Honeywell BW Ultra Quick Start Guide

Quick
| Gas Detection
Reference Guide
Honeywell BW ™ Ultra
Portable Five-gas Detector with Internal Pump
Introduction
The Honeywell BW ™ Ultra gas detector warns of hazardous gas at levels above user-defined alarm setpoints.
The detector is a personal safety device. It is your responsibility to respond properly to the alarm.
Monitored Gases
The detector can monitor up to five gases at a time. Refer to the Honeywell BW ™ Ultra User Manual for the full
list of gases.
What’s in the Box
• Honeywell BW ™ Ultra gas detector
• Battery (factory-installed) • Screwdriver telescope with double end
• Charging adapter • 3m PVC tube
• 1 Dust porous filter 7/16” • 2 Fitting mini quick connector to 1/8”
• 5 Pump filters • 2 Fitting male LuerLock to 1/8”
• USB memory stick containing user manuals
• Quick Reference Guide
• 1 screen protector (factory-installed)
1
Safety Information
Use the detector only as specified in this manual, otherwise, the protection provided by the detector may be impaired.
• Only the instrument capable of sounding the alarms and showing readings on a display should be used for immediate safety critical use. Wireless communication and infrastructure are only for informational monitoring.
• Use only Honeywell HUBAT battery pack. Using any other battery can cause an explosion or fire.
• Deactivating the detector by removing the battery pack may cause improper operation and harm the detector.
• Use only Honeywell approved battery chargers such as the MultiUnit Cradle Charger.
• If using the detector near its upper or lower operating temperature, Honeywell recommends zeroing or activating the detector in that environment.
• Charge the detector before first-time use. Honeywell recommends the detector also be charged after every workday.
• Calibrate the device on a regular schedule, depending on use and sensor exposure to poisons and contaminants. Honeywell recommends calibrating at least once every six months.
• For optimal performance, periodically zero the sensor in a normal atmosphere (20.9% v/v O) that is free of hazardous gas.
• The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different combustible gas in the % LEL range, calibrate the sensor using the appropriate gas.
• Only the combustible gas detection portion of this detector has been assessed for performance by CSA International.
• Honeywell recommends that the combustible sensor is checked with a known concentration of calibration gas after any exposure to contaminants/poisons such as sulfur compounds, silicone vapors, halogenated compounds, etc.
• Honeywell recommends that the sensors be bump tested before each day’s use to confirm their ability to respond to gas. Manually verify that the audible, visual, and vibrator alarms are activated. Calibrate if the readings are not within the specified limits.
• Portable safety gas detectors are life safety devices. The accuracy of ambient gas reading(s) is dependent upon factors such as accuracy of the calibration gas standard used for calibration and frequency of calibration.
• The detector is designed for use only in potentially explosive atmospheres where oxygen concentrations do not exceed 20.9% (v/v). Oxygen deficient atmospheres (<10% v/v) may suppress some sensor outputs.
• Extended exposure of the detector to certain concentrations of combustible gases and air may stress the detector element and seriously affect its performance. If an alarm occurs due to a high concentration of combustible gases, a calibration should be performed. If necessary, replace the sensor.
• High concentrations of certain toxic gases, for example, H2S, may hurt the LEL sensor. This effect, known as inhibition, is usually temporary but in extreme circumstances can impair the sensitivity of the LEL sensor after any gas exposure that causes alarm in the toxic gas sensors.
• The Honeywell BW ™ Ultra is provided with an antistatic coating over the LCD window to minimize the risk of ignition due to electrostatic discharge. Periodic inspection of this coating is required to ensure no degradation, delamination, abrasions or other deformities to this surface.
• Care must be taken to avoid exposure to excessive heat, harsh chemicals or solvents, sharp edges and abrasive surfaces. Clean the exterior with a soft, damp cloth.
• The Canadian Standards Association (CSA) requires the LEL sensor to be bump tested before each day’s use with calibration gas containing between 25% and 50% LEL. The instrument must be calibrated if the displayed LEL value during a bump test fails to fall between 100% and 120% of the expected value for the gas.
3
a WARNINGS
• Warning: Substitution of components may impair Intrinsic Safety.
• To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing.
• Protect the combustible sensor from exposure to lead compounds, silicones, and chlorinated hydrocarbons. Although certain organic vapors (such as leaded gasoline and halogenated hydrocarbons) can temporarily inhibit sensor performance, in most cases the sensor will recover after calibration.
