Honeywell MIDAS-M User Manual

MIDAS-M MULTIGAS TRANSMITTER
Fixed Single Point Extracted
Multi Gas Transmitter
3016M5001 Rev A ECO HAA190080
Table Of Contents
C
HAPTER 1 —
About Us
Product Description
Overview
User Interface
Default Configuration
Installation
C
HAPTER 2
Mount the Transmitter Device
Mount the Relay Module
Sample and Exhausting Tubing Calculations 17
7
7
7
8
9
10
11
11
15
In-Line Filters
Local Detector Option
Electrical Installation
Electrical Connection
Wiring The Relay Module
Mount The Sensor Cartridge
Device & Web Operations
C
HAPTER 3
Comissioning
Monitoring Mode
Review Information
18
18
19
23
26
28
29
29
30
32
Setup Operations 34
Test Mode
36
Calibration Mode
Calibration & Bump Test 39
Bump Test 40
Internal Web Server 41
Physical Network Components 41
Internet Settings 41
Running the Web Server 42
C
HAPTER 4
Maintenance
System Leak Check
C
HAPTER 5
Additional Information
EU Directive 2012/19/EU: Waste Electrical and Electronic
Equipment (WEEE)
38
65
66
69
69
Troubleshooting and Fault Diagnosis
Specifications
Ordering
Transmitter 77
Cartridge 77
Smart Sensor 78
Accesories 78
Gas Table
Combination Index
Reflex
A Modbus/TCP Interface
70
74
77
82
84
85
86
Security 99
Modbus TCP 99
DOS attack 99
Install Web Server Certificate in Google Chrome
Install Web Server Certificate in MS Browsers
101
103
Warranty 105
Sensor Cartridge Warranty 106
107Contact Us
CHAPTER

Introduction

1
Learn what you need to know about the Honeywell Midas®-M Gas Detector before operating.

About Us

Honeywell Analytics: the global leader in gas detection
While you build a better world, we’re building smarter safety.
For nearly 50 years, Honeywell Analytics has been protecting people who put their lives on the line every day. It takes a special level of drive and dedication to work in hostile environments, and we believe your workers deserve the same kind of commitment from us when it comes to securing their health and safety.
Right from the start, our founders were determined to find better ways to design, build and manufacture reliable gas detection solutions. Through advanced technology and pure industry know-how, Honeywell Analytics continues to improve and evolve our products. Today, we’re proud to partner with customers worldwide who share our vision of vigilance and rely on us to help them protect what matters most.
From custom-designed solutions to unmatched training and support, Honeywell Analytics is simply the smartest choice when you demand the best in safety.

Product Description

The Honeywell Midas®-M is a Fixed Extractive Single Point with 4-in-1 Multi Gas Detector that draws a sample locally or from a remote point to a sensor cartridge that is located inside the detector’s chassis. A wide range of toxic, flammable and oxygen gas sensor cartridges are available that enable detection of gases used or generated in the Semiconductor and other manufacturing industries.
Midas-M is shipped from the factory with the security function disabled. We strongly recommend enabling this function for safe use of the detector. To enable it, select Set up > Security and type an 8 character password.
Midas-M is wall mounted and displays gas concentration, alarm, fault and status information via its backlit LCD and LEDs. A simple to use 4-button keypad located under the display provides the facility to set-up, review, operate and make changes to the detector’s configuration.
Midas-M 7 User Manual
Midas-M has flexible power and communications capabilities built in as standard. These include 3 on board relays, 0-21 mA analog output, Modbus/TCP outputs for signal and service connectivity as well as Power over Ethernet (PoE) connection that enables a single Ethernet connection to be made for all power, control and communication requirements.

Overview

The Midas-M gas detector comprises of 4 parts: the main module, the mounting bracket assembly, the sensor cartridge and the pump module.
Midas-M 8 User Manual

