Invacare Platinum XL, Platinum 5, Platinum 10 User manual

Operator’s Manual
Platinum™Series
XL, 5, 10
Oxygen Concentrators Standard,
with
SensO
PART ONE
®
and HomeFill™II
2
Concentrator Information
Invacare 5
Model IRC5LXO2
Invacare 10
Model IRC10LXO2
Dealer: This manual MUST be given to the end user. User:
Before using this product, read this manual
and save for future reference.
For more information regarding
Invacare products, parts, and services,
with
SensO
with
SensO
2
2
Invacare 5
Model IRC5LX
Invacare 10
Model IRC10LX
WARNING
WARNING
DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT INVACARE TECHNICAL SERVICES BEFORE ATTEMPTING TO USE THIS EQUIPMENT ­OTHERWISE SERIOUS INJURY OR PROPERTY DAMAGE MAY RESULT.
ACCESSORIES
There are many different types of humidifiers, oxygen tubing, cannulas and masks that can be used with this device. You should contact your local home care provider for recommendations on which of these devices will be best for you. They should also give you advice on the proper usage, maintenance, and cleaning.
WARNING
Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.
Platinum™Series 2 Part No 1118353
TABLE OF CONTENTS
ACCESSORIES .................................................. 2
SPECIAL NOTES ............................................... 4
SECTION 1—GENERAL GUIDELINES .................. 5
SECTION 2—FEATURES .................................... 9
SECTION 3—HANDLING ................................. 10
Unpacking............................................................................................10
Inspection............................................................................................10
Storage.................................................................................................10
SECTION 4—SPECIFICATIONS ......................... 11
SECTION 5—OPERATING INSTRUCTIONS ....... 21
Introduction........................................................................................21
Select a Location ...............................................................................21
Recommended Guidelines for Optimum Performance............22
Plug in Power Cord ..........................................................................22
Connect Humidifier (If So Prescribed).........................................22
Power Switch .....................................................................................25
Flowrate...............................................................................................26
SensO2 Oxygen Purity Indicator (If Your Unit Features
the O2 Sensor Feature)....................................................................27
Initial Startup of the Concentrator ...............................................27
Explanation of Oxygen Purity Indicator Lights........................... 27
Indicators (If Your Unit Does Not Feature the O
Sensor) ... 28
2
SECTION 6—MAINTENANCE .......................... 29
Routine Maintenance........................................................................29
SECTION 7—TROUBLESHOOTING GUIDE ......... 32
LIMITED WARRANTY .................................... 35
Part No 1118353 3 Platinum™Series
SPECIAL NOTES
SPECIAL NOTES
SIGNAL WORD MEANING
Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious
DANGER
WARNING
CAUTION
The information contained in this document is subject to change without notice.
Users MUST not smoke while using this device. Keep all matches, lighted cigarettes or other sources of ignition out of the room in which this product is located. No smoking signs should be prominently displayed. Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result in severe fire, property damage and cause physical injury or death.
injury.
Warning indicates a potentially hazardous situation which, if not avoided, could result in death or seri­ous injury.
Caution indicates a potentially hazardous situation which, if not avoided, may result in property dam­age.
NOTICE
DANGER
Platinum™Series 4 Part No 1118353
SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
In order to ensure the safe installation, assembly and operation of the Platinum concentrator these instructions MUST be followed.
CAUTION
“Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.”
WARNING
SECTION 1 - GENERAL GUIDELINES contains important information for the safe operation and use of this product. DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as Owner’s Manuals, Service Manuals or Instruction Sheets supplied with this product or optional equipment. If you are unable to understand the Warnings, Cautions or Instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.
Risk of electric shock. DO NOT disassemble. Refer servicing to qualified service personnel.
The use of oxygen therapy requires that special care be taken to reduce the risk of fire. Any materials that will burn in air, and some that will not, are easily ignited and burn rapidly in high concentrations of oxygen. For safety concerns, it is necessary that all sources of ignition be kept away from the product and preferably out of the room in which it is being used. NO SMOKING signs should be prominently displayed.
Part No 1118353 5 Platinum™Series
SECTION 1—GENERAL GUIDELINES
A spontaneous and violent ignition may occur if oil, grease or greasy substances come in contact with oxygen under pressure. These substances MUST be kept away from the oxygen concentrator, tubing and connections, and all other oxygen equipment. DO NOT use any lubricants unless recommended by Invacare.
For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life.
If the concentrator has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water, call qualified technician for examination and repair.
Keep the cord away from heated or hot surfaces.
DO NOT move or relocate concentrator by pulling on the cord.
NEVER drop or insert any object into any opening.
NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air opening may be blocked. Keep the openings free from lint, hair and the like. Fill humidifier with water to the level shown by the manufacturer. DO NOT overfill.
Keep unit at least three (3) inches away from walls, draperies, furniture, and the like.
Invacare recommends that Crush-Proof oxygen tubing be used with this product and not exceed 50 ft. (15.2m) in length. Recommended use for Platinum 10 is up to 25 ft. in length of high flow tubing.
Platinum™Series 6 Part No 1118353
SECTION 1—GENERAL GUIDELINES
MAINTENANCE
The Invacare Oxygen Concentrator was specifically designed to minimize routine preventive maintenance. Only professionals of the healthcare field or persons fully conversant with this process such as factory trained personnel should perform preventive maintenance or performance adjustments on the oxygen concentrator.
RADIO FREQUENCY INTERFERENCE
Most electronic equipment is influenced by Radio Frequency Interference (RFI). Caution should be exercised with regard to the use of portable communications equipment in the area around such equipment.
TO REDUCE THE RISK OF BURNS, ELECTROCUTION, FIRE OR INJURY TO PERSONS.
Avoid using while bathing. If continuous usage is required by the physician’s prescription:
The concentrator MUST be located in another room at least seven (7) feet from the bath.
DO NOT come in contact with the concentrator while wet.
DO NOT place or store product where it can drop into water or other liquid.
DO NOT reach for product that has fallen into water. Unplug IMMEDIATELY.
A product should NEVER be left unattended when plugged in.
This device is to be used only in accordance with the prescription of a physician and this Owner’s Manual. If at any time the patient or attendant conclude that the patient is receiving an insufficient amount of oxygen, the supplier and/or physician should be contacted immediately.
DO NOT change the L/min. setting on the flowmeter unless a change has been prescribed by your physician or therapist.
Part No 1118353 7 Platinum™Series
SECTION 1—GENERAL GUIDELINES
Close supervision is necessary when this product is used near children or physically-challenged individuals.
Use this product for only intended use as described in this manual.
DO NOT connect the concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices.
Use of some administration accessories or certain humidifiers not specified for use with oxygen concentrator may impair the performance.
In certain circumstances oxygen therapy can be hazardous. Invacare recommends that you seek medical advice before using this machine.
This equipment is to be used as an oxygen supplement and is not considered life supporting or life sustaining.
Avoid creation of any spark near medical oxygen equipment. This includes sparks from static electricity created by any type of friction.
POLARIZED PLUG INSTRUCTION
As a safety feature, this appliance has a polarized plug (one blade is wider than the other). This plug will fit in a polarized outlet only one way. If the plug does not fit fully in the outlet, reverse the plug. If it still does not fit, contact a qualified electrician. DO NOT attempt to defeat this safety feature.
