1.0 Intended Use Statements ...................................................................................................................... xv
1.1 Intended Use .................................................................................................................................. xv
1.2 Intended Use (Tomosynthesis Option) ............................................................................................ xv
2.0 System Capabilities ............................................................................................................................... xv
3.0 Users .................................................................................................................................................... xv
4.0 Skills Needed for System Use .............................................................................................................. xvi
5.0 Training Requirements ......................................................................................................................... xvi
6.0 Quality Control Requirements ............................................................................................................. xvi
7.0 Product Complaints ............................................................................................................................. xvi
8.0 Hologic Cybersecurity Statement ......................................................................................................... xvi
9.0 Warnings, Cautions, and Notes .......................................................................................................... xvii
10.0 Terms and Definitions ..................................................................................................................... xviii
11.0 International Symbols ........................................................................................................................ xix
12.0 Document Standards ......................................................................................................................... xix
Chapter 1—General Information 1
1.0 System Description .................................................................................................................................1
2.0 Safety Information .................................................................................................................................. 3
2.1 General Safety ..................................................................................................................................3
2.4 Data Loss .........................................................................................................................................6
4.3 Log In ............................................................................................................................................ 18
5.0 How to Change the Language .............................................................................................................. 18
6.0 Perform the Functional Tests ................................................................................................................ 19
7.0 How to Turn Off the System ................................................................................................................. 25
8.0 How to Remove All Power from the Acquisition Workstation .............................................................. 25
Chapter 3—The User Interface 27
1.0 Select the Function to Perform ............................................................................................................. 27
2.0 How to Perform the Quality Control Tasks ........................................................................................... 28
3.0 The Select Patient Screen ..................................................................................................................... 29
3.1 How to Open a Procedure ............................................................................................................. 29
3.2 How to Add a New Patient ............................................................................................................ 30
3.3 How to Edit the Patient Information ............................................................................................... 30
3.4 How to Delete a Patient ................................................................................................................ 30
3.5 The Patient Filter Screen ................................................................................................................ 31
3.6 How to Refresh the Worklist .......................................................................................................... 32
3.7 How to Query the Worklist ........................................................................................................... 33
3.8 About the Admin Button ................................................................................................................ 33
3.9 How to Log Out ............................................................................................................................ 33
4.0 The Procedure Screen .......................................................................................................................... 34
4.1 How to Set the Exposure Parameters .............................................................................................. 34
4.2 How to Use the Implant Present Button ......................................................................................... 35
4.3 How to Acquire an Image .............................................................................................................. 35
4.4 How to Add or Remove a View ..................................................................................................... 36
4.5 How to Add a Procedure ............................................................................................................... 37
4.6 How to Edit a View ....................................................................................................................... 37
4.7 How to Close a Procedure ............................................................................................................. 38
5.0 How to Access Image Review Features ................................................................................................ 38
6.0 How to Use the Output Sets ................................................................................................................. 39
6.1 How to Select an Output Set ......................................................................................................... 39
6.2 How to Add or Edit an Output Set ................................................................................................. 39
7.0 How to Use the On-Demand Outputs .................................................................................................. 39
7.1 How to Archive ............................................................................................................................. 39
7.2 How to Print .................................................................................................................................. 40
7.3 How to Export ............................................................................................................................... 41
8.0 How to Use the Paddle Shift Feature .................................................................................................... 41
