SAFETY STEPS TO FOLLOW IF SOMEONE
IS THE VICTIM OF ELECTRICAL SHOCK
DO NOT TRY TO PULL OR GRAB THE lNDl-
VIDUAL
IF POSSIBLE, TURN OFF THE ELECTRICAL
POWER
IF YOU CANNOT TURN OFF THE ELECTRICAL
POWER, PULL, PUSH, OR LIFT THE PERSON TO
SAFETY USING A WOODEN POLE OR A ROPE
OR SOME OTHER INSULATING MATERIAL
SEND FOR HELP AS SOON AS POSSIBLE
AFTER THE INJURED PERSON IS FREE OF CONTACT WITH THE SOURCE OF ELECTRICAL
SHOCK, MOVE THE PERSON A SHORT
DISTANCE AWAY AND IMMEDIATELY START
ARTIFICIAL RESUSCITATION
Page 3
This manual includes copyright material reproduced by permission of the HEWLETT-PACKARD Company.
TM 11-6625-2969-14&P
TECHNICAL MANUAL
HEADQUARTERS
DEPARTMENT OF THE ARMY
No. 1l-6625-2969-14&P)
W
ASHINGTON
, DC,
9 May 1980
OPERATOR’S, ORGANIZATIONAL,
DIRECT SUPPORT AND GENERAL SUPPORT
MAINTENANCE MANUAL
(INCLUDING REPAIR PARTS
AND SPECIAL TOOLS LISTS)
POWER METER TS-3793/U
(HEWLETT-PACKARD MODEL 436A)
(NSN 6625-01-033-5050)
REPORTING OF ERRORS
You can improve this manual by recommending improvements using DA Form 2028-2 located in the back
of the manual. Simply tear out the self-addressed form, fill it out as shown on the sample, fold it where shown,
and drop it in the mail.
If there are no blank DA Forms 2028-2 in the back of your manual, use the standard DA Form 2028
(Recommended Changes to Publications and Blank Forms) and forward to the Commander, US Army Com-
munications and Electronics Materiel Readiness Command, ATTN: DRSEL-ME-MQ, Fort Monmouth, NJ
07703.
In either case a reply will be furnished direct to you.
SERIAL NUMBER
This manual applies directly to instruments with serial numbers prefixed 1606A,
1611A and 1629A.
With changes described in section VII, this manual also applies to instruments with
serial numbers prefixed 1447A, 1448A, 1451A, 1501A, 1503A, 1504A, 1505A,
1538, and 1550A.
For additional important information about serial numbers, see INSTRUMENTS
COVERED BY MANUAL in section I.
This manual is an authentication of the manufacturer’s commercial literature which, through usage, has been found to
cover the data required to operate and maintain this equipment. Since the manual was not prepared in accordance with
military specifications, the format has not been structured to consider levels of maintenance.
This product and related documentation must be
reviewed for familiarization with safety markings
and instructions before operation. This product has
been designed and tested in accordance with international standards.
SAFETY SYMBOLS
Instruction manual symbol: the
product will be marked with this
symbol when it is necessary for
the user to refer to the instruction manual (refer to Section II
of this manual.
Indicates hazardous voltages.
Indicates earth (ground) terminal.
The WARNING sign denotes a
hazard. It calls attention to. a
procedure, practice, or the like,
which, if not correctly performed
or adhered to, could result in
personal injury. Do not proceed
beyond a WARNING sign until
the indicated conditions are fully
understood and met.
The CAUTION sign denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like, which, if not correctly
performed or adhered to, could
result in damage to or destruction of part or all of the product.
Do not proceed beyond a CAUTION sign until the indicated
conditions are fully understood
and met.
SAFETY EARTH GROUND
This is a Safety Class I product (provided with a
protective earthing terminal). An uninterruptible
safety earth ground must be provided from the
main power source to the product input wiring terminals, power cord, or supplied power cord set.
Whenever it is likely that the protection has been
impaired, the product must be made inoperative
and be secured against any unintended operation.
BEFORE APPLYING POWER
Verify that the product is configured to match the
available main power source per the input power
configuration instructions provided in this manual.
If this product is to be energized via an auto-trans-
former make sure the common terminal is connected to the neutral (grounded side of mains
supply).
SERVICING
Any servicing, adjustment, maintenance,
or repair of this product must be performed only by qualified personnel.
Adjustments described in this manual
may be performed with power supplied
to the product while protective covers
are removed. Energy available at many
points may, if contacted, result in personal injury.
Capacitors inside this product may still
be charged even when disconnected from
its power source.
To avoid a fire hazard, only fuses with
the required current rating and of the
specified type (normal blow, time delay,
etc.) are to be used for replacement.
vii
Page 10
Page 11
SECTION 0
INSTRUCTIONS
TM ll-6625-2969-14&P
0-1.
operation and maintenance instructions.
TS-3793/U is referred to as the Hewlett-Packard Model 436A Power Meter.
0-2.
whether there are new editions, changes, or additional publications pertain-
ing to the equipment.
modification work orders (MWO’S) pertaining to the equipment.
0-3.
forms, records, and reports which are to be used by maintenance personnel
at all levels of maintenance are listed in and prescribed by TM
NAVSUPINST 4030.29/AFR 71-13/MCO P4030.29A and DSAR 4145.8.
SCOPE .
This manual describes Power Meter TS-3793/U (fig. 1-1) and provides
Throughout this manual, the
INDEXES OF PUBLICATIONS.
a.
DA Pam 310-4.
b.
DA Pam 310-7.
FORMS AND RECORDS.
a.
Reports of Maintenance and Unsatisfactory Equipment. Maintenance
Refer to the latest issue of DA Pam
Refer to DA Pam 310-7 to determine whether there are
310-4
to determine
38-750.
Report of Packaging and Handling Deficiencies. Fill out and forward
b. Report of Packaging Improvement Report) as prescribed in AR 700-58/
c. Discrepancy in Shipment Report (DISREP) (SF 361). Fill out and
forward Discrepancy in Shipment Report (DISREP) (SF 361) as prescribed in
AR 55-38/NAVSUPINST 4610.33A/AFR 75-18/MCO P4610.19B and DSAR 4500.15.
0-4.
Instructions for preparing EIR’s are provided in TM 38-750, The
Maintenance Management System.EIR’s should be mailed directly to
Commander, US Army Communications and Electronics Materiel Readiness
Command and Fort Monmouth, ATTN:
A reply will be furnished directly to you.
0-5.
shall be in accordance with TM 740-90-1.
0-6.
accordance with TM 750-244-2.
REPORTING OF EQUIPMENT IMPROVEMENT RECOMMENDATIONS (EIR).
EIR’s will be prepared using DA Form 2407, Maintenance Requiest.
Army
DRSEL-ME-MQ, Fort Monmouth, NJ 07703.
ADMINISTRATIVE STORAGE.
Administrative storage of equipment issued to and used by Army
DESTRUCTION OF ARMY ELECTRONICS MATERIEL.
Destruction of Army Electronics materiel to prevent enemy use shall be in
activities
0-1
Page 12
Page 13
Page 14
General Information
Model 436A
1-0
Figure 1-1. HP Model 436A Power Meter and Accessories Supplied
Page 15
Model 436A
General Information
SECTION I
GENERAL INFORMATION
1-1. INTRODUCTION
1-2. This manual provides information pertaining
to the installation, operation, testing, adjustment
and maintenance of the HP Model 436A Power
Meter.
1-3. Figure 1-1 shows the Power Meter with
accessories supplied.
1-4. Packaged with this manual is an Operating
Information Supplement. This is simply a copy of
the first three sections of this manual. This
supplement should be kept with the instrument for
use by the operator. Additional copies of the
Operating
ordered through your nearest Hewlett-Packard
office. The part numbers are listed on the title page
of this manual.
1-5. On the title page of this manual, below the
manual part number, is a “Microfiche” part
number. This number may be used to order
4x6-inch microfilm transparencies of the manual.
The microfiche package also includes the latest
Manual Changes supplement as well as all pertinent
Service Notes.
1-6. SPECIFICATIONS
1-7. Instrument
Table 1-1. These specifications are the performance
standards or limits against which the instrument
may be tested.
1-8. INSTRUMENTS COVERED BY MANUAL
1-9. Power Meter Options 002, 003, 009, 010,
011, 012, 013, 022, and 024 are documented in
this manual. The differences are noted in the
appropriate location such as OPTIONS in Section
I, the Replaceable Parts List, and the schematic
diagrams.
Information Supplement may be
specifications
are listed in
number prefix(es) aslisted under SERIAL
NUMBERS on the title page.
1-11. An instrument manufactured after the
printing of this manual may have a serial prefix
that is not listed on the title page. This unlisted
serial prefix indicates that the instrument is
different from those documented in this manual.
The manual for this instrument is supplied with a
yellow Manual Changes supplement that contains
change
differences.
1-12. In addition to change information, the
supplement may contain information for correcting errors in the manual. To keep this manual
as current and accurate as possible, HewlettPackard recommends that you periodically request
the latest Manual Changes supplement. The
supplement for this manual is keyed to the
manual’s print date and part number, both of
which appear on the title page. Complimentary
copies
Hewlett-Packard.
1-13. For information concerning a serial number
prefix not listed on the title page or in the Manual
Changes supplement, contact your nearest HewlettPackard office.
1-14. DESCRIPTION
1-15. The Power Meter is a precision digitalreadout instrument capable of automatic and
manual measurement of RF and Microwave power
levels. It is designed for interconnection with a
compatible Power Sensor (refer to Table 1-1,
Specifications) to form a complete power measurement system. The frequency and power range of
the system are determined by the particular Power
Sensor selected for use. With the Power Sensors
available, the overall frequency range of the system
is 100 kHz to 18 GHz, and the overall power range
is -70 to +35 dBm.
information
of the supplement are available from
that
documents
the
1-10. This instrument has a two-part serial
number. The first four digits and the letter
comprise the , serial number prefix. The last five
digits form the sequential suffix that is unique to
each instrument. The contents of this manual
apply directly to instruments having the same serial
1-16. Significant operating features of the Power
Meter are as follows:
●
Digital Display:
seven-segment LED, plus a sign when in the
dBm or dB (REL) mode. It also has under- and
The display is a four-digit,
1-1
Page 16
General Information
Model 436A
Table 1-1. Specifications
SPECIFICATIONS
Frequency Range:
100 kHz to 18 GHz (depending on power sensor used).
Power Range:
(display calibrated in watts, dBm, and dB relative to
reference power level).
With 8481A, 8482A, or 8483A sensors:
50 dB with 5
full scale ranges of –20, –10, O, 10, and 20 dBm
(l0pW
to 100 mW).
With 8481H or 8482H sensors:
45 dB with 5 full scale
ranges of 0, 10, 20, 30 and 35 dBm (1 mW to 3W).
With 8484A sensor:
50 dB with 5 full scale ranges of
-60, -50, -40, -30, and -20 dBm (1 nW to
10/Jw).
Accuracy:
Instrumentation1:
Watt mode: ±0.5%.
dBm mode: ±0.02 dB ±0.001 dB/°C.
dB [REL] mode2: ±0.02 dB ±0.001 dB/°C.
Zero:
Automatic, operated by front panel switch.
Zero set:
±0.5% of full scale on most sensitive range.
typical, ±1 count on other ranges.
Zero carry over:
±0.2% of full scale when zeroed on
the most sensitive range.
Noise
(typical, at constant temperature, peak change
over any one-minute interval): 20 pW (8484A);
40 nW (8481A, 8482A, 8483A); 4 pW (8481H,
8482H).
Drift
(1 hour, typical, at constant temperature after
24-hour warm-up); 20 pW (8484A); 10 nW (8481A,
8482A, 8483A); 1.0 0W (8481H, 8482 H).
Power Reference:
Internal 50 MHz oscillator with
Type N Female connector on front panel (or rear
panel, Option 003 only).
Power output: 1.00 mW.
Factory set to ±0.7%, traceable to the National
Bureau of Standards.
Accuracy: ±1.2% worst case (±0.9% rss) for one
year (0°C to 55°C).
Response Time:
(0 to 99% of reading, five time constants)
Range 1 (most sensitive) <10 seconds.
Range 2
Range 3-5
<1 second
<100 milliseconds.
(Typical, measured at recorder output).
Cal Factor:
16-position switch normalizes meter reading to account for calibration factor or effective efficiency,
Range 85% to 100% in 1% steps.
Cal Adjustment:
Front panel adjustment provides capability to adjust
gain of meter to match power sensor in use.
Recorder Output:
Proportional to indicated power with 1 volt corre-
sponding to full scale and 0.316 volts to -5 dB;
kfl
output impedance, BNC connector.
1
RF Blanking Output:
Open collector TTL; low corresponds to blanking
when auto-zero mode is engaged.
Display:
Digital display with four digits, 20% over-range capa-
bility on all ranges. Also, uncalibrated analog peak-
ing meter to see fast changes.
Power Consumption:
100, 120, 220, or 240 V + 5%, -10%, 48 to 440 Hz,
less than 20 watts (<23 watts with Option 022, or
024).
Dimensions:
134 mm High (5-1/4 inches).
213 mm Wide (8-3/8 inches).
279 mm Deep (11 inches).
Net Weight:
4.5 kg (10 lbs).
1-2
1
Includes sensor non-linearity. Add +1.5 -1.0% on top range when using the 8481A, 8482A, or 8483A power sensors.
2
Specifications are for within range measurements. For range-to-range accuracy add the range uncertainties.
Page 17
Model 436A
General Information
DESCRIPTION (cont’d).
overrange indicators.
overrange capability
There is a 20 percent
in all ranges.
Large
10 mm (0.375 inch) digits are easy to see even
in a high glare environment.
● Auxiliary Meter:
Complements the digital
display by showing fast changes in power level.
Ideal for
“peaking”
transmitter output or
other variable power devices.
● Choice of Display in Watts, dBm or dB:
Absolute power can be read out in watts or
dBm. Relative power measurements are made
possibile with the dB [REF] switch. Pressing
this switch zeros the display for any applied
input power and any deviation from this
reference is shown in dB with a resolution of
±0.01 dB. This capability is particularly useful
in frequency response testing.
●
Power Units and Mode Annunciator:
annunciator
provides
error-free
The units
display
interpretation by indicating appropriate power
units in the watt mode. The mode annunciator
indicates the mode of operation: dBm, dB
(REL), ZERO or REMOTE.
● Completely Autoranging: The Power Meter
automatically switches through its 5 ranges to
provide completely
“hands off” operation.
The RANGE HOLD switch locks the Power
Meter in one of its ranges when autoranging is
not desired.
● Automatic Sensor Recognition: The Power
Meter continually decodes the sensitivity of
the Power Sensor to which it is connected.
This information is then used to automatically
control the digital display decimal point
location and, when WATT MODE operation is
selected, to light the appropriate power units
annunciator.
●
Auto Zero:
Zeroing the meter is accomplished
by merely depressing the SENSOR ZERO
switch and waiting until the display shows all
zeros before releasing it. The meter is ready to
make measurements as soon as the zero light
in the mode annunciator goes off.
●
RF Blanking Output:
Open collector TTL; low
corresponds to blanking when the sensor zero
is engaged, ” May be used to remove the RF
input signal connected to the power sensor.
● Calibration Accuracy: A 1.00 mW, 50 MHz
reference output is available at the front panel
for calibrating the Power Meter and the Power
Sensor as a system. Calibration is accomplished using the CAL ADJ and CAL FACTOR
% controls. The CAL ADJ control compensates for slight differences in sensitivity
associated with a particular type of Power
Sensor and the CAL FACTOR % control
compensates for mismatch losses and effective
efficiency over the frequency range of the
Power Sensor.
