Horiba LAQUAtwin Salt-22 Operating Manual

Instruction Manual (Operation)
COMPACT WATER QUALITY METER LAQUAtwin-Salt-22

Specifications

Model LAQUAtwin-Salt-22 Target Salinity
Measurement principle ISE method Minimum sample volume More than 0.3 mL
Measure­ment range
Resolution (default)
Calibration Up to 2 points
Accuracy Waterproof IP67 (no failure when immersed in
Display Custom (monochrome) digital LCD
Operating environment
Power CR2032 batteries (2) Battery life Approx. 400 h continuous operation
Material ABS epoxy (main material) Dimensions 164 29 20 mm (excluding projec-
Mass Approx. 50 g (excluding batteries)
0.01% to 25% by weight Display range: Resolution
0.00% to 0.99%: 0.01%
1.0% to 9.9%: 0.1% 10% to 25%: 1%
Default: 0.5% and 5.0%
*2
±10% of reading value
water at a depth of 1 m for 30 min)
with backlight Temperature: 5C to 40C
Humidity: 85% relative humidity or less (no condensation)
(backlight off mode)
*4
tions)
*1 0.05 mL or more if sampling sheet B (sold sepa-
rately) is used.
*2 The closeness of agreemen t between a mea-
sured value and an actual value of standard solu­tion after two-point calibration using 0.5% and
5.0% standard solutions.
The standard solution used for the later cali-
bration was measured after two-point calibra­tion at 0.5% and 5.0%.
The calibration and measurement are per-
formed at the same temperature.
The error of standard solutions and rounding
error (1 digit) are not included.
*3 The meter cannot be used underwater.
*1
*4 When the backlight is used, battery life will
shorten.
Items Quantity
Sensor S021 1 Meter 1 Storage case 1 Batteries CR2032 2 Standard solution 0.5% 1
5.0% 1
*3
Pipette 1 Sampling sheet B (5 sheet-pack) 1 Instruction manual (Operation) 1 Instruction manual (Before use) 1
Consumable parts sold separately
Items Specifications Part No.
Sensor Standard
solution Sampling
S021, Salt
3200459866 514-05, NaCl 0.5% 3999960112 514-50, NaCl 5.0% 3999960113
Y046, 100 sheet-pack 3200053858

Part Names

2
1
4
3
5
6
9
Note
Press the switches 0.5 seconds or more u nless otherwise specified.
1 Flat sensor 2 Light shield cover 3 Liquid junction
7
10
4 Response mem-
brane
8
5 Lithium batteries 6 MEAS switch 7 ON/OFF switch 8 CAL switch 9 Waterproof gasket 10 Strap eyelet

Initial Setup

Attaching/detaching the sensor
Attaching the sensor
1. Power OFF the meter.
2. Confirm that the waterproofing gasket is
clean and undamaged.
3. Slide the sensor onto the meter so that catch "A" on the back of the meter fits into hole "a" on the sensor tongue as shown.
Note
Be careful not to twist the waterproof gasket.
Waterproof gasket
a
Detaching the sensor
1. Power OFF the meter.
2. Lift the sensor tongue
tip and slide the sensor a little away from the meter.
3. Pull out the sensor all the way from the meter.
Inserting/removing batteries
Inserting the batteries
1. Power OFF the meter.
2. Slide both batteries into
the battery case as shown.
Be sure to use two CR2032 batteries, and put them with the plus sides
Battery clip
(+) upwards.
Removing the batteries
1. Power OFF the meter.
2. Use a ball-point pen or
other tool to pry the batter­ies out from the clips as shown.
Electrode conditioning
Note
Before using the sensor for the first time or
after several days of disuse, perform electrode conditioning.
Perform calibration after electrode condition-
A
ing.
1. Place some drops of 5.0% standard solu­tion to the flat sensor.
2. Wait a few hours before use.
There is no need to switch the meter ON.
3. Clean the flat sensor with running water.
Battery
sheet B
LAQUAtwin-Salt-22 Operation
CODE: GZ0000460290
March, 2017
1 2017 HORIBA Advanced Techno Co., Ltd.

