Honeywell 96D User Manual 2

Page 1
User Manual
96d Controller Unit
Page 2
Page 3
Controller Unit
96d
User Manual
M-511401
8/10
Page 4
Page 5
Notices and Trademarks
Copyright by Honeywell International Inc.
August 2010
While this information is presented in good faith and believed to be accurate, Honeywell disclaims the implied warranties of merchantability for a particular purpose and makes no express warranties except as may be stated in its written agreement with and for its customers.
In no event is Honeywell liable to anyone for any indirect, special or consequential damages. The information and specifications in this document are subject to change without notice.
This manual covers software version 3.086.
Honeywell Analytics
405 Barclay Boulevard
Lincolnshire, Illinois 60069
M-511401 96d User Manual iii 8/10 Honeywell
Page 6
Page 7
Symbol Definitions
The following table lists the symbols used in this document to denote certain conditions:
Symbol Definition
ATTENTI ON: Identifies information that requires
special consideration
TIP: Identifies advice or hints for the user, often in terms of performing a task
REFERENCE _ INTERNAL: Identifies an additional source of information within the bookset.
Indicates a situation which, if not avoided, may
CAUTION
result in equipment or work (data) on the system being damaged or lost, or may result in the inability to properly operate the process.
CAUTION: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. CAUTION: Symbol on the equipment refers the user to the product manual for additional information. The symbol appears next to required information in the manual.
WAR NING: Indicates a potentially hazardous situation which, if not avoided, could result in serious injury or death. WAR NING symbol on the equipment refers the user to the product manual for additional information. The symbol appears next to required information in the manual.
M-511401 96d User Manual v 8/10 Honeywell
Page 8
Page 9
INTRODUCTION ..............................................................9
Intended Use ........................................................................................... 9
Receiving and Unpacking........................................................................ 9
Installation Instructions ...................................................................... 10
Basic Guidelines .................................................................................... 10
Surface Mount Installation ..................................................................... 11
Wiring Details .................................................................................... 12
Power Connections ........................................................................... 14
Communication Connections ............................................................ 14
Settings for Specific Transmitters ...................................................... 14
Relay Output ..................................................................................... 15
Jumper Use Instructions........................................................................ 16
GETTING STARTED ......................................................17
Initial Startup ......................................................................................... 17
Datalogger (SDcard) ............................................................................. 17
Programming Interface...................................................................... 18
Keypad Functions.................................................................................. 18
LED Definitions ...................................................................................... 19
System Operation .................................................................................. 19
SYSTEM PROGRAMMING ............................................20
Tx Info Menu ..................................................................................... 23
Ident Menu ............................................................................................ 24
Product and Sensor Types .................................................................... 26
COM Menu ............................................................................................ 27
Scale Menus (1 and 2) .......................................................................... 28
Detection Menu ..................................................................................... 29
Display Menu ......................................................................................... 30
Alarm A, B and C Menus ....................................................................... 31
Servicing and Operating Menus ............................................................ 32
Status Code........................................................................................... 33
Erase Current Tx ................................................................................... 33
Change Tx Address ............................................................................... 34
Groups Menu......................................................................................... 35
Creating Groups .................................................................................... 36
Deleting Groups .................................................................................... 37
M-511401 96d User Manual vii 8/10 Honeywell
Page 10
Event Menu .......................................................................................38
Action Menu ........................................................................................... 40
Delays Menu ..........................................................................................41
Latch Mode ............................................................................. 42
Conditions .............................................................................................. 43
Status..................................................................................................... 47
Database................................................................................................ 47
Acquisition Menu ...............................................................................48
Starting and Stopping Tx Logging ..........................................49
Starting and Stopping Event Logging ..................................... 50
Copy Menu ........................................................................................ 51
Configuration .....................................................................................51
Parameters ............................................................................................ 52
System Log Menu ..................................................................................53
Config Menu ...................................................................................... 54
Network Menu ...................................................................................59
Remote Calibration ................................................................................61
Tests Menu........................................................................................ 63
Test Sequence....................................................................................... 65
Normal Mode .........................................................................................67
Single Tx Mode ...................................................................................... 67
Debug Mode .......................................................................................... 67
Simulation Mode ....................................................................................68
BACnet Menu .................................................................................... 69
Wireless Network Menu .....................................................................70
Specifications ....................................................................................71
WARRANTY ...................................................................73
Warranty Statement ...............................................................................73
Warranty Conditions ..............................................................................73
Consumer Claims ..................................................................................74
viii 96d User Manual M-511401
Honeywell 8/10
Page 11
Page 12
Page 13
Introduction
Introduction
The 96D controllers act as nerve centers for gas detection networks, providing continuous monitoring for up to 96 connected units (plus 1 301ADI). Since the controllers are factory programmed to the owner’s specifications, installation is limited simply to mounting and connecting them. Once installed and connected, the controllers allow the user to monitor, adjust, or reconfigure an entire network of units.
Intended Use
The controller is intended to monitor an entire gas detection network around the clock. The unit offers logging capabilities, creating log files of all transmitter concentrations and alarms for analysis. The unit is also equipped with grouping or zoning capabilities that allow users to query and monitor specific groups of transmitters or specific transmitter zones.
Receiving and Unpacking
Upon receiving the controller unit:
Check that the package is undamaged
Carefully open the package.
Locate the packing slip or purchase order and verify that all items on the order are present and undamaged
Note: If the package or any of its contents are damaged, please refer to
the Warranty section at the back of the manual for instructions.
M-511401 96d User Manual 9 8/10 Honeywell
Page 14
Introduction
Installation Instructions
Installation Instructions
Basic Guidelines
Follow these instructions to the letter to ensure that the equipment will function properly. Failure to respect these guidelines will release Honeywell Analytics from any responsibility in the event of improper functioning:
Locate all units in areas easily accessible for service.
Avoid locations where instruments are subject to vibrations
Avoid locating units near sources of electromagnetic interference
Avoid locating units in areas subject to significant temperature swings
Verify local requirements and existing codes that may impact choice of location.
10 96d User Manual M-511401
Honeywell 8/10
Page 15
Introduction
6
5/16
13
1/4
16 cm
33.6 cm
1.52 m
5.0 ft
Installation Instructions
Surface Mount Installation
It is recommended that controllers be installed 5 feet (1.5 m) above the floor, at approximate eye level.
Mark the holes as shown:
Height markers 6 13/32” apart
Width markers 10 9/16” apart
Pre-drill 1/4” mounting holes as needed
Securely mount the 96D using the appropriate screws
Wiring for the unit must be passed through the knock-outs provided at the bottom of the unit.
M-511401 96d User Manual 11 8/10 Honeywell
Page 16
Introduction
COMMUNICATION
Communication Wire Gauge: 2-24 AWG (Belden 9841) Twisted and shielded cable 2000 feet (600 m) per channel T-tap: 65 feet (20 m) / T-tap 130 feet (40 m) total
Channel Specifications :
Channel 1-2: Modbus,Vulbus protocol Channel 3: Modbus protocol only Communicates only with Vulcain transmitters Channel 4: Modbus output Communicates only with VA301 BDCM
Channel 4 (not used)
Channel 3
A 4
B 4
A 3
B 3
Channel 2
Channel 1
B 2
A 2
A 1
B 1
V-
NEXT
V+
PREVIOUS
V-
V+

5(6(7
3&%&:5$
6+,(/'
6+,(/'
6+'1
9
%$%$
%$
5,*+7
/()7
5;
7;
$/$50B$
%
$
9
$/$50B%
$/$50B&
32:(5
)$8/7
'2:1
(17(5
(6&
6,/(1&(
1&
12
12 1&
83
0$'(,1&$1$'$
5&
5&
5&
5&
5
5
5
5
5(/$<
5(/$<
5(/$<
5(/$<
9LQ 9RXW
(7+(51(702'8/(
99
-
-
-
-
-
5
/('
/('
/('
/('
/('
&
6:
6:
6:
6:
6:
6:
/('
/('
/('
/('
/('
/('
/('
.
