The following general safety precautions must be observed during operation of this instrument. Failure tocomply with these precautions or with specific warnings elsewhere in this manual violates safety standardsof design, manufacture, and intended use of the instrument. Hewlett-Packard Company assumes no liabilityfor the customer's failure to comply with these requirements.
GROUND THE INSTRUMENT.
minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The
To
instrument is equipped with a three-conductor ac power cable. The power cable must either be plugged intoan approved three-contact outlet or used with a three-contact to two-contact adapter with the grounding wire
(green) firmly connected to an electrical ground (safety ground) at the power outlet. The power jack andmating plug of the power cable meet International Electrotechnical Commission (IEC) safety standards.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE.
Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electricalinstrument in such an environment constitutes a definite safety hazard.
DO NOT REMOVE INSTRUMENT COVERS.
Operating personnel must not remove instrument covers. Component replacement and internal adjustments
must be made by qualified maintenance personnel.Service instructions for this instrument are provided in aseparate Operating and Service Manual.
DANGEROUS PROCEDURE WARNINGS.
Warnings such as the example below, precede potentially dangerous procedures throughout this manual!
Instructions contained in the warnings must be followed.
WARNING
Dangerous voltages, capable of causing death, are present in this instrument.
Use extreme caution when handling, installing or operating.
EliminatingFlicker Storinga Trace Display Brightness Control
MeasurementTechniques
Introduction
VoltageMeasurements Absolute Voltage Measurements Peak-to-peak Voltage Measurements Average Voltage Measurements Amplitude Comparison Measurements Method1 Method2 CommonMode Rej ection
This Operators Guide will acquaint you with the Model1744A features, capabilities, accessories, power require-ments, and controls. To aid in operating the oscillo-scope, initial turn-on instructions, calibration proce-dures, and a performance check are provided. Detailedexplanations in the Applications Section show how youcan use varied capabilities of the 1744A to best ad-vantage in a variety of electrical measurements. Speci-fiications and general characteristics for the 1744A arelisted in tables 1 and 2. Service information is availablein a separate service manual.
DESCRIPTION.
The Hewlett-Packard Model 1744A is a dual-channel,
100-MHz, delayed-sweep variable persistence storageoscilloscope designed for single-shot transients and lowduty cycle waveform analysis in the bench or fieldenvironment. The writing speed specification of 1800cm/µs permits full screen capture of transients at themaximum slew rate of the vertical amplifier system. The
dual-channel dc to 100 MHz vertical deflection systemhas 12 calibrated deflection factors from 5 mV/div to 20V/div. A maximum sensitivity of 1 mV/div to 30 MHz isprovided an both channels by means of a 5X verticalmagnification feature. Selectable input impedance of 50ohms or 1 megohm permits selection of that impedancewhich best meets measurement requirements. The hori-
zontal deflection system has calibrated sweep ratesfrom 2 s/div to 0.05 µs/div and delayed sweep rates from20 ms/div to 0.05 gs/div. A 10X magnifier expands allsweeps by a factor of 10 and extends the fastest sweep to5 ns/div. In alternate or chop modes, a trigger-viewcontrol will display three signals: the channel A signal,the channel B signal, and the trigger signal. This allowsyou to correlate time between the trigger signal and thechannel A and channel B signals. In trigger-viewoperation, center screen represents the trigger thresholdpoint, allowing you to see the triggering level location.With the A VS B control, an X-Y mode of operation ispossible. The channel A input (Y-axis) is plotted versusthe channel B input (X-axis).
A trigger-view control provides capability for observing
the channel A signal, channel B signal, and an external
Operators Guide
Model 1744A
Table 1. Specifications
VERTICAL AMPLIFIERS (2)
Bandwidth and Rise Time at all deflection factorsfrom 0°C to +55°C.
BANDWIDTH: DC-Coupled:
input modes.
AC-Coupled:
divider probes.
BANDWIDTH LIMIT:
20 MHz.
RISE TIME:
of a 5 div input step.
DEFLECTION FACTORRanges:
in 1, 2, 5 sequence, accurate within 3%.
Vernier:
extends maximum deflection factor to at least 50V/div. UNCAL light indicates when vernier is not inCAL position.
POLARITY:
pushbutton.
INPUT RC (selectable)AC or DC: 50 Ohm: MAXIMUM INPUTAC or DC:
(p-p ac at 1 kHz or less).
3 dB down from 6 div reference signal.
dc to 100 MHz in both 50f1 and 1 Mn
approx 10 Hz to 100 MHz; 1 Hz with 10:1
li
mits upper bandwidth to approx
<3.5 ns, measured from 10% to 90% points
5 mV/div to 20 V/div (12 calibrated positions)
continuously variable between all ranges,
channel B may be inverted, front panel
1 MCi ±2% shunted by approx 20 pF.
50n ±3%.
250 V (dc + peak ac at 1 kHz or less) or 500 V
50 Ohms: 5 A+B OPERATION
Amplifler: bandwidth and deflection factors are un-
changed; channel B may be inverted for A-B opera-tion.
Differential (A-B) Common Mode:
dB from dc to 20 MHz. Common mode signal ampli-tude equivalent to 6 divisions with one vernieradjusted for optimum rejection.
V rms.
CMRR is at least 20
VERTICAL MAGNIFICATION (X5)
BANDWIDTH: DC-Coupled:
AC-Coupled:
RISE TIME:
5 div input step).
DEFLECTION FACTOR:
and 10 mV/div deflection factor settings by a factorof 5 for a maximum sensitivity of 1 mV an channelsA and B.
3 dB down from 6 div reference signal.dc to approx 30 MHz.approx 10 Hz to 30 MHz.
-<12 ns (measured from 10% to 90% points of
increases sensitivity of the 5
TRIGGER SOURCE
Selectable from channel A, channel B, composite, orline frequency.
CHANNEL A: all display modes triggered by channel A
signal.
Mode! 1744A
Operators Guide
Table 1. Specifications (Cont'd)
CHANNEL B:
all display modes triggered by channel B
signal.
COMPOSITE:
all display modes triggered by displayedsignal except in Chop. In Chop mode, trigger signalis derived from channel A.
LINE FREQUENCY:
trigger signal is derived from
power line frequency.
TRIGGER VIEW
Displays internal or external trigger signal. In Alter-nate or Chop mode, channel A, channel B, and thetrigger signals are displayed. In Channel A or Bmode, Trigger View overrides that channel. Internaltrigger signal amplitude approximates vertical sig-
nal amplitude. External trigger signal deflectionfactor is approx 100 mV/div or 1 V/div in EXT =10.Triggering point is approx center screen. With id-entically timed signals to a vertical input and theexternal trigger input, trigger signal delay is <3.5 ns.
MAIN AND DELAYED TIME BASES
RANGES
Main: 50 ns/div to 2 s/div (24 ranges) in 1, 2, 5 sequence.
Delayed:
sequence.
50 ns/div to 20 ms/div (18 ranges) in 1, 2 5
Accuracy:
Sweep Time/Div
50 ns to 20 ms
*Add 1% for 50 ms to 2 s ranges.
1
MAIN SWEEP VERNIER:
(over center 8 div)
*Accuracy X1
±3% ±4% 0°C to +15°C
±2%±3%
Temp Range
1
X10
±3% +15°C to +35°C ±4% +35°C to +55°C
continuously variable be-tween all ranges, extends slowest sweep to at least 5s/div. UNCAL light indicates when vernier is not inCAL position.
MAGNIFIER (X10):
expands all sweeps by a factor of
10, extends fastest sweep to 5 ns/div.
CALIBRATED SWEEP DELAY
DELAY TIME RANGE:
settings of 100 ns to 2 s (minimum delay 150 ns).
DIFFERENTIAL TIME MEASUREMENT ACCURACY:
Main Time Base Setting
100 ns/div to 20 ms/div
50 ms/div to 2 s/div
*Add 1% from 0°C to +15°C and +35°C to +55°C.
0.5 to 10 X Main Time/Div
*Accuracy
I
(
+
15°C
to
+35°C)
±(0.5% +0.1% of full scale)
±(1% +0.1% of füll scale)
1
3
Operators Guide
DELAY JITTER:
delay in each step from +15°C to +35°C; <0.005% (1part in 20 000) from 0°C to +15°C and +35°C to
+55°C.
<0.002% (1 part in 50000) of maximum
TRIGGERING
INTERNAL:
division or more vertical deflection, increasing to 1division of vertical deflection at 100 MHz in alldisplay modes (required signal level is increased by 2when in Chop mode and by 5 when X5 verticalmagnifier is used). Triggering on Line frequency isalso selectable.
EXTERNAL:
more increasing to 100 mV p-p at 100 MHz (requiredsignal level is increased by 2 when in Chop mode).
EXTERNAL INPUT RC:
prox 20 pF.
MAXIMUM EXTERNAL INPUT:
kHz or less) or 500 V (p-p ac at 1 kHz or less).
LEVEL and SLOPEInternal:
of the displayed waveform.
External:
either slope of the trigger signal, +10 V to -10 V individe by 10 mode (=10).
COUPLING:
dc to 25 MHz on signals causing 0.3
dc to 50 MHz on signals of 50 mV p-p or
approx 1 Mfl shunted by ap-
250 V (dc + peak ac at 1
at any point on the positive or negative slope
continuously variable from +1 V to -1 V on
AC, DC, LF REJ, or HF REJ.
Table 1. Specifications (Cont'd)
AC:
attenuates signals below approx 20 Hz.
LF Reject (Main Sweep):
approx 4 kHz.
HF Reject (Main Sweep):
approx 4 kHz.
CALIBRATED MIXED TIME BASE
Dual time base in which the main time base drivesthe First portion of sweep and the delayed time basecompletes the sweep at the faster delayed sweep.Also operates in single sweep mode. Accuracy, add2% to main time base accuracy.
A
vs
B OPERATION
BANDWIDTHChannel A (Y-AXIS): Channel B (X-AXIS): DEFLECTION FACTOR:
calibrated positions) in 1, 2, 5 sequence.
PHASE DIFFERENCE BETWEEN CHANNELS:
to 100 kHz.
CATHODE-RAY TUBE AND CONTROLS
Z-AXIS INPUT (INTENSITY MODULATIONS):
X50 ns width pulse blanks trace of any intensity
usable to 10 MHz for normal intensity. Input R,1 kf1 ±10%. Maximum input ±20 V (dc ± peak ac).
Model 1744A
attenuates signals below
attenuates signals above
same as channel A.dc to 5 MHz.
