Honeywell AA919 Installation And Technical Manual

Sensing and Internet of Things
Installation and Technical Manual for the
Summing In-Line Amplifier,
Issue A
Order Code AA919
m WARNING
MISUSE OF DOCUMENTATION
Failure to comply with these instructions could result in death or serious injury.
m WARNING
PERSONAL INJURY
DO NOT USE these products as safety or emergency stop devices or in any other application where failure of the product could result in personal injury.
Failure to comply with these instructions could result in death or serious injury.
008-0749-00
m AVERTISSEMENT MAUVAIS USAGE DE LA DOCUMENTATION
Les informations complètes d’installation, d’utilisation et d’entretien sont fournies avec les instructions accompagnant chaque produit.
L’inobservation de ces instructions risque d’entraîner des blessures graves, voire mortelles.
m AVERTISSEMENT PERSONNELS BLESSURES CORPORELLES
NE PAS UTILISER ces produits en tant que dispositifs d’arrêt d’urgence ou de sécurité, ni dans aucune autre application où la défaillance du produit pourrait entraîner des personnels blessures corporelles.
L’inobservation de ces instructions risque d’entraîner des blessures graves, voire mortelles.
Sensing and Internet of Things
Installation and Technical Manual for the
ISSUE A
Summing In-Line Amplifier, Order Code AA919
TABLE OF CONTENTS
About this Document ......................................................... 1
1 | PRODUCT DESCRIPTION .......................................... 1
1.1 | General ................................................................................ 1
1.2 | Product Specifications ............................................... 1
2 | MODULE DETAILS .....................................................2
2.1 | External Connections - Sensor’s Side ................. 2
2.2 | External Connections - Rear Side ......................... 2
3 | OPERATING GUIDE ....................................................3
3.1 | Installation of Software and Drivers .................... 3
3.1.1 | System Requirements..................................... 3
3.1.2 | Loading Software and Device Drivers .... 3
3.2 | Making the Connections ............................................ 5
3.2.1 | Connect Amplifier to Power ......................... 5
3.2.2 | Connect the Sensors ....................................... 5
3.3 | Connect Amplifier to PC and Launch Desktop
Application ............................................................................ 5
3.3.1 | Connect Mini USB cable .............................. 5
3.4 | Output Scaling ............................................................... 6
3.5 | Save Configuration ...................................................... 7
3.6 | Load Configuration ..................................................... 7
3.7 | Tare and UnTare ............................................................. 8
3.8 | Shunt Calibration .......................................................... 9
3.9 | Disconnect Option ........................................................ 9
008-0749-00
4 | WARNING & ERROR MESSAGES ...........................10
4.1 | Output Scaling ............................................................. 10
4.2 | Taring with More Than 30 % Load .................... 10
5 | DIAGNOSTIC FEATURES .........................................11
Notices and Trademarks .................................................12
ii sensing.honeywell.com
Installation and Technical Manual for the
70,0
[0.2]
116,0
ISSUE A
Summing In-Line Amplifier, Order Code AA919

About this Document

This document provides details for using the User Interface Software that comes with the Summing In-Line Amplifier.
User Interface software can be used to:
Communicate with the Device
Configuring/Output Scaling of the Sensor’s
Preload Tare
Shunt Calibration

1 | PRODUCT DESCRIPTION

1.1 | General

Honeywell’s Summing In-Line Amplifiers are housed in a small plastic package, which is connected between the transducer and a readout instrument/UI running on a computer or laptop. The Summing In-Line Amplifier supplies a regulated bridge
MOUNTING DIMENSIONS mm [in]
008-0749-00
excitation voltage for the transducer and converts the millivolt signal from the transducer to a 0-10 Vdc amplified signal. The Summing In-Line Amplifier features include auto excitation voltages, programmable gain settings, and a wide adjustment range on the Span and offset value.
1.2 | Product Specifications
Power Requirements 15 Vdc to 24 Vdc
Output Voltage 0 Vdc to 10 Vdc
Accuracy ±0.25 % full scale span (FSS)
Operating Temperature Range -10 °C to 60 °C [14 °F to 140 °F]
[4.6]
2 X Ø 7,0
[0.3]
3,0
[0.1]
[2.8]
2 X Ø 4,0
[0.1]
THRU
SEE DETAIL A
26 GA TWISTED LEADS TEFLON INSULATED
0,91 m length – cable
[3 ft length – cable]
127,0
[5.0]
EARTH (SHIELD)
DETAIL A
VCC OUTPUT GND
GREEN
RED WHITE BLACK
6,0
21,0 [0.8]
Sensing and Internet of Things 1
Installation and Technical Manual for the
ISSUE A
Summing In-Line Amplifier, Order Code AA919

