IMI Cornelius, Inc. QUEST NT 4000 Service Manual

QUEST NT
4000
Service Manual
®
Release Date: February 19, 2004
Publication Number: 720901117SER
Revision Date: November 23, 2004
Revision: C
Visit the IMI Cornelius web site at www.cornelius.com
QUEST NT 4000
SERVICE MANUAL
The products, technical information, and instructions contained in this manual are subject to change without notice. These instructions are not intended to cover all details or varia­tions of the equipment, nor to provide for every possible contingency in the installation, operation or maintenance of this equipment. This manual assumes that the person(s) working on the equipment have been trained and are skilled in working with electrical, plumbing, pneumatic, and mechanical equipment. It is assumed that appropriate safety precautions are taken and that all local safety and construction requirements are being met, in addition to the information contained in this manual.
To inquire about current revisions of this and other documentation or for assistance with any Cornelius product contact:
IMI Cornelius Inc.
Corporate Headquarters
One Cornelius Place
Anoka, MN 55303-6234
U.S.A.
tech.service@cornelius.com
In the U.S.A.:
phone:763-421-6120
800-238-3600
FAX:800-535-4231
Trademarks and copyrights:
Aurora, Cornelius, Decade, Hydro Boost, Sitco, Spirit, UF-1, Vanguard, Venture, Olympus, and Vista are registered trademarks of IMI Cornelius Inc.
Optifill trademark is pending.
This document contains proprietary information and it may not be
reproduced in any way without permission from Cornelius.
Printed in U.S.A.
Internet:
www.cornelius.com
Email:
Outside the U.S.A.:
phone:763-421-6120
FAX:763-422-3297
Copyright © 2004, All Rights Reserved, IMI Cornelius Inc.
TABLE OF CONTENTS
UNIT SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Concentrate Handling & Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Loading Concentrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Changing Concentrate Containers: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Brixing Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Checking/Adjusting the Brix Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Planned Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Daily . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Flush System: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Clean Splash Zones & Dispense Nozzles: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Weekly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Sanitize the Juice Dispenser: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Semi-Annual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Clean Water Inlet Strainer: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Clean Chassis Interior: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Check and Top-Off Water Ice Bath: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Annually . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Replace Pump Tubing: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Removing Pump Platform(s): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Replacing Pump Tubing: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
System Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Main Electrical Box Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Quest Series Juice Dispenser - 115VAC (based on 10 machines) . . . . . . . . . . . . . 15
Quest Series Juice Dispenser - 230VAC (based on 10 machines) . . . . . . . . . . . . . 16
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
IMI CORNELIUS INC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
UNIT SPECIFICATIONS
Quest Service Manual
Nameplate Data: Model QST 4000, 115 VAC, 4.5 amps, 1 phase 60 hertz, 6.75 oz. (191-193g) R-134a refrigerant. Test press: High side 400 psi (27.6 bar). Low side 100 psi (6.9 bar). Model QST 4000, 230 VAC, 3 amps, 1 phase 60hertz, 6.75 oz. (191-193g) R-134a refrigerant. Test press: High side 400 psi (27.6 bar. Low side 100 psi (6.9 bar).
Concentrate Storage: Four 0.8 gallon (3.0 liter) disposable bottles.
Clearance Recommended: 12 (30.48 cm) on top and 4 (10.16 cm) required in back for air circulation.
Electrical Connection: 6 ft. long (1.83 m) power cord with 3-prong plug attached to dispenser. Export
models have line cord less plug.
Power Supply: 15 amps at 120 volts dedicated power supply. 10 amps at 230 volts dedicated power supply.
Water Connection: 3/8 in. (0.95 cm) SAE male flare fitting on dispenser.
Water Supply Requirements: 80 psi (5.5 bar) maximum static pressure. 20 psi (1.4 bar) minimum
dynamic pressure; i.e., flowing pressure measured at dispenser water inlet with 3.0 ounces (88.7 ml) per second water flow, see part number 720901017 Rev. D for reference.
