Page 1

Invacare®Platinum®Mobile
POC1-100B,POC1-100C
enOxygenConcentrator
UserManual
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,thismanualMUSTbereadandsavedfor
futurereference.
Page 2

©2018InvacareCorporation
Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibited
withoutpriorwrittenpermissionfromInvacare.T rademarksareidentiedby™and®.Alltrademarks
areownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
PlatinumisaregisteredtrademarkintheU.S.A.,EuropeandAustralia.
MakingLife’sExperiencesPossibleisaregisteredtrademarkintheU.S.A.
Pine-sol,LestoilandCloroxareregisteredtrademarksofTheCloroxCompany.
DawnisaregisteredtrademarkofTheProcterandGambleCompany.
IOSisatrademarkorregisteredtrademarkofCiscointheU.S.andothercountriesandisused
underlicensetoApple,Inc.
AndroidisatrademarkofGoogle,LLC.
BluetoothisaregisteredtrademarkofBluetoothSIG,Inc.
Page 3

Contents
1General.........................................5
1.1QuickStartGuide..............................5
1.2Symbols.....................................9
1.3IndicationsForUse.............................11
1.4IntendedUse.................................11
1.5Contraindications...............................11
1.6Features.....................................12
1.6.1ControlPanel...............................12
1.6.2Input/OutputConnections.....................13
1.6.3Filters....................................13
1.6.4PowerOptions..............................14
1.6.5CarryBag..................................14
1.7OptionalAccessoriesandReplacementParts..........14
2Safety..........................................15
2.1LabelLocations................................15
2.2GeneralGuidelines.............................17
2.3RadioFrequencyInterference.....................22
2.4TravelingwiththeConcentrator....................22
3Setup...........................................25
3.1Unpacking....................................25
3.2Inspection....................................25
3.3PoweringtheOxygenConcentrator.................25
3.3.1InstallingtheBattery(s).......................26
3.3.2ChargingtheBattery(s)........................27
3.3.3ReadingtheBatteryGaugeforInstalledand
UninstalledBattery(s)........................28
3.3.4RemovingtheBattery(s).......................29
3.3.5ExternalACPower...........................30
3.3.6ExternalDCPower...........................30
3.4CarryBagSetup...............................31
3.4.1InstallingtheCarryBag.......................31
3.4.2RemovingtheCarryBag.......................32
3.4.3InstallingShoulderStraps......................32
3.5Connecting/PositioningtheNasalCannula............33
4Usage..........................................35
4.1Location.....................................35
4.2ConcentratorOperatingPosition...................35
4.3UsingtheCarryBag............................35
4.4TurningOntheConcentrator......................36
4.5TurningOfftheConcentrator......................37
4.6WarmUpPeriod...............................37
4.7BreathingwiththeUseoftheConcentrator...........37
4.8ReadingtheDisplayScreen.......................37
4.9AdjustingthePulseFlowSetting...................38
4.10BatteryLifeandManagement....................39
5Maintenance.....................................41
5.1Maintenance..................................41
5.2ServiceLife...................................41
5.3CleaningtheCabinet............................41
5.4CleaningtheAirIntakeFilterScreen................42
5.5CleaningtheCarryBag..........................42
5.6CleaningandDisinfectionBetweenPatients...........43
5.7ViewingHourMeter............................44
5.8PreventativeMaintenanceChecklist.................45
6AfterUse........................................47
6.1Storage......................................47
6.2Disposal.....................................47
6.3WearandT ear................................48
7Troubleshooting...................................49
7.1Troubleshooting...............................49
7.2AlarmConditions...............................51
7.2.1OperationalAlarms..........................52
7.2.2ShutDownAlarms...........................60
8T echnicalData....................................65
8.1TechnicalDescription............................65
8.1.1PneumaticDiagram..........................65
Page 4

8.2Specications.................................66
8.3RegulatoryListing..............................70
8.4ElectromagneticCompliance(EMC).................71
9Warranty........................................76
9.1LimitedWarranty..............................76
Page 5

General
1General
1.1QuickStartGuide
IMPORTANT!
–TheQuickStartGuideisforreferenceONLY .
Itisimperativetoreadthemanualinits
ENTIRETYforfullinstructions.
DANGER!
RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,death
orinjurytopersons:
–DONOTdisassemble.Referservicingto
qualiedservicepersonnel.Therearenouser
serviceableparts.
WARNING!
RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduse
andspecicationshasnotbeentestedandmay
leadtoproductdamage,lossofproductfunction,
orinjury.
–DONOTusethisproductinanywayotherthan
describedinthespecicationsandintended
usesectionsofthismanual.
DANGER!
RiskofDeath,InjuryOrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,
greasysubstancesandothercombustiblesare
easilyignitedandburnwithgreatintensityin
oxygenenrichedairandwhenincontactwith
oxygenunderpressure.Smokingduringoxygen
therapyisdangerousandislikelytoresultin
burnsordeath.T oavoidre,death,injuryor
damage:
–DONOTSMOKEwhileusingthisdevice.DO
NOTusenearOPENFLAMEorIGNITION
SOURCES.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic
electricitycreatedbyanytypeoffriction.
–Useonlyoxygencompatiblewater-basedlotions
orsalvesbeforeandduringoxygentherapy.
Toverify,refertothelotion/salvecontainer
foroxygencompatiblewater-basedstatement.
Ifnecessary,contactthemanufacturer.DO
NOTuseanylubricantsonconcentratorunless
recommendedbyInvacare.
–Keeptheoxygentubing,cord,ACadapter,
andconcentratoroutfromundersuchitems
asblankets,bedcoverings,chaircushions,
clothing,andawayfromheatedorhotsurfaces
includingspaceheaters,stoves,andsimilar
electricalappliances.
–Makesureconcentratorisoffwhennotinuse.
1194969-C5
Page 6

Invacare®Platinum®Mobile
DANGER!
RiskOfDeath,InjuryOrDamage
Improperuseoftheproductmaycausedeath,
injuryordamage.Thissectioncontainsimportant
informationforthesafeoperationanduseofthis
product.
–DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely
readingandunderstandingtheseinstructions
andanyadditionalinstructionalmaterialsuch
asusermanuals,servicemanualsorinstruction
sheetssuppliedwiththisproductoroptional
equipment.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare
professional,dealerortechnicalpersonnel
beforeattemptingtousethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct
isnotworkingcorrectly,contactatechnician
orInvacareforrepair .
–Thisproductisintendedtobeusedbyadults
onlyafterreadingandunderstandingthe
instructionsandwarningsofthisusermanual.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
61194969-C
Page 7

General
1.
Openthebatterycompartmentdoor.Withbattery
fuelgaugelabelfacingtowardthefrontoftheoxygen
concentrator ,slowlylowerbatteryAstraightdowninto
oneofthetwoslotsinthebatterycompartment.
RefertoInstallingtheBattery(s)inSetup.
2.
ConnecttheACPowerAdaptertotheexternalpower
inputconnectorB.PlugtheACcordintotheelectrical
outlet.
RefertoExternalACPowerinSetup.
Usingtheconcentratorforthersttimerequires
thebattery(s)tobecharged.RefertoCharging
theBattery(s)inSetup.
3.
Connectthenasalcannulatotheconcentratoroxygen
outletportonyourconcentratorC.Oxygenwillonly
bedeliveredwhileyouarebreathingthroughthenasal
cannulaconnectedtothisdevice.
RefertoConnecting/PositioningtheNasalCannulain
Setup.
1194969-C
7
Page 8

Invacare®Platinum®Mobile
4.
TurnontheconcentratorbypressingtheOn/OffButton
Duntilthecontrolpanelindicatorslightup.
RefertoTurningontheConcentratorinUsage.
5.
Adjustthepulseowsettingtotheoneprescribedby
yourclinicalprofessional.Thepulseowsettingcanbe
changedbypressingtheowselectionbuttonE.Place
thenasalcannulaoveryourearsandpositiontheprongs
inyournoseasinstructedbyyourhealthcareprovider.
RefertoAdjustingthePulseFlowSettinginUsage.
81194969-C
Page 9

General
1.2Symbols
Signalwordsareusedinthismanualandapplytohazards
orunsafepracticeswhichcouldresultinpersonalinjury
orpropertydamage.Refertotheinformationbelowfor
denitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardous
situationwhich,ifnotavoided,willresultin
deathorseriousinjury .
WARNING!
–Warningindicatesapotentiallyhazardous
situationwhich,ifnotavoided,couldresultin
deathorseriousinjury .
CAUTION!
–Cautionindicatesapotentiallyhazardous
situationwhich,ifnotavoided,mayresultin
propertydamageorminorinjuryorboth.
IMPORTANT!
–Importantindicatesahazardoussituationthat
couldresultindamagetopropertyifitisnot
avoided.
Givesusefultips,recommendationsandinformation
forefcient,trouble-freeuse.
Refertothefollowingtableforthemeaningofsymbols
markedontheequipmentand/orpackaging.
GeneralWarningSign
Thebackgroundcolorinsidethe
triangleisyellowonproductlabels.
FollowInstructionsforUse
Thecolorofthesymbolbackgroundis
blueonproductlabels.
Noopename
Thecolorofthecirclewithdiagonal
barisredonproductlabels.
DONOTSmoke
Thecolorofthecirclewithdiagonal
barisredonproductlabels.
ConsultInstructionsforUse
Thissymbolislocatedonthecannula
packing.
DirectCurrent
TypeBFAppliedPart
1194969-C9
Page 10

Invacare®Platinum®Mobile
ClassIIEquipment
PowerOn/Off
DataOutputTerminal
DONOTdisposeofinhouseholdwaste
Manufacturer
DateofManufacture
Recycle
ReferenceNumber
SerialNumber
DONOTReuse(SinglePatientUseOnly)
Thissymbolreferstothecannula
whichisforsinglepatientuseonly.
IndoorUseONL Y
ThissymbolisincludedontheAC
poweradapter.
DONOTuseoilorgrease
Keepdry
Protectedagainstsolidforeignobjectsof12.5
mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops
whenenclosuretiltedupto15°.
Protectedagainstsolidforeignobjectsof12.5
mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops.
TransportandStorageT emperature
TransportandStorageHumidity
ElectricalSafetyAgencyCerticationMark
LatexFree
101194969-C
Page 11

General
1.3IndicationsForUse
TheInvacare®Platinum®MobileOxygenConcentratoris
intendedtoprovidesupplementaloxygentopatientswith
respiratorydisorders.Thedevicecanbeusedinahome,
institution,vehicle,orotherenvironmentsoutsidethehome.
TheInvacareMobileMedicalApplicationAccessory
POC1-CONNECTisintendedforusewiththeInvacare
PlatinumMobileOxygenConcentratorPOC1-100Bor
POC1-100Cdevice,toallowpatientsviatheirAndroid™or
iOS™mobilephoneortablettodisplaydevicesettings,and
tocollectdeviceperformanceandusageinformationfor
maintenance/servicingpurposesonly.
1.4IntendedUse
TheInvacare®Platinum®MobileOxygenConcentratoris
intendedtoprovidesupplementaloxygentopatientswith
respiratorydisorders.TheInvacare®Platinum®Mobile
OxygenConcentratorcanbeusedinahome,institution,
vehicle,orotherenvironmentsoutsidethehome.The
deviceisnotintendedtobelife-supportingorlife-sustaining.
TheInvacareMobileMedicalApplicationAccessory
POC1-CONNECTisintendedtoallowpatientsviatheir
AndroidoriOSmobilephoneortablettodisplaydevice
settings,andtocollectdeviceperformanceandusage
informationformaintenance/servicingpurposesonly.
WARNING!
RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduse
andspecicationshasnotbeentestedandmay
leadtoproductdamage,lossofproductfunction,
orpersonalinjury.
–DONOTusethisproductinanywayotherthan
describedinthespecicationsandintended
usesectionsofthismanual.
1.5Contraindications
WARNING!
RiskofInjury
Thisproductistobeusedasanoxygen
supplementandisnotintendedtobe
life-supportingorlife-sustaining.ONLYusethis
productifthepatientiscapableofspontaneous
breath,abletoinhaleandexhalewithouttheuse
ofamachine
–DONOTuseinparallelorserieswithother
oxygenconcentratorsoroxygentherapy
devices.
1194969-C
11
Page 12

Invacare®Platinum®Mobile
WARNING!
RiskofMinorInjuryorDiscomfort
Theconserving,orpulsedose,oxygendelivery
techniqueusedbythisdeviceiscontraindicated
inpersonswhosebreathingduringnormalresting
wouldbeunabletotriggerthedevice.
–Properdevicetriggering,setupandoperation
–Notforpediatricuse.
–Notforusebytracheotomizedpatients.
1.6Features
1.6.1ControlPanel
mustbeconrmedbyanexperiencedclinician
orotherrespiratoryprofessional.
On/OffButton
A
Pressforonesecondtoturnconcentratoronoroff.
FlowSelectionButton
Presstoadjustthepulsedoxygenowtoyour
B
prescribedsetting.Theselectedowsettingis
shownonthedisplayscreen.RefertoAdjustingthe
PulseFlowSettinginUsageformoredetails.
DisplayBacklightButton
C
Pressoncetoilluminatethedisplayscreenfor5
seconds.
AudioOffButton
Presstomutetheaudiblesignal(beep)forboth
alarmsandstatusindications.Theblueindicator
D
lightnexttothebuttonilluminateswhenactivated.
Audiooffcanbeactivatedduringorpriortoan
alarmconditionoccurring.
DisplayScreen
E
Displaysinformationabouttheoperatingstatusof
theconcentrator .
AlarmIndicator
Theyellowindicatorlightatthecenterofthe
F
triangularsymbolilluminatestoindicateabnormal
operatingconditions.RefertoAlarmConditionsin
Troubleshootingformoredetails.
12
1194969-C
Page 13

General
BatteryGauge
Indicatesthechargestatusofthebattery(s).Refer
G
toReadingtheBatteryGaugeforInstalledand
UninstalledBattery(s)inSetupformoredetails.
ExternalPowerIndicator
Theorangeindicatorlightnexttotheplugsymbol
H
illuminateswhenanexternalpowersourceis
connectedtotheconcentrator .
BatteryCompartmentDoor
I
Usedtoaccesstheremovablebattery(s).
AudibleSignal(Beep)
Not
Shown
Indicatesachangeinoperatingstatusor
aconditionwhichrequirestheoperator’s
attention.
1.6.2Input/OutputConnections
A
B
C
OxygenOutletPort
Usedtoconnectthenasalcannulatothe
concentrator .
ExternalPowerInputConnector
Usedtoconnectanexternalpowersourcetothe
concentrator .
USBPort
Usedforservice.LocationofPOC1–CONNECT
Dongle.
1.6.3Filters
Airenterstheconcentratorthroughanairintakelter
locatedonthecarrybag.Thislterpreventshairandother
largeparticlesintheairfromenteringtheunit.Beforeyou
1194969-C13
Page 14

Invacare®Platinum®Mobile
operatetheconcentrator,makesurethislteriscleanand
dry,andtheconcentratorisproperlyinstalledinthecarry
bag.
Tocleantheairintakelter ,refertoCleaningtheAirIntake
FilterScreeninMaintenance.
Toensuretheconcentratorisproperlyinstalledinthecarry
bag,refertoInstallingtheCarryBaginSetup.
1.6.4PowerOptions
WARNING!
RiskofInjuryorDamage
Toavoidinjuryordamagewhichwillvoid
warranty:
–UseonlyInvacarespeciedpowersupplies.
Battery(s):Uptotworechargeablebatteriescanbeinstalled
inthebatterycompartmentoftheconcentrator .Whenfully
charged,asinglebatterysuppliespowerforuptovehours
(uptotenhoursfortwobatteries).Visibleandaudible
alarmsignalsoccurwhenthebatterypowerisgettinglow.
RefertoAlarmConditionsinTroubleshootingandCharging
theBattery(s)inSetup.
ACPowerAdapter:AnACpoweradapterallowsthe
concentratortobeconnectedtoa100–240volt50–60
hertzoutlet.UseoftheACpoweradapterwillallowthe
concentratortooperateandsimultaneouslyrechargethe
installedbattery.RefertoChargingtheBattery(s)inSetup.
theconcentratortooperateandsimultaneouslyrechargean
installedbattery(s).RefertoChargingtheBattery(s)inSetup.
1.6.5CarryBag
Thecarrybagprovidesaconvenientmeanstoholdthe
concentratorwhileperformingdailyactivities.
Itcanbeconguredasamessengerbagorbackpackusing
theshoulderstrapsprovided.RefertoInstallingtheCarry
BaginSetup.
1.7OptionalAccessoriesandReplacementParts
Thefollowingoptionalaccessoriesareavailablefrom
Invacare:
•Additionalbatteries,modelnumber:POC1–110
•Externalbatterychargerwithpowersupply:POC1–115
ThefollowingreplacementpartsareavailablefromInvacare:
•Battery,modelnumber:POC1–110
•CarryBag,modelnumber:POC1–150
•DCpowercable,modelnumber:POC1–140
•ACpoweradapterwithpowercord:POC1–130
•USBDongle,modelnumber:POC1–CONNECT
Thesepartsarebasemodels.PleasecontactInvacare
oryourproviderforcountryspecicmodels.
DCPowerCable:ADCpowercableallowstheconcentrator
tobeconnectedtoanautomobile’s(boat,motorhome,
etc.)12-voltDCoutlet.UseoftheDCpowercablewillallow
14
1194969-C
Page 15

