’s responsibility to safeguard this (sensitive) patient
information.
• Measured values can be dierent with dierent client weight.
• The iShear is not a product selection tool
• Friction in the back system could create a dierent TSF value
• The values measured by the iShear do not provide an indication of
the risk of tissue damage.
• Because of internal friction measured values between -2 kg and 2 kg
are less accurate.
• A (re-)calibration of the sensors may only be performed after specic
instructions from Vicair.
| iShear28
Page 28
5.3.1 Before measuring
Before measuring the TSF in dierent wheelchair setups, set the TSF to zero in a neutral sitting
position:
1. Let the client sit upright in the wheelchair, without leaning against the back support
(see Figure 19).
2. Tap ‘Reset’. A popup with ‘Reset value to 0’ appears (see Figure 19).
Figure 19 - Sit upright and reset
Please note: Only reset before a measurement session. Do not reset once you have started
measuring. This will alter the measurement data.
user manual | 29
Page 29
5.3.2 Single measurement
To take a single measurement of TSF:
1. Let the client lean against the back support (see Figure 20).
2. Wait until the real time measurement is stabilized.
Figure 20 - Lean against back support
To save the measurement:
3. Tap ‘Save’ in the measurement screen (see Figure 21). The Save screen opens.
4. Tap the empty text eld next to ‘Patient ID’ and ll in a patient ID (see Figure 22).
5. Tap the empty text eld underneath ‘comments’ and ll in any relevant information about
this measurement or wheelchair setup (see Figure 22) There is space for 200 characters max.
6. Tap ‘Save’ and then ‘OK’ in the popup screen (see Figure 23).
The measurement is now saved.
7. Tap ‘Back’ to leave the Save screen.
| iShear30
Page 30
Figure 21 - Tap Save
Figure 23 - Tap ‘OK’ to save
Figure 22 - Save screen
user manual | 31
Page 31
5.3.3 Over-time measuring (Log session)
To perform an over-time measurement of the TSF in a specic wheelchair setup over a set period
of time:
1. Let the client lean against the back support (see Figure 24).
2. Tap the ‘Log’ button in the Measurement screen (see Figure 25). This takes you to the Log session
screen (Figure 26).
3. Tap the empty text eld next to ‘Patient ID’ and ll in a patient ID (see Figure 26).
4. Swipe up or down next to ‘Duration’ to set the desired duration of the measurement in hours
and/or minutes (see Figure 26).
5. Tap the the desired log interval next to ‘Type’ to set the time interval of the measurement
(see Figure 26).
– ‘Trend’ = 1 measurement per 60 seconds.
– ‘Precise’= 1 measurement per 10 seconds.
– ‘Ultra’= 1 measurement per second.
– ‘Continuous’= 4 measurements per second.
6. Tap the empty text eld underneath ‘Comments’ and ll in any relevant information about
this measurement or wheelchair setup (see Figure 26). There is space for 200 characters max.
7. Tap ‘Start’ and then ‘OK’ in the popup screen with the Logging alert (see Figure 26a). Below in the
Log session screen, next to ‘Session Running’, the left Log session time is counting down (see
Figure 26).
8. Tap ‘Back’ to go to the measurement screen. The Log button will contain red lettering during
a Log session (see Figure 27).
9. The red light of the Log session indicator on the side of the iShear Sensor bar is ashing (red)
during the Log session.
10. You do not need to keep your device (phone/tablet) connected or close to the iShear that is
performing the Log session. The Log session will go on for the set duration and stop
automatically when the time is up.
| iShear32
Page 32
Figure 24 - Lean against back support Figure 25 - Tap Log
Figure 26 - Log session screen Figure 26a - Tap OK
user manual | 33
Page 33
Figure 27 - Active Log session
To stop a Log session before the end of the set duration:
You do not need to keep your device (phone/tablet) connected or close to the iShear that is
performing the Log session. Therefore it might be necessary to reconnect before you can stop
the Log session before the end of the set duration. If your device and the iShear are still
connected you can immediately perform step 4 and 5.
1. Log in (see chapter 5.1)
2. Tap ‘Scan Done scan to rescan’ (see chapter 5.1)
3. Connect with the iShear that is performing the Log session (see chapter 5.1)
4. Tap the ‘Log’ button in the Measurement screen (see Figure 25). This will take you to the Log
session screen.
