Impact 754 Operation Manual

MANUAL,
PORTABLE,
(VENTILATOR,
UNI-VENT™
700
MODEL
SERIES
754/754M
OPERATION
SELF-CONTAINED
COMPRESSOR,
Eagle™
&
VENTILATION
AIR/OXYGEN
SERVICE
MIXER)
Impact 27 West
©
1997
P/N
Instrumentation,
Fairfield
Caldwell,
Impact
906-0754-02
Place
Instrumentation,
New
Jersey
Inc.
Inc.
07006
REV.
1.85C
(03/00)
TABLE
LIST
OF
CONVENTIONS,
ABBREVIATIONS
A
WORD
SHIPPING
ACCESSORIES
LIMITED
CALIBRATION/PREVENTATIVE
UNPACKING
LOCATION
WARNINGS
ASSEMBLY,
OF
ILLUSTRATIONS
ABOUT
CONTENTS
LIST
COPYRIGHT
OF
USE
AND
INTERCONNECTIONS
CONTENTS
SUBJECT
TERMINOLOGY,
AS
EAGLES
CAUTIONS
USED
RELEASE
DEFINITIONS
IN
THIS
MANUAL
(MODEL
MAINTENANCE
REGARDING
AND
754M
USE
INITIAL
AND
only)
NOTICE
ADJUSTMENTS
PAGE
vi
vii
vii
vii
vili
viti
viii
viii
SECTION
INTRODUCTION
FEATURES
DESCRIPTION
CONTROLS
EXTERNAL OFF/ON-set PRESSURE
LOW INSPIRATION AIR/OXYGEN
MANUAL
LCD
STATUS
I.
OPERATION
OF
CONTROLS,
AIR
OFF/ON
Switch - PRESSURE
ALARM/PEAK
PRESSURE
BREATH/TRIGGER
VISUAL
MODE - V CENTER - PEAK DIGITAL SETPOINTS - P.y - EXTERNAL PLATEAU - HIGH AIRWAY TIME/L:E
BAR
PRESSURE
RATIO - TIDAL
ALARM TIME/T:E MIXER
INDICATORS
min - INSPIRATION/EXHALATION
AIRWAY
GRAPH - HIGH
AIRWAY
VISUAL
Switch - SIGH
INSPIRATORY
Control - VENTILATION
RATIO
Control - ALARM
Switch - MODE
PRESSURE - MEAN
PRESSURE
ALARM
(setpoint) - RATE - INSPIRATION
VOLUME - FIO,
INDICATORS
PLATEAU
Control - TIDAL
and
LOW
AIR - SIGH - PEEP - PRESSURE
OFF/ON
MUTE/CANCEL
AIRWAY
ALARM
Switch - PEEP OFF/ON PRESSURE
RATE
VOLUME
SELECTOR
-
POWER
AIRWAY
PRESSURE
(setpoint) - LOW
AND
CONNECTIONS
Switch - HIGH
RELIEF
Control
INFORMATION
Control
Control
Switch
Switch
PRESSURE
ALARM
-
-
-
-
-
-
TABLE
LED
VISUAL
STATUS
CHARGE
ALARM
ALARM SYSTEM
LCD
VISUAL
ALARMS - OPERATING
BATTERY DISCONNECT - HIGH LOW/FAIL - EXT SETTINGS - Vy - COMPRESSOR
ALARMS - NON-OPERATING
INVERSE SYSTEM
OF
INDICATORS
FAILURE
INDICATORS
LOW - EXTERNAL POWER
I:E - TRANSDUCER
FAILURE - VENTILATOR
CONTENTS,
SUBJECT
AIR
(cont'd)
LOW - LOW
PRESSURE - APNEA - HIGH
LOW/FAIL - FIO) - PRESSURE
CALIBRATION
FAILURE
ABORT
PRESSURE
PEEP - ©,
ALARM
-
PAGE
17
-
ALARMS - ADVISORY
INSPIRATION Vy
SETTINGS - PREVENTATIVE EXTENDED TOTAL
CONNECTIONS
OXYGEN EXHALATION COMMUNICATIONS
OPERATING
INITIAL
INITIAL
SELF-CHECK
TRANSDUCER
USER
SET-UP,
SET-UP
AUTOMATIC MANUAL
PROGRAMS
TIME
NON-USE/STORAGE
FLOW
BACKUP
INLET - AIR
VALVE - EXTERNAL
POWER
SELF-CHECK
CALIBRATION
CALIBRATION
CALIBRATION
TRUNCATED
INLET - GAS
PORT
SELECTION & STOPPING
TO
3-SEC - PLATEAU
MAINTENANCE
-
EXTERNAL POWER
OUTLET - TRANSDUCER
POWER
AND
TRANSDUCER
(AUTO
CAL)
-
JACK
-
CALIBRATION
VOLUME
FAILURE
-
-
-
sie
TABLE
MODES
ASSIST-CONTROL
SYNCHRONIZED
CONTINUOUS
CONTROL
USING
USING
BACKUP
HUMIDIFIERS
OPERATOR
ALARM
OPERATING
OF
OPERATION
VENTILATION
POSITIVE
PRESSURE
VENTILATOR
PERFORMANCE
FUNCTIONS
OF
AND
ALARMS
CONTENTS,
SUBJECT
VENTILATION
INTERMITTENT
POSITIVE
END
PLATEAU
HEAT
PRESSURE
DURING
EXPIRATORY
MOISTURE
CHECKS
(cont'd)
(ACV)
MANDATORY
VENTILATION
APNEA
PRESSURE
EXCHANGERS
VENTILATION
(CPAP)
(PEEP)
(HME'S)
(SIMV)
PAGE
4-1
42
43
4-5
4-5
4-6
4-6
47
4-7
4-7
5-1
5-1
BATTERY LOW PEEP - O, ALARM
NON-OPERATING
INVERSE SYSTEM
ADVISORY
INSPIRATION VT
SETTINGS - PREVENTATIVE EXTENDED TOTAL
ALARM
LOW - EXTERNAL
PRESSURE - DISCONNECT - HIGH
LOW/FAIL - EXT
SETTINGS - V; - COMPRESSOR
ALARMS
I:E - TRANSDUCER
FAILURE - VENTILATOR
ALARMS
TIME
NON-USE/STORAGE
FLOW
MUTING
BACK
AND
POWER
AIR
CALIBRATION
TRUNCATED
UP
CANCELLING
LOW
-
PRESSURE - APNEA - HIGH
LOW/FAIL - FIO; - PRESSURE
FAIL
TO
3-SEC - PLATEAU
MAINTENANCE
-
EXTERNAL
ABORT
-
POWER
-
VOLUME
FAILURE
54
5-6
-
-
5-7
-iii-
TABLE
ROUTINE
PREVENTATIVE
CALIBRATION
CLEANING
PRESSURE COMPRESSOR
PREVENTATIVE
BATTERY
IN
CASE
OPERATOR
OPERATOR
STORAGE
SPECIFICATIONS.
CARE:
HOSE
CARE
OF
DIFFICULTY
CORRECTABLE
PROBLEMS
INFORMATION
OF
CONTENTS,
SUBJECT
CALIBRATION,
MAINTENANCE
INLET
MAINTENANCE
AND
RECHARGING
FILTER
PROBLEMS
REQUIRING
CLEANING
SERVICE
(cont'd)
PAGE
AND
6-1
6-1
6-1
6-1 6-1
6-1
7-1
8-1
8-1
8-1
9-1
10-1
LIMITED
LIST
WARRANTY
OF
ILLUSTRATIONS
FIGURE
1
2
3
4
5a Initial
Sb
Model
Interconnection
Control
Disposable
Setup
Initial
Setup
DESCRIPTION
754/754M
Diagram
Panel
Ventilator
Main
Features
Circuit
11-1
PAGE
v
x
1-2
1-11
3-1
3-2
-iv-
12
VDC
POWER
CABLE
UNIVERSAL
POWER
AC
SUPPLY
LCD
DISPLAY
BATTERY
COMPARTMENT
DOOR
m
P
=
X
CLEAR
PROTECTIVE
DOOR
\
DISPOSABLE
VENTILATOR
CIRCUIT
CARRY
HANDLE
INSTRUCTION
MANUAL
OPERATION
&
SERVICE
Figure
HOSE
AIR
DIS.S.XDIS.S.
1.
Model
754/754M
-V-
\
\
OXYGEN
DIS.S.XDIS.S.
Main
HOSE
Features
\
FUNCTIONAL,
EASY-TO-USE
CONTROL
PANEL
CONVENTIONS, ABBREVIATIONS
CONVENTIONS
[
WARNING
[ A WARNING
[
CAUTION
[ A CAUTION
[NOTE
[
Information
TERMINOLOGY
Air/Oxygen
Model
754/754M - Uni-Vent™
Pressure
Uni-Vent™ - Model
Ventilator-Compressor-Air/Oxygen
message
statement
immediately
Mixer - Blender,
Plateau - Plateau
TERMINOLOGY,
AS
identifies
identifies
following
Gas
754/754M,
Ventilator,
USED
conditions
conditions
Blender
IN
THIS
that
could
have
that
could
is
of
sufficient
Mixer - Ventilator
importance
Eagle™
DEFINITIONS
MANUAL
an
damage
adverse
this
that
effect
device.
emphasis
upon
is
AND
the
made.
patient
or
operator.
DEFINITIONS
Gas - Air Ventilator - Any
# - Denotes
ABBREVIATIONS
Airway
Alarm
Assist-Control - A/C
Assist-Control
Breaths
Centimeters Continuous
delivered - del
End
Fraction Light Liquid
liters - L Liters
milliliters - ml
Oxygen - O,
Peak Positive Pounds
Power
Synchronized
Tidal
Work-of-Breathing - WOB
or
oxygen
that
Pressure - Paw
Message
Ventilation(s) - ACV
Per
Minute - BPM
of
Water - cmH,O
Positive
Tidal
Carbon
of
Inspired
Emitting
Crystal
Per
Inspiratory
Information
Volume - V
Diode - LED
Display - LCD
Minute - LPM
End
Expiratory
Per
Square
Intermittent
or
the
resultant
reference
display
Center - AMC
Dioxide - ETCO,
Pressure - PIP
to
the
will
indicate
Airway
Oxygen - FIO,
Inch - PSI
Center - PIC
Pressure - CPAP
Pressure - PEEP
Mandatory
Model
mixture
of
these
two
754/754M,
some
number
Ventilation(s) - SIMV
Uni-Vent™,
between 0 an 9 for
gases
-vi-
Ventilator-Compressor-Air/
each
appearance
of
A
WORD
This
very
ventilators.
contained never
before
sees
a
future
ABOUT
special
Uni-Vent™
Where
portable
found
that
no
EAGLES
the
eagle
has
ventilator
in
other
a
portable
portable
available
was
named
come
ventilator
ventilator
Eagle™,
to
symbolize
today.
Where
this
small.
has
to
represent
power
the
ever
seen.