Platinum™Series 8 Part No 1118353
SECTION 2—FEATURES
SECTION 2—FEATURES
Oxygen Purity Indicator Lights
(SensO
Oxygen Outlet
and Power Indicator Lights
model only)/ Fault
2
(LX model only)
Flowmeter
Power
Cord
REAR VIEW
Outlet
Fitting
Power Switch
Circuit Breaker
Elapsed Time Meter
Cabinet Filter
(one of two shown)
ACCESSORIES (NOT SHOWN):
Remote Humidifier/Flowmeter Accessory -
IRCRF5
PreciseRX™ Pediatric Humidifier/Flowmeter
Accessory - IRCPF16
Venture® HomeFill II home oxygen
compressor - IOH200
NOTE: This outlet fitting is to be used only for filling oxygen cylinders with the Venture HomeFill II home oxygen compressor. The outlet fitting does not affect concentrator performance. Refer to the HomeFill II owner's manual, part number 1100873, for connection and operating instructions. When not in use, the plug provided with the concentrator should be inserted into the outlet fitting. For more information about the HomeFill II, contact your Invacare dealer.
Part No 1118353 9 Platinum™Series
SECTION 3—HANDLING
SECTION 3—HANDLING
The concentrator should always be kept in the upright position to prevent cabinet damage while being transported. The shipping container has been designed to assure maximum protection of the concentrator.
If the concentrator is to be reshipped by common carrier, additional cartons are available from Invacare.
Unpacking
NOTE: For this procedure, refer to FIGURE 3.1.
1. Check for any obvious damage to the carton or its contents. If damage is evident, notify the carrier, or your local dealer.
2. Remove all loose packing from the carton.
3. Carefully remove all the components from the carton.
NOTE: Unless the oxygen concentrator is to be used immediately, retain containers and packing materials for storage until use of the concentrator is required.
FIGURE 3.1 Unpacking
Inspection
1. Inspect/examine exterior of the oxygen concentrator for nicks, dents, scratches or other damages. Inspect all components.
Storage
1. Store the repackaged oxygen concentrator in a dry area.
2. DO NOT place objects on top of repackaged concentrator.
Platinum™Series 10 Part No 1118353
SECTION 4—SPECIFICATIONS
SECTION 4—SPECIFICATIONS
Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60
Hz
Rated Current Input: Platinum 5 and XL: 4.3 A
Platinum 10: 4.7 A
Sound Level: Platinum 5 and XL: 50 dB Average
Platinum 10: 58 dB Average
Altitude: Platinum 5 and XL - Up to 6,000 ft. (1828
meters) above sea level without degradation of concentration levels. From 6,000 ft. (1828 meters) to 13,129 ft. (4000 meters) below 90% efficiency
Platinum 10 - Up to 4,000 ft. (1230 meters) above sea level without degradation of con­centration levels. From 4,000 ft. (1230 meters) to 13,129 ft. (4000 meters) below 90% efficiency
*Oxygen Output Concentration Levels:
Maximum Outlet Pressure: 5 ± 0.5 psi. (34.5 ± 3.45 kPa.)
All 5LXO2/5LX models 95.6% to 87% at 0.5 to 5 L/min. and all 10LXO2/10LX models
95.6% to 87% at 2 L to 10 L/min.
*NOTE: Stated concentration levels achieved after initial warm-up period (approximately 30 minutes).
Part No 1118353 11 Platinum™Series
SECTION 4—SPECIFICATIONS
5LXO2 INDICATORS:
Label
Symbol
O
2
O2 Purity
SYSTEM FAILURE
Below 73%
73% to 85%
SYSTEM OKAY
Over 85%
Indicator Lights
RED - Continuous Audible
Alarm Sieve-GARD Compressor Shutdown
YELLOW
GREEN
5LX INDICATORS:
Label
Symbol
I/O
Flow Range: Platinum 5 and XL - 0.5 to 5 L/min. For
Status
SYSTEM FAILURE
SYSTEM OKAY
flowrates less than 1 L/min., we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory.
Indicator Lights
RED
GREEN
Platinum 10 - 2 L/min. to 10 L/min
Low Flow Alarm: Platinum 5 - 0 L/min. to 0.5 L/min.,
Platinum 10 - 0 L/min. to 1 L/min. Rapid Audible Alarm Beeping (No Accessories Connected)
Average Power Consumption:
Pressure Relief Mechanism
Platinum 5 and XL: 400W Platinum 10: 585W
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)
Operational at:
Change in maximum recom-
0.7 L/min. mended flow when back pres­sure of 7kPa is applied:
Platinum™Series 12 Part No 1118353
SECTION 4—SPECIFICATIONS
Filters: Cabinet (2), Outlet HEPA and Compressor
Inlet.
Safety System: Current overload or line surge shutdown.
High temperature compressor shutdown. High Pressure Alarm w/compressor shut­down. Low Pressure Alarm w/compressor shutdown. Battery Free Power Loss Alarm. SensO
Oxygen System (SensO2 Model)
2
Low Flow Alarm.
Width: 18 3/8 in. ± 3/8 in. (46.7 cm ± 1 cm)
Height: 26 3/8 in. ± 3/8 in. (67.0 cm ± 1 cm)
Depth: 14 3/8 in. ± 3/8 in. (36.5 cm ± 1 cm)
Weight: Platinum 5 and XL: 52 lbs ± 2 lbs (23.6 kg ±
1 kg)
Platinum 10: 53 lbs ± 2 lbs (24.0 kg ± 1 kg)
Shipping Weight: Platinum 5 and XL: 58 lbs ± 2 lbs (26.3 kg ±
1 kg)
Platinum 10: 60 lbs ± 2 lbs (27.2 kg ± 1 kg)
Operating Ambient Temperature:
50°F - 95°F (10°C - 35°C) at 20-60% relative humidity
Exhaust: Platinum 5 and XL: Less than Ambient
+ 35°F (+19°C).
Platinum 10: Less than Ambient + 45°F (+25°C).
Oxygen Output:
Platinum 5 and XL: Less than Ambient + 4°F (+2°C).
Platinum 10: Less than Ambient + 8°F (+4.4°C).
Cabinet: Impact Resistant flame-retardant plastic cab-
inet that conforms to UL 94-V0.
Regulatory Listing: Double Insulated Product ETL certified
complying with UL 1097 ETL certified to CSA C22.2 No. 68
Electrical: No extension cords.
Part No 1118353 13 Platinum™Series
SECTION 4—SPECIFICATIONS
Placement: No closer than three (3) inches from any
wall, furniture, draperies, or similar surfaces.
Tubing: 7 ft cannula with a maximum 50 ft of
Crush-Proof Tubing (DO NOT pinch). Platinum 10: Recommended use up to 25 ft high flow tubing with high flow cannula at all flow rates.
Relative Humidity: 20 to 60%
Time of Operation: Up to 24 hours per day.
Recommended Storage and Shipping Temperature:
-20°F to 150°F (-29°C to 65°C), at 15-95%
relative humidity
Platinum™Series 14 Part No 1118353
NOTES
NOTES
Part No 1118353 15 Platinum™Series
NOTES
NOTES
Platinum™Series 16 Part No 1118353
Please Seal with
Tape Before Mailing
REGISTER YOUR PRODUCT!
The benefits of registering:
1. Safeguard your investment.
2. Ensure long term maintenance and servicing of your purchase.
3. Receive updates with product information, maintenance tips, and industry
news.
4. Invacare can contact you or your provider, if servicing is needed on your
product.
5. It will enable Invacare to improve product designs based on your input
and needs.
Register ONLINE at www.invacare.com
- or -
Complete form and mail.
Any registration information you submit will be used by Invacare Corporation
only, and protected as required by applicable laws and regulations.
