Catalog No.
122 20 bis 122 23S
122 25, 122 25S
122 27 bis 122 43
122 75
123 22
Test Leck for Sniffer
Leak Testing
Content
1 Description 3
1.1General 3
1.1.1Purpose 4
1.1.2Validity 4
1.2Safety information 5
1.3Technical data 6
2 Operation 9
2.1Description 9
2.2Handling 9
3 Maintenance 10
3.1Cleaning the outside 10
3.2Repair 11
3.3Returning the Produ ct 11
3.4Decommissioning 11
2Content
1Description
1.1General
INFICON covers no responsibility nor warranty if the user or
third parties
•disregard the information in this document
•utilize the product not according to the defined use
•make any kind of changes (mod ific at ion s, alterations,
etc.) to the product.
Important remarks concerning oper ational safety and pro tection are emphasised as follows:
Indicates procedures that must be strictly observed to prevent hazards to
persons
Indicates procedures that must strictly be observed to prevent damage, or
destruction.
Notice:Information on correct handling or use. Disregard may
lead to malfunctions or minor equipment damage.
The references to diagrams consist of the Fig. No. and the
item No. in that order. For example: (2/1) refers to item 1 in
the second figure, i.e. the Manometer
3Description
1.1.1Purpose
The test leak is designed as bench-top unit and use d for calibrating and testing sniffing leak detectors. The safety regulations have to be taken into account.
Cat. No.122 20SN and 122 40SN for special gases can be
ordered on special request only.
The Cat. No. is shown on the name plate on the back side of
the calibrated test leak.
1.2Safety information
Take into account the relevant regulations and safety measures
for the calibration gases used
( safety data sheets for the calibration gases).
The test leak is pressurized. Protect it from solar heat, temperatures above 50 8C, and damag e .
)
2
5Description
Calibration gas: Avoid eye-contact and inhalation in high concentrations.
Pass on the safety information to other users.
1.3Technical data
Flammable: Keep away from ignition sources.
Smallest deviation from calibration± 10 %
pressure range0.5 bar to 8 bar
temperature range+15°C to +35°C
Storage temperature0°C to +50°C
Service life (depending on the type of gas
and the leak rate)1 - 5 years
Weightapprox. 3.5 kg
Fig. 1 Dimensional drawing
Description6
Gas typeLeak rate, typicalCat. No.
g/aoz/y
R 222 … 50,07 … 0,18122 25
R 222 … 50,07 … 0,18122 25S
R 232 … 50,07 … 0,18122 26
R 322… 80,07 …0,28122 36S
R 134a2 … 50,07 …0,18122 20
R 134a2 … 110,07 ... 0,39122 20S
R 134a10 … 140,36 … 0,49122 40
R 152a2 … 50,07 … 0,18122 27
R 152a2 … 80,07 … 0,28122 27S
R 2907 … 80,25 … 0,28122 31
R 401a2 … 50,07 … 0,18122 30
R 401a2 … 150,07 … 0,53122 30S
R 404a2 … 50,07 … 0,18122 22
R 404a10 … 150,35 … 0,53122 22S
R 404a13 … 170,46 … 0,60122 42
R 407c2 … 50,07…0,18122 28
R 407c2 … 150,07…0,53122 28S
R 410a2 … 50,07…0,18122 29
R 410a2 … 50,07…0,18122 30S
R 502a11 … 150,39…0,53122 43
R 502a2 … 50,07…0,18122 23
R 502a2 … 150,07…0,53122 23S
R 600a2 … 50,07…0,18122 21
R 600a2 … 200,07…0,71122 21S
R 600a14 … 180,49…0,63122 41
7Description
Gas typeLeak rate, typicalCat. No.
g/aoz/y
HFO1234yf
2 … 50,07…0,18122 35
R 13012 … 170,07 … 0,60122 34S
R7442 … 3,50,07…0,12122 32
R74410 … 140,36…0,50122 75
He1,00 - 1,20 x 10
He1,00 - 1,20 x 10
He2,00 - 6,00 x 10
He6,00 - 8,00 x 10
10% H2,
90% He
1-1,2x10
100% H21 - 1,1 x 10
customer
specific*
customer
specific*
2 … 5122 20SF
2 … 5122 20SN
-4
mbar l/s122 37
-4
mbar l/s122 37S
-5
mbar l/s122 38
-6
mbar l/s122 39
-4
mbar l/s122 33
-4
mbar l/s123 22
customer
specific*
customer
specific*
10 … 14122 40SF
10 … 14122 40SN
* Test leaks are only suitable for specific gases. If you want
to have a test leak with a gas that is not on the list please contact us. We will tell you if the test leak with the gas you reqest
can be build.
For additional information please see the safety data sheet of
the calibration gas and the quality test certificate that was
added to the test leak.
Description8
2Operation
2.1Description
The calibrated test leak is installed in a rectangular housing.
The front view is shown in Fig. 2.
Fig. 2 Front view
2.2Handling
The calibration test leak is continually in operation.
Pos.
1
2
3
4
Description
Manometer
Leak rate of the calibrated test leak.
(Example: Gas R152a, leak rate 2,64 g/a )
Calibration gas outlet
Date of calibration
Flammable gases! The gases R152a, R290, R50 2, R600a and HFO-1234 yf
may form explosive gas mixtures.
Care for sufficient ventilation and keep away from ignition source.
Preconditions
Ambient temperature +15°C to +35°C
Displayed pressure range 0.5 bar to 8 bar
Notice:p < 0.5 bar: Test leak no longer ready for operation.
p > 8 bar: Test leak too hot.
Operation9
Sniffing
Keep probe in contact with the gas outl et until the measured
value is stable.
Fig. 3 Sniffing Probe
3Maintenance
The calibrated test leak requires no maintenance.
When the calibration gas is used up (pressure < 0.5 bar) the
test leak has to be refilled and calibrated. Please contact your
nearest INFICON Service Center.
Notice:If the pressure is < 0.5 bar a proper calibratio n of the leak
detector is not assured. Therefore an accurate
measurement of the test sample is not guaranteed.
3.1Cleaning the outside
For cleaning the outside of the test leak, a slightly humid cloth
should normally do.
Do not use under any circumstances any aggressive or scouring cleaning agents.
10Maintenance
3.2Repair
Defective products must not be opened.
Please contact your nearest INFICON-Service Center.
3.3Returning the Product
Calibration gas
Test leaks sent for repair may still contain calibration gas.
Take account of the national regulations for shipping the respective calibra-
tion gas.
3.4Decommissioning
Substances detrimental to the environment
Products, operating material, fluids, consumables etc. may have to be spe-
cially decommissioned.
For professional decommissioning of the product, please contact your nea-