-
-
-
(2/
(2/
(2/
(2/
%=
-
-
-
-
-
-
-
.
.
/
5
-
6:
-
.
0DGHLQ&DQDGD
9XOFDLQ,QF

9$&:35%
5(6(7
-
-
'
'
-3
/
1&
12
12
Ethernet: 10/100-compa tible with 10Base-T interface, RJ-45
Visual Indicators: Green LED LINK Yellow LED ACT
BacNet/IP MODULE (-BIP option)
Always respect minimum voltage requirements at device
LCD screen
Wireless communication module
Wireless Communication : ISM worldwide Indoor range 30m
RELAY OUTPUTS
3+5 4+6
1+3 2+4
Normally open
Normally closed
5
31
642
RISK OF ELECTRIC SHOCK
RISK OF ELECTRIC SHOCK
End-of-line specification : The E.O.L. jumper for channels 1-2-3-4 must always be in E.O.L. position.
End-of-line jumper
position
5&
5
SD Card
Grounding screw
Wiring Details
Wiring Details
The diagram below provides the details required to connect the 96d to a gas detection network.
Details concerning power supply, cables, capacities, etc., are provided in the Specifications section at the back of this manual.
12 96d User Manual M-511401
Honeywell 8/10
Page 17
Introduction
Wiring Details
The connectors, or ports, on the PCB allow various wiring to be connected to the controller. The wiring includes power, communication, BACNet and relays, each with an assigned position (and number) on the board:
J22 Power Input: Connect the power supply to the controller
(see Wiring Details for cabling diagrams) J23, J24 Communication inputs: Connect communication cables to channels 1
through 3. Relay Outputs 1-4: Depending on the desired configuration,
connect the relay cables to either N.O. or N.C. SHDN jumper Place the jumper over the Shutdown header
pins to reset or restart the system. EOL Resistors 1-4: Place the jumper over the header pins to
create the connection to attenuate
communication echoes.
M-511401 96d User Manual 13 8/10 Honeywell
Page 18
Introduction
Power Connections
Power Connections
The 96D requires a power range of 17-27 Vac, 50/60 Hz (8.64 VA), 18­36 Vdc, 350 mA @24 Vdc (8.4 VA). Polarization is not important in either AC or DC mode. The system must be grounded on the transformer and a dedicated circuit breaker must be used.
Communication Connections
Communication cables must be grounded using the shield terminal, using twisted and shielded pair Belden 2-24 AWG #9841 cable (or equivalent).
The network cabling can extend up to a limit of 2000 feet (609 m) per channel.
The length of a T-tap can reach 65 feet (20 m), up to a maximum of 130 feet (40 m) for all T-taps.
All wireless devices associated to the controller will automatically be detected and communicate with the controller using wireless communication protocols.
Settings for Specific Transmitters
Honeywell Sensepoint XCD Transmitters must be configured for 9600 baud, no parity, and a unique address. Honeywell XNX Universal Transmitters must be configured for 9600 baud and a unique address. Information on configuring each transmitter is in the associated
technical manual.
14 96d User Manual M-511401
Honeywell 8/10
Page 19
Introduction
RELAY OUTPUTS
N.O.
N.C.
COMM
Normally closed
Normally open
Relay Output
Relay Output
The relay output can withstand up to 5A at 30Vdc or 250Vac (resistive load only). Relays can be used to activate horns and strobes. Although each relay is programmed with a default setting (below), they can be configured using the controller programming menu.
If relays are set to Normally closed, the relay is powered up with the controller and the device linked to the relay is functioning. The relay will shut down when the specified alarm condition is reached.
If the relay is set to normally open, the relay will remain off when the controller is powered up and the device connected to the relay will only be activated when the specified alarm condition is reached.
Note: These functions are reversed if the controller Failsafe mode has
been activated.
M-511401 96d User Manual 15 8/10 Honeywell
Page 20
Introduction
Relay Output
Jumper Use Instructions
The jumpers on the controller PCB allow a variety of operations to be performed manually:
EOL 1-4: Enables the user to add End-Of-Line jumpers that improve
communication signals. Put the jumper in R position (as shown on wiring diagram) to activate the End-of-Line termination. (R provides a resistance termination and RC provides resistance and condensator termination.)
SHDN: Enables the microcontroller to be reset or temporarily shut
down. This function is used mainly when system wiring
adjustments are needed (power off for safety). Relays These jumpers allow the relay to be tested by activating it J29-J32 without having any effect on Events.
16 96d User Manual M-511401
Honeywell 8/10
Page 21
Getting Started
Relay Output
Getting Started
The controller units are customized to the purchaser’s specifications but they can be further programmed using the following menu options.
Initial Startup
Make sure that all wiring has been completed according to specifications in the wiring details before powering up the unit. When all is secure, remove the SHDN jumper to power-up the unit. It should only take a few seconds until the unit is fully operative.
Datalogger (SDcard)
The DLC (Data Logger Card) option for the controller collects data and stores it on a digital Flash memory card (SDCard). In the event that the card memory becomes full:
Information logging is stopped
No SDcard flag is displayed on-screen
The SDcard LED blinks
See the Acquisition section for more details on starting and stopping the datalogging function.
CAUTION
M-511401 96d User Manual 17 8/10 Honeywell
Always deactivate datalogging function before removing the SDcard. Never remove the card when its LED is on.
Page 22
Getting Started
Programming Interface
Programming Interface
The front panel of the 96D provides a programming keypad (buttons) and LEDs.
96D front panel Keypad
Keypad Functions
Each unit has 7 keypad keys, or buttons:
Arrows: Used to move the cursor through the various programming
fields (Up, Down, Left and Right), or to adjust the display
contrast (press and hold the up or down arrow until desired
contrast is reached and release). ESC: Used to exit the programming menu or to cancel a change or
input. Enter: Used to access the programming menu and to modify
programming fields. Silence: Turns off the controller’s buzzer.
18 96d User Manual M-511401
Honeywell 8/10
Page 23
System Programming
Programming Interface
System Programming
The system’s Normal programming mode offers several menu options that are accessible from the main menu screen:
Tx Info: Allows transmitter parameters to be programmed Groups: Allows groups of transmitters to be set up Events: Allows events and event behavior to be programmed Acqui: Allows the datalogging feature to be activated or deactivated Copy: Allows data or parameters to be copied from the (controller)
configuration to parameters Config: Allows system parameters and password to be set Network: Allows actions on the network to be performed,
communication statistics to be consulted, and remote
calibrations to be performed Tests: Allows each device to be tested sequentially (inputs, outputs,
communications, events, etc.) and operation of various
parameters to be validated BACNet: Allows a device’s BACNet parameters to be set Wireless: Allows a network of wireless gas detection devices to be
configured, monitored, and modified.
Note: Access to the programming functions is password protected.
The default password is 2967.
The screen display shown below appears initially. This display can be configured to scroll among the information screens for each device connected to the controller.
If one or more of the connected devices is in an alarm mode, the controller will only scroll between the main information screen and the screens for device(s) in alarm mode. In this case, you must scroll manually to view screens for other devices.
20 96d User Manual M-511401
Honeywell 8/10
Page 24
System Programming
Programming Interface
The information screen also displays icons representing certain system functions. Here is a list of possible icons and their meaning:
Icon Description
BACnet is not offered with the model 96d controller unit.
BACnet is not offered with the model 96d controller unit.
Debug: Indicates whether the controller is in debug mode (Single TX, Debug or Simulation modes). When in simulation mode, SIM appears next to this icon.
Log: Indicates that either “Tx Logging” or “Event logging” is enabled.
Log error: Indicates that an error occured during TX or Event logging. All logging functions are stopped.
SDC: Indicates that an SDcard is present and functionning. The icon “fills” (from white to black) progressively as memory is used. A white icon indicates empty memory and black indicates full memory.
SDC error: If this symbol persists for more than 5 seconds, an SD card card is present but not functioning properly.
Wireless networking is not offered with the model 96d controller unit.
Wireless networking is not offered with the model 96d controller unit.
