HP 8447D, 2944A, 8447E, 2944A, 8447F, 2944A, 8447F H64, 3113A Operation And Service Manual

Printed in USA March 2000
Notice
Hewlett-Packard to Agilent Technologies Transition
This documentation supports a product that previously shipped under the Hewlett­Packard company brand name. The brand name has now been changed to Agilent Technologies. The two products are functionally identical, only our name has changed. The document still includes references to Hewlett-Packard products, some of which have been transitioned to Agilent Technologies.
Advanced Test Equipment Rentals
www.atecorp.com 800-404-ATEC (2832)
®
E
s
t
a
b
l
i
s
h
e
d
1
9
8
1
Contacting Agilent
By internet, phone, or fax, get assistance with all your test and measurement needs.
Table 1-1 Contacting Agilent
Online assistance: www.agilent.com/find/assist United States
(tel) 1 800 452 4844
New Zealand
(tel) 0 800 738 378 (fax) (+64) 4 495 8950
Latin America
(tel) (305) 269 7500 (fax) (305) 269 7599
Japan
(tel) (+81) 426 56 7832 (fax) (+81) 426 56 7840
Canada
(tel) 1 877 894 4414 (fax) (905) 282-6495
Australia
(tel) 1 800 629 485 (fax) (+61) 39210 5947
Asia Call Center Numbers
Country Phone Number Fax Number
Singapore 1-800-375-8100 (65) 836-0252 Malaysia 1-800-828-848 1-800-801664 Philippines (632) 8426802
1-800-16510170 (PLDT Subscriber Only)
Thailand (088) 226-008 (outside Bangkok)
(662) 661-3999 (within Bangkok) Hong Kong 800-930-871 (852) 2506 9233 Taiwan 0800-047-866 (886) 2 25456723 People’s Republic
of China
800-810-0189 (preferred)
10800-650-0021
(632) 8426809 1-800-16510288 (PLDT Subscriber Only)
(66) 1-661-3714
10800-650-0121
Europe
(tel) (+31) 20 547 2323 (fax) (+31) 20 547 2390
India 1-600-11-2929 000-800-650-1101
2 Chapter1
Operation and Service Manual
HP 8447D/E/F and HP 8447F H64
Amplifiers
Serial Numbers: HP 8447D, 2944A
HP 8447E, 2944A HP 8447F, 2944A
HP 8447F H64, 3113A
Copyright Hewlett-Packard Company 1970-1996. All Rights Reserved. Reproduction, adaptation, or translation
without prior written permission is prohibited, except as allowed under the copyright laws.
1400 Fountaingrove Parkway, Santa Rosa, CA 95403-1799, USA
HP Part No. 08447-90066
Supersedes 08447-90033, 08447-90063, 08447-90059
Printed in USA
Print Date: July 1996
Legal
Legal
The information contained in this document is subject to change without notice. Hewlett-Packard makes no warranty of any kind with regard to this material, in­cluding but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, per­formance, or use of this material.
Safety Info
WARNING: Warning denotes a hazard. It calls attention to a procedure which, if not correctly
performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.
WARNING: This is a Safety Class I product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the instrument, is likely to make the instrument dangerous. Intentional interruption is prohibited.
CAUTION: Caution denotes a hazard. It calls attention to a procedure that, if not correctly performed or
adhered to, would result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
CAUTION: This product is designed for use in Installation Category II and Pollution Degree 2.
This instrument has been designed and tested in accordance with IEC Publication 348, Safety Requirements for Electronic Measuring Apparatus, and has been sup­plied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to main­tain the instrument in a safe condition.
”CSA” The CSA mark is a registered trademark of the Canadian Standards Asso­ciation.
Notice for Germany: Noise Declaration LpA %< 70 dB am Arbeitsplatz (operator position) normaler Betrieb (normal position) nach DIN 45635 T. 19 (per ISO 7779)
2 DRAFT
Operation and Service Manual
DRAFT 3
Operation and Service Manual
General Information
General Information
The HP 8447D, 8447E and 8447F are general purpose, wideband amplifiers. Each instrument consists of a power supply and one or two thin film, hybrid, integrated circuit amplifiers. The thin film amplifiers are hermetically sealed and feature low noise, low distortion, flat frequency response, and long term stability and reliabil­ity.
