HT instruments SIRIUS89N User Manual

SIRIUS89N
User’s Manual
© Copyright HT ITALIA 2008 Release 2.02 - 21/01/2008
SIRIUS89N
TABLE OF CONTENTS:
1. SAFETY PRECAUTIONS AND PROCEDURES.......................................................................................... 4
1.1. Forwards..................................................................................................................................................................4
1.2. Preliminary Instruction.............................................................................................................................................4
1.3. During Use...............................................................................................................................................................5
1.4. After Use..................................................................................................................................................................5
2. GENERAL DESCRIPTION........................................................................................................................... 6
2.1. Introduction..............................................................................................................................................................6
2.2. Functions...................................................................................................................... ...........................................6
3. PREPARATION FOR USE........................................................................................................................... 7
3.1. Initial Control............................................................................................................................................................7
3.2. Power Supply...........................................................................................................................................................7
3.3. Calibration................................................................................................................................................................8
3.4. Storage....................................................................................................................................................................8
4. INSTRUMENT DESCRIPTION..................................................................................................................... 9
4.1. Display Description................................................................................................................................................10
4.2. Initial Screen..........................................................................................................................................................10
4.3. Backlight function...................................................................................................................................................10
5. INITIAL SETTINGS......................................................................................................................................11
5.1. How to Adjust the Contrast....................................................................................................................................11
5.2. How to Set Date and Time.....................................................................................................................................11
5.3. How to Set the Language......................................................................................................................................11
5.4. Reset.....................................................................................................................................................................12
6. SAFETY TEST FUNCTIONS.......................................................................................................................13
6.1. LOW: Continuity Test with 200mA Test Current.................................................................................................13
6.1.1. Calibrating the test leads ("CAL" Mode)..........................................................................................................................14
6.1.2. Measurement Procedure.................................................................................................................................................15
6.1.3. Results of "AUTO" mode.................................................................................................................................................16
6.1.4. Results of "RT+" and "RT-" modes..................................................................................................................................16
6.1.5. "AUTO", RT+", "RT-" faulty cases....................................................................................................................................17
6.2. M: Insulation resistance Measurement with 50V, 100V, 250V, 500V, 1000V Test Voltage......................................19
6.2.1. Measurement Procedure.................................................................................................................................................19
6.2.2. Results of "MAN" mode...................................................................................................................................................21
6.2.3. Results of "TMR" mode....................................................................................................................................................22
6.2.4. "MAN" and "TIMER" mode faulty cases...........................................................................................................................23
6.3. RCD: Test on "A" and "AC" RCDs Type................................................................................................................24
6.3.1. Tripping times for the general and selective RCDs..........................................................................................................26
6.3.2. Measurement procedure..................................................................................................................................................27
6.3.3. RCD Faulty cases............................................................................................................................................................32
6.4. LOOP : Measurement of Line Impedance, Fault loop Impedance, Prospective Short Circuit Current
Calculation and Phase Sequence Indicator...........................................................................................................37
6.4.1. High resolution Impedance measurement (0.1mΩ)........................................................................................................37
6.4.2. Measurement procedure and results of "P-N" mode.......................................................................................................38
6.4.3. Measurement procedure and results of "P-P" mode........................................................................................................39
6.4.4. Measurement procedure and results of "P-PE" mode.....................................................................................................40
6.4.5. Measurement procedure and results of " RA" mode...................................................................................................42
6.4.6. Measurement procedure and results of " " mode.......................................................................................................43
6.4.7. LOOP Faulty Cases..................................................................................................................................................44
6.5. EARTH: Soil Resistance and Resistivity Measurements.......................................................................................49
6.5.1. Measurement procedure and results of "2-W"and "3-W" mode.......................................................................................50
6.5.2. Measurement procedure and results of "ρ" mode...........................................................................................................51
6.5.3. "2-W", "3-W" and "ρ" faulty cases....................................................................................................................................52
7. AUX: MEASUREMENT WITH EXTERNAL PROBES.................................................................................55
7.1. Environmental parameter and leakage current: REAL TIME MEASUREMENT....................................................56
7.2. Environmental parameter and leakage current: RECORDING..............................................................................58
SIRIUS89N
7.2.1. AUX Basic setting: RECORDER CONFIG.......................................................................................................................58
7.2.2. RECORDING: setting of Typical Configurations..............................................................................................................60
7.3. Sound level measurement procedure....................................................................................................................61
8. ANALYZER .................................................................................................................................................63
8.1. Basic Setting: ANALYZER CONFIG......................................................................................................................64
8.1.1. Type of electrical system under test................................................................................................................................64
8.1.2. How to set the fundamental frequency............................................................................................................................64
8.1.3. How to set the current range ...........................................................................................................................................64
8.1.4. Clamp Type .....................................................................................................................................................................64
8.1.5. How to set the value of the transformer voltage ratio (TV RATIO)..................................................................................65
8.1.6. How to enable/disable the password...............................................................................................................................65
8.2. Basic Setting: RECORDER CONFIG ....................................................................................................................66
8.3. Analyzer functions..................................................................................................................................................73
8.4. "VOLTAGE" Function.............................................................................................................................................73
8.4.1. Symbols...........................................................................................................................................................................73
8.4.2. "METER" mode................................................................................................................................................................74
8.4.3. "HARM" mode..................................................................................................................................................................75
8.4.4. "WAVE" mode..................................................................................................................................................................76
8.5. "CURRENT" Function............................................................................................................................................77
8.5.1. Symbols...........................................................................................................................................................................77
8.5.2. “METER" mode................................................................................................................................................................78
8.5.3. “HARM" mode..................................................................................................................................................................79
8.5.4. "WAVE" mode..................................................................................................................................................................80
8.6. "POWER" Function................................................................................................................................................81
8.6.1. Symbols...........................................................................................................................................................................81
8.6.2. "METER" mode................................................................................................................................................................82
8.6.3. "WAVE" mode..................................................................................................................................................................83
8.7. "ENERGY" Function..............................................................................................................................................84
8.7.1. Symbols...........................................................................................................................................................................84
8.7.2. "METER" mode................................................................................................................................................................85
8.8. Measuring Procedures ...........................................................................................................................................86
8.8.1. Using the Instrument in a Single Phase System..............................................................................................................86
8.8.2. Using the Instrument in a Three Phase System..............................................................................................................87
9. SAVING RESULTS......................................................................................................................................88
9.1. Saving Safety Test Results....................................................................................................................................88
9.2. Saving Displayed Values of ANALYZER Function.................................................................................................88
10. RECORDINGS.............................................................................................................................................89
10.1. Start a Recording...................................................................................................................................................89
10.2. Setting Typical Configurations...............................................................................................................................90
10.2.1. Default Configuration.......................................................................................................................................................90
10.2.2. Typical Configurations.....................................................................................................................................................91
10.3. During a Recording................................................................................................................................................94
10.3.1. MENU key........................................................................................................................................................................94
10.3.2. Rotary Switch during a recording.....................................................................................................................................95
10.4. Stopping a Recording or an Energy Measurement................................................................................................95
11. INSTRUMENT'S MEMORY.........................................................................................................................96
11.1. Safety test memory................................................................................................................................................96
11.2. Analyzer memory...................................................................................................................................................97
12. CONNECTING THE INSTRUMENT TO A PC.............................................................................................98
13. MAINTENANCE ..........................................................................................................................................99
13.1. General Instruction ................................................................................................................................................99
13.2. Battery Replacement.............................................................................................................................................99
13.3. Instrument Cleaning...............................................................................................................................................99
13.4. End of life...............................................................................................................................................................99
14. TECHNICAL SPECIFICATIONS...............................................................................................................100
14.1. Technical Features ..............................................................................................................................................100
14.1.1. Safety Test functions.....................................................................................................................................................100
14.1.2. ANALYZER and AUX functions.....................................................................................................................................101
SIRIUS89N
14.2. Standards ............................................................................................................................................................103
14.2.1. General..........................................................................................................................................................................103
14.2.2. Safety Test.....................................................................................................................................................................103
14.2.3. ANALYZER....................................................................................................................................................................103
14.2.4. AUX ...............................................................................................................................................................................103
14.3. General Specifications.........................................................................................................................................103
14.3.1. Mechanical Data............................................................................................................................................................103
14.3.2. Power supply.................................................................................................................................................................103
14.3.3. Display...........................................................................................................................................................................104
14.3.4. Memory..........................................................................................................................................................................104
14.4. ENVIRONMENT..................................................................................................................................................104
14.5. ACCESSORIES...................................................................................................................................................104
15. SERVICE...................................................................................................................................................105
15.1. WARRANTY CONDITIONS.................................................................................................................................105
15.2. SERVICE.............................................................................................................................................................105
16. PRACTICAL REPORTS FOR ELECTRICAL TESTS ...............................................................................106
16.1. Continuity Test On Protective Conductors...........................................................................................................106
16.2. Insulation Resistance Measurement of the Electrical Installations (250Vdc, 500Vdc, 1000Vdc).........................107
16.3. Check of the Circuit Separation................................................................................................ ...........................111
16.4. Earth Resistance Measurement in TT Systems ...................................................................................................114
16.5. Working Test of RCDs (Rcd, Rcd/Dc, Rcd S, Rcd/Dc S).....................................................................................115
16.6. Test of RCD Tripping Time (Rcd, Rcd/Dc)...........................................................................................................116
16.7. Measurement of Short-Circuit Fault Impedance (Zpn, Zpp)....................................................................................117
16.8. Fault Loop Impedance Measurement (Phase – Earth)........................................................................................117
16.9. Earth Resistivity Measurement.................................................................................................. ..........................118
16.10. Voltage Anomalies (Voltage S and Surge)..................................................................................................120
16.11. Voltage and current Harmonics...................................................................................................................120
16.11.1.
Theory..................................................................................................................................................................120
16.11.2.
Limit values for harmonics...................................................................................................................................122
16.11.3.
Presence of harmonics: causes...........................................................................................................................122
16.11.4.
Presence of harmonics: consequences...............................................................................................................123
16.12. Power and Power Factor definition..............................................................................................................123
16.12.1.
Conventions on powers and power factors..........................................................................................................125
16.12.2.
3 Phase 3 Wire System .......................................................................................................................................126
16.13. Measuring Method: outlines........................................................................................................................127
16.13.1.
Integration periods...............................................................................................................................................127
16.13.2.
Power factor calculations.....................................................................................................................................127
17. APPENDIX 1 – MESSAGES DISPLAYED................................................................................................128
18. APPENDIX 2 – RECORDABLE PARAMETERS: SYMBOLS ...................................................................129
SIRIUS89N
1. SAFETY PRECAUTIONS AND PROCEDURES
1.1. FORWARDS
This apparatus conforms with safety standards EN61557 and EN 61010-1 relating to electronic measuring instruments.
For your own safety as well as that of the apparatus you are recommended to follow the procedures described in this instruction manual and carefully read all
the notes preceded by the symbol .
Strictly keep to the following instructions before and during measurements:
Do not take measurements in wet environments or dusty places. Do not effect measurements in environments with explosive gas, fuels. Keep you insulated from the object under test waiting for measuring. Avoid any contact with exposed metal parts, ends of test leads not in use, circuits, etc. Do not effect any measurement in case of unusual conditions of the instrument such as
deformation, breakage, leakage of substances, absence of display reading etc.
Do not use the External power supply adapter (optional code A0050) if you notice
deformation, or breakage in the case, in the wire or in the plugs.
Pay careful attention when measuring voltages exceeding 25V in particular places
(building yards, swimming pools, etc.) and 50V in ordinary places because of the risk of electric shock.
Use only cables and accessories approved by HT Italia. The following symbols are used in this manual:
Caution: refer to the instructions reported on this manual; improper use may damage the apparatus or its components.
CAUTION
AC Voltage or Current.
Unidirectional pulsating Voltage or Current.
Rotary switch of the instrument.
1.2. PRELIMINARY INSTRUCTION
This instrument has been designed for use in environments with a pollution level 2 and
up to (and no more than) 2000 meters altitude.
It can be used for Safety Test on Installation with Overvoltage Category III 300V~
(phase to earth) and for voltage and current measurements on installations with overvoltage category III 600 V~ phase to phase / 300 V~ phase to earth or CATII 350 V phase to earth.
SIRIUS89N
Please keep to the usual safety standards aimed at:
Protecting against dangerous currents; Protecting the instrument against incorrect operations.
Only the accessories supplied with the instrument guarantee compliance with the
safety standards. Accordingly, they must be in good conditions and, if necessary, they must be replaced with identical models.
Do not take measurements on circuits exceeding the specified current and voltage
limits.
Before connecting cables, crocodiles and clamps to the circuit under test, make sure
that the right function has been selected.
Do not effect any measurement under environmental conditions beyond the limits
specified in paragraph 14.4.
Check that batteries are not weak and placed correctly. Before connecting test leads to the circuit under test, check that rotary switch position
is correct.
1.3. DURING USE
Please read carefully the following recommendations and instructions:
CAUTION
No compliance with the Warnings and/or Instructions may damage the apparatus and/or its components or injure the operator.
Before selecting any function disconnect the test leads from the circuit under test. When the instrument is connected to the circuit under test do not touch any unused
terminal.
Avoid to effect resistance measurements in the presence of external voltages; even
though the instrument is protected a too high voltage may cause malfunctions.
When measuring current, other currents located near the leads may affect the
measuring accuracy.
When measuring current, always position the wire in the very middle of the jaws in
order to obtain the highest accuracy.
A measured value remains constant if the "HOLD" function is active. Should you notice
that the measured value remains unchanged, disable the “HOLD” function.
CAUTION
The symbol " " shows the battery charge: When it is completely black the battery are full charge, while the " " symbol indicate weak batteries. When the batteries are too low to execute the test the instrument will show a warning message. In the case interrupt testing and replace batteries following the procedure described under paragraph 13.2. The instrument is capable of
keeping the data stored even though batteries are not installed. The Instrument Date and Time settings aren't lost if you change the batteries
within 24hours.
1.4. AFTER USE
After use, turn off the instrument by pressing ON/OFF for a few seconds. Remove batteries when the apparatus remains unused for long periods. Please follow
the storage instructions described at paragraph 14.4.
SIRIUS89N
2. GENERAL DESCRIPTION
2.1. INTRODUCTION
Dear Customer, we thank you for your patronage. The instrument you have just purchased will grant you accurate and reliable measurements provided that it is used according to the present manual’s instructions. The instrument was designed to grant the user the utmost safety conditions thanks to a new concept assuring double insulation and over voltage category III.
2.2. FUNCTIONS
The instrument is able to perform eh following test: LOW: Continuity Test of Protection and Equalising conductors with test current
higher than 200mA and open circuit voltage ranging from 4V to 24V.
M: Measurement of insulation resistance with DC test voltage 50V, 100V,
250V, 500V or 1000V.
RCD: Measurement on common and/or selective RCDs AC type (
type ( ) of the following parameters:
Tripping time. Tripping current. Contact voltage (U
t
). Global earth resistance (Ra). Under this mode the instrument can measure the overall earth resistance without causing RCD tripping.
) and A
LOOP : Measurement of line and fault loop impedance with calculation of
prospective short circuit current, Measurement of fault loop impedance between phase and earth and Global Earth resistance measurement without RCD tripping and calculation of prospective short circuit current, Indication of phase rotation sequence
EARTH Measurement of Earth Resistance and Resisivity using Earth rods.  AUX: Measurement and Recording of leakage current or environmental values
(temperature, humidity, Air Speed, Illuminance and Sound level).
ANALYZER: The Instrument allows the following operations:
Display in real time the electrical parameters of a single phase
systems and the harmonic analysis of voltage and current.
Conduct a direct Energy measurement (without memorizing).  Memorize (pressing SAVE key) the sampled values of the Parameters
present at instrument input generating a "Smp" record inside instrument memory. It will be possible to analyze the memorized data ONLY by transferring it to a PC.
Record simultaneously (pressing the START key after a proper
setting): RMS values of voltage, current, corresponding harmonics, active, reactive and apparent powers, power factors and cosϕ, active, reactive and apparent energies, voltage anomalies (voltage sag and surge) with 10ms resolution. It will be possible to analyze the
recorded data ONLY by transferring them to a PC.
CAUTION
Please note the difference between memorize and record. These terms will be used repeatedly in this manual. Please focus on their definitions and distinctions.
SIRIUS89N
3. PREPARATION FOR USE
3.1. INITIAL CONTROL
This instrument has been checked mechanically and electrically prior to shipment. Any care has been taken to ensure that the instrument reaches you under safe conditions.
You are recommended, however, to carry our a rapid check to detect any possible damage which might have been caused during transport. Should this be the case, immediately contact HT Italia.
Check also that the packaging contains all the parts listed under paragraph 14.5. In case of discrepancies contact the dealer.
In case you have to send the instrument back please follow the instructions reported in paragraph 15.
3.2. POWER SUPPLY
The instrument can be powered by:
6 batteries 1.5V AA - LR6 series located in the compartment on the back of the
instrument (not included in the package). For battery life see paragraph 14.3.1.
An external power supply adapter (optional code A0050) to be used only for
ANALYSIS and AUX function. We recommend You to use only A0050 HT Power
Supply adapter.
For your own safety it's not allowed to use the external power supply adapter during Safety Test (LOW, M, RCD, LOOP, EARTH rotary Switch positions). If you press the START button the Instrument will
show the message "o REMOVE POWER".