Basic Operation

Power ON
1. Press and hold the ON/ OFF switch.
The power is switched ON, and the meter model number is displayed on the LCD.
Power OFF
1. Press and hold the ON/OFF switch.
The power is switched OFF.

Calibration

Calibration is required before measurement. Use standard solution within the measurement range in the specifications.
Tip
Calibration values are saved even if the meter
is switched OFF.
Calibration value is rewritten if calibration is
repeated using the same standard solution.
Calibration points
The number of calibration points is up to 2.
Multi-point calibration
1. Set the salinity of standard solution for cal­ibration referring to " 1st calibration point setting" (page 5) and " 2nd calibra­tion point setting" (page 5).
The 1st point is set to 0.5% and the 2nd point is set to 5.0% by the default.
2. Open the light shield cover and place some drops of the standard solution on the flat sensor taking care to cover the entire flat sensor.
Rinsing the sensor with the standard solution beforehand will provide a more accurate calibration as it will reduce sample crossover contamination.
3. Close the light shield cover and press the CAL switch.
The meter enters the CAL mode and blinks the display of the set 1st-point salinity. Pressing the MEAS switch switches the displayed value between the set salinity.
Light shield cover
4. With the set salinity of the 1st point dis­played, press the CAL switch.
and blink, and the calibration value is displayed. After the calibration is complete, and stop blinking and the measured value is displayed. The calibration value at 25C is displayed for 1 s and the display returns to the measurement mode automatically.
5. Open the light shield cover and remove the standard solution. Then remove moisture on the sensor by gently dabbing with a so ft tissue.
This completes the 1st point calibration.
6. To perform 2nd point calibration, repeat steps 2. to 5.
Calibratio n error
If blinks and Er4 (error dis­play) appears, the calibration has failed. Perform electrode conditioning. Check that the correct standard solution is used, and repeat calibration after cleaning the sensor. If the calibration repeatedly fails when using the correct standard solution(s), the sensor may have deteriorated. Replace the sensor with new one.
Tip
For accurate measurement, calibrate using 2 points of st anda rd solution with a t enfold or mo re salinity difference encompassing the target salin­ity. When the salinity to be measured is very high or very low, accuracy may be poorer.

Measurement

Sample setting
1. Open the light shield cover and put some drops of sample on the flat sensor to cover the entire flat sensor.
2. Close the light shield cover.
Measurement mode
The auto stable (AS) mode and the auto hold (AH) mode can be selected. Refer to " Mea­surement mode change" (page 5) for the opera­tion to set the measurement mode.
Auto stable (AS) mode
This is the default setting. appears when the measured value meets the stability criteria. If the value changes, disappears.
1. Confirm that the meter is in the measure­ment mode, and place a sample on the sen­sor.
When the read value meets the stability criteria, appears and the reading is locked.
2. Document the displayed value when appears.
If the read value does not meet the stability criteria, disappears and the reading changes with time.
Auto hold (AH) mode
appears when the measured value meets the stability criteria. The reading then locks and will not change until the MEAS switch is pressed for the next measurement.
1. Confirm that the meter is in the measure­ment mode, and place a sample on the sen­sor.
2. Press the MEAS switch.
The auto hold function is activated.
blinks until the measured value has stabilized.
When the measured value is stable, stops blinking and the displayed value is locked with
and displayed simultaneously.
3. Document the displayed value.
4. Press the MEAS switch.
The auto hold function is deactivated and disappears. Be sure to perform this step before starting the next measurement. Or, you may mistake the displayed hold value for the next measured value.
LAQUAtwin-Salt-22 Operation 2 CODE: GZ0000460290
Loading...
+ 3 hidden pages