The HP 8447D Amplifier is a pre-amplifier that provides 25 dB of gain to signals from 100 kHz to 1.3 GHz. It can be used to increase the sensitivity of any lab, shop or field device operating within its frequency range.
The HP 8447E Amplifier is a power amplifier that provides 22 dB of gain to sig­nals from 100 kHz to 1.3 GHz. It can be used to increase the output power of sig­nal generators, swept sources and similar devices operating within its frequency range.
The HP 8447F Amplifier contains a preamplifier and a power amplifier. They are the same thin film amplifiers used in the HP 8447D and HP 8447E. The input and output ports of both amplifiers are available on the front panel. They can be used separately or cascaded to provide 47 dB of gain.
The HP 8447F Option H64 Amplifier is a general purpose wide band amplifier with Type-N connectors. It contains two preamplifiers. One provides 26 dB mini­mum gain to signals from 9 kHz to 50 MHz. The other provides 25 dB minimum gain to signals from 100 kHz to 1300 MHz.
Table 1 HP 8447D and HP 8447E Specifications
Specifications
Frequency range 0.1 to 1300 MHz 0.1 to 1300 MHz
Typical 3 dB bandwidth (Std, Opt 001 and Opt 011) .075 to 1700 MHz .075 to 1400 MHz
Mean Gain (20° to 30°C) 25 dB minimum 22 dB ± 2.0dB
Gain flatness across frequency range ± 1.5 dB ± 1.5 dB
Noise Figure
Output power for 1 dB gain compression
<8.5 dB <11 dB (typical)
>+7 dBm (typical) ≥+12.5 dBm
HP 8447D
Preamp
(0.1 to 1000 MHz)
HP 8447E
Power Amp
4 DRAFT
Table 1 HP 8447D and HP 8447E Specifications
Operation and Service Manual
General Information
Specifications
Harmonic distortion
Typical output for <60 dB harmonic distortion 30 dBm 20 dBm
VSWR, 1 to 1300 MHz
30 dB for 0 dBm output (typi-
cal)
<2.0:1 Input
HP 8447D
Preamp
30 dB for +8 dBm output
−−
HP 8447E
Power Amp
<2.2:1 Output
VSWR, 1 to 1400 MHz
−− <2.2:1 Input <2.5:1 Output
Impedance 50 50
Reverse isolation
Maximum DC input voltage ±10 V ±10 V
Typical risetime 400 ps 420 ps
Typical group delay
Net weight 3 lbs 8 oz (1.59 kg) 3 lbs 8 oz (1.59 kg)
Option 001 3 lbs 14 oz (1.75 kg)
Option 010 3 lbs 1 oz (1.84 kg) 3 lbs 10 oz (1.64 kg)
>40 dB >40 dB
1.1 ns
a
1.5 ns
−−
b
Option 011 4 lbs 10 oz (1.64 kg)
Dimensions 8 1/2 in (216 mm) by 5-1/2 in (130 mm) by 3-3/8 in (85.8 mm)
Power Requirements 115 or 230 Vac ±10%, 48 to 440 Hz, 15 Watts, 27 VAmax
Environmental
Operation: 0 Storage:
°C to +55°C
40°C to +75°C
−−
a. Variation over any 50 MHz band from 0.1-1300 MHz is typically <0.15 ns. b. Variation over any 50 MHz band from 0.1-1300 MHz is typically <0.25 ns.
