Hewlett-Packard 22A, 24A, HP 83620A User Manual

Page 1
HP
83620A/22A/24A
3420A 3250A 3245A 3213A 3145A 3143A 3119A 3108A 3104A 3102A 3050A 3044A 3036A
HP 8360 Series Synthesized Sweepers (Including Options 001, 003, 004, 006,
and 008)
User’s Handbook
SERIAL NUMBERS
This manual applies directly to any synthesized sweeper with serial number prefix combinations. You may have to modify this manual so that it applies directly to your instrument version. Refer to the
HP
HP
83623A 3420A 3339A 3250A 3245A 3213A 3145A 3143A 3119A 3108A 3104A 3102A 3050A 3044A 3036A
HP
83630A 3420A 3250A 3245A 3213A 3145A 3143A 3119A 3108A 3104A 3102A 3050A 3044A 3036A
HP
83640A 3420A 3339A 3250A 3245A 3213A 3145A 3143A 3101A
83650A 3420A 3250A 3245A 3213A 3145A 3143A 3052A
HP Part No. 08380-90070 Microfiche Part No. Printed in USA
08380-90073
November 1995
Edition 9
Page 2

Notice

The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Restricted Rights
Legend
Use, duplication, or disclosure by the U.S. Government is subject to restrictions as set forth in subparagraph (c) (1) (ii) of the Rights of Technical Data and Computer Software clause at DFARS
252.227-7013 for DOD agencies, and subparagraphs (c) (1) and (c) (2) of theCommercial clause at FAR 52.227-19 for other agencies.
Computer Software Restricted Rights
@ Copyright Hewlett-Packard Company 1992, 1995
All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws. 1400 Fountaingrove Parkway, Santa Rosa, CA 95403-1799, USA
Page 3

Certification

Hewlett-Packard Company certifies that this product met its
published specifications at the time of shipment from the factory.
Hewlett-Packard further certifies that its calibration measurements
are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International
Standards Organization members.

Warranty

This Hewlett-Packard instrument product is warranted against defects in material and workmanship for a period of one year from date of shipment. During the warranty period, Hewlett-Packard
Company will, at its option, either repair or replace products which prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by Hewlett-Packard. Buyer shall prepay shipping charges to Hewlett-Packard and Hewlett-Packard shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Hewlett-Packard from another country.
Hewlett-Packard warrants that its software and firmware designated by Hewlett-Packard for use with an instrument will execute its programming instructions when properly installed on that instrument. Hewlett-Packard does not warrant that the operation of the instrument, or software, or firmware will be uninterrupted or error-free.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. HEWLETT-PACKARD SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. HEWLETT-PACKARD SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
. . .
III
Page 4

Assistance

Product maintenance agreements and other customer assistance agreements are available for Hewlett-Packard products. For any assistance, contact your nearest Hewlett-Packard Sales and Service
Ofice.

Safety Notes

WARNING
CAUTION
The following safety notes are used throughout this manual. Familiarize yourself with each of the notes and its meaning before operating this instrument.
Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.
Caution denotes a hazard. It calls attention to a procedure that, if not correctly performed or adhered to, would result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
iv
Page 5

General Safety Considerations

WARNING
l
No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
n
For continued protection against fire hazard replace line fuse only
with same type and rating (F
5A/25OV).
The use of other fuses or
material is prohibited.
n
This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the instrument, is likely to make the instrument dangerous. Intentional interruption is prohibited.
n
This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the instrument, is likely to make the instrument dangerous. Intentional interruption is prohibited.
n
If this instrument is used in a manner not specified by Hewlett-Packard Co., the protection provided by the instrument may be impaired. This product must be used in a normal condition (in
which all means for protection are intact) only.
V
Page 6
CAUTION
H
Before switching on this instrument, make sure that the line
voltage selector switch is set to the voltage of the power supply and the correct fuse is installed.
n Always use the three-prong ac power cord supplied with this
instrument. Failure to ensure adequate earth grounding by not using this cord may cause instrument damage.
n Before switching on this product, make sure that the line voltage
selector switch is set to the voltage of the power supply and the correct fuse is installed. Assure the supply voltage is in the specified range.
w
Ventilation Requirements: When installing the instrument in a
cabinet, the convection into and out of the instrument must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the instrument by 4 “C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
n This product is designed for use in Installation Category II and
Pollution Degree 2 per IEC 1010 and 664, respectively.
Note
The detachable power cord is the instrument disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the instrument. The front panel switch is only a standby switch and is not a LINE switch.
vi
Page 7

