HoMedics TE-350-EU Service Manual

HTD8816C TE-350-EU
2 YEAR GUARANTEE
FOREWORD
The Non-Contact Infrared Body Thermometer operating Instructions intend to provide the necessary information for proper operation of the thermometer model. Only body mode was reviewed and certified by notified body. General knowledge of Infrared thermometer and an understanding of the features and functions of the thermometer model are prerequisites for proper use. The Non-Contact Infrared Body Thermometer is a medical device, and can be used repeatedly for the product’s recommended lifetime of 5 years. Please read the manual first before using it, if not fully understand the usages, please stop using the thermometer.
Do not operate the thermometer without completely reading and understanding these instructions.
Notice
Purchase or possession of this device does not carry any express or implied license to use with replacement parts which would, alone or in combination with this device, fall within the scope of one of the relating patents.
For further information contact:
HoMedics Group Ltd, Somerhill Business Park, Five Oak Green Road, Tonbridge, Kent TN11 0GP. Customer Service: service@homedics.co.uk
Company Name: Wellkang Ltd Company Address: 16 Castle St, Dover, CT16 1PW, UK Tel: +44 (20)30869438, 32876300 Fax: +44(20)76811874 Web: www.CEmark.com, www.CE-marking.com, www.CE-marking.eu, Email: AuthRep@CE-marking.eu
GB | 2
SAFETY INFORMATION
This device may only be used for the purposes described in these instructions. The manufacturer cannot be held liable for damage caused by incorrect application. The Non-Contact Infrared Body Thermometer is designed to minimize the possibility of hazards from errors in the software program by following sound and light engineering design processes, Risk Analysis and Software Validation.
WARNING
Warnings are identified by the WARNING symbol shown above.
• The Non-Contact Infrared Body Thermometer is to be operated by consumers in the home setting and primary care setting as screening tool. This manual, accessories, Directions for Use, all precautionary information, and specifications should be read before use.
• This product is designed to measure human body temperature on the forehead. Do not use it for any other purpose.
• This product is intended for use in the home setting and primary care setting as screening tool.
• Do not use the thermometer if it malfunctions or has been damaged in any matter.
• When the ambient temperature of the thermometer changes too much, such as moving the Thermometer from one place of lower temperature to another place of higher temperature, allow the thermometer to remain in a room for 30 minutes where the temperature is between 15°C to 40°C (59°F- 104°F).
• Remove primary batteries if equipment is not likely to be used for long time.
• This product is not waterproof, do not immerse in water or other liquid; If cleaning and disinfecting, please follow the “Care and Storage” section requirements.
• Do not touch the sensor of infrared detection with your fingers.
• Avoid using the thermometer to take a body temperature reading while using a cold compress or other measure to reduce fever.
• If measuring human forehead temperature, please select “body” mode; for measuring other objects, liquids, food and other temperature please select “surface” mode.
• This product must be operated in a stable environment. If the ambient environment is unstable, please check whether there is fog on the sensor and remove this before using in accordance with the “Care and Storage” section.
GB | 3
• Do not near strong electrostatic field or strong magnetic fields, thus avoiding the impact on the accuracy of the measurement data.
• Do not mix the old and new batteries to avoid damage to the product.
• It may aect the accuracy of measurements when the forehead is covered by hair, perspiration, cap or scarf.
• The measuring result of this product is only for your reference. If you have any doubt, please measure the temperature in other methods.
The device should be kept out of the reach of children/pets. When not in use, store
the device in a dry room and protect it against extreme moisture, heat, lint, dust and direct sunlight. Never place any heavy objects on the thermometer body.
Do not throw batteries into fire. Only use recommended batteries. Do not use rechargeable batteries. This thermometer does not replace a hospital diagnosis. Do not drop, disassemble or modify the device. Do not use this device if you think it is damaged or notice anything unusual. This device comprises sensitive components and must be treated with caution.
Observe the storage and operating conditions described in the ‘Technical
Specifications’ section.
Not servicing/maintenance while the thermometer is in use. When using, shall not touch battery and the patient simultaneously. Do not use the device if it is damaged/ degraded/loosened in any way. The continuous
use of a damaged unit may cause injury, improper results, or serious danger.
Based on the current science and technology, other potential allergic reactions are
unknown.
This equipment needs to be installed and put into service in accordance with the
information provided in the ACCOMPANYING DOCUMENTS.
OVERVIEW
Intended Use
The HoMedics Non-Contact Infrared Body Thermometers are designed to be used for intermittent measurement and monitoring of human body temperature by consumers in the home setting and primary care setting as screening tool.