M-511401 96d User Manual 21 8/10 Honeywell
Page 25
System Programming
PASSWORD
_ _ _ _
MENU
1. Tx Info 3. Events
2. Groups 4. Aqui
MENU
5. Copy 7. Network
6. Config 8. Tests
MENU
9. BACNet
10. Wireless
Programming Interface
Since the controller’s programming functions are password protected, it is necessary to access the login screen:
Press Enter to access the programming options. The password
screen appears:
Use the keypad Up or Down arrows to increase or decrease the
value, one digit at a time, starting with the first digit
When all the digits of the password are correct, press Enter to
access the programming functions.
The first MENU options screen appears. Use the keypad arrows to navigate through multiple screens to the desired function and press Enter to access it.
22 96d User Manual M-511401
Honeywell 8/10
Page 26
System Programming
Tx Info Menu
Tx Info Menu
Tx Info is the menu option that is dedicated to transmitter information and contains several sub-menu options. The exact list of screens will vary depending on the transmitter type. A summary is presented below with details on the following pages
Ident: Allows the network component’s
identification information to be viewed.
COM: Allows the communication
protocol to be viewed or changed
Detection Allows the detection range and or Scale(1) the unit of measurement and parameters to be viewed and Scale(2): changed
Display: Allows the label (or name) of a
specific component to be changed
Alarms: Allow alarm thresholds to be
viewed and sometimes changed. There can be significant variations in this screen depending on transmitter type.
Status Transmitter and node status Displays: (in hexadecimal values)
Erase or Erases or changes the Current Tx: Tx address
M-511401 96d User Manual 23 8/10 Honeywell
Page 27
System Programming
1. Tx Info 001
-Ident­NotVul CO2 (IR)
Tx Info Menu
Ident Menu
The Ident, or identification menu allows a component’s network ID to be consulted:
The upper right corner of the screen shows the component’s address. If the address of the device whose information must be viewed is known:
Use the arrows to move the cursor arrow to the on-screen address
Press Enter (the value can be edited while the number is flashing)
Use the up or down arrows to increase or decrease the value
Press Enter again to validate the entry and display the information for the desired device.
The bottom left corner of the display shows the transmitter name (ex.: 301D2 - product name) and the sensor type (ex.: CH4 - methane sensor). These values can also be changed for Group or Vulbus product types. The procedure is identical for both fields:
Programming or changing a product or sensor type
Use the arrows to move the cursor to the product type field.
Press Enter to select the field (the value can be modified when flashing)
Use the arrows to scroll through the list of product types and press Enter when the desired product or sensor appears
24 96d User Manual M-511401
Honeywell 8/10
Page 28
System Programming
1. Tx Info 121
-Ident­301W CO2 (IR)
Tx Info Menu
Product and Sensor Types
This is a list of all the (preprogrammed) product types available from the Identification option in the Tx Info menu.
The order of the products below is not necessarily the order in which they will appear in the controller
Product Types
Addresses 1-96
VA301D2 VA301AP XNX 90DM3R VASQN8X XCD VA301IRF VA301EM VA201R IRT100 VA301EMRP VA201T EC-F9 GsPnt2 90DM2 VA201T2 S301RLC VA201D VA301T (Std.Device)* VA301D GsPnt VA901T NotVul RgRt3
* E3Point and 420MDBS are examples of standard devices.
Note: When Group is selected as a product type, the remaining Tx
INFO screens are not accessible (because each product in the group has already been individually programmed). Only the Ident and Erase current Tx screens will be available.
M-511401 96d User Manual 25 8/10 Honeywell
Page 29
System Programming
Tx Info Menu
The sensor type list applies to address ranges 1-96 and is not dependent on the type of product selected. Devices in the address range from 97-170 will display a BACNet object identifier, rather than a sensor type.
*An additional Product Type, simply called “Group”, represents a group created in the Groups Menu in the controller. When scrolling through the available product type list, this name will appear as many times as there are groups created in the controller (example: Group 1, Group 2, Group 3, etc.). If a group is selected as the product type, then the sensor type options are limited to MIN, MAX and MEAN.
26 96d User Manual M-511401
Honeywell 8/10
Page 30
System Programming
1. Tx Info 001
-COM- Mdbs RTU 8D 1S EP 9600 bauds
1. Tx Info 001
-COM­Mdbs RTU 8D 1S EP 9600 bauds
Tx Info Menu
COM Menu
This screen displays the selected communication protocol for device addresses from 1 to 96. Each transmitter’s protocol is defined by the controller (see Network Auto-configuration section)
.
If a transmitter is compatible with several different protocols, it can be modified using to one of the following options:
•Vulbus
Mdbs ASCII 7D 2S NP 9600 bauds
Mdbs RTU 8D 2S NP 9600 bauds
Mdbs RTU 8D 1S NP 9600 bauds
Mdbs RTU 8D 1S OP 9600 bauds
When a transmitter is configured with the Modbus communication protocol, the transmitters automatically sends the programmable parameters to the controller.
Note: Vulbus transmitter parameters must be programmed manually.
M-511401 96d User Manual 27 8/10 Honeywell
Page 31
System Programming
1. Tx Info 001
-Scale(1)-
Max 00250
Min 00000
1. Tx Info 001
-Scale(2)-
Factor 00010
Units %
Tx Info Menu
Scale Menus (1 and 2)
These menus appear only for certain devices. Scale(1) allows the detection range, minimum and maximum, to be defined for the selected device. Whatever value is specified is the value that will appear at the device display (if applicable). The Minimum value is generally left at 0. Parameters for the XNX and XCD gas detectors can be viewed here but can be changed only at the transmitter.
Scale(2) allows the factor by which to divide the scale (between 0 and
65535) and the unit of measurement for the selected scale to be determined.
The factor allows precise scale limits for detection to be set. By dividing the maximum scale value in the first Scale screen (250 in this example) by 10, a scale value of 25.0 can then be displayed.
The “Units” allow the unit of measurement for the device to be selected:
°F: Sets degree Farenheit as the unit of measurement °C Sets degree Centigrade as the unit of measurement %RH Sets Relative Humidity as the unit of measurement mV Sets millivolts as the unit of measurement V Sets volts as the unit of measurement mA Sets milliamps as the unit of measurement % Sets the percentage of gas as the point unit of measurement ppm Sets parts per million of gas as the point unit of
measurement
28 96d User Manual M-511401
Honeywell 8/10
Page 32
System Programming
1. Tx Info 001
-Detection­scale: 0-0.00 unit: %
Tx Info Menu
Detection Menu
The detection menu (available only for devices with addresses between 1 and 96) displays the detection range (scale: 0-100.0) and the unit of measurement (unit: %) for the selected component. If a transmitter uses the Modbus protocol, the detection parameters are automatically defined during network configuration and are not editable. Vulbus protocols must be manually defined by the programmer.
The detection scale is between 0 and the maximum value (0.00) and the unit of measurement is either ppm or percent (% for oxygen and % LEL for combustibles).
The detection menu is not available for the VA301R or VA301AP.
Programming or modifying the scale range or unit:
Use the arrows to move the cursor to the scale or unit option
Press Enter and use the arrow to increase or decrease the value
Press Enter when the desired value is obtained
M-511401 96d User Manual 29 8/10 Honeywell
Page 33
System Programming
1. Tx Info 001
-Display-
edit (20 char.max)
301D2 CH4 Ad001
Tx Info Menu
Display Menu
This option allows a specific label or name to be assigned to the selected component (transmitters, relay modules, annunciators). Up to 20 characters, including spaces, can be used in the label (example: BOILER ROOM). The default Modbus transmitter labels are composed of the component (or transmitter) name, sensor type and address.
Vulbus transmitter labels contain 20 blank characters (spaces).
30 96d User Manual M-511401
Honeywell 8/10
Page 34
System Programming
Tx Info Menu
Alarm A, B and C Menus
The screens for viewing alarm thresholds are combined in this manual. There will be either two or three levels, depending on transmitter type.