5. Tap ‘Stop’ in the Log session screen (see Figure 26). The Log session is now stopped.
| iShear34
Page 34
5.4 Removing the iShear after measuring
• Do not remove the iShear when the client is seated on the cushion.
• Do not pull at the sensor bar and/or use excessive force when
removing the iShear.
• Always store the iShear in the original iShear carrying case after each
session.
• Do not fold the iShear in any other way than in the way it is originally
placed in the storage case.
• Do not place any objects on top of a folded iShear.
Remove the iShear from the wheelchair:
1. Transfer the client out of the wheelchair.
2. Remove the seat cushion from the wheelchair.
3. Remove the iShear and press the on/o button on the sensor bar for 2 seconds to turn the
iShear o. The blue light rst ashes rapidly before it turns o.
4. Place the seat cushion back in the wheelchair.
Make sure the cushion is used as described in the cushion operational manual
5. Transfer the client back into the wheelchair.
6. Place the iShear in the carrying case (see 6.3).
5.5 Working with saved measurements
• Measurement data and patient data are not in any way collected or
stored by Vicair.
• Within the app you have the option to store patient information. It is
the user’s responsibility to safeguard this (sensitive) patient
information.
• Measured values can be dierent with dierent client weight.
• The iShear is not a product selection tool
• Friction in the back system could create a dierent TSF value
• The values measured by the iShear do not provide an indication of
the risk of tissue damage.
• Because of internal friction measured values between -2 kg and 2 kg
are less accurate.
user manual | 35
Page 35
5.5.1 Viewing saved measurements
You can view saved measurements via the Menu screen.
5.5.1.1 Viewing saved single measurement les
1. Open the iShear app on your device.
2. Swipe to the right or tap the ‘Menu’ symbol in the top left corner of the screen to go to the
Menu screen.
3. Choose ‘Saved les’ (see Figure 28). This takes you to the list of saved Files (see Figure 29).
4. Tap the File you wish to view and tap ‘Open’ (see Figure 29a)
5. If necessary you an add extra information to ‘Comments’ or edit the ‘Patient ID’. Press ‘Save’
to save new information.
Figure 28 - Menu screen Figure 29 - Saved Files
| iShear36
Page 36
Figure 29a - Open selected File
user manual | 37
Page 37
5.5.1.2 Viewing saved over-time measurement Logs
Over-time measurements (Logs) are rst saved locally on the iShear. Only the account that saved
the Log can access the Log data to synchronise, view, share or delete.
A saved Log is a CSV le. You can only open (view) a Log by choosing to share it and then open in
with a suitable app on your device.
1. Open the iShear app on your device.
2. Make sure you are connected to the iShear (see chapter 5.1).
3. Swipe to the right or tap the ‘Menu’ symbol to go to the Menu screen.
4. Choose ‘Saved Logs’ (see Figure 30). This takes you to the list of saved Logs (see Figure 31).
5. To make sure you see all the Logs created by your User account, you rst have to synchronise
the Log data. ‘Sync’ will be blue if you need to synchronise. Tap ‘Sync’ (see Figure 31). Your
saved Logs are then synchronised (see Figure 32).
6. Tap the Log you wish to view and tap ‘share’ (see Figure 33). You can now choose the app in
which you would like to open and view the Log (see Figure 33a).
Figure 30 - Tap Saved Logs Figure 31 - Saved Logs
| iShear38
Page 38
Figure 32 - Syncing Logs Figure 33 - Tap Share
Figure 33a - Choose app to open or share
5.5.2 Deleting saved Logs to create free memory space
Over-time measurements (Logs) are rst saved locally on the iShear. Only the account that saved
the Log can access the Log data to synchronise, share or delete. The ‘free memory indicator’ in
your Saved Logs screen will show you when you need to delete saved Logs to create free memory
space.
user manual | 39
Page 39
Please Note: Share your Logs before deleting!
If you wish to save created Logs for your own administration, make sure you store your saved
Logs locally on your device (phone/tablet) by sharing them with yourself via email or another
suitable app on your device. Otherwise this data will be lost. See 5.5.3 ‘Sharing saved
measurements’ for instructions.
To delete a Saved Log:
1. Open the iShear app on your device.
2. Make sure you are connected to the iShear (see chapter 5.1).
3. Swipe to the right or tap ‘Menu’ to go to the Menu screen.
4. Choose ‘Saved Logs’ (see chapter 5.5.1.2 – Figure 30). This takes you to the list of saved Logs
(see chapter 5.5.1.2 – Figure 31).
