Imtmedical Bellavista 1000E, Bellavista 1000, Bellavista NEO User Manual

4.8 (4)

the art of ventilation

User Manual bellavista Ventilator

User Manual bellavista Ventilator

Thank you very much for choosing bellavista. You have purchased a top-quality ventilator that meets the highest standards.

imtmedical ag Gewerbestrasse 8 9470 Buchs (SG) Switzerland

sales@imtmedical.com

Table of Contents

1 Introduction

7

1.1

Intended use

7

1.2

Supported ventilation modes

7

1.3

Validity of this User Manual

8

1.4

Technical support

8

2 Important safety instructions

9

2.1

Legend

9

2.2

Liability

9

2.3

Introduction

9

2.4

Storing the documentation

9

2.5

Training documentation

9

2.6

Staff qualification

9

2.7

Correct User Manual

9

2.8

Use of a functional bellavista

9

2.9

Working safely with bellavista

10

2.10

Contraindications

10

2.11

Note about potential errors

10

2.12

Start-up

11

2.13

Setting up ventilation

15

2.14

Stopping ventilation, shutdown

16

2.15

Servicing and maintenance

16

2.16

Transport

16

3 Description of the device

17

3.1

Overview of bellavista 1000

17

3.2

Overview of bellavista neo

18

3.3

Overview of bellavista 1000 e

19

3.4

System overview

20

3.5

Air and oxygen supply

20

4 Preparing for ventilation

21

4.1

Checking delivery

21

4.2

First steps

21

4.3

Connecting supply lines

22

4.4

Battery operation

23

4.5

Oxygen connector

24

4.6

Inlet filter

25

4.7

Switching on bellavista

26

4.8

Start screen

26

4.9

Selecting the breathing circuit

28

4.10

Selecting the patient type

28

4.11

Safety instructions about breathing circuits

28

4.12

Connecting the breathing circuit

29

4.13

Connecting the humidifier

30

4.14

Connecting the patient

31

4.15

Flow sensor

31

4.16

Connecting the nebuliser

32

4.17

Dual limb adapter

34

4.18

Integrated expiratory valve

34

4.19

Capnography CO2 respiratory gas sensors

35

4.20

SpO2 pulse oximeter

36

5

NIV, non-invasive ventilation

39

6

Ventilation of neonates

41

7

Operation

43

 

7.1

Index of applications

44

 

7.2

Changing the monitoring settings

46

 

7.3

Changing the curve display

47

 

7.4

Zooming gestures

48

 

7.5

Safety-protected settings

48

 

7.6

Login, user level

49

 

7.7

Configuration Assist

50

 

7.8

Data Assist

51

 

7.9

iVista

52

8

Setting up ventilation

53

 

8.1

Making ventilation settings

53

 

8.2

beMode Assist

54

 

8.3

Starting ventilation (ventilation menu)

59

 

8.4

Introduction to the ventilation modes

60

 

8.5

ATC Automatic Tube Compensation

64

 

8.6

PLV Pressure Limited Ventilation

64

 

8.7

Sigh

65

 

8.8

Ramp

66

 

8.9

CPAP

67

 

8.10

nCPAP

68

 

8.11

nIPPV

69

 

8.12

PCV

70

 

8.13

P-A/C

70

 

8.14

T

70

 

8.15

PC-SIMV

71

 

8.16

PSV

72

 

8.17

S

72

 

8.18

S/T

72

 

8.19

beLevel

73

 

8.20

APRV

74

 

8.21

VCV

75

 

8.22

V-A/C

75

 

8.23

VC-SIMV

76

 

8.24

AVM

77

 

8.25

HFOT

78

9

During ventilation

81

 

9.1

Monitoring

81

 

9.2

Cockpit

81

 

9.3

Lung Recruitment Tool

82

 

9.4

Hold manoeuvre

87

 

9.5

Esophageal pressure monitoring

87

 

9.6

Chameleon

88

 

9.7

Trending

89

 

9.8

Alarms

89

10

Stopping ventilation, shutdown

91

 

10.1

bellavista shutdown

92

11

Servicing and maintenance

93

 

11.1

Factory repair

93

 

11.2

Changing fuses

93

 

11.3

Batteries

93

 

11.4

Disposal

93

 

11.5

Reprocessing,cleaning,disinfection

94

 

11.6

Calibration Assist

95

12

Specifications

97

 

12.1

Standards

97

 

12.2

Classification

97

 

12.3

Device data

98

 

12.4

Ambient conditions

99

 

12.5

Units and languages

99

 

12.6

Pressure, flow

99

 

12.7

Connection data

100

 

12.8

Trending

101

 

12.9

SpO2 pulse oximeter (optional)

101

 

12.10

CO2 respiratory gas sensor (optional)

101

 

12.11

Breathing circuit and flow sensor

102

 

12.12

Drug nebuliser (optional)

104

 

12.13

Noise generation

105

 

12.14

Supported ventilation modes

106

 

12.15

Ventilation settings

107

 

12.16

Curves and loops

111

 

12.17

Monitoring parameters

113

 

12.18

Alarm limits

119

 

12.19

Pneumatic block diagram

120

 

12.20

Models accessories, consumables, spare parts

121

 

12.21

Symbols on device and packaging

124

13

Appendix

127

 

13.1

Network / data sharing

127

 

13.2

Start screen settings

131

 

13.3

List of alarms

135

 

13.4

Manufacturer's EMC declaration in accordance with EN60601-1-2:2007

143

 

13.5

ESD safety measures

147

 

13.6

bellavista training certificate

148

 

13.7

Servicing and maintenance checklist for bellavista

151

 

13.8

bellavista quick check

153

14

Index

 

155

Passwords for bellavista ventilator

159

1 Introduction

1.1Intended use

1.2Supported ventilation modes

Introduction

1

Welcome to bellavista. This User Manual will show you how to put your bellavista ventilator into operation and use all its functions.