If present, separate “MIN” and “MAX”levels permit manual control of the hysteresis of each alarm. Normally, the “MAX” level is set greater than “MIN.” However, alarms can be made to trigger on falling concentration (as with oxygen) by setting the “MAX” threshold smaller than the “MIN” threshold.
With certain transmitters, only one threshold will be displayed.
Additionally, with certain transmitters, the alarm thresholds are read­only at the controller. These thresholds can be set only at the transmitter.
These are typical screens:
Typical screen for viewing alarms A or B on XCD and XNX transmitters
Typical screen for viewing or changing alarm A, B, or C thresholds on other transmitters.
M-511401 96d User Manual 31 8/10 Honeywell
Page 35
System Programming
Tx Info Menu
Servicing and Operating Menus
These functions vary depending on the transmitter type. These displays show the total time the device has been in service and the amount of time remaining until the next required calibration or replacement.
32 96d User Manual M-511401
Honeywell 8/10
Page 36
System Programming
1. Tx Info 001
erase current Tx Change Tx Address
Tx Info Menu
Status Code
These screens display transmission or node status and sensor status for the selected transmitter. This read-only information can assist service personnel in troubleshooting.
The XNX and XCD gas detectors will report the warning or fault number (iFaultWarnNumber) in hexidecimal on the third line. These transmitters will also report the monitoring state (iMonitoringState) in the fourth line. See the transmitter documentation for interpretation of fault numbers and monitoring states.
Typical display for XNX or XCD
Typical display for other transmitters
Erase Current Tx
This function allows the configuration to be erased or the Tx address for the displayed component to be changed.
Note: Selecting erase current Tx only erases the current device entry
Tx Info configuration. No other data is erased.
M-511401 96d User Manual 33 8/10 Honeywell
Page 37
System Programming
1. Tx Info 001
erase current Tx Change Tx Address
1. Tx Info 001
To address: 001
1. Tx Info 001
Processing...
Tx Info Menu
Change Tx Address
Selecting Change Tx Address allows users to move a device from one TX address to another:
Use the arrows to scroll to Change Tx Address and press Enter to select
In the next screen, scroll to the address number and press Enter to select
Use the up or down arrows to increase or decrease the address value and press Enter to validate the new address.
The Change Tx address option is only available (active) for device address 1 to 96 and if there is a Modbus device connected.
If the address is valid, the screen will display “Processing”.
If the address is invalid, the screen will display “Invalid Tx” and return to the Change Tx Address screen (the address for GasPoint devices cannot be changed)
A final screen will display either “Error” or “Success” (re-start procedure if Error is displayed).
Note: If a device address is changed to one already associated with
another device, the existing data will be overwritten. Customers should know their network’s address assignments and be careful when changing a Tx address. Delete the original Tx address to avoid duplicate entries.This feature is not supported with XNX and XCD transmitters.
34 96d User Manual M-511401
Honeywell 8/10
Page 38
System Programming
MENU
1. TxInfo 3. Events
2. Groups 4. Acqui
2. Groups 001 <end> [----] [----] [----] [----] [----] [----] [----]
empty all groups
Groups Menu
Groups Menu
Programming groups of transmitters allows several units to be combined which then enables actions (events) to be taken based on a series of units rather than each unit, individually.
A group is a stack containing the addresses from each of the transmitters included in the group.
Groups are displayed in a single line; if a group contains more than four components, the arrows must be used to scroll left and right of the display window to view all members of a group.
The cursor in the Group screen is represented by the blinking brackets (<end>). Any information between the brackets can be edited.
M-511401 96d User Manual 35 8/10 Honeywell
Page 39
System Programming
Groups Menu
Creating Groups
Use the arrows to move the cursor to a group line and press Enter
The field can be edited when the brackets stop blinking and the word “end” blinks
Use the up or down arrows to scroll through the list of all units connected to the 96D, until the desired address is displayed .
Press Enter again to validate the address.
The address is added to the group and the <end> bracket is shifted one position to the right.
The process can be repeated until all the desired transmitters in the group (up to 126) have been added. The address for each transmitter added in the Tx Info menu is available when creating groups.
Note: Groups created in the Groups menu will appear in the product
type list (Tx Info - Ident screen) as “Group xx” (the number assigned to the group when it was created).
36 96d User Manual M-511401
Honeywell 8/10
Page 40
System Programming
Groups Menu
Deleting Groups
Use the empty all groups command to delete all groups previously programmed in the controller.
Single groups can be deleted with a simple procedure:
Scroll to the first transmitter in the group list,
Select the transmitter (its address blinks) and scroll to <del> (<del> erases the entry and <end> marks the end of the stack)
Press enter and the group is emptied.
This procedure makes it possible to delete one, several or all entries previously included in a group.
Note: Up to 126 groups, with a maximum of 128 members each, can be
created.
M-511401 96d User Manual 37 8/10 Honeywell
Page 41
System Programming
Event Menu
Event Menu
Though the controller has a default setting for Event configuration (as shown below), the Event menu is programmable. Event programming lets specific actions to be defined:
Action: What will be done if programmed criteria are reached Delay: Defines the length of time to wait before taking an
action on an event and time to wait after an event has returned to normal before the action output is returned to normal state.
Conditions: AND, OR or none (---); equations that allow more
detailed control of an event Coverage Determines the period during which the event is period: applicable Status: Disables or enables a programmed Event disabled: Database: Erases the selected event or all events
38 96d User Manual M-511401
Honeywell 8/10
Page 42
System Programming
Event Menu
M-511401 96d User Manual 39 8/10 Honeywell
Page 43
System Programming
3. Events 001
-Action­Target: Ctrl Relay: #01
Event Menu
Action Menu
Actions are comprised of two parameters:
Target Indicates which component is responsible for the action to
be taken; Tx (transmitter) Re (Relay/Annunciator) Ctrl (Controller)
Relay Indicates which of three possible outputs will be activated
when the event is true; #XX (activates the component’s #xx relay), Buzzer (activates the component’s audible alarm) ALL (activates the relays and audible alarms)
Example: Tx 007 detects a concentration exceeding the set values. The target (controller) triggers relay 1 connected to that event (a fan perhaps).
40 96d User Manual M-511401
Honeywell 8/10
Page 44
System Programming
3. Events 001
-Delays-
Before: 99min After: 01min
Event Menu
Delays Menu
This option allows Before and After settings that will delay the activation or deactivation of an action to be programmed.
Before Delays the action for the specified length of time. If the
condition persists beyond this delay, the defined action is executed.
After The time to wait after an event has returned to normal before
returning action output to normal state. The after delay also offers a Latch option, described below.
Before and After delays can be configured at either 30 or 45 seconds or from 1 to 99 minutes, in one minute increments. Five dashes (-----) indicates that no delay has been programmed.
Use the keypad arrows to scroll to the desire option
Press Enter to select the option
Use the keypad arrows to scroll through the second or minute settings
Press Enter at the desired setting. The delay is set.
M-511401 96d User Manual 41 8/10 Honeywell
Page 45
System Programming
Event Menu
Latch Mode
The Latch function is executed on an Event state
It is possible to select the Latch mode by changing the after delay to “latch
The Event stays active until the Silence keypad button is pressed
The Silence keypad button has two functions: Silence the buzzer and unlatch the event.
When the Silence keypad button is pressed, events in Latch mode are unlatched and reevaluated. If the Event condition persists, the Event remains active and returns to Latch mode. If the condition does not persist, the event is deactivated.
Note: If the Event has a Before delay and the Silence button is pressed
while the Event conditions are still true, the buzzer will be silenced only for the length of the programmed delay.
42 96d User Manual M-511401
Honeywell 8/10
Page 46
System Programming
3. Events 001 1/3 Gr036 >2.01% AND
3. Events 001 all Gr003
AND > 2.99%
Event Menu
Conditions
Conditions are the parameters that define what makes an Event true. Each condition is defined by four elements and can be combined with other conditions to provide greater flexibility. A condition, as in the example provided below, defines:
IF at least 1/3 of group 36 detects concentrations greater than 2.01% of specified gas AND all of group 03 detects a concentration greater than
2.99% of gas, then the specified action (Actions were set at the first Event screen) for that Event will be triggered.
