HYDAC HMG 3000 User Manual

PORTABLE DATA RECORDER
HMG 3000
User Manual
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 2
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
Table of Contents
1 Tasks and Functions of the HMG 3000 .......................................................4
2 About Your HMG 3000..................................................................................7
2.1 Included with the HMG as Supplied......................................................7
2.2 Power Supply........................................................................................7
3 Technical Safety............................................................................................7
3.1 General Safety Precautions..................................................................8
4 Starting the HMG 3000 Up............................................................................9
4.1 Operating Elements ..............................................................................9
4.2 Connectors .........................................................................................10
4.3 Switching the HMG 3000 On ..............................................................11
5 Getting Started Guide.................................................................................13
6 Detailed Operating Instructions.................................................................17
6.1 Settings Menu.....................................................................................17
6.1.1 Changing the Settings of the Measurement Channels ...........17
6.1.2 Changing Display Settings .....................................................27
6.1.3 Zeroing Measurement Channels ............................................29
6.1.4 Changing the Initial Setup ......................................................31
6.1.5 Administer settings (save, load, ...) .......................................33
6.2 Recording Menu .................................................................................35
6.2.1 Starting a Recording ..............................................................35
6.2.2 Viewing Recordings ...............................................................51
6.2.3 Deleting Recordings...............................................................62
6.3 Extras Menu........................................................................................63
6.4 SMART Sensors .................................................................................65
6.4.1 Connecting SMART Sensors .................................................65
6.4.2 SMART Sensors Menu ..........................................................66
6.5 Technical Specifications................................................................................69
7 Cleaning.......................................................................................................70
8 Accessories.................................................................................................71
9 Connection of HMG Components..............................................................73
9.1 Pin assignment....................................................................................74
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
1 Tasks and Functions of the HMG 3000
The HMG 3000 portable data recorder is a hand-held unit for simple measurement and data capturing tasks involving hydraulic and pneumatic systems. Applications extend primarily to maintenance and servicing, troubleshooting and test rigs.
The HMG 3000 can concurrently evaluate signals from up to 10 sensors. The unit features input jacks for connecting the sensors; if necessary this number can be doubled using a adapter for measurement operations involving more than 5 sensors. HSI sensors (HYDAC Sensor Interface) are HYDAC sensors for pressure, temperature and flow rate with automatic sensor recognition.
You also have the option of connecting SMART sensors to the HMG 3000. These are the new generation sensors from HYDAC, capable of outputting several different measured variables. These sensors are also automatically recognized by the HMG 3000 via HIS, and the measured values including the measurement unit are shown in the display.
The HMG 3000 can render up to 32 measurement channels.
Depending on the sensor type the sensor features an internal memory providing for the storage of recorded measurement data on a long-term basis. In addition, depending on the sensor type preset parameters can be changed and stored in the internal memory as a sensor configuration.
Older model HYDAC sensors or standard sensors of other manufacturers commercially available can also be used. However, these sensors do not feature any automatic sensor detection, consequently the initial set-up has to be entered by hand.
The HMG 3000 is user-friendly by virtue of its easy-to-access selection menus leading to all of the unit’s functions and settings. The unit features a combination keypad for entering numeric values and text in a fashion similar to that of mobile phones.
The HMG 3000 was designed in particular for capturing typical measurement values (pressure, temperature, volumetric flow rate) in hydraulic and pneumatic systems. A variety of other measurement tasks can be performed by virtue of additional inputs for voltage measurement. Example: checking the actuation of a switching valve or plotting the characteristic curve of a valve setpoint. In addition, it is also possible to determine differential values between the measured values of individual sensors. One example of this is taking a volumetric flow measurement using a differential pressure orifice plate.
In addition to the analog measurement inputs, the HMG features two digital outputs, enabling frequencies or speeds to be recorded, thus expanding the unit’s range of potential applications.
When taking measurements of rapid, dynamic machine processes, all 8 analog input signals can be concurrently captured at a rate of 0.5 ms.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
One special feature of the HMG 3000 is its ability to record measurements of the highly dynamic processes in a machine. To this end, two input channels are featured which are capable of recording measured values at a rate of 0.1 ms. It goes without saying that this feature is dependent on the use of suitable, fast-acting sensors.
