HYDAC HDA 5500-0-0-DC-000 User Manual

Page 1
Digital Display Unit HDA 5500
User Manual
HDA 5500-0-0-AC-000 HDA 5500-0-0-DC-000
Dated 14.06.2005
Page 2
HDA 5500 User Manual
CONTENTS
1.1 European standards compatibility 3
2 Safety instructions 3
6.1 Mechanical installation 5
6.2 Electrical connection 5
6.3 Supply voltage 6
7 Programming 7
7.1 Adjusting the basic settings 7
7.2 Overview of basic settings menu 8
8.1 HDA 5500 with one transmitter 10
9 Digital display 11
9.1 Standard display 11
9.2 Special display 11
10 Output response 12
10.1 Analogue output 12
10.1.1 Analogue output set to 4..20 mA 12
10.1.2 Analogue output set to 0..10 V 12
10.2 Switching outputs 13
10.3 Connection of up to four switching outputs 13
11 Adjusting switching points and hystereses 13
12 Programming enable 13
12.1 Altering the operating programming enable 13
12.2 Altering the main programming enable 14
13 Technical specifications 15
14 Model code HDA 5500 16
15 Unit dimensions 17
16 Electrical connections 18
17 Error messages 19
18 HYDAC Service 19
- 2 -
Page 3
HDA 5500 User Manual
1 Introduction
The individual components and the final assembly of the Digital Display Unit HDA 5500 are subject to strict quality controls. Each HDA 5500 is individually calibrated and subjected to a final test. In this way we can guarantee that the unit is fault-free on despatch and conforms to the given specifications. However, if there is a cause for complaint, please return the unit to us outlining the fault. The HDA 5500 digital display unit is maintenance-free and operates perfectly when used according to the specifications. If faults do nonetheless arise, please contact HYDAC Serv­ice. Incorrect installation or interference by anybody other than HYDAC personnel will invalidate all warranty claims.
1.1 European standards compatibility
The HDA 5500 digital display unit carries the approval requirements currently applicable in Germany and to the European standards for operating these units. The applicable directives on electromagnetic compatibility and the safety regulations according to the low voltage directive are therefore guaranteed. This product complies with the regulations of the following European Directive:
EN 61000–6–1 / 2 / 3 / 4
We reserve the right to make technical modifications.
- mark and therefore corresponds to the
2 Safety instructions
The HDA 5500 digital display unit presents no safety concerns when installed and operated in accordance with this user manual. However, in order to avoid any risk to the operator or any damage due to incorrect handling of the unit, please adhere strictly to the following safety instructions:
The HDA 5500 must not be put into service if any known defects, either electrical or
mechanical, are apparent.
The unit must be installed exactly according to the instructions.
Read the information on the type code label.
Fault investigation and repairs must only be carried out by HYDAC Service.
All relevant and generally recognised safety requirements must be adhered to.
3 Functions
Depending on the model, the following functions are available:
Digital display of analogue signals
Temperature measurement (by means of external Pt100 temperature sensor)
Frequency measurement
• Display of a unit, e. g. bar, psi, MPa, kg/cm²,
oC,o
F, l/min (and others)
Display of a lowest or peak value (slave pointer function)
Display of a pre-set switch point
Display of differential value and switching on differential value
Counter function
Switching outputs via relay switching contacts
Analogue output 4..20 mA or 0..10 V can be selected
Basic settings menu for adapting the HDA 5500 to a particular measuring task
Programming enable functions can be disabled to prevent subsequent unauthorised
adjustment of the settings
RS 232 serial interface
- 3 -
Page 4
HDA 5500 User Manual
MODE
MODE
132
4
123
o
f active
relays
4 Operating keys
LED displays
Relais
4 digit
igital display
d
ensor
S
HDA5500
Keys for adjusting the switching points, switch-back points and the menu functions
LED displays
of active sensor
Display of
unit of measurement
or sensor 1, 2 or 3
f
5 Selection and display of the unit of measurement
The operator can select the appropriate unit label from a printed sheet and this is then displayed with background lighting. The label can be changed at any time. Procedure:
Disconnect the unit from supply voltage.
