Horizon Elite User manual

NOTE: The programmable run-time
has been preset for
10 Minutes
Instructions to change this
setting are located on page 7.
P/N 7711017
Model 755V Shown
Model 755 - 24-Place, 755V - 24-Place, 755 - 12-Place, 755 Microplate
Operator’s / Service Manual
2
Table of Contents
Model Description pg. 3 Supplied Equip. pg. 4 Available Features pg. 5 Specifications pg. 6-7 Control Panel pg. 8 Setup Location pg. 9 Initial Setup Procedure pg. 9-10 Additional Setup Procedure pg. 10-11 Operation pg. 11 Balanced Loads pg. 12 Tube Holder Configurations pg. 13 Force Chart—12 place pg. 14 Care and Preventative Service pg. 15 Rotor Removal and Installation pg. 16 Safety pg. 17 Emergency Rotor Chamber Entry pg. 17 Troubleshooting pg. 18 Available Accessories pg. 19-20 Warranty pg. 24
WARNING: Care should be taken when handling substances that are known to be toxic, radioactive or contaminated with pathogenic microorganisms when using this centrifuge for the safety of both the operator and service personnel. When Risk Group II materials are used (as identified in the World Health Organization “Laboratory Bio-Safety Manual”), a Bio-Seal should be employed. More than one level of protection must be provided in the case of materials of a higher group. The use of flammable or explosive materials as well as those materials which chemically react vigorously is prohibited.
Model Description:
The Horizon Model 755 is a continuous-duty, electronically-controlled, single or variable-speed, horizontal
laboratory centrifuge with a lid safety interlock system. The unit is controlled by an electronic push-button timer that is variable from 1 to 30 minutes, for precise spin times and ease of use. Samples can be safely viewed through the transparent lid. The imbalance detection system safely terminates a run cycle in the event that a severely imbalanced load is present. Entry into the machine is restricted during operation by the safety interlock system. The Horizon features a lighted control panel that displays the status of the machine, easily viewable from a distance. Finally, a cycle counter keeps a count of the total number of machine runs for servicing and maintenance records.
Intended Use:
General purpose laboratory centrifuge for sample preparation.
3
Supplied Equipment*:
Rev. A
P/N 7711017
Rev. A
The following items come standard with each Horizon 755V-24 Centrifuge:
1. One (1) six-place horizontal rotor
2. Six (6) 100 mm four-place tube carriers p/n 7786022
3. One (1) Operator's Manual (not shown) p/n 7713023
4. One (1) Tutorial CD (not shown) Optional:
• 50mLsingle-placetubecarrier  p/n7713037  • Carriercap p/n7713035
1. 2.
* The rotor and rotor accessories are rated for a rotation frequency of 3500 RPM (a force of 2100 xg).
Supplied Equipment*:
The following items come standard with each Horizon 755V-12 Centrifuge:
1. One (1) Horizon centrifuge
2. One (1) twelve-place horizontal rotor p/n 7786024
3. Twelve (12) 100 mm test tube holders p/n 7713031
4. Twelve (12) 125mm test tube holders p/n 7713032
5. Twelve (12) 75 mm test tube p/n 7713033
6. One (1) Operator’s manual
1. 2. 3. 4. 5.
Model 755V Shown
NOTE: The programmable run-time
has been preset for
10 Minutes
Instructions to change this
setting are located on page 7.
P/N 7711017
NOTE: Your centrifuge may
* The rotor and rotor accessories are rated for a rotation frequency of
have been shipped with additional
or alternate accessories.
Model 755 - 24-Place, 755V - 24-Place, 755 - 12-Place, 755 Microplate
Operator’s / Service Manual
4,000 RPM, (a force of 2450 xg with the 100mm tube holders).
Supplied Equipment*:
The following items come standard with each Horizon 755 Microplate Centrifuge:
1. One (1) Microplate rotor p/n 7786051
2. One (1) Operator’s manual p/n 7711017
1. 2.
NOTE: The programmable run-time
has been preset for
10 Minutes
Instructions to change this
setting are located on page 7.
Model 755V Shown
Model 755 - 24-Place, 755V - 24-Place, 755 - 12-Place, 755 Microplate
* This rotor is rated for a rotation frequency of 1500 RPM.