The symbol charge, while the
shows the battery charge: If it is completely "black" the battery are full
symbol indicate weak batteries. When the batteries are too low to execute the test the instrument will show a warning message. In the case interrupt testing and replace batteries following the procedure described under paragraph 13.2. The instrument is capable of keeping the data stored even though
batteries are not installed. The Instrument Date and Time settings aren't lost if you
change the batteries within 24hours.
CAUTION
SIRIUS89N
CAUTION
For recordings (ANALYSIS and AUX function) use ALWAYS the external power supply adapter (optional code A0050) even the instrument allows the operator to perform a recording using internal batteries. If during a recording the external power supply adapter is de-energised, the instrument will continue the recording using the internal battery power until the batteries are exhausted (the data stored up to the point the instrument shuts down won’t get lost). For this we recommend you ALWAYS insert a new set of batteries before a long recording.
The instrument uses sophisticated algorithms to prolong the battery life. Particularly:
The instrument switches OFF the backlight Automatically after 5 seconds. If the instrument is displaying in real time (and the external power supply adapter is not
connected), after about 5 minutes from the last key pressure or switch rotation the instrument turns off automatically ("AUTOPOWER OFF" procedure).
If the instrument is recording or is measuring energy (and the external power supply is
not connected), after about 5 minutes from the last key pressure or switch rotation the instrument starts a special procedure to save the batteries ("ECONOMY MODE"): the instrument keeps recording but the display is turned off.
3.3. CALIBRATION
The instrument fulfils the technical specifications listed in this manual. The performance of the specifications are guaranteed for one year.
3.4. STORAGE
In order to grant the accuracy of the measurements, after a period of storage in extreme environmental conditions, wait for the time necessary so that the apparatus is back to normal measuring conditions (see environmental specifications listed in paragraph 14.4).
SIRIUS89N
4. INSTRUMENT DESCRIPTION
LEGEND:
1
2
F1 F3F4F2
1.
Display
2.
Function Keys
3.
Rotary switch
3
F1
ON
START
STOP
SAVE
HOLD
ENTER
MENU
F2 F4 F3
/
OFF
ESC
START STOP
HOLD
SAVE
ENTER
MENU ESC
Multifunction Keys.
ON/OFF and backlight key. Press it for few seconds to switch OFF the
instrument, press it briefly to activate the backlight function.
This key start (and stop). the measurement.
This key allows to save the result displayed.
This key has a double function: it is the confirmation key inside the
configuration menu and it allows to freeze the displayed results using the ANALYZER function.
This key open the General Configuration Menu.
This key quit the .modification in the configuration menu or the selected
working mode.
SIRIUS89N
4.1. DISPLAY DESCRIPTION
The display is a graphic module with a resolution of128 x 128 pixels. The first line of the display shows date and time. If not correct, you can set the exact ones according to the procedure described at paragraph 5.2. On the top right corner of the display you can always see the battery indicator and, if the external power supply adapter (optional code A0050) is connected, the corresponding symbol.
LOW 05.06.01
R+ R-
---- ----
---mA ---mA
AUTO 0.11
----
FUNC CAL
27.09.00 17:35:12
SINGLE PHASE
V1 = 230.2 V Vpk1 = 325.5 V ThdV = 0.0 % freq = 50.0 Hz
HARM WAVE
VOLTAGE
These symbols will be omitted in the following illustrations.
4.2. INITIAL SCREEN
When turning on the instrument by pressing ON/OFF, this screen will appear for a few seconds:
SIRIUS89N HT ITALIA
SN:00000000 V: X.XX
BAUD RATE 57600
CALIBRATION DATE
Here you can see:
serial number of the instrument (SN.:)
firmware software release (V.X.XX:)
transmission speed through serial RS232 (Baud Rate)
calibration date (CALIBRATION:)
4.3. BACKLIGHT FUNCTION
01.01.02
When instrument is turned on, pressing briefly the ON button, the backlight will be enabled. The light will be automatically turned off after 5 seconds. If the batteries are too low the instrument will disable automatically the backlight function.
EN - 10
SIRIUS89N
5. INITIAL SETTINGS
By pressing the MENU key the following screen will be displayed:
MENU GENERAL
SAFETY TEST MEMORY
ANALYZER MEMORY RESET
ANALYZER CONFIG RECORDER CONFIG
CONTRAST DATE&TIME LANGUAGE
It’s not possible to enter the MENU during a recording or a Real Time Energy measurement. Pressing this button during a recording the display will show main recording parameter (see paragraph 10.3).
5.1. HOW TO ADJUST THE CONTRAST
By pressing the multifunction keys F1 and F2, position the cursor on the CONTRAST item and confirm it by pressing the ENTER key. By pressing the multifunction keys F3 and F4, adjust the contrast (higher values correspond to a higher contrast while lower values correspond to a lower contrast) and press the ENTER key to SAVE the change or press ESC to quit the modification. This setting will remain unchanged after turning off the instrument.
5.2. HOW TO SET DATE AND TIME
By pressing the multifunction keys F1 and F2, position the cursor on the DATE&TIME item and confirm it by pressing the ENTER key.
The time is expressed as hh:mm (2 digit for hours, 2 digit for minutes) military time. Press the ENTER key to SAVE the change or press ESC to quit the modification. This setting will remain unchanged also after turning off the instrument.
5.3. HOW TO SET THE LANGUAGE
By pressing the multifunction keys F1 and F2, position the cursor on the LANGUAGE (EN) or LINGUA (IT) item and confirm it by pressing the ENTER key. By pressing the multifunction keys F1 and F2, position the cursor on the desired language and press the ENTER key to SAVE the change or press ESC to quit the modification. This setting will remain unchanged after turning off the instrument.
EN - 11
SIRIUS89N
5.4. RESET
This option re-establishes the initial settings of the instrument in ANALYZER function. The “Current Range” parameter it is not modified by Reset command The initial settings of the instrument consist of: ANALYZER CONFIG:
System: SINGLE Frequency: not modified Current range: not modified Clamp type: not modified Transforming ratio of voltmetric transformers: 1 Password: OFF
RECORDER CONFIG:
Start: MANU (the recording starts at
00 sec mark on clock after
pressing the START/STOP key) Stop: MAN Integration period: 15min Recording of harmonics: ON Recording of Voltage anomalies (Sag and Surge: ON Voltage Reference for Sag and Surge detection: 230V Upper Limit for Sag and Surge detection: 6% Lower Limit for Sag and Surge detection: 10% Selected voltages: V1 Selected voltage harmonics: THD, 01, 03, 05, 07 Selected currents: I1 Selected current harmonics: THD, 01, 03, 05, 07 CO-GENERATION: OFF Powers, Pf and cosϕ selected: P1 Q1i Q1c S1 Pf1 DPf1 Energie: Ea1 Eri1 Erc1
The RESET command will not
erase the instrument’s memory.
EN - 12
SIRIUS89N
6. SAFETY TEST FUNCTIONS
6.1. LOW: CONTINUITY TEST WITH 200mA TEST CURRENT
The measurement is effected according to EN 61557-2 and VDE 0413 part 4.
Before carrying out the continuity test make sure that there is no voltage at the ends of the conductor under test.
Turn the switch on LOW position.
This key permits to select one of the following measuring modes: Mode "AUTO" (the instrument carries out two measurements with
reversed polarity and displays their average value). This mode is recommended for the continuity test.
Mode "RT+" (measurement with positive polarity and possibility of setting
the duration time of the test). In this case the operator can set a measuring time long enough to permit him to move the protective conductors while the instrument is carrying out the test so detecting any bad connection.
Mode "RT-" (measurement with negative polarity and possibility of setting
the duration time of the test). In this case the operator can set a measuring time long enough to permit him to move the protective conductors while the instrument is carrying out the test so detecting any bad connection.
This key permits to execute the "CAL" mode (compensation of the resistance of the cables used for the measurement).
CAUTION
CAUTION
If the resistance is lower than 5Ω (including the resistance of the calibration) the continuity test is executed by the instrument with a test current higher than 200mA. If the resistance is higher than 5 the continuity test is executed by the instrument with a current lower than 200mA.
We recommend you to check the Calibration of the test leads before executing a measurement according to next paragraph.
EN - 13
SIRIUS89N
6.1.1. Calibrating the test leads ("CAL" Mode)
1. Connect the black and blue test leads to B1 and B4 input terminals respectively.
Connection of instrument terminals during calibration procedure.
2. If the test leads supplied with the instrument are not long enough for the measurement you can extend the blue cable.
3. Short-cir c u i t the measuring cable ends making sure that the conductiv e parts of the crocodiles make a good contact to each other (see previous picture).
4. Press the F2 key. The instrument carries out the calibration.
CAUTION
Never disconnect the test leads when the message "MEASURING" is displayed.
LOW 05.06.01
----
R+ R-
---- ----
---mA ---mA
AUTO 0.11
FUNC CAL
A numerical value in this field means that the instrument has been calibrated; this value remains on
the display for any further measurement
even though the unit is switched off and on again
5. At the end of the test the result is stored and used as OFFSET (that is to say that it is subtracted from any continuity test carried out) for all the subsequent measurements.
Note: The instrument effects the calibration only if the resistance of the test leads is
lower than 5.
TEST LEADS Before each measurement always make sure that
the calibration is referred to the cables in use. During a continuity test, if the resistance value free of calibration (that is the resistance value less the calibration offset value) is negative, the symbols o is displayed. Probably the calibration resistance value stored in the instrument memory is not referred to the cable in use, therefore a new calibration must be effected.
EN - 14
SIRIUS89N
6.1.1.1. Procedure to reset test leads calibration parameters
To cancel calibration parameters it is necessary to effect a
calibration procedure with a resistance of test leads higher than
5 (for example with
open test leads). When a cancellation is effected the screen nearby is displayed first, followed
LOW 05.06.01
o
>
99.9
R+ R-
---- ----
---mA ---mA
AUTO 0.11
FUNC CAL
Message >99.9: means that the instrument detected a resistance higher than 5 therefore it will proceed with Reset procedure.
by the screen below:
6.1.2. Measurement Procedure
1. Select the desired mode by means of the F1 key.
2. Connect the black and blue test leads to B1 and B4 input terminals respectively
Connection of the test leads during LOW test.
3. If the cables supplied with the instrument are not long enough for the measurement you can extend the blue cable.
4. Short-circuit the test leads making sure that the conductive parts of the crocodiles make a good contact to each other. Press the START key. If the displays doesn't show 0.00 repeat the test leads calibration (see paragraph 6.1.1).
5. Connect the instrument terminals to the ends of the conductor under test (see previous picture).
6. If the mode "RT+" or "RT-" was selected use the F3, F4 keys to set the duration of the test.
START
STOP
7. Press the START key. The instrument will execute the measurement. In RT+/RT- (Timer mode) you can press START key again if you want to stop the test before the duration set is expired.
Never disconnect the test leads when the message "MEASURING" is displayed.
CAUTION
EN - 15
SIRIUS89N
A
A
6.1.3. Results of "AUTO" mode
t the end of the test, if
the average resistance
value Ravg is lower
than 5 the instrument emits a double sound signal indicating the
positive outcome of
the test and displays
one screen similar to the screen alongside.
LOW 05.06.01
1.05
R+ R-
1.07 1.03 219mA 219mA
AUTO 0.11
FUNC CAL
verage resistance value
(Ravg)
Resistance values and corresponding test current got exchanging the polarities of test leads
The displayed result can be stored pressing the SAVE key twice (according to paragraph 9.1).
6.1.4. Results of "RT+" and "RT-" modes
if a resistance value
RT+ or RT+ lower than 5 is detected, the
instrument emits a
double sound signal
indicating the positive
outcome of the test
displays one screen similar to the screen alongside.
LOW 05.06.01
1.07
219mA
RT+ 0.11 TIME: 10s
FUNC CAL
Max Resistance value of R+ or R-.
Test current
Duration of the Test
Note: We recommend to use crocodiles and to check if the crocodiles make a good
contact with the conductor under test. Indeed in this test the instrument gives as final result the maximum measured value of R+ or R- and suing test leads instead of crocodiles could give you faulty result due to faulty contact between the test leads and conductor under test
The displayed result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 16
SIRIUS89N
6.1.5. "AUTO", RT+", "RT-" faulty cases
If the instrument detect the External Power supply adapter connected to instrument will show the message displayed to side.
LOW 05.06.01
-.- -
R+ R-
--- ---
---mA ---mA
o REMOVE POWER
AUTO 0.11Ω! FUNC CAL
Disconnect the External Power Supply Adapter
If the terminal voltage is higher than 15V, the instrument does not carry out the test and displays the screen alongside for 5 seconds.
LOW 05.06.01
R+ R-
-.-- -.--
---mA ---mA
AUTO 0.11
FUNC CAL
-.- -
o VOLT IN INPUT
ATTENTION: the test was not effected because of
voltage at the terminal ends.
In case that:
R
CALIBRATION>RMEASURED
the instrument displays the screen alongside.
LOW 05.06.01
o 0.00
R+ R-
0.00 0.00 219mA 219mA
AUTO 0.11
FUNC CAL
CAL > RES
ATTENTION: R
CALIBRATION >RMEASURED
THE PREVIOUS RESULTS CAN'T BE SAVED.
EN - 17
SIRIUS89N
If the value of
Resistance is higher
than 5 (but lower than
99.9Ω) the instrument
emits a long sound
signal and displays one
screen similar to the screen alongside
LOW 05.06.01
o 5.17
R+ R-
5.17 5.17
209mA 209mA
AUTO 0.11
FUNC CAL
Resistance value higher than 5
Test current
The displayed result can be stored pressing the SAVE key twice (according to paragraph 9.1).
If the value of
Resistance is higher
than 99.9 the instrument emits a long sound signal and
displays the screen alongside.
LOW 05.06.01
o> 99.9
R+ R-
-.-- -.--
---mA ---mA
AUTO 0.11
FUNC CAL
Resistance value higher than 99.9
ATTENTION: Value of Resistance Out of Range
The displayed result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 18
SIRIUS89N
6.2. M: INSULATION RESISTANCE MEASUREMENT WITH 50V, 100V, 250V,
500V, 1000V TEST VOLTAGE
The measurement comply with IEC 61557-2 and VDE 0413 part 1.
CAUTION
Before effecting the insulation test make sure that the circuit under test is not energised and all the loads are disconnected.
Turn the switch on M position.
The key F1 permits to select one of the following measuring modes:
Mode "MAN" (Manual mode). Recommended test. Mode "TMR" (Timer mode: test duration depends on the selected interval
from 10 to 999 seconds). This test can be executed when the test required a defined duration.
6.2.1. Measurement Procedure
1. Select the desired mode by means of the F1 key.
2. Connect the test leads to the instrument input terminals B1 and B4 respectively,
Example: insulation measurement between phase and earth in an electrical installation using untied cables.
3. If the cables supplied with the inst rument are not long enough for the measurement you can extend the blue cable.
4. Connect the instrument terminals to the object which is to be submitted to the insulation test after de-energizing the circuit under test and all the relative loads (see previous picture).
5. B y means of F2 select the test voltage suitable for the type of test to be carried out (see Table1). The values to be selected are:
50V (test on telecommunication systems)
100V
250V
500V
1000V
EN - 19
Example: insulation measurement between phase and earth in an electrical installation using the shuko cable.
SIRIUS89N
Standard Brief description Test voltage Maximum limit value
> 0.250M > 0.500M
> 1.0M
> 50k (se V<500V)
> 100k (se V>500V)
> 230k
> 1M
CEI 64-8/6
CEI 64-8/4
Floor and wall insulation in civil installation s
Floor and wall insulation in systems over 500V
Systems SELV or PELV
Systems up to 500V (Civil installations)
Systems over 500V
250VDC 500VDC
1000VDC
500VDC
1000VDC EN60439 Electrical panel boards 230/400V 500VDC EN60204 Electrical equipment of machines 500VDC
Table1: Table reporting the test voltage and the corresponding limit values for
few Guidelines.
Rated voltage selected
for the test
50VDC 100VDC 250VDC 500VDC
1000VDC
R
= Maximum resistance
MAX
value
99.9M
199.9M 499M 999M
1999M
Table2: Table of maximum resistance values which can be measured under M
mode depending on the rated voltage selected.
6. If the "TMR" mode was selected use the F3, F4 keys to set the duration time of the test:
CAUTION
Never disconnect the test leads from the circuit under test when the message "MEASURING" is displayed as the circuit under test may remain charged at a dangerous voltage. The instrument has an internal "safety resistor" which is connect to output terminal before end of test in order to discharge the parasite capacities of the installation
START
STOP
7. Press the START key.
The instrument will start the test. MAN Mode: The test will take 4 seconds (maximum). If you kee p
the START key pressed longer than 4 seconds the test go on until the key is released.
TMR mode: The test will take the time set. If you want to stop
the test when it's running, press the START STOP key again.
EN - 20
SIRIUS89N
A
6.2.2. Results of "MAN" mode
t the end of the test if the Insulation resistance is lower
than R and the instrument generated the
Nominal test Voltage,
the instrument emits a
double sound signal
indicating the positive
outcome of the test
(see Table2)
MAX
M 05.06.01
1.07
514V 15s
MAN 500V
FUNC VNOM
M
Insulation Resistance
Voltage during the Test
Duration of the Test
Test mode
Test voltage set
and displays one screen similar to the screen alongside.
In order to evaluate the test you must compare the result with the limits indicated in the Guidelines (see Table1).
