Human Combilyer Plus User manual

User Manual
!"#$%&'()*+,-./
Cat.-No.: 17601P
!
CONTENTS
1 INTRODUCTION 3
1.1 Methodology of Urine Testing 3
1.2 Test Strips 3
2 DESCRIPTION OF COMBILYZER PLUS 4 3 INSTALLATION 6 4 OPERATION GUIDE 8
4.1 Control Menu System 8
4.2 Setup Procedure (Menu/Setup) 8
4.2.1 Setting the Clock 8
4.2.2 Setting Serial Output Port 9
4.2.3 Setting the Language 9
4.3 Strip Setup Options (Menu/Strip) 10
4.3.1 Setting Measurement Unit 10
4.3.2 Setting the Sensitivity Attribute of the Image Processing Program 10
4.3.3 Setting Pad Sequence in the Printout 11
4.4 Service Menu Options (Menu/Service) 12
4.4.1 Code Card 12
4.4.2 Selecting Measuring Mode 13
4.5 Measurement 13
4.5.1 Setting Patient ID for Measurement 13
4.5.2 Performing Measurement 13
4.6 Memory/Data Management 15
4.6.1 Result Printout 15
4.6.2 Sending data to Host Computer 16
4.6.3 Enter Patient ID for Memory Record 16
4.6.4 Clearing Test Data from Memory 16
4.7 Printer Station 17
4.8 Serial Port 17
5 MAINTENANCE 18 6 ERROR MESSAGES AND TROUBLESHOOTING 19 7 TECHNICAL FEATURES 20 8 PACKING LIST 20 9 SYMBOLS 21 10 APPENDIX 22
This page left blank
2/22 User Manual COMBILYZER Plus
1 INTRODUCTION
1.1 Methodology of Urine Testing
Urinalysis is one of the medical diagnosis methods frequently used by medical
doctors in laboratories in order to reveal diseases. This micro-chemistry system has been available for many years and allows qualitative and semi-quantitative analysis within one minute by simple but careful observation. The colour change occurring on each segment of the strip is compared to a colour chart to obtain the result. However, misreading or misinterpreting of results may happen, caused by individual handling habits of different users or different light conditions.
1.2 Test Strips
The COMBILYZER Plus works with COMBINA 11S urine test strips:
Tested parameters are as follows:
Bilirubin Urobilinogen Ketones Ascorbic Acid Glucose Protein Blood pH Nitrite Leucocytes Specific Gravity
User Manual COMBILYZER Plus 3/22
2 DESCRIPTION OF COMBILYZER Plus
Figure
1
. Parts of COMBILYZER Plus
The COMBILYZER Plus is
designed for urine test strip reading. The instrument includes two main functional units:
- portable, battery-operated
Reader Unit for urine analysis
- bench-top, AC-powered Printer
Station, which docks with the portable Reader Unit and produces printed reports.
The Reader is menu-driven, so the operator needs to set up the Reader Unit first according to the operational conditions with the help of the function menus. After placing the moistened urine strip on the strip holder the instrument reads and evaluates the strips automatically.
Figure 2. The Reader Unit
The urine test strip containing chemical reagents is positioned under the reader's sensor. After the pads' colours have been evaluated, the strip should be discarded by the operator. The testing corresponds with the reaction time of the chemical reagents - about 60 seconds. The COMBILYZER Plus works using the reflectance photometry method: inside
the reader, the strip is illuminated by white light, and the light reflected from the strip is detected by a colour CCD sensor. The RGB signals are digitised using analogue-digital conversion. The digitised image is processed by a built-in program: it locates the pads and, based on the reflected light, calculates the results. The results, together with the current date, time and patient ID, are
4/22 User Manual COMBILYZER Plus
stored and can be printed out by connecting the reader to the Printer Station. The instrument is designed specifically for in vitro diagnostic professional use and to meet the requirements of medical professionals. The measurement capacity is up to 80 test strip evaluations per hour.
User Manual COMBILYZER Plus 5/22
3 INSTALLATION
-
COMBILYZER Plus
-
-
-
-
The COMBILYZER Plus is shipped in a cardboard box. Prior to unpacking, clear the area where the instrument is to be operated: a 30 x 50 cm size space will be needed. Please note the shipping marks on the box while handling. Open the box. Cut the tape only, leave the carton material intact. (It is recommended to keep the packaging materials)
Prepare the accessories: power adapter, rechargeable batteries and printer paper.
Place
operating position.
Remove the Reader Unit from its
docked position, locate its battery holder on the bottom, remove the cover-plate and load the batteries (included) as shown in figure 3. Reconnect the Reader Unit in its docked position.
Using the standard RS232 serial
cable (included) to connect the Reader Unit's serial port (see figure
2) to the computer's serial port if you intend to upload data to an external device i.e. Computer
Connect the power adapter first to
the DC socket on the reader, then to the mains.
Switch the COMBILYZER Plus on
by pressing the Power-On button. In a few seconds the HUMAN logo will appear on the LCD, followed by the main menu, which offers the Memory/Meas/Menu options.
in its
6/22 User Manual COMBILYZER Plus
Figure 3. Installation of the batteries
Note: If during power on process the user keeps any part of the LCD pressed (touched), the boot up process will be interrupted and the system information can be viewed (serial number, software version, etc.). After releasing the screen the boot up process will continue.
- Take a roll of printer paper and load it
into the printer paper holder (see fig.
4) Insert the paper into the paper feeder slot and close the cover plate. The printer paper will be feed automatically to its start position. Now, the COMBILYZER Plus is
ready to operate using the Main Menu options (see chapter 9).
Figure 4. Loading the printer paper
Note: After installation check whether the instrument is operating accurately.
User Manual COMBILYZER Plus 7/22