HP 400D, 4001, 400H, H02-400D Organizational, Ds, Gs, And Depot Maintenance Manual

TM 11-6625-1514-15
DEPARTMENT OF THE ARMY TECHNICAL MANUAL
ORGANIZATIONAL, DS, GS,
AND DEPOT MAINTENANCE MANUAL
HEWLETT-PACKARD
VOLTMETER
400D, 400H, 4001
This copy is a reprint which includes current pages from Changes 1.
VACUUM TUBE
AND H02-400D
HEADQUARTERS, DEPARTMENT OF THE ARMY
MAY 1967
WARNING
DANGEROUS VOLTAGES
EXIST IN THIS EQUIPMENT
Be careful when working on the power supplies and
their circuits, or on the 230 or 115-volt ac line
connections.
DO NOT TAKE CHANCES
TM 11-5625-1514-15
C1
C
HANGE
HEADQUARTERS
DEPARTMENT OF THE ARMY
No. 1
Washington DC, 28 September 1982
Organizational, Direct Support, General Support and
Depot Maintenance Manual
HEWLETT–PACKARD VACUUM TUBE VOLTMETER MODELS
400D, 400H, 400L, and H02-400D
(NSN 6625-00-643-1670)
TM 11-6625-1514-15, 23 May 1967, is changed as follows:
1. Title of manual is changed as shown above.
2. New or changed material is indicated by a vertical bar in the margin.
3. Added or revised illustrations are indicated by a vertical bar next to the figure caption.
4. Remove old pages and insert new pages as indicated below:
Remove pages
None. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
i and 1-0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i, ii and 1-0
1-0.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-0.1
10-5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
5. File this change sheet in front of the publication for reference purposes.
. Warning page a/(b blank)
Insert pages
By Order of the Secretary of the Army:
Official:
ROBERT M. JOYCE
Major General, United States Army
The Adjutant General
Distribution:
To redistributed in accordance with special list.
E. C. MEYER
General, United States Army
Chief of Staff
TM 11-6625-1514-15
SAFETY STEPS IS THE VICTIM
TO FOLLOW IF OF ELECTRICAL
SOMEONE
SHOCK
DO NOT TRY TO PULL OR GRAB THE INDIVIDUAL
IF POSSIBLE , TURN OFF THE ELECTRICAL POWER
IF YOU CANNOT TURN OFF THE ELECTRICAL POWER, PULL, PUSH, OR LIFT THE PERSON TO
SAFETY USING A WOODEN POLE OR A ROPE OR
SOME OTHER INSULATING MATERIAL
SEND FOR HELP AS SOON AS POSSIBLE
AFTER THE INJURED PERSON IS FREE OF
CONTACT WITH THE SOURCE OF ELECTRICAL SHOCK, MOVE THE PERSON A SHORT DISTANCE AWAY AND IMMEDIATELY START ARTIFICIAL
RESUSCITATION
Change 1 a/(b blank)
This manual contains copyright material reproduced by permission of the Hewlett-Packard Company.
TM 11-6625-1514-15
Technical Manual
HEADQUARTERS
DEPARTMENT OF THE ARMY
No. 11-6625-1514-15
Washington, DC, 23 May 1967
ORGANIZATIONAL, DIRECT SUPPORT, GENERAL SUPPORT AND
DEPOT MAINTENANCE MANUAL
HEWLETT-PACKARD VACUUM TUBE VOLTMETER MODELS
400D, 400H, 400L, AND H02-400D
(NSN 6625-00-643-1670)
Paragraph
Section
Section
Section
Section
Section
I.
GENERAL DESCRIPTION
Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Index of Technical Publications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance Forms, Records, and Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reporting Errors and Recommending Improvements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reporting Equipment Improvement Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Administrative Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Destruction of Army Electronics Materiel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
II.
INSTALLATION
Unpacking and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Line Voltage Requirement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power Line Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operation Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
III.
OPERATING INSTRUCTIONS
Instrument Turn-On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
General Operating Information.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Low-Level Measurements and Ground Currents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measurement of Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measurement of Decibels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Impedance Correction Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Use of Voltmeter Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
IV.
