HYDAC HMG 3010 User Manual

Portable Data Recorder
HMG 3010
Operating Manual
(Original Manual)
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 2
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 3
1 Tasks and Functions of the HMG 3010 .......................................................4
1.1 Standard functions of the HMG 3010..............................................................4
1.2 CAN functions of the HMG 3010.....................................................................6
2 About Your HMG 3010 ..................................................................................7
2.1 Items supplied.................................................................................................7
2.2 Power supply ..................................................................................................7
3 Technical Safety............................................................................................7
3.1 Safety information...........................................................................................8
4 Commissioning .............................................................................................9
4.1 Operating Elements ........................................................................................9
4.2 Connections..................................................................................................10
4.3 Switching on the instrument..........................................................................11
5 Getting Started Guide.................................................................................13
6 Detailed Operating Instructions................................................................18
6.1 Settings Menu...............................................................................................18
6.1.1 Change settings of measurement channels ...........................18
6.1.2 Changing Display Settings .....................................................29
6.1.3 Zeroing Measurement Channels ............................................31
6.1.4 Changing the Initial Setup ......................................................33
6.1.5 Administer settings (save, load, ...) .............................36
6.2 Recording Menu............................................................................................38
6.2.1 Starting a Recording ..............................................................38
6.2.2 Viewing Recordings ...............................................................56
6.2.3 Deleting Recording.................................................................67
6.3 Extras Menu..................................................................................................68
6.4 SMART sensors............................................................................................70
6.4.1 Connecting SMART Sensors .................................................70
6.4.2 SMART Sensors Menu ..........................................................71
6.5 CAN Functions..............................................................................................74
6.5.1 Electrical connections ............................................................74
6.5.2 Main menu CAN bus ..............................................................75
6.6 Technical data...............................................................................................78
7 Cleaning.......................................................................................................80
8 Accessories.................................................................................................80
9 Examples of Connections ..........................................................................82
9.1 Pin assignment .............................................................................................83
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 4
1 Tasks and Functions of the HMG 3010
1.1 Standard functions of the HMG 3010
The HMG 3010 portable data recorder is a hand-held unit for measurement and data acqui­sition tasks on hydraulic and pneumatic systems. Applications are primarily in maintenance and servicing, troubleshooting and test rigs.
The HMG 3010 can monitor signals from up to ten sensors simultaneously. The unit has 5 input jacks for connecting the sensors; if necessary this number can be doubled using a Y adapter (if more than 5 sensors are required). HYDAC ELECTRONIC provides suitable HSI sensors (HYDAC Sensor Interface) for pressure, temperature and flow rate which are automatically recognised and for which the measuring range and unit are set by the HMG 3010 - the easiest choice for customers.
You also have the option of connecting SMART sensors to the HMG 3010. These represent a sensor generation from HYDAC which can output several different measured variables. These sensors are also automatically recognized by the HMG 3010 via HSI, and the meas­ured values including the measurement unit are shown in the display. The HMG 3010 can render up to 32 measurement channels in total. Depending on the sensor type, the sensor has an internal memory for the storage of re­corded measurement data on a long-term basis. In addition, depending on the sensor type, preset parameters can be changed and stored in the internal memory as a sensor configu­ration.
Older model HYDAC sensors or commonly available sensors made by other manufacturers can also be used. However, these sensors do not have any automatic sensor detection, consequently the initial set-up has to be entered manually.
The operator can access all of the unit’s functions and settings by means of clearly pre­sented selection menus. The HMG 3010 has a combination keypad for entering numeric values and text, similar to a mobile phone.
The HMG 3010 is designed in particular for recording typical measurements (pressure, temperature, flow rate) in hydraulic and pneumatic systems. A variety of other measurement tasks can be performed using additional inputs for voltage measurement. For example, checking the actuation of a switching valve or plotting the characteristic curve of a valve setpoint. In addition, it is also possible to determine differential values between the measured values of individual sensors. One example of this is taking a flow rate meas­urement using a differential pressure orifice.