● Recorder Output:
Provides a linear output
with respect to the input power level. For
each range, a +1.00 Vdc output corresponds
to a full scale input power level. Refer to
Table 1-1, Specifications, for the full-scale
range values associated with the various types
of Power Sensors available.
1-17. Two programming interfaces are available as
options for the Power Meter - a Hewlett-Packard
Interface Bus (HP-IB) Option 022; and a BCD
Interface, Option 024. Both interfaces allow full
remote control of all the power meter functions
(CAL FACTOR can be programmed to either
100% or the CAL FACTOR which has been
manually set on the front panel). These options
may be added by the user at a later time as his
requirements grow.
1-18. OPTIONS
1-19. Input-Output Options
1-20. Option 002.
A rear panel input connector is
connected in parallel with the front panel input
connector.
1-21. Option 003. A rear panel input connector
replaces the standard front panel input connector;
a rear panel POWER REF OUTPUT connector
replaces the standard front panel connector.
1-22. Cable Options
1-23. A 1.5
metre (5 ft.) Power Sensor Cable is
normally supplied. The 1.5 metre cable is omitted
with any cable option. The options and cable
lengths are shown in the table below.
Option
009
010
011
012
013
I
Cable Length
3.0 m (10 ft)
6.1 m (20 ft)
15.2 m (50 ft)
30.5 m (100 ft)
61.0 m (200 ft)
1-3
Page 18
General Information
Model 436A
1-24. Remote Control Options
1-25. Options 022 and 024 add remote interface
capability to the Power Meter. Option 022 is
compatible with the Hewlett-Packard Interface Bus
(AH1, C0, DC2, DT1, L2, LE0, PP0, RL2, SH1,
SR0, T3, TE0); Option 024 uses dedicated
input/output lines to enable remote programming
and to provide parallel, BCD-coded output data.
1-26. Option 022 or 024 may be ordered in kit
form under HP part numbers 00436-60035 and
00436-60034 respectively. Each kit contains a
control assembly printed-circuit board, an input/
output assembly printed circuit board, and a data
cable for interconnection.
1-27. ACCESSORIES SUPPLIED
1-28. The accessories supplied with the Power
Meter are shown in Figure 1-1.
a.
The 1.5 metre (5 ft.) Power Sensor Cable,
HP 00436-60026, is used to couple the Power
Sensor to the Power Meter. The 1.5 metre cable is
omitted with any cable option.
b. The line power cable may be supplied in
one of four configurations. Refer to the paragraph
entitled Power Cables in Section II.
c. An alignment tool for adjusting the CAL
ADJ front panel control (HP Part No. 8710-0630).
1-29. EQUIPMENT REQUIRED BUT NOT
SUPPLIED
1-30. To form a complete RF power measurement
system, a Power Sensor such as the HP Model
8481A must be connected to the Power Meter via
the Power Sensor cable.
1-31.
EQUIPMENT AVAILABLE
1-32.
The HP Model 11683A Range Calibrator is
recommended for performance testing, adjusting,
and troubleshooting the Power Meter. The Power
Meter’s range-to-range accuracy and auto-zero
operation can easily be verified with the Calibrator.
It also has the capability of supplying a full-scale
test signal for each range.
1-33. Two extender boards (HP Part Numbers
5060-0258, and 5060-0990; 24 and 44 pins
respectively) enable the Power Meter printed
circuit assemblies to be accessed for service. Rubber
bumpers (HP Part No. 0403-0115) should be installed on the extender boards to prevent the boards
from touching.
1-34. RECOMMENDED TEST EQUIPMENT
1-35. The test equipment shown in Table 1-2 is
recommended for use during performance testing,
adjustments,
optimum performance of the Power Meter, the
specifications of a substitute instrument must
equal or exceed the critical specifications shown in
the table.
1-36. SAFETY CONSIDERATIONS
1-37. The Power Meter is a Safety Class I
instrument. This instrument has been designed
according to international safety standards.
1-38. This operating and service manual contains
information, cautions, and warnings which must be
followed by the user to ensure safe operation and
to retain the instrument in safe condition.
and troubleshooting. To ensure
1-4
Page 19
Model 436A
General Information
Table 1-2. Recommended Test Equipment
Instrument
Type
Range Calibrator
Digital Voltmeter
Power Meter
Thermistor
Mount
Counter
Oscilloscope
Critical Specifications
Chopped dc output for each range referenced
Function: DC, resistance
Range Resistance: 200 ohms
Vdc: 100 m Vdc, 1000 mVdc, 10 Vdc, 100 Vdc
10MQ
input impedance
6-digit resolution (±0.05% of reading, +0.02% of range)
Range: 1 mW
Transfer Accuracy (input -to-output): 0.2%
SWR: 1.05,50 MHz
Accuracy:
Frequency Range: 220 Hz, 50 MHzHP 5245LA
Sensitivity: 100 m Vrms
Accuracy: 0.01%
Bandwidth: dc to 50 MHzHP 180C/T
Vertical Sensitivity: 0.2 V/division
Horizontal Sensitivity: 1 ms/division
±0.5% at 50 MHz **
to 1 mW range
Suggested
Model
HP 11683A
HP 3490A
HP 432A
HP 478A-H75
1801A/1821A
P,A,T
P,A,T
P, A
P, A
Use *
Logic Analyzer
*P = Performance Tests; A = Adjustments; T = Troubleshooting
sary to install the Power Meter. Covered in the
section are initial inspection, power requirements,
line voltage selection, interconnection, circuit
options, mounting, storage, and repackaging for
shipment.
2-3. INITIAL INSPECTION
2-4.
Inspect the shipping container for damage. If
the shipping container or cushioning material is
damaged, it should be kept until the contents of
the shipment have been checked for completeness
and the instrument has been checked mechanically
and electrically. The contents of the shipment
should be as shown in Figure 1-1. Procedures for
checking electrical performance are given in
Section IV. If the contents are incomplete, if there
is mechanical damage or defect, or if the instrument does not pass the electrical performance test,
notify the nearest Hewlett-Packard office. If the
shipping container is damaged, or the cushioning
material shows signs of stress, notify the carrier as
well as the Hewlett-Packard office. Keep the
shipping materials for the carrier’s inspection.
2-5. PREPARATION FOR USE
2-6. Power Requirements
2-7. The Power Meter requires a power source of
100, 120, 220, or 240 Vac, +5%, -0%, 48 to
440 Hz single phase. Power consumption is approximately 20 watts.
2-9. Figure 2-1 provides instructions for line
voltage and fuse selection. The line voltage
selection card and the proper fuse are factory
installed for 120 Vac operation.
SELECTION OF OPERATING VOLTAGE
1. Open cover door, pull the FUSE PULL lever
end rotate to left. Remove the fuse.
2. Remove the Line Voltage Selection Card.
Position the card so the line voltage appears
at top-left corner. Push the card firmly into
the slot.
3. Rotate the Fuse Pull lever to its normal position. Insert a fuse of the correct value in the
holder. Close the cover door.
Figure 2-1. Line Voltage Selection
If this instrument is to be energized via an
autotransformer for voltage reduction,
make sure the common terminal is
connected to the earthed pole of the
power source.
2-8. Line Voltage Selection
BEFORE SWITCHING ON THIS IN-
STR UMENT, make sure the instrument is
set to the voltage of the power source.
2-10. Power Cable
BEFORE SWITCHING ON
THIS IN-
STRUMENT, the protective earth termi-
nals of this instrument must be connected
to the protective conductor of the (mains)
power cord. The mains plug shall only be
inserted in a socket outlet provided with a
protective earth con tact. The protective
action must not be negated by the use of
an extension cord (power cable) without a
protective conductor (grounding).
2-1
Page 22
Installation
Model 436A
Power Cable (cont’d)
2-11. In accordance with international safety
standards, this instrument is equipped with a threewire power cable.
When connected to an
appropriate ac power receptacle, this cable grounds
the instrument cabinet. The type of power cable
plug shipped with each instrument depends on the
8481H Power Sensors provide identical full scale
outputs in response to input signal levels of 100
milliwatts and 3 watts, respectively.
erence in their sensitivity codes is detected by the
Power Meter, however, and the Power Meter digital
readout is automatically configured to indicate the
appropriate value.
country of destination. Refer to Figure 2-2 for the
part numbers of the power cable plugs available.
2-16. Hewlett-Packard Interface Bus Option 022.
Interconnection data for Hewlett-Packard Interface
Bus Option 022 is provided in Figure 2-3. Power
Meter programming and output data format is
described in Section III, Operation.
2-17. BCD Interface Bus Option 024. Inter-
connection data for BCD Interface Option 024 is
provided in Figure 2-4. Power Meter programming
8120-1378
8120-1351
and output data format is described in Section III,
Operation.
nectors for Options 022 and 024 are indicated in
Figures 2-3 and 2-4, respectively.
The dif-
Interface mating con-
8120-1369
8120-1689
2-20. Coaxial Connectors. Coaxial mating con-
nectors used with the Power Meter should be US
MIL-C-39012-compatible type N male or 50-ohm
Figure 2-2. Power Cable HP Part Numbers
Versus Mains Plugs Available
2-12. Circuit Options
2-13. Jumper options are available for selecting a
BNC male.
2-21. Operating Environment
2-22. The operating environment should be within
the following limitations:
filtered or unfiltered dc RECORDER OUTPUT,
for changing the TALK and LISTEN addresses
when Hewlett-Packard Interface Bus Option 022 is
installed, and for selecting the desired pro-
Temperature . . . . . . . . . . . . 0°C to +55°C
Humidity . . . , . . . . . . . . . <95% relative
Altitude . . . . . . . . . .
gramming of the SENSOR ZERO function when
BCD Interface Option 024 is installed. Table 2-1
lists the factory installed jumper connections and
indicates how they may be reconnected to select
the options.
2-23. Bench Operation
2-24. The instrument cabinet has plastic feet and a
fold-away tilt stand for convenience in bench
operation. (The plastic feet are shaped to ensure
self-aligning of the instruments when stacked. ) The
2-14. Interconnections
2-15. Power Sensor.
For proper system operation,
the Power Sensor must be connected to the Power
Meter using either the Power Sensor cable supplied
with the Power Meter or any of the optional Power
Sensor cables specified in Section I. Each of these
cables employs a sensitivity line to enable the
Power Meter to determine the operating range of
the Power Sensor and thus, the true value of the
input signal. For example, the 8481A and
tilt stand raises the front of the instrument for
easier viewing of the control panel.
2-25. Rack Mounting
2-26. Instruments that are narrower than full rack
width may be rack mounted using Hewlett-Packard
sub-module cabinets. If it is desired to rack mount
one Power Meter by itself, order half-module kit,
HP Part Number 5061-0057. If it is desired to rack
mount two Power Meters side by side, order the
following items:
<4570 m (15,000 ft)
2-2
Page 23
Model 436A
Installation
Rack Mounting (cont’d)
Rack Mount Flange Kit (two provided)
HP Part Number 5020-8862.
Front Horizontal Lock Links (four pro-
b.
vided) HP Part Number 0050-0515.
Rear Horizontal Lock Links (two pro-
vided HP Part Number 0050-0516.
2-27 In addition to the rack mounting hardware,
a front handle assembly (two provided) is also
available for the Power Meter. The part number is
HP 5060-9899.
2-28. STORAGE AND SHIPMENT
2-29. Environment
2-30. The instrument should be stored in a clean
dry environment. The following environmental
limitations apply to both storage and shipment:
Temperature . . . . . . . . . . . -40°C to +75°C
Humidity . . . . . . . . . . . . . <95% relative
Altitude . . . . . . . . . .
2-31. Packaging
2-32. Original Packaging.
<7620 m (25,000 ft)
Containers and materials
identical to those used in factory packaging are
available through Hewlett-Packard offices. If the
instrument is being returned to Hewlett-Packard
for servicing, attach a tag indicating the type of
service required,
return address, model number,
and full serial number. Also mark the container
FRAGILE to assure careful handling. In any
correspondence refer to the instrument by model
number and full serial number.
2-33. Other Packaging.
The following general
instructions should be used for re-packaging with
commercially available materials:
Wrap the instrument in heavy paper or
a.
plastic. (If shipping to a Hewlett-Packard office or
service center, attach a tag indicating the service
required, return address, model number, and full
serial number.)
b. Use a strong shipping container. A doublewall carton made of 275-lb test material is
adequate.
c.
Use enough shock-absorbing material (3
to 4-inch layer) around all sides of instrument to
provide firm cushion and prevent movement in the
container.
Protect the control panel with
cardboard.
d. Seal the shipping container securely.
Mark the shipping container FRAGILE to
e.
assure careful handling.
2-3
Page 24
Installation
Model 436A
Table 2-1. Circuit Options
Assembly
A-D Converter
Assembly A3
BCD Interface
Control Assembly
A6 (Option 024)
Hewlett-Packard
Interface Bus Control Assembly A6
(Option 022)
Service
Sheet
8
13
11
Jumper Functions
The factory-installed jumpers provide a filtered dc RECORDER OUTPUT
which corresponds to the average power input to the Power Sensor. If external filtering is desired, reconnect the jumpers to provide the optional unfiltered dc RECORDER OUTPUT as shown on Service Sheet 8.
The factory-installed jumper enables the SENSOR ZERO function to be
programmed only when the REMOTE ENABLE input to the Power Meter
is low. If it is desired to program the SENSOR ZERO function independently of the remote enable input, reconnect the jumper to provide +5 V
to U12C-9 as shown on Service Sheet 8.
The factory installed jumpers select TALK address M and LISTEN address
-
(minus sign) for the Power Meter. As shown on Service Sheet 11, either
of these addresses causes a high enable output at U2C-10. If it is desired
to change these addresses, refer to Service Sheet 11 and Table 2-2 and
reconnect the jumpers to decode the appropriate ASCII characters. For
example, to change to TALK address E and LISTEN address 70, the
jumpers would be reconnected as follows.
ASCII code (logic 1=0V)
DDDDDDD
1111111
0000000
7654321
M
E
-
%
Jumpers
M
-
1001101
1000101
0101101
0100101
U1B-13
U1B-12
U1B-10
U1B- 9
U2C- 9
HI01
LI02
HI03
HI04
HI05
Note: DI07 and DI06 must always be
1 and 0, respectively, for TALK address.
Note: DI07 and DI06 must always be
0 and 1, respectively, for LISTEN address.
E, % , Disconnect jumper from HI04
and reconnect to LI04.
2-4
Page 25
Model 436A
Installation
Table 2-2. USA Standard Code for Information Interchange (ASCII)
The Hewlett-Packard Interface Bus logic levels are TTL compatible,
i.e., the true (1) state is 0.0 Vdc to 0.4 Vdc and the false (0) state is
+2.5 Vdc to +5.0 Vdc.
Programming and Output Data Format
Refer to Section III, Operation.
Mating Connector
HP 1251-0293; Amphenol 57-30240.
Mating Cables Available
HP 10631A, 1.0 metre (3 ft.); HP 10631B, 2.0 metres (6 ft.)
HP 10631C, 4.0 metres (12 ft.); HP 10631D, 0.5 metre (1.5 ft.)
Cabling Restrictions
A Hewlett-Packard Interface Bus System may contain no more
1.
than 1.8 metres (6 ft.) of connecting cable per instrument.
2.
The maximum accumulative length of connecting cable for any
Hewlett-Packard Interface Bus System is 20.0 metres (65.6 ft.)