PREFACE

This manual provides user information for the HP 8360 Series Synthesized Sweepers.
Instruments Covered
By This Manual
This manual applies to instruments having a serial number prefix listed on the title page (behind the “Documentation Map” tab). Some changes may have to be made to this manual so that it applies directly to each instrument; refer to Chapter 5, “Instrument History”, to see what changes may apply to your instrument.
A serial number label (Figure O-l) is attached to the instrument’s rear panel. A prefix (four digits followed by a letter), and a suffix (five digits unique to each instrument), comprise the instrument serial number.
SERIAL NUMBER
PREFIX SUFFIX
-f----l
7
INSTALLED
OPTIONS
1234A
i
SER
12345
Figure O-l. Typical Serial Number Label
An instrument’s prefix that is not listed on the title page may indicate that the instrument is different from those documented in this manual. For serial number prefixes before those listed on the title page, refer to the HP 8360 Series Synthesized
Sweepers Instrument History (to order, see “Replaceable Parts” in
Assembly-Level Repair).
vii
Page 8
User’s Handbook
Organization
Tabs divide the major chapters of this manual. The contents of each chapter is listed in the “Table of Contents.”

HP 8360 Series Documentation

Typeface Conventions

Documentation Map
For a pictorial representation of the HP 8360 series documentation,
see the “Documentation Map” at the front of this manual.
Ordering Manual
A manual part number is listed on the title page of this manual. You may use it to order extra copies of this manual.
Parts”
documentation and ordering numbers.
The following conventions are used in the HP 8360 series documentation:
Italics Italic type is used for emphasis, and for titles of manuals and
other publications. Computer Computer type is used for information displayed on the
instrument. For example: In this sequence, POWER LEVEL is displayed.
(-1
instructed to press a
in Assembly-Level Repair for a complete list of HP 8360
Instrument keys are represented in “key cap.” You are
hardkey.
See “Replaceable
Softkeys
functions depend on the current display. These keys are represented in “softkey.” You are instructed to select a softkey.
Softkeys are located just below the display, and their
Page 9

Regulatory Information

Manufacturer’s Declaration
This product has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the instrument in a safe condition.
Note
Note
This is to certify that this product meets the radio frequency interference requirements of Directive FTZ 1046/1984. The German Bundespost has been notified that this equipment was put into circulation and has been granted the right to check the product type for compliance with these requirements.
Note: If test and measurement equipment is operated with unshielded cables and/or used for measurements on open set-ups, the user must insure that under these operating conditions, the radio frequency interference limits are met at the border of his premises.
Model HP 8360 Series Synthesized Sweepers
Hiermit wird bescheinigt, dass dieses
Ubereinstimmung funkentstijrt
Der Deutschen Bundespost wurde das Inverkehrbringen dieses
GerHtes/Systems
der Serie auf Einhaltung der Bestimmungen Zustzinformation Werden Mess- und Testgerate mit ungeschirmten Kabeln
in offenen Messaufbauten verwendet, so ist vom Betreiber sicherzustellen, dass die Funk-Entstorbestimmungen unter Betriebsbedingungen an seiner Grundstiicksgrenze eingehalten werden.
mit den Bestimmungen von
ist .
angezeight und die Berechtigung
fur
Mess-und Testgerate:
Gerat/System
Postverfiigung
eingeraumt.
in
1046/84
zur fiberpriifung
und/oder
ix
Page 10

Notice for Germany: Noise Declaration

Declaration of
Conformity
LpA < 70 dB
am Arbeitsplatz (operator position) normaler Betrieb (normal position)
nach
DIN 45635 T. 19 (per
IS0
7779)
X
Page 11