Description of Non-Contact Infrared Body Thermometer
• Device principle and introduction
GB | 4
• The HoMedics Non-Contact Infrared Body Thermometer is a hand-held, reusable,
40
battery operated device, which can measure human body temperature on forehead, the skin temperature on one’s forehead.
• The operation principle is based on Infrared Sensor technology. The IR sensor can output dierent signal when measuring dierent object temperature or in dierent ambient temperature, and the ASIC can turn the signal from IR Sensor to a digital value and display it on the LCD.
• Description on Controls, Indicators, and Symbols
Figure 1: Non-Contact Infrared Thermometer
1. IR sensor
2. Liquid crystal display (LCD)
3. MODE button
4. MEMO button
5. ON/scan button
6. Battery cover
7. Surface mode
8. Body mode
9. Data indicator
10. Indicator of measurement results
11. Low Battery indicator
12. Voice on/o indicator
13. Memory number
14. Fahrenheit
15. Celcius
16. Memory indicator
GB | 5
THERMOMETER APPLICATIONS
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
40
1cm-5cm
1cm-5cm
40
1cm-5cm
1cm-5cm
40
1cm-5cm
1cm-5cm
40
1cm-5cm
1cm-5cm
40
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
Thermometer Model Number
HTD8816C
TE-350-EU
EQUIPMENT SYMBOLS
40
Thermometer Style
Non-Contact Infrared Body Thermometers
Adult Pediatric
Ear Forehead Ear Forehead
ü ü
Warning Restriction of Hazardous Substances
Non-sterile packaging Batch code
Refer to operating instructions Compliance with WEEE Standard
Operating Temperature
DO NOT THROW AWAY Intended for multiple use
Operating Humidity Operating atmospheric pressure
This device compiles with Part 15 of FCC (Federal Communications Commission) Rules.
Manufacturer
Recyclable Serial number
IP22: The first number 2: Protected against solid foreign objects of Ф 12.5 mm and greater. The second number: Protected against vertically falling water drops when enclosure
GB | 6
Indicates this device is in compliance with MDD 93/42/EEC. 0598 is the Notified Body Number
Authorized Representative in the European community
TECHNICAL SPECIFICATIONS
Measurement unit °C/°F
Operating mode Adjusted mode (Body mode)
Reference Body Site Axillary
Rated output range Body mode:
Output range Body mode:
Laboratory Accuracy Body mode:
Display Resolution 0.1°C/0.1°F
Three-colour Backlight (Colour Alarm)
Auto Power O Time ≤18s
Measuring Time ≤ 2s
Measuring Distance 1 - 5cm (0.4 - 2 in)
Memories 50
Direct mode (surface mode)
34.0 - 43.0°C / 93.2 - 109.4°F Surface mode: 0 - 100.0°C / 32 - 212°F
34.0 - 43.0°C / 93.2 - 109.4°F Surface mode: 0 - 100.0°C / 32 - 212°F
34.0 - 34.9°C: ± 0.3°C / 93.2 - 94.8°F: ± 0.5°F
35.0 - 42.0°C: ± 0.2°C / 95.0 - 107.6°F: ± 0.4°F
42.1 - 43.0°C: ± 0.3°C / 107.8 - 109.4°F: ± 0.5°F
Surface mode: ± 2°C / ± 3.6°F
35.5 - 37.3°C / 95.9 - 99.1°F: Green (Normal Temperature)
37.4 - 38.0°C (Alarm point) / 99.3 - 100.4°F: Yellow (Slight Fever)
38.1 - 43.0°C / 100.6 - 109.4°F: Red (High Fever) Note: 1. Surface mode is always with Green backlight
2. In body mode 34.0-35.4°C is with green backlight
GB | 7
Power Supply Requirements
Batteries 2 x 1.5V (AAA) Alkaline battery (IEC Type LR03)
Adaptable Range 2.6V - 3.6V
Environmental
Operating Condition Operating Temperature:
Transport and Storage Condition
Dimension and Weighting
Weight (without batteries) 80g
Size L:122mm X W:96mm X H:42mm
Compliance
Item Compliant with
Equipment classification Safety Standards:
Type of protection Internally powered equipment (on battery power)
Degree of protection Non-Applied part
Front panel and case labelling EN ISO15223-1:2016
Temperature EN ISO80601-2-56:2017
Home healthcare environment EN 60601-1-11:2015
15 - 40°C (59 - 104°F)
Storage Temperature:
-20 - 55°C (-4 - 131°F)
Relative Humidity: ≤ 85%
Relative Humidity: ≤ 93%
EN 60601-1: 2006+A1:2013, EN 60601-1-2: 2015
Atmospheric pressure: 70-106Kpa
Atmospheric pressure: 70-106Kpa
GB | 8
Calculated values of the indicators according to ISO 80601-2-56
Indicators Calculated value
Clinical bias, rcb
Standard deviation, σj 0.14
Limits of agreement, LA 0.26
Clinical repeatability, σr 0.07
Note: the above value is calculated from clinical data of HTD8818A.
