Contents
1General...................4
1.1Symbols...............4
1.2IntendedUse...........4
1.3ConditionsforUse......5
1.4ServiceLife............5
1.5Features..............5
2Safety.....................6
2.1GeneralGuidelines.......6
2.2Operating
Information............6
General.............6
PinchPointsand
Positioning...........7
2.3LoadLimits............8
2.4PowerSupply...........8
2.5RadioFrequency
Interference............8
2.6ProductLabeling........9
3Setup.....................11
3.1IncludedItems..........11
3.2InstallingtheHangerBar
BeforeUse............11
3.3BeforeUse............12
4Operation.................13
4.1Indicatorlampsandaudio
signals................13
4.2Operatingthe
Hoist.................13
4.3Chargingthe
Battery...............14
4.4Emergencystopand
loweringstrap..........14
Emergencystop.......15
Emergency
lowering............16
Resettingthe
emergencystop.......17
5LiftingthePatient...........18
5.1SlingInformation........18
5.2AttachingtheSling.......18
5.3LiftingInformation.......19
5.4Liftingtoandfroma
seatedposition.........19
5.5Liftingtoandfromalying
position...............19
6AfterUse..................20
6.1Transportand
Storage...............20
6.2Disposal..............20
7Troubleshooting............22
7.1Troubleshooting
Table................22
8Maintenance...............23
8.1SafeMaintenance........23
8.2MaintenanceandSafety
Inspection.............23
8.3CleaningtheSlingand
Hoist.................24
9TechnicalData.............25
9.1TypicalProduct
Parameters............25
10Warranty..................27
10.1LimitedWarranty—
Europe...............27
10.2LimitedWarranty—
NorthAmerica.........27
Page 4
I-Transia™

1General

1.1Symbols

Signalwordsareusedinthismanualandapplytohazardsorunsafepracticeswhichcould resultinpersonalinjuryorpropertydamage.Seetheinformationbelowfordefinitions ofthesignalwords.
WARNING!
–Warningindicatesapotentiallyhazardoussituationwhich,ifnotavoided,could
resultindeathorseriousinjury.
CAUTION!
–Cautionindicatesapotentiallyhazardoussituationwhich,ifnotavoided,may
resultinpropertydamageorminorinjuryorboth.
IMPORTANT
–Indicatesahazardoussituationthatcouldresultindamagetopropertyifitis
notavoided.
Givesusefultips,recommendationsandinformationforefficient,trouble-freeuse.
EmergencyStop
EmergencyLoweringFunction
Resetemergencystop
Danger—rotatingparts
ThisproductcomplieswithDirective93/42/EECconcerningmedicaldevices.
ThelaunchdateofthisproductisstatedintheECdeclarationofconformity.
TypeBinaccordancewithIEC/EN60601–1
Readthemanualbeforeuse.

1.2IntendedUse

I-Transiaisahoistwhichcoverstheneedforliftingormovingapersonwithdisabilitiesinthe homecaresector,nursinghomes,rehabilitationcentres,andinstitutions.
4
1183377-B
Page 5

1.3ConditionsforUse

TheuseoftheI-Transiaceilinghoistissubjecttothefollowing:
TheI-Transiashouldonlybeusedbytrainedpersonnel.
Thesafeworkingload(SWL),205kg(450lbs)or250kg(550lbs),mustnotbe
exceeded.Referto2.3LoadLimits,page8
.
TheinstructionofferedbyInvacaretoallcustomergroupsinconnectionwiththe
purchaseofaceiling-mountedhoisthasbeenreceived.
Thecaregiverpaysattentiontothewell-beingoftheuserwhenusingthehoist.
Thehoistisusedinrailsystemswhichareinstalled,testedandapprovedaccordingto
DS/EN10535andInvacare’sstipulations.
OnlytechnicianswhohavebeencertifiedbyInvacaremayinstallandtesttherailsystems.
ThehoistisusedwithanInvacarehangerbar.
ThehoistisusedwithanInvacareliftingslingorwithothersuitableslings.Referto
5.1SlingInformation,page18
.

1.4ServiceLife

TheI-Transiahasanexpectedlifetimeof15years,ontheconditionofcorrectuseand correctserviceinspections.RefertotheI-Transiausermanualformoreinformation.

