F-3030 Temperature Bath Addendum: Switching between Internal and External Sensors J810045 (28 Jan 2014)
Note:
The radio button alone in
FluorEssence™ software does not
control switching between internal
and external probes. The temperature
bath’s controller MUST be changed
first.
F-3030 Temperature Bath
Addendum: Switching between
Internal and External Sensors
To switch between internal and external sensors, use the Temperature
Bath’s front panel:
1 Click the right-arrow ► button
to move to the sub-menu:
2 Scroll up and down with the up
arrow ▲ and down arrow ▼
buttons to choose Settings,
then click the central Enter
button .
3 Scroll up and down with the up
arrow ▲ and down arrow ▼
buttons to choose Application
Settings, then click the central
Enter button .
![](/html/b0/b0f7/b0f7315f89bcb26ede45fd19df8203885681fa407f7dced441646d4dfbeced4c/bg2.png)
F-3030 Temperature Bath Addendum: Switching between Internal and External Sensors J810045 (28 Jan 2014)
4 Scroll down with the down arrow ▼
to choose Remote Sensor, then
click the central Enter button .
5 Follow the instructions on the
screen to complete the switch-over
to the other sensor.
6 Use the corresponding radio button
in the Accessories page of a FluorEssence™
experiment.
In the Accessories icon, under the Temperature Controller tab, in the General area,
choose the Sensor: either T-bath (internal) or Probe (external).