User Interface

Detector's User Interface overview.
Midas-M 9 User Manual

Default Configuration

The Midas-M gas detector is factory-configured.
NOTE: Oxygen levels are 20.9% v/v in a normal atmosphere, equivalent to 17.3 mA. Use caution when integrating an oxygen Midas-M unit using the 4-20 mA output since fault, inhibit, and no power conditions are below 4 mA and, by default, an oxygen depletion alarm is triggered on a falling alarm (default 19.5% v/v). In that case, Honeywell Analytics recommends one of the following:
• Use Modbus TCP digital communications
• Use discrete relay inputs instead of a 4-20 mA signal to trigger gas alarms
• Program the control system of the 4-20 mA input with logic and a delay (e.g., 1 second) before triggering an alarm to determine if the 4-20 mA output is at one of the 0-4 mA conditions described above.
Contact Honeywell Analytics for further information.
Factory Default Configuration
Toxic Gas flammable gas
1
oxygen
Full Scale (FS) Typically4 x Threshold
LimitValue (TLV)
Alarm 1 (Relay 1)
Alarm 2 (Relay 2)
Fault (Relay 3)
Latching Latching. Alarm and fault relays DO NOT automatically reset when reading
Passcode No Passcode set. Address 169.254.60.47 subnet mask 255.255.255.0
Current source with:
1.0 mA Fault
2.0 mA Inhibit
3.0 mA Maintenance Fault
4.0 to 20.0 mA Gas reading (normal operation)
21.0 mA Over range
1/2 TLV 10% LEL 23.5% v/v (Rising) Normally de-energized, energizeson alarm. TLV 20% LEL 19.5% v/v (F alling) Normally de-energized, energizeson alarm. Normally energizes, de-energizedon fault.
fallsbelow alarm thresholds.Relays MUST be manually reset.
100% Lower Explosive Limit(LEL)
25% V olume (v/v)
1
Not yet released
Midas-M 10 User Manual
CHAPTER
2
For ease of installation Midas-M has been designed to allow the installation of the mounting bracket assembly and main module separately from the other parts of the detector. The detector location and hard wiring can therefore be completed before fitting the detector’s main module and sensor cartridge.
WARNING Midas-M is designed for installation and use in indoor safe area non-explosive atmospheres. Installation must be in accordance with the recognized standards of the appropriate authority in the country concerned. Prior to carrying out any installation ensure local regulations and site procedures are followed. Ensure the connection of all modules (e,g pump, and cartridge module) with the main module. Otherwise, Midas-M can have flow leakage or malfunction.

Installation

Mount the Transmitter Device

The Midas-M gas detector has an integral mounting bracket assembly that is easily mounted to a suitable vertical surface such as a wall, tool housing, mounting plate on a pole etc.
Midas-M outline dimensions
Midas-M 11 User Manual
NOTE Notice the space required to install the instrument, open the door, push in/pull out modules, and tubing.
Mounting procedure
Midas-M 12 User Manual
1. the door.
2. Un-tight the two screws.
3. Carefully pull the main module forward.
4. Drill two holes 2.2 in (55.9 mm) Use 2 x M4 Screws or equivalent for mounting (head size 7-11 mm).
Midas-M 13 User Manual
NOTE
This drill template is not to scale.
5. a. Partially screw the fixings into the mounting surface.
b. Place the mounting bracket assembly over the screws, so they pass
through the mounting holes and then slide down to locate in the slots.
c. Tighten the screws to secure the mounting bracket assembly.
6. a. Align the rounded corner at the lower of the main module with the similar
rounded corner at the bottom of the mounting bracket assembly
b. Slide the main module backward while pushing the main module up after
connecting with a mounting bracket so that the PCB connector and tubes engage simultaneously.
7. Ensure the PCB, connector, and tubes are fully engaged by firmly pushing the
main module horizontally backward on the mounting bracket assembly. DO NOT PUSH ON THE LCD AS THIS MAY DAMAGE IT.
8. a. Align the two fixing screws located at the bottom of the module with the
screw threads on the mounting bracket assembly.
b. Tighten the screws to secure the module to the mounting bracket
assembly.
9. Remove the internal packing card securing the pump. Failure to remove this
packing will result in damage to the Midas-M detector.
Midas-M 14 User Manual

Mount the Relay Module

The Relay Module has an integral mounting bracket assembly that is easily mounted to a suitable vertical surface such as a wall, tool housing, mounting plate on a pole etc.
1. Pull out the four screws and remove the plate cover.
2. a. Partially screw the fixings into the mounting surface. b. Place the Relay module over the screws, so they pass through the holes to locate in
the slots.
c. Tighten the screws to fix the Relay Module to the surface.
3. Connect the wire through the cable gland to the terminal block.
4. Place the plate cover in position and screw the four screws you removed in Step 1.
5. Push the mounting bracket assembly horizontally and engage the Relay Module connector.
6. a. Align the screw holes and the slot-boundaries, and tight the machine screws
(3016D0355) provided in packaging.
b. Connect the mounting bracket assembly to the Relay Module.
Midas-M 15 User Manual
Midas-M 16 User Manual