The displayed result can be stored pressing the SAVE key twice (according to paragraph 9.1).
If the Insulation resistance is higher
than R
the instrument emits a double sound signal at the end of the test
indicating the positive
(see Table2),
MAX
outcome of the test
and displays one screen similar to the screen alongside.
M 05.06.01
> 999
523V 15s
MAN 500V
FUNC VNOM
M
Maximum resistance value which can be measured (999is displayed if a rated voltage of 500V was selected see Table2).
The symbol ">" means that the resistance value is higher than R
MAX
.
Test duration
The displayed result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 21
SIRIUS89N
A
6.2.3. Results of "TMR" mode
t the end of the test if the Insulation resistance is lower
than R and the instrument generated the
Nominal test Voltage,
the instrument emits a
double sound signal
indicating the positive
outcome of the test
(see Table2)
MAX
M 05.06.01
1.07
514V 15s
TMR 500V TIME:60s
FUNC VNOM
M
Insulation Resistance
Voltage during the Test
Duration of the Test
Test mode
Setting Time
and displays one screen similar to the screen alongside.
The displayed result can be stored pressing the SAVE key twice (according to paragraph 9.1).
If the Insulation resistance is higher
than R
the instrument emits a double sound signal at the end of the test
indicating the positive
(see Table2),
MAX
outcome of the test
and displays one screen similar to the screen alongside.
M 05.06.01
> 999
523V 15s
TMR 500V TIME:60s
FUNC VNOM
M
Maximum resistance value which can be measured (999is displayed if a rated voltage of 500V was selected see Table2).
The symbol ">" means that the resistance value is higher than R
MAX
.
Test duration
The displayed result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 22
SIRIUS89N
6.2.4. "MAN" and "TIMER" mode faulty cases
If the instrument detect the External Power supply adapter connected to instrument will show the message displayed to side.
M 05.06.01
-.- -
---V 15s
o REMOVE POWER
MAN 500V!
FUNC VNOM
M
Disconnect the External Power Supply Adapter
If the instrument detect a Voltage between Input terminals higher than 15V, the instrument does not effect the test and displays the screen alongside for 5 seconds.
M 05.06.01
-.- -
---V 15s
o VOLT IN INPUT
MAN 500V
FUNC VNOM
M
ATTENTION: the test can't be executed. Check that the circuit is not energised.
These result can't be saved
If the instrument can't generate the Nominal Test Voltage it will emits a long acoustic signal and displays a screen similar to the screen alongside.
M 05.06.01
o 1.17
107V 15s
MAN 500V
FUNC VNOM
M
Insulation Resistance
ATTENTION: the test of resistance R
a voltage value lower than the
set rated voltage. Low
insulation case. This case occurs under low insulation conditions or in the presence of capacity on the installation.
ISO
Test Time
was effected at
The displayed result can be stored pressing the SAVE key twice (according to
paragraph 9.1).
EN - 23
SIRIUS89N
6.3. RCD: TEST ON "A" AND "AC" RCDS TYPE
The test is executed according to IEC61557-6, EN61008, EN61009, EN60947-2 B 4.2.4 and VDE 0413 part 6.
CAUTION
The automatic check of the RCD features causes the tripping of the RCD itself. Therefore check that all devices connected downstream the RCD under test are not damaged by power off. Possibly disconnect all the loads connected downstream the RCD as they could add additional leakage currents to the instrument ones and so making the test results void.
Turn the switch on RCD position:
The F1 key permits to select one of the following measuring mode (which can be shown cyclically when pressing the key):
Mode "AUTO" (the instrument effects the test automatically with a
leakage current equal to half, once and five times the value of the rated current set). Recommended test.
Mode "x ½" (the instrument effects the test with a leakage current equal
to half the value of the rated current set).
Mode "x 1" (the instrument effects the test with a leakage current equal
to once the value of the rated current set ).
Mode "x 2" (the instrument effects the test with a leakage current equal
to twice the value of the rated current set).
Mode "x 5" (the instrument effects the test with a leakage current equal
to five times the value of the rated current)
Mode " " (the instrument effects the test with a ramp growing leakage
current. Use this test to measure the tripping current).
Mode "R
" (the instrument effects the test with a leakage current
A
equal to half the value of the selected rated current and calculates the contact voltage as well as the Ra earth resistance).
N.B. The AUTO mode execute automatically test with phase 0° and 180°
RCD Type 0° Current Waveform 180° Current Waveform
AC type
A type
According to standard praxis it is recommended to effect RCD test both with phase 0° and with phase 180° even with no AUTO modes. If the RCD under test is A type (which means sensitive to both AC and unidirectional pulsing
EN - 24
SIRIUS89N
leakage currents) it is advisable to effect the test both with sine wave and unidirectional pulse current with phase 0° and 180°. The F2 key permits to select one of the following rated tripping currents of the RCD (which can be shown cyclically when pressing the key):
10mA. 30mA. 100mA. 300mA. 500mA.
The F3 key permits to select the RCD type (which can be shown cyclically when pressing the key):
 "b": general RCD AC type (sensitive to sine leakage current)  "l": general RCD A type (sensitive to pulsating leakage current)  "
bn
": selective RCD AC type (sensitive to sine leakage current)
"
ln
": selective RCD A type (sensitive to pulsating leakage current)
Note if the test is effected on general RCDs the symbol n is NOT displayed Note according to EN61008 the test on the selective RCDs requires an interval between
the tests of 60 seconds (30 seconds in case of tests at ½ I∆n). A timer is displayed indicating the waiting time for each step.
Example: Test with AUTO mode on a RCD with In=30mA.
a) the instrument effects the test at ½ I∆n 0°. The RCD must not
trip.
b) The instrument effects the test at ½ I∆n 180°. The RCD must
not trip. For a Selective RCD a 30 seconds timer starts before executing next test.
c) The instrument effects the test at I∆n 0°. If the RCD passed
the test, it must trip and the instrument shows the message "RESUME RCD". The operator shall resume the RCD. For a Selective RCD a 60 seconds timer starts before executing next test.
d) The instrument effects the test at I∆n 180°. Follow the same
procedure as described under c).
e) The instrument effects the test at 5I∆n 0°. Follow the same
procedure as described under c).
f) The instrument effects the test at 5I∆n 180°. Follow the same
procedure as described under c). The test is completed.
The F4 key permits to select one of the following limit values for the contact voltage (which can be shown cyclically when pressing the key):
50V (default) 25V.
CAUTION
Never disconnect the test leads when the message "MEASURING" is displayed.
EN - 25
SIRIUS89N
6.3.1. Tripping times for the general and selective RCDs
Table of tripping times for I
x1, I∆N x2, I∆N x5 and AUTO tests.
N
If the parameters set on the instrument comply with the type of RCD under test (and if the latter works properly) the test x1, x2, x5 SHALL cause the RCD tripping within the times shown in the following table:
RCD type
I
N
x 1 I
x 2 I
N
N
x 5
Description
General 0.3s 0.15s 0.04s Max tripping time in seconds
0.5s 0.20s 0.15s Max tripping time in seconds
Selective S
0.13s 0.05s 0.05s Minimum tripping time in seconds
* For rated values I∆N ≤ 30mA the test current at five times is 0.25A. For currents equal to ½
Table 3: Table of tripping times for tests with leakage currents I
I
the RCD shall not trip in any case.
N
x1, I
N
x5 and AUTO.
Table of tripping times for ramp tests " ".
This test is not used to be effected to compare the RCD tripping time at the tripping current, while the standards refer to the maximum tripping times in case the RCD is checked with a leakage current equal to the rated current. The limits value for the tripping current are indicated in the following Table:
RCD Type
A
AC
I
10mA
N
1,4 x I
I
1,4 x I∆N
N
I∆N
N
I
>
10mA
N
x2, I∆N
N
Table 4: Current limit value for "Ramp" Test
EN - 26
SIRIUS89N
6.3.2. Measurement procedure
1. Select the desired test parameter by means of the F1, F2, F3, F4 key.
2. Connect the Black, Green and Blue connectors of the three-terminal shuko cable or of the split cables to the corresponding input terminals of the instrument B1, B3, B4
Instrument connection for 230V single-phase
RCD check
Instrument connection for 400V + N (no PE) three-
phase RCD check
3. Connect the shuko plug or the Test leads to the System under test according with one of the picture above.
Instrument connection for 400V + N + PE three-
phase RCD check
Instrument connection for 400V + PE (no N)
three-phase RCD check
EN - 27
SIRIUS89N
6.3.2.1. Results of "x½" mode
START
STOP
4. Press the START key once to execute a test with 0° Current waveform.
or
Press the START key once to execute a test with 180° Current waveform.
CAUTION
Never disconnect the test leads when the message "MEASURING" is displayed.
If the RCD does NOT trip the instrument emits a double sound signal
indicating the positive
outcome of the test and
displays the screen alongside.
Working mode
RCD 05.06.01
> 999
FRQ=50.0Hz Ut= 1V VP-N=231V VP-PE=231V
x1/2 30mA 50V! FUNC IdN RCD UL
RCD nominal current
ms
RCD OK
The symbol ">" means that the RCD did not trip.
Value of contact voltage U detected referred to the rated value of the RCD current set.
OK: RCD passed the test.
Contact Voltage Limit value RCD type
t
The test can be stored pressing the SAVE key twice (according to paragraph 9.1).
6.3.2.2. Results of "x1, x2, x5" mode
STOP
4. Press the START key once to execute a test with Current waveform.
START
or
Press the START key once to execute a test with 180° Current waveform.
CAUTION
Never disconnect the test leads when the message "MEASURING" is displayed.
If the tripping time is within the limits reported in Table 3, the instrument emits a
double sound signal indicating the positive
outcome of the test and
displays the screen alongside.
Working mode
RCD 05.06.01
49
FRQ=50.0Hz Ut= 2V VP-N=231V VP-PE=231V
x1 30mA 50V! FUNC IdN RCD UL
RCD nominal current
ms
RCD OK
Tripping time (expressed in milliseconds).
Value of contact voltage U detected referred to the rated value of the RCD current set.
OK: RCD passed the test.
Contact Voltage Limit value RCD type
t
The test can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 28
SIRIUS89N
6.3.2.3. Results of "AUTO" mode
START
4. Press the START key once to execute the test. The instrument carries out the
STOP
following six tests with different values of rated current:
1/2In with 0° current waveform (the RCD shall not trip).  1/2In with 180° current waveform (the RCD shall not trip).  In with 0° current waveform (the RCD trips, message "RESUME RCD").  In with 180° current waveform (the RCD trips, message "RESUME RCD").  5In with 0° current waveform (the RCD trips, message "RESUME RCD").  5In with 180° current waveform (the RCD trips, end of the test).
The test is good if all values of tripping times are within the limits reported in Table 3.
CAUTION
Never disconnect the test leads when the message "MEASURING" is displayed.
At the end of the test if
all six tests resulted to
be positive, the
instrument displays the screen alongside relative to the last measurement effected.
RCD 05.06.01
0° 180° x1/2 >999ms >999ms
x1 55ms 65ms x5 20ms 30ms FRQ=50.0Hz Ut= 1V
VP-N=231V VP-PE=231V
AUTO 30mA 50V! FUNC IdN RCD UL
RCD OK
Tripping time (expressed in milliseconds).
Value of contact voltage U detected referred to the rated value of the RCD current set.
OK: RCD passed the test.
t
Contact Voltage Limit value
Working mode
RCD nominal current
RCD type
The test can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 29
SIRIUS89N
A
6.3.2.4. Results of "RAMP " mode
START
4. Press the START key once to execute a test with Current waveform.
STOP
or
Press the START key once to execute a test with 180° Current waveform.
The instrument generates a leakage current growing step by step for a given time interval.
CAUTION
Never disconnect the test leads when the message "MEASURING" is displayed.
t the end of the test if the RCD tripping current is lower than I∆n (Type AC) or 1.4I∆n (Type A with I∆n >10mA) or 2I∆n (Type A with I∆n 10mA), the instrument emits a double sound signal indicating the positive outcome of the test and displays the screen alongside.
RCD 05.06.01
27
FRQ=50.0Hz Ut= 1V VP-N=231V VP-PE=231V
30mA 50V!
FUNC IdN RCD UL
RCD OK
mA
35ms
Tripping Current
Tripping time (expressed in milliseconds).
Value of contact voltage U detected referred to the rated value of the RCD current set.
OK: RCD passed the test.
Contact Voltage Limit value
t
Working mode
RCD nominal current
RCD type
The test can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 30
SIRIUS89N
6.3.2.5. Results of " RA" mode
START
4. Press the START key once: the instrument carries out the test.
STOP
CAUTION
Never disconnect the test leads when the message "MEASURING" is displayed.
The RCD must NOT trip and the instrument emits a double sound
signal indicating the
positive outcome of
the test and displays the screen alongside.
RCD 05.06.01
12
FRQ=50.0Hz Ut= 1V VP-N=231V VP-PE=231V
RA 30mA 50V! FUNC
IdN
Ut OK
RCD UL
Global Earth Resistance Value).
Value of contact voltage U detected referred to the rated value of the RCD current set.
OK:Contact Voltage Not Dangerous.
t
Contact Voltage Limit value
Working mode
RCD nominal current
RCD type
The test can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 31
SIRIUS89N
6.3.3. RCD Faulty cases
6.3.3.1. Connection troubles
If the instrument detect the External Power supply adapter connected to instrument will show the message displayed to side.
RCD 05.06.01
- - -
FRQ=50.0Hz Ut= ---V VP-N=230V Vp-PE=230V
o REMOVE POWER
x1 30mA 50V! FUNC IdN RCD UL
ms
Disconnect the External Power Supply Adapter
Should the instrument detect that the phase and/or neutral cables are not connected to an installation, screen alongside is displayed when pressing START.
RCD 05.06.01
- - -
FRQ=50.0Hz Ut= ---V VP-N= 0V Vp-PE= 0V
o LOW VOLTAGE
x1 30mA 50V! FUNC IdN RCD UL
ms
NO VOLTAGE Detected
Should the instrument detect a voltage between phase and neutral higher of 250V, for example in case the blue cable is connected to an installation phase conductor of a 400V three-phase system, the screen alongside is displayed.
RCD 05.06.01
- - -
FRQ=50.0Hz Ut= ---V VP-N=401V VP-PE= 230V
o HIGH VOLTAGE
x1 30mA 50V! FUNC IdN RCD UL
ms
HIGH VOLTAGE Detected
EN - 32
SIRIUS89N
This screen is displayed when the phase conductor has been exchanged with the neutral one. The instrument does not effect the test. Reverse the shuko plug or exchange the black cable with the blue one. Repeat the test
RCD 05.06.01
- - -
FRQ=50.0Hz Ut= ---V VP-N=231V VP-PE= 0V
o CHANGE P-N
x1 30mA 50V! FUNC IdN RCD UL
ms
Phase and Neutral Conductors are exchanged.
This screen is displayed when the phase conductor has been exchanged with the Protection Conductors. The instrument does not effect the test. Reverse the phase to earth connection in the plug or exchange the black cable with the green one
RCD 05.06.01
- - -
FRQ=50.0Hz Ut= ---V VP-N= 2V VP-PE= 230V
o CHANGE P-PE
x1 30mA 50V! FUNC IdN RCD UL
ms
Phase and Protection Conductors are exchanged.
This screen is displayed when in a 230V Phase to Phase System the blue conductor was reversed with respect to the green one. The instrument does not effect the test. Reverse the blue and green conductors.
RCD 05.06.01
- - -
FRQ=50.0Hz Ut= ---V VP-N=130V VP-PE= 227V
o CHANGE N-PE
x1 30mA 50V! FUNC IdN RCD UL
ms
Neutral and Protection Conductors are exchanged.
EN - 33
SIRIUS89N
If a contact voltage U
higher than the selected limit (U
detected, the instrument interrupts the test and emits a long sound signal at the end of the test and displays the screen alongside.
) is
L
RCD 05.06.01
t
- - -
FRQ=50.0Hz Ut= ---V VP-N=234V VP-PE= 234V
o UT DANGEROUS
x1 30mA 50V! FUNC IdN RCD UL
ms
The instrument doesn't detect an efficient protection circuit.
If the instrument detects that the earth cable (green) is not connected, the screen alongside is displayed for 5 seconds then the initial display is back. Check the connections of PE conductor under test.
RCD 05.06.01
- - -
FRQ =50.0Hz Ut= ---V VP-N=234V VP-PE= 34V
o NO PE
x1 30mA 50V! FUNC IdN RCD UL
ms
Message “NO PE”: the instrument doesn't detect an efficient protection circuit.
If the instrument gets overheated, tests cannot be carried out and the message alongside is displayed. Wait until the initial screen is back in order to proceed with measurements.
RCD 05.06.01
- - -
FRQ=50.0Hz Ut= ---V VP-N=231V VP-PE= 230V
x1 500mA 50V! FUNC IdN RCD UL
ms
o HOT
Message “hot”: the instrument got overheated.
THE PREVIOUS RESULTS CANNOT BE SAVED
.
EN - 34
SIRIUS89N
!
Using the R if a contact voltage U
higher than the selected limit (U
detected the instrument emits a long sound signal at the end of the test and displays the screen alongside.
function,
A
) is
L
RCD 05.06.01
t
1800
FRQ=50.0Hz Ut= 54V VP-N=234V VP-PE= 34V
R
FUNC IdN RCD UL
o UT NOT OK
A
30mA 50V
The instrument doesn't detect an efficient protection circuit.