CIRCUIT DESCRIPTION
Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Input Voltage Divider and Step Attenuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Broadband Voltmeter Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Indicating Meter Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
V.
MAINTENANCE
Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Precautions . . . . . . . . . . . . . . . . . . . . . . . .
Test Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Meter Zero Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cabinet Removal . . . . . . . . . . . .
Tube Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacement of Special Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Trouble Shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Testing the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Testing Voltmeter Performance
Calibration and Frequency Response Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-A.1 1-A.2 1-A.3 1-A.4 1-A.5 1-A.6 1-A.7
1-1 1-4
2-1 2-3 2-5 2-8
2-10
3-1
3-3
3-12
3-14 3-17 3-20 3-22
4-1 4-3 4-7
4-10 4-14
5-1 5-3 5-5 5-7 5-9
5-10 5-13 5-17 5-20 5-22 5-24
Page
1-0.1 1-0.1 1-0.1 1-0.1 1-0.1 1-0.1 1-0.1 1-1 1-1
2-1 2-1 2-1 2-1 2-1
3-1 3-1 3-2 3-2 3-3 3-3 3-4
4-1 4-1 4-1 4-1 4-2
5-1 5-1 5-1 5-1 5-2 5-2 5-2 5-3 5-3 5-5 5-8
Change 1 i
TM 11-6625-1514-15
Section
Section
Section
Section Section
ILLUSTRATED PARTS BREAKDOWN
VI.
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GROUP ASSEMBLY PARTS BREAKDOWN
VII.
Vacuum Tube Voltmeter 400D/H/L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Main Chassis Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Range Switch Assembly 400D-19A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Printed Circuit Board Assembly 400D-75G . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Printed Circuit Board Assembly 400D-75F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Printed Circuit Board Assembly 400D-65C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
VIII.
NUMERICAL INDEXES
Part No. Numerical Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Hewlett-Packard Stock No. Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
REFERENCE DESIGNATION INDEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
IX.
AUXILIARY EQUIPMENT
X.
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Line Matching Transformer Model 11004A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Bridging Transformer Model 11005A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Final Performance Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Paragraph
6-1 7-1
7-2 7-3 7-4 7-5 7-6
10-1 10-3 10-7 10-8
Page
6-1 7-1
7-3
7-8 7-9 7-10 7-12
8-1 8-2 9-1
10-1 10-1 10-5 10-5
ii
Change 1
1-0
Change 1
TM 11-6625-1514-15
Figure 1-1.

SECTION I

GENERAL DESCRIPTION
TM 11-6625-1514-15
1-A.1. Scope
This manual includes installation and operation in­structions and covers operator’s, organizational, direct support (DS), general support (GS), and depot maintenance. It describes Hewlett-Packard (Federal Supply Code 28480) Vacuum Tube Voltmeter Models 400D and H02-400D, serial numbers 310-45571 and higher; and Models 400H and 400L, serial numbers 313-22177 and higher. A basic issue items list for this equipment is not included in this manual.
1-A.2. Index of Technical
Publications
Refer to the latest issue of DA Pam 310-4 to deter­mine whether there are new editions, changes or ad­ditional publications pertaining to the equipment.
1-A.3. Maintenance Forms, Records,
and Reports
a. Reports of Maintenance and Unsatisfactory
Equipment.
cedures used for equipment maintenance will be those prescribed by TM 38-750, The Army Maintenance Management System (Army). Air Force personnel will use AFR 66-1 for maintenance repor­ting and TO-00-35D54 for unsatisfactory equipment reporting.
b. Report of Packaging and Handling Deficien-
cies. Fill out and forward SF 364 (Report of
Discrepancy (ROD)) as prescribed in AR 735-11-2/ DLAR 4140.55/NAVMATINST 4355.73/AFR 400­54/MCO 4430.3E.
c. Discrepancy in Shipment Report (DISREP)
(SF 361).
ment Report (DISREP) (SF 361) as prescribed in AR 55-38/NAVSUPINST 4610.33B/AFR 75-18/MCO P4610.19C/DLAR 4500.15.
