Invacare POC1-100C, POC1-100B User Manual

Page 1
Invacare®Platinum®Mobile
POC1-100B,POC1-100C
enOxygenConcentrator
UserManual
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,thismanualMUSTbereadandsavedfor futurereference.
Page 2
©2018InvacareCorporation Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibited withoutpriorwrittenpermissionfromInvacare.T rademarksareidentiedby™and®.Alltrademarks areownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted. PlatinumisaregisteredtrademarkintheU.S.A.,EuropeandAustralia. MakingLife’sExperiencesPossibleisaregisteredtrademarkintheU.S.A. Pine-sol,LestoilandCloroxareregisteredtrademarksofTheCloroxCompany. DawnisaregisteredtrademarkofTheProcterandGambleCompany. IOSisatrademarkorregisteredtrademarkofCiscointheU.S.andothercountriesandisused underlicensetoApple,Inc. AndroidisatrademarkofGoogle,LLC. BluetoothisaregisteredtrademarkofBluetoothSIG,Inc.
Page 3
Contents
1General.........................................5
1.1QuickStartGuide..............................5
1.2Symbols.....................................9
1.3IndicationsForUse.............................11
1.4IntendedUse.................................11
1.5Contraindications...............................11
1.6Features.....................................12
1.6.1ControlPanel...............................12
1.6.2Input/OutputConnections.....................13
1.6.3Filters....................................13
1.6.4PowerOptions..............................14
1.6.5CarryBag..................................14
1.7OptionalAccessoriesandReplacementParts..........14
2Safety..........................................15
2.1LabelLocations................................15
2.2GeneralGuidelines.............................17
2.3RadioFrequencyInterference.....................22
2.4TravelingwiththeConcentrator....................22
3Setup...........................................25
3.1Unpacking....................................25
3.2Inspection....................................25
3.3PoweringtheOxygenConcentrator.................25
3.3.1InstallingtheBattery(s).......................26
3.3.2ChargingtheBattery(s)........................27
3.3.3ReadingtheBatteryGaugeforInstalledand
UninstalledBattery(s)........................28
3.3.4RemovingtheBattery(s).......................29
3.3.5ExternalACPower...........................30
3.3.6ExternalDCPower...........................30
3.4CarryBagSetup...............................31
3.4.1InstallingtheCarryBag.......................31
3.4.2RemovingtheCarryBag.......................32
3.4.3InstallingShoulderStraps......................32
3.5Connecting/PositioningtheNasalCannula............33
4Usage..........................................35
4.1Location.....................................35
4.2ConcentratorOperatingPosition...................35
4.3UsingtheCarryBag............................35
4.4TurningOntheConcentrator......................36
4.5TurningOfftheConcentrator......................37
4.6WarmUpPeriod...............................37
4.7BreathingwiththeUseoftheConcentrator...........37
4.8ReadingtheDisplayScreen.......................37
4.9AdjustingthePulseFlowSetting...................38
4.10BatteryLifeandManagement....................39
5Maintenance.....................................41
5.1Maintenance..................................41
5.2ServiceLife...................................41
5.3CleaningtheCabinet............................41
5.4CleaningtheAirIntakeFilterScreen................42
5.5CleaningtheCarryBag..........................42
5.6CleaningandDisinfectionBetweenPatients...........43
5.7ViewingHourMeter............................44
5.8PreventativeMaintenanceChecklist.................45
6AfterUse........................................47
6.1Storage......................................47
6.2Disposal.....................................47
6.3WearandT ear................................48
7Troubleshooting...................................49
7.1Troubleshooting...............................49
7.2AlarmConditions...............................51
7.2.1OperationalAlarms..........................52
7.2.2ShutDownAlarms...........................60
8T echnicalData....................................65
8.1TechnicalDescription............................65
8.1.1PneumaticDiagram..........................65
Page 4
8.2Specications.................................66
8.3RegulatoryListing..............................70
8.4ElectromagneticCompliance(EMC).................71
9Warranty........................................76
9.1LimitedWarranty..............................76
Page 5
General
1General
1.1QuickStartGuide
IMPORTANT!
–TheQuickStartGuideisforreferenceONLY .
Itisimperativetoreadthemanualinits ENTIRETYforfullinstructions.
DANGER! RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,death orinjurytopersons:
–DONOTdisassemble.Referservicingto
qualiedservicepersonnel.Therearenouser serviceableparts.
WARNING! RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduse andspecicationshasnotbeentestedandmay leadtoproductdamage,lossofproductfunction, orinjury.
–DONOTusethisproductinanywayotherthan
describedinthespecicationsandintended usesectionsofthismanual.
DANGER! RiskofDeath,InjuryOrDamageFromFire
Textiles,oilorpetroleumsubstances,grease, greasysubstancesandothercombustiblesare easilyignitedandburnwithgreatintensityin oxygenenrichedairandwhenincontactwith oxygenunderpressure.Smokingduringoxygen therapyisdangerousandislikelytoresultin burnsordeath.T oavoidre,death,injuryor damage:
–DONOTSMOKEwhileusingthisdevice.DO
NOTusenearOPENFLAMEorIGNITION SOURCES.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic electricitycreatedbyanytypeoffriction.
–Useonlyoxygencompatiblewater-basedlotions
orsalvesbeforeandduringoxygentherapy. Toverify,refertothelotion/salvecontainer foroxygencompatiblewater-basedstatement. Ifnecessary,contactthemanufacturer.DO NOTuseanylubricantsonconcentratorunless recommendedbyInvacare.
–Keeptheoxygentubing,cord,ACadapter,
andconcentratoroutfromundersuchitems asblankets,bedcoverings,chaircushions, clothing,andawayfromheatedorhotsurfaces includingspaceheaters,stoves,andsimilar electricalappliances.
–Makesureconcentratorisoffwhennotinuse.
1194969-C5
Page 6
Invacare®Platinum®Mobile
DANGER! RiskOfDeath,InjuryOrDamage
Improperuseoftheproductmaycausedeath, injuryordamage.Thissectioncontainsimportant informationforthesafeoperationanduseofthis product.
–DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely readingandunderstandingtheseinstructions andanyadditionalinstructionalmaterialsuch asusermanuals,servicemanualsorinstruction sheetssuppliedwiththisproductoroptional equipment.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare professional,dealerortechnicalpersonnel beforeattemptingtousethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct isnotworkingcorrectly,contactatechnician orInvacareforrepair .
–Thisproductisintendedtobeusedbyadults
onlyafterreadingandunderstandingthe instructionsandwarningsofthisusermanual.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
61194969-C
Page 7
General
1.
Openthebatterycompartmentdoor.Withbattery fuelgaugelabelfacingtowardthefrontoftheoxygen concentrator ,slowlylowerbatteryAstraightdowninto oneofthetwoslotsinthebatterycompartment.
RefertoInstallingtheBattery(s)inSetup.
2.
ConnecttheACPowerAdaptertotheexternalpower inputconnectorB.PlugtheACcordintotheelectrical outlet.
RefertoExternalACPowerinSetup.
Usingtheconcentratorforthersttimerequires thebattery(s)tobecharged.RefertoCharging theBattery(s)inSetup.
3.
Connectthenasalcannulatotheconcentratoroxygen outletportonyourconcentratorC.Oxygenwillonly bedeliveredwhileyouarebreathingthroughthenasal cannulaconnectedtothisdevice.
RefertoConnecting/PositioningtheNasalCannulain Setup.
1194969-C
7
Page 8
Invacare®Platinum®Mobile
4.
TurnontheconcentratorbypressingtheOn/OffButton Duntilthecontrolpanelindicatorslightup.
RefertoTurningontheConcentratorinUsage.
5.
Adjustthepulseowsettingtotheoneprescribedby yourclinicalprofessional.Thepulseowsettingcanbe changedbypressingtheowselectionbuttonE.Place thenasalcannulaoveryourearsandpositiontheprongs inyournoseasinstructedbyyourhealthcareprovider.
RefertoAdjustingthePulseFlowSettinginUsage.
81194969-C
Page 9
General
1.2Symbols
Signalwordsareusedinthismanualandapplytohazards orunsafepracticeswhichcouldresultinpersonalinjury orpropertydamage.Refertotheinformationbelowfor denitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardous
situationwhich,ifnotavoided,willresultin deathorseriousinjury .
WARNING!
–Warningindicatesapotentiallyhazardous
situationwhich,ifnotavoided,couldresultin deathorseriousinjury .
CAUTION!
–Cautionindicatesapotentiallyhazardous
situationwhich,ifnotavoided,mayresultin propertydamageorminorinjuryorboth.
IMPORTANT!
–Importantindicatesahazardoussituationthat
couldresultindamagetopropertyifitisnot avoided.
Givesusefultips,recommendationsandinformation forefcient,trouble-freeuse.
Refertothefollowingtableforthemeaningofsymbols markedontheequipmentand/orpackaging.
GeneralWarningSign
Thebackgroundcolorinsidethe triangleisyellowonproductlabels.
FollowInstructionsforUse
Thecolorofthesymbolbackgroundis blueonproductlabels.
Noopename
Thecolorofthecirclewithdiagonal barisredonproductlabels.
DONOTSmoke
Thecolorofthecirclewithdiagonal barisredonproductlabels.
ConsultInstructionsforUse
Thissymbolislocatedonthecannula packing.
DirectCurrent
TypeBFAppliedPart
1194969-C9
Page 10
Invacare®Platinum®Mobile
ClassIIEquipment
PowerOn/Off
DataOutputTerminal
DONOTdisposeofinhouseholdwaste
Manufacturer
DateofManufacture
Recycle
ReferenceNumber
SerialNumber
DONOTReuse(SinglePatientUseOnly)
Thissymbolreferstothecannula whichisforsinglepatientuseonly.
IndoorUseONL Y
ThissymbolisincludedontheAC poweradapter.
DONOTuseoilorgrease
Keepdry
Protectedagainstsolidforeignobjectsof12.5 mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops whenenclosuretiltedupto15°.
Protectedagainstsolidforeignobjectsof12.5 mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops.
TransportandStorageT emperature
TransportandStorageHumidity
ElectricalSafetyAgencyCerticationMark
LatexFree
101194969-C
Page 11
General
1.3IndicationsForUse
TheInvacare®Platinum®MobileOxygenConcentratoris intendedtoprovidesupplementaloxygentopatientswith respiratorydisorders.Thedevicecanbeusedinahome, institution,vehicle,orotherenvironmentsoutsidethehome.
TheInvacareMobileMedicalApplicationAccessory POC1-CONNECTisintendedforusewiththeInvacare PlatinumMobileOxygenConcentratorPOC1-100Bor POC1-100Cdevice,toallowpatientsviatheirAndroid™or iOS™mobilephoneortablettodisplaydevicesettings,and tocollectdeviceperformanceandusageinformationfor maintenance/servicingpurposesonly.
1.4IntendedUse
TheInvacare®Platinum®MobileOxygenConcentratoris intendedtoprovidesupplementaloxygentopatientswith respiratorydisorders.TheInvacare®Platinum®Mobile OxygenConcentratorcanbeusedinahome,institution, vehicle,orotherenvironmentsoutsidethehome.The deviceisnotintendedtobelife-supportingorlife-sustaining.
TheInvacareMobileMedicalApplicationAccessory POC1-CONNECTisintendedtoallowpatientsviatheir AndroidoriOSmobilephoneortablettodisplaydevice settings,andtocollectdeviceperformanceandusage informationformaintenance/servicingpurposesonly.
WARNING! RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduse andspecicationshasnotbeentestedandmay leadtoproductdamage,lossofproductfunction, orpersonalinjury.
–DONOTusethisproductinanywayotherthan
describedinthespecicationsandintended usesectionsofthismanual.
1.5Contraindications
WARNING! RiskofInjury
Thisproductistobeusedasanoxygen supplementandisnotintendedtobe life-supportingorlife-sustaining.ONLYusethis productifthepatientiscapableofspontaneous breath,abletoinhaleandexhalewithouttheuse ofamachine
–DONOTuseinparallelorserieswithother
oxygenconcentratorsoroxygentherapy devices.
1194969-C
11
Page 12
Invacare®Platinum®Mobile
WARNING! RiskofMinorInjuryorDiscomfort
Theconserving,orpulsedose,oxygendelivery techniqueusedbythisdeviceiscontraindicated inpersonswhosebreathingduringnormalresting wouldbeunabletotriggerthedevice.
–Properdevicetriggering,setupandoperation
–Notforpediatricuse. –Notforusebytracheotomizedpatients.
1.6Features
1.6.1ControlPanel
mustbeconrmedbyanexperiencedclinician orotherrespiratoryprofessional.
On/OffButton
A
Pressforonesecondtoturnconcentratoronoroff.
FlowSelectionButton
Presstoadjustthepulsedoxygenowtoyour
B
prescribedsetting.Theselectedowsettingis shownonthedisplayscreen.RefertoAdjustingthe PulseFlowSettinginUsageformoredetails.
DisplayBacklightButton
C
Pressoncetoilluminatethedisplayscreenfor5 seconds.
AudioOffButton
Presstomutetheaudiblesignal(beep)forboth alarmsandstatusindications.Theblueindicator
D
lightnexttothebuttonilluminateswhenactivated. Audiooffcanbeactivatedduringorpriortoan alarmconditionoccurring.
DisplayScreen
E
Displaysinformationabouttheoperatingstatusof theconcentrator .
AlarmIndicator
Theyellowindicatorlightatthecenterofthe
F
triangularsymbolilluminatestoindicateabnormal operatingconditions.RefertoAlarmConditionsin Troubleshootingformoredetails.
12
1194969-C
Page 13
General
BatteryGauge
Indicatesthechargestatusofthebattery(s).Refer
G
toReadingtheBatteryGaugeforInstalledand UninstalledBattery(s)inSetupformoredetails.
ExternalPowerIndicator
Theorangeindicatorlightnexttotheplugsymbol
H
illuminateswhenanexternalpowersourceis connectedtotheconcentrator .
BatteryCompartmentDoor
I
Usedtoaccesstheremovablebattery(s).
AudibleSignal(Beep)
Not
Shown
Indicatesachangeinoperatingstatusor aconditionwhichrequirestheoperator’s attention.
1.6.2Input/OutputConnections
A
B
C
OxygenOutletPort
Usedtoconnectthenasalcannulatothe concentrator .
ExternalPowerInputConnector
Usedtoconnectanexternalpowersourcetothe concentrator .
USBPort
Usedforservice.LocationofPOC1–CONNECT Dongle.
1.6.3Filters
Airenterstheconcentratorthroughanairintakelter locatedonthecarrybag.Thislterpreventshairandother largeparticlesintheairfromenteringtheunit.Beforeyou
1194969-C13
Page 14
Invacare®Platinum®Mobile
operatetheconcentrator,makesurethislteriscleanand dry,andtheconcentratorisproperlyinstalledinthecarry bag.
Tocleantheairintakelter ,refertoCleaningtheAirIntake FilterScreeninMaintenance.
Toensuretheconcentratorisproperlyinstalledinthecarry bag,refertoInstallingtheCarryBaginSetup.
1.6.4PowerOptions
WARNING! RiskofInjuryorDamage
Toavoidinjuryordamagewhichwillvoid warranty:
–UseonlyInvacarespeciedpowersupplies.
Battery(s):Uptotworechargeablebatteriescanbeinstalled inthebatterycompartmentoftheconcentrator .Whenfully charged,asinglebatterysuppliespowerforuptovehours (uptotenhoursfortwobatteries).Visibleandaudible alarmsignalsoccurwhenthebatterypowerisgettinglow. RefertoAlarmConditionsinTroubleshootingandCharging theBattery(s)inSetup.
ACPowerAdapter:AnACpoweradapterallowsthe concentratortobeconnectedtoa100–240volt50–60 hertzoutlet.UseoftheACpoweradapterwillallowthe concentratortooperateandsimultaneouslyrechargethe installedbattery.RefertoChargingtheBattery(s)inSetup.
theconcentratortooperateandsimultaneouslyrechargean installedbattery(s).RefertoChargingtheBattery(s)inSetup.
