The information contained in this Manual is confidential and proprietary to Hologic, Inc. This
information is provided only to authorized representatives of Hologic's customers solely for the
purpose of facilitating the use of Hologic's products. No information contained herein may be
disclosed to any unauthorized person for any purpose whatsoever without the prior written
consent of Hologic, Inc.
Hologic
products and company names, used in this manual, are trademarks and registered trademarks of
other manufacturers.
The procedures described in this document are intended solely for use by direct employees of
Hologic, Inc. or authorized Hologic Equipment Resellers and their trained technicians. Any
unauthorized or untrained persons performing these procedures may affect the warranty of the
Hologic Model QDR 4500.
Exercise proper caution when servicing the system. There are dangerous and potentially lethal
voltages accessible within the QDR 4500 system. To avoid exposure to shock hazards, the Main
circuit breaker should be switched off, and the power cord removed, before working inside any
part of the system.
The QDR 4500 Fan Beam X-ray Bone Densitometer produces ionizing radiation in the form of
X-rays. It may be dangerous to the patient, operator or service technician unless safe exposure
factors and operating instructions are observed. To avoid unsafe exposure, do not attempt to
service this equipment unless you are a Hologic, Inc. certified service technician. Exercise proper
caution when servicing the system. A dosimeter (film badge) should always be worn while on
site. Dose and scatter radiation measurements must be taken after each service call to ensure that
these parameters are still within specifications.
Hologic, Inc. has made all reasonable effort to ensure that the information in this manual is
accurate and complete. Hologic, Inc. shall not, however, be liable for any technical or editorial
errors or omissions contained herein, or for incidental, special or consequential damages in
connection with the furnishing or use of this material. The information contained in this manual
is subject to change without notice.
QDR 4500
and the Hologic logo are registered trade marks of Hologic, Inc. All other
Printed in U.S.A.
Copyright
ii
1998, 1999 by Hologic, Inc., All rights reserved
RECORD OF REVISIONS
Document # 080- 0462
Date:Revision:Description:
3/95ARelease document per ECO 2501.
5/95BRevised to include information about models QDR 4500W and QDR
4500C per ECO 2597
12/96CRevised per ECO 3279
6/98DAdd FDA required information about system specifications (Section 1).
Add new Leakage Testing Procedure on page 3-25.
Add Reproducibility Test on page 3-27.
Add information about Radiation Measurement Report on page 3-28.
Other revisions per ECO 3822.
SYSTEM OVERVIEW.............................................................................................................................................1-1
Measured Half Value Layer (HVL) At Different Operating Potentials.............................................................. .1-11
Line Voltage and Maximum Line Current..........................................................................................................1-11
Technique Factors for Maximum Line Current...................................................................................................1-11
Maximum Deviation............................................................................................................................................1-11
Measurement Criteria for Technique Factors......................................................................................................1-11
DISTRIBUTION BOARD.........................................................................................................................................2-3
Power ....................................................................................................................................................................2-3
MOTOR CONTROLLER BOARD...........................................................................................................................2-5
Power ....................................................................................................................................................................2-5
Service Switches....................................................................................................................................................2-7
Power ....................................................................................................................................................................2-8
CONTROL PANEL CONTROLLER BOARD................................................................................................. ......2-11
Power ..................................................................................................................................................................2-11
Power ..................................................................................................................................................................2-15
I/O and Logic Board.......................................................................................................................................2-20
Low Voltage Power Supply............................................................................................................................2-20
Power Factor Regulator (PFR) Substitution Board.........................................................................................2-21
DATA ACQUISITION SYSTEM...........................................................................................................................2-26
Solid State Detector.............................................................................................................................................2-27
Analog To Digital Board.....................................................................................................................................2-30
POWER MODULE .................................................................................................................................................2-32
ROOM AND DOORWAY SIZE...............................................................................................................................3-2
ARRANGE FOR HELP.............................................................................................................................................3-4
INSPECT FOR VISIBLE SHIPPING DAMAGE .....................................................................................................3-4
INSPECT FOR HIDDEN SHIPPING DAMAGE .....................................................................................................3-6
TAKE INVENTORY.................................................................................................................................................3-6
MEASURE PATH TO FINAL DESTINATION.......................................................................................................3-6
Short Doorway.......................................................................................................................................................3-6
REMOVE TABLE TOP (IF NECESSARY).............................................................................................................3-7
REMOVE QDR 4500A, OR SL, C-ARM (IF NECESSARY)..................................................................................3-8
