The Honeywell® RAEGuard 2 PID is a fixed photoionization detector (PID) that measures
a broad range of Volatile Organic Compounds (VOCs). The RAEGuard 2 PID monitor
operates in flow through mode. Housed in an explosion-proof stainless-steel enclosure,
the RAEGuard 2 PID can be used in most environments, including hazardous
environment applications for measuring toxic gases. Therefore, correct installation and
maintenance is required to ensure effectiveness as a VOC monitor.
This technical note describes guidelines for installation, commissioning, testing,
calibration and maintenance of the RAEGuard 2 PID and the factors that need to be
considered.
INSTALLATION
This instrument can be installed as a stand-alone unit or as part of a multi-point VOC
monitoring system.
Preface
1998-1001 Rev 1 11/16
It is vital to understand the gas and conditions to be measured. Many factors can impact the
performance of an active sampling system, including dust, water condensation, gas
condensation, and pressure within a duct.
Dust
Unless the particle size is less than 1 micron, the only impact on the performance is the filter
replacement interval.
Water condensation
The RAEGuard 2 PID system is able to operate within a relative humidity range of 0 to 95%,
non-condensing, without distortion of the actual VOC concentration reading. Nevertheless,
water vapors are very often present within VOC samples, especially in many duct sampling
applications. The sample temperature very often is high enough, and is higher than at the
measurement point. On the way from the sampling point to the instrument, the sample gas
is cooling. As temperature decreases, relative humidity (RH) of the water vapors in the VOC
samples increases and may reach 100% or the dew point. If the temperature at the measuring
point is lower than the dew point, water vapor may condense in the tubing, on the PID lamp,
and on the PID electrodes. The amount of condensed water may be significant and may affect
the accuracy of the measurement.
For example
If the sampling temperature is 122° F (50° C) and 40% RH, and at the measurement point the
temperature is 77° F (25° C), about 10 g of water may condense from 1m3 of the sampling air
within two (2) to three (3) days.
1
Technical Note
Installation and Maintenance
of a RAEGuard 2 PID
Presence of water must be considered at the lowest ambient temperature and the highest RH
at the highest temperature in the sample line. This calculation must be performed when
sampling from ducts (additional sampling equipment may be required). Honeywell
recommends using a water-collecting filter; housing part number 490-0163-000 and filter
replacement 490-0164-010 (pack of 10).
Figure 1. Water-collecting filter
Note: Filter housing and filter replacement elements are currently available only in North
America. Contact Honeywell for more information. The filter is installed between the sample
source and the instrument, at a location where the temperature does not exceed the
instrument’s rated operating temperature.
1998-1001 Rev 1 11/16
Target and background gases condensation
Many toxic gases have a low vapor pressure. Therefore, gas may become a liquid or a crystal
even at very low concentrations if the temperature at the measurement point is lower than
at the sampling point. This factor must be considered before selecting an active sampling
system.
The following example illustrates the fundamentals of the vapor pressure issue:
Phenol
The Honeywell® RAEGuard 2 PID can be used in phenol factories all over the world. Phenol
can crystallize from the gas phase to the solid phase when its vapor pressure in the gas state
is higher than that of the sublimation vapor of the solid phenol. This process depends greatly
on the temperature variations, since vapor pressures depend on them.
The approximate maximum Phenol gas concentrations in air at different temperatures are as
follows:
• At 122° F (50° C): The maximum concentration of phenol is around 3600 ppm.
• At 77° F (25° C): The maximum concentration of phenol is around 440 ppm.
Pressure
Until pressure variation is within a specified range of 90 to 110 kPa, a monitoring instrument
gives a correct reading. Make sure the pressure in the sampling duct does not exceed these
values. Slightly exceeding them is acceptable, but the reading may be distorted. That
distortion does not relate to the PID itself, which is rather stable in terms of pressure variation,
but it does relate to sampling gas density variation.
2
Technical Note
Installation and Maintenance
of a RAEGuard 2 PID
Outdoor Installation
When installing a RAEGuard 2 PID outdoors, use a rain shield to protect the monitor. It is
necessary to prevent water from being drawn into the sample stream.