DECLARATION OF CONFORMITY

accor&g to ISOIIEC Quide
22 and EN 45014
Manufacturer’s Name:
Manufacturer’s Address:
declares
conforms to the
that
the
product
Product Name: Model Numbers:
Product Options:
following Product specifications:
Safety: IEC 348:1978/HD 401 Sl:l981
CANZSA-C22.2
No. 231 (Series M-89)
Hewlett-Packard Co. Microwave Instruments Division
1400 Fountaingrove Parkway
Santa Rosa, CA 95403-1799
USA
Synthesized Sweeper HP 8362OA, HP
8362lA, HP
83622A, HP 83623A, HP 83624A, HP 8363OA, HP
8363lA, HP
8364OA, HP 36642A,
HP 8365OA, HP 83651A
This
declaration
above
products.
covers all
options of
the
EMC: ClSPR 11:199O/EN 55011:1991 Group 1, Class A
/EC 801-2:1984/EN 50082-1:1992 4
kV
CD, 8 kVAD /EC 801-3:1984/EN 50082-1:1992 3 V/m, 27-500 MHz /EC
80
f-4: 1988/EN
50082- 1:
1992
0.5
kV
Sig. Lines, 1
kV
Power Lines
Supplementary Information:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC
Santa Rosa, California, USA
European Contact:
and
the EMC Directive 89/336/EEC.
2 Oct. 1995
Your local Hewlett-Packard
Herrenberger
Sttasse
&/es
130,
D-71034 Biiblingen
and Service Office or Hewlett-Packard
Germany(FAX+4&7031-163143)
QmbH,
Department HQ-TRE,
xi
Page 12

Instrument Markings

A
!
The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
Cd
“ISMl-A”
I
I
0
I
The CE mark is a registered trademark of the European Community.
The CSA mark is a registered trademark of the
Canadian Standards Association.
This is a symbol of an Industrial Scientific and Medical Group 1 Class A product.
This is an ON symbol. The symbol ON is used to mark the position of the instrument power line switch.
This is an ON symbol. The symbol ON is used to mark the position of the instrument power line switch.
This is a STANDBY symbol. The STANDBY symbol is used to mark the position of the instrument power line switch.
This is an OFF symbol. The OFF symbol is used to mark the position of the instrument power line switch.
This is an AC symbol. The AC symbol is used to indicate the required nature of the line module input power.
xii
Page 13