PREPARING THE TABLE TOP FOR MOVING (SL only)..................................................................................3-10
MOVE UNIT TO DESTINATION (VERTICAL POSITION)................................................................................3-10
MOVE UNIT TO DESTINATION (HORIZONTAL POSITION)..........................................................................3-11
SET UP UNIT..........................................................................................................................................................3-13
INSTALL QDR 4500A, OR QDR 4500SL, UPPER C-ARM .................................................................................3-15
INSTALL QDR 4500W, OR QDR 4500C, C-ARM................................................................................................3-16
CHECK POWER LINE VOLTAGE........................................................................................................................3-18
Measure Line Voltage.........................................................................................................................................3-18
CHECK BELT TENSION.......................................................................................................................................3-23
ADJUST C-ARM Y BELT......................................................................................................................................3-23
ADJUST A/D GAIN CONTROL.............................................................................................................................3-24
MEASURE X-RAY DOSE TO PATIENT ..............................................................................................................3-24
CHECK HVPS/S (TANK) FOR RADIATION LEAKAGE....................................................................................3-25
CALIBRATE FOR AREA, BMD AND BMC.........................................................................................................3-26
TEST SCAN MODES.............................................................................................................................................3-26
Table Edge to T-Rail (“A” Dimension) Adjustment.........................................................................................4-1
Front to Back T-Rail and Table Edge/Rail Gap Adjustment .............................................................................4-3
C-ARM PARALLELISM ADJUSTMENT (A and SL only).....................................................................................4-3
X-RAY BEAM ALIGNMENT (A and SL only) .......................................................................................................4-4
X-RAY BEAM ALIGNMENT (C and W only)........................................................................................................4-8
v
APERTURE CALIBRATION (A and SL only) ........................................................................................................4-9
MOTOR CALIBRATION.........................................................................................................................................4-9
MOTOR$TZ (QDR 4500A and SL) ...................................................................................................................4-10
C-Arm Y Belt ........................................................................................................................................................5-2
C-Arm Y Motor or Gearcase.................................................................................................................................5-3
C-Arm Y Encoder................................................................................................................................................. 5-4
CONTROL PANEL AND TABLE Y FRUS.............................................................................................................5-4
Control Panel.........................................................................................................................................................5-5
PCBs Under Right-Side of the Table....................................................................................................................5-5
Table Y Belt..........................................................................................................................................................5-6
Table Y Motor or Gearcase...................................................................................................... .............................5-6
Table Y Encoder....................................................................................................................................................5-7
TABLE X FRUS........................................................................................................................................................5-8
Table X Motor Controller PCB.............................................................................................................................5-8
Table X Belt..........................................................................................................................................................5-9
Table X Motor or Gearcase...................................................................................................... ...........................5-10
Table X Encoder .................................................................................................................................................5-10
TABLE Z FRUS (A and SL only)...........................................................................................................................5-11
Linear Potentiometer (Encoder- Obsolete)..........................................................................................................5-13
The Linear Rotary String (Encoder)....................................................................................................................5-13
ARM R FRUS .........................................................................................................................................................5-15
Motor Controller Board.......................................................................................................................................5-15
Arm R Belt..........................................................................................................................................................5-16
Arm R Motor, Gearcase, Encoder or Encoder Belt.............................................................................................5-16
Gas Spring...........................................................................................................................................................5-18
Tank Assembly....................................................................................................................................................5-21
Analog to Digital Converter Board......................................................................................................................5-26
POWER MODULE FRUS.......................................................................................................................................5-27
28 Volt Power Supply.........................................................................................................................................5-27
±
15 Volt Power Supply.......................................................................................................................................5-28