And
-
this
eagle
where
The
its
leadership
Eagle™
is
known
the
eagle
Uni-Vent™
role
in
the
represents
for
its
is
skills
known
Eagle™
the
- for
-
world
most
this
its
its
not
Eagle™
keen
of
portable
powerful,
just
possesses
sight
-
another
self-
this
bird!
skills
Eagle™
SHIPPING
Each
Uni-Vent™
lea. lea. lea. lea. lea.
Tea.
lea. lea.
Each
Uni-Vent™
lea. lea. lea. lea.
lea.
1
ea.
lea.
2
ea.
2ea. | Instruction
ACCESSORIES
The
Accessories
Accessories
quantity
may
required.
CONTENTS
Ventilator,
Ventilator High High
Universal
Universal
12
Operator's
Ventilator,
Ventilator
High
High
Universal
12
Case,
Strap,
List
be
ordered
Circuit, Pressure Pressure
AC
AC
VDC
Power
Circuit, Pressure Pressure
AC
VDC
Power
Padded,
Velcro®,
LIST
contains
Manual,
direct
Eagle™
Compressor,
Eagle™
Compressor,
Manual,
common
Model
Disposable,
Hose,
D.LS.S.
Hose,
D.1.S.S.
Power
Supply
Power
Supply
Cable
Model
Model
Disposable,
Hose,
D.I.S.S.
Hose,
D.I.S.S.
Power
Supply
Cable
Ventilator & Accessories
24"
Long
Operation
items,
from
754
Air/Oxygen
Single
Oxygen X D.LS.S.
Air X D.LS.S.
(standard)
with
754/754M.
754M
Air/Oxygen
Single Oxygen X D.I.S.S. Air X D.LS.S.
with
& Service
required
Impact.
When
is
16-30V
16-30V
shipped
Mixer
Patient
Air,
DC-DC
is
shipped
Mixer
Patient
Air,
DC-DC
(Model
from
time
ordering,
with
the
Use
Oxygen, 6'
Long
Converter
with
Use
Oxygen, 6'
Long
Converter
754M)
to
time.
please
following
6'
Long
(optional)
the
following
6'
Long
Each
item
include
the
components:
components:
is
preceded
part
number,
by
its
part
number.
description
and
Send
written
PART
334-0020-00
402-0014-00
402-0016-00
820-0067-00
825-0002-00
825-0004-00
703-0754-09
703-0754-08
703-0754-03 906-0754-01
906-0754-02
purchase
NUMBER
orders
to:
Impact P.O. West
Telephonic Fax
Strap,
Velcro®,
Case,
Padded,
Case,
Padded,
Ventilator
High
High
Universal
Universal
12
Operator's
Instruction
Pressure
Pressure
VDC
Circuit,
AC
AC
Power
Manual
Manual,
Instrumentation,
Box
508
Caldwell,
orders:
24"
Ventilator & Accessories
Ventilator
Hose,
Hose,
Power
Power
Cable
New
orders:
DESCRIPTION
Long
Disposable,
D.I.S.S.
D.I.S.S.
Supply
Supply
(Model
Operation & Service
>
vii
-
Inc.
Jersey
Single
Oxygen X D.I.S.S.
Air X D.I.S.S.
with
16-30V
754/754M)
07007-0508
973/882-1212 973/882-4993
Patient
Use
Oxygen,
Air,
6'
DC-DC
(754M)
6'
Long
Converter
Long
LIMITED
Permission
for
use
in
COPYRIGHT
is
hereby
a
military
granted
service
to
the
training
Department
program
CALIBRATION/PREVENTATIVE
This
device
should
be operating year calibration sensitive next should
maintenance
754/754M Alarm
specifications
unless
significant
check
Calibration
calibration
be
examined,
contains
messages
should
is
due.
checks
internal
will
incorporated
(see
usage
be
made
label
is
DO
NOT
operationally
can
be
found
appear
into
LIMITED
warrants
following
attached
remove
tested,
in
clocks
that
in
the
Alarm
a
to
the

UNPACKING

Compare shipping before
LOCATION
shipping
damage.
attempting
case
contents
If
there
is
operation.
OF
USE
against
no
apparent
RELEASE
of
and
other
a
regular
WARRANTY
shorter
the
and
SERVICE
monitor
Message
SHIPPING
sign
each outer
this
label. its
of
preventative
period
2000
batteries
cumulative
mechanical
(Model
Defense
to
technical
MAINTENANCE
statement).
between
hours
case,
Following
section
Center
CONTENTS
of
indicating
recharged
of
the
use
when
damage,
754M
reproduce
training
only)
all
material
programs.
furnished
maintenance
Calibration
preventative
cumulative
when
6-months
Model
and
storage/non-use
calibration/preventative
list.
the
of
before
754M
Examine
read
instructions
program
maintenance use last
continuous patient-use
to
measurements
or
12-month
calibration
Instruction
periods.
instrument
under
this
insure
compliance
should
inspections.
period.
was
storage/non-use,
is
attempted.
Manual.
Appropriate
maintenance
for
contained
be
A
A
pressure-
performed
Recommended
The
is
any
obvious
within
contract
with
made
each
complete
and
the
this
device
Model
Advisory
required.
signs
this
Manual
of
The
Uni-Vent™
operated
barrier
in a wet
(small
WARNINGS
[
WARNING:
[
WARNING:
[
WARNING/CAUTION:
WARNING/CAUTION:
port
(see
[
CAUTION:
[
WARNINGS
CAUTION:
SECTION
CAUTION: entitled
[
CAUTION:
CLEANING
Eagle™,
tarp,
Electric
Do
Figure
2).
Federal
AND
Service
entitled
Do
not
Internal
Model
environment,
plastic
sheet,
AND
not
CAUTIONS
SERVICE).
allow
CAUTIONS
shock
hazard,
operate
law
is
for
components
this
Possible
DO
NOT
restricts
to
be
performed
oil
additional
754/754M,
users
should
etc.).
do
instrument
explosion
block
this
device
REGARDING
and
grease
instructions.
are
susceptible
is a transportable
take
precautions
REGARDING
not
remove
or
connect
by
qualified
to
enter
equipment
prior
to
hazard
to
sale
USE,
system.
to
if
the
by
damage
device,
and protect
covers.
reading
biomedical
instructions
used
in
the
ventilator
or
on
the
(cont'd)
See
ROUTINE
from
therefore,
this
USE
contained
presence
circuit
hose
order
of a physician.
equipment
CARE
static
discharge.
its
location
device
by
within
of
flammable
to
the
Internal
technicians
AND
of
use
will
covering
only
MAINTENANCE
it
with a protective
this
Manual.
anesthetics.
Compressor
(for
Model
vary.
Air
754M,
section
When
Filter
see
-
vili
-
ASSEMBLY,
ASSEMBLY:
Model
754
Model
754M
check
your
RECHARGING
INTERCONNECTIONS:
CAUTION:
NOTE: 440
Hz.
Model
includes a 16-30V
the
secondary
The
black
NOTE:
the
11-15
negative
Follow
©
1.
a
50-PSI
e
2. external
©
3. connectors
disposable
+
4. JACK
The
standard
Voltage
754M),
is
input
input
For
11-15
VDC
ground.
Foruse
For
Connect a disposable
Connect
operable
lead
power
INTERCONNECTIONS
-
No
assembly
-
Depending
contract
section
interconnection
with
external
use with
air
source.
on
the
ventilator
Universal
and
external
Universal
and
line
freguency
from
DC-DC
VDC
Converter.
leads
provided.
is
positive,
operation,
source. A 12
is
required
upon
accordingly.
for
instructions.
instructions
external
external
oxygen:
oxygen
Model
source,
air:
Use
only
ventilator
754/754M
circuit
AC
Power
power
source
AC
Power
sensing
115/230
It
Attachment
white
is
connect
VDC
before
contract
(see
VAC,
accepts
negative.
requirements,
If
battery
prior
connect Use
only
connect
medical-grade
Supply
Power
high
circuit
Connector
Figure
Supply,
(see
Figure
is
is
automatic.
50
to
external
to a mating
Do
an
appropriate
Cable
AND
placing
pack
4).
operable
400
this
installation
to
placing
high
pressure
medical-grade
pressure
compressed
to
its
respective Panel.
or
12
VDC
5b).
from
An
optional
Hz
(voltage
DC
voltages
connector
not
attach
power
is
provided
INITIAL
device
battery
pack
is
required,
this
device
oxygen
air
hose
air
gas
Observe
Power
AC
Universal
and
ranging
is
the
braided
cord
between
for
attachment
into
operation
installation
into
hose
oxygen
between
(see
Figures
outlet,
directions
Cable,
voltages
line
frequency
from
required
shield.
EXTERNAL
ADJUSTMENTS
may
see
BATTERY
service
between
(see
Figures
AIR
transducer,
between
between
AC
Power
16
and
to
an
be
(see
Figure
OXYGEN
5a
inlet
port
Sa
and
5b).
and
included
EXTERNAL
90
and 265
Supply
sensing
to
30
volts
polarity
must
POWER
automotive
necessary.
CARE
2).
and
5b).
and
exhalation
with
(standard
is
automatic),
which
be
power
Please
AND
inlet
port
a
50-PSI
each
POWER
volts,
47
connect
observed.
JACK
source,
and
valve
to
with
and
to
and
INITIAL
Before VISUAL
patient
ADJUSTMENTS:
placing
use.
this
INDICATORS
device
into
operation,
AND
CONNECTIONS.
read
section
Make
entitled
OPERATION:
control
settings
DESCRIPTION
and verify
device
OF
CONTROLS,
performance
prior
to
uNiversaL
AC
POWER
SUPPLY
CONNECTOR
EXTERNAL
AIR
INLET
FITTING
MALEDISS.
.-
3
GAS
OUTLET
HOSE
CONNECTOR
\
EXTERNAL
OXYGEN
FITTING
MALE
D.LS.S.
/
INLET
“VALVE
TRANSDUCER
HOSE
_>
EXHALATION
HOSE
BARB
(REAR)
BARB
(FRONT)
DISPOSABLE
VENTILATOR
CIRCUIT
INTERNAL
COMPRESSOR
AIR
FILTER
Figure
2.
Interconnection
“Ke
Diagram
SECTION
I.
NTRODUCTION
User's
will
find
this
instrument
device
features. A complete
many
features.
Your
Model
754/754M controlled settings, are
Plateau. ventilatory control
The
blended
and
filtered.
to
internal
Uni-Vent™
may
12 265 directly permits provided. positive, operating
by
alarm
operable
All
limits
Model
from
external
Acceptable
componentry.
be
recharged
VDC
Power
volts,
47
to
AC
use
Attachment
the
power
an
internal
parameters,
with
or
modes
are
backup
assures
peak
inspiratory
754/754M
external
compressed
is
operable
within
Cable
to
440
mains
from
external
white
is
- all
without
input
Hz.
providing
negative.
gas
OPERATION
quite
easy-to-learn
understanding
is a portable, microprocessor
gas
source(s),
PEEP,
with
PEEP
and
altitude
continued
can
provide
oxygen
air.
External
gas
It
is
from
the
range
is
provided.