PRODUCT REGISTRATION FORM
NAME _____________________________________________________
ADDRESS __________________________________________________
CITY _______________________ STATE/PROVINCE _____________
ZIP/POSTAL CODE ________________
EMAIL ___________________________ PHONE NO. ____________
INVACARE MODEL NO. _____________________________________
SERIAL NO. ______________________
PURCHASED FROM _______________
DATE OF PURCHASE: _____________
1. Method of purchase: (check all that apply) Medicare Insurance Medicaid Other
2. This product was purchased for use by: (check one) Self Parent Spouse Other
3. Product was purchased for use at: Home Facility Other
4. I purchased an Invacare product because: Price Features (list features)
5. Who referred you to Invacare products? (check all that apply)
Doctor Therapist Friend Relative Dealer/Provider OtherAdvertisement (circle one): TV, Radio, Magazine, Newspaper No Referral
6. What additional features, if any, would you like to see on this product?
_________________________________________________________________
7. Would you like information sent to you about Invacare products that may be available for a particular medical condition? Yes No
If yes, please list any condition(s) here and we will send you information by email and/or mail about any available Invacare products that may help treat, care for or manage such condition(s):
_________________________________________________________________
8. Would you like to receive updated information via email or regular mail about the Invacare home medical products sold by Invacare's dealers? Yes No
_________________________________________________________________
9. What would you like to see on the Invacare website?
_________________________________________________________________
10. Would you like to be part of future online surveys for Invacare products? Yes No
11. User's Year of birth: _________________
If at any time you wish not to receive future mailings from us, please contact us at Invacare Corporation, CRM Department, 39400 Taylor Parkway, Elyria, OH 44035, or fax to
877-619-7996 and we will remove you from our mailing list.
TO FIND MORE INFORMATION ABOUT OUR PRODUCTS, VISIT
WWW.INVACARE.COM
SECTION 5—OPERATING INSTRUCTIONS
SECTION 5—OPERATING
INSTRUCTIONS
Introduction
Your oxygen concentrator is intended for individual use in the home. It is an electronically operated device that separates oxygen from room air. It provides high concentration of oxygen directly to you through a nasal cannula. Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems.
Your provider will show you how to use your oxygen concentrator. He/She should be contacted with any questions or problems regarding your oxygen concentrator. This owner’s manual will tell you about your concentrator and will serve as a reference as you use your concentrator.
Select a Location
You may select a room in your house where using your oxygen concentrator would be most convenient. Your concentrator can be easily rolled from room to room on its casters.
Your oxygen concentrator will perform best when operated under the following conditions. Usage in environments other than those described may result in the need for increased equipment maintenance. The air intake of the unit should be located in a well ventilated area to avoid airborne pollutants and/or fumes.
Part No 1118353 21 Platinum™Series
SECTION 5—OPERATING INSTRUCTIONS
Recommended Guidelines for Optimum Performance
Temperature: 50°F - 95°F (10°C - 35°C)
Electrical: No extension cords.
Placement: No closer than three (3) inches from the
wall, furniture, draperies, or similar surfaces.
Tubing and Cannula: 7 ft cannula with a maximum 50 ft of
Crush-Proof Tubing (DO NOT pinch).
Platinum 10: Recommended use up to 25 ft high flow tubing with high flow cannula at all flow rates.
Environment: Smoke and soot-free. No confined spaces
(Example: No closets).
Relative Humidity: 20 to 60%
Time of Operation: Up to 24 hours per day.
Flow Rate Platinum 5 and XL: From 0.5 L/min. to 5
L/min. For flow rates less than 1 L/min., use the Pediatric Flowmeter Accessory.
Platinum 10: From 2 L/min. to 10 L/min
Minimum Operating Time: 30 Minutes
NOTE: Ensure that your concentrator is at least three (3) inches away from walls, draperies or furniture to assure sufficient air flow. Avoid deep pile carpets and heaters, radiators or hot air registers.
Plug in Power Cord
1. Plug in power cord to an electrical outlet.
Connect Humidifier (If So Prescribed)
NOTE: For this procedure, refer to FIGURE 5.1 - FIGURE 5.4.
Platinum™Series 22 Part No 1118353
SECTION 5—OPERATING INSTRUCTIONS
1. Fill humidifier with distilled water to the level indicated by the manufacturer. (For Platinum 5,XL - recommended humidifiers are: Invacare 3260, 003-40, or 006-40).
WARNING
DO NOT overfill humidifier.
Replace the humidifier cap and securely tighten.
DO NOT reverse the oxygen input and output connections. Water from the humidifier bottle will travel through the cannula back to the patient.
Humidifier Bottle
FIGURE 5.1 Filling the Humidifier
2. Place the humidifier bottle in the humidifier compartment.
Humidifier Bottle
FIGURE 5.2 Humidifier Compartment
3. Remove the filter access door located on the side of the concentrator. Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door open. Refer to Detail “A” in FIGURE 5.3.
Part No 1118353 23 Platinum™Series
SECTION 5—OPERATING INSTRUCTIONS
4. The humidifier adapter is next to the inlet filter. Pull up and remove the humidifier bottle adapter. Refer to Detail “B” in FIGURE 5.3.
5. Attach it to the humidifier by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened. Refer to Detail “C” in FIGURE 5.3.
DETAIL “A”
Plate Groove (Insert Flat Head Screwdriver Here)
Filter Access Door
DETAIL “B”
Humidifier Bottle
Adapter
DETAIL “C”
Filter
Access
Door
Humidifier Bottle
Adapter
FIGURE 5.3 Attaching the Humidifier Bottle Adapter
6. Attach oxygen tubing from the humidifier bottle to oxygen outlet connector on the oxygen concentrator.
Platinum™Series 24 Part No 1118353
SECTION 5—OPERATING INSTRUCTIONS
7. Attach the cannula/patient supply tubing to the humidifier bottle outlet.
8. After assembly, ensure that oxygen is flowing through the cannula.
Oxygen
Tubing
Humidifier Outlet
Oxygen Outlet
Connector
FIGURE 5.4 Attaching the Oxygen Tubing
WARNING
For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life.
Power Switch
NOTE: For this procedure, refer to FIGURE 5.5.
1. Press power switch to on position. All the panel lights and the audible
On/Off (I/O)
Power Switch
alarm will come on for one second, indicating that the unit is functioning properly. After one second, only the GREEN system ok/power light will stay
FIGURE 5.5 Power Switch
on.
Part No 1118353 25 Platinum™Series
SECTION 5—OPERATING INSTRUCTIONS
Flowrate
NOTE: For this procedure, refer to FIGURE 5.6.
1. Turn the flowrate knob to the setting prescribed by your physician or therapist.
WARNING
DO NOT change the L/min. setting on the flowmeter unless a change has been prescribed by your physician or therapist.
NOTE: To properly read the flowmeter, locate the prescribed flowrate line on the flowmeter. Next, turn the flow knob until the ball rises to the line. Now, center the ball on the L/min. line prescribed.
2. If the flowrate on the flowmeter ever falls below 0.5 L/min. for Platinum 5 (1 L/min. for Platinum 10) for more than about one minute, the LOW FLOW alarm will be triggered. This is a rapid beeping of the audible alarm. Check your tubing or accessories for blocked or kinked tubing or a defective humidifier bottle. After rated flow is restored, more than 0.5 L/min. for Platinum 5 (1 L/min. for Platinum 10), the LOW FLOW audible alarm will go off.
NOTE: The use of some accessories such as the pediatric flowstand and the HomeFill II compressor will deactivate the Low Flow Alarm.
Flow Knob
Ball
FIGURE 5.6 Flowrate
Platinum™Series 26 Part No 1118353
SECTION 5—OPERATING INSTRUCTIONS
SensO2 Oxygen Purity Indicator (If Your
Unit Features the O
This feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate.
