HP TS-3793/U Service and user manual

Page 1
TM 11-6625-2969-14&P
TECHNICAL MANUAL
OPERATOR’S, ORGANIZATIONAL, DIRECT SUPPORT
AND GENERAL SUPPORT MAINTENANCE MANUAL
(INCLUDING REPAIR PARTS AND SPECIAL TOOLS LISTS)
POWER METER TS-3793/U
(HEWLETT-PACKARD MODEL 436A)
(NSN 6625-01-033-5050)
9 MAY 80
Page 2
SAFETY STEPS TO FOLLOW IF SOMEONE IS THE VICTIM OF ELECTRICAL SHOCK
DO NOT TRY TO PULL OR GRAB THE lNDl-
VIDUAL
IF POSSIBLE, TURN OFF THE ELECTRICAL POWER
IF YOU CANNOT TURN OFF THE ELECTRICAL POWER, PULL, PUSH, OR LIFT THE PERSON TO SAFETY USING A WOODEN POLE OR A ROPE OR SOME OTHER INSULATING MATERIAL
SEND FOR HELP AS SOON AS POSSIBLE
AFTER THE INJURED PERSON IS FREE OF CON­TACT WITH THE SOURCE OF ELECTRICAL
SHOCK, MOVE THE PERSON A SHORT DISTANCE AWAY AND IMMEDIATELY START
ARTIFICIAL RESUSCITATION
Page 3
This manual includes copyright material reproduced by permission of the HEWLETT-PACKARD Company.
TM 11-6625-2969-14&P
TECHNICAL MANUAL
HEADQUARTERS
DEPARTMENT OF THE ARMY
No. 1l-6625-2969-14&P)
W
ASHINGTON
, DC,
9 May 1980
OPERATOR’S, ORGANIZATIONAL,
DIRECT SUPPORT AND GENERAL SUPPORT
MAINTENANCE MANUAL
(INCLUDING REPAIR PARTS
AND SPECIAL TOOLS LISTS)
POWER METER TS-3793/U
(HEWLETT-PACKARD MODEL 436A)
(NSN 6625-01-033-5050)
REPORTING OF ERRORS
You can improve this manual by recommending improvements using DA Form 2028-2 located in the back of the manual. Simply tear out the self-addressed form, fill it out as shown on the sample, fold it where shown, and drop it in the mail.
If there are no blank DA Forms 2028-2 in the back of your manual, use the standard DA Form 2028
(Recommended Changes to Publications and Blank Forms) and forward to the Commander, US Army Com-
munications and Electronics Materiel Readiness Command, ATTN: DRSEL-ME-MQ, Fort Monmouth, NJ
07703.
In either case a reply will be furnished direct to you.
SERIAL NUMBER
This manual applies directly to instruments with serial numbers prefixed 1606A, 1611A and 1629A. With changes described in section VII, this manual also applies to instruments with serial numbers prefixed 1447A, 1448A, 1451A, 1501A, 1503A, 1504A, 1505A, 1538, and 1550A. For additional important information about serial numbers, see INSTRUMENTS COVERED BY MANUAL in section I.
This manual is an authentication of the manufacturer’s commercial literature which, through usage, has been found to cover the data required to operate and maintain this equipment. Since the manual was not prepared in accordance with military specifications, the format has not been structured to consider levels of maintenance.
i
Page 4
Page 5
Model 436A
CONTENTS
Section
0
INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . 0-1
0-1
Scope
0-2
Indexes of Publications . . . . . 0-1
0-3
Forms and Records . . . . . . . . 0-1
0-4
0-5
Administrative Storage . . . . . 0-1
0-6
Destruction of Army Electronics
GENERAL INFORMATION . . . . . . . 1-1
I 1-1.
Introduction . . . . . . . . . . . . . . . . . 1-1
1-6.
Specifications . . . . . . . . . . . . 1-1
1-8.
Instruments Covered by Manual . . . . . . 1-1
1-14. Description . . . . . . . . . . . . 1-1
1-18. Options . . . . . . . . . . . . . . . . . . . . . . 1-3
1-19. 1-22. 1-24.
1-27. Accessories Supplied . . . . . . . . . 1-4
1-29. Equipment Required but not Supplied. . . . 1-4
1-31. Equipment Available . . . . . . . . . 1-4
1-34. Recommended Test Equipment . . . . . . 1-4
1-36. Safety Considerations . . . . . . . . . 1-4
II
2-1. Introduction . . . . . . . . . . . . 2-1
2-3. Initial Inspection . . . . . . . . . . . 2-1
2-5. Preparation for Use . . . . . . . . . . 2-1
2-6. Power Requirements . . . . . . . . 2-1
2-8. Line Voltage Selection . . . . . . . . 2-1
2-10. Power Cable . . . . . . . . . . . . . . . . . 2-1
2-12.
2-14. Interconnections . . . . . . . . . . 2-2
2-18.
2-21. Operating Environment. . . . . . . . 2-2
2-23. Bench Operation . . . . . . . . . . 2-2
2-25.
2-28. Storage and Shipment . . . . . . . . . 2-3
2-29. 2-31.
III OPERATION . . . . . . . . . . . . 3-1
3-1. Introduction . . . . . . . . . . . . 3-1
3-4. Panel Features. . . . . . . . . . . . 3-1
3-6. Operator’s Maintenance . . . . . . . . 3-1
3-8. Operator’s Checks . . . . . . . . . . 3-1
3-10. Local Operating Instructions . . . . . . . 3-1
3-12. Hewlett-Packard Interface Bus Remote
3-14.
3-18. DataMessages. . . . . . . . . . .3-20
3-21. Receiving Data Messages . . . . . . .3-20
3-32. Sending Data Messages from the Power
Input-Output Options . . . . . . . . 1-3
Cable Options. . . . . . . . . . . 1-3
Remote Control Options . . . . . . . 1-4
INSTALLATION . . . . . . . . . . . 2-1
Circuit Options . . . . . . . . . . 2-2
Mating Connectors . . . . . . . . . 2-2
Rack Mounting . . . . . . . . . . 2-2
Environment . . . . . . . . . . . 2-3
Packaging. . . . . .. . . . . . . . . . . . . . . . . . . . . 2-3
Operation . . . . . . . . . . . .3-20
Compatibility . . . . . . . . . . .3-20
Meter . . . . . . . . . . . .3-24
Page
0-1
0-1
0-1
Contents
Section
3-40. 3-42. 3-44. 3-46. 3-48.
3-50. 3-52.
3-54. 3-56. 3-58. 3-60.
3-62. Remote BCD Interface Operation . . . . . 3-29
3-65. 3-67.
3-73. Power Measurement Accuracy . . . . . 3-37
3-75. Sources of Error and Measurement .
3-81.
3-86.
IV 4-1. 4-3. 4-5. 4-7. 4-10. 4-11. 4-12. 4-13.
V 5-1. 5-4. 5-10. 5-12. 5-14. 5-16. 5-17. 5-18.
5-19. 5-20. 5-21.
5-22.
VI 6-1. 6-3. 6-5. 6-7.
Receiving the Trigger Message . . .
Receiving the Clear Message . . . .
Receiving the Remote Message . . .
Receiving the Local Message . . . .
Receiving the Local Lockout and Clear
Lockout Set Local Messages . . . Receiving the Pass Control Message . Sending the Required Service Message . Sending the Status Byte Message . .
Sending the Status Bit Message . . .
Receiving the Abort Message . . . .
Test of HP-IB Operation . . . . .
Output Data Format . . . . . . . . 3-36
BCD Remote Programming . . . . . . 3-36
Uncertainty . . . . . . . . . . . 3-37
Corrections for Error . . . . . . . . 3-38
Calculating Total Uncertainty . . . . . 3-38
PERFORMANCE TESTS . . . . . . . . 4-1
Introduction . . . . . . . . . . . . 4-1
Equipment Required . . . . . . . . . 4-1
Test Record . . . . . . . . . . . .. . . . . . . . . . . . 4-1
Performance Tests . . . . . . . . . . 4-1
Zero Carryover Test . . . . . . . . . . 4-2
Instrument Accuracy Test . . . . . . . . 4-3
Calibration Factor Test . . . . . . . . 4-6
Power Reference Level Test . . . . . . . 4-7
ADJUSTMENTS . . . . . . . . . . . 5-1
Introduction . . . . . . . . . . . . . . . . .. . . . . . . . . . . . 5-1
Safety Considerations . . . . . . . . . 5-1
Equipment Required . . . . . . . . . 5-1
Factory Selected Components . . . . . . 5-1
Adjustment Locations . . . . . . . . . 5-1
DC Offset Adjustment . . . . . . . . . 5-2
Auto Zero Offset Adjustment . . . . . . 5-3
Spike Balance Adjustment . . . . . . . 5-4
Multivibrator Adjustment . . . . . . . . 5-5
A-D Converter and Linear Meter Adjustment. . 5-6 Power Reference Oscillator
Frequency Adjustment . . . . . . . . 5-8
Power Reference Oscillator Level
Adjustment . . . . . . . . . . . . . . . . . . . . . . 5-9
REPLACEABLE PARTS . . . . . . . . 6-1
Introduction . . . . . . . . . . . . 6-1
Abbreviations . . . . . . . . . . . . 6-1
Replaceable Parts List . . . . . . . . . 6-1
Ordering Information . . . . . . . . . 6-1
. . . . 3-29
. . . . 3-29
. . . . 3-29
. . . .
. . . .
. . . . 3-29
. . . 3-29
. . . . 3-29
.. . . 3-29
. . .
. . . . 3-29
iii
Page
3-29
3-29
3-29
Page 6
Contents
CONTENTS (Cont’d)
Section
6-10. Parts Provisioning . . . . . . . . . . 6-1
6-12. Direct Mail Order System . . . . . . . . 6-1
6-13 Cross Reference Index . . . . . 6-21
MANUAL CHANGES . . . . . . . . . 7-1
VII 7-1.
Introduction . . . . . . . . . . . . 7-1
Manual Changes . . . . . . . . . . . 7-1
7-3. 7-6.
Manual Change Instructions . . . . . . . 7-1
VIII
SERVICE . . . . . . . . . . . . .8-1
8-1.
Introduction . . . . . . . . . . . 8-1
8-3.
Safety Considerations . . . . . . . . 8-1
8-9.
Recommended Test Equipment . . . . . 8-5
8-11.
Service Adds . . . . . . . . . . . 8-5
8-16.
Repair . . . . . . . . . . . . . 8-5
8-17. 8-19.
Factory Selected Components . . . . 8-5
Disassembly and Reassembly Procedures . 8-5
Page
APPENDIXES
Model 436A
Section
8-24. Basic Circuit Descriptions . . . . . . . 8-7
8-25. Linear Integrated Circuits . . . . . . 8-7
8-30. Digital Integrated Circuits and Symbols . 8-7
8-55. Troubleshooting . . . . . . . . . .
8-59. Standard Instrument Checkout . . . .
8-64. 8-68. 8-70. Block Diagram Circuit Descriptions . . . 8-71. 8-75. 8-87. 8-114. 8-117.
8-130. Data Mode Operation . . . . . .
8-155. Service Sheet 5 . . . . . . . . .
HP-IB Instrument Checkout . . . . .
BCD Instrument Checkout . . . . .
Service Sheet 1 . . . . . . . . .
Service Sheet 2 . . . . . . . . .
Service Sheet 3 . . . . . . . . .
Service Sheet 4 . . . . . . . . .
Command Mode Operation . . . . .
Page
8-16 8-16 8-69 8-117 8-123 8-123
8-124 8-128 8-150 8-150 8-153 8-163
APPENDIX A.
B.
Section I.
II.
APPENDIX C.
D.
Section I.
II.
III.
IV.
APPENDIX E.
Figure
HP Model 436A Power Meter and Accessories
1-1.
Supplied . . . . . . . . . . .1-0
Line Voltage Selection . . . . . . . . 2-1
2-1.
Power Cable HP Part Numbers Versus
2-2.
Mains Plugs Available . . . . . . . 2-2
Hewlett-Packard Interface Bus Connection . 2-6
2-3.
BCD Interface Connection. . . . . . . 2-7
2-4.
Front and Rear Panel Controls, Connectors,
3-1.
and Indicators. . . . . . . . . . 3-2
Operator’s Checks . . . . . . . . . 3-6
3-2.
Operating Instructions . . . . . . . . 3-16
3-3.
Power Meter Response Curve (Settling Time for
3-4.
Analog Circuits) . . . . . . . . . 3-26
Measurement Timing Flow Chart (Settling
3-5.
Time for Digital Circuitry) . . . . . 3-27
Operating Program Simplified Flow Chart. . 3-28
3-6.
References .. .. .. .. .. .. .. .. ..
Components of End Item List
Introduction
Integral Components of End Item . . . . . . . . . . B-2
Additional Authorization List Maintenance Allocation
Introduction
Maintenance Allocation Chart
Tools and Test Equipment Requirements .. .. .. D-5
Remarks (N/A) Expendable Supplies and Materials List (N/A)
ILLUSTRATIONS
Page
Figure
3-7. Test of HP-IB Operation Flowchart . . . 3-30
3-8. 436A Quick Programming Guide . . . . 3-31
3-9. Calculating Measurement Uncertainties . . 3-42 3-10. The Effect of Power Sensor Mismatch
3-11. Calculating Measurement Uncertainty
4-1. 4-2. 4-3. 4-4.
5-1. 5-2. 5-3. 5-4. 5-5.
Page A-1
B-1
(N/A)
D-1 D-3
Page
on Measurement Accuracy . . . . . . 3-44
(Uncertainty in dB Known) . . . . . 3-45
Zero Carryover Test Setup. . . . . . . 4-2
Instrument Accuracy Test Setup . . . . 4-4
Calibration Factor Test Setup . . . . . 4-6
Power Reference Level Test Setup . . . . 4-8
DC Offset Adjustment Setup . . . . . . 5-2
Auto Zero Offset Adjustment Setup . . . 5-3
Spike Balance Adjustment Setup . . . . 5-4
Multivibrator Adjustment Setup . . . . . 5-5
A-D Converter and Linear Meter
Adjustment Setup . . . . . . . . 5-6
iv
Page 7
Model 436A
Figure
Power Reference Oscillator Frequency
5-6.
Adjustment Setup . . . . . . .
Power Reference Oscillator Level
5-7.
Adjustment Setup . . . . . . .
Cabinet Parts . . . . . . . . . .
6-1.
Schematic Diagram Notes . . . . . .
8-1.
Front Panel Removal . . . . . . .
8-2.
Operational Amplifier Functional Circuits.
8-3.
Dual D-Type Flip-Flop . . . . . . .
8-4.
Four-Bit Bistable Latch. . . . . . .
8-5.
Dual J-K Master/Slave Flip-Flop and Gate
8-6.
Pulse Timing. . . . . . . . .
Dual J-K Edge-Triggered Flip-Flop . . .
8-7.
Programmable Counters . . . . . .
8-8.
3-Line to 8-Line Decoder . . . . . .
8-9.
8-Input Data Selector (Multiplexer) . . .
8-10.
LED Display Driver . . . . . . . .
8-11.
Numeric Display . . . . . . . . .
8-12.
MOS and TTL ROMs . . . . . . .
8-13.
Power Meter Operating Cycle . . . . .
8-14.
Operating Program Flow Chart . . . .
8-15.
HP-IB Verification Program
8-16.
(HP 9830A Calculator) . . . . . .
HP-IB Verification Program
8-17.
(HP 9820A Calculator) . . . . . .
Analog-to-Digital Converter Simplified
8-18.
Diagram and Waveforms . . . . .
HP-IB Listen Handshake Timing . . . .
8-19.
Data Valid Status Generator Timing . .
8-20.
Overall Block Diagram . . . . . . .
8-21.
AC Gain, A-D Converter and Display
8-22.
Circuits Block Diagram . . . . . .
Controller and Counters Block Diagram .
8-23.
HP-IB (Option 022) Circuit Block Diagram
8-24.
BCD Interface (Option 024) Circuit
8-25.
Block Diagram . . . . . . . .
AlAl Display Assembly Component
8-26.
Locations . . . . . . . . . . . 8-177
A1A2 Pushbutton Assembly
8-27.
Component Locations . . . . . . . 8-177
Front Panel Assembly
8-28.
Schematic Diagram . , . . . . . . 8-177
A2 AC Gain Assembly Component, Test
8-29.
Point, and Adjustment Locations . . . 8-179
ILLUSTRATIONS (Cont’d)
Page
Figure 8-30. AC Gain Assembly Schematic Diagram . .
5-8
8-31. A1A3 CAL FACTOR % Switch Assembly
Component Locations . . . . . . .
5-9
8-32. A3 A-D Converter Assembly
Components, Test Point, and
6-18
8-2 8-6 8-8 8-9 8-9
8-33. A-D Converter Assembly
8-34. A4 Counter Relative Assembly
8-35. Counter Relative Assembly
Adjustment Locations . . . . . . .
Schematic Diagram . . . . . . . .
Component and Test Point
Locations . . . . . . . . . . . . . . . . . . . . . . . . .
Schematic Diagram . . . . . . . .
8-10 8-10 8-11 8-12 8-13 8-14 8-15 8-15 8-17 8-25
8-70
8-36. A5 Controller Assembly Component
and Test Point Locations . . . . . .
8-37. Controller Assembly
Schematic Diagram . . . . . . . .
8-38. A6 HP-IB (Option 022) Control
Assembly Component and
Test Point Locations . . . . . . .
8-39. HP-IB (Option 022) Control Assembly
Schematic Diagram . . . . . . . .
8-40. A7 HP-IB (Option 022) Input/Output
Assembly Component and Test
Point Locations . . . . . . . . .
8-41. HP-IB (Option 022) Input/Output Assembly
Schematic Diagram . . . . . . . .
8-95
8-42. A6 BCD Interface Control (Option 024)
Assembly Component and
8-126 8-151 8-160 8-167
8-43. BCD Interface Control (Option 024)
8-44. A8 Power Reference Oscillator Assembly
Test Point Locations . . . . . . .
Assembly Schematic Diagram . . . .
Component, Test Point, and
8-169
8-171 8-173
8-175
8-45. Power Reference Oscillator Assembly
8-46. A9 Power Supply Rectifier and Regulator
Adjustment Locations . . . . . . .
Schematic Diagram . . . . . . . .
Assembly Component and
Test Point Locations . . . . .
8-47. Rear Panel Mounted Power Supply
Component Locations . . . . .
8-48. Power Supply Rectifier and Regulator
Assembly Schematic Diagram . . 8-49. Rear View of Front Panel (Removed) . 8-50. Top Internal View Standard Instrument 8-51. Top Internal View HP-IB or
BCD Interface . . . . . . .