9.0 About the Taskbar ................................................................................................................................ 42
4.8 FAST Compression Mode ...............................................................................................................56
5.0 Magnification Stand .............................................................................................................................. 57
5.1 How to Install and Remove the Magnification Stand ......................................................................57
6.1 How to Install and Remove the Localization Crosshair Device .......................................................58
6.2 How to Use the Localization Crosshair Device ..............................................................................58
6.3 How to Install and Remove the Magnification Crosshair Device .................................................... 59
6.4 How to Align the Crosshair Device ................................................................................................ 59
Chapter 6—Clinical Procedures 61
1.0 Standard Workflow ............................................................................................................................... 61
1.2 At the Gantry ................................................................................................................................. 61
1.3 At the Acquisition Workstation ....................................................................................................... 61
2.0 Screening Procedure Example .............................................................................................................. 62
2.1 How to Position the Patient ............................................................................................................ 62
2.2 Set the Exposure Techniques .......................................................................................................... 62
2.3 How to Acquire the Exposure ......................................................................................................... 63
2.4 How to Automatically Store the Image ........................................................................................... 63
2.5 How to Accept a Rejected Image ...................................................................................................63
2.6 How to Accept or Reject a Pended Image ...................................................................................... 63
Part Number MAN-01964 vii
User Manual
DRAFT
Table of Contents
Chapter 7—Maintenance and Cleaning 65
1.0 General Information ............................................................................................................................. 65
1.1 For General Cleaning .................................................................................................................... 65
1.2 To prevent Possible Injury or Equipment Damage .......................................................................... 66
2.2 System Messages ........................................................................................................................... 84
viiiPart Number MAN-01964
User Manual
DRAFT
Table of Contents
Appendix C—Dimensions Mobile 85
1.0 General Information ............................................................................................................................. 85
2.0 Conditions for Safety and Other Precautions ......................................................................................... 85
3.0 Mobile Specifications ........................................................................................................................... 86
3.1 Shock and Vibration Limits ............................................................................................................86
5.0 Prepare the System for Travel ...............................................................................................................87
6.0 Prepare the System for Use ................................................................................................................... 87
7.0 Test the System after Travel ..................................................................................................................88
7.1 Selenia Dimensions Controls and Functional Tests ......................................................................... 88
8.0 Quality Control Tests ............................................................................................................................88
Table A-1: Maximum mA Setting as a Function of kV .................................................................................. 81
Table B-1: System Messages ........................................................................................................................84
Part Number MAN-01964 xiii
User Manual
DRAFT
List of Tables
xivPart Number MAN-01964
Preface
DRAFT
1.0Intended Use Statements
United States Federal Law restricts this device to use by, or on the order of, a physician.
1.1Intended Use
The Selenia® Dimensions® Full Field Digital Mammography system generates digital
mammographic images that can be used for screening and diagnosis of breast cancer. The
Selenia Dimensions Full Field Digital Mammography system is intended for use in the same
clinical applications as traditional screen-film mammographic systems. Mammographic
images can be interpreted on either hard copy film or soft copy review workstations.
1.2Intended Use (Tomosynthesis Option)
The Selenia® Dimensions® Full Field Digital Mammography system acquires the digital
mammography images which can be used for screening and diagnosis of breast cancer. The
Selenia Dimensions system is intended for the clinical methods used with conventional Full
Field Digital Mammography systems. The Selenia Dimensions system can acquire
conventional full field digital mammograms in two dimensions and tomosynthesis
mammograms in three dimensions. The screening examination has a conventional image
set, or a conventional image set and a tomosynthesis image set.
User Manual
Preface
Intended Use Statements
Note…In Canada, Tomosynthesis is not approved for screening, and must be
2.0System Capabilities
The system provides the user interfaces for the performance of screening and diagnostic
mammograms:
•Conventional mammography with a digital image receptor equivalent in size to large
mammography film.
•Tomosynthesis scan with a digital image receptor equivalent in size to large
mammography film (Tomosynthesis option).
•Conventional digital mammogram and tomosynthesis scan during one compression
(Tomosynthesis option).
3.0Users
•A Technologist to acquire and review images
•A Technologist to perform the Quality Assurance
•A system administrator to enable permissions
•A Medical Physicist to perform the Quality Control tests
•A Radiologist can use the system with a Technologist
•The service personnel to install the system, set the site system configurations and
calibrations, and find faults
used in conjunction with conventional mammography (2D image set).
Part Number MAN-01964 xv
User Manual
DRAFT
Preface
Skills Needed for System Use
4.0Skills Needed for System Use
You must know how to do the following:
•Perform the trackball operations, like click, drag, and/or select
•Perform the touchscreen operations
•Select from menus
•Type information in text fields
•Select the options in the screens
•Select the entries from drop-down lists
•Use scroll bars
5.0Training Requirements
Hologic™ does not accept the responsibility for injury or damage from incorrect system
operation.