Sensor Feature)
2
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm-up time (approximately 30 min.) while waiting for the O2 purity to
reach maximum.
When the unit is turned on, the GREEN light will come on (SYSTEM OK/O2 greater than 85%). After five (5) minutes, the oxygen sensor will be operating normally and will control the indicator lights depending on oxygen concentration values.
Explanation of Oxygen Purity Indicator Lights
NOTE: For this procedure, refer to FIGURE 5.7.
GREEN light (O2) - Normal Operation.
YELLOW light ( ) - Call supplier IMMEDIATELY. You may continue to use the concentrator unless instructed otherwise by your supplier. Be certain that backup oxygen is nearby.
RED light ( )- Total unit shutdown. Switch IMMEDIATELY to backup oxygen supply. Call supplier IMMEDIATELY.
GREEN light - with YELLOW light flashing - Call supplier IMMEDIATELY. Oxygen sensor malfunctioning; you may continue to use the concentrator.
Part No 1118353 27 Platinum™Series
SECTION 5—OPERATING INSTRUCTIONS
OXYGEN PURITY
Auto Shut - Down
Use Backup
Call Supplier
RED
Below Normal
Normal
O
2
YELLOW
GREEN
FIGURE 5.7 SensO2 Oxygen Purity Indicator (If Your Unit
Features the O2 Sensor Feature)
Indicators (If Your Unit Does Not Feature the O
NOTE: For this procedure, refer to FIGURE 5.8.
LX models indicator Light Explanation
RED light ( ) - Total Unit Shut-Down. Switch immediately to a back-up oxygen supply. Call supplier IMMEDIATELY.
GREEN light (I/O) - On/Off. System okay.
Auto Shut - Down
Use Backup
Sensor)
2
RED
Call Supplier
Normal
I/O
GREEN
FIGURE 5.8 Indicators (If Your Unit Does Not Feature the O2
Sensor)
Platinum™Series 28 Part No 1118353
SECTION 6—MAINTENANCE
SECTION 6—MAINTENANCE
NOTE: The Invacare concentrators are specifically designed to minimize routine preventive maintenance. In places with high dust or soot levels, maintenance may need to be performed more often. Only authorized or factory-trained personnel should perform preventive maintenance on the concentrator. Power should be disconnected before beginning preventive maintenance on the Invacare concentrator.
NOTE: At a minimum, preventive maintenance MUST be performed according to the maintenance record guidelines. Refer to Preventive Maintenance Record on page 31.
Routine Maintenance
NOTE: The following routine maintenance should be performed more frequently. See recommended intervals in each section and preventive maintenance record.
Cleaning the Cabinet Filter
NOTE: For this procedure, refer to FIGURE 6.1.
NOTE: There are two (2) cabinet filters one (1) located on each side of the cabinet.
1. Remove each filter and clean at least once a week depending on environmental conditions.
NOTE: Environmental conditions that may require more frequent cleaning of the filters include but are not limited to: high dust, smoking, air pollutants, etc.
2. Clean the cabinet filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly.
3. Dry the filters thoroughly before reinstallation.
Part No 1118353 29 Platinum™Series
SECTION 6—MAINTENANCE
CAUTION
DO NOT operate the concentrator without the filters installed.
Cabinet Filter
(located on both
sides of
concentrator)
FIGURE 6.1 Cleaning the Cabinet Filter
Cabinet
WARNING
UNPLUG the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet.
1. Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge.
Humidifier
NOTE: To clean the oxygen humidifier, follow the instructions provided by the manufacturer. If none are provided, follow these steps:
1. Clean the humidifier every day.
2. Wash it in soapy water and rinse it with a solution of ten (10) parts water and one (1) part vinegar.
3. Rinse thoroughly with hot water and refill with distilled water to the level shown on the humidifier.
Platinum™Series 30 Part No 1118353
SECTION 6—MAINTENANCE
WARNING
DO NOT overfill humidifier bottle.
MODEL NO. IR C SERIAL NO.
PREVENTIVE MAINTENANCE RECORD
)
2
INVACARE PLATINUM OXYGEN CONCENTRATOR
MODELS)
2
ANNUALLY, DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
Clean/Replace Cabinet Filter(s)
Check Outlet HEPA Filter*
Check Compressor Inlet Filter*
Check Oxygen Concentration
(SensO
Check Power Loss Alarm
AS REQUIRED
Clean Heat Exchanger
Replace Exhaust Muffler
Record Date of Service
Record Elapsed Hours On Hour Meter
Clean Cabinet Filter(s)
Check Prescribed L/min. Flow Rate
EVERY 180 DAYS FOR IRC5LX (EVERY 90 DAYS FOR IRC10LX)
ON EACH INSPECTION
Check Oxygen Concentration
(UNITS WITHOUT SensO
Rebuild Top End of Compressor
*Refer to Preventive Maintenance section of Service Manual.
FIGURE 6.2 Preventive Maintenance Record
Part No 1118353 31 Platinum™Series
SECTION 7—TROUBLESHOOTING GUIDE
SECTION 7—TROUBLESHOOTING GUIDE
SYMPTOM: PROBABLE CAUSE: SOLUTION:
Alarm: Main Power Loss: 1. Insert plug into outlet.
SHORT BEEPS, LONG PAUSE
Concentrator NOT operating, power switch ON.
Beep......Beep...... 3. Tripped circuit
Alarm:
CONTINUOUS
Concentrator NOT operating, Power Switch ON.
Beeeeeeeep....
1. Power cord NOT plugged in.
2. No power at outlet. 3. Call Service Provider.
breaker.
System Failure:
1. Unit overheating due to blocked air intake.
2. Insufficient power at outlet.
3. Internal repairs required
2. Inspect house circuit breakers or fuses. If problem recurs, use a different outlet.
1a. Remove and clean cabinet filters.
1b. Move oxygen concentra­tor at least three (3) inches away from walls, draperies or furniture.
2. Do not use extension cords. Move to another elec­trical outlet or circuit.
3. Call Service Provider.
Platinum™Series 32 Part No 1118353
SECTION 7—TROUBLESHOOTING GUIDE
SYMPTOM: PROBABLE CAUSE: SOLUTION:
YELLOW or RED LIGHT Illumi­nated.
Alarm: CONTINUOUS On red light only.
* Only applies to 5LXO
and
2
10LXO
SensO
2
2
models.
1. Low oxygen purity.*
2. Kinked or blocked tub­ing, cannula or humidi­fier.*
3. Flowmeter set at 0.5 L/min. for Platinum 5 and 1 L/min. for Platinum 10 or less.*
4. Unit overheating due to blocked air intake.
5. Insufficient power at outlet.
6. Internal repairs required.
1. Clean or Replace filters.
2. Inspect for kinks or block­ages. Correct, clean or replace item. Once cor­rected, turn power OFF for 60 seconds and then turn power back ON.
3. Check flowmeter is set to
1.0 L/min. or more. (Refer to Recommended Guidelines for Optimum Performance on page 22.
4.a. Remove and clean cabinet filters.
4b. Move concentrator at least three (3) inches from walls, draperies, and furni­ture.
5. Do not use extension cords. Move to another elec­trical outlet or circuit.
6. Call Service Provider.
GREEN LIGHT with YELLOW
1. Internal repairs required.
1. Call Service Provider.
Light Flashing.
Part No 1118353 33 Platinum™Series
SECTION 7—TROUBLESHOOTING GUIDE
SYMPTOM: PROBABLE CAUSE: SOLUTION:
Alarm:
RAPID
Beep....Beep...
