SERVICE MANUAL
FOR
MODEL 902 AUTOMATIC
ANALYZER
Copyright Hitachi, Ltd. 1997. All rights reserved. Printed in Japan. Part No. 713-9039 KN-K (H-LT)
MODEL 902 AUTOMATIC ANALYZER
CONTENTS
1. PRODUCT SPECIFICATIONS................................................................................. 1-1
2. TROUBLESHOOTING ............................................................................................. 2-1
3. FLOW PATH DIAGRAMS ........................................................................................ 3-1
4. BLOCK DIAGRAMS................................................................................................. 4-1
5. ISE ........................................................................................................................... 5-1
6. SWITCH SETTING................................................................................................... 6-1
7. EXPLANATION OF FUNCTION............................................................................... 7-1
8. MOTORS, DETECTORS AND FUSES .................................................................... 8-1
9. ANALYTICAL METHODS......................................................................................... 9-1
10. CIRCUIT DIAGRAMS............................................................................................... 10-1
11. CROSS WIRING REFERENCE ............................................................................... 11-1
12. INSTALLATION........................................................................................................ 12-1
13. MAINTENANCE AND INSPECTION ........................................................................ 13-1
14. ADJUSTMENT SPECIFICATIONS........................................................................... 14-1
15. OEM......................................................................................................................... 15-1
16. SYSTEM INTERFACE.............................................................................................. 16-1
- i -
1.PRODUCT SPECIFICATIONS
1.1 Overall System.........................................................................................................1-1
1.2 Sampling System.....................................................................................................1-2
1.3 Reagent System.......................................................................................................1-3
1.4 Reaction System......................................................................................................1-4
1.5 Photometic System..................................................................................................1-5
1.6 ISE (option)..............................................................................................................1-5
1.7 Installation Conditions..............................................................................................1-6
1.8 User Interface...........................................................................................................1-6
1.9 Dionized Water........................................................................................................1-6
1.10 Analytical Methods...................................................................................................1-7
1.11 Accuracy Methods....................................................................................................1-7
1.12 Data Storage............................................................................................................1-7
1.13 System Interface......................................................................................................1-8
1.14Options for Analyzer.................................................................................................1-8
1 - 0
1.PRODUCT SPECIFICATIONS
1.1 Overall System
Configuration : Console type automatic analyzer, with analytical and operation units
integrated
Principle : Discrete type, random access, single-line multi-analysis system;
entire reaction monitoring system
Analytical methods : Colorimetry (absorbance measurement), ion selective electrode
method (ISE; option)
Throughput : Colorimetry ; 200 tests/hr (Photometry only max. 300 tests/hr
inclusive of ISE)
No. of analysis items: Colorimetry ; 36
ISE ; 3 (Na, K, Cl)
Serum indexes ; 3
Calculation items; 8
Reaction time : 10 minutes max.; 3, 4, 5, 10 minutes for concentration calculation
(same as Model 7070/7170)
Measurable samples : Serum (or blood plasma) and urine; one sample kind per channel
Application : Emergency (stat) and routine analyses
Test item selection : Via entry from operation panel, entry from system interface, entry
from no. of items settable (11 kinds)
1 - 1
1.2 Sampling System
Sample container : Hitachi standard sample cup
Sample tube ;diameter 13 to 17 mm
length 75 to 100 mm
Sample disk : 60 positions
Outer row;35 positions (routine and stat samples)
(with free adapter, barcode reader attachable)
Inner row;25 positions (standard + control samples (22),
rinse solutions (3))
Detergents : 3 kinds
(W1;for sample probe rinsing) HITERGENT
(W2;for sample carryover evasion) ISE RINSE SOLUTION
(W3;for sample carryover evasion) HICARRYNON
Sampling mechanism : Pipetter driven by stepping motor
2 to 50 µ L of sample/test (in 0.1 µ L steps)
Pipetting mode : Normal ; prescribed volume discharged at cell
bottom
Aqueous sample; system water solution for sample probe
internal rinsing usable as STD-1
Sample pre-dilution not performed
Insufficient sample detection: Resistance detection method
Sample ID : Barcode reader (option)
Following code shemes are usable in combination; CODE 39,
CODE 128, INTERLEAVED 2 OF 5, NW7 (Modulas 10,
Modulas 16)
Automatic rerun : Auto rerun function not provided
Stat sample analysis : Routine sample analysis interruptible for stat sample analysis
Sample carryover : Sample carryover evasion function available; 2 kinds of
detergents settable on sample disk inner row
Sample conveyance : Can be sampled directly from belt-line
1 - 2
1.3 Reagent System
Reagent disk : One disk, 40 positions
(20 positions each on inner and outer rows)
Reagent cooling : Cooling water circulating system (3 to 15° C) (option for China-
destined instrument)
Reagent bottle : 50 mL, 20 mL (adapter necessary)
(without barcode, concentrated reagent unusable)
Detergents : HITERGENT (50 mL); position 40 (fixed)
(for addition to reaction bath)
Detergent 1; for reagent probe, stirring rod rinsing, reagent
probe carryover evasion, reaction cell carryover
evasion
Detergent 2; for reagent probe carryover evasion, reaction cell
carryover evasion
(settable on reagent disk inner row)
ISE reagent : Internal standard solution(position 37 (fixed))
Diluent (position 38 (fixed))
Pipetting mechanism : Pipetter driven by stepping motor; 20 to 350 µ L/test (in 1 µ L
steps)
Reagent addition timing: 3 steps (0, 1.5 and 5 minutes); throughput is maintained even
when adding 3 reagents.
