Horiba LAQUAtwin NO3-11 Operating Manual

Instruction Manual (Operation)
COMPACT WATER QUALITY METER LAQUAtwin-NO3-11, LAQUAtwin-NO3-11C, LAQUAtwin-NO3-11S

Specifications

Model LAQUAtwin-NO3-11, LAQUAtwin-
NO3-11C, LAQUAtwin-NO3-11S
Target Nitrate ion (NO Measurement principle ISE method
Minimum sample volume More than 0.3 mL Intended
use
LAQUAtwin-NO3-11: General measurement LAQUAtwin-NO3-11C: Crop measurement LAQUAtwin-NO3-11S: Soil measurement
Measure­ment range
Resolution (default)
NO
: 6 to 9900 [unit: ppm or mg/L]
3
-N: 0.7 to 1100 [unit: kg/10a]
NO
3
-N: 1.4 to 2200 [unit:
NO
3
Display range (NO 6 to 99: 1 100 to 990: 10 1000 to 9900: 100
Calibration Up to 2 points
Default: LAQUAtwin-NO3-11: 150 ppm and 2000 ppm LAQUAtwin-NO3-11S: 30 ppm and 300 ppm LAQUAtwin-NO3-11C: 300 ppm and 5000 ppm
Accuracy
LAQUAtwin series (NO
*2
±10% of reading value
3
-
)
3
*1
ppm
or
mg/L
-
): Resolution
3
[unit: ppm or mg/L]
-
) Operation 1 2017 HORIBA Advanced Techno Co., Ltd.
Waterproof IP67 (no failure when immersed in
water at a depth of 1 m for 30 min)
Display Custom (monochrome) digital LCD
with backlight
Operating environment
Temperature: 5C to 40C Humidity: 85% relative humidity or
less (no condensation) Power CR2032 batteries (2) Battery life Approx. 400 h continuous operation
(backlight off mode)
*4
Material ABS epoxy (main material) Dimensions 164 29 20 mm (excluding projec-
tions) Mass Approx. 50 g (excluding batteries)
*1 0.05 mL or more if sampling sheet B (sold sepa-
rately) is used.
*2 The closeness of agreemen t between a mea-
sured value and an actual value of standard solu­tion after two-point calibration using the standard solutions of the same concentration as provided in the package.
The standard solution used for the later cali-
bration was measured.
]
The calibration and measurement are per-
formed at the same temperature.
The error of standard solutions and rounding
error (1 digit) are not included. *3 The meter cannot be used underwater. *4 When the backlight is used, battery life will
shorten.
Items in package
Quantity
Items
11* 11C* 11S*
Sensor S040 1 Meter 1 Storage case 1 Batteries CR2032 2
*3
Items
Quantity
11* 11C* 11S*
Standard solution 150 ppm 1 0 0
2000 ppm 1 0 0 30 ppm 0 0 1 300 ppm 0 1 1
5000 ppm 0 1 0 Pipette 1 Sampling sheet B 5
sheets
Instruction manual (Operation)
Instruction manual (Before use)
0 100
sheets
1
1
* The model names are abbreviated on the column

Part Names

2
1
4
3
5
6
9
Note
Press the switches 0.5 seconds or more u nless otherwise specified.
1 Flat sensor 2 Light shield cover 3 Liquid junction
7
10
4 Response mem-
brane
8
5 Lithium batteries 6 MEAS switch 7 ON/OFF switch 8 CAL switch 9 Waterproof gasket 10 Strap eyelet
headers. 11: LAQUAtwin-NO3-11 11C: LAQUAtwin-NO3-11C 11S: LAQUAtwin-NO3-11S
Consumable parts sold separately
Items Specifications Part No.
-
Sensor Standard
solution
Sampling
S040, NO Y045, NO Y043, NO Y044, NO
Y042, NO Y041, NO
Y046, 100 sheet-pack 3200053858
3
-
150 ppm 3200053536
3
-
2000 ppm
3
-
30 ppm
3
-
300 ppm
3
-
5000 ppm
3
3200459870
3200053532 3200053535 3200053514 3200053433
sheet B

