Invacare Poseidon User Manual

POSEIDON HUMIDIFIER
Assembly, Installation and Operating Instructions
SAVE THESE INSTRUCTIONS
NOTE: Check all parts for shipping damage. If damage is found, DO NOT use. Contact Carrier/Invacare for further instructions.
SAFETY SUMMARY
In order to ensure the safe installation, assembly and operation of the POSEIDON HUMIDIFIER these instruc­tions MUST be followed:
CAUTION
Federal law restricts this device to sale by or on order of a physician, or any other practitioner licensed by the law of the State in which he practices to use or order the use of this device.
The Poseidon humidifier was designed to be used with various CPAP units. ALWAYS check with your physician/clinician to determine if it can be used
with your sleep unit/flow
GENERAL WARNINGS
DO NOT install this equipment without first read­ing and understanding this instruction sheet AND your CPAP owner’s manual. If you are unable to understand these instructions, contact Invacare technical support at the number listed on the back page before attempting to install this equip­ment - otherwise, injury or damage may occur.
The humidifier is for SINGLE PERSON USE ONLY. This device is not to be used with CPAP flow
generators that have automatic pressure titra­tion capabilities.
This device will not alter compliance parameters when operated with a Bi-level or CPAP flow generators.
DO NOT allow water from the humidifier to spill through the air tubing into the flow generator as this could result in electrical shock hazard and could result in personal injury or equipment dam­age. To prevent this:
DO NOT overfill the humidifier; Disconnect the tubing at the flow generator
when not in use; DO NOT fill the humidifier while it is
attached to the flow generator; If water does enter the flow generator, switch
OFF the power and disconnect the flow gen­erator. Return it to your equipment supplier for evaluation.
Ensure that all items are properly and securely installed and the humidifier is posi­tioned under the flow generator.
generator.
OVERVIEW OF THE INVACARE POSEIDON HUMIDIFIER
The Poseidon Humidifier helps alleviate symptoms associated with the use of Continuous Positive Airway Pressure (CPAP) treatment. These symptoms include a dry nose and throat. It is intended for use with Invacare nasal CPAP systems as well as many other manufacturers CPAP systems.
The Poseidon Humidifier is a container that is filled with water and placed beneath the flow generator. Air from the flow gen­erator passes over the water in the humidifier picking up mois­ture before delivering it through your nasal cannula/mask. This process helps to relieve irritation of the nose and throat from use of the CPAP system.
The Poseidon Humidifier Kit Contains:
DESCRIPTION QTY.
Humidifier (blue-tinted plastic) 1 Flexible air tubing (short length) 1
FILLING THE HUMIDIFIER (FIGURE 1)
1. Rinse each item of the humidifier kit in clean water and drain completely.
WARNING
DO NOT overfill the humidifier - otherwise, in­jury or damage may occur.
2. Holding the humidifier upright, fill the humidifier with dis­tilled or demineralized water through either port to the level indicated by the fill line.
Inlet Port
FIGURE 1 - FILLING THE HUMIDIFIER
Outlet Port
1
CONNECTING THE HUMIDIFIER TO THE FLOW GENERATOR (FIGURE 2)
1. Place the humidifier securely on a level surface where it is to be used.
2. Position the flow generator on top of the humidifier be­tween the humidifier ports.
3. Push one (1) end of the shorter tubing onto the flow gen­erator outlet.
4. Push the other end of the shorter tubing onto the humidifier inlet port.
5. Push the longer tubing onto the humidifier outlet port.
6. Attach nasal mask to the other end of the longer tubing. Refer to and follow the instructions supplied with nasal mask.
7. Check that all tubing is securely connected to the humidifier. The humidifier is now ready for use.
Flow Generator
Humidifier Outlet
Port
Longer
Tubing
CLEANING AND DISINFECTING
CAUTION
DO NOT use hot water. DO NOT use bleach, chlorine-, alcohol-, or aro-
matic-based solutions on the tubing or con­nector as these could cause deterioration of the materials.
DO NOT hang the tubing in direct sunlight as the tubing may harden over time and even­tually crack.
Daily
Clean the humidifier daily as follows:
1. Turn the sleep unit/flow generator switch OFF.
2. Disconnect all tubing from the sleep unit/flow generator, hu­midifier and patient nasal mask.
3. Use a mild detergent and wash the humidifier and tubing in warm water.
4. Partly fill the humidifier and shake vigorously.
5. Rinse thoroughly with clean water and let air dry.
Weekly
Disinfect the humidifier weekly as follows:
1. Soak the humidifier in a combination of disinfectant solu­tion and cold water for 30 minutes.
2. Rinse thoroughly with clean water and let air dry completely.
Flow Generator
Outlet
Shorter
Tubing
Nasal Mask
Humidifier
Inlet Port
Humidifier
FIGURE 2 - CONNECTING THE HUMIDIFIER TO
THE FLOW GENERATOR
Invacare Corporation www.invacare.com
Respiratory Products Canada
699 Aero Lane 5970 Chedworth Way Invacare is a registered trademark of Invacare Corporation. Sanford, Florida USA Mississauga, Ontario Yes, you can. is a trademark of Invacare Corporation. 32771 L5R 3T9, Canada © (407) 328-5900 (905) 890-8838 Form No. 00-142 Part No. 1102385 Rev B (2) 09/01
Periodically
CAUTION
DO NOT mix the vinegar and water solution with the disinfectant solution.
Calcium deposits may build up inside your humidifier over time. Remove periodically as follows:
1. Soak the humidifier and shorter tubing in a solution of one (1) part vinegar and one (1) part warm water for 30 min­utes.
2. Rinse thoroughly with clean water until the smell of vin­egar has gone. Let air dry completely.
3. Drain any remaining water from the humidifier.
2001 Invacare Corporation