In addition to the analogue measurement inputs, the HMG 3010 features two digital inputs (e.g. for measuring frequency or speed), thus expanding the unit’s range of potential appli­cations. When taking measurements of rapid, dynamic machine processes, all 8 analogue input signals can be recorded simultaneously at a rate of 0.5 ms.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 5
One special feature of the HMG 3010 is its ability to also monitor highly dynamic processes in a machine. To achieve this, two input channels can record measured values at a rate of 0.1 ms. This function of course requires suitable fast sensors to be used.
The most attractive feature of the HMG 3010 is its ability to record the dynamic processes of a machine in the form of a measurement curve and render them as a graph — and, more­over, online and in real time.
The HMG 3010 is equipped with specially developed software providing for fast data collec­tion and processing. A measurement curve can consist of up to 500,000 measured values. The HMG 3010 memory is capable of storing at least 100 such measurement curves.
In addition to enabling simple measurement curves to be recorded, the HMG 3010 also has other functions enabling event-driven measurements to be taken and event logs to be re­corded. Various trigger options are available for triggering events.
Apart from measurement curves, the HMG 3010 can store user-specific settings (user pro­files). The main advantage of this is to enable identical measurements of various equipment items to be repeated for the purpose of preventive maintenance. All the user has to do is retrieve the relevant user profile from the HMG 3010’s memory.
The HMG 3010 has a 3.5" color display with graphics capabilities, which can show the measured values either as text or as measurement curve. Individual measured values can be displayed in a large format (7-segment format), enabling them to be read from far away.
The HMG 3010 also has a variety of user-friendly features for displaying, evaluating and processing measured values:
Table
Graph
Scaling
Ruler
Tracker
Zoom
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 6
The HMG 3010 communicates with a computer via a USB or serial port. HYDAC software HMGWIN which is specific to the HMG 3010, is supplied for convenient post-processing, rendering and evaluation of measurements on your computer. It also enables the HMG 3010 to be operated directly from your computer. The HYDAC software CMWIN is also supplied as standard with the unit. This software enables you to communicate directly from your computer with SMART sensors con­nected to the HMG 3010.
1.2 CAN functions of the HMG 3010
Option of connection to CAN-bus using CAN-adapter ZBE 3010 (accessory)
Read-out and evaluation of up to 32 measured values from up to 32 CAN messages
Configuration of the measured values from CAN-sensors
Parameterisation of HYDAC CAN-sensors (in direct connection when bus is inactive)
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 7
2 About Your HMG 3010
2.1 Items supplied
HMG 3010
Power supply for 90 ... 230 V
User Manual
CD-ROM containing USB drivers and HMGWIN and CMWIN software
(software for analyzing measurement data and sensor communication) in addition to other product information
USB connector cable
2.2 Power supply
The HMG 3010 has an internal NiMH battery. To charge the battery, simply connect the power supply with the HMG 3010 (see HMG 3010 connections, Section 9). The battery is fully charged after approx. 2 hours.
A direct current supply (12 ... 30 V DC) can be also connected be to the power input socket to operate the HMG or charge the battery.
When connecting the power supply to the input socket make sure that +U inner contact of the plug and 0 V to the outer contact. A JEITA RC 5320-5 plug, for example, is suitable for this purpose and is available in most electrical/electronics shops.
is applied to the
B
3 Technical Safety
The individual components of the HMG 3010 portable data recorder and the assembled unit are subject to strict quality assurance inspection and testing. Each HMG 3010 undergoes a final test. This ensures that when supplied the unit is free of defects and complies with the published specifications. The series HMG 3010 portable data recorders are maintenance-free and work perfectly when used under the conditions specified. If, however, you do encounter problems, please contact your HYDAC representative. Incorrect installation or tampering with the unit shall invalidate all warranty claims.
Compliance with European Standards The HMG 3010 series portable data recorders have the CE mark and thus comply with all current German regulatory requirements and European standards relating to the operation of these units. This ensures compliance with the applicable directives relating to electromag­netic compatibility and the safety provisions according to the low-voltage directive. This product complies with the provisions of the following European directives: EN 61000-6-1 / 2 / 3 / 4.