Make sure that you receive training on the Selenia Dimensions before you use this system on
patients. Hologic training programs address MQSA training requirements for any
Technologist or Physician.
Refer to this manual for directions on how to use Selenia Dimensions.
6.0Quality Control Requirements
The facilities in the United States must use the Quality Control Manual to create a Quality
Assurance and Quality Control program. The facility must create the program to meet the
requirements of the Mammography Quality Standards Act or to be accredited by ACR or
another accreditation body.
The facilities outside the United States can use the Quality Control Manual as a guide to
create a program to meet the local standards and regulations.
7.0Product Complaints
Report any complaints or problem in the quality, reliability, safety, or performance of this
product to Hologic. If the device has caused or added to patient injury, immediately report
the incident to Hologic. (See the title page for contact information.)
8.0Hologic Cybersecurity Statement
Hologic continuously tests the current state of computer and network security to examine
possible security problems. When necessary, Hologic provides the updates to the product.
For Cybersecurity Best Practices documents for Hologic products, refer to the Hologic
Internet site.
xviPart Number MAN-01964
9.0Warnings, Cautions, and Notes
DRAFT
Descriptions of Warnings, Cautions, and Notes used in this manual:
WARNING!The procedures that you must follow accurately to prevent
possible dangerous or fatal injury.
User Manual
Preface
Warnings, Cautions, and Notes
Warning:
Caution:The procedures that you must follow accurately to prevent the damage
Note…Notes indicate additional information.
The procedures that you must follow accurately to prevent injury.
to equipment, loss of data, or damage to files in software applications.
Part Number MAN-01964 xvii
User Manual
DRAFT
Preface
Ter ms and Definitions
10.0 Terms and Definitions
ACRAmerican College of Radiology
AECAutomatic Exposure Control
AnnotationsGraphic or text marks on an image to indicate an area of interest.
CollimatorDevice at the x-ray tube to control the area of the receptor
Combo ProcedureAn image acquisition procedure for which the system takes a
Conventional MammographySingle projection x-ray images of views for screening and
Diagnostic WorkstationSoftcopy workstation for diagnoses from digital images.
DICOMDigital Imaging and Communications in Medicine
EMCElectromagnetic Compatibility
GantryA part of the Selenia Dimensions that has the Detector,
that is exposed.
conventional mammography image and a tomosynthesis scan
during a single patient compression (Tomosynthesis option).
diagnostic purposes.
Generator and X-Ray Source, Positioning/Compression,
Power Distribution, and Accessories Subsystems.
GridElement within the Digital Image Receptor that reduces
scatter radiation during the exposure.
HISHospital Information System
HTC™High Transmission Cellular Grid
Image ReceptorAssembly of x-ray detector, x-ray scatter reduction grid,
and carbon fiber cover.
MQSAMammography Quality Standards Act
NoticeAnnotations and comments per image communicated
between Diagnostic Review Workstations, Technologist
Workstations, and Acquisition Workstations.
PACSPicture Archiving and Communications System. A
computer and network system for the transfer and archive
of digital medical images.
PendA mark on the image to indicate the Technologist is not
positive about the image quality. Pended images must be
Accepted or Rejected before the procedure is closed.
Projection ImagesThe group of x-ray images for tomosynthesis taken at different
projection angles through the breast (Tomosynthesis option).
RFRadio Frequency
RISRadiology Information System
ROIRegion of Interest
SIDSource to Image Distance
TomosynthesisAn imaging procedure which combines a number of
projections taken at different angles. The tomosynthesis
images can be reconstructed to show planes or slices
within the object (Tomosynthesis option).
UPSUninterruptible Power Supply
xviiiPart Number MAN-01964
11.0 International Symbols
DRAFT
This section describes the International Symbols on the Selenia Dimensions.
User Manual
Preface
International Symbols
Potential Equalization
terminal
Protective Earth
terminal
OffPower disconnected from the main power source.
OnPower connection to the main power source.
Off
On
WEEE
Dangerous VoltageIdentifies an area of possible lethal voltage.
