Honeywell IQ Express, MultiPro IQ Express Reference Manual

Reference
Manual
IQ Express
Docking Station
for
Honeywell Analytics, Inc. 405 Barclay Boulevard Lincolnshire, IL 60069 (800) 711-6776 (847) 955-8200 Fax (800) 995-4992 Part number 13-287 Version 4 27 August 2012
THE MULTIPRO IQ EXPRESS IS DESIGNED TO INTERFACE WITH HONEYWELL ANALYTICS GAS DETECTORS.
HONEYWELL ANALYTICS GAS DETECTORS HAVE BEEN DESIGNED FOR THE DETECTION OF DEFICIENCIES OF OXYGEN, ACCUMULATIONS OF FLAMMABLE GASES AND VAPORS AND ACCUMULATIONS OF TOXIC VAPORS.
IN ORDER TO ENSURE THAT THE USER IS PROPERLY WARNED OF POTENTIALLY DANGEROUS ATMOSPHERIC CONDITIONS, IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS MANUAL AND THE OPERATIONS AND/OR REFERENCE MANUAL(S) FOR THE GAS DETECTOR(S) BE READ, FULLY UNDERSTOOD, AND FOLLOWED.
THIS MANUAL IS NOT INTENDED TO REPLACE THE OPERATIONS AND/OR REFERENCE MANUALS FOR THE GAS DETECTOR. THIS MANUAL IS ONLY DESIGNED TO AID IN THE INSTALLATION AND OPERATION OF THE IQ EXPRESS SYSTEM AND SHOULD BE USED IN CONJUNCTION WITH THE INSTRUMENT REFERENCE OR OPERATIONS MANUAL AT ALL TIMES.
MultiPro IQ Express
Reference Manual
Honeywell Analytics Part Number 13-287
Version 4
Copyright 2012
by
Honeywell Analytics, Inc.
Lincolnshire, IL 60069
All rights reserved.
No page or part of this operation manual may be reproduced in
any form without written permission of the copyright owner
shown above.
Honeywell Analytics reserves the right to correct typographical
errors.
1
Table of Contents
Warnings and Cautions 3
A. Signal Words 3
1.Overview 4
1.1 Tests and record keeping 4
1.2 PC Connection 4
1.2.1 USB Connection 4
1.2.2 Ethernet Connection and E-mail 4
1.3 PC Requirements 5
1.4 Instrument Firmware Requirements 5
1.5 Power Requirements 5
1.5.1 Dock with direct USB Connection to PC 5
1.5.2 Multi-Dock Setup with USB Connection 5
1.6 Sensor compatibility 5
1.7 Default Calibration Gas Settings 5
1.8 Calibration Gas Cylinder Regulator Requirements 5
1.9 Dock Location 5
1.10 Delivery 5
2.Installation 5
2.1 Installation Overview 6
2.2 IQ Administrator Pro and PostgreSQL Database Server Installation 6
2.3 IQ Database Manager Software Installation 6
2.4 MultiPro IQ Express PC Software Installation 6
2.5 Install IQ Administrator Pro and PostgreSQL database 8
2.6 Power and Connectivity 8
2.6.1 PC – Single Dock via USB 8
2.6.2 PC – Multi Dock via USB 8
2.7 Connecting the Calibration Gas Cylinder to the Dock(s) 9
2.7.1 Single Cylinder to Single Dock 9
2.7.2 Single Cylinder to Multiple Docks 9
2.7.3 Fresh Air Port Instructions 10 Fresh Air Filter 10 Providing Fresh Air in a Contaminated Environment 11
Option 1: Using a Gas Cylinder to Provide Fresh Air to the Dock 11 Option 2: Using a Sealed Conduit to Provide Fresh Air to the Dock 11 Multi-Dock Setups with Fresh Air Manifold 11
2.8 Initial PC- Dock Connection 12
2.8.1 PC screens 12
2.8.2 Dock screens 13
3.Software 13
3.1 Dock Status (USB Connection) 14
3.2 Dock Controls and Instrument Status 14
3.2.1 Dock Configuration 14 Hardware Configuration 15 Connection Speed 15 Audible Alarm Sensitivity Controls 15 AC Power 15 Pump Diagnostics 15
2
LCD Contrast Control 15
Status After Test 16 Battery 16 Calibration Interval 16 Gas Configuration 16 Saving new settings 16 Dock LCD Display 17 Ethernet Controls 17
3.2.2 Instrument Tab 17
3.3 Advanced Settings / Passcode 18
4.Dock Use 19
4.1 Test Sequence 20
4.2 IrDA Connection Issues 21
4.2.1 Connection Lost 21
4.2.2 Outright Failure to Connect 21
4.3 Test Failures 21
4.3.1 No gas detected 21
4.3.2 Span calibration failure 21
4.3.3 Alarm test failure 21
4.3.4 PC Aborted Test 21
5.Menu Items 21
5.1 File Menu 21
5.2 Tools Menu 22
5.2.1 Dock 22
5.2.2 MultiPro 22
5.2.3 Certificates 23
5.2.4 Session Downloading 23
5.3 Help Menu 23
5.3.1 Check for Update 23
5.3.2 About IQ Express 23
6.Software Upgrades 24 Appendix A: USB Architecture with Network Access 25 Appendix B: Calibration Frequency 26 Honeywell Analytics Warranty Gas Detection Products 27
Warnings and Cautions
A. Signal Words
The following signal words, as defined by ANSI Z535.4-1998, are used in the MultiPro IQ Express Reference Manual.