Carryover evasion : Rinsing for carryover evasion possible; throughput may decrease
by specifying carryover evasion.
1 - 3
1.4 Reaction System
Reaction disk : Turntable type disk (10-minute reaction per rotation)
Reaction cell : Optical path length 6 mm (area 6 × 5 mm)
No. of reaction cells : 48
Reaction cell control : One rotation + one pitch feed (18 seconds)
Sample pipetting position: One position (fixed)
Reagent pipetting position: One position (fixed)
Reaction temperature : 37 ± 0.1° C (warm water circulating system)
Stirring : By means of stirring rod rotation
Timing for stirring; after each addition of R1,R2,R3
Stirring possible for a minimum volume of 100 µL
Stirring position : One position (fixed)
Reacting solution volume: Minimum; 250 µ L (minimum required volume for photometry)
Maximum; 500 µ L (temperature control, rinsing upper limit)
Photometric position : One position (fixed)
Cell rinsing : After completion of photometry (rinsing 3 times; rinsing with
detergent not performed)
Carryover evasion : Rinsing for carryover evasion possible; throughput may decrease
by specifying cell carryover evasion
1 -4
1.5 Photometric System
Photometer : Multiwavelength photometer (absorptiometry)
Wavelengths : 12 wavelengths
(340, 376, 415, 450, 480, 505, 546, 570, 600, 660, 700, 800 nm)
Monochromator : Grating
Detector : Photodiode
Linearity : Up to 2.5 Abs (10 mm cell conversion)
Photometric method: Direct photometry of reaction cell (at one or two wavelengths)
Correction : Cell blank correction prior to analysis
(passed cell blank measurement alone)
1.6 ISE (option)
Electrodes : Flow cell type, liquid-membrane ISE cartridge
Reference electrode flow path: 1 MKCL, liquid flow path
Measuring temperature : 36° C ± 2° C (warm water circulating system)
System : Indirect (dilution) potentiometry, 50-times dilution
Measuring cycle : 36 sec/sample (18 sec for sample, 18 sec for internal
standard solution)
Measured items : Na, K, Cl
Measurable samples : Serum and urine
Linearity : Na;10 to 250 mmol/L
K ;1 to 100 mmol/L
Cl;10 to 250 mmol/L
Reagent bottle : Internal standard solution ; 50 mL max.
Diluent ; 50 mL max.
Reference electrode solution; 500 mL max.
1 - 5
1.7 Installation Conditions
Power requirement : 230 V, 50/60 Hz, less than 1.5 kVA
Deionized water consumption: Less than 15 L/hr
Waste liquid drain : 2 systems (for concentrated and diluted liquids)
Ambient temperature/humidity: Temperature; 18 to 30° C
Humidity ; 20 to 80% (non-condensing)
Analyzer dimensions : 720 W × 720 D × 1085 H mm
Analyzer weight : Within 200 kg
BTU : 1300 kcal/hr max.
Noise (mean) in operation : < 55 dB
1.8 User Interface
Application : Routine analysis for Asia version
Display : Backlighted LCD; 256 × 128 dots, graphic
Keyboard : Touch screen keys (72 keys)
Printer : Thermal roll-paper printer (20 digits)
Multi-language
compatibility : Display; Japanese/English/Chinese/German/Spanish applicable
Printer; English alone
1.9 Deionised Water
Pressure : 0.5 to 3.5 kgf/cm
2
Conductivity: 1 µ s/cm or less, germ-free
1 - 6
1.10Analytical Methods
Assay modes : One-point
One-point end (+ prozone check)
Two-point rate
Two-point end (+ prozone check)
Three-point two-item
One-point rate two-item
Rate A (+ sample blank correction)
Rate A (+ serum indexes)
Rate B two-item (same wavelength)
Rate B two-item (different wavelengths)
ISE
Data alarms : Based on Model 7070/7170
Standard solution : 22 kinds max. (positions to be shared with control sera)
Calibration types : Linear (2-point linear)
K factor
4 parameter LOGIT-LOG
5 parameter LOGIT-LOG
Spline
Segmented line
Calibration method : At startup only; all points (FULL) and reagent-blank-corrected
calibration
Calculation channels : For 8 channels
Test-to-test compensation : For 8 channels
1.11Accuracy Control
Control serum : 5 kinds max. (positions shared with standard solutions)
1.12Data Storage
Routine sample data : 400 samples (in data disk)
Stat sample data : 50 samples (in data disk)
Control sample data : 5 kinds × 30 (in SRAM)
1 - 7
1.13System Interface
Interface : RS-232C and current loop
Communication protocol: Based on Model 7070/7170
Communication details : Communication with host; communication details based on Model
7070/7170
1.14Options for Analyzer
ISE
Sample ID accessory : Model ; BL180
Maker ; Keyence
Barcode spec.; CODE 39, ITF, NW7 (Modulas 10, Modulas 16),
CODE 128
1 - 8
2. TROUBLESHOOTING
2.1 Alarm Code Table ....................................................................................................2-1
2.1.1 LCD Display Alarm .....................................................................................2-24
2.2 Motor Control Alarms................................................................................................2-26
2.2.1 Operation Check Procedure at Occurrence of Alarm .................................2-27