Department of the Army forms and pro-
Fill out and forward Discrepancy in Ship-
1-A.4. Reporting Errors and Recom-
mending Improvements
You can help improve this manual. If you find any
mistakes or if you know of a way to improve the pro­cedures, please let us know. Mail your letter or DA Form 2028 (Recommended Changes to Publications and Blank Forms) direct to Commander, US Army Communications-Electronics Command and Fort Monmouth, ATTN: DRSEL-ME-MQ, Fort Mon­mouth, NJ 07703. A reply will be furnished direct to you.
1-A.5. Report Equipment Improve-
ment Recommendations (EIR)
If your vacuum tube voltmeter needs improvement, let us know. Send us an EIR. You, the user, are the only one who can tell us what you don’t like about your equipment. Let us know why you don’t like the design. Tell us why a procedure is hard to perform. Put it on an SF 368 (Quality Deficiency Report). Mail it to Commander, US Army Communications­Electronics Command and Fort Monmouth, ATTN: DRSEL-ME-MQ, Fort Monmouth, NJ 07703. We’ll send you a reply.
1-A.6. Administrative Storage
Administrative storage of equipment issued to and used by Army activities will have preventive maintenance performed before storing. When removing storage, the PMCS should be performed to assure operational readiness.
1-A.7. Destruction of Army Elec-
Destruction of Army electronics materiel to prevent
enemy use shall be in accordance with TM 750­244-2.
the equipment from administrative
tronics Materiel
Change 1
1-0.1
TM 11-6625-1514-15
Paragraphs 1-1 to 1-5
Section I
1-1. INTRODUCTION.
1-2. This manual contains operating and servicing instructions, and a parts breakdown, for the Models 400D, 400H, and 400L Vacuum Tube Voltmeters manu­factured by the Hewlett- Packard Company. The Model 400D Voltmeter is similar to a military counterpart, Electronic Voltmeter ME-30A/U, in appearance and operation, but contains modified electrical circuits to obtain improved performance. Applicable Federal Stock Numbers for the voltmeters are as follows:
Model 400D: 6625-643-1670 Model 400H: 6625-557-8261 Model 400L: 6625-729-8360
1-3. The Models 400D, 400H, and 400L Voltmeters are the same except for the differences listed in Fig­ure 1-2.
a. Voltage Range: 400D/H - 0.1 millivolt to 300 volts; 400L - 0.3 millivolt to 300 volts, in 12 ranges providing full-scale readings of the following voltages:
0.001
0.003 0.300 30.00
0.010
0.030 3.000
b. Decibel Range: -72 to +52 db, in 12 ranges. c. Frequency Range: 10 cps to 4 mc.
d. Input Impedance: 10 megohms shunted by 15 pf
(15 µµf) on ranges 1.0 volt to 300 volts; 25 pf on ranges
0.001 volt to 0.3 volt.
e. Stability: Line voltage variations of ±10% do not reduce the specified accuracy, and line voltage transients are not reflected in the meter reading. Electron tube deterioration to 75% of normal transconductance affects accuracy less than 0.5% from 20 cps to 1 mc.
f. Amplifier: OUTPUT terminals are provided so that the voltmeter can be used to amplify small signals or to enable monitoring of waveforms under test with an oscilloscope. Output voltage is approximately 0.15
volt rms on all ranges with full-scale meter deflection. Amplifier frequency response is same as the voltmeter. Internal impedance is approximately 50 ohms over entire frequency range.
(See figure l-l. )
0.100
1.000
10.00
100.00
300.00
a. The front panel meters are different in each
model, as described in paragraph 1-6.
b. The accuracy specifications are different for
each model, as described in figure 1-2.
1-4. DESCRIPTION.
1-5. The Hewlett-Packard Models 400D, 400H, and 400L Vacuum Tube Voltmeters are general purpose, portable electronic a-c voltmeters of high sensitivity and stability. They are suited to both laboratory and. field use. Models 400 D/H measure a-c voltages from
0.001 to 300 volts and Model 400L from .003 to 300 volts rms full scale, with a frequency bandwidth cover­ing 10 cps to 4 megacycles. The voltmeters are com­pact, accurate, and rugged and have fast meter re­sponse, high input impedance, stable calibration ac­curacy, and freedom from the effects of normal line
voltage variations.
long instrument life with a minimum of servicing.
g. Accuracy: Model 400D -
± 2% of full scale, 20 cps to 1 mc; ± 3% of full scale, 20 cps to 2 mc; ± 5% of full scale, 10 cps to 4 mc.