5. To make sure you see all the Logs created by your User account, you rst have to synchronise
the Log data. Tap the ‘Sync’ button (see chapter 5.5.1.2 – Figure 32). Your saved Logs are then
synchronised.
6. Tap the Log(s) you wish to delete and tap ‘Delete’ (see Figure 34).
7. Tap ‘OK’ in the pop-up screen to permanently delete the Log(s) from you device (phone/
tablet) and from the iShear and the iShear app(see Figure 34a).
Figure 34 - Tap Delete Log Figure 34a - Tap OK
| iShear40
Page 40
5.5.3 Sharing saved measurements
Please note: When you choose to share multiple Files/Logs, the data will be combined in one CSV
le. If you wish to receive a separate le per File/Log, then you will have to share them one at a
time.
To share a saved File or Log via email or another suitable app on your device (phone/tablet):
1. Go to your list of Files or Logs (as explained in 5.5.1 ‘Viewing saved measurements’).
2. Tap the File(s) or Log(s) that you want to share. For multiple selection simply tap more than
one saved measurement (for Files there is a maximum of 5 that you can share at once).
3. Tap ‘Share’ (see Figure 35).You can now choose the app with which you would like to share
your File(s) or Log(s) (see Figure 35a).
Figure 35 - Share File(s) or Log(s) Figure 35a - Choose app to share
user manual | 41
Page 41
5.5.4 Comparing single measurements
You can only compare saved single measurements (Files). It is not possible to compare Logs.
To compare Files:
1. Swipe to the right or tap the ‘Menu’ symbol
2. Tap ‘Saved Files’
3. Tap the Files that you want to compare (see Figure 36). There is a maximum selection of 5Files.
4. Tap ‘Compare’ (see Figure 36).
You can now compare the selected Files (see Figure 37).
The wide light-blue graphs show the TSF of the Files that you are comparing. The two narrow dark-
blue graphs (placed inside the light-blue graphs) show the left-right balance of the measurements.
You can tap one of the graphs for the specic File details.
Figure 36 - Select Files to compare Figure 37 - Compare Files screen
| iShear42
Page 42
6 Maintenance, cleaning and storage
6.1 Maintenance and cleaning
• Before each session check the iShear for any visible damage to the
sensor bar and/or mat (holes, tears, torn seams, dents etc.). In case of
any visible damage, do not use the iShear. Contact Vicair.
• Do not soak the iShear in uids, keep away from water, (bodily) uids
and humidity
• Do not continue to use the iShear when it is soiled with urine or
soaked with any other kind of uid. Contact Vicair.
• Do not remove the mat from the sensor strips of the iShear.
• Do not pinch or wrinkle the iShear.
• Do not fold the iShear in any other way than in the way it is originally
placed in the storage case.
• Do not place any objects on top of a folded iShear.Do not remove the
mat from the sensor strips of the iShear.
• Do not use undiluted bleach or Hydrogen Peroxide to clean the
iShear.
• Cleaning products or disinfectants must be thoroughly wiped o the
iShear mat after cleaning/disinfecting. Allow the iShear mat to dry
thoroughly before use or storage.
Clean the iShear mat daily with a damp (not wet) cloth using a mild detergent and water
solution.
For institutional cleaning use a general disinfectant cleaner. Mild alcohol based cleaners and
cleaners containing no more than 10% bleach may also be used.
Do not use undiluted bleach or Hydrogen Peroxide to clean the iShear.
Cleaning products or disinfectants must be thoroughly wiped o the iShear mat after cleaning/
disinfecting. Allow the iShear mat to dry thoroughly before use or storage.
user manual | 43
Page 43
6.2 Disinfecting
• Do not soak the iShear in uids, keep away from water, (bodily) uids
and humidity
• Do not continue to use the iShear when it is soiled with urine or
soaked with any other kind of uid. Contact Vicair.
• Do not remove the mat from the sensor strips of the iShear.
• Do not use undiluted bleach or Hydrogen Peroxide to clean the
iShear.
• Cleaning products or disinfectants must be thoroughly wiped o the
iShear mat after cleaning/disinfecting. Allow the iShear mat to dry
thoroughly before use or storage.
Before starting a session with a new client, the iShear must be disinfected to avoid cross
contamination. Clean the iShear as instructed above. Preferably use an antibacterial detergent.