• Caution: High off-scale LEL readings may indicate an explosive concentration.
• Any rapid up scaling reading followed by a declining or erratic reading may indicate a gas concentration beyond the upper scale limit, which can be hazardous.
• Products may contain materials that are regulated for transportation under domestic and international dangerous goods regulations. Return product in compliance with appropriate dangerous goods regulations. Contact freight carrier for further instructions.
• The lithium battery may present a risk of fire or chemical burn hazard if misused. Do not disassemble, incinerate, or heat above 212°F (100°C), or incinerate.Batteries must only be charged in a safe area free of hazardous gas. Batteries eexposed to heat at 266°F (130°C) for 10 minutes can cause fire and explosion.
• Dispose of used lithium cells immediately. Do not mix batteries with the solid waste stream. Spent batteries should be disposed of by a qualified recycler or hazardous materials handler.
a MISES EN GARDE
• Avertissement : Le remplacement d’un composant de l’appareil peut altérer sa sécurité intrinsèque.
• Pour éviter l’inflammation d’atmosphères inflammables ou combustibles, couper l’alimentation électrique avant tout entretien.
• Avertissement: Pour réduire le risque d’ignition dans les atmosphères inflammables, les piles doivent être chargées in a dans une zone sûre, exempte de gaz dangereux.
• Protégez le capteur de gaz combustibles contre toute exposition aux composés de plomb, aux silicones et aux hydrocarbures chlorés. Bien que certaines vapeurs organiques (comme l’essence au plomb ou les hydrocarbures halogénés) puissent neutraliser provisoirement les performances du capteur, dans la plupart des cas, le capteur retrouvera son fonctionnement normal après l’étalonnage.
• Attention : Des valeurs LIE hors échelle élevées peuvent indiquer la présence d’une concentration explosive.
• Toute mesure en rapide augmentation suivie d’une diminution ou d’une mesure fantaisiste peut indiquer une concentration de gaz au-delà de la limite d’échelle supérieure, risquant donc d’être dangereuse.
• Les produits peuvent contenir des matériaux qui sont réglementés pour le transport en vertu des règlements nationaux et internationaux de marchandises
• dangereuses. Retourner le produit conformément à la réglementation sur les marchandises dangereuses appropriées. Contactez transporteur pour plus d’instructions.
• Avertissement: La batterie au lithium peut présenter un risque d’incendie ou de brûlure chimique en cas de mauvaise utilisation. Elle ne doit jamais être démontée, incinérée ni chauffée au-delà de 100 °C.
5
• Avertissement: Les piles au lithium polymère exposées à une température supérieure à 130 °C pendant plus de 10 minutes peuvent provoquer un incendie et/ou une explosion.
• Mettez immédiatement au rebut les batteries au lithium usagées. Ne pas les mélanger aux autres déchets solides. Les piles usagées doivent être éliminées par un centre de recyclage agréé ou un centre de traitement des matières dangereuses.
Appearance
1
1. Visual alarm indicator
2. Exhaust port
3. Screen
4. Button
5. Beeper Aperture
6. Pump inlet
7. Alligator clip
8. Pump assembly
9. Charging connector and IR Interface
2
1
3
5
6
1
4
7
8
9
7
Screen Icons
Several icons are displayed in the detector’s screen as conditions dictate.
BW ULTRA
Screen Icons
BLE Pump/header
BLE Pairing Error
Paired Pump critical fail
Pump passed
Pairing failed Warning/failure/
Stealth mode enabled
Battery -three levels
Low battery Hole watch
IR LINK connection
Calibration passed
Calibration failed
Calibration cancelled
error/low battery
Hole watch mode high
Hole watch mode low
Oxygen bar
Correction factor
Peak gas exposure
Press button
Press and hold button
9
Screen Icons
Target gas Inert mode
Bump test cancelled
Bump test passed
Bump test failed
Over limit alarm
High alarm Sensor disabled
Low alarm Heartbeat
Sensor failure
STEL alarm
TWA alarm
Firmware update in progress
Activate the Detector
Activate the detector in a safe area with an atmosphere of
20.9% oxygen and free of hazardous gas.
1. For first time use, charge the battery for up to 8 hours or until LED light turns green using the charging adapter provided. Refer to Charge the battery for more information.
2. Press and hold the button for three seconds.
3. For first time use, a Warming sensors message is displayed and a 30 minutes countdown is displayed. In most cases, this countdown only lasts a couple of minutes.