The bellavista ventilator was developed for ventilating adult and paediatric patients and, optionally, neonatal patients as of a tidal volume of ≥ 2 mL 1. The device is intended for use in clinics and institutional facilities where medically trained professionals are available for attending to the patient. The device can be used at the bedside, as well as for transferral within a facility, when a patient is in need of oxygen.

bellavista is only intended for use by trained personnel under the supervision of a licensed doctor.

Carefully read the instruction for use before using bellavista. Use bellavista only if you are a trained professional.

This instruction for use only applies to the ventilator type and software version specified.

Abbreviation

Description

 

 

CPAP

Continuous Positive Airway Pressure

 

 

nCPAP

Nasal Continuous Positive Airway Pressure

 

 

nIPPV

Nasal Intermittent Positive Pressure Ventilation

 

 

PCV

Pressure Controlled Ventilation

 

 

P-A/C

Pressure Assist / Control Ventilation

 

 

PC-SIMV

Pressure Controlled Synchronised Intermittent Mandatory Ventilation

 

 

PSV

Pressure Support Ventilation, Spontaneous

 

 

S

Spontaneous without backup rate

 

 

S/T

Spontaneous/timed with backup rate

 

 

T

Timed

 

 

beLevel

Biphasic ventilation at two pressure levels and additional pressure

 

support

 

 

APRV

Airway Pressure Release Ventilation

 

 

VCV

Volume Controlled Ventilation

 

 

V-A/C

Volume Assist / Control Ventilation

 

 

VC-SIMV

Volume Controlled Synchronised Intermittent Mandatory Ventilation

 

 

AVM

Adaptive Ventilation Mode

 

 

HFOT

High Flow Oxygen Therapy

 

 

beModes

Special modes for Day / Night, DualVent, MaskFit, TargetVent, Apnea

 

Backup ventilation

 

 

TargetVent

Pressure-regulated, volume-controlled ventilation mode

 

 

1  Neonatal option not available in all countries.

imtmedical ag

7

Introduction

1

1.3

Validity of this User Manual

This User Manual applies to

 

 

 

 

 

 

 

bellavista 1000 (G6)

 

 

 

 

SN: MB200000 and higher

 

 

 

 

Software version 6.0

 

 

 

• bellavista 1000 neo (G6)

 

 

 

 

SN: MB160000 and higher

 

 

 

 

Software version 6.0

 

 

 

In the following document the device “bellavista 1000 neo” will be described with the

 

 

 

following name: “bellavista neo”.

 

 

 

bellavista 1000 e (G6)

 

 

 

 

SN: MB170000 and higher

 

 

 

 

Software version 6.0

 

 

 

Where necessary, distinctions are made accordingly in this User Manual.

 

1.4

Technical support

Should you encounter any unexpected problems with bellavista, please notify your

 

 

 

local distributor or contact imtmedical ag directly:

 

 

 

Technical support

 

 

 

techsupport@imtmedical.com

 

 

 

https://www.imtmedical.com/Support/OpenTicket

8

imtmedical ag

Important safety instructions

2

Important safety instructions

 

2.1

Legend

 

 

 

Indicates a potential danger to life and limb. Or indicates a hazard

 

 

 

 

that can cause damage to bellavista.

 

2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Notes and measures with which operation of bellavista can be made easier

 

 

 

 

and more efficient.

 

 

 

 

 

 

 

 

 

Definition:

 

 

 

The operator is the facility which, after purchase of the medical device, holds respon-

 

 

 

sibility for operation of that device.

 

2.2

Liability

The manufacturer does not assume liability for any damage that occurs owing to

 

 

 

non-compliance with this User Manual. The terms of warranty and liability contained

 

 

 

in the manufacturer's terms and conditions of sale and delivery are not extended by

 

 

 

the following provisions.

 

 

 

If the device is not used as intended, liability for the performance of bellavista shall

 

 

 

always pass to the owner or operator.

 

 

 

Modifications to the device are prohibited.

 

2.3

Introduction

Read through the User Manual carefully before you put bellavista into operation. This

 

 

 

User Manual only applies to the device types and software version referred to on the

 

 

 

first pages.

 

2.4Storing the documentation Always keep the documentation in a readily accessible place near bellavista.

2.5

Training documentation

This User Manual serves as training documentation for teaching the main user control

 

 

functions.

2.6

Staff qualification

bellavista is designed to be operated by qualified medical and technical personnel.

 

 

After suitable instruction, the patient cockpit can be operated by persons without a

 

 

medical qualification, e.g. patients and nursing staff.

 

 

Liability for the performance of bellavista shall pass to the owner or operator if bel-

 

 

lavista is improperly serviced or repaired by persons who are not authorised to do so

 

 

by the manufacturer.

2.7

Correct User Manual

Make sure you have the appropriate User Manual that belongs with the device and

 

 

the software to avoid any risk to the user, patient or device.

2.8

Use of a functional

If a defective bellavista is used, malfunctions may directly or indirectly endanger the

 

bellavista

patient's health.

Before putting bellavista into operation, always check to make sure it is functioning properly. Never use bellavista if it has been found to be faulty. Have any defects repaired without delay.

imtmedical ag

9

Important safety instructions

2.9

Working safely with

Before each use of bellavista, perform a quick check. bellavista has to be repro-

 

 

bellavista

cessed after each use on a patient. Check the ventilation settings before connecting

 

 

 

the patient up to bellavista. Check apnea backup settings carefully before activating

2

 

 

backups. Adjust the alarm volume so that the alarm can be heard. Nurse call is only

 

 

designed as an additional alarm. Check that it is working after installation.

 

 

 

The patient monitoring system is also active when on standby so pneumatograms

 

 

 

and measurements displayed are no indication of active ventilation by the device.

 

 

 

Check the oxygen saturation and CO2 concentration in the blood regularly using

 

 

 

pulse oximetry and capnography or blood gas analysis.

 

 

 

Before using a defibrillator, remove the SpO2 pulse oximeter and CO2 capnography

 

 

 

sensor from the patient or disconnect from the device.

