Honeywell APT2000, APT2000TC-0-00, APT2000TC-0-IS, APT2000TC-H-00, APT2000TC-H-IS User Manual

Page 1
APT2000 Series
70-82-25-96
MU1I-6251
Revision 1 – 03/00
2-Wire Toroidal
Conductivity Transmitters
User Manual
Page 2
TA-194.350-HWE01 210700
Copyright, Notices, and Trademarks
E Copyright 1999 by Honeywell Inc.
Revision 1 – 03/00
While this information is presented in good faith and believed to be accurate, Honeywell disclaims the implied warranties of merchantability and fitness for a
particular purpose and makes no express warranties except as may be stated in
its written agreement with and for its customer.
In no event is Honeywell liable to anyone for any indirect, special or consequen­tial damages. The information and specifications in this document are subject to
change without notice. Honeywell
Industrial Automation and Control Automation College 1100 Virginia Drive Ft. Washington, PA. 19034
Honeywell S. A. Espace Industriel Nord rue André Durouchez 80084 Amiens Cedex 2 France
Contacts
The following list identifies important contacts within Honeywell.
Organization Telephone Address
Honeywell Technical Assistance Center
1-800-423-9883 (USA and Canada)
1100 Virginia Avenue Fort Washington, PA 19034
Honeywell S.A. 33-3-22-54-56-56
(Europe)
80084 Amiens Cedex 2 France
Page 3
Warning
Warning
Warning
Information
3
Safety Precautions
Be sure to read and observe the following requirements!
The APT2000TC-0(H)-00 Transmitter is approved for operation in safe areas and hazardous locations DIV 2 (USA/Canada only). Before connecting the Transmitter to a power supply unit, make sure that this is not capable of outputting more than 40 Vdc (safe areas) / 30 Vdc (DIV 2).
The APT 2000PH-0(H)-IS Transmitter is approved for operation in hazardous locations DIV 1 (USA/Canada) / Zone 1 (Europe). Before connecting the Transmitter to a power supply unit, make sure that this is an associ­ated apparatus.
The measuring inputs of the APT 2000PH-0(H)-IS Transmitter may be led into Zone 0 (Europe). However, be sure to observe the national regulations concerning Zone 0 applications. The Transmitter itself is not approved for operation in Zone 0!
Whenever it is likely that the protection has been impaired, the Transmitter shall be made inoperative and secured against unintended operation. The protection is likely to be impaired if, for example:
the Transmitter shows visible damagethe Transmitter fails to perform the intended measurementsafter prolonged storage at temperatures above 70 °Cafter severe transport stresses
Before recommissioning the T ransmitter, a professional routine test must be performed. This test should be carried out at our factory.
The Transmitter shall not be used in a manner not specified by this manual.
Page 4
CAL
Note
Warning
CONF
CAL
Information
4
Information on this Instruction Manual
ITALICS are used for texts which appear in the APT2000TC Transmitter display.
Bold print is used to represent keys, e.g. CAL.
Keys for which the functions are explained are frequently shown in the left-hand column.
Notes provide important information that should be strictly followed when using the Transmit­ter.
Warning means that the instructions given must always be followed for your own safety. Failure to follow these instructions may result in injuries.
Mode Codes
After pressing CONF or CAL you can enter one of the following codes to access the desig- nated mode:
CONF, 0000: Error info CONF, 1200: Configuration CONF, 5555: Current source
CAL, 0000: Cal info CAL, 1001: Zero point calibration CAL, 1015: Temp probe adjustment CAL, 1 100: Cell factor calibration CAL, 1 125: Input/adjustment of sensor factor CAL, 2222: Test mode
Page 5
Contents
5
Contents
Safety Precautions 3. . . . . . . . . . . . . . . . . . . . . .
Information on this Instruction Manual 4. . .
Mode Codes 4. . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 Assembly 6. . . . . . . . . . . . . . . . . . . . . . . . . . .
Package Contents and Unpacking 6. . . .
Assembly 6. . . . . . . . . . . . . . . . . . . . . . . . . .
2 Installation, Connection
and Commissioning 10. . . . . . . . . . . . . . . . .
Proper Use 10. . . . . . . . . . . . . . . . . . . . . . . .
Overview of the
Conductivity Transmitter 1 1. . . . . . . . . . . . .
Terminal Assignment 12. . . . . . . . . . . . . . . .
Installation and Commissioning 13. . . . . . .
Typical Wiring 14. . . . . . . . . . . . . . . . . . . . . .
3 Operation 15. . . . . . . . . . . . . . . . . . . . . . . . . . .
User Interface 15. . . . . . . . . . . . . . . . . . . . . .
Display 16. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Keypad 16. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Safety Functions 17. . . . . . . . . . . . . . . . . . . .
Outputs 18. . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuration 19. . . . . . . . . . . . . . . . . . . . . . .
Calibration 21. . . . . . . . . . . . . . . . . . . . . . . . .
Measurement 25. . . . . . . . . . . . . . . . . . . . . .
4 Diagnostics, Maintenance and Cleaning 26
Sensoface
, Sensocheck
26. . . . . . . . . .
Error Messages 26. . . . . . . . . . . . . . . . . . . .
Diagnostics Functions 28. . . . . . . . . . . . . . .
Maintenance and Cleaning 29. . . . . . . . . . .
5 Appendix 30. . . . . . . . . . . . . . . . . . . . . . . . . . .
Product Line 30. . . . . . . . . . . . . . . . . . . . . . .
Specifications 31. . . . . . . . . . . . . . . . . . . . . .
Type Examination Certificate 35. . . . . . . . .
Calibration Solutions 38. . . . . . . . . . . . . . . .
Concentration Curves 39. . . . . . . . . . . . . . .
Index 45. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Page 6
Assembly
6
1 Assembly
Package Contents and Unpacking
Unpack the unit carefully and check the shipment for transport damage and completeness. The package contains:
–Front unit of Transmitter –Lower case –Short instruction sheet –This instruction manual –HART description
(only for Model APT2000TC-H-..)