Ice Bank/Pull Down: Weight 14-16 lbs. Pull Down: 3.5 - 5 hours at 75°F (24°C)
© 2004, IMI Cornelius Inc. - 1 - Publication Number: 720901117SER
Quest Service Manual
CONCENTRATE HANDLING & LOADING
It is recommended that the concentrate be thawed in a refrigerated 35°F-40°F (1.6°C-4.4°C) compart­ment for a minimum of 48 hours prior to loading into the Quest Juice Dispenser.
WARNING: Concentrate must be completely thawed and within the temperature range of 35°F- 40°F (1.6°C-4.4°C) prior to loading. Failing to supply concentrate inside the recommended tem­perature range, especially below 35°F (1.6°C), will cause an out of brix drink (refer to the Brixing Procedure section for details).
LOADING CONCENTRATE
The Quest Juice Dispenser is designed to use either disposable juice concentrate containers or the optional Cornelius generic refillable container sold separately.
1. Thoroughly shake concentrate prior to use.
2. Place concentrate containers on the dispensing platform shelf inside the refriger­ated cabinet.
3. Engage the concentrate container by pressing it downward into the bottle adapter opening on the dispensing platform.
NOTE: Be sure to lubricate the o-ring seal on the container nozzle. This will ensure a good seal and allow the pumps to draw concentrated from the containers more easily. Failure to create a good seal at this connection may result in weak drinks and/or seepage of concentrate.
4. Prime each pump by closing the cabinet door and press each dispense button until concentrate flows from the dispense nozzles.
CHANGING CONCENTRATE CONTAINERS:
1. Open the cabinet door and move the valve handle from “Dispense” to “Flush”
2. Close the door. Depress and hold the dispense button until clear water flows from the dispense noz­zle.
3. Open the cabinet door and return the handle to “Dispense”.
4. Depress and hold the dispense button for 1-2 seconds. This will relieve water pressure from the concentrate pump system.
5. Load concentrate container (see
Loading Concentrate
).
Publication Number: 720901117SER - 2 - © 2004, IMI Cornelius Inc.
NOTE: If concentrate is not properly thawed, it will adversely affect the amount of concentrate dispensed. Thawed product should be between 35°F/1.6°C to 40°F/4.4°C.
SUPPLIES
1-Small 12 oz. cup (354.8 ml)
1-Large 21 oz. cup (621.1 ml)
1-Straw
Quest Service Manual
BRIXING PROCEDURE
Paper Toweling
1-Thermometer
1-Refractometer
Your will also need a flat blade screwdriver to turn a screw if brix adjustments are required.
NOTE: The refractometer shown above, P/N 511004000, is available through your local Cornelius Distributor.
CHECKING/ADJUSTING THE BRIX SETTING
The following instructions are for use with a refractometer.
1. Dispense approximately 8 oz. (236 ml) of drink and discard. Now draw a second 8 oz. drink.
2. Check drink temperature with a reliable thermometer (target is 35-45°F, or 1.6-7.2°C). Discard this drink after checking temperature.
NOTE: If drink temperature is not within the target range, refer to the basic troubleshooting sec­tion.
3. Dispense a 12 oz (354.8 ml) drink sample into a clean, dry cup. Thoroughly stir the sample using a straw.
4. Using a straw, transfer a small sample of finished drink to the refractometer lens (refer to operating instructions supplied with your refractometer). Check the brix reading against the brix chart below.
NOTE: The following brix chart is generic and intended for reference use only. Contact your fro­zen concentrate supplier for specific brix readings.