Initial Setup

Attaching/detaching the sensor
Attaching the sensor
1. Power OFF the meter.
2. Confirm that the waterproofing gasket is
clean and undamaged.
3. Slide the sensor onto the meter so that catch "A" on the back of the meter fits into hole "a" on the sensor tongue as shown.
Note
Be careful not to twist the waterproof gasket.
CODE: GZ0000460288
Waterproof gasket
a
March, 2017
A
Detaching the sensor
1. Power OFF the meter.
2. Lift the sensor tongue
tip and slide the sensor a little away from the meter.
3. Pull out the sensor all the way from the meter.
Inserting/removing batteries
Inserting the batteries
1. Power OFF the meter.
2. Slide both batteries into
the battery case as shown.
Be sure to use two CR2032 batteries, and put them with the plus sides
Battery clip
Battery
(+) upwards.
Removing the batteries
1. Power OFF the meter.
2. Use a ball-point pen or
other tool to pry the batter­ies out from the clips as shown.
Electrode conditioning
Note
Before using the sensor for the first time or
after several days of disuse, perform electrode conditioning.
Perform calibration after electrode condition-
ing.
1. Place some drops of 2000 ppm standard solution to the flat sensor.
2. Wait a few hours before use.
There is no need to switch the meter ON.
3. Clean the flat sensor with running water.

Basic Operation

Power ON
1. Press and hold the ON/ OFF switch.
The power is switched ON, and the meter model num­ber is displayed on the LCD.
Power OFF
1. Press and hold the ON/OFF switch.
The power is switched OFF.

Calibration

Calibration is required before measurement. Use standard solution within the measurement range in the specifications. See also the tip of “How to select standard solution” below.
Tip
Calibration values are saved even if the meter
is switched OFF.
Calibration value is rewritten if calibration is
repeated using the same standard solution.
Calibration points
The number of calibration points is up to 2.
Multi-point calibration
1. Set the concentrations of standard solution for calibration referring to " 1st calibra­tion point setting" (page 5) and " 2nd cal­ibration point setting" (page 5).
The 1st point is set to 150 ppm and the 2nd point is set to 2000 ppm by the default.
2. Open the light shield cover and place so me drops of the standard solution on the flat sensor taking care to cover the entire flat sensor.
Rinsing the sensor with the standard solution beforehand will provide a more accurate cali­bration as it will reduce sample crossover con­tamination.
3. Close the light shield cover and press the CAL switch.
The meter enters the CAL mode and blinks the display of the set 1st-point concentration. Pressing the MEAS switch switches the dis­played value between the set concentrations.
Light shield cover
4. With the set concentration of the 1st point displayed, press the CAL switch.
and blink, and the calibration value is displayed. After the calibration is complete, and stop blinking and the measured value is dis­played. The calibration value at 25C is displayed for 1 s and the display returns to the measurement mode automatically.
5. Open the light shield cover and remove the standard solution. Then remove moisture on the sensor by gently dabbing with a soft tissue.
This completes the 1st point calibration.
6. To perform 2nd point calibration, repeat steps 2. to 5.
Calibration error
If blinks and Er4 (error dis­play) appears, the calibration has failed. Perform electrode conditioning. Check that the correct standard solution is used, and repeat calibration after cleaning the sensor. If the calibration repeatedly fails when using the correct standard solution(s), the sensor may have deteriorated. Replace the sensor with new one.
Tip
How to select standard solution: The following table show an example of the rela­tionship between the concentrations of targeted samples and standard solution used for calibra­tion.
[unit: ppm]
Targeted sample
concentration
Standard solution
Low High
2000 2000 9000
150 150 2000
15 15* 150
*How to prepare 15 ppm standard solution: Mix one part 150 ppm standard solution to 9 parts purified water (1:9 volume ratio).
For accurate measurement, calibrate using 2 points of st andard solution wi th a tenfold or more concentration difference encompassing the tar­get concentration. When the concentration to be measured is very high or very low, accuracy may be poorer.

Measurement

Sample setting
1. Open the light shield cover and pu t some drops of sample on the flat sensor to cover the entire flat sensor.
2. Close the light shield cover.
Measurement mode
The auto stable (AS) mode and the auto hold (AH) mode can be selected. Refer to " Mea­surement mode change" (page 5) for the opera­tion to set the measurement mode.
Auto stable (AS) mode
This is the default setting. appears when the measured value meets the stability criteria. If the value changes, disappears.
LAQUAtwin series (NO
-
) Operation 2 CODE: GZ0000460288
3