4. When the detector displays Pump test Block inlet, block the pump inlet with a finger, and then after a couple of seconds unblock the pump inlet. The detector performs a quick pump test. A Pump Test passed message is displayed. If you do not block the pump inlet, the detector will turn off after two minutes. If necessary, you will be instructed by screen prompts to calibrate newly-installed sensors. The detector will then perform a self-test, including testing the sensors. This process could take several minutes.
5. When the self-test is complete, press and hold the button to zero sensors. After the autozero is complete, the detector then checks the sensors for calibration and bump test. If the detector identifies sensors requiring calibration or bump testing, press the button and follow the on-screen instructions.
11
Alarms
When the detector goes into alarm it flashes, vibrates, and produces a loud siren noise. Depending on the type of alarm, these flashes, vibrations, and noises will be different.
There are ten types of alarms.
• Low alarm • Sensor Failure alarm
• High alarm • OverLimit alarm
• TWA alarm • Normal Deactivation
• STEL alarm • Low battery alarm
• Multi alarm • Critical battery alarm
NOTE: In Stealth mode the Honeywell BW ™ Ultra only vibrates.
IMPORTANT: Regardless, when a detector goes into alarm, always take appropriate action. Never ignore or dismiss an alarm.
For more information on the alarms, refer to the Honeywell BW ™ Ultra User Manual.
Navigate the Menu
There are four main menu items.
See Information
Start Bump Test
Zero Sensors
Start Calibration
1. Double pressing the button displays all four options on the screen. See Information is selected and highlighted by default.
2. Press the button to transition the selection to the next choice.
3. Press and hold for three seconds to enter the selected option.
4. Follow on-screen instructions for your selected operation. Most of the detector’s procedures are described in this guide.
13
Standards and Certifications
The Honeywell BW ™ Ultra gas detector is in conformity with the following standards and certifications:
Approvals: Approved by CSA to both U.S. and Canadian Standards
CAN/CSA C22.2 No. 157 and C22.2 No. 152 ANSI/UL - 913 and ANSI/ISA - 12.13.01 Part 1
CSA: Class I, Division 1, Group A, B, C, and D
ATEX: DEMKO 17 ATEX 1833X
II 1 G Ex da ia IIC T4 Ga EN 600790:2012, EN 600791:2014, EN 6007911:2012 Ex ia I Ma, Ex ia IIC T4 Ga (Tamb: 40 °C to +50 °C) Ex ia I Ma, Ex ia d IIC T4 Gb (Tamb: 20 °C to +50 °C) – when
infrared sensor is assembled.
IECEx: Ex da ia IIC T4 Ga IECEx UL 17.0010X IEC 600790:2011, IEC 600791:2014, IEC 60079
11:2011 Ex ia I Ma, Ex ia IIC T4 Ga (Tamb: 40 °C to +50 °C) Ex ia I Ma, Ex ia d IIC T4 Gb (Tamb: 20 °C to +50 °C) – when
infrared sensor is assembled.
KTL: GasAlertMicroClip XT: 12KB4BO00531
Inmetro: DNV 12.0134
cULus: Class I, Div. 1, Groups A, B, C and D with a
Temperature Class of T4
15
Contact Honeywell
Honeywell Analytics 405 Barclay Blvd. Lincolnshire, IL. USA 60069 Toll free. 18887498878
Bwa.customerservice@honeywell.com www.honeywellanalytics.com
Factory Calibration Certificate
Place
Factory
Calibration
Certificate
here
17
© 2018, All rights reserved
50122982349 ENA
Kurzanleitung
| Gasüberwachung
Honeywell BW ™ Ultra
Tragbarer 5Gasdetektor mit interner Pumpe
DEUTSCH
Einleitung
Der Honeywell BW ™ Ultra Gasdetektor dient zur Warnung bei Konzentrationen gefährlicher Umgebungsgase, die über den benutzerseitig festgelegten Alarmeinstellungen liegen.
Der Detektor ist ein personenbezogenes Schutzgerät. Die angemessene Reaktion auf einen Alarm liegt in der Verantwortung des jeweiligen Nutzers.
Überwachte Gase
Der Detektor kann bis zu fünf Gase gleichzeitig überwachen.
Die vollständige Liste der Gase finden Sie im Honeywell BW ™ Ultra-Benutzerhandbuch.