5 mV/div to 20 V/div (12
<3°, dc
+4 V,
Model 1744A
PERSISTENCEVariable: STORAGE WRITING SPEED:
6 x 8 div (with viewing hood).
STORAGE TIMEDisplay Mode: Store Mode: Wait Time: ERASE TIME:
approx 100 ms to 1 minute.
>,1800 cm/µs over center
at least 10 s at 22°C.
at least 30 s at 22°C.
at least 60 s at 22°C.
approx 300 ms.
GENERAL
REAR PANEL OUTPUTS:
V to >+2.5 V capable of supplying approx 5 mA.
main and delayed gates, +0.8
Table 1. Specifications (Cont'd)
AMPLITUDE CALIBRATOR (0°C to +55°C)
Output Voltage
Rise Time
Frequency
POWER:
VA max.
WEIGHT: net, 13.8 kg (30.4 lb); shipping, 16.6 kg (36.6
lb).
OPERATING ENVIRONMENT
Temperature:
Humidity: Altitude: Vibration:
0.254 mm (0.010 in.) excursion, 10 to 55 Hz.
100,120, 220, 240 Vac, ±10%, 48 to 440 Hz;100
to 4600 m (15 000 ft).
Operators Guide
1 V
p-p into <-1 MSl
0.1 V p-p into 50f111
0°C to +55°C.
to 95% relative humidity at +40°C.
vibrated in three planes for 15 min. each with
0.1 µs
approx 1.4 kHz
±1%
Operators Guide
VERTICAL DEFLECTION
VERTICAL DISPLAY MODES
Channel A; channel B; channels A and B displayedalternately on successive sweeps (ALT); channels Aand B displayed by switching between channels atan approximate 250 kHz rate with blanking duringswitching (CHOP); channel A plus channel B (al-gebraic addition); and trigger view.
DELAY LINE:
view leading edge of input pulse without advancedtrigger.
INPUT COUPLING:
ground. Ground position disconnects input connect-or and grounds amplifier input.
input signals are delayed sufficiently to
selectable AC or DC, 50(1 (dc), or
HORIZONTAL DISPLAY MODES
Main, main intensified, mixed, delayed, mag X10,and A vs B.
TRIGGERING
MAIN SWEEP
Normal: sweep is triggered by internal or external
signal.
Automatic:
input signal. Above 45 Hz, triggering is same asnormal. For stable triggering at approx 45 Hz andbelow, use Normal trig
bright baseline displayed in absence of
gering.
Table 2. General Characteristics
Single: automatically switches triggering to Normal
and the sweep occurs once with same triggering asNormal, reset pushbutton arms sweep and lightsindicator. Single sweep is also initiated with Erasepushbutton, sweep is armed after the erase cycle.
DELAYED SWEEP (SWEEP AFTER DELAY)Auto:
delayed sweep automatically starts at end of
delay.
Trig:
delayed sweep is armed and triggerable at end of
delay period.
TRIGGER HOLDOFF (Main Sweep):
holdoff time in all ranges.
CATHODE-RAY TUBE AND CONTROLS
TYPE: Hewlett-Packard, 12.7 cm (5 in.) rectangular
CRT, post accelerator, approx 7.5 kV acceleratingpotential, aluminized P31 phosphor.
GRATICULE: 8
nonparallax graticule, 0.2 subdivision marking onmajor horizontal and vertical axes, with markingsfor rise time measurements. Graticule illuminationis achieved with Persistence control set to minimum.
BEAM FINDER:
of setting of horizontal and vertical controls.
OPERATING MODES:
and auto erase.
REAR PANEL CONTROLS:
and display brightness.
DIMENSIONS:
Model 1744A
increases sweep
x 10 div (1 div = 0.72 cm) internal,
returns trace to CRT screen regardless
write, store, display, auto store,
astigmatism, trace align,
see outline
drawing.
6
Model 1744A
NOTES:
1.
2.
Table 2. General Characteristics (Contil)
DIMENSIONS ARE FOR GENERAL
INFORMATION ONLY. IF DIMENSIONSARE REQUIRED FOR BUILDINGSPECIAL ENCLOSURES, CONTACTYOUR HP FIELD ENGINEER.
DIMENSIONS ARE IN MILLIMETRES AND(I
NCHES).
Operators Guide
33513 3 1
1847 a1
B71T
1
a
=1
REAR
O
Operators Guide
Model 1744A
trigger signal on the same display when operating inALT or CHOP modes. Automatic storage is provided forcapturing single-shot data that occurs at random times.Automatic erase allows continuously repeated erasecycles with continuously variable viewing time betweenerase cycles. The CRT screen has 8 x 10 major divisions(0.72 cm/div) on an internal graticule.
ACCESSORIES FURNISHED.
A blue light filter, front-panel cover, power cord, vinylaccessory storage pouch, two 10:1 divider probes, a
metal mesh contrast filter, a B-scan jumper filter,
and a collapsible viewing hood are provided with the1744A.
ACCESSORIES AND OPTIONS AVAILABLE.
Several divider probes are available with various voltagedivision ratios and cord lengths. Refer to 1744A Operat-ing and Service Manual for model or part numbers.
Option 001
power cord.
Option 090
supplied. You may specify other probes that are moresuitable for your requirements.
Option 091
the two divider probes normally supplied.
Option
the two divider probes normally supplied.
has a fixed power cord in lieu of detachable
deletes the two 10:1 divider probes normally
supplies two Model 10042A probes in lieu of
092
supplies two Model 10040A probes in lieu of
Option
the two divider probes normally supplied.
Option 580
dicating compliance with CSA Bulletin 556B. (Cana-dian Standards Association)
Option
uals instead of one.A camera adapter permits use of an oscilloscope camera
with the 1744A. Various testmobiles are available which
accept the 1744A and provide convenient, mobile standsfor the oscilloscope. Refer to 1744A Operating andService Manual for model numbers.
096
supplies two Mode! 10006D probes in lieu of
supplies instrument with CSA Label in-
910
furnishes two Operating and Service Man-
PREPARATION FOR USE.
WARNING
Read the Safety Summary at the front of this
guide before installing or operating the in-
strument.
POWER CORD.
The power cord required depends on the ac input voltageand the country in which the instrument is to be used.Figure 1 illustrates standard power receptacle (walloutlet) configurations. The HP part number shownabove each receptacle drawing specifies the power cordequipped with the appropriate mating plug for that
8
Mode! 1744A
Operators Guide
HP POWER CABLE PART NUMBERS
8120-1692
OPTION 902
•___
8120-1703
OPTION 900
INPUT POWER RECEPTACALE TYPES
Figure 1. Power Receptacles
receptacle. If the appropriate power cord is not includedwith your instrument, notify the nearest HP Sales/Service Office and a replacement cord will be provided.
8120-2296
OPTION 906
8120-1369
OPTION 901
8120-1521
STANDARD
POWER REQUIREMENTS.
The oscilloscope requires a power source of 100,120, 220,or 240 volts ac ±10%, single phase 48 to 440 Hz that candeliver at least 100 VA.
LAUTION
Instrument damage may result if the linevoltage selection switch is not correctly set forthe proper input power source.
The instrument is normally set at the factory for 120 Voperation. To operate the instrument from any other acpower source, proceed as follows:
a.
Verify that 1744A power cable is not connected
to a power source.
b.
Stand instrument an rear-panel legs and useblade-type screwdriver to position power selector switch-es through opening in bottom cover (figure 2 showsswitches set for 120 V operation).
c.
For 220 V -240 V inputs, replace fuse Fl with 0.5A slow-blow fuse supplied with instrument (HP Part No.2110-0202).
d.
Connect 1744A input power Gable to input power
source.
9
Operators Guide
t
LL
Figure 2. Line Voltage Selection Switch Settings
^
100
1
r
100V I
—240V
120]
1
t
—arm
II
1
I
1741A-036-ß1-77
CONTROLS AND CONNECTORS.
Front- and rear-panel photographs (see figure 32) are
located at the rear of this guide on a fold-out page for
easy reference while you are reading any section. Con-
trol and connector descriptions have index numbersthat are keyed to the panel photographs. The followingparagraphs provide detailed descriptions of control andconnector functions. Refer to the Applications Sectionfor information an using the Model 1744A.
Mode! 1744A
CRT CONTROLS
0 LINE OFF/ON. Switch turne power on and off.
Indicator lights when power is ON.
Q
BEAM FIND. Restricts the display to the view-ing area regardless of control settings. Permitsthe operator to determine what action is ne-
button to select either AUTO STORE or AUTOERASE mode of operation. Respective lampslight to indicate which mode is selected. TheAUTO ERASE lamp is green.AUTO ERASE. In auto erase, the 1744A operatesin a repetitive single-shot mode, even when acontinuous signal is available. This mode is
STORE.Two-functionpush-
10
1
Model 1744A
Operators Guide
also helpful during setup for capture of single-
shot events by making it easier to obtain timum focus and intensity for a particular sig-nal. In addition, if you are viewing more thanone trace, such as two or three channels, the
1744A will wait for the required number ofsweeps to be displayed before automaticallyerasing the display. Operating at high drivelevels in variable persistence and storage modesmay cause the storage mesh to retain residuali
mages. These residual images may appear as acluttered display when BRIGHTNESS 0 is ator near maximum. Residual traces are conven-
iently removed by operating the 1744A in theauto erase mode for a few minutes with IN-TENSITY Q fully ccw.
NOTE
Pressing ERASE Q terminates theAUTO ERASE cycle allowing the erator to vary the cycle time at slowerrates.
AUTO STORE.
Used to set up the 1744A for
applications requiring capture of random sin-
gle-shot events. AUTO STORE is selected by
Op-
Op-
pressing SINGLE m, AUTO STORE Q, andWRITE 0 to prevent self-triggering. Whenthe random event occurs, the 1744A automati-cally triggers, sweeps and switches from WRITEmode to STORE mode. To capture another
event, press write 0 and RE SET m or ERASE0 to arm trigger circuit. AUTO STORE allowsthe operator to capture high-speed, single-shot,random transients. After triggering occursthe 1744A switches to store. At slower writingrates, the time the operator can view the imagebefore it fades will increase. Therefore, it isto the operator's advantage to use the lowestBRIGHTNESS setting in viewing and storingtraces.
Q PERSISTENCE/VIEW TIME.
Two-function po-tentiometer to vary persistence or viewing timeof signal, depending an operating mode selected.
PERSISTENCE.