Hewlett-Packard Sales and Service Offices

US FIELD
Headquarters Hewlett-Packard Co. 19320 Pruneridge Avenue Cupertino, CA 95014 (800) 752-0900
Colorado Hewlett-Packard Co. 24 Inverness Place, East Englewood, CO 80112 (303)
649-5512
New
Jersey
Hewlett-Packard Co. 150 Green Pond Rd. Rockaway, NJ 07866 (201) 586-5400
California, Northern Hewlett-Packard Co. 301 E. Evelyn Mountain View, CA 94041 (415) 694-2000
Georgia Hewlett-Packard Co. 2000 South Park Place Atlanta, GA 30339 (404)
955-1500
Texas
Hewlett-Packard Co. 930 E. Campbell Rd. Richardson, TX 75081
(214)231-6101
EUROPEAN
Headquarters Hewlett-Packard S.A. Hewlett-Packard France 150, Route du Naut-d’Avri1 1 Avenue Du Canada 1217 Meyrin 2JGeneva Zone Switzerland
(41
22)780.8111
Great
Britain
Hewlett-Packard Ltd. Eskdale Road, Winnersh Triangle
Wokingham, Berkshire RG41
England (44 734) 696622
France
D’Activite F-91947 Les Ulis Cedex
France
(33
1)69
82 60 60
5DZ
OPERATIONS
FIELD
De Courtaboeuf
California, Southern Hewlett-Packard Co. 1421 South Manhattan Ave. Fullerton, CA 92631 (714) 999-6700
Illinois Hewlett-Packard Co. 5201 Rolling Meadows, IL 60008 (708) 255-9800
OPERATIONS
Germany
Hewlett-Packard GmbH Hewlett-Packard Strasse 61352 Bad Homburg v.d.H
Germany (49 6172)
Tollview
16-O
Drive
INTERCON
Headquarters
Hewlett-Packard Company 3495 Deer Creek Road Palo Alto, California, USA 94304-1316 (415) 857-5027
Australia
Hewlett-Packard Australia Ltd.
31-41 Joseph Street
Blackburn, Victoria 3130
(61 3) 895-2895
China Japan
China Hewlett-Packard Company 38
Bei
San Huan Xl Road
Shuaug
Yu Shu
Hai
Dian District Beijing, China (86 1) 2566888
Hewlett-Packatid
1-27-15 Yabe,
Kauagawa 229, Japan
(81
427)59-1311
Taiwan
Hewlett-Packard Taiwan
8th
Floor, H-P Building 337 Fu Hsing North Road Taipei, Taiwan
:886 2) 712-0404
FIELD
Japan, Ltd.
Sagamihara
OPERATIONS
Canada
Hewlett-Packard (Canada) Ltd. 17500 South Service Road Tram-Canada Highway Kirkland, Quebec Canada (514) 697-4232
Singapore
Hewlett-Packard Singapore (Pte.) Ltd. 150 Beach Road
#29-00
Singapore 0718 (65) 291-9088
H9J 2X8
Gateway West
. . .
XIII
Page 14
Contents
1.
GETTING STARTED
What Is In This Chapter How To Use This Chapter
Equipment Used In Examples Introducing the HP 8360 Series Synthesized Sweepers Display Area Entry Area CW Operation and Start/Stop Frequency Sweep . .
CW Operation
Start/Stop Frequency Sweep Center Frequency/Span Operation Power Level and Sweep Time Operation
Power Level Operation
Sweep Time Operation Continuous, Single, and Manual Sweep Operation Marker Operation Saving and Recalling an Instrument State Power Sweep and Power Slope Operation
Power Sweep Operation
Power Slope Operation Getting Started Advanced Externally Leveling the Synthesizer
Leveling with Detectors/Couplers /Splitters ...
External Leveling Used With the Optional Step
Leveling with Power Meters
Leveling with MM-wave Source Modules Working with Mixers/Reverse Power Effects
Working with Spectrum Analyzers/Reverse Power
Effects
Optimizing Synthesizer Performance
Creating and
Array
Creating a User Flatness Array Automatically,
Creating a User Flatness Array, Example 2 . . Swept mm-wave Measurement with Arbitrary
Scalar Analysis Measurement with User Flatness
Using Detector Calibration Using the Tracking Feature
..................
..................
.................
...............
Attenuator
...................
..................
Example 1
Correction Frequencies, Example 3
Corrections, Example 4
...............
Applymg
...............
............
............
.........
..........
........
......
............
............
............
............
............
........
..........
.......
the User Flatness Correction
.........
..........
..........
.
.....
.....
.....
....
....
l-l
l-2 l-2 l-3 l-4 l-5 l-6 l-6 l-6
l-8 l-10 l-10 l-10
1-12 1-14 1-16 1-18 1-18 1-19 1-21
l-23 l-23
l-26 l-27 l-28 l-30
l-32 l-33
l-33
l-34 l-36
l-39
l-43 l-47 l-49
HP 8360 User’s Handbook
Contents-l
Page 15
Peaking..................
Tracking
ALC Bandwidth Selection............
Using Step Sweep
Creating and Using a Frequency List .......
Using the Security Features ...........
Changing the Preset Parameters .........
Getting Started Programming..........
HP-IB General Information ...........
Interconnecting Cables ............
Instrument Addresses.............