2-6
Figure 2-3. Hewlett-Packard Interface Bus Connection
Page 27
Model 436A
Installation
Logic Levels
The BCD Interface logic levels are TTL
compatible, i.e., the true state is 0.0 Vdc to
0.4 Vdc and the false state is +2.5 Vdc to
+5.0 Vdc.
Figure 2-4. BCD Interface Connection
Programming and Output Data Format
Refer to Section III, Operation
Mating Connectors
Mating Cables Available
- HP 1251-0086
- HP 562A-16C
for 5055A Printer
2-7/2-8
Page 28
Page 29
Model 436A
Operation
SECTION Ill
OPERATION
3-1. INTRODUCTION
3-2. This section provides complete operating
information for the Power Meter. Included in the
section are a description of all front- and rear-panel
controls,
features), operator’s checks, operating instructions,
power measurement accuracy considerations, and
operator’s maintenance.
3-3. Since the power Meter can be operated locally
as well as remotely via Hewlett-Packard Interface
Bus Option 022 or BCD Remote Interface Option
024, respectively, the information in this section is
arranged accordingly. All information unique to a
particular operating configuration is designated as
such; where no distinction is made, the informaion is applicable to both standard and optional
instrument operation.
3-4. PANEL FEATURES
3-5. Front and rear panel features of the Power
Meter are described in Figure 3-1. This figure
contains a detailed description of the controls,
connectors and indicators.
3-6. OPERATOR’S MAINTENANCE
3-7. The only maintenance the operator should
normally perform is replacement of the primary
power fuse located within Line Module Assembly
A11. For instructions on how to change the fuse,
refer to Section II, Line Voltage Selection.
Make sure that only fuses with the required
rated current and of the specified
connectors, and
indicators (panel
type (normal blow, time delay, etc.) are
used for replacement. The use of repaired
fuses and the short-circuiting of fuseholders must be avoided.
3-8. OPERATOR’S CHECKS
3-9. A procedure for verifying the major functions of the Power Meter is provided in Figure 3-2.
parts: Local Operation, Remote BCD Operation,
and Remote Hewlett-Packard Interface Bus Operation. For a standard instrument it is only necessary
to perform the Local Operation procedure. For
units equipped with either of the remote options,
the Local Operation procedure should be performed first to establish a reference against which
remote operation can be verified. Information
covering remote programming of the Power Meter
is provided in the following paragraphs, and a
Hewlett-Packard Interface Bus Verification Program is provided in Section VIII, Service.
3-10. LOCAL OPERATING INSTRUCTIONS
The procedure is divided into three
3-11. Figure 3-3 provides general instructions for
operating the Power Meter via the front-panel
controls.
Any
interruption
of the protective
(grounding) conductor (inside or outside
the instrument) or disconnecting the
protective earth terminal is likely to make
this instrument dangerous. Intentional
interruption is prohibited.
3-1
Page 30
Operation
Model 436A
FRONT AND REAR PANEL FEATURES
3-2
Figure 3-1. Front and Rear Panel Controls, Connector, and Indicators (1 of 4)
Page 31
Figure 3-1.
3-3
Page 32
Operation
Model 436A
FRONT AND REAR PANEL FEATURES
3-4
Figure 3-1. Front and Rear Panel Controls, Connector, and lndicators (3 of 4)
Page 33
Figure 3-1.
Figure 2-1
Table 2-2
3-5
Page 34
Operation
LOCAL OPERATION
Model 436A
OPERATOR’S CHECKS
3-6
BEFORE CONNECTING LINE POWER TO THIS INSTRUMENT, ensure that all devices connected to this instrument
are connected to the protective (earth) ground.
BEFORE SWITCHING ON THIS INSTRUMENT, ensure that
the line power (mains) plug is connected to a three-conductor
line power outlet that has a protective (earth) ground. (Ground-
ing one conductor of a two-conductor outlet is not sufficient.)
Figure 3-2. Operator’s Checks (1 of 10)
Page 35
Figure 3-2.
Section II
3-7
Page 36
Figure 3-2.
3-8
Page 37
Figure 3-2.
3-9
Page 38
Operation
Model 436A
OPERATOR’S CHECKS
LOCAL OPERATION
(cont’d)
3-10
25. Set the
that the
26. Set the POWER REF switch to ON (in) and adjust the
Readout indictes –2.00 dBm.
POWER REF switch to off (out). Verify that the
Digital Readout blanks
Figure 3-2. Operator’s Checks (5 of 10)
UNDER RANGE lamp lights and
CAL ADJ control so that the
Digital
Page 39
Figure 3-2.
3-11
Page 40
Operation
REMOTE BCD OPERATION
Model 436A
OPERATOR’S CHECKS
(cont’d)
3-12
8.
Set the POWER REF switch to ON. Verify that the
Digital Readout blanks (1_ . _ _
pW).
OVER RANGE lamp lights and the
NOTE
Underscore (_) indicates blanked digit.
9.
Program the Power Meter to Range 3. Verify that the mW lamp lights and that the OVER
RANGE lamp goes out.
Figure 3-2. Operator’s Checks (7 of 10)
Page 41
Figure 3-2.
3-13
Page 42
Figure 3-2.
3-14
Page 43
Figure 3-2.
Section VIII
3-15
Page 44
Figure 3-3.
3-16
Page 45
Figure 3-3.
Section II
3-17
Page 46
Figure 3-3.
3-18
Page 47
Figure 3-3.
Section II
3-19
Page 48
Operation
Model 436A
3-12. HEWLETT-PACKARD INTERFACE BUS
REMOTE OPERATION
NOTE
For a quick and easy programming guide
see Figure 3-8; for detailed information
study paragraphs 3-12 through 3-61.
3-13. Hewlett-Packard Interface
Bus (HP-IB)
Option 022 adds remote programming and digital
output capability to the Power Meter. For further
information about the HP-IB, refer to IEEE Standard 488 and the Hewlett-Packard Catalog. Power
Meter compatibility, programming, and data format is described in detail in the paragraphs which
follow.
3-14. Compatibility
3-15. The Power Meter controls that can be programmed via the Hewlett-Packard Interface Bus
are the MODE and SENSOR ZERO switches. The
controls not programmable are the POWER REF
and LINE switches. The CAL FACTOR % switch
can be enabled and disabled via the interface bus
but, when enabled, the calibration factor entered
at the front-panel of the Power Meter is used.
takes to make a valid measurement, study the rest
of the information in this section.
3-21. Receiving Data Messages
3-22. The Power Meter is configured to listen (receive data) when the controller places the interface
bus in the command mode (ATN and REN lines
low; IFC line high) and outputs listen address “-”
(minus sign). The Power Meter then remains configured to listen (accept programming inputs when
the interface bus is in the data mode) until it is unaddressed by the controller. To unaddress the
Power Meter, the controller can either send the
Abort Message (set tine IFC line low) or send the
Local Message (set the REN line high), or it can
place the interface bus in the command mode and
generate a universal unlisten command.
3-23. Data Input Format. The Power Meter does
not require any particular data input format. It is
capable of responding to each of the programming
codes listed in Table 3-3 on an individual basis.
Because it responds to these codes in the order it
receives them, we recommend that the code for
measurement rate be sent last.
3-16. In addition, specific ranges can be set and
various triggering options are available to the programmer. This will be described in detail later.
3-17. The programming capability of the Power
Meter will be described in terms of the twelve bus
messages found in Table 3-1.
3-18. Data Messages
3-19. The Power Meter communicates on the bus
primarily through data messages. It receives data
messages that tell it what range to use, what mode
to use, whether or not cal factor should be enabled, and what the measurement rate should be. It
sends data messages that tell the measurement
value, the mode and range the value was taken at,
and what the instrument’s status (see Table 3-4)
was when it took the measurement.
3-20. Table 3-2 outlines the key elements involved
in making a measurement. Indeed the Power Meter
can be programmed to make measurements via the
HP-IB by following only the sequence suggested in
the table, and briefly referring to Tables 3-3, 3-4,
(input and output data), and Fig. 3-8. However, to
take advantage of the programming flexibility built
into the Power Meter and minimize the time it
3-24. Program Codes. Table 3-3 lists the program
codes that the Power Meter responds to and the
functions that they enable. In the listen mode, the
Power Meter can handshake in 0.5
IJS.
The time required for the Power Meter to respond to the programming command, however, depends on where
the Power Meter is in the operating program (see
Figure 3-6). The overall worst case time for Power
Meter response to a programming command is
2.5 seconds, the minimum response time is approximately 100 microseconds.
NOTE
In addition to the program codes listed
in Table 3-3, Power Meter operation will
be affected by all other program codes
shown in columns 2, 3, 4, and 5 of Table
2-2, except (SP!”#$%&*). Thus care
should be taken to address the Power
Meter to unlisten before sending these
programming commands to other instruments on the interface bus.
3-25. Programming the Range. Remote range pro-
gramming is slightly different than Local range
selection. For Local operation the Power Meter
auto-ranges. For Remote operation, the program
codes have provision for direct selection of the de-
3-20
Page 49
Model 436A
Operation
Table 3-1. Message Reference Table
Message and Identification
Data
Trigger (DTO)No
Clear (DC2)
Remote (RL2)
Local (RL2)
Local Lockout (RL2)
Applicable
Yes
Yes
No
Yes
Yes
No
No
Command and Title
T3 Talker, L2 Listener,
AH1 Acceptor Handshake
SH1 Source Handshake.
Device Trigger
DCL Device Clear
SDC Selected Device Clear
REN Remote Enable
REN Remote Disable
GTL Go to LocalPower Meter does not respond to GTL
REN Remote DisablePower Meter does not respond to LLO
Power Meter changes mode, range, measurement rate, and Cal Factor enable or disable.
It outputs status and measurement data.
The Power Meter does not respond to a
Group Execute Trigger. However, remote
trigger capability is part of the Data message (measurement rate).
Upon receipt of DCL command, Power
Meter functions are set for Watt Mode,
Auto Range, Cal Factor Disable and Measurement rate Hold.
Power Meter goes to remote when addressed
to listen, and REN is true (low).
Power Meter goes to local when REN is
false (high).
command.
command.
Response
Clear Lockout/
Set Local (RL2)
Pass Control/Take
Control
Require Service
Status Byte
Status Bit
Abort
(C(3)
(SR~
(PP@)
Yes
No
No
No
No
Yes
Complete HP-IB capability as defined in IEEE Std. 488 is AH1, CO, DC2, DTO,
LEO, PPO, RL2, SH1, SR0, T3, TEO.
REN Remote Disable
ControllerPower Meter cannot act as bus controller.
SRQ Service Request
SPE Serial Poll Enable
SPD Serial Poll Disable
PP Parallel Poll
IFC Interface Clear
NOTE
Returns all devices on bus to local
operation.
Power Meter does not request service.
Power Meter does not respond to a
Serial Poll
Power Meter does not respond to a parallel
poll.
Power Meter stops talking or listening.
3-21
Page 50
Operation
Model 436A
Table 3-2. Measurement Sequence
MEASUREMENT SEQUENCE
Event 1
Event 2
Event 3
Event 4
Event 5
{controller talk and Power Meter listen}, {Program Codes}
See controller manual.Program codes to configure one or more of the
Power Meter Listen address
factory set to “-” (see
Tables 2-1 and 2-2).
e.g., CMD "?U-","9D+V"
wrt “pmrd”, “9D+V”
following (see Table 3-3):
1. Range
2. Remote mode (Watt, dBm, dB [Ref]
3. Cal Factor
4. Measurement Rate (and trigger)
Response time for meter’s digital (operating program) circuitry (see Table 3-5 and Figures 3-5 and 3-6).
Meter takes measurement; data available.
Additional delay to allow analog circuits to settle; necessary only if on Range 1 (most sensitive) or if
settling time measurement rates are not being used (see Figure 3-4). Here are some suggestions: *
1.
Load reading into controller (event five) and check data string for range (look at character number 1
or check measured value).
2.
If Power Meter is on Range 1, wait 10 seconds and take another reading.
3.
If settling time measurement rates are being used and meter is not on Range 1, use the first reading.
If settling time measurement rates are not being used, determine the range and branch to an appro-
4.
priate delay: Range 2, one second; Ranges 3-5, 0.1 second.
{universal unlisten, controller listen and Power Meter talk} , {variable name}
See controller manual. Power Meter Talk address factory set to “M” (see Tables 2-1 and 2-2).
*There are other ways to ensure that readings are not affected by analog circuit settling time. Also,
these recommended delays are worst case. A thorough understanding of the material in this section
will allow you to optimize measurement time for your particular application. For example, if the
power level is not changing, the controller can average at least two consecutive readings to see if the
result is still settling.
EXAMPLE PROGRAM SEQUENCE:
Line 1 {controller talk and power meter listen}, “9D+T”
Measurement Rate: Trigger with settling time.
Cal Factor Disable (100%)
dBm
Mode
Auto Range
Line 2 {universal unlisten, controller listen and power meter talk} , {variable name}
Power meter outputs measured value to controller.
Line 3
Line 4
Line 5
Controller checks value in variable for Range 2 threshold (e.g., <-20 dBm for Model 8482A
Power Sensor). If value is below threshold, program branches to line 4. If value is above thres-
hold, program branches to line 5.
{wait 10 seconds, then go to line 1} .
{continue}.
3-22
Page 51
Model 436A
Operation
Table 3-3. Hewlett-Packard Interface Bus
Input Program Codes
Function
Range
Least sensitive
Most sensitive
Auto
MODE
Watt
dB (Rel)
dB [Ref]
dBm
Sensor auto-zero
CAL FACTOR
Disable (100%)
Enable (front-panel
switch setting)
Measurement Rate
Hold
Trigger with set-
ling time
Trigger, immediate
Free Run at maxi-
mum rate
Free Run with set-
ling time
Program Codes
ASC IIDECIMAL
5
4
3
2
1
9
A
B
C
D
Z
+
–
H
T
I
R
V
53
52
51
50
49
57
65
66
67
68
69
43
45
72
84
73
82
86
program steps that should be written. Specific
examples are provided later in this Section. (Refer
to Tables 3-3 and 3-4 for Power Meter input and
output strings. Refer to controller manual for
programming syntax.)
3-28. Programming Cal Factor. While the setting
of the front panel CAL FACTOR switch cannot be
remotely changed, the programmer does have a
choice. If CAL FACTOR enable is programmed,
then the Power Meter uses the Cal Factor set by
the switch. If CAL FACTOR Disable is programmed, then the Power Meter uses a Cal Factor of
100%, but the program can correct for cal factor
by computing the corrected reading from the
actual reading and the cal factor (a Cal Factor table
must be stored in an array).
3-29. Programming Measurement Rate. A feature
that is only available via remote programming is
selection of standby, triggered, or free running operation of the Power Meter. (During Local operation, the Power Meter is allowed to free run with
approximately 133 milliseconds allowed for settling time between measurements.) The specific re-
mote triggering capabilities are:
a. HoId (H) - when the power meter is pro-
grammed to Hold, it is inhibited from taking
measurements and from outputting data. Thus, it
is set to a predetermined reference condition from
which a measurement can be triggered synchron-
ously to some external event.
sired range as well as for selection of the autorange
function.
b. Trigger Immediate (I) - this programming
command directs the Power Meter to make one
measurement and output the data in the minimum
possible time, then to go into Hold until the next
3-26. Programming the Mode. Remote mode pro-
gramming is similar to Local mode selection. The
sequence shown in Example 1 is recommended for
taking dB (Rel) readings from a dB [Ref] reference.
triggering command is received. It does not allow
settling time prior to the measurement.
c. Trigger with Delay (T) - this trigger com-
mand is identical to the trigger immediate com-
3-27. Programming Auto-Zero. The Power Meter is
remotely zeroed the same way it is zeroed in local.