Line Filter .................................................................................................................... ........................................5-29
Power Controller Board.......................................................................................................................................5-29
Aperture Motor PCB...........................................................................................................................................5-33
Aperture Position Belt.........................................................................................................................................5-33
Stepper Motor Assembly.....................................................................................................................................5-36
Figure 5-2. Control Panel and Table Y FRUs.....................................................................................................5-41
Figure 5-3. Left Side Table Y FRUs....................................................................................................................5-41
Figure 5-4. Table X FRUs...................................................................................................................................5-42
Figure 5-5. Table Z FRUs....................................................................................................................................5-42
Figure 5-6. Installing the Rotary String Encoder.................................................................................................5-42
Figure 5-7 C-Arm R FRUs (Outside View).........................................................................................................5-42
Figure 5-8. C-Arm R FRUs (Inside View) ..........................................................................................................5-43
Special Tools.......................................................................................................................................................5-48
BEFORE STARTING................................................................................................................................................6-1
POWER PROBLEMS ............................................................................................................................................... 6-1
No Display.............................................................................................................................................................6-7
No X-Rays.............................................................................................................................................................6-9
The Torque Specifications...................................................................................................................................6-13
Tank Top Cover Components and Screw Location.............................................................................................6-13
Tightening the Lexan Cup Screws.......................................................................................................................6-14
Tightening the Bladder Gasket Screws................................................................................................................6-14
Tightening the Transformer Seal Screws.............................................................................................................6-14
Tightening the Tank Cover Gasket Screws..........................................................................................................6-15
NO A/C LINE INTERRUPTS.................................................................................................................................6-17
FIELD SERVICE PREVENTIVE MAINTENANCE...............................................................................................7-1
Guide Rail and Bearing Maintenance....................................................................................................................7-2
SECTION 8 PCB SUMMARY INFORMATION ...........................................................................8-1
Power Distribution................................................................................................................................................8-1
Signal Distribution............................................................................................................ ....................................8-2
Stepper Motor Controller......................................................................................................................................8-4
Control Panel Controller.......................................................................................................................................8-4
The Hologic QDR4500 X-ray Bone Densitometer (Figure 1-1) estimates the bone mineral
content (BMC) and bone mineral density (BMD) of selected areas of the body, or of the entire
skeleton. It does so using X-rays of two different energy levels. This dual-energy scheme allows
soft tissue within the selected area to be subtracted out, leaving only bone to be scanned and
estimated.
Note:
The patient lies face up on the table and, with the aid of a cross-hair laser, the operator positions
the scanning arm over the region of interest. After entering patient data and selecting the type
and size of scan desired, the operator initiates the scan with a single keystroke.
The operator is not required to select technique factors, as tube current and voltage are preselected and fixed. Since testing is performed by fan beam method rather than by flooding the
area as in conventional radiography, the scanning time is a function of the dimensions of the area
to be measured, the desired resolution and the desired precision.
This manual uses "QDR 4500" to refer to all models in the QDR 4500 series systems.
Information presented in this manual, that applies only to a particular model, or models,
will be noted as such.
BMC results are expressed in grams of calcium hydroxyapatite, and BMD is reported in
grams/cm2 of the same compound.
Note:
The QDR 4500 employs a patented Automatic Internal Reference System, which continuously
calibrates the machine to eliminate the effects of variations in temperature, tube flux, etc. No
daily calibration is required. The daily scanning of a quality control phantom is required to
provide assurance that the system is functioning correctly, and to aid in the detection of any longterm drift.
The X-ray scans produced by the QDR 4500, and displayed on the monitor, are intended only to
locate anatomical sites for measurement, and to assure the operator that the machine is operating
properly. They are not intended as a substitute for conventional film-based diagnostic scans.
In most cases, no additional shielding is necessary for patient, operator or room, and the
QDR 4500 can be placed in any convenient non-shielded examination room. Contact your
state regulatory agency for details about additional shielding requirements, if any.
1-1
Operator's Console
QDR
®
4500 Technical Manual
Scanner Unit
Figure 1-1. QDR
4500
X-RAY SCANNING THEORY
An X-ray source, consisting of a generator and tube in a common, shielded enclosure, is mounted
beneath the patient. It generates a narrow, tightly collimated, fan shaped beam of X-rays which
alternate, at power line frequency, between 100kVp and 140kVp. The source is at one end of a
C-arm. At the other end, above the patient, is a crystal/solid state detector. During a scan, the Carm and table move under computer control to guide the beam over the desired scan area.