HP-IB Instrument Nomenclature........
Listener
Talker...................
Controller.................
Programming the Synthesizer .........
HP-IB Command Statements .........
Abort
Remote..................
Local Lockout ...............
Local...................
Clear...................
output..................
Enter...................
Getting Started with SCPI ...........
Definitions of Terms ..............
Standard Notation ..............
Command Mnemonics ...........
Angle Brackets...............
How to Use Examples.............
Command Examples ............
Response Examples.............
Essentials for Beginners.............
Program and Response Messages .......
Forgiving Listening and Precise Talking....
Types of Commands ............
Subsystem Command Trees ..........
The Command Tree Structure........
Paths Through the Command Tree......
Subsystem Command Tables..........
Reading the Command Table ........
More About Commands...........
Query and Event Commands........
Implied Commands............
Optional Parameters ...........
Program Message Examples.........
Parameter Types..............
Numeric Parameters ...........
Extended Numeric Parameters.......
Discrete Parameters ...........
Boolean Parameters ...........
.................
...............
..................
...................
l-49 l-49 l-50
1-51
l-52 l-53 l-54 l-55 l-56 l-56 l-56 l-56 l-56 l-56 l-56 l-56 l-57 l-57
l-58
1-58
l-59 l-59 l-60
1-61
l-63 l-63 l-64 l-64 l-64 l-64 l-64 l-65 l-66 l-66 l-66 l-67 l-67 l-67 l-68
1-71 1-71
l-72 l-72 l-72 l-72 l-72 l-73 l-73 l-74 l-74 l-75
Contents-2
User’s Handbook
HP 8360
Page 16
Reading Instrument Errors . . . . . . . . . .
Example Programs ..............
Example Program .............
Description ...............
Program Listing .............
Program Comments ...........
Details of Commands and Responses .......
In This Subsection ..............
Program Message Syntax ...........
Subsystem Command Syntax ........
Common Command Syntax .........
Response Message Syntax ...........
SCPI Data Types ..............
Parameter Types ..............
Numeric Parameters ...........
Extended Numeric Parameters .......
Discrete Parameters ...........
Boolean Parameters ...........
Response Data Types ............
Real Response Data ...........
Integer Response Data ..........
Discrete Response Data ..........
String Response Data ...........
Programming Typical Measurements .......
In This Subsection ..............
Using the Example Programs .........
Use of the Command Tables .........
HP-IB Check, Example Program 1 .......
Program Comments ............
Local Lockout Demonstration, Example Program 2
Program Comments ............
Setting Up A Typical Sweep, Example Program 3
Program Comments ............
Queries, Example Program 4 ..........
Program Comments ............
Saving and Recalling States, Example Program 5 .
Program Comments ............
Looping and Synchronization, Example Program 6
Program Comments ............
Using the
Program Comments ............
Using the User Flatness Correction Commands,
Example Program 8 ............
Programming the Status System .........
In This Subsection ..............
General Status Register Model .........
Condition Register .............
Transition Filter ..............
Event Register ...............
Enable Register ..............
An Example Sequence ...........
*WA1
Command, Example Program 7 .
l-75 l-76 l-76 l-76 l-76 l-77 l-79
l-79 l-79 l-80 l-80
1-81
l-82 l-82 l-82 l-83 l-84 l-84 l-84 l-84 l-85 l-85 l-85 l-86 l-86 l-86 l-87
1-88
l-89 l-89 l-90 l-90
1-91
l-92 l-92 l-93 l-94
l-95 l-95 l-96 l-97
l-98 l-101 l-101 l-101 l-101 l-102 l-102 l-102 l-102
HP 8360 User’s Handbook
Contents-3
Page 17
Programming the Trigger System .........
In This Subsection ..............
Generalized Trigger Model ...........
Overview .................
Details of Trigger States ...........
Inside the Idle State ...........
Inside the Initiate State ..........
Inside Event Detection States .......
Inside the Sequence Operation State ....
Common Trigger Configurations ........
The
INIT
Configuration ...........
The TRIG Configuration ..........
Description of Triggering in the HP 8360 Series
Synthesizers ...............
Advanced Trigger Configurations .......
Trigger Keyword Definitions ..........
ABORt IMMediate ODELay