2.3 Parameter Check .....................................................................................................2-29
2.3.1 Processing Flow .........................................................................................2-29
2.3.2 Details of Parameter Check........................................................................2-30
2.3.3 Details of Twin Test Simultaneous Analysis ...............................................2-34
2.4 Data Alarm...............................................................................................................2-35
2.4.1 Data Alarm Registratin Flow.......................................................................2-35
2.4.2 Data Alarm Code List..................................................................................2-36
2.4.3 Data Alarm Codes ......................................................................................2-37
2.4.4 ISE Data Alarms.........................................................................................2-49
2.4.5 Alarm Check Method..................................................................................2-52
2.4.6 Check and Set Alarm of Each Data............................................................2-60
2.4.7 Details of Data and Alarm Outputs Resulting from Calibration...................2-61
2.4.8 Output Check List for Each Photometry Assay CALIB. METHOD .............2-62
2.5 Retry Code Table.....................................................................................................2-64
2.5.1 Logging Program List......................................................................................2-65
2.6 Daily Alarm Trace.....................................................................................................2-67
2.6.1 Cumulative Alarm Trace.............................................................................2-69
2.6.2 Parameter Code List...................................................................................2-72
2.6.3 Communication Trace.................................................................................2-74
2.6.4 Cumulative Instrument Operation List.........................................................2-78
2.6.5 FD File Management..................................................................................2-79
2 - 0
2. TROUBLESHOOTING
2.1 Alarm Code Table
Category Alarm Name Category Alarm Name
1 STIRRER 38 VAC. TANK
2 39
3 RINSE 40
4 41 LAMP
5 R.DISK 42
6 S.PROBE 43
7 44
8 S.DISK 45 CELL BNK1
9 46 CELL BNK2
10 47 ADC1 ?
11 S. SHORT 48 ADC2 ?
12 49 ADC3 ?
13 50 BARCODE 1
14 51
15 S. SYRINGE 52
16 REAG. PROBE 53
17 54 BARCODE 2
18 REAG. DISK 55
19 56
20 57 REAG. SHORT
21 58
22 R. SYRINGE 59 REAG. LEVEL
23 60
24 ISE SIPPER 61
25 62
26 ISE SYRING. 63
27 ISE STOP ? 64
28 TEMP CONT. 65
29 INC. WATER 66
30 REF. WATER 67 TS OVER
31 DIST. WATER 68 PATNT OVER
32 DIST. SENS. 69 SAMP. END
33 70
34 RESERVOIR 71 DC POWER
35 72 FUSE
36 73 POWER FAIL
37 SIPPER 74
2 - 1
(cont’d)
Category Alarm Name Category Alarm Name
75 MOTOR CONT. 114 ACI ERROR
76 MOTOR TOUT 115
77 116
78 117
79 118
80 119 FD WRITE ?
81 120 FD READ ?
82 121 NO FD
83 STANDARD ? 122
84 CALIB. 123 FD PROTECT
85 CALIB. SD ? 124
86 SENS. ? 125 PRINTER
87 126 SYSTEM I/F
88 127
89 128
90 ISE LEVEL 129
91 ISE NOISE 130 WATER EXG.
92 ISE PREP. 131
93 ISE SLOPE 132
94 ISE I. STD 133
95 REF. SHORT 134
96 TWIN TEST ? 135
97 136
98 CHEM. PARAM ? 137
99 CLB. PARAM ? 138
100 VOLUME ? 139
101 140 PANEL I/F
102 141 REAGNT ?
103 CMP. TEST ? 142
104 S. INDEXES ? 143
105 144
106 ON BOARD ? 145 CELL C. O.
107 REAG. POS ? 146 ISE C. O.
108 CLB (IS) POS 147
109 148
110 149
111 150
112 151
113 152
2 - 2
Alarm
Control
No.
1 to 8 STIRRER 1 1 STOP In ascending action of the
21,22 RINSE 3 1 STOP The rinsing mechanism does
Alarm Category
1 2 STOP In ascending action of the
1 3 STOP In descending action of the
1 4 STOP When the stirrer moves
1 5 STOP The stirrer does not come to
1 6 STOP At resetting, the stirrer does
1 7 STOP At resetting, the stirrer does
1 8 STOP In rotation of the stirrer, it is
3 2 STOP The rinsing mechanism does
Subcode
Level Description Remedy
stirrer, it does not reach the
upper dead point (on the
rinsing bath side).
(Alarm at the first upper
dead point after resetting will
be issued from other than
the cell side.)
stirrer, it does not reach the
upper dead point (on the cell
side).
stirrer, it does not leave the
upper dead point.
toward the rinsing bath, it
does not reach the rinsing
bath position.
the cell position.
not reach the rinsing bath
position (home position) in
its return movement to the
home position.
not leave the rinsing bath
position (home position) in
its departing movement
from the home position.
not set at the upper dead
point.
not reach the upper dead
point in ascending motion.
not leave the upper dead
point in descending motion.