Before passing through the patient, the beam is filtered through a rotating drum, in which
alternating segments have radio-opacities equivalent to tissue, bone and air. When finally
intercepted by the detector, the beam contains information about the X-ray absorbing
characteristics of both the patient and the calibration materials in the filter drum. An A/D
converter, fed by the detector, supplies a complex digital signal to the computer, which uses that
signal both to construct the screen display and as the basis for its computations of BMC and
BMD.
The QDR 4500 computer algorithm is based on the principle that bone attenuates the X-ray beam
differently at high and low energies. The bone mineral content of any sample point can be
computed from:
Q = L - kH
where H and L are the logarithms of the sample attenuation at high (140kVp) and low (100kVp)
energies, respectively, and the constant k depends on the tissue attenuation characteristics of the
beam. In the QDR 4500, k is continuously measured using the “tissue” segment in the filter
wheel.
1-2
Section 1 - Introduction
The program works in the following manner:
1. Load preliminary scan and obtain regions of interest from operator.
2. Estimate k as an average value of:
- L
k = [L
tissue
air
] / [H
tissue
- H
air
]
Note:
where L
interposed by the filter drum, and L
The subscript "
indicates a low-energy measurement with tissue-equivalent material
tissue
, H
air
" designates the filter drum segment that is empty (i.e., contains neither
air
tissue
and H
are similarly defined.
air
bone- nor tissue-equivalent material).
3. Using this value of k, calculate Q for each point scanned using the formula given above
(Q = L - kH). This array of Q values constitutes a "Q scan". Displays the Q scan.
4. Compile a histogram of the Q values. Because a large portion of the scan contains soft
tissue only, this histogram will have a large peak. Choose a threshold value just above
this peak, and apply that value to discriminate, point by point in the Q scan, between
"bone" points (whose Q is above the threshold) and "non-bone" points (whose Q is
below the threshold).
Figure 1-2. Q Scan Plot
5.Use the "non-bone" points to calculate a baseline value for each scan line. Using these
points, form a new histogram and repeat steps 4 and 5 until the results converge.
6.Smooth the segment boundaries to eliminate isolated noise-generated "bone" points.
7.Display the "bone" and "non-bone" points for operator approval.
0
8.Determine the constant of proportionality (d
(grams). That constant is determined by measuring how much Q shifts when boneequivalent material is interposed by the filter drum.
9.Calculate the total bone mineral values by adding up the Q values for all "bone" points
in each region of interest (e.g., each vertebra), and multiplying by d
10. Determine the bone areas by counting the number of "bone" points in each region of
interest.
11. Calculate bone mineral density as:
) that relates the Q values to actual BMC
0
.
1-3
QDR
(
)
4500 Technical Manual
®
BMD = BMC / area
12. Display the calculated results and print the report.
FUNCTIONAL OVERVIEW
This section provides a block diagram of the QDR 4500 system along with a brief functional
overview description of each block. A detailed functional description along with interconnection
diagrams and interconnection descriptions is provided in Section 2.
ComputerControls and commands all QDR 4500 hardware modules.
Communications
Controller
Controls the flow of commands to and from the Scanner modules via the
communications bus.
Distribution BoardProvides the interconnections between the QDR 4500 Operator's Console
and the Scanner.
Control Panel
Controller
Interfaces the Scanner’s Control Panel to the Operator's Console
computer software.
Control PanelProvides switches (with visual indicators) for moving the Scanner’s C-
Arm and Patient Table. Also provides an Emergency Stop switch.
TZ Drive Motor
Controller
Controls the motion of the Patient’s Table left and right pedestal motors
based on commands from the computer software.
1-6
Section 1 - Introduction
Table Up/Down
Motor and Position
Encoder
TX Stepper Motor
Driver
Table In/Out Motor
and Position
Encoder
TY Stepper Motor
Driver
Table Left/Right
Motor and Position
Encoder
AR Stepper Motor
Driver
C-Arm Rotation
Motor and Position
Encoder
AY Stepper Motor
Driver
Raises or lowers the Patient Table and provides position monitoring.
Controls the motion of the Patient’s Table in and out motor based on
commands from the computer software.