Variable persistence allows youto adjust the trace retention for optimum dis-play when viewing low repetition rate, fast risetime signals. By adjusting persistence to matchthe sweep rate, you can cause the trace torefresh and provide a continuous display ofhard-to-view signals such as low duty-cyclepulse trains.
11
Operators Guide
Model 1744A
VIEW TIME.
VIEW TIME control establishes the time thatthe trace is retained on the display beforeanother erase cycle is initiated.
Q ERASE.
cycle to remove stored traces from the CRT
storage mesh. Inoperative when STORE/DIS-PLAY indicators are on to prevent accidentalerasure of the stored signal. In auto store mode,
ERASE may be pressed to arm the trigger cir-cuit.
Q BRIGHTNESS.
i
mages. Store time is inversely proportional toBRIGHTNESS setting, BRIGHTNESS is alsoused in the WRITE p mode to vary the writing
rate.
Q WRITE.
signals on the display in storage and variable
persistence operation. WRITE lamp lights whenWRITE is selected.
STORE/DISPLAY.
when the 1744A is operating in the WRITE 0mode stores the trace being written, and lightsthe STORE lamp. Pressing the pushbutton
In AUTO ERASE Q mode, the
Pushbutton which initiates the erase
Adjusts brightness of stored
Conditions the 1744A to write applied
Engaging this pushbutton
again causes both STORE and DISPLAY lampsto light and displays the stored trace on screen.Brightness of the display is increased by rotat-ing the BRIGHTNESS the trace has been examined, press the push-button again to enter STORE mode to achievelongest store time.
VERTICAL AMPLIFIER CONTROLS
m CAL 1 V.
square-wave signal with a frequency of ap-proximately 1.4 kHz (100 mV p-p when ter-minated in 50 ohms).
® 1".
ground.
CHAN A (B) VOLTS/DIV.
®
deflection factor in a 1, 2, 5 sequence from 0.005V/div to 20 V/div, accurate within 3% withvernier m in CAL detent.
m
Vernler.
flection factor between calibrated ranges. Ver-nier range is at least 2.5:1.
Provides a 1 V peak-to-peak (±1%)
Provides convenient front-panel chassis
Provides continuous control of the de-
control cw. After
43
Selects the vertical
12
Mode! 1744A
®
UNCAL.
or CHAN B) is out of CAL detent. Indicates that
VOLTS/DIV setting is uncalibrated.
m
Coupling.
pedance for the vertical amplifiers. In the ACposition, the dc component of the input signalis blocked. The lower 3-dB limit is approximately10 Hz.
GND.
amplifier, and the amplifier input is grounded.
DC.
to the vertical amplifier. The input impedance isapproximately 1 megohm shunted by 20 pF.
50fl.
input impedance is 5051. Pull the lever forwardand down to select this position. Do not applymore than 5 V rms to the input connector.
m
INPUT.
nals to the channel A (Y) or channel B (X)amplifier. Impedance and coupling are select-able by m. Do not apply more than 250 V (dc +peak ac at 1 kHz or less) or more than 500 V (p-pac at 1 kHz or less).
m
POSN $ .
display.
Lights when Vernier (either CHAN A
Selects the input coupling and im-
The input Signal is disconnected from the
All elements of the input signal are passed
The input signal is dc coupled, and the
BNC connector to apply external sig-
Controls the vertical position of the
Operators Guide
m
ALT.
Channel A and B signals are displayed
alternately an consecutive sweeps.m ®
Channel
Channel
A +
A. Displays the channel A input signal.
B.
Displays the channel B input signal.
Pressing both channel A m and channel
B.
B ® displays the algebraic sum of the channel
A and channel B input signals. If the channel Bdisplay is inverted (press CH B INVT m),an A minus B display results.
CHOP.
Channel A and B signals are displayedsimultaneously by switching between channelsat 250-kHz rate.
®
TRIGGER
signal as the trigger signal when INT
A. Selects a sample of the channel A
/EXTm
is INT.
m
TRIGGER
B.
When in INT m, a sample of thechannel B signal is selected as the triggersignal.
CO M P.
When display mode is set for channel A,
channel B, A + B, or ALT, and both ® and
m are pressed, the sweep is triggered by the
displayed signal. When display is set to CHOP,sweep is triggered by channel A signal only.
13
Operators Guide
Model 1744A
TRIG VIEW.
ger signal with approximately the same sensi-tivity indicated on the corresponding channelVOLTS/DIV ® control. A selected externaltrigger signal is displayed with a fixed sensi-tivity of approximately 100 mV/div with INT/
EXT m set to EXT (1 V/div if EXT = 10® engaged). TRIGGER LEVEL ® positionsthe trigger signal vertically about the center
horizontal graticule line. Center screen indi-cates the trigger threshold level with respect tothe trigger signal. In ALT m or CHOPmodes, three signals appear on the same dis-play: channel A, the selected trigger signal (atcenter screen), and channel B, with no need forerasing between each display. This is extremelyuseful in applications such as digital circuitswhere it is necessary to use external triggersources to maintain proper timing relation-ships and to know the time relationship of thetrigger signal to the displayed events. TRIG
VIEW is also helpful in establishing discretetrigger levels.
m MAG
five times, and increases the maximum sensi-tivity to 1 mV/div. The bandwidth is decreasedto 30 MHz.
Displays the selected internal trig-
X5.
Magnifies the vertical presentation
BW LIMIT.
A and channel B to approximately 20 MHz.
CH B 1 NVT.
m
signal. In A + B m & ® mode, pressing CH BINVT m results in an A minus B display.
MAIN TIME BASE CONTROLS
m & m POSITION 4P.
ments position the trace horizontally.
®
MAIN TRIGGER LEVEL.
level on the input trigger signal where the mainsweep is triggered. With external trigger sig-nals, the trigger level is continuously variablefrom +1.0 V to -1.0 V on either slope of theinput trigger signal; +10 V to -10 V in EXT = 10
m mode. With internal trigger signals, the
trigger level control selects any point on thedisplayed vertical waveform.
®
POS/NEG.
select either the positive or negative slope of thetrigger signal as the starting point for the sweep.
®
LF REJ.
signals below approximately 4 kHz. This is
Attenuates internal or external trigger
Reduces the bandwidth of channel
Inverts the polarity of the channel B
Coarse m and FINE m adjust-
Selects the voltage
Two-position pushbutton used to
14
Model 1744A
Operators Guide
useful to condition high-frequency signals forbest synchronization by eliminating unwantedlow-frequency signals such as power line inter-ference.
m
HF REJ.
signals above approximately 4 kHz. This is
useful to condition low-frequency signals forbest synchronization by eliminating unwantedhigh-frequency signals such as RF.
LINE.
m removes all EXT m input or INT m dis-
played signals from the trigger circuit andapplies a power line frequency signal for trig-
gering.
® AC/DC.
applied to the trigger circuit. The DC positionmust be selected for signals below approxi-mately 20 Hz.
m
INT/EXT.
vertical signal chosen by the TRIGGER source
® or m while EXT selects the signal at the
EXT TRIGGER m input for application to themain trigger circuit. Internal signals from dc to25 MHz displaying 0.3-div amplitude or moreare sufficient for stable triggering, increasing
Attenuates internal or external trigger
Selecting both LF REJ ® and HF REJ
Selects ac or dc coupling of the signal
INT selects a sample of the internal
to 1 div of amplitude at 100 MHz. Externallyapplied signals 50 mV p-p from dc to 50 MHz,increasing to 100 mV p-p at 100 MHz are suf-ficient for stable triggering.
®
EXT
+
10.
Attenuates EXT TRIGGER m input
signal by a factor of 10.
m
EXT TRIGGER.
ger input. Input impedance is approximately
one megohm shunted by approximately 20 pF.
Do not apply more than 250 V (dc + peak ac at 1
kHz or less) or 500 V (p-p ac at 1 kHz or less).
m
AUTO/NORM.
out). A free-running sweep provides a brightdisplay in the absence of a trigger signal. Atrigger signal input (internal or external) of 45
Hz or more overrides AUTO operation andsweep triggering is the same as in the NORM
mode.
RESET.
trigger circuit in the single-sweep mode. AfterRESET, the sweep can be triggered by an in-ternal trigger signal or by rotating the TRIG-GER LEVEL control ® through zero. RESETlamp lights to indicate circuit is armed for next
Momentary pushbutton that arms the
BNC connector for external trig-
AUTO sweep mode (pushbutton
15
Operators Guide
Model 1744A
trigger signal. Pressing ERASE 0 will also
reset the sweep.
SINGLE.
triggering as in NORM. After each sweep, thetrigger circuit raust be manually RESET m.SINGLE must also be pressed in conjunctionwith WRITE 0 and AUTO STORE 0 to con-dition the 1744A for AUTO STORE operation.
m
MAG X10.
times, and expands the fastest sweep time to5 ns/div, maintaining a sweep accuracy within3% at room temperature.
m
UNCAL.
is out of the CAL detent, and indicates that thesweep is not calibrated.
m
MAIN TIME/DIV.
main sweep rate, which is indicated by thenumbers displayed in the knob skirt opening.Sweep accuracy is within 2% (unmagnified) atroom temperatures.
SWEEP VERNIER.
ment of main sweep TIME/DIV between cali-brated positions, extending the slowest sweepto 5 s/div.
Sweep occurs once with the same
Magnifies the horizontal display 10
Lights when SWEEP VERNIER m
The inner knob controls the
Provides continuous adjust-
m
TRIGGER HOLDOFF.
tween sweeps and aid triggering an complexdisplays such as digital words.
CO MAIN.
play. Sweep rate and triggering are selected bythe main-sweep controls m - m.
m
A VS B.
channel A input (Y-axis) plotted versus channelB input (X-axis). Vertical positioning is ad-
justed by channel A POSN m, and horizontal
positioning is adjusted by POSITION m andFINE m.
m
DLY'D. Selects delayed sweep for horizontaldisplay.
m
MIXED.
horizontal display. The first portion of the sweepis at the main sweep rate, and the secondportion of the sweep (starting point chosen byDELAY ® is at the delayed-sweep rate. SeeMixed Sweep Display under Obtaining BasicDisplays for more information.
Selects main sweep for horizontal dis-
Selects an X-Y mode of operation with
DELAYED TIME BASE CONTROLS
Selects main and delayed sweeps for the
Increases the time be-
16
Model 1744A
®
DLY'D TIME/DIV.