Manufacturer
Connection for a conductor, except the Protective Earth
terminal, for a direct connection between two or more
pieces of electrical equipment.
Connector used for connection to ground of the line cord
or ground cable of the equipment and no other purpose.
Only a part of the equipment is disconnected from the
main power source.
Only a part of the equipment is connected to the main
power source.
Shows the compliance to the EC Directive on Waste
Electrical and Electronic Equipment (WEEE).
Date of Manufacture
Radio IconThis system transmits non-ionizing radiation
X-ray RadiationCaution—Radiation
12.0 Document Standards
When prompted to add text, enter the text written in monospaced fo nt exactly as
shown.
Part Number MAN-01964 xix
User Manual
DRAFT
Preface
Document Standards
xxPart Number MAN-01964
Chapter 1—General Information
DRAFT
1.0System Description
1.1Tubestand
User Manual
Chapter 1—General Information
System Description
Figure 1-1: Selenia Dimensions
Legend forFigure 1-1
1. Tubestand (Gantry and C-Arm)
2. Gantry
3. C-Arm (Tube Arm and Compression Arm)
4. Tube Arm
5. Compression Arm
Part Number MAN-01964 1
User Manual
DRAFT
Chapter 1—General Information
System Description
1.2Acquisition Workstation
Figure 1-2: Acquisition Workstations
Legend forFigure 1-2
1. Standard Acquisition Workstation
2. Premium Acquisition Workstation
3. Mobile Acquisition Workstation
2Part Number MAN-01964
2.0Safety Information
DRAFT
Read and understand this manual before you use the system. Keep the manual available
during the patient procedures.
Always follow all the instructions in this manual. Hologic does not accept the responsibility
for injury or damage from wrong system operation. Hologic can arrange for training at your
facility.
The Selenia Dimensions has protective devices, but the Technologist must understand how
to safely use the system. The Technologist must remember the health hazards of x-rays.
2.1General Safety
The Selenia Dimensions system is classified as CLASS I, TYPE B APPLIED PART, IPX0,
permanently connected equipment, continuous operation with short term loading per IEC
60601-1. There are no special provisions to protect the system from flammable anesthetics
or ingress of liquids.
WARNING!Do not open any of the panels. This system contains lethal
User Manual
Chapter 1—General Information
Safety Information
voltages.
WARNING!Per North American electrical safety requirements, you
must use a Hospital Grade receptacle to provide a correct
Ground.
WARNING!Electrical equipment used near flammable anesthetics can
cause an explosion.
WARNING!The user must correct problems before the system is used.
The user must arrange for preventive maintenance by an
authorized Service Engineer.
Warning:This device contains dangerous material. Return to Hologic all
material removed from service.
Warning:If a paddle touches possible infectious materials, call your
Infection Control Representative for decontamination
instructions.
Part Number MAN-01964 3
User Manual
DRAFT
Chapter 1—General Information
Safety Information
Caution:The system is a medical device and not a normal computer. Do not
Note…Hologic does not provide the Gantry power cable for some countries. If
2.2Patient Safety
WARNING!After power failure, remove the patient from the system
make changes to the hardware or software that are not authorized.
Install this device behind a firewall for network security. The
computer virus protection or network security for this medical device
is not provided (for example, a computer firewall). The network
security and anti-virus provisions are the responsibility of the user.
the power cable is not provided, the installed cable must meet the
following requirements and all local codes that apply: 3 conductor, 8
AWG (10 mm
before you apply power.
2
) copper not more than 25 feet (7.62 meters) in length.
WARNING!To keep the isolation quality for the system, attach only
approved accessories or options to the system. Only the
authorized personnel can make changes to the
connections.
WARNING!Keep a 1.5 meter safe distance between the patient and
any non-patient devices.
Non-patient system components (like the Workflow
Manager, the diagnostic review workstation, or the hard
copy printer) must not be installed in the Patient Area.
1.5m
Warning:Never leave the patient during the procedure if in contact with
the mammography system.