After cleaning spray the iShear mat with a disinfectant. Cleaning products or disinfectants must
be thoroughly wiped o the iShear mat after cleaning/disinfecting. Allow the iShear cover to dry
thoroughly before use or storage.
6.3 Storage
• Always store the iShear in the original iShear carrying case after each
session.
• Cleaning products or disinfectants must be thoroughly wiped o the
iShear mat after cleaning/disinfecting. Allow the iShear mat to dry
thoroughly before use or storage.
• Do not fold the iShear in any other way than in the way it is originally
placed in the storage case.
• Do not place any objects on top of a folded iShear.
• If the iShear is not in use for an extended period, remove the
batteries from the battery compartment of the sensor bar.
For temporarily storing the iShear after each session, make sure the iShear is turned o and
place it in the provided carrying case. Store in a dry place.
Store the iShear by gently folding the two lateral aps downwards underneath the section of
the iShear with the sensor strips. Place the iShear in the carrying case with the top of the iShear
facing up (See Figure 18)
| iShear44
Page 44
Figure 18 - Store the iShear correctly in its case
Remove the batteries before you store the iShear for a longer period of time. This will prevent
battery leakage and consequential damage to the product.
Storage and transport conditions
Ambient temperature Minimum: -10 °C / 14 F - Maximum: 40 °C / 104 F
Relative humidity Storage: 10 to 100% RH (noncondensing)
Store the iShear folded as described in this paragraph. Do not place any objects on top of a
folded iShear.
user manual | 45
Page 45
7 Troubleshooting
In case the iShear or iShear App is not working as expected, please try one of the following
solutions:
• Close down all running apps, except the iShear app.
• Re-start the iShear app.
• Check for app-updates (updates will be available in your app store / play store)
• Disconnect the iShear and reconnect.
• Try new batteries.
If this does not solve the problem, call technical support for assistance:
Vicair B.V.
Bruynvisweg 5
1531 AX, Wormer
The Netherlands
Worldwide Technical Support
+31 (0)75 642 9999
Monday to Friday
e-mail: support@ishear.com
Visit our website at:
www.ishear.com
| iShear46
Page 46
8 Disposing of the iShear
• For safe disposal of the iShear: please contact Vicair for instructions
The meaning of the symbol on the product, its accessory or packaging indicates that this product
shall not be treated as household waste. Please dispose of this product and/or accessory by
returning it to the manufacturer. Contact Vicair for information. See 7.2 for contact details.
Do not dispose of batteries as household waste. Take them to a battery recycling facility.
user manual | 47
Page 47
9 Warranty
Vicair oers a limited warranty against defects in workmanship and materials for a period of
two (2) years from the original date of purchase, provided the product has been used normally.
Please retain your receipt as proof of purchase. Before using the iShear, the user must ensure
that it is functional and in proper working conditions. The user must know how to operate the
iShear.
These operating instructions are part of the iShear. Compliance with these operating
instructions is mandatory in order to ensure the correct function and operation of the iShear.
Any damage arising from abnormal use, or caused by improper handling, cleaning or negligence
is excluded from this warranty. No responsibility shall be accepted for damage caused by any of
the following: unsuitable or improper storage or use, incorrect installation or putting into
operation by the owner or third parties, natural wear and tear, changes or modications,
incorrect or negligent handling, overuses, chemical, electrochemical or electrical interference or
humidity, unless this is attributable to negligence on the part of Vicair. If operating, climatic or
any other inuences lead to a major change in conditions or material quality, the warranty for
perfect functioning of the iShear shall be rendered null and void.
This warranty is void if failure of the software and/or hardware has resulted from accident,
abuse, or misapplication. Any modications by a third party to the software and/or hardware
will void the manufacturer’s warranty and any obligations to provide maintenance services.
Should a defect in materials or workmanship occur within two (2) years from the original date of
purchase, Vicair will, at its own choice, repair or replace the product free of charge.
Remedies for breach of express warranties herein are limited to repair or replacement
of the product. In no event shall damages for breach of any warranty include any
consequential damages or exceed the cost of non-conforming goods sold.
For additional product information, see our website: www.ishear.com.
| iShear48
Page 48
Page 49
Vicair B.V.
8 717261 042749
Bruynvisweg 5 1531 AX Wormer The Netherlands
080000-REV05/2017
www.ishear.com
shear genius.
ishear.com
Loading...