 

 

 

In non-invasive ventilation (NIV) the ventilator is intended to support a patient breath-

 

 

 

ing spontaneously. It is not intended to ensure complete ventilation of a patient.

 

 

 

A change in the patient's position may make it necessary to adjust the ventilator

 

 

 

settings. Make sure the patient is properly supplied at all times and that lung damage

 

 

 

and mistriggering are avoided.

 

• Check tidal volume and minute volume.

 

• Check the ventilation pressures.

 

• Adjust the ventilation pressure if necessary.

 

• Check the trigger reaction.

 

• Adjust the trigger settings.

 

• Always adapt the alarm limits to any new ventilation settings.

 

Only use your fingers to operate the touch screen. Pointed objects can damage the

 

touch screen.

2.10 Contraindications

• Complications with airway intubation

Complications from positive pressure ventilation

Barotrauma

Oxygen toxicity

Cardiovascular complications

Breathing effort and patient-respirator asynchrony

Adverse effects of sedation and paralysis

Other disease-related complications

2.11Note about potential errors If you have any concerns about the performance or the behaviour of bellavista, you

should take it out of service without delay. Please report your observations with respect to potential errors or ambiguities in bellavista or the accompanying documentation. The safety of the patient being ventilated is the responsibility of the attending physician or nursing staff. Their judgement takes priority over the content of this User Manual. Always ensure that appropriate measures are taken to adequately monitor the patient.

10

imtmedical ag

Important safety instructions

2.12 Start-up

Only use earthed cables in order to avoid the risk of an electric shock.

 

 

Do not use:

 

 

2

 

• a power cord set longer than 3 m

 

 

 

 

 

• a defective power cord set

 

 

• a power cord set that is not earthed

 

 

• extension cables

 

 

• double plugs or adapters

 

 

Keep the power cord set away from the patient to avoid strangulation. Keep the power

 

 

supply accessible to make it possible to disconnect the device from the mains if neces-

 

 

sary.

 

 

The absence of a ventilation alternative such as a self-priming, user-operated resus-

 

 

citator with a breathing mask (ventilation bag as specified in ISO 10651-4) can lead

 

 

to the patient's death if the ventilator fails.

 

 

If there are concerns about the condition or set-up of the earth wire, bellavista may

 

 

only be operated in battery mode.

 

 

Earthing is only reliable if the device is plugged into a socket marked as being suitable

 

 

for medical equipment.

 

 

 

 

 

 

Breathing circuit C must not be used for life-supporting ventilati-

 

 

 

on owing to the absence of expiratory volume monitoring.

 

 

 

 

 

 

 

Avoid occlusion of the expiratory valve.

 

 

Omitting a bacterial filter or using an incorrect bacterial filter can bring about trans-

 

 

mission of pathogens to the patient and contamination of bellavista.

 

 

Change the bacterial filter according to the manufacturer’s instructions and before

 

 

each new patient. Used filters must be disposed of as medical waste.

 

 

To avoid electric shocks to the patient and/or user, refrain from using antistatic or

 

 

electrically conductive tubes or lines.

 

 

Additional accessories in the breathing circuit can significantly increase the flow resist-

 

 

ance or dead space volume and, as a result, negatively impact ventilation performance.

 

 

Comply with the instructions provided by the manufacturer of the humidifier in order

 

 

to prevent exposure of the patient to a hazard and damage to bellavista. Place the

 

 

humidifier at a lower level than bellavista and the patient in order to prevent aspiration

 

 

of water and flooding of bellavista.

 

Calibrate flow sensor in the event of

a new patient

a new flow sensor

“Calibrate flow sensor” alarm

In the case of non-invasive ventilation, the expiratory measurements of volumes and capnography values can differ considerably from the actual expiration on account of leakage.

imtmedical ag

11

Important safety instructions

2

2.12.1Installation and ambient conditions

The installations for supplying power to bellavista must be earthed and must comply with the applicable standards.

Any external DC power supply must comply with IEC 60601-1 or IEC 60950-1 providing 2 MOOP between primary and secondary circuit and be of class II without functional earth.

Use oxygen only in well ventilated rooms. Not in hazardous areas or near/with combustible materials or gases. bellavista batteries and accessories must not be operated in an explosive environment or in the vicinity of combustible materials or gases.

bellavista must not be used for combustible mixtures of anaesthetic gases or anaesthetic agents with air or oxygen and/or nitrous oxide.

bellavista must never be operated in areas subject to splashing (e.g. near bath tubs, showers) or in the vicinity of an open flame (e.g. candle).

Do not cover bellavista and do not position it such that the openings for suctioning patient air or the device fan are covered or blocked (risk of overheating, inadequate supply to the patient).

Do not place vessels filled with liquid or any other objects on top of bellavista.

In order to avoid accidental excessive supply of oxygen to the patient, do not use bellavista in the vicinity of free-flowing oxygen.

Place bellavista in an upright position so that it cannot tip over.

bellavista has been tested for electromagnetic interference in accordance with the EN 60601-1-2 standard. Operation of bellavista can be influenced by electromagnetic interference. bellavista must therefore not be operated in the vicinity of magnetic resonance imaging scanners, mobile phones or any other potentially disruptive devices or systems, for example.

To avoid interference with bellavista, please refer to the tables in the appendix: Manufacturer's EMC declaration in accordance with EN 60601-1-2.

Do not position bellavista too close to other equipment or place it on top of other equipment.

Do not expose bellavista to ionising radiation. It must therefore not be used in the vicinity of devices that generate diagnostic or therapeutic radiation (X-ray equipment, radiotherapy devices).

bellavista must not be used in the vicinity of devices that generate diagnostic or therapeutic sound pressure.

bellavista is not a portable device and must not be used while being carried.

Do not use bellavista in a hyperbaric chamber (pressurised chamber).

12

imtmedical ag

Important safety instructions

2.12.2Connecting up to oxygen

Use medical oxygen only. Do not connect nitrogen oxide, helium, heliox or other gases. To avoid inadequate or excessive supply of oxygen, use the bellavista oxygen monitoring system and alarm options. The use of oxygen can lead to serious compli- 2 cations. The impact on the processes of peripheral and cerebral respiratory control

can cause respiratory pauses.