-0.027
Safety classification of ME EQUIPMENT
Protection against electric shock Internally powered ME equipment
Applied part Non-Applied part
Protection against harmful ingress of water or particulate matter
Mode of operation Continuous operation
Note: Not intended to be sterilized. Not for use in an OXYGEN RICH ENVIRONMENT.
IP22
GB | 9
OPERATION
1cm-5cm
Battery installation
Caution: The Non-Contact Infrared Body Thermometer does not operate with dead batteries or external power sources. Install new batteries.
1. Slide the battery door downward, toward the bottom of the Non-Contact Infrared Body Thermometer, and remove the battery access door;
2. Insert two pieces AAA size batteries according to the “+” and “-”;
3. Close the battery cover.
How to operate
Before applying the Thermometer, be sure to read and understand all warnings listed of the instructions before use.
• The thermometer is aligned with the middle of the forehead to measure body temperature between the eyebrows above and keep the vertical distance, press the On/scan button, temperature will display immediately, see figure 2.
Figure 2: Measuring position and distance
• When the ambient temperature of the thermometer changes too much, such as moving the Thermometer from one place of lower temperature to another place of higher temperature, allow the thermometer to remain in a room for 30 minutes where the temperature is between 15°C to 40°C.
• The ambient temperature around the test person should be stable, should keep away from the larger flow fan, air-conditioning vents and so on.
• When people moving from one place of lower temperature to another place of higher temperature, should at least remain in the test environment more than 5 minutes, to be consistent with the ambient temperature after the re-measurement.
GB | 10
• Wait at least 1 second for the next measurement. If the continuous measurement
1cm-5cm
of five times, it is recommended to wait at least 30 seconds and then continue measurement.
• You cannot use the thermometer in place where the sun is strong.
• If for some reason the forehead temperature measurement is low, try measurement behind the ears. See figure 3.
1cm-5cm
Figure 3: Align behind the ears to measurement
GENERAL SET UP AND USE
Start measuring
1. Turn on the thermometer by pressing the On/scan button. The thermometer will perform self-test with all segments displayed for 2 seconds.
2. Align the thermometer with the forehead, keeping the recommended distance, and then press the ON/scan button to start the measurement. Data reading will be displayed.
Note:
1. After measurement you will hear a “beep beep beep” three times which means that the measurement is complete. The temperature is displayed on the LCD. The LCD backlight colour will change red, green or yellow according to the measurement. Green means Normal Temperature (35.5-37.3°C / 95.9-99.1°F) ready for next measurement. Yellow means Slight Fever Warning (37.4-38.0°C /
99.3-100.4°F) Please pay attention to body temperature. Red means High Fever (38.1-43.0°C / 100.6-109.4°F) Please take action to cool down or go to the doctor.
2. To ensure the accuracy of the measurement, wait at least 30 seconds after 5 consecutive measurements.
GB | 11
Mode conversion
• When the device is running, pressing the MODE button to cycle conversion between “body” mode and “surface” mode.
• “Body” mode is used for measuring human body temperature, the “surface “ mode is used to measure the surface temperature. (The factory default is “body” mode)
Recalling and Erasing Memory Data
The last temperature taken before the thermometer powers o is stored in memory, up to 50.
1. In the boot or shutdown state, short press the MEMO button to view the history of measured values.
2. An empty memory cell shows “---°C” or “---°F”.
3. Temperature readings can be stored in memory. Up to 50 temperature readings can be stored into the memory cells and automatically overwrite historical data.
4. In boot mode, press the MEMO button until the LCD display “CLR”, which means that all stored data is cleared completely.
Parameter settings
As below step:
1. Unit Set-F1 Under the boot mode, long press the “MODE” button to enter F1, press the “MODE” or “MEMO”button to switch Celsius and Fahrenheit temperature units, press the On/ Scan button to confirm the unit settings (factory default is Celsius).
