Instrumentarium Dental Focus User Manual

ENGLISH
FOCUS
User Manual
51774-IMG rev. 13
Copyright Code: 51774-IMG rev 13 Date: 2 November 2015
Copyright © 11/2015 by Instrumentarium Dental, PaloDEx Group Oy. All rights reserved.
FOCUS™ and CLINIVIEW™ are either a registered trademark or trademark of Instrumentarium Dental in the United States and/or other countries.
All other trademarks are property of their respective owners.
Documentation, trademark and the software are copyrighted with all rights reserved. Under the copyright laws the documentation may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine readable form in whole or part, without the prior written permission of Instrumentarium Dental.
The original language of this manual is English.
Instrumentarium Dental reserves the right to make changes in specification and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your Instrumentarium Dental representative for the most current information.
Manufactured by Instrumentarium Dental, PaloDEx Group Oy
Nahkelantie 160 FI-04300 Tuusula FINLAND
Tel. +358 10 270 2000 www.instrumentariumdental.com
For service, contact your local distributor.
Table of Contents
1 Introduction.................................................................................................................. 1
1.1 FOCUS™ .............................................................................................................. 1
1.2 Intended use ......................................................................................................... 1
1.3 User profile............................................................................................................1
1.4 Symbols that may appear on the unit.................................................................... 2
1.5 Type and version................................................................................................... 3
1.6 Labels on FOCUS .................................................................................................4
1.7 Configurations ....................................................................................................... 5
1.8 Radiation protection guidelines ............................................................................. 7
1.9 Manufacturer's liability...........................................................................................8
1.10 Disposal ................................................................................................................ 9
2 Unit description .........................................................................................................11
2.1 Main parts ........................................................................................................... 11
2.2 Cones..................................................................................................................12
2.3 Symbols on remote control ................................................................................. 13
2.4 FOCUS with SmartBox option............................................................................. 14
2.5 Unit mount model FOCUS ..................................................................................15
3 Preparations for the exposure.................................................................................. 17
3.1 Precautionary actions for safe use...................................................................... 17
3.2 Switching the power on/off ..................................................................................17
3.3 Selecting the cone............................................................................................... 18
3.4 Selecting the exposure parameters .................................................................... 18
3.5 Positioning the patient......................................................................................... 20
3.6 AEC Automatic Exposure Control ....................................................................... 21
3.7 FOCUS with SmartBox ....................................................................................... 21
3.8 Taking an exposure............................................................................................. 22
3.9 Monitoring of dose levels ....................................................................................22
4 Program mode ...........................................................................................................23
4.1 Programmable anatomical time settings ............................................................. 24
5 Error messages..........................................................................................................25
6 Maintenance............................................................................................................... 27
6.1 Cleaning .............................................................................................................. 27
6.2 Disinfecting.......................................................................................................... 27
6.3 Periodic maintenance.......................................................................................... 28
6.4 Radiation dose measurement ............................................................................. 29
6.5 Changing the fuses ............................................................................................. 30
7 Technical data............................................................................................................ 31
7.1 Technical specifications ...................................................................................... 31
7.2 EMC declaration.................................................................................................. 35
8 User's statement........................................................................................................ 39
9 Recommended exposure times................................................................................45
51774-IMG rev 13 Instrumentarium Dental i
ii Instrumentarium Dental 51774-IMG rev 13
1 Introduction
1.1 FOCUS™
1 Introduction
FOCUS is a microprocessor controlled intraoral x-ray unit with a HF DC generator. FOCUS produces high quality dental images with film or digital sensors.
The well-balanced support arm is easy to move and very stable, keeping the unit motionless during the exposure. FOCUS' proprietary design has the VHF DC generator built into the horizontal part of the units support arm enabling greater reliability and ease of installation and service.
The VHF DC generator keeps the patient dose to the absolute minimum. The user friendly remote control features pre-programmed anatomical time settings making the exposure selection quick and effortless. These settings can be reprogrammed if needed.
Other selections include 60 or 70 kV, exposure times between 0.02 and 3.2 seconds, pediatric or adult modes, and film and digital modes. Exposures can be made directly from the remote control panel or with the optional remote exposure button. With a choice of arm lengths and ability to mechanically mount the unit in different configurations, the FOCUS is a fully customizable x-ray system.
As the manufacturer we strongly recommend that you read this manual before placing the unit into service.
NOTE! FOCUS must be installed according to the FOCUS Installation manual by a qualified technician. Only trained personnel should be allowed to operate FOCUS.
1.2 Intended use
FOCUS is intended to be used for producing diagnostic x­ray radiographs of dentition, jaws and other oral structures.
1.3 User profile
The unit is intended only for professionally qualified dental or medicinal personnel. The typical user is a dental nurse with specific training for using dental X-ray units.
51774-IMG rev 13 Instrumentarium Dental 1
1 Introduction
1.4 Symbols that may appear on the unit
The following symbols are used in FOCUS
Name and address of the manufacturer
Serial number
X-ray source assembly: emitting
Radiation warning
Focal spot
Filtration
Connector for remote control
Protective ground
Type B applied part
Dangerous voltage
ON or enabled
OFF or enabled
Operating instructions
Refer to operating instructions for more information. The operating instructions can be supplied electronically or in paper format
General caution
2 Instrumentarium Dental 51774-IMG rev 13
1 Introduction
3155129
Caution: Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner.
Do not reuse
Recyclable
CE (0537) symbol
MDD 93/42/EEC
ETL Mark
Conforms to UL STD 60601-1 Certified to CSA
This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment.
1.5 Type and version
The type and version of FOCUS is defined in the main label of the unit located on the under side of the horizontal arm and in the tube / HV generator housing assembly label on the tube head. The unit is class I, type B and with IP-20 protection.