Beep...Beep
LOW FLOW ALARM
1a. Kinked or blocked tubing, cannula or humid­ifier.
1b. Flowmeter set at 0.5 L/min. for Platinum 5 and 1 L/min. for Platinum 10 or less.
1a. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power OFF for 60 seconds and then turn power back ON.
1b. Reset flowmeter to pre­scribed flowrate. If prescribed flowrate is below 0.5 L/min., use the Pediatric Flowmeter IRCPF16.
Platinum™Series 34 Part No 1118353
LIMITED WARRANTY
LIMITED WARRANTY
NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to original purchaser of our products.This warranty gives you specific legal rights and you may also have other legal rights which may vary from state to state.
Invacare Corporation warrants its Platinum Oxygen Concentrator to be free from defects in materials and workmanship from date of purchase: Platinum 5 and XL for a period of five (5) years, Platinum 10 for a period of three (3) years.The 4-way valve body carries a lifetime warranty from date of purchase. DO NOT open or ATTEMPT TO service valve body. This will void any and all warranty applicable to the valve body. If within such warranty period any such product shall be proven to Invacare Corporation's satisfaction to be defective, such product shall be repaired or replaced, at Invacare Corporation's option. This warranty only applies to the labor for repairs performed by the Invacare Service Department or Invacare Service Centers. It does not apply to the labor performed by the purchaser or user. This warranty does not include normal wear and tear or shipping charges. Invacare Corporation's sole obligation and your exclusive remedy under this warranty shall be limited to such repair or replacement. Routine maintenance items, such as filters, are excluded from this warranty.
For warranty service, please contact Invacare Corporation's service department at the toll free number on the back cover.
Upon receiving notice of an alleged defect in a product, Invacare Corporation will issue a serialized return authorization. It shall then be the responsibility of the purchaser to return the entire unit or remove, at purchaser's cost, the defective component part(s) identified, pack the component part(s) in a manner to avoid shipping damage and to ship the component part(s) to either Invacare Corporation's plant or service center as specified by Invacare Corporation in advance. Defective component part(s) must be returned for warranty inspection using the serial number as identification within thirty (30) days of return authorization date. Do not return products to our factory without prior consent. C.O.D. shipments will be refused; please prepay shipping charges.
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, SOOT OR SMOKE-FILLED ENVIRONMENTS, OR OTHER THAN NORMAL APPLICATION, USE OR SERVICE, OR TO PRODUCTS MODIFIED WITHOUT INVACARE CORPORATION'S EXPRESS WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO, MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS) OR TO PRODUCTS
Part No 1118353 35 Platinum™Series
DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE CORPORATION OR TO PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE CORPORATION'S CONTROL. THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER EXPRESS OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTIBILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PERSUANT TO THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER.
SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE, OR LIMITATION OF HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE EXCLUSION AND LIMITATION MAY NOT APPLY TO YOU.
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/PROVINCIAL LAWS AND REQUIREMENTS.
This product may be manufactured under one or more of the following patents: US 5,906,672 and US 5,917,135. Other patents pending.
Invacare Corporation
USA
One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 Technical Services 800-832-4707
www.invacare.com
Invacare, Yes, you can., SensO Venture and the Medallion Design are
registered trademarks of Invacare Corporation. Platinum, HomeFill and Precise R
trademarks of Invacare Corporation. © 2005 Invacare Corporation
Part No 1118353 Rev E - 09/05 Asm. No. 1106664
,
2
are
X
Operator’s Manual
Platinum™Series
XL, 5, 10
II
with
/SensO
Oxygen Concentrators Standard,
HomeFill
PART TWO
HomeFill II Connection and Use with the
Concentrator
Invacare 5
Model IRC5LX
Invacare 10
Model IRC10LX
User: Before using the product, read this manual and save for future reference. Dealer:
This manual MUST be given to the end user.
Invacare 5
Model IRC5LX02
Invacare 10
Model IRC10LX02
with
/SensO2
with
/SensO
and
2
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
WARNING
WARNING
DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT INVACARE TECHNICAL SERVICES BEFORE ATTEMPTING TO USE THIS EQUIPMENT ­OTHERWISE, SERIOUS INJURY OR PROPERTY DAMAGE MAY RESULT.
ACCESSORIES
There are many different types of humidifiers, oxygen tubing, cannulas and masks that can be used with this device. You should contact your local home care provider for recommendations on which of these devices will be best for you. They should also give you advice on the proper usage, maintenance, and cleaning.
WARNING
Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.
NOTICE
The information contained in this document is subject to change without notice.
NOTE: Updated versions of this manual are available on www.invacare.com.
Platinum™Series 2 Part No 1118354
TABLE OF CONTENTS
ACCESSORIES .................................................. 2
SPECIAL NOTES ............................................... 4
HOMEFILL II INFORMATION ............................. 5
SECTION 1—GENERAL GUIDELINES .................. 7
Handling................................................................................................. 9
SECTION 2—INITIAL SETUP ............................ 10
Compressor Initial Setup Checklist ..............................................10
Choosing a Location.........................................................................10
Transporting the Compressor .......................................................10
Installing the Power Cord ...............................................................11
Connecting the Compressor to the Concentrator ..................12
SECTION 3—COMPRESSOR OPERATION .......... 19
Compressor Operation Checklist.................................................19
Cylinder Prefill Inspection...............................................................19
Connecting/Disconnecting the Cylinder to the
Compressor........................................................................................22
Turning the Compressor On .........................................................26
Setting the Flow Knob to the Prescribed Setting ......................28
CONCENTRATOR WARRANTY ...................... 30
Part No 1118354 3 Platinum™Series
SPECIAL NOTES
SPECIAL NOTES
Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words.
SIGNAL WORD MEANING
DANGER Danger indicates an imminently hazardous
situation which, if not avoided, will result in death or serious injury.
WARNING Warning indicates a potentially hazardous
situation which, if not avoided, could result in death or serious injury.
CAUTION Caution indicates a potentially hazardous
situation which, if not avoided, may result in property damage.
DANGER
Users MUST NOT SMOKE while using this device. Keep all matches, lighted cigarettes or other sources of ignition out of the room in which this product is located. NO SMOKING signs should be prominently displayed. Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result in severe fire, property damage and cause physical injury or death.
Platinum™Series 4 Part No 1118354
HOMEFILL II INFORMATION
HOMEFILL II INFORMATION
HomeFill II is an accessory device to the oxygen concentrator that enables the patient to refill custom portable oxygen cylinders for personal use.
WARNING
DO NOT install, assemble or operate the HomeFill II compressor without first reading and understanding the compressor owner’s manual, the concentrator owner’s manual, and the oxygen cylinder warning label. If you are unable to understand the warnings, cautions and instructions, contact a dealer or technical personnel if applicable before attempting to install this equipment ­otherwise, injury or damage may occur.
Check all parts for shipping damage. In case of damage, DO NOT use. Contact Invacare for further instruction.
NOTICE
The HomeFill II compressor is for use with specially designed Invacare Platinum 5, 10 and XL HomeFill II compatible concentrators.
The concentrator can be set to the following maximum flowrates when operating with HomeFill II compressor: Platinum 5 and XL: 3 L/min. or less. Platinum 10: 5 L/min. or less. To determine if your concentrator is HomeFill II compatible, examine the lower right corner of the rear of the concentrator for an outlet fitting, refer to illustration on next page.
Part No 1118354 5 Platinum™Series
HOMEFILL II INFORMATION
Outlet Fitting
TOP VIEW
Power Cord
Concentrator
Platinum™Series 6 Part No 1118354
SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
WARNING
SECTION 1 - IMPORTANT SAFEGUARDS contains important information for the safe operation and use of this product. DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as Owner’s Manuals, Service Manuals or Instruction Sheets supplied with this product or optional equipment. If you are unable to understand the Warnings, Cautions or Instructions, contact a health care professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.