We reserve the right to make technical modifications.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 8
3.1 Safety information
The HMG 3010 series Portable Data Recorder can be operated safely provided it is used in accordance with its proper, designated use. However, in order to avoid any risk to the operator or any damage due to incorrect handling of the unit, please adhere strictly to the following safety instructions:
The HMG 3010 may not be used unless it is in perfect condition/working order.
The instructions for use must be strictly adhered to. Functions which involve adjust-
ing the zero point in particular can lead to hazardous situations. For more informa­tion, please refer to Point 6.1.3 "Zeroing of Measurement Channels".
The information on the type code label must be noted.
Troubleshooting and repair work may only be carried out by the HYDAC Service
Department.
All relevant and generally recognised safety requirements must be adhered to.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Seite 9
4 Commissioning
4.1 Operating Elements
1
2
3
4
5
6
7
On/Off button
Brightness setting of the display
ESC key
To cancel an entry or go Back step
by step
Shift key (↑)
Switches the number-pad to a textpad when pressed; the textpad is active only as long as the Shift key is pressed.
Text/number keypad
Numbers and letters can be entered via the combination keypad similar to that of mobile phones.
Numerals: 0 to 9; “.” (decimal separator) and “-“ (minus) Text entry: a, b, c, ... x, y, z and A, B, C, ... X, Y, Z; ins = insert; del = delete; To insert special characters: shift () + "1" (if necessary, several times for /, %, $ ...) To insert spaces: shift () + ins (simultaneously) To delete characters: shift () + del (simultaneously) Line break: shift () + OK (simultaneously)
Tip:
To accept characters: release the Shift key () or press the right arrow () of the 5-way navigation key.
To jump directly to the function line: simultaneously press the Shift key () and the down arrow () of the 5-way navigation key.
Graphic display
Display of the menu and operating functions, measured values and curves
5-way navigation key
To navigate step by step in the display: OK key for inputting, finishing, accepting or
storing an entry
1
2
3
4
6
7
5
Stand 11/2011 - V04 R01 Mat.-Nr.: 669855
HMG 3010 Seite 10
4.2 Connections
E F G H J
DC IN
USB
RS232
A B C D I Y adapter
A - E
4 sensor input sockets* each with two measurement channels (channel A – D or E – H) to connect up to 8 analogue sensors or up to 4 SMART sensors, e.g. for
B - F
C - G
pressure, temperature or flow rate sensors.
The four input sockets can be doubled, if required, by plugging in Y adapters (ZBE
D - H
38).
D / H
For connection to CAN-bus:
For connection to CAN-bus, use connection adapter ZBE 3010 (see Point 6.5).
Caution!
When connecting to CAN-bus, a Y-adapter cannot be used on this input socket.
I / J
1 input socket for
-2 digital signals, e.g. for frequency or speed measurements (channel I, J)
DC IN
Female jack for power supply
USB
1 USB connector for PC communication
RS232
1 serial interface for PC communication
*
For overview table of measuring inputs / input signals, see Point 6.6 Technical Specifications
Stand 11/2011 - V04 R01 Mat.-Nr.: 669855
HMG 3010 Page 11
4.3 Switching on the instrument
When starting the unit up for the first time the Welcome start-up screen appears. Select your language using the 5-way navigation key (select or checkmark your option by pressing OK) and set the date and time. Select the de­sired format, and then enter the current date and time via the keypad. Save your entries by pressing OK, followed by Next.
The following appears in the display when starting the unit up in normal operation:
Manufacturer
Designation of the instrument
HMG version
Date and time
Battery charge status
The screen switches to measured value display mode in a few seconds. If sensors are already connected, the current meas­ured values are shown in the Current Measured Values menu (see title bar at top of display). To skip through the start­up screen more quickly and proceed di­rectly to the measured values menu, press Esc or OK.
The following is listed in the measured values screen (see screenshot): Input channel – Des­ignation – Value – Unit (of measurement) The menu bar at the bottom of the screen (function bar) displays the main menu and next to that, on the right, the battery charge status. The main menu contains the follow­ing functions which you can access using the navigation key (arrow key) plus the OK key.