The most attractive feature offered by the HMG 3000 is its ability to record the dynamic processes of a machine in the form of a measurement curve and render them as a graph — on line and in real time.
The HMG 3000 is equipped with specially developed software providing for fast data collection and processing. A measurement curve can comprise up to 500,000 measured values. The HMG 3000’s measured value memory is capable of storing at least 100 of these measurement curves.
In addition to enabling simple measurement curves to be recorded, the HMG 3000 also features other functions enabling event-driven measurements to be taken and event logs to be recorded. Various trigger options are available for triggering events.
Apart from measurement curves, the HMG 3000 can store user-specific settings (user profiles). The main advantage of this is enabling identical measurements of various
equipment items to be repeated for the purpose of preventive maintenance. All the user has to do is retrieve the respective user profile from the HMG 3000’s memory.
In addition to the analog measurement inputs, the HMG features two digital outputs, enabling frequencies or speeds to be recorded, thus expanding the unit’s range of potential applications. Individual measured values can be displayed in a large format (7-segment format), enabling them to be read at an extended distance.
The HMG 3000 also provides for a variety of user-friendly features for displaying, evaluating and processing measured values:
Table
Graph
Scaling
Ruler
Tracker
Zoom
The HMG 3000 communicates with a computer via a USB or serial port. HYDAC offers
HMGWIN 3000, the matching software for the HMG 3000 which comes standard with the
unit, for convenient post-processing, rendering and evaluation of measurements at your computer. It also enables the HMG 3000 to be operated directly from your computer.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
The HYDAC software CMWIN also comes standard with the unit. This software enables
you to communicate directly from your computer with SMART sensors connected to the HMG 3000.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 7
2 About Your HMG 3000
2.1 Included with the HMG as Supplied
HMG 3000
Power supply providing for 90 – 230 V
Manual
CD-ROM containing USB drivers and HMGWIN 3000 and CMWIN
(software for analyzing measurement data and sensor communication) in addition to other product information
USB connector cable
2.2 Power Supply
The HMG 3000 comes with an internal NiMH battery. To charge the battery, simply connect the power supply furnished with the unit (see connectors of the HMG 3000). The battery is fully charged after ca. 5 hours.
A direct current supply (12 – 30 V DC) can be connected to the power input socket for operating the HMG or charging the battery.
When connecting the power supply to the input socket make sure that +UB is present at the inner contact of the plug and 0 V at the outer contact. A JEITA RC 5320-5 plug, for example, is suitable for this purpose; it is available in most electrical/electronics shops.
3 Technical Safety
The individual components of the HMG 3000 portable data recorder and the assembled unit are subject to strict quality assurance inspection and testing. Each HMG 3000 undergoes outgoing testing. This ensures that when supplied the unit is free from defects and complies with the designated specifications.
The series HMG 3000 portable data recorders are maintenance-free and work perfectly when used under the conditions specified. If, however, you do encounter problems, please contact your HYDAC representative. Any tampering with the data recorder will cause all warranty claims to become null and void.
Compliance with European Standards
The HMG 3000 series portable data recorders feature the CE mark and thus comply with all relevant German regulatory requirements and European standards currently in effect pertaining to the operation of these units. This ensures compliance with currently valid regulations pertaining to electromagnetic compatibility and the provisions of the Low-Voltage Directive (73/23/EEC). This product complies with the provisions of the following European standards: EN 61000-6-1 / 2 / 3 / 4
All technical details are subject to change without notice.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 8
3.1 General Safety Precautions
The HMG 3000 series Portable Data Recorder can be operated safely provided it is used in accordance with its proper, designated use. Please maintain strict adherence to the following safety guidelines in order to preclude hazards to the user and damage to property and surroundings as the result of improper use:
The HMG 3000 may not be used unless it is in perfect condition/working order.
Strict adherence is to be maintained to the usage instructions. Operations involving
adjusting the zero point in particular can lead to dangerous situations. For more
information, please refer to Zeroing of Measurement Channels.
Heed the specifications on the rating plate.
Troubleshooting and repair work may only be done by HYDAC’s Service Department.
Adherence is to be maintained to all relevant generally recognized safety regulations.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 9
4 Starting the HMG 3000 Up
4.1 Operating Elements
1
2
3
4
5
6
7
On/Off button
Brightness setting of the display
Esc
For cancelling an entry or going Back
step by step
Shift key
Switches the keypad to a textpad when pressed; the textpad is active only as long as the Shift key is pressed.