NOTICE
Refer to HomeFill II Compressor Owner’s Manual, 1100873, for detailed information on all warnings, use, care and maintenance of the HomeFill II compressor.
In order to ensure the safe installation, assembly and operation of the HomeFill II Compressor these instructions must be followed.
“Caution: Federal law restricts this device to sale by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.”
WARNING
The use of this device is limited to the oxygen patient. Cylinders MUST be used ONLY by the oxygen patient and are not to be distributed to any other individual for any purpose.
Part No 1118354 7 Platinum™Series
SECTION 1—GENERAL GUIDELINES
WARNING
The maximum patient flow while filling cylinders with the IOH200 is 3 L/min for Platinum 5 and XL, 5 L/min. for Platinum 10. Patients with a higher prescribed flow should not transfill while receiving supplemental oxygen treatment from the Platinum Concentrator.
NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air opening may be blocked. Keep the openings free from lint, hair and the like.
DO NOT disassemble. There are minimal serviceable parts on the HomeFill II Compressor. If any service is required, contact Invacare Technical Service.
DO NOT allow oil from your hands or other source to come into contact with the regulator or cylinder valve connection. These solutions may be flammable and cause injury.
The HomeFill II is equipped with a high pressure relief valve to ensure the user’s safety. When activated, this safety feature will make an extremely loud noise. If this noise occurs, turn the unit off and contact your Home Health Care Provider or Service Representative.
Before moving or repositioning either the compressor or concentrator, always disconnect the AC power cords and the interconnect hose between the compressor and concentrator. Failure to do so may result in damage to the unit or personal injury.
Children should always be supervised around the HomeFill II unit. When used with the HomeFill table or similar elevated support surfaces, a child could pull over any such table causing the unit to fall. Failure to observe this warning may result in personal injury or product damage.
Platinum™Series 8 Part No 1118354
SECTION 1—GENERAL GUIDELINES
WARNING
Handling
Use extreme care when handling and filling an oxygen cylinder. Full oxygen cylinders are under pressure and can become a projectile if dropped or mishandled.
NEVER transport a compressor with a cylinder connected to or stored on the compressor, otherwise injury or damage can occur.
DO NOT store filled oxygen bottles near a furnace, hot water tank, in the trunk of a car or other high temperature area. Storage in areas such as this can result in bursting of the bottle or fire.
Part No 1118354 9 Platinum™Series
SECTION 2—INITIAL SETUP
SECTION 2—INITIAL SETUP
NOTE: Dealers Only - Refer to Dealer Information in Owner’s Manual, part number 1100873, before performing this procedure.
Compressor Initial Setup Checklist
Before operating the HomeFill II Compressor for the first time, complete the following checklist:
Choose a location for the compressor.Transport the compressor to the desired location.Connect the power cord to the compressor.Connect the compressor to the concentrator.
Choosing a Location
The compressor should be at least three (3) inches away from walls, draperies or furniture to ensure sufficient air flow.
Avoid placing the compressor next to heaters, radiators or hot air registers.
Do not use in a closet.
The compressor should sit on a hard flat stable surface, such as a table.
Transporting the Compressor
WARNING
NEVER transport a compressor with a cylinder connected to or stored on the compressor, otherwise injury or damage can occur.
Platinum™Series 10 Part No 1118354
SECTION 2—INITIAL SETUP
CAUTION
When transporting the compressor, be careful not to bump the compressor or the connectors into obstacles. Otherwise, damage to the compressor may occur.
NOTE: For this procedure, refer to FIGURE 2.1.
1. There are two (2) methods to safely transport the compressor:
A. One hand - Grasp the
hand grip at the bottom rear of the compressor.
B. Two hands - Grasp
the hand grips on the sides of the compressor.
Hand Grip
Hand Grip
FIGURE 2.1 Transporting
the Compressor
Hand Grip
Installing the Power Cord
NOTE: For this procedure, refer to FIGURE 2.2.
1. Ensure the compressor is on a hard flat stable surface, such as a table.
2. Plug the power cord into the electric connector in the back of the compressor.
3. Plug the opposite end of the power cord into the wall outlet.
Part No 1118354 11 Platinum™Series
SECTION 2—INITIAL SETUP
Wall
Outlet
Power
Cord
Compressor
(REAR)
Electrical Connection
NOTE: This product has a polarized plug (one blade is wider than the other). As a safety feature, this plug will fit in a polarized outlet only one way. If the plug does not fit fully in the outlet, reverse the plug. If it still does not fit, contact a qualified electrician. DO NOT attempt to defeat this safety feature.
FIGURE 2.2 Installing the Power Cord
Connecting the Compressor to the Concentrator
NOTE: For this procedure, refer to FIGURE 2.3.
1. Insert one (1) end of the interconnect hose into the outlet fitting on the back of the concentrator (Detail “A”).
2. Insert the opposite end of the interconnect hose into the inlet fitting on the side of the compressor (Detail “B”).
NOTE: The metal tabs on the concentrator outlet fitting and the compressor inlet fitting will pop out with an audible “click” when the end of the interconnect hose is properly installed (Detail “C”).
3. Loop any excess interconnect hose and secure to the back of the concentrator with the fastening straps (not shown).
Platinum™Series 12 Part No 1118354
SECTION 2—INITIAL SETUP
DETAIL “A”
Interconnect
Interconnect Hose into
Hose
CONCENTRATOR
TOP VIEW
Outlet Fitting
Insert End of
Inlet/Outlet Fitting
Power
Cord
DETAIL “B”
Compressor
Inlet Fitting
DETAIL “C”
Inlet/Outlet Fitting
Metal
Tab
FIGURE 2.3 Connecting the Compressor to the Concentrator
Part No 1118354 13 Platinum™Series
NOTES
NOTES
Platinum™Series 14 Part No 1118354
Please Seal with
Tape Before Mailing
REGISTER YOUR PRODUCT!
The benefits of registering:
1. Safeguard your investment.
2. Ensure long term maintenance and servicing of your purchase.
3. Receive updates with product information, maintenance tips, and industry news.
4. Invacare can contact you or your provider, if servicing is needed on your product.
5. It will enable Invacare to improve product designs based on your input and needs.
Register ONLINE at www.invacare.com
- or -
Complete form and mail.
Any registration information you submit will be used by Invacare Corporation
only, and protected as required by applicable laws and regulations.
PRODUCT REGISTRATION FORM
NAME _____________________________________________________
ADDRESS __________________________________________________
CITY _______________________ STATE/PROVINCE _____________
ZIP/POSTAL CODE ________________
EMAIL ___________________________ PHONE NO. ____________
INVACARE MODEL NO. _____________________________________
SERIAL NO. ______________________
PURCHASED FROM _______________
DATE OF PURCHASE: _____________
1. Method of purchase: (check all that apply)
Medicare Insurance Medicaid Other
2. This product was purchased for use by: (check one)
Self Parent Spouse Other
3. Product was purchased for use at:
Home Facility Other
4. I purchased an Invacare product because:
Price Features (list features)
5. Who referred you to Invacare products? (check all that apply)
Doctor Therapist Friend Relative Dealer/Provider OtherAdvertisement (circle one): TV, Radio, Magazine, Newspaper No Referral
6. What additional features, if any, would you like to see on this product?
_________________________________________________________________
7. Would you like information sent to you about Invacare products that may be
available for a particular medical condition? Yes No
If yes, please list any condition(s) here and we will send you information by email and/or mail about any available Invacare products that may help treat, care for or manage such condition(s):
_________________________________________________________________
8. Would you like to receive updated information via email or regular mail about the
Invacare home medical products sold by Invacare's dealers? Yes No
_________________________________________________________________
9. What would you like to see on the Invacare website?
_________________________________________________________________
10. Would you like to be part of future online surveys for Invacare products? Yes No
11. User's Year of birth: _________________
If at any time you wish not to receive future mailings from us, please contact us at Invacare Corporation, CRM Department, 39400 Taylor Parkway, Elyria, OH 44035, or fax to
877-619-7996 and we will remove you from our mailing list.