1.5Features

General
InformationpanelontheI-Transiabottomsurface
A
B
C
D
1183377-B5
Typeandmaximumloaddesignation
Green/yellowbatteryindicatorlamp
Strapforactivatingemergencystopand emergencyloweringdevice
Buttonforresettingtheemergencystop
Page 6
I-Transia™

2Safety

2.1GeneralGuidelines

WARNING!
–Readandunderstandtheseinstructions.DONOTusethisproductwithout
readingandunderstandingtheseinstructions.Ifyouneedhelpwiththese instructions,contactahealthcareprofessionaloranInvacaredealer. Otherwiseinjuryordamagemayoccur.
ACCESSORIESWARNING
–InvacareproductscanonlybeusedwithaccessoriesmadebyInvacare.
AccessoriesnotmadebyInvacarehavenotbeentestedbyInvacare. AccessoriesnotmadebyInvacareshouldnotbeused.
Checkallpartsforshippingdamagebeforeusing.Incaseofdamage,DONOTusethe equipment.ContactyourInvacaredealerforfurtherinstructions.
Theinformationcontainedinthisdocumentissubjecttochangewithoutnotice.

2.2OperatingInformation

Thissectionofthemanualcontainsgeneralsafetyinformationaboutyourproduct.For specificsafetyinformation,refertotheappropriatesectionofthemanualandprocedures withinthatsection.

General

WARNING! RiskofFalling
DONOTattemptanytransferwithouttheconsentofthepatient'shealthcare providers.Readthisusermanualandobserveotherpersonnelliftingpatients. Thenpracticeliftingundersupervisionwithacapablepersonactingasapatient. –Usecommonsenseinalllifts.SpecialcareMUSTBEtakenwithpeoplewith
disabilitieswhocannotcooperatewhilebeinglifted.
–Checkthattheslingattachmentsaresecureonthehangerbareachtimethe
slingisremovedandreplaced,toensurethattheslingisproperlyattached beforethepatientisremovedfromastationaryobject.
WARNING!
–ReadtheinstructionscarefullybeforeusingtheI-Transiaandinconnection
withcleaningandserviceofthehoist.
TheI-Transia’smaximumloadmustneverbeexceeded.TheI-Transiamayonlybeusedtoliftormoveaperson.Theredstrapfortheemergencystopandtheemergencyloweringmustbe
adjustedtothecaregiver’sreach,andmustnotberemoved.
–IftheI-Transiafailswhileinuse,stopusingthehoist,transferthepatientusing
adifferentmethodandcontactanInvacaredealerforrepairbyacertified technician.
–TheI-TransiaiscontrolledbyamicroprocessorPCB,whichcanbedamaged
bystaticelectricityiftouchedwithoutthenecessaryprecautions,
61183377-B
Page 7

PinchPointsandPositioning

Safety
WARNING!
TheelectronicsmayonlybeservicedbycertifiedservicetechniciansForsafetyreasonsthesidecovermayonlybedismantledwhentheemergency
stopisactivated.
Thisproductismechanicallyprotectedagainstderailingandjamming.Thehandcontrolisprotectedagainstwatersplashes(IPX4).DONOT
immersethehandcontrolinliquids,evenforashorttime.Avoidanycontact withwaterorotherliquids.
WARNING! RiskofInjury
PinchpointsarepresentonthestrapcarabinerAandfingerscouldbepinched. ThehangerbarBcanmovesuddenlyandcauseinjury.
ALWAYSkeephandsandfingersclearofmovingpartstoavoidinjury.Whenpositioningthehoist,beawareofthepositionofthehangerbarB
andthepatient.Injurycouldoccur.
1183377-B7
Page 8
I-Transia™

2.3LoadLimits

Readthelabelwhichindicatesthemaximumloadlimitsforeachcomponent.
Thecomponent,e.g.hangerbar,liftingsling,etc.labelledwiththelowestloadlimit determinesthemaximumloadlimitfortheentiresystem.
Thismaximumloadlimitmustnotbeexceeded.
Pleasenotethatthemaxloadmaychangewhendifferentcomponentsareused,such ashangerbars,liftingslings,etc.

2.4PowerSupply

I-Transiaisequippedwithabatterythatrequiresregularrecharging.Thepowersupplyfor thetransformerandthechargingstationmustbeconnectedbyacertifiedtechnician.
Thetransformersuppliedmustalwaysbeused.
Refertoformoreinformationaboutthepowersupply.

2.5RadioFrequencyInterference

WARNING!
–MostelectronicequipmentisinfluencedbyRadioFrequencyInterference
(RFI).CAUTIONshouldbeexercisedwithregardtotheuseofportable communicationequipmentintheareaaroundsuchequipment.IfRFIcauses erraticbehavior,stoptheunitusingthehandcontroloremergencystop.DO NOTusetheunitwhiletransmissionisinprogress.
81183377-B
Page 9

2.6ProductLabeling

Hoist
Safety
Onlyoneoftheweightlimitlabelsareonthebottomofthehoist.
A=SerialNumber
Powersupply
1183377-B9
Page 10
I-Transia™
Handcontrol
HangerBar
Thehangerbarlabelislocatedneartheendofthehangerbar,totheinside ofthehooks.
101183377-B
Page 11

3Setup

3.1IncludedItems

1.I-Transiahoist*
2.Handcontrol
3.Transformer
4.Chargingstation
5.Sidecovers
6.Hangerbar
Setup
*Thehoistincludestheliftingstrap,carabinerandtrolley.