Brix Reference Chart
Flavor Ratio Brix Orange Juice 4+1 11.8 Grapefruit Juice 5+1 10.6 Cranberry Cocktail 4+1 13.5 Apple Juice 5+1 12.0 Grape 5+1 13.0 Lemonade 5+1 10.5 Tropical Punch 5+1 11.8 Sweetened Iced Tea 7+1 6.0 Pineapple Juice 4+1 12.8 Prune Juice 2+1 16.0
© 2004, IMI Cornelius Inc. - 3 - Publication Number: 720901117SER
Quest Service Manual
5. To change the brix setting, simply readjust the water flow rate. Located on each of the valve assemblies inside the refrigerated compartment are the adjusting screws for the water flow rate (one per valve).
If the brix reading is too high or low, rotate the appropriate water flow control according to the diagram below. Repeat steps 1-5 until the brix setting is achieved.
Lowers Brix by increasing water
Raises Brix by reducing water
IMPORTANT: When making changes to the water flow control, do not rotate more than 1/4 turn per adjustment. Additionally, prior to taking your next brix reading, “tap” the corresponding dis­pense button several times prior to drawing a sample. This will clear remnant drink from the dis­pense nozzle AND help move the flow control to its new setting.
Publication Number: 720901117SER - 4 - © 2004, IMI Cornelius Inc.
PLANNED MAINTENANCE SCHEDULE
DAILY
Flush System:
1. Move Dispense/Flush levers located on the platform assembly in the refrigerated cabinet to the “Flush” position. Place an empty cup on the drip tray below each dispense nozzle.
2. Close the door and depress each dispense button for 2-3 sec­onds or until clear water flows from each dispense nozzle.
3. Return the Dispense/Flush levers to the “Dispense” position.
4. Press each dispense button for 1-2 seconds to release the water pressure present in the concentrate pump system.
Clean Splash Zones & Dispense Nozzles:
1. On a daily basis, clean the external cabinet and splash areas using a clean damp cloth. Remove and wash the cup rest and drip tray using a mild dish soap.
2. Remove the dispense nozzles and static mixers by rotating each 90° and pulling down. Remove the mixing chambers by pulling straight forward. Wash using a mild dish soap.
Quest Service Manual
IMPORTANT: DO NOT wash nozzles, static mixers, or mixing chambers in a dish washer. This will distort the plastic and damage the o-rings. Additionally, do not soak them in sanitizing solu­tion longer than 2 minutes.
WEEKLY
Check concentrate to water brix ratio (refer to the Brixing Procedure in this manual).
Sanitize the Juice Dispenser:
Step 1 - Rinse With Hot Water
1. Prepare two 2 oz. packets of Stera-Sheen solving each packet in 1 gallon (3.8L) of potable water to insure 200 ppm of available chlorine.
IMPORTANT: Use potable water at 80°F-100°F (26.7°C-37.8°C). Water above this range will break­down the chlorine count and minimize sanitation.
2. Remove the juice concentrate containers and place them in separate refrigerated compartment.
3. Flush the system by following the instructions in the Daily Section.
4. Fill a clean empty concentrate container with one quart of extremely hot tap water, approximately 140°F (60°C) and place into unit. Dispense all of the hot water into a large container. Repeat for the remaining dispense valves.
5. Remove the mixing chambers, nozzles, and static mixers. Rinse in hot water to remove excess pulp and concentrate.
6. Place the mixing chambers, nozzles, and static mixers in a separate container of sanitizing solution and agitate vigorously. Allow the parts to soak for two minutes. Rinse thoroughly with fresh tap water.
7. Reinstall the static mixer, nozzles, and mixing chambers.
Green Label sanitizing solution (or similar brand) by dis-
© 2004, IMI Cornelius Inc. - 5 - Publication Number: 720901117SER
Quest Service Manual
Step 2 - Sanitize Pump System
1. Fill a clean concentrate container with 2 quarts (1.9L) of fresh sanitizing solution.
2. Place handles in the “dispense” position and close the door.
3. Press and hold the dispense button for 90 seconds then stop. Allow sanitizing solution to remain in the lines for 5 minutes.