Example 2 ahown on the next page outlines the
EXAMPLE 1 (dB Rel/dB Ref)
1 {controller talk and Power Meter listen}, “CT”
{controller talk and Power Meter listen}, “BT”Takes first reading relative to set reference
2
{universal unlisten, controller listen and Power Meter talk}, {Variable name} Power Meter outputs reading to controller
3
{controller talk and Power Meter listen} , “T”
4
5 {universal unlisten, controller listen and Power Meter talk}, {Variable name} Power Meter outputs reading to controller
mand except that it causes the Power Meter to
execute a settling-time delay subroutine before
taking a measurement and outputting data.
Sets reference at present RF input level.
Takes subsequent readings
3-23
Page 52
Operation
Model 436A
Receiving Data Messages (cont’d)
d. Free run at maximum rate (R) - this pro-
gramming command is normally used for asynchronous operation of the Power Meter. It directs the
Power Meter to continuously take measurements
and output data in the minimum possible time. It
does not allow settling time prior to each measure-
ment.
e. Free run with delay (V) - this program-
ming command is identical to the previous command except that it causes the Power Meter to execute a settling-time delay subroutine prior to each
measurement.
3-30. When programming the Power Meter for
synchronous triggered operation, there are two
factors that the programmer must consider to
ensure the validity of the output measurement
data. The first factor is the time that it takes the
Power Meter to respond to a full scale change in
input power level. A typical Power Meter response
curve is shown in Figure 3-4. By comparing this
curve with the measurement timing cycle shown in
Figure 3-5 and summarized in Table 3-5, the
validity of the Power Meter output can be tabu-
lated according to operating range and triggering
interval versus change in input power level. A general summary of this information is as follows:
a. When the Power Meter is programmed for
trigger with settling time operation, sufficient time
is provided for the Power Meter to settle to the input power level on all ranges except Range 1
(most sensitive range). On Range 1 approximately
10 seconds (9-10 measurements) are required for
the Power Meter to settle to the input power level.
b. When the
trigger immediate
Power Meter is programmed for
operation, the desired amount of
settling time can be incorporated into the program.
3-31. Programming the Local to Remote Mode
Change.
The second factor that must be considered
when programming the Power Meter for synchron-
ous triggered operation is whether the first trigger
is sent immediately after terminating local opera-
tion. As illustrated in Figure 3-6, the Power Meter
will not respond to the first trigger following a
local to remote transition until it completes the
previously initiated measurement and display
cycle. Thus, the first data output of the Power
Meter may not be valid. The options available to
the programmer are:
1.
Send a trigger command (Data Message)
and discount the first data output. Upon
outputting the data, the Power Meter will
go to Hold and operate synchronously
starting with the next trigger command.
Wait approximately 2.5 seconds after plac-
2.
ing the Power Meter in remote and sending
the first program trigger command (Data
Message).
3.
Send a Clear Message (DCL) immediately
after placing the Power Meter in remote.
This will
restart the Power Meter operating
program.
3-32.
3-33.
Sending
Meter
The TALK ONLY/NORMAL switch
Data Messages from the Power
(see Figure 3-3) enables the Power Meter to func-
EXAMPLE 2 (Auto Zero)
1
Remove RF power from power sensor (or set it at least 20 dB below the lowest range of the sensor).
2 {controller talk and Power Meter listen}, “Z1T”
Send zero trigger program codes.
3 {universal unlisten, controller listen and Power Meter talk} , {variable name} Read measured value data from meter
(characters 4, 5, 6, and 7).
If absolute value of measured data is not <2 (0000 ± 0002) then branch to step 2; if it is, then continue. (Although
4
this step averages three seconds, it may take as long as 10 seconds to execute.)
5 {controller talk and Power Meter listen}, “9+Dl”
Send normal measurement mode program
codes.
6 {universal unlisten, controller listen and Power Meter talk} , {variabie name} Read status character (number 0) from
meter’s output data string.
7
Check status character for an auto zero loop enabled condition (character 0> decimal 84). If loop is enabled then
branch to step 5. If not, then continue. (This step takes approximately four seconds to execute.)
3-24
Page 53
Model 436A
Operation
Sending Data Messages (cont’d)
tion as a basic talker or in the talk only mode. If
the basic talker function is selected, the Power
Meter is configured to talk when the controller
places the interface bus in the command mode and
outputs talk address M. The Power Meter then remains configured to talk (output data when the interface bus is in the data mode), until it is unaddressed to talk by the controller. To unaddress the
Power Meter, the controller can either send an
Abort Message (generate an interface clear), or it
can place the interface bus in the command mode
and output a new talk address or a universal untalk
command. Examples of addressing and unaddressing the Power Meter to talk are provided in Table
3-2 and Figure 3-8.
3-34. Talk Only Mode. When the Power Meter
functions in the Talk Only Mode, it is automatically configured to TALK when the interface bus
is in the Data Mode and there is at least one listener. Since there can only be one talker at a time per
interface bus, this function is normally selected
only when there is no controller connected to the
system (e.g., when the Power Meter is interconnected to an HP 5150A recorder).
Table 3-4. Hewlett-Packard Interface Bus
Output Data String
Definition
Measured value valid
Watts Mode under Range
Over Range
Under Range dBm or dB
[REL] Mode
S
Power Sensor Auto Zero
Loop Enabled; Range 1
T
Under Range (normal for auto
A
T
zeroing on Range 1)
U
S
Power Sensor Auto Zero Loop
Enabled; Not Range 1, Under
Range (normal for auto zeroing
on Range 2-5)
Power Sensor Auto Zero Loop
Enabled; Over Range (error
condition - RF power applied
to Power Sensor; should not be)
Most Sensitive
R
A
N
G
E
Least Sensitive
1
2
3
4L
5
Character
ASC II
P
Q
R82
S
T84
U
V
I
J
K
M
Decimal
80
81
83
85
86
73
74
75
76
77
3-35. Output Data Format.
The output data
format of the Power Meter is shown and described
in Table 3-4.
3-36. The output data is a fourteen character
string that is provided once at the end of each
measurement cycle. It is a good idea to read at
least part of this string into the controller after
each measurement cycle, even if it will not be used.
This will avoid the possibility of incorrect data
being read after some future measurement.
3-37. The string begins with a status character
and ends with a carriage return and a line feed.
Measured value is formatted as a real constant:
plus or minus four digits (leading zeros not sup-
pressed) followed by an exponential multiplier.
The decimal point is not provided because it is
understood that it follows the four “measured
value” digits. The two-digit exponent is always
negative.
3-38. Data Output Time. Figure 3-6 provides a
simplified flow chart of Power Meter operation.
As shown in the figure, the Power Meter operates
according to a stored program and can only output
M
Watt
O
dB REL
D
dB REF (switch pressed)
E
dBm
S
I
space (+)
G
- (minus)
N
D
I
G
I
T
A
B
C
D
SP
-
0
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6
7
8
9
65
66
67
68
32
45
48
49
50
51
42
53
54
55
56
57
3-25
Page 54
Operation
Model 436A
Table 3-5. Power Meter Remote Access Time to First Output Data Character
Measurement
Triggering
Free Run at maximum rate,
Trigger immediately
Mode
WATT
dBm
dB (REL)
db [REF]
Range 1 or 2
70 ms
90 ms
160 ms
160 ms
Worst Case Access
Range 3,4 or 5
70 ms
90 ms
160 ms
160 ms
Time to First Output Character
Auto Range
Compute measurement times from Figure 3-5 and
add measurement time of each range that Power
Meter steps through to delay time listed below.
From ToDelay
2 1070 ms
1
1 1070 ms
2
23
Examples: Starting at block labeled “HOLD” in
Figure 3-5, worst case access time for range 1-3,
and range 3-1 changes with WATT MODE se-
lected are:
Range 1
1-2 Delay 1070 ms
Range 253 ms
2-3 Delay 133 ms
Range 3
133 ms
70 ms
53 ms
1379 ms2256 ms
From To
3
4
54
Range 3
3-2 Delay 1070 ms
Range 233 ms
2-1 Delay 1070 ms
Range 1
Delay
2 1070 ms
3,5 133 ms
133 ms
50 ms (33+17)
33 ms
Free Run with settling time or Trigger
with settling time.
WATT
dBm
dB (REL)
db [REF]
1130 ms
1130 ms
1200 ms
160 ms
190 ms
190 ms
260 ms
160 ms
Compute worst case Auto Range access times
from Figure 3-5.
Examples: Starting at block labeled “HOLD” in
Figure 3-5; worst case access times for range 1-3
Figure 3-6. Operating Program Simplified Flow Chart
3-28
are associated
decision is
trigger is
reverts
to
yes until
triggering,
withbothhold
always yes. When
received
and
receipt of
it
wiEl
provideoutPut
and externaE triggered
triggeredoperation
continuing
the
next
unitl
the
trigger. Thus,
data OIIIY
is
dig-
Page 57
Model 436A
Operation
Sending Data Messages (cont’d)
data after taking a measurement. Thus, when the
interface bus is placed in the data mode after the
Power Meter has been addressed to talk, the time
required to access the first output data character
depends on where the Power Meter is in the
operating program, and on how the Power Meter
has been previously programmed (see Programming
Codes above.) Worst case access times for each of
the Power Meter operating configurations are listed
in Table 3-5.
3-39. After the first output character is sent, the
remaining characters are sent at either a 10-kHz
rate (infinitely fast listener) or at the receive rate
of the slowest listener.
3-40. Receiving the Trigger Message
3-41. The Power Meter has no provision for re-
sponding to a Trigger Message (bus command GET).
Power Meter triggering is done with the Data Mes-
sage (through the Measurement Rate Program
Codes).
3-42. Receiving the Clear Message
3-48. Receiving the Local Lockout and Clear
Lockout Set Local Messages
3-49. The Power Meter does not respond to the
Local Lockout Message (LLO bus command). It
responds to the Clear Lockout/Set Local Message
in that when the REN bus line goes false, it will
revert to local operation.
3-50. Receiving the Pass Control Message
3-51. The Power Meter has no provision for operation as a controller.
3-52. Sending the Required Service Message
3-53. The Power Meter does not have provision
for requesting service.
3-54. Sending the Status Byte Message
3-55. The Power Meter does not respond
to a
Serial Poll.
3-56. Sending the Status Bit Message
3-57. The Power Meter does not respond
to a
Parallel Poll.
3-43. The Power Meter has provision for responding to the DCL bus command but not the SDC bus
command. Upon receipt of the DCL command, the
Power Meter operating program is reset causing the
Power Meter to enter the Hold state shown at the
top of Figure 3-6, and the HP-IB circuits are configured to provide Watt Mode, Auto Range, and
Cal Factor Disable outputs.
3-44. Receiving the Remote Message
3-45. When the Power Meter recieves the Remote
Message (REN line low) it completes the rest of its
current measurement cycle (see Figure 3-6) and
then goes to remote. See the Local to Remote
Mode Change (paragraph 3-31) for information
about how to program the local to remote mode
change.
3-46. Receiving the Local Message
3-47. The Power Meter does not respond to the
GTL (go to local) bus command. It reverts to local
operation when the REN (remote enable) bus
line goes false (high).
3-58. Receiving the Abort Message
3-59. When the Power Meter receives an Interface Clear command (IFC), it stops talking or
listening.
3-60. Test of HP-IB Operation
3-61. Figure 3-7 outlines a quick check of the
436A remote functions. This gives the user two
alternatives for testing the power meter: 1, write
a program corresponding to Figure 3-7 for a quick
check or 2, use the program in Section VIII for
complete testing and troubleshooting.
3-62. REMOTE BCD INTERFACE OPERATION
3-63. BCD Option 024 adds remote programming
and digital output capability to the Power Meter.
There are two basic methods for operating the
Power Meter with this option. It can be operated
locally with an external instrument used to record
output data, or it can be operated remotely by
sending remote programming inputs to the Power
Meter.
3-29
Page 58
Operation
Model 436A
3-30
Figure 3-7. Test of HP-IB Operation Flowchart
Page 59
Model 436A
Operation
436A QUICK PROGRAMMING GUIDE
This guide will help set up and program simple HP-IB instrumentation systems, thereby freeing you
from making an in-depth study of system design and BASIC or HPL programming languages.
THE SYSTEM:
I.
* HP-IB cables shown with dotted lines are used only if the Source and Device under test are programmable.
** Signal Source and Device under Test may be the same, e.g., checking Sig. Gen. Flatness.
II.
THE PROGRAM: If the power meter is the only part of the system to be programmed, use
the program statements in the order given. For more complex systems or programs, include
statements derived from the information in the optional (dashed line) flow chart boxes.
When it is necessary to write more statements, refer to Table 3-2.
Figure 3-8. 436A Quick Programming Guide (1 of 5)
3-31
Page 60
Operation
Model 436A
436A QUICK PROGRAMMING GUIDE (Cont’d)
page 8-34
page 8-34
page 8-33
page 8-33
3-32
Figure 3-8. 436A Quick Programming Guide (2 of 5)
Page 61
Model 436A
Operation
Figure 3-8. 436A Quick Programming Guide (3 of 5)
3-33
Page 62
Operation
Model 436A
436A QUICK PROGRAMMING GUIDE (cont’d)
(HPL)
“pmz”
Subroutines for 9825
- Power meter zero subroutine
“pmz”:
“remove source":dsp“disconnect sensor from source";stp
3-64. Figure 3-3 provides instructions for operating the Power Meter with the BCD option installed.
In order to follow these instructions, the operator
must be familiar with Power Meter programming
and output data format. This information is provided in detail in the paragraphs which follow.
NOTE
The Power Meter BCD option is designed
to interface directly with an HP 5055A
Digital Recorder. When it is used with
this recorder, it can only be operated in
the Local mode (unless a special cable is
fabricated), as the BCD interface bus
lines that are normally used to program
the Power Meter, are used instead to pre-
set the digital recorder print format. In
the paragraphs which follow, differences
in Power Meter output data format for
digital recorder and “universal” interfac-
ing are noted as applicable.
3-65. Output Data Format
3-66. When the Power Meter is interfaced with an
HP 5055A Digital Recorder, the output data print-
out is as described in Table 3-4. When the Power
Meter is interfaced with other controller or re-
corder instruments, data format is selected by the
user. Refer to Table 3-5 for a description of the
function and coding of the Power Meter output
data lines.
3-67. BCD Remote Programming
3-68. Remote programming of the Power Meter
is enabled when a 0.0 to +0.4 Vdc level is applied
to remote enable input line J7-21. The Power
Meter controls that can be programmed remotely
are the MODE and SENSOR ZERO switches. The
controls not programmable are the POWER REF
and LINE switches. The CAL FACTOR % switch
can be enabled and disabled via the remote interface but, when enabled, the calibration factor
entered at the front panel of the Power Meter is
used.
NOTE
Jumper options are provided to enable
remote programming of the SENSOR
ZERO switch when the remote enable in-
put is high (+2.5 to +5.0V level is applied
to J7-21). See Section II, Installation.
3-69. Remote range programming is slightly
different than Local Range selection. For Local
operation, a particular range is selected by allowing
the Power Meter to autorange to the desired range,
then pressing the RANGE HOLD switch to hold
the range. For Remote operation, the programming
codes have providion for direct selection of the
desired range as well as selection of the autorange
function.