3.2InstallingtheHangerBarBeforeUse

WARNING! RiskofInjury
PinchpointsarepresentonthestrapcarabinerBandfingerscouldbepinched. ThehangerbarAcanmovesuddenlyandcauseinjury.
ALWAYSkeephandsandfingersclearofmovingpartstoavoidinjury.Whenpositioningthehoist,beawareofthepositionofthehangerbarB
andthepatient.Injurycouldoccur.
1183377-B
11
Page 12
I-Transia™
ThehangerbarAcanbeinstalledtotheliftingstrapcarabinerBwithouttheuseoftools.
1.PullandholdtheretainerCuptowardthepinD.
2.MovethepinandretainerdowntoopenthecarabinerB.
3.InstallthehangerbarAontothecarabiner.
4.Releasetheretainerandpin.
5.Beforeuse,ensurethehangerbarisinstalledcompletelyontothecarabinerandthe retainerisinplace,asshowninthefigure.

3.3BeforeUse

I-Transiaisdeliveredwiththeemergencystopactivatedinordertoensurethatthebatteryis notbeingdischargedduringlong-termstorage(morethansixmonths).Resettheemergency stopbypressingthe“RESET”buttononthebottomofthehoist.RefertoResettingthe emergencystop,page17.
12
1183377-B
Page 13

4Operation

4.1Indicatorlampsandaudiosignals

Operation
Status
Off-standbyOff
AllOKGreen
LowbatteryYellow
FaultonhoistYellowBeepsat
Batterycritical low
OverloadGreen
Indicator
lamps
button activation
Yellow
Beepsat button activation

4.2OperatingtheHoist

Audio
signals
PossibleI-TransiaFunctions
UpDownEmergency
Lowering
üüü
üüü
ü
üü
üü
TheI-Transiaisswitchedonautomaticallywhenabuttononthehandcontrolis pressed.
TheI-Transiawillswitchoffautomaticallyafterapproximately8minutesof non-operation.
1.PresstheUPbuttonAonthehandcontrolBtoliftthepatient.
2.PresstheDOWNbuttonCtolowerthepatient.
Theremustbealoadontheliftingstrapcorrespondingtotheweightofthehanger barbeforeI-Transia’sloweringfunctionwilloperate.
1183377-B13
Page 14
I-Transia™

4.3ChargingtheBattery

TheI-Transiabatterychargeswhenthehandcontrolisplacedinthecharging station.
Theindicatorlamponthebottomofthehoistturnsyellowifthechargestatus becomeslow.Finishthecurrentliftinprogressandchargethebatterieswhen finished.
Thehoistischargedwhentheindicatorlamponthebottomofthehoistisgreen.
AlwaysleavethehandcontrolinthechargingstationwhentheI-Transiaisnotin use.ThisguaranteesI-Transiafunctionalityandmaintainsthebatterytoensurea longlifetime.
Thetransformermustbeconnectedandswitchedonbeforechargingcantake place.Agreenindicatorlamponthetransformerindicatesitisconnectedand switchedon.
1.TiltthehandcontrolAandplaceitonthetwopinsBonthechargingstationC.
2.PushtherubberhandleDintotheopeningonthechargingstation.
Anaudible“beep”indicatesthatthehandcontrolisplacedcorrectly.

4.4Emergencystopandloweringstrap

WARNING!
–Iftheemergencystoporloweringstrapmustbeused,thereisaproblem
withthesystem.ContactyourInvacaredealerforservicebeforeusingthe systemagain.
14
1183377-B
Page 15
TheredstrapAontheundersideofthehoistBhasthefollowingfunctions:
Emergencystop
Emergencylowering

Emergencystop

Operation
PullandreleasetheredstrapAtoactivatetheemergencystopfunctionifthehoistBdoes notstoporreacttothehandcontrolduringuse.
Lifting/loweringfunctionsaredeactivated.Emergencyloweringwillnotstop.
Thehoistwillnotfunctionuntiltheemergencystopisreset.
ThegreenlampCwillturnoff.
1183377-B15
Page 16
I-Transia™

Emergencylowering

Theemergencyloweringfunctionisusedtolowerthepatientifthehoiststopsworking.
1.PullandholdtheredstrapAtolowerthepatient.
2.Releasetheredstraptostoploweringthepatient.
Theemergencystopfeatureisactivatedwhenthestrapisreleased.
3.Ifthehoistsystemneedsservice,contactyourInvacaredealerforservice.
4.Resettheemergencystoptoresumeusingthehoist.RefertoResettingtheemergency stop,page17.
161183377-B
Page 17