Model 400H -
± 1% of full scale, 50 cps to 500 kc; ± 2% of full scale, 20 cps to ± 3% of full scale, 20 cps to
+ 5% of full scale, 10 cps to
Model 400L -
±2% of reading or ±1% of full scale,
whichever is more accurate, 50 cps to 500 kc.
±3% of reading or ±2% of full scale,
whichever is more accurate, 20 cps to 1 mc.
±4% of reading or ±3% of full scale,
whichever is more accurate,
20 cps to 2 mc.
±5% of reading 10 cps to 4 mc.
h. Power Requirement:
1000 cps, approximately 100 watts.
i. Size: 11-3/4 in. high, 7-1/2 in. wide, 12 in. deep. j. Weight: 18 lbs; shipping weight approximately
23 lbs.
The voltmeters are designed for
1 mc; 2 mc; 4 mc.
115/230 volts ±10%, 50 to
00102-3
Figure 1-2. Table of Specifications
1-1
TM 11-6625-1514-15
Section I
Paragraphs 1-6 to 1-10
1-6. Each model voltmeter has three calibrated scales on the panel meter. two linear VOLTS scales, 0 to 1 and 0 to 3, and one DECIBELS scale, the Models 400H and 400L are larger and include a mirror to eliminate parallax in viewing and to facilitate use of the higher scale calibration accuracy of these models. The Model 400L VOLTS scales are logarithmic in calibration, from 0.3 to 1 and 0.8 to 3; and the DECIBELS scale is linear. In all models, the VOLTS scales are calibrated to indicate the root-mean-square (rms) value of an applied sine wave. Actual meter deflection is proportional to the average value of the applied signal, thereby minimizing additional meter deflection due to noise and harmonic distortion.
1-7. A voltmeter output signal is provided at the front panel OUTPUT terminals. to the meter reading and has a waveshape similar to the applied signal.
rms for a full-scale meter reading, regardless of the input signal level. The internal impedance at the OUTPUT terminal is 50 ohms over the full frequency range. adversely affect the accuracy of the voltmeter. This output is valuable for increasing the sensitivity of bridges, etc., where distortion added to the waveform
is not a factor.
1-8. The voltmeter chassis is constructed of aluminum alloy throughout. The panel is finished in non-reflecting, light-grey baked enamel; the cabinet is finished in dark-blue, baked wrinkle paint. The cabinet is equipped with rubber feet and a leather carrying handle. Control markings on the front panel are engraved and black filled. INPUT and OUTPUT terminals are special binding posts which accept either bare wire or banana plugs; the 3/4-inch spacing between binding posts accepts standard dual-banana plugs. The “ground” side of the INPUT and OUTPUT terminals is connected to the instrument chassis which is in turn connected to the
power line ground through the third (round) prong of the plug on the power cable.
High-impedance loads (above 100K) will not
The Models 400D and 400H have
-12 to +2 db. The meters used in
This output is proportional
This signal level is about 0.15 volts
1-9. The voltmeter is equipped with a non-detachable power cord. Test leads, which may be plain wire leads or coaxial cable, and test probes must be supplied by the user.
1-10. Instruments designated Models 400DR, 400HR,
and 400LR are rack mount configurations of the 400D,
400H, and 400L, respectively. They are identical to their cabinet model counterparts in every other re-
spect. They are designed to be mounted in a stan­dard 19 inch wide x 7 inch high relay rack space. Re­fer to Appendix C for Replacement Parts information.
1-2
2-1. UNPACKING AND INSPECTION.