–Bag containing:
2 plastic plugs 1 hinge pin 5 hexagon nuts 3 cable ties 3 Pg cable glands 3 filler plugs 1 rubber reducer 3 sealing rings 1 Pg plug
11 1 metal plate
4 set screws for conduit
12
1 jumper
Assembly
Fig. 1 Assembling the case
(For sealing in case
of wall mounting)
(Can be inserted from either side)
11
For conduit mounting: Place washer between enclosure and nut.
11
12
Page 7
Assembly
7
80 [3.15]
32 [1.26]
72 [2.83]
6.2 [0.24]
approx. 14
84 [3.31]
42 [1.65]
Pg 13.5 (3 pcs.)
144 [5.67]
144 [5.67]
21
43 [1.69]
105 [4.13]
27
Control panel 1 – 22 mm
max. 25
78 [3.07]
27
Control panel cutout 138 x 138 mm (DIN 43700)
Panel-mount kit consisting of:
4 screws 4 span pieces 4 threaded sleeves 1 seal
➀ ➁ ➂ ➃
Holes for post mounting
Holes for wall mounting
(4 x)
(2 x)
[0.55]
[1.06]
[0.83]
[0.98]
[1.06]
[5.43 x 5.43]
[0.04 – 0.87]
Note: All dimensions
in mm [inches]
(not included in supply)
2 holes 21.5 mm dia. [8.47] for Pg 13.5
I
or 1/2 conduit
Fig. 2 Dimension drawing for Transmitter, mounting diagram and P/N 51205990-001 panel-mount kit
Page 8
Assembly
8
horizontal post/
40–60 mm dia.
51205989-001
(if required)
For vertical or pipe mounting
Pipe-mount kit consisting of:
4 self-tapping screws 1 pipe mounting plate
2 hose clamps with
➀ ➁
worm gear drive to DIN 3017
[1.57–2.36]
protective hood
Fig. 3 P/N 51205988-001 pipe-mount kit
165 [6.5]
173 [6.81]
132 [5.2]
Fig. 4 P/N 51205989-001 protective hood for wall and pipe mounting
Page 9
Assembly
9
280 [11.03]
7 [0.28]
Recommended stripping lengths for
multi-core cables
Recommended stripping length for
coaxial cable
Pulling out the terminals using a
screwdriver (also see
➆)
Cable laying in the Transmitter
Connecting lines for loop current Cover for sensor and temperature
probe terminals
Areas for placing the screwdriver to
pull out the terminals
Connection of handheld terminal
Be sure not to
notch the cable cores when stripping the insulation!
Dimensions in mm [inches].
➄ ➅
Coaxial cable
13 mm [0.5]
7 mm [0.28]
Fig. 5 Installation information
Page 10
Warning
Warning
Warning
Capabilities, Connection
10
2 Installation, Connection and Commissioning
Proper Use
The APT2000TC Transmitter is used for conductivity and temperature measurement in chemical, pulp and paper , biotechnology, food processing, pharmaceuti­cal, electroplating, and water/wastewater industries. It can either be mounted on site or in a control panel.
The APT2000TC-0(H)-00 Transmitter is approved for opera­tion in safe areas and hazardous locations DIV 2 (USA/Canada only). Before connecting the Transmitter to a power supply unit, make sure that this is not capable of outputting more than 40 Vdc (safe areas) / 30 Vdc (DIV 2).
The APT2000TC-0(H)-IS Transmit­ter is approved for operation in haz­ardous locations DIV 1 (USA/Can­ada) / Zone 1 (Europe). Before connecting the Transmitter to a power supply unit, make sure that this is an associated appara­tus.
The measuring inputs of the APT 2000PH-0(H)-IS Transmitter may be led into Zone 0 (Europe). However, be sure to observe the national regulations concerning Zone 0 applications. The Transmitter itself is not ap­proved for operation in Zone 0!
Page 11
Capabilities, Connection
11
Overview of the Conductivity Transmitter
mS/cm, % by wt, SAL
Temperature Pt 100/1000/ NTC 100 kW
mA
APT2000TC Transmitter
Fig. 6 System functions of APT2000TC Transmitter
Inputs for toroidal conductivity sensorsInput for temperature probeCurrent loop 4 – 20 mA,
transports power to and output signal from the transmitter,
with APT2000TC-H-.. Transmitter also for HARTr communication
Equipotential bonding (only with
APT2000TC-0(H)-IS Transmitter)
Page 12
Capabilities, Connection
12
Terminal Assignment
8 7 6 5 4 2 1
3
T1
temp
n.c.
+ 11 10
4 to
+
9
output
HART
20 mA
supply/
do not
connect
toroidal
conductivity sensor
lo hilohi
send receive
WT
BL
YW
GN
shield
RD
Fig. 7 Terminal assignment of APT2000TC-0(H)-00 Transmitter
NI, Class 1, Div 2, Group A – D, T4
8 7 6 5 4 2 1
3
T1
temp
n.c.
+ 11 10
4 to
+
9
output
HART
20 mA
supply/ toroidal
conductivity sensor
lo hilohi
send receive
WT
BL
YW
GN
shield
RD
Fig. 8 Terminal assignment of APT2000TC-0(H)-IS T ransmitter
IS, Class I, Div 1, Group A – D, T4 II 2(1) G EEx ib [ia] IIC T6
Page 13
Warning
Warning
Warning
Warning
Warning
Capabilities, Connection
13
Installation and Commissioning
Installation and commissioning of the Transmitter may only be carried out in accordance with this instruc­tion manual and per applicable local and national codes. Be sure to ob­serve the technical specifications and input ratings.
Before connecting the APT2000TC-0(H)-00 Transmitter to a power supply unit, make sure that this is not capable of outputting more than 40 Vdc (safe areas) / 30 Vdc (DIV 2).