2. Fever alert set-F2 Under F1 state, press ON/scan button to enter the F2, press the “MEMO” button to decrease 0.1°C, press the “MODE” button plus 0.1°C, long press to accelerate the speed of temperature regulation, and finally press the ON/scan button to save. (The factory default is 38.1°C)
3. Prompt sound settings-F3 Under F2 state, short press ON/scan button to enter F3, press MODE button or MEMO button to set voice switch, and press the ON/scan button to confirm the settings. (The factory default is the voice Prompt to open).
GB | 12
• Prompt sound settings ON/OFF function
Under the boot mode. Short press on/o button, at the same time press the “MODE” button to take on or take o the Prompt sound function.
• Restore to factory setting function Under the boot mode, short press on/o button, at the same time press the MEMO button until LCD display “rst”. Two seconds later, former F1-F3 parameter back to factory setting.
GB | 13
TROUBLESHOOTING
Message Situation Solution
Temperature taken in not within Typical human temperature range. (34.0°C ~ 43.0°C or 93.2°F ~ 109.4°F).
Measured over the distance: 1-5cm (0.4-2in) Optimum measurement distance is 1cm
Incorrect test position. See figure 2 Measuring position and distance.
Subjects forehead hair, Antipyretic stickers, head with sweat, etc.
Some people’s body temperature is lower than the general population.
Operating temperature exceeds the range of specified temperature.
The screen flicker, automatic turn o.
Battery capacity is too low. Taking Temperature is not allowed.
Ambient temperature changes too fast Wait until the ambient temperature is stable.
1. Power is o.
2. Improper battery installation.
3. The battery is exhausted.
4. Display remains blank.
Make sure the forehead thermometer is for forehead measurement, not other human body site.
Subjects sit quietly 5-10 minutes before the test.
The main concern fever temperature
Move to a room within the operating range wait 30 minutes before taking temperature.
Replace battery or the product has been damaged, needs repairs.
Install a new battery
1. Press ON button again.
2. Check the battery polarity.
3. Replace with a new battery.
4. Contact the retailer or service center.
REPLACING THE BATTERY
1. Open and release battery cover following indicator on the surface of battery cover. Before changing the battery be sure the thermometer is already turned o.
2. Remove the battery and replace with 2 new type AAA batteries, Make sure to align properly as indicated inside the battery cover.
3. Slide the battery cover back in until it snaps in place. Do not dispose of used batteries in household waste. Take them to special local collection sites.
GB | 14
4. If the thermometer appears frozen after changing the batteries, remove batteries, wait 30 seconds and replace batteries again.
WARNING Do not recharge, disassemble or dispose of in fire.
1. The typical service life of the new and unused batteries is 2000 measurements for the operation time is 18s.
2. Only use the recommended batteries, do not recharge non-rechargeable batteries and do not burn them.
3. Remove the batteries if the thermometer is not to be used for a long period.
CLEANING, CARE AND STORAGE
• The lens is very delicate.
• It is very important to protect the lens from dirt and damage.
• Use a clean, soft cloth to clean the surface of the device and LCD. Do not use solvents or immerse the device into water or other liquids.
• Always keep the thermometer within the storage temperature range (- 20°C to 55°C or – 4°F to 131°F) and humidity range (≤93% non-condensing).
• It is recommended to store the thermometer in a dry location free from dust. Do not expose the thermometer to direct sunlight, high temperature/ humidity or any extreme environment, otherwise the function will be reduced.
• When the ambient temperature of the thermometer changes too much, such as moving the thermometer from one place of lower temperature to another place of higher temperature, allow the thermometer to remain in a room for 30 minutes where the temperature is between 15°C to 40°C.
DISPOSAL
• Used batteries should not be disposed of in the household rubbish. Used Batteries should be deposited at a collection point.
• At the end of its life, the appliance should not be disposed of in household rubbish.
• Enquire about the options for environment-friendly and appropriate disposal. Take local regulations into account.
GB | 15
WARRANTY
Our company warrants Non-Contact Infrared Body Thermometer at the time of its original purchase and for the subsequence time period of 2 years.
The warranty does not cover the followings:
• The device series number label is torn o or cannot be recognized.
• Damage to the device resulting from misconnection with other devices.
• Damage to the device resulting from accidents.
• Changes performed by users without the prior written authorization of the company.
• Batteries and packaging are not covered under warranty
When asked to provide warranty service, you must have a purchase date and purchase stamp dealers (including dealers name and address) of the warranty card. Be sure to ask the dealer to purchase this product signature on the warranty card. When asked to provide warranty service, please put the product to get our distribution points for repair. Products outside the warranty expires, will be charged accordingly.
Note:
1. If you have any problems with this device, such as setting up, maintaining or using, please contact with SERVICE PERSONNEL of FKA Brands Ltd. Don’t open or repair the device by yourself.
