Honeywell RAE Systems NeutronRAE II User Manual

Page 1
NeutronRAE II
Personal Radiation Monitor
User’s Guide
www.raesystems.com
Revision B, January 2013
P/N 047-4204-100
Page 2
Read Before Operating
This manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturers instructions.
CAUTION!
!
To reduce the risk of electric shock, turn off power and remove batteries before removing the monitor cover. Never operate this monitor while the cover is removed. Remove monitor cover and sensor modules only in an area known to be non-hazardous.
WARNING
!
To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing.
WARNING
!
Do not mix old batteries with used batteries or mix batteries from different manufacturers.
WARNING
!
Substitution of components may impair performance.
This product complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including intereference that may cause undesired operation.
Page 3
Warnings
!
!
!
!
!
For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read and understand the user manual completely before operating or servicing.
Computer Interface
Do not transfer data by means of the Bluetooth® interface in hazardous atmospheres.
Static Hazard
Clean only with a damp cloth.
Long-Term Storage
Reliable performance of this radiation detector is based upon regular usage. For long-term storage, the battery should be disconnected. Preparation for use after long storage requires installation of the batteries and a warm-up period of at least 10 minutes for the sensors to equilibrate. The user should recognize that sensor life is based upon the purchase date.
www.raesystems.com
1
Page 4
Avertissements
!
!
!
!
!
Pour des raisons de sécurité, cet équipement doit être utilisé, entretenu et réparé uniquement par un personnel qualifié. Étudier le manuel d’instructions en entier avant d’utiliser, d’entretenir ou de réparer l’équipement.
Câble de Computer
Ne transférez pas les données au moyen de l’interface de bluetooth en atmosphères dangereuses.
Danger Risque D’origine Electrostatique
Nettoyer uniquement avec un chiffon humide.
Stockage à Long Term
Le fonctionnement durable de ce détecteur de rayonnement est conditionné par une utilisation régulière de celui ci. Lors d’un stockage à long terme, la batterie doit être déconnectée. Le rédémarrage aprés une longue période d’arrêt, nécessite la réinstallation de la batterie, et une période de chauffage de 10 mn afin que les capteurs se mettent à l’équilibre. L’utilisateur doit être conscient que la durée de vie indiquée pour le capteur démarre à sa date d’achat.
2
NeutronRAE II
Page 5
Contents
NeutronRAE II Features .......................................4
General Information ...........................................5
Inserting & Replacing Batteries ..................................6
User Interface ...............................................7
User Interface Icons ...........................................8
Turning the NeutronRAE II On and Off ...........................8
Flipping the Screen ..........................................12
Operation ..................................................13
Normal Operating Mode ...................................13
NORM ................................................14
CAL ..................................................14
PEAK .................................................15
MIN ..................................................15
DOSE .................................................16
STAT .................................................16
COMM ................................................17
TEST .................................................18
Programming Mode .......................................20
Alarm Set ..............................................21
Datalog Set .............................................22
Monitor Set ............................................23
Back Lite .............................................24
Buzz & Lite ...........................................25
Vibrate mode ..........................................25
Temp Unit ............................................25
Gamma Unit ..........................................26
Change Date ..........................................26
Change Time ..........................................26
Quit .................................................27
Default Settings and Parameter Ranges ..........................27
Specications ...............................................29
Limited Warranty ............................................34
www.raesystems.com
3
Page 6
LED alarm
LCD showing operation parameters and radiation measurements
LED alarm
MODE key for selecting mode and function
Buzzer alarm
Battery compartment
4
Front View
Battery compartment cover screw
NeutronRAE II
SET key for selecting or setting parameters
Sensor location
Belt clip
Rear/Side View
Page 7
General Information
The NeutronRAE II is a rapid detector
of γ-ray (gamma) and neutron radia-
tion sources.
sensitivity, it can alert rst responders
to the presence of a radiation threat well before they might be exposed to health-threatening levels. personal radiation detector features a
loud audible alarm, big, bright ashing
LEDs, and a vibration alarm. For stealth operations, law enforcement personnel can disable each alarm separately.