Check the upper dead
point detector.
Same as above
Same as above
Check the home
detector.
Check the detector on
the cell side.
Check the home
detector.
Same as above
Check the upper dead
point detector.
Check the upper dead
point detector.
Same as above
2 - 3
Alarm
Control
No.
Alarm Category
Subcode
Level Description Remedy
61 to 65 R. DISK 5 1 STOP The reaction disk cannot
recognize its stop position.
5 2 STOP The reaction disk does not
stop at the specified
position.
5 3 STOP At resetting, the reaction
disk cannot recognize its
home position.
5 4 STOP At resetting, the first cell
on reaction disk does not
stop at the specified
position.
5 5 STOP When the reaction disk
turns, the serum probe,
reagent probe, stirrer or
rinsing mechanism is not
set at the upper dead point
(on the cell side).
(When this alarm is
issued, another alarm (on
rinsing or stirring
mechanism) may concur.)
71 SAMPLE
PROBE
6 1 S.STOP/
STOP
(Note 1)
The serum probe does not
reach the upper dead point
in ascending motion (on
other than the cell side).
(Alarm at the first upper
dead point after resetting
will be issued from other
than the cell side.)
72 to 85 SAMPLE
PROBE
6 2 STOP The serum probe does not
reach the upper dead point
in ascending motion (on
the cell side).
6 3 S.STOP/
STOP
(Note 1)
The serum probe moves
down abnormally in
descending action (on
other than the cell side).
(±3 mm from cup bottom,
0 to 4 mm from cell
bottom)
6 4 STOP The serum probe moves
down abnormally in
descending action (on the
cell side).
6 5 S.STOP/
STOP
(Note 1)
The serum probe does not
go down from the upper
dead point in descending
motion.
NOTE: 1. S.STOP may be issued only during operation.
(cont’d)
Check the detectors for
stop positions on the
inner and outer tracks.
Same as above
Check the home
detector.
Check the home
detector or inner/outer
track detector.
(1) Perform resetting.
(2) Check the upper
dead point detector
of the mechanism
which caused
alarm.
Check the upper dead
point detector.
Same as above
(1) Check the liquid
level detector.
(2) Replace the liquid
level detector PC
board.
Same as above
Check the upper dead
point detector.
2 - 4
Alarm
Control
Alarm Category
No.
72 to 85 SAMPLE
6 6 STOP The serum probe does not
Sub-
code
Level Description Remedy
PROBE
6 7 S.STOP/
STOP
(Note 1)
6 8 S.STOP/
STOP
(Note 1)
6 9 S.STOP/
STOP
(Note 2)
6 11 S.STOP/
STOP
(Note 1)
6 12 WARNINGThe serum probe moves
6 13 S.STOP/
STOP
(Note 2)
6 14 STOP In rotation of the serum
6 15 STOP In rotation of the serum
101 to
106
SAMPLE
DISK
8 1 S.STOP/
STOP
(Note 2)
8 2 S.STOP/
STOP
(Note 2)
8 3 S.STOP/
STOP
(Note 2)
8 4 S.STOP/
STOP
(Note 2)
NOTE: 2. S.STOP may be issued only during operation.
go down from the upper
dead point in descending
motion (on the cell side).
Detection of abnormal
descending motion of the
serum probe remains on.
When serum probe turns to
the cell side, the cell position
cannot be detected.
When the serum probe turns
from the cell side to other
position, it does not come off
the cell position.
Before the probe goes down,
the liquid level detector is
already turned on.
down abnormally in
descending action (only at
turning of the serum probe in
adjustment).
In rotation of the serum
probe, it is not set at the
upper dead point.
probe from the reset
position, it does not come off
the reset position.
probe to the reset position, it
cannot detect the reset
position.
The sample disk cannot
detect the stop position on
outer track.
The sample disk does not
stop at the specified position
on outer track.
The sample disk cannot
detect the stop position on
inner track.
The sample disk does not
stop at the specified position
on inner track.
(cont’d)
Check the upper dead
point detector.
Refer to alarm code
6-3.
Check the home
detector.
Same as above
Refer to alarm code
6-3.
Same as above
Check the upper dead
point detector.
Check the home
detector.
Same as above
Check the outer track
detector.
Same as above
Check the inner track
detector.
Same as above
2 - 5
(cont’d)
Alarm
Control
Alarm Category
No.
101 to
106
SAMPLE
DISK
107 SAMPLE
DISK
251 to
310
551 to
552
561 to
568
SAMPLE
SHORT
SAMPLE
SYRINGE
REAGENT
PROBE
Subcode
Level Description Remedy
8 5 STOP At resetting, the sample disk
cannot detect the home
position.
8 6 STOP At resetting, the sample disk
does not stop at the specified
point of home position.
8 7 STOP At resetting, the sample disk
does not leave the home
position.
11 1 to 60WARNING In sipping from the sample
cup, sample on the sample
disk is inadequate.