TO FIND MORE INFORMATION ABOUT OUR PRODUCTS, VISIT
WWW.INVACARE.COM
SECTION 3—COMPRESSOR OPERATION
SECTION 3—COMPRESSOR OPERATION
Compressor Operation Checklist
NOTE: For this procedure, refer to FIGURE 3.1.
Each time the HomeFill II Compressor is used to fill a cylinder, complete the following checklist:
Ensure the concentrator has been on for at least thirty (30)
minutes. Refer to the concentrator Owner’s Manual, Part One.
Perform the prefill inspection on the cylinder.Connect the cylinder to the compressor.Turn the compressor on.Examine the indicator lights on the control panel.Disconnect and remove the full cylinder.Turn the compressor off.If filling another cylinder, repeat this checklist.
Cylinder Prefill Inspection
WARNING
ONLY use cylinders that have the Invacare HomeFill II connection and the label shown in FIGURE 3.1.
All cylinders MUST be inspected before attempting to fill. Otherwise, injury or damage may occur.
Part No 1118354 19 Platinum™Series
SECTION 3—COMPRESSOR OPERATION
The Hydrostatic Testing Date
Aluminum cylinders MUST undergo testing every five (5) years.
WARNING
DO NOT fill cylinders that have not been tested in the past five (5) years. Otherwise, serious personal may result. Contact your dealer for replacement.
External Examination
1. Examine the outside of the cylinder for the following conditions, and replace the cylinder if they exist:
Dents or dings
•Arc burns
Oil or grease
Any other signs of damage that might cause a cylinder
to be unacceptable or unsafe for use.
2. Examine the cylinder for evidence of fire or thermal damage. Evidence includes charring or blistering of the paint, or other protective coating or heat sensitive indicator. If fire or thermal damage is found, replace the cylinder.
3. Inspect the regulator assembly for the following, and replace if found:
Debris, oil or grease
Noticeable signs of damage
Signs of corrosion inside the valve
Signs of excessive heat or fire damage
Platinum™Series 20 Part No 1118354
SECTION 3—COMPRESSOR OPERATION
DANGER
Improper filling or abuse of this cylinder, or failure to follow these safety instructions may cause serious bodily injury or death. High pressure oxidizing gas vigorously accelerates combustion. DO NOT alter or modify cylinder or related components. DO NOT use Oil or Grease. Keep cylinder, cylinder valve, all equipment and connections free of oil and grease to avoid violent ignition resulting in serious bodily injury or death.
For medical applications. Use only practitioner. Uninterrupted use of over a long duration, without content of arterial blood, may be harmful.
Oxygen Cylinders filled by the Venture HomeFill shall be used for personal use only. "Not to be filled for resale or use by professional users".
This cylinder is to be used with the Invacare Venture HomeFill system. This cylinder contains oxygen produced by an oxygen concentrator using the pressure swing adsorption process.
DO NOT handle cylinder or use contents until you are professionally trained, including emergency procedures.
Use in accordance Open valve slowly. S
ecure cylinder during s
area.
Keep away from heat, flame and spark. Keep out of
reach of children. DO NOT drop.
Transfilling of this gas is performed by Venture HomeFill Only.
Cylinders that have been refinished and/or exposed to elevated temperatures must be hydrostatically tested before filling.
Cylinder temperature Aluminum cylinders subject temperatures in excess of 350°F service and condemned by trained caustic paint strippers. Valve and safety relief devices must be removed and replaced by trained and authorized personnel. DO NOT alter or change valve assembly.
Invacare Corporation 2101 E. Lake Mary Blvd. Sanford, FL 32773 1-800-832-4707
!
!
WARNING
as directed by a
high concentrations
monitoring its effect on
93 + 3%
with Venture HomeFill Operator’s
Close valve after each use
torage and use. No smoking in cylinder
should not exceed 125°F (52°C).
to the action of
must be
withdrawn from
personnel. DO NOT use
licensed
oxygen
OXYGEN
+
Manual.
and when empty.
fire or heated to
DO NOT REMOVE THIS PRODUCT LABEL
Not for Commercial Resale
Part Number 1125090 Rev B - 11/04
of oxygen
2
FIGURE 3.1 Cylinder Prefill Inspection
Part No 1118354 21 Platinum™Series
SECTION 3—COMPRESSOR OPERATION
Connecting/Disconnecting the Cylinder to the Compressor
Connecting the Cylinder to the Compressor
WARNING
DO NOT modify ANY connections on the HomeFill II compressor. NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur.
DO NOT drop oxygen cylinders. Use both hands when handling/transporting oxygen cylinders. Otherwise, injury or damage may occur.
DO NOT use a liquid leak detector to test for leaks.
NOTE: For this procedure, refer to FIGURE 3.2 and FIGURE 3.3.
1. Examine the cylinder gauge. If the cylinder pressure is less
than 1500 psig., proceed to STEP 2 (FIGURE 3.3).
NOTE: If the cylinder pressure is greater than 1500 psig., DO NOT attempt to top off this cylinder. It may not fill.
2. Set the regulator flow dial on the cylinder to Off (FIGURE
3.3).
3. Remove the cylinder and connector fillport covers (if present).
CAUTION
DO NOT connect the outlet (cannula) connection to the HomeFill II unit, otherwise the conserving cylinder will not work properly.
4. Momentarily push DOWN on the outer ring (sleeve) of the connector fillport to reset the connector as shown in FIGURE 3.2.
Platinum™Series 22 Part No 1118354
SECTION 3—COMPRESSOR OPERATION
NOTE: If the outer ring (sleeve) is in the UP position (GREEN dots not visible), the connector fillport will not be able to accept the cylinder fillport. Pushing DOWN momentarily will reset the connector fillport (GREEN dots visible) to accept the cylinder fillport.
PUSH
DOWN
GREEN Dots
Outer Ring (Sleeve)
Connector Fillport
BEFORE COUPLING CYLINDER, PUSH DOWN
ON SLEEVE
UNTIL GREEN DOTS ARE VISIBLE.
FIGURE 3.2 Resetting Connector Fillport
5. Grasp the cylinder/regulator assembly in the area behind the cylinder gauge (FIGURE 3.3).
6. Position the cylinder in the compressor cradle (FIGURE
3.3).
7. Align the cylinder fillport with the connector fillport (FIGURE 3.3).
8. Pull UP on the outer ring (sleeve) of the connector fillport while pushing DOWN on the cylinder/regulator assembly to couple the cylinder fillport into the connector fillport (FIGURE 3.3).
NOTE: The cylinder is properly connected when an audible “click” is heard.
Part No 1118354 23 Platinum™Series
SECTION 3—COMPRESSOR OPERATION
Compressor
Cylinder/Regulator
Cylinder
Gauge
PULL UP
Assembly
Cylinder
Fillport
Connector
Fillport
Outer Ring
(Sleeve)
Regulator
Cylinder
Cylinder
Gauge
Compressor
Cradle
Connector
Fillport
Regulator Flow Dial
Outlet (Cannula)
Connection
Cylinder Fillport
Power Switch
Outer Ring
(Sleeve)
NOTE: Fillport covers not shown for clarity.