NeutronRAE II’s water-resistant design makes for reliable operation in wet environments. Its sensitive Cesium Iodide (CsI) and Lithium Iodide (LiI) scintillators provide fast response to radiological threats.
The NeutronRAE II accumulates dosage. Stored dosage data can be cumulative or cleared and reset for each use period.
Because of its high
This
www.raesystems.com
5
Page 8
NeutronRAE II Features
Inserting & Replacing Batteries
NeutronRAE II uses two AA alkaline batteries as its power source (use only Duracell MN1500 or Energizer E91). To change the batteries:
1. Use a coin or screwdriver to open the cover on the side of the unit. Turn the screw counterclockwise to loosen the screw, and then tilt the cover off.
2. Insert batteries into the compartment according to the legend on the rear of the NeutronRAE II, making sure the batteries’ polarity is correct.
3. Replace the cover and turn the screw clockwise to secure the cover.
6
NeutronRAE II
Page 9
User Interface
NeutronRAE II’s user interface consists of the display, LEDs, alarm transducer and two buttons, labeled MODE and SET. The LCD display provides visual feedback that includes time, functional mode, battery condition, and gamma and neutron radiation measurements.
LCD display
Gamma alarm information
Current time
Functional mode
Neutron reading
sign
Gamma radiation
measurement
MODE and SET switches
M
O
D
E
www.raesystems.com
Neutron alarm information
Datalog indication
Neutron radiation
T
E
S
Battery status
Unit of measure (gamma)
Unit of measure (neutron)
measurement
7
Page 10
User Interface Icons
Icon Indication
Battery Voltage low Battery Low alarm triggered
Battery fully charged
Low Gamma alarm triggered
High Gamma alarm triggered
Gamma Over-Range (reading over 20 mR/h)
Gamma Overload
Low Neutron alarm triggered
Neutron Over-Range (reading over 100 cps)
Neutron Overload
Bluetooth® communication active
Datalogging active (ashing)
8
NeutronRAE II
Page 11
Turning NeutronRAE II On and Off
Turning NeutronRAE II On
Press and hold the MODE button for 3 seconds. As NeutronRAE II starts up, the following occur:
1. A long beep sounds (if the buzzer is set to On), and
unit information and the current rmware version are
displayed. The LEDs and vibrator undergo a self-test.
2. Unit self-check. The time and battery status are displayed:
During the 20-second self-check, the detector checks
that the gamma and neutron sensors are functioning properly. If so, the messages “Gamma sensor OK!”
and “Neutron sensor OK!” are displayed.
www.raesystems.com
9
Page 12
or
If the sensor is damaged or not functioning properly,
the message “Gamma sensor failed” or “All sensors failed” is displayed, turn the unit off and contact your distributor or the RAE Systems service department.
or
or
Important! If your NeutronRAE II displays
“Gamma sensor failed,” “Neutron sensor failed,” or “All sensors failed!” contact your distributor or RAE Systems service at 408-952-8200. Do not attempt to use or repair the unit. There are no user-serviceable parts in the NeutronRAE II.
3. The datalog type is displayed, followed by the gamma alarm mode. These parameters can be changed in
10
NeutronRAE II
Page 13
Programming Mode, described later in this guide.
or
or
4.
Gamma background calibration reading (only if the NeutronRAE II is in Search Mode). The NeutronRAE II performs a 36-second background reading countdown.
When background calibration reading is complete, datalogging starts (if Continuous Datalogging is enabled) and the unit is in Normal Operating Mode. The display looks like this:
www.raesystems.com
11
Page 14
Turning NeutronRAE II Off
Press the MODE button and hold for 5 seconds. The detector counts down 5 seconds and then turns itself off. Do not release the MODE button until the unit counts down to zero
and the display reads “Save cong...” and “Stop unit...”
Flipping the Screen
The NeutronRAE II is easy to read, whether held in the hand
or clipped to a belt. To ip the screen, press the SET button
and hold it down for 3 seconds. When the image inverts, release the button.