Sub-code groups
Routine/
stat samples:1 to 35
Control STD:36 to 57
W1 to W3 :58 to 60
15 1 S.STOP/
STOP
15 2 S.STOP/
STOP
The serum syringe does not
reach the upper dead point.
The serum syringe does not
go down from the upper dead
point.
16 1 STOP The reagent probe does not
reach the upper dead point in
ascending motion.
16 2 STOP The reagent probe moves
down abnormally in
descending motion.
16 3 STOP The reagent probe does not
go down from the upper dead
point in descending motion.
16 4 WARNINGDetection of abnormal
descending motion of the
reagent probe remains on.
Check the home
detector.
Same as above
Check the home
detector.
Add sample.
Check the upper
dead point detector.
Same as above
Check the upper
dead point detector.
Check liquid level
detection.
Check the upper
dead point detector.
Check the
descending error
detector.
2 - 6
Alarm
Control
No.
561 to
568
Alarm Category
REAGENT
PROBE
16 5 STOP When the reagent probe
Subcode
Level Description Remedy
turns toward the cell, it
cannot detect the cell
position.
16 6 STOP When the reagent probe
turns from the cell side to
other position, it does not
leave the cell position.
16 7 STOP Before the probe goes down,
the liquid level detector is
already turned on.
16 8 STOP In rotation of the reagent
probe, it is not set at the
upper dead point.
581 to
593
REAGENT
DISK
18 1 STOP The stop position of reagent
disk cannot be detected.
18 2 STOP The reagent disk does not
stop at the specified position.
18 3 STOP The home position of reagent
disk cannot be detected.
621 to
622
REAGENT
SYRINGE
22 1 STOP The reagent syringe does not
reach the upper dead point.
22 2 STOP The reagent syringe does not
move down from the upper
dead point.
641 ISE
SIPPER
24 1 STOP The sipper nozzle does not
reach the upper dead point
(during resetting/
operation).
642 ISE
SIPPER
661 to
662
ISE
SYRNG
671 ISE STOP
OK ?
24 2 WARNING/
STOP
(Note 4)
26 1 WARNING/
STOP
(Note 4)
26 2 WARNING/
STOP
(Note 4)
27 1 WARNING/
STOP
(Note 4)
The sipper nozzle does not
leave the upper dead point.
The sipper syringe does not
reach the upper dead point.
The sipper syringe does not
leave the upper dead point.
The ISE function is stopped
due to alarm.
(This warning is indicated
when restart in the sampling
stop status was attempted.)
NOTES: 3. S.STOP may be issued only during operation.
4. Though the photometry assay function works, the ISE function does not work.
(cont’d)
Check the home
detector.
Same as above
Check liquid level
detection.
Check the upper
dead point detector.
Check home
detection or count
detector.
Same as above
Check home
detection.
Check the upper
dead point detector.
Same as above
Check the upper
dead point detector of
ISE sipper.
Same as above
Check the upper
dead point detector of
ISE Syringe.
Check alarm log and
deal with the alarm.
2 - 7
(cont’d)
Alarm
Control
Alarm Category
No.
681 TEMP
CONTROL
28 1 WARNING The water temperature of
Subcode
Level Description Remedy
incubation bath is higher
than 45.0°C.
28 2 WARNING The water temperature of
incubation bath is outside a
range of 37 ± 0.5°C.
(This is checked only at
operation.)
691 INCUBA-
TOR
29 1 WARNING The water level of
incubation bath is too low.
WATER
701 REF
WATER
30 1 WARNING A period of 24 hours has
passed since exchange of
incubation bath water.
711 DISTILLED
WATER
31 1 STOP The water level of distilled
water tank is too low.
(This alarm will not be
issued during initialization
and water exchange.)
712 DIST
WATER
721 DIST
WATER
SENSOR
724 RESER-
VOIR
31 2 WARNING The water level of distilled
water tank is too low.
32 1 WARNING The water level sensor in
distilled water tank is
abnormal.
34 1 WARNING The waste solution reservoir
is full.
771 SIPPER 37 1 STOP The negative pressure of
vacuum pump is too low.
781 VACUUM
TANK
811 to
813
PHOTOMETER LAMP
38 1 WARNING Water is accumulated in the
vacuum tank.
41 1 WARNING In passed cell blank
measurement, a value of
more than 3.3 Abs is
indicated in any one of 4
measurements.
(3 times or less)
41 2 S.STOP
(Restart
unallowable
41 3 S.STOP
(Restart
un-
In passed cell blank
measurement, a value of
more than 3.3 Abs is
indicated for all of ADC1,
)
ADC2 and λ
to λ 12.
1
The above alarm code 41-1
is issued 10 times
consecutively.
allowable)
Check the thermistor
or thermostat of
heater.
Same as above
Replenish water or
check the drain
solenoid valve.
Exchange incubation
bath water.
Check the water level
sensor.
Same as above
Check float switch.
Check waste solution
tank detection.
Check vacuum level
at vacuum suction.
Check SV12 or SV14.
(1)Replace light
source lamp.
(2)Check 12 V
power supply.
2 - 8
(cont’d)
Alarm
Control
Alarm Category
No.