Operator’s / Service Manual
4
Available Features:
• Swing-outhorizontalrotordesignincorporatingauniquetesttubeholderthatproduceshorizontally separated samples while requiring no additional parts, (optional fixed-angle rotors are also available)
• Variable-speed(Model755V,seespeedrangebelow)
• Cool-Flowairowdesignthatpreventsoverheatingofsamples
• Heavygaugesteelconstructionforsafetyanddurability
• Lidsafetyswitchthatpreventsthecentrifugefromoperatingunlessthelidisclosedandlatched
• Removablerotorforeasycleaning
• Lockinglidthatallowsentryintothecentrifugeonlyaftertherotorhascompletelystopped
• BrushlessDCMotor;requiresnoroutinemaintenance,coolrunning
• Clearlidforsafeobservationofsamplesandopticalcalibrationofspeed
• Electronicallycontrolledtimedoperationvariablefrom1to30minutes
• Push-buttonoperation
• Indicatorlights:
‘RUN’ green -lights when power is applied to the motor ‘LATCH’ yellow -lights when the lid is closed and latched ‘UNLOCKED’ red -lights when the lock system is deactivated
• Cyclecountertomonitormachineusage
• Audibleindicatorattheendofeachrun
Specifications: 12-Place
Speed Range (Model 755V):
755V w/12-Place Horizontal Rotor: 500 to 3,200 (± 100) RPM* 755V w/Fixed-Angle Rotors: 500 to 4,000 (± 100) RPM* Max Speed (Model 755): (This speed is set by the factory, see page 7) with 12-Place Horizontal Rotor 3,600 (± 100) RPM with Fixed-Angle Rotors 3,600 (± 100) RPM
Maximum capacity:
12-Place Horizontal Rotor: 180 mL (12 x 15 mL) 20-Place Fixed-Angle Rotor: 300 mL (20 x 15 mL)
Overall Dimensions (H x W x D): 9 in. x 14.5 in. x 17 in. Centrifuge Motor: 1/2 H.P. Brushless DC Nominal Acceleration Time: 30 seconds Nominal Braking Time: 15 seconds Protection Breaker: 4 Amp. re-settable Timer (electronic): 1 to 30 min ± 1% Current Requirement: 2.0 A @ 115 VAC
or 1.0 A @ 230 VAC
Voltage Requirement: 115 or 230 (±10%) Volts AC Frequency: 50 / 60 Hz Weight: 39 lbs.
* Any use other than those specified by the manufacturer is explicitly prohibited. Maximum speed shown is for the given rotor with 100mm red tube holders. The use of other tube holders
will alter the maximum speed.
Specifications: 24-Place
Speed Range (V.S.V.): 500 to 3,150 (± 5%) RPM
755V w/24-Place Horizontal Rotor: 3,150 (± 100) RPM* 755V w/20-Place Fixed-Angle Rotors: 3,200 (± 100) RPM*
Force Range (variable speed version): 45 to 1,760 (± 10%) xg Single Speed Setting:
(This speed is set by the factory, see page 7) 3,000 (± 5%) RPM
Force (@ 3000RPM): 1,600 x g Maximum capacity: 240 mL (24 x 10 mL)*
300 mL (6 x 50 mL)*
Overall Dimensions (H x W x D): 9 in. x 14.5 in. x 17 in. Centrifuge Motor: 1/2 H.P. Brushless DC Nominal Acceleration Time: 30 seconds Nominal Braking Time: 15 seconds Protection Breaker: 4 Amp. re-settable Timer (electronic): 1 to 30 min ± 1% Current Requirement: 2.0 @ 115 VAC
or 1.0 A @ 230 VAC
Voltage Requirement: 115 or 230 (±10%) Volts AC Frequency: 50 / 60 Hz Weight: 39 lbs.
6
Specifications: Microplate Rotor
Speed: 1500RPM (±100) RPM Force: 322 (±50) xg Maximum capacity: 200g
Overall Dimensions (H x W x D): 9 in. x 14.5 in. x 17 in. Centrifuge Motor: ½ H.P. Brushless DC Nominal Acceleration Time: 65 seconds Nominal Deceleration (Breaking) Time: 50 seconds Protection Breaker: 4 Amp. re-settable Timer (electronic): 1 to 30 min ±1% Current Requirement: 2.0 A @ 115 VAC
or 1.0 A @ 230 VAC
Voltage Requirement: 115 or 230 (±10%) Volts AC Frequency: 50 / 60 Hz Weight: 39 lbs.
Any use other than those specified by the manufacturer is explicitly prohibited.
* Maximum speed density is 1.15 grams/mL (water density = 1.00 grams/mL).
Control Panel:
Lid Safety
Switch
‘RUN’
Indicator
‘LATCH’
Indicator
Speed Control (755V Only)
‘RUN’ Indicator
‘LATCH’ Indicator
‘UNLOCKED’ Indicator
Power (On/Off) Button
Speed Control (755V ONLY)
Power (On/Off) Button
Lights up when the machine is in operation, (power is being applied
to the motor).
Lights up when the lid has been closed and
latched properly.
Lights up to indicate that the locking mechanism has been deactivated, allowing access to the
rotor chamber.
This button controls the power to the centrifuge. The power is on when
the button is illuminated.
Use this input to set the running speed of the centrifuge. The speed may also be adjusted while the centrifuge is running.
‘START’ Button
‘OPEN / EMERGENCY STOP’ Button
Cycle Counter
Lid Safety Switch
‘UNLOCKED’
Indicator
START’
Button
‘OPEN/
EMERGENCY
STOP’
Button
Cycle
Counter
Begins a new run, (the lid must be closed, see pg. 6).
Allows access into the rotor chamber by disengaging the locking mechanism. Entry is only permitted when the rotor is stopped. Pressing this button during operation will terminate the run and unlock the lid after the
rotor has come to a stop.
This counter increments once for each run of the centrifuge. This count is used for both warranty purposes and for routine maintenance.
The lid safety switch prevents the centrifuge from operating while the lid is open. If the knob is not turned completely clockwise to its stop position the centrifuge cannot operate. Programming functions are accessed by first opening the lid safety switch.
8
Loading...
+ 16 hidden pages