Remove the front frame of the unit by hand.
Loosen the 4 front plate screws.
Press lightly on the back panel of the unit to push out the front side (display section) by a
few millimetre.
Cut the label with the required unit from the sheet supplied.
Insert the label into the slot on the side.
Press lightly on the front panel to push it back on.
Screw the front plate back in place.
Put the front frame back on.
Indication of unit using label from sheet
Front frame
- 4 -
Page 5
HDA 5500 User Manual
6 Installation and commissioning
6.1 Mechanical installation
The HDA 5500 is a control panel unit with a standard mounting housing with the following measures for mounting:
Control panel cut-out of 92 x 45 mm
Panel thickness: at least 1.25 mm
Mounting depth: at least 150 mm
Panel clips
Panel thickness
Clamping guides
6.2 Electrical connection
The electrical connection must be carried out by a qualified electrician in accordance with relevant regulations (in Germany: VDE 0100).
Important:
In order to fulfil the relevant EMC standards, subject to the ambient conditions, the screening effect of the control box must comply with the EMC standard.
Additional installation notes which, from experience, reduce the effect of electromagnetic interference:
Make line connections as short as possible.
Use screened lines (e. g. LIYCY 4 x 0.5 mm²).
The cable screening must be fitted by qualified personnel subject to the ambient condi-
tions and with the aim of suppressing interference.
Direct proximity to interference must be avoided as far as possible.
If inductive loads are to be switched using the relays, varistors must be used on the load to prevent high switch-off surges.
- 5 -
Page 6
HDA 5500 User Manual
12..32 VDC
85..265 VAC 50 / 60 Hz
6.3 Supply voltage
The supply voltage is connected to terminal X1.
HDA 5500–X–X–DC–000 (see type code label)
HDA 5500–X–X–AC–000 (see type code label)
After switching on the supply voltage, the unit displays HdA for approximately 2 s. Then the number of the sensor set as the primary display is shown. After a further approximately 2 s the actual value is displayed.
2 s
2 s
Important: Before sensors are connected, the HDA 5500 must be set appropriately.
If there is no input signal, then the following will be displayed, depending on the type of signal:
Starting value of the measuring range - flashing (for sensors with 4..20 mA)
Starting value of the measuring range (for sensors with 0..10 V)
- 6 -
Page 7
HDA 5500 User Manual
7 Programming
In order to adapt the unit to a particular application, the HDA 5500 is programmed by varying the basic settings. The basic settings are combined in a menu.
Important: When the menu is activated, no switching operations are carried out.
7.1 Adjusting the basic settings
To change the basic settings, the menu must be activated. Procedure for activating the menu:
During normal operation, press the MODE key and hold down for at least 5 s.
MEnu (Menu) is displayed.
After approximately 2 s the display changes and the first menu point is displayed (varies
according to type).
Press MODE until the required menu point appears in the display.
After approximately 2 s the display changes and the actual setting is displayed.
Use
Press MODE again until the next required menu point appears in the display.
or to change the setting.
mode
5 s
Press Mode key and hold down until Menu appears
Procedure for adopting and storing changes and to close the menu:
Press MODE until the menu point End is displayed.
After approximately 2 s the display changes and No is displayed.
Use
Confirm with MODE.
If the basic settings have been changed and the menu has been closed with Yes, then for approximately 2 s ProG (Programming) is displayed when leaving the menu. The changed settings are then saved in the HDA 5500. Afterwards the display returns to normal operation.
Important:
If no key is pressed for longer than approximately 25 s, then the menu automatically closes down. Any changes which may have been made will not be adopted or saved.
or to change the setting to Yes.
Menu is displayed Release mode key
mode
Press mode key until the required menu point is displayed
Use the setting, then select the next menu point
or to change
- 7 -
Page 8
HDA 5500 User Manual
7.2 Overview of basic settings menu
Menu point Setting Setting range Default setting
Primary display (PriM)
Value which is normally displayed:
Measured value sensor 1
Lowest value sensor 1
Peak value sensor 1
SEn.1 Min.1 ToP.1
SEn.1
Display filter (diSP)
Display reacts slowly to measurement variations (display delay approximately 1.3 s).