2. Please report to FKA Brands Ltd if any unexpected operation or events occur.
3. Calibration needed to ensure proper function for every two years or after device impact.
4. The patient is an intended operator. The patient can measure and change battery. Under normal circumstances and maintain the device and its accessories according to the user manual.
EMC DECLARATION
1. This equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCUMENTS; This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be aected by portable and mobile RF communications equipment.
GB | 16
2. * Caution: Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
3. *Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!
4. * Caution: this machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used.
Guidance and manufacturer’s declaration-electromagnetic emissions
The Non-Contact Infrared Body Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Thermometer should assure that it is used in such an environment.
Emissions test Compliance
RF emissions CISPR 11 Group 1
RF emissions CISPR 11 Class B The Non-Contact Infrared Body
Harmonic emissions IEC 61000-3-2
Voltage fluctuations Flicker emissions IEC 61000-3-3
Not applicable
Not applicable
Electromagnetic environment guidance
The Non-Contact Infrared Body Thermometer uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Thermometer is suitable for use in all establishments other that domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
GB | 17
Guidance and manufacturer’s declaration-electromagnetic immunity
The Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Thermometer should assure that it is used in such an environment.
Compliance
Immunity test IEC 60601 test level
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient/burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
±6kV contact ±15kV air
±2 kV for power supply lines ±1 kV air for input/ output lines
+/- 1kV line(s) to line(s) =/- 2kV line(s) to earth
<5% UT (>95% dip in UT for 0.5 cycle 40% UT (60% dip in UT for 5 cycles 70% UT (30% dip in UT for 25 cycles 5% UT (>95% dip in UT for 5 sec
30 A/m 30 A/m
Electromagnetic environment-
level
guidance Floor should be wood, concrete or
±6kV
ceramic tile. If floors are covered
contact
with synthetic material, the relative
±15kV air
humidity should be at least 30%.
Mains power quality should be that
Not
of a typical commercial or hospital
applicable
environment.
Mains power quality should be that
Not
of a typical commercial or hospital
applicable
environment.
Mains power quality should be that of a typical commercial or hospital environment. If the use of the Thermometer requires continued
Not
operation during power mains
applicable
interruptions, it is recommended that the Thermometer be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
GB | 18
Guidance and manufacturer’s declaration-electromagnetic immunity
The Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Thermometer should assure that it is used in such an environment.
IEC 60601
Immunity test
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations from radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can not be predicted theoretic call with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Thermometer is used exceeds the applicable RF compliance level above, the Thermometer should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Thermometer. b. Over the frequency range 150kHz to 80MHz, field strengths should be less than 10 V/m.
test level 3Vrms
150kHz to 80 MHz
10 V/m 80MHz to
2.5 GHz
Compliance level
Not applicable
10 V/m
Electromagnetic environment-guidance
Portable and mobile RF communications equipment should be used no closer to any part of the Thermometer, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance. d=1.2 d=1.2 80 MHz to 800 MHZ d=2.3 800 MHz to 2,5 Ghz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,” a. should be less than the compliance level in each frequency range.” b. Interference may occur in the vicinity of equipment marked with the following symbol:
GB | 19
Recommended separation distances between portable and mobile RF communications equipment and the Thermometer
The Thermometer is intended for use in an electromagnetic environment in which radiated RF distances are controlled. The customer or the user of the Thermometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Thermometer as recommended below according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter / W
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73 1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter / m
150 kHz to 80 MHZ
d=1.2
80 MHz to 800 MHZ
d=1.2
GB | 20
800 MHz to 2.5 GHz
d=2.3
BODY TEMPERATURE
• Body temperature varies from person to person and fluctuates during the course of the day. For this reason, it is suggested to know one’s normal, healthy forehead temperature to correctly determine the temperature.
• Body temperature runs approximately from 35.5°C to 37.8°C (95.9°F-100°F). To determine if one has a fever, compare the temperature detected with the person’s normal temperature. A rise over the reference body temperature of 1°C (1°F) or more is generally indication of fever.
• Dierent measurement sites (rectal, axillary, oral, frontal, auricular) will give dierent readings. Therefore it is wrong to compare the measurement taken from dierent sites.
• Here below are typical temperatures for adults, based on dierent measurement sites:
Rectal: 36.6°C to 38°C /97.9°F-99.1°F Axillary: 34.7°C to 37.3°C /94.5°F-99.1°F
ASTM laboratory accuracy requirements in the display range of 37°C (98°F to 102°F) for IR thermometers is ±0.2.(±0.4°F), whereas for mercury in-glass and electronic thermometers, the requirement per ASTM Standards E667-86 and E1112-86 is ±0.1.(±0.2°F).