An
optional
DC
to a mating
is
dedicated
mechanical
pressures
gas
(compressed
pressures
operable
internal,
of
either
The
Universal
115/230
voltages,
connector
Do
not
for
and
of
its
capabilities
electronically
(CPU)
gas
or
without
compensable
and
high
mixtures
gas
compressed
may
in
any
rechargeable
of
Universal
volts,
ranging
attach
patient
controlled
which
flows,
gas
SIGH.
support
pressure alarm
with
cylinder
air,
range
up
position;
the
aforementioned
AC
AC
Power
50
to
400
from
is
required
the
braided
care.
operate.
continuously
blends, and
to
oxygen
or
delivered
to
upright,
batteries;
Power
16
and
and
ACV
minimize
if
the
content
PTLOX)
80-PSI
Supply
Supply
Hz
(voltage
to
30
volts
polarity
shield.
The
ventilator,
patient
on
11-15
following
limitations
monitors
power
and
SIMV
your
becomes
setpoint.
ranging
and
from
an
without
its
side,
volts
DC
voltages. A Universal
connects
(standard
and
which
must
The
Model
text
will
compressor,
and
signals.
are
operable
patient's
apneic.
from
internal
electric
effecting
or
lying
DC
(negative
directly
with
line
frequency connect be
observed.
754/754M
presents a brief
allow you
displays
ACV,
work-of-
21 compressor, compressor,
measurement
flat.
the
to
to
take
air/oxygen
airway
SIMV
with
or
breathing
An
adjustable
to
100%. Gases
or
must
ground).
AC
to
AC
mains
Model
754M)
sensing
the
secondary
The
black
does
not
overview
advantage
mixer. pressure,
and
CPAP
without
external
accuracy
is
consume
Pressure
and
pressure
may
oxygen
be
oil-less
Its
battery
Power
Supply
providing
connects
automatic),
input
leads
input
lead
of
of
its
It
is
control
modes
automatic
limit
be
and
or
danger
pack
and
90
to
and
is
gas
for
Your
ventilatory
Advisory
operating,
The polycarbonate aeromedical, transport,
or
Model
deployment
754/754M
homecare,

FEATURES

Microprocessor
power
Extensive
©
«
Contemporary
©
©
e
system
employs a comprehensive
section
(see
extremely
is
is
material.
Gas-efficient
on
gas
Rechargeable
Self-contained
It
prehospital
and
and
storage.
sources,
alarm
electronic
for
operating
batteries,
entitled
durable
appropriate
control
internal
design
system,
of
compressor
monitoring
facilitate
to
control
power.
fully
may
ALARM
and use
for
ATLS,
(ALS,
functions
all
operating,
of
circuitry
compatible
operated
be
alarm
FUNCTIONS
designed
adults,
with
ACLS)
including
external
and
non-operating
transport
eliminates
with
without
system.
all
for
children
settings.
automatic
gases.
placement.
and
vehicular
attachment
Alarms
for
environments.
all
are
complete
infants
and
small
Its
monitoring
advisory
and
pneumatic-logic
electrical
external
of
categorized
descriptions
is
case
Its
clinical,
in
- size
systems
weight
and
internal
of
conditions.
circuits,
and
gas(es).
as
Operating,
each
of
injection
hospital,
field
facilitates
battery
and any
airborne
Non-
alarm).
molded
from
external
and
dependency
environments.
FEATURES,
Numerical
+
Graphics digital
+
Automatically
25,000
DESCRIPTION CONNECTIONS
Uni-VentTM
facilitate
display
control
CONTROLS
Control
Where
3).
contains
ease
of
use,
of
graphics,
panel
components
descriptions
applicable,
corresponding
(cont'd)
panel
markings
display
bar
ft.
visibility,
control
are
graph.
compensates
OF
various
settings,
from
listed
includes
CONTROLS,
controls,
and
interconnection. A large,
weather,
based
LCD
that
indicate
12-second
pressure
indicators
operating
fluids
upon
their
information
sequence-of-operation
pressure
transducer
and
conditions,
and
control
waveform,
to
VISUAL
connectors.
liquid
and
alarm
inadvertent
panel
location,
is
provided.
altitude-related
contact.
steps
its
amplitude
barometric
INDICATORS
Their
placement
crystal
display
status
information. A clear
Each
reading
is
from
described
to
simplify
is
calibrated
has
(LCD)
left
in
the
and
pressure
AND
been
carefully
provides
to
right,
following
speed
start-up.
to
the
adjacent
changes
chosen
continuous
door
protects
top
to
text
up
to
to
bottom.
(see
Figure
|
|
|
orana.
|
|
|
VENTILATION
|
pe
|
даун”
Rabat
Vat
Manas
par
ün
=
© © ©
RATE
e
MUTE/
CANCEL
rer
EB
INSPRATION
tive e RATIO
an
m
+ +
TRIGGER
Gk
ο
saa
κά
M
volume
PRESSURE
IM
-+
ON
Mad
=
пом.
YY
а
새야
a
one
a
ans
system
FARE
ALARMS
iow
1
CN
ee
ema
|
ножен
a
©
Me
Tome
y
MODE
Figure
3.
Control
1-2
Panel
CONTROLS,
EXTERNAL
manually
"samples"
pressure
compressor. from from sound.
switch.
SIGH or ventilations equals 50%. SIGH. seconds.
"OFF"
selected switch.
PEEP internal Pressing EXPIRATORY pushbutton
select
the
is
sensed,
the
external
"ON"
to
The
The
OFF/ON
without
SIGH.
150%
Proportionately
Asa
safety
Manually
is
its
("ON").
The
OFF/ON-SET
PEEP
the
switch.
(cont'd)
AIR
OFF/ON
external
respective
the
LCD
If a pressure
air
supply.
"OFF".
LCD
EXTERNAL
are
of
default
SIGH
control.
pushbutton
Attempting
display
Pushbutton
When
delivered once
the
INSPIRATION
increasing
precaution,
pressing
value.
The
LCD
OFF/ON
PRESSURE).
The
Pushbutton
compressed
gas
fitting
will
display
exceeding
Manually
area
for
EXTERNAL
AIR
OFF/ON
Switch:
activated,
every
both
Uni-Vent™
this
pushbutton
SIGH
becomes
display
display
Pushbutton
When
the
ventilator
allows a PEEP
The
PEEP
OFF/ON-SET
air
as
and
looks
"OFF"
40-PSI
pressing
to
select
The
the
first
100
TIME
the
area
for
will
Switch:
value
LCD
Switch:
your
"ON"
display
SIGH ventilator
ventilations
setting
inspiration
does
disabled
SIGH
blank
is
display
The
primary
for
an
air
(default
(+/-
2-PSI)
this
pushbutton
with
AIR
is
will
OFF/ON
(and
not
allow
will
toggle
("OFF")
is
on
when a SYSTEM
The
PEEP
turned
to
be
manually
area
display
EXTERNAL
air
source.
pressure,
value),
is
sensed,
less
than
on
the top
blank
when a SYSTEM
Pushbutton
generated
or
7-minutes,
exhalation
and
exhalation
the
the
display
in
the
top
OFF/ON-SET
ON,
PEEP
entered
for
PEEP
will
blank
AIR
When
greater
and
will
35-PSI
breath
inspiratory
the
line,
has a default
is when a SYSTEM
than
the
CPU
the
LCD
toggle
(+/-
line,
immediately
Switch
will
whichever
period),
time
allows
portion
from
"OFF"
CPAP
immediately
FAILURE
Pushbutton
(see
section
on
the top
Pushbutton
the
ventilator
40-PSI,
allows
operation
will
display
the
display
2-PSI)
present
below
FAILURE
permits
be a SIGH.
occurs
which
increases
Uni-Vent™
of a SIGH
to
"ON",
mode,
or
when
below
Alarm
occurs.
Switch
value
of 0 cmH,O
entitled
line,
immediately
FAILURE
Switch
permits
is
turned
Ifa
lower
from
"ON",
from
"OFF"
will
its
respective
Alarm
Uni-Vent™
Additional
first.
Each
delivered
to
breath
or
from
PRESSURE
its
respective
activates
USING
Alarm
you
ON,
its
pressure,
the
internal
and allow
to
"ON",
cause
an
occurs.
to
operate
SIGH
SIGH
volume
maintain
to
exceed
"ON"
PLATEAU
pushbutton
Uni-Vent™'s
(no
PEEP).
POSITIVE
below
its
occurs.
to
CPU
or
no
operation
or
alarm
to
pushbutton
with
ventilation
by
I:E
during
3-
to
"OFF".
is
END
respective
PRESSURE permits
(default
PRESSURE
cmH;0.
PLATEAU
display section SYSTEM
HIGH select absolute
INSPIRATORY
HIGH
immediately
a
LOW
has
PRESSURE
the
PRESSURE
SYSTEM
PRESSURE
absolute
an Control Control Control
ACV
value).
When
area entitled
FAILURE
HIGH
range
setpoint setpoint setting
PLATEAU
or
SIMV
Pressing
ALARM/PEAK
pushbutton
this
"OFF".
to
value
PRESSURE
for
USING
ALARM/PEAK
PRESSURE
15
from
PRESSURE
ALARM/PEAK its
below
FAILURE
ALARM
range
value
on
is
display
OFF/ON
operation
pushbutton
this
INSPIRATORY
PRESSURE
PLATEAU
PRESSURE
Alarm
occurs.
ALARM
100
to
RELIEF
respective
Alarm
occurs.
to
O
from
its
equals
line,
top
the
blank
will
Pushbutton
with
a
pressed,
is
PLATEAU).
INSPIRATORY
cmH;0.
INSPIRATORY
control.
Control:
cmH;0.
50
current
immediately
when
pressure
activates
PRESSURE
will
it
PLATEAU
the
on
is
PEAK
and
the
When
Control
The
control
This
When
position.
SYSTEM
a
Switch: plateau.
PLATEAU
a
toggle
becomes
line,
top
The
INSPIRATORY
ventilator
setpoint
PRESSURE PRESSURE
HIGH
is
the
The
below
FAILURE
PRESSURE
The When
the
value
RELIEF
display
the
disabled
immediately
PRESSURE
PRESSURE
turned
is
equals
value
select
to
used
ventilator
LCD
its
is
display
respective
Alarm
ventilator
is
that
Control
"OFF"
from
("
below
PLATEAU
RELIEF
PRESSURE
the
ON,
current
its
RELIEF
ALARM
LOW
the
ON,
turned
for
area control.
occurs.
PLATEAU
is
turned
referenced
setpoint.
its
to
the
in
")
respective
its
OFF/ON
Control:
RELIEF
HIGH
position.
Control
PRESSURE
LOW
the
LOW
the
LOW
The
setpoint
Control
OFF/ON
ON,
cmH;0
10
absolute
Its
PLATEAU
CPAP
pushbutton
display
This
setpoint.
PRESSURE
The
display
setting
PRESSURE
PRESSURE
PRESSURE
Pushbutton
PLATEAU
below
range
value,
mode.
will
control
It
ALARM/PEAK
display
LCD
the
on
is
will
ALARM
ALARM
is
OFF
the
5
is
from
or
LCD
The
switch
when
blank
used
is
an
has
area
line,
top
blank
setpoint.
