Urinalysis is one of the medical diagnosis methods frequently used by medical
doctors in laboratories in order to reveal diseases.
This micro-chemistry system has been available for many years and allows
qualitative and semi-quantitative analysis within one minute by simple but careful
observation. The colour change occurring on each segment of the strip is
compared to a colour chart to obtain the result. However, misreading or
misinterpreting of results may happen, caused by individual handling habits of
different users or different light conditions.
1.2 Test Strips
The COMBILYZER Plus works with COMBINA 11S urine test strips:
Tested parameters are as follows:
Bilirubin
Urobilinogen
Ketones
Ascorbic Acid
Glucose
Protein
Blood
pH
Nitrite
Leucocytes
Specific Gravity
User Manual COMBILYZER Plus3/22
2 DESCRIPTION OF COMBILYZER Plus
Figure
1
. Parts of COMBILYZER Plus
The COMBILYZER Plus is
designed for urine test strip reading.
The instrument includes two main
functional units:
- portable, battery-operated
Reader Unit for urine analysis
- bench-top, AC-powered Printer
Station, which docks with the
portable Reader Unit and
produces printed reports.
The Reader is menu-driven, so
the operator needs to set up the
Reader Unit first according to the
operational conditions with the
help of the function menus. After
placing the moistened urine strip
on the strip holder the instrument
reads and evaluates the strips
automatically.
Figure 2. The Reader Unit
The urine test strip containing chemical reagents is positioned under the reader's
sensor. After the pads' colours have been evaluated, the strip should be
discarded by the operator. The testing corresponds with the reaction time of the
chemical reagents - about 60 seconds.
The COMBILYZER Plus works using the reflectance photometry method: inside
the reader, the strip is illuminated by white light, and the light reflected from the
strip is detected by a colour CCD sensor. The RGB signals are digitised using
analogue-digital conversion. The digitised image is processed by a built-in
program: it locates the pads and, based on the reflected light, calculates the
results. The results, together with the current date, time and patient ID, are
4/22User Manual COMBILYZER Plus
stored and can be printed out by connecting the reader to the Printer Station.
The instrument is designed specifically for in vitro diagnostic professional use
and to meet the requirements of medical professionals. The measurement
capacity is up to 80 test strip evaluations per hour.
User Manual COMBILYZER Plus5/22
3 INSTALLATION
-
COMBILYZER Plus
-
-
-
-
The COMBILYZER Plus is shipped in a cardboard box. Prior to
unpacking, clear the area where the instrument is to be operated: a 30 x
50 cm size space will be needed. Please note the shipping marks on the
box while handling. Open the box. Cut the tape only, leave the carton
material intact. (It is recommended to keep the packaging materials)
Prepare the accessories: power adapter, rechargeable batteries and
printer paper.
Place
operating position.
Remove the Reader Unit from its
docked position, locate its battery
holder on the bottom, remove the
cover-plate and load the batteries
(included) as shown in figure 3.
Reconnect the Reader Unit in its
docked position.
Using the standard RS232 serial
cable (included) to connect the
Reader Unit's serial port (see figure
2) to the computer's serial port if
you intend to upload data to an
external device i.e. Computer
Connect the power adapter first to
the DC socket on the reader, then
to the mains.
Switch the COMBILYZER Plus on
by pressing the Power-On button.
In a few seconds the HUMAN logo
will appear on the LCD, followed by
the main menu, which offers the
Memory/Meas/Menu options.
in its
6/22User Manual COMBILYZER Plus
Figure 3. Installation of the batteries
Note: If during power on process the user keeps any part of the LCD pressed
(touched), the boot up process will be interrupted and the system information can
be viewed (serial number, software version, etc.). After releasing the screen the
boot up process will continue.
- Take a roll of printer paper and load it
into the printer paper holder (see fig.
4) Insert the paper into the paper
feeder slot and close the cover plate.
The printer paper will be feed
automatically to its start position.
Now, the COMBILYZER Plus is
ready to operate using the Main Menu
options (see chapter 9).
Figure 4. Loading the printer
paper
Note: After installation check whether the instrument is operating accurately.
User Manual COMBILYZER Plus7/22