Display reacts at standard rate to measurement variations (display delay approximately 0.6 s).
Display reacts quickly to measurement variations (display delay approximately 0.3 s).
Reset time (r.TiM)
Duration in seconds of for how long the last lowest / peak value is displayed.
Input signal sensor 1 (inP.1)
0..5 V
0..10 V
4..20 mA source
SLoW MEdi FAST
0..3600 0
5 V 10 V mA.r. mA.S.
MEdi
mA.S.
4..20 mA sink
Decimal place sensor 1 (dEc.1)
No. of decimal place(s) of the display of measured value of sensor 1
Lower limit sensor 1 (LOr.1)
Lower limit of the display range of sensor 1
Upper limit sensor 1 (Hir.1)
Upper limit of the display range of sensor 1
Background lighting for the unit label for sensor 1 (uni.1)
Lighting behind unit label is switched on when sensor 1 value is displayed.
Lighting behind unit label is switched off when sensor 1 value is displayed.
0..0.000 0.0
-999..9899 0
-899..9999 100.0
on oFF
on
- 8 -
Page 9
HDA 5500 User Manual
Menu point Setting Setting range Default setting
Calibration of sensor zero point (CALi.)
The actual measurement is saved as the new zero point. This is possible in the range +/- 2.5 % of the display range.
is displayed if a calibration has been carried out in the permitted range.
is displayed if a calibration has not been possible.
The function is useful for example, if the output signal always has an offset which should be displayed as the zero point.
Actual measured value is not to be saved.
YES no
no
Analogue output signal (ouT.M)
4..20 mA through analogue output signal
0..10 V through analogue output signal
Software version (VErS)
Display of the software version (for reference only)
To close basic settings (End)
Basic settings menu closes. Changed settings are saved.
Basic settings can still be altered.
MA VoLT
YES no
VoLT
no
- 9 -
Page 10
HDA 5500 User Manual
123
4
AGND
7
5
AGND
6
7
123
4
5
123
4
AGND
7
5
8 Connecting sensors
One analogue input for one transmitter is always available as standard.
8.1 HDA 5500 with one transmitter
Connection example for transmitter with a 4..20 mA sink as input signal, e. g. HDA 5500-0-0-DC-000 with HDA 4445-A-100-000 (refer to chapter 16 Pin configurations):
IN1
HDA 4445-A-...
Connection example for transmitter with a 4..20 mA source as input signal, e. g. HDA 5500-0-0-DC-000 with EDS 3446-3-0100-000 (refer to chapter 16 Pin configurations):
HDA 5500-...
Input signal sensor 1
EDS 3446-3-...
Connection example for transmitter with a 0..10 V as input signal, e. g. HDA 5500-0-0-DC-000 with HDA 4746-B-100-000 (refer to chapter 16 Pin configurations):
HDA 4746-B-...
IN1
+12 V
IN1
HDA 5500-...
Input signal sensor 1:
HDA 5500-...
Input signal sensor 1:
- 10 -
Page 11
HDA 5500 User Manual
9 Digital display
9.1 Standard display
According to the particular type, the following list of values can be set as the primary display. The pre-set value concerned will be continuously displayed.
SEn.1 (measured value from sensor 1)
The displayed value corresponds to the measured value and is proportional to the input signal of the corresponding sensor. The values are linear between the lower limit and the upper limit.
Min.1 (lowest value from sensor 1)
The displayed value corresponds to the lowest value which has been measured since the supply voltage was switched on or since the last reset.
ToP.1 (peak value from sensor 1)
The displayed value corresponds to the highest value which has been measured since the supply voltage was switched on or since the last reset.
Notes:
If the measured value exceeds the upper display range of the HDA 5500, then the
measured value can no longer be displayed. The upper display range flashes.
If the measured value is below 1 % of the lower display range, then the measured value
can no longer be displayed. The lower display range flashes.