Since the display screens offer limited space, scroll left and right to view and edit further information.
Condition programming screens
The portion of the Events condition screen that is within the brackets is divided into four editable list fields:
M-511401 96d User Manual 43 8/10 Honeywell
Page 47
System Programming
Event Menu
The top left portion contains the statistic quantifier (available only for Groups) that take only the specified part of the group into the equation. Options available in this field are:
all: includes all transmitters in the group mean: includes the average concentration for the group’s transmitters max: includes the group’s maximum concentration min: includes the group’s minimum concentration 1/4: includes at least a quarter of the group’s transmitters that
meets set conditions
1/3: includes at least a third of the group’s transmitters that meets
set conditions
1/2: includes at least half of the group’s transmitters that meets set
conditions
2/3: includes at least two thirds of the group’s transmitters that
meets set conditions
3/4: includes at least three quarters of the group’s transmitters that
meets set conditions
1 or +: at least one or more than one of the group’s transmitters that
meets set conditions
44 96d User Manual M-511401
Honeywell 8/10
Page 48
System Programming
Event Menu
The bottom left portion contains the logic, or operator, quantifier that determines how conditions are calculated. Options available in this field are:
Operator
Symbol
--- No operator
= Equal to <= Equal to or smaller than < Smaller than >= Equal to or larger than > Larger than != Not equal to
max
min
When the maximum value is reached, an action is triggered. It will not be deactivated until levels fall below minimum value
When concentrations fall below minimum value, an action is triggered. It will not be deactivated until concentrations rise above set maximum value
Meaning
The top right portion contains the source, which defines what device or group of devices the Event will be based on. The list provides the following options:
GrAll: Includes all transmitters (see note) Gr_ _ _: Includes only the devices in the specified group (see note) Tx000: Includes only the specified transmitter (connected to the con-
troller)
Clock: Includes only information gathered between the specified
times. Selecting clock sets a condition that is applied only between the start and end time frame. It is possible to set one condition screen to specific parameters and the second to clock, which means that the specified condition will trigger an event only if it occurs during the set time period.
Note: Clicking on the magnifying glass to the right of a Group number
on the display opens a view of the Group for consultation or editing. Press Esc to close the group view and return to the Event condition screen.
M-511401 96d User Manual 45 8/10 Honeywell
Page 49
System Programming
3. Events 001
-Coverage Period­All day All week
Event Menu
The bottom right portion contains the operand, which defines what device or group of devices on which the Event will be based. The list provides the following options:
OFF Used for status on binary inputs (ex.: used with 301ADI) ON: Used for status on binary inputs (ex.: used with 301ADI) Fault: Bases trigger on maintenance alarm, communication failure or
device failure
Alrm A: If the chosen device or group has an Alarm A or Alarm 1, an
event will be triggered.
Alrm B: If the chosen device or group has an Alarm B or Alarm 2, an
event will be triggered.
Alrm C: If the chosen device or group has an Alarm C, an event will be
triggered.
The Coverage Period screen allows the period that will be covered by the Event to be defined. (The time frames for each of these periods can be defined in the controller Config menu.) This option provides two further selection fields:
Day definition field: allows All day, Daytime, or Nighttime to be selected
Week definition field: Weekend, Working Days, All week
1. Use the keypad up or down arrows to scroll to either All day or All
week
2. Press Enter to select. The value can now be changed
3. Use the keypad up ro down arrows to scroll through options (see
above)
4. Press Enter to select.
46 96d User Manual M-511401
Honeywell 8/10
Page 50
System Programming
3. EVENT 001
-Status : disabled­Enable event .
3. Events 001
-Database­Erase current event Erase all events
Event Menu
Status
This screen displays the current event status and allows it to be either enabled or disabled, depending on the current status.
Enable event:This is a “toggle” switch; press Enter to change the value
from Enable to Disable and vice versa
After going through all the steps and programming an event, this screen will display “Enable event”. Press Enter to activate all the parameters and enable the Event.
If an existing Event is being consulted, this screen would display “Disable event”. Press Enter to disable an Event (it will not be deleted but will not function). The programming of this Event is always present, which means that it easily can be reactivated by scrolling to this screen and pressing Enter.
Database
This screens displays the options linked to the database:
Erase current event:Lets user erase the current event Erase all events:Lets user erase all events
M-511401 96d User Manual 47 8/10 Honeywell
Page 51
System Programming
4. Acquisition 10% threshold mode Start Tx logging Start event logging
2005-04-27 11:05:20;1_CO2_ppm;574;-normal-: 2005-04-27 11:06:02;1_CO2_ppm;503;-normal-: 2005-04-27 11:06:15;1_CO2_ppm;562;-normal-: 2005-04-27 11:06:28;1_CO2_ppm;645;-normal-: 2005-04-27 11:06:39;1_CO2_ppm;557;-normal-: 2005-04-27 11:30:45;1_CO2_ppm;715;-normal-:
Acquisition Menu
Acquisition Menu
The Acquisition mode is only accessible when there is an SD card present (controllers with the Data Logging, or DLC function). It is used to enable or disable the logging of system Events or transmitter information. The information is logged (or recorded) on an SD card. Intervals or conditions must be defined before using this option.
The first line of the Acquisition screen offers either :
Delay mode: Allows for delay intervals of 10 to 59 seconds
or 1 to 60 minutes.
Threshold mode: Allows log values to be set according to set
variation thresholds (based on last reading) of 3% or more, 5% or more or 10% or more of last detected concentration.
If a 3% threshhold is selected, the system will not log a value at 3% but will log a value of 3.1%. Remember that the sampling rate (system refresh rate) may have an impact on logging.
Here is an example of threshold logging. The logs a semi-colon delineated text files.
48 96d User Manual M-511401
Honeywell 8/10
Page 52
System Programming
2004-01-23 17;54;25; 001_CO_ppm;0;-normal-;002_NO2_ppm;1.5;-normal-;003_CO_ppm;0;-normal­2004-01-23 17;55;25; 001_CO_ppm;0;-normal-;002_NO2_ppm;0.5;-normal-;003_CO_ppm;0;-normal­2004-01-23 17;56;25; 001_CO_ppm;0;-normal-;002_NO2_ppm;0.5;-normal-;003_CO_ppm;0;-normal­2004-01-23 17;57;25; 001_CO_ppm;0;-normal-;002_NO2_ppm;1.0;-normal-;003_CO_ppm;0;-normal­2004-01-23 17;58;25; 001_CO_ppm;0;-normal-;002_NO2_ppm;1.5;-normal-;003_CO_ppm;0;-normal-
Acquisition Menu
Starting and Stopping Tx Logging
In the previous step, “Acquisition”, the frequency at which Tx logs would be recorded can be configured. To start the logging function:
When “Start Tx logging” appears on the display, it indicates that the acquisition, or logging, mode is inactive. When “Stop Tx logging” appears, it indicates that Tx data is being logged. The log message is displayed on the screen according to the chosen mode and LED 1 will light up.
Press the Enter keypad button to stop or start Tx logging.
When Tx data is logged, the system creates files named tayymmdd.log, tbyymmdd.log and tcyymmdd.log, each representing one third of the network. The record includes the transmitter’s date, time and address, the sensor type, the concentration read, as well as the alarm status. Here is a sample of what a Tx log looks like:
The first column of the Tx log displays the date (yyyy-mm-dd) and the time (hh:mm:ss) of the log. In this example, the “Delay mode” was set to one minute intervals.
The third column of the Tx log displays the transmitter address and the fourth displays the gas type, gas concentration and unit of measurement.
The display then lists the next transmitter address with its gas type, concentration and unit of measurement, and so on until all the transmitters have been listed.