3-70. An additional feature that is only available
via remote programming is selection of standby,
triggered, or free running operation of the Power
3-36
Table 3-6. Power Meter Output Data Printout for HP 5055A Digital Recorder
ColumnInterpretation
1 (right)
2
3
4
5
6
7
8
9
10 (left)Exponent Tens Digit
Intrepret measured value as
XXXX . 10
Units Digit
Tens Digit
Hundreds Digit
Thousands Digit
Sign
Range*
Mode**
Status***
Exponent Units Digit
*Range Code
**Mode Decode
***Status
1 = Range 1 (most sensitive)
2 = Range 2
3= Range 3
4 = Range 4
5 = Range 5 (least sensitive)
V = dB [REF]
A = dB (REL)
$2=
Watts
* = dBm
0 = In Range
1 = Underrange (WATT Mode)
2 = Overrange
3 = Underrange (dBm Mode)
4 = ZERO Mode
Page 65
Model 436A
Operation
BCD Remote Programming (cont’d)
Meter. (During Local operation, the Power Meter is
allowed to free run with approximately 133 milliseconds allowed for settling time between measurements.) The specific remote triggering capabilities
are:
Hold -
a.
when the power meter is pro-
trigger with settling time operation, sufficient time
is provided for the Power Meter to settle to the
input power level on all ranges except range 1. On
range 1 approximately 10 seconds (0-10 measurements) are required for the Power Meter to settle
grammed to Hold, it is inhibited from taking
measurements and from outputting data. Thus, it is
set to a predetermined reference condition from
which a measurement can be triggered synchrously
trigger immediate operation, the desired amount of
settling time can be incorporated into the operating program.
to some external event.
b. Trigger Immediate - this programming
command directs the Power Meter to make one
3-72. The programming codes that the Power
Meter will respond to are listed in Table 3-8.
measurement and output the data in the minimum
possible time, then to go into Hold until the next
Triggering command is recieved. It does not allow
settling time prior to the measurement.
3-73. POWER MEASUREMENT ACCURACY
3-74. A power measurement is never free from
error or uncertainty. Any RF system has RF losses,
c.
Trigger
command is identical to the trigger immediate
command except that it causes the Power Meter to
execute a settling-time delay subroutine before
taking a measurement and outputting data.
with Delay - this trigger
mismatch
instrumentation
uncertainty. Measurement errors as high as 50% are
not only possible, they are highly likely unless the
error sources are understood and, as much as
possible, eliminated.
d. Free run at maximum rate - this
programming
asynchronous operation of the Power Meter. It
directs the Power Meter to continuously take
measurements and output data in the minimum
possible time. It does not allow settling time prior
to each measurement.
command is normally used for
3-75. Sources of Error and Measurement Uncertainty
3-76. RF Losses. Some of the RF power that
enters the Power Sensor is not dissipated in the
power sensing elements. This RF loss is caused by
dissipation in the walls of waveguide power
e.
Free run with Delay - this programming
command is identical to the previous command
except that it causes the Power Meter to execute a
sensors, in the center conductor of coaxial power
sensors, in the dielectric of capacitors, connections
within the sensor, and radiation losses.
settling-time delay subroutine prior to each
measurement.
3-77. Mismatch. The result of mismatched impe-
dances between the device under test and the
3-71. When programming the Power Meter for
synchronous triggered operation, there are two
factors that the programmer must consider to
ensure the validity of the output measurement data.
The first factor is the time that it takes the Power
Meter to respond to a full scale change in input
power level. A typical Power Meter response curve
is shown in Figure 3-4. By comparing this curve
with the measurement timing cycle shown in
Figure 3-6 and summarized in Table 3-5, the
validity of the Power Meter output can be
tabulated according to operating range and
triggering interval - versus change in input power
level. A general summay of this information is as
follows:
power sensor is that some of the power fed to the
sensor is reflected before it is dissipated in the
load. Mismatches affect the measurement in two
ways. First, the initial reflection is a simple loss
and is called mismatch loss. Second, the power
reflected from the sensor mismatch travels back up
the transmission line until it reaches the source.
There, most of it is dissipated in the source
impedance, but some of its re-reflected by the
source mismatch. The re-reflected power returns to
the power sensor and adds to, or subtracts from,
the incident power. For all practical purposes, the
effect the re-reflected power has upon the power
measurement is unpredictable. This effect is called
mismatch uncertainty.
When the Power Meter is programmed for
a.
b. When the Power Meter is programmed for
losses,
mismatch
uncertainty and
uncertainy,
calibration
3-37
Page 66
Operation
Model 436A
Sources of Error and Measurement Uncertainty
(cont’d)
3-78. Instrumentation Uncertainty. Instruments:
tion uncertainty describes the ability of the metering circuits to accurately measure the dc output
from the Power Sensor’s power sensing device. In
the Power Meter this error is ± 0.5% for Ranges 1
through 5. It is important to realize, however, that
these uncertainty specifications do not indicate
overall measurement accuracy.
3-79. Power Reference Uncertainty. The output
level of the Power Reference Oscillator is factory
set to 1 mW ± 0.70% at 50 MHz. This reference is
normally used to calibrate the system, and is,
therefore, a part of the system’s total measurement
uncertainty.
3-80. Cal Factor Switch Resolution Error. The
resolution of the CAL FACTOR % switch
contributes a significant error to the total
measurement because the switch has 2% steps. The
maximum error possible in each position is ± 0.5%
ment and record the reading. Use the reflection
coefficient, magnitude and phase angle from the
table supplied with the Power Sensor to calculate
the corrected power level.
b. Set the CAL FACTOR % switch to the
nearest position above and below the correction
factor given on the table. Interpolating between
the power levels measured provides the corrected
power level.
3-86. Calculating Total Uncertainty
3-87. Certain errors in calculating the total
measurement uncertainty have been ignored in this
discussion because they are beyond the scope of
this manual. Application Note AN-64, “Microwave
Power Measurement”, delves deeper into the calcu-
lation of power measurement uncertainties. It is
available, on request, from your nearest HP office.
3-88. Known Uncertainties. The known uncer-
tainties which account for part of the total power
measurement uncertainty are:
3-81. Corrections for Error
3-82. The two correction factors basic to power
meters are
calibration factor and effective
efficiency. Effective efficiency is the correction
factor for RF losses within the Power Sensor.
Calibration factor takes into account the effective
efficiency and mismatch losses.
3-83. Calibration factor is
expressed as a
percentage with 100% meaning the power sensor
has no losses. Normally the calibration factor will
be 100% at 50 MHz, the operating frequency of
the internal reference oscillator.
3-84. The Power Sensors used with the Power
Meter have individually calibrated calibration
factor curves placed on their covers. To correct for
RF and mismatch losses, simply find the Power
Sensor’s calibration factor at the measurement
frequency from the curve or the table that is
supplied with the Power Sensor and set the CAL
FACTOR % switch to this value. The measurement
error due to this error is now minimized.
3-85. The CAL FACTOR % switch resolution
error of ± 0.5% may be reduced by one of the
following methods:
a.
Instrumentation uncertainty
± 0.02 dB (Range 1 through 5).
Power reference uncertainty
b.
± 0.03 dB.
c. CAL FACTOR switch resolution ±0.5% or
±0.02 dB.
The total uncertainty from these sources is ±1.7%
or ±0.07 dB.
3-89. Calculating Mismatch Uncertainty. Mismatch
uncertainty is the result of the source mismatch
interacting with the Power Sensor mismatch. The
magnitude of uncertainty is related to the magnitudes of the source and Power Sensor reflection
coefficients, which can be calculated from SWR.
Figure 3-9 shows how the calculations are to be
made and Figure 3-10 illustrates mismatch uncertainty and total calculated uncertainty for two
cases. In the first case, the Power Sensor’s SWR =
1.5, and in the second case, the Power Sensor’s
SWR = 1.26. In both cases the source has a SWR of
2.0. The example shows the effect on power
measurement accuracy a poorly matched power
sensor will have as compared to one with low
mismatch.
Leave the CAL FACTOR % switch on
a.
100% after calibration, then make the measure-
3-38
3-90. A faster, easier
tainty is to use the
way to find mismatch uncerHP Mismatch Error (uncer-
Page 67
Model 436A
Operation
Calculating Total Uncertainty (cont’d)
tainty) Limits/Reflectometer Calculator. The
calculator may be obtained, on request, from your
nearest Hewlett-Packard office by using HP Part
Number 5952-0448.
3-91. The method of calculating measurement
uncertainty from the uncertainty in dB is shown
by Figure 3-11. This method would be used when
the initial uncertainty calculations were made with
the Mismatch Error/Reflectometer Calculator.
NOTE
The BCD output data levels are TTL compatible. A false
(0) state is defined as 0.0 to +0.4 Vdc and a true state is
defined as +2.5 to +5.0 Vdc
Table 3-7. BCD Output Data Codes (1 of 2)
Function
MEASURED VALUE - The Power Meter format for outputting
the measured value is SIGN, Four BCD DIGITS, and a negative
EXPONENT. It is interpreted as:
± XXXX
Code
NOTES
Pin numbers refer to connec-
tor J7 on the rear panel.
When used with 5055A, a four
line format is established by
the following pins:
34 (ground)
10 (measurement rate;
floats high)
35 (cal factor disable;
floats high)
Sign
space (+)
Digits
Units
Tens101A
Hundreds
Thousands
10°A
10°B
10°C
10°D
101B
101C
1
10
D
2
10
A
102 B
102 C
2
10
D
3
10
A
103 B
103 C
103 D
PIN 9
Weight
11
2
4
8
1
2
4
8
1
2
4
8
1
2
4
8
0
1
Pin Number
2,
26
27
3
4
28
29
5
6
30
31
7
8
32
33
3-39
Page 68
Operation
Model 436A
Table 3-7. BCD Output Data Codes (2 of 2)
Function
STATUS OUTPUTS
In Range
Underrange (WATT Mode)
Overrange
Underrange (dBm Mode)
Zero Mode
RANGE - indicates range on which last measurement made.
1 (most sensitive)
2
3
4
5 (least sensitive)
EXPONENT
Units EX0A
EXOB
EXOC
O
EX
D
Tens EX
1
A
Code
Pin 40Pin 16 Pin 15
0
0
0
0
1
Pin 36Pin 12
00
0
0
1
1
Weight
2
1
4
8
1
00
0
1
1
0
Pin 11
1
1
0
0
Pin
17
18
42
43
19
1
0
1
0
1
0
1
0
1
MODE
PRINT
dB [REF]
dB (REL)
WATT
dBM
Note: when used with
5055A, four line format is established by
following pins:
20 (ground)
44 (ground)
45 (ground)
Pin 14
0
0
1
1
Pin 13
0
1
0
1
Note: when used with
5055A, four line format
is established by following pins:
38 (floats high)
39 (floats high)
High to low transition
on pin 48 when output
data is valid.
3-40
Page 69
Model 436A
Operation
Table 3-8. BCD Programming Commands
Commands
Remote enable
Input Pin
J7-21When high, enables local operation of Power Meter via front-
panel controls. When low, enables remote operation of Power
Meter via programming commands listed below.
Range Bit 1
J7-24Select Power Meter measurement range when Remote Enable
Range Bit 2J7-25input is low.
Range Bit 3
J7-23
*Standby range: Power Meter operating program is
held at Power Up address 000
Function
NOTE
When equipped with the BCD Option 024, the
Power Meter generates a Print command and
provides valid output data after each measure-
ment for both Local and Remote operation.
Range
0*
1
2
3
4
5
Auto
Pin 24
0
0
0
0
1
1
1
Pin 25
0
0
1
1
0
0
1
.
8
Pin 23
0
1
0
1
0
1
X (0 or 1)
Rate
Inhibit
Cal Factor
Disable
Mode Bit 1
Mode Bit 2
SENSOR
Zero Select
J7-10
J7-47
J7-35
Selects Power Meter triggering when remote enable input is low
Pin 10Pin 47Pin 49
Hold
Trigger Intermediate 0
Trigger with Delay 1
Free Run Fast
Free Run with Delay 1
X (0 or 1) 0
0
0
Positive-to-negative
0
1
1
When low disables front-panel CAL FACTOR % switch (same as
100% position). When high, enables switch.
J7-38
J7-39
Select mode when remote enable input is low.
Mode
dB [REF]
dB (REL)
WATT
dBm
Pin 38Pin 39
0
1
0
0
0
1
11
J7-46When low, enables power sensor auto zero circuit.
NOTE
When programming this function, allow the circuit about 7 seconds to settle before applying
input power to Power Sensor. If RF input power
is applied while ZERO lamp is on, it will introduce
an offset that will affect future measurements.
X (0 or 1)
transition
X (0 or 1)
X (0 or 1)
3-41
Page 70
Operation
CALCULATING MEASUREMENT UNCERTAINTY
1. Calculate the reflection coefficient from the given SWR.
SWR - 1
=
P
SWR + 1
Model 436A
Power Sensor #1
Power Sensor #2
Power Source
2. Calculate the relative power and percentage power mismatch uncertainties from the reflection
coefficients. An initial reference level of 1 is assumed.
Reletive Power Uncertainty
3-42
%PU =
%PU
1
%PU
2
Percentage Power Uncertainty
(PU - 1) 100% for PU >1
=
(1.138 - 1) 100%
and-(1 - PU) 100% for PU <1
and-(1 - 0.870) 100%
= (0.138) 100%and
=
13.8%
=(1.073 - 1) 100%
= (0.073) 100%
=
7.3%
Figure 3-9. Calculating Measurement Uncertainties (1 of 2)
Figure 3-9. Calculating Measurement Uncertainties (2 of 2)
3-43
Page 72
Operation
Model 436A
3-44
Figure 3-10. The Effect of Power Sensor Mismatch on Measurement Accuracy
Page 73
Model 436A
Operation
CALCULATING MEASUREMENT UNCERTAINTY
1. For this example the known values are:
source SWR, 2.2 and power sensor SWR, 1.16.
From the Mismatch Error Calculator the mismatch uncertainty is found to be +0.24, -0.25 dB.
2. Add the known uncertainties from paragraph 3-73, (± 0.10 dB). Our total measurement uncertainty is +0.34, -0.35 dB.
3. Calculate the relative measurement uncertainty from the following formula:
dB =
If dB is positive then:
> P0; let P0 = 1
P
1
P
MU =
1
If dB is negative then:
P
< P0; let P1 = 1
1
4. Calculate the percentage Measurement Uncertainty.
For P
> P
1
0
%MU =(P1 - P0) 100
=
(1.081 - 1) 100
=
+8.1%
Figure 3-11. Calculating Measurement Uncertainty (Uncertainty in dB Known)
For P1 < P
%MU =
=
=
0
– P0) 100
- (P
1
- (1 - 0.923) 100
-7.7%
3-45/3-46
Page 74
Page 75
Model 436A
Performance Tests
SECTION IV
PERFORMANCE TESTS
4-1.
INTRODUCTION
4-2.
The procedures in
ical performance of the Power Meter using the
specifications of Table 1-1 as performance standards. All tests can be performed without access to
the interior of the instrument. A simpler operational test is included in Section III under Operator’s Checks.
4-3. EQUIPMENT REQUIRED
4-4. Equipment required for the performance tests
is listed in Table 1-2, Recommended Test Equipment. Any equipment that satisfies the critical
specifications given in the table may be substituted
for the recommended model(s).
4-5. TEST RECORD
4-6. Results of the performance tests may be
tabulated on the Test Record at the end of the test
procedures. The Test Record lists all of the tested
specifications and their acceptable limits. Test
results recorded at incoming inspection can be used
for comparison in periodic maintenance, troubleshooting, and after repairs or adjustments.
this section test the electr-
4-7. PERFORMANCE TESTS
4-8. The performance tests given in this section are
suitable for incoming inspection, troubleshooting,
or preventive maintenance. During any performance test, all shields and connecting hardware
must be in place. The tests are designed to verify
published instrument specifications. Perform the
tests in the order given and record the data on the
test card and/or in the data spaces provided at the
end of each procedure.