Moves the Patient Table in and out and provides position monitoring.
Controls the motion of the Patient’s Table left and right motor based on
commands from the computer software.
Moves the Patient Table left and right and provides position monitoring.
Controls the motion of the C-Arm rotation motor based on commands
from the computer software.
Rotates the C-Arm and provides position monitoring.
Controls the motion of the C-Arm left and right motor based on
commands from the computer software.
C-Arm AY Motor
and Position
Encoder
C-Arm InterfaceControls the Aperture and Filter Drum motors, generates timing signals
X-Ray ControllerControls the operation of the X-ray source.
X-Ray Source UnitGenerates the X-ray beam.
Solid State DetectorsConverts the X-rays into electrical signals.
Integrator/
Multiplexor
Analog to Digital
Converter
Positioning LaserProvides a laser beam to assist in positioning the patient on the Patient
Moves the C-Arm left and right and provides position monitoring.
for the X-Ray Controller and the Data Acquisition System, and provides
power to the Positioning Laser.
Integrates the signals from the Solid State Detectors and applies them to
the Analog to Digital Converter.
Converts the analog signals from the Integrator/Multiplexor to a digital
format.
Table.
1-7
QDR
PRODUCT SPECIFICATIONS
SPECIFICATIONMODEL DEFINITION
Scanning Method:A,SLMultidetector array, Indexing table, and motorized C-arm
W,CMultidetector array, Indexing table, and Arm
Switched Pulse Dual-Energy X-ray tube, operating at 100 and
X-ray System:All
Detector System:
SL
C,W
Scanning Sites:
SLLumbar spine (in AP and lateral projections), proximal femur (hip)
WLumbar spine, proximal femur (hip), and whole body
Scan Region:A,W195cm (76.77") x 65cm (25.59")
SL96cm (38") x 65cm (25.59")
Scatter Radiation:
All
Leakage
Radiation:
External Shielding
Requirement:
Calibration:All
System Weight:ScannerConsole
(installed)
(shipping)
All
AllContact state regulatory agency.
SL365kg800lb
W310kg680lb
SL660kg1450lb
W622kg1370lb
140kV, 5mA avg. at 50% duty cycle, 2.5mA avg. at 25% duty cycle,
30s maximum, Tungsten target
216 multichannel detector consisting of CdWO4 scintillators
A
coupled to Silicon diodes
128 multichannel detector consisting of CdWO4 scintillators
coupled to Silicon diodes
64 multichannel detector consisting of CdWO4 scintillators coupled
to Silicon diodes
Lumbar spine (in AP and lateral projections), proximal femur (hip),
A
and whole body
CLumbar spine, proximal femur (hip)
C96cm (38") x 51cm (20")
Less than 10µGy/h (1mrad/h) at 2m (79 in.) from the center of the
X-ray beam for all scans except images, which is less than
10µGy/h (1mrad/h) at 3.5m (138 in.) from the center of the X-ray
beam.
The QDR-4500 meets the requirements of 21 CFR 1020.30(k) for
leakage from the X-ray source
Self Calibrating using HOLOGIC Automatic Internal Reference
System. Operator calibration NOT required.
50% for all scan modes except Whole Body
100% for Whole Body scans
Leakage Technique Factors
The leakage technique factors for all models of QDR 4500’s are the same. It is the maximum
continuous current at the maximum peak potential. This is X-ray mode #3. Peak potential
140/100kVp. (dual energy), current 10mA peak 25% duty cycle or 2.5mA average.
Minimum Beam Filtration
The minimum filtration permanently in the beam is 3.7mm Al equivalent @80kV.
1-10
Section 1 - Introduction
Measured Half Value Layer (HVL) At Different Operating Potentials
Measured operating potentialMeasured Half Value Layer
QDR4500A/SL
80kV3.7mm Al equivalent
100kV4.7mm Al equivalent
140kV7.2mm Al equivalent
QDR4500C/W
80kV3.7mm Al equivalent
100kV5.0mm Al equivalent
140kV6.5mm Al equivalent
Line Voltage and Maximum Line Current
Power
Requirements:
All
100VAC 16A 50/60Hz, Max apparent resistance = 0.32 ohm
120VAC 14A 50/60Hz, Max apparent resistance = 0.32 ohm
230VAC 8A 50/60Hz, Max apparent resistance = 1.28 ohm
Technique Factors for Maximum Line Current
Peak Potential 140kVp
Tube Current 10mA peak, 50% duty factor or 5mA average.