2 | MODULE DETAILS

2.1 | External Connections - Sensor’s Side

The front view below shows the connections for the 4 sensors to be connected to the Summing In-Line Amplifier. There are 4 mating connectors provided.
Figure 2.1a. External Connections

2.2 | External Connections - Rear Side

The rear view below shows the power input cable, as well as the USB mini connector for Programming interface to connect the Summing In-Line Amplifier to host PC.
008-0749-00
Figure 2.2. External Connections - Cable and Mini USB
Figure 2.1b. External Connector Pinout
Pin Number Connection
1 + Vexc 2 -Signal 3 + Signal 4 -Vexc/Gnd
pin 1 2 3 4
Note: Pin numbers are in the same order for all: Sensor-1, Sen­sor-2, Sensor-3, Sensor-4.:
Cable Wire Color Connection
Red (+) Supply Black Supply Return /
(-) Output White (+) Output Green Earth
The Summing In-Line Amplifier can be powered from a 15-24V DC power supply.
The power cable used for the Summing In-Line Amplifier is shielded 4 core cable. There are 4 conductors available
for the connections.
Note: Green Wire is connected to the Shield of the cable
Note Mini USB connector is used to configure the sensor’s
using the host PC/Laptop
2 sensing.honeywell.com
Installation and Technical Manual for the
ISSUE A
Summing In-Line Amplifier, Order Code AA919

3 | OPERATING GUIDE

3.1 | Installation of Software and Drivers

The Summing In-Line Amplifier requires installation of the Desktop Application and accompanying software drivers.
Please follow these instructions to obtain and load the software on a compatible PC or laptop.
Before installing, be sure to remove any earlier versions by using the Uninstall programs feature on your computer.

3.1.1 | System Requirements

Operating System – Windows 7 or Windows 10, 32 & 64 bits
The screen must support and be set for a resolution of 800 x 600 or higher

3.1.2 | Loading Software and Device Drivers

1. Download the required Summing In-Line Amplifier Soft­ware and Device Drivers from the following URL:
https://sensing.honeywell.com/test-measurement­products/summing-amplifiers/aa919
2. Unzip the downloaded file to a location on your computer or laptop.
008-0749-00
4. Select the directory for the installation and click the Next Button.
Figure 3.1.2b. Identify Installation Directory
5. Review the details and click Next.
Figure 3.1.2c. Continue Installation
3. Navigate to that location and run Setup.exe by right-click­ing the file and selecting Run as administrator.
Figure 3.1.2a. Start Setup.exe in Administrator Mode
Sensing and Internet of Things 3
Installation and Technical Manual for the
ISSUE A
Summing In-Line Amplifier, Order Code AA919
6. When you see Installation Complete, click the Next button.
Figure 3.1.2d Desktop Application Install Completes
008-0749-00
8. After extraction, click Next to continue with the Device Driver installation.
Figure 3.1.2f. Device Driver Welcome Screen
7. Install the Future Technology Devices International (FTDI) Combined Driver Model (CDM) drivers. Click on the Extract button.
Figure 3.1.2e. FTDI CDM Driver Installation
9. Accept the agreement by selecting the radio button, and click Next.
Figure 3.1.2g. Device Driver License Agreement
4 sensing.honeywell.com
Installation and Technical Manual for the
ISSUE A
Summing In-Line Amplifier, Order Code AA919
10. Allow the Device Driver installation to run. Drivers will dis­play as Ready to use when installation is complete, .
Figure 3.1.2h. Device Driver Installation Completion
008-0749-00
3.3 | Connect Amplifier to PC and Launch Desktop Application

3.3.1 | Connect Mini USB cable

1. Connect the smaller end of the provided Mini USB cable to the port on the rear of the Summing In-Line Amplifier as shown in 2.2. Then connect the larger end of the mini USB cable to a USB port on the computer..
2. Launch the Load Cell Summing In-Line Amplifier desktop application on the PC.
Figure 3.3.1a. Honeywell Splash Screen (momentarily shown)

3.2 | Making the Connections

3.2.1 | Connect Amplifier to Power
Refer to 1.3 and 2.2 to properly connect and apply power to the Summing In-Line Amplifier.
When you apply power to the Summing In-Line Amplifier, the POWER LED on the front of the device will glow green. The four sensor LEDs will blink red.