TM 11-6625-1514-15

Section II

Paragraphs 2-1 to 2-11
SECTION II
INSTALLATION
2-2. There are no special precautions for unpacking the voltmeter. Save the shipping carton and packing materials for possible storage or reshipment. When unpacking, inspect instrument and packing materials for signs of damage in shipment. Make an operation check as directed in paragraph 2-10 to determine if performance is satisfactory. If there is any indication of damage or deficiency, refer to
paragraph 1-A.3.
2-3. LINE VOLTAGE REQUIREMENT.
2-4. The voltmeter is wired at the factory for use on
115-volt a-c power. This voltage may vary ±10% without adverse effect upon voltmeter performance. The volt­meter can be wired for use on 230-volt a-c power by reconnecting the dual primary windings on the power transformer as shown in the schematic diagram in Section V. When using 230-volt power, change from a 1-amp to a 1/2-amp slow-blow fuse. If necessary, provide an adapter for attaching the standard 115-volt plug on the voltmeter to the 230-volt outlet.
2-5. POWER LINE CONNECTION.
2-6. The three-conductor power cable on the voltmeter
is terminated in a polarized three-prong male connector. The third contact is an offset round pin added to a stand­ard two-blade connector, which grounds the instrument chassis when used with the appropriate receptacle. To connect this plug in a standard two-contact receptacle,
use an adapter. The chassis ground connection is
brought out of the adapter in a green pigtail lead for
connection to a suitable ground.
The lower INPUT and OUTPUT signal terminals on the panel of the voltmeter are connected directly to the chassis of the voltmeter. Any voltage applied to the lower terminal will be shorted directly to ground. If the ground con­nection in the power cord is disconnected by use of an adapter, the entire voltmeter cabinet will carry whatever potential is applied to the
lower terminal and may be a hazard to the operator.
2-8. INSTALLATION.
2-9. The voltmeter is a portable instrument requiring no permanent installation. The voltmeter is for bench­top operation, standing on its rubber feet with its front
panel near the vertical plane. A bail is provided for
raising the front of the cabinet to obtain a better viewing
angle.
2-10. OPERATION CHECK.
2-11. The voltmeter is ready for use as received from the factory. The simple check described below can be made by incoming inspectors to determine if electrical
damage was incurred in shipment. If more complete
proof of instrument performance is required, the over-all performance check described in paragraph 5-22 must be used. Make a simple performance check as follows:
a. Connect voltmeter to the power line through a variable transformer. Set transformer for 115 volts, turn on and allow a five-minute warmup.
2-7. The power plug normally supplied with the volt-
meter is made of molded rubber and is an integral part of the power cable. On certain military contracts, a modification of the Model 400D, termed the H02-400D,
is equipped with a removable plug having the same pin configuration but constructed of corrosion-resistant
material. In all other respects the H02-400D is the
same as the Model 400D and carries the same Federal Stock Number.
00102-2
b. Measure any sine wave voltage, excepting the power
line,
from 0.01 to 300 volts whose exact voltage is known. Note that the lower INPUT terminal is connected to the power line ground.
c. While making the above measurement, adjust the line voltage from 103 to 127 volts. The reading on the meter must not change by more than the width of the pointer.
2-1
TM 11-6625-1514-15
Section III
3-0
Figure 3-1. Voltmeter Front Panel, Showing Controls and Connectors
00102-2

SECTION III

OPERATING INSTRUCTIONS
TM 11-6625-1514-15
Section III
Paragraphs 3-1 to 3-9
3-1. INSTRUMENT TURN-ON.
3-2. The voltmeter is ready for use as received from the factory and will give specified performance after a few minutes warmup. See Section II for information regarding connection to the power source and to the voltage to be measured. Controls are shown in figure 3-1.
3-3. GENERAL OPERATING INFORMATION.
3-4. METER ZERO CHARACTERISTIC. When the Model 400D and 400H Voltmeters are turned off, the
meter pointer should rest exactly on the zero calibration mark on the meter scale. If it does not, zero-set the meter as instructed in paragraph 5-7. The meter
supplied in the Model 400L Voltmeter is not provided with a mechanical meter zero adjustment. When the voltmeter is turned on with the INPUT terminals
shorted, the meter pointer may deflect upscale slightly;
this deflection does not affect the accuracy of a reading.