DEUTSCH
1
Lieferumfang
• Honeywell BW ™ Ultra­Gasdetektor
• Batterie (werkseitig installiert)
• Ladeadapter • 3 m PVCRohr
• 1 poröser Staubfilter 7/16"
• 5 Pumpenfilter • 2 Verschraubungen für männlichen
• USBStick mit Benutzerhandbüchern
• Kurzbeschreibung
• Teleskopschraubendreher mit doppeltem Einsatz
• 2 Verschraubungen für Mini­Schnellanschlüsse auf 1/8"
LuerLock auf 1/8"
• 1 Bildschirmschutz (werkseitig installiert)
Sicherheitsinformationen
Den Detektor nur in Übereinstimmung mit diesen Bedienungshinweisen verwenden. Bei Nichtbeachtung wird möglicherweise die Schutzfunktion des Gerätes beeinträchtigt.
• Nur das Gerät, das die Alarmtöne ausgibt und Werte auf einem Display anzeigt, darf für die sofortige sicherheitsrelevante Nutzung verwendet werden. Drahtlose Kommunikation und Infrastruktur sind nur für die Überwachung von Informationen vorgesehen.
• Nur Honeywell HUBAT Akkupack verwenden. Die Verwendung einer anderen Batterie kann zu einer Explosion oder einem Feuer führen.
• Eine Deaktivierung des Detektors durch Entnehmen des Akkupacks kann zu Funktionsstörungen und Beschädigungen des Geräts führen.
• Nur von Honeywell genehmigte Batterieladegeräte verwenden, wie etwa die Ladestation für mehrere Geräte.
• Wenn Sie das Gerät in der Nähe seiner oberen oder unteren Betriebstemperatur verwenden, empfiehlt Honeywell die Nullpunktjustage oder das Einschalten des Gaswarngeräts in dieser Umgebung.
• Laden Sie den Detektor vor dem ersten Einsatz auf. Honeywell empfiehlt, den Detektor nach jedem Arbeitstag aufzuladen.
• Kalibrieren Sie das Gerät je nach Gebrauch und Höhe der gemessenen toxischen Gas- und Schadstoffkonzentrationen in regelmäßigen Abständen. Honeywell empfiehlt eine Kalibrierung mindestens alle sechs Monate.
• Um optimalen Betrieb zu gewährleisten, setzen Sie den Sensor in regelmäßigen Abständen und in normaler Umgebung (20,9 % v/v O) ohne gefährliches Gas zurück.
• Der Sensor ist werkseitig auf 50 % UEG Methan kalibriert. Kalibrieren Sie den Sensor vor der Überwachung eines anderen brennbaren Gases im % UEGBereich mit dem entsprechenden Gas.
• Es wurde nur die Leistung für den zur Detektion brennbarer Gase bestimmten Teil dieses Detektors durch CSA International beurteilt.
DEUTSCH
3
• Honeywell empfiehlt, den Sensor für brennbare Gase nach jedem Kontakt mit Gefahrenstoffen/toxischen Gasen wie schwefelhaltigen Verbindungen, Siliziumdämpfen, halogenhaltigen Verbindungen usw. mit einem Kalibriergas bekannter Konzentration zu überprüfen.
• Honeywell empfiehlt Funktionstests der Sensoren vor der täglichen Nutzung, um ihre Reaktion auf Gas zu überprüfen. Prüfen Sie manuell, ob die akustischen, optischen und Vibrationsalarme tatsächlich aktiviert werden. Wenn die Messwerte nicht innerhalb der angegebenen Grenzwerte liegen, sollte eine Kalibrierung durchgeführt werden.
• Tragbare Gasdetektoren sind lebensrettende Sicherheitseinrichtungen. Die Genauigkeit der Umgebungsgasmesswerte hängt von Faktoren ab wie etwa der Genauigkeit des für die Kalibrierung verwendeten Kalibriergasstandards und der Häufigkeit der Kalibrierung.
• Der Detektor ist nur für den Einsatz in explosionsgefährdeten Atmosphären
mit einer Sauerstoffkonzentration von maximal 20,9 Vol.-% vorgesehen. Sauerstoffarme Atmosphären (weniger als 10 Vol.-%) können einige
Sensorausgaben unterdrücken.