4. After 5 minutes, dispense the remaining sanitizing solution.
Step 3 - Prepare Dispenser for Use
1. Replace sanitizing solution container with a concentrate container and close the door.
2. Press and hold the dispense button until juice appears from the nozzle. Next dispense and discard at least two 8 oz. (236.6ML) cups of juice in order to prime the system and prepare it for operation.
SEMI-ANNUAL
CAUTION: The following procedures require removal of the dispenser side panel(s). Disconnect the power cord from the receptacle prior to proceeding.
Clean Water Inlet Strainer:
1. Remove the right side panel from the dispenser.
2. Turn off the water supply to the dispenser.
3. Remove the access port from the “Y” shaped water inlet solenoid located on the right side of the dispenser.
4. Clean and reinstall the stainless steel water strainer.
Clean Chassis Interior:
1. Clean the condenser cooling fins.
2. Clean the air inlet grilles located on the rear and top panels of the dis­penser.
3. Clean the interior base.
4. Wipe the fan blade clean.
5. Reinstall the right side panel, turn on the water supply, and plug the dispenser into the power recep­tacle.
Check and Top-Off Water Ice Bath:
1. Remove the drip tray and lower splash panel.
2. If the Ice bath level is below the “Full” indicator, top it off with water. Refer to the Filling the Ice Bath procedure in the Installation section of this manual.
Publication Number: 720901117SER - 6 - © 2004, IMI Cornelius Inc.
ANNUALLY
Replace Pump Tubing:
A replacement pump tubing kit, part#45098, is available. The kit consists of one pre-cut length of pump tubing, two white plastic hose clamps, and instructions.
Removing Pump Platform(s):
1. Remove the concentrate containers from the dispenser and place them in a refrigerator.
2. Remove the cabinet shelf that the con­centrate containers sit on.
3. Flush the system prior to removing the pump platform (refer to the Daily sec­tion located at the beginning of the Planned Maintenance Schedule). Remove the dispense nozzles and static mixers.
4. Unplug the water line quick disconnect by pressing the gray button (see Fig. 1&2).
5. Slide the locking latch forward. Lift the platform slightly and pull forward to gain access to the electrical connector (see Fig.1).
6. Unplug the electrical connector by squeezing the locking tabs on either side and pulling out the con­nector. Lift and remove the pump platform (see Fig.1).
Quest Service Manual
Replacing Pump Tubing:
1. Remove pump platform (refer to the Removing Pump Platform(s) procedure located earlier in this section).
2. Remove the two white plastic hose clamps from the pump tube con­nections (see Fig. 1&2). Remove the concentrate delivery tubes from the hose ends.
3. Loosen and remove the four screws from the pump body (see fig 3).
4. Remove the rear pump body half only to reveal the pump tube and rollers.
5. Remove the old pump tube from the pump body. If the roller assem­bly comes out with the tubing, place it back into the pump housing being sure to align the roller assembly shaft keyway to the motor shaft so that the two interlock.
6. Firmly press the new hose into the pump body around the roller assembly, being sure to keep the protruding ends even with each other.
7. Once the tubing is in place, hold the tubing with one hand, cap­ture the lower part of the tubing with the outer housing, then proceed to capture the shaft of the roller assembly and push the rear pump housing into place. Make sure to capture the tubing within the body and not pinch it between the halves. Do not use any tool other than your finger tips to manipulate the tubing into the housing or you may damage the tube.
Fig.1
Fig. 2
© 2004, IMI Cornelius Inc. - 7 - Publication Number: 720901117SER
Quest Service Manual
8. While holding the pump halves together with your hand, reinsert the four screws and tighten using a criss-cross pattern as shown. The screws should be tightened about 1/4 turn beyond snug.
9. Insert the two concentrate delivery tubes into the pump tubing ends and secure them using the new hose clamps supplied in the kit. Be sure to use pliers to squeeze and tighten the hose clamps.
Fig. 3
Publication Number: 720901117SER - 8 - © 2004, IMI Cornelius Inc.