2Safety
2.1LabelLocations
Thelabelislocatedonthebackoftheconcentrator.Theserialnumberislocatedontherightsideoflabelshownhere.
ModelPOC1-100B
Theconcentratorhasfourpulseowsettings.
Safety
1194969-C15
Page 16

Invacare®Platinum®Mobile
ModelPOC1-100C
Theconcentratorhasvepulseowsettings.
161194969-C
Page 17

Safety
2.2GeneralGuidelines
Inordertoensurethesafeinstallation,assemblyand
operationoftheconcentratortheseinstructionsMUSTbe
followed.
DANGER!
RiskOfDeath,InjuryOrDamage
Improperuseoftheproductmaycausedeath,
injuryordamage.Thissectioncontainsimportant
informationforthesafeoperationanduseofthis
product.
–DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely
readingandunderstandingtheseinstructions
andanyadditionalinstructionalmaterialsuch
asusermanuals,servicemanualsorinstruction
sheetssuppliedwiththisproductoroptional
equipment.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare
professional,dealerortechnicalpersonnel
beforeattemptingtousethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct
isnotworkingcorrectly,contactatechnician
orInvacareforrepair .
–Thisproductisintendedtobeusedbyadults
onlyafterreadingandunderstandingthe
instructionsandwarningsofthisusermanual.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER!
RiskofDeath,InjuryOrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,
greasysubstancesandothercombustiblesare
easilyignitedandburnwithgreatintensityin
oxygenenrichedairandwhenincontactwith
oxygenunderpressure.Smokingduringoxygen
therapyisdangerousandislikelytoresultin
burnsordeath.T oavoidre,death,injuryor
damage:
–DONOTSMOKEwhileusingthisdevice.DO
NOTusenearOPENFLAMEorIGNITION
SOURCES.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Keepallopenames,matches,lighted
cigarettes,electroniccigarettesorothersources
ofignitionatleast10ft(3m)awayfromthis
concentratororanyoxygencarryingaccessories
suchascannulasortanks.
–Ifyoudisregardthesewarningsaboutthe
severehazardofoxygenusewhileyoucontinue
tosmoke,youmustalwaysturnoffthe
concentrator ,removethecannulaandthen
waittenminutesbeforesmokingorleavethe
roomwhereeithertheconcentratororany
oxygencarryingaccessoriessuchascannulas
ortanksarelocated.
–DONOTSMOKEwhileusingtheconcentrator .
1194969-C
17
Page 18

Invacare®Platinum®Mobile
CAUTION!
Federal(statutory)lawrestrictsthisdeviceto
salebyorontheorderofamedicalpractitioner
licensedbyagovernmentalagencywherehe/she
practices.
–ONL Yalicensedmedicalpractitionermayorder
thepurchaseoruseofthisdevice.
DANGER!
RiskofDeath,InjuryOrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,
greasysubstancesandothercombustiblesare
easilyignitedandburnwithgreatintensityin
oxygenenrichedairandwhenincontactwith
oxygenunderpressure.Toavoidre,death,
injuryordamage:
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic
electricitycreatedbyanytypeoffriction.
–Useonlyoxygencompatiblewater-basedlotions
orsalvesbeforeandduringoxygentherapy.
Toverify,refertothelotion/salvecontainer
foroxygencompatiblewater-basedstatement.
Ifnecessary,contactthemanufacturer.DO
NOTuseanylubricantsonconcentratorunless
recommendedbyInvacare.
–Keeptheoxygentubing,cord,ACadapter,
andconcentratoroutfromundersuchitems
asblankets,bedcoverings,chaircushions,
clothing,andawayfromheatedorhotsurfaces
includingspaceheaters,stoves,andsimilar
electricalappliances.
–Makesureconcentratorisoffwhennotinuse.
181194969-C
Page 19

Safety
CAUTION!
RiskofMinorInjuryorDiscomfort
Itisimportanttoplanaheadfortravelandother
situationsinwhichyoumaynothaveaccessto
additionaloxygenorpowersupplies.T oprevent
oxygendeprivation:
–Carrysupplementalbatterieswithyou.
–Stowtheconcentratorproperlyduringtravel.
Forproperstorageinstructionswhentravelling
refertothetravelguide.
WARNING!
RiskofInjuryorDamage
Topreventinjuryordamagefrommisuse:
–NEVERleaveconcentratorunattendedwhen
connectedtopower.
–Makesureconcentratorisoffwhennotinuse.
–OutdooruseMUSTbeconductedwithinternal
batterypoweronly.
WARNING!
RiskOfInjuryOrDamage
–Invacareproductsarespecicallydesigned
andmanufacturedforuseinconjunction
withInvacareprovidedorInvacarespecied
accessories.Anyotheraccessorieshave
notbeentestedbyInvacareandarenot
recommendedforusewithInvacareproducts.
–Nomodicationofthisequipmentisallowed.
CAUTION!
RiskofMinorInjuryorDiscomfort
Achangeinaltitudemayaffecttotaloxygen
availabletoyou.Topreventoxygendeprivation:
–Consultyourphysicianbeforetravelingto
higherorloweraltitudestodetermineifyour
owsettingsshouldbechanged.
DANGER!
RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,death
orinjurytopersons:
–DONOTdisassemble.Referservicingto
qualiedservicepersonnel.Therearenouser
serviceableparts.
–Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the
concentratormustbelocatedinanotherroom
atleast7ft(2.5m)fromthebath.
–DONOTcomeincontactwiththeconcentrator
whilewet.
–DONOTplaceorstoreconcentratorwhereit
candropintowaterorotherliquid.
–DONOTreachforconcentratorthathasfallen
intowater .UnplugIMMEDIATELY.
–DONOTblockaccesstopoweroutletneeded
tounplugtheACpowercord.
–DONOTusefrayedordamagedpowercords.
–DONOTuseACpoweradapterifitshousingis
crackedorseparated.
1194969-C19
Page 20

Invacare®Platinum®Mobile
WARNING!
RiskofInjuryorDeath
Topreventinjuryordeathfromproductmisuse:
–Closelysupervisewhenthisconcentratoris
usedbyimpairedindividualsornearchildren
and/orimpairedindividuals.
–Monitorpatientsusingthisdevicewhoare
unabletohearorseealarmsorcommunicate
discomfort.
CAUTION!
RiskofMinorInjuryorDiscomfort
WhileInvacarestrivestoproducethebestoxygen
concentratorinthemarkettoday,thisoxygen
concentratorcanfailtoproduceoxygendueto
powerfailureordevicemalfunction.
–ALWA YShaveabackupsourceofoxygenreadily
available.
–Intheeventtheconcentratorfailstoproduce
oxygen,theconcentratorwillbrieyalarm
signalingthepatienttoswitchtotheirbackup
sourceofoxygen.RefertoT roubleshootingfor
moredetail.
–Invacarerecommendskeepingatleastone
batteryinstalledintheconcentratorevenwhen
operatingfromanexternalpowersource.
WARNING!
RiskofDeath,InjuryorDamage
Topreventinjuryordamagefromcordmisuse:
–DONOTmoveorrelocateconcentratorby
pullingonthecord.
–DONOTuseextensioncordswithACpower
adaptersprovided.
–Properlystoreandpositionelectricalcords
and/ortubingtopreventatrippingand
strangulationhazards.
WARNING!
RiskofInjuryorDamage
Topreventinjuryordamageduringuse:
–Ifyoufeelilloruncomfortable,orifthe
concentratordoesnotsignalanoxygenpulse
andyouareunabletohearand/orfeelthe
oxygenpulse,consultyourequipmentprovider
and/oryourphysicianIMMEDIA TEL Y .
–Foroptimumperformance,Invacare
recommendsthateachuseoftheconcentrator
beaminimumof30minutes.Shorterperiods
ofoperationmayreducemaximumproductlife.
–Theconcentratorshouldbeusedinanupright
position.
–Theconcentratorcannotbeusedinconjunction
withPAP ,Bi-Level,mechanicalventilatoror
othersuchdevices.
201194969-C
Page 21

Safety
CAUTION!
RiskofMinorInjury,DiscomfortorDamage
–Useofthisdeviceatanaltitudeabove10,000
ft(3048m)oroutsideatemperatureof41°F
to104°F(5°Cto40°C)orarelativehumidity
above90%isexpectedtoadverselyeffectthe
owrateandthepercentageofoxygenand
consequentlythequalityoftherapy.
WARNING!
RiskofInjuryorDamage
Asasafetyfeature,thisappliancemayhavea
polarizedplug(onebladeiswiderthantheother).
Toavoidinjuryordamagefromelectricalshock:
–Thisplugwilltinapolarizedoutletonlyone
way.Iftheplugdoesnottfullyintheoutlet,
reversetheplug.Ifitstilldoesnott,contact
aqualiedelectrician.
–DONOTalterthepluginanyway.
–DONOTuseifthecordorplugaredamaged.
Callaqualiedelectricianforrepair .
CAUTION!
RiskofDamage
Topreventdamagefromtemperaturevariations:
–DONOToperateintemperaturesbelow41°F
(5°C)orabove104°F(40°C).
–Whenyourautomobileisturnedoff,disconnect
thecaraccessorypowercordandremove
theconcentratorfromtheautomobile.DO
NOTstoretheconcentratorinaveryhot
orcoldautomobileorinothersimilar,high
orlow,temperatureenvironments.Referto
SpecicationsinTechnicalData.
WARNING!
RiskofInjuryorDamage
Topreventdamagefromliquidingress:
–Iftheconcentratorisnotworkingproperly,if
ithasbeendroppedordamaged,ordropped
intowater ,callequipmentprovider/qualied
technicianforexaminationandrepair.
–NEVERdroporinsertanyobjectorliquidinto
anyopening.
–Invacarerecommendstheconcentratornotbe
usedintherain.
–Theconcentratorisnotdesignedtobeused
withahumidier .Useofthisdevicewitha
humidiermayimpairperformanceand/or
damagetheequipment.
–DONOTuseifcabinetofconcentratoror
housingofACpoweradapteriscrackedor
separated.
1194969-C
21
Page 22

Invacare®Platinum®Mobile
2.3RadioFrequencyInterference
MedicalElectricalEquipmentneedstobeinstalledandused
accordingtotheEMCinformationinthismanual.
Thisequipmenthasbeentestedandfoundtocomplywith
EMClimitsspeciedbyIEC/EN60601-1-2.Theselimits
aredesignedtoprovideareasonableprotectionagainst
electromagneticinterferenceinatypicalhomehealthcare
environment.
PortableandmobileRFcommunicationsequipmentcan
affecttheoperationofthisequipment.Useofthis
equipmentadjacenttoorstackedwithotherequipment
shouldbeavoidedasitcouldresultinimproperoperation.
Theconcentratoroperationshouldbeobservedtoverify
normaloperationinthepresenceofsuchequipment.
Otherdevicesmayexperienceinterferencefromeven
thelowlevelsofelectromagneticemissionspermittedby
theabovestandards.T odetermineiftheemissionsfrom
theconcentratorarecausingtheinterference,turnthe
concentratoroff.Iftheinterferencewiththeotherdevice(s)
stops,thentheconcentratoriscausingtheinterference.In
suchrarecases,interferencemaybereducedorcorrected
byoneofthefollowingmeasures:
•Reposition,relocate,orincreasetheseparationbetween
theequipment.
•Connecttheequipmentintoanoutletonacircuit
differentfromthattowhichtheotherdevice(s)is
connected.
2.4TravelingwiththeConcentrator
BeforeTraveling
1.Contactyourtravelcarrierand/ortourguideabout
yourtravelplansandprovideinformationaboutyour
intenttouseaportableoxygenconcentratorduring
yourtravels.Thecarrierwillletyouknowifyouwillbe
permittedtouseyourconcentratorduringyourtravel
andifthereareanyrestrictionsforuse.
Ifyouarenotpermittedtouseyourconcentrator
onthecarrierofchoiceandstillwanttotake
itwithyou:
•Carryitandanysupplementalbatterieson
boardwithyou.
•Ensuretheunitisturnedoff.
•Stowitproperlyforthetrip.
•DONOTplacetheconcentratororany
supplementalbatteriesincheckedbaggage.
2.Contactthecarriertoseeiftheyhaveseatswitha
poweroutletthatyoucanusetopowertheconcentrator
duringtravel.Ifapoweroutletisnotavailable,besure
tobringenoughsupplementalbatteriestolast150%of
thetraveltime.
Theremaybeunexpecteddelaysbeyond
scheduledtraveltimes.Carryyourpower
adapterswithyousoyoucanrechargebatteries
beforeandafterthetrip.
RefertoElectromagneticCompliance(EMC)inT echnical
Dataforadditionalinformation.
22
1194969-C
Page 23

Safety
3.Chargetheinstalledbatteryoftheconcentratorand
anysupplementalbatteriesfullybeforeyoudepart,thus
extendingtheoperatingtimeoftheconcentratorduring
travel.RefertoChargingtheBattery(s)inSetup.
4.Ensuretheunitisfreeofallgrease,oil,orother
petroleumproductsandthattheunitisingoodworking
order,isfreeofdamage,andtheairlterisclean.
RefertothePreventativeMaintenanceChecklistin
Maintenance.
5.Contactyouroxygensuppliertomakearrangementsif
youneedtohavebackupoxygenatyourdestination.
TravelingByAir
ArrivingattheAirport
1.Allowthesecurityagenttoinspectyourconcentrator ,
evenifyouareusingit,whengoingthroughthesecurity
checkpoint.
Theyarerequiredtoallowyoutotravelthrough
thecheckpointwithyourconcentrator,butthey
MUSTinspectitforsecurityreasons.
2.UseyourACpoweradapterwhileyouareintheairport,
ifpossible.Thiswillkeepthebattery(s)atfullcharge
andgiveyouthemostbatterypoweredoperatingtime
ontheaircraft.
1.Ifyouplanonusingtheconcentratortheentiretime
youareontheplane,verifythatyourconcentratordoes
notblockaccesstoanemergencyexitoraisleway.Ifit
does,requestaseatchange.
2.Inorderfortheightcrewtoensurethesafetyofyou
andalloftheotherpassengersontheight,allowthem
toinspecttheconcentratortoconrmthatitisFAA
approvedforuseontheight.
3.Ifyouarenotplanningtouseyourconcentrator
duringtaxi,takeoff,orlanding,youMUSTstoreitin
anapprovedstorageareasoitdoesnotblocktherow
oraisleaccess.
Contacttheightcrewaboutlocationof
approvedstorageareas.
OnthePlane
YouMUSTstorethesupplementalbatteryeither
securelyconnectedtotheconcentratororinyour
carry-onbaggage.ThesupplementalbatteryMUST
beprotectedfromdamageandshortingoutthe
externalpowerconnector.
BoardingthePlane
Youarenotpermittedtositinanexitrowifyou
plantouseyourconcentratoratanytimeduring
theight.
1194969-C23
Page 24