Text/numeric keypad
Numbers and letters can be entered via the combination keypad in a fashion similar to that of mobile phones.
Numerals: 0 to 9; “.” (decimal separator) and “-“ (minus) Text entry: a, b, c, ... x, y, z and A, B, C, ... X, Y, Z; To insert special characters: To insert spaces: To delete characters: Line break:
Tip:
SHIFT + OK (simultaneously)
SHIFT + INS (simultaneously)
SHIFT + DEL (simultaneously)
SHIFT + "1" (if necessary, several times for /, %, $ ...)
To accept characters: release the Shift key or press the right arrow of the
5-way navigation key.
To jump directly to the function line: simultaneously press the Shift key and the
down arrow of the 5-way navigation key.
Graphic display
Display of the menu and operating functions, measured values and measurement curves
5-way navigation key
To navigate step by step in the display: OK key for inputting, concluding, accepting or
storing an entry
1
2
3 4
5
INS = insert; DEL = delete;
6
7
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 10
4.2 Connectors
USB
DC IN
A B C D I I
E F G H J Y adapter
RS232
A - D
4 sensor input jacks for up to 8 analog sensors or up to 4 SMART sensors
(channel A – D and E – H*), e.g. for sensors for measuring pressure,
E - H
temperature or flow rate. The four input jacks can be doubled by plugging in Y adapters.
I / J
1 or 2 input jacks for
-2 digital signals, e.g. for frequency or speed measurements (channel I, J)
-1 voltage input ( - 10 V to + 10 V, channel H*)
DC IN
Female connector for power supply
USB
1 USB connector
RS232
1 serial port
H*: Channel H can be used for sensors with an analog signal as well as for voltage
measurements of -10 V to +10 V.
(for pin assignment, see section 9.)
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Seite 11
4.3 Switching the HMG 3000 On
When starting the unit up for the first time the
Welcome start-up screen appears. Select
your language using the 5-way navigation key (Select or checkmark your option by pressing
OK) and set the current date and time. Select
the desired format, followed by entering the
current date and time via the keypad. Save your entries via OK, followed by pressing Continue.
The following appears in the display when starting the unit up in normal operation:
Manufacturer
Designation of the unit
HMG version
Date and time
Battery charge status
The screen switches to measured value display mode in a few seconds. When sensors are connected you are shown the current
measured values in the Current Measured Values menu (see title bar in display). To skip
through the start-up screen more quickly and proceed directly to the measured values
menu, press Esc or OK.
The following is listed in the measured values screen (see screenshot): Input channel – Designation – Value – (Measurement) Unit
The menu bar at the bottom of the screen (function bar) displays the main menu and next to that, on the right, the battery charge status. The main menu contains the following functions which you can access via the
navigation key (arrow key) plus the OK key.
Settings — for entering or referring to
setting data used for performing measurements
Recording — for recording a
measurement and graphically rendering measurement data, e.g. as a curve
Note: Generally speaking, functions which can be selected and
executed are blue.
Black means the functions
can’t be selected or executed.
Warnings are shown in red
when incorrect entries are
made.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Seite 12
Extras — menu featuring the stopwatch,
among other things
plus a function for displaying the Min/Max
values.
Special significance is accorded the
measurement of the Min/Max values, which
is why it is explained in detail below.
Apart from the current measured values, the HMG captures the minimum and maximum values (Min/Max values) too, except when recording a measurement curve. The minimum value is the smallest value to which a measured variable has dropped during the time period under consideration, the maximum value being the largest value achieved by a measured variable during this period. Resetting causes the Min/Max values to be set to the current measured value. This is done jointly for all sensors.
The time period always begins when the Min/Max values are reset. Resetting the values is prompted by the following situations:
Switching the unit on
Selecting and activating Reset in the
menu bar
Connecting or removing an HSI sensor
Modifying the sensor settings
While recording an event log
When concluding recording
Before resetting the Min/Max values it is frequently necessary to determine when the Min/Max values were last reset. To do this, go
to Extras in the function bar and then About HMG 3000.