The test can be stored pressing the SAVE key twice (according to paragraph 9.1).
6.3.3.2. RCD tripping "faulty cases"
If the RCD trips during the preliminary test performed before the test (independently from the working mode) the instrument displays the screen alongside.
RCD 05.06.01
- - -
FRQ=50.0Hz Ut= ---V VP-N=231V VP-PE= 230V
x1 30mA 50V! FUNC IdN RCD UL
o
RCD TRIPS
ms
The RCD trips too early. Check if the RCD Nominal Current is correct. Some leakage currents may be present in the installation. Disconnect all the loads connected downstream the RCD.
THE PREVIOUS RESULTS CAN'T BE SAVED.
In the tests MAN x1, x2, x5 and AUTO (during x1 and x5 tests), if the RCD trips to separate the circuit within a time not complying with the limits reported in Table 3, the instrument emits a long acoustic signal at the
RCD 05.06.01
o
487
FRQ=50.0Hz Ut= 1V VP-N=231V VP-PE= 230V
o
x1 30mA 50V!
TIME NOT OK
ms
end of the test and displays the values alongside.
FUNC IdN RCD UL
RCD tripping time.
ATTENTION: the tripping time is higher than the standard limit.
The test can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 35
SIRIUS89N
If the RCD tripping time is higher than the instrument’s measuring limits, the instrument emits a long sound signal at the end of the test and displays the values alongside.
RCD 05.06.01
o
>999
FRQ=50.0Hz Ut= 1V VP-N=231V VP-PE= 230V
o
x1 30mA 50V! FUNC IdN RCD UL
TIME NOT OK
ms
RCD tripping time is bigger than the maximum measurable time (it depends on type of test, see following table).
ATTENTION: the tripping time is higher than the standard limit.
The maximum duration depends on the test type:
Test type General RCD Selective RCD
MAN x1 test 999ms 999ms MAN x2 test 200ms 250ms MAN x5 test 50ms 160ms
" " test 300ms
During the ramp if the RCD tripping time is higher than the limit, the instrument emits a long sound signal at the end of the test and displays the values alongside.
test
RCD 05.06.01
27
Freq=50.0Hz Ut= 1V VP-N=231V Vp-PE=230V
o TIME NOT OK
30mA 50V!
FUNC IdN
mA
>300ms
RCD
UL
Tripping Current.
Tripping Time exceeding limit value
During the ramp if the RCD tripping current is higher than I (Type AC) or 1.4 I∆n (Type A with I∆n>10mA) or 2 I∆n (Type A with I∆n≤10mA) , the instrument emits a long sound signal at the end of the test and displays the values alongside.
test
RCD 05.06.01
n
o > 42
>300ms
FRQ=50.0Hz Ut= 1V VP-N=231V VP-PE=230V
o CURRENT NOT OK
30mA 50V!
FUNC IdN RCD UL
mA
Maximum current generated by the instrument during the test for general RCDs (the value indicated is referred to an AC type 30mA RCD, in this case the maximum current supplied is equal to
1.4xI
ATTENTION: the RCD tripping current is higher than the Nominal
Value (I in the example
N
=30mA was set
N
).
The previous results can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 36
SIRIUS89N
6.4. LOOP : MEASUREMENT OF LINE IMPEDANCE, FAULT LOOP IMPEDANCE, PROSPECTIVE SHORT CIRCUIT CURRENT CALCULATION AND PHASE SEQUENCE INDICATOR
Turn the switch on LOOP position.
The F1 key permits to select one of the following measuring modes: Mode "P-N" (the instrument measures the impedance between the phase
and neutral conductors and calculates the phase to neutral prospective short circuit current).
Mode "P-P" (the instrument measures the impedance between two
phase conductors and calculates the phase to phase prospective short circuit current).
Mode "P-PE" (the instrument measures the impedance between the
phase and protective conductors and calculates the phase to earth prospective short circuit current).
Mode "RA" (the instrument measures the impedance between the
phase and protective conductors with a test current of 15mA in order to avoid RCD tripping and calculates the phase to earth prospective short circuit current).
Mode " " (the instrument detect the Phase Sequence in a three-phase
system).
CAUTION
Never disconnect the test leads when the message "MEASURING" is displayed.
6.4.1. High resolution Impedance measurement (0.1m)
The instrument is connectable to an external optional accessory (IMP57) useful for high resolution impedance measurement close to a power transformer. The high Resolution impedance measurement is available inside the LOOP P-P, P-N, P-PE modes by mean Un/In key. If You enable the High Resolution Impedance Measurement mode without connecting IMP57 the following screen will be displayed (e.g: Loop P-N):
LOOP 29.01.04
V1-2= 0V FRQ= 0.0Hz
P-N Z2 FUNZ ZSTD ICAL RMT
Z=-----
R=----- X=-----
IkSTD=----A
NO IMP57
For further details regarding IMP57 use and technical characteristics please refer to IMP57 user's manual or HT web site www. htitalia.com.
EN - 37
SIRIUS89N
A
6.4.2. Measurement procedure and results of "P-N" mode
1. Select P-N mode by means of the F1 key.
2. Connect the Black, Green and Blue connectors of the three-terminal shuko cable or of the split cables to the corresponding input terminals of the instrument B1, B3, B4
Instrument connection for P-N test in a 230V
single-phase System
Instrument connection for P-N in a 400V three-
phase system
3. Connect the shuko plug into a 230V 50Hz socket or the crocodiles to the conductors of the three-phase system (see previous pictures).
4. If possible disconnect all low impedance loads downstream the point at which the measurement is to be taken, as such impedances would be in parallel with
START
STOP
the line impedance to be measured.
5. Press the START key. The instrument starts the test.
CAUTION
The measurement in a 230V System make flow a test current of 6A approx. This may cause the tripping of magnetic protection switch with nominal value lower than 10A. If necessary effect the test upstream the switch. Never disconnect the test leads when the message "MEASURING" is displayed.
t the end of the test
the instrument emits a
double sound signal
indicating that the test is correctly
terminated and
displays the values alongside.
Working mode
LOOP 05.06.01
1.07
FRQ=50.0HZ VP-N=231V VP-PE=231V
P-N!
215A
FUNC Z2
Value of phase to neutral line impedance expressed in Ω.
Value of the phase to neutral prospective short circuit current expressed in Ampere calculated according to the following formula.
U
Formula for calculation of prospective short circuit current:
I =
where UN = Nominal phase to neutral voltage = 127 if V
230 if 150V< V
CC
meas
Z
PN
150
N
meas
250
This result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 38
SIRIUS89N
A
6.4.3. Measurement procedure and results of "P-P" mode
1. Select P-P mode by means of the F1 key.
2. Connect the Black, Green and Blue connectors of the three-terminal shuko cable or of the split cables to the corresponding input terminals of the instrument B1, B3, B4
Instrument connection for P-P test in a 400V three-phase system
3. Connect the shuko plug into a 230V 50Hz socket or the crocodiles to the conductors of the three-phase system (see previous pictures).
4. If possible disconnect all low impedance loads downstream the point at which the measurement is to be taken, as such impedance would be in parallel with
START
STOP
the line impedance to be measured.
5. Press the START key. The instrument starts the test.
CAUTION
The P-P measurement in a 400V system make flow a test current of 11.5A approx. This may cause the tripping of magnetic protection switch with nominal value lower than 10A. If necessary effect the test upstream the switch. Never disconnect the test leads when the message "MEASURING" is displayed.
t the end of the test
the instrument emits a
double sound signal
indicating that the test is correctly
terminated and
displays the values alongside.
Working mode
LOOP 05.06.01
0.57
FRQ=50.0HZ VP-P=402V VP-PE=231V
P-P! FUNC Z2
701A
Value of phase to phase impedance expressed in Ω.
Value of the phase to phase prospective short circuit current expressed in Ampere calculated according to the following formula.
Formula for calculation of prospective short circuit current:
U
N
Z
PP
150
meas
230 if 150V< V 400 if V
meas
> 260
meas
260
where U
= phase to phase voltage 127 if V
N
I =
CC
This result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 39
SIRIUS89N
6.4.4. Measurement procedure and results of "P-PE" mode
1. Select P-PE mode by means of the F1 key.
2. Connect the Black, Green and Blue connectors of the three-terminal shuko cable or of the split cables to the corresponding input terminals of the instrument B1, B3, B4.
Instrument connection for P-PE test in a 230V
single-phase System
Instrument connection for P-PE in a 400V three-phase system without Neutral conductor
3. Connect the shuko plug into a 230V 50Hz socket or the crocodiles to the conductors of the three-phase system (see previous pictures).
4. The key F4 permits to select one of the following limit values for the contact voltage (which can be shown cyclically when pressing the key):
50V (default).
START
STOP
5. Press the START key once to execute a test injecting a current in phase with
25V.
positive half wave of the voltage.
Press the START key twice to execute a test injecting a current in phase with negative half wave of the voltage.
The P-PE measurement in a 230V System make flow a test current of 6A approx. This may cause the tripping of magnetic protection switch with nominal value lower than 10A and will cause the tripping of RCD device. If necessary effect the test upstream the switch or RCD.
Never disconnect the test leads when the message "MEASURING" is displayed.
Instrument connection for P-PE in a 400V three-
phase system
CAUTION
CAUTION
EN - 40
SIRIUS89N
A
t the end of the test
the instrument emits a
double sound signal
indicating that the test is correctly
terminated and
displays the values alongside.
LOOP 05.06.01
1.07
Freq=50.0HZ VP-N=231V Vp-PE=231V
P-PE! FUNC Z2
215A
Value of phase to Earth Resistance expressed in Ω.
Value of the phase to Earth prospective short circuit current expressed in Ampere calculated according to the following formula.
Working mode
Formula for calculation of prospective short circuit current:
U
I =
CC
where UN = Nominal phase to neutral voltage = 127 if V
N
Z
PE
150
meas
230 if 150V< V
meas
250
This result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 41
SIRIUS89N
6.4.5. Measurement procedure and results of " RA" mode
1. Select RA mode by means of the F1 key.
2. Connect the Black, Green and Blue connectors of the three-terminal shuko cable or of the split cables to the corresponding input terminals of the instrument B1, B3, B4.
Instrument connection for P-PE test in a 230V
single-phase System
Instrument connection for P-PE in a 400V three-phase system without neutral
3. Connect the shuko plug into a 230V 50Hz socket or the crocodiles to the conductors of the three-phase system (see previous pictures).
4. If possible disconnect all low impedance loads downstream the point at which the measurement is to be taken, as such impedance would be in parallel with the line impedance to be measured.
5. The key F4 permits to select one of the following limit values for the contact voltage (which can be shown cyclically when pressing the key):
50V (default) 25V.
START
6. Press the START key once to execute a test with “” test current.
STOP
Press the START key twice to execute a test with “180°” test current.
The RA measurement make flow a test current of 15mA. This may cause the tripping of 10mA. If necessary effect the test upstream the switch. Never disconnect the test leads when the message "MEASURING" is displayed.
Instrument connection for P-PE in a 400V three-
phase system
or
CAUTION
EN - 42
SIRIUS89N
A
A
A
t the end of the test
the instrument emits a
double sound signal
indicating that the test is correctly
terminated and
displays the values alongside.
Working mode
LOOP 05.06.01
2
Freq=50.0HZ VP-N=231V Vp-PE=231V
R
!
A
FUNC
115A
Formula for calculation of prospective short circuit current: where U
= Nominal phase to neutral voltage = 127 if V
N
230 if 150V< V
Value of phase to Earth Resistance expressed in Ω.
Effective value of the phase to Earth prospective short circuit current expressed in
mpere calculated according
to the following formula..
U
I =
CC
meas
Z
150
N
PE
meas
250
This result can be stored pressing the SAVE key twice (according to paragraph 9.1).
6.4.6. Measurement procedure and results of " " mode
1. Select mode by means of the F1 key.
2. Connect the Black, Red and Green connectors of the split cables to the corresponding input terminals of the instrument B1, B2, B3.
Instrument connection for Phase Sequence Detection in a 400V three-phase system
START
3. Press the START key to execute a test.
STOP
t the end of the test
the instrument emits a
double sound signal
indicating that the test is correctly
terminated and
displays the values alongside.
LOOP 05.06.01
RST
FRQ=50.0HZ VR-S=391V VS-T=401V VT-R=399V
!
FUNC
OK
Phase Sequence OK
Value of the Phase to Phase Voltage
Working mode
This result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 43
SIRIUS89N
!
!
6.4.7. LOOP Faulty Cases
If the instrument detect the External Power supply adapter connected to instrument will show the message displayed to side.
LOOP 05.06.01
- - -
FRQ=50.0HZ VP-N=231V VP-PE=230V
o REMOVE POWER
P-N! FUNC Z2
---A
Disconnect the External Power Supply Adapter
Should the instrument detect that the phase and/or neutral cables are not connected to an installation, screen alongside is displayed when pressing START.
LOOP 05.06.01
- - -
FRQ=50.0HZ VP-N= 1V VP-PE= 0V
P-PE 50V FUNC Z2 UL
NO VOLTAGE
---A
NO VOLTAGE Detected
Should the instrument detect a voltage between phase and neutral higher of 250V, for example in case the blue cable is connected to an installation phase conductor of a 400V three-phase system, the screen alongside is displayed.
LOOP 05.06.01
- - -
Freq=50.0HZ VP-N=401V VP-PE=230V
HIGH VOLTAGE
P-PE 50V FUNC Z2 UL
---A
HIGH VOLTAGE Detected
EN - 44
SIRIUS89N
!
This screen is displayed when the phase conductor has been exchanged with the neutral one. The instrument does not effect the test. Reverse the schuko plug or exchange the black cable with the blue one.
LOOP 05.06.01
- - -
FRQ=50.0HZ VP-N=231V VP-PE= 0V
CHANGE P-N
P-PE 50V FUNC Z2 UL
---A
Phase and Neutral Conductors are exchanged.
This screen is displayed when the phase conductor has been exchanged with the Protection Conductors. The instrument does not effect the test. Reverse the phase to earth connection in the plug or exchange the black cable with the green one
LOOP 05.06.01
- - -
FRQ=50.0HZ VP-N= 1V VP-PE= 230V
CHANGE P-PE
P-N! FUNC Z2
---A
Phase and Protection Conductors are exchanged.
This screen is displayed when in a 230V Phase to Phase System the blue conductor was reversed with respect to the green one. The instrument does not effect the test. Reverse the blue and green conductors.
LOOP 05.06.01
- - -
FRQ=50.0HZ VP-N=131V VP-PE= 227V
CHANGE N-PE
P-N! FUNC Z2
---A
Neutral and Protection Conductors are exchanged.
EN - 45
SIRIUS89N
!
!
If a contact voltage U
higher than the selected limit (U
detected the instrument interrupts the test and emits a long sound signal at the end of the test and displays the screen alongside.
) is
L
LOOP 05.06.01
t
- - -
FRQ=50.0HZ VP-N= 1V Vp-PE= 0V
R
FUNC UL
Ut DANGEROUS
50V
A
---A
Message “Ut”: the instrument doesn't detect an efficient protection circuit.
Should the instrument detect an Earth Resistance extremely high to consider no presence of PE conductor or Earth installation ones, it shows the message on the right picture. Please check the efficiency of PE conductor and Earth installation.
LOOP 05.06.01
- - -
FRQ=50.0HZ VP-N=231V Vp-PE= 40V
P-PE 50V FUNC Z2 UL
---A
NO PE
Message “NO PE”: the instrument doesn't detect an efficient protection circuit.
If the instrument gets overheated, tests cannot be carried out and the message alongside is displayed. Wait until the initial screen is back in order to proceed with measurements.
LOOP 05.06.01
- - -
FRQ=50.0HZ VP-N=231V Vp-PE= 40V
o
P-P! FUNC Z2
---A
HOT o
Message “hot”: the instrument got overheated.
EN - 46
SIRIUS89N
Using the " " mode, if a Phase to Phase voltage is lowert than 100V, the instrument displays the screen indicated on side.
LOOP 05.06.01
- - -
FRQ =50.0HZ VR-S=391V VS-T= 0 V VT-R= 0V
o LOW VOLTAGE T
PHASE ROTATION
FUNC
!
Phase "T2 Voltage is lower than 100V
Using the " " mode, if the instrument detect two phases connected together displays the screen indicated on side.
LOOP 05.06.01
- - -
FRQ =50.0HZ VR-S=407V VS-T= 0 V VT-R=407V
o PHASE DOUBLED
PHASE ROTATION
FUNC
!
Two phases are connected together.
THE PREVIOUS RESULTS CANNOT BE SAVED.
EN - 47
SIRIUS89N
!
In the mode P-P, P-N mode the instrument carries out the test and detects a resistance to
higher than 199.9
screen alongside is displayed.
, the
LOOP 05.06.01
>199.9
FRQ =50.0HZ VP-N= 1V Vp-PE= 0V
P-N! FUNC Z2
---A
Message ">199.9" means that the Resistance measured is higher than the maximum measurable
This result can be stored pressing the SAVE key twice (according to paragraph 9.1).
In the mode P-PE, RA mode the instrument carries out the test and detects a resistance to
higher than 1999
screen alongside is displayed.
, the
LOOP 05.06.01
>1999
FRQ =50.0HZ VP-N= 1V Vp-PE= 0V
RA 50V FUNC UL
---A
Message ">1999" means that the Resistance measured is higher than the maximum measurable
This result can be stored pressing the SAVE key twice (according to paragraph 9.1).