NOTE
The Power Meter must have a half-hour
warmup and the line voltage must be
within +5%,
-10% of nominal if the
performance tests are to be considered
valid.
4-9. Each test is arranged so that the specification
is written as it appears in Table 1-1. Next, a
description of the test and any special instructions
or problem areas are included. Each test that
requires test equipment has a setup drawing and a
list of the required equipment. The initial steps of
each procedure give control settings required for
that particular test.
4-1
Page 76
Performance Tests
4-10. ZERO CARRYOVER TEST
Model 436A
PERFORMANCE TESTS
SPECIFICATION:
DESCRIPTION:
EQUIPMENT:
PROCEDURE:
±0.2% of full scale when zeroed on the most sensitive range.
After the Power Meter is initially zeroed on the most sensitive range, the change in the
digital readout is monitored as the Power Meter is stepped through its ranges. Thus,
this test also takes noise and drift into account because noise , drift, and zero carryover readings cannot be separated.
Figure 4-1. Zero Carryover Test Setup
Range Calibrator . . . . . . . HP 11683A
1.
Set the Power Meter switches as follows:
CAL FACTOR % . . . . . 100
POWER REF . . . . . .off (out)
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . . ON (in)
off (out)
Set the Range Calibrator switches as follows:
2.
FUNCTION . . . . . . . STANDBY
POLARITY . . . . . . . NORMAL
100
RANGE . . . . . . . .
LINE . . . . . . . . .
Connect the equipment as shown in Figure 4-1.
3.
4.
Press and hold the Power Meter SENSOR ZERO switch and wait for the digital
readout to stabilize. Then verify that the Power Meter ZERO lamp is lit and that
the digital readout indicates 0.00 ± 0.02.
Power Meter is now zeroed on most sensitive range (10 p W).
Release the Power Meter SENSOR ZERO switch and wait for the ZERO lamp to
5.
go out before proceeding to the next step.
6.
Set the Range Calibrator FUNCTION switch to CALIBRATE and verify that the
Power Meter autoranges to the 100
‘7.
Set the Power Meter RANGE HOLD switch to on (in) and the Range Calibrator
FUNCTION switch to standby.
#w
ON (in)
NOTE
KW
range.
4-2
Page 77
Model 436A
PERFORMANCE TESTS
4-10. ZERO CARRYOVER TEST (cont’d)
8.
Wait for the Power Meter’s digital readout to stabilize and verify that the indication
observed is within the limits shown on the table below. Then set the POWER
Meter RANGE HOLD switch to off (out).
9.
Repeat steps 6, 7, and 8 with the Range Calibrator RANGE switch set, in turn, to
1 mW, 10 mW, and 100 mW. Verify that the Power Meter autoranges properly,
and that the indication observed on each range is within the limits shown in
Table 4-1.
Table 4-1. Zero Carryover Autorange Digital Readout Results
Performance Tests
Range Calibrator
and
Power Meter
Range
10 /lw
100
/Jw
1 mW
10 mW
100 mW
4-11. INSTRUMENT ACCURACY TEST
SPECIFICATION: WATT MODE:
dBm MODE:
dB (REL) MODE:
±0.570 in Ranges 1 through 5.
±0.02 dB ±0.001 dB/°C in Ranges 1 through 5.
±0.02 dB ±0.001 dB/°C in Ranges 1 through 5.
The dB (REL) specifications are for within-range measurements.
For range-to-range accuracy, add the uncertainty associated with
the range in which the reference was entered, to the uncertainty
associated with the range in which the measurement was made.
For example, if a reference is entered in Range 1 and a measurement is made in Range 5, the total uncertainty is ±0.04 (Range 1
±0.02 + Range 5 ± 0.02 = ± 0.04).
Min
-0.02
-0.2
-.002
-0.02
-00.2
Results
ActualMax
0.02
0.2
.002
0.02
00.2
NOTE
DESCRIPTION:
After the Power Meter is initially calibrated on the 1 mW range, the digital readout is
monitored as the Range Calibrator is adjusted to provide reference inputs corresponding to each of the Power Meter operating ranges.
4-3
Page 78
Performance Tests
PERFORMANCE TESTS
4-11. INSTRUMENT ACCURACY TEST (cont’d)
Figure 4-2. Instrument Accuracy Test Setup
Model 436A
EQUIPMENT:
PROCEDURE:
Range Calibrator . . . . . . . HP 11683A
1.
Set the Power Meter switches as follows:
CAL FACTOR % . . . . . 100
POWER REF . . . . . . off (out)
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
Set the Range Calibrator switches as follows:
2.
off (out)
ON (in)
FUNCTION . . . . . . . STANDBY
POLARITY . . . . . . . NORMAL
RANGE . . . . . . . . 1mW
LINE . . . . . . . . .
3.
Connect the equipment as shown in Figure 4-2.
ON (in)
4. Press and hold the Power Meter SENSOR ZERO switch and wait for the digital
readout to stabilize. Then verify that the Power Meter ZERO lamp is lit and that
the digital readout indicates 0.00 ± 0.02.
NOTE
Power Meter is now zeroed on the most sensitive range (10 v W).
4-4
5.
Release the Power Meter SENSOR ZERO switch and wait for the ZERO lamp to
go out before proceeding to the next step.
6. Set the Range Calibrator FUNCTION switch to CALIBRATE and verify that the
Power Meter autoranges to the 1 mW range.
7.
Observe the Power Meter digital readout and, if necessary, adjust the front-panel
CAL ADJ control to obtain a 1.000 ± 0.002 indication.
NOTE
The Range Calibrator output level is adjustable in 5 dB increments.
Thus, the 3
M
W,
30p
W,
300p
W, 3 mW, and 30 mW legends on the
RANGE switch are approximations. The true outputs for these set-
tings are
3.161.t W, 31.61.JW,
316PW, 3.16 mW and 31.6 mW.
Page 79
Model 436A
PERFORMANCE TESTS
4-11.INSTRUMENT ACCURACY TEST (cont’d)
Performance Tests
8.
Set the Range Calibrator RANGE switch, in turn, to 10
100 mW. Verify that the Power Meter autoranges properly and that the indication
observed on each range is within the limits specified in the table below.
Set the Power Meter MODE switch to dBm.
9.
10.
Set the Range Calibrator RANGE switch, in turn, to -20 dBm, -10 dBm, 0 dBm,
+10 dBm, and +20 dBm. Verify that the Power Meter autoranges properly and
that the indication observed on each range is within the limits specified in
Table 4-2.
Table 4-2. Instrument Accuracy Test Results
Range Calibrator
and
Power Meter
Range
10
/..lw
100
#w
1 mW
10 mW
100 mW99.0
Results
Min Actual Max
9.95
99.5
0.995
9.95
10.05
100.5
1.005
10.05
101.0
Range Calibrator
and
Power Meter
Range
-20 dBm-20.02
-10 dBm-10.02
0 dBm-0.02
+10 dBm
+20 dBm
11. Set the Range Calibrator RANGE switch to -10 dBm.
uW,
100
Min Actual
9.98
19.96
PW,
10 mW, and
Results
-19.98
-9.98
0.02
10.02
20.04
12. Set the Power Meter MODE switch to dB [REF] and verify that the digital readout
indicates 0.00 ± 0.01.
13. Set the Range Calibrator RANGE switch, in turn, to -20 dBm, -5 dBm, and
+10 dBm. Verify that the Power Meter autoranges properly, and that the indica-
tion observed on each range is within the limits specified in Table 4-3.
Table 4-3. Instrument Accuracy Test Results for dB [REF] Mode
Range Calibrator
and
Power Meter
Ranges
-20 dBm
-5 dBm
+10 dBm
Min
-9.96
+4.96
+19.96
Results
ActualMax
-10.04
+5.04
20.04
4-5
Page 80
Performance Tests
4-12. CALIBRATION FACTOR TEST
Model 436A
PERFORMANCE TESTS
SPECIFICATION:
DESCRIPTION:
PROCEDURE:
16-position switch normalizes meter reading to account for calibration factor. Range
85% to 100% in 1% steps. 100% position corresponds to calibration factor at 50 MHz.
After the Power Meter is zeroed on the most sensitive range, a 1 mW, input level is applied to the Power Meter and the CAL ADJ control is adjusted to obtain a 1.000 mW
indication. Then the CAL FACTOR % switch is stepped through its 16 positions and
the digital readout is monitored to ensure that the proper indication is obtained for
each position.
Figure 4-3. Calibration Factor Test Setup
1.
Set the Power Meter switches as follows:
CAL FACTOR % . . . . . 100
POWER REF . . . . . .
off (out)
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
off (out)
ON (in)
Set the Range Calibrator switches as follows:
2.
FUNCTION . . . . . . . STANDBY
POLARITY . . . . . . . NORMAL
RANGE . . . . . . . . 1mW
LINE . . . . . . . . .
Connect the equipment as shown in Figure 4-3.
3.
Press and hold the Power Meter SENSOR ZERO switch and wait for the digital
4.
ON (in)
readout to stabilize. Then verify that the Power Meter ZERO lamp is lit and that
the digital readout indicates 0.00 ± 0.02.
NOTE
Power Meter is now zeroed on most sensitive range (10 p W)
Release the Power Meter SENSOR ZERO switch and wait for the ZERO lamp to
5.
go out before proceeding to step 6.
Set the Range Calibrator FUNCTION switch to CALIBRATE and verify that the
6.
Power Meter autoranges to the 1 mW range.
Adjust the Power Meter CAL ADJ control to obtain a 1.000 ± 0.002 indication
7.
on the digital readout.
4-6
Page 81
Model 436A
PERFORMANCE TESTS
4-12. CALIBRATION FACTOR TEST (cont’d)
8.
Set the CAL FACTOR % switch, in turn, to each position and verify that the indications observed are within the limits specified in Table 4-4.
Table 4-4. Calibration Factor Test Results
Performance Tests
CAL FACTOR
Switch
Position
100
99
98
97
96
95
94
93
4-13. POWER REFERENCE LEVEL TEST
Min.
0.994
1.004
1.014
1.025
1.036
1.047
1.058
1.069
Results
ActualMax.
1.006
1.016
1.026
1.037
1.048
1.059
1.070
1.081
CAL FACTOR
Switch
Position
92
91
90
89
88
87
86
85
Min.
1.081
1.093
1.105
1.118
1.130
1.143
1.157
1.170
Results
Actual
SPECIFICATION: Internal 50 MHz oscillator factory set to 1 mW ± 0.7% traceable to the National
Bureau of Standards.
Accuracy: ±1.2% worst case (±0.9% rms) for one year (0°C to 55°C).
DESCRIPTION:
The power reference oscillator output is factory adjusted to 1 mW ± 0.7%. To achieve
this accuracy, Hewlett-Packard employs a special measurement system accurate to 0.5%
(traceable to the National Bureau of Standards) and allows for a transfer error of ±0.2%
in making the adjustment. If an equivalent measurement system is employed for verification, the power reference oscillator output can be verified to 1 mW ±1.9% (±1.2%
accuracy + ±0.5% verification system error + ±0.2% transfer error = 1.9% maximum
error). The power reference oscillator can be set to ±0.7% using the same equipment
and following the adjustment procedure in paragraph 5-22. To ensure maximum accuracy in verifying the power reference oscillator output, the following procedure provides
step-by-step instructions for using specified Hewlett-Packard test instruments of known
capability. If equivalent test instruments are used, signal acquisition criteria may vary
and reference should be made to the manufacturer’s guidelines for operating the
instruments.
Max.
1.093
1.105
1.117
1.130
1.142
1.155
1.169
1.182
NOTE
The Power Meter may be returned to the nearest
Hewlett-Packard office to have the power reference
oscillator checked and/or adjusted. Refer to Section II,
PACKAGING.
4-7
Page 82
Performance Tests
PERFORMANCE TESTS
4-13. POWER REFERENCE LEVEL TEST (cont’d)
Figure 4-4. Power Reference Level Test Setup
Model 436A
EQUIPMENT:
PROCEDURE:
Power Meter . . . . . . . . HP 432A
Thermistor Mount . . . . . . HP 478A-H75
Digital Voltmeter (DVM). . . . HP 3490A
1.
Set up the DVM to measure resistance and connect the DVM between the V
connector on the rear panel of the 432A, and pin 1 on the thermistor mount end
of the 432A interconnect cable.
Round off the DVM indication to two decimal places and record this value as the
2.
internal bridge resistance (R) of the 432A (approximately 200 ohms).
Connect the 432A to the Power Meter as shown in Figure 4-4.
3.
Set the Power Meter LINE switch to ON (in) and the POWER REF switch to off
4.
(out). Then wait thirty minutes for the 432A thermistor mount to stabilize before proceeding to the next step.
Set the 432A RANGE switch to COARSE ZERO and adjust the front-panel
5.
COARSE ZERO control to obtain a zero meter indication.
Fine zero the 432A on the most sensitive range, then set the 432A RANGE switch
6.
to 1 mW.
NOTE
Ensure that DVM input leads are isolated from chassis
ground when performing the next step.
RF
4-8
7.
Set up the DVM to measure microvolt and connect the positive and negative input
leads, respectively, to the V
COMP
and V
connectors on the rear panel of the
RF
432A.
Observe the indication on the DVM. If less than 400 microvolt, proceed to the
8.
next step. If 400 microvolt or greater, press and hold the 432A FINE ZERO
switch and adjust the COARSE ZERO control so that the DVM indicates 200
microvolt or less. Then release the FINE ZERO switch and proceed to the next
step.
Round off the DVM indication to the nearest microvolt and record this value as V
9.
.
0
Page 83
Model 436A
PERFORMANCE TESTS
4-13. POWER REFERENCE LEVEL TEST (cont’d)
Performance Tests
10.
Set the Power Meter POWER
observed on the DVM as V
11.
Disconnect the DVM negative input lead from the V
REF switch to ON (in) and record the indications
.
1
connector on the 432A
RP
and reconnect it to 432A chassis ground. Record the new indication observed on
the DVM a
Calculate the power reference oscillator output level (P
12.
S V
COMP
.
) from the following
RF
formula:
Where:
P
= power reference oscillator output level
RF
= previously recorded value
v
COMP
= previously recorded value
V
1
= previously recorded value
V
0
R = previously recorded value
CALIBRATION FACTOR = value for thermistor mount at 50 MHz (traceable to
the National Bureau of Standards)
Verify that the P
13.
is within the following limits:
RF
4-9
Page 84
Performance Tests
Model 436A
Table 4-5. Performance Test Record (1 of 2)
Hewlett-Packard Company
Model 436A
Power Meter
Serial Number
Para.
No.
4-10.
ZERO CARRYOVER
Test
100
10
/..lw
/.lw
Tested By
Date
1 mW
10 mW
100 mW
4-11.INSTRUMENTATION ACCURACY
WATT MODE
10
/.lw
100
#w
1 mW
10 mW
100 mW
Results
Min
-0.02
-0.2
/.fw
/Jw
-0.002 mW
ActualMax
0.02
0.2
/.lw
0.002 mW
/.lw
-0.02 mW0.02 mW
-0.2 mW
9.95
/.lw
99.5
/.lw
0.2 mW
10.05
100.5
/Jw
/.fw
0.995 mW1.005 mW
9.95 mW10.05 mW
99.5 mW100.5 mW
dBm MODE
-20 dBm
-10 dBm
0 dBm-0.02 dBm0.02 dBm
10 dBm
20 dBm19.96 dBm20.04 dBm
dB (REL) MODE
-20 dBm
- 5 dBm
+10 dBm
4-12.CALIBRATION FACTOR
100
99
98
971.025 mW
96
95
94
93
-20.02 dBm-19.98 dBm
-10.02 dBm
-9.98 dBm
9.95 dBm10.02 dBm
-9.96 dBm-10.04 dBm
+4.96 dBm
+5.04 dBm
+19.96 dBm20.04 dBm
0.994 mW
1.004 mW
1.014 mW
1.006 mW
1.016 mW
1.026 mW
1.037 mW
1.036 mW
1.047 mW
1.058 mW
1.069 mW
1.048 mW
1.059 mW
1.070 mW
1.081 mW
4-10
Page 85
Model 436A
Performance Tests
Table 4-5. Performance Test Record (2 of 2)
Para.