To ip the display again, press and hold SET for 3 seconds.
12
NeutronRAE II
Page 15
Operation
The NeutronRAE II has two modes: Normal Operating Mode and Programming Mode. Normal Operating Mode is the default; see Programming Mode, on page 20, for details on entering and using Programming Mode.
Normal Operating Mode
In Normal Operating Mode, the NeutronRAE II detects gamma and neutron radiation and accumulates gamma radiation dosage data. In Normal Operating Mode, pressing the MODE button sequentially steps the unit through the following functions:
NORM Normal Operating Mode CAL PEAK Maximum radiation levels detected since last cleared MIN Minimum radiation levels detected since last cleared DOSE TIME Time, date, and time unit has been on STAT Detector status (battery voltage and internal temp.) COMM TEST Self test (if Diagnostic Flag is set during start-up)
Each function remains active for 60 seconds before automatically returning to the Normal Operating Mode. Pressing the MODE button changes the selection from one function to the next one in the sequence.
Gamma background calibration (only in Search Mode)
Accumulated gamma radiation dosage since last cleared
Open BlueTooth communication with computer
www.raesystems.com
13
Page 16
NORM
CAL
Normal Operating Mode of the NeutronRAE II.
Displays a measurement of the ambient gamma and neutron radiation. It can display gamma in counts per second (cps) or µR/h or µSv/h, depending on how the detector is set up (see Programming Mode) and neutron in counts per second. Switch back and forth between gamma units by pressing the SET button.
Press the MODE button to step to the next function.
Background calibration. This option is only active if the NeutronRAE II in Search Mode. This option is skipped if the detector is in Safety Mode.
Press the SET button to have the detector update the background radiation reading.
14
Press the MODE button to step to the next function.
NeutronRAE II
Page 17
PEAK
MIN
The maximum radiation levels detected since last cleared.
Press the SET button twice to clear the PEAK values. If you press SET once but do not want to clear the readings, press MODE to return to the PEAK readings. The PEAK values are also cleared when the detector is turned off.
Press the MODE button to step to the next function.
The minimum radiation levels detected since last cleared.
Press the SET button twice to clear the MIN values. If you press SET once but do not want to clear the readings, press MODE to return to the MIN readings. The MIN values are also cleared when the detector is turned off.
www.raesystems.com
15
Page 18
DOSE
TIME
Press the MODE button to step to the next function.
The accumulated gamma radiation dosage since last cleared.
Press the SET button twice to clear the DOSE value. If you press SET once but do not want to clear the DOSE reading, press MODE to return to the DOSE reading.
Press the MODE button to step to the next function.
Time, date, and run time. Date is displayed in YYYY-MM-DD format (can be set in Programming Mode). Time is displayed in HH:MM:SS format (can be set in Programming Mode).
On Run time from when the detector was turned
on (shown in hours and minutes, 0000:00)
16
NeutronRAE II
Page 19
STAT
Press the MODE button to step to the next function.
Detector status.
BattV Battery voltage Temp Internal temperature (can be displayed in °C
or °F – see Programming Mode)
If the internal temperature exceeds the
maximum operating temperature, the unit will alarm and display:
Cool the unit down to resume normal operation.
Press the MODE button to step to the next function.
www.raesystems.com
17
Page 20
COMM
TEST
Communicate with a computer using Bluetooth® radio. Press SET to open the connection on the NeutronRAE II unit end. Use the ProRAE Studio II Software on your computer to download datalogs.
Self test. This consists of tests to make sure all alert functions are working properly. This function is only enabled if the Diagnostic Flag was set at start-up. To set the Diagnostic Flag, press and hold SET and MODE simultaneously when turning the unit on, instead of pressing MODE only.
Press the SET button to accept and to initiate testing.
The Buzzer is tested. Press the SET button to start the next test.
18
NeutronRAE II
Page 21
The Vibrator is tested. Press the SET button to start the next test.
The LEDs are tested. Press the SET button to start the next test.
The back light is tested. Press the SET button to start the next test.
The Bluetooth® radio is tested.