indicates an imminently hazardous situation which, if not
avoided, will result in death or serious injury.
indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
indicates a potentially hazardous situation, which if not
avoided, may result in moderate or minor injury.
CAUTION used without the safety alert symbol indicates a potentially
hazardous situation which, if not avoided, may result in property damage.
3
1. Overview
The MultiPro IQ Express Docking Station is an automatic calibration station for use with MultiPro Gas Detectors. The MultiPro IQ Express Docking Station automatically performs up to eight critical tests including sensor identification, instrument performance, bump and alarm tests and record-keeping procedures in 1 minute. The station also retains a historical record of testing and instrument maintenance in the system's onboard memory.
The MultiPro IQ Express Docking Station may be operated as a stand-alone calibration station, or may be connected to a PC via USB cable for increased control over system operations.
When the Dock is operated without a PC interface, it will operate with the last set of calibration gas settings that were uploaded into the instrument.
Unless otherwise specified at the time of purchase, all MultiPro IQ Express Docks shipped from Honeywell Analytics are configured for instruments containing LEL, O
, CO and H2S sensors with the
2
standard calibration gas settings. In the event that the dock is to be used as a stand-alone calibration station with non­standard alarm settings, the dock must be reprogrammed with the MultiPro IQ Express Software.
The IQ Express Configuration Software must be loaded onto the PC prior to completing the USB connection.
Note: Do not connect the MultiPro IQ Express Dock to the PC until the software has been installed.
1.1 Tests and record keeping
The MultiPro IQ Express Docking Station automatically performs the following procedures whenever an instrument is placed in the dock.
Instrument identification Battery test Audible alarm test Visual alarm test Vibrating alarm test (if applicable) Fresh air calibration Bump test Record-keeping
Datalogger and Eventlogger download
(Dock must be connected to PC)
In the event that an instrument fails the bump test, or if the instrument’s calibration due date has passed, the MultiPro IQ Express Docking Station will automatically proceed to a full calibration without further intervention from the user.
Note: The MultiPro IQ Express Dock only tests the sensors that are currently detected by the MultiPro. The MultiPro only displays readings for sensors that are actively detected by the instrument. If the MultiPro does not show gas readings for a specific sensor on the instrument display, then it is not detecting that sensor.
If the instrument fails any of the tests listed above, the MultiPro IQ Express Dock will notify the user through the display on the dock, and through the PC if the dock is connected to a computer and the software is running.
Results of tests and calibration attempts are stored in the dock and uploaded to the PC if the software has been loaded and the USB connection is active.
1.2 PC Connection
PC Software must be installed prior to connecting the MultiPro IQ Express Dock to the PC. An overview of the installation procedure is given below in section 2.1.
1.2.1 USB Connection
Every MultiPro IQ Express Dock can be connected to the PC via the PC’s USB port.
Note: USB cable length from PC to dock may not exceed 10 feet.
1.2.2 Ethernet Connection and E-mail
MultiPro IQ Express Docks with the Ethernet upgrade may be connected to the PC via USB or via a network. For Ethernet Connection instructions, follow the directions given in the IQ Express Ethernet Instructions booklet.
Docks that are connected by Ethernet have additional controls at the PC to ping the dock, to view the Ethernet settings and to view and test the e-mail system.