The outer rotating ring selectsthe delayed sweep rate, which is indicated bythe marker an the ring. Sweep accuracy is thesame as with MAIN TIME/DIV m. An inter-
lock is incorporated so the delayed sweep is
always faster than the main sweep. Whenrotated from the OFF position in the MAIN mode, a portion of the main sweep is intensifiedin brightness indicating the length and delayedposition of the delayed sweep with respect to themain sweep.
DELAY.
The DELAY control provides a vari-able delay time from 0.5 to 10 X the MAINTIME/DIV m settings of 100 ns to 2 s. Seethe Application Section for more information.
®
DELAYED TRIGGER LEVEL.
level on the input trigger signal where thedelayed sweep is triggered. With external trig-ger Signals, the trigger level is continuously
variable from +1.0V to -1.0V on either slope of
the input trigger signal; +10 V to -10 V in
EXT = 10 m mode. With internal trigger sig-nals, the trigger level selects any point on thedisplayed vertical waveform.
Selects the voltage
CO
m
SWEEP AFTER DELAY AUTO/TRIG.
method of starting the delayed sweep when inmain intensified, delayed, or mixed mode
eration. In AUTO (pushbutton released), de-layed sweep starts immediately after the delayinterval, which is the product of the DELAY® dial reading (div) and the MAIN TIME/DIV m reading. In TRIG (pushbutton pressed),the delayed trigger circuit is armed after thedelay interval and delayed sweep must be trig-gered by either an internal or external trigger
signal. See pulse jitter in the Application Sec-tion for more information.
®
POS/NEG.
m
AC/DC.
®
INT/EXT.
m
EXT = 10.
m
EXT TRIGGER.
Refer to POS/NEG ®.
Refer to AC/DC ®.
Refer to INT/EXT m.
Refer to EXT = 10 ®.
Refer to EXT TRIGGER m.
Operators Guide
Selects the
Op-
17
Operators Guide
Model 1744A
REAR PANEL CONTROLS
DISPLAY BRIGHTNESS.
NORMAL position, the CRT floodgun is pulsed
on and off. In MAX position, the CRT floodgun
is always on.
m
Z-AXIS INPUT.
put of a signal to modulate CRT beam intensity.
TRACE ALIGN.
to the horizontal graticule lines.
ASTIGMATISM.
cd spot. (Interacts with FOCUS 0.)
m
Line Input.
m
MAIN GATE OUTPUT.
output of approximately +2.5 V coincident withthe main gate.
m
DLY'D GATE OUTPUT.
output of approximately +2.5 V coincident withthe delayed gate.
m
LINE FUSE.
A BNC connector allowing In-
Aligns horizontal trace parallel
Controls roundness of display-
Power cord connector.
AC power-input fuse.
When switch is in
Provides a rectangular
Provides a rectangular
TURN-ON PROCEDURE
WARNING
Before you turn on the oscilloscope:
1)
Read safety summary at front of thisguide;
2)
Observe all safety precautions;
3)
Be sure power selector switches are setproperly for power source you are using toavoid instrument damage;
4)
Familiarize yourself with controls andconnectors by reading Controls and Con-nectors section and referring to panelillustrations at back of this guide.
To turn on the Model 1744A, follow these steps to avoid
CRT damage and achieve a useful display.
1. Turn all control knobs to 12 o'clock positionexcept: PERSISTENCE/VIEW TIME 0, BEAM IN-TENSITY Q, and BRIGHTNESS 0 fully ccw, allVERNIERS m, m to CAL (detent) position, TRIGGERHOLDOFF D to MIN, and MAIN TIME/DIV fully cw.
2. All pushbuttons should be disengaged exceptDISPLAY A m, TRIGGER A ®, and MAIN D.
18
Model 1744A
3.
Engage LINE 0 switch; LINE indicator lamp
should light.
4.
Press WRITE 0 pushbutton; WRITE indicator
lamp should light.
5.
Press ERASE Q pushbutton.
6.
After CRT warmup, increase BEAM INTEN-SITY 0 to comfortable viewing level; adjust FOCUS 0for sharpest trace.
NOTE
The 1744A is equipped with a high perfor-formance CRT using an expansion storagetechnique. Because of this technique and
under certain operating conditions a second-
ary trace will appear an screen. This secondarytrace is called "write through" and is a normalcharacteristic of the expansion storage tech-nique. To minimize write through, reducebeam intensity.
CHECKING INSTRUMENT PERFORMANCE.
LAUTION
The CRT is capable of very high writing speed.
Operators Guide
When operating at high intensity setting,write gun electrons may start charging thestorage mesh faster than floodgun electronscan discharge it. If this happens, a slow-moving trace or spot may become extremelybright, outlined by a dark area. Should youobserve this condition, immediately turn
BEAM INTENSITY fully ccw and pressAUTO ERASE 0. Allow the instrument tooperate in this mode for 5 minutes to removethe residual trace. Otherwise, the CRT may bepermanently damaged.
You can make a few checks and simple adjustments toensure that the Model 1744A is operating properly. If theoscilloscope is moved from one electromagnetic environ-ment to another, the trace alignment control may needadjustment to align the horizontal trace with the grati-cule. Astigmatism and focus controls may needadjustment to obtain the sharpest display. Probe com-pensation may be required because the total inputresistance and capacitance varies slightly from one
oscilloscope to another.
TRACE ALIGNMENT.
1.
Obtain trace as described in initial turn-on
procedure.
19
Operators Guide
Model 1744A
With vertical POSN m, align trace with center
2.
graticule line.
Using nonmetallic alignment tool, adjust
3.TRACE ALIGN m (on rear panel) until trace alignswith horizontal graticule line.
ASTIGMATISM AND FOCUS.
Select A VS B m and set BEAM INTENSITY
1.
to low level.
2.
Position spot near center of CRT with POSN m
and POSITION m controls.
Adjust FOCUS 0 and ASTIGMATISM m (on
3.rear panel) for smallest, round spot.
PROBE COMPENSATION.
1.
Perform initial turn-on procedure.
Connect divider probe cable to channel A IN-
2.
PUT
S
connector.
I
3.
Connect probe tip to CAL 1 V m output.
4.
Set channel A input coupling m to DC position.
Set main TIME/DIV m for horizontal display
5.
of at least two full square waves.
Set channel A VOLTS/DIV ®for square-wave
6.
display with two or three divisions of vertical deflection.
7.
Adjust TRIGGER LEVEL ®for stable display.
B. Adjust divider probe compensation for correct
display (see figure 3).
OVERSHOOT
CORRECT
UNDERSHOOT
1740A-003
Figure 3. Probe Compensation
20
Model 1744A
Operators Guide
VERTICAL ACCURACY CHECK.
1.
Accomplish initial turn-on procedure.
2.
Connect CAL 1 V m output to channel A
INPUT D using BNC to banana plug adapter and test
lead with alligator
3.
Set channel A input coupling m to DC position.
4.
position.
should be five major divisions (±4%).
SWEEP TIME ACCURACY.
channel A INPUT m connector.
set one marker on second left vertical graticule line.
Set channel A VOLTS/DIV ® to 0.2 V/DIV
5.
Set main TIME/DIV m to 0.2 mSEC position.
6.
Square-wave amplitude of displayed waveform
1.
Accomplish initial turn-on procedure.
2.
Apply an accurate 0.5 µSEC time mark signal to
3.
Set main TIME/DIV m to 0.5 µSEC position.
4.
Using horizontal POSITION m, ®i controls,
Clip.
5.
Time markers should line up approximately
with each vertical graticule line across CRT.
6.
Marker on tenth graticule (second from right
side of CRT) of CRT should be within 0.16 division.
TRIGGER SELECTION TABLE.
Table 3 will help you in determining whether a triggermode is unusable, usable, good, or the best mode forvarious signal conditions.
OBTAINING BASIC DISPLAYS.
To familiarize yourself with the 1744A, you may wish touse the following procedures for producing some basicdisplays. Start by referring to the turn-on procedure and
control settings. In addition,
® Channel A VOLTS/DIV mChannel A Vernier mChannel A Coupling m MAIN TIME/DIV ® DELAY
NORMAL SWEEP DISPLAY.
1.
Connect 10:1 divider probe (provided with
1744A) between channel A INPUT m and CAL 1 V m
output.
........................
set
controls as follows:
............
.................
.................
............
0.05
CAL
DC
0.5 mSECfully ccw
21
Operators Guide
Model 1744A
Table 3. Display and Trigger Selection Table
TRIGGER SELECTION
SIGNAL CONDITIONS
I. Single Signals
Applied to ChannelA or B
II. Time Related Signals
Applied to ChannelsA & B
III. Nontime Related
Signals Applied toChannels A & B
1
Assume time related signal applied.
2
Time relation displayed.
No time relation displayed.
4
If COMP is selected in CHOP, switching
overrides and selects A.
5
Signal is only displayed on one channel.
DISPLAY MODE
A or B
ALTS or CHOP
ALT
CHOP
A+B (A-B)
ALT
A
OK
S
OK
or
or
1i
OKZ
OK
NG
Triggers on algebraic sum
6
signals.
OK
Usable trigger mode.
Good trigger mode.
OK
Best trigger mode.
OK
Unusable trigger mode.
NG
B
OK
OK
OKZ
OK2
OK
NG
COMP
OK
NG
NG3
NG
OK6
OK
or difference of
EXT
OK'
OK'
OK2
OK2
OK
NG
oha
Model 1744A
2.
Connect divider probe grounding strap to
ground post ®.
3.
Adjust main TRIGGER LEVEL ® for stable
display.
4.
Adjust channel A POSN m to align base of
square wave on center graticule line.
5.
Observe square-wave display of five to ninepositive going Pulses with amplitude of two divisions(see figure 4).
MAGNIFIED SWEEP DISPLAY.
1.
Accomplish normal sweep display procedure.
2.
Using horizontal POSITION m, place wave-
form to be magnified on center graticule line.
3.
Engage MAG X10 m pushbutton.
4.
Adjust horizontal POSITION mi for precise
placement of magnified display (see figure 5).
Operators Guide
Figure 4. Normal Display
1740A-004
1740A-005
Figure 5. Magnified Display
23
Operators Guide
DELAYED SWEEP DISPLAY.
1.
Accomplish normal sweep display procedure.
2.
Set delayed TIME/DIV ®to 50 µSEC position.
Observe portion of square wave that is intensified.
3.
Adjust BEAM INTENSITY 0 for comfortable
viewing level.
4.