Sample and Exhausting Tubing Calculations

Description of flow rate, and tubing vacuum at the inlet and exhaust points,
Inlet sample specifications
Description Maximum
Tubing Length, m (ft) 30 (100)
Sample Point Vacuum -25.4 cm H2O (-10 in H2O)
Transport Time (sec), ID 1/8”
Transport Time (sec), ID 3/16” 53
Flow rate, cc/min. 600 (Flow is constant)
Tubing OD, mm (in) 6.35 (0.25)
Tubing ID, mm (in) 3.18 (0.125)
1
Honeywell Analytics recommends thick-wall tubing (1/8” ID) for best speed of response. Due to
its lower surface area, thick- wall tubing may require less conditioning than thin-wall tubing.
2
The flow rate is electronically maintained at approximately 600 cc/min and may vary within
acceptable tolerances.
Tubing lengths vary among gases. See the Gas Chart for recommended lengths.
NOTE Honeywell Analytics recommends the use of Teflon FEP (Fluorinated Ethylene Polymer) tubing to assure proper sample transport. The properties of Teflon FEP make it the best choice for transporting sample toxic gases to instruments when compared with the properties of other similar tubing materials.
1
25
2
Outlet sample specifications
description maximum
Tubing Length, m (ft) 30 (100)
Back Pressure at Exhaust Point 20.3 cm H2O (8 in H2O)
Flow rate, cc/min 600 (Flow is constant)
Midas-M 17 User Manual
description maximum
Tubing OD, mm (in) 6.35 (0.25)
Tubing ID, mm (in) 4.76 (0.188)

Prepare Tubing

1. Cut the tube squarely and remove any burrs.
2. Mark from end of tube the length of insertion. The insertion lenght of Midas-Mfrom the end of tube is 15.5 mm.
NOTE When inserting the tube into the inlet/outlet port of the Midas-M, make sure that it is inserted up to the marked position.

In-Line Filters

External filters must be used to protect the tubing from contamination. Use particulate filter part number 780248 for normal gases and 1830-0055 or 1991-0147 for corrosive gases. Replace the filter every 3 months. Refer to the Gases Table for specific gases.

Local Detector Option

The Midas-M gas detector can also be used to monitor for gas at the location of the detector. To do this, an inline filter is simply connected to the sensor cartridge gas inlet port. The external dust filter part number is 780248 for normal gases and 1830-0055 or 1991-0147 for corrosive gases. The area around the detector is then being monitored as opposed to a sample being drawn from a remote location.
Midas-M 18 User Manual

Electrical Installation

Access for the electrical wires to the terminal module is made via the PG16 cable gland located at the bottom of the mounting bracket assembly. The cable gland can be removed and replaced with a suitable conduit fitting if required.
Typical wire routing
The terminals used are suitable for conductors of 24 to 14 AWG (0.5 to 1.8mm Dia.). We recommend using the 16 AWG (1.5 mm Dia.) conductors.
If Power over Ethernet (PoE) is used to power the device, then 24 VDC power must not also be connected to the device, (or conversely, if 24 VDC is used to power the Midas-M, then electrical power via the Ethernet port must not be applied). Failure to observe this requirement may cause damage to the gas detection system and will not be covered by the standard warranty.
When connecting the wires ensure that the power switch is in the off position.
NOTE Earthing Requirements: If the Midas-M unit’s metal chassis is not connected directly to a metal surface for earthing purposes, an additional earth wire will be required. Connect a
Midas-M 19 User Manual
wire via the PG16 gland to the dedicated earth tag (screw terminal) located on the bottom bracket and connect the other end of the wire to a dedicated external earthing point. If Power over Ethernet (PoE) power supply is being used, shielded CAT5 Ethernet cable is recommended. Please ensure that your wiring avoids earth ground loops that may affect the performance of your equipment.
NOTE Instrument grounding is required to ensure stable performance and to limit the effects of radiofrequency interference before installation.
The following diagram shows the terminal module layout and terminal identification as well as the jumper locations.
Transmitter Terminals
Midas-M 20 User Manual
Transmiter Terminals
terminal number function description
1 24Vdc Input 0Vdc
2 24Vdc Input +24Vdc
3 mA output - Channel 1 COM
4 mA output - Channel 1 mA-
5 mA output - Channel 1 mA+
6 mA output - Channel 2 COM
7 mA output - Channel 2 mA-
8 mA output - Channel 2 mA+
9 mA output - Channel 3 COM
10 mA output - Channel 3 mA-
11 mA output - Channel 3 mA+
12 mA output - Channel 4 COM
13 mA output - Channel 4 mA-
14 mA output - Channel 4 mA+
15 Relay 1 Normally Closed
16 Relay 1 Common
17 Relay 1 Normally Open
18 Relay 2 Normally Closed
19 Relay 2 Common
20 Relay 2 Normally Open
21 Relay 3 Normally Closed
Midas-M 21 User Manual
terminal number function description
22 Relay 3 Common
23 Relay 3 Normally Open
Relay function table
Display Description Relay 1 Relay 2 Relay 3
INSTRUMENT F ONLY
SEPARATE F
COMBINED F
Midas-M 22 User Manual
Instrument fault only
Separate fault relays
Combined fault relay
Alarm 1 Alarm 2
Any alarm
Alarm 1 Alarm 2 Any fault
Maintenance fault
Instrument fault
Instrument fault