Note: Even if the Min/Max values aren’t shown in the display HMG captures them in the background.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Seite 13
5 Getting Started Guide
This Getting Started Guide outlines the basic steps for conducting a simple measurement (measured values shown in display) and recording a measurement curve (graph of measured values). These examples involve pressure measurement using 2 pressure transmitters with
sensor detection (HSI sensors). For a detailed description of all the settings, see Detailed Operating Instructions.
The examples are based on the factory settings of the HMG 3000. Simply follow the steps described here. If you don’t know the status of your HMG 3000 because there might be more than one user in your company using the HMG, you can simply restore
the unit to its condition as supplied (see Administration Settings — Restore factory settings).
Preparation in Advance of Performing a Measurement First charge the batteries or connect the power supply as described in Power Supply. Two
HYDAC HDA 4748-H-0400-000 pressure transmitters with automatic sensor detection are used in the following example.
Example of a Simple Measurement Using Two Pressure Transmitt er s
Integrate the two pressure transmitters hydraulically in your equipment and connect the two pressure transmitters using the sensor connector cables, e.g. to jacks A and B of the HMG 3000.
After the HMG 3000 is switched on it jumps
to the Current Measured Values menu. The
menu bar (function bar) contains various functions providing you direct access to the three main menus:
- Settings - Recording - Extras
and to the display of the Min and Max
values.
Title bar
Function bar
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 14
When HSI sensors featuring automatic sensor detection are used, the HMG automatically detects the measurement range and the unit of measurement of the two connected pressure transmitters and shows them in the display with the current measured values. You can also have the minimum and maximum values for the sensors displayed.
To hide or unhide the Min/Max display, actuate the right arrow of the 5-way
navigation key until Min/Max is highlighted in blue and then press the OK key. This causes
the Min/Max values of the sensors to be
additionally displayed. Pressing the OK key
again causes the Min/Max display to be switched off.
When the Min/Max values are shown, Reset
appears in the function bar, enabling you to reset the Min/Max values as needed. Select them using the arrow keys (blue highlighted)
and then press the OK key: the Min/ Max
values are reset.
The measured values can also be displayed
in a 7-segment pane. To do this, go to
Settings in the function bar and then press OK. In the Settings menu, navigate to Change display settings using the 5-way
navigation key, and then confirm via OK (cf.
screenshot of two-pane display on the next page).
Now select the second item, Measured value rendering. Measured value table is the default setting here. Press OK. A drop-
down list appears in which you navigate to 2 display panels using the arrow keys,
followed by confirming via OK. To return to the measured value display press Esc twice.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 15
The HMG now shows the measured values in the 7-segment display panes. Depending on the setting, the Min/Max values are shown in addition to the current measured values.
The Recording menu is used for recording
measurement curves. During recording, the measured values are recorded for a certain period of time, stored and then shown as a graph.
Example of Recording a Measurement Curve:
The function bar of the Current Measured Values menu contains the item Recording.
Select it using the 5-way navigation key,
followed by pressing OK. In the Recording menu that follows, select Start recording and confirm via OK.
In the menu that follows, Start Recording,
you are shown the default settings for recording a measurement curve. In our
example, we want to generate a Simple measurement curve with two sensors at channels A and B with a sampling rate of
1 ms and a Recording time of 30 s. Then measurement is to be concluded.
Simple measurement curve is already set
in the menu. For the other settings, proceed as follows:
Using the arrow keys of the 5-way navigation
key, select the third item, Sampling rate. OK causes a drop-down list to open in which you select a sampling rate of 1 ms, followed by confirming via OK. This causes you to automatically proceed to the Recording time line. Press OK. Scroll in the drop-down list using the arrow keys to 30 seconds, followed by confirming via OK.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 16
In the Recording option line, select End when time has elapsed in the same
manner.
Your screen now shows the settings for recording measured values made by you (cf. screenshot at right). To double-check your
settings, refer to the recorded data selected
by you above the function bar.
Now you can start recording via Start in the function bar + OK. A progress bar shows
you the status of your measurement. The function bar offers you the option of having the measurement curve plotted or the current value shown while recording.
Graphic rendering provides for a faster
visual impression of the changes in the measured values. To have the measurement curve displayed while recording, select
Graph + OK. The screenshot at the right
shows you the curve plotted for our example.
When recording is concluded, you have the option of storing or discarding the measurement. To store the recording, select
Save and confirm via OK. The HMG
automatically assigns a name and a consecutive number, which can be subsequently changed by you as needed. For our example, accept the name
suggested by the HMG and save via OK.