DRAFT 5
Operation and Service Manual
General Information
Table 2 HP 8447F Specifications
Specifications
Frequency range 0.1 to 1300 MHz 0.1 to 1300 MHz 0.1 to 1300 MHz 9 kHz to 50 MHz
Typical 3 dB bandwidth (Std, Opt 001 and Opt 011)
Mean Gain (20° to 30°C) 25 dB minimum 22 dB ± 2.0dB 25 dB minimum 26 dB minimum
Gain flatness across fre­quency range
Noise Figure
Output power for 1 dB gain compression
Harmonic distortion
Typical output for <60 dB harmonic distortion
VSWR, 9 kHz to 50 MHz
VSWR, 1 to 1300 MHz
Preamp Power Amp
.075 to 1700 MHz .075 to 1400 MHz .075 to 1700 MHz 6 kHz to 400 MHz
± 1.5 dB ± 1.5 dB ± 1.5 dB ± 2 dB
<8.5 dB <11 dB (typical) <8.5 dB −−
>+7 dBm (typical) ≥+12.5 dBm
30 dB for 0 dBm
output (typical)
30 dBm 20 dBm 30 dBm 2 dBm
−− −− −− <2.0:1
<2.0:1 Input <2.2:1 Output
HP 8447F
(0.1 to 1000 MHz)
30 dB for +8 dBm
output
−− <2.0:1 Input
Preamp Low Freq. Preamp
>+7 dBm (typical) >+15 dBm
30 dB for 0 dBm
output (typical)
<2.2:1 Output
HP 8447F H64
(0.1 to 1000 MHz)
35 dBc for +10
dBm output
VSWR, 1 to 1400 MHz
−− <2.2:1 Input
−−
<2.5:1 Output
Impedance 50 50 50 50
Reverse isolation
Maximum DC input volt­age
Typical risetime 400 ps 420 ps 400 ps
Typical group delay
Net weight 3 lbs 14 oz (1.75 kg) 3 lbs 14 oz (1.75 kg)
Option 001
>40 dB >40 dB >40 dB −−
±10 V ±10 V ±10 V ±10 V
−−
1.1 ns
a
1.5 ns
b
1.1 ns
a
−−
−− −−
6 DRAFT
Table 2 HP 8447F Specifications
Operation and Service Manual
General Information
Specifications
Option 010 4 lbs 1 oz (1.84 kg) 4 lbs 1 oz (1.84 kg)
Dimensions 8 1/2 in (216 mm) by 5-1/2 in (130 mm) by 3-3/8 in (85.8 mm)
Power Requirements 115 or 230 Vac ±10%, 48 to 440 Hz, 15 Watts, 27 VAmax
Environmental
Preamp Power Amp
Operation: 0 Storage:
40°C to +75°C
HP 8447F
Preamp Low Freq. Preamp
°C to +55°C
a. Variation over any 50 MHz band from 0.1-1300 MHz is typically <0.15 ns. b. Variation over any 50 MHz band from 0.1-1300 MHz is typically <0.25 ns.
Table 3 HP 8447F H64 Characteristics
Characteristics
Noise figure
9 kHz to 20 kHz 20 kHz to 100 kHz 100 kHz to 50 MHz
Typical mean gain (20° to 30°C)
Preamp Low Freq. Preamp
−−
−−
HP 8447F H64
<12.0 dB <9.0 dB <7.0 dB
28 dB
HP 8447F H64
DRAFT 7
Operation and Service Manual
General Information
Figure 1 Typical Preamplifier Performance Curves
8 DRAFT
Operation and Service Manual
General Information
Figure 2 Typical Power Amplifier Performance Curves
DRAFT 9
Operation and Service Manual
General Information
Options
The HP 8447D is a single preamplifier with BNC connectors. However, to pro­vide flexibility, three options are offered:
a Option 001 dual preamplifier, BNC connectors.
b Option 010 single preamplifier, Type-N connectors.
c Option 011 dual preamplifier, Type-N connectors.
The HP 8447E is a power amplifier with BNC connectors. One option is offered:
Option 010, Type-N connectors.
The HP 8447F is a preamplifier and a power amplifier with BNC connectors. One option is offered:
Option 010, Type-N connectors.
Option H64, Two preamplifiers with Type-N connectors. This option is not offered in conjunction with any other options.
NOTE: Dual amplifier option 010 is furnished with a rigid coaxial cable (W6), with
Type-N connectors, that can be used to cascade the amplifiers with minimum loss. The option H64 dual amplifier is not supplied with this cable.
