Hopkins IMPACT 526018 User Manual

U ser Manual
Hopkins® IMPACT Digital Arm BP
Thank you for selecting the Hopkins® IMPACT Digital Arm BP #526018.
manual thoroughly and keep it for your reference.
User Manual
Blood Pressure Monitor
TMB-1018
Arm Type
Version:1.0
Item #526018
5 Greenwood Place • Baltimore, MD 21208
1-800-835-1995 • FAX: 410-484-4036
www.hmponline.com
RELATED ITEMS
• Replacement Adult Size Cuff #526019
• Large Adult Size Cuff #526020
• Hopkins® Flex Temp Thermometer #579420
• Hopkins® Wavetm Pulse Oximeter #594029
User Manual
Blood Pressure Monitor
TMB-1018
Arm Type
Version:1.0
Hopkins® IMPACT Digital Arm BP
INTRODUCTIONINDEX
1
• Measures Systolic Blood Pressure
• Measures Diastolic Blood Pressure
• Measures Pulse Rate
• Retains historic records of up to 60 previous measurements
• Large 5.5” x 1.5” bright LCD display
The following symbols may be used in the user manual, product labeling, or on other components. They are the requirement of standard and use.
Caution: consult accompanying documents
CE Mark: conforms to essential requirements of the Medical Device Directive 93/42/EEC
Manufacturer
Species serial number
Type B applied part DISPOSAL: Do not dispose this
product as unsorted municipal waste. Collection of such waste separately for special treatment is necessary.
Direct current Authorized Representative in the
European Community
Features
Safety Information
INTRODUCTION ................................................................................................. 1
• Safety information
• LCD display
• Monitor components
BEFORE YOU START ..................................................................................... 4
• Installing and replacing batteries
• Setting date and time
MEASUREMENT ................................................................................................. 7
• Positioning the cuff
• Taking a measurement
RECALLING RECORDED MEASUREMENTS .............................................. 9
• Recalling records
• Deleting records
INFORMATION FOR THE USER ..................................................................... 11
• Tips for operation
• Maintenance
ABOUT BLOOD PRESSURE ........................................................................... 13
• What are systolic and diastolic pressure?
• What is normal blood pressure?
Why does my blood pressure uctuate throughout the day?
• Why is the reading I get from a healthcare provider different from the ones I take at home?
• Which arm should I use to take my blood pressure?
TROUBLESHOOTING ...................................................................................... 15
SPECIFICATIONS ............................................................................................. 16
WARRANTY INFORMATION ......................................................................... 21
Thank you for selecting the Hopkins® Digital IMPACT Arm BP (Item #526018). This monitor fea­tures blood pressure measurement, pulse rate measurement, and an auto-save function.
Readings taken by the #526018 are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method.
This manual contains important safety and care information, and provides step by step instructions for using this product. Read this manual thoroughly before using this product
CAUTION
!
This device is for adult use only.
This device is intended for non-invasive measuring and monitoring of arterial blood pressure. It is not intended for use on extremities other than the arm or for functions other than obtaining a blood pressure measurement and pulse rate.
Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood pressure. Do not begin or end medical treatment based solely on self-monitoring, consult a physician for treatment advice.
If you are taking medication, consult your physician to determine the most appropriate time to measure your blood pressure. Never change a prescribed medication without consulting your physician. This unit is not suitable for continuous monitoring during medical emergencies or operations.
If the cuff pressure exceeds (300 mmHg), the unit will automatically deate. Should the cuff not deate when
pressures exceeds (300 mmHg), detach the cuff from the arm and press the START/ST OP button to stop
ination.
To avoid measurement errors, carefully read this manual before using this device.
This device is not AP/APG equipment and not suitable for use in the presence of a ammable anesthetic
mixture with air, or with oxygen, or nitrous oxide.
Please avoid strong electromagnetic elds as they may interfere with or damage this device.
Always check this device to make sure that it is in proper working condition before use.
INTRODUCTIONINTRODUCTION
32
LCD Display
SYMBOL DESCRIPTION EXPLANATION
SYS
Systolic BP High pressure result
DIA
Diastolic BP Low pressure result
Pul/min
Pulse Pulse per Minute (Beats per Minute)
Monitor components
SYMBOL DESCRIPTION
Systolic blood pressure High pressure result Diastolic blood pressure Low pressure result
EXPLANATION
Pulse Pulse/minute Deflating CUFF air is exhausting of deflating Time (hour:minute) Currently time
Memory If “M” shows, the displayed measurement
values is from the memory. For instructions,refer to Page 9
mmHg Measurement Unit of the blood pressure
(1mmHg=0.133kPa)
kPa Measurement Unit of the blood pressure
(1kPa=7.5mmHg)
Low battery Batteries are low and need to be replaced
CUFF
AIR HOSE
AIR CONNECTOR PLUG
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
BATTERY COMPARTMENT
List
Component list of pressure measuring system 1 Cuff 2 Micro Control Unit 3 Amplifier 4 Air pipe 5 Pump 6 Valve
Low Battery Batteries are low and need to be replaced
mmHg
Millimeters Mercury Measurement unit of blood pressure
Monitor components
SYMBOL DESCRIPTION
Systolic blood pressure High pressure result Diastolic blood pressure Low pressure result
EXPLANATION
Pulse Pulse/minute Deflating CUFF air is exhausting of deflating Time (hour:minute) Currently time
Memory If “M” shows, the displayed measurement
values is from the memory.
