Hologic PERILYNX, LABELWRITER450 TURBO User Manual

User Manual
IMPORTANT: Read the entire manual before operating the PeriLynx™ System
Hologic, Inc. 250 Campus Drive Marlborough, MA 01752 USA Tel: For Technical Support (USA and Canada) 1-888-PRETERM (1-888-773-8376) 1-800-442-9892 Fax: 1-508-263-2967 Tel: For Technical Support (Outside the USA and Canada)
Asia +852 3526 0718 Netherlands: 0800 0226782
Australia: +61 2 9888 8000 Norway: 800 15564
Austria: 0800 291919 Portugal: 800 841034
Belgium: 0800 77378 Spain: 900 994197
Denmark: 8088 1378 South Africa: 0800 980 731
Finland: 0800 114829 Sweden: 020 797943
France: 0800 913659 Switzerland: 0800 298921
Germany: 0800 1830227 UK: 0800 0323318
Ireland (Rep): 1 800 554144 Rest of the world: 00800.800.29892
Italy: 800 786308 Intl Fax number: 0041.21.633.39.10
©2016 Hologic, Inc. All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form or by any means in whole or in part without the prior written permission of Hologic, Inc. This product may be covered by one or more U.S. patents identified at http://hologic.com/patentinformation. Hologic, PeriLynx, Rapid fFN, and QCette are trademarks and/or registered trademarks of Hologic, Inc. and/ or its subsidiaries in the United States and/or other countries. All other trademarks, registered trademarks and product names are the property of their respective owners.
English AW-14693-001 Rev. 001 4-2016
i
Hologic Ltd. Heron House, Oaks Business Park, Crewe Road, Wythenshawe, Manchester, M23 9HZ, UK +44 (0)161 946 2206
TABLE OF CONTENTS
Section Page
Section 1 — Introduction 1-1
Intended Use 1-1
General Description 1-1
Components of the Analyzer 1-2
Touch screen 1-2 Cassette insertion site 1-2
Displayed/Printed Results 1-3
Specications 1-4
Cautions and Warnings 1-6
Disposal of Electrical & Electronic Equipment 1-10
Symbols Used on the Instrument 1-12
PeriLynx™ System User Manual Table of Contents
ii
TABLE OF CONTENTS
Section Page
Section 2 — Installation 2-1
General 2-1
Environmental Factors 2-1
Unpacking 2-2
System Setup 2-5
Getting Started 2-7
Setting the Date and Time 2-9
Factory Default Settings 2-10
Shutdown 2-11
Extended Shutdown 2-11
PeriLynx™ System User Manual Table of Contents
iii
TABLE OF CONTENTS
Section Page
Section 3 — General Operating/Testing Instructions 3-1
Starting the System 3-1
Overview of PeriLynx Analyzer Main Menu 3-2
Typical Menu Item Sequences 3-3
Using the Touch Screen 3-4
Run QCette QC — Quick Reference 3-8
Test Patient — Quick Reference 3-9
Enter New Calibration Code — Quick Reference 3-11
Run Liquid QC — Quick Reference 3-12
Section 4 — Software Functions — Detailed Descriptions 4-1
Startup Screen 4-1
Main Menu 4-2
Enter New Calibration Code 4-3
Test Patient 4-6
PeriLynx™ System User Manual Table of Contents
iv
TABLE OF CONTENTS
Section Page
Run QCette QC 4-15
Liquid Controls 4-20
Adjust Settings 4-28
Adjust Settings — Date/Time 4-28 Adjust Settings — Sound Settings 4-30 Adjust Settings — Auto Print 4-31 Adjust Settings — Password Settings 4-32 Adjust Settings — QCette Setup 4-35 Adjust Settings — Update Software 4-41
View Reports 4-42
Monthly Usage 4-42
Test Counts 4-44
Access Data 4-45
Access Data — View/Print Data 4-45 Access Data — Data Transfer 4-47
PeriLynx™ System User Manual Table of Contents
v
TABLE OF CONTENTS
Section Page
Section 5 — Care of the Analyzer 5-1
Section 6 — Printer 6-1
Loading Printer Labels 6-1
Removing an Empty Label Roll 6-4
Clearing Label Jams 6-5
Section 7 — Troubleshooting 7-1
General Information 7-1
Error/Invalid Codes 7-10
Section 8 — Service 8-1
Technical Support 8-1
Replacement Parts 8-3
PeriLynx™ System User Manual Table of Contents
vi

Section 1 — Introduction

For In Vitro Diagnostic Use Only
To be used by trained medical personnel

Intended Use

The Hologic PeriLynx™system is an in vitro diagnostic device intended to be used in conjunction with the RapidfFN®10Q cassette, the RapidfFN® control kit, and the PeriLynx™QCette® for the detection of fetal fibronectin (fFN) in cervicovaginal secretions. Refer to the directional insert for the Rapid fFN 10Q cassette for detailed intended use information.