FIGURE 3.3 Connecting the Cylinder to the Compressor
Platinum™Series 24 Part No 1118354
SECTION 3—COMPRESSOR OPERATION
Disconnecting the Cylinder from the Compressor
WARNING
NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur.
NOTE: For this procedure, refer to FIGURE 3.4.
1. Press the compressor power switch to the Off (O) position.
2. Grasp the cylinder/regulator assembly in the area behind the cylinder gauge.
3. With the other hand, grasp the outer ring (sleeve) of the connector fillport and push DOWN.
4. Lift up on the cylinder/regulator assembly to remove from the connector fillport.
Cylinder/Regulator
Assembly
PUSH DOWN
Cylinder
Gauge
Compressor
Connector
Fillport
Cylinder
Fillport
Outer Ring
(Sleeve)
FIGURE 3.4 Disconnecting the Cylinder from the Compressor
Part No 1118354 25 Platinum™Series
SECTION 3—COMPRESSOR OPERATION
5. When the cylinder fillport is disconnected from the connector fillport, release the outer ring (sleeve) of the connector fillport and use two hands to remove the cylinder from the compressor cradle.
6. Place the fillport covers onto the connector and cylinder fillports.
NOTE: The fillport covers should be used whenever cylinders are not being filled.
Turning the Compressor On
NOTE: For this procedure, refer to FIGURE 3.5.
1. Make sure the concentrator is On. Refer to the concentrator Owner’s Manual.
CAUTION
The concentrator flow rate to the patient MUST be set at maximum: 3 L/min. or less for Platinum 5, XL and 5 L/min. for Platinum 10. Otherwise, the oxygen level to the compressor will be <90% and the compressor will not fill the cylinder.
2. Set the concentrator flow rate to the patient to 3 L/min. or less for Platinum 5, XL and 5 L/min. for Platinum 10. Refer to the concentrator Owner’s Manual, Part One.
3. Make sure the cylinder is connected to the compressor. Refer to Connecting the Cylinder to the Compressor page 22.
4. Push the power switch on the control panel to the On position.
5. Examine the control panel. The following sequence of events should occur:
A. 0-3 minutes - The WAIT (YELLOW) light will be lit
while the compressor warms up.
Platinum™Series 26 Part No 1118354
on
SECTION 3—COMPRESSOR OPERATION
B. After three (3) minutes - The FILLING (GREEN) light
will be on while the cylinder is filling.
C. The FULL (GREEN) light will be on when the cylinder
is finished filling. Proceed to STEP 6.
NOTE: The O2 BELOW NORMAL (YELLOW) light may come on if the oxygen level from the concentrator has not yet reached or drops
below 90%. The compressor will begin or resume filling when the oxygen level reaches 90%. If light does not go Off within ten (10) minutes, refer to the troubleshooting section on page.
NOTE: If the ATTENTION (RED) light and audible alarm are on. Refer to Troubleshooting/ Cylinder Fill times on page 31 of the HomeFill operator’s manual, part number 1100873.
6. Remove the full cylinder. Refer to Disconnecting the Cylinder from the Compressor on page 25.
Part No 1118354 27 Platinum™Series
SECTION 3—COMPRESSOR OPERATION
ATTENTION
(RED Light and Audible Alarm)
WAIT
(YELLOW Light)
BELOW
O
2
NORMAL
(YELLOW Light)
O2 BELOW NORMAL
FILLING
(GREEN Light)
FULL
(GREEN Light)
Control
Panel
USE NO OIL
I
O
Power Switch
FIGURE 3.5 Turning the Compressor On
Setting the Flow Knob to the Prescribed Setting
NOTE: For this procedure, refer to FIGURE 3.6.
1. Attach the nasal cannula to the outlet fitting of the cylinder.
2. Turn the flow knob to the L/min. setting prescribed by your physician or therapist.
Platinum™Series 28 Part No 1118354
SECTION 3—COMPRESSOR OPERATION
WARNING
As changing the L/min. setting on the flow knob will affect the dose of oxygen delivered, DO NOT readjust the L/min. setting unless directed by your physician or therapist.
3. To turn the flow of oxygen Off, turn the flow knob counter-clockwise to the Off position.
Outlet Fitting
Flow Knob
Prescribed Setting
FIGURE 3.6 Setting the Flow Knob to the Prescribed Setting
NOTICE
Refer to HomeFill II Compressor Owner’s Manual, 1100873, for detailed information on all warnings, use, care and maintenance of the HomeFill II compressor.
Part No 1118354 29 Platinum™Series
CONCENTRATOR WARRANTY
CONCENTRATOR WARRANTY
NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to original purchaser of our products.This warranty gives you specific legal rights and you may also have other legal rights which may vary from state to state.
Invacare Corporation warrants its Platinum Oxygen Concentrator to be free from defects in materials and workmanship from date of purchase: Platinum 5 and XL for a period of five (5) years, Platinum 10 for a period of three (3) years.The 4-way valve body carries a lifetime warranty from date of purchase. DO NOT OPEN OR ATTEMPT TO SERVICE VALVE BODY. This will void any and all warranty applicable to the valve body. If within such warranty period any such product shall be proven to Invacare Corporation's satisfaction to be defective, such product shall be repaired or replaced, at Invacare Corporation's option. This warranty only applies to the labor for repairs performed by the Invacare Service Department or Invacare Service Centers. It does not apply to the labor performed by the purchaser or user. This warranty does not include normal wear and tear or shipping charges. Invacare Corporation's sole obligation and your exclusive remedy under this warranty shall be limited to such repair or replacement. Routine maintenance items, such as filters, are excluded from this warranty.
For warranty service, please contact Invacare Corporation's Service Department at the toll free number on the back cover.
Upon receiving notice of an alleged defect in a product, Invacare Corporation will issue a serialized return authorization. It shall then be the responsibility of the purchaser to return the entire unit or remove, at purchaser's cost, the defective component part(s) identified, pack the component part(s) in a manner to avoid shipping damage and to ship the component part(s) to either Invacare Corporation's plant or service center as specified by Invacare Corporation in advance. Defective component part(s) must be returned for warranty inspection using the serial number as identification within thirty (30) days of return authorization date. Do not return products to our factory without prior consent. C.O.D. shipments will be refused; please prepay shipping charges.
Platinum™Series 30 Part No 1118354
CONCENTRATOR WARRANTY
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, SOOT OR SMOKE-FILLED ENVIRONMENTS, OR OTHER THAN NORMAL APPLICATION, USE OR SERVICE, OR TO PRODUCTS MODIFIED WITHOUT INVACARE CORPORATION'S EXPRESS WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO, MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS) OR TO PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE CORPORATION OR TO PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE CORPORATION'S CONTROL.
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER, WHETHER EXPRESS OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTIBILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PERSUANT TO THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER.
SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE, OR LIMITATION OF HOW LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE EXCLUSION AND LIMITATION MAY NOT APPLY TO YOU.
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/PROVINCIAL LAWS AND REQUIREMENTS.
Part No 1118354 31 Platinum™Series
Invacare Corporation
USA
One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 Technical Services 800-832-4707
www.invacare.com
Invacare, Yes, you can, the Medallion Design and SensO2 are registered
trademarks of Invacare Corporation. Platinum and HomeFill are trademarks of Invacare Corporation. © 2005 Invacare Corporation
Part No 1118354 Rev D - 09/05 Asm. No. 1106444