1.6.5CarryBag
Thecarrybagprovidesaconvenientmeanstoholdthe concentratorwhileperformingdailyactivities.
Itcanbeconguredasamessengerbagorbackpackusing theshoulderstrapsprovided.RefertoInstallingtheCarry BaginSetup.
1.7OptionalAccessoriesandReplacementParts
Thefollowingoptionalaccessoriesareavailablefrom Invacare:
Additionalbatteries,modelnumber:POC1–110
Externalbatterychargerwithpowersupply:POC1–115
ThefollowingreplacementpartsareavailablefromInvacare:
Battery,modelnumber:POC1–110
CarryBag,modelnumber:POC1–150
DCpowercable,modelnumber:POC1–140
ACpoweradapterwithpowercord:POC1–130
USBDongle,modelnumber:POC1–CONNECT
Thesepartsarebasemodels.PleasecontactInvacare oryourproviderforcountryspecicmodels.
DCPowerCable:ADCpowercableallowstheconcentrator tobeconnectedtoanautomobile’s(boat,motorhome, etc.)12-voltDCoutlet.UseoftheDCpowercablewillallow
14
1194969-C
Page 15
2Safety
2.1LabelLocations
Thelabelislocatedonthebackoftheconcentrator.Theserialnumberislocatedontherightsideoflabelshownhere.
ModelPOC1-100B
Theconcentratorhasfourpulseowsettings.
Safety
1194969-C15
Page 16
Invacare®Platinum®Mobile
ModelPOC1-100C
Theconcentratorhasvepulseowsettings.
161194969-C
Page 17
Safety
2.2GeneralGuidelines
Inordertoensurethesafeinstallation,assemblyand operationoftheconcentratortheseinstructionsMUSTbe followed.
DANGER! RiskOfDeath,InjuryOrDamage
Improperuseoftheproductmaycausedeath, injuryordamage.Thissectioncontainsimportant informationforthesafeoperationanduseofthis product.
–DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely readingandunderstandingtheseinstructions andanyadditionalinstructionalmaterialsuch asusermanuals,servicemanualsorinstruction sheetssuppliedwiththisproductoroptional equipment.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare professional,dealerortechnicalpersonnel beforeattemptingtousethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct isnotworkingcorrectly,contactatechnician orInvacareforrepair .
–Thisproductisintendedtobeusedbyadults
onlyafterreadingandunderstandingthe instructionsandwarningsofthisusermanual.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER! RiskofDeath,InjuryOrDamageFromFire
Textiles,oilorpetroleumsubstances,grease, greasysubstancesandothercombustiblesare easilyignitedandburnwithgreatintensityin oxygenenrichedairandwhenincontactwith oxygenunderpressure.Smokingduringoxygen therapyisdangerousandislikelytoresultin burnsordeath.T oavoidre,death,injuryor damage:
–DONOTSMOKEwhileusingthisdevice.DO
NOTusenearOPENFLAMEorIGNITION SOURCES.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Keepallopenames,matches,lighted
cigarettes,electroniccigarettesorothersources ofignitionatleast10ft(3m)awayfromthis concentratororanyoxygencarryingaccessories suchascannulasortanks.
–Ifyoudisregardthesewarningsaboutthe
severehazardofoxygenusewhileyoucontinue tosmoke,youmustalwaysturnoffthe concentrator ,removethecannulaandthen waittenminutesbeforesmokingorleavethe roomwhereeithertheconcentratororany oxygencarryingaccessoriessuchascannulas ortanksarelocated.
–DONOTSMOKEwhileusingtheconcentrator .
1194969-C
17
Page 18
Invacare®Platinum®Mobile
CAUTION!
Federal(statutory)lawrestrictsthisdeviceto salebyorontheorderofamedicalpractitioner licensedbyagovernmentalagencywherehe/she practices.
–ONL Yalicensedmedicalpractitionermayorder
thepurchaseoruseofthisdevice.
DANGER! RiskofDeath,InjuryOrDamageFromFire
Textiles,oilorpetroleumsubstances,grease, greasysubstancesandothercombustiblesare easilyignitedandburnwithgreatintensityin oxygenenrichedairandwhenincontactwith oxygenunderpressure.Toavoidre,death, injuryordamage:
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic electricitycreatedbyanytypeoffriction.
–Useonlyoxygencompatiblewater-basedlotions
orsalvesbeforeandduringoxygentherapy. Toverify,refertothelotion/salvecontainer foroxygencompatiblewater-basedstatement. Ifnecessary,contactthemanufacturer.DO NOTuseanylubricantsonconcentratorunless recommendedbyInvacare.
–Keeptheoxygentubing,cord,ACadapter,
andconcentratoroutfromundersuchitems asblankets,bedcoverings,chaircushions, clothing,andawayfromheatedorhotsurfaces includingspaceheaters,stoves,andsimilar electricalappliances.
–Makesureconcentratorisoffwhennotinuse.
181194969-C
Page 19
Safety
CAUTION! RiskofMinorInjuryorDiscomfort
Itisimportanttoplanaheadfortravelandother situationsinwhichyoumaynothaveaccessto additionaloxygenorpowersupplies.T oprevent oxygendeprivation:
–Carrysupplementalbatterieswithyou. –Stowtheconcentratorproperlyduringtravel.
Forproperstorageinstructionswhentravelling refertothetravelguide.
WARNING! RiskofInjuryorDamage
Topreventinjuryordamagefrommisuse:
–NEVERleaveconcentratorunattendedwhen
connectedtopower. –Makesureconcentratorisoffwhennotinuse. –OutdooruseMUSTbeconductedwithinternal
batterypoweronly.
WARNING! RiskOfInjuryOrDamage
–Invacareproductsarespecicallydesigned
andmanufacturedforuseinconjunction
withInvacareprovidedorInvacarespecied
accessories.Anyotheraccessorieshave
notbeentestedbyInvacareandarenot
recommendedforusewithInvacareproducts. –Nomodicationofthisequipmentisallowed.
CAUTION! RiskofMinorInjuryorDiscomfort
Achangeinaltitudemayaffecttotaloxygen availabletoyou.Topreventoxygendeprivation:
–Consultyourphysicianbeforetravelingto
higherorloweraltitudestodetermineifyour owsettingsshouldbechanged.
DANGER! RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,death orinjurytopersons:
–DONOTdisassemble.Referservicingto
qualiedservicepersonnel.Therearenouser serviceableparts.
–Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the concentratormustbelocatedinanotherroom atleast7ft(2.5m)fromthebath.
–DONOTcomeincontactwiththeconcentrator
whilewet.
–DONOTplaceorstoreconcentratorwhereit
candropintowaterorotherliquid.
–DONOTreachforconcentratorthathasfallen
intowater .UnplugIMMEDIATELY.
–DONOTblockaccesstopoweroutletneeded
tounplugtheACpowercord. –DONOTusefrayedordamagedpowercords. –DONOTuseACpoweradapterifitshousingis
crackedorseparated.
1194969-C19
Page 20
Invacare®Platinum®Mobile
WARNING! RiskofInjuryorDeath
Topreventinjuryordeathfromproductmisuse:
–Closelysupervisewhenthisconcentratoris
usedbyimpairedindividualsornearchildren and/orimpairedindividuals.
–Monitorpatientsusingthisdevicewhoare
unabletohearorseealarmsorcommunicate discomfort.
CAUTION! RiskofMinorInjuryorDiscomfort
WhileInvacarestrivestoproducethebestoxygen concentratorinthemarkettoday,thisoxygen concentratorcanfailtoproduceoxygendueto powerfailureordevicemalfunction.
–ALWA YShaveabackupsourceofoxygenreadily
available.
–Intheeventtheconcentratorfailstoproduce
oxygen,theconcentratorwillbrieyalarm signalingthepatienttoswitchtotheirbackup sourceofoxygen.RefertoT roubleshootingfor moredetail.
–Invacarerecommendskeepingatleastone
batteryinstalledintheconcentratorevenwhen operatingfromanexternalpowersource.
WARNING! RiskofDeath,InjuryorDamage
Topreventinjuryordamagefromcordmisuse:
–DONOTmoveorrelocateconcentratorby
pullingonthecord. –DONOTuseextensioncordswithACpower
adaptersprovided. –Properlystoreandpositionelectricalcords
and/ortubingtopreventatrippingand
strangulationhazards.
WARNING! RiskofInjuryorDamage
Topreventinjuryordamageduringuse:
–Ifyoufeelilloruncomfortable,orifthe
concentratordoesnotsignalanoxygenpulse
andyouareunabletohearand/orfeelthe
oxygenpulse,consultyourequipmentprovider
and/oryourphysicianIMMEDIA TEL Y . –Foroptimumperformance,Invacare
recommendsthateachuseoftheconcentrator
beaminimumof30minutes.Shorterperiods
ofoperationmayreducemaximumproductlife. –Theconcentratorshouldbeusedinanupright
position. –Theconcentratorcannotbeusedinconjunction
withPAP ,Bi-Level,mechanicalventilatoror
othersuchdevices.
201194969-C
Page 21
Safety
CAUTION! RiskofMinorInjury,DiscomfortorDamage
–Useofthisdeviceatanaltitudeabove10,000
ft(3048m)oroutsideatemperatureof41°F to104°F(5°Cto40°C)orarelativehumidity above90%isexpectedtoadverselyeffectthe owrateandthepercentageofoxygenand consequentlythequalityoftherapy.
WARNING! RiskofInjuryorDamage
Asasafetyfeature,thisappliancemayhavea polarizedplug(onebladeiswiderthantheother). Toavoidinjuryordamagefromelectricalshock:
–Thisplugwilltinapolarizedoutletonlyone
way.Iftheplugdoesnottfullyintheoutlet, reversetheplug.Ifitstilldoesnott,contact
aqualiedelectrician. –DONOTalterthepluginanyway. –DONOTuseifthecordorplugaredamaged.
Callaqualiedelectricianforrepair .
CAUTION! RiskofDamage
Topreventdamagefromtemperaturevariations:
–DONOToperateintemperaturesbelow41°F
(5°C)orabove104°F(40°C).
–Whenyourautomobileisturnedoff,disconnect
thecaraccessorypowercordandremove theconcentratorfromtheautomobile.DO NOTstoretheconcentratorinaveryhot orcoldautomobileorinothersimilar,high orlow,temperatureenvironments.Referto SpecicationsinTechnicalData.
WARNING! RiskofInjuryorDamage
Topreventdamagefromliquidingress:
–Iftheconcentratorisnotworkingproperly,if
ithasbeendroppedordamaged,ordropped intowater ,callequipmentprovider/qualied technicianforexaminationandrepair.
–NEVERdroporinsertanyobjectorliquidinto
anyopening.
–Invacarerecommendstheconcentratornotbe
usedintherain.
–Theconcentratorisnotdesignedtobeused
withahumidier .Useofthisdevicewitha humidiermayimpairperformanceand/or damagetheequipment.
–DONOTuseifcabinetofconcentratoror
housingofACpoweradapteriscrackedor separated.
1194969-C
21
Page 22
Invacare®Platinum®Mobile
2.3RadioFrequencyInterference
MedicalElectricalEquipmentneedstobeinstalledandused accordingtotheEMCinformationinthismanual.
Thisequipmenthasbeentestedandfoundtocomplywith EMClimitsspeciedbyIEC/EN60601-1-2.Theselimits aredesignedtoprovideareasonableprotectionagainst electromagneticinterferenceinatypicalhomehealthcare environment.
PortableandmobileRFcommunicationsequipmentcan affecttheoperationofthisequipment.Useofthis equipmentadjacenttoorstackedwithotherequipment shouldbeavoidedasitcouldresultinimproperoperation. Theconcentratoroperationshouldbeobservedtoverify normaloperationinthepresenceofsuchequipment.
Otherdevicesmayexperienceinterferencefromeven thelowlevelsofelectromagneticemissionspermittedby theabovestandards.T odetermineiftheemissionsfrom theconcentratorarecausingtheinterference,turnthe concentratoroff.Iftheinterferencewiththeotherdevice(s) stops,thentheconcentratoriscausingtheinterference.In suchrarecases,interferencemaybereducedorcorrected byoneofthefollowingmeasures:
Reposition,relocate,orincreasetheseparationbetween theequipment.
Connecttheequipmentintoanoutletonacircuit differentfromthattowhichtheotherdevice(s)is connected.
2.4TravelingwiththeConcentrator
BeforeTraveling
1.Contactyourtravelcarrierand/ortourguideabout yourtravelplansandprovideinformationaboutyour intenttouseaportableoxygenconcentratorduring yourtravels.Thecarrierwillletyouknowifyouwillbe permittedtouseyourconcentratorduringyourtravel andifthereareanyrestrictionsforuse.
Ifyouarenotpermittedtouseyourconcentrator onthecarrierofchoiceandstillwanttotake itwithyou:
Carryitandanysupplementalbatterieson boardwithyou.
Ensuretheunitisturnedoff.
Stowitproperlyforthetrip.
DONOTplacetheconcentratororany supplementalbatteriesincheckedbaggage.
2.Contactthecarriertoseeiftheyhaveseatswitha poweroutletthatyoucanusetopowertheconcentrator duringtravel.Ifapoweroutletisnotavailable,besure tobringenoughsupplementalbatteriestolast150%of thetraveltime.
Theremaybeunexpecteddelaysbeyond scheduledtraveltimes.Carryyourpower adapterswithyousoyoucanrechargebatteries beforeandafterthetrip.
RefertoElectromagneticCompliance(EMC)inT echnical Dataforadditionalinformation.
22
1194969-C
Page 23
Safety
3.Chargetheinstalledbatteryoftheconcentratorand anysupplementalbatteriesfullybeforeyoudepart,thus extendingtheoperatingtimeoftheconcentratorduring travel.RefertoChargingtheBattery(s)inSetup.
4.Ensuretheunitisfreeofallgrease,oil,orother petroleumproductsandthattheunitisingoodworking order,isfreeofdamage,andtheairlterisclean. RefertothePreventativeMaintenanceChecklistin Maintenance.
5.Contactyouroxygensuppliertomakearrangementsif youneedtohavebackupoxygenatyourdestination.
TravelingByAir
ArrivingattheAirport
1.Allowthesecurityagenttoinspectyourconcentrator , evenifyouareusingit,whengoingthroughthesecurity checkpoint.
Theyarerequiredtoallowyoutotravelthrough thecheckpointwithyourconcentrator,butthey MUSTinspectitforsecurityreasons.
2.UseyourACpoweradapterwhileyouareintheairport, ifpossible.Thiswillkeepthebattery(s)atfullcharge andgiveyouthemostbatterypoweredoperatingtime ontheaircraft.
1.Ifyouplanonusingtheconcentratortheentiretime youareontheplane,verifythatyourconcentratordoes notblockaccesstoanemergencyexitoraisleway.Ifit does,requestaseatchange.
2.Inorderfortheightcrewtoensurethesafetyofyou andalloftheotherpassengersontheight,allowthem toinspecttheconcentratortoconrmthatitisFAA approvedforuseontheight.
3.Ifyouarenotplanningtouseyourconcentrator duringtaxi,takeoff,orlanding,youMUSTstoreitin anapprovedstorageareasoitdoesnotblocktherow oraisleaccess.
Contacttheightcrewaboutlocationof approvedstorageareas.
OnthePlane
YouMUSTstorethesupplementalbatteryeither securelyconnectedtotheconcentratororinyour carry-onbaggage.ThesupplementalbatteryMUST beprotectedfromdamageandshortingoutthe externalpowerconnector.
BoardingthePlane
Youarenotpermittedtositinanexitrowifyou plantouseyourconcentratoratanytimeduring theight.
1194969-C23
Page 24
Invacare®Platinum®Mobile
1.Youmayuseyourconcentratorwhenmovingabout thecabin,ONL Yafterthepilotturnsoffthe“Fasten SeatBelt”sign.
2.Turnofftheunitiftheconcentratoralarmsduringthe ight,unlessthealarmresetsitself.Ifthealarmdoes notreset,theconcentratorMUSTbeturnedoffand storedinanapprovedstoragearea.