Invacare®Platinum®Mobile
1.Youmayuseyourconcentratorwhenmovingabout
thecabin,ONL Yafterthepilotturnsoffthe“Fasten
SeatBelt”sign.
2.Turnofftheunitiftheconcentratoralarmsduringthe
ight,unlessthealarmresetsitself.Ifthealarmdoes
notreset,theconcentratorMUSTbeturnedoffand
storedinanapprovedstoragearea.
3.IfanACpoweroutletwithaminimum100Wcapacityis
availableontheight,performthefollowing:
a.Connectthepoweradaptertotheconcentrator .
b.Pluginthepoweradapterintotheairlinepower
outlet.
c.Turntheuniton.
Ifyouhaveanytroublemakingtheconnectionsto
theplane’spoweroutlet,contacttheightcrewfor
assistance.
AftertheFlight
1.Ensureyouhaveenoughpowertorunyourconcentrator
whileleavingtheairport.Rechargethebattery(s),if
necessary.RefertoChargingtheBattery(s)inSetup.
2.Contactyourmedicalgassuppliertoobtainyourbackup
oxygensupply,ifnecessary .
TravelingbyBoat
TravelingbyTrain
Contactthetrainauthoritiesatleasttwelvehoursaheadof
arrivalthatyouaregoingtouseyourconcentrator.There
shouldbenorestrictionsonyouruseoftheconcentrator ,
butpoweronthetrainmaynotbeavailableforyouruse
duringtraveltimes.Ensurethereisenoughbatterylifefor
yourtripbeforeyouleave.
TravelingbyBus
Contactthebuslineaboutusingaconcentrator.There
shouldbenorestrictionsonyouruseoftheconcentrator
duringyourtrip,butpowermaynotbeavailableforyour
useduringtraveltimes.Ensurethereisenoughbatterylife
foryourtripbeforeyouleave.
TravelingbyCar
Thereshouldbenorestrictionsonyouruseofthe
concentratorduringyourtrip,butpowermaynotbe
availableforyouruseduringtraveltimes.Ensurethereis
enoughbatterylifeforyourtripbeforeyouleave.
Contactthecruiselineandinformthemyouaretraveling
withaconcentrator .Thereshouldbenorestrictionsonyour
useoftheconcentratorduringyourtravel,butitiswiseto
checkwiththemaheadoftime.Powershouldbeavailable
foruseofyourACpoweradapterduringyourtravels,but
checkrst.
24
1194969-C
Page 25

Setup
3Setup
3.1Unpacking
1.Checkforanyobviousdamagetothecartonorits
contents.Ifdamageisevident,notifythecarrieror
yourlocaldealer .
2.Removeallloosepackingfromthecarton.
3.Carefullyremoveallthecomponentsfromthecartons.
TheInvacareconcentrator’spackagingcontainsthe
followingitems(asshownbelow)inadditiontothis
usermanual.Ifanypartsaremissing,pleasecontact
yourequipmentprovider.
•Concentrator(ModelPOC1–100B)withcarrybag
(ModelPOC1–150)installed
•Cannula(Westmed#0194)
•Twoshoulderstraps(PartNumber1187483)
•Battery(ModelPOC1–110)
•ACpoweradapter(PartNumber1187452)
•ACpowercord(PartNumber1187454formodel
POC1–100B)
•DCpowercable(ModelPOC1–140)
Retainallcontainersandpackingmaterialsfor
storageorreturnshipment.
3.3PoweringtheOxygenConcentrator
WARNING!
RiskofInjuryorDamage
Toavoidinjuryanddamagethatwillvoidthe
warranty:
–UseonlyInvacarespeciedpowersupplieswith
theoxygenconcentrator .
–PriortousingtheDCpowercable,the
automobile(boat,motorhome,etc.)engine
shouldberunninginordertoeitheroperateor
chargetheoxygenconcentrator.
IMPORTANT
Theconcentratorwillnotturnonunlessthe
batteryischarged.
–ConnectACpowertotheconcentratorand
allowbatterytofullychargepriortorstuse.
Theoxygenconcentratorallowsthefreedomtochoosefrom
thefollowingpowersourcestoenableusebothinsideand
outsideofthehome:
•Rechargeablebattery(s)
•ACpoweroutlet
•12voltDCpowercable
3.2Inspection
Inspect/examineexterioroftheoxygenconcentratorand
accessoriesfordamage.Inspectallcomponents.Ifdamage
isfound,DONOTuseconcentrator.Contactyourprovider
forserviceand/orrepairorforassistancewiththesetup
procedureandtoreportunexpectedoperationorevents.
1194969-C25
Page 26

Invacare®Platinum®Mobile
3.3.1InstallingtheBattery(s)
DANGER!
RiskofDeath,InjuryorDamage
Toavoiddeath,injuryordamagefromre:
–DONOTheatabove140°F(60°C),incinerate,
disassemble,orshortterminals.
–Disposeinaccordancewithalllocalregulations.
Theoxygenconcentratorcomesequippedwithasingle
rechargeablelithiumbattery.Uptotwobatteriescanbe
installedintheconcentrator .Whenfullycharged,asingle
batterysuppliespowerforuptovehours,dependingon
thepulseowsetting.Twobatteriescansupplypowerfor
uptotenhourswhenfullycharged,dependingonthepulse
owsetting.
Abatterycanbeinstalledwhiletheconcentratorisoperating
fromanexternalpowersourceorasecondchargedbattery.
1.OpenthebatterycompartmentdoorF.
2.WithbatteryfuelgaugelabelAfacingtowardthefront
oftheoxygenconcentratorB,slowlylowerbatteryC
straightdownintooneofthetwoslotsinthebattery
compartmentD.
UsetheliftringEtoholdthebatterywhile
installingit.
Thebatterywillnottproperlyinthebattery
compartmentifnotorientedasdescribedabove.
261194969-C
Page 27

Setup
3.Ensurethebatteryfullyengagesthebatteryconnections
locatedatthebottomofthebatterycompartment.
Whenfullyinserted,thetopofthebatteryislevelwith
thetopedgeofthebatterycompartment.
CAUTION!
RiskofDamage
Toavoiddamagetobatteryconnectionsfrom
excessiveforceormisuse:
–DONOTslamorforcebattery(s)intoplace.
–Orientthebattery(s)asdescribedinthe
instructionspriortoinstalling.
Toavoiddamagetothebatterydoorfrom
excessiveforceormisuse:
–DONOTliftconcentratorusingthebattery
door.
–Makesurebattery(s)arefullyinstalled
beforeclosingthisbatterydoor.
–DONOTuseconcentratorwithbattery
dooropen.
4.Ifasecondbatteryisbeingused(soldseparately),install
itinthesamemannerintheunoccupiedbatteryslot.
5.ClosethebatterycompartmentdoorF.
3.3.2ChargingtheBattery(s)
upthebatteries.Tochargethebattery(s),performthe
followingsteps:
1.Installthebattery.RefertoInstallingtheBattery(s)in
Setup.
2.ConnectACpowertotheconcentrator .RefertoExternal
ACPowerinSetup.
3.Monitorthebatterychargelevelandchargeuntilfull.
RefertoBatteryGaugeforInstalledandUninstalled
Battery(s)inSetup.
4.DisconnecttheACpoweradapterfromtheconcentrator
forportableoperation.
IMPORTANT
Aconcentratorwillnotturnonunlessthebattery
ischarged.
–ConnecttheACpowertotheconcentratorand
allowbatterytofullychargepriortorstuse.
–RefertoSpecicationsinT echnicalDatafor
approximatechargetimes.
ChargingtheBattery(s)AfterInitialCharging
Theinstalledbattery(s)canbechargedwhetherthe
concentratorisonoroff.
InitialChargingoftheBattery(s)
Usingtheconcentratorforthersttimerequiresthe
battery(s)tobecharged.Batteriesareshippedina“sleep”
modeandaninitialchargecycleisrequiredto“wake”
1194969-C
27
Page 28

Invacare®Platinum®Mobile
1.Forchargingofthebattery(s)afterinitialcharging,do
oneofthefollowing:
•ConnecttheACpoweradapter(ifchargingfroma
walloutlet).RefertoExternalACPowerinSetup.
•ConnecttheDCpowercable(ifchargingfroma
vehicle).RefertoExternalDCPowerinSetup.
•Removebatteryandchargewiththeoptional
ExternalBatteryChargeraccessory,Model
POC1–115.
2.Monitorthebatterychargelevelandchargeuntilfull.
RefertoBatteryGaugeforInstalledandUninstalled
Battery(s)inSetup.
3.Disconnectexternalpowerfromtheconcentratorfor
portableoperation.
3.3.3ReadingtheBatteryGaugeforInstalledand
UninstalledBattery(s)
BatteryGaugeforInstalledBattery(s)
ThebatterygaugeAshowsthestatusoftheinstalled
batterychargelevelwhentheconcentratoristurnedonand
whentheconcentratorisoffbutconnectedtoanexternal
powersource.ThenumberofilluminatedsegmentsB
ofthebatterygaugeAindicatesthechargelevelofthe
installedbattery .Iftwobatteriesareinstalledtheindicated
chargelevelisthatoftheleastchargedbattery.
NumberofLitSegmentsB
1
2
3
4
ChargeLevel
0–25%charge
26–50%charge
51–75%charge
76%-100%charge
Whentheconcentratorisconnectedtoanexternalpower
source(whethertheconcentratoristurnedonoroff)the
“last”or“highest”segmentBofthebatterygaugeAalso
281194969-C
Page 29

Setup
ashestoshowthebatteryischarging.Whenthebatteryis
3.3.4RemovingtheBattery(s)
fullycharged,thefourthsegmentBofthebatterygaugeA
stopsashingandremainslit.
BatteryGaugeforUninstalledBattery(s)
1.OpenthebatterycompartmentdoorFofthe
concentratorB.
2.PullstraightupontheliftringEtoremovethebattery
CfromthebatterycompartmentD.
3.Ifasecondbatteryisbeingused(soldseparately),
removeitinthesamemanner .
4.Closethebatterycompartmentdoor.
Asteadyaudiblebeepwilloccurwhenallpower
sourcesareremovedfromtheconcentrator .Referto
ShutDownAlarmsinTroubleshooting.
WhenthebatteryDisremovedfromtheconcentrator ,
pressthebatterystatusbuttonCtoshowthestatusofthe
chargelevelonthebatterygaugeA.Refertothecharge
levelBtableinReadingtheBatteryGaugeforInstalledand
Abatterycanberemovedwithoutaffectingtheoperation
oftheconcentratoraslongasanexternalpowersourceor
asecondchargedbatteryisinstalled.
UninstalledBattery(s)inSetup.
1194969-C29
Page 30

Invacare®Platinum®Mobile
3.3.5ExternalACPower
TheACpoweradapterAallowstheoxygenconcentratorto
beconnectedtoa100–240volt50–60Hertzoutlet(i.e.
walloutletofyourhome).ThepoweradapterconvertsAC
voltagetoaDCvoltagethatcanbeusedtopowerthe
oxygenconcentrator .UseoftheACpoweradapterwillallow
theoxygenconcentratortooperateandsimultaneously
rechargethebattery(s).
CompletethefollowingstepstoconnecttheACpower
adaptertotheconcentrator:
1.Ifnotalreadyassembled,attachtheACpowercordE
totheACpoweradapterA.
2.InserttheACpoweradapteroutletconnectorBintothe
externalpowerinputconnectorCoftheconcentrator .
3.InsertplugDintoanACoutlet.
Theexternalpowerinputconnectorofthe
concentratorisangledslightlyupward.Theoutlet
connectorofthepowercordneedstobeangled
slightlyupwardwheninstalled.
Whenabatteryisinstalled,externalACorDCpower
canbeconnectedordisconnectedwhetherthe
concentratorisofforon.
Whenanexternalpowersourceisproperly
connectedtotheconcentrator,theorangeindicator
lightnexttotheplugsymbolonthecontrolpanel
willilluminateandtherewillbeanaudiblebeep.
Whenexternalpowerisdisconnected,theorange
lightwillturnoffandtherewillbeanaudiblebeep.
3.3.6ExternalDCPower
TheDCpowercableGallowstheoxygenconcentratorto
beconnectedtoanautomobile’s(boat,motorhome,etc.)
12–voltDCoutlet.UseoftheDCpowercablewillallow
301194969-C
Page 31

Setup
theoxygenconcentratortobeoperatedandsimultaneously
3.4.1InstallingtheCarryBag
rechargethebattery.
CompletethefollowingstepstoconnecttheDCpowercable
totheconcentrator:
1.InserttheDCpowercableoutletconnectorFintothe
externalpowerinputconnectorCoftheconcentrator .
2.InsertplugHintoaDCoutletaftertheautomobile
(boat,motorhome,etc.)isrunning.
3.4CarryBagSetup
CAUTION!
RiskofDamage
–Improperinstallationoftheconcentratorinto
thecarrybagcancauseahightemperature
alarmconditionandresultintheconcentrator
shuttingdown.
–Thecarrybagisnotdesignedtoprotectthe
concentratorfromexcessivedrops,impacts,
orabuse.
–T oinstallorremovethecarrybagthe
concentratormustbeoffanddisconnected
fromexternalpowerandcannula.
Concentratorcanbeoperatedwithorwithout
thecarrybag.Whenoutofthecarrybag,the
concentratorislimitedtostationaryuse.Referto
UsingtheCarryBaginUsage.
1194969-C31
1.Fullyunzipthebaseofthecarrybag.
2.Positiontheconcentratorstandinguprightonaat
surfaceandorientedwithairintakeopeningAfacing
you.Ensuretheconcentratorisoffanddisconnected
fromexternalpowerandcannula.
3.Positionthecarrybagabovetheconcentratorwiththe
airintakelterscreenBfacingyou.
4.HoldtheunzippedbottomofthecarrybagopenC,
andslidethecarrybagdownontotheconcentrator
untilthecarrybagcompletelycoverstheconcentrator .
Page 32

Invacare®Platinum®Mobile
5.Laytheconcentratorandcarrybagdownwithairintake
lterscreenfacingupward.
6.Pushtheconcentratorintothecarrybaguntilitisfully
seatedagainstthetopretainer .
7.Rotatethebaseofthecarrybagontothebottomof
theconcentratorandfullyzipupthebottomofthe
carrybag.
8.Returntheconcentratorandcarrybagtotheupright
position.
9.Checkthatthecutoutsinthetopretainerarecorrectly
alignedwiththeoxygenoutletportandexternalpower
inputconnector.Ifnotaligned,removethecarrybag
andreinstallasdescribedinsteps1–9.
3.4.2RemovingtheCarryBag
1.Ensuretheconcentratorisoffanddisconnectedfrom
externalpowerandcannula.
2.Laytheconcentratoronaatsurfacewiththeairintake
lterscreenfacingupwardandfullyunzipthebase.
3.Rotatethebaseofthecarrybagoffthebottomofthe
concentrator .
4.Fromthebottom,pulltheconcentratoroutofthecarry
bag.Ifneeded,thetopsurfaceoftheconcentrator
canbepushedtohelpremoveitfromthebottomof
thecarrybag.
5.Returntheconcentratortotheuprightposition.
6.Fullyzipupthebottomofthecarrybag.
3.4.3InstallingShoulderStraps
Thecarrybagcanbeconguredasamessengerbag
usingoneshoulderstraporasabackpackusing
twoshoulderstraps.
MessengerBagConguration
1.UsingasingleshoulderstrapA,attachaclipBfrom
eachendofthestraptoringsCatthetopofthecarry
bagD.
2.Adjustshoulderstraptodesiredlength.
3.SlidepadEonshoulderstraptothedesiredlocation.
321194969-C
Page 33

BackPackConguration
Setup
3.5Connecting/PositioningtheNasalCannula
DANGER!
RiskofInjuryorDeath
Thecannulacancausetripping,falling,orother
injuryifimproperlypositionedandsecured.To
avoidinjuryordeath:
–ThecannulaMUSTberoutedandsecured
properly.
–DONOTpositionthecannulaaroundtheneck.
Ensurethepatientcanmovefreelywhile
wearingthecannula.
–Avoidpositioningcannulasacrossareasofhigh
foottrafc(i.e.aisles,doorways,hallways,etc.)
1.AttacheachclipBattheendsofasingleshoulderstrap
AtothetworingsClocatedatthetopandbottom
ofthecarrybagD.
2.Repeatforthesecondstrapontheoppositeside.
3.Adjusteachshoulderstraptodesiredlength.
4.SlidepadEoneachshoulderstraptothedesired
location.
WARNING!
RiskofInjuryorDeath
Toavoidchokingand/orstrangulationfromtubing
entanglement:
–Keepchildrenandpetsawayfromnasal
cannulaandtubing.
–Closesupervisionisnecessarywhenthenasal
cannulaisusedbyimpairedindividualsornear
childrenand/orimpairedindividuals.
1194969-C33
Page 34