NOTE
When the voltmeter RANGE switch is set to the lowest ranges and the INPUT terminals are not terminated or shielded, noise pickup can be enough to produce up to full-scale meter deflec­tion. This condition is normal and is caused by stray voltages in the vicinity of the instru­ment. For maximum accuracy on the .001-volt range, the voltage under measurement should be applied to the voltmeter through a shielded test lead.
3-5. METER SCALES. The two voltage scales on each
of the voltmeter models are related to each other by a factor of 1 10 (10 db). In conjunction with the calib-
rated RANGE switch steps, this provides an intermediate
range step spaced 10 db between “power of ten” ranges, which are 20 db apart. The relationship of the DECIBELS scale to the 0 to 1 VOLT scale is determined by making 0 db on the DECIBELS scale equal to the voltage required to produce 1 milliwatt in 600 ohms (0.775 volts). Thus,
the DECIBELS scale reads directly in dbm (decibels
referred to one milliwatt) across a 600-ohm circuit, and can be used to measure absolute level of sine wave signals. It can also be used to measure relative levels of any group of signals which have the same waveform,
across any constant circuit impedance. The RANGE
switch changes voltmeter sensitivity in 10-db steps
accurate to within ± 1/8 db. The RANGE switch position
indicates the value of a full-scale meter reading.
3-6. CONNECTIONS. Voltmeter test leads must be provided by the user. The type of leads and probes used will depend upon the application, as listed below:
a For connection to low-impedance signal sources,
plain wire leads often are sufficient. 00102-2
b. For high-impedance sources, or where noise pickup
is a problem, low-capacity shielded wire must be used with a shielded, dual banana plug for connection to the voltmeter terminals.
c. If a probe is used, it should also be shielded to
prevent pickup from the hand.
d. For signals above a few hundred kilocycles, the capacity of the test leads must be kept to a minimum by using very short leads, preferably unshielded. An alligator clip should be used at the test end so that connection can be made without adding the capacity of the user’s hands.
3-7. MAXIMUM INPUT VOLTAGE. Do not apply more than 600 volts de to the INPUT terminals. To do so ex­ceeds the voltage rating of the input capacitor.
3-8. If an applied voltage momentarily exceeds the selected full-scale voltmeter sensitivity, a few seconds may be required for circuit recovery, but no damage
will result. 3-9. INPUT VOLTAGE WAVEFORM. The voltmeter
is calibrated to indicate the root-mean-square value of a sine wave; however, meter pointer deflection is proportional to the average value of whatever waveform is applied to the input.
is not a sine wave, the reading will be in error by an amount dependent upon the amount and phase of the harmonics present, as shown in figure 3-2 below. When harmonic distortion is less than about 10%, the error which results is negligible.
INPUT VOLTAGE
CHARACTERISTICS
Fundamental = 100 Fundamental +10%
2nd harmonic
Fundamental +20%
2nd harmonic
Fundamental +50%
2nd harmonic
Fundamental +10%
3rd harmonic
Fundamental +20%
3rd harmonic
Fundamental +50%
3rd harmonic Note: This chart is universal in application since these errors are inherent in all average-respond­ing type voltage-measuring instruments.
Figure 3-2. Effect of Harmonics on Voltage
Measurements
If the input signal waveform
TRUE
RMS
VALUE
100
100.5 102 112
100.5 102 112
METER
INDICATION
100
100 100-102 100-110
96-104 94-108 90-116
3-1
TM 11-6625-1514-15
Section III
Paragraphs 3-10 to 3-16
Figure 3-3. Test Setup for Avoiding Ground Loop
3-10. Since the voltmeter meter deflection is propor-
tional to the average value of the input waveform, it
is not adversely affected by moderate levels of random
noise. The effect that noise has on the accuracy of the meter reading depends upon the waveform of the noise and upon the signal-to-noise ratio. A square wave has the greatest effect, a sine wave intermediate effect, and
“White” noise has the least effect on the meter reading. 3-11. If the noise signal is a 50% duty cycle square wave
and the signal-to-noise ratio is 10:1 (between peak voltages), the error will be about 1% of the meter
reading. If the noise signal is “white” noise and the signal-to-noise ratio 10:1, the error is negligible.