No.
4-12.
4-13
Test
CALIBRATION FACTOR (cont’d)
921.081 mW
91
90
89
88
87
86
85
POWER REFERENCE
P
RF
Results
Min.
ActualMax
1.093 mW
1.093 mW
1.105 mW
1.118 mW
1.105 mW
1.117 mW
1.130 mW
1.130 mW1.142 mW
1.143 mW
1.155 mW
1.157 mW1.169 mW
1.170 mW
1.182 mW
0.981 mW1.019 mW
4-11/4-12
Page 86
Page 87
Model 436A
SECTION V
ADJUSTMENTS
INTRODUCTION
5-1.
This section describes the adjustments which
5-2.
return the Power Meter to peak operating
will
condition after repairs are completed.
5-3. If the adjustments are to be considered valid,
the Power Meter must have a half-hour warmup
and the line voltage must be within +5 to -10% of
nominal.
5-4. SAFETY CONSIDERATIONS
5-5. Although this instrument has been designed in
accordance with international safety standards, this
manualcontainsinformation,
warnings which must be followed to ensure safe
operation and to retain the instrument in safe
condition (see Sections II and III). Service and
adjustments should be performed only by qualified
service personnel.
cautions,
and
Adjustments
5-9. Whenever it is likely that the protection
offered by fuses has been impaired, the instrument
must be made inoperative and secured against any
unintended operation.
Adjustments described herein are performed with power supplied to the instru-
ment while protective covers are removed.
Energy available at many points may, if
contacted, result in personal injury.
5-10. EQUIPMENT REQUIRED
5-11. The test equipment required for the adjustment procedures is listed in Table 1-2, Recommended Test Equipment. The critical specifications
of substitute test instruments must meet or exceed
the standards listed in the table if the Power Meter
is to meet the standards set forth in Table 1-1,
Specifications.
Any
(grounding) conductor (inside
the instrument) or disconnection of the
protective earth terminal is likely to make
the instrument dangerous. Intentional
interruption is prohibited.
5-6. Any adjustment, maintenance, and repair of
the opened instrument with voltage applied should
be avoided as much as possible and, when inevitable, should be carried out only by a skilled person
who is aware of the hazard involved.
5-7. Capacitors inside the instrument may still be
charged even if the instrument has been disconnected from its source of supply.
5-8. Make sure that only fuses with the required
rated current and of the specified type (normal
blow, time delay, etc.) are used for replacement.
The use of repaired fuses and the shortcircuiting of
fuseholders must be avoided.
interruption ofthe
protective
or outside
5-12. FACTORY SELECTED COMPONENTS
5-13. Factory selected components are indicated
on the schematic and replaceable parts list with an
asterisk immediately following the reference designator. The nominal value of the component is
listed. Table 5-1 lists the parts by reference
designator and provides an explanation of how the
component is selected, the normal value range, and
a reference to the appropriate service sheet. The
Manual Changes supplement will update any
changes tofactoryselected
information.
5-14. ADJUSTMENT LOCATIONS
5-15. The last foldout in this manual contains a
table which cross-references all pictorial and
schematic locations of the adjustment controls.
The accompanying figure shows the locations of
the adjustable controls, assemblies, and chassismounted parts.
component
5-1
Page 88
Adjustments
Model 436A
ADJUSTMENTS
Table 5-1. Factory Selected Components
Reference
Designator
A2R18
A display readout of 100.0 mW if the Power
Selected For
Meter, after being properly adjusted, passes
all of the Instrumentation Accuracy Tests
specified in Section IV except for the high
range (100 mW/20 dBm)
A2R50
Adjust A2R69 FREQ (Frequency Adj) for
maximum indication on digital readout,
then check frequency of 220 Hz Multivibrater. If out of specification (220 ± 16 Hz)
select value for A2R50 to produce maximum
indication on digital readout while 220 Hz
Multivibrator frequency is in specification.
A8R5A Power Reference Oscillator output of
1 mW if this value falls outside the range of
adjustment available with LEVEL
ADJUST potentiometer A8R5.
Normal Value
Range
196K
(150K~ to
250Kfl
)
13.3Kf2
(1OK!2
to
17.8KC?)
7100
(7100s’2
to
7500Q)
Service
Sheet
7
7
14
5-16. DC OFFSET
REFERENCE:
DESCRIPTION:
EQUIPMENT:
PROCEDURE:
ADJUSTMENT
Service Sheet 8.
DC OFF potentiometer A3R2 is adjusted to remove any dc voltage introduced by the
dc amplifier
Figure 5-1. DC Offset Adjustment Setup
Range Calibrator . . . . . . . Hp 11683A
Set the Power Meter Switches as follows:
1.
CAL FACTOR % . . . . . 100
POWER REF . . . . . .
off (out)
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
off (out)
ON (in)
5-2
Page 89
Model 436A
5-16. DC OFFSET ADJUSTMENT (cont’d)
2.
Set the Range Calibrator switches as follows:
FUNCTION . . . . . . . CALIBRATE
POLARITY . . . . . . . NORMAL
RANGE . . . . . . . . 100mW
LINE . . . . . . . . .
3.
Connect the equipment as shown in Figure 5-1.
4.
Verify that the Power Meter autoranges to the 100 mW range, then set the RANGE
HOLD switch to ON (in).
5.
Set the Range Calibrator FUNCTION switch to STANDBY.
Remove the Power Meter top cover and adjust DC OFF potentiometer A3R2 so
6.
that the digital readout indicates 00.0 with a blinking minus sign.
Adjustments
ADJUSTMENTS
ON (in)
5-17. AUTO ZERO OFFSET ADJUSTMENT
REFERENCE:Service Sheet 8.
DESCRIPTION:ZERO OFF potentiometer A3R47 is adjusted to remove any dc offset that is intro-
duced when the SENSOR ZERO switch is pressed.
Figure 5-2. Auto Zero Offset Adjustment Setup
EQUIPMENT:
PROCEDURE: 1.
Range Calibrator . . . . . . . HP 11683A
Set the Power Meter switches as follows:
CAL FACTOR % . . . . . 100
POWER REF . . . . . . off (out)
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
off (out)
ON (in)
2.
Set the Range Calibrator switches as follows:
FUNCTION . . . . . . . STANDBY
POLARITY . . . . . . . NORMAL
LINE . . . . . . . . .
3.
Connect the equipment as shown in Figure 5-2.
ON (in)
5-3
Page 90
Adjustments
ADJUSTMENTS
5-17. AUTO ZERO OFFSET ADJUSTMENT (cont‘d)
4.
Verify that the Power Meter autoranges to the 10
Meter top cover.
If specified indication cannot be obtained in next step,
perform DC Spike Balance Adjustment. Then repeat
this procedure.
5.
Press and hold the Power Meter SENSOR ZERO switch and adjust ZERO OFF
potentiometer A3R47 so that the digital readout indicates 0.00 with blinking
minus sign.
5-18. SPIKE BALANCE ADJUSTMENT
NOTE
Model 436A
PW
range, and remove the Power
REFERENCE:
DESCRIPTION:
EQUIPMENT:
PROCEDURE:
Service Sheets 7 and 8.
A reference signal is applied to the Power Meter from the Range Calibrator to force
the sensor zero circuit to its negative extreme. The SENSOR ZERO switch is then
held pressed while BAL potentiometer A3R65 is adjusted to center the sensor zero
circuit output voltage range.
Figure 5-3. Spike Balance Adjustment Setup
Range Calibrator . . . . . . . HP 11683A
1.
Set the Power Meter switches as follows:
CAL FACTOR % . . . . 100
POWER REF. . . . . .
off (out)
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
off (out)
ON (in)
5-4
2.
Set the Range Calibrator switches as follows:
FUNCTION . . . . . . . CALIBRATE
POLARITY . . . . . . . NORMAL
RANGE . . . . . . . .
LINE . . . . . . . . .
ON (in)
Page 91
Model 436A
ADJUSTMENTS
5-18. SPIKE BALANCE ADJUSTMENT (cont’d)
3.
Remove the Power Meter top cover and adjust the front-panel CAL ADJ control
so that the digital readout indicates 100.0
4.
Press and hold the Power Meter SENSOR ZERO switch and adjust BAL poteniometer A3R65 so that the display readout indicates 60.0 ± 0.2
The Power Meter sensor zero circuit must be
re-zeroed as described in the following steps
before valid power measurements can be made.
5.
Set the Range Calibrator FUNCTION switch to standby. Then press the Power
Meter SENSOR ZERO switch and wait for the digital readout to stabilize.
6.
Release the Power Meter SENSOR ZERO switch and wait for the ZERO lamp to
go out.
Adjustments
PW
MW.
NOTE
5-19. MULTIVIBRATOR ADJUSTMENT
REFERENCE:
DESCRIPTION:
EQUIPMENT:
PROCEDURE: 1.Set the Power Meter switches as follows:
Service Sheet 7.
FREQ potentiometer A2R69 is adjusted to set the reference frequency of the multi-
vibrator which drives the phase detector and the FET power sensor.
Figure 5-4. Multivibrator Adjustment Setup
Range Calibrator . . . . . . . HP 11683A
Counter . . . . . . . . . . HP 5245L
CAL FACTOR % . . . . . 100
POWER REF . . . . . .
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
off (out)
off (out)
ON (in)
5-5
Page 92
Adjustments
ADJUSTMENTS
5-19. MULTIVIBRATOR ADJUSTMENT (cont’d)
Set the Range Calibrator switches as follows:
2.
FUNCTION . . . . . . . CALIBRATE
POLARITY . . . . . . . NORMAL
LINE . . . . . . . . . ON (in)
3.
Connect the equipment as shown in Figure 5-4.
4.
Remove the Power Meter top cover, adjust FREQ potentiometer A2R69 to obtain
maximum indication on the digital readout, and verify that the counter indicates
220 ± 16 Hz.
5.
Perform the Instrument Accuracy Test described in Section IV to verify overall
Power Meter accuracy. If all indications are obtained as specified, the adjustment
is complete. If any indication cannot be obtained as specified, perform the A-D
Converter and Linear Meter Adjustment.
Model 436A
5-20. A-D CONVERTER AND LINEAR METER ADJUSTMENT
REFERENCE:Service Sheets 7 and 8.
DESCRIPTION:The A-D converter circuit is adjusted to obtain the specified digital
and the meter circuit is adjusted for a corresponding indication.
Figure 5-5. A-D Converter and Linear Meter Adjustment Setup
EQUIPMENT:
Range Calibrator . . . . . . . HP 11683A
Digital Voltmeter (DVM). . . . HP 3490A
PROCEDURE:
1.
Set the Power Meter switches as follows:
CAL FACTOR %. . . . . 100
POWER REF . . . . . .
off (out)
MODE . . . . . . . . .. WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
off (out)
ON (in)
readout accuracy
5-6
Page 93
Model 436A
ADJUSTMENTS
5-20. A-D CONVERTER AND LINEAR METER ADJUSTMENT (cont’d)
2.
Set the Range Calibrator switches as follows:
FUNCTION . . . . . . . STANDBY
RANGE . . . . . . ..1 mW
POLARITY . . . . . . . NORMAL
LINE . . . . . . . . . ON (in)
3.
Connect the equipment as shown in Figure 5-5.
4.
Remove the Power Meter top cover and set the DVM to the 1000 mV range.
Press the Power Meter SENSOR ZERO switch and wait for the display readout to
5.
stabilize. Then release the SENSOR ZERO switch and wait for ZERO led to go
out before proceeding to the next step.
6.
Set the Range Calibrator FUNCTION switch to CALIBRATE and adjust the
Power Meter front-panel CAL ADJ control to obtain a 1.000 Vdc indication on
the DVM.
Adjustments
7.
Adjust the Power Meter LIN potentiometer A3R37 so that the digital readout indicates 1.000 mW.
8.
Set the Power Meter MODE and
RANGE HOLD switches to dBm and on (in),
respectively.
NOTE
The next step sets the
A-D log threshold. When the specified indication (-10.00 dBm) is obtained, the digital-readout should be just on the verge of blanking, i.e., the readout
may randomly alternate between -10.00 and UNDER
RANGE, -1.
9.
Set the Range Calibrator RANGE switch to -10 dBm and adjust the power
meter LZR,
10.
Set the Power Meter RANGE HOLD switch to off (out) and the Range Calibrator
-
A3R59, for -10 dBm.
RANGE switch to 1 mW.
11.
Adjust Power Meter LFS potentiometer A3R48 so that the digital readout indicates -0.00.
12.
Set the Power Meter MODE switch to WATT and adjust MTR potentiometer
A3R17 so that the pointer is aligned half way between the last two marks on the
meter face.
5-7
Page 94
Adjustments
ADJUSTMENTS
5-21. POWER REFERENCE OSCILLATOR FREQUENCY ADJUSTMENT
NOTE
Adjustment of the Power Reference Oscillator fre-
quency may also affect the output level of the
oscillator. Thus after the frequency is adjusted to
50.0 ± 0.5 MHz, the output level should be checked
as described in Section IV. A procedure for adjust-
ing the output to the specified level is provided in
the next paragraph.
Model 436A
REFERENCE:
DESCRIPTION:
EQUIPMENT:
PROCEDURE:
Service Sheet 14.
Variable inductor A8L1 is adjusted to set the power reference oscillator output fre-
quency to 50.0 ± 0.5 MHz.
Figure 5-6. Power Reference Oscillator Frequency Adjustment Setup
Counter . . . . . . . . . . HP 5245L
1.
Set the Power Meter LINE switch to ON (in) and the POWER REF switch to off
(out).
Set up the counter to measure frequency and connect the equipment as shown in
2.
Figure 5-6.
Set the Power Meter POWER REF switch to ON (in) and observe the indication
3.
on the counter. If it is 50.0 ± 0.5 MHz, no adjustment of the power reference
oscillator frequency is necessary. If it is not within these limits, adjust the power
reference oscillator frequency as described in steps 4 through 9.
5-8
4.
Remove the Power Meter top cover.
Take care not to ground the +15V or -15V inputs to the
power reference oscillator when performing the following
steps. Grounding either of these inputs could damage the
power reference oscillator, and/or the power supply.
Grasp the power reference oscillator assembly firmly, and remove the four screws
5.
which secure it to the Power Meter chassis.
Page 95
Model 436A
ADJUSTMENTS
5-21. POWER REFERENCE OSCILLATOR FREQUENCY ADJUSTMENT (cont’d)
Tilt the power reference oscillator assembly to gain access to the circuit board
6.
underneath the metal cover, and adjust A8L1 to obtain a 50.00 ± 0.5 MHz indication on the counter.
Reposition the power reference oscillator on the Power Meter chassis but do not
7.
replace the mounting screws.