In the voltage of one or more phase is too low, one or more phase has a low voltage the instrument will show a screen similar to the along side displayed.
mode, if the
LOOP 05.06.01
RTS
FRQ =50.0HZ VR-S=391V VS-T= 0V VT-R= 0V
NOT CORRECT
!
FUNC
Phase Sequence not correct
Message “LOW Voltage Phase T”: means that Phase T has a low voltage value. Similar message for Phase R and S.
This result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 48
SIRIUS89N
6.5. EARTH: SOIL RESISTANCE AND RESISTIVITY MEASUREMENTS
Turn the switch on EARTH position.
The F1 key permits to select one of the following measuring modes (which can be shown cyclically when pressing the key):
Mode "2-W" (the instrument measures the resistance between 2 points). Mode "3-W" (the instrument measures the resistance using two auxiliary
earth rods).
Mode "ρ" (the instrument measures the ground resistivity).
At the end of each test, instrument shows automatically the medium value of Earth Resistance or Earth Resistivity calculated on the base of any measured values until that moment.
F2 key clear medium value of Earth Resistance or Earth Resistivity and the counter of executed Resistance measures.
Never disconnect the test leads from the circuit under test when the message "MEASURING" is displayed
CAUTION
EN - 49
SIRIUS89N
A
A
6.5.1. Measurement procedure and results of "2-W"and "3-W" mode
1. Select "2-W" or "3-W" Earth measurement mode by means of the F1 key.
2. Connect the Black, Red, Green and Blue cables to the corresponding input terminals of the instrument B1, B2, B3, B4 (see possible connections in the following pictures).
Connection for 3 point Earth resistance
measurement
Connection for 2 point Earth Resistance
measurement
Connection for measuring the resistance between an extraneous conductive part and the earth system
START
STOP
3. Press the START key. The instrument starts the test.
t the end of the test
the instrument emits a
double sound signal
indicating that the test is correctly
terminated and
displays the values alongside.
"3 Point" Working mode
EARTH 05.06.01
0.77
Test:04 RAVG=0.74
3-W! FUNC CLR
Vd= 1V
Earth Resistance value expressed in Ω.
Voltage value of electrical noise
Number of Test
verage Value of Earth Resistance calculated over the Number of Test displayed.
4. The instrument will show automatically the Average value of the Earth resistance calculate over the tests performed. Press F2 to RESET this value and the number of Test.
This result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 50
SIRIUS89N
A
A
6.5.2. Measurement procedure and results of "ρ" mode
1. Select ρ measurement mode by means of the F1 key.
2. Select the distance d between the earth rods by means the F3 and F4 keys.
3. Connect the 4 Blac k, Red, Green and Blue connectors of the single cables in the corresponding input terminals of the instrument B1, B2, B3, B4.
Instrument connection for Earth resistivity measurement
START
4. Press the START key. The instrument starts the test.
STOP
t the end of the test
the instrument emits a
double sound signal
indicating that the test is correctly
terminated and
displays the values alongside.
"ρ" Working mode
EARTH 05.06.01
1.77
Test:04 ρAVG=0.74
ρ DIST= 2m! FUNC CLR
m
Vd= 1V
Earth Resistivity value expressed in Ωm.
Voltage value of electrical noise
Number of Test
verage Value of Earth Resistivity calculated over the Number of Test displayed.
5. The instrument will show automatically the Average value of the Earth Resistivity calculate over the tests performed. Press F2 to RESET this value and the number of Test.
This result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 51
SIRIUS89N
6.5.3. "2-W", "3-W" and "ρ" faulty cases
If the instrument detect the External Power supply adapter connected to instrument will show the message displayed to side.
EARTH 05.06.01
- - -
Test:04 RAVG=0.74
3-W! FUNC CLR
Vd= ---V
o REMOVE POWER
Disconnect the External Power Supply Adapter
If the Instrument detect a voltage values higher than 5V the instrument will shows the screen displayed alongside.
EARTH 05.06.01
o - - -
Test:04 RAVG=0.74
3-W! FUNC CLR
Vd= 230V
o VOLT IN INPUT
Warning symbol: Voltage noise in input.
The message "Rc high" indicates that the instrument can't flow the minimum current necessary for measurement. Check that the terminals are correctly connected and the Auxiliary earth rod connected to B4 (blue conductor) has not been inserted in a pebbly or poor conductive ground. If necessary pour some water around the rod.
EARTH 05.06.01
- - -
Test:04 RAVG=0.74
3-W! FUNC CLR
Vd= 1V
o Rc HIGH
Message "Rc": Check Auxiliary Earth rods.
EN - 52
SIRIUS89N
The message "Rp high" indicates that the
instrument can't measure correctly the Voltage from Auxiliary Earth rod. Check that the terminals are correctly connected and the Auxiliary earth rod connected to B2 (red conductor) has not been inserted in a pebbly or poor conductive ground.
If necessary pour some
water around the rod.
EARTH 05.06.01
- - -
Test:04 RAVG=0.74
o Rp HIGH
3-W! FUNC CLR
Vd= 1V
Message "Rp": Check Auxiliary Earth rods.
The message "Rp and Rc high" indicates that
the instrument can't measure correctly the Voltage from Auxiliary Earth rod and can't flowe the minimum current for measurement. Check that the terminals are correctly connected. Check if the Auxiliary earth rod connected to B2 (red conductor) and B4(blue conductor) has not been inserted in a pebbly or poor
EARTH 05.06.01
- - -
Test:04 RAVG=0.74
o Rp and Rc HIGH
3-W! FUNC CLR
Vd= 1V
Message "Rp and Rc HIGH": Check Auxiliary Earth rods.
conductive ground. If
necessary pour some
water around the rod.
THE PREVIOUS RESULTS CANNOT BE SAVED.
EN - 53
SIRIUS89N
If the Instrument detect a Resistance value higher than 1999Ω, the instrument will show the screen alongside.
EARTH 05.06.01
> 1999
Test:04 RAVG=0.74
3-W! FUNC CLR
Vd= 1V
Message ">1999" means that the resistance value is higher than the maximum measurable.
This result can be stored pressing the SAVE key twice (according to paragraph 9.1).
If the Instrument detect a Resistivity value higher than 1999k the instrument will show the screen alongside.
m,
EARTH 05.06.01
> 1999
Test:04
ρAVG=0.74km
ρ DIST=5m! FUNC CLR
km
Vd= 1V
Message ">1999" means that the resistivity value is higher than the maximum measurable.
This result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 54
SIRIUS89N
7. AUX: MEASUREMENT WITH EXTERNAL PROBES
Turn the switch on AUX position.
The F4 function key assume the following operations: Pressing this key the instrument shows one of the following working
mode displayed circularly:
Environmental Parameter & Leakage Current (mA, °C, °F, HR%, m /s, mV, Lux) Sound Level Measurement
The "Environmental Parameter & Leakage Current" mode allows the following operations:
display in real time the values coming from external probes or clamps.  memorize the values displayed (pressing SAVE key).  record (pressing the START key after a proper setting) an Input signal
coming from an External Probe or clamp among the possibilities illustrated above. It will be possible to analyze the recorded data ONLY by transferring them to a PC.
record (pressing the START key) an Input signal coming from an
External Probe or clamp, using “Typical Configurations” with pre­programmed parameters inside meter (see paragraph 7.2.2).It will be
possible to analyze the recorded data ONLY by transferring them to
a PC.
CAUTION
Please focus the difference between memorize and record: the former means that the instrument store in the memory only the actual values displayed while the latter means that you want to store the course of the input signals during a recording time (typically long).
The "Sound Level Measurement" mode allows the following operations:
display in real time the values coming from external probe of sound
pressure level (Type 1).
Calculate at the end of measurement the Equivalent Level of noise
LeqT.
CAUTION
The instrument has the only I1 active auxiliary input, so please pay attention while connecting the external probes. The instrument does not recognize any auxiliary probe which is connected to the I2 or I3 inputs.
EN - 55
SIRIUS89N
7.1. ENVIRONMENTAL PARAMETER AND LEAKAGE CURRENT: REAL TIME MEASUREMENT
This working mode allows to execute real time measurement and recording of Environmental Parameter and Leakage current
1. Press this key to access to "AUX" mode.
2. Pressing this function key you will change the measuring unit of the instrument's input. The following possibilities will be displayed circularly:
- - - (Disabled instrument's input) mA (Leakage current) °C (Celsius Temperature) °F (Fahrenheit Temperature) HR% (Relative humidity) m/s (Air Speed) mV (Voltage) LUX (20) (Illuminance: Full Scale 20Lux) LUX (2k) (Illuminance: Full Scale 2kLux) LUX (20k) (Illuminance: Full Scale 20kLux)
3. Connect the External probe or clamp to the input I1 of the instrument.
4. Check if the selector on the probe or clamp is adjusted according to the range set in the instrument. Both range must be set to the same value.
Example of External probes connections
CAUTION
The OFF position of few probes send to probe's output the Battery voltage (approx 9V which is over the expect full scale). This could influence the measurement of the other instrument's inputs. So NEVER left connect to instrument probes with the Selector placed in OFF position.
EN - 56
SIRIUS89N
Indirect Leakage Current measurement in a
single phase system
Indirect Leakage Current measurement in a three
phase system
Direct Leakage Current measurement in a single
phase system
Direct Leakage Current measurement in a three
phase system
5. The Instrument shows in real time the values present in inputs.
Example of screen.
05.06.01 11:43:04
In1= 23°C
AUX
Example of Input Signal 1
In1 PG+
6. Press this key to enable/disable the HOLD function (updating interruption of the displayed data). When the HOLD function is enabled, the word HOLD is displayed. This key is disabled during a recording. It's not possible to run a recording if this function is enabled.
7. The displayed result can be stored pressing the SAVE key twice (according to paragraph 9.1).
EN - 57
SIRIUS89N
7.2. ENVIRONMENTAL PARAMETER AND LEAKAGE CURRENT: RECORDING
Before starting a recording we recommend You to check that real time values are correct. To this purpose follow the measurement procedure described in paragraph 7.1.
In addition it's fundamental that Instrument settings correspond to the accessories is use. For this we recommend you to check instrument's setting before execute an AUX recording. To this purpose please check the RECORDER CONFIG settings.
MENU: to enter in the MENU mode and change the instrument settings. It’s
not possible to enter the configuration MENU during a recording or an energy measurement.
START/STOP: to record the selected parameters according to the instrument’s
settings (see chapter 10).
7.2.1. AUX Basic setting: RECORDER CONFIG
Place the rotary switch in the AUX position, press the MENU key, using the F1/F2 keys select the RECORDER CONFIG item and press the ENTER Key.
It’s not possible to enter the MENU during a recording or a Real Time Energy measurement.
CAUTION
MENU GENERAL
SAFETY TEST MEMORY ANALYZER MEMORY RESET
ANALYZER CONFIG
RECORDER CONFIG
CONTRAST DATE&TIME LANGUAGE
Inside RECORDER CONFIG mode it’s possible to execute the following operations:
Setting of Start/Stop recording mode (AUTO or MANUAL) and the date/hour of
Start/Stop (in AUTO mode).
Setting of Integration Period (see paragraph 16.13.1) selectable from 5 sec to 60 min
Press ENTER to confirm each settings or ESC to escape from modifies and exit from RECORDER CONFIG menu.
The various pages of the "RECORDER CONFIG" can be schematised as follows:
EN - 58
SIRIUS89N
R
MENU
Press ENTER to confirm and leave the Menu keeping the settings made. Press ESC to leave the Menu without modifying the existing parameters.
ECORDER CONFIG
START
MANU
STOP MANU
INT. PERIOD: 15min
ENTER
+ -
MENU GENERAL
SAFETY TEST MEMORY ANALYZER MEMORY RESET
ANALYZER CONFIG
RECORDER CONFIG
CONTRAST DATE&TIME LANGUAGE
To Select MANUAL or AUTOMATIC start/stop mode, place the cursor on MANU or AUTO using the multifunction key F1 or F2 and select the desired mode using F3 or F4.
Use the multifunction keys F1, F2 to position the cursor on the desired word and use the multifunction keys F3 / F4 to modify the value. Press ENTER to confirm and proceed inside the Menu the Menu keeping the settings made. Press ESC to leave the Menu without modifying the existing parameters.
In the following table is shown the means of each items inside RECORDER CONFIG menu:
Symbols Description Advised settings
START:MAN
STOP:MAN
START:AUTO
STOP:AUTO
INT. PERIOD
For eventual messages displayed see Appendix 1 – MESSAGES DISPLAYED
The recording of all the selected parameters will start at 00 seconds after pressing START/STOP. The recording of all the selected parameters will be interrupted manually by pressing START/STOP. The recording of all the selected values will be started / interrupted at the set dates and times. In order to start the recording the user will have to press START/STOP to set the instrument in Stand-by mode until the start date and time previously set.. The value of this parameter determines every how many seconds the values of the selected parameters will be memorised (see paragraph 16.13.1). Available choices: 5sec,10sec,30sec,1min, 2min 5min, 10min, 15min, 30min, 60min.
☺ ☺
15min
EN - 59
SIRIUS89N
7.2.2. RECORDING: setting of Typical Configurations
The following “Typical Configurations” are selectable inside meter:
Standard Configuration Description
LEAKAGE (I1) Setting of measuring and recording mode of Leakage current on I1 channel.
TEMP. °C(I1) Setting of measuring and recording mode of Temperature (°C) on I1 channel.
HUM %HR(I1)
Setting of measuring and recording mode of Relativity Humidity %HR on I1 channel.
To activate the above configurations, follow this procedure:
1. Turn the rotary switch on “AUX”.
2. Press MENU key. The instrument shows the following screen:
MENU GENERAL
SAFETY TEST MEMORY ANALYZER MEMORY RESET
ANALYZER CONFIG
RECORDER CONFIG
CONTRAST DATE&TIME LANGUAGE
3. Press MENU key again. The instrument displays a screen as shown below (picture on the left). In this way is possible select the desired configuration on I1 channel with F1 or F2 key.
TYPICAL CONFIG.
LEAKAGE(I1)
TEMP.°C(I1)
HUM %HR(I1)
Configuration selection Confirmation configuration selection
TYPICAL CONFIG.
LEAKAGE (I1)
TEMP.°C(I1)
HUM %HR(I1)
Data saved
4. Press ENTER key. The instrument displays the message “Data saved” for a while to confirm the desired configuration (see above picture on the right). The instrument return on the measure mode and you can start recording by pressing START/STOP key.
EN - 60
SIRIUS89N
7.3. SOUND LEVEL MEASUREMENT PROCEDURE
This working mode allows to execute real time measurement of the Sound Pressure Level and to get
1. Press this key to access to "SOUND" mode.
2. Connect the Sound Level probe (class 1) to instrument using the original Serial Optical cable (C2001) and the adapter.
F1 F3F4F2
START STOP
HOLD
SAVE
ENTER
MENU ESC
Connection between HT55 and the instrument
3. Set the switch on the Sound Level probe to ON position.
The noise probe HT55 is not equipped with auto-power-off device, in order
CAUTION
to allow also long measurements. To maximize the battery duration put the instrument ON only during the measurements.
START
STOP
4. Press START/STOP to run the measurement. The message "HT55 no RS232" means that the instrument is not correctly connected with the noise probe HT55. Please check:
- the Sound Level Probe batteries.
- if the Switch on the Sound Level Probe in the ON position.
- if the cable and the adapter are correctly connected.
During the measurement the instrument shows the screen indicated to side. The Peak value will be available only at the end of measurement.
SOUND 05.06.01
SPL 84.2dB Peak ----dB
Duration 0000:00:00
MEASURING...
Sound Pressure Level
Measurement Duration
PG+
EN - 61
SIRIUS89N
START
STOP
5. Press START/STOP to stop the measurement and display the message "waiting for final SPL" and then the following screen will be displayed:
.
05.06.01 11:43:04
SOUND
Leq 94.2dB Peak 121.7dB
Duration 0000:00:00
MEASURING...
PG+
Sound Pressure Level
Peak Value (dBA)
Measurement Duration
It contains:
- Duration: duration of the measurement expressed in hours : minutes : seconds
- LEq: value of equivalent noise level during the measurement.
+ 01
L
eq
1
10
0
t
log10
tt
[]
=
1
t
0/)(
2
dtptp
⎥ ⎦
- Peak: peak value of SPL.
6. The displayed result can be stored pressing the SAVE key twice (according to paragraph 9.1).
7.3.1.1. SOUND LEVEL MEASUREMENT "faulty cases"
If the instrument communicate with the probe, the message "HT55: NO RS232" will be displayed.
SOUND 05.06.01
SPL ----dB Peak ----dB Duration 0000:00:00
o
HT55:NO RS232
!
PG*
The instrument can't communicate with Probe. Check if the Batteries aren't weak, the cable and the adapter are correctly connected.
THE PREVIOUS RESULTS CAN'T BE SAVED.