3.IfanACpoweroutletwithaminimum100Wcapacityis availableontheight,performthefollowing:
a.Connectthepoweradaptertotheconcentrator . b.Pluginthepoweradapterintotheairlinepower
outlet.
c.Turntheuniton.
Ifyouhaveanytroublemakingtheconnectionsto theplane’spoweroutlet,contacttheightcrewfor assistance.
AftertheFlight
1.Ensureyouhaveenoughpowertorunyourconcentrator whileleavingtheairport.Rechargethebattery(s),if necessary.RefertoChargingtheBattery(s)inSetup.
2.Contactyourmedicalgassuppliertoobtainyourbackup oxygensupply,ifnecessary .
TravelingbyBoat
TravelingbyTrain
Contactthetrainauthoritiesatleasttwelvehoursaheadof arrivalthatyouaregoingtouseyourconcentrator.There shouldbenorestrictionsonyouruseoftheconcentrator , butpoweronthetrainmaynotbeavailableforyouruse duringtraveltimes.Ensurethereisenoughbatterylifefor yourtripbeforeyouleave.
TravelingbyBus
Contactthebuslineaboutusingaconcentrator.There shouldbenorestrictionsonyouruseoftheconcentrator duringyourtrip,butpowermaynotbeavailableforyour useduringtraveltimes.Ensurethereisenoughbatterylife foryourtripbeforeyouleave.
TravelingbyCar
Thereshouldbenorestrictionsonyouruseofthe concentratorduringyourtrip,butpowermaynotbe availableforyouruseduringtraveltimes.Ensurethereis enoughbatterylifeforyourtripbeforeyouleave.
Contactthecruiselineandinformthemyouaretraveling withaconcentrator .Thereshouldbenorestrictionsonyour useoftheconcentratorduringyourtravel,butitiswiseto checkwiththemaheadoftime.Powershouldbeavailable foruseofyourACpoweradapterduringyourtravels,but checkrst.
24
1194969-C
Page 25
Setup
3Setup
3.1Unpacking
1.Checkforanyobviousdamagetothecartonorits contents.Ifdamageisevident,notifythecarrieror yourlocaldealer .
2.Removeallloosepackingfromthecarton.
3.Carefullyremoveallthecomponentsfromthecartons. TheInvacareconcentrator’spackagingcontainsthe followingitems(asshownbelow)inadditiontothis usermanual.Ifanypartsaremissing,pleasecontact yourequipmentprovider.
Concentrator(ModelPOC1–100B)withcarrybag (ModelPOC1–150)installed
Cannula(Westmed#0194)
Twoshoulderstraps(PartNumber1187483)
Battery(ModelPOC1–110)
ACpoweradapter(PartNumber1187452)
ACpowercord(PartNumber1187454formodel POC1–100B)
DCpowercable(ModelPOC1–140)
Retainallcontainersandpackingmaterialsfor storageorreturnshipment.
3.3PoweringtheOxygenConcentrator
WARNING! RiskofInjuryorDamage
Toavoidinjuryanddamagethatwillvoidthe warranty:
–UseonlyInvacarespeciedpowersupplieswith
theoxygenconcentrator .
–PriortousingtheDCpowercable,the
automobile(boat,motorhome,etc.)engine shouldberunninginordertoeitheroperateor chargetheoxygenconcentrator.
IMPORTANT
Theconcentratorwillnotturnonunlessthe batteryischarged.
–ConnectACpowertotheconcentratorand
allowbatterytofullychargepriortorstuse.
Theoxygenconcentratorallowsthefreedomtochoosefrom thefollowingpowersourcestoenableusebothinsideand outsideofthehome:
Rechargeablebattery(s)
ACpoweroutlet
12voltDCpowercable
3.2Inspection
Inspect/examineexterioroftheoxygenconcentratorand accessoriesfordamage.Inspectallcomponents.Ifdamage isfound,DONOTuseconcentrator.Contactyourprovider forserviceand/orrepairorforassistancewiththesetup procedureandtoreportunexpectedoperationorevents.
1194969-C25
Page 26
Invacare®Platinum®Mobile
3.3.1InstallingtheBattery(s)
DANGER! RiskofDeath,InjuryorDamage
Toavoiddeath,injuryordamagefromre:
–DONOTheatabove140°F(60°C),incinerate,
disassemble,orshortterminals.
–Disposeinaccordancewithalllocalregulations.
Theoxygenconcentratorcomesequippedwithasingle rechargeablelithiumbattery.Uptotwobatteriescanbe installedintheconcentrator .Whenfullycharged,asingle batterysuppliespowerforuptovehours,dependingon thepulseowsetting.Twobatteriescansupplypowerfor uptotenhourswhenfullycharged,dependingonthepulse owsetting.
Abatterycanbeinstalledwhiletheconcentratorisoperating fromanexternalpowersourceorasecondchargedbattery.
1.OpenthebatterycompartmentdoorF.
2.WithbatteryfuelgaugelabelAfacingtowardthefront oftheoxygenconcentratorB,slowlylowerbatteryC straightdownintooneofthetwoslotsinthebattery compartmentD.
UsetheliftringEtoholdthebatterywhile installingit. Thebatterywillnottproperlyinthebattery compartmentifnotorientedasdescribedabove.
261194969-C
Page 27
Setup
3.Ensurethebatteryfullyengagesthebatteryconnections locatedatthebottomofthebatterycompartment. Whenfullyinserted,thetopofthebatteryislevelwith thetopedgeofthebatterycompartment.
CAUTION! RiskofDamage
Toavoiddamagetobatteryconnectionsfrom excessiveforceormisuse:
–DONOTslamorforcebattery(s)intoplace. –Orientthebattery(s)asdescribedinthe
instructionspriortoinstalling.
Toavoiddamagetothebatterydoorfrom excessiveforceormisuse:
–DONOTliftconcentratorusingthebattery
door.
–Makesurebattery(s)arefullyinstalled
beforeclosingthisbatterydoor.
–DONOTuseconcentratorwithbattery
dooropen.
4.Ifasecondbatteryisbeingused(soldseparately),install itinthesamemannerintheunoccupiedbatteryslot.
5.ClosethebatterycompartmentdoorF.
3.3.2ChargingtheBattery(s)
upthebatteries.Tochargethebattery(s),performthe followingsteps:
1.Installthebattery.RefertoInstallingtheBattery(s)in Setup.
2.ConnectACpowertotheconcentrator .RefertoExternal ACPowerinSetup.
3.Monitorthebatterychargelevelandchargeuntilfull. RefertoBatteryGaugeforInstalledandUninstalled Battery(s)inSetup.
4.DisconnecttheACpoweradapterfromtheconcentrator forportableoperation.
IMPORTANT
Aconcentratorwillnotturnonunlessthebattery ischarged.
–ConnecttheACpowertotheconcentratorand
allowbatterytofullychargepriortorstuse.
–RefertoSpecicationsinT echnicalDatafor
approximatechargetimes.
ChargingtheBattery(s)AfterInitialCharging
Theinstalledbattery(s)canbechargedwhetherthe concentratorisonoroff.
InitialChargingoftheBattery(s)
Usingtheconcentratorforthersttimerequiresthe battery(s)tobecharged.Batteriesareshippedina“sleep” modeandaninitialchargecycleisrequiredto“wake”
1194969-C
27
Page 28
Invacare®Platinum®Mobile
1.Forchargingofthebattery(s)afterinitialcharging,do oneofthefollowing:
ConnecttheACpoweradapter(ifchargingfroma walloutlet).RefertoExternalACPowerinSetup.
ConnecttheDCpowercable(ifchargingfroma vehicle).RefertoExternalDCPowerinSetup.
Removebatteryandchargewiththeoptional ExternalBatteryChargeraccessory,Model POC1–115.
2.Monitorthebatterychargelevelandchargeuntilfull. RefertoBatteryGaugeforInstalledandUninstalled Battery(s)inSetup.
3.Disconnectexternalpowerfromtheconcentratorfor portableoperation.
3.3.3ReadingtheBatteryGaugeforInstalledand
UninstalledBattery(s)
BatteryGaugeforInstalledBattery(s)
ThebatterygaugeAshowsthestatusoftheinstalled batterychargelevelwhentheconcentratoristurnedonand whentheconcentratorisoffbutconnectedtoanexternal powersource.ThenumberofilluminatedsegmentsB ofthebatterygaugeAindicatesthechargelevelofthe installedbattery .Iftwobatteriesareinstalledtheindicated chargelevelisthatoftheleastchargedbattery.
NumberofLitSegmentsB
1
2
3
4
ChargeLevel
0–25%charge
26–50%charge
51–75%charge
76%-100%charge
Whentheconcentratorisconnectedtoanexternalpower source(whethertheconcentratoristurnedonoroff)the “last”or“highest”segmentBofthebatterygaugeAalso
281194969-C
Page 29
Setup
ashestoshowthebatteryischarging.Whenthebatteryis
3.3.4RemovingtheBattery(s)
fullycharged,thefourthsegmentBofthebatterygaugeA stopsashingandremainslit.
BatteryGaugeforUninstalledBattery(s)
1.OpenthebatterycompartmentdoorFofthe concentratorB.
2.PullstraightupontheliftringEtoremovethebattery CfromthebatterycompartmentD.
3.Ifasecondbatteryisbeingused(soldseparately), removeitinthesamemanner .
4.Closethebatterycompartmentdoor.
Asteadyaudiblebeepwilloccurwhenallpower sourcesareremovedfromtheconcentrator .Referto ShutDownAlarmsinTroubleshooting.
WhenthebatteryDisremovedfromtheconcentrator , pressthebatterystatusbuttonCtoshowthestatusofthe chargelevelonthebatterygaugeA.Refertothecharge levelBtableinReadingtheBatteryGaugeforInstalledand
Abatterycanberemovedwithoutaffectingtheoperation oftheconcentratoraslongasanexternalpowersourceor asecondchargedbatteryisinstalled.
UninstalledBattery(s)inSetup.
1194969-C29
Page 30
Invacare®Platinum®Mobile
3.3.5ExternalACPower
TheACpoweradapterAallowstheoxygenconcentratorto beconnectedtoa100–240volt50–60Hertzoutlet(i.e. walloutletofyourhome).ThepoweradapterconvertsAC voltagetoaDCvoltagethatcanbeusedtopowerthe oxygenconcentrator .UseoftheACpoweradapterwillallow theoxygenconcentratortooperateandsimultaneously rechargethebattery(s).
CompletethefollowingstepstoconnecttheACpower adaptertotheconcentrator:
1.Ifnotalreadyassembled,attachtheACpowercordE totheACpoweradapterA.
2.InserttheACpoweradapteroutletconnectorBintothe externalpowerinputconnectorCoftheconcentrator .
3.InsertplugDintoanACoutlet.
Theexternalpowerinputconnectorofthe concentratorisangledslightlyupward.Theoutlet connectorofthepowercordneedstobeangled slightlyupwardwheninstalled.
Whenabatteryisinstalled,externalACorDCpower canbeconnectedordisconnectedwhetherthe concentratorisofforon.
Whenanexternalpowersourceisproperly connectedtotheconcentrator,theorangeindicator lightnexttotheplugsymbolonthecontrolpanel willilluminateandtherewillbeanaudiblebeep. Whenexternalpowerisdisconnected,theorange lightwillturnoffandtherewillbeanaudiblebeep.
3.3.6ExternalDCPower
TheDCpowercableGallowstheoxygenconcentratorto beconnectedtoanautomobile’s(boat,motorhome,etc.) 12–voltDCoutlet.UseoftheDCpowercablewillallow
301194969-C
Page 31
Setup
theoxygenconcentratortobeoperatedandsimultaneously
3.4.1InstallingtheCarryBag
rechargethebattery.
CompletethefollowingstepstoconnecttheDCpowercable totheconcentrator:
1.InserttheDCpowercableoutletconnectorFintothe externalpowerinputconnectorCoftheconcentrator .
2.InsertplugHintoaDCoutletaftertheautomobile (boat,motorhome,etc.)isrunning.
3.4CarryBagSetup
CAUTION! RiskofDamage
–Improperinstallationoftheconcentratorinto
thecarrybagcancauseahightemperature alarmconditionandresultintheconcentrator shuttingdown.
–Thecarrybagisnotdesignedtoprotectthe
concentratorfromexcessivedrops,impacts, orabuse.
–T oinstallorremovethecarrybagthe
concentratormustbeoffanddisconnected fromexternalpowerandcannula.
Concentratorcanbeoperatedwithorwithout thecarrybag.Whenoutofthecarrybag,the concentratorislimitedtostationaryuse.Referto UsingtheCarryBaginUsage.
1194969-C31
1.Fullyunzipthebaseofthecarrybag.
2.Positiontheconcentratorstandinguprightonaat surfaceandorientedwithairintakeopeningAfacing you.Ensuretheconcentratorisoffanddisconnected fromexternalpowerandcannula.
3.Positionthecarrybagabovetheconcentratorwiththe airintakelterscreenBfacingyou.
4.HoldtheunzippedbottomofthecarrybagopenC, andslidethecarrybagdownontotheconcentrator untilthecarrybagcompletelycoverstheconcentrator .
Page 32
Invacare®Platinum®Mobile
5.Laytheconcentratorandcarrybagdownwithairintake lterscreenfacingupward.
6.Pushtheconcentratorintothecarrybaguntilitisfully seatedagainstthetopretainer .
7.Rotatethebaseofthecarrybagontothebottomof theconcentratorandfullyzipupthebottomofthe carrybag.
8.Returntheconcentratorandcarrybagtotheupright position.
9.Checkthatthecutoutsinthetopretainerarecorrectly alignedwiththeoxygenoutletportandexternalpower inputconnector.Ifnotaligned,removethecarrybag andreinstallasdescribedinsteps1–9.
3.4.2RemovingtheCarryBag
1.Ensuretheconcentratorisoffanddisconnectedfrom externalpowerandcannula.
2.Laytheconcentratoronaatsurfacewiththeairintake lterscreenfacingupwardandfullyunzipthebase.
3.Rotatethebaseofthecarrybagoffthebottomofthe concentrator .
4.Fromthebottom,pulltheconcentratoroutofthecarry bag.Ifneeded,thetopsurfaceoftheconcentrator canbepushedtohelpremoveitfromthebottomof thecarrybag.
5.Returntheconcentratortotheuprightposition.
6.Fullyzipupthebottomofthecarrybag.
3.4.3InstallingShoulderStraps
Thecarrybagcanbeconguredasamessengerbag usingoneshoulderstraporasabackpackusing twoshoulderstraps.
MessengerBagConguration
1.UsingasingleshoulderstrapA,attachaclipBfrom eachendofthestraptoringsCatthetopofthecarry bagD.
2.Adjustshoulderstraptodesiredlength.
3.SlidepadEonshoulderstraptothedesiredlocation.
321194969-C
Page 33
BackPackConguration
Setup
3.5Connecting/PositioningtheNasalCannula
DANGER! RiskofInjuryorDeath
Thecannulacancausetripping,falling,orother injuryifimproperlypositionedandsecured.To avoidinjuryordeath:
–ThecannulaMUSTberoutedandsecured
properly.
–DONOTpositionthecannulaaroundtheneck.
Ensurethepatientcanmovefreelywhile wearingthecannula.
–Avoidpositioningcannulasacrossareasofhigh
foottrafc(i.e.aisles,doorways,hallways,etc.)
1.AttacheachclipBattheendsofasingleshoulderstrap AtothetworingsClocatedatthetopandbottom ofthecarrybagD.
2.Repeatforthesecondstrapontheoppositeside.
3.Adjusteachshoulderstraptodesiredlength.
4.SlidepadEoneachshoulderstraptothedesired location.
WARNING! RiskofInjuryorDeath
Toavoidchokingand/orstrangulationfromtubing entanglement:
–Keepchildrenandpetsawayfromnasal
cannulaandtubing.
–Closesupervisionisnecessarywhenthenasal
cannulaisusedbyimpairedindividualsornear childrenand/orimpairedindividuals.
1194969-C33
Page 34
Invacare®Platinum®Mobile
WARNING! RiskofInjuryorDeath
Toreducetheriskofinjuryordeathfromillness:
–Replacethenasalcannulaonaregularbasis.