-
Contents
Page
8-179
8-181
8-181
8-181
8-183
8-183
8-185
8-185
8-186
8-187
8-189
8-189
8-191
8-191
8-193
8-193
. . 8-195
. . 8-195
8-195
. . . . 8-197
8-197
. .
. . 8-197
v
Page 8
Contents
Table
1-1. 1-2.
2-1. 2-2.
3-1. 3-2. 3-3.
3-4.
3-5.
3-6.
3-7. 3-8.
4-1.
4-2.
TABLES
Page
Specifications . . . . . . . . . . . . 1-2
Recommended Test Equipment . . . . . . 1-5
Circuit Options . . . . . . . . . . . 2-4
USA Standard Code for Information
Interchange (ASCII) . . . . . . . . 2-5
Message Reference Table . . . . . . . . 3-21
Measurement Sequence . . . . . . . . 3-22
Hewlett-Packard Interface Bus
Input Program Codes . . . . . . . . 3-23
Hewlett-Packard Interface Bus
Output Data String . . . . . . . . . 3-25
Power Meter Remote Access Time to
First Output Data Character . . . . . . 3-26
Power Meter Output Data Printout for
HP 5055A Digital Recorder . . . . . . 3-36
BCD Output Data Codes . . . . . . . . 3-39
BCD Programming Commands . . . . . . 3-41
Zero Carryover Autorange Digital
Readout Results . . . . . . . . . . 4-3
Instrument Accuracy Test Results . . . . . 4-5
Model 436A
Table 4-3.
Instrument Accuracy Test Results
for dB [REF] Mode . . . . . . . . 4-5
4-4.
Calibration Factor Test Results . . . . . . 4-7
Performance Test Record . . . . . . . . 4-10
4-5.
Factory Selected Components . . . . . . 5-2
5-1. 6-1.
Reference Designators and Abbreviations . .
Replaceable Parts . . . . . . . . .
6-2. 6-3.
Code List of Manufacturers . . . . . .
Manual Changes by Serial Number . . . .
7-1. 8-1.
Logic Levels and Power Requirements . . .
8-2.
Program Mnemonic Descriptions . . . .
8-3.
Standard Instrument Checkout . . . . .
8-4.
HP-IB Circuit Troubleshooting . . . . .
8-5.
BCD Interface Option 024 Checkout . . .
8-6.
Operating Program Description . . . . .
8-7.
Up/Down Count Control Logic Steering . .
8-8.
Function Decoder Clock Selection . . . .
Programming Command Logic
8-9.
Operating Summary . . . . . . . .
Power Meter Talk HP-IB Output
8-10.
Data Format . . . . . . . . . . 8-161
Page
. ..6-2 . .6-4 . 6-19
. .7-1 . .8-9
.8-20 .8-52 .8-99 .8-117 .8-130
8-145
.8-155
8-155
SERVICE SHEETS
Service Sheet
1
Overall Block Diagram . . . . . . . . . . . . . . . . .8-166
2 AC Gain, A-D Converter, and Display Circuits
Block Diagram (A1A1, A2, A3) . . . . . . . .8-168
3 Controller and Counters Block Diagram
(A1A2, A4, A5) . . . . . . . . . . . . . . . . . . . . .8-170
4 HP-IB (Option 022) Circuit Block
Diagram (A6, A7) . . . . . . . . . . . . . . . . .8-172
5 BCD Interface (Option 024) Circuit
Block Diagram (A6, A7). . . . . . . . . . . . .8-174
Front Panel Assembly (A1A1, A1A2) . . . . . . .8-176
6
7 AC Gain Assembly (A2) . . . . . . . . . . . . . . . .8-178
Page
Service Sheet 8
A-D Converter Assembly (A1A3, A3) . . . . . . .8-180
9
Counter Relative Assembly (A4) . . . . . . . . . .8-182
10
Controller Assembly (A5). . . . . . . . . . . . . . .8-184
HP-IB (Option 022) Control Assembly (A6) . . .8-186
11
HP-IB (Option 022) Input/Output
12
Assembly (A7) . . . . . . . . . . . . . . . . . . . . . . .8-188
BCD Interface Control (Option 024)
13
Assembly (A6, A7) . . . . . . . . . . . . . . . .8-190
Power Reference Oscillator Assembly (A8). . . .8-192
14
Power Supply Rectifier and Regulator
15
Assembly (A9, A10) . . . . . . . . . . . . . . .8-194
Page
vi
Page 9
Model 436A
Safety Considerations
SAFETY CONSIDERATIONS
GENERAL
This product and related documentation must be reviewed for familiarization with safety markings and instructions before operation. This product has been designed and tested in accordance with inter­national standards.
SAFETY SYMBOLS
Instruction manual symbol: the
product will be marked with this symbol when it is necessary for the user to refer to the instruc­tion manual (refer to Section II of this manual.
Indicates hazardous voltages.
Indicates earth (ground) terminal.
The WARNING sign denotes a hazard. It calls attention to. a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.
The CAUTION sign denotes a hazard. It calls attention to an operating procedure, practice, or the like, which, if not correctly performed or adhered to, could result in damage to or destruc­tion of part or all of the product. Do not proceed beyond a CAU­TION sign until the indicated conditions are fully understood and met.
SAFETY EARTH GROUND
This is a Safety Class I product (provided with a protective earthing terminal). An uninterruptible
safety earth ground must be provided from the main power source to the product input wiring ter­minals, power cord, or supplied power cord set. Whenever it is likely that the protection has been impaired, the product must be made inoperative and be secured against any unintended operation.
BEFORE APPLYING POWER
Verify that the product is configured to match the available main power source per the input power configuration instructions provided in this manual.
If this product is to be energized via an auto-trans-
former make sure the common terminal is con­nected to the neutral (grounded side of mains supply).
SERVICING
Any servicing, adjustment, maintenance,
or repair of this product must be per­formed only by qualified personnel.
Adjustments described in this manual
may be performed with power supplied
to the product while protective covers are removed. Energy available at many points may, if contacted, result in per­sonal injury.
Capacitors inside this product may still
be charged even when disconnected from
its power source.
To avoid a fire hazard, only fuses with
the required current rating and of the specified type (normal blow, time delay,
etc.) are to be used for replacement.
vii
Page 10
Page 11
SECTION 0
INSTRUCTIONS
TM ll-6625-2969-14&P
0-1.
operation and maintenance instructions. TS-3793/U is referred to as the Hewlett-Packard Model 436A Power Meter.
0-2.
whether there are new editions, changes, or additional publications pertain-
ing to the equipment.
modification work orders (MWO’S) pertaining to the equipment.
0-3.
forms, records, and reports which are to be used by maintenance personnel at all levels of maintenance are listed in and prescribed by TM
NAVSUPINST 4030.29/AFR 71-13/MCO P4030.29A and DSAR 4145.8.
SCOPE .
This manual describes Power Meter TS-3793/U (fig. 1-1) and provides
Throughout this manual, the
INDEXES OF PUBLICATIONS.
a.
DA Pam 310-4.
b.
DA Pam 310-7.
FORMS AND RECORDS.
a.
Reports of Maintenance and Unsatisfactory Equipment. Maintenance
Refer to the latest issue of DA Pam
Refer to DA Pam 310-7 to determine whether there are
310-4
to determine
38-750.
Report of Packaging and Handling Deficiencies. Fill out and forward
b. Report of Packaging Improvement Report) as prescribed in AR 700-58/
c. Discrepancy in Shipment Report (DISREP) (SF 361). Fill out and
forward Discrepancy in Shipment Report (DISREP) (SF 361) as prescribed in
AR 55-38/NAVSUPINST 4610.33A/AFR 75-18/MCO P4610.19B and DSAR 4500.15.
0-4.
Instructions for preparing EIR’s are provided in TM 38-750, The
Maintenance Management System. EIR’s should be mailed directly to
Commander, US Army Communications and Electronics Materiel Readiness Command and Fort Monmouth, ATTN:
A reply will be furnished directly to you.
0-5.
shall be in accordance with TM 740-90-1.
0-6.
accordance with TM 750-244-2.
REPORTING OF EQUIPMENT IMPROVEMENT RECOMMENDATIONS (EIR).
EIR’s will be prepared using DA Form 2407, Maintenance Requiest.
Army
DRSEL-ME-MQ, Fort Monmouth, NJ 07703.
ADMINISTRATIVE STORAGE.
Administrative storage of equipment issued to and used by Army
DESTRUCTION OF ARMY ELECTRONICS MATERIEL.
Destruction of Army Electronics materiel to prevent enemy use shall be in
activities
0-1
Page 12
Page 13
Page 14
General Information
Model 436A
1-0
Figure 1-1. HP Model 436A Power Meter and Accessories Supplied
Page 15
Model 436A
General Information
SECTION I
GENERAL INFORMATION
1-1. INTRODUCTION
1-2. This manual provides information pertaining to the installation, operation, testing, adjustment and maintenance of the HP Model 436A Power Meter.
1-3. Figure 1-1 shows the Power Meter with accessories supplied.
1-4. Packaged with this manual is an Operating Information Supplement. This is simply a copy of the first three sections of this manual. This supplement should be kept with the instrument for use by the operator. Additional copies of the
Operating ordered through your nearest Hewlett-Packard office. The part numbers are listed on the title page of this manual.
1-5. On the title page of this manual, below the
manual part number, is a “Microfiche” part number. This number may be used to order
4x6-inch microfilm transparencies of the manual.
The microfiche package also includes the latest Manual Changes supplement as well as all pertinent
Service Notes.
1-6. SPECIFICATIONS
1-7. Instrument Table 1-1. These specifications are the performance standards or limits against which the instrument may be tested.
1-8. INSTRUMENTS COVERED BY MANUAL
1-9. Power Meter Options 002, 003, 009, 010, 011, 012, 013, 022, and 024 are documented in this manual. The differences are noted in the appropriate location such as OPTIONS in Section I, the Replaceable Parts List, and the schematic diagrams.
Information Supplement may be
specifications
are listed in
number prefix(es) as listed under SERIAL NUMBERS on the title page.
1-11. An instrument manufactured after the printing of this manual may have a serial prefix that is not listed on the title page. This unlisted serial prefix indicates that the instrument is different from those documented in this manual. The manual for this instrument is supplied with a yellow Manual Changes supplement that contains change differences.
1-12. In addition to change information, the supplement may contain information for cor­recting errors in the manual. To keep this manual as current and accurate as possible, Hewlett­Packard recommends that you periodically request the latest Manual Changes supplement. The supplement for this manual is keyed to the manual’s print date and part number, both of which appear on the title page. Complimentary copies Hewlett-Packard.
1-13. For information concerning a serial number prefix not listed on the title page or in the Manual Changes supplement, contact your nearest Hewlett­Packard office.
1-14. DESCRIPTION
1-15. The Power Meter is a precision digital­readout instrument capable of automatic and manual measurement of RF and Microwave power levels. It is designed for interconnection with a compatible Power Sensor (refer to Table 1-1,
Specifications) to form a complete power measure­ment system. The frequency and power range of the system are determined by the particular Power
Sensor selected for use. With the Power Sensors available, the overall frequency range of the system is 100 kHz to 18 GHz, and the overall power range
is -70 to +35 dBm.
information
of the supplement are available from
that
documents
the
1-10. This instrument has a two-part serial number. The first four digits and the letter comprise the , serial number prefix. The last five digits form the sequential suffix that is unique to each instrument. The contents of this manual apply directly to instruments having the same serial
1-16. Significant operating features of the Power
Meter are as follows:
Digital Display:
seven-segment LED, plus a sign when in the
dBm or dB (REL) mode. It also has under- and
The display is a four-digit,
1-1
Page 16
General Information
Model 436A
Table 1-1. Specifications
SPECIFICATIONS
Frequency Range:
100 kHz to 18 GHz (depending on power sensor used).
Power Range:
(display calibrated in watts, dBm, and dB relative to reference power level).
With 8481A, 8482A, or 8483A sensors:
50 dB with 5 full scale ranges of –20, –10, O, 10, and 20 dBm
(l0pW
to 100 mW).
With 8481H or 8482H sensors:
45 dB with 5 full scale
ranges of 0, 10, 20, 30 and 35 dBm (1 mW to 3W).
With 8484A sensor:
50 dB with 5 full scale ranges of
-60, -50, -40, -30, and -20 dBm (1 nW to
10/Jw).
Accuracy:
Instrumentation1:
Watt mode: ±0.5%. dBm mode: ±0.02 dB ±0.001 dB/°C. dB [REL] mode2: ±0.02 dB ±0.001 dB/°C.
Zero:
Automatic, operated by front panel switch.
Zero set:
±0.5% of full scale on most sensitive range.
typical, ±1 count on other ranges.
Zero carry over:
±0.2% of full scale when zeroed on
the most sensitive range.
Noise
(typical, at constant temperature, peak change over any one-minute interval): 20 pW (8484A); 40 nW (8481A, 8482A, 8483A); 4 pW (8481H,
8482H).
Drift
(1 hour, typical, at constant temperature after 24-hour warm-up); 20 pW (8484A); 10 nW (8481A, 8482A, 8483A); 1.0 0W (8481H, 8482 H).
Power Reference:
Internal 50 MHz oscillator with Type N Female connector on front panel (or rear panel, Option 003 only).
Power output: 1.00 mW.
Factory set to ±0.7%, traceable to the National Bureau of Standards. Accuracy: ±1.2% worst case (±0.9% rss) for one year (0°C to 55°C).
Response Time:
(0 to 99% of reading, five time constants) Range 1 (most sensitive) <10 seconds.
Range 2 Range 3-5
<1 second <100 milliseconds.
(Typical, measured at recorder output).
Cal Factor:
16-position switch normalizes meter reading to ac­count for calibration factor or effective efficiency, Range 85% to 100% in 1% steps.
Cal Adjustment:
Front panel adjustment provides capability to adjust gain of meter to match power sensor in use.
Recorder Output:
Proportional to indicated power with 1 volt corre-
sponding to full scale and 0.316 volts to -5 dB;
kfl
output impedance, BNC connector.
1
RF Blanking Output:
Open collector TTL; low corresponds to blanking
when auto-zero mode is engaged.
Display:
Digital display with four digits, 20% over-range capa-
bility on all ranges. Also, uncalibrated analog peak-
ing meter to see fast changes.
Power Consumption:
100, 120, 220, or 240 V + 5%, -10%, 48 to 440 Hz,
less than 20 watts (<23 watts with Option 022, or
024).
Dimensions:
134 mm High (5-1/4 inches). 213 mm Wide (8-3/8 inches). 279 mm Deep (11 inches).
Net Weight:
4.5 kg (10 lbs).
1-2
1
Includes sensor non-linearity. Add +1.5 -1.0% on top range when using the 8481A, 8482A, or 8483A power sensors.
2
Specifications are for within range measurements. For range-to-range accuracy add the range uncertainties.
Page 17
Model 436A
General Information
DESCRIPTION (cont’d).
overrange indicators. overrange capability
There is a 20 percent in all ranges.
Large 10 mm (0.375 inch) digits are easy to see even in a high glare environment.
Auxiliary Meter:
Complements the digital display by showing fast changes in power level. Ideal for
“peaking”
transmitter output or
other variable power devices.
Choice of Display in Watts, dBm or dB:
Absolute power can be read out in watts or dBm. Relative power measurements are made possibile with the dB [REF] switch. Pressing this switch zeros the display for any applied input power and any deviation from this reference is shown in dB with a resolution of ±0.01 dB. This capability is particularly useful in frequency response testing.
Power Units and Mode Annunciator:
annunciator
provides
error-free
The units
display interpretation by indicating appropriate power units in the watt mode. The mode annunciator indicates the mode of operation: dBm, dB (REL), ZERO or REMOTE.
Completely Autoranging: The Power Meter
automatically switches through its 5 ranges to provide completely
“hands off” operation. The RANGE HOLD switch locks the Power Meter in one of its ranges when autoranging is
not desired.
Automatic Sensor Recognition: The Power
Meter continually decodes the sensitivity of the Power Sensor to which it is connected. This information is then used to automatically control the digital display decimal point location and, when WATT MODE operation is selected, to light the appropriate power units annunciator.
Auto Zero:
Zeroing the meter is accomplished by merely depressing the SENSOR ZERO switch and waiting until the display shows all zeros before releasing it. The meter is ready to make measurements as soon as the zero light in the mode annunciator goes off.
RF Blanking Output:
Open collector TTL; low corresponds to blanking when the sensor zero is engaged, ” May be used to remove the RF input signal connected to the power sensor.
Calibration Accuracy: A 1.00 mW, 50 MHz
reference output is available at the front panel
for calibrating the Power Meter and the Power Sensor as a system. Calibration is accomp­lished using the CAL ADJ and CAL FACTOR % controls. The CAL ADJ control compen­sates for slight differences in sensitivity associated with a particular type of Power Sensor and the CAL FACTOR % control compensates for mismatch losses and effective efficiency over the frequency range of the Power Sensor.
Recorder Output:
Provides a linear output with respect to the input power level. For each range, a +1.00 Vdc output corresponds to a full scale input power level. Refer to Table 1-1, Specifications, for the full-scale range values associated with the various types
of Power Sensors available.
1-17. Two programming interfaces are available as options for the Power Meter - a Hewlett-Packard Interface Bus (HP-IB) Option 022; and a BCD
Interface, Option 024. Both interfaces allow full remote control of all the power meter functions (CAL FACTOR can be programmed to either
100% or the CAL FACTOR which has been manually set on the front panel). These options may be added by the user at a later time as his requirements grow.
1-18. OPTIONS
1-19. Input-Output Options
1-20. Option 002.
A rear panel input connector is connected in parallel with the front panel input connector.
1-21. Option 003. A rear panel input connector
replaces the standard front panel input connector; a rear panel POWER REF OUTPUT connector replaces the standard front panel connector.
1-22. Cable Options
1-23. A 1.5
metre (5 ft.) Power Sensor Cable is normally supplied. The 1.5 metre cable is omitted with any cable option. The options and cable lengths are shown in the table below.
Option
009 010 011 012 013
I
Cable Length
3.0 m (10 ft)
6.1 m (20 ft)
15.2 m (50 ft)
30.5 m (100 ft)
61.0 m (200 ft)
1-3
Page 18
General Information
Model 436A
1-24. Remote Control Options
1-25. Options 022 and 024 add remote interface capability to the Power Meter. Option 022 is compatible with the Hewlett-Packard Interface Bus (AH1, C0, DC2, DT1, L2, LE0, PP0, RL2, SH1, SR0, T3, TE0); Option 024 uses dedicated input/output lines to enable remote programming
and to provide parallel, BCD-coded output data.