SOURce ..................
Related Documents ...............
The International Institute of Electrical and
Electronics Engineers. ...........
Hewlet
..................
................
.................
t-Packard Company ...........
l-104 l-104 l-104 l-104 l-105 l-105 l-106 l-107 l-109 l-109 l-109 l-110
l-111
1-112 1-112 1-113 1-113
1-113 1-113 1-114
1-114
l-114
2.
OPERATING AND PROGRAMMING REFERENCE
How To Use This Chapter . . . . . . . . . . . .
A.
Address ..........
1
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
..........
Adrs Menu
LJ
. . . . . . . . . .
@-
,C 3andwidth
AI
hLC
Bandwidth Select High
ALC
Bandwidth
AI..C BW
.tmate
81
hM
BW Cal Always
........
Select
Selsct
Menu
.......
Regs
Autc
Low
......
. . . .
AM BW Cal Once . . . . . .
AYI Cal Menu . . . . . . .
AMMenu hM
AM On/Off
. . . . . . . . .
On/Off 10
dB/V
100%/V
. . . .
. . . .
AM On/Off Ext . . . . . .
hM
On/Off Int . . . . . .
Amp1
Markers . . . . . . .
AM Type IO
dB/V
. . . . .
2-l
A-l A-l
A-3
A-10
A-10 A-11 A-11 A-12 A-12 A-13 A-13 A-14 A-15 A-15 A-16 A-17 A-17 A-18
Contents-4
HP 8360
User’s Handbook
Page 18
hM
Type
100%fV
................
ANALYZER STATUS REGISTER
ArrowKeys
(ASSIGN) .- . . . .
Auto Fill Auto Fill Auto Fill Start Auto Fill Stop . Auto Track . . .
..................
. . . .
Incr %Pts
. .
. . . .
. . . .
. . . .
. . . .
. . . .
B.
Blank Disp . .
C.
[CENTER). . . .
Center=Marker
Clear Fault . Clear
Memory
Clear Point .
CONNECTORS
(CONT)
.....................
Copy List
CorPair
Disable
Coupling Factor
@g
CW/CF
.....................
Coupled
. . . . . . . . . . . . . . . .
................
................
................
.
................
................
................
..................
...............
...............
................
........
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . .
. . .. .
. . . . . .
. . . . . .
. . . . . .
. . .
. . .
. . .
A-19 A-19
A-21 A-22
.
A-23 A-24
A-25 A-25 A-26
B-l
C-l
c-2 c-2 c-3
c-4 c-4
c-12
c-13
c-13
c-14
c-14
c-15
HP 8360 User’s Handbook
D.
Dblr Amp
DeepAM
Delay Menu Delete Menu Delete All Delete Current Delete Undef Delta Delta Mkr Ref Disp Status
Doubler Amp
Menu
................
...................
..................
.................
..................
................
.................
Ma&e+
.................
................
.................
Mode
Doubler Amp Mode Off Doubler Amp
Dwell Coupled
Hode
................
AUTO
............
............
On
.............
D-l
D-2
D-2 D-3 D-3
D-4 D-5 D-5
D-6 D-6 D-8
D-9
D-10 D-10
Contents-6
Page 19
E.
8360
Adz-s
Enter
Cum
EnterFreq
..................
..................
..................
Enter List Dwell Enter List Freq Enter List Offset
ENTRY KEYS
[ENTR~~N/OFF]
ExtDetCal
F.
Fault Menu
..................
Fault Info 1 Fault Info 2 Fault Info 3 Fltness Menu
-
[FLTNESS ON/OFF)
FM Coupling
1OOkHz
FM Coupling DC . .
FM Menu . . . . .
FM On/Off AC . . . FM On/Off DC . . . FM
On/Off
Ext . . FM On/Off Int . . Freq Cal Menu . . Freq Follow . . .
FREQUENCY
@ki-)
Freq Mult . . . .
Freq Offset . . .
FullUsr
Cal . . .
..............
...............
..............
.................
.................
.................
.................
.................
.................
.................
. .
..............
.............
..............
..............
..............
..............
..............
..............
..............
..............
..............
..............
..............
. . . . . . . . . . . . . .
E-l
E-l
E-2 E-2 E-3 E-4
E-4 E-5
E-5
F-l
F-2
F-3
F-4
F-5
F-11 F-11
F-12 F-13 F-14 F-14 F-15 F-16 F-16 F-17
F-17 F-18
F-19 F-20
Contents-6
G.
Global Dwell
Global Offset
.................
................
User’s Handbook
G-l
G-l
HP 8360
Page 20
H.
HP-IB Address .................
HP-IB
I.
Menu
..................
Internal AM Depth . . . . . . . . . . . . . .
Internal Internal AM Waveform Noise Internal AM Waveform Ramp Internal AM Waveform Sine Internal AM Internal AM Waveform Triangle
Intexnal
Internal FM Rate Internal FM Waveform Noise Internal FM Waveform Ramp Internal FM Waveform Sine Internal FM Waveform Square Internal FM Waveform Triangle Internal Menu Internal Pulse Generator Period Internal Pulse Generator Rate Internal Pulse Generator Width Internal Pulse Mode Auto
Internal Pulse Mode Gate Internal Pulse Mode Trigger Invert Input