Checkwithyourequipmentprovideror physiciantodeterminehowoftenthecannula shouldbereplaced.
–DONOTsharecannulasbetweenpatients.
Cannulasareforsinglepatientuseonly.
CAUTION! RiskofDamage
–Useofsterilizationorcleaningsolutionsmay
leavethisdevicenonfunctional.
CAUTION! RiskofMinorInjuryorDiscomfort
Toensureproperbreathdetectionandoxygen delivery:
–Ensurethecannulaprongsarepositioned
properlyinyournose.Thisiscriticaltothe effectivenessoftheoxygentherapy.
–DONOTusetubing/cannulalengthexceeding
25ft(7,6m). –Usecrush-proofoxygentubing. –UseONLYsinglelumencannulaswiththe
owcharacteristicsstatedinSpecicationsin
TechnicalData. –Avoiduseofconcentratorinwindyconditions. –Checkforgasowattheoutletofthecannula.
1.Removethecannulafromitspackaging.
2.Connectthecannulatotheconcentratoroxygenoutlet portA.Ensuretheconnectionissecure.
3.PlacethecannulaBoveryourearsandpositionthe prongsinyournoseasinstructedbyyourhealthcare providerorcannulamanufacturer .
Gasowattheoutletofthecannulacanbechecked whiletheconcentratoriswarmingup.Waveyour handinfrontofthenasalprongs.Youshouldbe abletohearandfeelthepulsedowofgas.If youdonotfeelthegaspulse,checkthecannula connectionforleaks.
341194969-C
Page 35
Usage
4Usage
4.1Location
Theconcentratorcanbeusedinahome,institution,vehicle orotherenvironmentsoutsidethehome.
WARNING! RiskofInjuryorDamage
Toavoidinjuryordamagefromairborne pollutantsand/orfumesandforoptimal performance:
–Locateandpositiontheconcentratorinawell
ventilatedspacesothattheairintakeandthe airexhaustopeningsarenotobstructed.
–NEVERblocktheairopeningsofthe
concentratororplaceitonasoftsurface,such asabedorcouch,wheretheairopeningmay beblocked.
–Keeptheopeningsfreefromlint,hairand
similarforeignitems.
–Keepconcentratoratleast12in(30.5cm)away
fromwalls,draperiesandfurniture.
–DONOTuseinpresenceofpollutants,smoke
orfumes,ammableanesthetics,cleaning agentsorchemicalvapors.
RefertoUsingtheCarryBaginUsagefordetailon thelocationoftheairintakeandexhaustopenings.
4.2ConcentratorOperatingPosition
Whennotbeingworn,theconcentratorshouldbeoperated intheuprightpositionandplacedonarm,levelsurface.
Theconcentratorshouldbeinalocationwherethe audiblesignalcanbeheard.
4.3UsingtheCarryBag
WARNING! RiskofInjuryorDamage
Toavoidinjuryordamagefromdroppingthe concentratororincorrectpositioningofthe concentrator:
–Thecarrybagmustbeusedforportable
operationoftheconcentrator .
–DONOTadjusttheshoulderstrapwhile
carryingtheconcentrator .
–Positionconcentratorsoexhaustventsare
directedawayfromthebodyandcontrolpanel facesupward.
–UseonlyInvacarecarrybag,modelPOC1–150.
CAUTION! RiskofMinorInjuryorDiscomfort
Toavoidinjuryfromconcentratormalfunctioning:
–Ensureyouareabletoheartheaudible
signalwarningswhenusingthecarrybagasa backpackincaseanalarmconditionoccurs.
Thecarrybagcanbeconguredasamessengerbagusing oneshoulderstraporasabackpackusingtwoshoulder straps.Thelengthoftheshoulderstrap(s)canbeadjusted
1194969-C35
Page 36
Invacare®Platinum®Mobile
tothedesiredposition.RefertoInstallingShoulderStraps inSetup.
MessengerBagConguration
BackpackConguration
Whencarryingtheconcentratorwiththeshoulderstrap(s) A,positiontheconcentratorwiththecontrolpanelfacing upwardandtheairintakeCopeningtowardsthebody. ThisorientstheconcentratorsothattheexhaustventB
isdirectedawayfromthebodyandthecontrolpanelis positionedforproperviewing.
ThecarrybagalsohasahandleDthatcanbeusedtomove andpositiontheconcentrator .
4.4TurningOntheConcentrator
1.ToturnontheconcentratorpresstheOn/OffbuttonA untilyouseethecontrolpanelindicatorslightup.
2.Theaudiblesignalwillbeepandalltheindicatorswill lightupforapproximatelytwosecondsaftertheunitis rstturnedon.Thispowerupsequenceprovidesthe opportunitytoensurethatallindicatorsarefunctioning properly.
3.Followingthepowerupsequenceawarmupperiod begins.Formoredetails,refertoWarmUpPeriodin Usage.
361194969-C
Page 37
Usage
4.5TurningOfftheConcentrator
1.ToturnofftheconcentratorpresstheOn/Offbutton until“PoweringDown”isshownonthedisplayscreen.
2.Thepowerdownsequencetakesaapproximatelythree seconds.
4.6WarmUpPeriod
CAUTION! RiskofMinorInjuryorDiscomfort
Duringthewarmupperiod(typicallyless thanveminutes)oxygenoutputisnotwithin specicationslistedinSpecicationsinT echnical Data.
–Concentratormaybeusedduringwarmup
period.
Afterturningontheconcentratorawarmupperiodofupto 15minutesisneededtoreachthespeciedoxygenpurity. “WarmingUp”isshownonthedisplayscreenduringthis period.
Whilewarmingup,theconcentratorwillautomatically deliverapulseofoxygenabouteveryfoursecondswhenno breathisdetected.Thenasalcannulashouldbeconnected totheconcentratorduringwarmupperiod.
Whentheconcentratorhasnishedwarmingup,thecurrent pulsesettingwillbeshownonthedisplayscreen.Ifno breathisdetectedafterwarmuphasended,theNoBreath Detectedalarmwilloccur .RefertoAlarmConditionsin Troubleshooting.
4.7BreathingwiththeUseoftheConcentrator
Asyoubreathewiththenasalcannulainstalled,apulse ofoxygenisdeliveredeachtimetheconcentratorsenses aninhalation.RefertoConnecting/PositioningtheNasal CannulainSetupforcannulainstallationinstructions.
Ifnobreathisdetectedfor15seconds,theNoBreath Detectedalarmwilloccur .RefertoAlarmConditionsin Troubleshooting.
4.8ReadingtheDisplayScreen
PulseFlowSettingandPulseIcon
Aftertheconcentratornisheswarmingup,thecurrent oxygenpulseowsettingBandpulseiconCareshownon thedisplayscreenA.
Thepulseiconashesoneachtimeanoxygenpulseis triggered.
1194969-C37
Page 38
Invacare®Platinum®Mobile
Thepulseowsettingandpulseiconarenotshown whenalarmtextisdisplayedonscreen.
InformationalText
Informationaltextisshownonthedisplayscreenduring certainperiodsofconcentratoroperation:
“WarmingUp”—displayedwhiletheconcentratoris warmingupafterrstbeingturnedon.
“PoweringDown”—displayedwhiletheconcentrator ispoweringdownafterbeingturnedoff.
AlarmText
Ifanalarmconditionexists,textdescribingthealarmwillbe shownonthedisplayscreen.
RefertoAlarmConditionsinTroubleshootingformoredetail onalarmconditionsandrelatedalarmtext.
4.9AdjustingthePulseFlowSetting
CAUTION! RiskofMinorInjuryorDiscomfort
Itisveryimportanttoselecttheprescribed oxygenowsetting.Thiswillensureyouwill receivethetherapeuticamountofoxygen accordingtoyourmedicalcondition:
–DONOTincreaseordecreasetheowsetting
unlessachangehasbeenprescribedbyyour physicianortherapist.
–Thetherapeuticeffectivenessoftheprescribed
oxygenowsettingshouldbeperiodically reassessed.
–Useonlythelengthofcannulathatwasusedto
determinetheprescribedoxygenowsetting.
–Thepulseowsettingsofthisconcentrator
mightnotcorrespondtocontinuousow oxygen.
–Thesettingsofothermodelsorbrandsof
oxygentherapyequipmentdonotcorrespond tothesettingsofthisconcentrator.
381194969-C
Page 39
1.Withtheconcentratorrunning,presstheowselection buttonBtoincreasethepulseowsettingbyone increment.
2.Presstheowselectionbuttonadditionaltimes,as needed,untilyourprescribedowsetting(P1toP4/P5) isshownonthedisplayscreenC.
Eachtimetheowsettingischanged,anaudible beepwilloccurandthedisplayscreenbacklightwill turnonforveseconds.
POC1-100BModels:Theconcentratorhasfour pulseowsettings.Whenthemaximumpulse owsettingP4isreached,thepulseowsetting returnstosettingP1withthenextbuttonpress.
POC1-100CModels:Theconcentratorhasve pulseowsettings.Whenthemaximumpulse owsettingP5isreached,thepulseowsetting returnstosettingP1withthenextbuttonpress.
Usage
Ifamessageotherthanthepulseowsettingis shownonthedisplayscreen,theadjustedpulseow settingwilltemporarilybedisplayedforveseconds afterpressingtheowselectionbutton.
Theconcentratorpulseowsettingatpower-upwill bethesameowsettingselectedthelasttimethe unitwasturnedoff .
4.10BatteryLifeandManagement
Timeawayfromhomeisgreatlyextendedbycombining theuseoftheACpoweradapter ,DCpowercableandthe concentrator’sbatteries.T oensurethebattery(s)maintain theiroptimalchargelevel,utilizetheACpoweradapter wheneveryouhaveaccesstoelectricpower .UtilizetheDC powercablewheneveryouareinavehicle.
CAUTION! RiskofMinorInjuryorDiscomfort
Batteryexhaustionwillresultinalossof supplementaloxygen.Toensureproper supplementaloxygendeliveryduringapower outage:
–Plugyourportableconcentratorintoan
alternatepowersource.
–Haveanalternatesourceofoxygenavailable
thatdoesnotrequireapowersource.
WhattoDo
Whenyourstreceiveyourconcentrator,fullycharge thebattery(s)overnight.
1194969-C39
Page 40
Invacare®Platinum®Mobile
Keepyourbattery(s)fullychargedwhenusingthe concentratoronadailybasis.
Y ourconcentratorbattery(s)canberechargedatany time.
Alwaysensuretheconcentratorbattery(s)arerecharged assoonaspossibleaftertheybecomefullydischarged. Thebattery(s)maybepermanentlydegradedifleftfully dischargedforanextendedlengthoftime.
Checkthestatusofyourconcentratorbattery(s)once amonthifyouarenotusingyourconcentratoron adailybasis.Battery(s)shouldbemaintainedattwo illuminatedbarsworthofchargeifnotusingthe concentratoronadailybasis.
Heatistheworstenemyofabattery .Allowplentyof airtocirculatearoundtheconcentratorsothatthe battery(s)iskeptascoolaspossiblewheninuseand whencharging.
WhatNottoDo
DONOTuseorleavetheconcentratororbattery(s) inexcessiveheatorcold.
DONOTstoreorleavetheconcentratororbattery(s)in cartrunks,etc.forextendedperiodsoftime.
DONOTstorebattery(s)fullycharged(fourbars illuminatedontheunit’sbatterygauge)ifyouaregoing tostoreyourconcentratorforanytimegreaterthan aday.Rechargeordischargethebatterytotwobars (50%charge)only.Storingabatterywithafullcharge maydegradeitsusefullife.
DONOTleaveyourbattery(s)installedintothe concentratorwhentheconcentratorisnotgoingto beusedformorethanaday.Thebattery(s)willlose chargewhilepluggedintotheconcentratorevenwith theconcentratorturnedoff.
401194969-C
Page 41
Maintenance
5Maintenance
5.1Maintenance
WARNING! RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecically designedtominimizeroutinepreventive maintenance.Topreventinjuryordamage:
–Onlyprofessionalsofthehealthcareeldor
personsfullyconversantwiththisprocesssuch asfactorytrainedpersonnelshouldperform preventivemaintenanceorperformance adjustmentsontheoxygenconcentratorand itsequipment,exceptfortasksdescribedin thismanual.
–UsersshouldcontactyourdealerorInvacare
forservice.
5.2ServiceLife
Theexpectedservicelifeofthisproduct,withtheexception ofwearcomponents(see6.3WearandTear,page48),is threeyearsofoperationwhenusedinaccordancewiththe safetyinstructions,maintenanceintervalsandcorrectuse statedinthismanual.Theeffectiveservicelifecanvary accordingtothefrequencyandintensityofuse.Refertothe proceduresin5Maintenance,page41
.
5.3CleaningtheCabinet
DANGER! RiskofInjuryorDamage
Liquidwilldamagetheinternalcomponentsof theconcentratoranditsequipment.T oavoid damageorinjuryfromelectricalshock:
–T urnOfftheconcentratorandunplugthe
powercordbeforecleaning.
–DONOTallowanycleaningagenttodripinside
theairinletandoutletopenings.
–DONOTsprayorapplyanycleaningagent
directlytothecabinet.
–DONOThosedowntheproduct.
CAUTION! RiskofDamage
Harshchemicalagentscandamagethe concentratoranditsequipment.T oavoid damage:
–DONOTcleanwithalcoholandalcoholbased
products(isopropylalcohol),concentrated chlorine-basedproducts(ethylenechloride), andoil-basedproducts(Pine-Sol®,Lestoil®)or anyotherharshchemicalagents.Onlyusemild liquiddishdetergent(suchasDawn®).
Performallmaintenanceaccordingtotherecommended scheduleinthismanual.
1194969-C
41
Page 42
Invacare®Platinum®Mobile
Periodicallycleantheconcentrator’scabinetasfollows:
1.Turnoffconcentrator .
2.Removetheconcentratorfromthecarrybagorback pack.
3.Useadampcloth,orsponge,withamilddetergent suchasDawndishwashingsoaptogentlycleanthe exteriorcase.
4.Allowtheconcentratortoairdry,oruseadrytowel, beforereturningtheconcentratortothecarrybagor back,andpriortooperatingtheconcentrator .
5.Returntheconcentratortoitscarrybagorbackpack.
TheACpoweradapterandACpowercordaretobe cleanedinasimilarmanner .
1.Turnoffconcentrator .
2.Useavacuumcleanerornylonbrushtoclean dust/debrisfromairintakelterscreenA.
5.5CleaningtheCarryBag
5.4CleaningtheAirIntakeFilterScreen
CAUTION! RiskofDamage
Toavoiddamagefromclogging:
–Replacethecarrybagiftheairintakelter
screenistornorfrayed.
Cleantheairintakelterscreenatleastonceeach week.Theairintakelterscreenispartofthecarry bag.
42
CAUTION! RiskofDamage
Toavoiddamagetotheproduct:
–DONOTmachinewashordrythecarrybag.
1.Turnoffconcentrator .
2.RemovetheconcentratorAfromthecarrybagB.
3.Connectconcentratortoanexternalpowersource.
1194969-C
Page 43
Maintenance
4.Wipeorbrushthecarrybagwithamildliquiddish detergent(suchasDawn)andwater.Rinsethoroughly.
5.Allowthecarrybagtoairdryaftercleaningandbefore using.
5.6CleaningandDisinfectionBetweenPatients
Cleaninganddisinfectionmustbeperformedbetween patientsandasnecessary .
WARNING! RiskofInjuryorDamage
Topreventinjuryfrominfectionordamageto concentrator:
–Onlyqualiedpersonnelshouldperform
cleaninganddisinfectionoftheoxygen concentratorandaccessoriesbetweenpatients.
Followtheseinstructionstoeliminatepossible pathogenexchangebetweenpatientsdueto contaminationofcomponentsoraccessories.
1.Disposeofandreplaceallpatientsideaccessoriesnot suitableformultiplepatientuse,includingbutnot limitedto:cannulaandoxygentubing.
2.Cleanconcentratorandaccessoriesasdescribedin Maintenance.