You can view any stored recording again or delete it as you like. For more information on
this, refer to Detailed Operating Instructions.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 17
6 Detailed Operating Instructions
6.1 Settings Menu
Submenus of the Settings menu:
Change settings of measurement
channels
Change display settings
Zero measurement channels
Change initial setup
Administer settings (save, load, ...)
Administer sensor configuration
(not visible unless the configuration of a SMART sensor has been read out, see section 6.4.2.5)
6.1.1 Changing the Settings of the Measurement Channels
Select Change settings of measurement channels to obtain a listing of all the
channels available in the HMG (see screenshot). There is a checkbox next to each of the channels enabling you to switch
the input on or off (+ OK key). Activated channels are marked by an “X”.
To the right of the checkbox is the
measurement range setting, the unit of
measurement, and, as applicable, a
designation for the sensor, e.g. "System
pressure".
The various options available for channels A to H are demonstrated here using channel A as an example. Select the
measurement range entry to the right of
the channel using the arrow keys followed
by OK. You now proceed to the menu
window of the channel selected. To enter a
designation, press OK again and then
enter it in the box using the combination
keypad, followed by OK.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 18
In the Mode menu you now have the option of Automatic detection (HSI) and Manual. The Automatic detection (HSI) option is
used for HYDAC sensors featuring automatic
sensor detection. The Manual setting is
used for adapting sensors featuring standard analog signals to the HMG (e.g. 4-20 mA).
However, there are applications in which entering settings by hand is useful, also for HSI sensors. For example, by entering settings manually for a pressure measurement you can have proportional variables, e.g. the ”force”, shown on the HMG.
Manual mode also provides for the following
settings:
Input signal
Upper and lower measurement range
Unit (of measurement)
Decimal format
As soon as you select one of these items, a drop-down list appears or a box appears for entering your settings.
For Input signal you can choose in the drop­down list between HSI (HYDAC Sensor Interface) and various standard analog
signals (using the arrow keys). To set the
Lower and Upper measurement range,
enter the respective limits via the combination keypad.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 19
A drop-down list is also provided for
selecting the Unit (of measurement). You also have the option of defining your “other unit” (uppermost line of the drop-down list). The cursor then jumps to the input box and a small triangular arrow appears as the insertion point. Enter the new unit using
the combination keypad, e.g. ”kPa” for kilopascal.
To enter additional characters, move the triangle insertion point to the desired position and enter the characters. Conclude
your entry via OK so that your unit of
measurement is stored.
Decimal format: Here you can specify how
many decimal places are to be used when recording and displaying the measured values.
In addition to the settings described above,
Channel B has two further measurement options. In the Mode drop-down menu you now also have the menu options Pressure
difference dP and Orifice measurement Q=f(dP). With Pressure difference dP the
HMG 3000 gives you the opportunity to display the difference in pressure between channel A – channel B with a very high degree of accuracy. This can be used, for example, to determine if a filter is clogged, by measuring the pressure differential.
The function Orifice measurement Q=f(dP) enables you to carry out a flow rate
measurement by measuring the differential pressure at a flow rate measurement orifice (e.g. HYDAC EVS 1000). For both these modes HYDAC HSI pressure transmitters with the same measurement range must be connected to channel A and channel B. The differential measurement range is produced automatically, being ± 10% of the measurement range of the pressure transmitters connected. The accuracy of the HSI pressure transmitters is ≤±0.5%FS. This means that one 100 bar pressure transmitter might measure 50 bar pressure as being 49.5 bar, and another might measure it as 50.5 bar. These pressure transmitter errors can be compensated for by recording a correction curve. After correctly recording a curve, the accuracy for differential measurement is ≤±0.1% of the differential measurement range.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 20
Example: For both 100 bar pressure transmitters there is a differential measurement range of
-10 – +10 bar, in other words a measurement margin of 20 bar. The accuracy of ≤±0.1%
refers to this measurement margin. The pressure differential can therefore be measured with an accuracy of ≤±0.02 bar.