3-12. LOW-LEVEL MEASUREMENTS AND GROUND CURRENTS.
3-13. When the voltmeter is used to measure signal levels below a few millivolts, ground currents in the meter test leads can cause an error in meter reading. Such currents are created when two or more ground connections are made between the instruments of a test setup and/or between the instruments and the power
line ground. Two ground connections complete an electrical circuit (ground loop) for the voltages which are generated across all instrument chassis by stray fields, particularly the fields of transformers. These ground currents can be minimized by disconnecting the ground lead in the power cord from either the voltmeter or the signal source being measured, at the power outlet as shown in figure 3-3, and by making
sure that in the test setup no other ground loop is formed that can cause a ground current to flow in the voltmeter test leads. Although the resultant voltage developed across a test lead is in the order of micro­volt, it is enough to cause noticeable errors in measurements of a few millivolts. The presence of
ground currents can sometimes be determined by
simply changing the grounds for the instruments in the
3-2
setup and watching for a change in meter reading. If changing the ground system causes a change in meter
reading, ground currents are present.
3-14. MEASUREMENT OF VOLTAGE.
3-15. The meter has two VOLTS scales, 0 to 1 and 0 to 3. When the RANGE switch is set to .001, .01,
.1, 1, 10, or 100 VOLTS, read the 0 to 1 scale. When the RANGE switch is set to .003, .03, .3, 3, 30, or 300 VOLTS, read the 0 to 3 scale.
The lower (black) signal INPUT and OUT­PUT terminals and the instrument case are connected to the power system ground when the instrument is used with a standard three­terminal (grounding) receptacle. Connect only ground-potential circuits to the black INPUT and OUTPUT terminals.
3-16. Operate the instrument as follows:
a. Connect the voltmeter to the a-c power source. b. Turn the Power switch ON and allow a warmup
period of approximately five minutes.
c. Disconnect any external equipment from the OUT-
PUT terminals.
d. Set the RANGE switch to the VOLTS range which will read the voltage to be measured at mid-scale or above. If in doubt, select a higher VOLTS range.
e. Connect the voltage to be measured to the INPUT terminals.
00102-2
AVOID A SHORT CIRCUIT ACROSS THE POW­ER LINE! To measure power line voltage, first
connect only the upper (red) INPUT terminal to
each side of the power line, in turn, leaving it connected to the side that causes meter indi­cation. Then connect the lower (black) INPUT
terminal (grounded internally) to the other side of the line. If this procedure is not followed, the power line may be short-circuited through the grounded INPUT terminal of the voltmeter.
TM 11-6625-1514-15
Section III
Paragraphs 3-17 to 3-21
f. Note the meter indication on the DECIBELS scale
(-12 to +2 db). The signal level is the algebraic sum of the meter indication and the db value indicated by the RANGE selector. Study the following examples:
Example 1 If the indication on the DECIBELS scale is +2 and the
RANGE switch is in the +20 DB position, the level is +22 dbm.
Example 2 If the indication on the DECIBELS scale is +1.5 and the RANGE switch is in the -40 DB position, the level is
-38.5 dbm.
f. Read the meter indication on the appropriate VOLTS
scale, in accordance with the full-scale value indicated on the RANGE switch. Evaluate the reading in terms of the full-scale value indicated on the RANGE switch. Study the following examples:
Example 1
When the RANGE switch is in the .1 VOLTS range, read
the 0 to 1 VOLTS scaIe. If the meter indicates .64 on
that scale, the voltage being measured is:
Example 2 When the WGE switch is in the 30 VOLTS range, read
the 0 to 3 VOLTS scale. If the meter indicates 1.6 on
that scale, the voltage being measured is:
3-17. MEASUREMENT OF DECIBELS.