AM
Rate
Waveform
..............
Square
FM Deviation
..............
................
.................
.........
..........
..........
.........
........
............
.........
..........
..........
.........
........
.......
........
.......
..........
..........
.........
H-l H-l
I-l
I-2 I-2 I-3 I-3 I-4 I-4 I-5 I-5 I-6 I-6 I-7
I-7 I-8 I-8 I-9
I-10 I-10
I-11
I-11 I-12 I-12
HP 8360 User’s Handbook
L.
Leveling
Leveling Leveling Leveling Leveling Pointlntml Leveling
Leveling
LINE SWITCH ................
ListMenu
List Mode Pt List Mode Pt List Mode Pt
(LOCAL)
.....................
ModeALCoff ModeNormal ModeSearch PointExtDet
............. L-l
.............
.............
............
............
PointModule PointPwrMtr
............
............
..................
TrigAuto TrigBus TrigExt
............
............
............
L-2 L-2 L-3 L-3 L-4 L-4
L-5 L-5
L-8 L-8 L-9
L-9
Contents-7
Page 21
M.
MI--M2
Manual Sweep
(MARKER)
....................
MarkerMi
Marker
MarkerM3
Marker
MarkerM5
Markers All Off Measure
Measure Corr Current Measure Corr Undef Meter
Meter
Meter On/Off FM
@
ModOut ModOut
Modulation Amplitude Modulation
FMModulation
Pulse Modulation
Module Menu Module Select AUTO Module Select Front Module Select None
Module Select Rear Monitor Menu morenfm.. Mtr
Sweep
..........
:
.................
..................
M2
..................
..................
M4
..................
..................
...............
Corx
All
..............
............
.............
Adxs
On/Off AM
..................
...............
...............
.....................
On/Off API On/Off FM
..............
..............
..................
.............
................
................
.................
.............
.............
.............
.............
.................
.................
Meas
Menu
................
......
M-l M-l
M-3
M-4 M-5
M-5 M-6 M-6 M-7 M-7 M-8 M-8 M-9 M-9
M-10 M-10 M-11
M-12 M-12 M-14 M-17 M-19 M-23
M-24 M-24 M-25
M-26 M-26 M-27 M-28
Contents-8
P.
PPPeak RF Always
PeakRFOnce
(POWER
(POWER
POWER(E)POWER(E)
Pawer
P Power Slope
P
LEVEL)
LEVEL)
.................
.................
Offset
Power Sweep
(PRESET)
;w)
......................
Preset Mode Factory . . . . . . . . . . . . .
Preset Mode Factory Preset Mode User
Preset Mode User . . . . . . . . . . . . . .
. .- . . . . . . . . . . . . . . . . . .
................
.................
..................................
.................
.................
.................
.............
..............
HP 8360
User’s Handbook
P-l
P-2 P-2
P-5 P-6
P-6 P-7
P-8 P-9
P-10
Page 22
Printer Adxs . . . . . . . . . . . . . . . . .
(PRIOR) . . . . . . . . . . . .
Programming Language Programming Language
Analyzr CIIL
. .
Programming Language SCPI . .
P
t Trig Menu . . . . . . . . .
Pulse Delay
Normal
. . . . .
Pulse Delay Txig'd . . . . .
Pulse Menu . . . . . . . . . .
Pulse Menu . . . . . . . . . .
Pulse Pulse Pulse
OnfOffExtrnl On/OffIntxnl OnfOffScalax
. . . . .
. . . . .
. . . . .
Pulse Period . . . . . . . . .
Pulse Rate . . . . . . . . . .
Pulse Rise Pulse Rise Pulse Rise
TimeAuto TimeFast TimeSlow
. . . . .
. . . . .
. . . . .
Pulse Width . . . . . . . . .
PwrMtr
Range . . . . . . . . . . . . . . . .
........
........
........
........
........
........
........
........
........
........
........
........
........
........
........
........
........
........
P-10 P-11 P-12
P-12 P-13 P-14 P-14 P-15 P-16
P-16 P-18 P-18 P-19 P-20 P-20 P-21 P-21 P-22 P-22 P-23
R.
(RECALL)
.....................
Ref Osc Menu
(-ON/OFF)
...................
ROTARY KNOB ................
S.
LSAVE)
.....................
SaveLock
Save User Preset
SCPI Conformance Information
SCPI COMMAND SUMMARY SCPI STATUS REGISTER
Security
Selftest
(Full)
SetAtten
(SINGLE)
.....................
SoftwareRev
.................
..................
..............
.........
STRU&iRi
Menu
................
...............
..................
.................
..........
..........
..........
..........
..........
’ : : : :
R-l R-l
R-2 R-2
S-l
s-2 s-2
s-3 s-14 S-56 S-58
s-59 s-59
S-60 S-60 S-61
S-61 S-62
S-63 S-63
HP 8360 User’s Handbook
Contents-9
Page 23
. . . . . . . . . . .
...........
...........
. . . . . . . . . . .
. . . . . . . . . . .
. . . . . . . . . . .
. . . . . . . . . . .
...........
...........
...........
...........