4Part Number MAN-01964
User Manual
DRAFT
Chapter 1—General Information
Safety Information
Warning:Keep the hands of the patient away from all buttons and
switches at all times.
Warning:The C-Arm movement is motorized.
Warning:You increase the patient dose to high levels if you increase the
AEC exposure adjustment setting. You increase the image noise
or decrease image quality if you decrease the AEC exposure
adjustment setting.
Warning:Put both footswitches away from the patient and C-Arm area to
prevent any accidental footswitch use. When the patient has a
wheelchair, put the footswitches away from the area.
Warning:Control the access to the equipment according to local
2.3Radiation Safety
WARNING!This x-ray system can be dangerous to the patient and the
WARNING!The disk drives installed in this system are a Class I Laser
Warning:For exposures except magnification case studies, always use the
Warning:The Face Shield does not protect from radiation.
regulations for radiation protection.
user. Always follow the safety precautions for x-ray
exposures.
Product. Prevent direct exposure to the beam. Hidden
laser radiation exists if the case to a disk drive is open.
Face Shield.
Part Number MAN-01964 5
User Manual
DRAFT
Chapter 1—General Information
Safety Information
Warning:The bar code scanner installed in this system is a Class II Laser
Warning:You must keep your complete body behind the radiation shield
2.4Data Loss
Warning:Do not move the C-Arm while the system retrieves the image.
Caution:Never turn off the Acquisition Workstation Circuit Breaker except in
Product. Prevent direct exposure to the beam. Hidden laser
radiation exists if the cover is opened.
for the time of the exposure for maximum protection from x-ray
exposure.
emergency. The circuit breaker can turn off the Uninterruptible Power
Supply (UPS) and risk data loss.
Caution:Do not put any magnetic media near or on devices that create any
magnetic fields, because stored data can be lost.
2.5Equipment Damage
Caution:Do not put any heat source on the image receptor.
Caution:To minimize possible damage from thermal shock to the Digital Image
Receptor, follow the recommended procedure to turn off the
equipment.
Caution:Do not make any brightness or contrast adjustments to the display
unless the SMPTE test pattern is on the screen.
Caution:Use the least possible amount of cleaning fluids. The fluids must not
flow or run.
Caution:To prevent damage to the electronic components, do not spray
disinfectant on the system.
6Part Number MAN-01964
2.6Emergency Off Switches
DRAFT
The Emergency Off switches remove the power from the Gantry and the Standard
Acquisition Workstation Lift Mechanism. Do not normally use the Emergency Off switches to
turn off the system. See Chapter 2, page 25, for complete information.
2.7Interlocks
The Selenia Dimensions has safety interlocks:
•The C-Arm vertical drive and rotation is disabled when 45 Newtons (10 pounds) or
greater of compression force is displayed.
•If the x-ray button is released before the end of the exposure, the exposure stops and an
alarm message appears.
•When in Tomo mode, the system does not allow the Grid in the x-ray field
(Tomosynthesis option).
•Mirror and Filter interlocks prevent the x-ray exposure when the Light Field Mirror or the
Filter is not aligned.
User Manual
Chapter 1—General Information
Safety Information
Part Number MAN-01964 7
User Manual
DRAFT
Chapter 1—General Information
Compliance
3.0Compliance
This section describes the mammography system compliance requirements and the
responsibilities of the manufacturer.
3.1Compliance Requirements
The manufacturer has the responsibility for the safety, reliability, and performance of this
equipment with the following provisions:
•The electrical installation of the room meets all requirements.
•The equipment is used according to Instructions for Use.
•The assembly operations, extensions, adjustments, changes, or repairs are performed
only by authorized persons.
•The network and communication equipment must be installed to meet IEC Standards.
The complete system (network and communications equipment and Selenia Dimensions
Mammography System) must be in compliance with IEC 60601-1 and IEC 60601-1-1.
Caution:Medical Electrical Equipment needs special precautions about EMC
and must be installed, put into service and used according to the EMC
information provided.
Caution:Portable and Mobile RF communications can affect Medical electrical
Equipment.
Caution:The use of unauthorized accessories and cables can result in increased
emissions or decreased immunity. To keep the isolation quality for the
system, attach only approved Hologic accessories or options to the
system.
