Invacare HF2PCM2KIT User Manual

HomeFill

Oxygen Conserving

Cylinder

HF2PCM2KIT

Dealer: This manual MUST be given to the end user. User:
Before using this cylinder, read this
manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
WARNING
WARNING
DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT A HEALTHCARE PROFESSIONAL, DEALER OR A QUALIFIED TECHNICIAN BEFORE ATTEMPTING TO USE THIS EQUIPMENT - OTHERWISE INJURY OR DAMAGE MAY RESULT.
WARNING

Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.

DANGER
Users MUST NOT SMOKE while using this device. Keep all matches, lighted cigarettes or other sources of ignition out of the room in which this product is located. Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result in severe fire, property damage and cause physical injury or death.

NOTICE

The information contained in this document is subject to change without notice.

NOTE: Updated versions of this manual are available on www.invacare.com.

Oxygen Conserving Cylinder 2 Part No 1143247
TABLE OF CONTENTS
TABLE OF CONTENTS
SPECIAL NOTES .............................................. 4
RECEIVING INSPECTION .................................. 4
INTENDED USE ................................................ 4
SECTION 1—GENERAL GUIDELINES .................. 5
SECTION 2—FEATURES .................................... 8
SECTION 3—TYPICAL PRODUCT
PARAMETERS ................................................. 9
SECTION 4—INSPECTION AND COMPRESSOR
INTERFACE ................................................... 10
Cylinder Prefill Inspection...............................................................10
Connecting/Disconnecting the Cylinder to/from the
Compressor........................................................................................13
Changing the Carrying Position of the Cylinder........................ 18
Cleaning the Cylinder Bag...............................................................19
SECTION 5—OPERATING INSTRUCTIONS ....... 21
SECTION 6—MAINTENANCE .......................... 23
SECTION 7—CYLINDER FILL TIMES ................. 24
Troubleshooting ................................................................................24
LIMITED WARRANTY .................................... 26
Part No 1143247 3 Oxygen Conserving Cylinder

SPECIAL NOTES

SPECIAL NOTES

Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words.

SIGNAL WORD MEANING

DANGER Danger indicates an imminently hazardous
situation which, if not avoided, will result in death or serious injury.
WARNING Warning indicates a potentially hazardous
situation which, if not avoided, could result in death or serious injury.
CAUTION Caution indicates a potentially hazardous
situation which, if not avoided, may result in property damage.

RECEIVING INSPECTION

Remove product from package and inspect for damage. If there is any damage, DO NOT use. Contact your dealer/ Home Care equipment provider.