Invacare®Platinum®Mobile
WARNING!
RiskofInjuryorDeath
Toreducetheriskofinjuryordeathfromillness:
–Replacethenasalcannulaonaregularbasis.
Checkwithyourequipmentprovideror
physiciantodeterminehowoftenthecannula
shouldbereplaced.
–DONOTsharecannulasbetweenpatients.
Cannulasareforsinglepatientuseonly.
CAUTION!
RiskofDamage
–Useofsterilizationorcleaningsolutionsmay
leavethisdevicenonfunctional.
CAUTION!
RiskofMinorInjuryorDiscomfort
Toensureproperbreathdetectionandoxygen
delivery:
–Ensurethecannulaprongsarepositioned
properlyinyournose.Thisiscriticaltothe
effectivenessoftheoxygentherapy.
–DONOTusetubing/cannulalengthexceeding
25ft(7,6m).
–Usecrush-proofoxygentubing.
–UseONLYsinglelumencannulaswiththe
owcharacteristicsstatedinSpecicationsin
TechnicalData.
–Avoiduseofconcentratorinwindyconditions.
–Checkforgasowattheoutletofthecannula.
1.Removethecannulafromitspackaging.
2.Connectthecannulatotheconcentratoroxygenoutlet
portA.Ensuretheconnectionissecure.
3.PlacethecannulaBoveryourearsandpositionthe
prongsinyournoseasinstructedbyyourhealthcare
providerorcannulamanufacturer .
Gasowattheoutletofthecannulacanbechecked
whiletheconcentratoriswarmingup.Waveyour
handinfrontofthenasalprongs.Youshouldbe
abletohearandfeelthepulsedowofgas.If
youdonotfeelthegaspulse,checkthecannula
connectionforleaks.
341194969-C
Page 35

Usage
4Usage
4.1Location
Theconcentratorcanbeusedinahome,institution,vehicle
orotherenvironmentsoutsidethehome.
WARNING!
RiskofInjuryorDamage
Toavoidinjuryordamagefromairborne
pollutantsand/orfumesandforoptimal
performance:
–Locateandpositiontheconcentratorinawell
ventilatedspacesothattheairintakeandthe
airexhaustopeningsarenotobstructed.
–NEVERblocktheairopeningsofthe
concentratororplaceitonasoftsurface,such
asabedorcouch,wheretheairopeningmay
beblocked.
–Keeptheopeningsfreefromlint,hairand
similarforeignitems.
–Keepconcentratoratleast12in(30.5cm)away
fromwalls,draperiesandfurniture.
–DONOTuseinpresenceofpollutants,smoke
orfumes,ammableanesthetics,cleaning
agentsorchemicalvapors.
RefertoUsingtheCarryBaginUsagefordetailon
thelocationoftheairintakeandexhaustopenings.
4.2ConcentratorOperatingPosition
Whennotbeingworn,theconcentratorshouldbeoperated
intheuprightpositionandplacedonarm,levelsurface.
Theconcentratorshouldbeinalocationwherethe
audiblesignalcanbeheard.
4.3UsingtheCarryBag
WARNING!
RiskofInjuryorDamage
Toavoidinjuryordamagefromdroppingthe
concentratororincorrectpositioningofthe
concentrator:
–Thecarrybagmustbeusedforportable
operationoftheconcentrator .
–DONOTadjusttheshoulderstrapwhile
carryingtheconcentrator .
–Positionconcentratorsoexhaustventsare
directedawayfromthebodyandcontrolpanel
facesupward.
–UseonlyInvacarecarrybag,modelPOC1–150.
CAUTION!
RiskofMinorInjuryorDiscomfort
Toavoidinjuryfromconcentratormalfunctioning:
–Ensureyouareabletoheartheaudible
signalwarningswhenusingthecarrybagasa
backpackincaseanalarmconditionoccurs.
Thecarrybagcanbeconguredasamessengerbagusing
oneshoulderstraporasabackpackusingtwoshoulder
straps.Thelengthoftheshoulderstrap(s)canbeadjusted
1194969-C35
Page 36

Invacare®Platinum®Mobile
tothedesiredposition.RefertoInstallingShoulderStraps
inSetup.
MessengerBagConguration
BackpackConguration
Whencarryingtheconcentratorwiththeshoulderstrap(s)
A,positiontheconcentratorwiththecontrolpanelfacing
upwardandtheairintakeCopeningtowardsthebody.
ThisorientstheconcentratorsothattheexhaustventB
isdirectedawayfromthebodyandthecontrolpanelis
positionedforproperviewing.
ThecarrybagalsohasahandleDthatcanbeusedtomove
andpositiontheconcentrator .
4.4TurningOntheConcentrator
1.ToturnontheconcentratorpresstheOn/OffbuttonA
untilyouseethecontrolpanelindicatorslightup.
2.Theaudiblesignalwillbeepandalltheindicatorswill
lightupforapproximatelytwosecondsaftertheunitis
rstturnedon.Thispowerupsequenceprovidesthe
opportunitytoensurethatallindicatorsarefunctioning
properly.
3.Followingthepowerupsequenceawarmupperiod
begins.Formoredetails,refertoWarmUpPeriodin
Usage.
361194969-C
Page 37

Usage
4.5TurningOfftheConcentrator
1.ToturnofftheconcentratorpresstheOn/Offbutton
until“PoweringDown”isshownonthedisplayscreen.
2.Thepowerdownsequencetakesaapproximatelythree
seconds.
4.6WarmUpPeriod
CAUTION!
RiskofMinorInjuryorDiscomfort
Duringthewarmupperiod(typicallyless
thanveminutes)oxygenoutputisnotwithin
specicationslistedinSpecicationsinT echnical
Data.
–Concentratormaybeusedduringwarmup
period.
Afterturningontheconcentratorawarmupperiodofupto
15minutesisneededtoreachthespeciedoxygenpurity.
“WarmingUp”isshownonthedisplayscreenduringthis
period.
Whilewarmingup,theconcentratorwillautomatically
deliverapulseofoxygenabouteveryfoursecondswhenno
breathisdetected.Thenasalcannulashouldbeconnected
totheconcentratorduringwarmupperiod.
Whentheconcentratorhasnishedwarmingup,thecurrent
pulsesettingwillbeshownonthedisplayscreen.Ifno
breathisdetectedafterwarmuphasended,theNoBreath
Detectedalarmwilloccur .RefertoAlarmConditionsin
Troubleshooting.
4.7BreathingwiththeUseoftheConcentrator
Asyoubreathewiththenasalcannulainstalled,apulse
ofoxygenisdeliveredeachtimetheconcentratorsenses
aninhalation.RefertoConnecting/PositioningtheNasal
CannulainSetupforcannulainstallationinstructions.
Ifnobreathisdetectedfor15seconds,theNoBreath
Detectedalarmwilloccur .RefertoAlarmConditionsin
Troubleshooting.
4.8ReadingtheDisplayScreen
PulseFlowSettingandPulseIcon
Aftertheconcentratornisheswarmingup,thecurrent
oxygenpulseowsettingBandpulseiconCareshownon
thedisplayscreenA.
Thepulseiconashesoneachtimeanoxygenpulseis
triggered.
1194969-C37
Page 38

Invacare®Platinum®Mobile
Thepulseowsettingandpulseiconarenotshown
whenalarmtextisdisplayedonscreen.
InformationalText
Informationaltextisshownonthedisplayscreenduring
certainperiodsofconcentratoroperation:
•“WarmingUp”—displayedwhiletheconcentratoris
warmingupafterrstbeingturnedon.
•“PoweringDown”—displayedwhiletheconcentrator
ispoweringdownafterbeingturnedoff.
AlarmText
Ifanalarmconditionexists,textdescribingthealarmwillbe
shownonthedisplayscreen.
RefertoAlarmConditionsinTroubleshootingformoredetail
onalarmconditionsandrelatedalarmtext.
4.9AdjustingthePulseFlowSetting
CAUTION!
RiskofMinorInjuryorDiscomfort
Itisveryimportanttoselecttheprescribed
oxygenowsetting.Thiswillensureyouwill
receivethetherapeuticamountofoxygen
accordingtoyourmedicalcondition:
–DONOTincreaseordecreasetheowsetting
unlessachangehasbeenprescribedbyyour
physicianortherapist.
–Thetherapeuticeffectivenessoftheprescribed
oxygenowsettingshouldbeperiodically
reassessed.
–Useonlythelengthofcannulathatwasusedto
determinetheprescribedoxygenowsetting.
–Thepulseowsettingsofthisconcentrator
mightnotcorrespondtocontinuousow
oxygen.
–Thesettingsofothermodelsorbrandsof
oxygentherapyequipmentdonotcorrespond
tothesettingsofthisconcentrator.
381194969-C
Page 39

1.Withtheconcentratorrunning,presstheowselection
buttonBtoincreasethepulseowsettingbyone
increment.
2.Presstheowselectionbuttonadditionaltimes,as
needed,untilyourprescribedowsetting(P1toP4/P5)
isshownonthedisplayscreenC.
Eachtimetheowsettingischanged,anaudible
beepwilloccurandthedisplayscreenbacklightwill
turnonforveseconds.
•POC1-100BModels:Theconcentratorhasfour
pulseowsettings.Whenthemaximumpulse
owsettingP4isreached,thepulseowsetting
returnstosettingP1withthenextbuttonpress.
•POC1-100CModels:Theconcentratorhasve
pulseowsettings.Whenthemaximumpulse
owsettingP5isreached,thepulseowsetting
returnstosettingP1withthenextbuttonpress.
Usage
Ifamessageotherthanthepulseowsettingis
shownonthedisplayscreen,theadjustedpulseow
settingwilltemporarilybedisplayedforveseconds
afterpressingtheowselectionbutton.
Theconcentratorpulseowsettingatpower-upwill
bethesameowsettingselectedthelasttimethe
unitwasturnedoff .
4.10BatteryLifeandManagement
Timeawayfromhomeisgreatlyextendedbycombining
theuseoftheACpoweradapter ,DCpowercableandthe
concentrator’sbatteries.T oensurethebattery(s)maintain
theiroptimalchargelevel,utilizetheACpoweradapter
wheneveryouhaveaccesstoelectricpower .UtilizetheDC
powercablewheneveryouareinavehicle.
CAUTION!
RiskofMinorInjuryorDiscomfort
Batteryexhaustionwillresultinalossof
supplementaloxygen.Toensureproper
supplementaloxygendeliveryduringapower
outage:
–Plugyourportableconcentratorintoan
alternatepowersource.
–Haveanalternatesourceofoxygenavailable
thatdoesnotrequireapowersource.
WhattoDo
•Whenyourstreceiveyourconcentrator,fullycharge
thebattery(s)overnight.
1194969-C39
Page 40

Invacare®Platinum®Mobile
•Keepyourbattery(s)fullychargedwhenusingthe
concentratoronadailybasis.
•Y ourconcentratorbattery(s)canberechargedatany
time.
•Alwaysensuretheconcentratorbattery(s)arerecharged
assoonaspossibleaftertheybecomefullydischarged.
Thebattery(s)maybepermanentlydegradedifleftfully
dischargedforanextendedlengthoftime.
•Checkthestatusofyourconcentratorbattery(s)once
amonthifyouarenotusingyourconcentratoron
adailybasis.Battery(s)shouldbemaintainedattwo
illuminatedbarsworthofchargeifnotusingthe
concentratoronadailybasis.
•Heatistheworstenemyofabattery .Allowplentyof
airtocirculatearoundtheconcentratorsothatthe
battery(s)iskeptascoolaspossiblewheninuseand
whencharging.
WhatNottoDo
•DONOTuseorleavetheconcentratororbattery(s)
inexcessiveheatorcold.
•DONOTstoreorleavetheconcentratororbattery(s)in
cartrunks,etc.forextendedperiodsoftime.
•DONOTstorebattery(s)fullycharged(fourbars
illuminatedontheunit’sbatterygauge)ifyouaregoing
tostoreyourconcentratorforanytimegreaterthan
aday.Rechargeordischargethebatterytotwobars
(50%charge)only.Storingabatterywithafullcharge
maydegradeitsusefullife.
•DONOTleaveyourbattery(s)installedintothe
concentratorwhentheconcentratorisnotgoingto
beusedformorethanaday.Thebattery(s)willlose
chargewhilepluggedintotheconcentratorevenwith
theconcentratorturnedoff.
401194969-C
Page 41

Maintenance
5Maintenance
5.1Maintenance
WARNING!
RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecically
designedtominimizeroutinepreventive
maintenance.Topreventinjuryordamage:
–Onlyprofessionalsofthehealthcareeldor
personsfullyconversantwiththisprocesssuch
asfactorytrainedpersonnelshouldperform
preventivemaintenanceorperformance
adjustmentsontheoxygenconcentratorand
itsequipment,exceptfortasksdescribedin
thismanual.
–UsersshouldcontactyourdealerorInvacare
forservice.
5.2ServiceLife
Theexpectedservicelifeofthisproduct,withtheexception
ofwearcomponents(see6.3WearandTear,page48),is
threeyearsofoperationwhenusedinaccordancewiththe
safetyinstructions,maintenanceintervalsandcorrectuse
statedinthismanual.Theeffectiveservicelifecanvary
accordingtothefrequencyandintensityofuse.Refertothe
proceduresin5Maintenance,page41
.
5.3CleaningtheCabinet
DANGER!
RiskofInjuryorDamage
Liquidwilldamagetheinternalcomponentsof
theconcentratoranditsequipment.T oavoid
damageorinjuryfromelectricalshock:
–T urnOfftheconcentratorandunplugthe
powercordbeforecleaning.
–DONOTallowanycleaningagenttodripinside
theairinletandoutletopenings.
–DONOTsprayorapplyanycleaningagent
directlytothecabinet.
–DONOThosedowntheproduct.
CAUTION!
RiskofDamage
Harshchemicalagentscandamagethe
concentratoranditsequipment.T oavoid
damage:
–DONOTcleanwithalcoholandalcoholbased
products(isopropylalcohol),concentrated
chlorine-basedproducts(ethylenechloride),
andoil-basedproducts(Pine-Sol®,Lestoil®)or
anyotherharshchemicalagents.Onlyusemild
liquiddishdetergent(suchasDawn®).
Performallmaintenanceaccordingtotherecommended
scheduleinthismanual.
1194969-C
41
Page 42

Invacare®Platinum®Mobile
Periodicallycleantheconcentrator’scabinetasfollows:
1.Turnoffconcentrator .
2.Removetheconcentratorfromthecarrybagorback
pack.
3.Useadampcloth,orsponge,withamilddetergent
suchasDawndishwashingsoaptogentlycleanthe
exteriorcase.
4.Allowtheconcentratortoairdry,oruseadrytowel,
beforereturningtheconcentratortothecarrybagor
back,andpriortooperatingtheconcentrator .
5.Returntheconcentratortoitscarrybagorbackpack.
TheACpoweradapterandACpowercordaretobe
cleanedinasimilarmanner .
1.Turnoffconcentrator .
2.Useavacuumcleanerornylonbrushtoclean
dust/debrisfromairintakelterscreenA.
5.5CleaningtheCarryBag
5.4CleaningtheAirIntakeFilterScreen
CAUTION!
RiskofDamage
Toavoiddamagefromclogging:
–Replacethecarrybagiftheairintakelter
screenistornorfrayed.
Cleantheairintakelterscreenatleastonceeach
week.Theairintakelterscreenispartofthecarry
bag.
42
CAUTION!
RiskofDamage
Toavoiddamagetotheproduct:
–DONOTmachinewashordrythecarrybag.
1.Turnoffconcentrator .
2.RemovetheconcentratorAfromthecarrybagB.
3.Connectconcentratortoanexternalpowersource.
1194969-C
Page 43

Maintenance
4.Wipeorbrushthecarrybagwithamildliquiddish
detergent(suchasDawn)andwater.Rinsethoroughly.
5.Allowthecarrybagtoairdryaftercleaningandbefore
using.
5.6CleaningandDisinfectionBetweenPatients
Cleaninganddisinfectionmustbeperformedbetween
patientsandasnecessary .
WARNING!
RiskofInjuryorDamage
Topreventinjuryfrominfectionordamageto
concentrator:
–Onlyqualiedpersonnelshouldperform
cleaninganddisinfectionoftheoxygen
concentratorandaccessoriesbetweenpatients.
Followtheseinstructionstoeliminatepossible
pathogenexchangebetweenpatientsdueto
contaminationofcomponentsoraccessories.
1.Disposeofandreplaceallpatientsideaccessoriesnot
suitableformultiplepatientuse,includingbutnot
limitedto:cannulaandoxygentubing.
2.Cleanconcentratorandaccessoriesasdescribedin
Maintenance.
3.Disinfectthesurfacesoftheconcentratorand
accessoriesusingClorox®4inOneDisinfectant&
Sanitizerorsubstantiallyequivalentproduct.
WARNING!
RiskofInjuryorDamage
–Followthecleaning/disinfectantproduct
manufacturer’sinstructionsforuse,
handling,storageanddisposalofthe
product.
–Useofinappropriatedisinfectingproducts
maydegradedevicematerialsand
negativelyaffectdevicesafetyand
performance.
Iftheaccessories,suchasthecarrybag,are
unabletobecleanedordisinfected,discardthe
accessoryandreplace.
4.PerformthetasksonPreventiveMaintenanceChecklist.
5.Beforerepackaginganddistributiontonewpatient,
ensurepackagingcontentscontaintheconcentrator,
battery,carrybagwithshoulderstraps,ACpower
adapter,ACpowercord,DCpowercable,labelsand
usermanual.
1194969-C43
Page 44