Use clean, intact oxygen tubing. Open the flow valve of the oxygen cylinder slowly and only when you have connected up the oxygen cylinder. An oxygen leak in the O2 supply or in bellavista can lead to a fire risk or an inadequate supply of oxygen to the patient.

Check the system regularly for leaks. In the event of a leak, turn off the oxygen source immediately. Use oxygen only in well ventilated rooms.

Do not use bellavista in the MR environment.

2.12.3Accessories, combination with other devices

The operator is responsible for ensuring that bellavista is only operated with accessories approved for it. The use of non-approved accessories or cables can:

jeopardise the safety of the patient and/or user

have a negative impact on the proper functioning of bellavista

reduce performance

have a negative impact on EMC protection

lead to non-compliance with legal regulations.

Combination with devices that are not mentioned in this User Manual is only permissible in agreement with the manufacturers.

Ventilation modes nCPAP and nIPPV may only be used with a compatible nasal interface, in order to ensure correct alarms.

Do not reuse single-use accessories because this can lead to a detrimental effect on sterility, functionality and general performance. Do not use antistatic breathing circuits!

imtmedical ag

13

Important safety instructions

2

2.12.4Nebulisation

Do not use a nebuliser together with the capnography sensor (risk of incorrect measurements). During nebulisation do not use any expiratory or HME filters.

Installation of a nebuliser between the Y-piece (or expiratory valve) and the patient increases dead space ventilation. Only nebulise medicinal products approved for nebulisation.

Check and clean, or replace the expiratory valve on a regular basis. Nebulisation can affect expiratory valve performance. Also bear in mind that nebulisation has an impact on the concentration of oxygen administered.

2.12.5External sensors (SpO2 and CO2)

According to EN/ISO 80601-2-12 bellavista must be provided, before start-up, with a monitoring device for measuring the expiratory carbon dioxide concentration in compliance with ISO 80601-2-55, e.g. in the expiratory section of the breathing circuit or at the patient port.

Only use the sensor in combination with other methods when monitoring the vital functions of a patient. Only use the sensor on a patient if you have the necessary expertise. Always bear in mind the dead space volume of the capnography airway adapter.

Do not use the sensor in conjunction with flammable anaesthetic gases.

Do not autoclave the sensor or immerse it in a liquid.

Do not pull on the sensor cable.

Comply with the operating temperature range.

2.12.6Communication interface

The data supplied via a network / data sharing system is provided for reference purposes only. Decisions on patient treatment should be made by the clinician on the basis of patient observation.

In the event of an alarm, check the patient and bellavista on the spot immediately because not all alarms are displayed in detail in the network / data sharing system. Only use recommended connecting cables. The devices connected must be approved medical devices conforming to EN 60601-1.

Connecting bellavista to a network / data sharing system that contains other devices can lead to previously unknown risks for the patient, user or third parties.

The following changes to the network / data sharing system can lead to risks and thus require additional analyses. Changes to the network / data sharing system particularly include the following:

Changes in configuration

Connection of additional elements

Update or upgrade of connected devices

14

imtmedical ag

 

Important safety instructions

2.13 Setting up ventilation

Ventilation does not start automatically but has to be started by the user pressing

 

the button.

Before connecting a patient:

2

 

• Perform a quick check.

 

 

• Using the start screen select the breathing circuit and the correct patient

 

category.

 

• Adjust the ventilation settings.

 

• Adjust the alarm settings to meet the specific requirements

 

Minimise rebreathing of CO2 by carefully selecting the settings for PEEP and expira-

 

tion time.

 

Monitor the patient carefully if his/her condition changes. Monitor the patient carefully

 

at the start of ventilation and when changing the settings or the breathing circuit.

 

Check the oxygen saturation and CO2 concentration regularly using pulse oximetry

 

and capnography or blood gas analysis.

 

Only use the Alarm Autoset function if the current ventilation situation is safe and stable.

 

Do not use any unadjusted alarm settings since this could prevent activation of the

 

alarm in an emergency.

 

Connectors and pins that bear the ESD warning symbol must not

 

be touched with the hand or a hand tool before the necessary safety measures have

 

been taken. Otherwise the device may be damaged.

 

2.13.1 HFOT High Flow Oxygen Therapy

 

Only use for patients breathing spontaneously. Only use with a special interface for

 

oxygen therapy (e.g. nasal cannula). Do not use nasal CPAP masks or prongs!

 

Only use with actively humidified breathing tubes. Always use an SpO2 monitoring.

 

2.13.2 Lung Recruitment Tool

 

Do not perform the lung recruitment and assessment manoeuvre with patients

 

breathing spontaneously. Always take into account the duration of manoeuvres for

 

lung recruitment and assessment in the context of the patient's condition.

 

In the event of a leak during manoeuvres the accuracy of measurements (parameters

 

and loops) is reduced. Bear in mind that ventilation after the recruitment manoeuvre

 

will continue with PEEPEnd.

 

imtmedical ag

15

Important safety instructions

2.14

Stopping ventilation,

Disconnect the patient from the bellavista before switching off the ventilator.

 

 

shutdown

 

 

2.15

 

 

2

Servicing and maintenance

Only trained professionals authorised by the manufacturer may perform maintenance

 

 

 

 

and repair work. Appropriate measuring equipment and testing devices must be

 

 

 

available.

 

 

 

Before subjecting bellavista to any maintenance work

 

 

 

• Switch off and unplug at the mains.

 

 

 

• Clean and disinfect it.

 

 

 

• Please return bellavista in a disinfected and cleaned condition.

 

 

 

If an error message appears during the self-test or the quick check, do not put bel-

 

 

 

lavista into operation.

 

 

 

For servicing, maintenance, cleaning and disposal refer to the instruction leaflet and/

 

 

 

or the operating instructions of the relevant accessory.