INTENDED USE

To store, regulate, and deliver supplemental oxygen to patients who may have difficulty extracting oxygen from the air they breathe. It is for patients who would normally receive the oxygen via a nasal cannula. The device delivers USP 93% oxygen at a number of different flow/pulse settings. It is intended to be used as an oxygen saving device that may reduce the drying out of the airways.
Oxygen Conserving Cylinder 4 Part No 1143247
SECTION 1—GENERAL GUIDELINES
SECTION 1—GENERAL GUIDELINES
WARNING
SECTION 1 - IMPORTANT SAFEGUARDS contains important information for the safe operation and use of this product. DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as Owner’s Manuals, Service Manuals or Instruction Sheets supplied with this product or optional equipment. If you are unable to understand the Warnings, Cautions or Instructions, contact a health care professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.
“Caution: Federal law restricts this device to sale by or on the order of a physician or any other practitioner licensed by the law of the state in which he/she practices to use or order the use of this device.”
WARNING
Only personnel instructed and trained in the use of the HomeFill oxygen conserving cylinder should operate this product.
Product contains magnetic, ferrous material that may affect the results of an MRI.
Be sure all connections are tight and leak free.
DO NOT use a liquid leak detector to test for leaks.
DO NOT sterilize the cylinder.
DO NOT gas sterilize with ethylene oxide.
DO NOT clean with aromatic hydrocarbons.
Part No 1143247 5 Oxygen Conserving Cylinder
SECTION 1—GENERAL GUIDELINES
WARNING
DO NOT attempt to repair product.
DO NOT immerse product in any kind of liquid.
Store product in a clean area when not in use.
Only use Invacare Corporation HF2PCM2 carrying bag.
NEVER smoke in an area where oxygen is being administered.
ALWAYS confirm prescribed dose before administering to patient and monitor on a frequent basis.
DO NOT use if dirt or contaminants are present on or around cylinder, valve, product or connecting devices.
NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere.
ALWAYS follow ANSI and CGA standards for Medical Gas Products and Flowmeters (E-7) and Oxygen Handling (G-4).
NO OXYGEN is delivered when the pointer is aligned with “Off”.
DO NOT block the outlet fitting or kink the cannula when the product is in use, this may damage the unit.
The use of this device is limited to the oxygen patient. Cylinders MUST be used only by the oxygen patient and are not to be distributed to any other individual for any purpose.
This product is NOT to be used by patients who breathe through their mouths.
DO NOT use oils, greases, organic lubricants or any combustible materials on or near this product. Wash and dry hands properly prior to usage.
DO NOT use a humidifier with this product.
Oxygen Conserving Cylinder 6 Part No 1143247
SECTION 1—GENERAL GUIDELINES
WARNING
DO NOT allow cylinders to tip or fall. Secure gas cylinders so they cannot fall. For storage, ensure cylinder is ALWAYS in upright position.
DO NOT store cylinders near sources of heat or flame.
This product is equipped with a relief valve. If you hear a loud hissing or popping sound coming from the product, discontinue use of product, turn the flow selector to the “Off” position, and contact your dealer/Home Care equipment provider.
DO NOT use while sleeping without consulting your healthcare provider.
This product is designed to operate with a standard single lumen, adult oxygen nasal cannula with a maximum length of seven feet.
The cannula is for single patient use only.
Cannula MUST be installed on outlet connection prior to use.
DO NOT place the unit under clothing while in use. When this product is in use a small amount of oxygen is vented. Wearing the unit under clothing may saturate fabrics with oxygen and cause them to burn rapidly if exposed to sparks or flame. It may take several hours for oxygen levels in fabrics to return to normal.
Keep the oxygen tubing and cylinder away from heated or hot surfaces, including space heaters, blankets, stoves and similar electrical appliances.
Part No 1143247 7 Oxygen Conserving Cylinder
SECTION 2—FEATURES
SECTION 2—FEATURES
DANGER
This product is not intended as a life-sustaining or life-supporting device.
Outlet
Connection
Cannula
Indicator
Flow Dial
Cylinder
Gauge
Pointer
Cylinder Fill
Port
FIGURE 2.1 Product Diagram
Oxygen Conserving Cylinder 8 Part No 1143247
SECTION 3—TYPICAL PRODUCT PARAMETERS
SECTION 3—TYPICAL PRODUCT PARAMETERS
CONSERVER S/N RANGE: 75,000+
INLET PRESSURE RANGE: 300 - 2100 psig
INTERNAL REGULATED PRESSURE: 50 psig
DIMENSIONS:
HF2PCM2KIT: Weight 2.0 lbs (.90 kg)
HF2PCM2KIT: Overall Length 8.9 in (22.2 cm)
HF2PCM2KIT: Diameter 3.2 in (8.13 cm)
PULSE SETTINGS: 1,2,3,4, & 5
FLOW SETTINGS: 2 LPM Continuous
SAVINGS RATIO: Up to 5.7:1
CANNULA REQUIREMENT:
OPERATING CONDITIONS TEMPERATURE:
ALTITUDE: 0 to 10,000 ft (0 to 3071 m)
STORAGE CONDITIONS TEMPERATURE:
MAXIMUM HUMIDITY: 95%
4 ft recommended (MS0204)
(7 ft long max) std. adult
single lumen oxygen nasal
cannula.
35°F (1.7°C) to 105°F
(40.6°C)
-40°F (-40°C) to 140°F (60°C)
Part No 1143247 9 Oxygen Conserving Cylinder