Before connecting the APT2000TC-0(H)-IS Transmitter to a power supply unit, make sure that this is an associated apparatus (for input ratings refer to the Control Drawing or the annex of the EC Type Examination Certificate).
Do not use alternating current or mains power supply!
When commissioning, a complete configuration must be carried out.
For easier installation, the terminal strips are of a plug-in design. The terminals are suitable for single wires and flexible leads up to 2.5 mm
2
(AWG 14) (see Pg. 9). A connection example is shown on Pg. 14.
Page 14
Warning
Note
Capabilities, Connection
14
Typical Wiring
Conductivity measurement with Honeywell 5000TC toroidal conductivity sensor
The Honeywell 5000TC toroidal conductivity sensor is used to measure low to highest conductivity val­ues. It can be used for measurements in safe areas.
WHITE
BLUE
YELLOW
GREEN
RED
SHIELD
toroidal
cond. sensor
5421
3
temp
lohi lo hi
sendreceive
WT
BL
YW
GN
shield
RD
87
6
T1
T2
n.c.
Fig. 9 Conductivity measurement with Honeywell
5000TC toroidal conductivity sensor
The Honeywell 5000TC toroidal conductivity sensor may only be used in combination with the APT2000TC-0(H)-00 Transmitter.
For special mounting conditions of the sensor, the cell factor can vary between 4.0 and 4.5. Therefore the user should perform a wet calibra­tion of each new sensor to deter­mine the exact cell factor.
Settings for Honeywell 5000TC toroidal conductivity sensor
Menu Setting
Temp probe conf 1200 Pt 1000 Cell factor cal 1100 4.44
Page 15
Operation
15
3 Operation
User Interface
MEAS
Alarm LED
Keypad
Model designation
Rating plate
Status fields:
- Measuring mode
- Calibration mode
- Alarm
- Configuration mode
CAL ALARM CONF
CONF ENTER
CAL
"
"
Fig. 10 Front view of Transmitter
Page 16
CAL
CONF
"
Y
ENTER
CAL
ENTER
CONF
ENTER
+
Y
"
Operation
16
Display
Mode code entry
Interval/
Response time
Hold state
active
Continue with
ENTER
Measurement symbols
Tempera-
ture
Loop
current
Alarm
setting
Senso­check
Calibration
Sensor
data
Sensoface
Wait
Display/
Process variable
Bar for instrument status
Fig. 11 Display of Transmitter
Keypad
Start, end calibration
Start, end configuration
Select digit position (selected position flashes)
Change digit
Prompt in display: continue in program sequence, Configuration:
Confirm entries, next configuration step, Measuring mode:
Display loop cur-
rent
Cal info, display cell factor and zero point (see Pg. 25)
Error info, display last output error message (see Pg. 25)
Start GainCheck
manual instru-
ment self-test (see Pg. 17)
Page 17
+
Y
"
Operation
17
Safety Functions
Sensoface sensor monitoring
Sensoface
provides information on the sensor condition. A sad “Smiley” indicates that there is a Sensocheck
message.
Sensocheck
signals a short cir­cuit of the primary coil and its lines as well as an interruption at the secondary coil and its lines. Senso­check
can be switched off. With
Sensocheck
switched off, no
friendly Smiley appears.
For more detailed information, see chapter “Diagnos­tic, Maintenance and Cleaning” (Pg. 26).
GainCheck
manual instrument self-test
Simultaneously pressing Y and " starts the manual instrument self­test.
A display test is carried out, the software version is displayed and the memory and measured value transmission are checked.
Automatic instrument self-test
The automatic unit self-test checks the memory and the measured-value transmission. It runs automati­cally in the background at fixed intervals.
Page 18
Operation
18
Outputs
Current loop (4 to 20 mA)
The current loop transports power to and output sig­nals from the Transmitter. The current is controlled by the process variable selected in the configuration. The current characteristic can be configured as lin­ear or logarithmic curve for conductivity and resistiv­ity. The current beginning and end can be set to repre­sent any desired value. If LIN (linear characteristic) is chosen, the minimum span is 5% of the selected process variable / mea­surement range. If LOG (logarithmic characteristic) is chosen, the minimum span is one decade within the chosen range. To check connected peripherals (e.g. limit switches, controllers), the loop current can be manually speci­fied (see Pg. 28).
HART
communication
The APT2000TC-H-.. Transmitter can be remote controlled via HART
communication. It can be con­figured using a handheld terminal or from the control room. Measured values, messages and instrument identification can be downloaded at any time. This allows easy integration also in fully automatic pro­cess sequences.
For more detailed information, refer to the HART
Command Specification.
Alarm
During an error message the alarm LED flashes. Alarm response time is permanently set to 10 sec.
Error messages can also be signaled with a 22 mA signal via the loop current (see Configuration, Pg. 20).
Page 19
CONF
Note
Operation
19
Configuration
The instrument arrives from the factory configured and ready to operate as a conductivity transmitter. This section provides detailed procedures for chang­ing operation values for specific applications.
Activate with CONF change parameter with Y and ", confirm/continue with ENTER, end with CONF.
Mode code “1200”
During configuration the Transmitter is in the Hold state, the loop current is frozen.
When the configuration mode is exited, the Transmit­ter remains in the Hold state for safety reasons. This prevents undesirable reactions of the connected pe­ripherals (e.g. limit switches, controllers) due to in­correct settings. The measured value and Hold are displayed alternately. Now you can check whether the measured value is plausible and specifically end the Hold state with ENTER. After a relax time of 20 sec (for measured value stabilization) the Trans­mitter returns to measuring mode.
The configuration parameters are checked during the input. In the case of an incorrect input “ERR” is displayed for 3 sec. The parame­ters cannot be stored with ENTER until the input has been repeated.
Configuration parameters
Before attempting any changes refer to the parameter setup list shown below. This table presents the possible options and the factory settings.