 

 

 

Calibrate the O2 sensor regularly. An uncalibrated O2 sensor can lead to incorrect

 

 

 

measurement and inadequate alarm signals. Subject the external CO2 sensor (op-

 

 

 

tional) to zero-point calibration if necessary. An incorrectly calibrated CO2 sensor may

 

 

 

lead to incorrect CO2 respiratory gas measurement.

 

 

 

A patient air filter that is soiled or inappropriate can cause inadequate supply to the

 

 

 

patient. Missing, incorrect or soiled air filters can cause bellavista to become contam-

 

 

 

inated or overheated.

 

 

 

Only use original spare parts.

 

 

 

2.15.1 bellavista quick check prior to start-up

 

 

 

Always perform the quick check in full. After the quick check, return the settings to

 

 

 

the correct values.

2.16

Transport

bellavista is a medical device and should be transported carefully. bellavista and ac-

 

 

 

cessories must be shipped in their original packaging. Keep the original packaging in

 

 

 

a dry place.

16

imtmedical ag

Description of the device

3 Description of the device

3.1Overview of bellavista 1000

1

2

6

11

19 21

20 22 8

23

28 29

26

24

imtmedical ag

4 3

30

5

3

7

9

12 8 10

14

15 13 17

18 16

27 25

No.

Description

 

 

 

 

 

1

Alarm lights

 

 

 

 

 

2

Screen with touch-screen function

 

 

 

 

 

3

Speaker

 

 

 

 

 

4

Attachment of expiratory valve or dual limb adapter

3

 

 

 

5

Handle

 

 

 

 

 

 

6

Cover for cooling air filter

 

 

 

 

 

7

Cover for patient air filter

 

 

 

 

 

8

Connection Assist button

 

9External device interface (blue)

Connector for SpO2 sensor

Data Communication Interface

10External device interface (yellow)

Connector for CO2 sensor

11

Expiratory valve or, alternatively, dual limb adapter

 

 

12

Inspiration patient connector

 

 

13

Cover for O2 sensor

 

 

14

Connector for bedside pressure measurement

 

 

15

Connector for flow sensor

 

 

16

Connector for expiratory valve

 

 

17

Nebuliser pressure connector (optional)

 

 

18

PAux additional pressure measurement connector (optional)

 

 

19

bellavista bus

 

 

20

Nurse call

 

 

21

2 × USB 2.0 *

 

 

22

100 Mbit Ethernet

 

 

23

On/Off

 

 

24

Power cord set strain relief

 

 

25

Oxygen connector

 

 

26

External 24 VDC power supply

 

 

27

Power indicator (green LED = battery charging)

 

 

28

Power plug

 

 

29

Device fuse 2 × T 6.3 AH, 250 V

30Display Port (for servicing purposes)

*It is recommended not to plug in or unplug any USB devices when ventilation is taking place.

17

Description of the device

3.2Overview of bellavista neo

1

3

2

4

3

30

5

3

6 7

9

12 8 10

14

15

11

13

17

18 16

19 21

20 22 8

23

28

29

25

27

26

 

 

 

24

No.

Description

 

 

1

Alarm lights

 

 

2

Screen with touch-screen function

 

 

3

Speaker

 

 

4

Attachment of expiratory valve or dual limb adapter

 

 

5

Handle

 

 

6

Cover for cooling air filter

 

 

7

Cover for patient air filter

 

 

8

Connection Assist button

9External device interface (blue)

Connector for SpO2 sensor

Data Communication Interface

10External device interface (yellow)

Connector for CO2 sensor

11

Expiratory valve or, alternatively, dual limb adapter

 

 

12

Inspiration patient connector

 

 

13

Cover for O2 sensor

 

 

14

Connector for bedside pressure measurement

 

 

15

Connector for flow sensor

 

 

16

Connector for expiratory valve

 

 

17

Nebuliser pressure connector

 

 

18

PAux additional pressure measurement connector

 

 

19

bellavista bus

 

 

20

Nurse call

 

 

21

2 × USB 2.0 *

 

 

22

100 Mbit Ethernet

 

 

23

On/Off

 

 

24

Power cord set strain relief

 

 

25

Oxygen connector

 

 

26

External 24 VDC power supply

 

 

27

Power indicator (green LED = battery charging)

 

 

28

Power plug

 

 

29

Device fuse 2 × T 6.3 AH, 250 V

30Display Port (for servicing purposes)

*It is recommended not to plug in or unplug any USB devices when ventilation is taking place.

18

imtmedical ag

Imtmedical Bellavista 1000E, Bellavista 1000, Bellavista NEO User Manual

Description of the device

3.3Overview of bellavista 1000 e

1

3

2

4

30

29

3

6

5

8

11 7 9

13

10

12

14

 

16

 

 

17

15

 

 

 

 

 

 

 

18 20

19 21 7

22

27

28

24

25

26

 

23

imtmedical ag

No. Description

1Alarm lights

2Screen with touch-screen function

3Speaker

4Display Port (for servicing purposes)

5Cover for cooling air filter

6

Cover for patient air filter

3

7Connection Assist button

8External device interface (blue)

Connector for SpO2 sensor

9External device interface (yellow)

Connector for CO2 sensor (optional)

10

Expiratory valve

 

 

11

Inspiration patient connector

 

 

12

Cover for O2 sensor

 

 

13

Connector for bedside pressure measurement

 

 

14

Connector for flow sensor

 

 

15

Connector for expiratory valve

 

 

16

Nebuliser pressure connector

 

 

17

PAux additional pressure measurement connector

 

 

18

bellavista bus

 

 

19

Nurse call

 

 

20

2 × USB 2.0 *

 

 

21

100 Mbit Ethernet network

 

 

22

On/Off

 

 

23

Power cord set strain relief

 

 

24

Oxygen connector

 

 

25

External 24 VDC power supply

 

 

26

Power indicator (green LED = battery charging)

 

 

27

Power plug

 

 

28

Device fuse 2 × T 6.3 AH, 250 V

 

 

29

Data communication interface

30Handle

*It is recommended not to plug in or unplug any USB devices when ventilation is taking place.