Caution: This infrared thermometer meets requirements established in ASTM Standard (E1965-98) Except of clause 5.2.2. It displays subject’s temperature over a range of
34.0°C ~43.0°C. Full responsibility for the conformance of this product to the standard
is assumed by (Hetaida Technology Co., Ltd. Add: 4F, Baishida High-Tech Park, Xiandong Industrial Area, Dalingshan Town, Dongguan City, Guangdong, China).
GB | 21
AVANT-PROPOS
Le mode d’emploi du thermomètre infrarouge sans contact a été rédigé afin de donner les informations nécessaires au bon fonctionnement du modèle de thermomètre. Seule la prise de mesure corporelle a été examinée et certifiée par l’organisme notifié. Une connaissance générale du thermomètre infrarouge et une compréhension des caractéristiques et fonctions du modèle de thermomètre sont des conditions préalables à toute utilisation appropriée. Le thermomètre corporel infrarouge sans contact est un appareil médical que vous pouvez utiliser de façon répétée et dont la durée de vie est de 5 ans. Veuillez lire ce mode d’emploi avant de l’utiliser ; si vous ne comprenez pas complètement les consignes données, veuillez cesser d’utiliser ce thermomètre.
N’utilisez aucun le thermomètre sans avoir lu et compris complètement ces instructions et consignes.
Important
L’achat ou la possession de cet appareil ne comporte aucune licence expresse ou implicite d’utilisation avec des pièces de rechange qui, seules ou en combinaison avec cet appareil, relèveraient de l’un des brevets associés.
Pour plus d’informations, contactez :
HoMedics Group Ltd, Somerhill Business Park, Five Oak Green Road, Tonbridge, Kent TN11 0GP. Customer Service: service@homedics.co.uk
Nom de la compagnie: Wellkang Ltd Adresse de la société: 16 Castle St, Dover, CT16 1PW, UK Tel: +44 (20)30869438, 32876300 Fax: +44(20)76811874 Web: www.CEmark.com, www.CE-marking.com, www.CE-marking.eu, Email: AuthRep@CE-marking.eu
F | 22
À PROPOS DE LA SÉCURITÉ
Cet appareil ne peut être utilisé qu’aux fins décrites dans ces instructions et consignes. Le fabricant ne peut pas être retenu pour responsable des dommages causés par une application inappropriée. Le thermomètre corporel infrarouge sans contact est conçu pour minimiser les risques d’erreurs dans le programme logiciel en suivant les processus de conception technique du son et de la lumière, l’analyse des risques et la validation du logiciel.
MISE EN GARDE
Les mises en garde sont identifiées par le symbole de MISE EN GARDE.
• Le thermomètre corporel infrarouge sans contact est destiné aux utilisateurs à domicile et en soins primaires comme outil de dépistage. Veuillez lire attentivement le mode d’emploi, les accessoires, les instructions d’utilisation, toutes les informations de précaution et les spécifications avant utilisation.
• Ce produit est conçu pour mesurer la température du corps humain sur le front.
• Ne l’utilisez pas à d’autres fins que celles prévues,
• Ce produit est destiné à l’usage domestique et pour les soins primaires comme outil
• de dépistage.
• N’utilisez pas le thermomètre s’il fonctionne mal ou a été endommagé de quelque manière que ce soit.
• Si l’écart de température entre l’endroit où le thermomètre est conservé et la pièce où la mesure sera prise est trop élevé, laissez le thermomètre à température ambiante dans cette dernière pendant un minimum de 30 minutes à une température comprise entre 15°C to 40°C (59°F- 104°F).
• Retirez les piles d’origine si l’équipement n’est pas susceptible d’être utilisé pendant une longue période.
• Ce produit n’est pas étanche, ne l’immergez pas dans l’eau ou tout autre liquide ; pour le nettoyer et le désinfecter, veuillez suivre les instructions et consignes de la section « Entretien et conservation ».
• Ne touchez pas la sonde de détection infrarouge avec vos doigts.
• Si une compresse froide a été posée sur les patients fébrile au front, ou si vous prenez d’autres mesures pour faire baisser la fièvre, les données de température seront basses ; évitez dans ce cas de mesurer la température corporelle.
F | 23
• Si vous mesurez la température du front d’un patient, veuillez sélectionner la « prise de mesure corporelle » ; pour mesurer d’autres objets, liquides, aliments et autres températures, veuillez sélectionner la « prise de mesure en surface ».