1231 CELL
BLANK
1241 to
1400
1441 to
PASS CELL
BLANK
ADC1? 47 1 WARNING Any one of I/O error,
1444
Subcode
45 1 S.STOP
Level Description Remedy
In passed cell blank
(Restart
unallowable)
measurement, any one of
ADC1, ADC2, λ x and λy
differs from the reference
value (Note 5) by more than
0.1 Abs through 10
consecutive cycles.
(Counting is not made in
use for stopped cell blank
test.)
46 1 to 48WARNING In 4 passed cell blank
measurements, any one of
ADC1, ADC2, λ x and λy
differs from the reference
value (Note 5) by more than
0.1 Abs twice or more.
(In case any one of those
values differs only once, the
average value of normal
data is treated as a value of
passed cell blank.)
boundary error of parameter
block pointer, I/O device
busy, channel error and I/O
device error (PC board not
mounted, device
inoperable, time-out) has
occurred.
47 2 WARNING • A/D count value remains
at 0.
• After A/D conversion,
interruption is impossible.
• The command or
parameter given to A/D
PC board is abnormal.
• A/D conversion cannot be
completed.
(Time-out occurs.)
(Reference voltage is
checked.)
47 3 WARNING The number of A/D starts
cannot be reset to 0.
(This is checked at reaction
measurement.)
(1)Carry out cell
washing.
(2)Replace the cell.
Same as above
Replace ECPU230 or
EMIO100 PC board.
(1)Replace Log Amp
PC board.
(2)Check 2 V and
6 V reference
voltages.
Same as above
2 - 9
(cont’d)
Alarm
Control
No.
1441 to
1444
1451 to
1453
1461 to
1462
1471 to
4506
Alarm Category
ADC1? 47 4 WARNING • A/D count for 2 V is
Subcode
Level Description Remedy
Same as above
abnormal.
•A/D count for 6 V is
abnormal.
Normal count for 2 V
(7547 < count value <
8341)
Normal count for 6 V
(22460 < count value <
25023)
(This is checked at
measurement of
reference voltage.)
ADC2? 48 1 WARNING Refer to alarm code 47-1. Replace ISE AMP PC
board.
48 2 WARNING Refer to alarm code 47-2. Same as above
48 3 WARNING Refer to alarm code 47-4.
Reference voltage for
ISE: 2 V and 8 V
Normal count for 2 V
(7547 < count value <
8341)
Normal count for 8 V
(30184 < count value <
33364)
Normal offset value
(22811 < count value <
40547)
ADC3? 49 1 WARNING Refer to alarm code 47-2. Replace EMIO100
49 2 WARNING • A/D count for 2 V is
abnormal.
Same as above
PC board.
Same as above
•A/D count for 6 V is
abnormal.
Normal count for 2 V
(12452 < count value <
13763)
Normal count for 6 V
(37356 < count value <
41288)
BARCODE1 50 1 to 35WARNING Data reception from the
barcode reader has not
been completed before ID
reception time-out.
(Sub-code indicates the
position No. on disk.)
Replace the label or
barcode reader.
2 - 10
(cont’d)
Alarm
Control
Alarm Category
No.
1801 to
BARCODE2 54 1 to 5WARNING An error has occurred in
1805
1931 to
1971
2231 to
2271
2781 to
2830
2851,
REAGENT
SHORT
REAGENT
LEVEL
UNASSIGNED
TS OVER 67 1 WARNING Because 400 samples are
2852
2861 PATIENT
SAMPLE
OVER
Subcode
Level Description Remedy
communication with the
barcode reader. (Parity
error, framing error or
overrun error)
Sub-code indication
1:Unassigned
2:Unassigned
3:Sample disk barcode
reader
4:Unassigned
5:Unassigned
57 1 to 40WARNING • The volume of reagent to
be sipped from a reagent
bottle is inadequate.
• The reagent volume is a
total amount of one kind
of reagent for each test
item.
(1)Sub-code indicates
position No.
(2)Positions 37 and 38
are used for ISE.
(3)Position 40 is used for
HITERGENT.
(4)Position 39 is used for
HIALKALI.
59 1 to 40WARNING Reagent volume is smaller
than the remaining reagent
check value specified with
system parameter.
registered in routine sample
test selecting information,
new test selecting
information cannot be
registered from the host.
68 1 WARNING Because 400 samples are
registered in routine sample
test selecting information, a
new routine sample cannot
be analyzed.
Check the
communication cable
of barcode reader.
Set new reagent.
Set reagent newly.
After completion of
analysis, register TS
again.
Same as above
2 - 11
(cont’d)
Alarm
Control
No.
2891 to
2894
Alarm Category
DC POWER 71 1 STOP 15 V DC power supply is
Subcode
Level Description Remedy
Replace the ±15 V
abnormal.
power supply
module.
71 2 STOP -15 V DC power supply is
Same as above
abnormal.
71 3 WARNING 12 V lamp power supply is
abnormal.
71 4 STOP 5 V power supply is abnormal.
71 11 E.STOP 24 V DC
Replace the 12 V
power supply.
5 V for other than
CPU board
2911 FUSE 72 1 E.STOP AC fuse has blown. Replace the F3
fuse.