Resettingtheemergencystop

Operation
TheyellowresetbuttonApopsoutwhentheredstrapBhasbeenpulled.Thehoistwill notfunctionuntiltheemergencystopisreset.
1.Presstheyellowbuttononthebottomofthehoisttoresettheemergencystopor loweringfeature.
2.Activatethehandcontroltwicetoresumeuseofthehoist.
1183377-B17
Page 18
I-Transia™

5LiftingthePatient

5.1SlingInformation

WARNING!
Invacareshallnotbeliablefordamageorinjurythatmayoccurasaresultof usingslingsmadebyothermanufacturers. Invacareshallnotbeliablefordamageorinjuryduetoincorrectuseofthesling orinadequateattentionbythecaregiveroruser. –Usetheslingthatisrecommendedbytheindividual’shealthcareteamforthe
comfortandsafetyoftheindividualthatisbeinglifted.Ifthereisanydoubt abouttheselectionoruseofasling,pleasecontactyourInvacaredealer.
Aslingwithfourtosixliftingstrapsdesignedformountingonhooksshouldbeusedwhen usingaInvacarehangerbar.
Refertotheslingusermanualformoredetailsabouttheslings.
UKCustomers—PleaseseetheInvacarebulletinconcerningslings.Acopycanbeobtained fromInvacareUK—01656776222oruk@invacare.com.

5.2AttachingtheSling

CAUTION!
Becarefulwhenattachingtheliftingslingonthehooks. –Checkthatthestrapshavebeenpositionedcompletelyontothehooksofthe
hangerbar.
–Whenpressingtheupbuttontolifttheuser,checkagainthatallthestraps
remaincorrectlyplacedinthehangerbar’shooks.
1.PlacetheuppermostsetofstrapsfromthebackoftheslingonthehooksAonthe hangerbarB.
2.Placethelowestsetofstrapsfromthelegsoftheslingonthehooksonthehangerbar.
Refertotheslingusermanualforproperpositioningofthepersonintheslingand properpositioningofthestrapsonthehangerbar.
181183377-B
Page 19
LiftingthePatient

5.3LiftingInformation

WARNING!
Onlypersonnelwhohavereceivedtrainingregardingtheuseofliftingequipment andfittingofslingsshouldusethehoist.
Planthemove.Avoidleavingtheuserintheslingunattended.TheI-Transialiftsquickly.Beforelifting,checkthattheuseriscompletely
freeofhis/hersurroundings.
–Theuser’shead,arms,handsandfeetmustnotbeindangerofbecoming
trapped.
Becarefulwithanytubesandwiresthatareattachedtotheuser.Checkthattheredemergencystrap,handcontrolandhandcontrolcableare
freeofthehangerbar,userandotherobjectsbeforethehoistisactivated andmovedupordown.
–Whenthehoistisusedcorrectly,theusershouldonlybeliftedtotheextent
thatshe/heisclearofthesurfaceandshouldbemovedatthisheight.

5.4Liftingtoandfromaseatedposition

Followthisprocedurewhenliftingapersonfromaseatedposition(wheelchair,commode, etc.).
1.MovetheI-Transiatowardsthepersontobelifted.
2.Positionthehangerbaratthesameheightastheperson’schestandnocloserthan mid-thigh.
3.Positionthehangerbarsoitisparalleltotheperson’sshoulders.
4.Positionthepersoninthesling.Refertotheslingusermanual.
5.Attachtheslingtothehangerbar.Referto5.2AttachingtheSling,page18.

5.5Liftingtoandfromalyingposition

Followthisprocedurewhenliftingapersonfromalyingposition(inbed).
1.MovetheI-Transiatowardsthepersontobelifted.
2.Positionthehangerbaroverthecenteroftheperson.
3.Positionthehangerbarsoitisparalleltotheperson’sshoulders.
4.Positionthepersoninthesling.Refertotheslingusermanual.
5.Elevatetheheadofthebed,sothattheuserissittingup.
6.Attachtheslingtothehangerbar.Referto5.2AttachingtheSling,page18.
1183377-B19
Page 20
I-Transia™

6AfterUse

6.1TransportandStorage

InvacarerecommendsthattheI-Transiaisalwaystransportedandstoredintheoriginal packaging.
Forlongtermstorage,theemergencystopmustbeactivated.RefertoEmergencystop,page
.Thiswillpreventthedischargeofthebattery.
15
TheI-Transiashouldbestoredasindicatedbythepackagingsymbolsinthetable.
Packaging
Symbol
Temperaturesbetween–10and+40°C(14and104°F)
RelativeAirHumiditybetween30and70%
AirPressurebetween700and1060hPa
ThisSideUp
StorageCondition