Picto­graph
Parameter Choices Factory
setting
Process variable / meas. range Selected process variable and
measuring range control loop cur-
rent, limit values and display. Complete configuration required after change.
oo.oo mS / ooo.o mS / oooo mS ooo.o % ooo.o SAL
ooo.o mS
Concentration (only with %)
-01- NaCl (0 – 28 % by wt)
-02- HCl (0 – 17 % by wt)
-03- NaOH (0 – 22 % by wt)
-04- H2SO4(0 – 35 % by wt)
-05- HNO
3
(0 – 28 % by wt)
-06- H
2SO4
(95 – 99 % by wt)
-01-
Temperature display °C
°F
°C
Page 20
Operation
20
Temperature probe Pt 100 / Pt 1000 /
NTC 100 kW
Pt 1000
T emperature compensation (not with % and SAL)
OFF LIN NLF (natural waters)
OFF
Temperature coefficient (only with tc LIN)
xx.xx %/K 02.00 %/K
Current characteristic (not with % and SAL)
LIN LOG
LIN
Current beginning (4 mA) (only with LIN)
mS / % / SAL 000.0 mS
Current end (20 mA) (only with LIN)
mS / % / SAL
100.0 mS
Current beginning (4 mA) (only with LOG)
mS *
0.1 mS
Current end (20 mA) (only with LOG)
mS *
100 mS
Hold state Last:Last current value
Fix: Current specified
Last
Hold value (only with Fix)
xx.xx mA 21.00 mA
22 mA signal for error message ON / OFF OFF Sensocheck
ON / OFF OFF
* 0.1 / 1 / 10 / 100 / 1000 mS Configuration is cyclical. To stop, press CONF.
Page 21
CAL
CAL
ENTER
Operation
21
Calibration
In the calibration mode the cell factor can be modi­fied in two ways. If the cell factor of the sensor in use is known under consideration of the installation con­ditions, it can be entered directly. Furthermore, the cell factor can be determined with a known calibra­tion solution under consideration of the temperature.
Activate with CAL, confirm/continue with ENTER, abort with CAL ENTER
During calibration the Transmitter is in the Hold state. The loop current is frozen.
When the calibration mode is exited, the Transmitter remains in the Hold state for safety reasons. This prevents undesirable reactions of the connected pe­ripherals (e.g. limit switches, controllers) due to in­correct settings. The measured value and Hold are displayed alternately. Now you can check whether the measured value is plausible and specifically end the Hold state with ENTER or repeat calibration with CAL. When you end the Hold state, the Transmitter will return to measuring mode after 20 sec (mea­sured value stabilization).
Calibration by input of cell factor
Activate calibration by pressing the CAL key. Using the Y , " keys enter mode code “1100” and then press ENTER.
Using the Y , " keys enter the cell factor. The lower display shows the conductivity value.
A change in the cell factor also changes the conductivity value.
When there has not been an entry for approx. 6 sec, conductivity and temperature are displayed alter­nately.
Press ENTER to confirm the cell factor.
The Transmitter remains in the Hold state. You can end the Hold state with ENTER. After 20 sec (mea­sured value stabilization) the Trans­mitter returns to measuring mode.
Page 22
Note
CAL
ENTER
Operation
22
Calibration with calibration solution
Be sure to use known calibration solutions and the respective tempe­rature-corrected table values (see Calibration Solutions, Pg. 38).
Activate calibration by pressing the CAL key. Using the Y , " keys enter mode code “1100” and then press ENTER.
Immerse the sensor in the calibra­tion solution.
After approx. 6 sec the lower dis­play alternately shows the conduc­tivity and temperature values. Read the conductivity value correspond­ing to the displayed temperature from the table of the used calibra­tion solution (for tables see Pg. 38).
Using the Y , " keys change the cell factor until the display shows the conductivity value from the table.
Make sure that the temperature is stable during the calibration proce­dure.
Press ENTER to confirm the cell factor.
The Transmitter remains in the Hold state. You can end the Hold state with ENTER. After 20 sec (mea­sured value stabilization) the Trans­mitter returns to measuring mode.
Page 23
Note
Note
CAL
ENTER
Note
Note
CAL
Operation
23
Zero point calibration in air
Zero point calibration is only re­quired when very low conductivity values are to be measured.
Before you start calibration, remove the sensor from the process, clean it and dry it up.
Activate calibration by pressing the CAL key. Using the Y , " keys enter mode code “1001” and then press ENTER.
Using the Y , " keys modify the zero point until the lower display reads 0 mS. If required, change the sign of the zero point!
When there has not been an entry for approx. 6 sec, the lower display alternately shows the zero-cor­rected conductivity value and the temperature value.
Press ENTER to confirm the zero point.
The Transmitter remains in the Hold state. You can end the Hold state with ENTER. After 20 sec (mea­sured value stabilization) the Trans­mitter returns to measuring mode.
Input and adjustment of sensor factor
This function should only be used by experts. Incorrectly set parame­ters may go unnoticed, but change the measuring properties.
The Transmitter comes with a preset sensor factor of
24.6 for the 5000TC sensor. Should you use another sensor , you must enter another sensor factor or de­termine it using a comparison resistor. After that, you can calibrate the sensor (see Pg. 21).
Resistance measurement in test mode can only show the correct value of the test resistor when the sensor factor has been correctly determined.
Activate calibration by pressing the CAL key. Using the Y , " keys enter mode code “1125” and then press ENTER.
Using the Y , " keys enter the sensor factor of the sensor in the main display .
If you do not know the sensor fac­tor, it can be determined using a comparison resistor (recommended resistance value: 100 W). The sen­sor factor must be adjusted until the corresponding resistance value is shown in the lower display.
Page 24
ENTER
Note
CAL
ENTER
Operation
24
Press ENTER to confirm the sen­sor factor.
The Transmitter remains in the Hold state. You can end the Hold state with ENTER. After 20 sec (mea­sured value stabilization) the Trans­mitter returns to measuring mode.
