F-3030 Temperature Bath Addendum: Switching between Internal and External Sensors J810045 (28 Jan 2014)
Note:
The radio button alone in
FluorEssence™ software does not
control switching between internal and external probes. The temperature bath’s controller MUST be changed first.
F-3030 Temperature Bath Addendum: Switching between Internal and External Sensors
To switch between internal and external sensors, use the Temperature Bath’s front panel:
1 Click the right-arrow ► button
to move to the sub-menu:
2 Scroll up and down with the up
arrow ▲ and down arrow ▼
buttons to choose Settings, then click the central Enter
button .
3 Scroll up and down with the up
arrow ▲ and down arrow ▼
buttons to choose Application Settings, then click the central
Enter button .
F-3030 Temperature Bath Addendum: Switching between Internal and External Sensors J810045 (28 Jan 2014)
4 Scroll down with the down arrow ▼
to choose Remote Sensor, then click the central Enter button .
5 Follow the instructions on the
screen to complete the switch-over to the other sensor.
6 Use the corresponding radio button
in the Accessories page of a FluorEssence™ experiment.
In the Accessories icon, under the Temperature Controller tab, in the General area, choose the Sensor: either T-bath (internal) or Probe (external).