Invacare®Platinum®Mobile
5.7ViewingHourMeter
Toviewtheelapsedhoursofconcentratorruntimeperform
thefollowingsteps:
1.Ensurethattheunitispoweredoffandconnectedto
anexternalpowersource.
2.Pressandholdtheowselectionbuttonforve
seconds.
3.Thecumulativehoursofconcentratoroperation
(compressorhours)willbeshownonthedisplayscreen
foraslongasthebuttoniscontinuallypressed.The
concentratorruntimeisdisplayedtothenearest0.1
hours.
4.Releasetheowselectionbutton.
44
1194969-C
Page 45

5.8PreventativeMaintenanceChecklist
ThefollowingPreventativeMaintenanceChecklistcontainsmaintenancetaskstobeperformedbytheusersofthisproduct,
exceptwhereotherwisenoted.Ifyouareunabletounderstandthesetasks,contactyourprovideroraqualiedtechnician.
WEEKLY
RecordDateofService.
RecordElapsedHoursonHourMeter.
(RefertoViewingHourMeterinMaintenance.)
CleanAirIntakeFilterScreen
InspectCabinetofConcentratorandACAdapterfor
Damage.
Donotuseifdamageisfound.ReturntoInvacarefor
repair .
InspectCarryBagStrap(s)andhardwareforDamage.
Replacestraporcarrybagifdamagefound.
InspectElectricalCordsforDamage.
ReplaceElectricalCordifdamageisfound.
CheckforpresenceandlegibilityofallLabels.
ReplaceLabelsasneeded.RefertoLabelLocations.
Maintenance
1194969-C45
Page 46

Invacare®Platinum®Mobile
EVERY4,380HOURS,EVERY3YEARS,ANDBETWEENP ATIENTS,WHICHEVERCOMESFIRST
CheckOxygenPurity*
PerformFunctionalTest*
DURINGPREVENTATIVEMAINTENANCESCHEDULEANDBETWEENPATIENTS
CheckIntakeFilterScreenfordamage.
Replacecarrybagifdamageisfound.
Check/ReplaceOutletFilter*
Inspectionperiodsshownashoursisinreferencetohoursofconcentratoroperationsincethelastdateofservice.Referto
ViewingHourMeterinMaintenance.Inspectionperiodsshownasmonthsoryearsisinreferencetodurationoftimesincethe
lastdateofservice.
*Tobeconductedbyproviderorqualiedtechnician.Refertoservicemanual.
461194969-C
Page 47

AfterUse
6AfterUse
6.1Storage
1.Removebattery(s)priortostorage.
2.Storetherepackagedoxygenconcentratorandbattery(s)
inacool,dryarea.
3.DONOTplaceobjectsontopofpackagedconcentrator .
RefertostorageconditionsinSpecicationsin
TechnicalData.
Afterremovingfromstorage,itmaytakeuptoone
houratroomtemperaturefortheproducttoreach
itsoperatingtemperatureandbereadyforuse.
6.2Disposal
RecycleDONOTdisposeofin
householdwaste
Thisproducthasbeensuppliedfromanenvironmentally
awaremanufacturerwhocomplieswiththeWasteElectrical
andElectronicEquipment(WEEE)Directive2012/19/EC.This
productmaycontainsubstancesthatcouldbeharmfulto
theenvironmentifdisposedofinplaces(landlls)thatare
notappropriateaccordingtolegislation.
Followlocalgoverningordinancesandrecyclingplans
regardingdisposaloftheconcentratororcomponents
normallyusedinoperation.Theconcentratordoesnot
generatewasteorresidueinoperation.
•DONOTdisposeoftheconcentratorinthenormal
wastestream.
•AnyaccessoriesnotpartoftheconcentratorMUST
behandledinaccordancewiththeindividualproduct
markingfordisposal.
•DONOTdisposeoftheinternalorsupplemental
batteries.Batteriesshouldbereturnedtoyour
dealer/provider .
1194969-C
47
Page 48

Invacare®Platinum®Mobile
6.3WearandTear
Invacarereservestherighttoaskforanyitembackthat
hasanallegeddefectinworkmanship.SeeWarrantythat
shippedwiththeproductforspecicwarrantyinformation.
RefertoMaintenanceinthismanualforproperpreventative
maintenancescheduleanduseoftheproduct.
Thisisonlyageneralguidelineanddoesnotincludeitems
damagedduetoabuseandmisuse.
Normalwearandtearitemsandcomponentsforthis
productarelistedbelow.
Description
Compressorinletlter1,460hours
Carrybag18months
Battery
Sievebedassembly
ExpectedLife
18months
1year
Expectedlifevaluesshownashoursisinreference
tohoursofconcentratoroperation.RefertoViewing
HourMeterinMaintenance.Expectedlifevalues
shownasmonthsoryearsisinreferencetoduration
oftimesincethedateofpurchase.
Sieveisaporouslteringmaterialandisconsidered
awearitem.Somefactorsthatcouldaffect
sievemateriallifeincludehumidity,temperature,
particulates,aircontaminates,airintake,vibration
andotherenvironmentalconditions.
481194969-C
Page 49

Troubleshooting
7Troubleshooting
7.1Troubleshooting
Ifyourconcentratorfailstooperateproperly,refertothefollowingchartforpossiblecausesandsolutions.Ifnecessary,
contactyourprovider.
OxygenConcentratorDoesNotTurnOnorDoesNotStayOn
Symptom
Concentratordoes
notoperatewhen
On/Offbuttonis
pressed.
OR
Concentrator
beginstooperate
whenOn/Off
buttonispressed,
butallLCDand
frontpanellights
godark.
ProbableCause
On/Offbuttonwasnothelddownlongenough.
Batteryisdischarged(orimproperlyconnectedto
theconcentrator).
Batteryisimproperlyconnectedtotheconcentrator .1.Removethebatteryandanyexternalpower
Batteryhasperformedaself-protectshutdown
whentherewasnoexternalpower .
Solution(s)
Trytopoweruptheconcentratoragainwhile
continuingtopresstheOn/Offbuttonuntilthe
frontpanellightsandLCDbegintoilluminate.This
takestypicallyonefullsecond.
1.Removebatteryandcheckbatterygauge.
2.Ifgaugeshowsbatteryischarged,thenreinstall
batteryandretry.
3.Ifgaugeshowsbatteryisdepleted,theneither
installachargedbatteryorconnectexternal
powerandretry.
connection.
2.Reinstallthebattery,makingsureitis
completelyinsertedandretry.
3.ConnecttoeitherACorDCexternalpower
andretry.
1.Waitoneminuteandretry.
2.Changetoanalternatebatteryorconnect
concentratortoeitherACorDCexternalpower
andretry.
1194969-C49
Page 50

Invacare®Platinum®Mobile
ConcentratorisBeepingwithNoLightsorDisplay
Symptom
Concentratorisbeepingwith
nolightsordisplay.
Batteryhasbeenremovedordisconnected
andthereisnoexternalpowerconnected.
ProbableCause
Beepingwilllastless
thantwominutesif
nopowersourceis
provided.
Whenoperatingwithoutabattery,external
powerislost.
BatteryNotChargingWhenExternalPowerisConnected
Symptom
WithACorDC
powerpluggedinto
theconcentrator,
theexternalpower
iconisilluminated
Powersourceisfaulty ,orthereisaloose
connection.
Batteryisoutsidetheallowedtemperaturerange
forcharging.
ProbableCauseSolution
andbatterycharge
levelindicatoris
notashing.
Batteryisnotfullyseated.1.Ensurethebatteryiscompletelyinserted.
Batterychargeiscomplete(allfoursegmentsof
chargelevelindicatorareilluminated).
Solution(s)
1.Ensurethebatteryiscompletelyinserted
andrestart.
2.Installadifferentchargedbatteryand
restart.
1.Tryanotherpoweroutletandcheck
connectionstoconcentratorandrestart.
1.Tryanotherpoweroutletandcheck
connectionstoconcentrator .
1.Allowconcentratorandbatterytocooldown
tolessthan104°F(40°C),orwarmupto41°F
(5°C).
1.Noactionrequired.
501194969-C
Page 51

NoExternalPowerSymbolIlluminatedonControlPanel
Symptom
WithACpower
pluggedinto
concentrator ,
externalpower
iconisnot
illuminatedonthe
Powersourceisfaulty ,orthereisaloose
connection.
ProbableCauseSolution
1.IfthegreenindicatorlightontheACadapter
islit,checkpowerconnectionsfromtheAC
adaptertoconcentrator .
2.IfthegreenindicatorlightontheACadapter
isnotlit,checkpowerconnectionsfromoutlet
toACadapterortryanotheroutlet.
controlpanel.
WithDCpower
pluggedinto
concentrator ,
externalpower
iconisnot
Powersourceisfaultyorthereisaloose
connection.
TheDCpowercablefusehasblown.1.SwitchtoACpowerandcontactyourprovider
1.Checkpowerconnectionstoconcentrator.
2.Tryanotherpoweroutlet.
toservicetheDCpowercable.
illuminatedonthe
controlpanel.
7.2AlarmConditions
Whenanyalarmconditionoccurs,theyellowalarmindicatorlightonthecontrolpanelilluminatesandtextdescribing
thealarmconditionisshownonthedisplayscreen.Refertothesolutionsectionofthealarmtablesforpotential
correctiveactions.Ifnecessary,contactyourprovider.
Inallcases,pressingandholdingtheOn/Offbuttonforonesecondwillturnoffandresettheconcentrator.
Troubleshooting
Ifthecauseofthealarmconditionisnotcorrected,itwillreappearwhentheconcentratoristurnedbackon.
AllalarmsareclassiedasLowPrioritytechnicalalarmconditions.
Whenmultiplealarmconditionsexist,thetextassociatedwiththehighestrankedalarmwillbeshownonthedisplay
screen.
1194969-C51
Page 52

Invacare®Platinum®Mobile
7.2.1OperationalAlarms
Theconcentratorcontinuestorunwhenthealarmconditionslistedinthissectionoccur.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
NoBreathDetected
CheckCannula
Oxygenconcentratorhasnotdetecteda
breathovera15secondperiod.
1.Verifycannulaisconnected,not
kinked,properlypositionedanduser
isbreathingthroughhis/hernose.
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
NOTES:
Alarmwillturnoffwhenabreathisdetected.
Thisalarmdoesnotapplyduringwarmup.
Ifnobreathisdetectedfortwominutes,thealarmwillescalatetoashutdownalarm.RefertoShutdownAlarms
inTroubleshooting.
Therstoccurrenceofthisalarmconditionwillcanceltheaudiooff(mute)feature.Audiooff(mute)canbereactivated.If
reactivated,thealarmaudiooff(mute)featurewillremainactivatedevenifthisalarmconditionoccursagain.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
BreathRateHigh
ReduceActivity
Theuserbreathratehasexceededthe
capacityoftheconcentratorformore
than15seconds.
1.Theusershouldimmediatelyreduce
his/heractivityleveltogethis/her
breathratetoslowdown.
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
521194969-C
Page 53

Troubleshooting
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
NOTES:
Asufcientreductionoftheuser’sbreathratewillautomaticallyturnoffthealarm.
Therstoccurrenceofthisalarmconditionwillcanceltheaudiooff(mute)feature.Audiooff(mute)canbereactivated.If
reactivated,thealarmaudiooff(mute)featurewillremainactivatedevenifthisalarmconditionoccursagain.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
LowBattery
ChargeConcentrator
INDICATORS:
Singleaudiblebeepnorepeat
YELLOWalarmindicatorilluminated
Batterygauge:
Remainingbatteryruntimeof
approximately30minutes.Battery
requirescharging.
1.Connectoxygenconcentratorto
eithertheACorDCpowerorinsert
analternatechargedbatteryinto
theunit,OR
2.Insertanalternatechargedbattery
intotheunit.Removethedepleted
batteryandchargeitwiththe
externalbatterychargerforfuture
use.
•PulseFlowSettingsP1–P3:onlythe
lowestsegmentisilluminated
•PulseFlowSettingP4–P5:only
thelowesttwosegmentsare
illuminated
NOTES:
Thealarmturnsoffwhenexternalpowerisconnectedorwhenachargedbatteryisinstalled.
Alarmwillturnoffwhenabreathisdetected.
1194969-C53
Page 54

Invacare®Platinum®Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
LowBattery
ChargeNow
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
Batterygauge:
Remainingbatteryruntimeof
approximately15minutesorless.
Batteryrequirescharging.
Lossofpowerisimminent.
1.Connectoxygenconcentratorto
eithertheACorDCpower ,OR
2.Insertanalternatechargedbattery
intotheunit.Removethedepleted
batteryandchargeitwiththe
externalbatterychargerforfuture
use,OR
3.Ifnootherpowersourceisavailable,
changetoanothersourceofoxygen.
•Onlythelowestsegmentis
illuminated
NOTES:
Thealarmturnsoffwhenexternalpowerisconnectedorwhenachargedbatteryisinstalled.
Therstoccurrenceofthisalarmconditionwillcanceltheaudiooff(mute)feature.Audiooff(mute)canbereactivated.If
reactivated,thealarmaudiooff(mute)featurewillremainactivatedevenifthisalarmconditionoccursagain.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
UnitHot
Allowtocool
INDICATORS:
Concentratorisnearitsmaximum
operatingtemperature.
1.Moveconcentratortocooler
surroundings.Allowconcentratorto
cooldowntolessthan104°F(40°C).
2.Cleanintakelterscreen.
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
NOTES:
Thealarmwillturnoffwhentherequiredinternaloperatingtemperatureisreached.
541194969-C
Page 55

Troubleshooting
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
UnitCold
Allowtowarm
Concentratorisnearitsminimum
operatingtemperature.
1.Moveconcentratortowarmer
surroundings.Allowconcentratorto
warmupto41°F(5°C).
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
NOTES:
Thealarmwillturnoffwhentherequiredinternaloperatingtemperatureisreached.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
BatteryHot
Allowtocool
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
Batteryisnearitsmaximumoperating
temperature.
1.Moveconcentratortocooler
surroundings.Allowconcentratorto
cooldowntolessthan104°F(40°C),
OR
2.UseACorDCpowerandremove
battery,OR
3.Insertanalternatechargedbattery
andremovethehotbattery .
NOTES:
Thealarmwillturnoffwhentherequiredbatterytemperatureisreachedorifexternalpowerisconnectedandbattery
isremoved.
1194969-C55
Page 56

Invacare®Platinum®Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
BatteryCold
Allowtowarm
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
Batteryisnearitsminimumoperating
temperature.
1.Moveconcentratortowarmer
surroundings.Allowconcentratorto
warmupto41°F(5°C),OR
2.UseACorDCpowerandremove
battery,OR
3.Insertanalternatechargedbattery
andremovethecoldbattery.
NOTES:
Thealarmwillturnoffwhentherequiredinternaloperatingtemperatureisreachedorifexternalpowerisconnectedand
batteryisremoved.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
LowOxygen
SeeManual
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
Theoxygenoutputpurityhasfallentoa
valuebetween73%and87%.
1.Verifyconcentratorisin
recommendedenvironmental
temperature.RefertoSpecications
inTechnicalData.
2.Cleanintakelterscreenandensure
bothintakeandexhaustarenot
blocked.
3.Turnconcentratoroff,thenonagain
toretry.
561194969-C
Page 57

LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
NOTES:
Thealarmwillturnonwhentheoxygenpurityhasdroppedbelow85%.
Thealarmwillturnoffwhentheoxygenoutputpurityhasincreasedaboveto87%.
Thealarmlimitshaveatoleranceof+/-2%.
Alarmsignalgenerationmaybedelayedbyupto128secondsduetothemonitoringalgorithmusedtopreventnuisancealarms.
Callyourproviderandreporttheproblemifthewarningcontinueswithnextuse.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
Cannotcharge
Allowtocool
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
Batteryistoohottoallowthebattery
tocharge.
ThisalarmonlyoccurswhenExternal
Powerisconnected.
1.Moveconcentratortocooler
surroundings.Allowconcentratorto
cooldowntolessthan104°F(40°C),
OR
2.Removebattery(withACorDC
powerconnected)toallowbattery
tocool.
NOTES:
Thealarmwillturnoffwhentherequiredbatterytemperatureisreached,whenthebatteryisremoved,orwhenexternal
powerisdisconnected.
Troubleshooting
1194969-C57
Page 58

Invacare®Platinum®Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
Cannotcharge
Allowtowarm
INDICATORS:
Singleaudiblebeepevery30seconds
Batteryistoocoldtoallowthebattery
tocharge.
Thisalarmonlyoccurswhenanexternal
powersourceisconnected.
1.Moveconcentratortowarmer
surroundings.Allowconcentratorto
warmupto41°F(5°C),OR
2.Removebattery(withACorDC
powerconnected)toallowbattery
towarm.
YELLOWalarmindicatorilluminated
NOTES:
Thealarmwillturnoffwhentherequiredbatterytemperatureisreached,whenthebatteryisremoved,orwhenexternal
powerisdisconnected.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
SensorFault
SeeManual
Oxygensensorfeedbackisout-of-range.
Sensorfailureislikely.
1.TurnconcentratorOff,thenOnagain
toretry.
INDICATORS:
Singleaudiblebeepevery30seconds
YELLOWalarmindicatorilluminated
NOTES:
Thealarmwillturnoffwhenthesensorfeedbackisbackinrange.
Callyourproviderandreporttheproblemifalarmcontinues.
581194969-C
Page 59

Troubleshooting
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
WarmingUp
INDICATORS:
Noaudiblebeep
YELLOWalarmindicatorilluminated
Theconcentratorisstillwarmingup
aftertwominutesfrominitialstart
up.Theoxygenoutputpurityhasnot
reachedtheminimumvaluelistedin
SpecicationsinTechnicalData.
1.Verifyconcentratorisin
recommendedenvironmental
temperature.RefertoSpecications
inTechnicalData.
2.Cleanintakelterscreenandensure
bothintakeandexhaustarenot
blocked.
3.Allowconcentratortocontinue
warmingupforatleast15minutes.
4.Turnconcentratoroff,thenonagain
toretry.
NOTES:
Dependingonavarietyoffactors,itcantakeupto15minutestoreachtheminimumspeciedoxygenpurity.
Thealarmwillturnoffwhentheminimumoxygenoutputpurityisreached.
Iftheminimumoxygenoutputpurityhasnotbeenreachedwithin15minutesfrominitialstartup,thealarmwillbecomea
LowPurityalarmcondition.
1194969-C59
Page 60

Invacare®Platinum®Mobile
7.2.2ShutDownAlarms
Theconcentratorshutsdownwhenthealarmconditionsinthissectionoccur .
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
NoBreathDetected
Oxygenconcentratorhasnotdetecteda
breathforoveratwominuteperiod.
1.Verifycannulaisconnected,not
kinked,properlypositionedanduser
isbreathingthroughhis/hernose.
2.Changetoanothersourceofoxygen
ifalarmcontinues.
NOTES:
Thisalarmdoesnotapplyduringwarmup.
Toclearalarm,turntheconcentratoroff.Completesolution1beforerestarting.PressandholdtheOn/Offbuttontopower
downtheconcentratorandrestart.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
Stuckbutton
INDICATORS:
Twoaudiblebeepsevery16seconds
Astuckbuttonisdetectedupon
connectiontopowersource.
1.Removeallpowersourcesfrom
concentrator .Presseachcontrol
buttonlookingforastuckbutton.
Reconnectpower .
2.Changetoanothersourceofoxygen
ifalarmcontinues.
YELLOWalarmindicatorilluminated
NOTES:
Toclearalarm,turntheconcentratoroff.Completesolution1beforerestarting.Callyourproviderandreporttheproblem
ifalarmcontinues.
601194969-C
Page 61

LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
BatteryDepleted
INDICATORS:
Remainingbatterycapacityhasbeen
completelyused.Batteryrequires
charging.
Concentratorfunctionhasshutdown.
1.Connectoxygenconcentratorto
2.Insertanalternatechargedbattery.
Twoaudiblebeepsevery16seconds
YELLOWalarmindicatorilluminated
3.Ifnootherpowersourceisavailable,
Batterygauge:
•Nosegmentsilluminated
NOTES:
Toclearthealarm,theconcentratorneedstobeturnedoff.Completesolution1or2beforerestarting.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
UnitTooHot
INDICATORS:
Twoaudiblebeepsevery16seconds
Concentratoristoohottoallowitto
continueoperating.
Concentratorfunctionhasshutdown.
Internalfanwillcontinuetoruntohelp
lowerinternaltemperature.
1.Moveconcentratortocooler
2.UseACorDCadapterforpower.
3.Cleanintakelterscreen.
4.Changetoanothersourceofoxygen
YELLOWalarmindicatorilluminated
NOTES:
Troubleshooting
eithertheACorDCpower ,OR
Removethedepletedbatteryand
chargewiththeoptionalexternal
batterychargeraccessoryforfuture
use,OR
changetoanothersourceofoxygen.
surroundings.Allowconcentratorto
cooldownto104°F(40°C).
whilewaiting.
Toclearthealarm,turntheconcentratoroff.Completesolutions1,2,and/or3beforerestarting.
1194969-C61
Page 62

Invacare®Platinum®Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
UnitTooCold
INDICATORS:
Twoaudiblebeepsevery16seconds
Concentratoristoocoldtoallowitto
continueoperating.
Concentratorfunctionhasshutdown.
Internalfanwillcontinuetoruntohelp
increaseinternaltemperature.
YELLOWalarmindicatorilluminated
NOTES:
Toclearthealarm,turntheconcentratoroff.Completesolutions1,2,and/or3beforerestarting.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
BatteryTooHot
INDICATORS:
Batteryistoohottoallowthebatteryto
continueoperating.
Concentratorfunctionhasshutdown.
Twoaudiblebeepsevery16seconds
YELLOWalarmindicatorilluminated
NOTES:
Toclearthealarm,turntheconcentratoroff .Completesolutions1,2,3,or4beforerestarting.
1.Moveconcentratortowarmer
surroundings.Allowconcentratorto
warmupto41°F(5°C).
2.UseACorDCadapterforpower.
3.Changetoanothersourceofoxygen
whilewaiting.
1.Moveconcentratortocooler
surroundings.Allowconcentratorto
cooldowntolessthan104°F(40°C),
OR
2.UseACorDCpowerandremove
battery,OR
3.Insertanalternatechargedbattery
andremovethehotbattery,OR
4.Changetoanothersourceofoxygen
ifnootherpowersourceisavailable.
621194969-C
Page 63

LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
BatteryCold
INDICATORS:
Batteryistoocoldtoallowthebattery
tocontinueoperating.
Concentratorfunctionhasshutdown.
Twoaudiblebeepsevery16seconds
YELLOWalarmindicatorilluminated
NOTES:
Toclearthealarm,turntheconcentratoroff .Completesolutions1,2,3,or4beforerestarting.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
OxygenFault
INDICATORS:
Theoxygenoutputpurityhasfallento
avaluebelow73%.
Concentratorfunctionhasshutdown.
Twoaudiblebeepsevery16seconds
YELLOWalarmindicatorilluminated
NOTES:
Troubleshooting
1.Moveconcentratortowarmer
surroundings.Allowconcentratorto
warmupto41°F(5°C),OR
2.UseACorDCpowerandremove
battery,OR
3.Insertanalternatechargedbattery
andremovethecoldbattery,OR
4.Changetoanothersourceofoxygen
ifnootherpowersourceisavailable.
1.Verifyconcentratorisin
recommendedenvironmental
temperature.RefertoSpecications
inTechnicalData.
2.Cleanintakelterandensureboth
intakeandexhaustarenotblocked.
3.Changetoanothersourceofoxygen
ifalarmcontinues.
Toclearthealarm,turntheconcentratoroff.CompleteSolution1and/or2beforerestarting.Thealarmlimithasa
toleranceof+/-2%
Alarmsignalgenerationmaybedelayedbyupto128secondsduetothemonitoringalgorithmusedtopreventnuisancealarms.
Callyourproviderandreporttheproblemifalarmcontinues.
1194969-C63
Page 64

Invacare®Platinum®Mobile
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
ShutDown
SystemFault
INDICATORS:
Twoaudiblebeepsevery16seconds
Abnormalsystemconditionhasbeen
detected.
Maybecausedbymalfunctionof
compressor,fan,pressureorsoftware.
Concentratorfunctionhasshutdown.
YELLOWalarmindicatorilluminated
NOTES:
Toclearthealarm,turntheconcentratoroff .Restartconcentrator.
Callyourproviderandreporttheproblemifalarmcontinues.
LCDDISPLAYTEXT:DESCRIPTION:SOLUTIONS:
Notapplicable.Allpowerlost,display
isblank.
INDICATORS:
Allpowersourceshavebeen
lost/removed.
Concentratorfunctionhasshutdown.
Audiblebeepseverysecondfor
approximately30to60seconds
Noothercontrolpanelindicators
illuminated
NOTES:
Thisalarmoccurseveniftheconcentratorisoffwhenallpowerisremoved.
1.TurntheconcentratorOff,thenOn
againtoretry.
2.Changetoanothersourceofoxygen
ifalarmcontinues
1.Ensurethebatteryiscompletely
inserted.
2.Checkexternalpowerconnections
totheconcentratorandtothe
poweroutlet.
3.Tryanotherpoweroutlet.
4.Changetoanothersourceofoxygen
ifalarmcontinues.
Toclearthealarm,completesolutions1,2,or3beforerestarting.
641194969-C
Page 65

TechnicalData
8TechnicalData
8.1TechnicalDescription
TheInvacarePlatinumMobileusesamolecularsieveand
pressureswingadsorptionmethodologytoproducethe
oxygengasoutput.Ambientairentersthedevice,isltered
andthencompressed.Thiscompressedairisthendirected
intoanitrogenadsorbingsievebed.Concentratedoxygen
exitstheoppositeendofthesievebedandisdirectedinto
anoxygenreservoirfromwhichitisdeliveredtothepatient.
Theoxygenpurityleveloftheoutputgasrangesfrom87%
to95.6%.Theoxygenisdeliveredtothepatientthrough
theuseofanasalcannula.Apulsedosedeliverymethod
isused.Theconcentratordetectsthestartofpatient
inhalationanddeliversameasuredpulseofoxygen.No
furtheroxygenisdelivereduntilthenextpatientinhalation
isdetected.Thevolumeofoxygendeliveredeachminute
isaxedamountbasedontheselectedpulseowsetting.
Thevolumeofeachoxygenpulsewillvarywiththepatient’s
breathratesuchthatthexedminutevolumeismaintained.
TheInvacarePlatinumMobileiscapableofoperationbythe
patientinahome,institution,vehicle,orotherenvironments
outsidethehome.Devicestandardpoweroptionsinclude
anACtoDCswitchingpoweradapteroperatingfroman
ACpoweroutlet(100–240VAC,50–60Hertznominal),aDC
powercableoperatingfromanaccessoryoutlettypically
foundinavehicletypeenvironment(12VDCnominal),and
uptotworechargeablebatteries.
Nospecicproductknowledgeortrainingisrequiredto
operatetheproductotherthanwhatiscontainedinthis
manual.
Serviceinformationwillbeavailableonrequesttoqualied
technicalpersonnelONLY .
8.1.1PneumaticDiagram
•A=RoomAirIn
•B=ExhaustAirOut
•C=PatientOutletFitting
•D=NasalCannula
1194969-C65
Page 66

Invacare®Platinum®Mobile
8.2Specications
ElectricalRequirements:ACPowerSupply:110–240VAC,50–60Hz
DCPowerSupply:11–16VDC
RatedCurrentInput:
PowerConsumption:
(Typical)
5Aat19VDC,10Aat11–16VDC
PulseSetting:
P1=18W
P2=24W
P3=35W
P4–P5=45W
Dataisforconcentratoroperationonly(nobatterycharging)utilizinganACpowersource.
OperatingEnvironmental
Conditions:
(Allpowersources)
StorageandT ransport
OperatingTemperature:41°Fto104°F(5°Cto40°C)
RelativeHumidity:15–90%non-condensingrelativehumidity,watervaporpressuresupto1.48
inHg(50hPa)
-13°Fto140°F(-25°Cto60°C)
Temperatures:
StorageandT ransport
Humidity:
OperatingAltitude:
OperatingAtmospheric
Upto90%non-condensingrelativehumidityfortemperaturesof41°Fto95°F(5°Cto35°C)
Watervaporpressureupto1.48inHg(50hPa)fortemperaturesgreaterthan95°F(35°C)
Upto10,000ft(3048m)abovesealevel
697–1060hPa
Pressure:
661194969-C
Page 67

TechnicalData
OxygenPurity:
87%to95.6%,atallowsettingsandovertheratedrangesforambienttemperature,humidity
andatmosphericpressure.
Afterinitialwarm-upperiod(typicallylessthan5minutes)
ConserverTrigger
Sensitivity:
<0.18cmH
Opressuredrop(Forallcannulalengths)
2
Factoryset—noadjustment,pressureactivated
Onlypatientrespiratoryeffortsthatachievethetriggerpressurewillresultinthedeliveryof
anoxygenbolus.
ConserverBreathRate
15–40BPM(breathsperminute)withoutreductionofbolusminutevolume
Capacity:
MaximumOutlet
28.5psig(197kPa)
Pressure:
CannulaRequirements:
Length:4–25ft(1.2–7.6m)includingalloxygentubing
Tubing:crush-proof,singlelumen
Adult,standardow(ratedforupto6L/mincontinuousow)forlengthsupto7ft)
Adult,highow(ratedforupto15L/mincontinuousow)forlengthsgreaterthan7ftto25ft)
ExampleofPossibleCannulaModel:WestmedInc.PartNumber0194(4ftlength)
BatterySpecications
Rechargeablelithium-ion,14.4V,5800mAh,83.5Wh,500fullcharge/dischargecyclelife
(eachbattery):
BatteryShelfLife:12monthsfromdateofmanufacture
1194969-C67
Page 68

Invacare®Platinum®Mobile
BatteryDuration:
(Timesareapproximate)
BatteryChargeTime:
(Timesareapproximate)
SoundPressureLevel:
SoundPowerLevel:
AudibleSignalSound
PressureLevel:
Dimensions:
Weight:
(Nominal)
Condition
OneBattery
TwoBatteries
PulseSettingP15hr5min10hr10min
PulseSettingP23hr30min7hr0min
PulseSettingP32hr20min4hr40min
PulseSettingP4–P51hr45min3hr30min
Condition
OneBattery
TwoBatteries
ConcentratorOn,PulseSettingP12hr20min4hr40min
ConcentratorOn,PulseSettingP22hr20min4hr40min
ConcentratorOn,PulseSettingP32hr30min5hr0min
ConcentratorOn,PulseSettingP4–P53hr10min6hr20min
ConcentratorOff2hr20min4hr40min
<40dBAweightedforowsettingP2(TestedperISO3744:2010withmicrophonelocationas
speciedinISO8359:1996subclause4.6)
65dBAweightedforowsettingP4–P5(TestedperISO80601–2–69subclause201.9.6.2.1.101)
<
<65dBAweightedforowsettingP4–P5(T estedperISO80601–2–69subclause201.9.6.2.1.101)
55dBA+/-5dBA
9.4inhighx7.4inwidex3.7indeep(23.9cmhighx18.8cmwidex9.4cmdeep)
4.9lbs(2.18kg)withsinglebatteryandnocarrybag
•Add0.75lbs(0.34kg)forcarrybag
•Add1.0lbs(0.45kg)forasecondbattery
681194969-C
Page 69