19

Description of the device

3.4System overview

3

3.5Air and oxygen supply

You will find the block diagram in the section “Specifications” under “Pneumatic block diagram”.

bellavista is an electronically controlled pneumatic ventilation system. It is powered by AC or DC and also provided with an internal battery set. The bellavista pneumatic system guarantees the supply of respiratory gas whilst the electrical systems control the pneumatics, monitor the alarms and provide the power supply.

The user can enter values or parameters in the bellavista microprocessor system via the touch screen. These inputs entail instructions for bellavista’s pneumatic system to ventilate the patient with a precisely controlled gas mixture. bellavista gathers readings from the proximal flow sensor and other sensors within the ventilator. Patient monitoring data can be displayed on the graphical user interface.

bellavista has two microprocessor systems, one to control the ventilation and one to control the user interface. The two processor systems cross-check each other and are able to trigger alarms independently of one another. This multiple check helps to prevent simultaneous failure of the main functions.

A comprehensive system of visual and audible alarms helps ensure the patient’s safety. Clinical alarms indicate abnormal physiological conditions. Technical alarms triggered by the ventilator’s self-tests, including ongoing background checks, can indicate hardware or software failures.

bellavista has several ways of ensuring a safe ventilation pressure. Maximum working pressures are ensured by appropriate alarm limits. If the set upper pressure alarm limit is reached, the expiratory valve opens.

bellavista uses ambient air and high-pressure oxygen. Ambient air is drawn in through the fresh air inlet and compressed by a turbine. Oxygen is supplied through a pressure port. An electronic blender mixes the oxygen and air to achieve the concentration entered by the user.

The gas is delivered to the patient through a microprocessor-controlled inspiratory valve. bellavista ventilates the patient through the inspiratory limb of the breathing circuit, which may include an inspiratory filter, flexible tubes, the humidifier system, a water trap, the Y-piece, the flow sensor and other components.

Gas exhaled by the patient is discharged through the flow sensor and the optional expiratory limb of the breathing circuit. Measurements taken by the flow sensor are used to determine the pressure, flow and volume.

An oxygen sensor monitors the oxygen concentration of the gas delivered to the patient.

20

imtmedical ag

Preparing for ventilation

4 Preparing for ventilation

4.1Checking delivery

4.2First steps

bellavista ventilator, completely packaged, with accessories, comprising:

Ventilator

Power plug set

Document box

Accessory bag

Bacterial filter

Breathing circuit

4

EasyLung test lung

 

• Set of filter mats (patient air filter and device filter)

Step Description

1Connecting the breathing circuit

2Switching on bellavista

3On the start screen

Configure breathing circuit

Enter patient type

Specify start parameters for ventilation

Start ventilation

You will find more options in the

 

During ventilation:

 

 

 

section “Operation”.

 

 

 

 

 

 

Cockpit for ventilation and settings

 

 

 

 

 

 

 

 

 

 

 

Select and set ventilation mode

 

 

 

 

 

 

 

Monitoring

 

 

 

 

 

 

 

Alarms

 

 

 

 

 

 

 

Shutdown

 

 

 

 

imtmedical ag

21

Preparing for ventilation

4.3

Connecting supply lines

Connect the power plug set to bellavista and an appropriate mains socket. bellavis-

 

 

ta can be operated at 100 to 240 VAC and from 50 to 60 Hz and automatically adjusts

 

 

to the respective voltage and frequency without the need for a manual switchover.

• Connect the oxygen supply, if available.

• Connect the nurse call, if available, to the connector.

• Always perform the quick check.

 

 

Secure the power plug set with a strain relief to prevent it from being acciden-

 

4

 

tally unplugged. Only pull on the plug, not the cable.

 

 

 

 

 

Keep the power plug set away from the patient to avoid strangula-

tion.

Power cable connection

The absence of a ventilation alternative such as a self-priming, user-operated resuscitator with a breathing mask (ventilation bag as specified in ISO 10651-4) can lead to the patient's death if the ventilator fails.

Only use earthed cables in order to avoid the risk of an electric shock.

Do not use:

a power plug set longer than 3 m

a defective power plug set

a power plug set that is not earthed

extension cables

double plugs or adapters

If there are concerns about the condition or set-up of the earth wire, bellavista may only be operated in battery mode.

Earthing is only reliable if the device is plugged into a socket marked as being suitable for medical equipment.

22

imtmedical ag

Preparing for ventilation

4.4Battery operation

Display the battery indicator by clicking on the top of the screen. The white dot on the right prevents the display from disappearing after a short period.

bellavista 1000 can be operated in battery mode for at least 4 hours. The ventilation settings will have a significant impact on battery life. Battery operation is monitored by several alarms.

Battery indicator

…%

Battery is charging or fully charged

calc

Calculation of new battery time

…h…min

Battery time remaining

4

Optimise battery time:

Always charge the batteries fully before running bellavista from the battery or if bellavista has not been used for any lengthy period.

Battery monitoring is fully automated. Keeping bellavista connected to the mains all the time has no negative effect on battery time.

Minimum charge: the optimum state of charge during lengthy periods of non-use is 40%. Charge for 4 hours every 6 months. The battery will also charge when bellavista is not switched on.

Lengthen the life of the battery:

Avoid battery depletion

Avoid high temperatures (whether in operation or in storage).

Never leave the patient unattended during battery operation! Keep an alternative ventilation option always on hand.

Continuously monitor the remaining battery life, especially when settings have been changed! Do not continue operating bellavista until the battery is completely dead. Connect to the mains power supply in good time!

After a storage period of ≥ 6 months connect bellavista to the power supply for at least 4 hours in order to charge the batteries.

4.4.1

Alarms

 

Event

 

Alarm

 

 

 

Interruption of the

Message, after two minutes an alarm that disappears as soon as:

power supply

the user confirms the message.

 

 

the power supply has been restored.

 

 

Remaining battery

Message, after two minutes an alarm that disappears as soon as:

life <1 h

 

the user confirms the message.