M-511401 96d User Manual 49 8/10 Honeywell
Page 53
System Programming
2004-01-23 17:54:25: Event logging enable 2004-01-23 17:55:25: Event logging enabled 2004-01-23 19:05:47; Simulation sequence activated 2004-01-23 19:05:48; Tx 6 communication no more in fault 2004-01-23 19:05:48; Tx 8 communication no more in fault
Acquisition Menu
Starting and Stopping Event Logging
The Acquisition menu offers an event logging option. Event Logging records controller transactions, events, Tx and alarm flags and relay status.
When “Start Event logging” appears on the display, it indicates that the acquisition, or logging, mode is inactive. When “Stop Event logging” appears, it indicates that Event data is being logged.
Press the Enter keypad button to stop or start Event logging.
When Event data is logged, the system creates a file named evyymmdd.log. The record includes the date, time and the event. Here is a sample of what an Event log looks like:
The first column of the Event log displays the date (yyyy-mm-dd) and time (hh:mm:ss) of the log. Column A displays the date and time of the log. In this example, the event’s “Delay mode” was set to one minute intervals.
The system logs the following types of events:
Event Log
Event status changed
Alarm A, B, C, Fault, and X status changed
Note: New log files are created when the existing files reach 32 000
lines or at the start of a new week (0h00 Sunday)
50 96d User Manual M-511401
Honeywell 8/10
Page 54
System Programming
5. Copy
-Configuration­Config to SDcard SDcard to Config
Copy Menu
Copy Menu
The Copy menu allows programmed parameters to be copied and transferred. Data from the SD card can be transferred to a controller or from a controller to the SD card or copy parameters from one device to the next. The Copy option offers three screens: Configuration, Parameters and System Log.
Configuration
If the controller is equipped with an SD card, the configuration function allows data to be transferred either from the 96D to the SD card or the reverse. This makes it possible to transfer the controller’s programming to a computer or from a computer to the controller.
The first option in the configuration screen is 96D to SDcard. Selecting this option copies the controller’s configuration and parameters into a
“config.ini” file
1
.
The second option is SDcard to 96D allows the configuration and parameters of the “config.ini” file on an SDcard to be copied to the controller.
Note: The “config.ini” file contents can be modified at any time and from
any computer.
1. When transferring data, the system will automatically search for an existing “config.ini” file before proceeding. If one exists, the system searches for a “config.bak” file. If found, the file is deleted. Then, the pre-existing “config.ini” file is renamed “config.bak”, making it possible to save the new “config.ini” file and keep a backup copy of the previ­ous one.
After inserting an SD card into the controller, the controller’s system looks for an exist-
ing “config.ini” file that contains an “autoload” tag equal to 1 (yes). If the tag is found, the system loads the contents of the file and resets “autoload” to 0 (no). This is a useful feature for editing the file on a computer without having any impact on the controller (such as recorded Events).
M-511401 96d User Manual 51 8/10 Honeywell
Page 55
System Programming
5. Copy
-parameters­TxInfo to TxInfo Event to Event
TxInfo to TxInfo source tx001 target tx113 COPY
Event to Event source #001 target #013 COPY
Configuration
Parameters
The “parameters” function allows one transmitter’s configuration to be
copied to another or one event’s parameters to be copied to another
event. This allows several devices that share identical or similar
parameters to be quickly configured.
The options within this screen are:
Tx Info to Tx Info copies transmitter parameters from one device to
another.
Event to Event copies parameters from Event to Event.
The process is identical for both options:
Select source, (the data to be copied) using the up/down keypad arrows and press Enter.
When the transmitter address is flashing, use the up/down keypad buttons to search for the desired device address.
Press Enter to select the new address
Select the target address (where the data is to be copied to) in exactly the same way as source
Select COPY and press Enter. The parameters have been copied.
52 96d User Manual M-511401
Honeywell 8/10
Page 56
System Programming
5. Copy
-System Log-
SysLog to SDcard
--- START of system log dump : 2007-04-18 13:19:05 --­2007-04-04 18:42:06;Accessing menu; 2007-04-04 18:43:47;Event 1 definition modified; 2007-04-04 18:48:12;Exiting menu; 2007-04-04 18:54:49;System power-down; 2007-04-04 18:56:40;System power-up; 2007-04-04 19:02:44;Accessing menu; 2007-04-04 19:03:07;Event 6 definition modified; 2007-04-04 19:03:21;Exiting menu; 2007-04-05 10:51:28;Accessing menu; 2007-04-05 10:54:59;Database reset; 2007-04-05 10:55:18;Tx 25 parameters modified; 2007-04-05 10:55:29;Group 0 definition modified; 2007-04-05 10:55:36;Group 0 definition modified; 2007-04-05 10:55:46;Group 0 definition modified; 2007-04-05 10:55:55;Group 4 definition modified; 2007-04-05 10:55:57;Exiting menu; 2007-04-05 10:56:02;Accessing menu; 2007-04-05 10:56:19;Tx 24 parameters modified;
Configuration
System Log Menu
The controller will record log information to its internal memory. If the controller is equipped with an SD card, the system log function allows users to save system log information to the memory card in text format.
When this option is selected, a log of all the last actions performed on the controller is copied to the SDcard, with the filename slyymmdd.log. This file can contain up to a maximum of 64Kb of information in text format. Once the memory card is full, the oldest log entries are erased and replaced by new entries.
Here is an example of a system log:
M-511401 96d User Manual 53 8/10 Honeywell
Page 57
System Programming
MENU
5. Copy 7. Network
6. Config 8. Tests
6. Config
Set User Password
- 2967 -
Slave Port Cnfg
6. Config Relay Configuration AP Brdcst: auto Honeywell
6. Config Day definition Week definition
6. Config Manual scroll mode Set Date and Time Menu français
Slave Port Cnfg Address: 001 38400 bauds Mdbs RTU 8D 2S NP
Relay Configuration
-Failsafe­All relays Inactive
Relay Configuration
-Silence­1234 dis ena dis dis
Day definition
-Daytime­sta: 08:00 end: 17:00
Week definition
-Working Days­sta: Monday end: Friday
Date and Time
2005-11-21 11:47:36 (ISO 8601 standard)
Config Menu
Config Menu
The Config menu contains several main configuration screens and is used to program the controller display mode, adjust the date and time, select the display language, change the controller access password, modify the Slave Port configuration, set the Relay Configuration, and select the AP Broadcast mode.
Each main screen offers further programming options, as shown.
54 96d User Manual M-511401
Honeywell 8/10
Page 58
System Programming
6. Config Manual scroll mode Set Date and Time Menu français
Date and Time
2005-11-21 11:47:36 (ISO 8601 standard)
Config Menu
Selecting the first line of the first screen allows selection from three display modes: Manual scroll, 3-second scroll, 5-second scroll. If Manual scroll mode is chosen, the screen will only advance if you press on the arrow keypad buttons. If either 3 or 5 second scroll mode is chosen, the screens will automatically scroll display readings for all devices connected to the controller after 3 or 5 seconds.
3 or 5 second scroll modes do not prevent the keypad arrows to be used to return to a previous screen or move ahead through the screens manually.
Selecting the second line allows the date and time in a new screen to be adjusted; Date and Time. When a number is flashing, the value can be changed using the up/down keypad arrows. The year, month, day and the hour, minute and second values can be changed.
The controller does not manage Daylight Savings Time, therefore, users must manually adjust any time changes.
Selecting the third line allows the display language to be changed. If the display is already in English, it will then display the Menu français option (and vice-versa). Simply scroll to the line and press Enter to change the language.
M-511401 96d User Manual 55 8/10 Honeywell
Page 59
System Programming
6. Config
Set User Password
- 2967 -
Slave Port Cnfg
Config Menu
The second main screen in the Config menu allows a new user password to be set and the slave port configuration to be modified.
The default password is 2967. Select Set User Password to change the password:
When the first digit blinks, change the value by using the up/down keypad arrows to increase or decrease the number
Use the left/right keypad arrows to move from one digit to the next.
When the desired password has been set, press Enter to validate it and exit the editing mode.
Note: Contact Honeywell technical support for help with lost passwords
at 1-800-563-2967.