ALARM
ALARM
Switch
HIGH
90
to
its
(see
a
to
for
when
It
CONTROLS,
VENTILATION
ventilation
breaths
equals
immediately
change
Alarm
INSPIRATION duration
maximum.
(inspiratory default ventilator Setting causes
time RATIO TIME/I:E selected.
TIDAL
maximum
of
the
rate for
per
minute
its
current
to
12
occurs,
of
all
Its
time
I:E
ratio
is
turned
INSPIRATION
the
INSPIRATION
and
IE
Control
RATIO
It
will
VOLUME
flow
following
(cont'd)
RATE
position.
above
BPM.
or
the
TIME/I:E
ventilator-delivered
usable
is
of
ratio
are
setpoint
blink
is
equivalent
gas
Control:
ACV
and
SIMV
(BPM).
its
greater
1:2
ON,
Control
When
The
respective
The
VENTILATION
CPAP
mode
RATIO
range
is
than
is
activated
the
INSPIRATION
TIME
TIME
simultaneously
is
on
setting
during
INVERSE
Control:
to
or
gas
combinations:
LCD
limited
expiratory
causes
the
The
approximately
The
VENTILATION
operation.
the
ventilator
display
control.
is
to
bottom
display
During
RATE
selected.
Control:
breaths.
by
the
when
this
the
resulting
be
calculated
displayed
line,
will
I-E
TIDAL
It is
not
is
turned
area
for
VENTILATION
APNEA,
Control
The
INSPIRATION
It
is
adjustable
VENTILATION
time)
are
not
control
TIME/I:E
conditions.
VOLUME
is
I:E
when
in
the
LCD.
immediately
blank
when a SYSTEM
1000mI/sec
External
RATE
operable
ON,
the
the
RATE
setting
in
0.1
RATE
permitted.
tumed
to
its RATIO RATIO
Control
Control
to
VENTILATION
The
above
allows
(60
LPM).
Air
(Cylinder)
control
in
CPAP
setting
determines
(RATE = 0).
VENTILATION
RATE
Control
Control
display
TIME/I;E
second
Control
(See fully
be
calculated. Setting
LCD
its
respective
gas
TIDAL
setting
will
blank
RATIO
increments
setting.
section counterclockwise
setpoint
display
FAILURE
RATE
to
be
entitled
value
settings
area
control.
delivered
VOLUME
the
Its
range
RATE
Control
setpoint
and
its
LCD
when a SYSTEM
Control
from
0.1
to
Inverse
ALARM
position.
equals
its
the
default
are
changed.
for
INSPIRATION
The
INSPIRATION
Alarm
occurs
over a wide
may
mechanical
is
from 1 to
setpoint
is
on
the
display
FAILURE
sets
the
inspiratory
3.0
seconds
I:E
Ratio's
FUNCTIONS).
When
current
be
I:E
or
CPAP
range.
obtained
position. RATIO Inspiration
TIMELE
150
value
bottom
will
the
is
Its
from
line,
A
(1:2)
any
volume
Tidal
VOLUME
initially
LCD whatever
del"
immediately SYSTEM
AIR/OXYGEN oxygen
gas
flow
TIDAL
FAILURE
be
to
or
gas
combinations:
External
calculated
is
Control
displays
corrections
VOLUME.
above
MIXER
delivered.
21
Oxygen
setpoint
"444
are
The
respective
its
Alarm
Control:
has
It
100%
to
to
21
and
displayed
and
equals
value
TIDAL
set"
required,
LCD
control.
occurs
range
a
21%
FIO»
100%
External
External
in
its
VOLUME
then
or
CPAP
The
of
21%
FIO,
21%
FIO;
External
area
The
AIR/OXYGEN
21%
from
from
display
from
Compressed
Internal
Air
Cylinder
External
milliliters
position.
current
the
is
FIO,
FIO,
for
alternately
LCD
TIDAL
for
TIDAL
selected.
air)
(all
from
External
from
Oxygen
External
Air
Air
Compressor
or
Compressor,
Oxygen
When
(ml).
Uni-Vent™'s
first
the
VOLUME
VOLUME
MIXER
100%
to
External
Compressed
Internal
External
and
Oxygen
(Compressor)
ventilator
the
two
or
one
displays
Control
Control
(all
Air
Air
and
"####
Control
setting
allows
oxygen)
(Cylinder)
(Compressor)
Air
Compressor
Air
Internal
or
Internal
is
monitors
CPU
delivered
set"
setpoint
display
air,
may
and
Cylinder
Compressor
Air
Air
Compressor
ON,
turned
and
breaths,
TIDAL
on
is
will
oxygen,
obtained
be
Compressor
or
the
adjusts
while
VOLUME
bottom
the
blank
mixtures
or
from
TIDAL
gas
CPU
the
and
when
the
flow.
makes
"####
line,
a
air
of
following
Its
and
100%
FIO,
from
External
Oxygen
CONTROLS,
When
the
Uni-Vent™'s mode,
FIO,
during
the
AIR/OXYGEN
AIR/OXYGEN
when
an
EXT
ALARM MUTE/CANCEL Pushbutton
Muting:
period
(see
predetermined
the
pre-existing
When
an message. Pressing will
Ifa
the
described
Canceling:
an
the apnea. settings.
the
resume
new
display
Advisory
audible
Cancellation
(cont'd)
ventilator
CPU
monitors
proportions
first
one
or
MIXER
MIXER
AIR
LOW/FAIL,
Switch
can
Depressing
ALARM
mute-period
alarm.
alarm
causing
The
ALARM
ALARM
flashing
alarm
condition
of
that
alarm.
when
the
ALARM
depressed,
When
condition
Alarm
visual
and
FUNCTIONS).
when
of
is
turned
ON,
and
are
based
two
delivered
Control
Control
Pushbutton
either
mute
this
switch
is
condition
LED
alternately
MUTE/CANCEL
the
mute
occurs
The
new
MUTE/CANCEL
the
(see
APNEA
an
alarms
APNEA
the
AIR/
adjusts
on
gas
breaths.
setpoint
setting
display
O,
LOW/FAIL,
an
audible
mutes
the
Alarm
reached
occurs,
period
during
alarm
ALARM
ALARM
and
alarm
OXYGEN
gas
flow
according
flowing
is
flashes,
Pushbutton
the
will
at
Delivered
on
the
bottom
will
or
Switch:
alarm
audible
muting
(audible
the
LCD
at
ends.
mute
period
cause
Pushbutton
MUTE/CANCEL FUNCTIONS).
controlled
the
Uni-Vent™
allows
MIXER
to
1000ml/sec.
FIO
is
line,
blank
when a SYSTEM
FIO;
Alarm
Depending
signal,
or
component
is
reset
when
alarm
will
resound). A new
ALARM
the
same frequency
Switch
the
causes
of a previous
ALARM
Switch
During
ventilations
to
Control
the
displayed
immediately
occurs.
upon
cancel a particular
of an
MESSAGE
Pushbutton
resume
setpoint value
FIO)
and
The
CPU
alarm
Operating
the
current
as
the
ALARM
alarm,
LED
to
is
pressed.
APNEA
an
that
operation
TIDAL
makes
in
the
LCD.
above
FAILURE
category,
Alarm
alarm
alarm
CENTER
the
accompanying
the
alternately
Switch
automatically
are
VOLUME
any
its
respective
the
alarm
condition
condition
will
LED
to
new
alarm
flash
cancels
alarm
the
at
equals
its
setpoints.
necessary
The
LCD
control.
Alarm
occurs.
ALARM
function.
condition
no
overrides a "muted"
display
audible
stay
on
continuously.
will
and
respond
audible
the
condition,
invoked
ACV,
preset
current
have
position.
In
CPAP
flow
corrections
display
for a predetermined longer
the
area
The
It
will
MUTE/CANCEL
applies
pertinent
alarm.
no
effect
as
previously
component
cancel
will
it
onset
the
at
or
SIMV
for
blink
or
alarm
It
upon
of
both
of
CPAP
MANUAL operable
Pressing is settings inspiratory until
audible
This MANUAL this (LPM). exceeding
MODE
power
a operation transducer
two
this
one
to
equal
and
airway
beep
control
pushbutton
A
40
MANUAL
The
Selector
the
in
Turning
"SELE-CHECK"
the
will
BREATH/TRIGGER
ways.
control
expiratory
time
pressure
is
functions
TRIGGER
pressure
cmHz0.
ACV,
MODE
permitted
is
during
complete
of
heard.
is
not
ventilatory
period.
1.67-seconds, has
as
contains
depressed,
relieving
BREATH/TRIGGER
Switch
SIMV,
SELECTOR
section
(see
until
begin
in
Pushbutton
normal
reached
MANUAL
a
the
(OFF-A/C-SIMV-CPAP-CAL):
CPAP,
the
CPU
cycle
CPAP
the
In
pressure
and
baseline
TRIGGER
dedicated
MANUAL
mechanism,
Pushbutton
CAL
and
Switch
entitled "SELF-CHECK" CAL
SELF-CHECK)
position
Switch:
operation,
consisting
(zero
circuitry
contained
positions.
from
a
mode,
limited
PEEP).
or
backup,
capable
TRIGGER
OFF
successfully
is
(see
only
The
delivers
current
the
of
manual
cmH,0.
40
at
Each
if
of
delivers
this
within
protected
is
AIC,
to
before
section
MANUAL
MANUAL
one
INSPIRATORY
is
breath
time
CPU
a
providing
continuous
a
circuit,
against
MODE
The
SIMV,
initiating
completed.
entitled
BREATH/TRIGGER
BREATH.
delivered
MANUAL
A
MANUAL
a
occurs
failure
manually
gas
peak
limits
accidental
SELECTOR
causes
CPAP
or
operating
Calibration
TRANSDUCER
TIME/TIDAL
flow
a
at
BREATH
BREATH
the
in
triggered
at
flow
inspiratory
contact
Switch
the
the
in
of CALIBRATION).
Pushbutton
MANUAL
Each
VOLUME
LPM,
30
of
rate
primary
breaths.
liters
30
a
by
microprocessor
selected
the
be
cannot
triggered,
is
system.
For
per
pressure
circular
provides
mode.
pressure
airway
minute
is
BREATH
an
delivered
an
The
long
as
from
guard.
operating
perform
to
Ventilator
as
VISUAL
When within control
illuminates
MODE SIMV,
Vmin
when
INSPIRATION/EXHALATION
phase
INSPIRATION/EXPIRATION
POWER
the
LCD's
and
the
the
power
will
INDICATORS
activated,
the
LCD
panel.
When
green.
STATUS
Indicator:
CPAP,
or
Indicator:
a
PLATEAU
of
mechanical
INFORMATION
lower
CPU
validates
check
blank
when
The
POWER
STATUS,
except
activated,
Based
CAL.