Quest Service Manual
TROUBLESHOOTING GUIDE
The following pages contain trouble-shooting information intended to aid an experienced service person in diagnosing operational problems that may occur. For further assistance, contact the IMI Cornelius Technical Services department at 1-888-248-5568 (630-539-6850 outside the United States) between the hours of 7:30A.M. and 5:00P.M. Central Standard Time. You must have the model and serial number (Located on the right side of the dispenser) prior to calling.
PROBLEM PROBABLE CAUSE REMEDY
Totally Inoperative No power to dispenser due to
tripped circuit breaker
Loose or broken power supply connection inside dispenser.
No cooling Line voltage is not within +/-
10% of nameplate specifications causing compressor overload to trip
No water in water ice bath or water level extremely low exposing the ice bank sensing probe
Defective Ice Bank Control or sensing probe
Cabinet fan is inoperative result­ing in warm concentrate (water continues to cool)
Compressor short cycles on overload
Reset circuit breaker. Confirm that breaker is correct size and no other equipment is operating on the same circuit. Also confirm that supply voltage is +/- 10% of name­plate specifications.
Repair connection.
Contact an electrician
Fill ice bath to proper water level
Replace
Replace
Excessively high discharge
pressure due to restricted con­denser or inoperative condenser fan motor
Compressor starts but hums and trips overload
Defective compressor overload or start capacitor
Compressor starts but does not switch off of start winding
Refrigerant leak Repair leak, evacuate and
No water dis­pensed, concen­trate only
© 2004, IMI Cornelius Inc. - 9 - Publication Number: 720901117SER
No water to dispenser Restore Water
Water supply line inside refriger-
ated cabinet disconnected from pump platform
Seized or shorted compressor, replace
Test and replace
Relay or compressor is defec-
tive. Test and replace faulty item
recharge system
Reconnect
Quest Service Manual
PROBLEM PROBABLE CAUSE REMEDY
No water dis­pensed, concen­trate only
(continued)
No water and no concentrate, refrigeration is working
Water solenoid located on pump platform clogged or defective
Main water solenoid/strainer located at the rear of dispenser is clogged, binding or defective
Disassemble and clean sole­noid. Replace if necessary.
Remove and clean strainer. Confirm 28VDC is present at sole­noid during dispense. Confirm solenoid coil is not open. Disas­semble and clean solenoid.
Water supply pressure is greater than 80 psi (5.5 bar) forc-
Add external regulator and lower pressure to 50 psi (3.5 bar)
ing the brix flow control closed
Freeze-up of water coil in ice bath
Unplug dispenser and allow 2-4 hours to thaw. Check operation of agitator motor and ice bank con­trol.
Refrigeration system may be low on charge resulting in a deformed ice bank and freeze-up of the water coil in the ice bath.
Black service switch located on
Tur n o n sw i t c h
the rear of the cabinet door in OFF position
No concentrate dispensed, water only
White door switch open Door switch must be closed in order to dispense. Check switch operation and replace if neces­sary.
6.25 amp fuse inside front elec-
trical box blown
Replace with 6.25, 250VAC slow blow fuse and test
No output from Transformer Confirm transformer output by of 26VAC +/- 2. Replace transformer if necessary.
Defective voltage regulator
board (VRB) located inside front electrical box
Measure across the VDC output of the board. There should be 28VDC present when the dispense button is pressed. Replaced VRB if necessary.
Defective dispense push button
Test and replace if necessary
or portion control board
Concentrate container not fully
engaged into receptacle on pump
Refer to Concentrate Loading section of this manual
platform
Dispense/Flush lever in FLUSH
position
Move lever to DISPENSE posi­tion
Concentrate too cold, not properly thawed
Publication Number: 720901117SER - 10 - © 2004, IMI Cornelius Inc.