Ensure SWEEP AFTER DELAY m is in AUTO
position.
I
NTENSIFIED
Model 1744A
5.
Adjust DELAY ® clockwise until intensified
portion of trace is over area tobe investigated (see figure
6).
6.
Engage DLY'D m pushbutton. Note intensifiedportion of trace is now displayed across entire CRT (seefigure 7).
NOTE
Other pulses in the pulse train may be ob-served by varying the position of the DELAYcontrol ®.
24
Figure 6. Normal Display With Intensified Area
1740A-007
Figure 7. Delayed Sweep Display
Model 1744A
MIXED SWEEP DISPLAY.
1.
Accomplish normal sweep display procedure.
2.
Set delayed TIME/DIV ®to 50 µSEC position.
3.
Adjust BEAM INTENSITY Q for comfortable
viewing level.
4.
Adjust DELAY ® clockwise until intensified
portion of waveform is in second half of CRT (see figure
8).
Operators Guide
1740A-009
Figure 9. Mixed Sweep Display
5. Engage MIXEDm pushbutton. Verify that firstportion of display is at main TIME/DIV m sweep rateand second portion is at delayed TIME/DIV ® sweep
rate (see figure 9).
X-Y DISPLAY.
1.
Engage A VS B m pushbutton.
I
NTENSI FI ED
1740A-008
Figure B. Normal Display With Intensified Area
2. Keep the beam intensity low with no deflection.
3.
Apply vertical (Y-axis) signal to channel A
INPUT m.
25
Operators Guide
4.
Apply horizontal (X-axis) signal to channel B
INPUT W.
NOTE
Model 1744A
up to 15 seconds, automatically erases, resets the sweep,then repeats the cycle. The display time is adjustable
from 600 ms to 15 s by the VIEW TIME control. To store
a waveform, engage the STORE pushbutton.
Channel A POSN m will adjust vertical Posi-tion of the waveform. Horizontal POSITIONm will adjust the horizontal position.
5.
Adjust channel A and channel B VOLTS/DIV
® controls as required.
6.
If display is not visible, engage BEAM FIND Qto locate display. Make necessary control adjustmentsto center display an CRT.
STORAGE OPERA-
VARIABLE PERSISTENCE
-
TION.
The 1744A variable persistence and storage capabilitiesmay be used through three different modes: AUTOERASE, AUTO STORE, and continuous scan. (Seetable 4 for summary of operation.)
AUTO ERASE.
operation automatically sets the persistance to maximum,insuring maximum writing speed. The 1744A waits for atrigger signal, sweeps once, displays the waveform for
26
Selecting the AUTO ERASE mode of
The AUTO ERASE mode is useful for applications
where very fast writing speeds are required and theoperator needs an updated display every few seconds.
Since the screen is erased automatically prior to each
display cycle, the waveform and its background are
presented in a high quality display. The short update
cycle (600 ms) is very helpful when immediate feedback
is desired after making adjustments to either the circuit
under test, or to the 1744A controls, such as INTEN-
SITY, focus, position, etc.
AUTO STORE MODE.
operation automatically selects maximum persistenceand the NORMAL trigger mode of operation. This modepermits the maximum "wait time" for a transient andautomatically places the 1744A into the STORE modeafter the occurrence of the transient. This optimizes thestore time of the waveform.
When using the AUTO STORE mode, the BRIGHT-
NESS control should be set to minimum (fully counter-clockwise) to ensure maximum "wait time". To view astored signal, select DISPLAY mode and rotate the
sweep events.Always a cleandisplay. Easy set-up of single sweepevent.
Operators Guide
MAXIMUM
WRITING
SPEED
1800 cm/s
AUTO STORE
CONTINUOUS
SCAN
Automaticallyset to maximum
variable-100ms minimum
10 seconds
after selectionof DISPLAYmode.
Continuous-
down to 10seconds de-pending anPERSISTENCEand BRIGHT-NESS controlsettings.
30 seconds,after up to 60seconds "waittime".
30 secondsminimum afterselection of Storemode.
NORMAL
(
Automati-
cally Selected
NORMAL
OR
AUTO
Maximizes "wait"ti
me" for tran-sient; thenswitches intomaximum storetime mode (30 sec-onds minimum).
For bright displayof low repetition
rate signals. Forcapture of non-periodic glitches.
1800 cm/µs
1800 cm/µs
27
Operators Guide
Model 1744A
BRIGHTNESS control slowly clockwise until the stored
trace is visible. As the BRIGHTNESS control is rotated
clockwise, the operator trades viewing time for bright-ness. A step-by-step procedure to accomplish the AUTOSTORE mode is as follows:
1.
Repeat setup used in storing a trace and turn
BRIGHTNESS 0 and PERSISTENCE Q fully ccw.
2.
Press AUTO ERASE 0 and adjust BEAM
INTENSITY p and FOCUS 0 for optimum display.
3.
Engage CHOP ® pushbutton for multitrace,
single-shot capture.
4.
Press AUTO STORE Q and WRITE Q push-
buttons.
5.
Engage SINGLE m pushbutton. Model 1744A
now set up for AUTO STORE operation.
6.
Turn main TRIGGER LEVEL ® fully ccw
7.
Trigger circuit can be armed either by pressing
RESET m or ERASF 0. Note that RESET lamp comes
on to indicate that circuit is armed. It is good practice toarm by pressing ERASE Q, since erase cycle is initiatedand ensures that previously stored traces are erased.
B. Rotate main TRIGGER LEVEL ® cw untilRESET lamp goes out. Note that STORE lamp is on toshow that trace has been captured and stored.
9. Press STORE m and note that DISPLAY lamp
also lights. Turn BRIGHTNESS Q slowly cw to view
trace.
CONTINUOUS SCAN.
scan mode when the WRITE mode is selected and AUTOERASE/AUTO STORE modes are not selected. In thismode, the operator may select either AUTO or NORMALtriggering and has the full range of persistence viathe PERSISTENCE control. The BRIGHTNESS con-trol may be used to enhance the writing speed. When thePERSISTENCE and BRIGHTNESS controls are fullycounter-clockwise, the 1744A will never require erasureto eliminate blooming and it will operate with minimumpersistence (100 ms). It is possible to store any displaythat occurs during the continuous scan mode by select-ing the STORE mode. Writing speed is maximized byusing maximum persistence along with some clockwiserotation of the BRIGHTNESS control.
The continuous scan mode is useful for a variety of
applications. For slow sweep speeds, the use of per-sistence helps eliminate flicker and provides a contin-uous display. For fast sweep applications the PER-
The 1744A is in the continuous
28
Model 1744A
SISTENCE and BRIGHTNESS controls permit theviewing of glitches which occur randomly. The con-
tinuous scan mode maximizes the time that the 1744A
acquires information, therefore, optimizing the probabilityof capture of events which do not occur an every sweep.To obtain a variable persistence display, procced as follows:
1.
Accomplish initial turn-on procedure.
2.
Apply fast transition, low repetition rate signal
to channel A INPUT m connector.
3.
Note waveform transition give very faint trace
(see figure 10).
4.
Turn PERSISTENCE 0 slowly clockwise. Ob-serve transition portion of waveform gradually inte-grates up to bright, easily observed trace (see figure 11).
Operators Guide
uuausuuuuu
Figure 10. Fast Rise Time, Low Repetition Rate Signal
(Normal)
ELIMINATING FLICKER.
viewed at sweep speeds of approximately 2 ms/div orslower, or whenever the trigger frequency is belowapproximately 30 Hz. This flicker can be eliminated byadjusting persistence so that one trace is just dis-appearing as the next trace is written.
1.
Turn PERSISTENCE control Q fully ccw. Ad-
just MAIN TIME/DIV m to 5 mSEC/DIV and press
MAG X10 m.
A trace will "flicker" when
Figure 11. Fast Rise Time, Low Rep Rate Signal
"Integrated Up" with PERSISTENCE Control
29
Operators Guide
Model 1744A
2. Slowly turn PERSISTENCE Q cw until flicker
ceases.
NOTE
If display "blooms" (entire screen becomes
illuminated), reduce beam intensity and press
ERASE Q until display no longer "blooms."
This may not occur until PERSISTENCE is set
to MAX.
STORING A TRACE.
to press the STORE/DISPLAY STORE indicator will light, signifying that the displayhas been stored. The 1744A will switch to the STOREmode of operation only after all sweeps are completed.For example, in the ALT-TRIG VIEW mode of operation(
with three waveforms displayed), the 1744A will switchto the STORE mode after the third waveform is written.To view the stored waveform(s), press the STORE/DISPLAY switch to the DISPLAY mode of operation.
m
In the STORE mode of operation, the left sideof the CRT will flash every second or so.This flashing is normal for the expansion-storage CRT used and has no adverse effecton the STORE mode of operation.
The procedure for storing a trace is
pushbutton again. The 1744A will then
NOTE
mi
pushbutton. The
DISPLAY BRIGHTNESS CONTROL.
CRT display can be increased by a rear-panel switch(
DISPLAY BRIGHTNESS ment is operated with the DISPLAY BRIGHTNESS switch set to NORMAL. In this mode of operation, theCRT floodgun is pulsed on and off. Dim displays can bebrightened by placing the DISPLAY BRIGHTNESS switch to its MAX position. In this mode of operation,the CRT floodgun is always on. This enhances thedisplay with a corresponding decrease in view time.
B-SCAN.
CRT storage mesh due to too high INTENSITY setting,long exposure to repetitive signals, etc., it is best toremove these images while in B-SCAN mode of Opera-tion. To accomplish B-SCAN, proceed as follows:
Using filter accessory (HP Part No. 01744-furnished with 1744A, connect output from 1744A cali-brator m to channel A input connector m.
3.
Adjust BEAM INTENSITY Q for medium in-
tensity (approximately eleven o'clock position).
4.
Adjust channel A VOLTS/DIV ® control and
channel A Vernier m so that vertical deflection isslightly greater than full screen.
5.
Allow oscilloscope to operate in this mode (B-SCAN) for one hour, checking in display mode forremoval of deeply written image.
6.
After removal of deeply written images, discon-nect filter (HP Part No. 01744-
62101)
from channel A
input connector m.
NOTE
Some deeply written images may require longerthan one hour of B-SCAN operation to be
removed.