Displays
VOLUME
and/or
left
section.
a
portion
a
SYSTEM
INFORMATION
1:
Line
and
for
the
ALARM,
upon
The
MODE
Minute
or
spontaneous
CENTER:
It
usable
of
SELF-CHECK
FAILURE
"EXT
ALARM
the
the
SYSTEM
Indicator
illuminates
source
PWR
INDICATORS
SYSTEM
ALARM
MODE
Indicator
Volume
Indicator:
breaths
display
The
of
power
Alarm
CENTER
ON"
and
SYSTEM
Selector
display
(in
liters)
FAILURE
Alternately
in
will
POWER
when
the
(see
(see
SELF-CHECK).
occurs.
or,
are
FAILURE,
Switch
will
in
the Alarm
each
operating
blank
INFORMATION
MODE
OPERATING
can
display
continuously
and
FAILURE
blank
ACV
displays
when
CHARGE
setting,
this
when
mode,
occurs.
the
mode,
a
SYSTEM
SELECTOR
POWER
The
the
following
displayed.
LED's
LED's
illuminate
indicator
a
SYSTEM
blanks during
"INSPIRATION"
blanks
during
FAILURE
CENTER
SWITCH
SELECTION
POWER
INFORMATION
messages:
All
indicators
which
red,
displays,
FAILURE
SIMV,
or
CAL.
Alarm
(PIC)
occupies
is
in
any
&
are
appear
the
respectively,
Alarm
CPAP
"EXHALATION"
The
position STOPPING)
displayed
elsewhere
CHARGE
occurs.
and
occurs,
a
2-line
except
CENTER
ASSIST,
CAL
on
the
LED
or
area
in
OFF
during
display
Line
Line
Line
Line
"EXT
PWR
denotes POWER external
Battery 30-minutes internal denotes
charge operation,
a
low
LOW/FAIL
power
Line
Line
Line
Line
Line
icon
batteries (1)
(see
"EXTPWR
1:
"EXT
1:
1:
"EXT
Blank
1:
ON"
denotes
external
is
connected.
Battery
2:
Battery
2:
2:
Battery
Battery
2:
Blank
2:
denotes
"OK"
remaining
of
can
operation
section
internal
batteries
PWR
CHK
operation
power
Alarm
icon
icon
icon
icon
(1)
charge
provide
internal
from
entitled
ALARMS)
are
LOW"
FAIL"
FUSE"
from
source
(see
section
or,
"OK
"LOW"
CHG"
"ON
"CHK
operation
or,
(2)
at
least
batteries
being
or,
or,
(flashing,
an
voltage
external
and
entitled
or,
or,
FUSE"
30-minutes
(flashing,
from
in
conjunction
having
or,
(2)
charged
internal
in and
see
description
power
works
in
ALARMS).
see
batteries
with
operation.
of
less
than
conjunction
currently
below)
source.
conjunction
description
external
have
"EXT
This
line
that
have power
Battery
30-minutes
with
external
less
or,
PWR
with
is
blank
below)
more
operation,
icon
of
power
than
30-minutes
LOW"
the
EXTERNAL
when
or,
than
"LOW"
remaining
no
VISUAL
INDICATORS,
of
operating
LED
illuminates,
blank
if
no
OPEN
or
or
MISSING
missing
1.
2.
3.
(cont'd)
time
capability. when
battery
fuses
is
sensed.
FUSES:
under
the
During normal &
Charge
Prior &
Charge
During Charge
Fuse
a.
EXT continues PIC
b. c.
When
EXT
PIC
d.
to
beginning
Fuse
a.
Operation
b.
PIC
normal
Fuse
a.
Operation
b.
PIC
Battery
charging
Line
Line
Line 1 flashes
opens
Line 2 flashes
current
The
following
PWR
Alarm
PWR
operation
POWER
operation
opens
(or
FAIL/DISCONNECT
from
displays
1
Mute/Cancel FAIL/DISCONNECT flashes
1
operation
opens
(or
begins
(or
is
continues
icon
"ON
CHG"
is
flowing
INFORMATION
circumstances:
from
external
is
removed):
"EXT
"EXT
from
is
removed):
from
internal
"CHK
from
external
from
battery
battery.
PWR pushbutton
CHK
external
EXT
external
icon
internal
removed):
is
displayed
into
the
power,
Alarm
FAIL".
Alarm
FUSE".
power,
battery. FUSE".
power,
the
power.
"CHK
battery
CENTER
the
External
activates,
is
pressed,
reset,
is
the
External
Battery
FUSE".
,
and
the
pack.
is
Power
operation
the
Operation
CHARGE
This
line
able
Operation
Power
will
to
identify
Operation
&
open
PEAK
AIRWAY
PEAK
AIRWAY
MEAN MEAN
DIGITAL pressure. markers Indicator
HIGH PRESSURE GRAPH.
The
AIRWAY
AIRWAY
MEAN
appear
and
AIRWAY
BAR
Its
absolute
at 0,
display
LOW
ALARM
Setpoint
4.
Prior
connected),
PRESSURE
PRESSURE
PRESSURE
PRESSURE
PRESSURE
GRAPH
range
50
and
will
blank
AIRWAY
Setpoint
indicators
to
beginning
a.
The
b.
The
tone
Indicator
to
(PIP-PEEP) X
Indicator:
is
from
100
when a SYSTEM
PRESSURE
Indicators
are
operation
and
the
Battery
ventilator
SYSTEM
sounds
(non-mutable).
Indicator:
display
Indicator:
following
the
Indicator
The
-10
to
cmH,O.
positioned
Vertical
from
Operation
will
not
operate,
FAILURE
Displays
Displays
TOTAL CYCLE
DIGITAL BAR
+100
ALARM
appear
according
the
will
blank
algorithm:
display
cmH,O.
resolution
FAILURE
Setpoint
as
small
internal
&
Charge
its
LED
illuminates
PEAK
when
the
MEAN
I
TIME
will
blank
GRAPH
Increment
is 2 cmH
Alarm
horizontal
to
their
respective
battery
(external
Fuse
LCD
will
not
and a continuous
AIRWAY
a
SYSTEM
AIRWAY
-
when a SYSTEM
occurs,
Indicators:
+
indicator
markings
O/bar.
lines
to
alarm
power
opens
(or
illuminate.
PRESSURE
FAILURE
PRESSURE.
PEEP
FAILURE
provides appear The
DIGITAL
The
HIGH
the
right
control
is
not
is
removed):
of
the
Alarm
Uni-Vent™
continuous
every
10
and
LOW
of
the
DIGITAL
settings
previous
occurs.
Alarm
display
emH3O,
BAR
GRAPH
AIRWAY
and
reposition
breath.
calculates
occurs.
of
airway
numerical
BAR
The
VISUAL
whenever
LOW
Alarm
Pay most adjacent
display
CHARGE current
Control
and
The called
Indicator OFF, each information OPERATING
period.
an
AIRWAY
is
triggered
Indicator: The
recent
digital
will
is
flowing.
Setting
lower
lines.
ALARM
LCD
display
the
ALARM
and
whenever
alarm.
During
INDICATORS,
alarm
control
PRESSURE
and
Pay
12-second
bar
graph.
blank
when a SYSTEM
Indicator: The
Indicators:
INDICATORS
allocates
MESSAGE
SYSTEM
an
alarm
Alarms
on
are
alarms,
ALARM,
an
ADVISORY
OPERATING
(cont'd)
is
adjusted.
ALARM
blanks
when a SYSTEM
indicator
period.
see
Airway
Markings
green
All
up
to
FAILURE
causing
categorized
section
the
ALARM
Setpoint
Setpoint
represents a continuous
pressure
along
FAILURE
CHARGE
control
4-lines
of
CENTER
LED's.
condition
as
OPERATING,
entitled
Indicator
ALARM
ALARMS
indicator
Indicator
FAILURE
amplitude
the
horizontal
Alarm
Indicator
settings,
alarm
(AMC).
pressing
The
other
ALARM
except
message
ALARM
than
LED
Each
FUNCTIONS).
MUTE/CANCEL
vertical
blinks
and
along
axis
occurs.
LED
MODE
information
alarm
Indicator SYSTEM NON-OPERATING,
remains
resolution
when a HIGH
Alarm
occurs.
updating
the
represent
illuminates
SELECTOR
is
displayed
FAILURE
illuminated,
is 2 cmH,O.
display
vertical
LED
axis
1-second
whenever
in a dedicated
in
illuminates,
occurs.
or
When
the
but
tums
the
PRESSURE
of
airway
is
calibrated
intervals.
sufficient
Switch,
the
ADVISORY
stops
ALARM
appear
area.
LCD
and alternately
An
audible
flashing,
The
respective
or
LOW
pressure.
to
coincide
The
P,y
battery
on
This
dedicated
except
for
audible
signal
Indicator
signal
(for
detailed
is
muted
during
PRESSURE
It
display's
Indicator
recharging
the
LCD's
the
ALARM
flashes
accompanies
during
the
mute
LED
OFF.
HIGH
with
upper
area
ON
or
the
the
is
and
an
BATTERY displays: BATTERY.
EXTERNAL
than SOURCE/CONNECTIONS
LOW consecutive PRESSURE
non-operating
setpoint
DISCONNECT displays: operating
HIGH breaths,
displays:
will
Alarm
APNEA deflections SPONTANEOUS seconds RATE
BATTERY
VDC.
10.9
PRESSURE
indicator,
DISCONNECT alarm
PRESSURE
2-seconds
or
HIGH
if
blank
LCD
Alarm:
(based
LESS
LOW/FAIL
AMC
The
POWER
When
Alarm:
breaths
have
and
PEAK
-
occurs.
alarm
during
Alarm:
occurs.
PRESSURE
non-operating
a
setpoint
ACV
been
BREATHING
a
on
THAN
Alarm:
LOW/FAIL
will
LOW
activated
causes
INSPIRATORY
a
Initiates
-
Alarm:
continuously,
indicator,
and
sensed.
30-second
6-BPM
- if
blank
Alarm:
the
EXT
or
Initiates
LCD
its
setpoint
The
PRESSURE
CHECK
Initiates
PEAK
-
alarm
during
SIMV:
When
or
moving
APNEA.
or
a
Initiates
when
when
RECHARGE/REPLACE
non-operating
a
Initiates
AMC PWR.
when
setpoint
PRESSURE
ALARM
a
CIRCUIT
when
causes
and
INSPIRATORY
occurs.
PRESSURE
a
initiates
activated
APNEA.
average).
when
displays:
AMC
The
fails
PIP
indicator
indicator
SETTINGS
disconnect
CONNECTIONS
exceeds
PIP
LCD
its
setpoint
The
when
the
CPAP:
AMC
The
low
alarm
EXTERNAL
to
TOO
will
ALARM
approximately
AMC
When
will
BATTERY
occurs.
external
will
is
power
blank
the
exceed
blink.
to
or
LOW
alternately
Alarm.
in
sensed
HIGH
the
setpoint
PRESSURE
indicator
SETTINGS
displays:
initiates
‘when
activated
blank
condition
battery
is
PACK,
sensed
(as
POWER
non-operating
a
if
PRESSURE
LOW
LOW
the or
indicator
will
19-seconds
a
if
activated
When
PRESSURE.
with
blink,
patient
DISCONNECT.
PRESSURE
to
TOO
alternately
APNEA
spontaneous
no
AMC
the
non-operating
sensed.
at
LOW
the
circuit.