6.2Disposal

WARNING! EnvironmentalHazard
Thisproducthasbeensuppliedfromanenvironmentallyawaremanufacturer thatcomplieswiththeWasteElectricalandElectronicEquipment(WEEE) Directive2002/96/CE. ThisdevicecontainsNiMHbatteries. Thisproductmaycontainsubstancesthatcouldbeharmfultotheenvironmentif disposedofinplaces(landfills)thatarenotappropriateaccordingtolegislation. –DONOTdisposeofbatteriesinnormalhouseholdwaste.TheyMUST
betakentoaproperdisposalsite.Contactyourlocalwastemanagement companyforinformation.
–Pleasebeenvironmentallyresponsibleandrecyclethisproductthroughyour
recyclingfacilityatitsendoflife.
201183377-B
Page 21
EnvironmentalPolicyStatement
Invacareiscontinuouslyworkingtowardsensuringthatthecompany'simpactonthe environment,locallyandglobally,isreducedtoaminimum.ItisInvacare’sgoalto:
Complywiththecurrentenvironmentallegislation(e.g.WEEEandREACHdirectives)
Ensurethatwe,atthewidestpossiblerange,useRoHScompliantmaterialsand
components
Ensurethatourproductsdonothaveanunnecessarynegativeimpactonthe
environmentregardinguse,recirculationordisposal
Ensurethatourproductscontributetoapositiveworkingenvironmentintheplaces
theyareutilised
InspectionsaremadeannuallybytheDepartmentforNatureandEnvironmentfromthe MunicipalityofAarhususingtheDanishEnvironmentalProtectionAct,section42asa reference.
AfterUse
1183377-B
21
Page 22
I-Transia™

7Troubleshooting

7.1TroubleshootingTable

SYMPTOMSPROBLEMSOLUTION
I-Transiahoistdoesnot respondtothehandcontrol’s keys.
Trolleydoesnottraverse.Railsrequireserviceor
Hoistwillnotcharge
Noindicatorlightonhoist
Theemergencyloweringcord doesnotturnhoistoff
Theemergencyloweringcord doesnotlowerthehoist.
Emergencystopisactivated.
Systemdoesnothavepower orbatteryisdischarged
cleaning. Systemdoesnothavepower
Systemdoesnothavepower orbatteryisdischarged.
Hoistrequiresservice.
Hoistrequiresservice.
Checkthattheemergency stopisnotactivated.
CheckthattheI-Transiahas powersupplyandthatthe batteryisrecharged.
Checkthatthetransformer isswitchedonandconnected tothechargingstation.
Placethehandcontrolinthe chargingstationandrecharge theI-Transia.
ContactanInvacaredealer forservice.
Placethehandcontrolinthe chargingstationandrecharge theI-Transia
Checkthatthepowersupply isswitchedonandconnected tothechargingstation.
ContactanInvacaredealer forservice.
CheckthattheI-Transiahas powersupplyandthatthe batteryisrecharged.
Checkthatthepowersupply isswitchedonandconnected tothechargingstation.
Placethehandcontrolinthe chargingstationandrecharge theI-Transia.
ContactanInvacaredealer forservice.
ContactanInvacaredealer forservice.
Ifproblemsarenotremediedbythesuggestedmeans,pleasecontactyour Invacaredealer.
22
1183377-B
Page 23
Maintenance

8Maintenance

8.1SafeMaintenance

WARNING!
Regularmaintenanceofpatientliftsandaccessoriesisnecessarytoassure properoperation. –MaintenanceMUSTbeperformedonlybyacertifiedtechnician.