Maximum Deviation
The maximum deviation from the preindication given by labeled technique factor control settings
or indicators are as follows:
The measurement criteria of the technique factors is as follows:
Peak Potential:The voltage peak is measured with an oscilloscope. Voltage is a square pulse.
Peak is defined as the peak voltage of the 4 millisecond pulse shape, not
counting any initial overshoot.
Current:Current is measured with an oscilloscope on the last millisecond of the 4
millisecond pulse.
Time:Time of each pulse is measured with an oscilloscope and defined as the time
between 50% rise and fall times of the peak potential pulse. Time of the scan
is measured by counting the number of AC line pulses from the start to the
end. X-ray pulses are synchronous with the AC line.
1-11
SECTION 2
FUNCTIONAL DESCRIPTION
This section provides a detailed functional description along with interconnection diagrams and
descriptions of the Hologic QDR 4500. Refer to Section 1 for a block diagram and a brief
functional description of each block.
COMPUTER
The QDR 4500 Scanner interfaces to an ISA Bus computer to control table and C-arm movement
and X-ray generation, perform all necessary calculations, and manage patient and QC database
information.
The computer is a Pentium-based (or higher) PC compatible that comes equipped with 3.5-inch
floppy disk and hard disk drives, keyboard, color monitor, and an optional Iomega 1GB JAZ
drive. For details pertaining to the computer and its associated components, please refer to the
documentation shipped with each unit.
COMMUNICATIONS CONTROLLER BOARD
The Communications Controller board handles all the communications between the Computer
and the Scanner’s C-Arm and Table assemblies. The board resides in one of the computer
internal slots and interfaces with the computer via the computer's I/O bus. It connects to the
Scanner’s Distribution board through a 50 conductor ribbon cable. This cable contains two
independent communications links (one asynchronous and one synchronous) and additional
system control signals. Each signal requires a pair of conductors for differential (RS422) noise
immunity.
The asynchronous communications link communicates with the Motor Controller boards (TX,
TY, AY and AR), the TZ Drive board, the C-Arm Interface board, and the Control Panel
Controller board. The synchronous communications link communicates with the Data
Acquisition System (DAS).
Interface Connections
Table 2-1 describes the interconnections between the Communications Controller board and the
Distribution board. The table also identifies the interconnection connector and pin assignments.
Figure 2-1 shows the interconnections between the Communications Controller board and the
Distribution board.
The Distribution board provides interconnections between the QDR 4500 Operator's Console and
the Scanner. It passes several signal and power lines from the Operator's Console directly to the
C-Arm Interface module. It also provides buffering and individual drivers and receivers for
various signal lines to and from individual Scanner modules and the Communications Controller.
The Distribution board is located in the Electronics Tray in the base of the Scanner.
Three cables connect the Operator's Console to the Distribution board. One cable connects the
communications buss to the Communications Controller located in the Operator's Console
computer. Two cables connect to the Operator's Console Power Module. These two cables
connect power to the Distribution board and connect the X-Ray On and Emergency signal lines
to the Power Module.
Eight cables connect the Distribution board to the various Scanner boards. Four cables connect
to the four Motor Controller boards (Table X, Table Y, Arm R, and Arm Y). Two (one signal
and one power) connect to the C-Arm Interface board. One connects to the TZ Drive board.
And, one connects to the Scanner's Control Panel Controller board.
The Distribution board has provision for three jumpers that can be installed to override the
EMERGENCY signal lines when trouble-shooting.
Power
The Distribution board receives +28, +24 and +/-15VDC from the Operator's Console Power
Module. The +28VDC is applied through four individual circuit breakers to the Table X, Table
Y, Arm R, and Arm Y Motor Controller boards. The +24 and +/-15VDC are passed to the CArm Interface board. +24VDC is also applied through a circuit breaker to the Table Z Drive
board. It is also reduced to +7 and + 5VDC by regulators to power op-amplifiers and analog
switches located on this board. The + 7VDC is passed to the Table Z Drive and Control Panel
Controller boards. The + 5VDC powers the digital section of the Distribution board. The +28
and +24VDC power supplies are not closely regulated and their outputs may range from +24V to
+ 35V under normal conditions.