To measure a pressure differential, select the item Pressure differential dP. Then you find an item Organise curves. Here you can
plot up to 3 curves for 3 pressure transmitter pairs. If a curve for a pressure transmitter combination already exists, you can quit the
menu again by selecting Back or the Esc
key. The HMG 3000 automatically selects the appropriate curve. If you quit the menu and the HMG 3000 detects that no curve exists for this combination, you will get the message shown on the right. You can now, with less accuracy, measure the differential or you call the menu item again to record a
new curve under Organise curves.
Upon selecting Organise curves, a screen
appears which allows you to record a new curve, or you see information on curves which already exist. At this point you can also delete existing curves if they are no longer required. If curve 1 already exists,
navigate to this field and press the OK key.
You can now, for example, select
Characteristic curve 2, Not available and
then record a curve. To record the curve you must connect both pressure transmitters to the HMG 3000 at channel A and channel B. When doing this, identify which pressure transmitter has been connected to channel A and which to channel B, since the pressure transmitters must be re-connected in the same way later when taking measurements.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 21
Connect the pressure transmitter hydraulically so that the two instruments are subjected to the same pressure. The ideal for recording curves is connecting to a test rig on which you can increase the pressure slowly from 0 bar to the full measurement range of the pressure transmitter and back again. However, the minimum requirement is to record the curves/measured values in the pressure range in which the differential measurement is to take place later.
In the example above (to measure the filter contamination level) both pressure transmitters might be fitted to the inlet side of the filter, e.g. using T-fittings and then, with the machine running, the curve could be recorded. When these pre-conditions are in place,
select Record characteristic curve. This
brings up a display showing 0 to 100% of the whole measurement range. If you increase the pressure you will see from the blue bar that measured values are being recorded in this pressure range.
If the measured values remain under the red line, this means that not enough values have been recorded in this range. To achieve maximum accuracy for differential measurement, the recorded measured values should at least be in the area between the red and the green line. After
this, select Stop recording and you can quit
the menu. The curve is now automatically taken into account for the differential measurement using these two pressure transmitters.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 22
In order to measure the flow rate using a differential pressure measurement at a flow
rate orifice, select the item Orifice measurement Q=f(dP). Then select the
number of decimal places and the unit (of measurement). A flow rate orifice is supplied with information showing the correlation between pressure differential and flow rate. For example, on a HYDAC EVS 1000 it says: calibration factor
4.87 bar at 350 l/min. Enter these values
under the flow rate and pressure.
Here again the menu Organise curves appears, (for a function description and handling, see Pressure differential dP above). Curve recording can be carried out exactly as described
above.
If you are using a HYDAC EVS 1000 flow rate measurement orifice, you can also use this to record curves. On the EVS 1000 there are 2 pressure ports before the orifice (ports 1 and 2) and one after (port 3).
To make a recording, connect the pressure transmitter from channel A to port 1, and the one from channel B to port 2. To start measurement you need only then to change the pressure transmitter from port 2 to port 3.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
HMG 3000 Page 23
Input channels I and J are for digital signals.
As the input signal you can choose between
Count pulses and Frequency signal. There
are many possibilities for this, e.g. speed sensors, proximity switches, pulse generators, flow rate measuring instruments with a frequency signal, HYDAC HDS 1000, to name only a few. As in the case of the analog sensor inputs, settings can be manually entered for the input signal, unit of measurement and decimal format.
In addition to these settings, a factor has to be entered which is required for conversion to the desired unit of measurement.
Another special feature is the inputting of the
Upper measurem. range. Frequencies of up
to 30000 Hz can be measured on channels I and J, consequently a scale from 0 to 30000 is possible in the graphical display. If, for example, you record a speed of up to 1450 rpm, an awkward Y-axis scale will result in
the graph. If you enter "1500" as the Upper measurement range for this speed
measurement, for example, the Y-axis in the graph will be scaled from 0 to 1500 rpm.
In this case entering the upper measurement range is just a means of limiting the scale of the Y-axis in the graphical display. You should set this value as close as possible to that of the expected measured value in order to get the best possible graphical display. If measured values exist which are larger than the upper measurement range, these are not lost. The scale of the Y-axis can be changed again in the graph at any time later on.
N.B.
The measurement range of the frequency entries is 1 Hz to 30000 Hz. If, for example, the rpm of a motor is recorded and this motor stops abruptly, the last rpm can still show in the display for up to 2 seconds, depending on the measurement system, before 0 is displayed.
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
Loading...
+ 53 hidden pages