Press the SET button to start the next test.
www.raesystems.com
19
Page 22
The Buzzer in Alarm is tested.
Press the SET button to nish the TEST functions.
This returns NeutronRAE II to Normal Operating Mode. During any test, if the SET button is not pressed within 60 seconds, the unit will automatically move on to the next test.
Programming Mode
Programming Mode is used to change alarm settings and detector settings (units of measure, time, etc.).
Note: Changes to some parameters cause the detector to automatically restart or recalibrate.
From Normal Operating Mode, simultaneously press the MODE button and SET button, and hold them down for 3 seconds. NeutronRAE II enters the Programming Mode and displays the Main Menu:
20
NeutronRAE II
Page 23
Press the MODE button to step through the menu of parameters.
Press the SET button to select a parameter.
Alarm Set
Select Alarm Set to set the alarm mode or type and the detection limits for the alarm:
Parameter Used In
Alarm Type Select Search or Safety Mode
Gamma CF Search Mode
Gamma Hi Safety Mode
Gamma Low Safety Mode
Quit
The Gamma CF (Correction Factor) is used to set the Search Mode alarms at a calculated threshold above background. The alarms are set according to:
Search Mode Alarm = CF*σbk
where σbk is the standard deviation of the stored
background reading.
www.raesystems.com
21
Page 24
Note: The default value is 005.8 for the Gamma CF. In Safety Mode, there are two absolute Gamma alarms: Gamma Low and Gamma Hi. These are set in µR/h, where 1µR/h is equal to 0.01µSv/h. The Neutron alarm remains in Search Mode at all times.
Press the MODE button to step through the digits from left to right. Use the SET button to increment each digit (from 0 through 9). Select Save to save changes or Quit to discard changes.
To exit Alarm Set, step through the options using the MODE button until you see Quit. Press the SET button to exit Alarm Set, and then step through the Main Menu until you see Quit. Press SET to exit Program mode and return to Normal Operating Mode.
Datalog Set
Select Datalog Set to change any of the following datalog parameters:
Clear Data Log interval Datalog Type Quit
22
NeutronRAE II
Page 25
Clear Data: Press the SET button to clear the datalog. Press the MODE button to return to the Datalog Set Menu.
Datalog Type: Set the type of datalogging. Choices are Continuous or Event-driven. Choose Continuous to log data at a set interval at all times (except when the unit is Programming Mode). Choose Event­driven to log data only when the dose rate readings increase or the unit goes into alarm.
Log Interval: Set the time interval between datalog points. You may set any time interval between 1 and 3600 seconds. Press the MODE button to step through the digits from left to right. Use the SET button to increment each digit (from 0 to 9). Select Save to save changes or Quit to discard changes.
Quit: To exit Datalog Set, step through the options using the MODE button until Quit is selected. Then press the SET button.
www.raesystems.com
23
Page 26
Monitor Set
Select Monitor Set to change any of the following parameters:
Press the MODE button to step through these options.
Select an option using the SET button. Once you have selected a menu option, step through the options using the MODE button, and select using the SET button.
Back Lite
Set the behavior of the display backlight. Choices are Automatic (a photosensor turns on the light in dark locations), Manual (light stays on for 15 seconds when you tap any button in any operating mode), or Off.
Back Lite Buzz & Lite Vibrate Mode Temp Unit Gamma Unit Change Date Change Time Quit
24
NeutronRAE II
Page 27
Buzz & Lite
Set the behavior of the audible and visible alarms. Choices are Both on (LEDs and audible alarm), Light only (LEDs only), or Both off.
Note: You cannot turn off all of the alarm alerts at the same time. If you select Both Off when the Vibration alarm is Off, “Invalid Setting” is displayed and the previous Buzz & Lite setting is restored.
Vibrate Mode
Set the vibration alarm. Choices are On or Off.
Note: You cannot turn off all of the alarm alerts at the same time. If you select Off when the Buzz & Lite is set to Both Off, “Invalid Setting” will be the previous Vibrate setting will be restored.