9.2 Special display
If a measured value is set as the primary display, then by pressing the following values can be displayed separately. These values are each displayed for approxi­mately 3 s:
Lowest value
Peak value
The relevant lowest and peak values are reset by pressing the ously during normal operation. If the reset has been successful, rES (reset) is displayed to confirm this.
and keys simultane-
or keys the
- 11 -
Page 12
HDA 5500 User Manual
Analog. Out
AGND
min
V
AGND
RL
max
= 400
Analog. Out
10 Output response
The HDA 5500 has a through analogue output which is proportional to the input signal. In addition, depending on the model and type, without switching outputs or two or four switching outputs are available.
10.1 Analogue output
On the analogue output a 4..20 mA or 0..10 V signal is available. This signal is proportional to the input signal of the sensor. It’s type can be adjusted during normal operation.
10.1.1 Analogue output set to 4..20 mA
Connection example for analogue output set to 4..20 mA:
I
= 4..20 mA
ut
o
10.1.2 Analogue output set to 0..10 V
Connection example for analogue output set to 0..10 V:
= 0..10 V
out
RL
2,000
- 12 -
Page 13
HDA 5500 User Manual
10.2 Switching outputs
Not available on this type.
10.3 Connection of up to four switching outputs
Not available on this type.
11 Adjusting switching points and hystereses
Procedure for HDA 5500 without switching outputs:
During normal operation, press mode key.
Since no switching outputs are available, noSP (no switch point) will be displayed.
After approximately 2 s the display returns to normal operation.
Note:
If the mode key is pressed for longer than approximately 5 s, then the Basic Settings Menu will be activated.
12 Programming enable
The unit has two types of programming enable, the operating programming enable and the main programming enable. Both types must be enabled to change the settings. The operating programming enable provides protection from unintentional changes to the settings during normal operation. The operating programming enable can be altered during operation. The main programming enable provides general protection from changes to the settings. The main programming enable can only be altered when the HDA 5500 is switched on.
12.1 Altering the operating programming enable
Procedure:
During normal operation, press the
and keys simultaneously and hold down for at
least 3 s.
OPrG (Operating Programming) is displayed.
After approximately 2 s the display changes and the actual setting is displayed:
Free (programming enabled) or Loc (programming disabled).
Within approx. 3 s the setting can be changed to Loc or Free by pressing
or .
If or is not pressed for approximately 3 s or longer, then the displayed setting is
accepted. If a setting has been altered, then ProG (programming) is displayed for ap­proximately 2 s. This means the changed setting is stored in the HDA 5500. Afterwards the display returns to normal operation.
Displays measured value
3 s
mode
Press both arrow keys simultaneously and hold down for 3 s
3 s
Display (release arrow keys)
2 s
Use to change setting
or
= programming enabled
FrEE Loc
= programming disabled
mode
- 13 -
Page 14
HDA 5500 User Manual
12.2 Altering the main programming enable
The main programming enable can only be altered when switching on the HDA 5500. If the HDA 5500 is already in operation, then the supply voltage must first be switched off or the HDA 5500 must be disconnected from the supply voltage. Procedure:
When switched off, press both arrow keys simultaneously and hold down.
Switch on or connect the supply voltage and at the same time continue to hold down the
arrow keys for at least 3 s.
MPrG (Main Programming) is displayed.
After approximately 2 s the display changes and the actual setting is displayed:
Free (programming enabled) or Loc (programming disabled).