MEASUREMENT TECHNIQUES
INTRODUCTION.
the Model 1744A Oscilloscope for various measurement
This section will assist you in using
62101)
Operators Guide
applications. In many cases, illustrations and examples
are provided for clarity. We do not attempt to cover everypossible application; if you have difficulty with anyprocedures or if you have questions about an applicationnot described here, please feel free to call your HPSales/Service representative for assistance. A list of HPSales/Service Offices is included at the rear of thisGuide. When measurements are made by scaling or
interpolating an the CRT graticule, you should use fiveor more major divisions of display between measure-ment points. Most observers will agree, that whenadequate care is used, most measurements can be keptwithin
±1/20
of a major division. This amounts to a
scaling error of ±1% for five divisions of separation.
VOLTAGE MEASUREMENTS.
Voltage measurementscan be made between a point an a waveform and a zero-volt reference (absolute voltage) or between any two
points an a waveform (voltage difference). These voltage
measurements are illustrated in figure
The 1744A vertical deflection system, with positions from 5 mV/div to
20
V div, allows you to make
12.
12
calibrated
voltage measurements which are accurate within 3%.
ABSOLUTE VOLTAGE MEASUREMENTS.
The follow-
ing procedure is used to make absolute voltage mea-
surements with respect to a zero-volt reference.
31
Operators Guide
VDIFFERENCE
Model 1744A
5.
Set the input coupling to DC, and adjust TRIG-
T
GER LEVEL for a stable display. Adjust the main
TIME/DIV control as required.
6.
Measure the distance in divisions between thereference line and the level on the waveform you want tomeasure. An example is shown in figure 13.
v
ABSOLUTE
OV REFERENCE
Figure 12. Absolute Voltage and Voltage Difference
Measurements
1.
Connect input signal to channel A or B INPUTconnector, and select channel A or B DISPLAY andTRIGGER.
2.
Adjust the appropriate VOLTS/DIV control for
six to seven divisions of display. The vernier should be
in the CAL detent position.
3.
Set input coupling to GND and AUTO/NORM
to AUTO.
4.
With the appropriate POSN control, set the
trace on a graticule line to establish a zero-volt re-
ference. Do not move POSN control after the zeroreference is set.
32
1740A-010
7
DIV
REFERENCE
LINE
Figure 13. Absolute Voltage Measurement
7.
You can determine the polarity of the signal bycomparing it to the reference line. If it is above thereference line, the voltage is positive; below the line it isnegative.
B.
Multiply the number of divisions in step 6 by theVOLTS/DIV setting. Include the attenuation factor ifyou are using a probe.
Mode! 1744A
Operators Guide
Voltage = 7 x 0.5
= +3.5 volts
Example: Assume the vertical distance is 7 div, thewaveform is above the reference line, and the VOLTS/DIV setting is 0.5 V/div (see figure 13).The waveform is above the reference line, so the voltageis positive.
PEAK-TO-PEAK VOLTAGE MEASUREMENTS. Oscillo-
scope displays of ac voltages contain errors in amplitudedue to the frequency response of the instrument. Withlow-Signal frequencies, there is less amplitude error.With increasing signal frequencies, the amplitude error
increases. To obtain displays with less than 10% ampli-
tude error, the frequency of the signal being measured
must be less than half of the specified bandwidth of theoscilloscope. A frequency equal to the specified band-width of the oscilloscope will display a voltage ampli-tude on the CRT that is somewhat less than the actualamplitude of the applied signal. The frequency rolloff ofthe instrument must be considered when making volt-age measurements with an oscilloscope. To measure thepeak-to-peak voltage of an input signal, proceed asfollows:
1.
Connect the signal to the channel A or B INPUT
connector.
2.
Set input coupling to AC and adjust main
TRIGGER LEVEL for a stable display.
3.
Adjust vertical POSN, VOLTS/DIV, and mainTIME/DIV for a centered display of at least three cyclesduration and at least three divisions of amplitude. Makesure that the VOLTS/DIV vernier is in the CAL detent.
4.
Using the vertical POSN control, place the
negative peaks of the input signal on a horizontalgraticule line near the bottom of the graticule.
5.
Using the horizontal POSITION control, placeone positive peak of the signal on the center verticalgraticule line.
6.
Count the number of vertical divisions from themost negative to the most positive portions of thewaveform (estimate to nearest tenth of division). (Seefigure 14.)
7.
Multiply the number of divisions noted in step 6by the setting of the VOLTS/DIV switch. If the signal isderived through a divider probe multiply the result of
this step by the attenuation factor of probe. Rememberto consider the amplitude attenuation caused by thefrequency rolloff of the oscilloscope.
33
Operators Guide
Model 1744A
4.
Switch coupling to AC and measure the absolute
voltage to the Same point an the waveform. (See figure
15B.)
PEAK-TO-PEAKVOLTAGE
1740A-012
Figure 14. Peak-to-peak Voltage Measurement
AVERAGE VOLTAGE MEASUREMENTS.
average voltage using the oscilloscope alone, proceed as
follows:
1.
Connect the Signal to the channel A or B INPUT
connector.
2.
Set input coupling to GND and AUTO/NORM
to AUTO. The trace level is zero volt.
3.
Switch input coupling to DC and measure theabsolute voltage at the point of interest an the wave-form. (See figure 15A.)
34
To measure
5.
The difference between the first and second
voltage measurements is the average voltage.
T
1.5V
0
1
REFERENCE
OV
0.5V
-
1
__J
AVERAGE VOLTAGE = 1.5V - 0.5V = 1 V
Figure 15. Average Voltage
11
B
OV REFERENCE
f-1_
1740A-013
Model 1744A
Operators Guide
AMPLITUDE COMPARISON MEASUREMENTS.METHOD NO. 1.
signal to a known (reference) amplitude, it may behelpful to use deflection factors not calibrated on theVOLTS/DIV control. With this method, a particularamplitude can be displayed by an exact number ofdivisions of deflection. This would be desirable whenyou are calibrating an instrument. You can also in-crease the accuracy of your measurements with thecomparison method; the accuracy of your measurementdepends on the reference signal accuracy, not on theoscilloscope accuracy.
1.
Apply the reference voltage to the channel AINPUT connector, and set DISPLAY and TRIGGER tochannel A.
2.
Adjust main TIME/DIV control for severalcycles of display and TRIGGER LEVEL for a stabledisplay.
3.
Set the appropriate VOLTS/DIV, vernier, andPOSN controls for exactly six, seven, or eight divisionsof amplitude. Do not readj ust the vernier after this step.
4.
You should now calculate a scale factor (sf) so
the amplitude of a known signal can be verified or the
amplitude of an unknown signal can be determined.
When you are comparing an unknown
Use the following formula:
Reference Signal Amplitude (volts)
sf —
-
Display Amplitude in Div (Step 3)
x VOLTS/DIV Setting
5.
Disconnect the reference signal and connect thesignal to be measured. Adjust the VOLTS/DIV controlfor enough amplitude to make an accurate measure-ment. Note this display amplitude. Do not readjust thevernier control.
6.
Use the following formula to calculate the am-
plitude of the signal being measured:
Amplitude = VOLTS/DIV Setting (Step 5) x sf (Step 4) x
Display Amplitude in Div (Step 5)
Example: Assume a reference signal amplitude of 40Volts, a VOLTS/DIV setting of 5, and a display ampli-tude of six divisions.
Substituting in the formula from Step 4:
40
sf=
6x5
=1.3
35
Operators Guide
Model 1744A
Now, if the signal to be measured has a display ampli-tude of five divisions with a VOLTS/DIV setting of 2,determine the amplitude from the formula in Step 6.Amplitude = 2 VOLTS/DIV x 1.3 x 5 DIV = 13 volts.
You can also calculate an unknown signal as a per-centage of a known signal.Example: Assume the reference signal has a deflectionof eight divisions. Therefore, each division represents
12.5%. If the unknown signal has a deflection of 6.2divisions, the amplitude of the unknown signal is:UNKNOWN SIGNAL AMPLITUDE = 6.2 DIV x 12.5%/DIV = 77.5% of the reference signal amplitude.
METHOD NO. 2.
Storage operation provides anothersimple method of amplitude comparison between twowaveforms.
1.
Apply the two signals of interest to the channel
A and channel B INPUT connectors.
2.
Select DISPLAY A and TRIGGER A, and setchannel A V/DIV as desired. Use MAIN sweep and setmain TIME/DIV as required.
3.
Use channel A POSN to locate trace in top halfof the display area.
4.
When desired trace is obtained in WRITE mode,
press STORE pushbutton.
5.
Repeat steps 1 through 4 using channel B.Position trace in bottom half of graticule using channelB POSN, with WRITE pushbutton pressed.
6.
Again press STORE.
7.
Pressing STORE a second time and turningBRIGHTNESS cw will display both signals and retainthem for comparison.
COMMON MODE REJECTION.
Frequently, signals ofinterest are offset by undesired dc or low-frequency accomponents that prevent use of vertical ranges sensitiveenough to make good measurements. Often a signalsimilar to the unwanted component can be connected tothe opposite channel, inverted, and added algebraicallyto the signal of interest to cancel the unwanted com-ponent.
True dc components can usually be eliminated byselecting ac input coupling. The ability of an oscillo-scope to cancel ac common-mode signals varies with theamplitude and frequency of the signals. Very highcommon-mode amplitudes may not be completely can-celled. Good common-mode rejection should be achieved
36
Model 1744A
with common-mode signal amplitudes of up to twoscreen diameters (16 CRT divisions). With high-fre-quency, common-mode signals, minor components maybe impossible to eliminate from the display. The lowerthe frequency of the common-mode signal, the better willbe the common-mode rejection in the oscilloscope.
To use the common-mode rejection technique, proceed asfollows:
1.
Apply the desired signal with unwanted com-ponents to the channel A INPUT connector and a signalsimilar to the unwanted components to the channel BINPUT connector.
2.
Set input coupling as required, and select ALT.Adjust the VOLTS/DIV and vernier controls so theunwanted components an the channel A and B signalsare approximately equal in amplitude.
Select TRIGGER A, CH B INVT, and DISPLAY
3.A+B. With either channel A or channel B verniercontrol, adjust for minimum deflection of the commonmode signal.
4. The resultant display will either subtract all theunwanted components in the desired signal or displaythe desired signal larger than the common mode signal.
Operators Guide
CH A
CH B
1740A-014
Figure 16. Channel A with Desired Signal
and Unwanted Components. Channel B
with Only Unwanted Components
A-B
Example: In figures 16 and 17 the common mode
rejection method is illustrated.
1740A-015
Figure 17. Resultant Display
37
Operators Guide
TIME-INTERVAL MEASUREMENTS.