1-26. Option 022 or 024 may be ordered in kit
form under HP part numbers 00436-60035 and 00436-60034 respectively. Each kit contains a control assembly printed-circuit board, an input/ output assembly printed circuit board, and a data
cable for interconnection.
1-27. ACCESSORIES SUPPLIED
1-28. The accessories supplied with the Power
Meter are shown in Figure 1-1.
a.
The 1.5 metre (5 ft.) Power Sensor Cable, HP 00436-60026, is used to couple the Power Sensor to the Power Meter. The 1.5 metre cable is
omitted with any cable option.
b. The line power cable may be supplied in
one of four configurations. Refer to the paragraph
entitled Power Cables in Section II.
c. An alignment tool for adjusting the CAL
ADJ front panel control (HP Part No. 8710-0630).
1-29. EQUIPMENT REQUIRED BUT NOT
SUPPLIED
1-30. To form a complete RF power measurement system, a Power Sensor such as the HP Model 8481A must be connected to the Power Meter via the Power Sensor cable.
1-31.
EQUIPMENT AVAILABLE
1-32.
The HP Model 11683A Range Calibrator is recommended for performance testing, adjusting, and troubleshooting the Power Meter. The Power Meter’s range-to-range accuracy and auto-zero
operation can easily be verified with the Calibrator. It also has the capability of supplying a full-scale
test signal for each range.
1-33. Two extender boards (HP Part Numbers
5060-0258, and 5060-0990; 24 and 44 pins respectively) enable the Power Meter printed circuit assemblies to be accessed for service. Rubber bumpers (HP Part No. 0403-0115) should be in­stalled on the extender boards to prevent the boards from touching.
1-34. RECOMMENDED TEST EQUIPMENT
1-35. The test equipment shown in Table 1-2 is recommended for use during performance testing, adjustments, optimum performance of the Power Meter, the specifications of a substitute instrument must equal or exceed the critical specifications shown in the table.
1-36. SAFETY CONSIDERATIONS
1-37. The Power Meter is a Safety Class I instrument. This instrument has been designed according to international safety standards.
1-38. This operating and service manual contains information, cautions, and warnings which must be followed by the user to ensure safe operation and to retain the instrument in safe condition.
and troubleshooting. To ensure
1-4
Page 19
Model 436A
General Information
Table 1-2. Recommended Test Equipment
Instrument
Type
Range Calibrator
Digital Voltmeter
Power Meter
Thermistor Mount
Counter
Oscilloscope
Critical Specifications
Chopped dc output for each range referenced
Function: DC, resistance
Range Resistance: 200 ohms Vdc: 100 m Vdc, 1000 mVdc, 10 Vdc, 100 Vdc
10MQ
input impedance
6-digit resolution (±0.05% of reading, +0.02% of range)
Range: 1 mW
Transfer Accuracy (input -to-output): 0.2% SWR: 1.05,50 MHz
Accuracy: Frequency Range: 220 Hz, 50 MHz HP 5245L A
Sensitivity: 100 m Vrms Accuracy: 0.01%
Bandwidth: dc to 50 MHz HP 180C/ T Vertical Sensitivity: 0.2 V/division Horizontal Sensitivity: 1 ms/division
±0.5% at 50 MHz **
to 1 mW range
Suggested
Model
HP 11683A HP 3490A
HP 432A
HP 478A-H75
1801A/1821A
P,A,T
P,A,T
P, A
P, A
Use *
Logic Analyzer
*P = Performance Tests; A = Adjustments; T = Troubleshooting
**Traceable to the National Bureau of Standards
Clock Input: 60 kHz Trigger Word: 8 Bits Bit Input: TTL
Display Word: 8 Bits
HP 1601L
T
1-5/1-6
Page 20
Page 21
Model 436A
Installation
SECTION II
INSTALLATION
2-1. INTRODUCTION
2-2. This section provides all information neces-
sary to install the Power Meter. Covered in the section are initial inspection, power requirements, line voltage selection, interconnection, circuit
options, mounting, storage, and repackaging for
shipment.
2-3. INITIAL INSPECTION
2-4.
Inspect the shipping container for damage. If the shipping container or cushioning material is damaged, it should be kept until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically. The contents of the shipment should be as shown in Figure 1-1. Procedures for checking electrical performance are given in Section IV. If the contents are incomplete, if there is mechanical damage or defect, or if the instru­ment does not pass the electrical performance test, notify the nearest Hewlett-Packard office. If the shipping container is damaged, or the cushioning material shows signs of stress, notify the carrier as well as the Hewlett-Packard office. Keep the shipping materials for the carrier’s inspection.
2-5. PREPARATION FOR USE 2-6. Power Requirements
2-7. The Power Meter requires a power source of 100, 120, 220, or 240 Vac, +5%, -0%, 48 to 440 Hz single phase. Power consumption is approx­imately 20 watts.
2-9. Figure 2-1 provides instructions for line voltage and fuse selection. The line voltage selection card and the proper fuse are factory installed for 120 Vac operation.
SELECTION OF OPERATING VOLTAGE
1. Open cover door, pull the FUSE PULL lever end rotate to left. Remove the fuse.
2. Remove the Line Voltage Selection Card. Position the card so the line voltage appears at top-left corner. Push the card firmly into the slot.
3. Rotate the Fuse Pull lever to its normal posi­tion. Insert a fuse of the correct value in the holder. Close the cover door.
Figure 2-1. Line Voltage Selection
If this instrument is to be energized via an autotransformer for voltage reduction,
make sure the common terminal is connected to the earthed pole of the
power source.
2-8. Line Voltage Selection
BEFORE SWITCHING ON THIS IN-
STR UMENT, make sure the instrument is set to the voltage of the power source.
2-10. Power Cable
BEFORE SWITCHING ON
THIS IN-
STRUMENT, the protective earth termi-
nals of this instrument must be connected to the protective conductor of the (mains)
power cord. The mains plug shall only be
inserted in a socket outlet provided with a
protective earth con tact. The protective
action must not be negated by the use of an extension cord (power cable) without a
protective conductor (grounding).
2-1
Page 22
Installation
Model 436A
Power Cable (cont’d)
2-11. In accordance with international safety standards, this instrument is equipped with a three­wire power cable.
When connected to an appropriate ac power receptacle, this cable grounds the instrument cabinet. The type of power cable plug shipped with each instrument depends on the
8481H Power Sensors provide identical full scale outputs in response to input signal levels of 100
milliwatts and 3 watts, respectively.
erence in their sensitivity codes is detected by the Power Meter, however, and the Power Meter digital
readout is automatically configured to indicate the
appropriate value.
country of destination. Refer to Figure 2-2 for the
part numbers of the power cable plugs available.
2-16. Hewlett-Packard Interface Bus Option 022.
Interconnection data for Hewlett-Packard Interface Bus Option 022 is provided in Figure 2-3. Power Meter programming and output data format is described in Section III, Operation.
2-17. BCD Interface Bus Option 024. Inter-
connection data for BCD Interface Option 024 is provided in Figure 2-4. Power Meter programming
8120-1378
8120-1351
and output data format is described in Section III, Operation.
2-18. Mating Connectors 2-19. Interface Connectors.
nectors for Options 022 and 024 are indicated in Figures 2-3 and 2-4, respectively.
The dif-
Interface mating con-
8120-1369
8120-1689
2-20. Coaxial Connectors. Coaxial mating con-
nectors used with the Power Meter should be US MIL-C-39012-compatible type N male or 50-ohm
Figure 2-2. Power Cable HP Part Numbers
Versus Mains Plugs Available
2-12. Circuit Options
2-13. Jumper options are available for selecting a
BNC male.
2-21. Operating Environment
2-22. The operating environment should be within the following limitations:
filtered or unfiltered dc RECORDER OUTPUT,
for changing the TALK and LISTEN addresses
when Hewlett-Packard Interface Bus Option 022 is installed, and for selecting the desired pro-
Temperature . . . . . . . . . . . . 0°C to +55°C
Humidity . . . , . . . . . . . . . <95% relative
Altitude . . . . . . . . . .
gramming of the SENSOR ZERO function when
BCD Interface Option 024 is installed. Table 2-1 lists the factory installed jumper connections and indicates how they may be reconnected to select the options.
2-23. Bench Operation
2-24. The instrument cabinet has plastic feet and a fold-away tilt stand for convenience in bench operation. (The plastic feet are shaped to ensure self-aligning of the instruments when stacked. ) The
2-14. Interconnections 2-15. Power Sensor.
For proper system operation, the Power Sensor must be connected to the Power Meter using either the Power Sensor cable supplied with the Power Meter or any of the optional Power Sensor cables specified in Section I. Each of these cables employs a sensitivity line to enable the Power Meter to determine the operating range of the Power Sensor and thus, the true value of the input signal. For example, the 8481A and
tilt stand raises the front of the instrument for easier viewing of the control panel.
2-25. Rack Mounting
2-26. Instruments that are narrower than full rack width may be rack mounted using Hewlett-Packard sub-module cabinets. If it is desired to rack mount one Power Meter by itself, order half-module kit, HP Part Number 5061-0057. If it is desired to rack mount two Power Meters side by side, order the following items:
<4570 m (15,000 ft)
2-2
Page 23
Model 436A
Installation
Rack Mounting (cont’d)
Rack Mount Flange Kit (two provided)
HP Part Number 5020-8862.
Front Horizontal Lock Links (four pro-
b.
vided) HP Part Number 0050-0515.
Rear Horizontal Lock Links (two pro-
vided HP Part Number 0050-0516.
2-27 In addition to the rack mounting hardware, a front handle assembly (two provided) is also available for the Power Meter. The part number is
HP 5060-9899.
2-28. STORAGE AND SHIPMENT 2-29. Environment
2-30. The instrument should be stored in a clean dry environment. The following environmental limitations apply to both storage and shipment:
Temperature . . . . . . . . . . . -40°C to +75°C
Humidity . . . . . . . . . . . . . <95% relative
Altitude . . . . . . . . . .
2-31. Packaging 2-32. Original Packaging.
<7620 m (25,000 ft)
Containers and materials identical to those used in factory packaging are available through Hewlett-Packard offices. If the instrument is being returned to Hewlett-Packard for servicing, attach a tag indicating the type of
service required,
return address, model number, and full serial number. Also mark the container FRAGILE to assure careful handling. In any correspondence refer to the instrument by model number and full serial number.
2-33. Other Packaging.
The following general instructions should be used for re-packaging with commercially available materials:
Wrap the instrument in heavy paper or
a. plastic. (If shipping to a Hewlett-Packard office or service center, attach a tag indicating the service required, return address, model number, and full serial number.)
b. Use a strong shipping container. A double­wall carton made of 275-lb test material is
adequate.
c.
Use enough shock-absorbing material (3 to 4-inch layer) around all sides of instrument to provide firm cushion and prevent movement in the container.
Protect the control panel with
cardboard.
d. Seal the shipping container securely.
Mark the shipping container FRAGILE to
e.
assure careful handling.
2-3
Page 24
Installation
Model 436A
Table 2-1. Circuit Options
Assembly
A-D Converter Assembly A3
BCD Interface Control Assembly A6 (Option 024)
Hewlett-Packard Interface Bus Con­trol Assembly A6
(Option 022)
Service
Sheet
8
13
11
Jumper Functions
The factory-installed jumpers provide a filtered dc RECORDER OUTPUT which corresponds to the average power input to the Power Sensor. If ex­ternal filtering is desired, reconnect the jumpers to provide the optional un­filtered dc RECORDER OUTPUT as shown on Service Sheet 8.
The factory-installed jumper enables the SENSOR ZERO function to be programmed only when the REMOTE ENABLE input to the Power Meter is low. If it is desired to program the SENSOR ZERO function independ­ently of the remote enable input, reconnect the jumper to provide +5 V to U12C-9 as shown on Service Sheet 8.
The factory installed jumpers select TALK address M and LISTEN address
-
(minus sign) for the Power Meter. As shown on Service Sheet 11, either of these addresses causes a high enable output at U2C-10. If it is desired to change these addresses, refer to Service Sheet 11 and Table 2-2 and reconnect the jumpers to decode the appropriate ASCII characters. For example, to change to TALK address E and LISTEN address 70, the jumpers would be reconnected as follows.
ASCII code (logic 1=0V)
DDDDDDD 1111111 0000000 7654321
M E
-
%
Jumpers
M
-
1001101 1000101
0101101 0100101
U1B-13 U1B-12 U1B-10 U1B- 9 U2C- 9
HI01 LI02 HI03 HI04 HI05
Note: DI07 and DI06 must always be 1 and 0, respectively, for TALK address.
Note: DI07 and DI06 must always be 0 and 1, respectively, for LISTEN address.
E, % , Disconnect jumper from HI04 and reconnect to LI04.
2-4
Page 25
Model 436A
Installation
Table 2-2. USA Standard Code for Information Interchange (ASCII)
NOTE
3
0 0
0 0 0 0 0 0 0 0
1
0
1
0
1
0
1
1
0 0
1
0 0
1
0
1
0 1 1 1 1 1
1
0
0
1
0
1
1
1
0
0
1
0
0
1
1
1
0 1
0
1
1
1
0
0
1
0
0
1
0
2 3
4
5 6 7 8
9
10 11 12 13 14
NUL
1
SOH STX DC2 ETX EOT ENQ
ACK
BEL BS HT LF SUB
VT FF CR GS SO
D LE DC1
DC3 DC4 NAK
SYN
ETB CAN EM
ESC FS
RS
SP !
“ # $ 4
% & ‘ ( ) *
+
,
0
1 A 2 3
5 E
6
7 8
9
< = >
@
B C S D
F G H
I
J
K
L M N
P Q R
T
UI V w
x Y Z
[ \
a b c d e f
g h
i
i
k
l m n
p q
r s t
u v w x
y z
(
;
)
1 1 1
1
15
SI
US
NOTE 3
NOTE 1: HP-IB valid LISTEN addresses NOTE 2: HP-IB valid TALK addresses NOTE 3: Logic 1 = OV
I
?
O
NOTE 1 NOTE 2
o
DEL
2-5
Page 26
Installation
Model 436A
Logic Levels
The Hewlett-Packard Interface Bus logic levels are TTL compatible, i.e., the true (1) state is 0.0 Vdc to 0.4 Vdc and the false (0) state is
+2.5 Vdc to +5.0 Vdc.
Programming and Output Data Format
Refer to Section III, Operation.
Mating Connector
HP 1251-0293; Amphenol 57-30240.
Mating Cables Available
HP 10631A, 1.0 metre (3 ft.); HP 10631B, 2.0 metres (6 ft.)
HP 10631C, 4.0 metres (12 ft.); HP 10631D, 0.5 metre (1.5 ft.)
Cabling Restrictions
A Hewlett-Packard Interface Bus System may contain no more
1. than 1.8 metres (6 ft.) of connecting cable per instrument.
2.
The maximum accumulative length of connecting cable for any Hewlett-Packard Interface Bus System is 20.0 metres (65.6 ft.)
2-6
Figure 2-3. Hewlett-Packard Interface Bus Connection
Page 27
Model 436A
Installation
Logic Levels
The BCD Interface logic levels are TTL compatible, i.e., the true state is 0.0 Vdc to
0.4 Vdc and the false state is +2.5 Vdc to
+5.0 Vdc.
Figure 2-4. BCD Interface Connection
Programming and Output Data Format
Refer to Section III, Operation
Mating Connectors Mating Cables Available
- HP 1251-0086
- HP 562A-16C
for 5055A Printer
2-7/2-8
Page 28
Page 29
Model 436A
Operation
SECTION Ill OPERATION
3-1. INTRODUCTION
3-2. This section provides complete operating information for the Power Meter. Included in the section are a description of all front- and rear-panel controls, features), operator’s checks, operating instructions, power measurement accuracy considerations, and operator’s maintenance.
3-3. Since the power Meter can be operated locally as well as remotely via Hewlett-Packard Interface Bus Option 022 or BCD Remote Interface Option 024, respectively, the information in this section is arranged accordingly. All information unique to a
particular operating configuration is designated as such; where no distinction is made, the informa­ion is applicable to both standard and optional instrument operation.
3-4. PANEL FEATURES
3-5. Front and rear panel features of the Power Meter are described in Figure 3-1. This figure contains a detailed description of the controls, connectors and indicators.
3-6. OPERATOR’S MAINTENANCE
3-7. The only maintenance the operator should normally perform is replacement of the primary power fuse located within Line Module Assembly A11. For instructions on how to change the fuse, refer to Section II, Line Voltage Selection.
Make sure that only fuses with the required rated current and of the specified
connectors, and
indicators (panel
type (normal blow, time delay, etc.) are
used for replacement. The use of repaired fuses and the short-circuiting of fuse­holders must be avoided.
3-8. OPERATOR’S CHECKS
3-9. A procedure for verifying the major func­tions of the Power Meter is provided in Fig­ure 3-2. parts: Local Operation, Remote BCD Operation, and Remote Hewlett-Packard Interface Bus Opera­tion. For a standard instrument it is only necessary to perform the Local Operation procedure. For units equipped with either of the remote options, the Local Operation procedure should be per­formed first to establish a reference against which remote operation can be verified. Information covering remote programming of the Power Meter is provided in the following paragraphs, and a Hewlett-Packard Interface Bus Verification Pro­gram is provided in Section VIII, Service.
3-10. LOCAL OPERATING INSTRUCTIONS
The procedure is divided into three
3-11. Figure 3-3 provides general instructions for operating the Power Meter via the front-panel controls.
Any
interruption
of the protective (grounding) conductor (inside or outside the instrument) or disconnecting the
protective earth terminal is likely to make
this instrument dangerous. Intentional interruption is prohibited.
3-1
Page 30
Operation
Model 436A
FRONT AND REAR PANEL FEATURES
3-2
Figure 3-1. Front and Rear Panel Controls, Connector, and Indicators (1 of 4)
Page 31
Figure 3-1.
3-3
Page 32
Operation
Model 436A
FRONT AND REAR PANEL FEATURES
3-4
Figure 3-1. Front and Rear Panel Controls, Connector, and lndicators (3 of 4)
Page 33
Figure 3-1.
Figure 2-1
Table 2-2
3-5
Page 34
Operation
LOCAL OPERATION
Model 436A
OPERATOR’S CHECKS
3-6
BEFORE CONNECTING LINE POWER TO THIS INSTRU­MENT, ensure that all devices connected to this instrument are connected to the protective (earth) ground.