 

 

the power supply has been restored

 

 

Remaining battery

Continuous high priority alarm. Maximum 10 min battery life

life <15 min

guaranteed at the time of occurrence of the event.

 

 

Remaining battery

Ventilation is stopped. Continuous high priority alarming with

life 0 min

 

additional buzzer alarm sound until the battery is completely

 

 

drained.

 

 

 

imtmedical ag

23

Preparing for ventilation

4.5Oxygen connector

4

1 2

Oxygen connection adapter

Oxygen can be fed through the oxygen supply connector, irrespective of the ventilation mode.

 

Adapter type

Pressure

Flow

 

 

 

 

1

DISS 301.259.000

0 –7 bar

0 –110 L/min

 

 

 

 

2

NIST 301.258.000

0 –7 bar

0 –110 L/min

 

 

 

 

If there is any uncertainty regarding the quality of the oxygen supply, use the optional O2 filter and water trap.

Do not connect nitrogen oxide, helium, heliox or other gases.

4.5.1O2 functions

The O2 blender allows accurate oxygen dosage

FiO2 concentration 21 – 100%

Oxygen suction

An alarm sounds if the desired concentration is not reached (e.g. in case of low pressure supply)

FiO2 monitoring and alarms

The battery status bar at the top of the screen displays the symbol “O2” as soon as oxygen is connected.

For configuration of O2 supply <1.5  bar or O2 supply <100 % O2 see Configuration Assist.

If there is an oxygen sensor failure, the oxygen options can be temporarily disabled in Configuration Assist

24

imtmedical ag

Preparing for ventilation

4.5.2Oxygen consumption

Patient/Mode

Formula for estimating O2 consumption

 

 

 

 

 

Adult, paediatric

(FiO2 - 21%) / 79% × (MVInsp + (base flow × 2/3)) and nebuliser

 

 

Assumptions: I:E = 1:2, PEEP > 2

 

 

Applicable to all modes apart from 1

 

 

Neonatal, CPAP,

(FiO2 - 21%) / 79% × (MVInsp + (base flow × 2/3)) and nebuliser

 

beLevel, APRV 1

Assumption: PEEP > 2

 

Base flow

Adult

6 L/min

 

 

 

 

Paediatric

15 L/min

4

 

If PEEP < 2 mbar, base flow is reduced to a minimum of 5 L/min

 

 

 

 

Nebuliser

8 L/min with active nebuliser (insp, exp, cont)

 

 

 

 

 

 

 

If the oxygen option is switched off, always use external O2 measurement.

Only use medical oxygen. To avoid inadequate or excessive supply of oxygen, use the bellavista oxygen monitoring system and its alarm options.

Possible complications:

The use of oxygen can lead to serious complications. The impact on the processes of peripheral and cerebral respiratory control can cause respiratory pauses.

Lung tissue can be damaged by the toxic effect of high concentrations.

Use clean, intact oxygen tubing.

Only after the oxygen cylinder has been connected should the flow valve of the oxygen supply be slowly opened.

An oxygen leak in the O2 supply or in bellavista can lead to a fire risk or an inadequate supply of oxygen to the patient. Check the system regularly for leaks. In the event of a leak, turn off the oxygen source immediately.

Use oxygen only in well ventilated rooms.

4.6

Inlet filter

An optional filter protects the patient air inlet of bellavista against external contamina-

 

 

tion. It is fitted instead of the standard inlet filter. Two different filter types are available.

 

 

See also the relevant set of instructions 302.505.000

• 302.303.000 Inlet filter HEPA H14

imtmedical ag

25

Preparing for ventilation

4.7

Switching on bellavista

Switch on bellavista with the “ON/OFF” button on the left-hand side of the device.

 

 

Start-up takes ≈1 min.

During the start-up procedure bellavista undergoes an automatic self-test. If there are any deviations an alarm will sound and if necessary the device will be deactivated.

4.8

Start screen

On the start screen the following settings can be made:

4

1a

2

4a

 

 

 

1b

4b

5

 

3

1c

4c

 

 

 

 

4d

6

7

6

 

Objective

 

Steps

 

 

 

To continue ventilation

 

Change ventilation setting if necessary (5)

with the same settings

 

Start ventilation (6)

 

 

 

In an emergency

 

Load predefined profile (1c)

 

 

Start ventilation (6)

 

 

 

To set a new patient

 

Select new patient (1b) or profile (1c)

 

 

Set patient type (2), with additional information (3) if nec-

 

 

essary

 

 

Set and check breathing circuit (4)

 

 

Adjust ventilation settings (5)

 

 

Start ventilation (6)

Selection

 

Description

 

 

 

 

1

 

Select basic settings/profile

 

 

 

1a

 

Continue ventilation with the same settings as previously.

 

 

 

1b

 

Configure ventilation for a new patient, based on the current

 

 

settings.

 

 

 

1c

 

Load predefined profile (see also Data Assist).

 

 

 

2

 

Select patient type, Adult / Paediatric / Neonatal.

 

 

 

3

 

Set gender, height and lung condition as optional to obtain

 

 

start settings for ventilation and alarms.

 

 

 

4

 

Select breathing circuit.

 

 

 

4a

 

Set invasive, non-invasive ventilation or HFOT (High Flow

 

 

Oxygen Therapy).

 

 

 

4b

 

ATC Automatic Tube Compensation

 

 

(disabled in Configuration Assist)

 

 

 

4c

 

Breathing circuit (1- / 2-tube, expiratory valve, flow sensor)

 

 

 

4 d

 

Perform breathing circuit test.

 

 

 

5

 

Ventilation adjustment parameters (See all: )

 

 

 

6

 

Start ventilation.

 

 

 

7

 

Access to the main menu without starting ventilation.

 

 

 

26

imtmedical ag

When a circuit test is successfully executed the circuit test button indicates dates and time in the start screen and ventilation menu. Not executed or aborted circuit tests will also be shown as “not performed” with a yellow background.