Alarm
blink.
or
HIGH
Alarm.
have
CHECK
-
displays:
When
BATTERY
External
the
CHECK
- alarm
ALARM
AMC
the
The
HIGH
When
The
setpoint
When
HIGH
blink,
elapsed
PATIENT
breathing
APNEA
alarm
activated
LOW/FAIL
Power
POWER
occurs.
setpoint
displays:
AMC
PRESSURE
activated
will
AMC
for
activated
PRESSURE.
the
with
no
and
is
CPAP
-
occurs.
AMC
the
Jack),
for
LOW
blank
will
Alarm
the blank
consecutive
four
AMC
the
PRESSURE
LOW
pressure FOR detected
AVERAGE
or
two
ifa
AMC
if
The
for
less
is
LCD
non-
a
AMC
10-
VISUAL
HIGH activated
HIGH
O, AMC will
EXT When LOW/FAIL
FIO,
to
meet
SOURCE/SETTINGS/CONNECTIONS
PRESSURE ALARM indicators, ALARM
SETTINGS
occurs.
V;
displays:
will
INDICATORS,
PEEP
Alarm:
the
AMC
PEEP.
LOW/FAIL
displays:
blank
AIR
activated
Alarm:
Alarm:
blank
The
if a non-operating
LOW/FAIL
or
EXT
Initiates
its
proportion
ALARM
setpoints
and
both
SETTINGS
REVERSED
Initiates
DELIVERED
-
Vr
in
the
Initiates
displays:
AMC
will
Alarm:
O,
the
CPAP
Initiates
LOW/FAIL - CHECK
Alarm:
AMC
displays:
AIR.
The
when
of
the
SETTINGS
are
reversed
indicators
Alarm.
or
when
mode,
(cont'd)
when
the
HIGH
PEEP - INSPIRATION
blank
if a non-operating
when
alarm
occurs,
Initiates
EXT
AMC
will
the
oxygen
gas
mixture.
Alarm:
(i.e.:
low
adjacent
When
activated
PRESSURE
delivered
tidal
TIDAL
VOLUME
or
if a non-operating
inspiratory
low
pressure
OXYGEN
when
low
AIR
LOW/FAIL - CHECK
blank
if a non-operating
component
When
activated
or
FIO,.
The
Initiates
setpoint
to
the
ALARM
volume
is
DIGITAL
the
AMC
SETTINGS.
does
DOES
cycle
begins
before
BEGAN
alarm
occurs.
is
sensed
SOURCE/CONNECTIONS,
pressure
or
set
alarm
is
the
air
the
AMC
will
when
the
higher
BAR displays:
not
equal
NOT
occurs.
AMC
EQUAL
BEFORE
from
an
sensed
AIR
alarm
component
displays:
blank
HIGH
than
the
GRAPH,
PRESSURE
The
AMC
set
tidal
end
expiratory
END
PRESSURE
external
from
if a non-operating
PRESSURE
high
alternately
SET
oxygen
O,
LOW/FAIL
an
external
SOURCE/CONNECTIONS,
occurs.
of
the
AIR/OXYGEN
FIO) - GAS
ALARM
setpoint).
blink
ALARM
will
blank
if a non-operating
volume.
TIDAL
When
VOLUME
pressure
supply.
source
Both
during a PRESSURE SETTINGS - ALARM
plateaus.
PLATEAU
When
or
of
compressed
MIXER
MIX
ERROR.
alarm
occurs.
and
LOW
Control
activated
or
Vr.
activated
0).
EXT
PRESSURE
setpoint
the
AMC
The
When
The
AIR
is
unable
CHECK
alarm
AMC
or
the
AMC
air.
COMP volume.
AMC
INVERSE AMC
The
TRANSDUCER
is
RECALIBRATE MEMORY
Alarm:
When
will
blank
L:E
displays:
will
AMC
stopped
SYSTEM
occurred.
VENTILATOR
conditions will will an
INSPIRATION time
TIME
SIGH
prematurely.
CHECK
FAILURE
This
will
the
cause display:
abbreviated
exceed
to
TRUNCATED
breath
Initiates
activated
if
NON-OPERATING
Alarm:
INVERSE
blank
CALIBRATION
TRANSDUCER
alarm
FAIL
not
ALARM
VENTILATOR
description
TIME
3-seconds.
that
when
the
a
non-operating
Initiates
the
in
When
SYSTEM
or
Alarm:
usually
is
Alarm:
cause
LED
ADVISORY
TRUNCATED
TO
exceed
would
AMC
I:E
CPAP
LCD
the
to
the
of
is
It
3-SEC
the
-
activated
FAILURE
internal displays:
alarm
when
INSPIRATORY
mode,
or
FAILURE
LED,
related
Initiates
to
illuminate.
failure.
ALARMS
disabled
NOTE
-
3.0-seconds
compressor
occurs.
ALARMS
inspiratory
if
or
ABORT
the
TRANSDUCER
illuminates
hardware
a
to
when
blank,
DETECTED
TO
the
in
L-TIME
COMP
time
TIME
SYSTEM
a
Alarm:
displays:
AMC
Alarm
when
any
some
and
not
will
It
3-SEC
CPAP
&
truncated
is
output
COMPRESSOR
-
becomes
LONGER
FAILURE
Initiates
TRANSDUCER
CALIBRATION
occurs.
is
CPU
problem
seven
of
one
be
may
the
cause
followed
Alarm:
mode.
The
LE.
but
When
will
does
when
forced
and
operator
Initiates
AMC
OUTPUT
longer
THAN
Alarm
the
to
cause
will
ventilator
correctable.
SYSTEM
FAILURE
by
when
activated
will
activate
not
produce
not
its
expiratory
than
EXHALATION
occurs.
TRANSDUCER
CALIBRATION
ABORT.
shutdown
the
failure
FAILURE
CODE
control
the
if
blank
this
intended
LOW/FAIL
The
operation
LCD
causing
A
AMC
non-operating
a
alarm.
contribution
time.
TIME
CALIBRATION
will
AMC
or
blank.
to
conditions
VENTILATOR
to
LED
number
(a
#
settings
would
displays:
to
COMP.
or
activated
When
INVERSE
or
ABORT
a
illuminate.
-
if
blank
failure
CPU
occur.
FAIL
1
from
cause
INSPIRATION
occurs.
alarm
tidal
The
the
I:E.
process
a
has
These
Alarm
LCD
The
and
7),
to
inspiration
A
VISUAL
PLATEAU
volume
LESS
Vr
SETTINGS liters
60 DELIVER
PREVENTATIVE whichever
CONTACT
EXTENDED
has
PATIENT
EXTERNAL external SOURCE/CONNECTIONS occurs.
TOTAL and
of
at
displays: operating
INDICATORS,
VOLUME
by
more
THAN
per
minute
SET
occurs
CUSTOMER
occurred.
USE.
power
FLOW
internal
least
TOTAL alarm
compressor)
10
than
SET
VOLUME.
Alarm:
VOLUME
first.
NON-USE/STORAGE
When
The
POWER
operation.
BACKUP
LPM
FLOW
occurs.
(cont'd)
Alarm:
5%,
When
Initiates
(LPM).
MAINTENANCE
activated
and a total
When
or
V;
When
activated
SERVICE.
the
AMC
will
FAILURE
When
or
Alarm:
exceeding
tidal
BACKUP - CONTACT
Initiates
activated
The
whenever
SETTINGS.
AMC
blank
activated
EXT
when
the
AMC
will
activated,
Alarm: the
AMC
The
AMC
Alarm:
displays:
if a non-operating
Alarm:
the
PWR
FAIL/DISCONNECT.
Initiates
set
flows
volume
of
delivered
AMC
displays:
blank
in
the
sum
of
the
AMC
The
AMC
Initiates
displays:
will
blank
Initiates
EXTENDED
Initiates
when
by
at
AMC
+/-
least
whenever
displays:
the
50%
200ml
CUSTOMER
PRESSURE
PLATEAU
the
CPAP
mode,
the
flows
of
the
displays:
will
after
at
alarm
backup
for
V;
SETTINGS - I-TIME X FLOW
blank
in
2000-hours
PREVENTATIVE
if a non-operating
power-up
5-consecutive
for
after
NON-USE/STORAGE
occurs.
external
EXT
PWR
The
flow
sensor
this
alarm
SERVICE.
PLATEAU
VOLUME - DELIVERED
or
if a non-operating
selected
the
CPAP
of
6-months
power
FAIL/DISCONNECT - CHECK
AMC
to
tidal
gases
mode.
cumulative
MAINTENANCE
alarm
fails,
will
blank
detects
breaths.
the
operate.
The
AMC
volume
would
occurs.
of
continuous
-
or
is
if a non-operating
sum
There must
When
will
is
alarm
exceed a flow
use,
or
TEST
disconnected
of
the
activated
blank
less
than
set
tidal
VOLUME
occurs.
rate
UNABLE
12-months,
DUE
-
non-use/storage
BEFORE
during
POWER
alarm
flows
(Oz,
Air,
be a total
flow
the
AMC
if a non-
rate
of
TO.
CONNECTIONS
OXYGEN
pressure
D.LS.S.
AIR
(medical-grade D.LS.S.
Filter/Moisture
GAS LD.
TRANSDUCER:
ventilator
EXHALATION disposable
EXTERNAL
Supply
INLET:
reducer
Oxygen X D.LS.S.
INLET:
6'
Air,
OUTLET:
corrugated
circuit
ventilator
external
or
[CAUTION
[
DO
The
connect
NOT
EXTERNAL
Nominal
(medical-grade
Nominal
only),
Long,
Trap
hose
POWER
50
electric
or
interconnection.
for
whenever
pressure,
Low
(see
Figure
pressure,
Low
transducer
VALVE:
circuit
JACK: volt
11-15
external
POWER
50
PSI
only),
Oxygen,
input,
PSI
compressor
external
22mm
4).
fits
(see
hose
pressure,
Low
exhalation
External
power
sources
power
JACK
input,
D.I.S.S.
PTLOX;
6'
Long,
D.I.S.S.
(oil-less
Note:
air
is
provided
tapered
male
I.D.
3/16"
4).
Figure
fits
valve
power
via
source
rated
includes
pins
oxygen,
or
on-board
for
interconnection.
male-thread.
air,
To
tubing.
This
1/4"
Figure
(see
source
12
higher
containing
filtered).
and
protect
by
an
connection.
Connects
connector
tubing.
LD.