3.Disinfectthesurfacesoftheconcentratorand accessoriesusingClorox®4inOneDisinfectant& Sanitizerorsubstantiallyequivalentproduct.
WARNING! RiskofInjuryorDamage
–Followthecleaning/disinfectantproduct
manufacturer’sinstructionsforuse, handling,storageanddisposalofthe product.
–Useofinappropriatedisinfectingproducts
maydegradedevicematerialsand negativelyaffectdevicesafetyand performance.
Iftheaccessories,suchasthecarrybag,are unabletobecleanedordisinfected,discardthe accessoryandreplace.
4.PerformthetasksonPreventiveMaintenanceChecklist.
5.Beforerepackaginganddistributiontonewpatient, ensurepackagingcontentscontaintheconcentrator, battery,carrybagwithshoulderstraps,ACpower adapter,ACpowercord,DCpowercable,labelsand usermanual.
1194969-C43
Page 44
Invacare®Platinum®Mobile
5.7ViewingHourMeter
Toviewtheelapsedhoursofconcentratorruntimeperform thefollowingsteps:
1.Ensurethattheunitispoweredoffandconnectedto anexternalpowersource.
2.Pressandholdtheowselectionbuttonforve seconds.
3.Thecumulativehoursofconcentratoroperation (compressorhours)willbeshownonthedisplayscreen foraslongasthebuttoniscontinuallypressed.The concentratorruntimeisdisplayedtothenearest0.1 hours.
4.Releasetheowselectionbutton.
44
1194969-C
Page 45
5.8PreventativeMaintenanceChecklist
ThefollowingPreventativeMaintenanceChecklistcontainsmaintenancetaskstobeperformedbytheusersofthisproduct, exceptwhereotherwisenoted.Ifyouareunabletounderstandthesetasks,contactyourprovideroraqualiedtechnician.
WEEKLY
RecordDateofService.
RecordElapsedHoursonHourMeter. (RefertoViewingHourMeterinMaintenance.)
CleanAirIntakeFilterScreen
InspectCabinetofConcentratorandACAdapterfor Damage. Donotuseifdamageisfound.ReturntoInvacarefor repair .
InspectCarryBagStrap(s)andhardwareforDamage. Replacestraporcarrybagifdamagefound.
InspectElectricalCordsforDamage. ReplaceElectricalCordifdamageisfound.
CheckforpresenceandlegibilityofallLabels. ReplaceLabelsasneeded.RefertoLabelLocations.
Maintenance
1194969-C45
Page 46
Invacare®Platinum®Mobile
EVERY4,380HOURS,EVERY3YEARS,ANDBETWEENP ATIENTS,WHICHEVERCOMESFIRST
CheckOxygenPurity*
PerformFunctionalTest*
DURINGPREVENTATIVEMAINTENANCESCHEDULEANDBETWEENPATIENTS
CheckIntakeFilterScreenfordamage. Replacecarrybagifdamageisfound.
Check/ReplaceOutletFilter*
Inspectionperiodsshownashoursisinreferencetohoursofconcentratoroperationsincethelastdateofservice.Referto ViewingHourMeterinMaintenance.Inspectionperiodsshownasmonthsoryearsisinreferencetodurationoftimesincethe lastdateofservice.
*Tobeconductedbyproviderorqualiedtechnician.Refertoservicemanual.
461194969-C
Page 47
AfterUse
6AfterUse
6.1Storage
1.Removebattery(s)priortostorage.
2.Storetherepackagedoxygenconcentratorandbattery(s) inacool,dryarea.
3.DONOTplaceobjectsontopofpackagedconcentrator .
RefertostorageconditionsinSpecicationsin TechnicalData.
Afterremovingfromstorage,itmaytakeuptoone houratroomtemperaturefortheproducttoreach itsoperatingtemperatureandbereadyforuse.
6.2Disposal
RecycleDONOTdisposeofin
householdwaste
Thisproducthasbeensuppliedfromanenvironmentally awaremanufacturerwhocomplieswiththeWasteElectrical andElectronicEquipment(WEEE)Directive2012/19/EC.This productmaycontainsubstancesthatcouldbeharmfulto theenvironmentifdisposedofinplaces(landlls)thatare notappropriateaccordingtolegislation.
Followlocalgoverningordinancesandrecyclingplans regardingdisposaloftheconcentratororcomponents normallyusedinoperation.Theconcentratordoesnot generatewasteorresidueinoperation.
DONOTdisposeoftheconcentratorinthenormal wastestream.
AnyaccessoriesnotpartoftheconcentratorMUST behandledinaccordancewiththeindividualproduct markingfordisposal.
DONOTdisposeoftheinternalorsupplemental batteries.Batteriesshouldbereturnedtoyour dealer/provider .
1194969-C
47
Page 48
Invacare®Platinum®Mobile
6.3WearandTear
Invacarereservestherighttoaskforanyitembackthat hasanallegeddefectinworkmanship.SeeWarrantythat shippedwiththeproductforspecicwarrantyinformation.
RefertoMaintenanceinthismanualforproperpreventative maintenancescheduleanduseoftheproduct.
Thisisonlyageneralguidelineanddoesnotincludeitems damagedduetoabuseandmisuse.
Normalwearandtearitemsandcomponentsforthis productarelistedbelow.
Description
Compressorinletlter1,460hours
Carrybag18months
Battery
Sievebedassembly
ExpectedLife
18months
1year
Expectedlifevaluesshownashoursisinreference tohoursofconcentratoroperation.RefertoViewing HourMeterinMaintenance.Expectedlifevalues shownasmonthsoryearsisinreferencetoduration oftimesincethedateofpurchase.
Sieveisaporouslteringmaterialandisconsidered awearitem.Somefactorsthatcouldaffect sievemateriallifeincludehumidity,temperature, particulates,aircontaminates,airintake,vibration andotherenvironmentalconditions.
481194969-C
Page 49
Troubleshooting
7Troubleshooting
7.1Troubleshooting
Ifyourconcentratorfailstooperateproperly,refertothefollowingchartforpossiblecausesandsolutions.Ifnecessary, contactyourprovider.
OxygenConcentratorDoesNotTurnOnorDoesNotStayOn
Symptom
Concentratordoes notoperatewhen On/Offbuttonis pressed.
OR
Concentrator beginstooperate whenOn/Off buttonispressed, butallLCDand frontpanellights godark.
ProbableCause
On/Offbuttonwasnothelddownlongenough.
Batteryisdischarged(orimproperlyconnectedto theconcentrator).
Batteryisimproperlyconnectedtotheconcentrator .1.Removethebatteryandanyexternalpower
Batteryhasperformedaself-protectshutdown whentherewasnoexternalpower .
Solution(s)
Trytopoweruptheconcentratoragainwhile continuingtopresstheOn/Offbuttonuntilthe frontpanellightsandLCDbegintoilluminate.This takestypicallyonefullsecond.
1.Removebatteryandcheckbatterygauge.
2.Ifgaugeshowsbatteryischarged,thenreinstall batteryandretry.
3.Ifgaugeshowsbatteryisdepleted,theneither installachargedbatteryorconnectexternal powerandretry.
connection.
2.Reinstallthebattery,makingsureitis completelyinsertedandretry.
3.ConnecttoeitherACorDCexternalpower andretry.
1.Waitoneminuteandretry.
2.Changetoanalternatebatteryorconnect concentratortoeitherACorDCexternalpower andretry.
1194969-C49
Page 50
Invacare®Platinum®Mobile
ConcentratorisBeepingwithNoLightsorDisplay
Symptom
Concentratorisbeepingwith nolightsordisplay.
Batteryhasbeenremovedordisconnected andthereisnoexternalpowerconnected.
ProbableCause
Beepingwilllastless thantwominutesif nopowersourceis provided.
Whenoperatingwithoutabattery,external powerislost.
BatteryNotChargingWhenExternalPowerisConnected
Symptom
WithACorDC powerpluggedinto theconcentrator, theexternalpower iconisilluminated
Powersourceisfaulty ,orthereisaloose connection.
Batteryisoutsidetheallowedtemperaturerange forcharging.
ProbableCauseSolution
andbatterycharge levelindicatoris notashing.
Batteryisnotfullyseated.1.Ensurethebatteryiscompletelyinserted.
Batterychargeiscomplete(allfoursegmentsof chargelevelindicatorareilluminated).
Solution(s)
1.Ensurethebatteryiscompletelyinserted andrestart.
2.Installadifferentchargedbatteryand restart.
1.Tryanotherpoweroutletandcheck connectionstoconcentratorandrestart.
1.Tryanotherpoweroutletandcheck connectionstoconcentrator .
1.Allowconcentratorandbatterytocooldown tolessthan104°F(40°C),orwarmupto41°F (5°C).
1.Noactionrequired.
501194969-C
Page 51
NoExternalPowerSymbolIlluminatedonControlPanel
Symptom
WithACpower pluggedinto concentrator , externalpower iconisnot illuminatedonthe
Powersourceisfaulty ,orthereisaloose connection.
ProbableCauseSolution
1.IfthegreenindicatorlightontheACadapter islit,checkpowerconnectionsfromtheAC adaptertoconcentrator .
2.IfthegreenindicatorlightontheACadapter isnotlit,checkpowerconnectionsfromoutlet toACadapterortryanotheroutlet.
controlpanel.
WithDCpower pluggedinto concentrator , externalpower iconisnot
Powersourceisfaultyorthereisaloose connection.
TheDCpowercablefusehasblown.1.SwitchtoACpowerandcontactyourprovider
1.Checkpowerconnectionstoconcentrator.
2.Tryanotherpoweroutlet.
toservicetheDCpowercable.
illuminatedonthe controlpanel.
7.2AlarmConditions
Whenanyalarmconditionoccurs,theyellowalarmindicatorlightonthecontrolpanelilluminatesandtextdescribing thealarmconditionisshownonthedisplayscreen.Refertothesolutionsectionofthealarmtablesforpotential correctiveactions.Ifnecessary,contactyourprovider.
Inallcases,pressingandholdingtheOn/Offbuttonforonesecondwillturnoffandresettheconcentrator.
Troubleshooting
Ifthecauseofthealarmconditionisnotcorrected,itwillreappearwhentheconcentratoristurnedbackon.
AllalarmsareclassiedasLowPrioritytechnicalalarmconditions.
Whenmultiplealarmconditionsexist,thetextassociatedwiththehighestrankedalarmwillbeshownonthedisplay screen.
1194969-C51
Page 52
Invacare®Platinum®Mobile
7.2.1OperationalAlarms
Theconcentratorcontinuestorunwhenthealarmconditionslistedinthissectionoccur.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
NoBreathDetected
CheckCannula
Oxygenconcentratorhasnotdetecteda breathovera15secondperiod.
1.Verifycannulaisconnected,not kinked,properlypositionedanduser isbreathingthroughhis/hernose.
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
NOTES:
Alarmwillturnoffwhenabreathisdetected.
Thisalarmdoesnotapplyduringwarmup.
Ifnobreathisdetectedfortwominutes,thealarmwillescalatetoashutdownalarm.RefertoShutdownAlarms inTroubleshooting.
Therstoccurrenceofthisalarmconditionwillcanceltheaudiooff(mute)feature.Audiooff(mute)canbereactivated.If reactivated,thealarmaudiooff(mute)featurewillremainactivatedevenifthisalarmconditionoccursagain.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
BreathRateHigh
ReduceActivity
Theuserbreathratehasexceededthe capacityoftheconcentratorformore than15seconds.
1.Theusershouldimmediatelyreduce his/heractivityleveltogethis/her breathratetoslowdown.
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
521194969-C
Page 53
Troubleshooting
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
NOTES:
Asufcientreductionoftheuser’sbreathratewillautomaticallyturnoffthealarm.
Therstoccurrenceofthisalarmconditionwillcanceltheaudiooff(mute)feature.Audiooff(mute)canbereactivated.If reactivated,thealarmaudiooff(mute)featurewillremainactivatedevenifthisalarmconditionoccursagain.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
LowBattery
ChargeConcentrator
INDICATORS:
Singleaudiblebeepnorepeat
YELLOWalarmindicatorilluminated
Batterygauge:
Remainingbatteryruntimeof approximately30minutes.Battery requirescharging.
1.Connectoxygenconcentratorto eithertheACorDCpowerorinsert analternatechargedbatteryinto theunit,OR
2.Insertanalternatechargedbattery intotheunit.Removethedepleted batteryandchargeitwiththe externalbatterychargerforfuture use.
PulseFlowSettingsP1–P3:onlythe lowestsegmentisilluminated
PulseFlowSettingP4–P5:only thelowesttwosegmentsare illuminated
NOTES:
Thealarmturnsoffwhenexternalpowerisconnectedorwhenachargedbatteryisinstalled.
Alarmwillturnoffwhenabreathisdetected.
1194969-C53
Page 54
Invacare®Platinum®Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
LowBattery
ChargeNow
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
Batterygauge:
Remainingbatteryruntimeof approximately15minutesorless. Batteryrequirescharging.
Lossofpowerisimminent.
1.Connectoxygenconcentratorto eithertheACorDCpower ,OR
2.Insertanalternatechargedbattery intotheunit.Removethedepleted batteryandchargeitwiththe externalbatterychargerforfuture use,OR
3.Ifnootherpowersourceisavailable, changetoanothersourceofoxygen.
Onlythelowestsegmentis illuminated
NOTES:
Thealarmturnsoffwhenexternalpowerisconnectedorwhenachargedbatteryisinstalled.
Therstoccurrenceofthisalarmconditionwillcanceltheaudiooff(mute)feature.Audiooff(mute)canbereactivated.If reactivated,thealarmaudiooff(mute)featurewillremainactivatedevenifthisalarmconditionoccursagain.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
UnitHot
Allowtocool
INDICATORS:
Concentratorisnearitsmaximum operatingtemperature.
1.Moveconcentratortocooler surroundings.Allowconcentratorto cooldowntolessthan104°F(40°C).
2.Cleanintakelterscreen.
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
NOTES:
Thealarmwillturnoffwhentherequiredinternaloperatingtemperatureisreached.
541194969-C
Page 55
Troubleshooting
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
UnitCold
Allowtowarm
Concentratorisnearitsminimum operatingtemperature.
1.Moveconcentratortowarmer surroundings.Allowconcentratorto warmupto41°F(5°C).
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
NOTES:
Thealarmwillturnoffwhentherequiredinternaloperatingtemperatureisreached.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
BatteryHot
Allowtocool
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
Batteryisnearitsmaximumoperating temperature.
1.Moveconcentratortocooler surroundings.Allowconcentratorto cooldowntolessthan104°F(40°C), OR
2.UseACorDCpowerandremove battery,OR
3.Insertanalternatechargedbattery andremovethehotbattery .
NOTES:
Thealarmwillturnoffwhentherequiredbatterytemperatureisreachedorifexternalpowerisconnectedandbattery isremoved.
1194969-C55
Page 56
Invacare®Platinum®Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
BatteryCold
Allowtowarm
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
Batteryisnearitsminimumoperating temperature.
1.Moveconcentratortowarmer surroundings.Allowconcentratorto warmupto41°F(5°C),OR
2.UseACorDCpowerandremove battery,OR
3.Insertanalternatechargedbattery andremovethecoldbattery.
NOTES:
Thealarmwillturnoffwhentherequiredinternaloperatingtemperatureisreachedorifexternalpowerisconnectedand batteryisremoved.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
LowOxygen
SeeManual
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
Theoxygenoutputpurityhasfallentoa valuebetween73%and87%.
1.Verifyconcentratorisin recommendedenvironmental temperature.RefertoSpecications inTechnicalData.
2.Cleanintakelterscreenandensure bothintakeandexhaustarenot blocked.
3.Turnconcentratoroff,thenonagain toretry.
561194969-C
Page 57
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
NOTES:
Thealarmwillturnonwhentheoxygenpurityhasdroppedbelow85%.
Thealarmwillturnoffwhentheoxygenoutputpurityhasincreasedaboveto87%.
Thealarmlimitshaveatoleranceof+/-2%.
Alarmsignalgenerationmaybedelayedbyupto128secondsduetothemonitoringalgorithmusedtopreventnuisancealarms.