3-18. The DECIBELS meter scale is provided for measuring dbm directly across 600 ohms and for measuring db ratio for comparison purposes when each measurement is made across the same circuit
impedance. To measure signal level directly in dbm
(0 dbm equals 1 milliwatt into 600 ohms) proceed as follows:
a. Connect the voltmeter to the a-c power source. b. Turn the Power switch ON and allow a warmup
period of approximately five minutes.
c. Disconnect any external equipment from the OUT-
PUT terminals.
d. Set the RANGE switch to the DB range which will give an upscale reading of the signal to be measured. If in doubt, select a higher-level scale.
e. Connect the voltage to be measured to the INPUT terminals.
00102-2
3-19. To measure db across impedances other than 600 ohms, follow the above procedure and evaluate the
results as follows:
NOTE
Since the measurement is made across other than 600 ohms, the level obtained in step f is in db, but not in dbm.
a. To obtain the difference in db between measure­ments made across equal impedances, algebraically subtract the levels being compared.
b. To obtain the reading of a single measurement
in dbm, note the impedance across which the measure­ment is made and refer to the Impedance Correction Graph, described in paragraph 3-20.
c. To obtain the difference in dbm between measure-
ments made across different impedances, convert each measurement to dbm using the Impedance Correction Graph described in paragraph 3-20. Then algebraically subtract the dbm levels being compared.
3-20. IMPEDANCE CORRECTION GRAPH.
3-21. As the voltmeter DECIBELS scale is calibrated to indicate dbm for measurements made across 600-ohm circuits, a correction factor must be used when meas­urements are made across circuit impedances other than 600. ohms, if absolute dbm levels are desired. The correction factor is not necessary in measuring relative db levels (not dbm) across the same impedance, but it is required for comparison of db levels measured across different impedances.
in figure 3-4 gives the correction factor for conversion of the meter reading to dbm when the impedance of the circuit under test is known. To use the graph, read the conversion factor corresponding to the test circuit impedance and add it to the meter reading determined by the method of paragraph 3-17. Observe the algebraic sign of the correction factor in making the algebraic addition. Use the following examples:
Example 1 If the measurement is made across 90 ohms, the indication on the DECIBELS scale is +2, and the RANGE switch is at the +30 DB position, the level in dbm is obtained as follows:
The Impedance Correction Graph
3-3
TM 11-6625-1514-15
Section III
Paragraphs 3-22 to 3-25
+ 2 (meter indication) +30
(RANGE switch position) +32 (sum) +8
(correction factor from the Impedance
+40 dbm
Example 2 For the same conditions as given above, except that the measurement is made across an impedance of 60,000 ohms, the level in dbm is obtained as follows:
+ 2 (meter indication) +30
(RANGE switch position) +32 (sum)
-20
(Correction factor from the Impedance
+12 dbm
3-22. USE OF VOLTMETER AMPLlFIER.
3-23. The amplifier in the voltmeter may be used for amplifying weak signals. With full-scale meter deflec­tion, the open-circuit output of the amplifier is approxi-
mately 0.15 volt rms regardless of the RANGE switch position. The impedance looking into the OUTPUT terminals is approximately 50 ohms. The frequency
Correction Graph)
Correction Graph)
response and calibration of the voltmeter may be affected by the impedance of a load applied to the OUTPUT terminals. To check the effect of the applied load: observe the meter reading obtained with no load connected to the OUTPUT terminals and then note any shift of reading when the external circuit is connected to the OUTPUT terminals. If the shift is negligible, the measurement is not being affected appreciably by the load. Whenever the input signal is changed, i.e., a different frequency or band of frequencies is applied, repeat the quick check described above.
3-24. Maximum gain from the amplifier is obtainable only on the lowest (.001 volts) range, since output level is the same for all bands. This is due to the 10-db amplification loss per step inserted by the RANGE switch as it is turned clockwise. Amplification may also be obtained on the .003, .01, .03, and 1 volt ranges.
3-25. When the voltmeter is used as an amplifier, select a range which gives a meter deflection near
full scale. Off-scale signals more than twice the value of the position of the RANGE switch will cause severe distortion.
3-4
00102-2
TM 11-6625-1514-15
Section III
Figure 3-4. Impedance Correction Graph
3-5
4-0
Section IV
TM 11-6625-1514-15
00102-2
Figure 4-1.
Loading...
+ 51 hidden pages