Temp Unit
Set the temperature unit. Choices are Fahrenheit and Celsius. Press MODE to select Fahrenheit or Celsius. Then press SET to save your choice and exit.
displayed and
www.raesystems.com
25
Page 28
Gamma Unit
Set the unit of measure (of gamma radiation). Choices are µR/h or µSv/h. Press MODE to select your choice, and then SET to save your choice and exit.
Change Date
Set the date (mm/dd/yyyy). Press MODE to step through month, date, and year, and press SET to increment through the digits (holding the SET button continuously speeds through the numbers).
Press MODE to select Save, and press SET to save your date changes. Otherwise, press MODE again to select Quit and press SET to exit.
Change Time
Set the time (hh/mm/ss). Press MODE to step through hours, minutes, and seconds, and press SET to increment through the digits (holding the SET
26
NeutronRAE II
Page 29
Quit
button continuously speeds through the numbers). Note: The clock operates in 24-hour mode only.
Press MODE to select Save, and press SET to save your time changes. Otherwise, press MODE again and select Quit. Press SET to exit.
To exit Monitor Set, step through the options of Monitor Set until you see Quit. Press the SET button to exit Monitor Set. Then step through Main Menu selections using the MODE button until you see Quit. Press SET to exit Program Mode and return to Normal Operating Mode.
www.raesystems.com
27
Page 30
NeutronRAE II Default Settings & Parameter Ranges
Main Menu
Alarm Set
Datalog Set
* Correction Factor
Sub Menu
Alarm Type Search Search/Safety
Gamma CF* 5.8 1.0-9.9
Gamma Hi 200µR/h 0 to 20mR/h Gamma Low 50µR/h 0 to 20mR/h
Clear Data N/A N/A
Change Type
Log Interval 60 sec 1-3600 sec
Default Settings
Continuous
Data Range
Continuous or Event-driven
28
NeutronRAE II
Page 31
Monitor Set
Automatic
Back Lite Manual
Buzz & Lite Both On
Vibrate On
Temp Unit Celsius
Gamma Unit µR/h
Change Date
Change Time HH/MM/SS
Manual Off
Both On Light Only
Both Off On
Off Fahrenheit Celsius µR/h
µSv/h
MM/DD/ YYYY
www.raesystems.com
29
Page 32
NeutronRAE II Specications*
Size 4.92" x 2.68" x 1.38" (125 x 68 x 35 mm) Weight Detectors
Battery
Operating Period
Display
Key Pad Direct Readout
*Specications are subject to change.
10 oz. (283g)
Gamma: 1cc CsI(TI)/Photodiode Neutron: 1cc LiI(Eu)/Photodiode
2 AA alkaline batteries easily accessible with a coin
500 hours
Graphic LCD display with 1.2" x 0.75" (30.5 mm x 19 mm) viewable area can
be ipped for view by wearer or by
others
2 operation/program buttons
Gamma dose rate (cps, µR/h, µSv/h) Neutron radiation intensity (cps) Gamma/Neutron peak Gamma/Neutron minimum Gamma dose Battery status Current time and date Current time since startup Internal temperature
30
NeutronRAE II
Page 33
Backlight Calibration
Automatic, manual, or off
Periodic functional test recommended using a
137
Cs check source. Factory calibration recommended every two years
Alarm Setting Programmable alarm sensitivity with
background compensation to minimize false alarms, absolute safety alarms for gamma
Alarm Modes
Search Mode: Alarm thresholds set at
a calculated value above background reference reading Safety Mode: User-set high and low absolute alarm levels (gamma only)
Alarm Alerts
Loud 85dB @ 12" (30 cm) for noisy environments
Built-in vibration alarm
Highly visible LED lights on both sides of LCD graphic display
Datalog
10,000 datapoints (20 days @ 180-sec. intervals), wraparound log
Specications continue on next page
www.raesystems.com
31
Page 34
Specications (continued)
Datalog Modes
Continuous: Logs data continuously Event Driven: Begins logging on alarm
Log Interval Communica-
User-programmable (1 to 3600 sec.)