General Description

The PeriLynx™ analyzer is an electronic optical reflectance device that converts a colorimetric reaction from a cassette into a digitized format. The data are analyzed using multiple parameters, including a comparison of sample data to calibration data. The analyzer reports the fFN concentration in the clinical specimen.
The analyzer reports fFN concentrations ranging from 0–500ng/mL. Concentrations greater than 500ng/mL are reported as >500ng/mL. The result is reported as invalid if specific internal test criteria have not been met.
PeriLynx™ System User Manual Section 1 — Introduction
1-1

Components of the Analyzer

The major components of the analyzer are the touch screen and the cassette insertion site.

Touch screen

Use the touch screen to enter data, select options, and move through the menus. See Section 3 — General Operating/Testing Instructions, for a full description of using the touch screen.

Cassette insertion site

The Cassette Insertion Site contains a slightly concave trough designed to capture any fluids that may have been spilled while applying sample to the cassette. This area of the instrument should be cleaned regularly (see Section 5 — Care of the Analyzer).
Note: For the detection of fetal fibronectin (fFN)
in cervicovaginal secretions, insert a RapidfFN10Q Cassette or PeriLynx QCette only.
PeriLynx™ System User Manual Section 1 — Introduction
1-2

Displayed/Printed Results

Each menu function result is displayed on the analyzer touch screen. With Auto Print ON, the result is automatically printed. Each printed result requires one printer label. Results can be printed from any data record screen either immediately after a test or in Access Data mode. The example below demonstrates the display and printout of a patient test.
Example: Displayed/Printed Results of Patient Test
Displayed Printed
PeriLynx™ System User Manual Section 1 — Introduction
1-3
Specifications
Power Supply UL +24 VDC listed power supply
Memory Capacity 50 Calibration Records
50 QCette Records
50 Control Records
1000 Patient Records
Touch screen 5-inch diagonal display with 480 x 800 resolution and
256 colors. Resistive interface.
Dimensions Length — 10 inches (25.4 cm)
Width — 7.5 inches (19.05 cm)
Height — 4 inches (10.16 cm)
Weight — 2.4 pounds (1090 g)
Operating Temperature 15° to 30°C
59° to 86°F
Operating Humidity 20% to 80%, non-condensing
PeriLynx™ System User Manual Section 1 — Introduction
1-4
Sound level, analyzer Maximum A-weighted sound pressure level at the typ-
ical user’s position and at a radius of 1 meter around the analyzer is 56 dB.
A.C. Supply 100-240VAC
50-60 Hz
2.1A Input Connector Coaxial power plug with positive center conductor Output Connectors 9-pin RS-232 connector
3 host USB connectors
Ethernet connector
Printer power connector
PeriLynx™ System User Manual Section 1 — Introduction
1-5

Cautions and Warnings

There are no known hazards associated with the PeriLynx system when it is operated in accordance with the instructions in this manual. However, you should be aware of situations that can result in serious injury.
WARNING! Ensure that the analyzer power adapter is connected to an AC electrical outlet that provides
voltage and current specified by Hologic. Use of an incompatible power receptacle can cause shock and fire hazard.
CAUTION! Use only the power adapter supplied by Hologic. Use of an incompatible power adapter can damage the internal components.
CAUTION! Always turn off the power and unplug the power adapter before cleaning the exterior of the analyzer. Fluid can damage internal components. DO NOT clean the power adapter.
CAUTION! Extreme heat can damage the display and other electronic components.
PeriLynx™ System User Manual Section 1 — Introduction
1-6
WARNING! Never apply cleaning reagents by spray as the liquid may leak into the analyzer causing
damage to the electrical components or possibly electrical shock to the user.
CAUTION! Do not immerse the analyzer in liquid. Fluid can damage internal components.
CAUTION! Do not clean the touch screen display with undiluted bleach solution or other solvents. Caustic
cleaning solutions can damage the touch screen.
CAUTION! Use appropriate laboratory procedures for handling biohazardous materials.
If this equipment is used in a manner not specified by the manufacturer, then the protection provided by the equipment may be impaired.
FCC Notice:
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
PeriLynx™ System User Manual Section 1 — Introduction
1-7
FCC Warning:
Changes or modification not expressly approved by the manufacturer responsible for compliance could void the user’s authority to operate the equipment.
Note: The use of a non-shielded interface cable with this equipment is prohibited.
CE Notice:
This equipment has been tested and found to be in compliance with the following standards per the IVD Directive:
EN61326-2-6 Electrical Equipment for Measurement, Control and Laboratory Use; EMC Requirements; In Vitro Diagnostic (IVD) Medical Equipment. EN61010-2-101 Electrical Equipment for Measurement, Control and Laboratory Use; Part 101; Particular
requirements for In Vitro Diagnostic (IVD) Medical Equipment. This equipment has been designed and tested to CISPR 11 Class A. In a domestic environment it may cause radio interference, in which case, you may need to take measures to mitigate the interference. The electromagnetic environment should be evaluated prior to operation of the device.
Do not use this device in close proximity to sources of strong electromagnetic radiation
(e.g. unshielded intentional RF sources), as these may interfere with the proper operation.
PeriLynx™ System User Manual Section 1 — Introduction
1-8
This equipment also complies with the following safety standards:
UL61010-1 Safety Requirements for Electrical Equipment for Laboratory Use Part 1: General
Requirements, with an ETL or equivalent Approval Mark
ICES-003 Industry Canada Regulation: Interference-Causing Equipment Standards - Digital
Apparatus
CAN/CSA C22.2 No. 61010-2-101
Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 1: General Requirements, with ETL or equivalent Approval Mark
PeriLynx™ System User Manual Section 1 — Introduction
1-9