Settings — for entering or retrieving set­ting data used for performing measure­ments Recording — for recording a measure­ment and graphically rendering meas­urement data, e.g. as a curve
Note: As a rule, functions which can be
selected and executed are in blue.
Functions which can’t be se­lected or altered are in Black.
Warnings appear in red when input data is incorrect.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 12
Extras — for other functions,
plus a function for displaying the Min/Max values.
Measuring the Min/Max values has a special significance and is explained in more detail below.
Apart from the current measured values, the HMG records the minimum and maximum values (Min/Max values) too, except when plotting a measurement curve. The minimum value is the lowest value to which a meas­ured variable has dropped during the time period under consideration, the maximum value being the highest value achieved by a measured variable during this period. Reset­ting causes the Min/Max values to be set to the actual measured value at the time. This will reset all sensors.
The time period always begins when the Min/Max values are reset. The values are reset in the following situations:
Switching the unit on
Note:
Even when the Min/Max values are not shown in the display, the HMG is nevertheless re­cording them in the back­ground.
Selecting and activating Reset function in the menu bar (only if Min/Max is active)
Connecting or removing an HSI sensor
Altering the sensor settings
While recording an event log
When finishing recording
Before resetting the Min/Max values it is fre­quently necessary to determine when the Min/Max values were last reset. To do this, go to Extras in the function bar and then About HMG 3010.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 13
5 Getting Started Guide
This Getting Started Guide outlines the basic steps for conducting a simple measurement (measured values shown in display) and plotting a measurement curve (graph of measured values). The examples involve pressure measurement using 2 pressure transmitters with sensor detection (HSI sensors). For a detailed description of all the settings, see Detailed Operating Instructions.
The examples are based on the factory settings of the HMG 3010. Simply follow the steps described here. If you don’t know the status of your HMG 3010 because there might be more than one user in your company using the HMG, you can simply restore the unit to its "as supplied" condition (see Manage Settings — Restore Factory Settings).
Preparation for Performing a Measurement First charge the batteries or connect the power supply as described in Power Supply. Two HYDAC HDA 4748-H-0400-000 pressure transmitters with automatic sensor detection are used in the following example.
Example of a Simple Measurement Us­ing Two Pressure Transmitters
Connect the two pressure transmitters hydraulically to your machine and connect the two pressure transmitters using the sen­sor connector cables, e.g. to sockets A and B of the HMG 3010.
After the HMG 3010 is switched on it jumps to the Current Measured Values menu. The function bar contains various functions pro­viding quick access to the three main menus:
- Settings
- Recording
- Extras
and to the display of the
- Min/Max values.
.
Title bar
Function bar
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 14
When HSI sensors with automatic sensor detection are used, the HMG automatically detects the measurement range and the unit of measurement of the two connected pres­sure transmitters and shows them in the dis­play with the actual measured values. You can also display the minimum and maximum values for the sensors.
To hide or show the Min/Max display, press the right arrow of the 5-way navigation key until Min/Max is highlighted in blue and then press the OK key. The Min/Max values of the sensors are now also displayed. Press the OK key again to hide the Min/Max display.
When the Min/Max values are shown, Reset appears in the function bar, enabling you to reset the Min/Max values as required. Select Reset using the arrow keys (blue highlight) and then press the OK key to reset the Min/Max values.
The measured values can also be displayed in a 7-segment pane. To do this, go to Set- tings in the function bar and then press OK. In the Settings menu, navigate to Change display settings using the 5-way navigation key, and then confirm with OK (see screen­shot of two-pane display on the next page).
Now select the second item, Measured value rendering. Measured value table is the default setting here. Press OK. A drop-
down list appears in which you navigate to 2 display panels using the arrow keys, fol-
lowed by confirming with OK. To return to the measured value display press Esc twice.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 15
The HMG now shows the measured values in the 7-segment display panes. Depending on the setting, the Min/Max values are shown in addition to the current measured values.