There is a special rain guard cap that is mounted to the instrument inlet. In most cases, it
prevents water from penetrating inside and allows correct instrument measurement.
Figure 2. Rain Guard Cap Figure 3. Air-Drying Pre-Filter Kit
In some special cases with very harsh humidity conditions, an additional air-drying pre-filter
kit. It can be added between an instrument and rain guard cap. The filter comprises a plastic
housing with holders and can contain 15 to 20 plastic bags with CaCl2 as a water absorbent.
Note: Although an air-drying pre-filter helps prevent water penetration inside the
instrument sample train, it may cause slower instrument response to VOCs and negatively
impact proper instrument reading. This filter should not be used if water-soluble gases are
monitored.
If the monitor is used in duct sampling applications, the risk of condensation is high and the
above pre-filter kit may need to be installed to provide proper instrument readings.
1998-1001 Rev 1 11/16
Note: Consult Honeywell before installation.
COMMISSIONING RAEGUARD 2 PID
Ensure the power is correctly connected and meets the nominal requirements. Double-check for
correct wiring. Be sure that instrument is grounded properly. Refer to the RAEGuard 2 PID User Guide
for information on wiring and testing.
Assemble a tubing system for gas delivery from the sampling area to the instrument. The
recommended tubing should be either stainless steel or Teflon. This type of tubing provides the
minimum gas absorbing to the walls and reduces instrument reading distortion. Attach the tubing to
the outlet to return the gas back to the gas stream or out of the zone of the personal working area in
order to prevent possible contamination with toxic gases. There are no requirements for this portion
of the tubing.
When all the necessary parts of the system are connected and the whole system is installed, perform
a check for leaks. Disconnect the tubing from the sampling duct and block the tube inlet, or simply
3
Technical Note
Installation and Maintenance
1998-1001 Rev 1 11/16
of a RAEGuard 2 PID
block the tube inlet from the sampling area if there is no connection for the short period of time
(usually for the several seconds). If there is no leak in the system, the pump stops working. Make
sure that you are observing that the pump stops due to your action rather than due to the pump duty
cycle. Resume pump work according to the User Guide. If there is a leak, perform failure analysis, fix
system and make a final leak check using the procedure described above.
SERVICE AND CALIBRATION INTERVAL
Honeywell® recommends sensor module calibration if it does not pass a bump test (calibration check),
no less than every six months, and filter replacement every six months. Note that these intervals are
application dependent and directly relate to the gas exposure. Therefore, intervals should be adjusted
accordingly. For the calibration procedure, follow the recommendations in the User Guide.
Note: A two-hour warm-up time is recommended before making critical measurements.
Honeywell recommends guidelines based on the correct use of filters:
1. Perform monthly bump test (calibration checks). If the system fails to pass, a full
calibration must be done.
2. Replace filter, part number is 490-0164-010, at 6-month intervals unless a filter
break-through is observed, requiring immediate replacement. Pre-filter filament
replacement at 1 to 3 month intervals, and depends on the severity of humidity
conditions.
3. Replace or clean the sensor and lamp at 12-month intervals.
While every effort has been made to ensure accuracy in this publication, no responsibility can be accepted for errors or omissions. Data
may change, as well as legislation, and you are strongly advised to obtain copies of the most recently issued regulations, standards and
guidelines. This publication is not intended to form the basis of a contract and the company reserves the right to amend the design and
specification without notice.
Find out more:
www.raesystems.com
Javastrasse 2
8604 Hegnau
Switzerland
Tel: +41 (0)44 943 4300
Customer Service:
(Freephone number)
Alternative number)
Fax: 00800 333 222 55
(Fixed Gas Detection)
(Portable Gas Detection)
Americas
405 Barclay Blvd.
Lincolnshire, IL 60069
USA
Tel: +1 847 955 8200
Fax: +1 847 955 8210
Phone: 408.952.8200
Fax: 408.952.8480
Asia Pacific
7F SangAm IT Tower,
Seoul 03922,
Korea
Technical Services
Technical Services
5
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.