There are several benefits to cold field emission and few detractors. Cold Cathode Field
Emission microscopy provides higher resolution, higher beam density (brightness), and longer
tip life than Thermal Tungsten wire SEMs and thermally assisted “Schottky” field emitters. The
following Table highlights those parameters responsible for the Cold Cathode Field Emission’s
higher performance at lower accelerating voltages.
Thermal Tungsten
Cold Cathode Field Emission
Source Diameter Energy Spread (eV) Brightness
50 – 100 kA◦ < 2.0 1 x
30 – 50 A◦ 0.2 1,000 x
HOW DOES IT WORK?
The field emission tip is made up of
a sharply etched piece of monocrystalline tungsten. A field is
applied to the tip causing electrons to
tunnel out of the tip and accelerate
down the column. This is a basic
diagram of the electron gun
assembly in a field emission system.
Notice that, in addition to the usual
accelerating voltage anode (V0), a
second voltage anode has been
placed in close proximity to the FE
tip. The desired accelerating voltage
is obtained by adjusting Vo to
accelerate or decelerate electrons
emitted at the extraction voltage V
It is the ultimate combination of these two anode potentials which sets the final electron speed
down the optical column. The computer automatically ratios these values for the operator.
12. Ensure that the BSE Detector is in the fully retra
cted (out of the column) position!!!
Nils Hasselmo Hall EM Lab 3 12/21/2011
COLD FINGER (Optional)
The built-in anti-contamination cold
finger, located on the right side of the
specimen chamber, can be filled with
liquid Nitrogen to reduce visible
contamination on the specimen by
collecting any grease, dirt, or impurities
that can impede image observation,
especially at high magnifications. The
cold finger has a capacity of 0.9 L and is
usable for nearly 5 hours once liquid
Nitrogen is injected. Fill the Cold Finger
with liquid Nitrogen. Pour in slowly at
first and allow the trap to chill down for
several minutes. Then add liquid Nitrogen
until overflow occurs. It is important to
maintain the trap fully cooled throughout
your session because trapped gases will release if warmed—increasing specimen contamination.
A sample mounted at the
appropriate height is shown
on the right.
The Sled is attached to the
Specimen Exchange Rod;
the Post connects the Sled
to the Sample Mount of
choice; the Locking Ring is
tightened to secure the
sample at the appropriate
height.
There are a variety of
Sample Mounts that fall
into two categories: those
to which the samples are
attached directly; those
which accept the standard Pin-Type stubs.
Nils Hasselmo Hall EM Lab 5 12/21/2011
SAMPLE INSERTION
1. Always check the sample height using the sample height gauge (if height exceeds the
gauge, consult with staff).
2. The Stage must be in the HOME position. Move the stage manually to the appropriate
Home Settings in the Table below. Current Home Setting can be obtained from the
“Stage Z Axis Setting Today” posting on the front of the column.
STAGE HOME POSITIONS
Stage Rod Z (mm) Rotate (°) Tilt (°) X (mm) Y (mm)
EMITECH cryo-stage EMITECH rod 15 0 0 12.5 12.5
EMITECH cryo-stage HITACHI rod 28.5 0 0 12.5 12.5
HITACHI stage HITACHI rod 12 0 0 12.5 12.5
3. Make sure the High Voltage is OFF*; S.C./S.E.C toggle switch to S.E.C position; and
the MV1 Chamber Valve is in the closed position.
*Consult “Computer Startup and Gun Flash Procedure” section if the computer/software is
not on/open.
4. Ensure that the STAGE LOCK is OFF.
5. Hit the AIR button to vent the exchange chamber.
6. Pull the door open by grabbing the SEC unit, not the rod.
7. Push the Specimen Exchange Rod slightly to unlock it and screw the sample holder onto
the end of the rod.
Nils Hasselmo Hall EM Lab 6 12/21/2011
Loading...
+ 13 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.