Adjustment of temperature probe
This function should only be used by experts. Incorrectly set parame­ters may go unnoticed, but change the measuring properties. Especially for Pt 100 temperature probe, it is advisable to perform an adjustment.
Activate calibration by pressing the CAL key. Using the Y , " keys enter mode code “1015” and then press ENTER.
Measure the temperature of the process medium using an external thermometer.
Using the Y , " keys enter the de­termined temperature value in the main display. If you take over the temperature value shown in the lower display, the correction is with­out effect.
Press ENTER to confirm the tem­perature value.
The Transmitter remains in the Hold state. You can end the Hold state with ENTER. After a relax time of 20 sec (for measured value stabi­lization) the Transmitter returns to measuring mode.
Page 25
Note
Operation
25
Measurement
Measuring mode
In the measuring mode the main display shows the configured process variable and the lower display the temperature.
Cal info
With CAL and mode code “0000” you can activate the cal info. Cal info shows the current calibration data for approx. 20 sec. The 20 sec can be reduced by pressing ENTER. During cal info the Transmitter is not
in Hold state.
Error info
With CONF and mode code “0000” you can activate the error info. Error info shows the most recent error message for approx. 20 sec. After that the message will be deleted. The 20 sec can be reduced by press­ing ENTER. During error info the Transmitter is not in Hold state.
Hold state
The Transmitter will enter the Hold state under the following conditions:
For calibration: Mode code 1001
Mode code 1015 Mode code 1100 Mode code 1125 Mode code 2222
configuration: Mode code 1200
Mode code 5555
The loop current is frozen at Last or Fix (configura­tion Pg. 20). If the calibration or configuration mode is exited, the Transmitter remains in the Hold state for safety rea­sons. This prevents undesirable reactions of the con­nected peripherals (e.g. limit switches, controllers) due to incorrect settings. The measured value and Hold are displayed alternately. Now you can check whether the measured value is plausible and specifi­cally end the Hold state with ENTER. After a relax time of 20 sec (for measured value stabilization) the Transmitter returns to measuring mode.
During error conditions the Hold state will not be active.
Page 26
CONF
ENTER
Troubleshooting, Cleaning
26
4 Diagnostics, Maintenance and Cleaning
Sensoface, Sensocheck
Sensoface provides information
on the sensor condition. A sad “Smiley” indicates that there is a Sensocheck
message.
Sensocheck
signals a short circuit of the primary coil and its lines as well as an interruption at the sec­ondary coil and its lines. Sensocheck
can be
switched off. With Sensocheck
switched off, no
friendly Smiley appears.
Error Messages
When one of the following error messages is output, the Transmitter can no longer correctly determine the process variable or output it via the loop current.
During an error message the alarm LED flashes. The alarm response time is permanently set to 10 sec.
Error messages can also be signaled with a 22 mA signal via the loop current (see Configuration, Pg. 20).
Error info
With CONF and mode code “0000” you can activate the error info. Error info shows the most recent error message for approx. 20 sec. After that the message will be dele­ted. The 20 sec can be reduced by pressing ENTER. During error info the Transmitter is not
in Hold state.
Error number
Display (flashing)
Problem Possible causes
Err 01 Sensor - Wrong cell factor
- Outside measurement range
- SAL > 45 ‰
- Sensor connection or cable defective Err 02 Sensor - Unsuitable sensor Err 03 Temperature probe - Outside temp range
- Outside temp range for TC
- Outside temp range for SAL
- Outside temp range for concentration Err 21 Loop current - Meas. value below configured current beginning
- Wrong configuration for current beginning (see Pg. 20)
Page 27
Troubleshooting, Cleaning
27
Error number
Possible causesProblemDisplay
(flashing)
Err 22 Loop current - Meas. value above configured current end
- Wrong configuration for current end (see Pg. 20) Err 23 Loop current - Configured current span too small
(Difference between current beginning and end)
Err 33 Sensocheck
- Short circuit in primary coil
- Short circuit of cable Err 34 Sensocheck
- Open circuit in secondary coil
- Cable interrupted Err 98 System error - Configuration or calibration data defective; completely
reconfigure and recalibrate the Transmitter
- Measured value transmission defective
- Memory error in Transmitter program (PROM defective)
Err 99 Factory settings - EEPROM or RAM defective
- Error in factory settings
This error message normally should not occur, as the data are protected from loss by multiple safety functions. Should this error message nevertheless occur, there is no remedy. The Transmitter must be repaired and recali­brated at the factory.
Page 28
Warning
Troubleshooting, Cleaning
28
Diagnostics Functions
Cal info
Pressing CAL and entering mode code “0000” is going to activate the cal info. Cal info shows the cur­rent calibration data for approx. 20 sec. During cal info the Transmitter is not
in Hold state.
Test mode
Pressing CAL and entering mode code “2222” is going to activate the test mode. In the test mode you can check the measuring equipment with a resistor. Sensoface
is disabled.
To do so, a comparison resistor is looped through the sensor. The comparison resistance value is indi­cated in the main display in kW. When the resistance value exceeds 2 kW, the display shows “– – – –”.
R: e.g. 100 W
Pressing ENTER ends the test mode. The T ransmit­ter goes to Hold state.
Error info
Pressing CONF and entering mode code “0000” is going to activate the error info. Error info shows the most recent error message for approx. 20 sec. After that the message will be deleted. During error info the Transmitter is not
in Hold state.
Display loop current
Pressing ENTER in measuring mode displays the loop current for a few seconds.
Current source
To check the connected peripherals (e.g. limit switches, controllers), the loop current can be manu­ally specified.
In the current source mode the loop current no longer follows the mea­sured value! It is manually speci­fied. Therefore, it must be ensured that the connected devices (control room, controllers, indicators) do not interpret the current value as a measured value!
Pressing CONF and entering mode code “5555” is going to activate the current source mode. Specify the loop current using ", Y and ENTER. The actu­ally flowing loop current is shown in the lower dis­play. Pressing CONF exits the current source mode again.