The focal length is defined in the cone label in addition to type and version.
The software version is momentarily displayed on remote control display after switching the unit on.
51774-IMG rev 13 Instrumentarium Dental 3
1 Introduction
Labels on the picture are for reference purposes only. Actual texts may not be accurate.
1.6 Labels on FOCUS
4 Instrumentarium Dental 51774-IMG rev 13
1.7 Configurations
WARNING! USE LIMITATION: The unit or its parts must
not be changed or modified in any way without approval and instructions from the manufacturer. The use of accessories, transducers, and cables other than those specified may result in degraded ELECTROMAGNETIC COMPATIBILITY of the EQUIPMENT and/or SYSTEM.
WARNING! If you suspect any electro-magnetical interference affecting or caused by the unit, call service. Portable and mobile RF communications equipment can interfere with operation of the medical electrical equipment.
FOCUS is delivered with one of the following configuration:
FOCUS BASIC SYSTEMS:
1 Introduction
FOCUS x-ray machine with short reach 176 cm/ 69.2", 115 V,
FOCUS x-ray machine with long reach 191 cm/ 75.1", 115 V
FOCUS x-ray machine with extra long reach 216 cm/ 84.9", 115 V
FOCUS x-ray machine with short reach 176 cm/ 69.2", 230 V
FOCUS x-ray machine with long reach 191 cm/ 75.1", 230 V
FOCUS x-ray machine with extra long reach 216 cm/ 84.9", 230 V
Each FOCUS basic system is equipped with remote control with 32.8 feet (10m) 8 wire cable with RJ-45 plugs and Installation manual.
One chosen item from each module below is delivered with the basic unit:
TO BE CHOSEN FOR BASIC SYSTEM:
Cones
Short cone round
Short cone rectangular / Short cone rectangular for Germany
Long cone round
Long cone rectangular
Short full metal rectangular cone
Long full metal rectangular cone
51774-IMG rev 13 Instrumentarium Dental 5
1 Introduction
Wall mount plates
Wall mount plate for single stud
Wall mount plate for 16” center studs
Manual language
English
Italian
German
French
Following accessories are approved items, and they can be ordered separately.
ACCESSORIES
Short cone round
Short cone rectangular
Long cone round
Long cone rectangular
Long full metal rectangular cone
Short full metal rectangular cone
Short cone rectangular for Germany
Short horizontal arm 500 mm / 19.7 in
Long horizontal arm 650 mm / 25.6 in
Extra long horizontal arm 900 mm / 35.4 in
Additional remote control (one FOCUS unit can be installed with two)
Remote exposure button (one FOCUS unit can be installed with two)
Wall mount plate for single stud.
Wall mount plate for 16" center studs.
NOTE! To maintain safe and correct operation of FOCUS, only the approved accessories should be used. All the standard and optional items and approved accessories are suitable for use within the patient environment.
6 Instrumentarium Dental 51774-IMG rev 13
1.8 Radiation protection guidelines
The device emits X-ray radiation for medical purposes. The unit may cause an injury if used improperly. The instructions contained in this manual must be read and followed when operating the FOCUS. All government and local regulations pertaining to radiation safety must be observed.
NOTE! for USA: Many provisions of these regulations are based on recommendations of the National Council on Radiation Protection and Measurements. Recommendations for dental x-ray protection are published in NCRP Report #35 available from NCRP Publications, 7910 Woodmont Avenue, Suite 1016, Bethesda, MD 20814.
Personal radiation monitoring and protective devices are available and recommended for staff members. It is also recommended to provide the patient with a protective apron. Consult the physician before taking images of pregnant patients.
1 Introduction
WARNING! FOCUS must not be used in rooms where explosive hazards exist.
Use FOCUS with radiation protection in accordance with IEC 60601-1-3 (and/or local requirements).
PROTECTION BY DISTANCE
In all examinations the user of the x-ray equipment should wear protective clothing. The operator does not need to be close to the patient during normal use. The protection against scatter radiation can be achieved by using the remote control or the remote exposure button not less than 7 feet (2 meters) from the focal spot and the x-ray beam. The cable length of the optional remote exposure button is approximately 32 feet (10 meters). The operator should maintain visible contact with the patient and technique factors. This allows immediate termination of radiation by the release of the exposure button in the event of a malfunction or disturbance.
51774-IMG rev 13 Instrumentarium Dental 7
1 Introduction
WARNING:
This x-ray unit may be
dangerous to patient and
operator unless safe
exposure factors and
operating instructions
are observed.
Caution information on remote control
1.9 Manufacturer's liability
As a manufacturer we can only assume liability of safe and reliable operation of this unit when
FOCUS installation was performed according to the
FOCUS Installation Manual supplied with the unit.
FOCUS is used according to the FOCUS User's
Manual
Maintenance and repairs are performed by a qualified
FOCUS dealer
Original or authorized spare parts are used
If service on the unit is performed, a work order describing the type and extent of repair must be provided by the service technician. This must contain information of changes of nominal data or work range performed. The work order must furthermore indicate the date of repair, the name of the company concerned and a valid signature. User should keep this work order for future references.
8 Instrumentarium Dental 51774-IMG rev 13
1.10 Disposal
When the unit does no longer meet the manufacturer's intended operational specifications, despite proper maintenance and repair, then the unit is no longer serviceable and should be replaced. Follow all regulations on disposal of waste parts. FOCUS has at least the following parts that should be regarded as non­environmentally friendly waste products:
- X-ray source assembly
- All electronic circuits
1 Introduction
51774-IMG rev 13 Instrumentarium Dental 9
1 Introduction
10 Instrumentarium Dental 51774-IMG rev 13