8.2MaintenanceandSafetyInspection

ServiceInterval
Atnormaldailyoperation,aservicecheck-upshouldtakeplaceeveryyear,accordingto theSafetyInspectionChecklistintheServiceManual.Whenperformingannualorregular maintenance,allpartsdesignedtocarryloadmustbe,asaminimum,testedwithload.All safetyfeaturesmustbecheckedaccordingtoENISO10535:2006AnnexB.Thisinspection mustbeperformedbyacertifiedtechnician.
LOLERStatement
TheUKHealthandSafetyExecutive’sLiftingOperationsandLiftingEquipmentRegulations 1998,requireanyequipmentthatisusedintheworkplacetoliftaloadbesubjectto safetyinspectiononasixmonthlybasis.PleaserefertotheHSEwebsiteforguidance www.hse.gov.uk.
ThepersonresponsiblefortheequipmentmustensureadherencetoLOLERregulations.
GeneralMaintenance
Followthemaintenanceproceduresdescribedinthismanualtokeepyourpatient liftincontinuousservice.
Thehoistisdesignedtoprovideasafe,efficientandsatisfactoryservicewithminimumcare andmaintenance.
Itisimportanttoinspectslings,hangerbar,carabinerandliftingstrapforsignsofcracking, fraying,deformationordeterioration.Ifdamageisfound,DONOTuse.Contactyour InvacaredealerIMMEDIATELY.
Ifyouquestionthesafetyofanypartofthelift,contactyourInvacaredealerimmediatelyand advisehim/herofyourproblem.
Replacementofpartsmustonlybeperformedbyacertifiedtechnician.
1183377-B23
Page 24
I-Transia™
DailyInspections
Thehoistshouldbecheckedeachtimeitisused.Performthefollowingchecksinadditionto thoseperformedintheannualserviceinspection.Ifyouquestionthesafetyofanypartofthe lift,DONOTuse.ContactyourInvacaredealerIMMEDIATELY.
qVisuallyinspectthesling,hangerbarandliftingstrap.Checkallpartsfordamageor
wear.Ifdamageisfound,DONOTuse.ContactyourInvacaredealerIMMEDIATELY.
qChecktheemergencystopandemergencyloweringfunction.Checkallpartsfor
externaldamageorwear.Ifdamageisfound,DONOTuse.ContactyourInvacare dealerIMMEDIATELY.
qCheckthatallhardwareandattachmentpointsfordamageorwear.Checkallpartsfor
externaldamageorwear.Ifdamageisfound,DONOTuse.ContactyourInvacare dealerIMMEDIATELY.
qVerifythatthehandcontrolisfunctional. qCheckthebatterycharger. qChargethebatteryeverydaytheliftisused. qCheckthecarabinerforwearatthebottomcontactpointA.Donotusethehoistifthe
carabinermeasureslessthan6mm(1/4in)atthispoint.
SafetyInspectionChecklist
TheannualsafetyinspectionchecklistisavailableintheI-Transiaservicemanual.

8.3CleaningtheSlingandHoist

CleaningtheSling
Refertothewashinginstructionsontheslingandtotheslingmanualforcleaningdetails.
CleaningandDisinfectingtheHoist
Topreventcross-infection,thepartsofthehoistthatcomeincontactwithcaregiversand patientsmustbecleanedanddisinfected.
Asoftcloth,dampenedwithwaterandasmallamountofdishliquid,isallthatisneeded tocleanthehoist,liftingstrapandcarabiner.Thehoistcanbecleanedwithnon-abrasive cleaners.
Neveruseacids,alkalineorsolventsforcleaningthehoist.Drythehoistcarefullyafter cleaning.
Thehoist,handcontrolandmountingpartscanbedamagedifthehoistiscleanedanyother waythanstatedabove.
Thehoistmustbewipedwithamoistened,firmlywrungclothwithordinaryhousehold disinfectants.Onlyusedisinfectiondetergentsapprovedbythefacility.
24
1183377-B
Page 25

9TechnicalData

9.1TypicalProductParameters

Classification
ClassIIequipment
Flammability
TechnicalData
Non-permanentinstallationwithoutprotectiveground.
MobileequipmentisclassIIequipment(markedwith double-encasedsymbol)andcanbeconnectedtothemains directbytheuser.
Equipmentisdisconnectedfromthesupplymainsby detachingthemainsplugfromthewalloutlet.
Flammability
Thisequipmentisnotsuitableforuseinthepresenceof flammablemixtures.
LiquidIngress
Liftingmodule Handcontrol
Chargingstation Powersupply
IP44 IP44
IP20 IP20
Functions
Liftingcapacity,SafeWorking Load(SWL) Operation
Soundlevel
205kg(450lb),250kg(550lb)
Handcontrol 52dB(A)
LiftingSpeed
85kg(187lb)load40mm/sec.(1.6in/sec) 150kg(330lb)load40mm/sec.(1.6in/sec) SafeWorkingLoad,SWL40mm/sec.(1.6in/sec) Max.5kg(11lb)load80mm/sec.(3.2in/sec)
Weightandmaterials
SafeWorkingLoad(SWL)205kg(450lb),250kg(550lb) Hoistweight Covers,topandbottom
8.0kg(17.6lb) Impact-resistantUL94V-0flameretardantrecyclableplastic
1183377-B25
Page 26
I-Transia™
Dimensions
A B C D E,min F,min G
Depthofhoist194mm(7.6in)
580mm(22.8in) 350mm(13.8in) 156mm(6.1in) 184mm(7.2in) 83mm(3.3in) 415mm(16.3in) 2500mm(98.4in)
Safety
Emergencystop Emergencyloweringdevice Controlofliftingstrap Cut-offangle45°alongtherail,10°acrosstherail.
Yes Yes,electrical Yes
Electronics
On/off Overloadprotection Lowbatteryprotection Powersupply Supplyvoltage,transformer Battery Continuousoperationwith shorttimeloadingwith3 hourswithoutrecharging
Maxnumberofliftsinseries with85kg(187lb)
Maxnumberofliftsinseries withSWL[205kg(450lb), 250kg(550lb)]
Maxchargingtimeat25°C (77°F) Operatingtemperature10°C—35°C(50°F—95°F)
Automaticwhenused.Softstar/stop Automatic Automatic 36VDC,0.83A 100–240VAC,47–63Hz 24VNiMH,2.8Ah
10/90%(2min.operation18min.pause)
55/1000mm
28/1000mm
3hrs
261183377-B
Page 27
Warranty