Concentrate should be 35-40°F (1.7-4.5°C) prior to loading
Quest Service Manual
PROBLEM PROBABLE CAUSE REMEDY
Defective pump motor Replace pump motor
Warm Drink Ambient air around dispenser is
too warm
Excessive demand on dispenser Add water pre-cooler or second
Dirty condenser coil Clean condenser coil
Inoperative condenser fan Replace condenser fan motor
Defective Ice Bank Control Test and replace if necessary
Loss of refrigerant charge due to
leak in system
Water continu­ously drips from nozzle when in OFF mode
Concentrate
Main water solenoid at base of unit or water solenoid on pump platform not shutting off tightly
Cabinet fan inoperable Check/replace fan
warm, water cold
Agitator motor/pump inoperable or restricted
Loss of refrigerant charge due to leak in system
Relocate dispenser
dispenser
Repair leak and recharge system
Clean solenoid(s), replace parts
as necessary (refer to the Planned Maintenance Section)
Check/replace agitator motor
Repair leak and recharge system
Brix Problem Water supply pressure too low,
less than 20 psi (1.4 bar) flowing water pressure fluctuates sharply
Water flow control binding or spring is defective
Improperly thawed concentrate. Brix changes as the concentrate
Correct water supply problem to ensure a constant 50 psi (3.5 bar) flowing to the dispenser
Clean and/or replace parts as necessary
Concentrate should be 35-40°F (1.7-4.5°C) prior to loading
temperature changes (concen­trate becomes thinner as temper­ature rises)
Pump Inoperative Pump motor defective 28VDC should be present at
pump motor during dispense. If voltage is present and motor does not start, replace pump motor
No power to transformer or no 28VAC output from transformer
Confirm transformer has line volt­age present on primary side. If no 28VAC output from the secondary replace transformer
© 2004, IMI Cornelius Inc. - 11 - Publication Number: 720901117SER
Quest Service Manual
PROBLEM PROBABLE CAUSE REMEDY
Machine contin­ues to dispense after dispense button is released or dispenses with­out operator input
Defective voltage regulator board (VRB) located inside front electri­cal box
Defective dispense control board (Push button or portion control)
Push button or portion control pad stuck in on position
Relay on voltage regulator board (VRB) stuck on.
Confirm board produces 28VDC present when the dispense button is pressed (refer to the Electrical Box Wiring Diagram for VDC out­put location). Replace VRB if nec­essary
Test and replace if necessary
Disconnect the wire harness from
the rear of the portion control and close the door. If unit does not dispense on its own the dispense control board is bad (stuck on)
Disconnect the 4-wire harness from the lower right corner of the VRB. If the unit continues to dis­pense on its own the VRB is defective (relay stuck on)
Publication Number: 720901117SER - 12 - © 2004, IMI Cornelius Inc.
SYSTEM WIRING DIAGRAM
Quest Service Manual
DIAGRAMS
© 2004, IMI Cornelius Inc. - 13 - Publication Number: 720901117SER
Quest Service Manual
MAIN ELECTRICAL BOX WIRING DIAGRAM
Publication Number: 720901117SER - 14 - © 2004, IMI Cornelius Inc.