BEFORE SWITCHING ON THIS INSTRUMENT, ensure that
the line power (mains) plug is connected to a three-conductor
line power outlet that has a protective (earth) ground. (Ground-
ing one conductor of a two-conductor outlet is not sufficient.)
Figure 3-2. Operator’s Checks (1 of 10)
Page 35
Figure 3-2.
Section II
3-7
Page 36
Figure 3-2.
3-8
Page 37
Figure 3-2.
3-9
Page 38
Operation
Model 436A
OPERATOR’S CHECKS
LOCAL OPERATION
(cont’d)
3-10
25. Set the that the
26. Set the POWER REF switch to ON (in) and adjust the Readout indictes –2.00 dBm.
POWER REF switch to off (out). Verify that the
Digital Readout blanks
Figure 3-2. Operator’s Checks (5 of 10)
UNDER RANGE lamp lights and
CAL ADJ control so that the
Digital
Page 39
Figure 3-2.
3-11
Page 40
Operation
REMOTE BCD OPERATION
Model 436A
OPERATOR’S CHECKS
(cont’d)
3-12
8.
Set the POWER REF switch to ON. Verify that the
Digital Readout blanks (1_ . _ _
pW).
OVER RANGE lamp lights and the
NOTE
Underscore (_) indicates blanked digit.
9.
Program the Power Meter to Range 3. Verify that the mW lamp lights and that the OVER RANGE lamp goes out.
Figure 3-2. Operator’s Checks (7 of 10)
Page 41
Figure 3-2.
3-13
Page 42
Figure 3-2.
3-14
Page 43
Figure 3-2.
Section VIII
3-15
Page 44
Figure 3-3.
3-16
Page 45
Figure 3-3.
Section II
3-17
Page 46
Figure 3-3.
3-18
Page 47
Figure 3-3.
Section II
3-19
Page 48
Operation
Model 436A
3-12. HEWLETT-PACKARD INTERFACE BUS
REMOTE OPERATION
NOTE
For a quick and easy programming guide see Figure 3-8; for detailed information study paragraphs 3-12 through 3-61.
3-13. Hewlett-Packard Interface
Bus (HP-IB) Option 022 adds remote programming and digital output capability to the Power Meter. For further information about the HP-IB, refer to IEEE Stand­ard 488 and the Hewlett-Packard Catalog. Power Meter compatibility, programming, and data for­mat is described in detail in the paragraphs which
follow.
3-14. Compatibility
3-15. The Power Meter controls that can be pro­grammed via the Hewlett-Packard Interface Bus are the MODE and SENSOR ZERO switches. The controls not programmable are the POWER REF and LINE switches. The CAL FACTOR % switch can be enabled and disabled via the interface bus
but, when enabled, the calibration factor entered at the front-panel of the Power Meter is used.
takes to make a valid measurement, study the rest of the information in this section.
3-21. Receiving Data Messages
3-22. The Power Meter is configured to listen (re­ceive data) when the controller places the interface bus in the command mode (ATN and REN lines low; IFC line high) and outputs listen address “-” (minus sign). The Power Meter then remains con­figured to listen (accept programming inputs when the interface bus is in the data mode) until it is un­addressed by the controller. To unaddress the Power Meter, the controller can either send the Abort Message (set tine IFC line low) or send the
Local Message (set the REN line high), or it can place the interface bus in the command mode and generate a universal unlisten command.
3-23. Data Input Format. The Power Meter does
not require any particular data input format. It is capable of responding to each of the programming codes listed in Table 3-3 on an individual basis.
Because it responds to these codes in the order it receives them, we recommend that the code for
measurement rate be sent last.
3-16. In addition, specific ranges can be set and various triggering options are available to the pro­grammer. This will be described in detail later.
3-17. The programming capability of the Power Meter will be described in terms of the twelve bus messages found in Table 3-1.
3-18. Data Messages
3-19. The Power Meter communicates on the bus primarily through data messages. It receives data messages that tell it what range to use, what mode to use, whether or not cal factor should be en­abled, and what the measurement rate should be. It sends data messages that tell the measurement value, the mode and range the value was taken at, and what the instrument’s status (see Table 3-4) was when it took the measurement.
3-20. Table 3-2 outlines the key elements involved in making a measurement. Indeed the Power Meter can be programmed to make measurements via the HP-IB by following only the sequence suggested in the table, and briefly referring to Tables 3-3, 3-4,
(input and output data), and Fig. 3-8. However, to
take advantage of the programming flexibility built
into the Power Meter and minimize the time it
3-24. Program Codes. Table 3-3 lists the program
codes that the Power Meter responds to and the functions that they enable. In the listen mode, the Power Meter can handshake in 0.5
IJS.
The time re­quired for the Power Meter to respond to the pro­gramming command, however, depends on where the Power Meter is in the operating program (see
Figure 3-6). The overall worst case time for Power
Meter response to a programming command is
2.5 seconds, the minimum response time is approx­imately 100 microseconds.
NOTE
In addition to the program codes listed in Table 3-3, Power Meter operation will
be affected by all other program codes
shown in columns 2, 3, 4, and 5 of Table 2-2, except (SP!”#$%&*). Thus care
should be taken to address the Power Meter to unlisten before sending these programming commands to other instru­ments on the interface bus.
3-25. Programming the Range. Remote range pro-
gramming is slightly different than Local range
selection. For Local operation the Power Meter auto-ranges. For Remote operation, the program codes have provision for direct selection of the de-
3-20
Page 49
Model 436A
Operation
Table 3-1. Message Reference Table
Message and Identification
Data
Trigger (DTO) No
Clear (DC2)
Remote (RL2)
Local (RL2)
Local Lockout (RL2)
Applicable
Yes
Yes
No
Yes
Yes
No
No
Command and Title
T3 Talker, L2 Listener, AH1 Acceptor Handshake SH1 Source Handshake.
Device Trigger
DCL Device Clear
SDC Selected Device Clear
REN Remote Enable
REN Remote Disable
GTL Go to Local Power Meter does not respond to GTL
REN Remote Disable Power Meter does not respond to LLO
Power Meter changes mode, range, measure­ment rate, and Cal Factor enable or disable. It outputs status and measurement data.
The Power Meter does not respond to a Group Execute Trigger. However, remote trigger capability is part of the Data mes­sage (measurement rate).
Upon receipt of DCL command, Power Meter functions are set for Watt Mode, Auto Range, Cal Factor Disable and Meas­urement rate Hold.
Power Meter goes to remote when addressed to listen, and REN is true (low).
Power Meter goes to local when REN is false (high).
command.
command.
Response
Clear Lockout/ Set Local (RL2)
Pass Control/Take Control
Require Service
Status Byte
Status Bit
Abort
(C(3)
(SR~
(PP@)
Yes
No
No
No
No
Yes
Complete HP-IB capability as defined in IEEE Std. 488 is AH1, CO, DC2, DTO,
LEO, PPO, RL2, SH1, SR0, T3, TEO.
REN Remote Disable
Controller Power Meter cannot act as bus controller.
SRQ Service Request
SPE Serial Poll Enable
SPD Serial Poll Disable
PP Parallel Poll
IFC Interface Clear
NOTE
Returns all devices on bus to local operation.
Power Meter does not request service.
Power Meter does not respond to a Serial Poll
Power Meter does not respond to a parallel poll.
Power Meter stops talking or listening.
3-21
Page 50
Operation
Model 436A
Table 3-2. Measurement Sequence
MEASUREMENT SEQUENCE
Event 1
Event 2
Event 3
Event 4
Event 5
{controller talk and Power Meter listen}, {Program Codes}
See controller manual. Program codes to configure one or more of the Power Meter Listen address factory set to “-” (see Tables 2-1 and 2-2). e.g., CMD "?U-","9D+V"
wrt “pmrd”, “9D+V”
following (see Table 3-3):
1. Range
2. Remote mode (Watt, dBm, dB [Ref]
3. Cal Factor
4. Measurement Rate (and trigger)
Response time for meter’s digital (operating program) circuitry (see Table 3-5 and Figures 3-5 and 3-6).
Meter takes measurement; data available.
Additional delay to allow analog circuits to settle; necessary only if on Range 1 (most sensitive) or if settling time measurement rates are not being used (see Figure 3-4). Here are some suggestions: *
1.
Load reading into controller (event five) and check data string for range (look at character number 1 or check measured value).
2.
If Power Meter is on Range 1, wait 10 seconds and take another reading.
3.
If settling time measurement rates are being used and meter is not on Range 1, use the first reading. If settling time measurement rates are not being used, determine the range and branch to an appro-
4. priate delay: Range 2, one second; Ranges 3-5, 0.1 second.
{universal unlisten, controller listen and Power Meter talk} , {variable name}
See controller manual. Power Meter Talk address factory set to “M” (see Tables 2-1 and 2-2).
*There are other ways to ensure that readings are not affected by analog circuit settling time. Also,
these recommended delays are worst case. A thorough understanding of the material in this section will allow you to optimize measurement time for your particular application. For example, if the power level is not changing, the controller can average at least two consecutive readings to see if the result is still settling.
EXAMPLE PROGRAM SEQUENCE:
Line 1 {controller talk and power meter listen}, “9D+T”
Measurement Rate: Trigger with settling time.
Cal Factor Disable (100%)
dBm
Mode
Auto Range
Line 2 {universal unlisten, controller listen and power meter talk} , {variable name}
Power meter outputs measured value to controller.
Line 3
Line 4 Line 5
Controller checks value in variable for Range 2 threshold (e.g., <-20 dBm for Model 8482A
Power Sensor). If value is below threshold, program branches to line 4. If value is above thres-
hold, program branches to line 5.
{wait 10 seconds, then go to line 1} . {continue}.
3-22
Page 51
Model 436A
Operation
Table 3-3. Hewlett-Packard Interface Bus
Input Program Codes
Function
Range
Least sensitive
Most sensitive Auto
MODE
Watt dB (Rel) dB [Ref] dBm Sensor auto-zero
CAL FACTOR
Disable (100%) Enable (front-panel
switch setting)
Measurement Rate
Hold
Trigger with set-
ling time
Trigger, immediate Free Run at maxi-
mum rate
Free Run with set-
ling time
Program Codes
ASC II DECIMAL
5 4 3 2
1
9
A B C D Z
+
H T
I R
V
53 52 51 50
49 57
65 66 67 68 69
43
45
72 84
73 82
86
program steps that should be written. Specific examples are provided later in this Section. (Refer to Tables 3-3 and 3-4 for Power Meter input and output strings. Refer to controller manual for programming syntax.)
3-28. Programming Cal Factor. While the setting
of the front panel CAL FACTOR switch cannot be remotely changed, the programmer does have a choice. If CAL FACTOR enable is programmed, then the Power Meter uses the Cal Factor set by the switch. If CAL FACTOR Disable is program­med, then the Power Meter uses a Cal Factor of
100%, but the program can correct for cal factor
by computing the corrected reading from the
actual reading and the cal factor (a Cal Factor table
must be stored in an array).
3-29. Programming Measurement Rate. A feature
that is only available via remote programming is
selection of standby, triggered, or free running op­eration of the Power Meter. (During Local opera­tion, the Power Meter is allowed to free run with approximately 133 milliseconds allowed for set­tling time between measurements.) The specific re-
mote triggering capabilities are:
a. HoId (H) - when the power meter is pro-
grammed to Hold, it is inhibited from taking measurements and from outputting data. Thus, it
is set to a predetermined reference condition from
which a measurement can be triggered synchron-
ously to some external event.
sired range as well as for selection of the autorange function.
b. Trigger Immediate (I) - this programming
command directs the Power Meter to make one measurement and output the data in the minimum possible time, then to go into Hold until the next
3-26. Programming the Mode. Remote mode pro-
gramming is similar to Local mode selection. The sequence shown in Example 1 is recommended for taking dB (Rel) readings from a dB [Ref] reference.
triggering command is received. It does not allow
settling time prior to the measurement.
c. Trigger with Delay (T) - this trigger com-
mand is identical to the trigger immediate com-
3-27. Programming Auto-Zero. The Power Meter is
remotely zeroed the same way it is zeroed in local. Example 2 ahown on the next page outlines the
EXAMPLE 1 (dB Rel/dB Ref)
1 {controller talk and Power Meter listen}, “CT”
{controller talk and Power Meter listen}, “BT” Takes first reading relative to set reference
2
{universal unlisten, controller listen and Power Meter talk}, {Variable name} Power Meter outputs reading to controller
3
{controller talk and Power Meter listen} , “T”
4
5 {universal unlisten, controller listen and Power Meter talk}, {Variable name} Power Meter outputs reading to controller
mand except that it causes the Power Meter to execute a settling-time delay subroutine before taking a measurement and outputting data.
Sets reference at present RF input level.
Takes subsequent readings
3-23
Page 52
Operation
Model 436A
Receiving Data Messages (cont’d)
d. Free run at maximum rate (R) - this pro-
gramming command is normally used for asynchro­nous operation of the Power Meter. It directs the Power Meter to continuously take measurements and output data in the minimum possible time. It does not allow settling time prior to each measure-
ment.
e. Free run with delay (V) - this program-
ming command is identical to the previous com­mand except that it causes the Power Meter to exe­cute a settling-time delay subroutine prior to each measurement.
3-30. When programming the Power Meter for synchronous triggered operation, there are two factors that the programmer must consider to ensure the validity of the output measurement data. The first factor is the time that it takes the
Power Meter to respond to a full scale change in
input power level. A typical Power Meter response curve is shown in Figure 3-4. By comparing this curve with the measurement timing cycle shown in
Figure 3-5 and summarized in Table 3-5, the
validity of the Power Meter output can be tabu-
lated according to operating range and triggering interval versus change in input power level. A gen­eral summary of this information is as follows:
a. When the Power Meter is programmed for trigger with settling time operation, sufficient time is provided for the Power Meter to settle to the in­put power level on all ranges except Range 1
(most sensitive range). On Range 1 approximately 10 seconds (9-10 measurements) are required for
the Power Meter to settle to the input power level.
b. When the
trigger immediate
Power Meter is programmed for
operation, the desired amount of
settling time can be incorporated into the program.
3-31. Programming the Local to Remote Mode Change.
The second factor that must be considered
when programming the Power Meter for synchron-
ous triggered operation is whether the first trigger is sent immediately after terminating local opera-
tion. As illustrated in Figure 3-6, the Power Meter
will not respond to the first trigger following a local to remote transition until it completes the previously initiated measurement and display cycle. Thus, the first data output of the Power Meter may not be valid. The options available to the programmer are:
1.
Send a trigger command (Data Message) and discount the first data output. Upon outputting the data, the Power Meter will go to Hold and operate synchronously starting with the next trigger command.
Wait approximately 2.5 seconds after plac-
2. ing the Power Meter in remote and sending the first program trigger command (Data Message).
3.
Send a Clear Message (DCL) immediately
after placing the Power Meter in remote.
This will
restart the Power Meter operating
program.
3-32.
3-33.
Sending Meter
The TALK ONLY/NORMAL switch
Data Messages from the Power
(see Figure 3-3) enables the Power Meter to func-
EXAMPLE 2 (Auto Zero) 1
Remove RF power from power sensor (or set it at least 20 dB below the lowest range of the sensor).
2 {controller talk and Power Meter listen}, “Z1T”
Send zero trigger program codes.
3 {universal unlisten, controller listen and Power Meter talk} , {variable name} Read measured value data from meter
(characters 4, 5, 6, and 7).
If absolute value of measured data is not <2 (0000 ± 0002) then branch to step 2; if it is, then continue. (Although
4
this step averages three seconds, it may take as long as 10 seconds to execute.)
5 {controller talk and Power Meter listen}, “9+Dl”
Send normal measurement mode program codes.
6 {universal unlisten, controller listen and Power Meter talk} , {variabie name} Read status character (number 0) from
meter’s output data string.
7
Check status character for an auto zero loop enabled condition (character 0> decimal 84). If loop is enabled then
branch to step 5. If not, then continue. (This step takes approximately four seconds to execute.)
3-24
Page 53
Model 436A
Operation
Sending Data Messages (cont’d)
tion as a basic talker or in the talk only mode. If the basic talker function is selected, the Power Meter is configured to talk when the controller places the interface bus in the command mode and outputs talk address M. The Power Meter then re­mains configured to talk (output data when the in­terface bus is in the data mode), until it is unad­dressed to talk by the controller. To unaddress the Power Meter, the controller can either send an
Abort Message (generate an interface clear), or it can place the interface bus in the command mode and output a new talk address or a universal untalk command. Examples of addressing and unaddress­ing the Power Meter to talk are provided in Table
3-2 and Figure 3-8.
3-34. Talk Only Mode. When the Power Meter
functions in the Talk Only Mode, it is automati­cally configured to TALK when the interface bus is in the Data Mode and there is at least one listen­er. Since there can only be one talker at a time per interface bus, this function is normally selected only when there is no controller connected to the
system (e.g., when the Power Meter is intercon­nected to an HP 5150A recorder).
Table 3-4. Hewlett-Packard Interface Bus
Output Data String
Definition
Measured value valid Watts Mode under Range Over Range
Under Range dBm or dB
[REL] Mode
S
Power Sensor Auto Zero
Loop Enabled; Range 1
T
Under Range (normal for auto
A T
zeroing on Range 1)
U S
Power Sensor Auto Zero Loop
Enabled; Not Range 1, Under Range (normal for auto zeroing on Range 2-5)
Power Sensor Auto Zero Loop
Enabled; Over Range (error condition - RF power applied to Power Sensor; should not be)
Most Sensitive
R
A
N
G
E
Least Sensitive
1 2 3 4 L
5
Character
ASC II
P Q
R 82 S
T 84
U
V
I J K
M
Decimal
80 81
83
85
86
73 74 75 76
77
3-35. Output Data Format.
The output data
format of the Power Meter is shown and described
in Table 3-4.
3-36. The output data is a fourteen character
string that is provided once at the end of each measurement cycle. It is a good idea to read at least part of this string into the controller after
each measurement cycle, even if it will not be used.
This will avoid the possibility of incorrect data
being read after some future measurement.
3-37. The string begins with a status character
and ends with a carriage return and a line feed.
Measured value is formatted as a real constant:
plus or minus four digits (leading zeros not sup-
pressed) followed by an exponential multiplier.
The decimal point is not provided because it is
understood that it follows the four “measured
value” digits. The two-digit exponent is always
negative.