Within approximately 3 s the setting can be changed to Loc or Free by pressing
the
If
setting is accepted. If a setting has been changed, then ProG (Programming) is dis­played for approximately 2 s. This means the changed setting is stored in the HDA 5500. Afterwards the display returns to normal operation.
or key.
or have not been pressed for approximately 3 s or longer, then the displayed
mode
Press both arrow keys simultaneously and hold down. Switch on supply voltage (hold down keys for 3 s)
3 s
Display (release arrow keys)
2 s
mode
Use to change setting
or
FrEE Loc
= programming enabled
= programming disabled
Displays
measured value
3 s
- 14 -
Page 15
HDA 5500 User Manual
13 Technical specifications
Display range
Display 4-digit 7-segment LED display, red, height of digits 14.2 mm
3 LED for active sensor
4 LED for switching relays Display range - 999..9999 (adjustable by user) Display units with background lighting bar, kg/cm², MPa, psi, °C, °F, l/min, mA, V, Hz, kN, m, mm, l, gal,
gal/min, 1/min, and more
Input data Analogue signal input(s)
Measuring range(s) (up to 3 analogue inputs) adjustable: 4..20 mA, 0..5 V or 0..10 V Accuracy class
PT100 input
Measuring range - 25..100 °C Accuracy class
Frequency / counter input
Signal threshold 0..0.6 V = LOW
Frequency range 15 Hz .. 4 kHz
Output data
Analogue output
Accuracy of the analogue output Rise time 70 ms
Switching outputs
Type 2 or 4 relays each with separate common supply Switching voltage 0.1..250 VAC Switching current 9 mA .. 2 A Switching capacity 400 VA, 50 W (for inductive loads, use varistors) Life expectancy of the switching contacts
Reaction time (with switching delay = 0 ms) approximately 20 ms Setting range of the switching points 1.5..100 % of the pre-set display range Setting range of the switching hystereses (switch-back points) Interface Serial interface RS232
Ambient conditions
Nominal temperature range 0 .. + 50 °C Operating temperature range 0 .. + 50 °C Storage temperature range - 40 .. + 80 °C
- mark
Other data
Housing Control panel housing 96 x 48 x 109 mm
Electrical connections Supply voltage: plug-in terminal block, 2 pole, RM 5.08
Supply voltage 85..265 VAC 50 / 60 Hz or
Power consumption 15 VA at 85..230 VAC, fuse protection 1 A inert Supply of the measurement transmitter Protection class to DIN 40050 IP 20 Reverse polarity protection of the supply voltage, excess voltage, override protection Weight approximately 320 g
± 0.5 % at 25 °C
± 0.5 % at 25 °C
3..24 V = HIGH
adjustable: 4..20 mA, Ohmic resistance 400 L or
0..10 V, Ohmic resistance  2 kL
± 0.5 % at 25 °C
20 million at minimum load
1 million at nominal load
0.5..99 % of the pre-set display range
Baud rate 19,200 bauds – 8 data bits – 1 start and stop bit –
no parity – no handshake
EN 50081-1 and –2 EN 50082-1 EN 61000-6-2
Control panel cut-out 92 (+0.8) x 45 (+0.6) mm
Front panel thickness 1.25.. 5 mm
Maximum installation depth 121 mm
Inputs / outputs: plug-in terminal block, 11 pole, RM 3.5
Relays: plug-in terminal block 6 pole, RM 5.08
Max. cross section of the connections 1.5 mm² for inputs / outputs
Max. cross section of the connections 2.5 mm² for supply voltage and
relays
12..32 VDC
12 ±1 % VDC max. 20 mA per analogue input
provided
- 15 -
Page 16
HDA 5500 User Manual
AC = 85..265 VAC
14 Model code HDA 5500
Order details
HDA 5 5 0 0 – X – X – X X – 0 0 0
Inputs
0 = one analogue input 1 = three analogue inputs 2 = one analogue input + frequency input / counter function 3 = one analogue input + Pt100 input
Outputs
0 = without switching outputs 1 = two relay outputs 2 = four relay outputs
Supply voltage
DC = 12..32 VDC
Modification
000 = standard
- 16 -
Page 17
HDA 5500 User Manual
15 Unit dimensions
Control panel thickness