3.2.2 | Connect the Sensors

Refer to 2.1 to properly connect required sensors to amplifier ports. Up to four sensors can be connected.
As sensors are connected, accompanying port LEDs on the front of the amplifier will glow green.
3. Select the appropriate Communication Port from the drop down. You can verify port availability by opening Device Manager> Ports(COM & LPT), then select the correct port from the dropdown.
Figure 3.3.1b. Select Communication Port
4. Click Connect. When the connection is complete the simu­lated LED on the Connect button will glow green as shown in “Figure 3.3.1b. Select Communication Port”
Sensing and Internet of Things 5
Installation and Technical Manual for the
ISSUE A
Summing In-Line Amplifier, Order Code AA919

3.4 | Output Scaling

1. Click on the Output Scaling button on the application interface that should now be active.
Figure 3.4a. Output Scaling Option
008-0749-00
4. Click on the Output Scale button to apply the changes back to the amplifier. The application interface asks for verification first.
Figure 3.4c. Verify Output Scaling Changes
5. Click OK to continue.
Note:
The Sensivity should be within the Range of 1.5 mV/V-12 mV/V. You will see a warning if the value is out of range.
Zero Output should be within ± 10 % of FS. You will see a warning if the value is out of range.
2. The user interface reads the default configuration values from the Summing In-Line Amplifier and displays them.
3. Modify desired values based on calibration sheets provid­ed by the sensor manufacturers. Fields available include product serial number, sensitivity in mV/V, and zero output of the sensor from the calibration sheet. Zero output has to be entered as % of FS.
Figure 3.4b Output Scaling User Interface
6 sensing.honeywell.com
Installation and Technical Manual for the
ISSUE A
Summing In-Line Amplifier, Order Code AA919
3.5 | Save Configuration
1. To save the configuration to the PC/laptop, click on the Save button.
Figure 3.5a. Save Configuration Option
2. Select the desired directory path and provide a file name. Then click Apply, and the interface will validate the save.
Figure 3.5b. Saving - Provide Path and File Name
008-0749-00
3.6 | Load Configuration
1. To load a configuration previously saved to the PC/laptop into the user interface, then to the amplifier, click on the Load button.
Figure 3.6a. Load Configuration Option
2. Select the path and file, and click on the Load Settings button to load the settings into the user interface.
Figure 3.6b. Load Configuration Option
3. Click on the Output Scale button to apply the settings to the amplifier.
Figure 3.6c. Apply Settings to Device
4. Verify that the settings were loaded into the amplifier by clicking Read From Device.
Figure 3.6d. Read from Device Option
5. Exit Output Scaling by clicking on Exit.
Figure 3.6e. Exit Output Scaling Option
Sensing and Internet of Things 7
Installation and Technical Manual for the
ISSUE A
Summing In-Line Amplifier, Order Code AA919

3.7 | Tare and UnTare

1. The Tare function is useful in zeroing the preload on setup. Click the Tare button to load Zero Value to the amplifier.
Note: Tare limit is 30% of FS. When the preload of load on the amplifier is more than 30% of FS, Tare command will show an error.
Figure 3.7a. Tare Option
008-0749-00
Figure 3.7c. UnTare Option
1. After UnTare, the output will return to original values..
Figure 3.7d. UnTare Results
2. Click Yes to continue with the Tare function.
Figure 3.7b Tare Validation
3. The Output Voltage is now set close to zero and the values are saved to flash memory on the Summing In-Line Ampli­fier.
4. Tared output can be reversed by clicking on the UnTare button.
8 sensing.honeywell.com
Installation and Technical Manual for the
ISSUE A
Summing In-Line Amplifier, Order Code AA919

3.8 | Shunt Calibration

1. An electrical signal equivalent to that produced by a known load can be obtained by activating the Shunt Calibration function.
The Shunt Calibration function is achieved by connecting a high-precision resistor of known value, in parallel (shunt) with one arm of the strain gage Wheatstone bridge. The connection is made by a solid-state switch, which will be activated by the Shunt Cal ON button.
In the Summing In-Line Amplifier, a fixed resistor with re­sistance of 61900 Ohms is used as the shunt resistor. The output of the Summing In-Line Amplifier during Shunt Cal ON depends on the bridge resistance and sensitivity of the sensor connected to the amplifier.
When a Sensor with the 350 Ohm Bridge resistance and
2.0mV/V sensitivity is connected to the amplifier, the output of the amplifier will saturate to ~70% of full scale during the shunt calibration.
Table 3.8a. Shunt calibration Output vs Bridge resis­tance & Sensitivity
008-0749-00
Figure 3.8a. Shunt Cal ON Option
1. Observe the output.
Figure 3.8b. Shunt Cal ON Output
SL No Bridge resistance
1 350 2 70%
2 350 4 35%
3 350 10 14%
4 700 10 28.2%
Shunt Cal On Output = (0.000004286* Bridge Resistance) *100 (Sensitivity)
Example: Consider Sensor with Bridge resistance of 700 Ohm & Sensitivity 10mV/V
Shunt Cal On Output = 0.000004286 * 700*100 = 28.2% of Full Scale = 2.82V
0.010
of Sensor in Ohms
Output of the Summing In-Line Amplifier under shunt calibration for any other values of the bridge resistance and sensitivity can be calculated using this equation:
Sensivity In mV/V
% of Full Scale Output of
Summing In-Line Amplier
when Shunt Cal ON
2. Turn off shunt calibration by clicking the Shunt Cal OFF button:
Figure 3.8c Shunt Cal OFF Option3.9 | Other Options