Disposal of Electrical & Electronic Equipment

Waste Electrical and Electronic Equipment (WEEE)
Hologic is dedicated to meeting country specific requirements associated with the environmentally sound treatment of our products. Our objective is to reduce the waste arising from our electrical and electronic equipment. Hologic realizes the benefits of subjecting such WEEE equipment to potential reuse, treatment, recycling or recovery to minimize the amount of hazardous substances entering the environment.
Your responsibility
As a Hologic customer, you are responsible for ensuring that devices marked with the symbol shown below are not placed into the municipal waste system unless authorized to do so by the authorities in your area. Please contact Hologic (see below) prior to disposing any electrical equipment provided by Hologic.
Symbol used on the instrument
The following symbol is used on this instrument:
Do not dispose in municipal waste.
Contact Hologic (see below) for information regarding proper disposal.
Reclamation
Hologic will provide for the collection and proper reclamation of electrical devices we provide to our customers. Hologic strives to reuse Hologic devices, subassemblies, and components whenever possible. When reuse is not appropriate, Hologic will ensure the waste material is properly disposed of.
PeriLynx™ System User Manual Section 1 — Introduction
1-10
Hologic Contact Information
Corporate Headquarters Hologic, Inc.
250 Campus Drive Marlborough, MA 01752 USA Tel: (USA and Canada) 1-888-PRETERM (1-888-773-8376) 1-800-442-9892 1-508-263-2900 Fax: 1-508-263-2967
Authorized Representative in the European Community
PeriLynx™ System User Manual Section 1 — Introduction
Hologic Ltd. Heron House, Oaks Business Park Crewe Road, Wythenshawe Manchester, M23 9HZ, UK Tel: +44 (0)161 946 2206
1-11

Symbols Used on the Instrument

30˚C
15˚
The following symbols may appear on this instrument or its packaging:
Caution, consult instructions for use
Waste Electrical and Electronic Equipment - contact Hologic for disposal of the instrument.
Catalogue Number
Serial Number Biological risks
For in vitro diagnostic testing Direct current (DC)
PeriLynx™ System User Manual Section 1 — Introduction
Manufacturer
Authorized Representative in the European Community
Store between 15°C and
C
30°C
1-12

Section 2 — Installation

General

This section provides detailed installation instructions for the PeriLynxsystem. Follow installation steps carefully to insure proper installation and operation.

Environmental Factors

The PeriLynxsystem has been designed to be safe under the following conditions:
Indoor use;
Altitudes up to 2000 m;
Maximum relative humidity of 80% for temperatures up to 30°C;
Pollution Degree II, in accordance with IEC 61010-1 Category II. The PeriLynx system is for use only in an
ofce or a clean laboratory environment.
The operating temperature should be held relatively constant. The optimum operating temperature is 15° to 30°C (59° to 86°F). Before operating, allow the instrument to equilibrate to room temperature.
Place the instrument away from direct sunlight and away from locations subject to extreme temperature variations (e.g., near open windows, ovens, hot plates, radiators, etc.).
PeriLynx™ System User Manual Section 2 — Installation
2-1