TechnicalData
ShippingWeight:
10.5lbs(4.8kg)
(Nominal)
Classications:ClassIIElectricalShockProtection,TypeBFAppliedPart,ContinuousOperation
IngressProtection
Rating:
Concentrator—IP22
ACPowerAdapter—IP21
Battery—KeepDry
AppliedParts:
Cannula/OxygenTubing,OxygenOutletPort,CarryBag
DeliveredOxygenPulseVolumes:
•ThenominalpulsevolumespublishedinthefollowingtableisinmillilitersatSTPD(standardtemperatureandpressuredry)
conditionsandapplyovertheratedrangesforambienttemperature,humidity ,andatmosphericpressure.
•Maximumvariationfromnominal:+/-15%
4ftto25ftCannula/OxygenTubingLengths
BreathsPerMinute
152025303540
PulseSetting=P1
PulseSetting=P2
PulseSetting=P3
PulseSetting=P4
PulseSetting=P5
14.711.08.87.36.3
29.322.017.614.712.611.0
44.033.026.422.018.916.5
58.744.035.229.325.122.0
66.750.040.033.328.625.0
5.5
1194969-C69
Page 70

Invacare®Platinum®Mobile
8.3RegulatoryListing
ETLcertied
complying
with:
EN/IEC60601-1;Ed:3.1
EN/IEC60601-1-2;Ed:4
AAMIES60601–1(UnitedStates)
CSA22.2No.60601–1(Canada)
ISO80601–2–69
ISO80601–2–67
IEC60601–1–6
IEC60601–1–8
IEC60601–1–11
RTCADO160G
701194969-C
Page 71

8.4ElectromagneticCompliance(EMC)
Guidanceandmanufacturer´sdeclaration—electromagneticemission
TheDeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.ThecustomerortheuseroftheDevice
shouldassurethatitisusedinsuchanenvironment.
TechnicalData
Emissionstest
RFemissions
CISPR11
RFemissions
CISPR11
Harmonic
emissions
IEC61000-3-2
ComplianceElectromagneticenvironment-guidance
TheDeviceusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissions
GroupI
areverylowandarenotlikelytocauseanyinterferenceinnearbyelectronic
equipment.
ClassB
ClassA
TheDeviceissuitableforuseinallestablishmentsincludingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
Voltageuctuations
/ickeremissions
Complies
IEC61000-3-3
Guidanceandmanufacturer´sdeclaration—electromagneticimmunity
TheDeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.ThecustomerortheuseroftheDevice
shouldassurethatitisusedinsuchanenvironment.
Immunitytest
Electrostatic
discharge(ESD)
IEC61000-4-2
IEC60601testlevelCompliancelevelElectromagneticenvironment–guidance
±8kVcontact
±8kVcontact
Floorsshouldbewood,concreteorceramictile.
Ifoorsarecoveredwithsyntheticmaterial,the
±15kVair
±15kVair
relativehumidityshouldbeatleast30%.
1194969-C
71
Page 72

Invacare®Platinum®Mobile
ElectricalFast
transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltagedips
IEC61000-4-11
Voltage
Interruptions
IEC61000-4-11
Powerfrequency
(50/60Hz)
magneticeld
IEC61000-4-8
±2kVforpowersupply
lines
±1kVforinput/output
lines
±1kVline(s)toline(s)
±2kVline(s)toground
0%UT;1cycle
and
70%UT;25/30cycles
Singlephase:at0°
±2kVforpowersupply
lines
±1kVforinput/output
lines
±1kVline(s)toline(s)
±2kVline(s)toground
0%UT;1cycle
and
70%UT;25/30cycles
Singlephase:at0°
0%UT;250/300cycle0%UT;250/300cycle
30A/m30A/m
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.Iftheuser
oftheDevicerequirescontinuedoperationduring
powermainsinterruptions,itisrecommendedthat
theDevicebepoweredfromanun-interruptible
powersupplyorabattery.
UTisthea.c.mainsvoltagepriortoapplication
ofthetestlevel.
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.Iftheuser
oftheDevicerequirescontinuedoperationduring
powermainsinterruptions,itisrecommendedthat
theDevicebepoweredfromanun-interruptible
powersupplyorabattery.
UTisthea.c.mainsvoltagepriortoapplication
ofthetestlevel.
Powerfrequencymagneticeldsshouldbeat
levelscharacteristicofatypicallocationinatypical
commercialorhospitalenvironment.
72
1194969-C
Page 73

ConductedRF
IEC61000-4-6
RadiatedRF
IEC61000-4-3
3V3V
10V/m10V /m
TechnicalData
PortableandmobileRFcommunicationsequipment
shouldbeusednoclosertoanypartofthe
Deviceincludingcables,thantherecommended
separationdistancecalculatedfromtheequation
applicabletothefrequencyofthetransmitter .
Recommendedseparationdistance:
d=1.2√P150kHzto80MHz
d=0.35√P
d=0.7√P
80MHzto800MHz
800MHzto2,5GHz
wherePisthemaximumoutputpowerrating
ofthetransmitterinwatts(W)accordingto
thetransmittermanufactureranddisthe
recommendedseparationdistanceinmetres(m).
FieldstrengthsfromxedRFtransmitters,as
determinedbyanelectromagneticsitesurvey,
shouldbelessthanthecompliancelevelineach
frequencyrange.
b
Interferencemayoccurinthevicinityofequipment
markedwiththefollowingsymbol:
a
Fieldstrengthsfromxedtransmitters,suchasbasestationsforradio(cellular/cordless)telephonesandlandmobileradios,
amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywithaccuracy.T oassess
1194969-C73
a
Page 74

Invacare®Platinum®Mobile
theelectromagneticenvironmentduetoxedRFtransmitters,anelectromagneticsitesurveyshouldbeconsidered.Ifthe
measuredeldstrengthinthelocationinwhichtheDeviceisusedexceedstheapplicableRFcompliancelevelabove,theDevice
shouldbeobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary,
suchasreorientingorrelocatingtheDevice.
b
Overthefrequencyrange150kHzto80MHz,eldstrengthsshouldbelessthan[V1]V/m.
At80MHzand800MHz,thehigherfrequencyrangeapplies.
RecommendedseparationdistancesbetweenportableandmobileRFcommunicationsequipmentandtheDevice
TheDeviceisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.The
customerortheuseroftheDevicecanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetween
portableandmobileRFcommunicationsequipment(transmitters)andtheDeviceasrecommendedbelow,accordingtothe
maximumoutputpowerofthecommunicationsequipment
Separationdistanceaccordingtofrequencyoftransmitter[m]
Ratedmaximumoutput
oftransmitter[W]
150kHzto80MHz
d=1.2√P
80MHzto800MHz
d=0.35√P
800MHzto2,5GHz
d=0.7√P
0.010.120.040.07
0.10.370.110.22
11.170.350.70
103.691.112.21
10011.673.507.00
74
1194969-C
Page 75

TechnicalData
Fortransmittersratedatamaximumoutputpowernotlistedabovetherecommendedseparation,distancedinmeters(m)
canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePisthemaximumoutputpower
ratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer .
At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.
Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreection
fromstructures,objectsandpeople.
1194969-C75
Page 76

Invacare®Platinum®Mobile
9Warranty
9.1LimitedWarranty
Refertothewarrantyinformationincludedwithyour
productforwarrantydetails.
761194969-C
Page 77

Notes
Page 78

Notes
Page 79

Notes
Page 80

InvacareCorporation
USA
Manufacturer
OneInvacareWay
Elyria,OhioUSA
44035
440–329–6000
800–333–6900
TechnicalServices
440-329-6593
800-832-4707
www.invacare.com
1194969-C2018-10-09
*1194969C*
InvacareCorporation
2101E.LakeMaryBlvd.
Sanford,FLUSA32773
407–321–5630
MakingLife’sExperiencesPossible®
Page 81

Invacare®Platinum®Mobile
POC1-100B,POC1-100C
enOxygenConcentrator
UserManual
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,thismanualMUSTbereadandsavedfor
futurereference.
Page 82

©2018InvacareCorporation
Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibited
withoutpriorwrittenpermissionfromInvacare.T rademarksareidentiedby™and®.Alltrademarks
areownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
PlatinumisaregisteredtrademarkintheU.S.A.,EuropeandAustralia.
MakingLife’sExperiencesPossibleisaregisteredtrademarkintheU.S.A.
Pine-sol,LestoilandCloroxareregisteredtrademarksofTheCloroxCompany.
DawnisaregisteredtrademarkofTheProcterandGambleCompany.
IOSisatrademarkorregisteredtrademarkofCiscointheU.S.andothercountriesandisused
underlicensetoApple,Inc.
AndroidisatrademarkofGoogle,LLC.
BluetoothisaregisteredtrademarkofBluetoothSIG,Inc.
Page 83

Contents
1General.........................................5
1.1QuickStartGuide..............................5
1.2Symbols.....................................9
1.3IndicationsForUse.............................11
1.4IntendedUse.................................11
1.5Contraindications...............................11
1.6Features.....................................12
1.6.1ControlPanel...............................12
1.6.2Input/OutputConnections.....................13
1.6.3Filters....................................13
1.6.4PowerOptions..............................14
1.6.5CarryBag..................................14
1.7OptionalAccessoriesandReplacementParts..........14
2Safety..........................................15
2.1LabelLocations................................15
2.2GeneralGuidelines.............................17
2.3RadioFrequencyInterference.....................22
2.4TravelingwiththeConcentrator....................22
3Setup...........................................25
3.1Unpacking....................................25
3.2Inspection....................................25
3.3PoweringtheOxygenConcentrator.................25
3.3.1InstallingtheBattery(s).......................26
3.3.2ChargingtheBattery(s)........................27
3.3.3ReadingtheBatteryGaugeforInstalledand
UninstalledBattery(s)........................28
3.3.4RemovingtheBattery(s).......................29
3.3.5ExternalACPower...........................30
3.3.6ExternalDCPower...........................30
3.4CarryBagSetup...............................31
3.4.1InstallingtheCarryBag.......................31
3.4.2RemovingtheCarryBag.......................32
3.4.3InstallingShoulderStraps......................32
3.5Connecting/PositioningtheNasalCannula............33
4Usage..........................................35
4.1Location.....................................35
4.2ConcentratorOperatingPosition...................35
4.3UsingtheCarryBag............................35
4.4TurningOntheConcentrator......................36
4.5TurningOfftheConcentrator......................37
4.6WarmUpPeriod...............................37
4.7BreathingwiththeUseoftheConcentrator...........37
4.8ReadingtheDisplayScreen.......................37
4.9AdjustingthePulseFlowSetting...................38
4.10BatteryLifeandManagement....................39
5Maintenance.....................................41
5.1Maintenance..................................41
5.2ServiceLife...................................41
5.3CleaningtheCabinet............................41
5.4CleaningtheAirIntakeFilterScreen................42
5.5CleaningtheCarryBag..........................42
5.6CleaningandDisinfectionBetweenPatients...........43
5.7ViewingHourMeter............................44
5.8PreventativeMaintenanceChecklist.................45
6AfterUse........................................47
6.1Storage......................................47
6.2Disposal.....................................47
6.3WearandT ear................................48
7Troubleshooting...................................49
7.1Troubleshooting...............................49
7.2AlarmConditions...............................51
7.2.1OperationalAlarms..........................52
7.2.2ShutDownAlarms...........................60
8T echnicalData....................................65
8.1TechnicalDescription............................65
8.1.1PneumaticDiagram..........................65
Page 84

8.2Specications.................................66
8.3RegulatoryListing..............................70
8.4ElectromagneticCompliance(EMC).................71
9Warranty........................................76
9.1LimitedWarranty..............................76
Page 85

General
1General
1.1QuickStartGuide
IMPORTANT!
–TheQuickStartGuideisforreferenceONLY .
Itisimperativetoreadthemanualinits
ENTIRETYforfullinstructions.
DANGER!
RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,death
orinjurytopersons:
–DONOTdisassemble.Referservicingto
qualiedservicepersonnel.Therearenouser
serviceableparts.
WARNING!
RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduse
andspecicationshasnotbeentestedandmay
leadtoproductdamage,lossofproductfunction,
orinjury.
–DONOTusethisproductinanywayotherthan
describedinthespecicationsandintended
usesectionsofthismanual.
DANGER!
RiskofDeath,InjuryOrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,
greasysubstancesandothercombustiblesare
easilyignitedandburnwithgreatintensityin
oxygenenrichedairandwhenincontactwith
oxygenunderpressure.Smokingduringoxygen
therapyisdangerousandislikelytoresultin
burnsordeath.T oavoidre,death,injuryor
damage:
–DONOTSMOKEwhileusingthisdevice.DO
NOTusenearOPENFLAMEorIGNITION
SOURCES.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic
electricitycreatedbyanytypeoffriction.
–Useonlyoxygencompatiblewater-basedlotions
orsalvesbeforeandduringoxygentherapy.
Toverify,refertothelotion/salvecontainer
foroxygencompatiblewater-basedstatement.
Ifnecessary,contactthemanufacturer.DO
NOTuseanylubricantsonconcentratorunless
recommendedbyInvacare.
–Keeptheoxygentubing,cord,ACadapter,
andconcentratoroutfromundersuchitems
asblankets,bedcoverings,chaircushions,
clothing,andawayfromheatedorhotsurfaces
includingspaceheaters,stoves,andsimilar
electricalappliances.
–Makesureconcentratorisoffwhennotinuse.
1194969-C5
Page 86

Invacare®Platinum®Mobile
DANGER!
RiskOfDeath,InjuryOrDamage
Improperuseoftheproductmaycausedeath,
injuryordamage.Thissectioncontainsimportant
informationforthesafeoperationanduseofthis
product.
–DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely
readingandunderstandingtheseinstructions
andanyadditionalinstructionalmaterialsuch
asusermanuals,servicemanualsorinstruction
sheetssuppliedwiththisproductoroptional
equipment.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare
professional,dealerortechnicalpersonnel
beforeattemptingtousethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct
isnotworkingcorrectly,contactatechnician
orInvacareforrepair .
–Thisproductisintendedtobeusedbyadults
onlyafterreadingandunderstandingthe
instructionsandwarningsofthisusermanual.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
61194969-C
Page 87

General
1.
Openthebatterycompartmentdoor.Withbattery
fuelgaugelabelfacingtowardthefrontoftheoxygen
concentrator ,slowlylowerbatteryAstraightdowninto
oneofthetwoslotsinthebatterycompartment.
RefertoInstallingtheBattery(s)inSetup.
2.
ConnecttheACPowerAdaptertotheexternalpower
inputconnectorB.PlugtheACcordintotheelectrical
outlet.
RefertoExternalACPowerinSetup.
Usingtheconcentratorforthersttimerequires
thebattery(s)tobecharged.RefertoCharging
theBattery(s)inSetup.
3.
Connectthenasalcannulatotheconcentratoroxygen
outletportonyourconcentratorC.Oxygenwillonly
bedeliveredwhileyouarebreathingthroughthenasal
cannulaconnectedtothisdevice.
RefertoConnecting/PositioningtheNasalCannulain
Setup.
1194969-C
7
Page 88

Invacare®Platinum®Mobile
4.
TurnontheconcentratorbypressingtheOn/OffButton
Duntilthecontrolpanelindicatorslightup.
RefertoTurningontheConcentratorinUsage.
5.
Adjustthepulseowsettingtotheoneprescribedby
yourclinicalprofessional.Thepulseowsettingcanbe
changedbypressingtheowselectionbuttonE.Place
thenasalcannulaoveryourearsandpositiontheprongs
inyournoseasinstructedbyyourhealthcareprovider.
RefertoAdjustingthePulseFlowSettinginUsage.
81194969-C
Page 89

General
1.2Symbols
Signalwordsareusedinthismanualandapplytohazards
orunsafepracticeswhichcouldresultinpersonalinjury
orpropertydamage.Refertotheinformationbelowfor
denitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardous
situationwhich,ifnotavoided,willresultin
deathorseriousinjury .
WARNING!
–Warningindicatesapotentiallyhazardous
situationwhich,ifnotavoided,couldresultin
deathorseriousinjury .
CAUTION!
–Cautionindicatesapotentiallyhazardous
situationwhich,ifnotavoided,mayresultin
propertydamageorminorinjuryorboth.
IMPORTANT!
–Importantindicatesahazardoussituationthat
couldresultindamagetopropertyifitisnot
avoided.
Givesusefultips,recommendationsandinformation
forefcient,trouble-freeuse.
Refertothefollowingtableforthemeaningofsymbols
markedontheequipmentand/orpackaging.
GeneralWarningSign
Thebackgroundcolorinsidethe
triangleisyellowonproductlabels.
FollowInstructionsforUse
Thecolorofthesymbolbackgroundis
blueonproductlabels.
Noopename
Thecolorofthecirclewithdiagonal
barisredonproductlabels.
DONOTSmoke
Thecolorofthecirclewithdiagonal
barisredonproductlabels.
ConsultInstructionsforUse
Thissymbolislocatedonthecannula
packing.
DirectCurrent
TypeBFAppliedPart
1194969-C9
Page 90