2921 POWER
FAIL
2941 to
2973
MOTOR
CONTROL-
73 1 WARNING Power supply to the
instrument is interrupted
75 1 to 18 STOP Data cannot be written into
the motor controller.
Check power
supply.
Replace ECPU230.
LER
2991 to
3023
MOTOR
TIMEOUT
76 1 to 18 E.STOP Motor operation was
abnormal. Error has been
detected in the time-out check
of motor controller.
(1) Replace
ECPU230.
(2) Check the
mechanism
which caused
Sub-code Motor
alarm.
1 Reaction disk
2 Sample disk
3 Reagent disk
4 Rinsing mechanism
up/down
5 Sample arm
up/down
6 Sample arm
rotation
7 Reagent arm
up/down
8 Reagent arm
rotation
9 Stirrer up/down
10 Stirrer rotation
11 ISE sipper up/down
12 (Unassigned)
13 Serum syringe
14 Reagent syringe
15 ISE syringe
16 (Unassigned)
17 GMCNT
18 (Unassigned)
2 - 12
(cont’d)
Alarm
Control
No.
3101 to
3189
3251 to
3291
3401 to
3438
Alarm Category
Subcode
STANDARD? 83 1 to 40
(ch.)
CALIBRATION
84 1 to 40
(ch.)
CALIB.SD? 85 1 to 36
(ch.)
Level Description Remedy
WARNING <<Photometry assay>>
(1) In calibration, the
STD absorbance data
is indicated with
alarm.
(2) In calibration, data
calculation is
disabled.
(Photometry assay)
• Replace STD
sample.
• Check the
concentration
parameter.
(ISE assay)
Same as above
<<ISE assay>>
(1) In calibration, the
potential data of
standard or internal
standard solution is
indicated with ADC
error, insufficient
sample alarm, noise
alarm or level alarm.
(2) In calibration, data
calculation is
disabled.
(The result of calibration
is not updated nor saved
onto FD.)
WARNING <<Photometry assay>>
The factor value 'K'
determined in calibration
(Photometry assay)
(ISE assay)
Same as above
differs from the previous
value by more than ±20%.
<<ISE assay>>
The calibrator
concentration value and
slope value determined in
calibration differ from the
previous values by more
than the compensation
limit (%).
WARNING The mean error determined
in multi-point calibration is
Replace STD and
check again.
larger than the SD limit
(input value).
2 - 13
(cont’d)
Alarm
Control
No.
3551 to
3588
Alarm Category
SENSITIVITY?
Subcode
86 1 to 36
(ch.)
Level Description Remedy
WARNING In linear (with 2 to 6 points) or
nonlinear calibration, a
difference between the mean
STD (1) absorbance and the
mean STD (N)
(Note 7 )
The result of
calibration is not
updated nor saved
onto FD.
absorbance is smaller than
the sensitivity limit (input
value).
NOTES :
7.N:=2 for linear
(2points)
=2 to 6 for nonlinear
and linear (3 to 6
points)
(Span point input value)
8.If either STD (1) or STD
(N) alone has been
measured, the
absorbance value of the
other STD is checked
using the previous data.
4151 to
4153
4161 to
4163
ISE LEVEL 90 1 WARNING The mean potential value
(EAV) at three out of five
measurement points of
internal standard solution is
outside the following range.
(Internal standard)
Na:-90.0mV ≤ EAV ≤ -10mV
⇒ OK
90 2 WARNING K:-90.0mV ≤ EAV ≤ -10mV
⇒ OK
90 3 WARNING Cl:100.0mV ≤ EAV ≤
180.0mV
⇒ OK
ISE NOISE 91 1 WARNING A difference (FIV) between
maximum and minimum
potential values at three of
five measurement points of
internal standard solution is
outside the following range.
(Internal standard, sample)
Na:0.7mV < FIV(2) - FIV(4)
91 2 WARNING K :1.0mV < FIV(2) - FIV(4) Same as above
(1)Replace STD
and carry out
calibration.
(2)Replace the Na
electrode.
(1)Same as above
(2)Replace the K
electrode.
(1)Same as above
(2)Replace the Cl
electrode.
Carry out reagent
priming and check
for bubble
formation.
91 3 WARNING Cl:0.8mV < FIV(2) - FIV(4) Same as above
2 - 14
Alarm
Control
No.
4171 to
4173
Alarm Category
Subcode
Level Description Remedy
ISE PREP. 92 1 WARNING Upon calibration, the slope
value is within the following
range.
Na : 45.0mV ≤ Slope value
≤ 49.9mV or
68.1mV ≤ Slope value
92 2 WARNING K : 45.0mV ≤ Slope value
≤ 49.9mV or
68.1mV ≤ Slope value
92 3 WARNING Cl : -39.9mV ≤ Slope value
≤ -35.0mV or
-68.1mV ≥ Slope value
(cont’d)
• Make sure that the
standard solution
and reagent are set
properly.
• Make sure that the
standard solution is
free from
concentration or
deterioration.
• Make sure that the
electrodes (Na, K,
Cl) are within their
guaranteed life.
