Honeywell Fendall 2000 Installation Manual

0 (0)
Honeywell Fendall 2000 Installation Manual

Fendall 2000Cartridge Quick Installation

1.Inspect label and review Expiration Date to determine if cartidge is current and ready for use. (See Fendall 2000 Installation and Operation Manual Section C for detailed instructions.)

2.Install Sterile Cartridge by holding the side handles, making sure the nozzle and tube are facing downward and the front label is facing out.

CAUTION: CARTRIDGE WEIGHS APPROX. 60 LBS (27 KG). Two people lifting cartridge is strongly recommended for installation.

3.Release the four straps by cutting the zip-ties. Ensure that the nozzle and tubing is fully extended from the cartridge.

Fig. 1

4.Hang cartridge on brackets and push cartridge all the way back. Fig. 2

5.Remove the blue poly-bag protecting the nozzle (let the nozzle hang in the basin below).

6.Compress the waste reservoir by pushing it upward. Slide all four straps (two front & two rear) into the hooks, making sure the straps are not twisted. Visually inspect all four corners of the Bottom Plate, making sure they are equal in height.

WARNING: If waste reservoir is not properly compressed or if the straps are not installed into the hooks, the unit will not function properly on activation. Make sure the area underneath the Eyewash Station is kept clear.

7.Make sure the drain spout is closed. Fig. 3

8.Place the front panel back on the unit and close side panels.

9.In event of alarm activating before station is armed, see ALARM SHUT-OFF PROCEDURE (Below).

10.Open the door and pull the head rest towards you.

11.Pull out the nozzle from the basin. Slide the nozzle into the nozzle compartment, making sure the tube is not twisted and that the small locking hub of the wire tie is facing towards the inside of the unit. Fig. 4

12.Push the headrest back to its original position until an audible ‘click’ is heard. Close the door by pressing the center of the door and a ‘click’ will be heard confirming the nozzle is engaged.

13.Unit is now armed and ready for use.

14.In 3 days come back to the unit to check for leaks or salt residue by the nozzle and at the base of the unit to ensure proper installment.

Alarm Shut-off Procedure

1.Locate small blue button inside door of station.

2.Press and hold button for approximately 3-5 seconds or until alarm shuts-off.

3.If still setting up station, complete Section C.

4.To rearm alarm, close station door. ALARM IS NOW ARMED.

Black Zip-tie

Fig. 1

Fig. 3

Fig. 4

 

Fig. 2

Sperian Eye & Face Protection, Inc. • 825 East Highway 151 • Platteville, WI 53818 USA • 1-800-336-3255

Part # 32-005631 Rev. C

Loading...
+ 2 hidden pages