GainCheck
manual instrument self-test
The manual instrument self-test is started by simulta­neously pressing Y and ".
A display test is carried out, the software version is displayed and the memory and measured-value transmission are checked.
Page 29
Troubleshooting, Cleaning
29
Automatic self-test
The automatic self-test checks the memory and the measured-value transmission. It runs automatically in the background at fixed intervals.
Maintenance and Cleaning
Maintenance
The Transmitter contains no user repairable compo­nents. If problems persist even after reviewing sec­tion 4, please contact the factory.
Cleaning
To remove dust, dirt and spots, the external surfaces of the Transmitter may be wiped with a damp, lint­free cloth. A mild household cleaner may also be used if necessary .
Page 30
Specifications
30
5 Appendix
Product Line
Units
Ref. No.
Toroidal Conductivity Trans­mitter for application in safe areas or hazardous locations DIV 2 (USA/Canada only)
APT2000TC-0-00
Toroidal Conductivity IS Transmitter for application in hazardous locations DIV 1 (USA/Canada) / Zone 1 (Europe)
APT2000TC-0-IS
Toroidal Conductivity Trans­mitter with HART
commu­nication for application in safe areas or hazardous locations DIV 2 (USA/Canada only)
APT2000TC-H-00
Toroidal Conductivity IS Transmitter with HART
com­munication, for application in hazardous locations DIV 1 (USA/Canada) / Zone 1 (Europe)
APT2000TC-H-IS
Mounting Accessories
Ref. No.
Pipe-mount kit 51205988-001 Panel-mount kit 51205990-001 Protective hood 51205989-001
Further Accessories
Ref. No.
HART
test socket, integrated in Pg cable gland (for APT2000TC-H-.. Transmitter only)
51205991-001
Page 31
Specifications
31
Specifications
APT2000TC-0(H)-00 Transmitter Cond input
Input for Series 5000 toroidal conduc­tivity sensor
Process vari­able/ranges**
00.00 to 99.99 mS/cm
000.0 to 999.9 mS/cm
0000 to 9999 mS/cm Concentration0.0 to 100.0 % by wt. Salinity
0.0 to 45.0 ‰ (0 to 35 °C)
Accuracy***
< 1 % of meas. value ± 0.02 mS/cm
Sensor monitoring
Sensocheck: monitoring of primary
and lines for short circuit and monitor-
ing of secondary for open circuit
(can be switched off)
Sensor stan­dardization
*
Entry of cell factor with display
of conductivity and temperature – Zero point adjustment – Temperature probe adjustment – Input of sensor factor
Permissible cell factor
0.100 to 19.999
Permissible sensor factor
1.00 to 99.99
Permissible offset
± 0.5 mS/cm
APT2000TC-0(H)-IS Transmitter Cond input
Input for Series 5000 toroidal conduc­tivity sensor
Process vari­able/ranges**
00.00 to 99.99 mS/cm
000.0 to 999.9 mS/cm
0000 to 9999 mS/cm Concentration0.0 to 100.0 % by wt. Salinity
0.0 to 45.0 ‰ (0 to 35 °C)
Accuracy***
< 1 % of meas. value ± 0.02 mS/cm
Sensor monitoring
Sensocheck: monitoring of primary
and lines for short circuit and monitor-
ing of secondary for open circuit
(can be switched off)
Sensor stan­dardization
*
Entry of cell factor with display
of conductivity and temperature – Zero point adjustment – Temperature probe adjustment – Input of sensor factor
Permissible cell factor
0.100 to 19.999
Permissible sensor factor
1.00 to 99.99
Permissible offset
± 0.5 mS/cm
Page 32
Specifications
32
APT2000TC-0(H)-00 Transmitter Temp input
Pt 100 / Pt 1000 / NTC 100 kW
Ranges
NTC –10.0 to +130.0 °C
+14 to +266 °F
Pt –20.0 to +150.0 °C
–4 to +302 °F Resolution 0.1 °C / 1 °F Accuracy < 0.5 K**** Temp com-
pensation
*
(Ref. temp 25 °C)
LIN 00.00 to 19.99 %/K NLF Natural waters to
EN 27888 (0 to 36 °C)
Concentra­tion deter­mination
-01- NaCl 0-26.3 % by wt (0 °C) ... 0-28.1 % by wt (100 °C)
-02- HCl 0-17 % by wt (-20 °C) ... 0-17 % by wt (50 °C)
-03- NaOH 0-12 % by wt (0 °C) ... 0-22 % by wt (100 °C)
-04- H
2SO4
0-25 % by wt (-17 °C) ... 0-35 % by wt (110 °C)
-05- HNO
3
0-28 % by wt (-20 °C) ... 0-28 % by wt (50 °C)
-06- H
2SO4
95-99 % by wt (-10 °C) ... 95-99 % by wt (110 °C)
Display LC display, alarm LED Loop current 4 to 20 mA, floating
22 mA for error message
*
supply voltage 14 to 40 V
Characteris­tic
*
Linear or logarithmic
Current error
< 0.3 % of current value ± 0.05 mA
APT2000TC-0(H)-IS Transmitter Temp input
Pt 100 / Pt 1000 / NTC 100 kW
Ranges
NTC –10.0 to +130.0 °C
+14 to +266 °F
Pt –20.0 to +150.0 °C
–4 to +302 °F Resolution 0.1 °C / 1 °F Accuracy < 0.5 K**** Temp com-
pensation
*
(Ref. temp 25 °C)
LIN 00.00 to 19.99 %/K NLF Natural waters to
EN 27888 (0 to 36 °C)
Concentra­tion deter­mination
-01- NaCl 0-26.3 % by wt (0 °C) ... 0-28.1 % by wt (100 °C)
-02- HCl 0-17 % by wt (-20 °C) ... 0-17 % by wt (50 °C)
-03- NaOH 0-12 % by wt (0 °C) ... 0-22 % by wt (100 °C)
-04- H
2SO4
0-25 % by wt (-17 °C) ... 0-35 % by wt (110 °C)
-05- HNO
3
0-28 % by wt (-20 °C) ... 0-28 % by wt (50 °C)
-06- H
2SO4
95-99 % by wt (-10 °C) ... 95-99 % by wt (110 °C)
Display LC display , alarm LED Loop current
EEx ib IIC
4 to 20 mA, floating 22 mA for error message
*
supply voltage 14 to 30 V, I