Callyourproviderandreporttheproblemifthewarningcontinueswithnextuse.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
Cannotcharge
Allowtocool
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
Batteryistoohottoallowthebattery tocharge.
ThisalarmonlyoccurswhenExternal Powerisconnected.
1.Moveconcentratortocooler surroundings.Allowconcentratorto cooldowntolessthan104°F(40°C), OR
2.Removebattery(withACorDC powerconnected)toallowbattery tocool.
NOTES:
Thealarmwillturnoffwhentherequiredbatterytemperatureisreached,whenthebatteryisremoved,orwhenexternal powerisdisconnected.
Troubleshooting
1194969-C57
Page 58
Invacare®Platinum®Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
Cannotcharge
Allowtowarm
INDICATORS:
Singleaudiblebeepevery30seconds
Batteryistoocoldtoallowthebattery tocharge.
Thisalarmonlyoccurswhenanexternal powersourceisconnected.
1.Moveconcentratortowarmer surroundings.Allowconcentratorto warmupto41°F(5°C),OR
2.Removebattery(withACorDC powerconnected)toallowbattery towarm.
YELLOWalarmindicatorilluminated
NOTES:
Thealarmwillturnoffwhentherequiredbatterytemperatureisreached,whenthebatteryisremoved,orwhenexternal powerisdisconnected.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
SensorFault
SeeManual
Oxygensensorfeedbackisout-of-range. Sensorfailureislikely.
1.TurnconcentratorOff,thenOnagain toretry.
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
NOTES:
Thealarmwillturnoffwhenthesensorfeedbackisbackinrange.
Callyourproviderandreporttheproblemifalarmcontinues.
581194969-C
Page 59
Troubleshooting
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
WarmingUp
INDICATORS:
Noaudiblebeep
YELLOWalarmindicatorilluminated
Theconcentratorisstillwarmingup aftertwominutesfrominitialstart up.Theoxygenoutputpurityhasnot reachedtheminimumvaluelistedin SpecicationsinTechnicalData.
1.Verifyconcentratorisin recommendedenvironmental temperature.RefertoSpecications inTechnicalData.
2.Cleanintakelterscreenandensure bothintakeandexhaustarenot blocked.
3.Allowconcentratortocontinue warmingupforatleast15minutes.
4.Turnconcentratoroff,thenonagain toretry.
NOTES:
Dependingonavarietyoffactors,itcantakeupto15minutestoreachtheminimumspeciedoxygenpurity.
Thealarmwillturnoffwhentheminimumoxygenoutputpurityisreached.
Iftheminimumoxygenoutputpurityhasnotbeenreachedwithin15minutesfrominitialstartup,thealarmwillbecomea LowPurityalarmcondition.
1194969-C59
Page 60
Invacare®Platinum®Mobile
7.2.2ShutDownAlarms
Theconcentratorshutsdownwhenthealarmconditionsinthissectionoccur .
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
NoBreathDetected
Oxygenconcentratorhasnotdetecteda breathforoveratwominuteperiod.
1.Verifycannulaisconnected,not kinked,properlypositionedanduser isbreathingthroughhis/hernose.
2.Changetoanothersourceofoxygen ifalarmcontinues.
NOTES:
Thisalarmdoesnotapplyduringwarmup.
Toclearalarm,turntheconcentratoroff.Completesolution1beforerestarting.PressandholdtheOn/Offbuttontopower downtheconcentratorandrestart.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
Stuckbutton
INDICATORS:
Twoaudiblebeepsevery16seconds
Astuckbuttonisdetectedupon connectiontopowersource.
1.Removeallpowersourcesfrom concentrator .Presseachcontrol buttonlookingforastuckbutton. Reconnectpower .
2.Changetoanothersourceofoxygen ifalarmcontinues.
YELLOWalarmindicatorilluminated
NOTES:
Toclearalarm,turntheconcentratoroff.Completesolution1beforerestarting.Callyourproviderandreporttheproblem ifalarmcontinues.
601194969-C
Page 61
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
BatteryDepleted
INDICATORS:
Remainingbatterycapacityhasbeen completelyused.Batteryrequires charging.
Concentratorfunctionhasshutdown.
1.Connectoxygenconcentratorto
2.Insertanalternatechargedbattery.
Twoaudiblebeepsevery16seconds
YELLOWalarmindicatorilluminated
3.Ifnootherpowersourceisavailable,
Batterygauge:
Nosegmentsilluminated
NOTES:
Toclearthealarm,theconcentratorneedstobeturnedoff.Completesolution1or2beforerestarting.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
UnitTooHot
INDICATORS:
Twoaudiblebeepsevery16seconds
Concentratoristoohottoallowitto continueoperating.
Concentratorfunctionhasshutdown.
Internalfanwillcontinuetoruntohelp lowerinternaltemperature.
1.Moveconcentratortocooler
2.UseACorDCadapterforpower.
3.Cleanintakelterscreen.
4.Changetoanothersourceofoxygen
YELLOWalarmindicatorilluminated
NOTES:
Troubleshooting
eithertheACorDCpower ,OR
Removethedepletedbatteryand chargewiththeoptionalexternal batterychargeraccessoryforfuture use,OR
changetoanothersourceofoxygen.
surroundings.Allowconcentratorto cooldownto104°F(40°C).
whilewaiting.
Toclearthealarm,turntheconcentratoroff.Completesolutions1,2,and/or3beforerestarting.
1194969-C61
Page 62
Invacare®Platinum®Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
UnitTooCold
INDICATORS:
Twoaudiblebeepsevery16seconds
Concentratoristoocoldtoallowitto continueoperating.
Concentratorfunctionhasshutdown.
Internalfanwillcontinuetoruntohelp increaseinternaltemperature.
YELLOWalarmindicatorilluminated
NOTES:
Toclearthealarm,turntheconcentratoroff.Completesolutions1,2,and/or3beforerestarting.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
BatteryTooHot
INDICATORS:
Batteryistoohottoallowthebatteryto continueoperating.
Concentratorfunctionhasshutdown.
Twoaudiblebeepsevery16seconds
YELLOWalarmindicatorilluminated
NOTES:
Toclearthealarm,turntheconcentratoroff .Completesolutions1,2,3,or4beforerestarting.
1.Moveconcentratortowarmer surroundings.Allowconcentratorto warmupto41°F(5°C).
2.UseACorDCadapterforpower.
3.Changetoanothersourceofoxygen whilewaiting.
1.Moveconcentratortocooler surroundings.Allowconcentratorto cooldowntolessthan104°F(40°C), OR
2.UseACorDCpowerandremove battery,OR
3.Insertanalternatechargedbattery andremovethehotbattery,OR
4.Changetoanothersourceofoxygen ifnootherpowersourceisavailable.
621194969-C
Page 63
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
BatteryCold
INDICATORS:
Batteryistoocoldtoallowthebattery tocontinueoperating.
Concentratorfunctionhasshutdown.
Twoaudiblebeepsevery16seconds
YELLOWalarmindicatorilluminated
NOTES:
Toclearthealarm,turntheconcentratoroff .Completesolutions1,2,3,or4beforerestarting.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
OxygenFault
INDICATORS:
Theoxygenoutputpurityhasfallento avaluebelow73%.
Concentratorfunctionhasshutdown.
Twoaudiblebeepsevery16seconds
YELLOWalarmindicatorilluminated
NOTES:
Troubleshooting
1.Moveconcentratortowarmer surroundings.Allowconcentratorto warmupto41°F(5°C),OR
2.UseACorDCpowerandremove battery,OR
3.Insertanalternatechargedbattery andremovethecoldbattery,OR
4.Changetoanothersourceofoxygen ifnootherpowersourceisavailable.
1.Verifyconcentratorisin recommendedenvironmental temperature.RefertoSpecications inTechnicalData.
2.Cleanintakelterandensureboth intakeandexhaustarenotblocked.
3.Changetoanothersourceofoxygen ifalarmcontinues.
Toclearthealarm,turntheconcentratoroff.CompleteSolution1and/or2beforerestarting.Thealarmlimithasa toleranceof+/-2%
Alarmsignalgenerationmaybedelayedbyupto128secondsduetothemonitoringalgorithmusedtopreventnuisancealarms.
Callyourproviderandreporttheproblemifalarmcontinues.
1194969-C63
Page 64
Invacare®Platinum®Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
SystemFault
INDICATORS:
Twoaudiblebeepsevery16seconds
Abnormalsystemconditionhasbeen detected.
Maybecausedbymalfunctionof compressor,fan,pressureorsoftware.
Concentratorfunctionhasshutdown.
YELLOWalarmindicatorilluminated
NOTES:
Toclearthealarm,turntheconcentratoroff .Restartconcentrator.
Callyourproviderandreporttheproblemifalarmcontinues.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
Notapplicable.Allpowerlost,display isblank.
INDICATORS:
Allpowersourceshavebeen lost/removed.
Concentratorfunctionhasshutdown.
Audiblebeepseverysecondfor approximately30to60seconds
Noothercontrolpanelindicators illuminated
NOTES:
Thisalarmoccurseveniftheconcentratorisoffwhenallpowerisremoved.
1.TurntheconcentratorOff,thenOn againtoretry.
2.Changetoanothersourceofoxygen ifalarmcontinues
1.Ensurethebatteryiscompletely inserted.
2.Checkexternalpowerconnections totheconcentratorandtothe poweroutlet.
3.Tryanotherpoweroutlet.
4.Changetoanothersourceofoxygen ifalarmcontinues.
Toclearthealarm,completesolutions1,2,or3beforerestarting.
641194969-C
Page 65
TechnicalData
8TechnicalData
8.1TechnicalDescription
TheInvacarePlatinumMobileusesamolecularsieveand pressureswingadsorptionmethodologytoproducethe oxygengasoutput.Ambientairentersthedevice,isltered andthencompressed.Thiscompressedairisthendirected intoanitrogenadsorbingsievebed.Concentratedoxygen exitstheoppositeendofthesievebedandisdirectedinto anoxygenreservoirfromwhichitisdeliveredtothepatient.
Theoxygenpurityleveloftheoutputgasrangesfrom87% to95.6%.Theoxygenisdeliveredtothepatientthrough theuseofanasalcannula.Apulsedosedeliverymethod isused.Theconcentratordetectsthestartofpatient inhalationanddeliversameasuredpulseofoxygen.No furtheroxygenisdelivereduntilthenextpatientinhalation isdetected.Thevolumeofoxygendeliveredeachminute isaxedamountbasedontheselectedpulseowsetting. Thevolumeofeachoxygenpulsewillvarywiththepatient’s breathratesuchthatthexedminutevolumeismaintained.
TheInvacarePlatinumMobileiscapableofoperationbythe patientinahome,institution,vehicle,orotherenvironments outsidethehome.Devicestandardpoweroptionsinclude anACtoDCswitchingpoweradapteroperatingfroman ACpoweroutlet(100–240VAC,50–60Hertznominal),aDC powercableoperatingfromanaccessoryoutlettypically foundinavehicletypeenvironment(12VDCnominal),and uptotworechargeablebatteries.
Nospecicproductknowledgeortrainingisrequiredto operatetheproductotherthanwhatiscontainedinthis manual.
Serviceinformationwillbeavailableonrequesttoqualied technicalpersonnelONLY .
8.1.1PneumaticDiagram
A=RoomAirIn
B=ExhaustAirOut
C=PatientOutletFitting
D=NasalCannula
1194969-C65
Page 66
Invacare®Platinum®Mobile
8.2Specications
ElectricalRequirements:ACPowerSupply:110–240VAC,50–60Hz
DCPowerSupply:11–16VDC
RatedCurrentInput:
PowerConsumption: (Typical)
5Aat19VDC,10Aat11–16VDC
PulseSetting:
P1=18W
P2=24W
P3=35W
P4–P5=45W
Dataisforconcentratoroperationonly(nobatterycharging)utilizinganACpowersource.
OperatingEnvironmental Conditions: (Allpowersources)
StorageandT ransport
OperatingTemperature:41°Fto104°F(5°Cto40°C)
RelativeHumidity:15–90%non-condensingrelativehumidity,watervaporpressuresupto1.48 inHg(50hPa)
-13°Fto140°F(-25°Cto60°C)
Temperatures:
StorageandT ransport Humidity:
OperatingAltitude:
OperatingAtmospheric
Upto90%non-condensingrelativehumidityfortemperaturesof41°Fto95°F(5°Cto35°C)
Watervaporpressureupto1.48inHg(50hPa)fortemperaturesgreaterthan95°F(35°C)
Upto10,000ft(3048m)abovesealevel
697–1060hPa
Pressure:
661194969-C
Page 67
TechnicalData
OxygenPurity:
87%to95.6%,atallowsettingsandovertheratedrangesforambienttemperature,humidity andatmosphericpressure.
Afterinitialwarm-upperiod(typicallylessthan5minutes)
ConserverTrigger Sensitivity:
<0.18cmH
Opressuredrop(Forallcannulalengths)
2
Factoryset—noadjustment,pressureactivated
Onlypatientrespiratoryeffortsthatachievethetriggerpressurewillresultinthedeliveryof anoxygenbolus.
ConserverBreathRate
15–40BPM(breathsperminute)withoutreductionofbolusminutevolume
Capacity:
MaximumOutlet
28.5psig(197kPa)
Pressure:
CannulaRequirements:
Length:4–25ft(1.2–7.6m)includingalloxygentubing
Tubing:crush-proof,singlelumen
Adult,standardow(ratedforupto6L/mincontinuousow)forlengthsupto7ft)
Adult,highow(ratedforupto15L/mincontinuousow)forlengthsgreaterthan7ftto25ft)
ExampleofPossibleCannulaModel:WestmedInc.PartNumber0194(4ftlength)
BatterySpecications
Rechargeablelithium-ion,14.4V,5800mAh,83.5Wh,500fullcharge/dischargecyclelife
(eachbattery):
BatteryShelfLife:12monthsfromdateofmanufacture
1194969-C67
Page 68
Invacare®Platinum®Mobile
BatteryDuration: (Timesareapproximate)
BatteryChargeTime: (Timesareapproximate)
SoundPressureLevel:
SoundPowerLevel:
AudibleSignalSound PressureLevel:
Dimensions:
Weight: (Nominal)
Condition
OneBattery
TwoBatteries
PulseSettingP15hr5min10hr10min
PulseSettingP23hr30min7hr0min
PulseSettingP32hr20min4hr40min
PulseSettingP4–P51hr45min3hr30min
Condition
OneBattery
TwoBatteries
ConcentratorOn,PulseSettingP12hr20min4hr40min
ConcentratorOn,PulseSettingP22hr20min4hr40min
ConcentratorOn,PulseSettingP32hr30min5hr0min
ConcentratorOn,PulseSettingP4–P53hr10min6hr20min
ConcentratorOff2hr20min4hr40min
<40dBAweightedforowsettingP2(TestedperISO3744:2010withmicrophonelocationas speciedinISO8359:1996subclause4.6)
65dBAweightedforowsettingP4–P5(TestedperISO80601–2–69subclause201.9.6.2.1.101)
<
<65dBAweightedforowsettingP4–P5(T estedperISO80601–2–69subclause201.9.6.2.1.101)
55dBA+/-5dBA
9.4inhighx7.4inwidex3.7indeep(23.9cmhighx18.8cmwidex9.4cmdeep)
4.9lbs(2.18kg)withsinglebatteryandnocarrybag
Add0.75lbs(0.34kg)forcarrybag
Add1.0lbs(0.45kg)forasecondbattery
681194969-C
Page 69
TechnicalData
ShippingWeight:
10.5lbs(4.8kg)
(Nominal)
Classications:ClassIIElectricalShockProtection,TypeBFAppliedPart,ContinuousOperation
IngressProtection Rating:
Concentrator—IP22
ACPowerAdapter—IP21
Battery—KeepDry
AppliedParts:
Cannula/OxygenTubing,OxygenOutletPort,CarryBag
DeliveredOxygenPulseVolumes:
ThenominalpulsevolumespublishedinthefollowingtableisinmillilitersatSTPD(standardtemperatureandpressuredry) conditionsandapplyovertheratedrangesforambienttemperature,humidity ,andatmosphericpressure.