In the MultiPro IQ Express, e-mail may be sent by the dock to a single e-mail address when a fault is detected at the dock, but for no other reason. The dock must be connected to a network via
4
Ethernet for the e-mail function to work. E-mail capability is controlled entirely through the IQ Database Manager Pro program. For more information on the e­mail function see Ethernet Controls in section 3.2.1 below or the IQ Database Manager Pro manual.
1.3 PC Requirements
Pentium II Processor 400MHz or
better or equivalent.
128MB RAM. Windows 98*/2000**/XP. 30MB hard drive disk space. USB Port.
*Honeywell Analytics discourages multiple IQ Express Dock setups with PC’s running Windows 98.
**Service Pack 3 for Windows 2000 is highly recommended.
1.4 Instrument Firmware Requirements
Instrument firmware version 1.32 or higher is required in the MultiPro. Firmware upgrades to the MultiPro may be made with a PC through the IrDA port.
1.5 Power Requirements
The MultiPro IQ Express Dock is delivered with an appropriate power supply and a variety of adapters.
The dock must be plugged into an appropriate electrical outlet using the power supply / wall cube that was included with the dock at the time of purchase.
1.5.1 Dock with direct USB Connection to PC
The MultiPro IQ Express must be plugged into an appropriate electrical outlet using the power supply / wall cube that was included with the dock at the time of purchase. A USB cable is included to connect the dock to the PC with the USB cable.
1.5.2 Multi-Dock Setup with USB Connection
For multi-dock configurations each dock must be powered using the power supply that was included with the dock. An alternate option is to purchase the optional power supply that is capable of running up to 4 MultiPro IQ Express docks. Under no circumstances can multiple docks be run from a single USB
port or from a USB hub regardless of whether it is a powered or non-powered hub.
Power and Connectivity are discussed in greater detail in section 2.6.
1.6 Sensor compatibility
The MultiPro IQ Express Dock is capable of processing instruments containing any combination of O sensors.
, LEL, CO and H2S
2
1.7 Default Calibration Gas Settings
Dock
Setting
O2 18% vol.
LEL 50% LEL
CO 50 PPM
H2S 25 PPM
Default Calibration Gas Settings and
Dock Requirements
Default Cal Gas
Setting
1.8 Calibration Gas Cylinder Regulator Requirements
A demand-flow regulator must be used with the IQ Express.
1.9 Dock Location
The MultiPro IQ Express Dock should be located in an area that is not exposed to direct sunlight.
1.10 Delivery
Each MultiPro IQ Express Calibration Station is delivered with the following items:
MultiPro IQ Express Docking Station. MultiPro IQ Express Configuration
Disc with MultiPro IQ Express Pro Software, IQ Database Manager Pro Software and PostgreSQL Database Server Software.
USB Cable. Power Supply. 2’ piece of calibration gas tubing with
pre-installed white quick disconnect fitting.
2. Installation
The MultiPro IQ Express Docking Station must be configured with a PC prior to use. Once it has been configured, it may be used as a stand-alone calibration station without a PC connection, or connected to
5
a PC for increased control over system operations.
Unless otherwise specified at the time of purchase, all MultiPro IQ Express Docks are shipped from Honeywell Analytics configured with the standard calibration gas settings, which are shown in section
1.7. Whenever the user chooses to activate
the MultiPro IQ Express without an active PC connection, the dock will operate with the last set of calibration gas settings that were uploaded into the instrument.
Note: Do not connect the MultiPro IQ Express to the PC until all three items on the MultiPro IQ Express Installation disc have been installed.
2.1 Installation Overview
A number of software programs and a PostgreSQL Database Server must be installed to control the MultiPro IQ Express.
1. Install the PostGreSQL Database Server and the IQ Administrator Pro Program as described in the IQ Administrator Pro Reference Manual.
2. Install the IQ Database Manager Pro program as described in the IQ Database Manager Pro Reference Manual.
3. Install the MultiPro IQ Express PC Software as described below in section 2.4.
4. Create a default PostgreSQL database as described in section 6 of the IQ Administrator Pro Reference Manual.
5. If connecting the dock to the PC via USB, follow the instructions in section
2.6. If connecting the dock via Ethernet, follow the Ethernet Instructions booklet.
6. Follow the instructions in section 2.7 for providing calibration gas and fresh air to the dock(s).
7. For docks that will be connected permanently by USB port, configure the docks by connecting them to the PC’s USB port. If the instrument is to be connected via Ethernet, follow the Ethernet Instructions that came with the dock.