4).
connection. Power
VDC
than
male-thread.
aircraft
the
15
generated
Connects
Use
ventilator
electric
Connects
ventilator
colored
is
Connects
connector
This
Cable
or
volts
signals
for
air
Connects
(provided).
lower
Connects
source.
of
output
to
pressure
high
the
and
dirt
from
compressor.
disposable
to
pressure
green.
ventilator
is
COMMUNICATIONS
the
exhalation
colored
ventilator
11
than
to
output
of
Use
the
cylinder
air
hose,
condensate,
ventilator
transducer
aluminum.
clear
Universal
to
(see
volts
oxygen
high
D.L.S.S.
to
valve
SPECIFICATIONS).
cylinder
pressure
Air
an
use
circuit
control
Power
AC
reducer
Air
22mm
pressure
disposable
PORT
hose,
X
Inlet
port
to
]
|
1-10
CONNECTIONS,
(cont'd)
EXHALATION
HOSE
VENTILATOR
(Pins a
DC
The
(CONNECTS
VALVE"
EXHALATION
2, COMMUNICATIONS
connector
"EXHALATION
CLEAR
BARB
FITTING)
VALVE
3
and
5).
is
VALVE
TO
HOSE
АГАТ
mi
User’s
wired
jon
requiring
CABLE
as
follows:
~
-
sil
Figure
4.
communications
from
Pin Pin
Pin
TRANSDUCER
VENTILATOR
GREEN
>
3
Disposable
capability
Impact’s
Power
Power
-
1
Ground
-
4
Shield
-
6
Customer
HOSE
“TRANSDUCER”
HOSE
y
/
/
<
Ventilator
during
Service
(+)
(CONNECTS
BARB)
=
N
operation
Department.
Communications
Pin Pin
Pin
TO
Circuit
on
Send
-
2
Receive
-
3
Ground
-
5
battery
GAS
“GAS
———
power
DELIVERY
TO
OUT”
CONNECTOR
(15mm
X
22mm
may
(CONNECTS
J
HOSE
VENTILATOR
FITTING)
PATIENT
FEMALE
MALE)
special-order
COMMUNICATIONS exchange.
of
the
Uni-Vent™’s
switches,
Contact for
communications
Supervisor.
The
port
Universal
Impact’s
AC
communications
controls,
Switch,
+
updated
Control
Customer
PORT:
consists
Power
alarms,
and
control,
every
signals
protocols
of
a
standard,
Supply.
port
pressure
and
100-milliseconds.
allow servicing
Service
must
The
Model
The
connector
permits
signals.
alarm
information
Department,
be
made
754/754M
9-Pin,
Sub-D,
remote
personnel
for
in
writing
provides
Female,
is
wired
as
Pin
2
-
Send
Pin
3
-
Receive
Pin 5 -
Ground
monitoring
is
monitored
to
perform
additional
and
mailed
1-11
a
communications
RS-232C
follows:
of
ventilator
and
updated
“factory-level”
information
to
the
attention
interface
connector
functions
on
communications
every
device
of
which
and
second.
the
port
for
data
is
located
capabilities,
Pressure
calibrations.
protocols.
Customer
on
including:
signals
Service
the
side
are
Requests
Department
OPERATING
Uni-Vent™ AC
Power
DC
EXTERNAL POWER
12-volt
An
VAC, ranging The
The
Uni-Vent™ When a live external and
activates
from
Two 2AG, fuse battery CONTROLS,
additional
is
Supply
operating
battery.
optional
black
MODE
internal
external
effects
Universal
50
to
400
from
input
is
external
power
the
10A
fuse.
battery
icon
"ON
information.
designed
automatically
voltages,
Hz
16
to
lead
Selector
designed
failure
EXTERNAL POWER
power
fuseholders
The
CHG"
VISUAL
POWER
to
operate
within
JACK. A 12
AC
Power
(voltage
30
volts.
is
positive,
Switch
to
give
power
source
occurs,
to
the
external
are
fuse
closest
operation
will
INDICATORS
SELECTION
from
accepts
the
range
VDC
Supply
and
line
The
input
white
acts
as a master
operating
is
Uni-Vent™
source.
located
to
the
and
charging.
not
display
various
input
voltages
of
11 to
Power
(standard
frequency
leads
provided
is
negative.
power
power
applied,
adjacent
internal
automatically
FAILURE
to
top
cover
POWER
if
their
AND
CONNECTIONS
&
voltages
sensing
the
respective
and
from
15
volts
Cable
is
with
is
require a mating
Do
not
switch.
priority
batteries
switches
Alarm.
battery
effects
external
INFORMATION
STOPPING
frequencies
90
to
265
(negative
provided
the
automatic),
attach
to
When
compartment
fuse(s)
for
Model
the
Use
this
external
are
automatically to external
power
is/are
-
POWER
(see
volts
AC,
ground),
operation
754M),
and
permits
connector
braided
switch
power
rather
its
internal
power
door.
operation
CENTER
blown
INFORMATION
SPECIFICATIONS).
47
to
440
Hz.
may
be
connected
from a vehicular
is
operable
shield.
to
initiate
battery
Each
or
from
use
from
and
polarity
or
cease
than
its
internal
placed
on
pack
reappears,
fuseholder
and
battery
messages
missing.
nominal
external
must
charge.
for
operation
charging.
for
See
CENTER,
Its
Universal
directly
operation.
operating
"EXT
DESCRIPTION
to
the
source
or
115/230
DC
voltages,
be
observed.
battery
pack.
When
an
power
will
switch
contains a type
The
PWR"
for
external
other
and
OF
INITIAL
INITIAL
Uni-Vent
for
is
particular
SET-UP,
SET-UP:
easily
configurable
uses.
Figures
BOTTLED
REGULATED
50-PSI
SELF-CHECK
(Refer
OUTPUT
to
to
suit
5a
and
Sb
TEST
OPERATION
GAS
Figures
ALL
5a
most
applications.
depict
common
CONFIGURATIONS
PRIOR
Г
EXTERNAL
AL
{
AIR
&
TRANSDUCER
and
5b).
Additional
applications.
TO
PATIENT
GAS
SOURCES
IL
(
|
| |
OXYGEN
hoses,
fittings
FOR
CORRECT
CONNECTION.
|
CALIBRATION
and
adapters
BOTTLED
REGULATED
50-PSI
OUTPUT
may
GAS
be
required
HIGH
D.1.S.S.
PRESSURE
X
D.1.S.S.
AIR
HOSE
HIGH
PRESSURE
D.1.S.S.
OXYGEN
TRANSPORTABLE
SETUP
X
D.1.S.S.
OPTIONS
HOSE
Figure
5a.
3-1
Initial
Setup
Do
BOTTLED
REGULATED
50-PSI
HIGH
PRESSURE
0.1.5.5.
OUTPUT
[
a
n
Х
0..5.5.
AIR
GAS
-OR
HOSE
AIR
COMPRESSOR
|
|
EXTERNAL
-OR-
^
GAS
PIPED
GAS
(AIR)
©
SOURCES
BOTTLED
REGULATED
OUTPUT
50-PSI
5
OXYGEN
]
GAS
-OR-
n
PIPED
GAS
(OXYGEN)
HIGH
PRESSURE
DISS.XDIS.S.
OXYGEN
|
HOSE
L
p
AC
POWER | -OR- | POWER
SUPPLY
12
VDC
Figure
5b. Initial
3-2
Setup
NON-TRANSPORTABLE
SETUP
OPTIONS
SELF-CHECK
Uni-Vent™
SIMV,
or
checking
The
SELF-CHECK
verify
external
settings.
NOTE
If
external
PSI)
at
the
WARNING
SELF-CHECK that
the
TRANSDUCER
SELF-
CHECK information SELF-CHECK
Operation
Tf
Uni-Vent™
to
its
OFF
-
DO
NOT
SELF-CHECK exceeds CHECK
undergoes a self-checking
CPAP;
is
+/- 1 cmH,
will
or
from
not
repeated
process
power/internal
oxygen
time
position
SELF-CHECK
must
process
stored
failure.
begins
fails
ATTEMPT
will
cause
and/or
be
in
immediately
SELF-CHECK, a VENTILATOR
and
automatically
an
CAL
to
if
the
operator
consists
battery
external
is
performed
connection
to
sense a residual
memory.
then PATIENT
from
audible
Accordingly,
following
repeat
its
last
tone
process
ACV,
SIMV,
turns
the
of
interaction
status,
memory
compressed
performed.
with
the
disposable
is
open
to
airway
any
SELF-CHECK.
this
procedure.
USE.
alert
attendant
calibration. A
and
the
AMC
every
time
or
CPAP.
MODE
between
check,
gas
is
connected,
ambient
pressure.
residual
If
SELF-CHECK
personnel
TRANSDUCER
to
display:
its
MODE
After
the
Selector
ventilator
atmosphere.
FAIL
Switch
Uni-Vent™'s
pressure
each
circuit
Current
pressure
Alarm
if
the
Selector
initial
transducer
gas
Ignoring information
would
will
fails
pressure
CALIBRATION
Switch
SELF-CHECK
to
another
microprocessor
pressure
disconnected
result
occur.
again,
transducer
is
operating
calibration
must
from
this
requirement
is
compared
in a false
Return
contact
calibration
Alarm
turned
from
is
performed,
mode
and
peripheral
and control
be
at
least
the
patient.
could
to
reading
the
MODE
qualified
activated
"OFF"
self-
position.
panel
40-PSI
This
cause
previous
leading
Selector
service
"zero"
during
to
ACV,
circuitry
(+/-
calibration
to
personnel
baseline
to
2-
insures
the
a
Switch
SELF-
]
]
TRANSDUCER
CALIBRATE
VENTILATOR
DETECTED
FAILURE
e
pressure
the
only
If
TRANSDUCER
TRANSDUCER
disposable
The transducer, stored then Transducer
The
pressure.
transducer
in
serial,
a
in
5-minutes
every
calibration
TRANSDUCER
During
which
CODE
SELF-CHECK
transducer
CALIBRATION
CALIBRATION
ventilator
provides
turn,
non-volatile
thereafter
is
CALIBRATION
operation,
compared
are
CALIBRATION
TRANSDUCER
FAILURE
1
FAILURE!
calibration
-
connects
circuit
stream
a
EEPROM
through
essential
for
Uni-Vent™'s
its
to
-
portion
NOT
DO
to
data
of
memory.
Eagle™'s
correct
process
microprocessor
control
OR
-
if
SELF-CHECK
Alarm
is
related
Transducer Calibration:
SELF-CHECK
of
ATTEMPT
pressure-sensing
a
Uni-Vent™'s
to
AUTO
operation.
calibrates
panel
PATIENT
information
This
CAL
Uni-Vent™'s
will
settings.
-
not
to
the
proceed
fails,
USE.
element
microprocessor.
function,
respond
(transducer)
updated
is
it
or,
internal
according
VENTILATOR
DETECTED
FAILURE
SELF-CHECK
*
section
the
to
the
in
Transducer
automatically
be
can
pressure
to
calibration
manually
transducer
pressure
FAILURE
CODE
ventilator.
1
entitled
during
updated
signals
FAILURE!
information
SELF-CHECK,
to from
The
is
user.
the
by
atmospheric
the
TRANSDUCER
AUTOMATIC
During Process triggering
unstable, Alarm,
[WARNING
TRANSDUCER
disconnected Ignoring The
patient's
operation,
maintains a consistent
accuracy.
Eagle™
CODE 7 is
MANUAL
this
residual
work-of-breathing
e
e
CALIBRATION,
CALIBRATION
Eagle™
If
will
activated
CALIBRATION
CALIBRATION,
from
the
requirement would
pressure
1.
Set
MODE
2.
Calibration
performs
the
baseline
retry
AUTO
to
patient.
becomes
by
Selector
MODE = CAL
will
(cont'd)
(AUTO
transducer
drifts
CAL
assure
like
This
insures
allow
the
new
the
residual
Switch
Calibration...Please
take
CAL)
an
automatic
baseline
significantly
1-minute
patient
safety.