Unpacking

PeriLynxanalyzer
Carefully remove the analyzer and accessories from the shipping carton.
Inspect the carton and the analyzer for visible signs of damage. If the analyzer is damaged, immediately contact the carrier and Hologic Customer Service.
The carton should contain the following parts/accessories:
PeriLynx analyzer
AC adapter and power cord (one or more for international compatibility)
User manual
PeriLynx QCette
®
Note: Retain the shipping carton for future use. If the analyzer needs to be
shipped, use the original shipping carton.
PeriLynx™ System User Manual Section 2 — Installation
2-2
Printer
Carefully remove the printer and accessories from the shipping carton. Inspect the carton and the printer for visible signs of damage. If the printer is damaged, immediately contact the carrier and Hologic Customer Service.
The printer carton should contain the following parts/accessories:
Printer
Printer labels (1 roll)
Power cord
USB cable
Note: Retain the shipping carton for future use. If the
printer needs to be shipped, use the original shipping carton.
PeriLynx™ System User Manual Section 2 — Installation
2-3
Barcode scanner (optional)
A barcode scanner is available from Hologic for entering data into the analyzer. This barcode scanner is the only barcode scanner for use with the PeriLynx system. If your order included the optional barcode scanner, carefully remove the barcode scanner and accessories from the shipping carton. Inspect the carton and the barcode scanner for visible signs of damage. If the barcode scanner is damaged, immediately contact the carrier and Hologic Customer Service.
Follow all of the instructions provided by the barcode scanner manufacturer regarding the safe use of the barcode scanner.
The barcode scanner carton should contain the following parts/accessories:
Barcode scanner
Stand
PeriLynx™ System User Manual Section 2 — Installation
2-4

System Setup

1. The analyzer and printer should be placed on a flat, level surface. Verify that the power switch on the back of the analyzer is set to the off (0) position.
Note: Do not install any of the Dymo Label software that comes with the printer.
2. Plug the power cable jack into the receptacle labeled 24V DC on the bottom of the printer. Plug the other end of the cable into the PeriLynxanalyzer.
3. Insert the USB cable into the receptacle labeled USB on the bottom of the printer. Route the USB cable and the power cable through the cable channel and set the printer upright on a flat surface. Plug the other end of the USB cable into any of the USB ports on the analyzer.
System
(Photo includes optional barcode scanner)
PeriLynx™ System User Manual Section 2 — Installation
Printer
Analyzer
2-5
4. Use the AC cord corresponding to the room power outlets. Plug the output jack from the power adapter into the rear of the analyzer. Plug the wall mount adapter into an AC power outlet.
Caution: Only the power adapter provided with the PeriLynxanalyzer may be used. Any substitutions can
result in damage to the PeriLynxanalyzer and printer.
5. To connect the optional barcode scanner, plug the USB cable of the scanner into any of the USB ports on the analyzer.
PeriLynx™ System User Manual Section 2 — Installation
2-6

Getting Started

Turn the analyzer power switch to the on ( l ) position. The power switch is located on the back of the instrument. (If the analyzer does not turn on, see Section 7 — Troubleshooting, Item 1.)
Upon power up, the analyzer displays the Hologic logo and then PeriLynxsystem logo with the message “Initializing system...” for 60 seconds. Then, the analyzer performs a self-test of the analyzer components.
If there is a problem after the self-test, a beep will sound to indicate an error and an error message will be displayed. If an error message is displayed, refer to the troubleshooting section of the manual.
If the printer is not connected, an error message will be displayed.
PeriLynx™ System User Manual Section 2 — Installation
2-7
Once the self-test is complete, the display will change to the Main Menu. The date and time may need to be reset for your time zone.
PeriLynx™ System User Manual Section 2 — Installation
2-8

Setting the Date and Time

1. From the Main Menu, select Adjust Settings.
2. Touch Date/Time Settings and follow the prompts.
For more details about setting the date and time, see Section 4 — Software Functions — Detailed Descriptions.
PeriLynx™ System User Manual Section 2 — Installation
2-9

Factory Default Settings

The PeriLynxsystem uses the following default settings. To customize the unit to your laboratory requirements, refer to Section 4 — Software Functions — Detailed Descriptions.
The default settings are as follows:
Auto Print
Factory setting is
Auto Print
ON. After every result, the printer will generate a printed result.
Date format
The default setting is MM/DD/YY.
Time format
The default setting is 24 hour time.
Sound settings
The default setting is for the audible tone (beep) to be turned on at a volume of 5.
Password settings
The default setting has the password disabled.
PeriLynx™ System User Manual Section 2 — Installation
2-10

Shutdown

The system may be left plugged in and idle when not in use. If the system is to be shut down make sure it is idle. Move the power switch on the back of the analyzer to the off position.

Extended Shutdown

If the system is to be turned off for a long period of time, or taken out of use, power off the analyzer as described above. Completely remove power from the device by unplugging the power cord from the electrical outlet.
PeriLynx™ System User Manual Section 2 — Installation
2-11
Loading...
+ 83 hidden pages