56 96d User Manual M-511401
Honeywell 8/10
Page 60
System Programming
6. Config Relay Configuration AP Brdcst: auto Honeywell
Relay Configuration
-Failsafe­All relays Inactive
Relay Configuration
-Silence-
1234 dis ena dis dis
Config Menu
Scroll through the main Config menu screens using the left (previous) or right (next) keypad arrows.
The third main screen in the Config menu allows the relay configuration to be set, the AP broadcast mode and to select from four separate manufacturers for the given controller.
When Relay Configuration is selected, two further options to configure the relays are available: The first screen, Failsafe, appears allowing the failsafe to be activated for all relays using the Enter keypad button. This function is used in case of a power failure: if power is cut, the relay will activate the connected device (ex. a light) to signal the problem.
Scrolling to the right displays the “Silence” screen that enables or disables the silence option for each relay, using the Enter keypad button.
M-511401 96d User Manual 57 8/10 Honeywell
Page 61
System Programming
6. Config Day definition Week definition
Day definition
-Daytime­sta: 08:00 end: 17:00
Week definition
-Working Days­sta: Monday end: Friday
Config Menu
The fourth screen in the Config menu allows a definition of a day and a week to be programmed.
Day and Week definition allows hours (time frames) to be defined for either Daytime and Working Days respectively.
Note: Remember, the controller uses a 24 hour clock (0:00 to 23:59).
Any time changes (DST) must be made manually or through BACNet time synchronisation (BACNet module required).
58 96d User Manual M-511401
Honeywell 8/10
Page 62
System Programming
7. Network
Reset Database Scan Network Reset and Scan
Network Menu
Network Menu
The Network menu allows network device information to be either scanned or reset.
This menu offers four options, divided into two screens; the first screen contains three options:
Reset Database: Resets all network device Tx information in the
database. This only resets the Tx infomation for the network device. It does not affect programmed Groups or Events.
Network Scan: Begins an auto-detect of all network devices
that allows the system to configure the Tx database for network devices (i.e. it will scan and add new devices but will not overwrite or erase the old database)
Reset and Scan: Performs both previous functions
simultaneously.
Note: Once one of these options has been set, wait until the controller
completes the process. Do not interrupt or stop the process once it has begun.
M-511401 96d User Manual 59 8/10 Honeywell
Page 63
System Programming
7. Network
Statistics
Calibration
Statistics 001
Valid 16 100% Errors 0 0% Timeouts 0 0%
Network Menu
The second Network screen offers the Statistics and Calibration options.
Selecting Statistics from the Network menu displays a screen containing the statistics for the selected device address.
Valid: Indicates the number of valid responses for the last 16
requests
Errors: Indicates the number of errors in the response for the last 16
requests
Timeouts: Indicates the number of timeouts (no response) for the last
16 requests
60 96d User Manual M-511401
Honeywell 8/10
Page 64
System Programming
7. Network
Statistics Calibration
Calib 007 S301M Status: Normal Set Zero Set Span 246 PPM
Calib 007 S301M Status: Normal Set Zero Set Span 246 PPM
Network Menu
Remote Calibration
The network menu also offers a Calibration option for use with devices that support network calibration.
The Calibration screen contains four (4) lines of information:
Line 1: Indicates the mode (Calib, meaning calibration), the
(Modbus) address of the device to calibrate (001) and the type of device to calibrate (301D2)
Line 2: Indicates the status (Normal or In calib…) of the specified
device Line 3: Displays the function to perform (Set Zero) Line 4: Displays the function to perform (Set Span) and the span
gas concentration value (246 ppm)
5. On the first line, scroll to the device address and press Enter
6. Scroll through the devices to display the desired device* and press
Enter to select.
7. The second line displays the device’s status
8. Scroll to select the desired function, Set Zero to set the device's
zero, and press Enter to select.
9. Upon pressing Set Zero, the controller requests confirmation.
*The device must be configured in the 96D’s database in order to be included in the device addresses displayed on screen.
M-511401 96d User Manual 61 8/10 Honeywell
Page 65
System Programming
Calib 001 : 301D2 Status : Normal Set Zero Set Span 2.60 ppm
Calib 001 : 301D2 Status : In calib... Set Zero Set Span 2.60 ppm
Network Menu
10. Press Enter to confirm or Esc to cancel. If confirmed, the controller calibrates the sensor’s Zero. This takes only a few moments and the display returns to the default calibration screen.
Note: Never calibrate any unit’s Zero with ambient air. Always use
Nitrogen (N
) at the calibration port to calibrate the Zero.
2
11. To calibrate the device, scroll to Set Span** and change the span gas calibration value using this procedure;
a. Using the right arrow, move the cursor to xxx PPM (span
value field). Press Enter to select the field (it is editable when flashing).
b. Use the up or down arrows to increase or decrease the
value, press Enter to validate the new value.
c. Move the cursor back to Set Span and press Enter to start
the calibration.
The device Span is being calibrated. The screen will display the device’s status as “In calib...” until the calibration is complete.
**When selecting Set Span, make sure that the device has been supplied with the appropriate calibration gas before and during the calibration process.
62 96d User Manual M-511401
Honeywell 8/10
Page 66
System Programming
MENU
5. Copy 7. Network
6. Config 8. Tests
8. Tests
Oprt Mode: normal
8. Tests Test sequence Start Sim Sequence Maximum load
Simulation Sequence Start with Tx001 Confirm
Test sequence Relays activated Any key to continue Esc to quit
Test sequence BUZZER activated Any key to continue Esc to quit
2
Test sequence LEDs activated Any key to continue Esc to quit
Test sequence Press ‘left’
Esc to quit
Test sequence Display test Any key to continue Esc to quit
Test sequence Test Network comm. 9600 bauds Channel 1 PASSED
Test sequence Test done Any key to continue
7
6
5
4
3
1
Tests Menu
The Tests menu allows a variety of tests to be performed on components and on the network communications. It also allows the system to be operated in four different modes which, in turn, provide different functionalities.
Tests Menu
M-511401 96d User Manual 63 8/10 Honeywell
Page 67
System Programming
8. Tests Test sequence Start Sim Sequence Maximum load
8. Tests Oprt Mode: normal
Tests Menu
The Tests menu provides four main options, divided between two screens. Each of these options offers different capabilities.
The first screen presents three options:
Test sequence: Enables each output to be activated and validates
operation of each controller keypad buttons, display
pixels, and various communication protocols. Start Sim Sequence: This options starts or stops the Simulation mode,
which allows a simulation of a gas concentration
over an associated scale range on all transmitters.
The simulated gas concentration values are local (on
the controller) and do not affect logging functions.
(Events will be activated for the simulation but
detection devices are not affected.) Maximum load: Activates all controller components
The second screen option is “Oprt Mode”, which offers three separate operation settings: Normal, Single Tx or Debug.
Normal Normal controller operation mode Single Tx: Activates the polling mode on a single transmitter. Debug: Activates the service mode to perform a calibration
and to test Events without triggering actions.
64 96d User Manual M-511401
Honeywell 8/10
Page 68
System Programming
Test sequence
Relays activated Any key to continue Esc to quit
Test sequence
BUZZER activated Any key to continue Esc to quit
Test sequence
Press ‘left’
Esc to quit
Tests Menu
Test Sequence
When test sequence is selected from the main Tests menu, the controller will display the Test sequence screen.
If Esc is pressed on the keypad, the main Tests menu screen will be displayed. However, to perform system tests, press any key to proceed to the first test screen.
This screen tests each component individually and will advance only to the next component when a key is pressed. This option will display 13 screens. Screens 1, 2 and 3 test Relays, BUZZER and LEDs.
The following six screens prompt the user to press the keypad buttons, in turn: left, right, up, down, Silence, Enter and Esc. The system will not advance until a key is pressed.
M-511401 96d User Manual 65 8/10 Honeywell
Page 69
System Programming
Test sequence
Display test Any key to continue Esc to quit
Test sequence
Test Network comm. 9600 bauds Channel 1 PASSED
Test sequence
Test done Any key to continue
Tests Menu
The system then moves to the Display test. When the blank screen is displayed, it is testing for display pixels. Press any key to proceed to the next step.