Instruments Covered by this Manual
This manual fully documents the HP 8447D, HP 8447E and the HP 8447F. Any references in the manual apply to all three amplifiers unless otherwise noted.
Each amplifier has a ten digit serial number on the serial plate on the rear panel. The first five digits of the serial number are a prefix. The contents of this manual apply directly to instruments with the same serial number prefix as listed on the inside title page.
Revisions required to adapt the manual to other serial number prefixes are con­tained in a yellow “Manual Changes” insert supplied with the manual. For infor­mation about serial number prefixes not listed on the title page or in the insert, contact your nearest Hewlett-Packard office.
10 DRAFT
Operation and Service Manual
General Information
DRAFT 11
Operation and Service Manual
Receiving and Operating
Receiving and Operating
Initial Inspection
Mechanical Check
If damage to the shipping carton is evident, ask the carrier’s agent to be present when the instrument is unpacked. Inspect the instrument for mechanical damage. Also check the cushioning material for signs of severe stress.
Performance Check
The electrical performance of an amplifier should be verified upon receipt. Perfor­mance checks suitable for incoming inspections are given in the performance test section.
Claims for Damage
If the instrument is mechanically damaged in transit, notify the carrier and the nearest Hewlett-Packard field office immediately. A list of field offices is con­tained in the back of this manual. Retain the shipping carton and padding material for the carrier’s inspection. The field office will arrange for replacement or repair of your instrument without delay for claim settlements against the carrier. Before shipment, this instrument was inspected and found free of mechanical and electri­cal defects. If there is any deficiency, or if electrical performance is not within specifications, notify your nearest Hewlett-Packard sales and service office.
Preparation For Use
Power Requirements
The amplifier operates from 115 or 230 volt ac line voltages, at any line frequency between 48 and 440 Hz. A slide switch on the rear panel is set to the correct posi­tion for the line voltage available.
Power Cable
To protect operating personnel, the National Electrical Manufacturer’s Associa­tion (NEMA) recommends that the instrument panel and cabinet be grounded. All Hewlett-Packard instruments are equipped with a three-conductor power cable which, when plugged into the appropriate receptacle, grounds the instrument. The
12 DRAFT
Operation and Service Manual
Receiving and Operating
offset pin on the power cable’s three-prong connector is the ground wire. The power cord and power input connector meet the specifications established by the International Electrotechnical Commission (IEC).
CAUTION: Always use the three-prong ac power cord supplied with this instrument. Failure to ensure
adequate earth grounding by not using this cord may cause instrument damage.
Mating Connectors
Mating connectors used with the amplifier should be 50 type BNC male. If the amplifier has the optional Type-N connectors, use Type-N male connectors that are compatible with US MIL-C-39012.
Operating Environment
The operating range of the amplifier is from 0°C to +55°C. The amplifier can be stored in a temperature range of
40°C to +75°C.
Bench Use
The amplifier is equipped with plastic feet and tilt stand in place, ready for use as a bench instrument.
WARNING: If this product is not used as specified, the protection provided by the equipment could
be impaired. This product must be used in a normal condition only (in which all means for protection are intact).
Rack Mounting
The amplifier may be rack mounted by using an adapter frame. The adapter frame is a rack frame that accepts several combinations of sub-modular units. For addi­tional information, address inquiries to your nearest Hewlett-Packard sales and service office.
Storage and Shipment
Packaging
The following paragraphs contain a general guide to repackaging of the instru­ment for shipment. Refer to them for information about using the original packag­ing, or using other packaging.
DRAFT 13
Operation and Service Manual
Receiving and Operating
If the instrument is to be shipped to Hewlett-Packard for service or repairs, attach a tag to the instrument identifying the owner and indicating the service or repair required. Include the model number and full serial number of the instrument.
Original Packaging
If the original container is to be used, proceed as follows:
a Place instrument in original container. If it is not available, a suitable container can be
purchased from your nearest Hewlett-Packard sales and service office.
b Be sure the container is well sealed with strong tape or a metal band.