Preparing for ventilation

4.8.1Breathing circuit test

The system test performs the following measurements and checks (for limits see section “Specifications” under “Breathing circuits and flow sensor”):

Step

Action, measurement, check

1Interrupt patient / test lung

Insp. flow resistance of breathing circuit

Flow sensor calibration

2Close the breathing circuit

Breathing circuit leakage

Exp. flow resistance of breathing circuit

4

Compliance of breathing circuit

 

3Calibration of proximal flow sensor in expiration direction (only with Adult/ Paediatric flow sensor)

4Calibration of proximal flow sensor in inspiration direction

This step is of major importance because the proper functioning of sensors, valves and alarms depends on it.

4.8.2Quick check on start-up

Perform the quick check each time the device is started.

Breathing circuit test

Circuit test successful/ventilation overlay

imtmedical ag

27

Preparing for ventilation

4.9Selecting the breathing circuit

4

4.10 Selecting the patient type

 

bellavista supports a very wide range of different breathing systems for various applications and preferences.

Passive 1-tube circuit for non-invasive ventilation

Active 1-tube circuit for invasive and non-invasive ventilation

Dual limb circuit for invasive and non-invasive ventilation

Neonatal breathing circuits and patient interfaces for nCPAP and nIPPV

Breathing circuits and patient interfaces for HFOT (High Flow Oxygen Therapy)

Patient type

Adult

Paediatric

Neonatal (optional) 1

Height

> 145 – 250 cm

50 – 171 cm

n. a.

 

 

 

 

IBW (Ideal Body Weight)

> 39 – 138 kg

6 – 63 kg

<10 kg

 

 

 

 

Tidal volume Vt

250 – 2500 mL

40 – 500 mL

2 – 250 mL

 

 

 

 

Tube (for invasive ven-

> 5 mm

3 – 7 mm

< 5 mm

tilation)

 

 

 

 

 

 

 

Tube diameter

22 mm

15 – 22 mm

10 – 12 mm

 

 

 

 

Flow sensor

301.328.010

301.328.010

301.470.010

 

 

 

 

1 Not available in the USA

4.11Safety instructions about breathing circuits

Avoid occlusion of the expiratory valve.

Omitting a bacterial filter or using an incorrect bacterial filter can bring about transmission of pathogens to the patient and contamination

of bellavista. Change the bacterial filter according to the manufacturer’s instructions and before each new patient. Used filters must be disposed of as medical waste.

To avoid electric shocks to the patient and/or user, refrain from using antistatic or electrically conductive breathing circuits or lines. Additional accessories in the breathing circuit can significantly increase the flow resistance and, as a result, negatively impact ventilation performance.

Comply with the instructions provided by the manufacturer of the humidifier in order to prevent exposure of the patient to a hazard and damage to bellavista. Place the humidifier at a lower level than bellavista and the patient in order to prevent aspiration of water and flooding of bellavista.

Calibrate flow sensor in the event of

a new patient

a new flow sensor

“Calibrate flow sensor” alarm

In the case of non-invasive ventilation, the expiratory measurements of volumes and capnography values can differ considerably from the actual expiration on account of leakage.

28

imtmedical ag

Preparing for ventilation

4.12Connecting thebreathing circuit

Breathing circuit C

Not to be used for life-sustaining ventilation owing to the absence of expiratory volume monitoring.

Various breathing circuits: see separate list of validated accessories. Configure breathing circuit using the start screen. When inserting or removing the tubes, grip by the sleeve.

Breathing circuit

Connection

Patient interface

4.12.1

Breathing circuit A

Non-invasive NIV

 

 

 

Ventilated mask with

 

1-tube circuit, passive, Adult/

leak and safety valve

 

Paediatric

 

 

 

 

Prox. pressure measurement

 

 

 

Expiratory flow calculated

 

 

 

 

 

 

 

 

4.12.2

Breathing circuit C

Non-invasive NIV

 

Non-ventilated mask

 

1-tube circuit, Adult/Paediatric

without leak (ventilat-

 

ed mask can be used

4

Proximal pressure measure-

as an alternative)

 

ment

 

Invasive

 

Expiratory flow calculated

 

Tube without leak,

 

 

 

 

 

 

tracheotomy cannula

 

 

 

 

 

 

4.12.3

Breathing circuit D

Non-invasive NIV

 

Non-ventilated mask

 

 

 

 

1-tube circuit, Adult/Paediatric

without leak (ventilat-

 

ed mask can be used

 

Proximal pressure measure-

 

as an alternative)

 

ment

 

 

 

 

 

 

Expiratory flow calculated

 

 

 

 

 

 

 

 

 

 

Invasive

 

 

 

Tube without leak,

 

 

 

tracheotomy cannula

 

 

 

 

 

 

 

 

Non-invasive NIV

 

4.12.4

Breathing circuit E

Non-ventilated mask

 

 

 

without leak (ventilat-

 

Dual limb circuit, Adult/Paediat-

ed mask can be used

 

ric/Neonatal

as an alternative)

 

Proximal pressure measure-

 

 

 

ment

 

 

 

 

 

Invasive

 

Expiratory flow calculated

 

Tube without leak,

 

 

 

 

 

 

tracheotomy cannula

 

 

 

 

 

 

4.12.5

Breathing circuit E

Non-invasive NIV

 

(Neonatal)

 

 

neo NIV

List of compatible

 

 

 

nasal interfaces

 

Dual limb circuit, Neonatal, for

(masks and prongs)

 

ventilation modes nCPAP and

 

 

 

nIPPV

 

 

 

 

Proximal pressure measure-

 

 

 

ment

 

 

 

 

Expiratory flow calculated

 

 

 

 

 

 

 

 

 

 

List of compatible

 

4.12.6

Breathing circuit

interfaces (nasal

 

 

for HFOT

cannulas)

 

 

 

 

 

1-tube circuit, for HFOT (High Flow Oxygen Therapy)

Proximal pressure measurement

Expiratory flow calculated

No breathing circuit test required!

imtmedical ag

29

Loading...
+ 134 hidden pages