PERIOD MEASUREMENTS.
1.
Apply signal to the channel A or B INPUT
connector, and select A or B DISPLAY and TRIGGER.
2.
Adjust the appropriate VOLTS/DIV control forfive to six divisions of display, if possible, and set themain TIME/DIV control to the fastest sweep speed that
will display at least one cycle within the 8 available
divisions an the CRT.
3.
Use appropriate vertical POSN control and
horizontal POSITION control to center the display.
4.
Measure the horizontal distance in divisions forone cycle. The SWEEP VERNIER should be in CALdetent.
5.
Multiply horizontal distance obtained in step 4times the main TIME/DIV setting. If you are using theMAG X10 switch, divide the product by 10.
Use the following formula:Period = Horizontal Distance for One Cycle in Div (step
4) x MAIN TIME/DIV Setting (step 2) : Magnifier (if used).
Model 1744A
^{
4 D I V
1 CYCLE
1740A-017
Figure 18. Waueform Period Measurement
Example: Assume one cycle of the waveform occurs infour divisions, the MAIN TIME/DIV setting is 0.2mSEC, and the MAG X10 switch is off (see figure 18).Substituting in the formula:Period = 4 div x 0.2 ms/div = 0.8 ms ±2% at room
temperature.
REPETITION RATE OR
FREQUENCY
MEASURE-
MENTS.
The repetition rate or frequency of a waveform is
1.
the reciprocal of the period.
2.
Use the procedure for period measurements to
calculate the period of your signal and take the
reciprocal to determine repetition rate or frequency.
38
Model 1744A
Operators Guide
Example: Using the period from the previous example of
0.8 ms, take the reciprocal to find the repetition rate orfrequency.
Repetition Rate or Frequency =
1 cvcle8x10
TRANSITION TIME MEASUREMENTS.
48
=0.125x10
cycle
4
1
Period
s
=1.25 kHz ±2%
Transition mea-
=
1 cycle
=
0.8ms
surements are normally made between the 10% and 90%points an the leading or trailing edge of the waveform.
The 1744A has 10/90% and 20/80% points conveniently
marked by lines for a five-division reference. The dotsare also spaced identically to the minor division mark-
ings an the major axis to assist you in interpolation.
1.
Apply the pulse to the channel A or B INPUT
connector, and select A or B DISPLAY and TRIGGER.
2.
Adjust the appropriate VOLTS/DIV controland vernier for at least five divisions of amplitude andthe MAIN TIME/DIV control to display enough pulsetop and baseline for measurement. Spread the 10% and
90% points as far apart as possible.
3.
Turn the horizontal POSITION control until the
10% point an the waveform interseets a 10% marking an
a vertical graticule line. The display should be centered
in the viewing ara.
4.
Count the number of divisions until the pulse-rise Grosses the 90% markings. The SWEEP VERNIERshould be in CAL detent.
5.
Multiply the number of divisions obtained inStep 4 times the MAIN TIME/DIV setting. This is therise time (RT). If you use the MAG X10 switch,divide the produce by 10.
Example: Assume the number of divisions between the
10% and 90% points is four and the MAIN TIME/DIVsetting is 2 µSEC (see figure 19).
= 4 x 2 µs=8 µs
RT
If you use the oscilloscope to measure a transition timethat is near to the 1744A rise time specifications (-<3.5ns), error correction may be required. For accurate
results, error correction should be used when the pulsetransition time is four times the oscilloscope rise time orfaster, Use the following formula:
R
(pulse) = JRT2 (observed) - R
T
2
(oscilloscope)
T
Example: Assume the 10% to 90% observed rise time is
7.5 ns and the oscilloscope rise time is 3.5 ns.Substituting in the formula:R
(pulse) = J75
T
2
-3.5
2
=6.6 ns
39
Operators Guide
4 D I V
140A-018
Figure 19. Rise Time Measurement
DELAYED SWEEP.
ments, delayed sweep will provide increased accuracy
and resolution. In this guide, we discuss three pro-
cedures using delayed sweep: magnification of a portion
of a complex waveform for closer investigation, mea-suring the time interval between two pulses, and mea-suring pulse jitter.The first procedure is discussed in the section: Ob-taining Basic Displays. The remaining two proceduresare explained in the following paragraphs.
Delayed Sweep Time Interval Measurements.
layed sweep mode can be used to increase the accuracy
For many time-interval measure-
The de-
Model 1744A
of your timing measurements. The following measure-ment determines the time interval between two pulsesdisplayed on the same trace. The procedure may also beused to measure the time interval between pulses fromtwo different channels or to make time duration mea-surements an a single pulse. To demonstrate the in-crease in accuracy, a measurement will first be madeusing only the main time base, and then the delayedtime base will be used to make the same measurement.
1.
Apply your signal to the channel A INPUTconnector, and set TRIGGER and DISPLAY to channelA.
2.
Set input coupling as desired, and adjustVOLTS/DIV for approximately four divisions of ampli-tude.
3.
Select INT main trigger, and MAIN sweep.
4.
Adjust the main TIME/DIV control to displaysix to eight divisions between pulses, and adjust mainTRIGGER LEVEL for a stable display.
5.
Using horizontal POSITION, place the 50%point of the first pulse on a convenient graticule line andcount the number of divisions to the 50% point of thesecond pulse (see figure 20).
40
Model 1744A
50%
Operators Guide
1
6.2 DI V
0
1740A-019
Figure 20. Time Interual Measurement Using Main
Time Base
6. To calculate time interval (t), use the following
formula:t = (Divisions between pulses x main TIME/DIVExample: Assume 6.2 divisions between pulses, and amain TIME/DIV setting of 0.5 mSEC.Substituting in the formula:t = (6.2 div x 0.5 ms/div) ±2% at room temperaturet = 3.1 ms ±0.062 ms
Now we will use delayed sweep to make the samemeasurement.
1.
Perform steps 1 through 4 of the previous pro-
cedure and select AUTO SWEEP AFTER DELAY.
1740A-020
I
NTENSIFIED
Figure 21. Intensified Area on First Pulse
2.
Set the delayed TIME/DIV control as required,
and turn the DELAY control to place the intensified
portion on the first pulse (see figure 21).
3.
Select DLY'D sweep and adjust the DELAYcontrol so the 50% amplitude point of the first pulse is onthe center vertical graticule line. Note the DELAYcontrol reading.
4.
Turn DELAY control clockwise until the secondpulse is positioned on the same point of the centervertical graticule line. You can verify this is the correctpulse by returning to MAIN sweep and observing theintensified portion. Again note the DELAY controlreading.
41
Operators Guide
Model 1744A
5. To calculate time interval (t), use the following
formula:
‚Second DELAY -First DELAY x main
t=(
reading
reading )TIME/DIV ±error
Example: Assume the First DELAY control reading is
1.31 and the second DELAY control reading is 7.58 with
main TIME/DIV control set to 0.5 ms/div.
Substituting in the formula:
=
(7.58 -1.31) x 0.5 ms/div ±error
tt = 3.14 ms ±error
The error is ±[(0.5% x t) + (0.1% x maximum delay
period)]. The maximum delay period is the main sweep
rate times the total length of the display (10 div in the
1744A).Therefore,error = ±[(0.5% x 3.14) + (0.1% x 5)] = ±0.021 ms.And,
t = 3.14 ms ±0.021 ms, an accuracy of <0.7%. For greatest
accuracy, use the fastest main sweep rate possible. This
reduces the maximum delay period.
Pulse Jitter Measurements. (Also see section on Ob-taining Basic Displays.)
Jitter is a time uncertainty in thewaveform caused by random noise, or spurious, non-periodic signals. To measure jitter use the followingprocedure.
1.
Apply the signal to channel A or B INPUT
connector, and select A or B DISPLAY and TRIGGER.
2.
Adjust the appropriate VOLTS/DIV control forfive or more divisions of vertical deflection, and set themain TIME/DIV control to show the complete wave-form.
AdjustTRIGGERLEVELuntilthedisplayisas
3.
stable as possible.
4.
Set the delayed TIME/DIV control as required,and turn the DELAY control to place the intensifieddisplay on the portion of the pulse showing jitter.
SWEEP VERNIER control should be in CAL detent
position.
Select the DLY'D mode and AUTO SWEEP
5.AFTER DELAY. The horizontal movement of the pulseis the pulse jitter. There is some inherent jitter in anydelayed sweep time base, and it should be included in themeasurement (jitter in the 1744A is 1:50 000, which isinsignificant in most measurements). Using the hori-
zontal POSITION control, place the leading edge of thepulse an the center vertical graticule line; with thevertical POSN control center the display.
6.
Measure horizontal displacement on the center
horizontal graticule line as shown in figure 22. Thisdisplacement times the delayed TIME/DIV settingis the pulse jitter in time.
42
Model 1744A
0.2 DIV
DELAY INTERVAL __^..l1
Operators Guide
VARIABLE
ARMED
D SWEEP
1740A-026
1740A-024
Figure 22. Pulse Jitter Measurement
Example: Assume the delayed TIME/DIV setting is 0.1
mSEC and the horizontal displacement is 0.2 div (seefigure 22).Pule Jitter = 0.2 div x 0.1 ms/div = 0.02 ms.
Eliminating Jitter. You can eliminate jitter from thedisplay by using the TRIG-SWEEP AFTER DELAYcontrol. In this mode, the delayed sweep is triggered the jittering pulse after the delay interval. So by trig-gering the delayed sweep after the delay period, theeffect of jitter
on
the display is eliminated, and you canmeasure pulse parameters. Remember, in this mode theDELAY dial is uncalibrated.
Figure 23. Pulse With Variable Time Duration
Viewing Pulses With Variable Time Duration. When the
time duration between the end of one pulse and the startof another pulse is variable, you can use the TRIG-SWEEP AFTER DELAY control and the DELAY dial toarm the delayed-trigger circuit after the last knownpulse. The delayed sweep will now be triggered by thepulse with variable time duration and its parameterscan be measured (see figure 23).
on
MEASURING PHASE DIFFERENCE BY TIME DELAY.
The phase difference between two Signals of the samefrequency can be determined up to the frequency limita-tion of the vertical amplifier. Use the following pro-cedure:
43
Operators Guide
Model 1744A
Select ALT, channel A TRIGGER, and main
1.
POS trigger.
2.
Apply the input signal to the channel A INPUTconnector and the output signal to the channel B INPUTconnector. The cables or probes used must either havethe Same electrical length or the delay differences mustbe accounted for to prevent measurement error.