The Recording menu is used for recording measurement curves. During recording, the measured values are recorded for a certain period of time, stored and then shown as a graph.
Example of Recording a Measurement Curve:
The function bar of the Actual Measured Values menu contains the item Recording.
Select it using the 5-way navigation key, fol­lowed by OK. In the Recording menu that follows, select Start recording and confirm with OK. In the menu that follows, Start Recording, you are shown the default settings for re­cording a measurement curve. In our exam­ple, we want to generate a Simple meas- urement curve with two sensors at chan­nels A and B with a sampling rate of 1 ms and a Recording time of 30 s. Then meas-
urement is to be finished.
Simple measurement curve is already set
in the menu. For the other settings, proceed as follows:
Using the arrow keys of the 5-way navigation key, select the third item, Sampling rate. OK opens up a drop-down list where you select a sampling rate of 1 ms, confirm with OK. This takes you automatically to the Re- cording time line. Press OK. Scroll down the list using the arrow keys to 30 seconds, and confirm with OK.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 16
In the Recording option line, select End when time has elapsed in the same man­ner.
In the Automatic operation line, select the
Deactivated option and in the Magnified section the 5 second option.
Your display now shows the settings you made for recording measured values (see screenshot). To double-check your settings, refer to the recorded data selected by you above the function bar.
Now you can start recording by pressing
Start in the function bar + OK. A progress bar shows you the status of your measure-
ment. The function bar offers you the option of having the measurement curve plotted or the actual value shown while recording.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 17
Graphic rendering provides a faster visual
impression of the changes in the measured values. To have the measurement curve displayed while recording, select Graph + OK. The screenshot on the right shows you the curve plotted for our example.
When recording is finished, you have the option of saving or discarding the measure­ment. To save the recording, select Save and confirm with OK. The HMG automati­cally assigns a name and a consecutive number, which can also be renamed by you. In our example, accept the name suggested by the HMG and save with OK.
You can view any stored recording again or delete it as you like. For more information on this, refer to Detailed Operating Instructions.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 18
6 Detailed Operating
Instructions
6.1 Settings Menu
Submenus of the Settings menu:
Change settings of measurement channels
Change display settings
Zero measurement channels
Change initial setup
Administer settings (save, load, ...)
Administer sensor configuration
(not visible unless the configuration of a SMART sensor has been read out, see section 6.4.2.5)
6.1.1 Change settings of measure­ment channels
Select Change settings of measurement channels to obtain a listing of all the chan-
nels available in the HMG (see screenshot). There is a checkbox next to each of the channels enabling you to switch the input on or off (+ OK key). Activated channels are marked by an “X”. To the right of the checkbox is the measurement range set- ting, the unit of measurement, and, as ap­plicable, a designation for the sensor, e.g. "System pressure".
For connection to CAN-bus:
If the connection adapter ZBE 3010 is plugged into D/H, the HMG 3010 automati­cally switches channel D to CAN-bus and channel H to off. For all the setting options for CAN-Bus, see Section 6.5‚ CAN Functions.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 19
The various options available for channels A to H are demonstrated here using channel A as an example. Select the measurement range entry on the right of the channel using the navigation keys and press OK. This takes you to the menu window for the se­lected channel. To enter a designation, press OK again and then enter it in the box using the com­bination keypad, followed by OK. In the Mode menu you now have the option of Automatic detection (HSI) and Manual. The Automatic detection (HSI) option is used for HYDAC sensors which have auto­matic sensor detection. The Manual setting is used for adapting sensors with standard analogue signals (e.g. 4 - 20 mA) to the HMG.
However, there are applications for which it is useful to enter settings manually, also in the case of HSI sensors with sensor recogni­tion. For example, by entering settings manually for a pressure measurement, you can have physical variables displayed which are proportional to the pressure, e.g. force.
Manual mode also provides the following further settings:
Input signal
Upper and lower measurement range
Unit (of measurement)
Decimal format
As soon as you select one of these items, a drop-down list appears or a box appears for entering your settings.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 20
For Input signal you can choose between HSI (HYDAC Sensor Interface) and various standard analogue signals in the drop-down list (using the arrow keys). To set the Lower and Upper measurement range, enter the relevant limits via the combination keypad.