CUFF
AIR HOSE
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
BATTERY COMPARTMENT
Component list of
Memory
If “M” shows, the displayed measurement value is from the memory of recorded measurements (See page 9 for instructions)
Inating
The unit is inating with air to obtain the
required level of pressure
Deating
The unit is deating and the air is being
exhausted from the cuff
Monitor components
LCD display signal
SYMBOL DESCRIPTION
Systolic blood pressure High pressure result Diastolic blood pressure Low pressure result
EXPLANATION
Pulse Pulse/minute Deflating CUFF air is exhausting of deflating Time (hour:minute) Currently time
CUFF
AIR HOSE
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
Time (hour:minute) The current time (set by user)
Monitor components
Systolic blood pressure High pressure result Diastolic blood pressure Low pressure result
EXPLANATION
Pulse Pulse/minute Deflating CUFF air is exhausting of deflating Time (hour:minute) Currently time
Memory If “M” shows, the displayed measurement
values is from the memory. For instructions,refer to Page 9
mmHg Measurement Unit of the blood pressure
(1mmHg=0.133kPa)
kPa Measurement Unit of the blood pressure
(1kPa=7.5mmHg)
Low battery Batteries are low and need to be replaced Shocking remainding Shocking will result in inaccurate
CUFF
AIR HOSE
AIR CONNECTOR PLUG
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
BATTERY COMPARTMENT
List
1.Blood Pressure Monitor
2.Cuff (Type B applied part) (AC2232-01)
Component list of pressure measuring system 1 Cuff 2 Micro Control Unit 3 Amplifier 4 Air pipe 5 Pump 6 Valve
Shock Warning
Inaccurate readings may occur if unit is hit, dropped, shaken, or otherwise shocked
Monitor components
Systolic blood pressure High pressure result Diastolic blood pressure Low pressure result
EXPLANATION
Pulse Pulse/minute Deflating CUFF air is exhausting of deflating Time (hour:minute) Currently time
Memory If “M” shows, the displayed measurement
values is from the memory. For instructions,refer to Page 9
mmHg Measurement Unit of the blood pressure
(1mmHg=0.133kPa)
kPa Measurement Unit of the blood pressure
(1kPa=7.5mmHg)
Low battery Batteries are low and need to be replaced Shocking remainding Shocking will result in inaccurate
Average The average of blood pressure
CUFF
AIR HOSE
AIR CONNECTOR PLUG
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
BATTERY COMPARTMENT
List
1.Blood Pressure Monitor (TMB-1018-A)
2.Cuff (Type B applied part) (AC2232-01)
Component list of pressure measuring system 1 Cuff 2 Micro Control Unit 3 Amplifier 4 Air pipe 5 Pump 6 Valve
Average The average of the blood pressure measurements
Monitor components
Systolic blood pressure High pressure result Diastolic blood pressure Low pressure result
EXPLANATION
Pulse Pulse/minute Deflating CUFF air is exhausting of deflating Time (hour:minute) Currently time
Memory If “M” shows, the displayed measurement
values is from the memory. For instructions,refer to Page 9
mmHg Measurement Unit of the blood pressure
(1mmHg=0.133kPa)
kPa Measurement Unit of the blood pressure
(1kPa=7.5mmHg)
Low battery Batteries are low and need to be replaced Shocking remainding Shocking will result in inaccurate
Average The average of blood pressure Recalling The erecords will be showed
CUFF
AIR HOSE
AIR CONNECTOR PLUG
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
BATTERY COMPARTMENT
List
1.Blood Pressure Monitor (TMB-1018-A)
2.Cuff (Type B applied part) (AC2232-01)
Component list of pressure measuring system 1 Cuff 2 Micro Control Unit 3 Amplifier 4 Air pipe 5 Pump 6 Valve
Recalling The stored records are being retrieved
Monitor components
Systolic blood pressure High pressure result Diastolic blood pressure Low pressure result
EXPLANATION
Pulse Pulse/minute Deflating CUFF air is exhausting of deflating Time (hour:minute) Currently time
Memory If “M” shows, the displayed measurement
values is from the memory. For instructions,refer to Page 9
mmHg Measurement Unit of the blood pressure
(1mmHg=0.133kPa)
kPa Measurement Unit of the blood pressure
(1kPa=7.5mmHg)
Low battery Batteries are low and need to be replaced Shocking remainding Shocking will result in inaccurate
Average The average of blood pressure Recalling The erecords will be showed
Irregular heartbeat
Arrhythmia
CUFF
AIR HOSE
AIR CONNECTOR PLUG
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
BATTERY COMPARTMENT
List
1.Blood Pressure Monitor (TMB-1018-A)
2.Cuff (Type B applied part) (AC2232-01)
Component list of pressure measuring system 1 Cuff 2 Micro Control Unit 3 Amplifier 4 Air pipe 5 Pump 6 Valve
Irregular Heartbeat Irregular heartbeat (arrhythmia) detected
Monitor components
Systolic blood pressure High pressure result Diastolic blood pressure Low pressure result
EXPLANATION
Pulse Pulse/minute Deflating CUFF air is exhausting of deflating Time (hour:minute) Currently time
Memory If “M” shows, the displayed measurement
values is from the memory. For instructions,refer to Page 9
mmHg Measurement Unit of the blood pressure
(1mmHg=0.133kPa)
kPa Measurement Unit of the blood pressure
(1kPa=7.5mmHg)
Low battery Batteries are low and need to be replaced Shocking remainding Shocking will result in inaccurate
Average The average of blood pressure Recalling The erecords will be showed
Irregular heartbeat
Arrhythmia Grade The grade of the blood pressure.