SELF-CHECK,
that
the
the
procedure
calibration
amount.
to
CAL.
approximately
calibration
over a wide
since
the
later.
If
AUTO
must
TRANSDUCER
to
sense
reference
3-seconds.
(albeit a false
The
AMC
Wait
of
its
pressure
temperature
prior
AUTO
CAL
be
performed
connection
any
residual
will
display:
When
finished,
transducer
range
to
CAL
fails
again, a VENTILATOR
with
the
is
airway
reference),
the
every
assure
display,
was
performed,
disposable
open
to
ambient
pressure
which
AMC
in
display
will
5-minutes.
monitoring, or
the
FAILURE
ventilator
atmosphere.
the
patient
increase
will
change
This
and
baseline
circuit
circuit.
your
to:
is
[NOTE
The PIC
AMC
messages.
TRANSDUCER
A
operating Selector
1,
steps RECALIBRATE
SYSTEM
USER
Model
The changed.
in
stored
particular
that
OFF.
turned
default,
VENTILATOR.
bottom
The
display
will
position
mode
Switch
2
PROGRAMS
Trigger
to
above.
3
and
FAILURE
754/754M
changes
Some
memory
characteristic.
User
Level
of
line
Hrs"
‘#44
MODE = CAL
information
CALIBRATION
before
position
any
When
TRANSDUCER
Alarm
contains
can
each
apply
programmable/selectable
Sensitivity,
menu
each
where
-
pertinent
CAL other
activated
or
occurs.
USER
a
stored
be
the
time
Temporary
Set
USER
in
"####"
Preventative
last
ABORT
completed.
is than
the
TRANSDUCER
PROGRAMS
in
ventilator
Spontaneous
PROGRAMS
represents
OK
operating
to
will
Alarm
The
then
and
CAL,
displays:
AMC
Eagle™'s
changes
memory
operated
is
not
are
characteristics
Flow,
cumulative
Maintenance
and battery
power
if
occur
procedure
CAL
returning
TRANSDUCER
CALIBRATION
allows
that
menu
temporary
- until
or,
in
stored
LCD
are:
Demonstration
contains
the
hours
performed.
was
of
charge
MODE
the
must
CAL,
to
it
ABORT.
certain
changes
user
the
memory
Backlight
Mode
following
operation
status
Selector
restarted
be
repeating
and
CALIBRATION
The
operating
not.
are
new
a
makes
last
will
and
Threshold
(DEMO),
information:
the
since
addition
in
Switch
by
AMC
characteristics
Program
program
until
default, and
to
turned
is
turning
process
the
ABORT
blank
will
changes
change
ventilator
LCD
TEST
above
the
an
to
MODE
the
described
- ifa
to
effecting
power
Contrast
BACKUP
be
that
in
get
is
USER
To Pushbutton
PROGRAMS,
enter
the
e
©
"##
"Version
Dys"
-
#.##"
USER
PROGRAMS
Switches
Follow
Press
adjacent
Press
pushbutton
mode
of
(cont'd)
where
"###"
Maintenance
-
where
while
tuming
on
screen
pushbutton
operation:
represents
was
#.##
indicates
menu,
simultaneously
the
MODE
prompts.
switches
switch
adjacent
cumulative
performed
the
software
Control
to
make
to
"EXIT"
A/C
SIMV
CPAP
CPAP
days
press
Switch
your
to
AIC
DEMO
SIMV
DEMO
DEMO
since
the
version.
MUTE
to
A/C,
selections.
leave
USER
last
Preventative
and
MANUAL
SIMV,
PROGRAMS
CPAP,
BREATH/TRIGGER or
CAL.
menu
and
return
to
selected
LCD
BACKLIGHT
defaults
can not
LCD may increase during "SAVE"
TRIGGER
breaths.
Trigger
6.0
be below
power
[NOTE
Under
carefully is cmH,0 work-of-breathing.
to
current
be
stored
pressed,
CONTRAST
choose
or
decrease
the
current
prompt
Eagle™'s
Level
cmH0
below
set
cmH,0
2
turned
is
most
consider
selected,
motion
to
compensate
in
memory
the
previously
to
increase
LEVEL
Sensitivity
in
0.5 cmH,0.
2
"OFF"
operating
THRESHOLD
ambient
operating
must
default
cmH,O
will
their
artifact
lighting
by
pressing
stored
DEFAULT:
or
decrease
to
the
setpoint
cycle.
be
pressed.
SENSITIVITY:
sensitivity
to
be
increments.
Sensitivity
affect
only
Trigger
conditions
operating
may
for
excessive
DEFAULT:
conditions
the
setpoint
Controls
color
is
seen
To
store
If
"SAVE"
is
temporarily
For
adjustments
mandatory
Sensitivity
Level
Eagle™'s
environment
cause
false
motion
to
pushbutton
remains.
the
contrast
on
the
changes
is
Controls
1.5
to
2.0
changed
spontaneous
breaths.
default
before
triggering
artifact,
CAL
Sets
the
assure
switch
LCD's
between
LCD.
in
not
pressed,
the
inspiratory
cmH,O
until
emH,0,
2
to
New
returned
is
Trigger
or
LCD's
adequate
adjacent
contrast
activated
Changes
memory,
the
below
operating
breaths
or
settings
to
Level
selecting a different of
breaths.
physiologic
backlight
display
to
setpoint.
LCD
can
be
the
pushbutton
previously
work-of-breathing
end
pressure.
power
occurring
higher,
cannot
default
its
Sensitivity
Conversely,
conditions,
setpoint.
illumination
the
on
screen
Depending
pixels
stored
stored
This
is
turned
in
SIMV
affect
will
saved
be
value.
should
Trigger
selecting a sensitivity
during
"SAVE"
upon
and
their
in
memory
switch
setpoint
setpoint
USER
OFF.
and
breaths.
all
memory.
in
be
Level
can
cause
When
programmed,
operation.
prompt.
viewing
background.
or
temporarily
adjacent
CPAP,
used.
Sensitivity.
to
remains.
for
patient-triggered
PROGRAM
It
is
adjustable
the
Adjustments
Once
User’s
excessive
it
Changes
If
"SAVE!
comfort,
the
sensitivity
are
inspiratory
user's
Any
used
on
screen
allows
from
made
operating
cautioned
If
1.0
cmH,O
above
2.0
1.0
to
cannot
to
USER
SET Spontaneous
adjustable power
DEMONSTRATION demonstration/training simulation Pressure
WARNING
DEMO
TEST
Activating VENTILATOR. Control
PROGRAMS,
SPONTANEOUS
Flow
from
10
is
turned
"OFF",
waveforms
Transducer
mode
is
not
BACKUP
Switch
the
BACKUP
VENTILATOR:
To
to
(for
to
"OFF"
(cont'd)
FLOW:
use
in
SIMV
60
LPM
in 5 LPM
Spontaneous
MODE purposes.
on
Eagle™'s
operation.
for
patient
VENTILATOR
leave
the
BACKUP
then
"ON"
Eagle™'s
or
Flow
(DEMO):
DEMO
LCD.
use.
Once
Permits
to
default
CPAP)
desired
to
increments.
is
returned
Permits
mode
allows
DEMO
operating
user
will
cause
VENTILATOR
operating
Spontaneous
be
temporarily
New
settings
to
its
default
user's
to
operate
user's
mode
does
power
to
test
the
BACKUP
it
to
function
function
mode.
Flow
changed
to
make
not
is
turned
is
cannot
value.
this
device
settings,
allow
Disconnect
"OFF"
VENTILATOR
as
described
and
return
60
LPM.
until
operating
be
saved
in a mode
adjustments,
DEMO
in
to
normal
This
in
Alarm
the
section
USER
power
memory.
that
trigger
activation
mode
is
and
MANUAL
entitled
operation,
PROGRAM
is
turned
Once
operating
is
quite
helpful
alarms,
or
Proximal
cancelled.
TRIGGER.
BACKUP
turn
the
allows
OFF.
and
MODE
It is
for
view
MODES
Your
Model
instruments.
simplify numbered
[NOTE
[IF
your
protocol
[NOTE
If
your
protocol
remains
Control
following
preset.
OF
754/754M
In
and
e
©
either
addition
speed
in
order
s
Step
e
Step
Step
calls
Step
Step
involves
fully
panel
settings
SELF-CHECK,
OPERATION
has
been
carefully
to
its
start-up.
of
use,
I:
Select
2:
Set
3:
Set
for
Set
4:
5:
Set
counterclockwise
clean,
Only
in
a
S-step
operating
VENTILATION
INSPIRATION
use
only
at
TIDAL
the
use
may
VOLUME
AIR/OXYGEN
without
be
adjusted
however,
designed
uncluttered
the
five
primary
sequence.
mode;
RATE
TIME
the
1:2
external
or
fully
at
any
the
1:2
I:E
appearance,
controls,
It
is
ACV,
LE
RATIO
MIXER
oxygen,
clockwise
time.
RATIO
to
ease
possible
SIMV
or
preset,
for
an
or
with
and
In
normal
and
the
learning
Uni-Vent™
common
to
initiate
CPAP
Step
FIO;
between
100%
control
use,
extents
transition
includes
to
most
applications,
operation
3
can
be
bypassed.
21
and
oxygen,
#5
adjustments
for
can
FIO,
the
be
bypassed.
and
commonly numerical
using
100
AIR/OXYGEN
are
typically
PRESSURE
as
are
few
associated
panel
made ALARMS
markings marked. as
3-controls.
MIXER
immediately
with
new
to
They
are
control
may
be
I
The
Model
[
WARNING
Functions
airway.
an
unprotected
tubes,
and
754/754M
CONTINUOUS
which
are
This
will
prevent
airway.
resuscitation
can
operate
ASSIST-CONTROL
ASSIST-CONTROL
SYNCHRONIZED
SYNCHRONIZED
WITH/WITHOUT
CONTROL
dependent
This masks where
WITH/WITHOUT
CONTROL
upon
"leaks"
applies
in
the
following
WITH/WITHOUT
WITH/WITHOUT
INTERMITTENT
INTERMITTENT
PEEP,
POSITIVE
VENTILATION
VENTILATION
accurate
from
primarily
the
pressure
distorting
to
face-to-mask-seal
modes:
VENTILATION:
VENTILATION:
PRESSURE
PEEP,
WITH/WITHOUT
AIRWAY
(apnea
pressure
use
MANDATORY
WITH/WITHOUT
MANDATORY
PRESSURE:
backup
readings
signals.
with
uncuffed endotracheal
integrity
WITH/WITHOUT
SIGH
WITH/WITHOUT
PLATEAU
PRESSURE
WITH/WITHOUT
of
ACV,
(backup
should
ventilator)
only
DO
NOT
is
frequently
PEEP,
PEEP,
VENTILATION:
SIGH
VENTILATION:
PLATEAU
SIMV
and
CPAP)
be
used
in
conjunction
use
pressure
tubes,
uncuffed
and
typically
PEEP
with a protected
dependent
tracheostomy
compromised.
functions
with
Loading...
+ 136 hidden pages