Quest Service Manual
SPARE PARTS
QUEST SERIES JUICE DISPENSER - 115VAC (BASED ON 10 MACHINES)
Part Number Description Qty
729011132 Kit, Parts - Stocking 1 560004360 Condenser Fan, Motor 115VAC 1 18071 Seat, Armature - Water Valve 2 18367 Spring, Coil S/S - Water Solenoid 2 31314 Fluorescent Bulb 2 45028 Transformer, 115VAC 1 45098 Pump Tubing Kit 4 45432 Inner Door Switch 1 7215323 Armature - Water Solenoid 1 7245699 Flow Control/Bonnet Assembly 2 620710107 Rubber Gasket 2 19695001 Guide, Water Solenoid 1 31525037 O-Ring - Dispensing Nozzle 2 45012102 Voltage Regulator Board 2 45016001 Pump Gear Motor 2 45026100 Bottle Adapter O-Ring Kit (all O-rings in bottle
adapter) 45091001 Main Water Solenoid Valve 1 45728001 Pump Rotor Assembly 1 48520001 Coil, 24VDC, Valve Block 1 48979103 Spool & Sleeve Asmb w/ 31525-060 O-Ring 2 59328001 Fuse, 6.25 Amp, 250VAC 4 440000902 Global Ice Bank Control (115VAC) 1 720500014 Drip Tray Grill 1 720500104 Drip Tray 1 720500710 Door Gasket 1 720202009 Catch w/switch 1 720502800 Agitator Motor 1 720503201 Compressor, 115V/60Hz, R-134a 1 720503206 Overload, Compressor, 115VAC 1 720503207 Relay, Compressor, 115VAC 1 720503208 Capacitor, Compressor, 115VAC 1 720506104 Cabinet Fan, 115VAC 2 720507101 Dispense Nozzle, Static Mixer, & O-Ring 1 720508801 Push Button Control Board 1 720508802 Portion Control Board 2 720511606 Ballast Board 1 720703502 Foam Gasket, Lower Door 1 720704001 Latch, Door 1 720501504 Condenser 1 720504904 Frame Assy, Base 1
1
© 2004, IMI Cornelius Inc. - 15 - Publication Number: 720901117SER
Quest Service Manual
QUEST SERIES JUICE DISPENSER - 230VAC (BASED ON 10 MACHINES)
Part Number Description Q
t
y 729011132 Kit, Parts - Stocking 560004361 Condenser Fan, Motor 230VAC 1 18071 Seat, Armature - Water Valve 2 18367 Spring, Coil S/S - Water Solenoid 2 31314 Fluorescent Bulb 2 720500794 Transformer, 230VAC 1 45098 Pump Tubing Kit 4 45432 Inner Door Switch 1 7215323 Armature - Water Solenoid 1 7245699 Flow Control/Bonnet Assembly 2 620710107 Rubber Gasket 2 19695001 Guide, Water Solenoid 1 31525037 O-Ring - Dispensing Nozzle 2 45012102 Voltage Regulator Board 2 45016001 Pump Gear Motor 2 45026100 Bottle Adapter O-Ring Kit (all O-rings in bottle
adapter) 45091001 Main Water Solenoid Valve 1 45728001 Pump Rotor Assembly 1 48520001 Coil, 24VDC, Valve Block 1 48979103 Spool & Sleeve Asmb w/ 31525-060 O-Ring 2 59328001 Fuse, 6.25 Amp, 250VAC 4 440000901 Global Ice Bank Control (230VAC) 1 720500014 Drip Tray Grill 1 720500104 Drip Tray 1 720500710 Door Gasket 1 720202009 Catch w/switch 1 720502803 Agitator Motor 1 720503202 Compressor, 230V/50Hz, R-134a 1 720503209 Overload, Compressor, 230VAC 1 720503210 Relay, Compressor, 230VAC 1 720506104 Cabinet Fan, 230VAC 2 720507101 Dispense Nozzle, Static Mixer, & O-Ring 1 720508801 Push Button Control Board 1 720508802 Portion Control Board 2 720511606 Ballast Board 1 720703502 Foam Gasket, Lower Door 1 720704001 Latch, Door 1 720501504 Condenser 1
1
Publication Number: 720901117SER - 16 - © 2004, IMI Cornelius Inc.
IMI CORNELIUS INC.
Certificate of Warranty
ONE YEAR LIMITED BEVERAGE EQUIPMENT WARRANTY
IMI Cornelius Inc. warrants to the original commercial purchaser/user, that any commercial product of its manufacture bearing the name Wilshire will be free from defect in material and/or factory workmanship, and that if properly installed, maintained, and serviced in accordance with the Service Manual furnished with the product, it will perform adequately under normal use. This product warranty shall be effective for a period of one year from the date of original installation or 15 months from the date of original shipment by IMI Cornelius, whichever period elapses first.