The final test that the system performs is a network communication test:
Once these tests have begun, do not interrupt or stop them. The system tests all five bauds (communication speed: 9600, 19200, 38400, 57600, 115200) and each of the 4 channels per baud.
When the system has completed the test, it displays the final Tests screen. Press any key to return to the main Tests menu.
66 96d User Manual M-511401
Honeywell 8/10
Page 70
System Programming
Tests Menu
Normal Mode
This is the system’s normal (default) operation mode. When the system is in normal mode, some values can be changed without interrupting services. When a value has been changed in any of the menu fields, the change will take effect upon returning to the main menu screen.
Single Tx Mode
This mode allows transmitters to be analyzed one at a time. The controller polls only the selected device, which subsequently has its information updated. This mode does not interfere with Event Evaluation functions.
Debug Mode
This mode allows complete system operation to be evaluated and tested without affecting operations (outside of debug mode). Events
are evaluated and displayed as necessary but no action is triggered.
M-511401 96d User Manual 67 8/10 Honeywell
Page 71
System Programming
Tests Menu
Simulation Mode
This mode deactivates network communication Information Updates. It can be combined with any of the three previous modes (example: using the Simulation mode when in Debug mode allows the user to test the entire system [groups, events, etc] without triggering any actions or using any additional material such as gases). It allows gas concentrations to be simulated over an associated scale for each transmitter, sequentially:
Alarm levels A, B and C are evaluated according to the simulated gas concentration and events are evaluated and actions are taken.
This type of alarm simulation at the controller does not work with certain transmitters with falling alarms. In these cases, an alarm can be simulated at the transmitter.
While in simulation mode, the controller is unaware of the device’s actual network status. This mode can be stopped at any time in the Test menu (see Normal System Operation).
If one of these modes has been activated, the system will automatically return to Normal Mode after 12 hours of inactivity. (No changes will be lost.)
68 96d User Manual M-511401
Honeywell 8/10
Page 72
System Programming
BACnet Menu
BACnet is not offered with the model 96d controller unit.
BACnet Menu
M-511401 96d User Manual 69 8/10 Honeywell
Page 73
System Programming
Wireless Network Menu
Wireless Network Menu
Wireless networking is not offered with the model 96d controller unit.
70 96d User Manual M-511401
Honeywell 8/10
Page 74
Specifications
Specifications
Power requirements 96D 17-27 Vac, 50/60 Hz, 8.64 VA
18-36 Vdc, 350mA @ 24 Vdc (8.4 VA) Operating temperature range -20°C to 50°C (-4°F to 122°F) Operating humidity range 0 to 95% RH (non-condensing) Operating altitude Up to 3000 m (9843 ft) Network capacity Up to 96 transmitters, 32 per channel
Channels 1, 2 = Modbus
Channel 3 = Modbus only Communication Up to 609m (2,000 ft) per channel (length of lines) T-tap maximum = 20m (65 ft) per t-tap,
overall total of 40m (130 feet) User interface Graphic 122 x 32 dot matrix backlit display
User friendly keypad Visual indicators Power Green LED
Warn, Alarm, High Alarm Red LED
Fault Amber LED
Tx Amber LED
Rx Green LED Outputs 4 DPDT relays Output rating 5A, 30Vdc or 250 Vac (resistive load) Audible alarm 65dBA at 1 m (3 ft) Time delays Programmable Before and After delays Battery 3 volt lithium battery Enclosure Painted aluminum Dimensions (HxWxD) 9.5” x 14” x 3.5” (24.1 x 35.6 x 8.9 cm) Certifications ANSI/UL 61010-1
CAN/CSA C22.2 No. 61010-1
ETL 116662
M-511401 96d User Manual 71 8/10 Honeywell
Page 75
Page 76
Warranty
Warranty Statement
All products are designed and manufactured to the latest internationally recognized standards by Honeywell Analytics under a Quality Management System that is certified to ISO
9001.
Service in the field or at the customer’s premises is not covered under these warranty terms. Time and travel expenses for on-site warranty services will be charged at Honeywell Analytics’ normal billing rates. Contact your Honeywell Analytics Service Representative for information on Service Contracts.
Warranty Conditions
The Honeywell Analytics (HA) Limited Product Warranty only extends to the sale of new and unused products to the original buyer where purchased from HA or from a HA authorized distributor, dealer or representative. Not covered are: consumable items such as dry-cell batteries, filters and fuses or routine replacement parts due to the normal wear and tear of the product; any product which in HA’s opinion has been altered, neglected, misused or damaged by accident or abnormal conditions of operation, handling, use or severe sensor poisoning; defects attributable to improper installation, repair by an unauthorized person or the use of unauthorized accessories/parts on the product
Any claim under the HA Product Warranty must be made within the warranty period and as soon as reasonably possible after a defect is discovered. If a Warranty claim is being sought it is the responsibility of the buyer to obtain a Service Event number (SE#) from HA and if practical return the product clearly marked with the SE# and a full description of the fault.
HA, at its sole discretion, may elect to send replacement goods to buyer prior to receipt of the defective goods. Buyer agrees to return defective goods with in 30 days or to pay for the replacement goods.
M-511401 96d User Manual 73 8/10 Honeywell
Page 77
Warranty
Buyer is responsible for transportation costs from the buyer’s location to HA. HA is responsible for transportation costs from HA’s location to the buyer.
If in the case of a fixed installation or when it is not practical to return the product, the buyer should submit a claim to HA Service Department. A service engineer will attend on site on a day rate basis. Where a valid warranty claim is identified, the faulty product will be repaired or replaced free of charge. A warranty claim will be accepted if all conditions contained within this Warranty are met.
When, in the opinion of HA, a warranty claim is valid, HA will repair or replace the defective product free of charge and send it or any replacement back to the buyer. If, in the opinion of HA the warranty claim is not valid, HA will, at the option of the buyer, return the unit unaltered at the buyer’s expense, repair the unit at the then prevailing rates, replace the unit with an appropriate replacement item at the then prevailing price, or discard the unit. HA reserves the right to charge for any attendance by its service engineer at the usual rates in force at the time the claim was received.
In no event shall HA’s liability exceed the original purchase price paid by the buyer for the product.
Consumer Claims
If you purchased your HA product as a consumer, the above warranty conditions do not affect your rights under any applicable consumer protection legislation.
Honeywell Analytics reserves the right to change this policy at any time. Contact Honeywell Analytics for the most current warranty information.
74 96d User Manual M-511401
Honeyewell 8/10
Page 78
Warranty
M-511401 96d User Manual 75 8/10 Honeywell
Page 79
Page 80
Find out more
www.honeywellanalytics.com
Contact Honeywell Analytics:
Americas
Honeywell Analytics Inc.
23500 W. 105th Street, MD 400
Olathe, KS 66061
Tel: +1 913 712 5576
Toll free: +1 800 444 9935
Toll free fax: +1 888 328 2417
detectgas@honeywell.com
Europe, Middle East, Africa
Life Safety Distribution AG
Wilstrasse 11-U11
CH-8610 Uster
Switzerland
Tel: +41 (0)44 943 4300
Fax: +41 (0)44 943 4398
gasdetection@honeywell.com
Technical Services
ha.global.service@honeywell.com
Asia Pacific
Honeywell Analytics Asia Pacific
#508, Kolon Science Valley (1)
187-10 Guro-Dong, Guro-Gu
Seoul, 152-050,
Korea
Tel: +82 (0)2 2025 0307
Fax: +82 (0)2 2025 0329
analytics.ap@honeywell.com
Please Note:
While every effort has been made to ensure accuracy in this publication, no responsibility can be accepted for errors or omissions. Data may change, as well as legislation, and you are strongly advised to obtain copies of the most recently issued regulations, standards and guidelines. This publication is not intended to form the basis of a contract.
www.honeywell.com
M-511401, 08/10
© 2010 Honeywell Analytics
Loading...