Invacare®Platinum®Mobile
ClassIIEquipment
PowerOn/Off
DataOutputTerminal
DONOTdisposeofinhouseholdwaste
Manufacturer
DateofManufacture
Recycle
ReferenceNumber
SerialNumber
DONOTReuse(SinglePatientUseOnly)
Thissymbolreferstothecannula
whichisforsinglepatientuseonly.
IndoorUseONL Y
ThissymbolisincludedontheAC
poweradapter.
DONOTuseoilorgrease
Keepdry
Protectedagainstsolidforeignobjectsof12.5
mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops
whenenclosuretiltedupto15°.
Protectedagainstsolidforeignobjectsof12.5
mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops.
TransportandStorageT emperature
TransportandStorageHumidity
ElectricalSafetyAgencyCerticationMark
LatexFree
101194969-C
Page 91

General
1.3IndicationsForUse
TheInvacare®Platinum®MobileOxygenConcentratoris
intendedtoprovidesupplementaloxygentopatientswith
respiratorydisorders.Thedevicecanbeusedinahome,
institution,vehicle,orotherenvironmentsoutsidethehome.
TheInvacareMobileMedicalApplicationAccessory
POC1-CONNECTisintendedforusewiththeInvacare
PlatinumMobileOxygenConcentratorPOC1-100Bor
POC1-100Cdevice,toallowpatientsviatheirAndroid™or
iOS™mobilephoneortablettodisplaydevicesettings,and
tocollectdeviceperformanceandusageinformationfor
maintenance/servicingpurposesonly.
1.4IntendedUse
TheInvacare®Platinum®MobileOxygenConcentratoris
intendedtoprovidesupplementaloxygentopatientswith
respiratorydisorders.TheInvacare®Platinum®Mobile
OxygenConcentratorcanbeusedinahome,institution,
vehicle,orotherenvironmentsoutsidethehome.The
deviceisnotintendedtobelife-supportingorlife-sustaining.
TheInvacareMobileMedicalApplicationAccessory
POC1-CONNECTisintendedtoallowpatientsviatheir
AndroidoriOSmobilephoneortablettodisplaydevice
settings,andtocollectdeviceperformanceandusage
informationformaintenance/servicingpurposesonly.
WARNING!
RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduse
andspecicationshasnotbeentestedandmay
leadtoproductdamage,lossofproductfunction,
orpersonalinjury.
–DONOTusethisproductinanywayotherthan
describedinthespecicationsandintended
usesectionsofthismanual.
1.5Contraindications
WARNING!
RiskofInjury
Thisproductistobeusedasanoxygen
supplementandisnotintendedtobe
life-supportingorlife-sustaining.ONL Yusethis
productifthepatientiscapableofspontaneous
breath,abletoinhaleandexhalewithouttheuse
ofamachine
–DONOTuseinparallelorserieswithother
oxygenconcentratorsoroxygentherapy
devices.
1194969-C
11
Page 92

Invacare®Platinum®Mobile
WARNING!
RiskofMinorInjuryorDiscomfort
Theconserving,orpulsedose,oxygendelivery
techniqueusedbythisdeviceiscontraindicated
inpersonswhosebreathingduringnormalresting
wouldbeunabletotriggerthedevice.
–Properdevicetriggering,setupandoperation
–Notforpediatricuse.
–Notforusebytracheotomizedpatients.
1.6Features
1.6.1ControlPanel
mustbeconrmedbyanexperiencedclinician
orotherrespiratoryprofessional.
On/OffButton
A
Pressforonesecondtoturnconcentratoronoroff.
FlowSelectionButton
Presstoadjustthepulsedoxygenowtoyour
B
prescribedsetting.Theselectedowsettingis
shownonthedisplayscreen.RefertoAdjustingthe
PulseFlowSettinginUsageformoredetails.
DisplayBacklightButton
C
Pressoncetoilluminatethedisplayscreenfor5
seconds.
AudioOffButton
Presstomutetheaudiblesignal(beep)forboth
alarmsandstatusindications.Theblueindicator
D
lightnexttothebuttonilluminateswhenactivated.
Audiooffcanbeactivatedduringorpriortoan
alarmconditionoccurring.
DisplayScreen
E
Displaysinformationabouttheoperatingstatusof
theconcentrator .
AlarmIndicator
Theyellowindicatorlightatthecenterofthe
F
triangularsymbolilluminatestoindicateabnormal
operatingconditions.RefertoAlarmConditionsin
Troubleshootingformoredetails.
12
1194969-C
Page 93

General
BatteryGauge
Indicatesthechargestatusofthebattery(s).Refer
G
toReadingtheBatteryGaugeforInstalledand
UninstalledBattery(s)inSetupformoredetails.
ExternalPowerIndicator
Theorangeindicatorlightnexttotheplugsymbol
H
illuminateswhenanexternalpowersourceis
connectedtotheconcentrator .
BatteryCompartmentDoor
I
Usedtoaccesstheremovablebattery(s).
AudibleSignal(Beep)
Not
Shown
Indicatesachangeinoperatingstatusor
aconditionwhichrequirestheoperator’s
attention.
1.6.2Input/OutputConnections
A
B
C
OxygenOutletPort
Usedtoconnectthenasalcannulatothe
concentrator .
ExternalPowerInputConnector
Usedtoconnectanexternalpowersourcetothe
concentrator .
USBPort
Usedforservice.LocationofPOC1–CONNECT
Dongle.
1.6.3Filters
Airenterstheconcentratorthroughanairintakelter
locatedonthecarrybag.Thislterpreventshairandother
largeparticlesintheairfromenteringtheunit.Beforeyou
1194969-C13
Page 94

Invacare®Platinum®Mobile
operatetheconcentrator,makesurethislteriscleanand
dry,andtheconcentratorisproperlyinstalledinthecarry
bag.
Tocleantheairintakelter ,refertoCleaningtheAirIntake
FilterScreeninMaintenance.
Toensuretheconcentratorisproperlyinstalledinthecarry
bag,refertoInstallingtheCarryBaginSetup.
1.6.4PowerOptions
WARNING!
RiskofInjuryorDamage
Toavoidinjuryordamagewhichwillvoid
warranty:
–UseonlyInvacarespeciedpowersupplies.
Battery(s):Uptotworechargeablebatteriescanbeinstalled
inthebatterycompartmentoftheconcentrator .Whenfully
charged,asinglebatterysuppliespowerforuptovehours
(uptotenhoursfortwobatteries).Visibleandaudible
alarmsignalsoccurwhenthebatterypowerisgettinglow.
RefertoAlarmConditionsinTroubleshootingandCharging
theBattery(s)inSetup.
ACPowerAdapter:AnACpoweradapterallowsthe
concentratortobeconnectedtoa100–240volt50–60
hertzoutlet.UseoftheACpoweradapterwillallowthe
concentratortooperateandsimultaneouslyrechargethe
installedbattery.RefertoChargingtheBattery(s)inSetup.
theconcentratortooperateandsimultaneouslyrechargean
installedbattery(s).RefertoChargingtheBattery(s)inSetup.
1.6.5CarryBag
Thecarrybagprovidesaconvenientmeanstoholdthe
concentratorwhileperformingdailyactivities.
Itcanbeconguredasamessengerbagorbackpackusing
theshoulderstrapsprovided.RefertoInstallingtheCarry
BaginSetup.
1.7OptionalAccessoriesandReplacementParts
Thefollowingoptionalaccessoriesareavailablefrom
Invacare:
•Additionalbatteries,modelnumber:POC1–110
•Externalbatterychargerwithpowersupply:POC1–115
ThefollowingreplacementpartsareavailablefromInvacare:
•Battery,modelnumber:POC1–110
•CarryBag,modelnumber:POC1–150
•DCpowercable,modelnumber:POC1–140
•ACpoweradapterwithpowercord:POC1–130
•USBDongle,modelnumber:POC1–CONNECT
Thesepartsarebasemodels.PleasecontactInvacare
oryourproviderforcountryspecicmodels.
DCPowerCable:ADCpowercableallowstheconcentrator
tobeconnectedtoanautomobile’s(boat,motorhome,
etc.)12-voltDCoutlet.UseoftheDCpowercablewillallow
14
1194969-C
Page 95

2Safety
2.1LabelLocations
Thelabelislocatedonthebackoftheconcentrator.Theserialnumberislocatedontherightsideoflabelshownhere.
ModelPOC1-100B
Theconcentratorhasfourpulseowsettings.
Safety
1194969-C15
Page 96

Invacare®Platinum®Mobile
ModelPOC1-100C
Theconcentratorhasvepulseowsettings.
161194969-C
Page 97

Safety
2.2GeneralGuidelines
Inordertoensurethesafeinstallation,assemblyand
operationoftheconcentratortheseinstructionsMUSTbe
followed.
DANGER!
RiskOfDeath,InjuryOrDamage
Improperuseoftheproductmaycausedeath,
injuryordamage.Thissectioncontainsimportant
informationforthesafeoperationanduseofthis
product.
–DONOTusethisproductoranyavailable
optionalequipmentwithoutrstcompletely
readingandunderstandingtheseinstructions
andanyadditionalinstructionalmaterialsuch
asusermanuals,servicemanualsorinstruction
sheetssuppliedwiththisproductoroptional
equipment.
–Ifyouareunabletounderstandthewarnings,
cautionsorinstructions,contactahealthcare
professional,dealerortechnicalpersonnel
beforeattemptingtousethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproduct
isnotworkingcorrectly,contactatechnician
orInvacareforrepair .
–Thisproductisintendedtobeusedbyadults
onlyafterreadingandunderstandingthe
instructionsandwarningsofthisusermanual.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER!
RiskofDeath,InjuryOrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,
greasysubstancesandothercombustiblesare
easilyignitedandburnwithgreatintensityin
oxygenenrichedairandwhenincontactwith
oxygenunderpressure.Smokingduringoxygen
therapyisdangerousandislikelytoresultin
burnsordeath.T oavoidre,death,injuryor
damage:
–DONOTSMOKEwhileusingthisdevice.DO
NOTusenearOPENFLAMEorIGNITION
SOURCES.
–NOSMOKINGsignsshouldbeprominently
displayed.
–Keepallopenames,matches,lighted
cigarettes,electroniccigarettesorothersources
ofignitionatleast10ft(3m)awayfromthis
concentratororanyoxygencarryingaccessories
suchascannulasortanks.
–Ifyoudisregardthesewarningsaboutthe
severehazardofoxygenusewhileyoucontinue
tosmoke,youmustalwaysturnoffthe
concentrator ,removethecannulaandthen
waittenminutesbeforesmokingorleavethe
roomwhereeithertheconcentratororany
oxygencarryingaccessoriessuchascannulas
ortanksarelocated.
–DONOTSMOKEwhileusingtheconcentrator .
1194969-C
17
Page 98

Invacare®Platinum®Mobile
CAUTION!
Federal(statutory)lawrestrictsthisdeviceto
salebyorontheorderofamedicalpractitioner
licensedbyagovernmentalagencywherehe/she
practices.
–ONL Yalicensedmedicalpractitionermayorder
thepurchaseoruseofthisdevice.
DANGER!
RiskofDeath,InjuryOrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,
greasysubstancesandothercombustiblesare
easilyignitedandburnwithgreatintensityin
oxygenenrichedairandwhenincontactwith
oxygenunderpressure.Toavoidre,death,
injuryordamage:
–Avoidcreationofanysparknearoxygen
equipment.Thisincludessparksfromstatic
electricitycreatedbyanytypeoffriction.
–Useonlyoxygencompatiblewater-basedlotions
orsalvesbeforeandduringoxygentherapy.
Toverify,refertothelotion/salvecontainer
foroxygencompatiblewater-basedstatement.
Ifnecessary,contactthemanufacturer.DO
NOTuseanylubricantsonconcentratorunless
recommendedbyInvacare.
–Keeptheoxygentubing,cord,ACadapter,
andconcentratoroutfromundersuchitems
asblankets,bedcoverings,chaircushions,
clothing,andawayfromheatedorhotsurfaces
includingspaceheaters,stoves,andsimilar
electricalappliances.
–Makesureconcentratorisoffwhennotinuse.
181194969-C
Page 99

Safety
CAUTION!
RiskofMinorInjuryorDiscomfort
Itisimportanttoplanaheadfortravelandother
situationsinwhichyoumaynothaveaccessto
additionaloxygenorpowersupplies.T oprevent
oxygendeprivation:
–Carrysupplementalbatterieswithyou.
–Stowtheconcentratorproperlyduringtravel.
Forproperstorageinstructionswhentravelling
refertothetravelguide.
WARNING!
RiskofInjuryorDamage
Topreventinjuryordamagefrommisuse:
–NEVERleaveconcentratorunattendedwhen
connectedtopower.
–Makesureconcentratorisoffwhennotinuse.
–OutdooruseMUSTbeconductedwithinternal
batterypoweronly.
WARNING!
RiskOfInjuryOrDamage
–Invacareproductsarespecicallydesigned
andmanufacturedforuseinconjunction
withInvacareprovidedorInvacarespecied
accessories.Anyotheraccessorieshave
notbeentestedbyInvacareandarenot
recommendedforusewithInvacareproducts.
–Nomodicationofthisequipmentisallowed.
CAUTION!
RiskofMinorInjuryorDiscomfort
Achangeinaltitudemayaffecttotaloxygen
availabletoyou.Topreventoxygendeprivation:
–Consultyourphysicianbeforetravelingto
higherorloweraltitudestodetermineifyour
owsettingsshouldbechanged.
DANGER!
RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,death
orinjurytopersons:
–DONOTdisassemble.Referservicingto
qualiedservicepersonnel.Therearenouser
serviceableparts.
–Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the
concentratormustbelocatedinanotherroom
atleast7ft(2.5m)fromthebath.
–DONOTcomeincontactwiththeconcentrator
whilewet.
–DONOTplaceorstoreconcentratorwhereit
candropintowaterorotherliquid.
–DONOTreachforconcentratorthathasfallen
intowater .UnplugIMMEDIATELY.
–DONOTblockaccesstopoweroutletneeded
tounplugtheACpowercord.
–DONOTusefrayedordamagedpowercords.
–DONOTuseACpoweradapterifitshousingis
crackedorseparated.
1194969-C19
Page 100

Invacare®Platinum®Mobile
WARNING!
RiskofInjuryorDeath
Topreventinjuryordeathfromproductmisuse:
–Closelysupervisewhenthisconcentratoris
usedbyimpairedindividualsornearchildren
and/orimpairedindividuals.
–Monitorpatientsusingthisdevicewhoare
unabletohearorseealarmsorcommunicate
discomfort.
CAUTION!
RiskofMinorInjuryorDiscomfort
WhileInvacarestrivestoproducethebestoxygen
concentratorinthemarkettoday,thisoxygen
concentratorcanfailtoproduceoxygendueto
powerfailureordevicemalfunction.
–ALWA YShaveabackupsourceofoxygenreadily
available.
–Intheeventtheconcentratorfailstoproduce
oxygen,theconcentratorwillbrieyalarm
signalingthepatienttoswitchtotheirbackup
sourceofoxygen.RefertoT roubleshootingfor
moredetail.
–Invacarerecommendskeepingatleastone
batteryinstalledintheconcentratorevenwhen
operatingfromanexternalpowersource.
WARNING!
RiskofDeath,InjuryorDamage
Topreventinjuryordamagefromcordmisuse:
–DONOTmoveorrelocateconcentratorby
pullingonthecord.
–DONOTuseextensioncordswithACpower
adaptersprovided.
–Properlystoreandpositionelectricalcords
and/ortubingtopreventatrippingand
strangulationhazards.
WARNING!
RiskofInjuryorDamage
Topreventinjuryordamageduringuse:
–Ifyoufeelilloruncomfortable,orifthe
concentratordoesnotsignalanoxygenpulse
andyouareunabletohearand/orfeelthe
oxygenpulse,consultyourequipmentprovider
and/oryourphysicianIMMEDIA TEL Y .
–Foroptimumperformance,Invacare
recommendsthateachuseoftheconcentrator
beaminimumof30minutes.Shorterperiods
ofoperationmayreducemaximumproductlife.
–Theconcentratorshouldbeusedinanupright
position.
–Theconcentratorcannotbeusedinconjunction
withPAP ,Bi-Level,mechanicalventilatoror
othersuchdevices.
201194969-C