Maximumvariationfromnominal:+/-15%
4ftto25ftCannula/OxygenTubingLengths
BreathsPerMinute
152025303540
PulseSetting=P1
PulseSetting=P2
PulseSetting=P3
PulseSetting=P4
PulseSetting=P5
14.711.08.87.36.3
29.322.017.614.712.611.0
44.033.026.422.018.916.5
58.744.035.229.325.122.0
66.750.040.033.328.625.0
5.5
1194969-C69
Page 70
Invacare®Platinum®Mobile
8.3RegulatoryListing
ETLcertied complying with:
EN/IEC60601-1;Ed:3.1 EN/IEC60601-1-2;Ed:4 AAMIES60601–1(UnitedStates) CSA22.2No.60601–1(Canada) ISO80601–2–69 ISO80601–2–67 IEC60601–1–6 IEC60601–1–8 IEC60601–1–11 RTCADO160G
701194969-C
Page 71
8.4ElectromagneticCompliance(EMC)
Guidanceandmanufacturer´sdeclaration—electromagneticemission
TheDeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.ThecustomerortheuseroftheDevice shouldassurethatitisusedinsuchanenvironment.
TechnicalData
Emissionstest
RFemissions CISPR11
RFemissions CISPR11
Harmonic emissions IEC61000-3-2
ComplianceElectromagneticenvironment-guidance
TheDeviceusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissions
GroupI
areverylowandarenotlikelytocauseanyinterferenceinnearbyelectronic equipment.
ClassB
ClassA
TheDeviceissuitableforuseinallestablishmentsincludingdomestic establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes. Voltageuctuations /ickeremissions
Complies
IEC61000-3-3
Guidanceandmanufacturer´sdeclaration—electromagneticimmunity
TheDeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.ThecustomerortheuseroftheDevice shouldassurethatitisusedinsuchanenvironment.
Immunitytest
Electrostatic discharge(ESD)
IEC61000-4-2
IEC60601testlevelCompliancelevelElectromagneticenvironment–guidance
±8kVcontact
±8kVcontact
Floorsshouldbewood,concreteorceramictile. Ifoorsarecoveredwithsyntheticmaterial,the
±15kVair
±15kVair
relativehumidityshouldbeatleast30%.
1194969-C
71
Page 72
Invacare®Platinum®Mobile
ElectricalFast transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltagedips
IEC61000-4-11
Voltage Interruptions
IEC61000-4-11
Powerfrequency (50/60Hz) magneticeld
IEC61000-4-8
±2kVforpowersupply lines
±1kVforinput/output lines
±1kVline(s)toline(s)
±2kVline(s)toground
0%UT;1cycle
and
70%UT;25/30cycles
Singlephase:at0°
±2kVforpowersupply lines
±1kVforinput/output lines
±1kVline(s)toline(s)
±2kVline(s)toground
0%UT;1cycle
and
70%UT;25/30cycles
Singlephase:at0°
0%UT;250/300cycle0%UT;250/300cycle
30A/m30A/m
Mainspowerqualityshouldbethatofatypical commercialorhospitalenvironment.
Mainspowerqualityshouldbethatofatypical commercialorhospitalenvironment.
Mainspowerqualityshouldbethatofatypical commercialorhospitalenvironment.Iftheuser oftheDevicerequirescontinuedoperationduring powermainsinterruptions,itisrecommendedthat theDevicebepoweredfromanun-interruptible powersupplyorabattery.
UTisthea.c.mainsvoltagepriortoapplication ofthetestlevel.
Mainspowerqualityshouldbethatofatypical commercialorhospitalenvironment.Iftheuser oftheDevicerequirescontinuedoperationduring powermainsinterruptions,itisrecommendedthat theDevicebepoweredfromanun-interruptible powersupplyorabattery.
UTisthea.c.mainsvoltagepriortoapplication ofthetestlevel.
Powerfrequencymagneticeldsshouldbeat levelscharacteristicofatypicallocationinatypical commercialorhospitalenvironment.
72
1194969-C
Page 73
ConductedRF
IEC61000-4-6
RadiatedRF
IEC61000-4-3
3V3V
10V/m10V /m
TechnicalData
PortableandmobileRFcommunicationsequipment shouldbeusednoclosertoanypartofthe Deviceincludingcables,thantherecommended separationdistancecalculatedfromtheequation applicabletothefrequencyofthetransmitter .
Recommendedseparationdistance:
d=1.2√P150kHzto80MHz
d=0.35√P
d=0.7√P
80MHzto800MHz
800MHzto2,5GHz
wherePisthemaximumoutputpowerrating ofthetransmitterinwatts(W)accordingto thetransmittermanufactureranddisthe recommendedseparationdistanceinmetres(m).
FieldstrengthsfromxedRFtransmitters,as determinedbyanelectromagneticsitesurvey, shouldbelessthanthecompliancelevelineach frequencyrange.
b
Interferencemayoccurinthevicinityofequipment markedwiththefollowingsymbol:
a
Fieldstrengthsfromxedtransmitters,suchasbasestationsforradio(cellular/cordless)telephonesandlandmobileradios,
amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywithaccuracy.T oassess
1194969-C73
a
Page 74
Invacare®Platinum®Mobile
theelectromagneticenvironmentduetoxedRFtransmitters,anelectromagneticsitesurveyshouldbeconsidered.Ifthe measuredeldstrengthinthelocationinwhichtheDeviceisusedexceedstheapplicableRFcompliancelevelabove,theDevice shouldbeobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary, suchasreorientingorrelocatingtheDevice.
b
Overthefrequencyrange150kHzto80MHz,eldstrengthsshouldbelessthan[V1]V/m.
At80MHzand800MHz,thehigherfrequencyrangeapplies.
RecommendedseparationdistancesbetweenportableandmobileRFcommunicationsequipmentandtheDevice
TheDeviceisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.The customerortheuseroftheDevicecanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetween portableandmobileRFcommunicationsequipment(transmitters)andtheDeviceasrecommendedbelow,accordingtothe maximumoutputpowerofthecommunicationsequipment
Separationdistanceaccordingtofrequencyoftransmitter[m]
Ratedmaximumoutput oftransmitter[W]
150kHzto80MHz
d=1.2√P
80MHzto800MHz
d=0.35√P
800MHzto2,5GHz
d=0.7√P
0.010.120.040.07
0.10.370.110.22
11.170.350.70
103.691.112.21
10011.673.507.00
74
1194969-C
Page 75
TechnicalData
Fortransmittersratedatamaximumoutputpowernotlistedabovetherecommendedseparation,distancedinmeters(m) canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePisthemaximumoutputpower ratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer .
At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.
Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreection fromstructures,objectsandpeople.
1194969-C75
Page 76
Invacare®Platinum®Mobile
9Warranty
9.1LimitedWarranty
Refertothewarrantyinformationincludedwithyour productforwarrantydetails.
761194969-C
Page 77
Notes
Page 78
Notes
Page 79
Notes
Page 80
InvacareCorporation
USA
Manufacturer
OneInvacareWay Elyria,OhioUSA 44035 440–329–6000 800–333–6900 TechnicalServices 440-329-6593 800-832-4707 www.invacare.com
1194969-C2018-10-09
*1194969C*
InvacareCorporation 2101E.LakeMaryBlvd. Sanford,FLUSA32773 407–321–5630
MakingLife’sExperiencesPossible®
Page 81
Invacare®Platinum®Mobile
POC1-100B,POC1-100C
enOxygenConcentrator
UserManual
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,thismanualMUSTbereadandsavedfor futurereference.
Page 82
©2018InvacareCorporation Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibited withoutpriorwrittenpermissionfromInvacare.T rademarksareidentiedby™and®.Alltrademarks areownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted. PlatinumisaregisteredtrademarkintheU.S.A.,EuropeandAustralia. MakingLife’sExperiencesPossibleisaregisteredtrademarkintheU.S.A. Pine-sol,LestoilandCloroxareregisteredtrademarksofTheCloroxCompany. DawnisaregisteredtrademarkofTheProcterandGambleCompany. IOSisatrademarkorregisteredtrademarkofCiscointheU.S.andothercountriesandisused underlicensetoApple,Inc. AndroidisatrademarkofGoogle,LLC. BluetoothisaregisteredtrademarkofBluetoothSIG,Inc.
Page 83
Contents
1General.........................................5
1.1QuickStartGuide..............................5
1.2Symbols.....................................9
1.3IndicationsForUse.............................11
1.4IntendedUse.................................11
1.5Contraindications...............................11
1.6Features.....................................12
1.6.1ControlPanel...............................12
1.6.2Input/OutputConnections.....................13
1.6.3Filters....................................13
1.6.4PowerOptions..............................14
1.6.5CarryBag..................................14
1.7OptionalAccessoriesandReplacementParts..........14
2Safety..........................................15
2.1LabelLocations................................15
2.2GeneralGuidelines.............................17
2.3RadioFrequencyInterference.....................22
2.4TravelingwiththeConcentrator....................22
3Setup...........................................25
3.1Unpacking....................................25
3.2Inspection....................................25
3.3PoweringtheOxygenConcentrator.................25
3.3.1InstallingtheBattery(s).......................26
3.3.2ChargingtheBattery(s)........................27
3.3.3ReadingtheBatteryGaugeforInstalledand
UninstalledBattery(s)........................28
3.3.4RemovingtheBattery(s).......................29
3.3.5ExternalACPower...........................30
3.3.6ExternalDCPower...........................30
3.4CarryBagSetup...............................31
3.4.1InstallingtheCarryBag.......................31
3.4.2RemovingtheCarryBag.......................32
3.4.3InstallingShoulderStraps......................32
3.5Connecting/PositioningtheNasalCannula............33
4Usage..........................................35
4.1Location.....................................35
4.2ConcentratorOperatingPosition...................35
4.3UsingtheCarryBag............................35
4.4TurningOntheConcentrator......................36
4.5TurningOfftheConcentrator......................37
4.6WarmUpPeriod...............................37
4.7BreathingwiththeUseoftheConcentrator...........37
4.8ReadingtheDisplayScreen.......................37
4.9AdjustingthePulseFlowSetting...................38
4.10BatteryLifeandManagement....................39
5Maintenance.....................................41
5.1Maintenance..................................41
5.2ServiceLife...................................41
5.3CleaningtheCabinet............................41
5.4CleaningtheAirIntakeFilterScreen................42
5.5CleaningtheCarryBag..........................42
5.6CleaningandDisinfectionBetweenPatients...........43
5.7ViewingHourMeter............................44
5.8PreventativeMaintenanceChecklist.................45
6AfterUse........................................47
6.1Storage......................................47
6.2Disposal.....................................47
6.3WearandT ear................................48
7Troubleshooting...................................49
7.1Troubleshooting...............................49
7.2AlarmConditions...............................51
7.2.1OperationalAlarms..........................52
7.2.2ShutDownAlarms...........................60
8T echnicalData....................................65
8.1TechnicalDescription............................65
8.1.1PneumaticDiagram..........................65
Page 84
8.2Specications.................................66
8.3RegulatoryListing..............................70
8.4ElectromagneticCompliance(EMC).................71
9Warranty........................................76
9.1LimitedWarranty..............................76
Page 85
General
1General
1.1QuickStartGuide
IMPORTANT!
–TheQuickStartGuideisforreferenceONLY .
Itisimperativetoreadthemanualinits ENTIRETYforfullinstructions.
DANGER! RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,death orinjurytopersons:
–DONOTdisassemble.Referservicingto
qualiedservicepersonnel.Therearenouser serviceableparts.
WARNING! RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduse andspecicationshasnotbeentestedandmay leadtoproductdamage,lossofproductfunction, orinjury.
–DONOTusethisproductinanywayotherthan
describedinthespecicationsandintended usesectionsofthismanual.
DANGER! RiskofDeath,InjuryOrDamageFromFire
Textiles,oilorpetroleumsubstances,grease, greasysubstancesandothercombustiblesare easilyignitedandburnwithgreatintensityin oxygenenrichedairandwhenincontactwith oxygenunderpressure.Smokingduringoxygen therapyisdangerousandislikelytoresultin burnsordeath.T oavoidre,death,injuryor damage:
–DONOTSMOKEwhileusingthisdevice.DO
NOTusenearOPENFLAMEorIGNITION SOURCES.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic electricitycreatedbyanytypeoffriction.
–Useonlyoxygencompatiblewater-basedlotions
orsalvesbeforeandduringoxygentherapy. Toverify,refertothelotion/salvecontainer foroxygencompatiblewater-basedstatement. Ifnecessary,contactthemanufacturer.DO NOTuseanylubricantsonconcentratorunless recommendedbyInvacare.
–Keeptheoxygentubing,cord,ACadapter,
andconcentratoroutfromundersuchitems asblankets,bedcoverings,chaircushions, clothing,andawayfromheatedorhotsurfaces includingspaceheaters,stoves,andsimilar electricalappliances.
–Makesureconcentratorisoffwhennotinuse.
1194969-C5
Page 86
Invacare®Platinum®Mobile
DANGER! RiskOfDeath,InjuryOrDamage
Improperuseoftheproductmaycausedeath, injuryordamage.Thissectioncontainsimportant informationforthesafeoperationanduseofthis product.
–DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely readingandunderstandingtheseinstructions andanyadditionalinstructionalmaterialsuch asusermanuals,servicemanualsorinstruction sheetssuppliedwiththisproductoroptional equipment.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare professional,dealerortechnicalpersonnel beforeattemptingtousethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct isnotworkingcorrectly,contactatechnician orInvacareforrepair .
–Thisproductisintendedtobeusedbyadults
onlyafterreadingandunderstandingthe instructionsandwarningsofthisusermanual.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
61194969-C
Page 87
General
1.
Openthebatterycompartmentdoor.Withbattery fuelgaugelabelfacingtowardthefrontoftheoxygen concentrator ,slowlylowerbatteryAstraightdowninto oneofthetwoslotsinthebatterycompartment.
RefertoInstallingtheBattery(s)inSetup.
2.
ConnecttheACPowerAdaptertotheexternalpower inputconnectorB.PlugtheACcordintotheelectrical outlet.
RefertoExternalACPowerinSetup.
Usingtheconcentratorforthersttimerequires thebattery(s)tobecharged.RefertoCharging theBattery(s)inSetup.
3.
Connectthenasalcannulatotheconcentratoroxygen outletportonyourconcentratorC.Oxygenwillonly bedeliveredwhileyouarebreathingthroughthenasal cannulaconnectedtothisdevice.
RefertoConnecting/PositioningtheNasalCannulain Setup.
1194969-C
7
Page 88
Invacare®Platinum®Mobile
4.
TurnontheconcentratorbypressingtheOn/OffButton Duntilthecontrolpanelindicatorslightup.
RefertoTurningontheConcentratorinUsage.
5.
Adjustthepulseowsettingtotheoneprescribedby yourclinicalprofessional.Thepulseowsettingcanbe changedbypressingtheowselectionbuttonE.Place thenasalcannulaoveryourearsandpositiontheprongs inyournoseasinstructedbyyourhealthcareprovider.
RefertoAdjustingthePulseFlowSettinginUsage.
81194969-C
Page 89
General
1.2Symbols
Signalwordsareusedinthismanualandapplytohazards orunsafepracticeswhichcouldresultinpersonalinjury orpropertydamage.Refertotheinformationbelowfor denitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardous
situationwhich,ifnotavoided,willresultin deathorseriousinjury .
WARNING!
–Warningindicatesapotentiallyhazardous
situationwhich,ifnotavoided,couldresultin deathorseriousinjury .
CAUTION!
–Cautionindicatesapotentiallyhazardous
situationwhich,ifnotavoided,mayresultin propertydamageorminorinjuryorboth.
IMPORTANT!
–Importantindicatesahazardoussituationthat
couldresultindamagetopropertyifitisnot avoided.
Givesusefultips,recommendationsandinformation forefcient,trouble-freeuse.
Refertothefollowingtableforthemeaningofsymbols markedontheequipmentand/orpackaging.
GeneralWarningSign
Thebackgroundcolorinsidethe triangleisyellowonproductlabels.