EN - 62
SIRIUS89N
8. ANALYZER
This function allows the following operations: display in real time the electrical parameters of a single phase systems (with and
without neutral wire) and the harmonic analysis of voltage and current.
conduct a direct Energy measurement (without memorizing). memorize (pressing SAVE key) the sampled values of the Parameters present at
instrument input generating a "Smp" record inside instrument memory. It will be possible to analyse the memorized data ONLY by transferring it to a PC.
record simultaneously (pressing the START key after a proper setting): RMS values
of voltage, current, corresponding harmonics, active, reactive and apparent powers,
ϕ
power factors and cos (voltage sag and surge) with 10ms resolution. It will be possible to analyse the recorded data ONLY by transferring them to a PC.
record simultaneously (pressing the START key) values of Voltage and Current
values coming from test leads and Clamp, using “Typical Configurations” with pre­programmed parameters inside meter (see paragraph 10.2 ).It will be possible to analyze the recorded data ONLY by transferring them to a PC.
It's fundamental the Instrument settings correspond to the Installation type under test and accessories is use. For this we recommend you to check instrument's setting before execute an ANALYSIS measurement.
Select the ANALYZER rotary switch position. By pressing the MENU key the following screen will be displayed:
, active, reactive and apparent energies, voltage anomalies
MENU GENERAL
SAFETY TEST MEMORY
ANALYZER MEMORY RESET
ANALYZER CONFIG RECORDER CONFIG
CONTRAST DATE&TIME LANGUAGE
It’s not possible to enter the MENU during a recording or a Real Time Energy measurement.
Generally to check instrument's settings you must check "ANALYZER CONFIG" and "RECORDER CONFIG" items.
EN - 63
SIRIUS89N
8.1. BASIC SETTING: ANALYZER CONFIG
Place the rotary switch in the ANALYZER position, press the MENU key, using the F1/F2 keys select the ANALYZER CONFIG item and press the ENTER Key. The following page will be displayed:
ANALYZER CONFIG
SYSTEM :SINGLE
FREQUENCY:50HZ
CURRENT RANGE:1000A
CLAMP TYPE: STD
TV RATIO:0001
PASSWORD:ON
+
-
This page of setting can be confirmed by pressing the ENTER key or cancelled by pressing the ESC key.
8.1.1. Type of electrical system under test
The parameter SYSTEM is fixed to "SINGLE" value as the instrument is able only to analyse Single Phase system.
8.1.2. How to set the fundamental frequency
Position the cursor on the corresponding word by pressing the multifunction keys F1 and F2 and select the network frequency between the possible values 50Hz and 60Hz by
pressing the multifunction keys F3 and F4. This parameter is important ONLY if the input voltage is not sufficient to recognise the value of the frequency (for example, only the clamps for the current measurement are connected). In this case the instrument generates an internal synchronism equal to the value of the set frequency.
8.1.3. How to set the current range
The value of this parameter must be always equal to the full scale of the current clamp used to take the measurement. In case multi-scale clamp is used, the value of this
parameter must be equal to the scale selected on the clamp. Set the desired value by pressing the multifunction keys F3 and F4.
8.1.4. Clamp Type
The value of this parameter must be always equal to the clamp type you are using. Three types of clamps are available, set the desired value by pressing the keys F3 and F4:
STD: for Standard clamps or Current Transformer FlexEXT: for Flexible clamps with External electronic control box FlexINT: for Flexible clamps without any electronic control box
CAUTION
If “FlexINT” is chosen the current range could be set only to 1000A or 3000A.
EN - 64
SIRIUS89N
CAUTION
The instrument has the only I1 active auxiliary input, so please pay attention while connecting the clamp. The instrument does not recognize any clamp which is connected to the I2 or I3 inputs.
8.1.5. How to set the value of the transformer voltage ratio (TV RATIO)
The instrument can also be interfaced with step-down transformers in the equipment under test: it can display the value of he voltages present on the primary winding of these transformers. To do this it will be necessary to set the value of the transformers’ windings ratio from 2:1 to 3000:1. The default is set at 1:1 for measurements of none transformer systems. Select “TV RATIO” in the ANALYZER CONFIG menu. Set the desired value by pressing the multifunction keys F3 and F4.
8.1.6. How to enable/disable the password
The instrument is provided with a protective routine to avoid the risk of being disturbed or interrupted during a recording or an energy measurement. Once a recording or a direct energy measurement has been started (with the option “PASSWORD” enabled), after about 3 minutes from the last key pressure or switch rotation it won’t be possible to press START/STOP to stop the recording, “PASSWORD” will be displayed and it will be necessary to insert the password. In order to insert the password (which is not changeable), press the multifunction keys in the following sequence (within 10 seconds):
F1, F4, F3, F2
If you wait more than about 10 seconds the display will return to the meter mode and the instrument will continue recording. If you insert a wrong password the message “Password error” will be displayed under “PASSWORD”. After a few seconds the display will return to meter mode and the instrument will continue recording. In order to enable/disable this option the correct password will have to be entered. The display will return to meter mode and START/STOP will have to be pressed again to stop the recording. You will then need to re-enter the “ANALYZER CONFIG” menu and scroll up or down to the item “PASSWORD: ON” using the multifunction keys F1 and F2. Then turn the password off by pressing the multifunction keys F3 and F4.
EN - 65
SIRIUS89N
8.2. BASIC SETTING: RECORDER CONFIG
Place the rotary switch in the ANALYZER position, press the MENU key, using the F1/F2 keys select the RECORDER CONFIG item and press the ENTER Key. This option allows you to check and eventually modify the recording parameters and the selected parameters (up to a maximum of 62+Frequency). If the number of selected values exceeds 63 the message "too many param" will be displayed. The RECORDER CONFIG mode is divided into 4 separate sub-pages:
1st page: This page allows you to set the START/ STOP mode (AUTO or MANUAL),
the START and STOP time if AUTO mode is selected, the Integration Period value, the Enabling/Disabling of Voltage Anomalies detection, the Enabling/Disabling of Harmonics detection. Press ENTER to confirm the settings and pass to the following page. Press ESC to leave the Menu without modifying the existing parameters.
2nd page: This page is devoted to the settings relevant to the VOLTAGE recording.
Press ENTER to confirm the settings and pass to the following page. Press ESC to leave this page without modifying the existing parameters. From this page you can enter the sub-page “Harmonics” which permits to select the voltage harmonics to be recorded. Press ENTER to confirm the settings and leave the “Menu Harmonics". Press ESC to leave the "Menu Harmonics" without modifying the existing parameters.
3rd page: This page is devoted to the settings relevant to the CURRENT recording.
Press ENTER to confirm the settings and pass to the following page. Press ESC to leave this page without modifying the existing parameters. From this page you can enter the sub-page “Harmonics” which permits to select the current harmonics to be recorded. Press ENTER to confirm the settings and leave the “Menu Harmonics". Press ESC to leave the "Menu Harmonics" without modifying the existing parameters.
4th page: Menu composed of two sub-pages devoted to the selection of the
POWERS and ENERGIES to be recorded. From this page you can enter the sub-page “POWER” and “ENERGY” which permits to select the parameters to be recorded. Selecting the active powers for the recording, the corresponding active energies will be automatically selected. Selecting the reactive powers for the recording, the corresponding reactive energies will be selected. Press ENTER to leave this page confirming the modifications made. Press ESC to leave the "Menu" without modifying the existing parameters.
The various pages of the "RECORDER CONFIG" can be schematised as follows:
EN - 66
SIRIUS89N
R
R
R
MENU
ECORDER CONFIG
START
MANU
STOP MANU
INT. PERIOD: 15min HARM REC: ON ANOM REC: ON
st
1
Page of the RECORDER
CONFIG MENU
ENTER
+ -
To Select MANUAL or AUTOMATIC start/stop mode, place the cursor on MANU or AUTO using the multifunction key F1 or F2 and select the desired mode using F3 or F4.
Use the multifunction keys F1, F2 to position the cursor on the desired word and use the multifunction keys F3 / F4 to modify the value. Press ENTER to confirm this page and proceed inside the Menu the Menu keeping the settings made. Press ESC to leave this page without modifying the existing parameters.
Use the multifunction keys F1, F2 to position the cursor on the desired symbol and use the multifunction keys F3 / F4 to select / deselect the desired parameter (it’s selected if marked in black). Press ENTER to confirm and leave the Menu keeping the settings made. Press ESC to leave this page without modifying the existing parameters.
ECORDER CONFIG
VOLTAGE:
V1
HARM. REC:Pg (ON)
Vref P-N: 230V LIM+: 06% (243.8V) LIM-: 10% (207.0V)
+ -
Example of 2nd page with ANOM
If you want to change Voltage Harm. Selection place the
cursor on the corresponding “Pg” symbol then Press F3
VOLTAGE HARMONICS
ENTER
Thd DC 01 02 03 04
05 06 07 08 09 10 11
12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49
flag enabled
ECORDER CONFIG
Example of sub-page
"VOLTAGE HARMONICS"
+ -
Use the multifunction keys F1, F2 to position the cursor on the desired voltage harmonic and use the multifunction keys F3 / F4 to select / deselect (it’s selected if highlighted in black). Press ENTER to confirm this page. Press ESC to leave the Menu without modifying the existing settings. The instrument will record the values of the selected harmonics corresponding to the voltages selected in one of the two pages of the Menu previously illustrated.
EN - 67
SIRIUS89N
R
R
of RECORDER CONFIG MENU
From 2nd page
ENTER
ECORDER CONFIG
CURRENT:
I1
HARM. REC:Pg (ON)
Use the multifunction keys F1, F2 to position the cursor on the desired word and use the multifunction keys F3 / F4 to modify the value or select / deselect the desired parameter (it’s selected if marked in black). Press ENTER to confirm. Press ESC to leave the Menu without modifying the existing settings.
If you want to change Current Harm. Selection place
Cursor on the corresponding “Pg” symbol then Press F3
ENTER
Thd DC 01 02 03 04
05 06 07 08 09 10 11
12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49
Example of 3
CURRENT HARMONICS
Example of sub-page
"CURRENT HARMONICS"
ECORDER CONFIG
+ -
rd
+ -
page
Use the multifunction keys F1, F2 to position the cursor on the
desired current harmonic and use the multifunction keys F3 / F4 to select / deselect (it’s selected if marked in black). Press ENTER to confirm this page. Press ESC to leave this page without modifying the existing settings. The instrument will record the values of the selected harmonics corresponding to the currents selected in one of the two pages of the Menu previously illustrated.
EN - 68
SIRIUS89N
R
R
of RECORDER CONFIG MENU
From 3rd page
ENTER
ECORDER CONFIG
CO-GENERATION:ON
POWER:Pg
ENERGY:Pg
Example of 4
+ -
th
page
ENTER
In order to select the POWER to be recorded use the multifunction keys F1, F2 to position the cursor on the corresponding “Pg” Symbol and then press F3
ECORDER CONFIG
P1 Q1i Q1c S1 Pf1 dPf1
Example of POWER sub-page
+ -
Selecting the active power for the recording, the corresponding active energy will
CAUTION
be automatically selected.
Selecting the reactive powers for the recording, the corresponding reactive
energies will be selected.
EN - 69
SIRIUS89N
R
R
of RECORDER CONFIG MENU
From 3rd page
ENTER
ECORDER CONFIG
CO-GENERATION:ON
POWER:Pg
ENERGY:Pg
ENTER
Selecting/deselecting the active energy for the recording, the corresponding
active power will be automatically selected/deselected.
Selecting/deselecting the reactive energies for the recording, the corresponding
reactive powers will be selected/deselected.
Ea1 Eri1 Erc1
Example of ENERGY sub-page
in single-phase mode
Example of 4
ECORDER CONFIG
+ -
th
page
In order to select the ENERGIES to be recorded use the multifunction keys F1, F2 to position the cursor on the corresponding “Pg” Symbol and then press F3
+ -
CAUTION
EN - 70
SIRIUS89N
Symbols Description Advised settings
START:MAN
STOP:MAN
START:AUTO
STOP:AUTO
INT. PERIOD
HARM REC.
ANOM REC.
V1 Voltage RMS value
The recording of all the selected parameters will start at 00 seconds after pressing START/STOP (see chapter 10.1). The recording of all the selected parameters will be interrupted manually by pressing START/STOP (see chapter 10.1). The recording of all the select ed value s will be sta rted / i nterrupted at the set dates and times. In order to start the recording the user will have to press START/STOP to set the instrument in Stand-by mode until the start date and time previously set (see chapte r 10.1). The value of this parameter determines every how many seconds the values of all the selected parameters will be memorised (see chapter 16.13.1). Available choices: 5sec,10sec,30sec,1min,2min,5min,10min,15min,30min,60min.
ON = the instrument will record the values of the selected voltage and current harmonics.
OFF = the instrument will not record any voltage or current
harmonic selected
ON = the Instrument will record Voltage Anomalies (voltage Sag
and Surge) (see paragraph 16.10)
OFF = the instrument will not record any voltage Sag and Surge
☺ ☺
15min
☺ ☺
V1
THD, DC, 01...49 Voltage Total Harmonic Distortion, DC Component, 01..49
Harmonics respectively RMS reference value for Voltage used in Voltage Anomalies
Vref
(only if ANOM. REC flag
has been set ON)
LIM+, LIM-
(only if ANOM. REC flag
has been set ON)
I1
THD, DC, 01..49 Current Total Harmonic Distortion, DC Component, 01..49
detection (Voltage Sag and Surge). The Reference is: a) Voltage Phase to Neutral for Single Phase and 4 wires three
phase system
b) Voltage Phase to Phase for 3 wires three phase system High and Low Voltage Percent threshold used in Voltage
Anomalies detection (Voltage Sag and Surge). These parameters can be adjusted in range 3% ÷ 30% (step 1%). Example: Three Phase System 4 wires. Vref = 230, LIM+= 6%, LIM-=10% => High Lim = 243.8V, Low Lim = 207.0V The Instrument will detect a voltage Anomalies if the RMS Voltage Values (calculated every 10ms) beyond the above calculated thresholds (see paragraph 16.10.
Current RMS value.
Harmonics respectively
THD,01,03,05,07
Single phase: 230V
3 Phase system:
3 wires: 400V
4 wires 230V
+6% / -10%
I1
THD,01,03,05,07
EN - 71
SIRIUS89N
ON = the instrument is able to face situations of CO-GENERATION
of electrical equipment (that is, the equipment under test is able to generate energy besides absorbing it). Accordingly, the
CO-GENERATION
P1 Active Power
instrument will record the powers and energies both absorbed and generated (see paragraph 16.12.1). If this flag is
enabled the maximum number of parameters which can
be selected decrease to 38.
OFF = the instrument will record ONLY the powers and energies
absorbed.
☺ ☺
P1
Q1i Inductive Reactive Power
Q1c Capacitive Reactive Power
S1 Apparent Power
Pf1 Power factor
DPft1
Ea1 Active energy
Eri1 Inductive reactive energy
Erc1 Capacitive reactive energy
The value of the network frequency is automatically selected if the voltage is selected.
The symbols "i" and "c" stand for reactive powers (Q), power factors (Pf) and cosϕ (dpf) inductive and capacitive respectively. Selecting a power factor (Pf) or a cosϕ (dPf) for the recording automatically their inductive value and their capacitive value will be
recorded separately. For eventual messages displayed see Appendix 1 – MESSAGES DISPLAYED.
cosϕ
Q1i Q1c
S1
Pf1 dPf1
Ea1
Eri1 Erc1
EN - 72
SIRIUS89N
8.3. ANALYZER FUNCTIONS
For a simple usage, the main working mode of the ANALYZER mode can be
selected by means of F3 and F4.
"VOLTAGE" function: to be used to display voltage and corresponding
harmonics (see paragraph 8.4)
"CURRENT" function: to be used to display current and corresponding
harmonics (see paragraph 8.5)
"POWER" function: it permits to display all the parameters measurable by
the instrument: voltage, current, active, reactive and
apparent power, power factor, cosϕ and energy (see paragraph 8.6)
"ENERGY" function: to be used to display active, reactive and apparent
power, power factor, cos
ϕ
and energy (see paragraph
8.7)
More practically, we may schematise the right procedure of use the ANALYZER function as follows:
1. Check and eventually modify the "ANALYZER CONFIG" settings of the instrument
2. Using F3 and F4, select the type of measurement to be taken
3. Connect the instrument to the electrical system to be tested
4. Evaluate the values of the parameters under test
5. If you want to record: a) Decide what to record b) Press MENU and check if the "RECORDER CONFIG" settings meet your
requirements.
c) Consider Typical Configuration recording (see paragraph 10.2).
6. Connect the External Power Supply A0050 (optional)
7. Start the recording by pressing START/STOP.
8.4. "VOLTAGE" FUNCTION
This function permits you to display in real time the RMS value of AC/DC voltage,
the peak, the Thd value (see paragraph 16.11), the waveform and the harmonic spectrum of the voltage.
8.4.1. Symbols
The VOLTAGE position has three working modes:
METER WAVE HARM
These modes will be described in detail in the next paragraphs. The used symbols are described below:
Symbol Description
V1 Voltage RMS value Vpk1 Peak value of the voltage h01 ÷ h49 Harmonic 01 ÷ Harmonic 49. ThdV Factor of total harmonic distortion of the voltage (see paragraph 16.11). freq Network frequency
Tab. 1: Symbols used in the position VOLTAGE
EN - 73
SIRIUS89N
8.4.2. "METER" mode
In this mode the instrument shows the below screens according to the settings made as per paragraph 8.1.