3-38. Data Output Time. Figure 3-6 provides a
simplified flow chart of Power Meter operation. As shown in the figure, the Power Meter operates according to a stored program and can only output
M
Watt
O
dB REL
D
dB REF (switch pressed)
E
dBm
S
I
space (+)
G
- (minus)
N
D
I
G
I
T
A B C D
SP
-
0 1 2 3 4 5 6 7 8 9
0
1
2
3
4
5
6
7 8
9
65 66 67 68
32 45
48 49 50 51 42 53 54 55 56 57
3-25
Page 54
Operation
Model 436A
Table 3-5. Power Meter Remote Access Time to First Output Data Character
Measurement
Triggering
Free Run at maxi­mum rate, Trigger immediately
Mode
WATT dBm dB (REL) db [REF]
Range 1 or 2
70 ms
90 ms 160 ms 160 ms
Worst Case Access
Range 3,4 or 5
70 ms
90 ms 160 ms 160 ms
Time to First Output Character
Auto Range
Compute measurement times from Figure 3-5 and add measurement time of each range that Power Meter steps through to delay time listed below.
From To Delay
2 1070 ms
1
1 1070 ms
2 23
Examples: Starting at block labeled “HOLD” in Figure 3-5, worst case access time for range 1-3, and range 3-1 changes with WATT MODE se-
lected are:
Range 1 1-2 Delay 1070 ms Range 2 53 ms 2-3 Delay 133 ms Range 3
133 ms
70 ms
53 ms
1379 ms 2256 ms
From To
3 4 54
Range 3 3-2 Delay 1070 ms Range 2 33 ms 2-1 Delay 1070 ms Range 1
Delay
2 1070 ms
3,5 133 ms
133 ms
50 ms (33+17)
33 ms
Free Run with set­tling time or Trigger with settling time.
WATT
dBm dB (REL) db [REF]
1130 ms 1130 ms 1200 ms
160 ms
190 ms
190 ms
260 ms
160 ms
Compute worst case Auto Range access times
from Figure 3-5.
Examples: Starting at block labeled “HOLD” in Figure 3-5; worst case access times for range 1-3
and range 3-1 with WATT MODE selected are:
1-3 (1070 + 53, + 1070 + 53 + 133 + 53) = 2432 ms 3-1 (133 + 33 + 1070 + 33 + 1070 + 33) = 2372 ms.
3-26
Figure 3-4. Power Meter Response Curve (Settling Time for Analog Circuits)
Page 55
Model 436A
Operation
Figure 3-5. Measurement Timing Flow Chart (Settling Time for Digital Circuitry)
3-27
Page 56
Operation
Model 436A
Note: The decidon blocks labeled HOLD operation, programmed,
,ital
when the Power Meter is programmed for external after receiving a trigger in the listen mode.
When
readout is
hold ie programmed,
the decEsion is
updated.
The
no,
dectsion
starting
then
the
when a
Figure 3-6. Operating Program Simplified Flow Chart
3-28
are associated
decision is
trigger is
reverts
to
yes until
triggering,
with both hold
always yes. When
received
and
receipt of
it
wiEl
provide outPut
and externaE triggered
triggered operation
continuing
the
next
unitl
the
trigger. Thus,
data OIIIY
is
dig-
Page 57
Model 436A
Operation
Sending Data Messages (cont’d)
data after taking a measurement. Thus, when the interface bus is placed in the data mode after the
Power Meter has been addressed to talk, the time required to access the first output data character depends on where the Power Meter is in the
operating program, and on how the Power Meter has been previously programmed (see Programming
Codes above.) Worst case access times for each of
the Power Meter operating configurations are listed
in Table 3-5.
3-39. After the first output character is sent, the
remaining characters are sent at either a 10-kHz
rate (infinitely fast listener) or at the receive rate
of the slowest listener.
3-40. Receiving the Trigger Message
3-41. The Power Meter has no provision for re-
sponding to a Trigger Message (bus command GET).
Power Meter triggering is done with the Data Mes-
sage (through the Measurement Rate Program Codes).
3-42. Receiving the Clear Message
3-48. Receiving the Local Lockout and Clear
Lockout Set Local Messages
3-49. The Power Meter does not respond to the Local Lockout Message (LLO bus command). It
responds to the Clear Lockout/Set Local Message
in that when the REN bus line goes false, it will
revert to local operation.
3-50. Receiving the Pass Control Message
3-51. The Power Meter has no provision for oper­ation as a controller.
3-52. Sending the Required Service Message
3-53. The Power Meter does not have provision for requesting service.
3-54. Sending the Status Byte Message
3-55. The Power Meter does not respond
to a
Serial Poll.
3-56. Sending the Status Bit Message
3-57. The Power Meter does not respond
to a
Parallel Poll.
3-43. The Power Meter has provision for respond­ing to the DCL bus command but not the SDC bus command. Upon receipt of the DCL command, the Power Meter operating program is reset causing the Power Meter to enter the Hold state shown at the top of Figure 3-6, and the HP-IB circuits are con­figured to provide Watt Mode, Auto Range, and Cal Factor Disable outputs.
3-44. Receiving the Remote Message
3-45. When the Power Meter recieves the Remote Message (REN line low) it completes the rest of its current measurement cycle (see Figure 3-6) and then goes to remote. See the Local to Remote Mode Change (paragraph 3-31) for information about how to program the local to remote mode change.
3-46. Receiving the Local Message
3-47. The Power Meter does not respond to the
GTL (go to local) bus command. It reverts to local
operation when the REN (remote enable) bus
line goes false (high).
3-58. Receiving the Abort Message
3-59. When the Power Meter receives an Inter­face Clear command (IFC), it stops talking or listening.
3-60. Test of HP-IB Operation
3-61. Figure 3-7 outlines a quick check of the 436A remote functions. This gives the user two alternatives for testing the power meter: 1, write a program corresponding to Figure 3-7 for a quick check or 2, use the program in Section VIII for complete testing and troubleshooting.
3-62. REMOTE BCD INTERFACE OPERATION
3-63. BCD Option 024 adds remote programming and digital output capability to the Power Meter. There are two basic methods for operating the Power Meter with this option. It can be operated locally with an external instrument used to record
output data, or it can be operated remotely by
sending remote programming inputs to the Power
Meter.
3-29
Page 58
Operation
Model 436A
3-30
Figure 3-7. Test of HP-IB Operation Flowchart
Page 59
Model 436A
Operation
436A QUICK PROGRAMMING GUIDE
This guide will help set up and program simple HP-IB instrumentation systems, thereby freeing you from making an in-depth study of system design and BASIC or HPL programming languages.
THE SYSTEM:
I.
* HP-IB cables shown with dotted lines are used only if the Source and Device under test are programmable.
** Signal Source and Device under Test may be the same, e.g., checking Sig. Gen. Flatness.
II.
THE PROGRAM: If the power meter is the only part of the system to be programmed, use the program statements in the order given. For more complex systems or programs, include statements derived from the information in the optional (dashed line) flow chart boxes. When it is necessary to write more statements, refer to Table 3-2.
Figure 3-8. 436A Quick Programming Guide (1 of 5)
3-31
Page 60
Operation
Model 436A
436A QUICK PROGRAMMING GUIDE (Cont’d)
page 8-34
page 8-34
page 8-33
page 8-33
3-32
Figure 3-8. 436A Quick Programming Guide (2 of 5)
Page 61
Model 436A
Operation
Figure 3-8. 436A Quick Programming Guide (3 of 5)
3-33
Page 62
Operation
Model 436A
436A QUICK PROGRAMMING GUIDE (cont’d)
(HPL)
“pmz”
Subroutines for 9825
- Power meter zero subroutine
“pmz”: “remove source":dsp “disconnect sensor from source";stp
wrt "pmrd", "Z1T"; fmt 2,3x,f5.0;red “pmrd.2”,Z
“verify zero”
:if abs(Z)>2;gto "remove source"
"unzero":wrt "pmrd", "9+AI";fmt 3,b;red "pmrd.3",Z “verify unzero”: if Z>34;gto “unzero” “preset/ret”:wrt "pmrd","9D+V";ret
“pmr”
- Power meter read subroutine
“pmr”:
fmt 1,1x,b,1x,f5.0,1x,f3.0
O
+R
for X=1 to 20 wrt “pmrd”, “9D+V” wait (R=73) 4000
red “pmrd.1’’,R,P,E
if X=1;g to “P1”
if abs(P-S)>l;gto “P1”
P10”E+P;ret
"P1":P+S
next X dsp “power meter not settled”
Note:
The next statement should be
should be used.
"end"
:end,
or if another subroutine follows then a
gto “end”
3-34
Figure 3-8. 436A Quick Programming Guide (4 of 5)
Page 63
Model 436A
Operation
436A QUICK PROGRAMMING GUIDE (cont’d)
Subroutines for 9830 (BASIC)
Figure 3-8. 436A Quick Programming Guide (5 of 5)
3-35
Page 64
Operation
Model 436A
REMOTE BCD INTERFACE OPERATION (cont’d)
3-64. Figure 3-3 provides instructions for operat­ing the Power Meter with the BCD option installed. In order to follow these instructions, the operator must be familiar with Power Meter programming and output data format. This information is pro­vided in detail in the paragraphs which follow.
NOTE
The Power Meter BCD option is designed to interface directly with an HP 5055A
Digital Recorder. When it is used with
this recorder, it can only be operated in the Local mode (unless a special cable is
fabricated), as the BCD interface bus
lines that are normally used to program the Power Meter, are used instead to pre-
set the digital recorder print format. In
the paragraphs which follow, differences in Power Meter output data format for
digital recorder and “universal” interfac-
ing are noted as applicable.
3-65. Output Data Format
3-66. When the Power Meter is interfaced with an
HP 5055A Digital Recorder, the output data print-
out is as described in Table 3-4. When the Power
Meter is interfaced with other controller or re-
corder instruments, data format is selected by the
user. Refer to Table 3-5 for a description of the
function and coding of the Power Meter output
data lines.
3-67. BCD Remote Programming
3-68. Remote programming of the Power Meter is enabled when a 0.0 to +0.4 Vdc level is applied to remote enable input line J7-21. The Power Meter controls that can be programmed remotely are the MODE and SENSOR ZERO switches. The controls not programmable are the POWER REF and LINE switches. The CAL FACTOR % switch can be enabled and disabled via the remote inter­face but, when enabled, the calibration factor entered at the front panel of the Power Meter is used.
NOTE
Jumper options are provided to enable remote programming of the SENSOR ZERO switch when the remote enable in-
put is high (+2.5 to +5.0V level is applied
to J7-21). See Section II, Installation.
3-69. Remote range programming is slightly
different than Local Range selection. For Local
operation, a particular range is selected by allowing
the Power Meter to autorange to the desired range,
then pressing the RANGE HOLD switch to hold the range. For Remote operation, the programming
codes have providion for direct selection of the
desired range as well as selection of the autorange
function.
3-70. An additional feature that is only available
via remote programming is selection of standby,
triggered, or free running operation of the Power
3-36
Table 3-6. Power Meter Output Data Printout for HP 5055A Digital Recorder
Column Interpretation
1 (right) 2 3 4 5
6 7 8 9 10 (left) Exponent Tens Digit
Intrepret measured value as XXXX . 10
Units Digit Tens Digit Hundreds Digit Thousands Digit Sign
Range* Mode** Status*** Exponent Units Digit
*Range Code
**Mode Decode
***Status
1 = Range 1 (most sensitive) 2 = Range 2 3= Range 3 4 = Range 4 5 = Range 5 (least sensitive)
V = dB [REF] A = dB (REL)
$2=
Watts
* = dBm
0 = In Range 1 = Underrange (WATT Mode) 2 = Overrange 3 = Underrange (dBm Mode) 4 = ZERO Mode
Page 65
Model 436A
Operation
BCD Remote Programming (cont’d)
Meter. (During Local operation, the Power Meter is allowed to free run with approximately 133 milli­seconds allowed for settling time between measure­ments.) The specific remote triggering capabilities are:
Hold -
a.
when the power meter is pro-
trigger with settling time operation, sufficient time is provided for the Power Meter to settle to the input power level on all ranges except range 1. On range 1 approximately 10 seconds (0-10 measure­ments) are required for the Power Meter to settle
grammed to Hold, it is inhibited from taking measurements and from outputting data. Thus, it is set to a predetermined reference condition from
which a measurement can be triggered synchrously
trigger immediate operation, the desired amount of settling time can be incorporated into the operat­ing program.
to some external event.
b. Trigger Immediate - this programming
command directs the Power Meter to make one
3-72. The programming codes that the Power Meter will respond to are listed in Table 3-8.
measurement and output the data in the minimum possible time, then to go into Hold until the next
Triggering command is recieved. It does not allow settling time prior to the measurement.
3-73. POWER MEASUREMENT ACCURACY
3-74. A power measurement is never free from
error or uncertainty. Any RF system has RF losses,
c.
Trigger command is identical to the trigger immediate command except that it causes the Power Meter to execute a settling-time delay subroutine before taking a measurement and outputting data.
with Delay - this trigger
mismatch instrumentation uncertainty. Measurement errors as high as 50% are not only possible, they are highly likely unless the error sources are understood and, as much as possible, eliminated.
d. Free run at maximum rate - this programming asynchronous operation of the Power Meter. It
directs the Power Meter to continuously take measurements and output data in the minimum possible time. It does not allow settling time prior
to each measurement.
command is normally used for
3-75. Sources of Error and Measurement Uncer­tainty
3-76. RF Losses. Some of the RF power that enters the Power Sensor is not dissipated in the power sensing elements. This RF loss is caused by dissipation in the walls of waveguide power
e.
Free run with Delay - this programming command is identical to the previous command except that it causes the Power Meter to execute a
sensors, in the center conductor of coaxial power sensors, in the dielectric of capacitors, connections
within the sensor, and radiation losses. settling-time delay subroutine prior to each measurement.
3-77. Mismatch. The result of mismatched impe-
dances between the device under test and the 3-71. When programming the Power Meter for synchronous triggered operation, there are two
factors that the programmer must consider to ensure the validity of the output measurement data. The first factor is the time that it takes the Power
Meter to respond to a full scale change in input
power level. A typical Power Meter response curve
is shown in Figure 3-4. By comparing this curve
with the measurement timing cycle shown in
Figure 3-6 and summarized in Table 3-5, the validity of the Power Meter output can be
tabulated according to operating range and
triggering interval - versus change in input power
level. A general summay of this information is as
follows:
power sensor is that some of the power fed to the sensor is reflected before it is dissipated in the load. Mismatches affect the measurement in two ways. First, the initial reflection is a simple loss and is called mismatch loss. Second, the power reflected from the sensor mismatch travels back up the transmission line until it reaches the source. There, most of it is dissipated in the source impedance, but some of its re-reflected by the source mismatch. The re-reflected power returns to the power sensor and adds to, or subtracts from, the incident power. For all practical purposes, the effect the re-reflected power has upon the power measurement is unpredictable. This effect is called mismatch uncertainty.
When the Power Meter is programmed for
a.
b. When the Power Meter is programmed for
losses,
mismatch
uncertainty and
uncertainy,
calibration
3-37
Page 66
Operation
Model 436A
Sources of Error and Measurement Uncertainty
(cont’d)
3-78. Instrumentation Uncertainty. Instruments:
tion uncertainty describes the ability of the meter­ing circuits to accurately measure the dc output
from the Power Sensor’s power sensing device. In the Power Meter this error is ± 0.5% for Ranges 1 through 5. It is important to realize, however, that these uncertainty specifications do not indicate
overall measurement accuracy.
3-79. Power Reference Uncertainty. The output
level of the Power Reference Oscillator is factory
set to 1 mW ± 0.70% at 50 MHz. This reference is
normally used to calibrate the system, and is,
therefore, a part of the system’s total measurement
uncertainty.
3-80. Cal Factor Switch Resolution Error. The
resolution of the CAL FACTOR % switch
contributes a significant error to the total
measurement because the switch has 2% steps. The
maximum error possible in each position is ± 0.5%
ment and record the reading. Use the reflection coefficient, magnitude and phase angle from the table supplied with the Power Sensor to calculate the corrected power level.
b. Set the CAL FACTOR % switch to the nearest position above and below the correction factor given on the table. Interpolating between the power levels measured provides the corrected power level.
3-86. Calculating Total Uncertainty
3-87. Certain errors in calculating the total measurement uncertainty have been ignored in this discussion because they are beyond the scope of this manual. Application Note AN-64, “Microwave
Power Measurement”, delves deeper into the calcu-
lation of power measurement uncertainties. It is available, on request, from your nearest HP office.
3-88. Known Uncertainties. The known uncer-
tainties which account for part of the total power measurement uncertainty are:
3-81. Corrections for Error
3-82. The two correction factors basic to power meters are
calibration factor and effective efficiency. Effective efficiency is the correction factor for RF losses within the Power Sensor. Calibration factor takes into account the effective efficiency and mismatch losses.
3-83. Calibration factor is
expressed as a percentage with 100% meaning the power sensor has no losses. Normally the calibration factor will be 100% at 50 MHz, the operating frequency of the internal reference oscillator.
3-84. The Power Sensors used with the Power Meter have individually calibrated calibration factor curves placed on their covers. To correct for RF and mismatch losses, simply find the Power Sensor’s calibration factor at the measurement frequency from the curve or the table that is supplied with the Power Sensor and set the CAL FACTOR % switch to this value. The measurement error due to this error is now minimized.
3-85. The CAL FACTOR % switch resolution error of ± 0.5% may be reduced by one of the following methods:
a.
Instrumentation uncertainty
± 0.02 dB (Range 1 through 5).
Power reference uncertainty
b.
± 0.03 dB.
c. CAL FACTOR switch resolution ±0.5% or
±0.02 dB.
The total uncertainty from these sources is ±1.7% or ±0.07 dB.
3-89. Calculating Mismatch Uncertainty. Mismatch
uncertainty is the result of the source mismatch
interacting with the Power Sensor mismatch. The magnitude of uncertainty is related to the magni­tudes of the source and Power Sensor reflection coefficients, which can be calculated from SWR.
Figure 3-9 shows how the calculations are to be
made and Figure 3-10 illustrates mismatch uncer­tainty and total calculated uncertainty for two
cases. In the first case, the Power Sensor’s SWR =
1.5, and in the second case, the Power Sensor’s
SWR = 1.26. In both cases the source has a SWR of
2.0. The example shows the effect on power
measurement accuracy a poorly matched power
sensor will have as compared to one with low
mismatch.
Leave the CAL FACTOR % switch on
a.
100% after calibration, then make the measure-
3-38
3-90. A faster, easier
tainty is to use the
way to find mismatch uncer­HP Mismatch Error (uncer-
Page 67
Model 436A
Operation
Calculating Total Uncertainty (cont’d)
tainty) Limits/Reflectometer Calculator. The calculator may be obtained, on request, from your nearest Hewlett-Packard office by using HP Part
Number 5952-0448.