• Veuillez utiliser ce thermomètre dans un environnement stable. Si cet environnement change et qu’il y a du brouillard sur le capteur, veuillez suivre les instructions et consignes indiquées à la section « Entretien et conservation » pour éliminer le brouillard.
• Ne vous approchez pas d’un champ électrostatique ou magnétique puissant, pour ne pas aecter la précision des données de mesure.
• Ne mélangez pas les anciennes et les nouvelles piles afin d’éviter d’endommager le produit.
• Cela peut aecter la précision des mesures lorsque le front est couvert de cheveux, de transpiration, d’une casquette ou d’un foulard.
• Le résultat de mesure de ce produit est uniquement pour votre référence.
• En cas de doute, veuillez mesurer la température par d’autres méthodes.
L’appareil doit être conservé hors de portée des enfants / animaux domestiques. Veuillez conserver ce thermomètre, lorsqu’il n’est pas utilisé, dans une pièce sèche et protégez-le contre l’humidité extrême, la chaleur, les peluches, la poussière et la lumière directe du soleil. Ne placez jamais d’objets lourds sur le corps du
thermomètre. Ne jetez pas les piles au feu. Utilisez uniquement les piles recommandées. N’utilisez pas de piles rechargeables. Ce thermomètre ne remplace pas un diagnostic hospitalier, Ne faites pas tomber, ne démontez pas et ne modifiez pas le thermomètre. N’utilisez pas ce thermomètre si vous pensez qu’il est endommagé ou si vous
remarquez quelque chose d’inhabituel.
Cet appareil comprend des composants sensibles et doit être traité avec prudence. Respectez les conditions de conservation et de fonctionnement décrites dans la
section « Caractéristiques techniques ».
Pas d’entretien / maintenance lorsque le thermomètre est en cours d’utilisation. Lors de l’utilisation, ne touchez pas la pile et le patient simultanément. N’utilisez pas le thermomètre s’il est endommagé / dégradé / desserré dans ses
parties de quelque façon que ce soit. L’utilisation continue d’un thermomètre
endommagé peut causer des blessures, des résultats incorrects ou un grave danger. D’autres réactions allergiques potentielles sont inconnues après la science et la
F | 24
technologie actuelles.
Cet équipement doit être installé et mis en service conformément aux informations
données dans les DOCUMENTS QUI L’ACCOMPAGNE.
PRÉSENTATION
Usage prévu
Les thermomètres infrarouges pour le corps HoMedics sans contact sont conçus pour être utilisés pour la mesure et la surveillance intermittentes de la température du corps humain par les consommateurs à domicile et en soins primaires comme outil de dépistage.
Description du thermomètre infrarouge sans contact
• Principe de l’appareil et introduction
• Le thermomètre corporel infrarouge HoMedics sans contact est un appareil portatif, réutilisable, fonctionnant sur batterie, qui peut mesurer la température du corps humain sur le front, la température de la peau sur le front.
• Le principe de fonctionnement se base sur la technologie des capteurs infrarouges.
• Le capteur IR peut émettre un signal diérent lors de la mesure de diérentes températures d’objet ou à diérentes températures ambiantes, et l’ASIC peut
• transformer le signal du capteur IR en une valeur numérique et l’acher sur l’écran LCD.
F | 25
Figure 1: Non-Contact Infrared Thermometer
40
1. Capteur IR
2. Achage à cristaux liquides (LCD)
3. Bouton de stockage en mémoire MODE
4. Bouton de mémorisation MEMO
5. Bouton ON/scan
6. Couvercle du compartiment de la pile
7. Prise en mesure en surface
8. Prise de mesure corporelle
APPLICATIONS DU THERMOMÈTRE
Numéro de
modèle
du thermomètre
HTD8816C
TE-350-EU
Type de thermomètre
Thermomètres infrarouges sans contact
9. Indicateur de données
10. Indicateur de résultat de mesure
11. ndicateur de batterie faible
12. Indicateur marche / arrêt du voix
13. Numéro de mémoire
14. Fahrenheit
15. Celsius
16. Indicateur de mémoire
Adulte Enfant
Oreille Fron t Oreille Front
ü ü
F | 26
SYMBOLES D’ÉQUIPEMENT
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
40
1cm-5cm
1cm-5cm
40
1cm-5cm
1cm-5cm
40
1cm-5cm
1cm-5cm
40
1cm-5cm
1cm-5cm
40
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
1cm-5cm
Mise en garde Restriction des substances dangereuses
Emballage non stérile Code lot
Veuillez consulter le mode d'emploi
40
Température de fonctionnement
Humidité ambiante
Cet appareil est conforme à la partie 15 des règles de la FCC (Federal Communications Commission).