CUFF
AIR HOSE
AIR CONNECTOR PLUG
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
BATTERY COMPARTMENT
List
1.Blood Pressure Monitor (TMB-1018-A)
4.User manual3. 4*AAA alkaline batteries
2.Cuff (Type B applied part) (AC2232-01)
Component list of pressure measuring system 1 Cuff 2 Micro Control Unit 3 Amplifier 4 Air pipe 5 Pump 6 Valve
BP Category Indicator
Blood pressure level reading (See page 13 for instructions)
Monitor Components
CUFF
AIR HOSE
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
BATTERY COMPARTMENT
2.Cuff (Type B applied part) (AC2232-01)
Component list of pressure measuring system 1 Cuff 2 Micro Control Unit 3 Amplifier 4 Air pipe 5 Pump 6 Valve
CUFF
AIR HOSE
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
BATTERY COMPARTMENT
4.User manual3. 4*AAA alkaline batteries
2.Cuff (Type B applied part) (AC2232-01)
Component list of pressure measuring system 1 Cuff 2 Micro Control Unit 3 Amplifier 4 Air pipe 5 Pump 6 Valve
AAA
AAA
AAA
AAA
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
Component list of pressure measuring system 1 Cuff 2 Micro Control Unit 3 Amplifier 4 Air pipe 5 Pump 6 Valve
Parts Included
#526018 Hopkins® IMPACT
Digital BP Monitor
Adult Digital Cuff
Cuff Circumference: 8.5” - 12.5”
(22cm - 32cm)
Four AAA Batteries
Monitor components
CUFF
AIR HOSE
Monitor components
CUFF
AIR HOSE
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
BP CATEGORY
INDICATOR
Monitor components
CUFF
AIR HOSE
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
BATTERY COMPARTMENT
Component list of
CUFF TUBING
AIR CONNECTOR
PLUG
CUFF
AIR HOSE
LCD DISPLAY
DC POWER SOCKET
MEM/DOWN BUTTON
START/OFF BUTTON
SET/UP BUTTON
BATTERY COMPARTMENT
Component list of pressure measuring system 1 Cuff 2 Micro Control Unit 3 Amplifier 4 Air pipe 5 Pump
1. Slide off the battery cover on the back of the unit.
2. Install the batteries by matching the correct polarity, as shown above.
Always use the correct battery type (4 alkaline AAA size)
3. Replace the battery cover.
BEFORE YOU STARTBEFORE YOU START
54
AC adaptor
BEFORE YOU STARTBEFORE YOU START
Setting Date, time and measurement unit
It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory. (year :2000—2050,time:24 H)
1.When the unit is off, hold pressing “SET” for 3 seconds to enter the mode for year setting.
2.Press the “MEM” to change the [YEAR].
START STOP
MEM SET
Installing the Batteries
• If this device will be stored for an extended period of time, please remove the batteries in order to avoid corrosion.
• Used batteries are harmful to the environment, so please dispose of appropriately. Remove the old batteries from the device and follow your local recycling guidelines
• Battery corrosion is not covered under your three year limited warranty.
CAUTION
!
Replace the batteries when any of the following occurs:
• The shows
• The display dims
• The display does not light up
AC adaptor
BEFORE YOU STARTBEFORE YOU START
Setting Date, time and measurement unit
It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory. (year :2000—2050,time:24 H)
1.When the unit is off, hold pressing “SET” for 3 seconds to enter the mode for year setting.
2.Press the “MEM” to change the [YEAR].
3.When you get the
START STOP
MEM SET
START STOP
MEM SET
Setting the Date and Time
It is important to set the clock before using your blood pressure monitor so that a time stamp can be assigned to each record stored in the memory. (The year can be set to anywhere between 2000 and 2050.)
1. When the unit is off,
press and hold the “SET” button for 3 seconds to enter the mode settings.
2. Press the “MEM” button to advance to the current [YEAR].
3. Press “SET” when you have reached the correct year to save your choice and advance to the next step.
BEFORE YOU STARTBEFORE YOU START
START STOP
MEM SET
START STOP
MEM SET
START STOP
MEM SET