Built-in Bluetooth® radio
tions Ergonomics Nonslip rubber housing with grippable
ridges securely ts hand or glove
Energy Range
Gamma: 0.06 to 3.0 MeV Neutron: thermal to 14 MeV
Sensitivity
Gamma: More than 0.3 cps per µR/hr
(>30 cps per µSv/hr) Neutron: 1 to 2 cps per 2.5 neutrons/
2
Dose Equivalent
second/cm Gamma:
1 µR to 20 mR/h (0.01 to 200 µSv/h)
Neutron: 1 to 100 cps
Rate (DER) range for
Accuracy of DER for
Dosage Range
137
137
Cs
±20%
Cs
1 µR to 999 R (gamma only)
Temperature -20° C to 50° C (-4° F to 122° F) Humidity 0% to 95% (non-condensing)
32
NeutronRAE II
Page 35
Shock Resistance
IP Rating Attachments Rugged metal belt clip and wrist strap
Passes drop test from 1.5m (59")
IP67
Proper Battery Disposal
The Waste Electrical and Electronic Equipment (WEEE) directive (2002/96/ EC) is intended to promote recycling of electrical and electronic equipment and their components at end of life. This symbol (crossed-out wheeled bin) indicates separate collection of waste electrical and electronic equipment in the EU countries. This product may contain one or more Nickel-metal hydride (NiMH), Lithium-ion, or Alkaline
batteries. Specic battery information is given in this user
guide. Batteries must be recycled or disposed of properly.
At the end of its life, this product must undergo separate collection and recycling from general or household waste. Please use the return and collection system available in your country for the disposal of this product.
www.raesystems.com
33
Page 36
Limited Warranty
RAE Systems Inc. warrants NeutronRAE II to be free from defects in materials and workmanship for a period of 1 year. This warranty is expressly limited to the original owner who purchases the equipment directly from RAE Systems or from an authorized RAE Systems reseller. To validate this warranty the original Warranty & Registration Card supplied with the product must be completed and returned to RAE Systems, or the product registered online via the RAE Systems, Inc. online registration system, within 30 days of purchase.
To maintain this limited warranty, the product must be operated, calibrated and maintained in accordance with the Operation and Maintenance Manual supplied with the product. Abuse, mechanical damage, alteration, and/or repair procedures not made in accordance with the Operation and Maintenance Manual voids the RAE Systems Standard Limited Warranty.
The obligation of RAE Systems under this limited warranty is limited to the repair or replacement of components deemed by the RAE Systems Instrument Service Department to have been defective under the scope of this Standard Limited Warranty. To receive consideration for warranty repair or replacement procedures, products must be returned to RAE Systems at its manufacturing location in San Jose, California, USA, with transportation and shipping charges prepaid. It is
34
NeutronRAE II
Page 37
necessary to obtain a return authorization number from RAE Systems prior to shipment.
This limited warranty is expressly in lieu of any and all representations, express or implied, including but not limited
to, the warranty of tness for a particular purpose. RAE
Systems will not be liable for loss or damage of any kind connected to the use of its products or failure of its products to function or operate properly.
www.raesystems.com
35
Page 38
RAE Systems Contacts
RAE Systems by Honeywell World Headquarters
3775 N. First St. San Jose, CA 95134-1708 USA Phone: 408.952.8200 Fax: 408.952.8480
E-mail: customerserv@raesystems.com Web Site: www.raesystems.com
RAE Systems Technical Support
Monday through Friday, 7:00AM to 5:00PM Pacic Time
+1.888.723.4800 (toll-free) email: tech@raesystems.com
RAE Systems Europe ApS
Ørestads Boulevard 69 DK-2300 Copenhagen, DENMARK Phone: +45.8652.5155 Fax: +45.8652.5177
RAE Systems (Hong Kong) Ltd.
Room 8, 6/F, Hong Leong Plaza 33 Lok Yip Road
Fanling, N.T. HONG KONG
Phone: +852.2669.0828 Fax: +852.2669.0803
Life-critical after-hours support is +1.408.752.0723 select option 9
available
P/N: 047-4204-100 REV B
Loading...