2.2 IQ Administrator Pro and PostgreSQL Database Server Installation
The IQ Administrator Pro Program is contained on the IQ Express Installation Disk. Place the disk in your PC’s CD tray and follow the installation instructions given in the IQ Administrator Pro Reference Manual that is included with the IQ Express.
2.3 IQ Database Manager Software Installation
The Database Manager Pro Program is contained on the IQ Express Installation Disk. Place the disk in your PC’s CD tray and follow the installation instructions given in the IQ Database Manager Software manual that is included with the IQ Express.
2.4 MultiPro IQ Express PC Software Installation
The PostgreSQL Database Server and IQ Database Manager Pro Programs should be installed before proceeding to step 1.
Note: Screens shown below may be slightly different depending on your PC’s operating system.
1. The Software Installation screen
should come up automatically once the installation of Database Manager is complete. If this screen does not come up, access the CD drive using Windows Explorer.
2. Select “MultiPro IQ Express Pro” and
click “Install”. The InstallShield Wizard will begin the installation.
6
3. Click “Next” to continue with the installation. The Licensing Agreement screen will be shown.
4. If you accept the terms of the License
Agreement, click “I accept.....” and
then click “Next” to continue. The upgrade information and PC requirements will be shown.
6. Enter the name of the user and the name of the organization. Then select whether this application may be used by “Anyone who uses this computer” or “Only for me”. Then click “Next. The destination folder screen will be shown.
7. To install IQ Express in the default directory at “C:\Program Files \ Biosystems \ IQ Pro \ MultiPro IQ Express Pro \” click “Next”. To install the files to another directory, click “Browse” and use Windows Explorer to specify the new location. Click “Next” once this is accomplished. The following screen will then be shown.
5. Click “Next” to continue. The Customer Information screen will be shown.
7
8. Click “Next” to continue. The status of the file decompression will be shown.
Once the installation is complete, the following screen will be shown:
9. Click Finish to conclude the installation. The IQ Express logo will appear on your desktop.
Note: Do not launch the MultiPro IQ Express Software again until the instructions in sections 2.5 – 2.8 below have been completed.
2.5 Install IQ Administrator Pro
and PostgreSQL database
Proceed to the IQ Administrator Pro Reference Manual section 6 and follow the instructions to create and specify the PostgreSQL database.
2.6 Power and Connectivity
Honeywell Analytics recommends that an appropriate surge suppressor be used to protect this equipment against destructive surge currents and transient voltages.
This section covers providing power to the IQ Express Docks and connecting them to your PC. The instructions given below pertain to standard connections via the PC’s USB port.
Note: Disregard this section if you are planning to connect the docks via Ethernet or if you have purchased the IQ Express Mounting Bracket.
IQ Express Docks that are Ethernet-ready are shipped with an Ethernet Instructions booklet. Proceed to the Ethernet Instructions booklet if you are planning to connect the dock to the PC via Ethernet. Once the instructions in the booklet have been followed and the IQ Express Dock is connected, return to section 2.7 for calibration gas cylinder connection instructions.
If you have purchased the IQ Express Mounting Bracket, proceed to the Assembly Instructions for IQ Express Mounting Bracket. Once the docks are correctly installed on the mounting bracket, proceed to section 2.7.3 for instructions concerning the Fresh Air Port(s).
2.6.1 PC – Single Dock via USB
The MultiPro IQ Express must be plugged into an appropriate electrical outlet using the power supply / wall cube that was included with the dock at the time of purchase. After connecting the dock to the PC with the USB cable that was included with the dock proceed to section
2.7 for instructions concerning calibration gas.
Note: USB cable length from PC to dock may not exceed 10 feet.
2.6.2 PC – Multi Dock via USB
For multi-dock configurations each dock must be powered using the power supply that was included with the dock. An alternative option is to purchase the optional power supply that is capable of running up to 4 MultiPro IQ Express docks. Under no circumstances can multiple docks be run from a single USB port or from a USB hub regardless of whether it is a powered or non-powered hub. Once the power requirement for each dock is satisfied the USB connection to the PC can be made by using a USB hub or by connecting the USB cable directly to the PC.
Note: With a USB hub, USB cable length from hub to dock may not exceed 10 feet.
8