Inspire IMC-734 Owner's Manual

Owner's manual
Blood Pressure
Monitoring System
IMC-734
1
IMPORTANT INFORMATION...............................................................2
CONTENTS AND DISPLAY INDICATORS...........................................3
INTENDED USE...................................................................................4
CONTRAINDICATION..........................................................................4
PRODUCT DESCRIPTION..................................................................4
SPECIFICATIONS..................................................................5
NOTICE.................................................................................6
SETUP AND OPERATING PROCEDURES.........................................7
1. BATTERY LOADING....................................................................7
2. CLOCK AND DATE ADJUSTMENT..............................................8
3. CONNECTING THE CUFF TO THE MONITOR...........................9
4. APPLYING THE CUFF..................................................................9
5. BODY POSTURE DURING MEASUREMENT............................10
6. TAKING YOUR BLOOD PRESSURE READING........................10
7. DISPLAYING STORED RESULTS..............................................11
8. DELETING MEASUREMENTS FROM THE MEMORY..............12
9. ASSESSING HIGH BLOOD PRESSURE FOR ADULTS............12
10. TROUBLESHOOTING (1)........................................................13
11. TROUBLESHOOTING (2).........................................................14
MAINTENANCE.................................................................................15
EXPLANATION OF SYMBOLS ON UNIT...........................................16
WARRANTY INFORMATION.............................................................16
SERVICE CENTER............................................................................16
ELECTROMAGNETIC COMPATIBILITY INFORMATION..................17
2
IMPORTANT INFORMATION
NORMAL BLOOD PRESSURE FLUCTUATION
All physical activity, excitement, stress, eating, drinking, smoking, body posture and many other activities or factors (including taking a blood
pressure measurement) will inuence blood pressure value. Because
of this, it is mostly unusual to obtain identical multiple blood pressure readings.
Blood pressure uctuates continually day and night. The highest value
usually appears in the daytime and lowest one usually at midnight. Typically, the value begins to increase at around 3:00AM, and reaches to highest level in the daytime while most people are awake and active.
Considering the above information, it is recommended that you measure your blood pressure at approximately the same time each day.
Please always relax a minimum moment of 1 to 1.5 minutes between measurements to allow the blood circulation in your arm to recover. It is rare that you obtain identical blood pressure readings each time.
3
CONTENTS AND DISPLAY INDICATORS
MEM button
WHO Color Bar
START button
Date/Time Display (Alternating)
Blood Pressure level
Classication Indicator
Diastolic Pressure/ Pulse Rate Display (Alternating)
Ready to
inate indicator
Irregular Heartbeat Symbol
Low Battery Indicator
Systolic Pressure
Cuff
LCD display
4
CONTRAINDICATION
It is inappropriate for people with serious arrhythmia to use this Electronic Sphygmomanometer.
INTENDED USE
Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-
invasive technique in which an inatable cuff is wrapped around the
wrist.
PRODUCT DESCRIPTION
Based on Oscillometric methodology and silicon integrated pressure sensor, blood pressure and pulse rate can be measured automatically and non-invasively. The most recent 60 measurements can be stored in the memory with date and time stamp. The Electronic Sphygmomanometer corresponds to the below standards: EN 60601-1:1990+A1:1993+A2:1995(Medical electrical equipment­Part1:General requirements for safety) , EN 60601-1-2:2007(Medical electrical equipment-Part1:General requirements for safety; Collateral Standard-Electromagnetic compatibility-Requirements and tests) , EN 1060-1:1995+A2:2009 (Non-invasive sphygmomanometers - Part 1: General requirements), EN 1060-3:1997+A2:2009(Non-invasive sphygmomanometers - Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems), ANSI/AAMI SP-10:2002+A1:2003+A2:2006.
5
SPECIFICATIONS
1.Product name: Blood Pressure Monitor
2.Model: IMC-734
3.Classication: Internally powered, Type B applied part,IPX0,No AP
or APG, Continuous operation
4.Machine size: 87mm x 66.6mm x 25.3mm
5.Cuff circumference: 14cm ~ 19.5cm
6.Weight: about 115g (exclude batteries)
7.Measuring method: oscillometric method, automatic air ination and
measurement
8.Memory volume: 60 times with time and date stamp
9.Power source: 2 ×1.5V SIZE AAA batteries
10.Cuff pressure range: 0-295mmHg
11.Range of measurement: 30-280mmHg
12.Measuring accuracy: ±3mmHg
13.Heart pulse rate range: 40-180times/min, accuracy: ±5%
14.Environmental temperature for operation: 5°C~40°C
15.Environmental humidity for operation:<90%
16.Environmental temperature for storage and transport: -20°C~55°C
17.Environmental humidity for storage and transport: <95%
18.Environmental pressure: 80KPa-105KPa
19.Battery life: Approx.3 month with alkaline batteries and 3-min.
usage per day.
20.a list of all components belonging to the pressure measuring
system, including accessories
1)Pump: KPM14A or CJP31-C03A1 or MPDC3V-03 or KP-PUMP-01
2)Value: KSV05A3 or CJV10-A03A1 or VLDC3V-02
3)LCD: IMC-734-L01
4)Wrist Cuff: 14cm ~ 19.5cm
5)Sensor: KD-2107-006G
If you want to change these components, please contact us.
6
NOTICE
1.Stay quiet, calm and rest for 5 minutes before blood pressure
measurement.
2.The cuff should be placed at the same level as your heart.
3.During measurement, neither speak nor move your body and arm.
4.Measuring on same wrist for each measurement.
5.Please always relax a minimum moment of 1 to 1.5 minutes
between measurements to allow the blood circulation in your arm to
recover. Prolonged over-ination (cuff pressure exceed 300 mmHg or
maintained above15 mmHg for longer than 3 minutes) of the bladder may cause ecchymoma of your arm.
6.Remove the batteries if the monitor will not be used for a month or
more to avoid damage of battery leakage.
7.This Electronic Sphygmomanometer is designed for adults and
should never be used on infants or young children. Consult your physician or other health care professionals before use on older children.
8.Blood pressure measurements determined by this monitor are
equivalent to those obtained by a trained observer using the cuff/ stethoscope auscultation method, within the limits prescribed by the American National Standard Institute, Electronic or automated sphygmomanometers.
9.Information regarding potential electromagnetic or other interference
between the blood pressure monitor and other devices together with advice regarding avoidance of such interference please see part ELECTROMAGNETIC COMPATIBILITY INFORMATION.
10.If Irregular Heartbeat (IHB) brought by common arrhythmias
is detected in the procedure of blood pressure measurement, a signal of will be displayed. Under this condition, the Electronic Sphygmomanometer can keep function, but the results may not be accurate, it’s suggested that you consult with your physician for accurate assessment.