FollowInstructionsforUse
Thecolorofthesymbolbackgroundis blueonproductlabels.
Noopename
Thecolorofthecirclewithdiagonal barisredonproductlabels.
DONOTSmoke
Thecolorofthecirclewithdiagonal barisredonproductlabels.
ConsultInstructionsforUse
Thissymbolislocatedonthecannula packing.
DirectCurrent
TypeBFAppliedPart
1194969-C9
Page 90
Invacare®Platinum®Mobile
ClassIIEquipment
PowerOn/Off
DataOutputTerminal
DONOTdisposeofinhouseholdwaste
Manufacturer
DateofManufacture
Recycle
ReferenceNumber
SerialNumber
DONOTReuse(SinglePatientUseOnly)
Thissymbolreferstothecannula whichisforsinglepatientuseonly.
IndoorUseONL Y
ThissymbolisincludedontheAC poweradapter.
DONOTuseoilorgrease
Keepdry
Protectedagainstsolidforeignobjectsof12.5 mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops whenenclosuretiltedupto15°.
Protectedagainstsolidforeignobjectsof12.5 mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops.
TransportandStorageT emperature
TransportandStorageHumidity
ElectricalSafetyAgencyCerticationMark
LatexFree
101194969-C
Page 91
General
1.3IndicationsForUse
TheInvacare®Platinum®MobileOxygenConcentratoris intendedtoprovidesupplementaloxygentopatientswith respiratorydisorders.Thedevicecanbeusedinahome, institution,vehicle,orotherenvironmentsoutsidethehome.
TheInvacareMobileMedicalApplicationAccessory POC1-CONNECTisintendedforusewiththeInvacare PlatinumMobileOxygenConcentratorPOC1-100Bor POC1-100Cdevice,toallowpatientsviatheirAndroid™or iOS™mobilephoneortablettodisplaydevicesettings,and tocollectdeviceperformanceandusageinformationfor maintenance/servicingpurposesonly.
1.4IntendedUse
TheInvacare®Platinum®MobileOxygenConcentratoris intendedtoprovidesupplementaloxygentopatientswith respiratorydisorders.TheInvacare®Platinum®Mobile OxygenConcentratorcanbeusedinahome,institution, vehicle,orotherenvironmentsoutsidethehome.The deviceisnotintendedtobelife-supportingorlife-sustaining.
TheInvacareMobileMedicalApplicationAccessory POC1-CONNECTisintendedtoallowpatientsviatheir AndroidoriOSmobilephoneortablettodisplaydevice settings,andtocollectdeviceperformanceandusage informationformaintenance/servicingpurposesonly.
WARNING! RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduse andspecicationshasnotbeentestedandmay leadtoproductdamage,lossofproductfunction, orpersonalinjury.
–DONOTusethisproductinanywayotherthan
describedinthespecicationsandintended usesectionsofthismanual.
1.5Contraindications
WARNING! RiskofInjury
Thisproductistobeusedasanoxygen supplementandisnotintendedtobe life-supportingorlife-sustaining.ONL Yusethis productifthepatientiscapableofspontaneous breath,abletoinhaleandexhalewithouttheuse ofamachine
–DONOTuseinparallelorserieswithother
oxygenconcentratorsoroxygentherapy devices.
1194969-C
11
Page 92
Invacare®Platinum®Mobile
WARNING! RiskofMinorInjuryorDiscomfort
Theconserving,orpulsedose,oxygendelivery techniqueusedbythisdeviceiscontraindicated inpersonswhosebreathingduringnormalresting wouldbeunabletotriggerthedevice.
–Properdevicetriggering,setupandoperation
–Notforpediatricuse. –Notforusebytracheotomizedpatients.
1.6Features
1.6.1ControlPanel
mustbeconrmedbyanexperiencedclinician orotherrespiratoryprofessional.
On/OffButton
A
Pressforonesecondtoturnconcentratoronoroff.
FlowSelectionButton
Presstoadjustthepulsedoxygenowtoyour
B
prescribedsetting.Theselectedowsettingis shownonthedisplayscreen.RefertoAdjustingthe PulseFlowSettinginUsageformoredetails.
DisplayBacklightButton
C
Pressoncetoilluminatethedisplayscreenfor5 seconds.
AudioOffButton
Presstomutetheaudiblesignal(beep)forboth alarmsandstatusindications.Theblueindicator
D
lightnexttothebuttonilluminateswhenactivated. Audiooffcanbeactivatedduringorpriortoan alarmconditionoccurring.
DisplayScreen
E
Displaysinformationabouttheoperatingstatusof theconcentrator .
AlarmIndicator
Theyellowindicatorlightatthecenterofthe
F
triangularsymbolilluminatestoindicateabnormal operatingconditions.RefertoAlarmConditionsin Troubleshootingformoredetails.
12
1194969-C
Page 93
General
BatteryGauge
Indicatesthechargestatusofthebattery(s).Refer
G
toReadingtheBatteryGaugeforInstalledand UninstalledBattery(s)inSetupformoredetails.
ExternalPowerIndicator
Theorangeindicatorlightnexttotheplugsymbol
H
illuminateswhenanexternalpowersourceis connectedtotheconcentrator .
BatteryCompartmentDoor
I
Usedtoaccesstheremovablebattery(s).
AudibleSignal(Beep)
Not
Shown
Indicatesachangeinoperatingstatusor aconditionwhichrequirestheoperator’s attention.
1.6.2Input/OutputConnections
A
B
C
OxygenOutletPort
Usedtoconnectthenasalcannulatothe concentrator .
ExternalPowerInputConnector
Usedtoconnectanexternalpowersourcetothe concentrator .
USBPort
Usedforservice.LocationofPOC1–CONNECT Dongle.
1.6.3Filters
Airenterstheconcentratorthroughanairintakelter locatedonthecarrybag.Thislterpreventshairandother largeparticlesintheairfromenteringtheunit.Beforeyou
1194969-C13
Page 94
Invacare®Platinum®Mobile
operatetheconcentrator,makesurethislteriscleanand dry,andtheconcentratorisproperlyinstalledinthecarry bag.
Tocleantheairintakelter ,refertoCleaningtheAirIntake FilterScreeninMaintenance.
Toensuretheconcentratorisproperlyinstalledinthecarry bag,refertoInstallingtheCarryBaginSetup.
1.6.4PowerOptions
WARNING! RiskofInjuryorDamage
Toavoidinjuryordamagewhichwillvoid warranty:
–UseonlyInvacarespeciedpowersupplies.
Battery(s):Uptotworechargeablebatteriescanbeinstalled inthebatterycompartmentoftheconcentrator .Whenfully charged,asinglebatterysuppliespowerforuptovehours (uptotenhoursfortwobatteries).Visibleandaudible alarmsignalsoccurwhenthebatterypowerisgettinglow. RefertoAlarmConditionsinTroubleshootingandCharging theBattery(s)inSetup.
ACPowerAdapter:AnACpoweradapterallowsthe concentratortobeconnectedtoa100–240volt50–60 hertzoutlet.UseoftheACpoweradapterwillallowthe concentratortooperateandsimultaneouslyrechargethe installedbattery.RefertoChargingtheBattery(s)inSetup.
theconcentratortooperateandsimultaneouslyrechargean installedbattery(s).RefertoChargingtheBattery(s)inSetup.
1.6.5CarryBag
Thecarrybagprovidesaconvenientmeanstoholdthe concentratorwhileperformingdailyactivities.
Itcanbeconguredasamessengerbagorbackpackusing theshoulderstrapsprovided.RefertoInstallingtheCarry BaginSetup.
1.7OptionalAccessoriesandReplacementParts
Thefollowingoptionalaccessoriesareavailablefrom Invacare:
Additionalbatteries,modelnumber:POC1–110
Externalbatterychargerwithpowersupply:POC1–115
ThefollowingreplacementpartsareavailablefromInvacare:
Battery,modelnumber:POC1–110
CarryBag,modelnumber:POC1–150
DCpowercable,modelnumber:POC1–140
ACpoweradapterwithpowercord:POC1–130
USBDongle,modelnumber:POC1–CONNECT
Thesepartsarebasemodels.PleasecontactInvacare oryourproviderforcountryspecicmodels.
DCPowerCable:ADCpowercableallowstheconcentrator tobeconnectedtoanautomobile’s(boat,motorhome, etc.)12-voltDCoutlet.UseoftheDCpowercablewillallow
14
1194969-C
Page 95
2Safety
2.1LabelLocations
Thelabelislocatedonthebackoftheconcentrator.Theserialnumberislocatedontherightsideoflabelshownhere.
ModelPOC1-100B
Theconcentratorhasfourpulseowsettings.
Safety
1194969-C15
Page 96
Invacare®Platinum®Mobile
ModelPOC1-100C
Theconcentratorhasvepulseowsettings.
161194969-C
Page 97
Safety
2.2GeneralGuidelines
Inordertoensurethesafeinstallation,assemblyand operationoftheconcentratortheseinstructionsMUSTbe followed.
DANGER! RiskOfDeath,InjuryOrDamage
Improperuseoftheproductmaycausedeath, injuryordamage.Thissectioncontainsimportant informationforthesafeoperationanduseofthis product.
–DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely readingandunderstandingtheseinstructions andanyadditionalinstructionalmaterialsuch asusermanuals,servicemanualsorinstruction sheetssuppliedwiththisproductoroptional equipment.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare professional,dealerortechnicalpersonnel beforeattemptingtousethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct isnotworkingcorrectly,contactatechnician orInvacareforrepair .
–Thisproductisintendedtobeusedbyadults
onlyafterreadingandunderstandingthe instructionsandwarningsofthisusermanual.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER! RiskofDeath,InjuryOrDamageFromFire
Textiles,oilorpetroleumsubstances,grease, greasysubstancesandothercombustiblesare easilyignitedandburnwithgreatintensityin oxygenenrichedairandwhenincontactwith oxygenunderpressure.Smokingduringoxygen therapyisdangerousandislikelytoresultin burnsordeath.T oavoidre,death,injuryor damage:
–DONOTSMOKEwhileusingthisdevice.DO
NOTusenearOPENFLAMEorIGNITION SOURCES.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Keepallopenames,matches,lighted
cigarettes,electroniccigarettesorothersources ofignitionatleast10ft(3m)awayfromthis concentratororanyoxygencarryingaccessories suchascannulasortanks.
–Ifyoudisregardthesewarningsaboutthe
severehazardofoxygenusewhileyoucontinue tosmoke,youmustalwaysturnoffthe concentrator ,removethecannulaandthen waittenminutesbeforesmokingorleavethe roomwhereeithertheconcentratororany oxygencarryingaccessoriessuchascannulas ortanksarelocated.
–DONOTSMOKEwhileusingtheconcentrator .
1194969-C
17
Page 98
Invacare®Platinum®Mobile
CAUTION!
Federal(statutory)lawrestrictsthisdeviceto salebyorontheorderofamedicalpractitioner licensedbyagovernmentalagencywherehe/she practices.
–ONL Yalicensedmedicalpractitionermayorder
thepurchaseoruseofthisdevice.
DANGER! RiskofDeath,InjuryOrDamageFromFire
Textiles,oilorpetroleumsubstances,grease, greasysubstancesandothercombustiblesare easilyignitedandburnwithgreatintensityin oxygenenrichedairandwhenincontactwith oxygenunderpressure.Toavoidre,death, injuryordamage:
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic electricitycreatedbyanytypeoffriction.
–Useonlyoxygencompatiblewater-basedlotions
orsalvesbeforeandduringoxygentherapy. Toverify,refertothelotion/salvecontainer foroxygencompatiblewater-basedstatement. Ifnecessary,contactthemanufacturer.DO NOTuseanylubricantsonconcentratorunless recommendedbyInvacare.
–Keeptheoxygentubing,cord,ACadapter,
andconcentratoroutfromundersuchitems asblankets,bedcoverings,chaircushions, clothing,andawayfromheatedorhotsurfaces includingspaceheaters,stoves,andsimilar electricalappliances.
–Makesureconcentratorisoffwhennotinuse.
181194969-C
Page 99
Safety
CAUTION! RiskofMinorInjuryorDiscomfort
Itisimportanttoplanaheadfortravelandother situationsinwhichyoumaynothaveaccessto additionaloxygenorpowersupplies.T oprevent oxygendeprivation:
–Carrysupplementalbatterieswithyou. –Stowtheconcentratorproperlyduringtravel.
Forproperstorageinstructionswhentravelling refertothetravelguide.
WARNING! RiskofInjuryorDamage
Topreventinjuryordamagefrommisuse:
–NEVERleaveconcentratorunattendedwhen
connectedtopower. –Makesureconcentratorisoffwhennotinuse. –OutdooruseMUSTbeconductedwithinternal
batterypoweronly.
WARNING! RiskOfInjuryOrDamage
–Invacareproductsarespecicallydesigned
andmanufacturedforuseinconjunction
withInvacareprovidedorInvacarespecied
accessories.Anyotheraccessorieshave
notbeentestedbyInvacareandarenot
recommendedforusewithInvacareproducts. –Nomodicationofthisequipmentisallowed.
CAUTION! RiskofMinorInjuryorDiscomfort
Achangeinaltitudemayaffecttotaloxygen availabletoyou.Topreventoxygendeprivation:
–Consultyourphysicianbeforetravelingto
higherorloweraltitudestodetermineifyour owsettingsshouldbechanged.
DANGER! RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,death orinjurytopersons:
–DONOTdisassemble.Referservicingto
qualiedservicepersonnel.Therearenouser serviceableparts.
–Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the concentratormustbelocatedinanotherroom atleast7ft(2.5m)fromthebath.
–DONOTcomeincontactwiththeconcentrator
whilewet.
–DONOTplaceorstoreconcentratorwhereit
candropintowaterorotherliquid.
–DONOTreachforconcentratorthathasfallen
intowater .UnplugIMMEDIATELY.
–DONOTblockaccesstopoweroutletneeded
tounplugtheACpowercord. –DONOTusefrayedordamagedpowercords. –DONOTuseACpoweradapterifitshousingis
crackedorseparated.
1194969-C19
Page 100
Invacare®Platinum®Mobile
WARNING! RiskofInjuryorDeath
Topreventinjuryordeathfromproductmisuse:
–Closelysupervisewhenthisconcentratoris
usedbyimpairedindividualsornearchildren and/orimpairedindividuals.
–Monitorpatientsusingthisdevicewhoare
unabletohearorseealarmsorcommunicate discomfort.
CAUTION! RiskofMinorInjuryorDiscomfort
WhileInvacarestrivestoproducethebestoxygen concentratorinthemarkettoday,thisoxygen concentratorcanfailtoproduceoxygendueto powerfailureordevicemalfunction.
–ALWA YShaveabackupsourceofoxygenreadily
available.
–Intheeventtheconcentratorfailstoproduce
oxygen,theconcentratorwillbrieyalarm signalingthepatienttoswitchtotheirbackup sourceofoxygen.RefertoT roubleshootingfor moredetail.
–Invacarerecommendskeepingatleastone
batteryinstalledintheconcentratorevenwhen operatingfromanexternalpowersource.
WARNING! RiskofDeath,InjuryorDamage
Topreventinjuryordamagefromcordmisuse:
–DONOTmoveorrelocateconcentratorby
pullingonthecord. –DONOTuseextensioncordswithACpower
adaptersprovided. –Properlystoreandpositionelectricalcords
and/ortubingtopreventatrippingand
strangulationhazards.
WARNING! RiskofInjuryorDamage
Topreventinjuryordamageduringuse:
–Ifyoufeelilloruncomfortable,orifthe
concentratordoesnotsignalanoxygenpulse
andyouareunabletohearand/orfeelthe
oxygenpulse,consultyourequipmentprovider
and/oryourphysicianIMMEDIA TEL Y . –Foroptimumperformance,Invacare
recommendsthateachuseoftheconcentrator
beaminimumof30minutes.Shorterperiods
ofoperationmayreducemaximumproductlife. –Theconcentratorshouldbeusedinanupright
position. –Theconcentratorcannotbeusedinconjunction
withPAP ,Bi-Level,mechanicalventilatoror
othersuchdevices.
201194969-C