max
= 100 mA, P
max
= 0.8 W
Characteris­tic
*
Linear or logarithmic
Current error
< 0.3 % of current value ± 0.05 mA
Page 33
Specifications
33
APT2000TC-0(H)-00 Transmitter Start/End of
scale
*
As desired within ranges for mS, %, SAL
Min. span LIN 5 % of selected range
LOG 1 decade
Current source
3.8 mA to 22.00 mA
HART
com-
munication
(HART trans­mitter only)
Digital communication via FSK modu­lation of loop current, reading of de­vice identification, measured values, status and messages reading and writing of parameters
Explosion protection
USA/Canada: NI, Class I, Div 2, Group A – D, T4
Data retention
> 10 years (EEPROM)
RFI suppres­sion
To EN 50081-1 and EN 50081-2
Immunity to interference
To EN 50082-1 and EN 50082-2
Temperature Operating/ambient temp
–20 to +55 °C Transport and storage temp –20 to +70 °C
Enclosure Material: thermoplastic polyester, re-
inforced (polybutylene terephthalate) Protection: IP 65 Color: bluish gray RAL 7031
Cable glands 3 breakthroughs for Pg 13.5
2 breakthroughs for NPT 1/2 “ or Rigid Metallic Conduit
APT2000TC-0(H)-IS Transmitter Start/End of
scale
*
As desired within ranges for mS, %, SAL
Min. span LIN 5 % of selected range
LOG 1 decade
Current source
3.8 mA to 22.00 mA
HART
com-
munication
(HART trans­mitter only)
Digital communication via FSK modu­lation of loop current, reading of de­vice identification, measured values, status and messages reading and writing of parameters
Explosion protection
USA/Canada: IS, Class I, Div 1, Group A – D, T4 Europe: II 2G EEx ib [ia] IIC T6 CE 0032 TÜV 99 ATEX 1430
Data retention
> 10 years (EEPROM)
RFI suppres­sion
To EN 50081-1 and EN 50081-2
Immunity to interference
To EN 50082-1 and EN 50082-2
Temperature Operating/ambient temp
–20 to +55 °C Transport and storage temp –20 to +70 °C
Enclosure Material: thermoplastic polyester, re-
inforced (polybutylene terephthalate) Protection: IP 65 Color: bluish gray RAL 7031
Cable glands 3 breakthroughs for Pg 13.5
2 breakthroughs for NPT 1/2 “ or Rigid Metallic Conduit
Page 34
Specifications
34
APT2000TC-0(H)-00 Transmitter Dimensions
See Dimension drawings, Pg. 7 ff
Weight approx. 1 kg * user defined
** displayed with 3 1/2 digits *** ± 1 count **** with Pt 100 < 1K, with NTC > 100 °C < 1 K
APT2000TC-0(H)-IS Transmitter Dimensions
See Dimension drawings, Pg. 7 ff
Weight approx. 1 kg * user defined
** displayed with 3 1/2 digits *** ± 1 count **** with Pt 100 < 1K, with NTC > 100 °C < 1 K
Page 35
Certificate of Conformity
35
Type Examination Certificate
Page 36
Certificate of Conformity
36
Page 37
Certificate of Conformity
37
Page 38
Calibration Solutions
38
Calibration Solutions
Potassium Chloride Solutions Electrical Conductivity in mS/cm
Temperature [°C]
Concentration
0.01 mol/l 0.1 mol/l 1 mol/l
0
0.776 7.15 65.41 5 0.896 8.22 74.14 10 1.020 9.33 83.19 15 1.147 10.48 92.52 16 1.173 10.72 94.41 17 1.199 10.95 96.31 18 1.225 11.19 98.22 19 1.251 11.43 100.14 20 1.278 11.67 102.07 21 1.305 11.91 104.00 22 1.332 12.15 105.94 23 1.359 12.39 107.89 24 1.386 12.64 109.84 25 1.413 12.88 111.80 26 1.441 13.13 113.77 27 1.468 13.37 115.74 28 1.496 13.62 29 1.524 13.87 30 1.552 14.12 31 1.581 14.37 32 1.609 14.62 33 1.638 14.88 34 1.667 15.13 35 1.696 15.39 36
15.64
Data source: K. H. Hellwege (Editor), H. Landolt, R. Börnstein: Zahlen-
werte und Funktionen .... Volume 2, Part. Volume 6
Data source: * K. H. Hellwege (Editor), H. Landolt, R. Börnstein: Zahlen-
werte und Funktionen .... Volume 2, Part. Volume 6
** Test solutions calculated according to IEC 746-3
Sodium Chloride Solutions Electrical Conductivity in mS/cm
Temperature [°C]
Concentration saturated* 0.1 mol/l** 0.01 mol/l**
0
134.5 5.786 0.631 1 138.6 5.965 0.651 2 142.7 6.145 0.671 3 146.9 6.327 0.692 4 151.2 6.510 0.712 5 155.5 6.695 0.733 6 159.9 6.881 0.754 7 164.3 7.068 0.775 8 168.8 7.257 0.796 9 173.4 7.447 0.818 10 177.9 7.638 0.839 11 182.6 7.831 0.861 12 187.2 8.025 0.883 13 191.9 8.221 0.905 14 196.7 8.418 0.927 15 201.5 8.617 0.950 16 206.3 8.816 0.972 17 211.2 9.018 0.995 18 216.1 9.221 1.018 19 221.0 9.425 1.041 20 226.0 9.631 1.064 21 231.0 9.838 1.087 22 236.1 10.047 1.111 23 241.1 10.258 1.135 24 246.2 10.469 1.159 25 251.3 10.683 1.183 26 256.5 10.898 1.207 27 261.6 11.114 1.232 28 266.9 11.332 1.256 29 272.1 11.552 1.281 30 277.4 11.773 1.306 31 282.7 11.995 1.331 32 288.0 12.220 1.357 33 293.3 12.445 1.382 34 298.7 12.673 1.408 35 304.1 12.902 1.434 36
309.5 13.132 1.460
Page 39
Concentration Curves
39
Concentration Curves
Conductivity [mS/cm]
700
600
500
400
300
200
100
0
0 5 10 15 20 25 30
Concentration [% by wt]
Concentration calculation error [% by wt]
< 0.5 < 2
100 C
°
91 C°
82 C
°
73 C°
63 C°
53 C°
46 C°
34 C°
25 C°
15 C°
10 C°
0 C°
Fig. 12 Concentration curves NaCl (configuration: concentration -01-)
Page 40
Concentration Curves
40
Conductivity [mS/cm]
1200
1000
800
600
400
200
0
024681012
Concentration [% by wt.]