27.09.00 17:35:12
VOLTAGE
SINGLE PHASE
V1 = 230.2 V Vpk1 = 325.5 V ThdV = 0.0 % Freq = 50.0 Hz
HARM. WAVE PG- PG+
Example of screen
The symbols used are described in Tab.1 For eventual messages displayed see Appendix 1 – MESSAGES DISPLAYED. Following keys are enabled:
F1: to pass to "HARMONIC" mode (see paragraph 8.4.3).  F2: to pass to "WAVE" mode (see paragraph 8.4.4).  F3/F4: to pass to previous/next function respectively.  SAVE: to save in the instrument memory a record of “Smp” type (see
paragraph 9.2) containing the instantaneous values of voltage and current present on the instrument inputs. This function is disabled during a recording.
ENTER/HOLD: to enable/disable the HOLD function (updating interruption) of the
displayed data. All the previous functions remain however available. When the HOLD function is enabled, the word HOLD is displayed. When this function is enabled it’s not possible to record or take an energy measurement. This function is disabled during a recording or an energy measurement.
MENU: to enter in the MENU mode and change the instrument settings (see
paragraph 8.1 and 8.2). It’s not possible to enter the configuration MENU during a recording or an energy measurement.
START/STOP: to record the selected parameters according to the instrument’s
settings (see chapter 10).
EN - 74
SIRIUS89N
8.4.3. "HARM" mode
Selecting the HARM mode the screen below will be displayed according to the settings made as per paragraph 8.1. The screen show the harmonics (see paragraph 16.11) of the voltage.
27.09.00 17:35:12
V1 = 230.2 V h03 = 10.2 V h03 = 4.3 % ThdV = 11.0 %
h49
Example of screen
The symbols used are described in Tab. 1. For eventual messages displayed see Appendix 1 – MESSAGES DISPLAYED. The displayed histograms represent the harmonic content of the voltage under test. The value of the first harmonic h01 (fundamental at 50Hz) is not represented in scale along with the other harmonics in order to maximize the display of the latter. In case both voltage and current are connected to the instrument inputs, eventual negative values of the harmonics (therefore represented under the horizontal axis), indicate that such voltage harmonics are “generated” by the load. Following keys are enabled:
F3, F4: to move the cursor of the selected harmonic leftwards and rightwards
respectively. At the same time the values relevant to the order no. of the selected harmonic and to the corresponding absolute and relative values (calculated on the basis of the fundamental) are updated.
F2: to display the page of the harmonics h01 ÷ h24 (symbol h24) or that
÷
of the harmonics h25
h49 (symbol h49).
ESC: to return back to METER mode (see paragraph 8.4.2).  SAVE: to save in the instrument memory a record of “Smp” type (see
paragraph 9.2) and the instantaneous values of voltage and current present on the instrument inputs. This function is disabled during a recording.
ENTER/HOLD: to enable/disable the HOLD function (updating interruption) of the
displayed data. All the previous functions remain however available. When the HOLD function is enabled, the word HOLD is displayed. When this function is enabled it’s not possible to record or take an energy measurement. This function is disabled during a recording or an energy measurement.
MENU: to enter in the MENU mode and change the instrument settings (see
paragraph 8.1 and 8.2). It’s not possible to enter the configuration MENU during a recording or an energy measurement.
START/STOP: to record selected parameters according to the instrument’s settings
(see chapter 10).
EN - 75
SIRIUS89N
8.4.4. "WAVE" mode
Selecting the WAVE mode the screen below will be displayed according to the settings made as per paragraph 8.1. The screen show the waveform of the voltage.
27.09.00 17:35:12
V1 = 230.2 V Vpk1 = 325.5 V freq = 50.0 Hz
Example of screen
The symbols used are described in Tab. 1. For eventual messages displayed see Appendix 1 – MESSAGES DISPLAYED. Following keys are enabled:
ESC: to return back to METER mode (see paragraph 8.4.2).  SAVE: to save in the instrument memory a record of “Smp” type (see
paragraph 9.2) containing the instantaneous values of voltage and current present on the instrument inputs. This function is disabled during a recording.
ENTER/HOLD: to enable/disable the HOLD function (updating interruption) of the
displayed data. All the previous functions remain however available. When the HOLD function is enabled, the word HOLD is displayed. When this function is enabled it’s not possible to record or take an energy measurement. This function is disabled during a recording or an energy measurement.
MENU: to enter in the MENU mode and change the instrument settings (see
paragraph 8.1 and 8.2). It’s not possible to enter the configuration MENU during a recording or an energy measurement.
START/STOP: to record selected parameters according to the instrument’s settings
(see chapter 10).
EN - 76
SIRIUS89N
8.5. "CURRENT" FUNCTION
This function permits to display in real time the RMS value of AC/DC current, the peak, the Thdl value (see paragraph 16.11), the waveform and the harmonic spectrum of the current.
8.5.1. Symbols
The CURRENT position has three working modes:
METER WAVE HARM
These modes will be described in detail in the next paragraphs. The symbols used are described below:
Symbol Description
I1 Current RMS value Ipk1 Peak value of the current h01 ÷ h49 Harmonic 01 ÷ harmonic 49. ThdI Total harmonic distortion factor of the current (see paragraph 16.11). freq Network frequency
Tab. 2: Symbols used in the position CURRENT
EN - 77
SIRIUS89N
8.5.2. “METER" mode
In this mode the instrument shows the below screens according to the settings made as per paragraph 8.1.
27.09.00 17:35:12
CURRENT
SINGLE PHASE
I1 = 30.21 A Ipk1 = 49.53 A ThdI = 23.06 % freq = 50.0 Hz
CLAMP TYPE: STD
HARM. WAVE PG- PG+
Example of screen
The symbols used are described in Tab. 2. For eventual messages displayed see Appendix 1 – MESSAGES DISPLAYED. Following keys are enabled:
F1: to pass to "HARMONIC" mode (see paragraph 8.5.3).  F2: to pass to "WAVE" mode (see paragraph 8.5.4).  F3/F4: to pass to previous/next function respectively.  SAVE: to save in the instrument memory a record of “Smp” type (see
paragraph 9.2) containing the instantaneous values of voltage and current present on the instrument inputs. This function is disabled during a recording.
ENTER/HOLD: to enable/disable the HOLD function (updating interruption) of the
displayed data. All the previous functions remain however available. When the HOLD function is enabled, the word HOLD is displayed. When this function is enabled it’s not possible to record or take an energy measurement. This function is disabled during a recording or an energy measurement.
MENU: to enter in the MENU mode and change the instrument settings (see
paragraph 8.1 and 8.2). It’s not possible to enter the configuration MENU during a recording or an energy measurement.
START/STOP: to record selected parameters according to the instrument’s settings
(see chapter 10).
EN - 78
SIRIUS89N
8.5.3. “HARM" mode
Selecting the HARM mode the screen below will be displayed according to the settings made as per paragraph 8.1. The screen shows the harmonics (see paragraph 16.11) of the current.
27.09.00 17:35:12
I1 = 230.2 A h03 = 10.2 A h03 = 4.3 % ThdI = 11.0 %
h49
Example of screen
The symbols used are described in Tab. 2. For eventual messages displayed see Appendix 1 – MESSAGES DISPLAYED. The displayed histograms represent the harmonic content of the current under test. The value of the first harmonic h01 (primary at 50Hz) is not represented in scale along with the other harmonics in order to maximise the display of the latter. In case both voltage and current are connected to the instrument inputs, eventual negative values (therefore represented under the horizontal axis), indicate that such current harmonics are “generated” by the load. Following keys are enabled:
F3, F4: to move the cursor of the selected harmonic leftwards and rightwards
respectively. At the same time the values relevant to the order no. of the selected harmonic and to the corresponding absolute and relative values (calculated on the basis of the fundamental) are updated.
F2: to display the page of the harmonics h01 ÷ h24 (h24 symbol) or that
÷
of the harmonics h25
h49 (h49 symbol).
ESC: to return back to METER mode (see paragraph 8.5.2)  SAVE: to store in the instrument memory a record of “Smp” type (see
paragraph 9.2) and the instantaneous values of voltage and current present on the instrument inputs. This function is disabled during a recording.
ENTER/HOLD: to enable/disable the HOLD function (updating interruption) of the
displayed data. All the previous functions remain however available. When the HOLD function is enabled, the word HOLD is displayed. When this function is enabled it’s not possible to record or take an energy measurement. This function is disabled during a recording or an energy measurement.
MENU: to enter in the MENU mode and change the instrument settings (see
paragraph 8.1 and 8.2). It’s not possible to enter the configuration MENU during a recording or an energy measurement.
START/STOP: to record selected parameters according to the instrument’s settings
(see chapter 10).
EN - 79
SIRIUS89N
8.5.4. "WAVE" mode
Selecting the WAVE mode the screen below will be displayed according to the settings made as per paragraph 8.1. The screen shows the waveform of the current.
27.09.00 17:35:12
I1 = 230.2 A Ipk1 = 325.5 A freq = 50.0 Hz
Example of screen
The symbols used are described in Tab. 2. For eventual messages displayed see Appendix 1 – MESSAGES DISPLAYED. Following keys are enabled:
ESC: to return back to METER mode (see paragraph 8.5.2).  SAVE: to save in the instrument memory a record of “Smp” type (see
paragraph 9.2) containing the instantaneous values of voltage and current present on the instrument inputs. This function is disabled during a recording.
ENTER/HOLD: to enable/disable the HOLD function (updating interruption) of the
displayed data. All the previous functions remain however available. When the HOLD function is enabled, the word HOLD is displayed. When this function is enabled it’s not possible to record or take an energy measurement. This function is disabled during a recording or an energy measurement.
MENU: to enter in the MENU mode and change the instrument settings (see
paragraph 8.1 and 8.2). It’s not possible to enter the configuration MENU during a recording or an energy measurement.
START/STOP: to record selected parameters according to the instrument’s settings
(see chapter 10).
EN - 80
SIRIUS89N
8.6. "POWER" FUNCTION
This function permits to display in real time the RMS value of AC/DC voltage, the peak and ThdV value and the waveform of the voltage, the RMS value of AC/DC currents, the peak, Thdl value and the waveform of the current. Furthermore, the instrument calculates and displays the value of the active, reactive and apparent power and the value of the power factors and cosϕ.
8.6.1. Symbols
The position POWER has two working modes:
METER WAVE
For voltage and current harmonics see paragraphs 8.4.3 and 8.5.3 respectively. These modes will be described in detail in the next paragraphs.
The symbols used are described below:
Symbol Description
V1 Voltage RMS value freq Network frequency I1 Current RMS value P1 Active power Q1 Reactive power S1 Apparent power Pf1 Power factor dpf1
cosϕ
Tab. 3: Symbols used in the position POWER
The symbols "i" and "c" stand for reactive powers (Q), power factors (Pf) and cosϕ (dpf) respectively inductive and capacitive.
EN - 81
SIRIUS89N
8.6.2. "METER" mode
In this mode the instrument shows the below screens according to the settings made as per paragraph 8.1.
27.09.00 17:35:12
POWER
SINGLE PHASE
V1 = 230.0 V I1 = 145.3 A P1 = 32.91 kW Q1 = 5.767 kVAR S1 = 33.41 kVA pf1 = 0.99 i dpf1 = 0.99 i
WAVE PG- PG+
Example of screen
The symbols used are described in Tab. 3. For eventual messages displayed see Appendix 1 – MESSAGES DISPLAYED Following keys are enabled:
F2: to pass to "WAVE" mode (see paragraph 8.6.3).  F3/F4: to pass to previous/next function respectively.  SAVE: to save in the instrument memory a record of “Smp” type (see
paragraph 9.2) and the instantaneous values of voltage and current present on the instrument inputs. This function is disabled during a recording.
ENTER/HOLD: to enable/disable the HOLD function (updating interruption) of the
displayed data. All the previous functions remain however available. When the HOLD function is enabled, the word HOLD is displayed. When this function is enabled it’s not possible to record or take an energy measurement. This function is disabled during a recording or an energy measurement.
MENU: to enter in the MENU mode and change the instrument settings (see
paragraph 8.1 and 8.2). It’s not possible to enter the configuration MENU during a recording or an energy measurement.
START/STOP: to record selected parameters according to the instrument’s settings
(see chapter 10).
EN - 82
SIRIUS89N
8.6.3. "WAVE" mode
Selecting the WAVE mode the screen below will be displayed according to the settings made as per paragraph 8.1. The screen shows the waveform of the current and the voltage.
27.09.00 17:35:12
V1 = 229.7 V I1 = 132.0 A Pf1 = 0.98 i
Example of screen
The symbols used are described in Tab. 3. For eventual messages displayed see appendix 1 – MESSAGES DISPLAYED. Following keys are enabled:
ESC: to return back to METER mode (see paragraph 8.6.2).  SAVE: to save in the instrument memory a record of “Smp” type (see
paragraph 9.2) containing the instantaneous values of voltage and current present on the instrument inputs. This function is disabled during a recording.
ENTER/HOLD: to enable/disable the HOLD function (updating interruption) of the
displayed data. All the previous functions remain however available. When the HOLD function is enabled, the word HOLD is displayed. When this function is enabled it’s not possible to record or take an energy measurement. This function is disabled during a recording or an energy measurement.
MENU: to enter in the MENU mode and change the instrument settings (see
paragraph 8.1 and 8.2). It’s not possible to enter the configuration MENU during a recording or an energy measurement.
START/STOP: to record selected parameters according to the instrument’s settings
(see chapter 10).
EN - 83
SIRIUS89N
8.7. "ENERGY" FUNCTION
This function permits to display the values of the active powers, capacitive and inductive reactive powers, power factor and cosϕ. Furthermore, the instrument is able to measure directly (see 8.7.2) the values of the energies and the values of the capacitive and inductive reactive energies.
8.7.1. Symbols
The position ENERGY has only one working mode:
METER
This mode will be described in detail in the next paragraphs. The symbols used are described below:
Symbol Description
P1 Active power Q1 Reactive power S1 Apparent power pf1 Power factor dpf1 Ea1 Active energy Eri1 Inductive reactive Energy Erc1 Capacitive reactive Energy
cosϕ
Tab. 4: Symbols used in the position ENERGY
The symbols "i" and "c" stand for reactive powers (Q) and energies (Er) inductive and capacitive respectively.
EN - 84
SIRIUS89N
8.7.2. "METER" mode
In this mode the instrument shows the below screens according to the settings made as per paragraph 8.1.
27.09.00 17:35:12
ENERGY
SINGLE PHASE
Ea1 = 0.000 kWh Erc1 = 0.000 kVARh Eri1 = 0.000 kVARh P1 = 36.38 kW Q1 = 6.375 kVAR S1 = 36.94 kVA dpf1 = 0.98 i Meas Time: 00:00:00
MEAS PG- PG+
Example of screen
The symbols used are described in Tab. 4. For eventual messages displayed see Appendix 1 – MESSAGES DISPLAYED. Following keys are enabled:
F2: to start / stop immediately a direct energy measurement. The
energy counters will start increasing proportionally to the active power absorbed by the load.
The results obtained cannot be memorised.
F3/F4: to pass to previous/next function respectively.
If the active power is negative the counters will not increase.
SAVE: to save in the instrument memory a record of “Smp” type (see
paragraph 9.2) containing the instantaneous values of voltage and current present on the instrument inputs. This function is disabled during a recording.
ENTER/HOLD: to enable/disable the HOLD function (updating interruption) of the
displayed data. All the previous functions remain however available. When the HOLD function is enabled, the word HOLD is displayed. When this function is enabled it’s not possible to record or take an energy measurement. This function is disabled during a recording or an energy measurement.
MENU: to enter in the MENU mode and change the instrument settings (see
paragraph 8.1 and 8.2). It’s not possible to enter the configuration MENU during a recording or an energy measurement.
START/STOP: to record selected parameters according to the instrument’s settings
(see chapter 10).
EN - 85
SIRIUS89N
8.8. MEASURING PROCEDURES
8.8.1. Using the Instrument in a Single Phase System
The maximum voltage between B1 and B4 inputs is 600 V~ (CATII) / 350V~
CAUTION
phase – earth or 600V~ (CATIII) / 300 V~ phase to earth. Do not measure voltages exceeding the limits prescribed by this manual. Should you exceed the voltage limits you could damage the instrument and/or its components or endanger your safety.
Instrument connection in a single-phase system
CAUTION
If possible, before connecting the instrument to the electrical equipment to be tested take the power supply off the electrical equipment.
1. Check, and if needed modify, the basic settings of the instrument (see paragraphs 8.1 and 8.2).
2. Select the working mode corresponding to the type of analysis desired. In case of doubts select the POWER working mode (see paragraph 8.6).
3. Connect the phase and neutral voltage wires respecting the connections shown in the picture.
4. If you want to measure current and power, connect the clamp meter to the phase conductor respecting the specifications shown on the clamp and the connections shown in the picture. In case of doubts select the POWER working mode and check if the active power P is positive. If it’s negative, remove current transducer from the wire and reconnect it so the transducer label faces the opposite direction.
5. Apply voltage to the electrical equipment under test (if previously shut off for the instrument connection).
6. The values of the available electrical parameters will be displayed on the display of the instrument. For further details see the paragraph relevant to the position of the switch.
7. You can press HOLD to interrupt the updating in real time of the displayed values.
8. You can press SAVE to save the displayed values (see paragraph 9.2).
9. If you want to record: a) Check, and if needed modify, the values of the basic parameters (see paragraphs
8.1 and 8.2).
b) Check, and if needed modify, the recording parameters by pressing MENU (see the
paragraph corresponding to the position of the rotary switch selected).
c) To start the recording press START (see chapter 6).