4181 to
4183
4191 to
4193
ISE SLOPE 93 1 WARNING (1) In the result of
calibration, the slope
value is within the
following range.
(2) The response
characteristic of
electrode is poor (in case
carry-over rate (A) is as
indicated below).
Na : (1)SLOPE < 45.0mV
(2)0.232 < A
93 2 WARNING K : (1)SLOPE < 45.0mV
(2)0.160 < A
93 3 WARNING Cl : (1)SLOPE < -35.0mV
(2)0.490 < A
ISE I.STD 94 1 WARNING The concentration of internal
standard solution (C(IS)) is
within the following range.
Na : C(IS) < 120.0mEq/L or
160.0mEq/L < C(IS)
94 2 WARNING K : C(IS) < 3.0mEq/L or
7.0mEq/L < C(IS)
94 3 WARNING Cl : C(IS) < 80.0mEq/L or
120.0mEq/L < C(IS)
Refer to alarm
codes 92-1 to 3.
Same as above
Same as above
(1) Replace STD
and carry out
calibration
again.
(2) Replace t he
internal
standard
solution.
Same as above
Same as above
2 - 15
Alarm
Control
Alarm Category
No.
4201 to
4203
ISE
REAGENT
SHORT
4211 TWLN
TEST?
Subcode
Level Description Remedy
95 1 Unassigned
(Liquid level detection for
IS/DIL solution)
95 2 Unassigned (Same as
above)
95 3 WARNING The volume of reference
electrode solution is 30 mL
or less.
96 1 to 36
(ch.)
WARNING (1) when analyzing two
tests at a time, the
assigned method for
the corresponding test
is inadequate.
(2) Assignment of the
corresponding test is
being done or not
being done when
analysis for two tests is
not being made at the
same time.
(3) When analyzing two
tests at the same time,
the analytical
parameters are not
identical.
Sub-codes (1 to 36)
signify the channel
numbers.
(cont’d)
Replace the
reference electrode
solution with new
one.
Check the ISE
reagent volume.
• Start setting in
order from the
tests that are
designated by the
Measured Point of
the first half of the
1 channel 2 Test
Analysis Method (3
Point, 1 Point and
Rate, Rate B)
• When not
analyzing two tests
at the same time,
do not designate
'Two Test Analysis'
for the Analysis
Parameter.
• With the Analysis
Parameter screen
for the applicable
test, unify all the
parameters below
for the
Simultaneous 2
Test Analysis Test.
• Analytical Method
• Reaction Time
• Sample Volume
• Reagent
Pipetting volume
(R1-R3)
• Calibration
Method
• Calibration Point
• Standard solution
volume
• Standard solution
position
2 - 16
Alarm
Control
No.
4511 to
4548
4661 to
4698
Alarm Category
CHEMISTRY
PARAMETER?
CALIBRATION
PARAMETER CALIB.
Sub-
code
98 1 to 36
(ch.)
99 1 to 36
(ch.)
Level Description Remedy
WARNING(1)The relationship
between assay code and
Correct the
parameter.
photometric point is
improper.
(2)The assigned
photometric point lags
behind the specified
reaction time.
(Operation is
impossible.)
WARNING (1)The relationship
between assay code and
Check parameter
and input it again.
calibration type is
improper.
(2)Necessary calibration
points for calibration
type are not input.
(3)Necessary standard
positions for calibration
are not input.
(4)The relationship
between calibration type
and calibration method
is improper.
(5)The standard
concentration values are
not set in ascending
order. (Except for STD
(3) and (4) for isozyme)
(6)Concentration value is
not zero when '99' is
entered for POS. of STD
(1).
1.The photometry assay
data in other than
manual mode is
checked.
2.Operation is
unallowable.
3.Check in (4) is made
with regard to
specifications of timeout calibration, lot-to-lot
calibration, bottle-tobottle calibration and
test selecting
information.
(cont’d)
2 - 17
(cont’d)
Alarm
Control
No.
4811 to
4848
Alarm Category
VOLUME
CHECK?
Sub-
code
100 1 to 36
(ch.)
Level Description Remedy
WARNING (1)The total reagent
volume up to the last
photometric point is
more than 250 to
500 µL.
(2)The volumes of
reagents 1 to 3 are all
Check parameter.
If the improper
condition indicated
by alarm can be
detected, correction
and reentry are
required.
zero.
(3)The reagent volume
having a timing behind
the reaction time is not
zero.
(4)The total liquid volume
of sample and reagent is
less than 250 µL.
1.When '999' (stirring
only) is specified for
reagent volume, 0 µL is
taken for the volume.
However, when '999' is
specified for the reagent
volume having a timing
behind the reaction time,
an error occurs.
2.Operation is
unallowable.
3.The final liquid volume
is a total volume of
sample and reagent
within 250 to 500 µL.
5261 CMP. TEST 103 1 to 8 WARNING The setting of formula
number corresponding to
the relevant code is
improper.
(1)An unmeasurable test is
specified for
compensation.
(2)A compensated test is
not included in the
formula.
(3)In photometry assay for
compensated test, the
electrolyte parameter is
specified.
2 - 18
Call up calculation
item screen and
check the
compensation
formula on it.