10Warranty

10.1LimitedWarranty—Europe

Termsandconditionsofthewarrantyarepartofthegeneraltermsandconditionsparticulartotheindividual countriesinwhichthisproductissold. InordertocomplywithHealth&SafetyregulationsandtoensurethatInvacareemployeesarenotexposed totheriskofinfection,wemustinsistthatallproductsreturnedforrepairarethoroughlycleaned.Any productsenttoInvacarewhichappearstobecontaminatedwillberejectedanddisposedofimmediatelyat thecustomer’sexpense.

10.2LimitedWarranty—NorthAmerica

PLEASENOTE:THEWARRANTYBELOWHASBEENDRAFTEDTOCOMPLY WITHFEDERALLAWAPPLICABLETOPRODUCTSMANUFACTUREDAFTER JULY4,1975.
Thiswarrantyisextendedonlytotheoriginalpurchaser/userofourproducts. Thiswarrantygivesyouspecificlegalrightsandyoumayalsohaveotherlegalrightswhichvaryfromstateto state. Invacarewarrantstheproductsmanufacturedtobefreefromdefectsinmaterialsandworkmanshipforaperiod ofoneyearfromthedateofpurchase.Ifwithinsuchwarrantyperiodanysuchproductshallbeproventobe defective,suchproductshallberepairedorreplaced,atInvacare’soption.Thiswarrantydoesnotincludeany labororshippingchargesincurredinreplacementpartinstallationorrepairofanysuchproduct.Invacare’ssole obligationandyourexclusiveremedyunderthiswarrantyshallbelimitedtosuchrepairand/orreplacement. Forwarrantyservice,pleasecontactthedealerfromwhomyoupurchasedyourInvacareproduct.Intheevent youdonotreceivesatisfactorywarrantyservice,pleasewritedirectlytoInvacareattheaddressontheback cover,providedealer’sname,address,dateofpurchase,indicatenatureofthedefect. InvacareCorporationwillissueaserializedreturnauthorization.ThedefectiveunitorpartsMUSTbereturned forwarrantyinspectionusingtheserialnumber,whenapplicableasidentificationwithin30daysofreturn authorizationdate.Donotreturnproductstoourfactorywithoutourpriorconsent.C.O.D.shipmentswillbe refused;pleaseprepayshippingcharges. LIMITATIONSANDEXCLUSIONS:THEFOREGOINGWARRANTYSHALLNOTAPPLYTOSERIAL NUMBEREDPRODUCTSIFTHESERIALNUMBERHASBEENREMOVEDORDEFACED,PRODUCTS SUBJECTEDTONEGLIGENCE,ACCIDENT,IMPROPEROPERATION,MAINTENANCEORSTORAGE, PRODUCTSMODIFIEDWITHOUTINVACARE’SEXPRESSWRITTENCONSENT(INCLUDING,BUTNOT LIMITEDTO,MODIFICATIONTHROUGHTHEUSEOFUNAUTHORIZEDPARTSORATTACHMENTS; PRODUCTSDAMAGEDBYREASONOFREPAIRSMADETOANYCOMPONENTWITHOUTTHE SPECIFICCONSENTOFINVACARE,ORTOAPRODUCTDAMAGEDBYCIRCUMSTANCESBEYOND INVACARE’SCONTROL,ANDSUCHEVALUATIONWILLBESOLELYDETERMINEDBYINVACARE.THE WARRANTYSHALLNOTAPPLYTOPROBLEMSARISINGFROMNORMALWEARORFAILURETO ADHERETOTHEINSTRUCTIONSINTHISMANUAL. THEFOREGOINGWARRANTYISEXCLUSIVEANDINLIEUOFANYOTHEREXPRESSWARRANTIES. IMPLIEDWARRANTIES,IFANY,INCLUDINGTHEIMPLIEDWARRANTIESOFMERCHANTABILITY ANDFITNESSFORAPARTICULARPURPOSE,SHALLNOTEXTENDBEYONDTHEDURATIONOF THEEXPRESSEDWARRANTYPROVIDEDHEREINANDTHEREMEDYFORVIOLATIONSOFANY IMPLIEDWARRANTYSHALLBELIMITEDTOREPAIRORREPLACEMENTOFTHEDEFECTIVEPRODUCT PURSUANTTOTHETERMSCONTAINEDHEREIN.INVACARESHALLNOTBELIABLEFORANY CONSEQUENTIALORINCIDENTALDAMAGESWHATSOEVER. SOMESTATESDONOTALLOWEXCLUSIONORLIMITATIONOFINCIDENTALORCONSEQUENTIAL DAMAGE,ORLIMITATIONONHOWLONGANIMPLIEDWARRANTYLASTS,SOTHEABOVE EXCLUSIONSANDLIMITATIONSMAYNOTAPPLYTOYOU. THISWARRANTYSHALLBEEXTENDEDTOCOMPLYWITHSTATEORPROVINCIALLAWSAND REQUIREMENTS.