EN - 86
SIRIUS89N
8.8.2. Using the Instrument in a Three Phase System
The maximum voltage between B1 and B4 inputs is 600 V~ (CATII) / 350V~
CAUTION
phase – earth or 600V~ (CATIII) / 300 V~ phase to earth. Do not measure voltages exceeding the limits prescribed by this manual. Should you exceed the voltage limits you could damage the instrument and/or its components or endanger your safety.
The instrument is a single phase system. So you can analyse a three phase system repeating 3 times a single phase analysis described in previous paragraph.
Instrument connection in a three-phase system
The Three Phase power values are given by:
Total Active Power = Pt = P1 + P2 + P3
Total Reactive Power = Qt= Q1 + Q2 + Q3
Total Apparent Power = St = √(Pt2 + Qt2)
EN - 87
SIRIUS89N
9. SAVING RESULTS
The SAVE button can be used to store the displayed results related to the rotary switch position:
SAFETY TEST and for AUX rotary switch position: pressing this key the instrument will
store the displayed result generating a corresponding record in the SAFETY TEST MEMORY (see paragraph 11.1)
ANALYZER rotary switch position: pressing this key the instrument will store the
displayed result generating a "Smp" record in the ANALYZER MEMORY (see paragraph 11.2)
Please note that Saving results is different from recording.
9.1. SAVING SAFETY TEST RESULTS
After a SAFETY TEST (function LOWΩ, MΩ, RCD, LOOP, Phase sequence, EARTH) or during a real time measurement in AUX position the user can press the SAVE button to store the displayed result.
The REMINDER PLACE parameter isn't related to Measurement Order Number and can help the user to remind the place where he performed the measurement.
SAVE IN LOC. 006
OK? (SAVE/ESC)
REMINDER PLACE:009
The following keys are available:
F3, F4: to adjust the REMINDER PLACE.  SAVE: to store the test result associating to the actual REMINDER PLACE  ESC: to quit this mode without saving.
9.2. SAVING DISPLAYED VALUES OF ANALYZER FUNCTION
05.06.01
Memory Location
REMINDER PARAMETER
During a Real Time measurement (in ANALYZER function) if the user press the SAVE button, a "Smp" record will be generate in the "ANALYZER MEMORY". This file contains the Voltage and Current values present at instrument's input when the user pressed the SAVE key. Downloading these values to a PC (using the management Software) the Power, Energy, Harmonics, etc values can be calculated and displayed as well.
EN - 88
SIRIUS89N
10. RECORDINGS
10.1. START A RECORDING
The recording function is available for ANALYZER and AUX rotary switch position. As you can read in the paragraphs 7.2.1 and 8.2, a recording can be started manually or automatically. Therefore, after setting all the parameters and leaving the Menu, the instrument will start to record:
MANUALLY: the recording will start when Instrument time reach the “00” seconds
value after pressing START/STOP.
AUTOMATICALLY: If the operator has pressed START/STOP the instrument will
remain in stand-by until the date and time previously set, then the recording will start. While if the operator doesn’t press START/STOP the recording will never start.
For recordings we recommend to use the external power supply adapter (optional code A0050) even the instrument allows the operator to perform a recording using internal batteries.
If you press Start a recording without the external power supply adapter (optional code A0050) the instrument will display a warning message “No ext supply”. Press START key again to run the recording or press ESC to quit.
If during a recording the external power supply adapter (optional code A0050) is de­energised, the instrument will continue the recording using the internal battery power until the batteries are exhausted (the data stored until the definitive turning off won’t get lost). For this we recommend you ALWAYS insert a new set of batteries before a long recording.
The instrument uses sophisticated algorithms to prolong the battery life. Particularly:
The instrument switches OFF the backlight Automatically after 5 seconds. If the Battery level is too low the Backlight function will be disabled. If the instrument is just displaying in real time (and the external power supply is not
connected), after about 5 minutes from the last key pressure or switch rotation the instrument turns off automatically ("AUTOPOWER OFF" function).
If the instrument is recording or is measuring energy (and the external power supply is
not connected), after about 5 minutes from the last key pressure or switch rotation the instrument starts a special procedure to save the batteries ("ECONOMY MODE"): the instrument keeps recording but the display is turned off.
Before starting a recording the operator should first evaluate the state of the equipment, decide what to record and set the instrument accordingly.
CAUTION
EN - 89
SIRIUS89N
10.2. SETTING TYPICAL CONFIGURATIONS
In order to facilitate this task, the instrument is provided of the following two pre-setting recording modes:
1. Default Configuration: it’s comprehensive which should fit most cases.
2. Typical Configuration: it’s possible to select recording with pre-setting parameters for
the following situations:
EN50160
SURGES & DIPS
HARMONICS
START-UP POWER & ENERGY
Setting parameters for Networks Quality in compliance with EN 50160 standard (see paragraph 16.11.2). Setting parameters for Voltage Anomalies detection (surges, dips, break, etc.) (see paragraph 16.10). Setting parameters for Harmonics Analysis of Voltage and Current (see paragraph 16.11). Setting parameters for Start-Up motors and electrical devices. Setting parameters for Power and Energy measures (see paragraph 16.12).
10.2.1. Default Configuration
The default configuration of instrument consists in the following parameters setting: ANALYZER CONFIG:
System: SINGLE Frequency: 50Hz Current Range: 1000A Clamp Type: FlexINT TV Ratio: 1 Password: enabled
RECORDER CONFIG:
Start: MANU (the recording starts at
00 sec mark on clock after
pressing the START/STOP key) Stop: MANU Integration period: 15min Recording of harmonics: ON Recording of Voltage anomalies (voltage Sag and Surge) : ON Voltage Reference for Sag and Surge detection: 230V Upper Limit for Sag and Surge detection: 6% Lower Limit for Sag and Surge detection: 10% Selected voltages: V1 Selected voltage harmonics: THD, 01, 03, 05, 07 Selected currents: I1 Selected current harmonics: THD, 01, 03, 05, 07 CO-GENERATION: OFF Powers, Pf and cosϕ selected: P1, Q1i, Q1c, S1, Pf1, dpf1 Energies: Ea1, Eri1, Erc1
If the user has changed the instrument’s settings, he can quickly resume the above configuration by using the RESET option (see paragraph 5.4).
By pressing START/STOP the recording of the selected parameters is started according to the settings made in the MENU (see paragraphs 8.1 and 8.2). The rotary switch position doesn’t affect the recording setting.
EN - 90
SIRIUS89N
As the default value of the integration periods is set at 15 minutes the instrument will store data in the temporary memory for 15 minutes. Afterwards the instrument will elaborate the results saved in the temporary memory and will save the result of this elaboration (min, avg, and max values) in the definitive memory. Therefore, if an integration period of 15 minutes has been set, the recording will continue for about 15 minutes before producing a series of recorded values. If the recording is interrupted before the selected integration period has completely elapsed the data stored in the temporary memory will not be elaborated and the corresponding series of values won’t be transferred to the definitive memory.
10.2.2. Typical Configurations
To activate the Typical Configurations, follow this procedure:
1. Turn the rotary switch on “ANALYZER”.
2. Press MENU key. The instrument shows the following screen:
MENU GENERAL
SAFETY TEST MEMORY ANALYZER MEMORY RESET
ANALYZER CONFIG
RECORDER CONFIG
CONTRAST DATE&TIME LANGUAGE
3. Press MENU key again. The instrument displays a screen as shown below (picture on the left). This way is possible to select the desired configuration with F1 or F2 key.
TYPICAL CONFIG.
EN50160
SURGES & DIPS
HARMONICS
START-UP
POWER & ENERGY
Configuration selection Confirmation configuration selection
TYPICAL CONFIG.
EN50160
SURGES & DIPS
HARMONICS
START-UP
POWER & ENERGY
Data saved
4. Press ENTER key. The instrument displays the message “Data saved” for a while to confirm the desired configuration (see above on the right). The instrument return to the measure mode.
Below you can find the parameters for each of 5 Typical Configurations:
EN - 91
SIRIUS89N
EN50160
ANALYZER CONFIG:
All the parameters: unmodified
RECORDER CONFIG:
Start: MANU (the recording starts at
00 sec mark on clock after
pressing the START/STOP key) Stop: MANU Integration period: 10min Recording of harmonics: ON Recording of Voltage anomalies (voltage Dips and Surge) : ON Voltage Reference for Dips and Surge detection: 230V Upper Limit for Dips and Surge detection: 6% Lower Limit for Dips and Surge detection: 10% Selected voltages: V1 Selected voltage harmonics: THD, DC, 01, 02, 03, 04, … 25 CO-GENERATION: OFF
SURGES & DIPS
ANALYZER CONFIG:
All the parameters: unmodified
RECORDER CONFIG:
Start: MANU (the recording starts at
00 sec mark on clock after
pressing the START/STOP key) Stop: MANU Integration period: 1min Recording of harmonics: OFF Recording of Voltage anomalies (voltage Dips and Surge) : ON Voltage Reference for Dips and Surge detection: 230V Upper Limit for Dips and Surge detection: 6% Lower Limit for Dips and Surge detection: 10% Selected voltages: V1 Selected current: I1 CO-GENERATION: OFF
HARMONICS
ANALYZER CONFIG:
All the parameters: unmodified
RECORDER CONFIG:
Start: MANU (the recording starts at
00 sec mark on clock after
pressing the START/STOP key) Stop: MANU Integration period: 10min Recording of harmonics: ON Recording of Voltage anomalies (voltage Dips and Surge) : OFF Selected voltages: V1 Sel. Harmonics voltage : THD,DC,01,03,05,07,09,11,13,15,17,19,21,23,25 Selected current: I1 Sel. Harmonics current: THD,DC,01,03,05,07,09,11,13,15,17,19,21,23,25 CO-GENERATION: OFF
EN - 92
SIRIUS89N
START-UP
ANALYZER CONFIG:
All the parameters: unmodified
RECORDER CONFIG:
Start: MANU (the recording starts at
00 sec mark on clock after
pressing the START/STOP key) Stop: MANU Integration period: 5sec Recording of harmonics: ON Recording of Voltage anomalies (voltage Dips and Surge) : ON Voltage Reference for Dips and Surge detection: 230V Upper Limit for Dips and Surge detection: 6% Lower Limit for Dips and Surge detection: 10% Selected voltages: V1 Selected voltage harmonics: THD, 01, 03, 05, 07, 09, 11, 13, 15 Selected currents: I1 Selected current harmonics: THD, 01, 03, 05, 07, 09, 11, 13, 15 CO-GENERATION: OFF Selected powers, Pf and cosϕ: P1, Q1i, Q1c, S1, Pf1, dpf1 Energies: Ea1, Eri1, Erc1
POWER & ENERGY
ANALYZER CONFIG:
All the parameters: unmodified
RECORDER CONFIG:
Start: MANU (the recording starts at
00 sec mark on clock after
pressing the START/STOP key) Stop: MANU Integration period: 15min Recording of harmonics: OFF Recording of Voltage anomalies (voltage Dips and Surge) : OFF Selected voltages: V1 Selected currents: I1 CO-GENERATION: ON Selected powers, Pf and cosϕ: P1, Q1i, Q1c, S1, Pf1, dpf1 Energies: Ea1, Eri1, Erc1
By pressing START/STOP the recording of the selected parameters is started according to the settings made for each Typical Configuration. The rotary switch position doesn’t affect the recording setting.
EN - 93
SIRIUS89N
10.3. DURING A RECORDING
If during a recording the external power supply is de-energised, the instrument will continue the recording using the internal battery power until the batteries are exhausted (the data stored up to the point the instrument shuts down won’t get lost). For this we recommend you ALWAYS insert a new set of batteries before a long recording.
The instrument uses sophisticated algorithms to prolong the battery life. Particularly if the instrument is recording or is measuring energy (and the external power supply is not connected), after about 5 minutes from the last key pressure or switch rotation the instrument starts a special procedure to save the batteries ("ECONOMY MODE"): the instrument keeps recording but the display is turned off.
During a recording the following are disabled:
AUTOPOWER OFF functionON/OFF keyHOLD keySAVE key
10.3.1. MENU key
If you press the MENU key during a recording the following screen will appear:
INFO REC n XX
START
09.18.01 11:35 STOP
13.18.01 12:00 INT PERIOD: 15min REC PERIODS:00004 REC TIME:139d.02h HARM REC: (ON) ANOM REC: (ON) N ANOMALIES: 00000
Recording Recording
This page includes:
1. START Date and Time
2. STOP Date and Time (or Manual).
3. Integration Period
4. Actual Number of Elapsed Integration Periods
5. Actual Rec ording Time
6. Status of Harmonic Flag
7. Status of Voltage Anomalies Flag
8. Number of Voltage anomalies occurred during the recording
EN - 94
SIRIUS89N
10.3.2. Rotary Switch during a recording
If You move the rotary switch during a recording the following screen will appear:
Recording Recording
This page means that a recording is running but the actual rotary switch position doesn't correspond to this.
The instrument will continue to record.
10.4. STOPPING A RECORDING OR AN ENERGY MEASUREMENT
The instrument uses a protective routine to avoid the risk of being disturbed or interrupted during a recording or an energy measurement. Once a recording or a direct energy measurement (see paragraph 8.7.2) has been started (with the option PASSWORD enabled), after about 3 minutes from the last key pressure or switch rotation it won’t be sufficient to press START/STOP (if a recording is running) or F2 (if an energy measuring is running) to stop the recording, it will be necessary to insert the password. In order to insert the password (which is not changeable), press the multifunction keys in the following sequence (within 10 seconds):
F1, F4, F3, F2 In order to enable/disable this option see paragraph 8.1. If a wrong password is inserted, the instrument will display an error message and will
repeat the request. If no key is pressed after about 10 seconds the instrument returns back to the original
screen.
EN - 95
SIRIUS89N
11. INSTRUMENT'S MEMORY
By pressing the MENU key the following screen will be displayed:
MENU GENERAL
SAFETY TEST MEMORY
ANALYZER MEMORY RESET
ANALYZER CONFIG RECORDER CONFIG
CONTRAST DATE&TIME LANGUAGE
It’s not possible to enter the MENU during a recording or a Real Time Energy measurement.
11.1. SAFETY TEST MEMORY
Selecting the SAFET YEST MEMORY item and pressing ENTER the instrument display the following screen:
SAFETY TEST MEMORY
MEM TYPE PLACE 001 LOW 003 002 EARTH 003 003 M 004 004 RCD 004 005 LOOP 001
TOT:005 FREE:994
Example of SAFETY TEST
MEMORY screen
LAST
ALL
MEM: Order Number of the measurement TYPE: Measurement TYPE PLACE: Mnemonic parameter associated by User to Measurement TOT: Total Number of Measurement FREE: Available Memory Location
Following keys are enabled:
F1, F2: (to select the Measurement).  F3: to cancel the last recording effected.  F4: to cancel all the recordings effected.  ENTER: to see the measurement results of the selected test  ESC: to quit this mode
EN - 96
SIRIUS89N
11.2. ANALYZER MEMORY
This option permits you to display:
The present content of the instrument memory The size of the memorised data The residual space available for future recordings (expressed in days and hours)
All the stored data can be displayed and analysed only downloading them into a PC with the operating software.
After selecting “ANALYZER MEMORY” from the Main Menu the screen below will be displayed
ANALYZER MEMORY
01 Smp 02.01 01:23 02 Rec 02.01-02.01 03 R&a 02.01-02.01 04 Rec 02.01-02.01 05 R&a 02.01-02.01 06 Rec 04.01-05.01
DATA SIZE:0.11Mb REC TIME: 0d.06h
Example of ANALYZER
MEMORY screen
LAST
ALL
Rec: recordings effected with respective Start and Stop dates expressed in the
format “day.month” (start) – “day.month” (stop) without Voltage Sag and Surge detection.
R&a: recordings effected with respective Start and Stop dates expressed in the
format “day.month” (start) – “day.month” (stop) with Voltage Anomalies (Sag and Surge) detection .
Smp: values of the samples of voltage and current stored by pressing SAVE. DATA SIZE: dimensions of the data saved in the instrument memory.  REC TIME: amount of memory available, calculated on the basis of the parameters
selected for recording, therefore the most complete one (expressed in the format
“days.hours”) to make recordings. The maximum quantity of Rec + R&a + Smp which can be contained by the instrument is 35. Following keys are enabled:  F1, F2: (only if the quantity of Rec+R&a+Smp is higher than 7) to run over all the
recordings stored in the instrument memory.
F3: to cancel the last recording effected.  F4: to cancel all the recordings effected.  ESC: to quit this mode
EN - 97
SIRIUS89N
12. CONNECTING THE INSTRUMENT TO A PC
In order to connect the instrument to a PC you must connect the Optical serial cable code C2001 shipped with the instrument to a PC COM port.
The available transmission speeds are the following:
9600, 19200, 57600 (default value)
The value of the transmission speed (Baud Rate) is displayed on the initial screen (immediately after turning on the instrument, see paragraph 4.2). The value of this parameter can be modified only with the management software.
For download instructions please refer to software help file. In order to transfer the memorized data from the instrument to the PC the following
procedure must be followed (after Sw installation):
1. Switch ON the instrument and wait that Initial screen disappears (the rotary
Switch position isn't relevant).
2. Connect the Optical serial output of the instrument to the serial output of the PC
through the Original C2001 serial cable
3. Run the program
4. Select the "Download" command
5. Refer to software help ON Line for further instructions.
EN - 98
Loading...