3.9 | Disconnect Option

1. To disconnect the device from the PC/laptop, click on the Disconnect button.
Figure 3.9. Disconnect Option
Sensing and Internet of Things 9
Installation and Technical Manual for the
ISSUE A
Summing In-Line Amplifier, Order Code AA919

4 | WARNING & ERROR MESSAGES

4.1 | Output Scaling

The Sensitivity should be within the Range of 1.5mV/V -12mV/ V. The user interface will pop up a warning when it is out of range
Figure 4.1. Sensivity Out of Range
008-0749-00

4.2 | Taring with More Than 30 % Load

The amplifier will allow Taring off 30 % of FS preload of the system.
Taring off with more than 30 % delivers an error message.
Here is an example.
Figure 4.2a. Tare Weight Limit Crossed
If the entered value is more than 12, a default value of 12 mV/V will be used.
If the entered value is less than 1.5, a default value of 1.5 mV/V will be used.
Similarily, the Value of Zero output should be within the range of ± 10 % FSS.
Because the example shows 50 % of load applied, the device would pop up a warning message.
10 sensing.honeywell.com
Installation and Technical Manual for the
ISSUE A
Summing In-Line Amplifier, Order Code AA919
Here is an example showing a successful Tare using 25 % load on Sensor 1.
Figure 4.2b. Successful Tare
The Output also shows appropriate value in a successful Tare..
008-0749-00

5 | DIAGNOSTIC FEATURES

The amplifier’s sensor LED indicators glow red or green de­pending on whether a sensor is connected to the amplifier port. Under normal operating conditions with four sensors connect­ed to the amplifier, all four LEDs will glow green.
1. When sensors are not conencted to an amplifier port, the LED for that port will glow red instead of green. Likewise when the amplifier is connected to a PC/laptop with the desktop application running, the sensor displays will also glow green or red.
Figure 5. Sensor LED and Desktop Interface
In this example you see that sensors 3 and 4 are not connected and glowing red on the amplifier and the user interface.
2. A red indicator can also mean the sensor is faulty, a bridge is open, or shorted out.
3. A red indicator also indicates a sensor is loaded more than 110 % of FS.
Example: When the load on an individual sensor is more than 110 % of FS, the LED will blink red to indicate poten­tial overload of the sensor.
Sensing and Internet of Things 11
Installation and Technical Manual for the
ISSUE A
Summing In-Line Amplifier, Order Code AA919
WARRANTY/REMEDY
Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell’s standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowl­edgement or consult your local sales office for specific war­ranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items that Honeywell, in its sole discretion, finds defective. The foregoing is buyer’s sole
remedy and is in lieu of all other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.
While Honeywell may provide application assistance personally, through our literature and the Honeywell web site, it is cus­tomer’s sole responsibility to determine the suitability of the product in the application.
Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, Honeywell assumes no responsibility for its use.
008-0749-00

Notices and Trademarks

While this information is presented in good faith and believed to be accurate, Honeywell disclaims the implied warranties of merchantability and fitness for a particular purpose and makes no express warranties except as may be stated in its written agreement with and for its customers.
In no event is Honeywell liable to anyone for any indirect, spe­cial or consequential damages. The information and specifica­tions in this document are subject to change without notice.
Honeywell is a registered trademarks of Honeywell Internation­al Inc.
Other brand or product names are trademarks of their respec­tive owners.
For more information
Honeywell Sensing and Internet of
Things services its customers through a
worldwide network of sales offices and
distributors. For application assistance,
current specifications, pricing or the
nearest Authorized Distributor, visit
sensing.honeywell.com or call:
Asia Pacific +65 6355-2828
Europe +44 (0) 1698 481481
USA/Canada +1-800-537-6945
Honeywell Sensing and Internet of Things
9680 Old Bailes Road
Fort Mill, SC 29707
www. honeywell.com
COAX-SEAL® is the registered trademark of Universal Electronics. Scotch® is the registered trademark of 3M. EthernNet/IP® is the registered trademark of ODVA, Inc.
008-0749-00-A-EN | 1 | 06/19 © 2019 Honeywell International Inc. All rights reserved.
Loading...