1183377-B27
Page 28
Notes
Page 29
Notes
Page 30
Notes
Page 31
UnitedKingdom
Ireland
AustraliaInvacareAustraliaPtyLtd.•1LentonPlace•NorthRocks•NSW
NewZealand
France
Switzerland
Deutschland
Portugal
España
Belgium&LuxemburgInvacarenv•Autobaan22•B-8210Loppem•Tel:(32)(0)50
Nederland
Norge
Österreich
USA
Canada
InvacareLimited•PencoedTechnologyPark,Pencoed,Bridgend CF355AQ•Tel:(44)(0)1656776222•Fax:(44)(0)1656776 220•www.invacare.co.uk•UK@invacare.com
InvacareIrelandLtd•Unit5SeatownBusinessCampus•Seatown Road•Swords•CountyDublin–Ireland•Tel:(353)18107084• Fax:(353)18107085•www.invacare.ie•ireland@invacare.com
2151•Australia•Tel:(61)288395333•Fax:(61)288395353• www.invacare.com.au•sales@invacare.com.au
InvacareNewZealand•4WestfieldPlace•Mt.Wellington• Auckland•NewZealand•Tel:(64)99173939•Fax:(64)9917 3957•www.invacare.co.nz•sales@invacare.co.nz
InvacarePoirierSAS•RoutedeStRoch•F-37230Fondettes•Tel: (33)(0)247626466•Fax:(33)(0)247421224•www.invacare.fr
•contactfr@invacare.com
InvacareAG•Benkenstrasse260•CH-4108Witterswil•Tel.:(41) (0)614877080•Fax.:(41)(0)614877081•www.invacare.ch• switzerland@invacare.com
InvacareGmbH•Alemannenstraße10•D-88316Isny•Tel:(49) (0)75627000•Fax:(49)(0)756270066•www.invacare.de• kontakt@invacare.com
InvacareLda•RuaEstradaVelha•949•P-4465-784LeçadoBalio
•Tel:(351)(0)225105946/47•Fax:(351)(0)225105739• www.invacare.pt•portugal@invacare.com
InvacareSA•c/Arenys/n•PolígonIndustrialdeCelrà•E-17460 Celrà(Girona)Tel:(34)(0)972493200•Fax:(34)(0)972493220
•www.invacare.es•contactsp@invacare.com
831010•Fax:(32)(0)50831011•belgium@invacare.com• www.invacare.be
InvacareBV•Celsiusstraat46•NL-6716BZEde•Tel:(31)(0)318 695757•Fax:(31)(0)318695758•nederland@invacare.com• www.invacare.nl•csede@invacare.com
InvacareAS•Grensesvingen9•Postboks6230•Etterstad• N-0603Oslo•Tel:(47)(0)22579500•Fax:(47)(0)22579501• www.invacare.no•norway@invacare.com•island@invacare.com
InvacareAustriaGmbH•HerzogOdilostrasse101•A-5310 Mondsee•Tel:(43)623255350•Fax:(43)623255354• www.invacare.at•info@invacare-austria.com
InvacareCorporationWorldwideHeadquarters•OneInvacare Way•Elyria,Ohio44036•Tel:(800)3336900•Fax:(877)619 7996•ConsumerCare@invacare.com•www.invacare.com
InvacareCanada•570MathesonBlvd.East,Unit8•Mississauga, OntarioL4Z4G4•Tel:(800)6685324•Fax:(800)6685478• www.invacare.ca•infocanada@invacare.com
Page 32
InvacareCorporation
USA
OneInvacareWay
Elyria,OhioUSA
44036–2125
800–333–6900
www.invacare.com
Manufacturer
V.GuldmannA/S
GrahamBellsVej21–23A
DK-8200AarhusN,Denmark
EURepresentative
InvacareLimited
PencoedTechnologyPark
Pencoed,BridgendCF355AQ
UnitedKingdom
1183377-B2013-06-26
*1183377B*
MakingLife’s
ExperiencesPossible™
Loading...