Fabricant
Conformité à la norme DEEE
NE PAS JETER Destiné à un usage multiple
Pression atmosphérique de fonctionnement
Indique que cet appareil est conforme à MDD 93/42/CEE. 0598 est le numéro de l’organisme notifié
Représentant agréé dans la Communauté Européenne
Recyclable Numéro de série
Le premier numéro 2 : Protégé contre les corps étrangers solides de ‘V 12,5 mm et plus. Le deuxième numéro 2 : Protégé contre les gouttes d’eau tombant verticalement sur l’enveloppe
F | 27
CARACTÉRISTIQUES TECHNIQUES
Unité de mesure °C/°F
Mode de fonctionnement
Zone corporelle de référence
Plage de sortie nominale
Plage de sortie étendue Prise de mesure corporelle : 34.0 - 43.0°C / 93.2 - 109.4°F
Précision de laboratoire Prise de mesure corporelle
Résolution d’achage 0.1°C/0.1°F
Rétroéclairage tricolore (alarme couleur)
Temps de mise hors tension automatique
Durée des mesures ≤ 2s
Distance de mesure 1 - 5cm (0.4 - 2 “)
Mémoires 50
Mode ajusté (prise de mesure corporelle) Mode direct (prise de mesure en surface)
Axillaire
Prise de mesure corporelle : 34.0 - 43.0°C / 93.2 - 109.4°F Prise de mesure en surface : 0 - 100.0°C / 32 - 212°F
Prise de mesure en surface : 0 - 100.0°C / 32 - 212°F
34.0 - 34.9°C: ± 0.3°C / 93.2 - 94.8°F: ± 0.5°F
35.0 - 42.0°C: ± 0.2°C / 95.0 - 107.6°F: ± 0.4°F
42.1 - 43.0°C: ± 0.3°C / 107.8 - 109.4°F: ± 0.5°F
Prise de mesure en surface ± 2°C / ± 3.6°F
35.5 - 37.3°C / 95.9 - 99.1°F: Vert (température normale) ;
37.4 - 38.0°C (point d’alarme) / 99.3 - 100.4°F: Jaune (légère fièvre)
38.1 - 43.0°C / 100.6 - 109.4°F: Rouge (fièvre élevée) Remarque : 1. La prise de mesure en surface se fait toujours avec un rétro-éclairage vert.
2. Le mode prise de mesure corporelle ache toujours la température comprise entre 34,0 et 35,4 °C avec une rétro-éclairage vert
≤18s
F | 28
Exigences d’alimentation
Piles 2 x 1.5V (AAA) Pile alcaline (IEC Type LR03)
Plage adaptable 2,6V 3,6V
Conditions environnementales
Conditions de fonctionnement
Température de fonctionnement et de stockage
Dimension et poids
Poids (sans piles) 80g
Taille L:122mm X W:96mm X H:42mm
Conformité
Article Conforme à
Classification des équipements Valeurs calculées des indicateurs selon
Type de protection Équipement alimenté en interne (sur batterie)
Degré de protection Partie non appliquée
Étiquettes du panneau avant et du boîtier
Température EN ISO80601-2-56:2017
Contexte : soins à domicile EN 60601-1-11:2015
Température de fonctionnement : 15 °C - 40 °C (59 - 104 °F), humidité relative 85 %, pression atmosphérique : 70 - 106 Kpa
Température de stockage : -20 - 55 °C / -4 - 131 °F, humidité relative 93 %, pression atmosphérique :70 - 106 Kpa
EN 60601-1: 2006+A1:2013, EN 60601-1-2: 2015
EN ISO15223-1:2016
F | 29
Valeurs calculées des indicateurs selon ISO 80601-2-56
Indicateurs Valeur calculée
Biais clinique, rcb
Deviazione standard, σj 0.14
Limites de l’accord, LA 0.26
Répétabilité clinique, σr 0.07
Remarque : la valeur ci-dessus est calculée à partir des données cliniques de HTD8818A.
-0.027
Classification de sécurité de ME EQUIPMENT
Protection contre les décharges électriques Équipement ME, alimenté en interne
Partie appliquée Partie non appliquée
Protection contre la pénétration nocive d’eau ou de particules
Mode of operation Fonctionnement continu
Remarque : Non destiné à être stérilisé. Ne pas utiliser dans un ENVIRONNEMENT RICHE EN OXYGÈNE.
IP22
F | 30