Digital Display Unit
HDA 5500
User Manual
HDA 5500-0-0-AC-000
HDA 5500-0-0-DC-000
Dated 14.06.2005
HDA 5500 |
User Manual |
CONTENTS
1 |
Introduction |
3 |
|
|
1.1 |
European standards compatibility |
3 |
2 |
Safety instructions |
3 |
|
3 |
Functions |
3 |
|
4 |
Operating keys |
4 |
|
5 |
Selecting and displaying the unit of measurement |
4 |
|
6 |
Installation and commissioning |
5 |
|
|
6.1 |
Mechanical installation |
5 |
|
6.2 |
Electrical connection |
5 |
|
6.3 |
Supply voltage |
6 |
7 |
Programming |
7 |
|
|
7.1 |
Adjusting the basic settings |
7 |
|
7.2 |
Overview of basic settings menu |
8 |
8 |
Connecting sensors |
10 |
|
|
8.1 |
HDA 5500 with one transmitter |
10 |
9 |
Digital display |
11 |
|
|
9.1 |
Standard display |
11 |
|
9.2 |
Special display |
11 |
10 |
Output response |
12 |
|
|
10.1 |
Analogue output |
12 |
|
|
10.1.1 Analogue output set to 4..20 mA |
12 |
|
|
10.1.2 Analogue output set to 0..10 V |
12 |
|
10.2 |
Switching outputs |
13 |
|
10.3 |
Connection of up to four switching outputs |
13 |
11 |
Adjusting switching points and hystereses |
13 |
|
12 |
Programming enable |
13 |
|
|
12.1 |
Altering the operating programming enable |
13 |
|
12.2 |
Altering the main programming enable |
14 |
13 |
Technical specifications |
15 |
|
14 |
Model code HDA 5500 |
16 |
|
15 |
Unit dimensions |
17 |
|
16 |
Electrical connections |
18 |
|
17 |
Error messages |
19 |
|
18 |
HYDAC Service |
19 |
- 2 -
HDA 5500 |
User Manual |
1 Introduction
The individual components and the final assembly of the Digital Display Unit HDA 5500 are subject to strict quality controls. Each HDA 5500 is individually calibrated and subjected to a final test. In this way we can guarantee that the unit is fault-free on despatch and conforms to the given specifications. However, if there is a cause for complaint, please return the unit to us outlining the fault.
The HDA 5500 digital display unit is maintenance-free and operates perfectly when used according to the specifications. If faults do nonetheless arise, please contact HYDAC Service. Incorrect installation or interference by anybody other than HYDAC personnel will invalidate all warranty claims.
1.1 European standards compatibility
The HDA 5500 digital display unit carries the - mark and therefore corresponds to the approval requirements currently applicable in Germany and to the European standards for operating these units. The applicable directives on electromagnetic compatibility and the safety regulations according to the low voltage directive are therefore guaranteed.
This product complies with the regulations of the following European Directive:
EN 61000–6–1 / 2 / 3 / 4
We reserve the right to make technical modifications.
2 Safety instructions
The HDA 5500 digital display unit presents no safety concerns when installed and operated in accordance with this user manual. However, in order to avoid any risk to the operator or any damage due to incorrect handling of the unit, please adhere strictly to the following safety instructions:
•The HDA 5500 must not be put into service if any known defects, either electrical or mechanical, are apparent.
•The unit must be installed exactly according to the instructions.
•Read the information on the type code label.
•Fault investigation and repairs must only be carried out by HYDAC Service.
•All relevant and generally recognised safety requirements must be adhered to.
3 Functions
Depending on the model, the following functions are available:
•Digital display of analogue signals
•Temperature measurement (by means of external Pt100 temperature sensor)
•Frequency measurement
•Display of a unit, e. g. bar, psi, MPa, kg/cm², oC, oF, l/min (and others)
•Display of a lowest or peak value (slave pointer function)
•Display of a pre-set switch point
•Display of differential value and switching on differential value
•Counter function
•Switching outputs via relay switching contacts
•Analogue output 4..20 mA or 0..10 V can be selected
•Basic settings menu for adapting the HDA 5500 to a particular measuring task
•Programming enable functions can be disabled to prevent subsequent unauthorised adjustment of the settings
•RS 232 serial interface
- 3 -
HDA 5500 |
User Manual |
4 Operating keys
LED displays of active relays
|
4 digit |
|
||
|
digital display |
|
||
|
|
|
|
LED displays |
|
|
|
|
of active |
Relais |
Sensor |
sensor |
||
|
||||
1 |
|
|
|
|
2 |
1 |
2 |
3 |
|
3 |
|
|||
|
|
|
|
|
4 |
|
|
|
|
MODE |
HDA5500 |
|
||
|
|
|
|
Display of |
|
|
|
|
unit of measurement |
|
|
|
|
for sensor 1, 2 or 3 |
Keys for adjusting the switching points, switch-back points and the menu functions
5 Selection and display of the unit of measurement
The operator can select the appropriate unit label from a printed sheet and this is then displayed with background lighting. The label can be changed at any time. Procedure:
•Disconnect the unit from supply voltage.
•Remove the front frame of the unit by hand.
•Loosen the 4 front plate screws.
•Press lightly on the back panel of the unit to push out the front side (display section) by a few millimetre.
•Cut the label with the required unit from the sheet supplied.
•Insert the label into the slot on the side.
•Press lightly on the front panel to push it back on.
•Screw the front plate back in place.
•Put the front frame back on.
Indication of unit using label from sheet
Front frame
- 4 -
HDA 5500 |
User Manual |
6 Installation and commissioning
6.1 Mechanical installation
The HDA 5500 is a control panel unit with a standard mounting housing with the following measures for mounting:
•Control panel cut-out of 92 x 45 mm
•Panel thickness: at least 1.25 mm
•Mounting depth: at least 150 mm
Panel clips
Panel thickness
Clamping guides
6.2 Electrical connection
The electrical connection must be carried out by a qualified electrician in accordance with relevant regulations (in Germany: VDE 0100).
Important:
In order to fulfil the relevant EMC standards, subject to the ambient conditions, the screening effect of the control box must comply with the EMC standard.
Additional installation notes which, from experience, reduce the effect of electromagnetic interference:
•Make line connections as short as possible.
•Use screened lines (e. g. LIYCY 4 x 0.5 mm²).
•The cable screening must be fitted by qualified personnel subject to the ambient conditions and with the aim of suppressing interference.
•Direct proximity to interference must be avoided as far as possible.
If inductive loads are to be switched using the relays, varistors must be used on the load to prevent high switch-off surges.
- 5 -
HDA 5500 |
User Manual |
6.3 Supply voltage
The supply voltage is connected to terminal X1.
12..32 VDC
HDA 5500–X–X–DC–000 (see type code label)
85..265 VAC 50 / 60 Hz
HDA 5500–X–X–AC–000 (see type code label)
After switching on the supply voltage, the unit displays HdA for approximately 2 s. Then the number of the sensor set as the primary display is shown. After a further approximately 2 s the actual value is displayed.
2 s |
2 s |
Important:
Before sensors are connected, the HDA 5500 must be set appropriately.
If there is no input signal, then the following will be displayed, depending on the type of signal:
•Starting value of the measuring range - flashing (for sensors with 4..20 mA)
•Starting value of the measuring range (for sensors with 0..10 V)
- 6 -