3-91. The method of calculating measurement uncertainty from the uncertainty in dB is shown by Figure 3-11. This method would be used when
the initial uncertainty calculations were made with the Mismatch Error/Reflectometer Calculator.
NOTE
The BCD output data levels are TTL compatible. A false
(0) state is defined as 0.0 to +0.4 Vdc and a true state is defined as +2.5 to +5.0 Vdc
Table 3-7. BCD Output Data Codes (1 of 2)
Function
MEASURED VALUE - The Power Meter format for outputting
the measured value is SIGN, Four BCD DIGITS, and a negative
EXPONENT. It is interpreted as:
± XXXX
Code
NOTES
Pin numbers refer to connec-
tor J7 on the rear panel.
When used with 5055A, a four
line format is established by
the following pins:
34 (ground) 10 (measurement rate;
floats high)
35 (cal factor disable;
floats high)
Sign
space (+)
Digits
Units
Tens 101A
Hundreds
Thousands
10°A 10°B 10°C 10°D
101B 101C
1
10
D
2
10
A 102 B 102 C
2
10
D
3
10
A 103 B 103 C 103 D
PIN 9
Weight
1 1 2 4 8 1 2 4 8 1 2 4 8 1 2 4 8
0 1
Pin Number
2, 26 27
3
4 28 29
5
6 30 31
7
8
32
33
3-39
Page 68
Operation
Model 436A
Table 3-7. BCD Output Data Codes (2 of 2)
Function
STATUS OUTPUTS
In Range Underrange (WATT Mode) Overrange Underrange (dBm Mode) Zero Mode
RANGE - indicates range on which last measurement made.
1 (most sensitive) 2 3 4 5 (least sensitive)
EXPONENT
Units EX0A
EXOB EXOC
O
EX
D
Tens EX
1
A
Code
Pin 40 Pin 16 Pin 15
0 0 0 0 1
Pin 36 Pin 12
0 0 0 0 1 1
Weight
2
1
4 8 1
0 0 0 1 1 0
Pin 11
1 1 0 0
Pin
17 18 42 43 19
1 0 1 0
1
0
1
0
1
MODE
PRINT
dB [REF] dB (REL) WATT dBM
Note: when used with
5055A, four line for­mat is established by following pins:
20 (ground) 44 (ground) 45 (ground)
Pin 14
0 0 1 1
Pin 13
0 1 0
1
Note: when used with
5055A, four line format is established by fol­lowing pins:
38 (floats high) 39 (floats high)
High to low transition
on pin 48 when output
data is valid.
3-40
Page 69
Model 436A
Operation
Table 3-8. BCD Programming Commands
Commands
Remote enable
Input Pin
J7-21 When high, enables local operation of Power Meter via front-
panel controls. When low, enables remote operation of Power
Meter via programming commands listed below.
Range Bit 1
J7-24 Select Power Meter measurement range when Remote Enable Range Bit 2 J7-25 input is low. Range Bit 3
J7-23
*Standby range: Power Meter operating program is
held at Power Up address 000
Function
NOTE
When equipped with the BCD Option 024, the
Power Meter generates a Print command and
provides valid output data after each measure-
ment for both Local and Remote operation.
Range
0* 1
2
3 4 5
Auto
Pin 24
0 0
0 0 1 1 1
Pin 25
0 0 1
1 0 0
1
.
8
Pin 23
0 1
0 1 0
1
X (0 or 1)
Rate
Inhibit
Cal Factor
Disable
Mode Bit 1 Mode Bit 2
SENSOR Zero Select
J7-10
J7-47
J7-35
Selects Power Meter triggering when remote enable input is low
Pin 10 Pin 47 Pin 49 Hold Trigger Intermediate 0
Trigger with Delay 1 Free Run Fast
Free Run with Delay 1
X (0 or 1) 0
0
0
Positive-to-negative
0
1 1
When low disables front-panel CAL FACTOR % switch (same as
100% position). When high, enables switch.
J7-38 J7-39
Select mode when remote enable input is low.
Mode
dB [REF] dB (REL) WATT dBm
Pin 38 Pin 39
0 1 0
0 0 1
1 1
J7-46 When low, enables power sensor auto zero circuit.
NOTE
When programming this function, allow the cir­cuit about 7 seconds to settle before applying input power to Power Sensor. If RF input power is applied while ZERO lamp is on, it will introduce
an offset that will affect future measurements.
X (0 or 1) transition X (0 or 1)
X (0 or 1)
3-41
Page 70
Operation
CALCULATING MEASUREMENT UNCERTAINTY
1. Calculate the reflection coefficient from the given SWR.
SWR - 1
=
P
SWR + 1
Model 436A
Power Sensor #1
Power Sensor #2
Power Source
2. Calculate the relative power and percentage power mismatch uncertainties from the reflection coefficients. An initial reference level of 1 is assumed.
Reletive Power Uncertainty
3-42
%PU =
%PU
1
%PU
2
Percentage Power Uncertainty
(PU - 1) 100% for PU >1
=
(1.138 - 1) 100%
and -(1 - PU) 100% for PU <1 and -(1 - 0.870) 100%
= (0.138) 100% and =
13.8%
= (1.073 - 1) 100%
= (0.073) 100% =
7.3%
Figure 3-9. Calculating Measurement Uncertainties (1 of 2)
and -13.0% and -(1 - 0.928) 100%
and -(0.072) 100% and -7.2%
-(0.130) 100%
Page 71
Model 436A
CALCULATING MEASUREMENT UNCERTAINTY
3. Calculate the Measurement Uncertainty in dB.
Operation
10 [log (10) (0.870) - log (10) (1)]
= 10 [0.031]
= +0.31 dB
and
and
and and
and and
and and
10 [log (8.70) - log (10)] 10 [0.94 - 1] 10 [- 0.060]
- 0.60 dB
10 [log (10) (0.928) -log (10) (1)]
10 [log (9.28) - log (10)] 10 [0.968 - 1]
10 [- 0.032]
- 0.32 dB
Figure 3-9. Calculating Measurement Uncertainties (2 of 2)
3-43
Page 72
Operation
Model 436A
3-44
Figure 3-10. The Effect of Power Sensor Mismatch on Measurement Accuracy
Page 73
Model 436A
Operation
CALCULATING MEASUREMENT UNCERTAINTY
1. For this example the known values are:
source SWR, 2.2 and power sensor SWR, 1.16.
From the Mismatch Error Calculator the mismatch uncertainty is found to be +0.24, -0.25 dB.
2. Add the known uncertainties from paragraph 3-73, (± 0.10 dB). Our total measurement uncer­tainty is +0.34, -0.35 dB.
3. Calculate the relative measurement uncertainty from the following formula:
dB =
If dB is positive then:
> P0; let P0 = 1
P
1
P
MU =
1
If dB is negative then:
P
< P0; let P1 = 1
1
4. Calculate the percentage Measurement Uncertainty. For P
> P
1
0
%MU = (P1 - P0) 100
=
(1.081 - 1) 100
=
+8.1%
Figure 3-11. Calculating Measurement Uncertainty (Uncertainty in dB Known)
For P1 < P %MU =
=
=
0
– P0) 100
- (P
1
- (1 - 0.923) 100
-7.7%
3-45/3-46
Page 74
Page 75
Model 436A
Performance Tests
SECTION IV
PERFORMANCE TESTS
4-1.
INTRODUCTION
4-2.
The procedures in
ical performance of the Power Meter using the
specifications of Table 1-1 as performance stand­ards. All tests can be performed without access to the interior of the instrument. A simpler opera­tional test is included in Section III under Opera­tor’s Checks.
4-3. EQUIPMENT REQUIRED
4-4. Equipment required for the performance tests
is listed in Table 1-2, Recommended Test Equip­ment. Any equipment that satisfies the critical specifications given in the table may be substituted for the recommended model(s).
4-5. TEST RECORD
4-6. Results of the performance tests may be tabulated on the Test Record at the end of the test procedures. The Test Record lists all of the tested specifications and their acceptable limits. Test results recorded at incoming inspection can be used for comparison in periodic maintenance, trouble­shooting, and after repairs or adjustments.
this section test the electr-
4-7. PERFORMANCE TESTS
4-8. The performance tests given in this section are suitable for incoming inspection, troubleshooting, or preventive maintenance. During any per­formance test, all shields and connecting hardware must be in place. The tests are designed to verify published instrument specifications. Perform the tests in the order given and record the data on the test card and/or in the data spaces provided at the end of each procedure.
NOTE
The Power Meter must have a half-hour warmup and the line voltage must be
within +5%,
-10% of nominal if the
performance tests are to be considered
valid.
4-9. Each test is arranged so that the specification
is written as it appears in Table 1-1. Next, a description of the test and any special instructions or problem areas are included. Each test that requires test equipment has a setup drawing and a list of the required equipment. The initial steps of each procedure give control settings required for that particular test.
4-1
Page 76
Performance Tests
4-10. ZERO CARRYOVER TEST
Model 436A
PERFORMANCE TESTS
SPECIFICATION:
DESCRIPTION:
EQUIPMENT: PROCEDURE:
±0.2% of full scale when zeroed on the most sensitive range. After the Power Meter is initially zeroed on the most sensitive range, the change in the
digital readout is monitored as the Power Meter is stepped through its ranges. Thus, this test also takes noise and drift into account because noise , drift, and zero carry­over readings cannot be separated.
Figure 4-1. Zero Carryover Test Setup
Range Calibrator . . . . . . . HP 11683A
1.
Set the Power Meter switches as follows:
CAL FACTOR % . . . . . 100
POWER REF . . . . . . off (out)
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . . ON (in)
off (out)
Set the Range Calibrator switches as follows:
2.
FUNCTION . . . . . . . STANDBY
POLARITY . . . . . . . NORMAL
100
RANGE . . . . . . . .
LINE . . . . . . . . .
Connect the equipment as shown in Figure 4-1.
3.
4.
Press and hold the Power Meter SENSOR ZERO switch and wait for the digital
readout to stabilize. Then verify that the Power Meter ZERO lamp is lit and that
the digital readout indicates 0.00 ± 0.02.
Power Meter is now zeroed on most sensitive range (10 p W).
Release the Power Meter SENSOR ZERO switch and wait for the ZERO lamp to
5. go out before proceeding to the next step.
6.
Set the Range Calibrator FUNCTION switch to CALIBRATE and verify that the Power Meter autoranges to the 100
‘7.
Set the Power Meter RANGE HOLD switch to on (in) and the Range Calibrator FUNCTION switch to standby.
#w
ON (in)
NOTE
KW
range.
4-2
Page 77
Model 436A
PERFORMANCE TESTS
4-10. ZERO CARRYOVER TEST (cont’d)
8.
Wait for the Power Meter’s digital readout to stabilize and verify that the indication observed is within the limits shown on the table below. Then set the POWER Meter RANGE HOLD switch to off (out).
9.
Repeat steps 6, 7, and 8 with the Range Calibrator RANGE switch set, in turn, to 1 mW, 10 mW, and 100 mW. Verify that the Power Meter autoranges properly, and that the indication observed on each range is within the limits shown in Table 4-1.
Table 4-1. Zero Carryover Autorange Digital Readout Results
Performance Tests
Range Calibrator
and
Power Meter
Range
10 /lw
100
/Jw
1 mW
10 mW
100 mW
4-11. INSTRUMENT ACCURACY TEST
SPECIFICATION: WATT MODE:
dBm MODE:
dB (REL) MODE:
±0.570 in Ranges 1 through 5. ±0.02 dB ±0.001 dB/°C in Ranges 1 through 5.
±0.02 dB ±0.001 dB/°C in Ranges 1 through 5.
The dB (REL) specifications are for within-range measurements. For range-to-range accuracy, add the uncertainty associated with the range in which the reference was entered, to the uncertainty
associated with the range in which the measurement was made.
For example, if a reference is entered in Range 1 and a measure­ment is made in Range 5, the total uncertainty is ±0.04 (Range 1
±0.02 + Range 5 ± 0.02 = ± 0.04).
Min
-0.02
-0.2
-.002
-0.02
-00.2
Results
Actual Max
0.02
0.2
.002
0.02
00.2
NOTE
DESCRIPTION:
After the Power Meter is initially calibrated on the 1 mW range, the digital readout is
monitored as the Range Calibrator is adjusted to provide reference inputs correspond­ing to each of the Power Meter operating ranges.
4-3
Page 78
Performance Tests
PERFORMANCE TESTS
4-11. INSTRUMENT ACCURACY TEST (cont’d)
Figure 4-2. Instrument Accuracy Test Setup
Model 436A
EQUIPMENT:
PROCEDURE:
Range Calibrator . . . . . . . HP 11683A
1.
Set the Power Meter switches as follows:
CAL FACTOR % . . . . . 100
POWER REF . . . . . . off (out)
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
Set the Range Calibrator switches as follows:
2.
off (out)
ON (in)
FUNCTION . . . . . . . STANDBY
POLARITY . . . . . . . NORMAL
RANGE . . . . . . . . 1mW
LINE . . . . . . . . .
3.
Connect the equipment as shown in Figure 4-2.
ON (in)
4. Press and hold the Power Meter SENSOR ZERO switch and wait for the digital readout to stabilize. Then verify that the Power Meter ZERO lamp is lit and that the digital readout indicates 0.00 ± 0.02.
NOTE
Power Meter is now zeroed on the most sensitive range (10 v W).
4-4
5.
Release the Power Meter SENSOR ZERO switch and wait for the ZERO lamp to go out before proceeding to the next step.
6. Set the Range Calibrator FUNCTION switch to CALIBRATE and verify that the
Power Meter autoranges to the 1 mW range.
7.
Observe the Power Meter digital readout and, if necessary, adjust the front-panel CAL ADJ control to obtain a 1.000 ± 0.002 indication.
NOTE
The Range Calibrator output level is adjustable in 5 dB increments. Thus, the 3
M
W,
30p
W,
300p
W, 3 mW, and 30 mW legends on the
RANGE switch are approximations. The true outputs for these set-
tings are
3.161.t W, 31.61.JW,
316PW, 3.16 mW and 31.6 mW.
Page 79
Model 436A
PERFORMANCE TESTS
4-11. INSTRUMENT ACCURACY TEST (cont’d)
Performance Tests
8.
Set the Range Calibrator RANGE switch, in turn, to 10 100 mW. Verify that the Power Meter autoranges properly and that the indication observed on each range is within the limits specified in the table below.
Set the Power Meter MODE switch to dBm.
9.
10.
Set the Range Calibrator RANGE switch, in turn, to -20 dBm, -10 dBm, 0 dBm,
+10 dBm, and +20 dBm. Verify that the Power Meter autoranges properly and that the indication observed on each range is within the limits specified in
Table 4-2.
Table 4-2. Instrument Accuracy Test Results
Range Calibrator
and
Power Meter
Range
10
/..lw
100
#w
1 mW
10 mW
100 mW 99.0
Results
Min Actual Max
9.95
99.5
0.995
9.95
10.05
100.5
1.005
10.05
101.0
Range Calibrator
and
Power Meter
Range
-20 dBm -20.02
-10 dBm -10.02 0 dBm -0.02
+10 dBm +20 dBm
11. Set the Range Calibrator RANGE switch to -10 dBm.
uW,
100
Min Actual
9.98
19.96
PW,
10 mW, and
Results
-19.98
-9.98
0.02
10.02
20.04
12. Set the Power Meter MODE switch to dB [REF] and verify that the digital readout indicates 0.00 ± 0.01.
13. Set the Range Calibrator RANGE switch, in turn, to -20 dBm, -5 dBm, and
+10 dBm. Verify that the Power Meter autoranges properly, and that the indica-
tion observed on each range is within the limits specified in Table 4-3.
Table 4-3. Instrument Accuracy Test Results for dB [REF] Mode
Range Calibrator
and
Power Meter
Ranges
-20 dBm
-5 dBm +10 dBm
Min
-9.96 +4.96
+19.96
Results
Actual Max
-10.04 +5.04
20.04
4-5
Page 80
Performance Tests
4-12. CALIBRATION FACTOR TEST
Model 436A
PERFORMANCE TESTS
SPECIFICATION:
DESCRIPTION:
PROCEDURE:
16-position switch normalizes meter reading to account for calibration factor. Range 85% to 100% in 1% steps. 100% position corresponds to calibration factor at 50 MHz.
After the Power Meter is zeroed on the most sensitive range, a 1 mW, input level is ap­plied to the Power Meter and the CAL ADJ control is adjusted to obtain a 1.000 mW
indication. Then the CAL FACTOR % switch is stepped through its 16 positions and
the digital readout is monitored to ensure that the proper indication is obtained for each position.
Figure 4-3. Calibration Factor Test Setup
1.
Set the Power Meter switches as follows:
CAL FACTOR % . . . . . 100
POWER REF . . . . . .
off (out)
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
off (out)
ON (in) Set the Range Calibrator switches as follows:
2.
FUNCTION . . . . . . . STANDBY
POLARITY . . . . . . . NORMAL
RANGE . . . . . . . . 1mW
LINE . . . . . . . . .
Connect the equipment as shown in Figure 4-3.
3.
Press and hold the Power Meter SENSOR ZERO switch and wait for the digital
4.
ON (in)
readout to stabilize. Then verify that the Power Meter ZERO lamp is lit and that the digital readout indicates 0.00 ± 0.02.
NOTE
Power Meter is now zeroed on most sensitive range (10 p W)
Release the Power Meter SENSOR ZERO switch and wait for the ZERO lamp to
5.
go out before proceeding to step 6. Set the Range Calibrator FUNCTION switch to CALIBRATE and verify that the
6.
Power Meter autoranges to the 1 mW range. Adjust the Power Meter CAL ADJ control to obtain a 1.000 ± 0.002 indication
7. on the digital readout.
4-6
Page 81
Model 436A
PERFORMANCE TESTS
4-12. CALIBRATION FACTOR TEST (cont’d)
8.
Set the CAL FACTOR % switch, in turn, to each position and verify that the indi­cations observed are within the limits specified in Table 4-4.
Table 4-4. Calibration Factor Test Results
Performance Tests
CAL FACTOR
Switch
Position
100
99 98 97 96 95 94 93
4-13. POWER REFERENCE LEVEL TEST
Min.
0.994
1.004
1.014
1.025
1.036
1.047
1.058
1.069
Results
Actual Max.
1.006
1.016
1.026
1.037
1.048
1.059
1.070
1.081
CAL FACTOR
Switch
Position
92
91
90
89 88 87 86 85
Min.