3.
Select AC input coupling for both channels, andadjust channels A and B VOLTS/DIV and verniercontrols for an equal amplitude an both channels.
4.
Adjust the main TIME/DIV and SWEEPVERNIER controls so a complete cycle for each wave-form is displayed within 8 horizontal divisions.
5.
Using the POSN controls center both wave-
forms vertically.
6.
Readjust SWEEP VERNIER for one completecycle of the input signal in an exact number of majordivisions. Six or eight divisions is suggested, whichwould equal 60°/Div and 45°/Div respectively. You canobtain additional resolution by using the MAG X10switch. In this Gase, six divisions would equal 6°/Divand eight divisions would equal 4.5°/Div.
7. Count the number of major plus minor divisionsbetween the reference signal .and the output signal atthe point where they both Cross the center horizontalgraticule line. Convert divisions to degrees and this isthe phase difference.
I
NPUT
3 MINOR DIV =
OUTPUT
3fi
1740A-027
Figure 24. Phase Difference Measurement
Example: Assume one cycle of the input signal occurs insix divisions and there are three minor divisions be-tween the input and output waveforms (see figure 24).Since one major division equals 60°, one minor divisionequals 12°. Phase Difference = 3 x 12 = 36°; the outputlags the input by 36°.
44
Model 1744A
Operators Guide
A VS B PHASE MEASUREMENTS.
The A VS B mode willallow you to measure the phase difference between twosignals of the same frequency up to 100 kHz. Thechannel A input signal provides deflection along the Y-axis, and the channel B input signal provides deflectionalong the X-axis. The phase difference can be measured
from the resulting Lissajous pattern using the followingprocedure:
Connect one signal to the channel A and the
1.
other to the channel B INPUT cc mectors.
2.
Select A VS B, and adjust the channelA VOLTS/DIV control for five to six divisions ofvertical deflection (Y-axis) and the channel BVOLTS/DIV control for seven to eight divisions ofhorizontal deflection (X-axis).
3.
Use the channel A POSN control and the
horizontal POSITION control to center the display.
Measure distances A and B as shown in figure
4.
25. A is the distance intersected by the trace an thecenter horizontal graticule line, and B is the totalhorizontal deflection of the trace.
«
1-
8 DIV
C = D
1740A-028
Figure 25. A VS B Phase Measurement
5.
The sine of the phase angle between the twosignals is A/B. Figures 26, 27, and 28 show signals inphase, 90° out of phase, and 180° out of phaserespectively. If the trace is rotating, the signals are notat the same frequency.
Example: In figure 25, A equals five divisions and Bequals eight divisions. Distance C is equal to distance D.The sine of the phase difference (4) is A/B which is
0.625. Therefore: Phase Angle (4) = arc sine of
0.625=38.7°.
45
Operators Guide
Mode! 1744A
CI:
Figure 26. Signals in Phase
Figure 27. Signals 90
1740A-029
0
1740A-030
Out of Phase
1740A-031
Figure 28. Signals 1800 Out of Phase
TRIGGERING.
TRIGGER VIEW.
channel A or B trace with the trigger Signal if channel A
or B is selected as the display mode. In this type ofoperation, the trigger sensitivity is approximately equal
to the VOLTS/DIV setting. In the ALT or CHOP
display mode, three signals are displayed: channel A,the selected main trigger signal, and channel B. InTRIG VIEW, the Center horizontal graticule linerepresents the trigger threshold level with respect to thetrigger signal.
The TRIG VIEW control replaces the
Model 1744A
It is frequently helpful to observe the trigger signalbeing applied to the external trigger input. When you usetrigger view in conjunction with CHOP or ALT, bothvertical channels plus the external main trigger signal
can be viewed simultaneously. This is useful in settingtriggering and observing time correlation between theexternal trigger Signal and the channel A and B signals.The deflection factor is approximately 100 mV/div.Delayed trigger cannot be viewed by TRIG VIEWcontrol.Example: We will now use trigger view to determine thetriggering level location.
1.
Connect the trigger signal to the main EXTTRIGGER input connector and select main EXTTRIGGER.
2.
Select TRIG VIEW. The trigger signal will bedisplayed near center screen. The point where thetrigger signal Grosses the center horizontal graticule is
the trigger point (see figure 29).
By adjusting the TRIGGER LEVEL control, you can
move the trigger level location. The center horizontal
graticule line indicates the trigger point. When you use
the POS position of the POS/NEG switch, the triggercircuit triggers on the positive-going portion of thetrigger signal; in NEG, it triggers on the negative-goingportion of the trigger signal.
Operators Guide
FRIGGER
PO I NT
Figure 29. Trigger Point Location
ELIMINATING MULTIPLE TRIGGERING ON COMPLEXWAVEFORMS.
triggering. To have a stable display, the period betweensweeps must match the period of the waveform beingdisplayed. In the example, the first sweep displaysthree bits of a four-bit word. The next sweep displays theremaining bit in the word. So on consecutive sweeps wesee different portions of the saure word causing theinstability in figure 30.
To eliminate the instability, the TRIGGER HOLDOFFcontrol can be adjusted to vary the time between the end
Figure 30 shows an example of multiple
47
Operators Guide
1740A-033
Figure 30. Multiple Triggering with Display Instability
Mode! 1744A
of one sweep and the beginning of the next. This is theholdoff period. In the example, if you increase theholdoff period long enough, the trigger from the fourthbit is held off, which eliminates the additional sweep
that caused the display instability (see figure 31).
MEASURING CRT WRITING SPEED.
Writing speed is a rate that defines how far the electron
beam can be deflected per unit of time and still provide astored signal for display. The writing speed of the 1744A
is specified at 1800 cm/µs, making it possible to capture
a 100 MHz, single-shot, sine wave with an amplitude ofeight major divisions. Single-shot pulses eight majordivisions in amplitude with transition timen as fast as
3.5 ns may also be captured. Although the 1744A mustbe overdriven to obtain an 8-division display atmaximum bandwidth, a fully specified trace is capturedand displayed within the quality area of the CRT.
With a specified writing speed of 1800 cm/µs, themaximum cw frequency that can be captured with anamplitude of eight major division can be calculated asfollows:
1740A-034
Figure 31. Multiple Triggering Eliminated with
Trigger Holdoff Control
48
Writing Speed
= 7r
Af centimeters/µs
Where: f = frequency in megahertz.
A = peak-to-peak amplitude in centimeters.
Model 1744A
Therefore: 1800 cm/µs = (3.1416)(8 div)(0.72 cm/div) f
_
(3.1416)(8)(0.72)(10 6)
1800
= 99.47 MHz
To determine the writing speed necessary to capture an8-division amplitude pulse with an observed transitiontime of 3.5 ns, use the following procedure:
Writing Speed =
Therefore: Writing Speed =
0.8A
r
0.8(div^
(
0.72 cm/div)
3.5 x 10 s
In the above example, it will be noted that the specifiedwriting rate of the 1744A (1800 cm/µs) is more thanadequate to capture pulses that approach the specifiedvertical deflection limits of the 1744A.
=1317 cm/ s
µ
5
6
7
8
9
10
16
13
1411
12
17
20 21 22
19
1
21 3
4
129130131142139140141
1
49NrnT
Operators Guide
62
61
63
1
0
•
1
66
Front and Rear-panel Controls and Connectors
Figure 32.
49
AFRICA, ASIA, AUSTRALIA
g
-142DT
Sprin
1611 dOpO Oaeensland
Tel
220-1544
Cable HEWPARO Bnsbarle
^td1 Potk
Med
G
uam Memcal
Jay
Base
P 0 Box 8947Temu
^g
Tel
6468A4t7
Gable EARMED Guam
KONG
HONG
Scnm^ot 8 Co IHon
P 0 Box 297
m
Lonnal^
939tn FloorConnau
9
Hong Kong
Tel
H-25529
74166
Telex
Gable SCHMIOTCO
INDIA
Blue 51ar LIeKastun Buddm s
Mp Tata Rd
Jam
et
Bombay Tel
29 50 21
001-2156
TNex
Gable BLUFFROSTBWe Star LIe.
Sands
Ort
414.2 Prabhade+iBombe 400 025
y
Tel
n5 7Telex 011-4093Gable. FROS
Blue Star Ltd .
BdBHouse
anox
Prxbnaoew
m
Bobey 400 025T¢I d5 73 01
011-3751
TelexGable BLUESTARBlue Star Llo
Hare SinterP 0 Box 506Cel700 001
23-01
Tel
23-0131
T
el
ex 0217655
Gable BLUESTAR
ANGOLATelectraEmpresa Tecmca de
EgwpamenrosEIl
dncos
$ A RL
R
Barboss Radnques.
Caua P.Stal. 6467
Luanda
TN 355151fi
Cable. tELECTRA L uanea
AUSTRALIA
Hewlett-Packard Australla
Ply
Ltd
.
31
41 Joseph Street
Blaekbuen. victona 3130
P 0 Box
He
Doncaet
er Eaet. Victon a 3109
Tel
89-6351
Telex
x
0
31-
24
Gable HEWPARD MelbourneHewlelt-PackardA usValia
P
Ltd
P
31 eehege Street
New
New Sautn WalesTel 449-6566TelexGable HEWPAflD Sydney
Hewlett Packard Ausvalu
153 Greenn
Parkei-e S A.. 5063TelTelex 82536 AOEIGable HEWPARD AOELAlO
tot $brtin
Netllantls. W A 6009
el
Ta
Gbe
Gable
HNPI
121
yehwrwkFTel
Telex
Cable
Hewl¢II Packard Au slralla
litt
Testhers Umon Buudln,
495-499 B.uhdary Sbeet
2013
21661
Pry Ltd
5
Road
27225911
NMewII-Packard Aootralra
Pty
Ltd
Hign
q
wa
y
B6-5455
PER7H
H EWP
EWPARO PERTH
kard A,Stralia
LId
P
ry
Lid
Slre¢1
g0"
hwiek A C T 2609
95-2733
62650 Canberra
CANBERRA
Fluor
raha
Ltd
Pro
CaltuMtors Onl
N
Inc
5 u
DDh
Builaing. Room 210
%9tt
Kan I Ltd
A
Il
Centre
Lentral
Road
5
t-
HR
SCHMC
Hong Kunq
q
020
2t
Savarkar Mar A
B
87
LUE
Blue Star Ltd7m 6 811 FloorBnandan Hnuse91 Nenru Plato