Caution:
Caution:This system is intended for use by healthcare professionals only. This
The Medical Electrical (ME) Equipment or ME System should not be used
adjacent to or stacked with other equipment. If adjacent or stacked use
is necessary, the ME Equipment or ME System should be observed to
verify normal operation in the configuration in which it is used.
system may cause radio interference or may disrupt the operation of
nearby equipment. It may be necessary to take mitigation measures,
such as re-orienting or relocating the equipment or shielding the
location.
Caution:Changes or modifications not expressly approved by Hologic could
void your authority to operate the equipment.
8Part Number MAN-01964
Caution:This equipment has been tested and found to comply with the limits
DRAFT
for a Class A digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is
likely to cause harmful interference in which case the user will be
required to correct the interference at his own expense.
3.2Compliance Statements
The manufacturer states this device is made to meet the following requirements:
•CAN/CSA ISO 13485:2003
•CAN/CSA: Medical Electrical Equipment Part 1: C22.2 No. 601.1–M90 (R2005)—
General Requirements for Safety
•EN 60601-1:1990 +A1+A11+A12+A2+A13 Medical Electrical Equipment—General
Requirements for Basic Safety and Essential Performance
•ETSI EN 300 330-1 V1.7.1(2010-02)—Electromagnetic compatibility and Radio spectrum
Matters (ERM); Short Range Devices (SRD); Radio equipment in the frequency range 9
kHz to 25 MHz and inductive loop systems in the frequency range 9 kHz to 30 MHz
•ETSI EN 301 489-1: V1.8.1 (2008-04)—Electromagnetic compatibility and Radio
spectrum Matters (ERM); ElectroMagnetic Compatibility (EMC) standard for radio
equipment and services
2.6Standard Acquisition Workstation Control Display
Use the Mouse to select items.
2.7Preview Display
See the images on the Preview Display.
Part Number MAN-01964 13
User Manual
DRAFT
Chapter 2—System Controls and Indicators
Tubestand Controls and Indicators
3.0Tubestand Controls and Indicators
Legend for Figure 2-4
1. Rotation Angle Displays (each side)
2. C-Arm Controls (each side)
3. Compression Device
4. Patient Handles (each side)
5. Emergency Off Switches (each side)
6. Compression Handwheels
7. Patient Face Shield
8. Tubehead Display
9. Footswitches
Figure 2-4: Tubestand Controls and Indicators
14Part Number MAN-01964
3.1C-Arm Controls
AE
C
P
O
S
I
TIO
N
DRAFT
The C-Arm Controls provide the
Collimator and C-Arm functions.
See Section 6.0, page 19.
User Manual
Chapter 2—System Controls and Indicators
Tubestand Controls and Indicators
Figure 2-5: C-Arm Controls
3.2Compression Device Controls and Displays
Legend for Figure 2-6
1.Manual Compression Handwheels
2.Paddle Shift Buttons
3.AEC Sensor Buttons
4.Compression Device Display
5.The FAST Compression Mode Slide
6.Paddle Clamp
Figure 2-6: Compression Device
The Display on the compression
device shows:
•AEC Sensor Position
•Compression Force (displays 0.0
when force is less than 4 pounds)
•Compression Thickness
•Angle of C-Arm after rotation (for
5 seconds)
Figure 2-7: Compression Display
Part Number MAN-01964 15
User Manual
COMPRESSION
C-ARM
DRAFT
Chapter 2—System Controls and Indicators
Tubestand Controls and Indicators
3.3Tubehead Display
The Tubehead Display shows:
•SID
•Filter Type
•Collimator Setting
•Paddle Position
SID: 70 cm Rh
Coll: 24x29 C
3.4Dual Function Footswitches
Warning:Put both footswitches away from the patient and C-Arm area to
prevent any accidental footswitch use. When the patient has a
wheelchair, put the footswitches away from the area.
To use the footswitches:
1. Press the footswitch to actuate.
2. Release the switch to stop the
movement.