Control panel cut-out
- 17 -
Page 18
HDA 5500 User Manual
16 Electrical connections
HDA 5500-X-0-XC-000
HDA 5500-X-1-XC-000
HDA 5500-X-2-XC-000
Supply voltage plug X1
UNIT PIN DESCRIPTION Models
HDA 5500 – X – X – AC – 000
HDA 5500 – X – X – DC – 000 + / - Supply voltage inputs 24 VDC 1 , 2 and 3
~ / ~
X1
Supply voltage inputs 85 VAC to 265 VAC 50 / 60 Hz 1 , 2 and 3
Signals plug X2
X2
UNIT 12 3 4567891011 Models
HDA 5500 – 0 – X – XX – 000 RXD TXD DGND
HDA 5500 – 1 – X – XX – 000 RXD TXD DGND
HDA 5500 – 2 – X – XX – 000 RXD TXD DGND
HDA 5500 – 3 – X – XX – 000 RXD TXD DGND
Analogue
output
Analogue
output
Analogue
output
Analogue
output
AGND +12 V
AGND +12 V
AGND +12 V
AGND +12 V
Sensor 1
input
Sensor 1
input
Sensor 1
input
Sensor 1
input
Sensor 2
Sensor 3
input
Freq Start Stop Clear 1 , 2 and 3
Pt100 Pt100 Pt100 Pt100 1 , 2 and 3
AGND 1 , 2 and 3
input
1 , 2 and 3
Relay plugs X3 / X4
UNIT 12 3 456123456Models
HDA 5500 – X – 0 – XX – 000 1
HDA 5500 – X – 1 – XX – 000
HDA 5500 – X – 2 – XX – 000
X3 X4
N/C
N/C
R1
R1
Common
supply
R1
R1
Common
supply
R1
N/O
R1
N/O
N/C
N/C
R2
R2
Common
supply
R2
Common
supply
R2
R2
N/O
R2
N/OR3N/C
R3
Common
supply
N/O
2
R4 N/C
R4
Common
supply
N/O
R4
3
R3
- 18 -
Page 19
HDA 5500 User Manual
17 Error messages
If an error is detected, then a corresponding error message appears which must be acknowl­edged by pressing any key. Possible error messages are as follows: Er.01 The switching points and hystereses have been set in such a way that the resulting
switch-back point is no longer within the permissible setting range. Example: Switching point is set to 180 bar, the hysteresis to 200 bar.
Action: Correct the hysteresis setting.
Er.02 The minimum difference in the given measuring ranges between lower display
range and upper display range is too low. Examples: Lower display range = 0.025 and upper display range = 0.075 Lower display range = 0.25 and upper display range = 0.75 Lower display range = 2.5 and upper display range = 7.5 Lower display range = 25 and upper display range = 75 Action: Correct measuring range settings as follows: Lower display range = 0.0XX => minimum value for upper display range = 0.1XX Lower display range = 0.XXX => minimum value for upper display range = 1.XXX Lower display range = X.XXX => minimum value for upper display range = 1X.XX Lower display range = XX.XX => minimum value for upper display range = 1XX.X
Er.03 The difference of the measuring values between sensor 1 and sensor 2 is outside
the measuring range. Action: Check all the settings (programming enables, switching points, switch­back points and basic settings) and correct them if necessary. If the errors occur frequently, please contact HYDAC Service.
E.10 A data error has been detected in the saved settings. Possible causes are strong
electromagnetic interference or a component fault. Action: Check all the settings (programming enables, switching points, switch­back points and basic settings) and correct them if necessary. If the errors occur frequently, please contact HYDAC Service.
The display shows four horizontal segments. This means that the sensor supply voltage (+12 V, plug X2.6) is overloaded. Action: Ensure that the above mentioned supply voltage of +12 V has not short­circuited against AGND. Check that the maximum current has not exceeded 100 mA.
18 HYDAC Service
HYDAC Service is available for enquiries about repairs or alterations:
Subject to technical modifications
HYDAC TECHNOLOGY GMBH
HYDAC Service Hauptstr. 27 D-66128 Saarbrücken Germany
Tel.: +49-(0)681-7099-237 Fax: +49-(0)681-7099-239
HYDAC ELECTRONIC GMBH
Hauptstr. 27 D-66128 Saarbrücken Germany
Tel.: +49-(0)681-7099-0 Fax: +49-(0)681-7099-202
- 19 -
Loading...