1.081
1.093
1.105
1.118
1.130
1.143
1.157
1.170
Results
Actual
SPECIFICATION: Internal 50 MHz oscillator factory set to 1 mW ± 0.7% traceable to the National
Bureau of Standards. Accuracy: ±1.2% worst case (±0.9% rms) for one year (0°C to 55°C).
DESCRIPTION:
The power reference oscillator output is factory adjusted to 1 mW ± 0.7%. To achieve this accuracy, Hewlett-Packard employs a special measurement system accurate to 0.5% (traceable to the National Bureau of Standards) and allows for a transfer error of ±0.2% in making the adjustment. If an equivalent measurement system is employed for veri­fication, the power reference oscillator output can be verified to 1 mW ±1.9% (±1.2% accuracy + ±0.5% verification system error + ±0.2% transfer error = 1.9% maximum error). The power reference oscillator can be set to ±0.7% using the same equipment and following the adjustment procedure in paragraph 5-22. To ensure maximum accur­acy in verifying the power reference oscillator output, the following procedure provides step-by-step instructions for using specified Hewlett-Packard test instruments of known capability. If equivalent test instruments are used, signal acquisition criteria may vary and reference should be made to the manufacturer’s guidelines for operating the instruments.
Max.
1.093
1.105
1.117
1.130
1.142
1.155
1.169
1.182
NOTE
The Power Meter may be returned to the nearest Hewlett-Packard office to have the power reference oscillator checked and/or adjusted. Refer to Section II, PACKAGING.
4-7
Page 82
Performance Tests
PERFORMANCE TESTS
4-13. POWER REFERENCE LEVEL TEST (cont’d)
Figure 4-4. Power Reference Level Test Setup
Model 436A
EQUIPMENT:
PROCEDURE:
Power Meter . . . . . . . . HP 432A
Thermistor Mount . . . . . . HP 478A-H75
Digital Voltmeter (DVM). . . . HP 3490A
1.
Set up the DVM to measure resistance and connect the DVM between the V connector on the rear panel of the 432A, and pin 1 on the thermistor mount end of the 432A interconnect cable.
Round off the DVM indication to two decimal places and record this value as the
2.
internal bridge resistance (R) of the 432A (approximately 200 ohms).
Connect the 432A to the Power Meter as shown in Figure 4-4.
3. Set the Power Meter LINE switch to ON (in) and the POWER REF switch to off
4. (out). Then wait thirty minutes for the 432A thermistor mount to stabilize be­fore proceeding to the next step.
Set the 432A RANGE switch to COARSE ZERO and adjust the front-panel
5. COARSE ZERO control to obtain a zero meter indication.
Fine zero the 432A on the most sensitive range, then set the 432A RANGE switch
6. to 1 mW.
NOTE
Ensure that DVM input leads are isolated from chassis ground when performing the next step.
RF
4-8
7.
Set up the DVM to measure microvolt and connect the positive and negative input
leads, respectively, to the V
COMP
and V
connectors on the rear panel of the
RF
432A. Observe the indication on the DVM. If less than 400 microvolt, proceed to the
8. next step. If 400 microvolt or greater, press and hold the 432A FINE ZERO switch and adjust the COARSE ZERO control so that the DVM indicates 200 microvolt or less. Then release the FINE ZERO switch and proceed to the next step.
Round off the DVM indication to the nearest microvolt and record this value as V
9.
.
0
Page 83
Model 436A
PERFORMANCE TESTS
4-13. POWER REFERENCE LEVEL TEST (cont’d)
Performance Tests
10.
Set the Power Meter POWER observed on the DVM as V
11.
Disconnect the DVM negative input lead from the V
REF switch to ON (in) and record the indications
.
1
connector on the 432A
RP
and reconnect it to 432A chassis ground. Record the new indication observed on the DVM a
Calculate the power reference oscillator output level (P
12.
S V
COMP
.
) from the following
RF
formula:
Where:
P
= power reference oscillator output level
RF
= previously recorded value
v
COMP
= previously recorded value
V
1
= previously recorded value
V
0
R = previously recorded value CALIBRATION FACTOR = value for thermistor mount at 50 MHz (traceable to
the National Bureau of Standards)
Verify that the P
13.
is within the following limits:
RF
4-9
Page 84
Performance Tests
Model 436A
Table 4-5. Performance Test Record (1 of 2)
Hewlett-Packard Company Model 436A Power Meter
Serial Number
Para.
No.
4-10.
ZERO CARRYOVER
Test
100
10
/..lw /.lw
Tested By
Date
1 mW
10 mW
100 mW
4-11. INSTRUMENTATION ACCURACY
WATT MODE
10
/.lw
100
#w
1 mW
10 mW
100 mW
Results
Min
-0.02
-0.2
/.fw
/Jw
-0.002 mW
Actual Max
0.02
0.2
/.lw
0.002 mW
/.lw
-0.02 mW 0.02 mW
-0.2 mW
9.95
/.lw
99.5
/.lw
0.2 mW
10.05
100.5
/Jw /.fw
0.995 mW 1.005 mW
9.95 mW 10.05 mW
99.5 mW 100.5 mW
dBm MODE
-20 dBm
-10 dBm 0 dBm -0.02 dBm 0.02 dBm
10 dBm 20 dBm 19.96 dBm 20.04 dBm
dB (REL) MODE
-20 dBm
- 5 dBm
+10 dBm
4-12. CALIBRATION FACTOR
100
99 98 97 1.025 mW 96 95 94 93
-20.02 dBm -19.98 dBm
-10.02 dBm
-9.98 dBm
9.95 dBm 10.02 dBm
-9.96 dBm -10.04 dBm +4.96 dBm
+5.04 dBm
+19.96 dBm 20.04 dBm
0.994 mW
1.004 mW
1.014 mW
1.006 mW
1.016 mW
1.026 mW
1.037 mW
1.036 mW
1.047 mW
1.058 mW
1.069 mW
1.048 mW
1.059 mW
1.070 mW
1.081 mW
4-10
Page 85
Model 436A
Performance Tests
Table 4-5. Performance Test Record (2 of 2)
Para.
No.
4-12.
4-13
Test
CALIBRATION FACTOR (cont’d)
92 1.081 mW
91
90
89 88 87
86
85
POWER REFERENCE
P
RF
Results
Min.
Actual Max
1.093 mW
1.093 mW
1.105 mW
1.118 mW
1.105 mW
1.117 mW
1.130 mW
1.130 mW 1.142 mW
1.143 mW
1.155 mW
1.157 mW 1.169 mW
1.170 mW
1.182 mW
0.981 mW 1.019 mW
4-11/4-12
Page 86
Page 87
Model 436A
SECTION V
ADJUSTMENTS
INTRODUCTION
5-1.
This section describes the adjustments which
5-2.
return the Power Meter to peak operating
will condition after repairs are completed.
5-3. If the adjustments are to be considered valid, the Power Meter must have a half-hour warmup and the line voltage must be within +5 to -10% of nominal.
5-4. SAFETY CONSIDERATIONS
5-5. Although this instrument has been designed in accordance with international safety standards, this manual contains information, warnings which must be followed to ensure safe
operation and to retain the instrument in safe condition (see Sections II and III). Service and adjustments should be performed only by qualified
service personnel.
cautions,
and
Adjustments
5-9. Whenever it is likely that the protection offered by fuses has been impaired, the instrument must be made inoperative and secured against any unintended operation.
Adjustments described herein are per­formed with power supplied to the instru-
ment while protective covers are removed. Energy available at many points may, if contacted, result in personal injury.
5-10. EQUIPMENT REQUIRED
5-11. The test equipment required for the adjust­ment procedures is listed in Table 1-2, Recom­mended Test Equipment. The critical specifications
of substitute test instruments must meet or exceed the standards listed in the table if the Power Meter is to meet the standards set forth in Table 1-1,
Specifications.
Any
(grounding) conductor (inside the instrument) or disconnection of the
protective earth terminal is likely to make
the instrument dangerous. Intentional
interruption is prohibited.
5-6. Any adjustment, maintenance, and repair of the opened instrument with voltage applied should be avoided as much as possible and, when inevit­able, should be carried out only by a skilled person who is aware of the hazard involved.
5-7. Capacitors inside the instrument may still be charged even if the instrument has been discon­nected from its source of supply.
5-8. Make sure that only fuses with the required rated current and of the specified type (normal blow, time delay, etc.) are used for replacement. The use of repaired fuses and the shortcircuiting of fuseholders must be avoided.
interruption of the
protective
or outside
5-12. FACTORY SELECTED COMPONENTS
5-13. Factory selected components are indicated on the schematic and replaceable parts list with an asterisk immediately following the reference desig­nator. The nominal value of the component is listed. Table 5-1 lists the parts by reference designator and provides an explanation of how the component is selected, the normal value range, and a reference to the appropriate service sheet. The Manual Changes supplement will update any changes to factory selected information.
5-14. ADJUSTMENT LOCATIONS
5-15. The last foldout in this manual contains a table which cross-references all pictorial and
schematic locations of the adjustment controls. The accompanying figure shows the locations of the adjustable controls, assemblies, and chassis­mounted parts.
component
5-1
Page 88
Adjustments
Model 436A
ADJUSTMENTS
Table 5-1. Factory Selected Components
Reference Designator
A2R18
A display readout of 100.0 mW if the Power
Selected For
Meter, after being properly adjusted, passes
all of the Instrumentation Accuracy Tests specified in Section IV except for the high
range (100 mW/20 dBm)
A2R50
Adjust A2R69 FREQ (Frequency Adj) for maximum indication on digital readout, then check frequency of 220 Hz Multivib­rater. If out of specification (220 ± 16 Hz) select value for A2R50 to produce maximum indication on digital readout while 220 Hz Multivibrator frequency is in specification.
A8R5 A Power Reference Oscillator output of
1 mW if this value falls outside the range of adjustment available with LEVEL
ADJUST potentiometer A8R5.
Normal Value
Range
196K
(150K~ to
250Kfl
)
13.3Kf2
(1OK!2
to
17.8KC?)
7100
(7100s’2
to
7500Q)
Service
Sheet
7
7
14
5-16. DC OFFSET
REFERENCE:
DESCRIPTION:
EQUIPMENT:
PROCEDURE:
ADJUSTMENT
Service Sheet 8.
DC OFF potentiometer A3R2 is adjusted to remove any dc voltage introduced by the dc amplifier
Figure 5-1. DC Offset Adjustment Setup
Range Calibrator . . . . . . . Hp 11683A
Set the Power Meter Switches as follows:
1.
CAL FACTOR % . . . . . 100
POWER REF . . . . . .
off (out)
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
off (out) ON (in)
5-2
Page 89
Model 436A
5-16. DC OFFSET ADJUSTMENT (cont’d)
2.
Set the Range Calibrator switches as follows:
FUNCTION . . . . . . . CALIBRATE
POLARITY . . . . . . . NORMAL
RANGE . . . . . . . . 100mW
LINE . . . . . . . . .
3.
Connect the equipment as shown in Figure 5-1.
4.
Verify that the Power Meter autoranges to the 100 mW range, then set the RANGE HOLD switch to ON (in).
5.
Set the Range Calibrator FUNCTION switch to STANDBY. Remove the Power Meter top cover and adjust DC OFF potentiometer A3R2 so
6. that the digital readout indicates 00.0 with a blinking minus sign.
Adjustments
ADJUSTMENTS
ON (in)
5-17. AUTO ZERO OFFSET ADJUSTMENT
REFERENCE: Service Sheet 8.
DESCRIPTION: ZERO OFF potentiometer A3R47 is adjusted to remove any dc offset that is intro-
duced when the SENSOR ZERO switch is pressed.
Figure 5-2. Auto Zero Offset Adjustment Setup
EQUIPMENT: PROCEDURE: 1.
Range Calibrator . . . . . . . HP 11683A
Set the Power Meter switches as follows:
CAL FACTOR % . . . . . 100
POWER REF . . . . . . off (out)
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
off (out) ON (in)
2.
Set the Range Calibrator switches as follows:
FUNCTION . . . . . . . STANDBY
POLARITY . . . . . . . NORMAL
LINE . . . . . . . . .
3.
Connect the equipment as shown in Figure 5-2.
ON (in)
5-3
Page 90
Adjustments
ADJUSTMENTS
5-17. AUTO ZERO OFFSET ADJUSTMENT (cont‘d)
4.
Verify that the Power Meter autoranges to the 10
Meter top cover.
If specified indication cannot be obtained in next step, perform DC Spike Balance Adjustment. Then repeat
this procedure.
5.
Press and hold the Power Meter SENSOR ZERO switch and adjust ZERO OFF potentiometer A3R47 so that the digital readout indicates 0.00 with blinking minus sign.
5-18. SPIKE BALANCE ADJUSTMENT
NOTE
Model 436A
PW
range, and remove the Power
REFERENCE: DESCRIPTION:
EQUIPMENT:
PROCEDURE:
Service Sheets 7 and 8.
A reference signal is applied to the Power Meter from the Range Calibrator to force the sensor zero circuit to its negative extreme. The SENSOR ZERO switch is then held pressed while BAL potentiometer A3R65 is adjusted to center the sensor zero
circuit output voltage range.
Figure 5-3. Spike Balance Adjustment Setup
Range Calibrator . . . . . . . HP 11683A
1.
Set the Power Meter switches as follows:
CAL FACTOR % . . . . 100
POWER REF. . . . . .
off (out)
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
off (out) ON (in)
5-4
2.
Set the Range Calibrator switches as follows:
FUNCTION . . . . . . . CALIBRATE
POLARITY . . . . . . . NORMAL
RANGE . . . . . . . .
LINE . . . . . . . . .
ON (in)
Page 91
Model 436A
ADJUSTMENTS
5-18. SPIKE BALANCE ADJUSTMENT (cont’d)
3.
Remove the Power Meter top cover and adjust the front-panel CAL ADJ control
so that the digital readout indicates 100.0
4.
Press and hold the Power Meter SENSOR ZERO switch and adjust BAL poteni­ometer A3R65 so that the display readout indicates 60.0 ± 0.2
The Power Meter sensor zero circuit must be
re-zeroed as described in the following steps
before valid power measurements can be made.
5.
Set the Range Calibrator FUNCTION switch to standby. Then press the Power Meter SENSOR ZERO switch and wait for the digital readout to stabilize.
6.
Release the Power Meter SENSOR ZERO switch and wait for the ZERO lamp to go out.
Adjustments
PW
MW.
NOTE
5-19. MULTIVIBRATOR ADJUSTMENT
REFERENCE: DESCRIPTION:
EQUIPMENT:
PROCEDURE: 1. Set the Power Meter switches as follows:
Service Sheet 7. FREQ potentiometer A2R69 is adjusted to set the reference frequency of the multi-
vibrator which drives the phase detector and the FET power sensor.
Figure 5-4. Multivibrator Adjustment Setup
Range Calibrator . . . . . . . HP 11683A
Counter . . . . . . . . . . HP 5245L
CAL FACTOR % . . . . . 100
POWER REF . . . . . .
MODE . . . . . . . . . WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
off (out) off (out)
ON (in)
5-5
Page 92
Adjustments
ADJUSTMENTS
5-19. MULTIVIBRATOR ADJUSTMENT (cont’d)
Set the Range Calibrator switches as follows:
2.
FUNCTION . . . . . . . CALIBRATE
POLARITY . . . . . . . NORMAL
LINE . . . . . . . . . ON (in)
3.
Connect the equipment as shown in Figure 5-4.
4.
Remove the Power Meter top cover, adjust FREQ potentiometer A2R69 to obtain maximum indication on the digital readout, and verify that the counter indicates 220 ± 16 Hz.
5.
Perform the Instrument Accuracy Test described in Section IV to verify overall Power Meter accuracy. If all indications are obtained as specified, the adjustment is complete. If any indication cannot be obtained as specified, perform the A-D Converter and Linear Meter Adjustment.
Model 436A
5-20. A-D CONVERTER AND LINEAR METER ADJUSTMENT
REFERENCE: Service Sheets 7 and 8. DESCRIPTION: The A-D converter circuit is adjusted to obtain the specified digital
and the meter circuit is adjusted for a corresponding indication.
Figure 5-5. A-D Converter and Linear Meter Adjustment Setup
EQUIPMENT:
Range Calibrator . . . . . . . HP 11683A
Digital Voltmeter (DVM). . . . HP 3490A
PROCEDURE:
1.
Set the Power Meter switches as follows:
CAL FACTOR %. . . . . 100
POWER REF . . . . . .
off (out)
MODE . . . . . . . . .. WATT
RANGE HOLD . . . . .
LINE . . . . . . . . .
off (out)
ON (in)
readout accuracy
5-6
Page 93
Model 436A
ADJUSTMENTS
5-20. A-D CONVERTER AND LINEAR METER ADJUSTMENT (cont’d)
2.
Set the Range Calibrator switches as follows:
FUNCTION . . . . . . . STANDBY
RANGE . . . . . . ..1 mW
POLARITY . . . . . . . NORMAL
LINE . . . . . . . . . ON (in)
3.
Connect the equipment as shown in Figure 5-5.
4.
Remove the Power Meter top cover and set the DVM to the 1000 mV range.
Press the Power Meter SENSOR ZERO switch and wait for the display readout to
5. stabilize. Then release the SENSOR ZERO switch and wait for ZERO led to go out before proceeding to the next step.
6.
Set the Range Calibrator FUNCTION switch to CALIBRATE and adjust the Power Meter front-panel CAL ADJ control to obtain a 1.000 Vdc indication on
the DVM.
Adjustments
7.
Adjust the Power Meter LIN potentiometer A3R37 so that the digital readout in­dicates 1.000 mW.
8.
Set the Power Meter MODE and
RANGE HOLD switches to dBm and on (in),
respectively.
NOTE
The next step sets the
A-D log threshold. When the speci­fied indication (-10.00 dBm) is obtained, the digital-read­out should be just on the verge of blanking, i.e., the readout may randomly alternate between -10.00 and UNDER
RANGE, -1.
9.
Set the Range Calibrator RANGE switch to -10 dBm and adjust the power
meter LZR,
10.
Set the Power Meter RANGE HOLD switch to off (out) and the Range Calibrator
-
A3R59, for -10 dBm.
RANGE switch to 1 mW.
11.
Adjust Power Meter LFS potentiometer A3R48 so that the digital readout indi­cates -0.00.
12.
Set the Power Meter MODE switch to WATT and adjust MTR potentiometer A3R17 so that the pointer is aligned half way between the last two marks on the meter face.