A drop-down list is also provided for selecting the Unit (of measurement). You also have the option of defining your own unit with other (top line of the drop-down list). The cursor then jumps to the input box and a small triangular arrow appears as the insertion point. Enter the new unit using the combination keypad, e.g. ”kPa” for kilopascal.
To enter additional characters, move the triangle insertion point to the desired posi­tion and enter the characters. Confirm your entry with OK so that your unit of meas­urement is stored.
Decimal format: Here you can specify how many decimal places are to be used when recording and displaying the measured val­ues.
In addition to the settings described above, Channel B has two further measurement options. In the Mode drop-down menu you now also have the menu options Pressure
difference dP and Orifice measurement Q=f(dP). With Pressure difference dP the
HMG 3010 gives you the opportunity to dis­play the difference in pressure between channel A – channel B with a very high de­gree of accuracy. This can be used, for ex­ample, to determine if a filter is clogged, by measuring the pressure differential.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 21
The function Orifice measurement Q=f(dP) enables you to carry out a flow rate measurement by measuring the differ­ential pressure at a flow rate measurement orifice (e.g. HYDAC EVS 1000). For both these modes HYDAC HSI pres­sure transmitters with the same measure­ment range must be connected to channel A and channel B. The differential meas­urement range is produced automatically, being ± 10% of the measurement range of the pressure transmitters connected. The accuracy of the HSI pressure transmitters is ± 0.5 % FS. This means that one 100 bar pressure transmitter might measure 50 bar pressure as being 49.5 bar, and another might measure it as 50.5 bar. These pressure transmitter errors can be compensated for by recording a correction curve. After correctly recording a curve, the accu­racy for differential measurement is ≤±0.1% of the differential measurement range.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 22
Example: For both 100 bar pressure transmitters there is a differential measurement range of
-10 – +10 bar, in other words a measurement margin of 20 bar. The accuracy of ≤ ± 0.1% refers to this measurement margin. The pressure differential can therefore be measured with an accuracy of ± 0.02 bar.
To measure a pressure difference, select the item Pressure differential dP. Then you find a menu point Organise curves. Here you can plot up to 3 curves for 3 pressure transmitter pairs. If a curve for a pressure transmitter combination already exists, you can quit the menu again by selecting Back or the Esc key. The HMG 3010 automatically selects the appropriate curve. If you quit the menu and the HMG 3010 detects that no curve exists for this combination, you will get the message shown on the right. You can now, with less accuracy, measure the differ­ential or you call the menu item again to re­cord a new curve under Organise curves.
Once Organise curves is selected, a screen appears which allows you to record a new curve, or you see information on curves which already exist. At this point you can also delete existing curves if they are no longer required. If curve 1 already exists, navigate to this field and press the OK key. You can now, for example, select Character­istic curve 2, Not available and then record a curve. To record the curve you must con­nect both pressure transmitters to the HMG 3010 at channel A and channel B.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 23
When doing this, identify which pressure transmitter has been connected to channel A and which to channel B, since the pressure transmitters must be re-connected in the same way later when taking measurements.
Connect the pressure transmitter hydraulically so that the two instruments are subjected to the same pressure. The ideal for recording curves is connecting to a test rig on which you can increase the pressure slowly from 0 bar to the full measurement range of the pressure transmitter and back again. However, the minimum requirement is to record the curves/measured values in the pressure range in which the differential measurement is to take place later.