• Häufiger oder lang andauernder Kontakt des Detektors mit bestimmten Konzentrationen brennbarer Gase kann zur Abnutzung von Detektorelementen und damit zu einer schwerwiegenden Beeinträchtigung der Funktionsbereitschaft führen. Nach Auftreten eines Alarms aufgrund erhöhter Konzentrationen brennbarer Gase sollte das Gerät kalibriert werden. Ggf. den Sensor ersetzen.
• Hohe Konzentrationen bestimmter toxischer Gase, zum Beispiel H2S, können den LELSensor beschädigen. Dieser als Hemmung bekannte Effekt ist gewöhnlich vorübergehend. In Extremsituationen kann er jedoch die Empfindlichkeit des LELSensors nach einem Kontakt mit einem Gas beeinträchtigen, der einen Alarm der Sensoren für toxische Gase auslöst.
• Der Honeywell BW ™ Ultra weist eine antistatische Beschichtung über dem LCD Bildschirm auf, um die Entzündungsgefahr durch elektrostatische Entladung zu minimieren. Eine regelmäßige Überprüfung dieser Beschichtung ist erforderlich, um zu gewährleisten, dass diese Oberfläche in keiner Weise beschädigt oder in ihrer Funktion beeinträchtigt ist (Zersetzung, Abblättern, Abrieb oder andere Fehler).
• Der Kontakt mit übermäßiger Wärme, aggressiven Chemikalien oder Lösungsmitteln, scharfen Kanten und rauen Oberflächen muss vermieden werden. Reinigen Sie das Gerät äußerlich mit einem feuchten, weichen Tuch.
• Die Canadian Standards Association (CSA) fordert Funktionstests des LELSensors vor der täglichen Nutzung mit einem Kalibriergas, das zwischen 25 % und 50 % LEL enthält. Das Gerät muss kalibriert werden, wenn der angezeigte LELWert während eines Funktionstests nicht zwischen 100 % und 120 % des erwarteten Werts für das Gas fällt.
DEUTSCH
5
a WARNUNGEN
• Warnung: Der Austausch von Komponenten kann die Eigensicherheit des Geräts beeinträchtigen.
• Trennen Sie vor der Durchführung von Wartungsarbeiten die Stromversorgung, um eine Entzündung brennbarer Atmosphären zu verhindern.
• Schützen Sie den Sensor vor Kontakt mit bleihaltigen Verbindungen, Silikonen und hohen Konzentrationen an chlorierten Kohlenwasserstoffen. Bestimmte organische Dämpfe (z.B. verbleites Benzin und halogenhaltige Kohlenwasserstoffe) können die Sensorfunktion vorübergehend beeinträchtigen. In den meisten Fällen ist der Sensor jedoch nach einer Kalibrierung wieder einsatzbereit.
• Achtung: Werte weit oberhalb des UEGMessbereichs können auf explosive Konzentrationen hinweisen.
• Ein schnell steigender Messwert gefolgt von fallenden oder unregelmäßigen Messwerten weist möglicherweise auf eine gefährliche Gaskonzentration über dem oberen Grenzwert hin.
• Produkte können Materialien enthalten, für deren Transport nationale und internationale Vorschriften zur Beförderung von Gefahrengütern gelten. Befolgen Sie bei der Rücksendung von Produkten die entsprechenden Vorschriften. Weitere Informationen erhalten Sie vom Transportunternehmen.
• Bei unsachgemäßem Gebrauch der Lithiumbatterie besteht möglicherweise Feuer- oder Verätzungsgefahr. Nicht zerlegen, verbrennen oder über 100°C erwärmen. Batterien dürfen nur in sicheren Bereichen ohne gefährliche Gase aufgeladen werden. Werden Batterien länger als 10 Minuten einer Temperatur von 130 °C ausgesetzt, besteht Brand- und Explosionsgefahr.
• Gebrauchte Lithiumzellen sind umgehend zu entsorgen. Entsorgen Sie Batterien nicht im normalen Hausmüll. Leere Akkus sind von einer qualifizierten Recyclingeinrichtung oder Sammelstelle für Sondermüll zu entsorgen.
DEUTSCH
7
Aufbau
1
2
1. Optische Alarmanzeige
2. Abluftanschluss
3. Abschirmungs-
4. Taste
5. Tonsignal­Öffnung
6. Pumpeneinlass
7. Krokodilclip
8. Pumpen­baugruppe
9. Anschluss des
Ladegeräts und der IRSchnittstelle
1
3
5
6
1
4
7
8
9
Loading...
+ 110 hidden pages