5-7
Page 94
Adjustments
ADJUSTMENTS
5-21. POWER REFERENCE OSCILLATOR FREQUENCY ADJUSTMENT
NOTE
Adjustment of the Power Reference Oscillator fre-
quency may also affect the output level of the
oscillator. Thus after the frequency is adjusted to
50.0 ± 0.5 MHz, the output level should be checked
as described in Section IV. A procedure for adjust-
ing the output to the specified level is provided in the next paragraph.
Model 436A
REFERENCE:
DESCRIPTION:
EQUIPMENT:
PROCEDURE:
Service Sheet 14. Variable inductor A8L1 is adjusted to set the power reference oscillator output fre-
quency to 50.0 ± 0.5 MHz.
Figure 5-6. Power Reference Oscillator Frequency Adjustment Setup
Counter . . . . . . . . . . HP 5245L
1.
Set the Power Meter LINE switch to ON (in) and the POWER REF switch to off (out).
Set up the counter to measure frequency and connect the equipment as shown in
2. Figure 5-6.
Set the Power Meter POWER REF switch to ON (in) and observe the indication
3.
on the counter. If it is 50.0 ± 0.5 MHz, no adjustment of the power reference oscillator frequency is necessary. If it is not within these limits, adjust the power reference oscillator frequency as described in steps 4 through 9.
5-8
4.
Remove the Power Meter top cover.
Take care not to ground the +15V or -15V inputs to the power reference oscillator when performing the following steps. Grounding either of these inputs could damage the power reference oscillator, and/or the power supply.
Grasp the power reference oscillator assembly firmly, and remove the four screws
5. which secure it to the Power Meter chassis.
Page 95
Model 436A
ADJUSTMENTS
5-21. POWER REFERENCE OSCILLATOR FREQUENCY ADJUSTMENT (cont’d)
Tilt the power reference oscillator assembly to gain access to the circuit board
6. underneath the metal cover, and adjust A8L1 to obtain a 50.00 ± 0.5 MHz indi­cation on the counter.
Reposition the power reference oscillator on the Power Meter chassis but do not
7. replace the mounting screws.
Observe the indication on the counter. If it is 50.0 ± 0.5 MHz, the adjustment
8.
procedure is complete. If it is not within these limits, repeat steps 6 and 7 except offset the power reference oscillator frequency as required to obtain a 50.0 ±
0.5 MHz indication on the counter when the power reference oscillator assembly is repositioned on the Power Meter chassis.
Replace the four screws which secure the power reference oscillator to the Power
9.
Meter chassis.
Adjustments
5-22. POWER REFERENCE OSCILLATOR LEVEL ADJUSTMENT
REFERENCE: Service Sheet 14. DESCRIPTION: The power reference oscillator output is factory-adjusted to 1 mW ± 0.7% using a special
measurement system accurate to 0.570 (traceable to the National Bureau of Standards) and allowing for a 0.2% transfer error. To ensure maximum accuracy in readjusting the power reference oscillator, the following procedure provides step-by-step instructions for using specified Hewlett-Packard instruments of known capability. If equivalent in-
struments are used, signal acquisition criteria may vary and reference should be made to the manufacturer’s guidelines for operating the equipment.
NOTE
The Power Meter may be returned to the nearest HP office to have the power reference oscillator checked and/or adjusted. Refer to Section II, PACKAGING.
EQUIPMENT:
Figure 5-7. Power Reference Oscillator Level Adjustment Setup
Power Meter . . . . . . . . HP 432A
Thermistor Mount . . . . . . HP 478A-H75
Digital Voltmeter (DVM) . . . . HP 3490A
5-9
Page 96
Adjustments
ADJUSTMENTS
5-22. POWER REFERENCE OSCILLATOR LEVEL ADJUSTMENT (cont’d)
Model 436A
PROCEDURE: 1.
Set up the DVM to measure resistance and connect the DVM between the V nector on the rear panel of the 432A and pin 1 on the thermistor mount end of the 432A interconnect cable.
Round off the DVM indication to two decimal places and record this value as the in-
2.
ternal bridge resistance (R) of the 432A (approximately 200 ohms). Connect the 432A to the Power Meter as shown in Figure 5-7.
3.
4.
Set the Power Meter LINE switch to ON (in) and the POWER REF switch to off (out). Then wait thirty minutes for the 432A thermistor mount to stabilize before proceed­ing to the next step.
Set the 432A RANGE switch to COARSE ZERO and adjust the front-panel COARSE
5.
ZERO control to obtain a zero meter indication.
Fine zero the 432A on the most sensitive range, then set the 432A RANGE switch to
6.
1 mW.
NOTE
Ensure that the DVM input leads are isolated from chassis
ground when performing the next step.
7.
Set up the DVM to measure microvolt and connect the positive and negative inputs leads, respectively, to the V 432A.
COMP
and V
connectors on the rear panel of the
RF
RF
con-
5-10
8.
Observe the indication on the DVM. If less than 400 microvolt, proceed to the next step. If 400 microvolt or greater, press and hold the 432A FINE ZERO switch and adjust the COARSE ZERO control so that the DVM indicates 200 microvolt or less. Then release the FINE ZERO switch and proceed to the next step.
9.
Round off the DVM indication to the nearest microvolt and record this value as V Disconnect the DVM negative input lead from the V
10,
connector on the 432A and
RF
reconnect it to chassis ground.
11.
Set the Power Meter POWER REF switch to ON (in) and record the indication ob­served on the DVM as V
Disconnect the DVM negative input lead from chassis ground and reconnect it to the
12.
V
connector on the rear panel of the 432A. The DVM is not setup to measure
RF
V
which represents the power reference oscillator output level.
1
13.
Calculate the value of V
1
.
COMP
equal to 1 milliwatt from the following equation:
.
0
Page 97
Model 436A
ADJUSTMENTS
5-22. POWER REFERENCE OSCILLATOR LEVEL ADJUSTMENT (cont’d)
where:
V
= previously recorded value
0
= previously recorded value
v
COMP
-3
10
= 1 milliwatt R = previously recorded value EFFECTIVE EFFICIENCY = value for thermistor mount at 50 MHz (traceable to the
National Bureau of Standards).
14. Remove the Power Meter top cover and adjust LEVEL ADJUST potentiometer
A8R4 so that the DVM indicates the calculated value of V
Adjustments
.
1
TYPICAL
CALCULATIONS: 1.
ACCURACY: DVM Measurements: (HP 3490A -90 days, 23°C ±5°C)
Math Assumptions: EFFECTIVE EFFICIENCY CAL (NBS):
MISMATCH UNCERTAINTY:
(Source & Mount SWR
MATH ASSUMPTIONS:
2.
Assume:
v
0
- (V1 - V0)2 V
Want:
0
2
– V
2
- V
<
1.05)
2
= (V1 - V0)
1
= -V
2
1
2
1
(V
COMP
(v
– V0) ±0.023%
l
2
+ 2V1 -V
(R)
0
2
)
±0.018%
±0.03% ±0.01%
±0.5%
±0.1%
<
±0.7%
if 2V0(V1-V0) < <2V
-4
v
COMP
(typically V
volts. If V0 <400
can be set to < 50 uV).
0
(V1 - V0) i.e., V0 << V
COMP
LLV,
error is < 0.01%.
, error is negligible.
COMP
5-11
Page 98
Adjustments
ADJUSTMENTS
5-22. POWER REFERENCE OSCILLATOR LEVEL ADJUSTMENT (cont’d)
3.
TYPICAL
Derivation of Formula for V1 - V
0
CALCULATIONS (cont’d)
Model 436A
Desired P
= 1mmW = 10
RF
-3
Let (4R) (EFFECTIVE EFFICIENCY) (10-3) = K
2
Substitute -(V Then 0 = (V
l
- V0)2 for V
1
- VO)2 - 2V
COMP (Vl
- V
0
2
(see Math Assumptions under Accuracy)
1
- V0) + K
5-12
Page 99
Model 436A
Replaceable Parts
SECTION VI
REPLACEABLE PARTS
6-1. INTRODUCTION
6-2. This section contains information for ordering parts. Table 6-1 lists abbreviations used in the parts list and throughout the manual. Table 6-2 lists all replaceable parts in reference designation order. Table 6-3 contains the names and addresses that correspond with the manufacturers’ code numbers.
6-3. ABBREVIATIONS
6-4. Table 6-1 lists abbreviations used in the parts list, schematics and throughout the manual. In some cases, two forms of the abbreviation are used, one all in capital letters, and one partial or no capitals. This occurs because the abbreviations in the parts list are always all capitals. However, in the schematics and other parts of the manual, other abbreviation forms are used with both lower case and upper case letters.
6-5. REPLACEABLE PARTS LIST
6-6. Table 6-2 is the list of replaceable parts and is organized as follows:
Electrical assemblies and their compo-
a. nents in alpha-numerical order by reference designation.
b. Chassis-mounted parts in alpha-numerical
order by reference designation.
Miscellaneous parts.
c.
The information given for each part consists of the
following:
a. The Hewlett-Packard part number.
b. The total quantity (Qty) used in the instrument.
c. The description of the part.
d. A typical manufacturer of the part in a
five-digit code.
6-7. ORDERING INFORMATION
6-8. To order a part listed in the replaceable parts table, quote the Hewlett-Packard part number, indicate the quantity required, and address the
order to the nearest Hewlett-Packard office.
6-9. To order a part that is not listed in the replaceable parts table, include the instrument model number, instrument serial number, the description and function of the part, and the number of parts required. Address the order to the nearest Hewlett-Packard office.
6-10. PARTS PROVISIONING
6-11. Stocking spare parts for an instrument is
often done to ensure quick return to service after a malfunction occurs. Hewlett-Packard has a Spare Parts Kit available for this purpose. The kit consists
of selected replaceable assemblies and components
for this instrument. The contents of the kit and the
Recommended Spares list are based on failure
reports and repair data, and parts support for one year. A complimentary Recommended Spares list
for this instrument may be obtained on request
and the Spare Parts Kit may be ordered through
your nearest Hewlett-Packard office.
6-12. DIRECT MAIL ORDER SYSTEM
6-13. Within the USA, Hewlett-Packard can supply
parts through a direct mail order system. Advan-
tages of using the system are:
a.
Direct ordering and shipment from the HP
Parts Center in Mountain View, California.
b. No maximum or minimum on any mail
order (there is a minimum order amount for parts
ordered through a local HP office when the orders
require billing and invoicing).
Prepaid transportation (there is a small
c.
handling charge for each order).
d. No invoices ­tages, a check or money order must accompany each order.
to provide these advan-
e.
The manufacturer’s number for the part.
The total quantity for each part is given only once at the first appearance of the part number in the list.
6-14. Mail order forms and specific ordering information is available through your local HP office. Addresses and phone numbers are located at the back of this manual.
6-1
Page 100
Replaceable Parts
Model 436A
Table 6-1. Reference Designations and Abbreviations (1 of 2)
REFERENCE DESIGNATIONS
A
. . . . . . . . . . .
AT. . B. . .
BT. , . . . . . . . ..
c
“CP”lllllIllj:.aupler
CR . . . . . . . dtode; diode
DC .,.
DO .,.......
Da
A
ac
ACCBSS . . . . . accaeeory
ADJ , . . . . . .
AID
AF . . . . .
AFC, . . . . . ..autanattc
‘AGC . . . . .
~ AL
ALC . . . .
AM . . .
AMPL . . . . . .
APC, ,..
AaaY . . . . . . ..ae8etnbm
AUX . . . . . . .
~S . . . . . . . . . ..avemge
AW(3 . . . .
BAL . . . . . . . . ..
BCD . . . . . .
BD . . . . . . . . . . .. boud
BE CU
BFO . . . .
BH . . . . . . . . binder head
*KDN . . . . . .
BP . . . . . . . . .. bartdpeee
BPF .,...
Bra . . . . . . . . . . .
BWO, .,..
CAL . . . . . . .
“8EWR: :......,.
CHAN
cm . . . . . . . .. centimeter
CMO . .
COAX, . . . . . . .
attenuator; isolator;
termination
...0.. ,
thyrktor; varactor
direettortel
. . . . . . .
dgnaltng
(audtbla or vieuat); lamp: LED
. . . . . . . . . .
... .
alternating cument
.,..
freauancy
control
. . . . . . . . . aluminum
.
control
amplttude modula-
tion
control
*uge
dectmat
. . . . . .
copper
.
Oeetllator
oesttttator
courtter-elockwtee
. . . . . . . . .
cabtnet
assembly
fan; motor
bettery
capacitor
coupler
delay
amtmeiator;
device
..ermae
adjustment
nalog-to-digital
audio frequency
cotttzol
automatic gabs
automatic level
..atttpltfter
automatic phase
..auxUtary
American wtre
balance
binuy coded
basyllium
beat
freauenc
breakdown
bandpau ftlter
bac~werd-wave
..caltbrate
ceramic
channel
mount
.
COSULtd
line
brass
only
E
. . . . . . . .
electricef pert F FL”::::::::::::” H HA”:::::::::” J. . .electrical connector
(stationary Portion);
jack
K. . . . . . . . . . . .
L. . . . . . .
M MP” ::::::: “r&~eihme%~
mecharttcaf part
COEF . . . . . .
COM . . . . . . . ..
COMP ,. ...
COMPL , . . . . .
CONN . . . . . . .
CP . . . . . .
CRT .,.
CTL . . . .
Cw . . . . .
Co w . . . . . . . .
cm.........centtmeter
D/A . . . .
da b . . . . . . . . . .
dBm . . . .
dc . . . . . . . direct current
deg , .
0
. . . . . . . . .
‘c ..,...
~F ,.. .
K DEPC . .
DEB T . . . . . . . .
dlam. ., . . . .
y
DIA . . .
DIFF
div. , . . . . . . . . ..dtvidon
DPDT . . . . .
DR . . . . . . . . . . . . drtve
DSB, . . .
DTL . . . .
DVM ECL .:::
EMF . ,
cathode-ray
complementary
traneiet.or logic
continuous
digttal-tc-analog
decibel refetred
to
1
mW
degree (temperature
interval or
ence)
angte ) (centigrade)
degree Fahrenheit
0 . . . . . .
depoeited
diameter (used in
Parfe list)
AMPL . . dffferentiat
arOPtifier
double-throw
double sideband
diode
Ioglc
digital
emitter coupled
logic
electromotive
miscellaneous
fuse
filter
herd ware
circulator
.
relay
colt;
tnductor
.
ABBREVIATIONS
.
coefficient
common
composition
.
complete
connector
eadmtum ptate
double-pole,
tube
wave
clockwise
decibel
diffe~
degree (ptane
degree Cefatue
degree Kelvin
carbon
detector
..diarneter
trandetor
voltmetm
force
P
R RT” :: ::::::: “thermistor
s
T TAB”::::::””
TC . . . . . .
TA P . . . . . . . . testpoint
EDP, . . . .
ELECT . . . . .
ENCAP , . . .
EXT ..,......
F FET” ::: :::: “
F/F flip-flop
AH.:::::::::
FIL H . . . . . fillister head
FM
FP . . . . . . . . front pane]
FREQ . . . . . . .
FAD . . . . . . . . . . . fixed
g
GE”: :::: :::” &na%%
GHz . . . . . , . . . gigahertz
GL . . . . . . . . . . . ..gbw.s
GRD . . . . . . .
H h HET
HEX . . . . . . . . hexagonaf
HD . . . . . . . . . . . .. head
HDW . . . . . . .. hardware
HF . . . . . .
HA . . . . . . . . . . mercury
HI . . . . . . . . . . . ..tdgh
HP HPF”: :::.
HR . . . . . . .
HV . . . . . . . .
Hz . . . . . . . . . . .. Hertz
IC . . . .
ID . . . . . .
IF . . , . . . .
IMPG ., . . . impregnated
irt . . . . . . . . . . . . .. inch
INCD ., . . . INCL INP . : :;::::...
INS . . . . . . . .
electrical
. . .
(movable portion): plug
. . . . . .
Q
transistor: SCR;
triode thyristor
. . . . . . . . . . . .
terminet board
thermocouple
electronic
processing
encapsulated
tranrdetor
. .
frequency modulation
. . . . . . . . . . . . .
:::::: : “
high
Hewlett-Packard
IrIgh pass filter
hour (ueadin
parts
list)
highvoltage
integrated circuit
tneide diameter
intermediate
frerIuenc y
incandescent
connector
resistor
switch
transformer
data
electrolytic
external
farad
tieid-etfect
flat head
frequency
ground(ed)
henry
hour
h&&odyne
frequent y
tnclude(e)
input
insulation
u
. . . . .
integrated circuit:
v electron tube
w. . . x
Y
INT. ..internafnternaf
kg . . . . . . . . .. kilogram
kHz k~.:::::::::.
kV . . . . . . . . . ..lrilovolt
lb . . . . . . . . . . . .
LC . , . . . . . ,
LED . .
LF . . . . . .
LG . . . . . . . . . . . .. long
LH . . . . . . . . ..
LIM . . . . . . . . . . .. Iimit
LIN . . .
lin . . . . . . . . . . ..
LK LO . . .
LOG . . . .
10g . . . . . . . .
LPF . . . . .
LV . . . . . . . . Iow voltage
m. . . . . .meter
mA . . . . . . . . milliampere
MAX M~. :::: ::. .megohm MEG ...,
MET FLM . . . . metal film
MET OX . .
MF . .
MFR . . . . . . manufacturer
reg . . . . . . . . .
MHz . . . . . . . .
mH . . . . . . . ..miUthenry
mho. , . . . . . . . . ..mho
MIN . . . . . . . .
miss . . . . .
. ..’....
MINAT . . . . . .
mm . . . . . . ..
microcircuit
All”:;;:’””
z. . .
WASH . .
voltage regulator;
breakdown
.
cable; transmission
path;
. . . . . . . . . . . .
. . . .
wire
crystal
electric or
.
tuned
circuit
capacitance
fight-em itting
low fr:quency
linear taper (used
in perta
low;
local oscillator
logarithmic taper
(used in parts
meg
in parts
.
medium frequency; mlcroferarl (ueed parts list )
.
angle)
diode
socket
unit (piezo-
quartz)
cavtty;
tuned
kilohertz
kilohm
pound
inductance-
diode
left
hand
list)
linear
.
lock
washer
list)
logarithm
low
pass filter
(distance
.
maximum (106) (used
list)
metalfic
oxide
in
mffligram
megehertz
minimum
minute (time) mtnute
(plane
.
miniature
millimeter
)
6-2
Aft
bbreviatlone in the
TfOTE
parts
liet wffl be
in upper-caee.
Loading...