In the example above (to measure the filter contamination level) both pressure transmitters might be fitted to the inlet side of the filter, e.g. using T-fittings and then, with the machine running, the curve could be re­corded. When these pre-conditions are in place, select Record characteristic curve. This brings up a display showing 0 to 100% of the whole measurement range. If you increase the pressure you will see from the blue bar that measured values are being recorded in this pressure range. If the measured values remain under the red line, this means that not enough values have been recorded in this range. To achieve maximum accuracy for differential measurement, the recorded measured val­ues should at least be in the area between the red and the green line. After this, select Stop recording and you can quit the menu. The curve is now automatically taken into account for the differential measurement using these two pressure transmitters.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 24
In order to measure the flow rate using a differential pressure measurement at a flow rate orifice, select the item Orifice meas- urement Q=f(dP). Then select the number of decimal places and the unit (of measure­ment). A flow rate orifice is supplied with information showing the correlation between pressure differential and flow rate. For example, on a HYDAC EVS 1000 it says: calibration factor
4.87 bar at 350 l/min. Enter these values under the flow rate and pressure.
Here again the menu Organise curves appears, (for a function description and handling, see Pressure differential dP above). Curve recording can be carried out exactly as de­scribed above.
If you are using a HYDAC EVS 1000 flow rate measurement orifice, you can also use this to record curves. On the EVS 1000 there are 2 pressure ports before the orifice (ports 1 and 2) and one after (port 3).
To make a recording, connect the pressure transmitter from channel A to port 1, and the one from channel B to port 2. To start measurement you need only then to change the pressure transmitter from port 2 to port 3.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 25
Input channels I and J are for digital (not analogue) signals. As the input signal you can choose between Count pulses and Frequency signal. There are many possi­bilities for this, e.g. speed sensors, proximity switches, pulse generators, flow rate meas­uring instruments with a frequency signal, HYDAC HDS 1000, to name only a few. As in the case of the analogue sensor inputs, settings can be manually entered for the input signal, unit of measurement and deci­mal format.
In addition to these settings, a factor has to be entered which is required for con­version to the desired unit of measure­ment.
Another special feature is that the Upper measurem. range must be input. Frequen­cies of up to 30000 Hz can be measured on channels I and J, consequently a scale from 0 to 30000 is possible in the graphical dis­play. If, for example, you record a speed of up to 1450 rpm, an awkward Y-axis scale will result in the graph. If you enter "1500" as the Upper measurement range for this speed measurement, for example, the Y-axis in the graph will be scaled from 0 to 1500 rpm.
In this case the reason for inputting the upper measurement range is just a means of limiting the scale of the Y-axis in the graphical display. You should set this value as close as possible to that of the expected measured value in order to get the best possible graphical display. If there are measured values higher than the upper measurement range, these are not lost. The scale of the Y-axis can be changed again in the graph at any time later on.
Caution!
The measurement range of the frequency entries is 1 Hz to 30000 Hz. If, for example, the rpm of a motor is recorded and this motor stops abruptly, the last rpm can still show in the display for up to 2 seconds, depending on the measurement system, before 0 is displayed.
Edition 11/2011 - V04 R01 Part No.: 669855
HMG 3010 Page 26
Examples of frequency measurements:
1) You would like to measure a frequency up to 1000 Hz. Frequencies are recorded by the HMG per minute. Enter "60" as the Factor to display the frequency in Hz, enter "1000" for the Upper measurem. range to scale the Y­axis from 0 to 1000 in the graphical display.
2) You would like to measure the speed of a machine from 0 to 2900 rpm, the speed sen­sor supplies 2 pulses per revolution. Enter "2" as the Factor and "3000" as the Upper measurem. range. The display is then in revolutions per minute; for the graphical dis­play, the Y-axis is scaled from 0 to 3000.
3) A flow rate measurement transmitter in a measurement range up to 450 l/min with a frequency output supplies 8 pulses per litre. Enter "8" as the Factor and "500" as the Up- per measurem. range. The display is then in l/min; for the graphical display, the Y-axis is scaled from 0 to 500.
Examples of count pulses:
1) The switching output of a proximity switch on a conveyor belt is connected to the fre­quency input of the HMG 3010. Enter "1" as the Factor and the expected maximum quantity to be counted (maximum 30000) as the Upper measurem. range.
2) A gear flow meter supplies 250 pulses per liter. Enter "250" as the Factor and as the Upper measurement range, the maximum no. of litres expected (maximum 30000).
Edition 11/2011 - V04 R01 Part No.: 669855
Loading...
+ 60 hidden pages