Legend for Figure 2-9
1.C-Arm Down
2.C-Arm Up
3.Compression Down
4.Compression Up
Figure 2-8: Tubehead Display
Figure 2-9: Dual Function Footswitches
16Part Number MAN-01964
Chapter 2—System Controls and Indicators
DRAFT
4.0How to Turn On the Selenia Dimensions
4.1Preparation
1. Reset all three Emergency Off switches.
2. Make sure that both system circuit breakers are in the On position.
3. Remove any obstructions to the C-Arm movement and to the view of the Operator.
4.2Startup
Figure 2-11: Standard Acquisition
Figure 2-10: Premium
Acquisition Workstation
Power Buttons
1. If the UPS was shut down, press the UPS power button (at the rear of the Premium
Acquisition Workstation or on the side of the Standard Acquisition Workstation).
2. Press the computer power button at the rear of the Acquisition Workstation.
Workstation Power Buttons
User Manual
How to Turn On the Selenia Dimensions
Legend for Figure 2-10
and Figure 2-11
1.Acquisition
Workstation Circuit
Breaker
2.Computer Power
Button
3.UPS Power Button
Figure 2-12: The Startup Screen
3. Select the Log In button.
Note…The Startup screen includes a Shutdown button that turns off the
system, and a Reboot button that restarts the system.
Part Number MAN-01964 17
User Manual
DRAFT
Chapter 2—System Controls and Indicators
How to Change the Language
Note…
4.3Log In
The system requires between five minutes and forty-five minutes to
prepare for image acquisition. The wait time depends on the detector
power configuration. A timer in the Taskbar displays the wait time before
the system is ready. Do not acquire clinical or QC images unless the
System Status Icon indicates the system is Ready.
Figure 2-13: How to Log In
When the user Log In screen displays, all Managers and Technologists show in the list of
Operators.
1. To display the Service, Applications, and Physicists user names, select the Show All
button.
2. Select your user name, enter your password, and select the Log In button.
Or
Validate your fingerprint.
5.0How to Change the Language
1. Select the Admin button.
2. Select the My Settings option.
3. From the Locale field, select a language from the drop-down menu.
4. Select the Save button, then select the OK button to the Update Successful message. The
selected language displays.
18Part Number MAN-01964
6.0Perform the Functional Tests
DRAFT
Legend for Figure 2-14
1.Compression Release
2.(Provisional use)
3.Light Field Lamp
4.(Provisional use)
5.Collimator Override
6.Clockwise C-Arm Rotation
7.C-Arm Up and Down
8.Counterclockwise C-Arm
Rotation
9.Compression Up
10. Compression Down
A C-Arm control panel is on
both the left and right sides
of the Gantry.
Perform the Functional Tests as part of your monthly visual checklists to make sure that the
control operates correctly.
User Manual
Chapter 2—System Controls and Indicators
Perform the Functional Tests
Figure 2-14: C-Arm Controls (left side shown)
Table 2-1: C-Arm Functional Tests
FunctionFunctional Test
Compression DownPress a Compression Down button:
•The compression brake engages.
•The light field lamp illuminates.
•The compression device lowers.
Note… When you press the Compression Down button, the
compression brake remains engaged until the Compression
Release button is pressed.
Compression down movement stops:
•When you release the button.
•When you reach the Down Force limit.
•When you reach the lower travel limit.
Compression UpPress a Compression Up button:
•The Compression Device moves toward the top.
•The Compression Up button does not release the
Compression Brake.
Compression Up movement automatically stops:
•When you release the button.
•When you reach the upper travel limit.
Part Number MAN-01964 19
User Manual
DRAFT
Chapter 2—System Controls and Indicators
Perform the Functional Tests
FunctionFunctional Test
Compression Release Press the Compression Release button:
C-Arm UpPress the C-Arm Up button:
Table 2-1: C-Arm Functional Tests
•The Compression Motor Brake releases.
•The Compression Device lifts.
•The C-Arm movement automatically stops when the
button is released.
•The C-Arm movement automatically stops when the
C-arm reaches the upper travel limit.
•The C-Arm movement is disabled when a compression
force of 45 N (10 pounds) or greater is applied.