2-3
®
Note:
QDR
+7VDC may measure anywhere from +6.25VDC to +7.25VDC. This is true everywhere
4500 Technical Manual
+7VDC is shown in this manual.
Six green LEDs indicate the status of the +28, +24, +15, -15, +7 and +5VDC (ON indicates the
respective voltage is present). Five red LEDs indicate the status of the five circuit breakers
applying voltage to the motor drivers/controller. ON indicates the circuit breaker has detected an
over-current condition.
Interface Connections
Figure 2-2 shows boards that connect to the Distribution board.
OPERATOR’S
CONSOLE
COMPUTER
COMMUNICATIONS
CONTROLLER
BOARD
OPERATOR’S
CONSOLE
POWER MODULE
+24 VDC, +/-15VDC,
EMERGENCY
SIGNAL L INES
DISTRIBUTION
BOARD
INSTRUMENT
C-ARM
INTERFACE
BOARD
CONTROL P ANE L
CONTROLLER
BOARD
TZ
DRIVE
BOARD
TX MOTOR
CONTROLLER
BOARD
TY MOTOR
CONTROLLER
BOARD
+28 VDC
Figure 2-2. Distribution Board High Level Interconnection Diagram
2-4
AR MOTOR
CONTROLLER
BOARD
AY MOTOR
CONTROLLER
BOARD
Section 2 - Functional Description
MOTOR CONTROLLER BOARD
The Motor Controller board is a microprocessor controlled power driver circuit for use with a
two coil bipolar stepper motor. It receives high level commands through the Distribution Board
from the host computer, and applies 28 volt pulses to the stepper motor windings. The QDR 4500
uses four identical Motor Controller boards to control and drive the Table X (Table In/Out),
Table Y (Table Left/Right), C-Arm Rotate, and C-Arm Y (C-Arm Left/Right) stepper motors.
The motor windings are driven by two integrated H-bridges. These integrated circuits provide
internal level conversion and power limiting. Their logic level control inputs are driven from a
stepper motor control microcircuit that receives commands from the microprocessor. The control
circuit senses the current in the motor windings and adjusts the duty cycle of the applied voltage
in such a way as to limit the maximum motor current. The maximum value is determined by an
8-bit control word at a Digital to Analog Converter.
Each Motor Controller board monitors the position of its respective mechanism using a signal
received from an associated Position Encoder connected to the mechanism. The Position Encoder
is a precision potentiometer that divides a +/-3V reference source into a sense amplifier in
proportion to the position of the mechanism driven by the motor. The sense amplifier output is
converted to digital code that provides position feedback to the microprocessor.
The Motor Controller boards receive movement commands from the QDR 4500 computer via the
communications bus. Each Motor Controller board contains an ID switch and four status
indicators (LEDs). The ID switch is a 16-position rotary encoded switch (SW1) that is read
during system initialization to determine the Motor Controller board's address for communicating
with the QDR 4500 computer. ID switch settings for the four Motor Controller boards are as
follows:
Table X drive4
Table Y drive5
C-Arm Rotate drive 6
C-Arm Y drive7
The four red status LEDs provide visual indications of motor drive power on, Stepper CPU
active, motor drive direction, and motor step pulses.
The Motor Controller board also has provision (JP2) for connecting limit switches to inhibit
motor operation when the mechanism goes beyond established mechanical limits. This feature is
not used in the QDR 4500 and therefore no cable is connected to JP2.
Power
Power input to the Motor Controller board is +28VDC. This voltage provides the motor drive
power and is converted down to +5VDC for use by logic circuits on the board. The +5VDC is
also converted to -5VDC. Two green LEDs provide visual indication of the power present on the
Motor Controller board.
2-5
QDR
4500 Technical Manual
®
Interface Connections
Figure 2-3 shows the interconnections between the Distribution board, the Motor Controller
boards, the Stepper Motors, and the Position Encoders. Table 2-1 describes the interconnections
between the Distribution board and the Motor Controller. Table 2-3 describes the
interconnections between each Motor Controller and its respective stepper motor and position
encoder. The tables also identify the interconnection connector and pin assignments.