Electrical Connection

Midas-M can be powered by either 24 VDC via traditional discrete wiring or by approximately 48 VDC delivered through the Ethernet cable from a PoE source. In either case, the 4-20 mA analog output can be used.
The analog output can be configured for fully isolated operation. With 24 VDC power the 4-20 mA output can be configured for sink, source or isolated output operations. Following are some schematic diagrams of typical electrical connection configuration.
NOTE When wiring the Midas-M Transmitter to a controller, program the controller for a 1-2 second delay before reporting to prevent false alarms.
Generic example Midas-M 3-wire sourcing
Midas-M 23 User Manual
Generic example Midas-M 4-Wire Isolated output
Generic example Midas-M 3-wire Sinking output
Midas-M 24 User Manual
Generic example Midas-M Isolated 4-20mA output w/PoE power
Midas-M 25 User Manual

Wiring The Relay Module

The Midas-M relay module contains:
Nine relays to activate external devices.
Eight gas alarm relays which indicate level 1 and level 2 alarms of each channel.
A common fault relay which indicate maintenance fault or instrument fault.
Specifications
Power Supply
Power consumption
Relay rating 1.0 A @ 30VDC or 0.5 A @ 125 VAC max, 10 uA @ 10 mV minimum
Wiring 14 AWG maximum per each channel
Midas-M 26 User Manual
No separated power supply required Supplied from the Midas-M transmitter
< 1.45 W
Default configuration
All alarm relays Normally de-energized, energized on alarm
Common fault relay Normally energized, de-energized on fault
Midas-M 27 User Manual

Mount The Sensor Cartridge

The Midas-M Sensor Cartridge enables the detection of a wide range of asphyxiant, toxic, flammable, pyrophoric, corrosive, and oxidizer (including Oxygen) gases. The plug & play Sensor Cartridge is located inside the detector’s chassis.
NOTE The Midas-M sensor cartridge is supplied separately. Before use, verify the part number and type of sensor cartridge is correct for your application.
1. Turn the power off.
2. Open the door and remove the BiAS battery module and the two plug caps from the sensor cartridge.
3. Align the Sensor cartridge pins with the socket in the sensor cartridge chamber and then push the sensor cartridge gently until fully seated.
4. Close the door.
5. a. Press the ‘O’ button to clear the “Detect New Cartridge ” message.
b. Confirm that the green LED is flashing, and that the yellow and red LEDs are off.
c. Confirm the display shows a concentration of zero as appropriate.
Midas-M 28 User Manual
CHAPTER
3
Learn what you can do with your Honeywell Midas®-M Detector, from commissioning to Maintenance

Device & Web Operations

Comissioning

Comissioning should be followed carefully and only performed by suitably trained personnel.
WARNING Before carrying out any work, ensure local and site procedures are followed. Ensure that the associated control panel is inhibited to prevent false alarms.
1. Ensure the detector is wired correctly.
2. Ensure that the correct sensor cartridge is fitted. (If the cartridge has not been stored at room temperature, allow one hour for equilibration.)
3. Ensure the ON/OFF switch on the bottom of the mounting bracket assembly is in the ON position.
4. Apply power to the system. Wait for a couple of minutes until the start-up routine completes.
5. After the start-up routine, the detector will display the Normal/Monitoring Operating Mode.
6. Perform a leak test to ensure all connections are secure. ‘Waiting…’ will be displayed on booting time.
7. Allow the detector to stabilize until the ‘Warm-up’ message is no longer displayed after booting. The maximum warm-up time is dependent on sensor type. Refer to individual cartridge data sheets. Warm-up times are typically much faster.
8. Ensure the correct ID code is selected in Set-up > alarm menu
9. If this is a first-time start-up, the “Detect New Cartridge ~” message could be displayed. Press the ‘O’ button to clear the message.
Midas-M 29 User Manual

Monitoring Mode

After commissioning, the Midas-M detector is set in the Monitoring Mode, which is the starting point for all of the device's operations.
The examples in the following table are for a linear 4-20 mA output over a full scale range of 2 ppm. The current output for a given gas concentration will be different for other full scale ranges (linear 4 mA = 0 % full scale to 20 mA = 100 % full scale). The alarm and fault relays are in default (latching) mode.
Normal Operation Display and Outputs.
This table is based on the default configuration of toxic sensor type.
Status Relay Status
Alarm 1
Midas-M 30 User Manual
Alarm relay 1 activated
4-20mA out­put
mA output to match the concentration
LEDs Backlight
Green flash Alarm 1 Red on
Red on of the channel
Loading...
+ 79 hidden pages