IMI Cornelius Inc.’s obligation under this warranty is strictly limited to the replacement of any parts which the purchaser/user returns to IMI Cornelius Inc.’s factory, transportation costs prepaid, and which IMI Cornelius finds to be defective in workmanship and/or material within the warranty period. The serial and model numbers and date of original installation of the product must be given. No part or assembly which has been subjected to accident, alteration, or misuse or which is not installed, maintained, or serviced in accordance with Service Manual furnished with the product, or which is from a machine on which the serial number has been removed, shall be covered by this warranty. This warranty does not provide for service calls from factory representatives or from any other agencies and shall not include charges of any nature.
IMI Cornelius Inc. will accept a part, parts, or equipment freight prepaid and return same freight collect to the sender within the continental U.S. or port of export within the continental limits of the U.S. IMI Corne­lius Inc. is not responsible for international freight, customs fees or duties at country of destination.
Quest Service Manual
WARRANTY
ADDITIONAL FOUR YEAR LIMITED WARRANTY ON COMPRESSOR
This warranty shall be effective for a period of four (4) years from the expiration of the above warranty.
The hermetically sealed refrigeration compressor is covered by the above one year limited warranty. In addition to that warranty, if the compressor fails because of a defect in materials or workmanship during the second through fifth year from the date of installation, IMI Cornelius Inc. will repair or, at its option, replace the compressor. Labor charges and the cost of relays, overloads, and capacitors are not included.
THIS WARRANTY DOES NOT COVER DAMAGED CAUSED BY LACK OF PREVENTATIVE MAINTE­NANCE, IMPROPER INSTALLATION, ACCIDENT, MISUSE, NEGLIGENCE, ALTERATION, FIRE, FLOOD, OR ACTS OF GOD. In those jurisdictions where liability for damages cannot be disclaimed, original purchasers recovery shall not exceed the cost of the warranted product.
IMI CORNELIUS INC. ASSUMES NO LIABILITY FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND, INCLUDING, BUT NOT LIMITED TO, SPOILED PRODUCT, LOST PROFITS, OR DAM­AGE TO OTHER PROPERTY.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER WRIT­TEN, ORAL, OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, AND SUPERSEDES AND EXCLUDES ANY ORAL WARRANTIES OR REPRESENTATIONS OR WRITTEN LANGUAGE IN ANY MANUAL, LITERATURE, ADVERTISING BROCHURE OR OTHER MATERIALS NOT EXPRESSLY DESIGNATED IN WRITING AS A “WAR- RANTY”.
© 2004, IMI Cornelius Inc. - 17 - Publication Number: 720901117SER
Quest Service Manual
The table below will be used by IMI Cornelius Inc. as a STANDARD service call guide to determine fair and reasonable labor charges for warranty repairs. Charges in excess of these rates will be subject to review and/or adjustments.
The labor warranty referenced in the
Certificate of Warranty
in this manual applies to the replacement of the defective part. IMI Cornelius Inc. will not accept labor warranty claims for water leaks applicable to the installation, clogged drains, adjustments of any kid including regulators, pumps, thermostats, ratio, or brix settings, stratification issues, preventative maintenance, sanitizing, etc. IMI Cornelius Inc. will con­sider only actual service time on the equipment. Charges for mileage, holiday pay, night charges, and overtime will not be considered. All labor claim that are fare, reasonable, and within the terms of the war­ranty and allotted repair times will be paid in U.S. Dollars.
Any questions regarding the warranty procedures can be directed to our Technical Services group at 1­800-238-3600 (763-421-6120 outside the United States).
ITEM # PAG E# DESCRIPTION MAX HOURS
*Recovery and pump down require at least two hours depending on contamination and is not included in the allotted repair time shown above.
Publication Number: 720901117SER - 18 - © 2004, IMI Cornelius Inc.
IMI Cornelius Inc.
www.cornelius.com
Loading...