Observe the indication on the counter. If it is 50.0 ± 0.5 MHz, the adjustment
8.
procedure is complete. If it is not within these limits, repeat steps 6 and 7 except
offset the power reference oscillator frequency as required to obtain a 50.0 ±
0.5 MHz indication on the counter when the power reference oscillator assembly
is repositioned on the Power Meter chassis.
Replace the four screws which secure the power reference oscillator to the Power
9.
Meter chassis.
Adjustments
5-22. POWER REFERENCE OSCILLATOR LEVEL ADJUSTMENT
REFERENCE:Service Sheet 14.
DESCRIPTION:The power reference oscillator output is factory-adjusted to 1 mW ± 0.7% using a special
measurement system accurate to 0.570 (traceable to the National Bureau of Standards)
and allowing for a 0.2% transfer error. To ensure maximum accuracy in readjusting the
power reference oscillator, the following procedure provides step-by-step instructions
for using specified Hewlett-Packard instruments of known capability. If equivalent in-
struments are used, signal acquisition criteria may vary and reference should be made
to the manufacturer’s guidelines for operating the equipment.
NOTE
The Power Meter may be returned to the nearest HP
office to have the power reference oscillator checked
and/or adjusted. Refer to Section II, PACKAGING.
EQUIPMENT:
Figure 5-7. Power Reference Oscillator Level Adjustment Setup
Power Meter . . . . . . . . HP 432A
Thermistor Mount . . . . . . HP 478A-H75
Digital Voltmeter (DVM) . . . . HP 3490A
5-9
Page 96
Adjustments
ADJUSTMENTS
5-22. POWER REFERENCE OSCILLATOR LEVEL ADJUSTMENT (cont’d)
Model 436A
PROCEDURE: 1.
Set up the DVM to measure resistance and connect the DVM between the V
nector on the rear panel of the 432A and pin 1 on the thermistor mount end of the
432A interconnect cable.
Round off the DVM indication to two decimal places and record this value as the in-
2.
ternal bridge resistance (R) of the 432A (approximately 200 ohms).
Connect the 432A to the Power Meter as shown in Figure 5-7.
3.
4.
Set the Power Meter LINE switch to ON (in) and the POWER REF switch to off (out).
Then wait thirty minutes for the 432A thermistor mount to stabilize before proceeding to the next step.
Set the 432A RANGE switch to COARSE ZERO and adjust the front-panel COARSE
5.
ZERO control to obtain a zero meter indication.
Fine zero the 432A on the most sensitive range, then set the 432A RANGE switch to
6.
1 mW.
NOTE
Ensure that the DVM input leads are isolated from chassis
ground when performing the next step.
7.
Set up the DVM to measure microvolt and connect the positive and negative inputs
leads, respectively, to the V
432A.
COMP
and V
connectors on the rear panel of the
RF
RF
con-
5-10
8.
Observe the indication on the DVM. If less than 400 microvolt, proceed to the next
step. If 400 microvolt or greater, press and hold the 432A FINE ZERO switch and
adjust the COARSE ZERO control so that the DVM indicates 200 microvolt or less.
Then release the FINE ZERO switch and proceed to the next step.
9.
Round off the DVM indication to the nearest microvolt and record this value as V
Disconnect the DVM negative input lead from the V
10,
connector on the 432A and
RF
reconnect it to chassis ground.
11.
Set the Power Meter POWER REF switch to ON (in) and record the indication observed on the DVM as V
Disconnect the DVM negative input lead from chassis ground and reconnect it to the
12.
V
connector on the rear panel of the 432A. The DVM is not setup to measure
RF
V
which represents the power reference oscillator output level.
1
13.
Calculate the value of V
1
.
COMP
equal to 1 milliwatt from the following equation:
.
0
Page 97
Model 436A
ADJUSTMENTS
5-22. POWER REFERENCE OSCILLATOR LEVEL ADJUSTMENT (cont’d)
where:
V
= previously recorded value
0
= previously recorded value
v
COMP
-3
10
= 1 milliwatt
R = previously recorded value
EFFECTIVE EFFICIENCY = value for thermistor mount at 50 MHz (traceable to the
National Bureau of Standards).
14. Remove the Power Meter top cover and adjust LEVEL ADJUST potentiometer
A8R4 so that the DVM indicates the calculated value of V
5-22. POWER REFERENCE OSCILLATOR LEVEL ADJUSTMENT (cont’d)
3.
TYPICAL
Derivation of Formula for V1 - V
0
CALCULATIONS
(cont’d)
Model 436A
Desired P
= 1mmW = 10
RF
-3
Let (4R) (EFFECTIVE EFFICIENCY) (10-3) = K
2
Substitute -(V
Then 0 = (V
l
- V0)2 for V
1
- VO)2 - 2V
COMP (Vl
- V
0
2
(see Math Assumptions under Accuracy)
1
- V0) + K
5-12
Page 99
Model 436A
Replaceable Parts
SECTION VI
REPLACEABLE PARTS
6-1. INTRODUCTION
6-2. This section contains information for ordering
parts. Table 6-1 lists abbreviations used in the parts
list and throughout the manual. Table 6-2 lists all
replaceable parts in reference designation order.
Table 6-3 contains the names and addresses that
correspond with the manufacturers’ code numbers.
6-3. ABBREVIATIONS
6-4. Table 6-1 lists abbreviations used in the parts
list, schematics and throughout the manual. In
some cases, two forms of the abbreviation are used,
one all in capital letters, and one partial or no
capitals. This occurs because the abbreviations in
the parts list are always all capitals. However, in
the schematics and other parts of the manual,
other abbreviation forms are used with both lower
case and upper case letters.
6-5. REPLACEABLE PARTS LIST
6-6. Table 6-2 is the list of replaceable parts and is
organized as follows:
Electrical assemblies and their compo-
a.
nents in alpha-numerical order by reference
designation.
b. Chassis-mounted parts in alpha-numerical
order by reference designation.
Miscellaneous parts.
c.
The information given for each part consists of the
following:
a.The Hewlett-Packard part number.
b. The total quantity (Qty) used in the
instrument.
c.The description of the part.
d. A typical manufacturer of the part in a
five-digit code.
6-7. ORDERING INFORMATION
6-8. To order a part listed in the replaceable parts
table, quote the Hewlett-Packard part number,
indicate the quantity required, and address the
order to the nearest Hewlett-Packard office.
6-9. To order a part that is not listed in the
replaceable parts table, include the instrument
model number, instrument serial number, the
description and function of the part, and the
number of parts required. Address the order to the
nearest Hewlett-Packard office.
6-10. PARTS PROVISIONING
6-11. Stocking spare parts for an instrument is
often done to ensure quick return to service after a
malfunction occurs. Hewlett-Packard has a Spare
Parts Kit available for this purpose. The kit consists
of selected replaceable assemblies and components
for this instrument. The contents of the kit and the
Recommended Spares list are based on failure
reports and repair data, and parts support for one
year. A complimentary Recommended Spares list
for this instrument may be obtained on request
and the Spare Parts Kit may be ordered through
your nearest Hewlett-Packard office.
6-12. DIRECT MAIL ORDER SYSTEM
6-13. Within the USA, Hewlett-Packard can supply
parts through a direct mail order system. Advan-
tages of using the system are:
a.
Direct ordering and shipment from the HP
Parts Center in Mountain View, California.
b. No maximum or minimum on any mail
order (there is a minimum order amount for parts
ordered through a local HP office when the orders
require billing and invoicing).
Prepaid transportation (there is a small
c.
handling charge for each order).
d. No invoices tages, a check or money order must accompany
each order.
to provide these advan-
e.
The manufacturer’s number for the part.
The total quantity for each part is given only once
at the first appearance of the part number in the
list.
6-14. Mail order forms and specific ordering
information is available through your local HP
office. Addresses and phone numbers are located at
the back of this manual.
6-1
Page 100
Replaceable Parts
Model 436A
Table 6-1. Reference Designations and Abbreviations (1 of 2)
REFERENCE DESIGNATIONS
A
. . . . . . . . . . .
AT. .
B. . .
BT. , . . . . . . . ..
c
“CP”lllllIllj:.aupler
CR . . . . . . . dtode; diode
DC .,.
DO .,.......
Da
A
ac
ACCBSS . . . . . accaeeory
ADJ, . . . . . .
AID
AF . . . . .
AFC, . . . . . ..autanattc
‘AGC . . . . .
~ AL
ALC. . . .
AM . . .
AMPL . . . . . .
APC, ,..
AaaY . . . . . . ..ae8etnbm
AUX. . . . . . .
~S . . . . . . . . . ..avemge
AW(3 . . . .
BAL. . . . . . . . ..
BCD . . . . . .
BD . . . . . . . . . . .. boud
BE CU
BFO. . . .
BH . . . . . . . . binder head
*KDN . . . . . .
BP . . . . . . . . .. bartdpeee
BPF .,...
Bra . . . . . . . . . . .
BWO, .,..
CAL. . . . . . .
“8EWR: :......,.
CHAN
cm . . . . . . . .. centimeter
CMO . .
COAX, . . . . . . .
attenuator;isolator;
termination
...0.. ,
thyrktor; varactor
direettortel
. . . . . . .
dgnaltng
(audtbla or vieuat);
lamp: LED
. . . . . . . . . .
... .
alternating cument
.,..
freauancy
control
. . . . . . . . . aluminum
.
control
amplttude modula-
tion
control
*uge
dectmat
. . . . . .
copper
.
Oeetllator
oesttttator
courtter-elockwtee
. . . . . . . . .
cabtnet
assembly
fan;motor
bettery
capacitor
coupler
delay
amtmeiator;
device
..ermae
adjustment
●
nalog-to-digital
audiofrequency
cotttzol
automatic gabs
automatic level
..atttpltfter
automatic phase
..auxUtary
American wtre
balance
binuy coded
basyllium
beat
freauenc
breakdown
bandpau ftlter
bac~werd-wave
..caltbrate
ceramic
channel
mount
.
COSULtd
line
brass
only
E
. . . . . . . .
electricef pert
F
FL”::::::::::::”
H
HA”:::::::::”
J. . .electricalconnector
(stationary Portion);
jack
K. . . . . . . . . . . .
L. . . . . . .
M
MP” ::::::: “r&~eihme%~
mecharttcaf part
COEF. . . . . .
COM . . . . . . . ..
COMP ,. ...
COMPL, . . . . .
CONN . . . . . . .
CP . . . . . .
CRT .,.
CTL . . . .
Cw . . . . .
Co w . . . . . . . .
cm.........centtmeter
D/A . . . .
dab . . . . . . . . . .
dBm . . . .
dc . . . . . . . direct current
deg , .
0
. . . . . . . . .
‘c ..,...
~F ,.. .
K
DEPC . .
DEB T . . . . . . . .
dlam. ., . . . .
y
DIA . . .
DIFF
div. , . . . . . . . . ..dtvidon
DPDT . . . . .
DR . . . . . . . . . . . . drtve
DSB, . . .
DTL . . . .
DVM
ECL .:::
EMF . ,
cathode-ray
complementary
traneiet.or logic
continuous
digttal-tc-analog
decibel refetred
to
1
mW
degree (temperature
interval or
ence)
angte )
(centigrade)
degree Fahrenheit
0 . . . . . .
depoeited
diameter (used in
Parfe list)
AMPL . . dffferentiat
arOPtifier
double-throw
double sideband
diode
Ioglc
digital
emitter coupled
logic
electromotive
miscellaneous
fuse
filter
herd ware
circulator
.
relay
colt;
tnductor
.
ABBREVIATIONS
.
coefficient
common
composition
.
complete
connector
eadmtum ptate
double-pole,
tube
wave
clockwise
decibel
diffe~
degree (ptane
degree Cefatue
degree Kelvin
carbon
detector
..diarneter
trandetor
voltmetm
force
P
R
RT” :: ::::::: “thermistor
s
T
TAB”::::::””
TC . . . . . .
TA P . . . . . . . . testpoint
EDP, . . . .
ELECT . . . . .
ENCAP , . . .
EXT ..,......
F
FET” ::: :::: “
F/Fflip-flop
AH.:::::::::
FIL H . . . . . fillister head
FM
FP . . . . . . . . front pane]
FREQ . . . . . . .
FAD. . . . . . . . . . . fixed
g
GE”: :::: :::” &na%%
GHz . . . . . , . . . gigahertz
GL . . . . . . . . . . . ..gbw.s
GRD . . . . . . .
H
h
HET
HEX . . . . . . . . hexagonaf
HD . . . . . . . . . . . .. head
HDW . . . . . . .. hardware
HF . . . . . .
HA . . . . . . . . . . mercury
HI . . . . . . . . . . . ..tdgh
HP
HPF”: :::.
HR . . . . . . .
HV . . . . . . . .
Hz . . . . . . . . . . .. Hertz
IC . . . .
ID . . . . . .
IF . . , . . . .
IMPG ., . . . impregnated
irt . . . . . . . . . . . . .. inch
INCD ., . . .
INCL
INP . : :;::::...
INS . . . . . . . .
electrical
. . .
(movableportion):
plug
. . . . . .
Q
transistor: SCR;
triodethyristor
. . . . . . . . . . . .
terminet board
thermocouple
electronic
processing
encapsulated
tranrdetor
. .
frequencymodulation
. . . . . . . . . . . . .
:::::: : “
high
Hewlett-Packard
IrIgh pass filter
hour(ueadin
parts
list)
highvoltage
integrated circuit
tneide diameter
intermediate
frerIuenc y
incandescent
connector
resistor
switch
transformer
data
electrolytic
external
farad
tieid-etfect
flat head
frequency
ground(ed)
henry
hour
h&&odyne
frequent y
tnclude(e)
input
insulation
u
. . . . .
integrated circuit:
velectrontube
w. . .
x
Y
INT. ..internafnternaf
kg . . . . . . . . .. kilogram
kHz
k~.:::::::::.
kV . . . . . . . . . ..lrilovolt
lb . . . . . . . . . . . .
LC . , . . . . . ,
LED . .
LF . . . . . .
LG . . . . . . . . . . . .. long
LH . . . . . . . . ..
LIM . . . . . . . . . . .. Iimit
LIN . . .
lin. . . . . . . . . . ..
LK
LO . . .
LOG . . . .
10g . . . . . . . .
LPF . . . . .
LV . . . . . . . . Iow voltage
m. . . . . .meter
mA . . . . . . . . milliampere
MAX
M~. :::: ::. .megohm
MEG ...,
MET FLM . . . . metal film
MET OX . .
MF. .
MFR . . . . . . manufacturer
reg . . . . . . . . .
MHz . . . . . . . .
mH . . . . . . . ..miUthenry
mho. , . . . . . . . . ..mho
MIN . . . . . . . .
miss . . . . .
. ..’....
MINAT. . . . . .
mm . . . . . . ..
microcircuit
All”:;;:’””
z. . .
WASH . .
voltage regulator;
breakdown
.
cable; transmission
path;
. . . . . . . . . . . .
. . . .
wire
crystal
electricor
.
tuned
circuit
capacitance
fight-em itting
lowfr:quency
linear taper (used
in perta
low;
local oscillator
logarithmic taper
(used in parts
meg
in parts
.
mediumfrequency;
mlcroferarl (ueed
parts list )
.
angle)
diode
socket
unit(piezo-
quartz)
cavtty;
tuned
kilohertz
kilohm
pound
inductance-
diode
left
hand
list)
linear
.
lock
washer
list)
logarithm
low
pass filter
(distance
.
maximum(106)(used
list)
metalfic
oxide
in
mffligram
megehertz
minimum
minute(time)
mtnute
(plane
.
miniature
millimeter
)
6-2
Aft
●
bbreviatlone in the
TfOTE
parts
liet wffl be
inupper-caee.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.