Concentration calculation error [% by wt]
< 0.2 < 1 < 2
14 16 18 20
–20 C°
–10 C°
50 C°
40 C°
30 C° 25 C°
20 C°
10 C°
0 C°
Fig. 13 Concentration curves HCl (configuration: concentration -02-)
Page 41
Concentration Curves
41
Conductivity [mS/cm]
1600
1400
1200
1000
800
600
400
200
0
0
5101520
25
Concentration [% by wt]
Concentration calculation error [% by wt]
< 0.2
100 C°
90 C°
80 C°
70 C°
60 C°
50 C°
40 C°
30 C°
20 C°
10 C°
0 C°
Fig. 14 Concentration curves NaOH (configuration: concentration -03-)
Page 42
Concentration Curves
42
Conductivity [mS/cm]
1600
1400
1200
1000
800
600
400
200
0
0
5101520
25 30
Concentration [% by wt]
< 0.5
100 C°
90 C°
80 C
°
70 C
°
60 C°
50 C°
40 C°
30 C°
20 C
°
–10 C°
10 C°
0 C°
35
1800
< 1
110 C°
Concentration calculation error [% by wt]
Fig. 15 Concentration curves H2SO4 (configuration: concentration -04-)
Page 43
Concentration Curves
43
Conductivity [mS/cm]
1200
1000
800
600
400
200
0
0 5 10 15 20 25
30
Concentration [% by wt.]
< 0.2 < 1 < 2
50 C°
40 C°
30 C°
–20 C
°
–10 C
°
25 C° 20 C
°
10 C°
0 C°
Concentration calculation error [% by wt]
Fig. 16 Concentration curves HNO3 (configuration: concentration -05-)
Page 44
Concentration Curves
44
Conductivity [mS/cm]
600
400
200
0
95.0 95.5 96.0 96.5
97.5
Concentration [% by wt]
Concentration calculation error [% by wt]
< 0.2
100 C°
90 C°
80 C°
70 C°
60 C
°
50 C° 40 C°
30 C° 20 C°
–10 C
°
10 C°
0 C
°
98.0
110 C°
300
500
98.5 99.0
100
97.0
Fig. 17 Concentration curves H2SO4 (range 95 to 99 % by wt), (configuration: concentration -06-)
Page 45
Index
45
Index
, 26
22 mA signal for alarm, 18, 26
configuring, 20
A
Alarm, 18
response time, 26 Alarm LED, 26 Alarm via loop current, 18, 26
configuring, 20 Assembly, 6
C
Cal info, 25, 28 Calibration, 21
input of cell factor, 21
sensor factor adjustment, 23
temp probe adjustment, 24
with calibration solution, 22
zero point, in air, 23 Calibration data, display , 28 Calibration solutions, 38 Cell factor, calibration, 22 Cleaning, 29 Concentration curves, 39 Conductivity measurement, 25 Configuration, 19
Connecting
handheld terminal, 9
lines, 9 Connecting cable, fixing, 9 Current characteristic, configur-
ing, 20 Current loop
Hold state, 20
Hold value, 20 Current source, 28
D
Diagnostics functions, 28 Dimension drawings, 7 Display, 16
E
Error info, 25, 26, 28 Error message, last, 25, 26, 28 Error message via loop current,
18, 26
configuring, 20 Error messages, 26–29 Explosion protection, APT2000,
33
G
GainCheck, 17, 28
H
HART communication, 18 Hold state, 25
I
Installation, 13 Instrument self–test
automatic, 17 manual, 17
K
Keypad, 16
L
Loop current, 18
display, instantaneous, 28 frozen, 25
M
Maintenance, 29 Measuring mode, 25 Messages, Sensoface, 26 Mode code, 4 Mounting diagram, 7
O
Outputs, 18
P
Packing list, 6
Page 46
Index
46
Pipe–mount kit, 8 Process variable, configuring, 19 Product line, 30 Protective hood, 8
S
automatic, 29 manual, 28
Sensocheck, on or off, 20
Sensoface, 17, 26
diagnostics, 26
messages, 26 Sensor factor , adjustment, 23 Sensor monitoring, Sensoface, 26 Sensors, monitoring, 17 Smiley, 26 Software version, display , 17, 28 Specifications, 31 Stripping lengths, 9
T
Terminals, pulling out, 9 Test mode, 28 Type Examination Certificate, 35
U
User Interface, 15
W
Wiring example, 14
Page 47
Page 48
Sensing and Control Honeywell Inc. 11 West Spring Street Freeport, IL 61032 USA
Honeywell S. A. Espace Industriel Nord Rue André Durouchez 80084 Amiens Cedex 2 France
Loading...