Other Packing Material
If the original container is not used, proceed as follows:
a Wrap the instrument in heavy paper or anti-static plastic before placing in inner anti-
static container.
b Place packing material around all sides of the instrument and protect panel face with
cardboard strips.
c Place instrument in a heavy carton and seal with strong tape. A double-wall carton
made of 200 pound test material is adequate.
d Mark shipping container; “Delicate Instrument. Fragile.”
Operating Instructions
Connect the line power cable to the three-prong rear panel socket and proceed as follows:
a Select the line power to be used (115V or 230V) with the rear panel SELECTOR
switch.
b Connect the line power cable to line power and press the LINE switch.
CAUTION: The input to the HP 8447D/E/F and the HP 8447F Option H64 amplifiers is very
susceptible to damage from electrostatic discharge. Before connecting any coaxial cable to the input jack of the amplifier, make certain that static electricity buildup in the cable is completely discharged.
Do not apply power levels to the amplifier that are higher than the following:
HP 8447D: 0 dBm HP 8447E: 10 dBm HP 8447F preamp: 0 dBm HP 8447F power amp: +10 dBm
14 DRAFT
Operation and Service Manual
Receiving and Operating
c Attach INPUT and OUTPUT cables.
CAUTION Power out of the HP 8447E and power out of the HP 8447F power amp can
exceed 100 mW. This may be enough power to damage sensitive circuits connected to the OUTPUT.
Operator Maintenance
Operator maintenance is limited to replacement of the front panel LINE switch light, the A1 power supply fuse, and the rear panel fuse. Clean the cabinet using a damp cloth only.
Fuses
To replace the rear panel fuse (F1), remove the rear panel fuse knob and replace the fuse (see rear panel for value of F1).
To replace the A1 power supply (A1F1) fuse, turn the instrument on its top. Lift the tilt stand and remove the bottom cover. Replace the fuse on the A1 Power Supply board (see parts list for the value of A1F1).
WARNING: For continued protection against fire hazard replace line fuse only with same type and
rating ([0.5A/250V]). The use of other fuses or material is prohibited.
Lamp Replacement
To replace the front panel line switch lamp (DS1), proceed as follows:
a Disconnect cord from rear panel receptacle.
b Pull the white cover portion of this switch from the instrument and then remove the
lamp from inside the cover.
c Replace old lamp with a new lamp (see parts list for part number of DS1).
d Place white cover into switch receptacle.
e Align tab on white cover with socket and push in.
WARNING: No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent
electrical shock, do not remove covers.
DRAFT 15
Operation and Service Manual
Performance Tests Procedures
Performance Tests Procedures
Test equipment and accessories required to perform maintenance are listed in Table 4. Equipment other than recommended models can be used provided the minimum specifications are satisfied.
The performance tests assume the use of a standard amplifier with BNC connec­tors. If an amplifier with Type-N connectors is to be tested, suitable adapters will have to be added to the equipment lists.
The performance tests and adjustment procedure cover all three amplifiers. Where required, test steps list equipment settings, and other additional information, for each amplifier. If a test step lists only one setting or specification, it is valid for all three amplifiers. Disregard references to any amplifier not being tested.
Table 4 Recommended Test Equipment and Accessories
Instrument Type Minimum Specifications Suggested Model
Generator/Sweeper
Generator/Sweeper
Sweep Oscillator
Frequency Range: 0.1 to 110 MHz Flatness: Output Level: Frequency Accuracy: 1% Compatible with network analyzer
Frequency Range: 9 kHz to 50MHz Flatness: Output Level: Frequency Accuracy: 1% Compatible with network analyzer
Frequency Range: 0.1 to 1.3 GHz Flatness: 0.7 dB over full range Output Level:
Frequency Accuracy: Compatible with network analyzer
±0.25 dB over full range
+10 dBm minimum
±100 kHz
±0.25 dB over full range
+10 dBm minimum
±100 kHz
+10 dBm minimum
±20 MHz
HP 8601A P,T
HP 3335A P,T
HP 8620A/8621A P,T
a
Use
(for testing opt H64)
16 DRAFT
Loading...
+ 41 hidden pages