Figure 2-3. Distribution Board/Motor Controller Board
Interconnection Diagram
Table 2-2. Distribution Board/Motor Controller Board
Interconnection Descriptions
SignalDescription
ARD+
Asynchronous Receive Data.JP7JP5J P8JP1111
ARDATD+
Asynchronous Transmit Data.JP7JP5JP8JP1114
ATDSYSRST+
SYSRST28V
28V_RET
System Reset. Resets the Motor
Controller board.
DC power for the Motor
Controller board.
TableXTableYC-Arm
JP7JP5JP8JP1117
JP7JP5JP8JP112,3,4,5
C-ArmYPin(s)
Rotate
12
15
18
1,6,7,8
2-6
Section 2 - Functional Description
Table 2-3. Motor Controller Board/Stepper Motor and Position Encoder
Interconnection Descriptions
SignalDescriptionPin
(No label)Motor drive signals (4).JP5-1 - JP5-4
+REFPrecision positive voltage to position potentiometer.JP3-1
(No label)Position encoder wiper return voltage.JP3-3
-REFPrecision negative voltage to position potentiometer.JP3-5
TZ DRIVE BOARD
The TZ Drive Board is a microprocessor controlled power driver circuit for the two pedestal
motors which raise and lower the patient table in the QDR 4500A and SL. This board is located
in the electronics tray in the bottom of the Scanner base assembly.
The TZ Drive board communicates with the communications controller, via the Distribution
board, to drive the pedestal motors under computer control. Manual raising of the pedestal may
be required in case of an emergency. In this case, manual control is provided through the Table
switch of the Scanner’s Operator Control Panel.
The TZ Drive board monitors the position of each pedestal using signals received from an
associated Position Encoder connected to the respective pedestal.
Service Switches
The TZ Drive board contains four service switches used during replacement of a defective
pedestal or Position Encoder. Table 2-4 describes these switches and their respective functions.
CAUTION:
The TZ drive motors are designed to run at a 5% duty cycle. If the motors overheat,
the built-in thermal cutouts may trip and cause the motors to stop functioning. If
this happens, you must wait about 20 minutes before functionality is restored. See
the heading
Unable to position device within specified tolerance…
on page 6-45.
Table 2-4. TZ Drive Service Switches
SwitchFunction
Mode
(Normal/Service)
Direction (Up/Down)When the TZ drive is in Service mode, determines the direction of pedestal
LeftWhen the TZ drive is in Service mode, moves the left pedestal in the
Determines whether the TZ Drive is in Normal or Service operation.
movement (not active in normal mode).
direction specified by the Direction switch (not active in normal mode).
2-7
QDR
4500 Technical Manual
®
RightWhen the TZ drive is in Service mode, moves the right pedestal in the
direction specified by the Direction switch (not active in normal mode).
Reset
Resets the board after manual operation.
The TZ Drive board must be
reset after any manual operation.
Power
Power input to the TZ Drive board is +24VDC from the Distribution board and 240VAC from
the Operator's Console Power Module. +24VDC powers circuitry located on this board and is
reduced to +5VDC. +5VDC is converted to -5VDC and +/-3VDC. The +5and -5VDC powers
logic circuitry on this board, while the +/- 3VDC provides the reference voltage for the position
sensors. Two green LEDs provide visual indication of the +24 and +5VDC status (ON indicates
the respective voltage is present).
The 240VAC power is connected through control relays to the pedestal motors.
Interface Connections
Figure 2-4 shows the interconnections between the Distribution board, the TZ Drive Board, the
Pedestal Motors, and the Pedestal Position Encoders. Table 2-5 describes the interconnections
between the Distribution board and the TZ Drive Board. Table 2-6 describes the line voltage
(240VAC, line to line) between the Operator's Console Power Module and the TZ Drive Board.
Table 2-7 describes the interconnections between the TZ Drive Board and the two pedestal
motors and their respective position encoders. The tables also identify the interconnection
connector and pin assignments.
2-8
Loading...
+ 220 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.