...........
. . . . . . . . . . .
...........
...........
...........
...........
S-64 S-64 S-66 S-67 S-67 S-68 S-69 s-70 s-70 s-71
s-71 S-72
s-73 s-73 s-74 s-75 s-75
S-76 S-76 s-77
T.
10 MHz Freq Std Auto 10
MHz
Freq Std
10 MHz Freq Std
Exts-nl Intrnl
10 MHz Freq Std None Tracking Menu
TrigOut
U.
Delay
Uncoupl Atten
Unlock Info Up/Down Power
Up]Dn
Size CW
Up/Dn
Size Swept
[UsER).
...................
USER DEFINED
UsrKey
UsrMenu
Clear
Clear
................
................
................
.................
................
................
[MENU)
.............
.................
................
............
...........
...........
............
..............
T-l T-l
T-2 T-2 T-3 T-3
U-l U-l
u-2 u-2 u-3
u-4 u-4
u-5 u-5
Contents-l 0
HP 8360
User’s Handbook
Page 24
Z.
ZeroFreq
Wavef Zoom
2a. ERROR MESSAGES
Introduction
ofllt
.....................
..................
Nenu
................
..................
Front Panel Error Messages in Alphabetical Order
SCPI Error Messages in Numerical Order .....
Synthesizer Specific SCPI Error Messages
Universal SCPI Error Messages ........
Error Messages From -499 To -400 .....
Error Messages From -399 To -300 .....
Error Messages From -299 To -200 .....
Error Messages From -199 to -100 ......
2b. Menu Maps
AX
Menu
Frequency Menu
Marker Menu
..................
................
.................
Modulation Menu ...............
Power Menu
Service Menu Sweep Menu System Menu
User Cal Menu
..................
.................
..................
.................
.................
.
....
Z-1 Z-1
z-2
2a-1 2a­2a-5 2a-5 2a-6 2a-6 2a-6 2a-6 2a-7
2b-2 2b-4 2b-6 2b-8
2b-10 2b-12 2b-14 2b-16 2b-18
1
2c. Specifications
Frequency
Range
....................
Resolution
Frequency Bands (for CW signals) .......
Frequency Modes: ..............
CW and Manual Sweep ............
Synthesized Step Sweep ............
Synthesized List Mode ............
Ramp Sweep Mode ..............
Internal 10 MHz Time Base ..........
RF Output
Output Power
Accuracy ( dB)4 ..............
Flatness
Analog Power Sweep .............
External Leveling ...............
Source Match
Spectral Purity
Spurious Signals ...............
Single-Sideband Phase Noise (dBc/Hz) .....
Offset from Carrier .............
Residual FM (RMS, 50 Hz to 15 kHz bandwidth) .
...................
..................
..................
................
(dB)
...............
................
.................
2c-2 2c-2 2c-2
2c-2 2c-2 2c-2 2c-3 2c-3 2c-3 2c-3 2c-4 2c-4 2c-5
2c-5 2c-6 2c-6
2c-6
2c-7 2c-7 2c-9
2c-9
2c-9
HP 8360
User’s Handbook
Contents-l 1
Page 25
Modulation
Pulse AM and Scan
FM .....................
Simultaneous Modulations
Internal Modulation Generator Option
AM,FM
Pulse Modulation Meter.
General
Environmental Warm-Up Time Power Requirements Weight & Dimensions
Adapters Supplied
Inputs & Outputs
Auxiliary Output RF Output External ALC Input Pulse Input/Output AM Input FM Input Trigger Input Trigger Output 10 MHz Reference Input 10 MHz Reference Output Sweep Output Stop Sweep Input/Output
Z-Axis Blanking/Markers Output .......
Volts/GHz
Source Module Interface Auxiliary Interface
Pulse Video Output (Option 002 only) ....
Pulse Sync Out (Option 002 only) ......
AM/FM Output (Option 002 only) ......
Models Options
Option 001 Add Step Attenuator .......
Option 002 Add Internal Modulation Generator
Option 003 Delete Keyboard/Display .....
Option 004 Rear Panel RF Output ......
Option 006 Fast Pulse Modulation ......
Option 008 1 Hz Frequency Resolution ....
Option 700 MATE System Compatibility ...
Option 806 Rack Slide Kit
Option 908 Rack Flange Kit
Option 910 Extra Operating & Service Manuals
Option 013 Rack Flange Kit
Option W30 Two Years Additional Return-To-HP
Service . . . . . . . . . . . . . . . . .
..................
....................
................
...........
002
..................
....................
..............
....................
................
...............
.............
.............
..............
..............
..............
................
............
............
.................
.................
...............
...............
...........
..........
...............
..........
Output
...................
...................
.............
...........
.............
..........
.........
.........
....
2c-10 2c-10 2c-11 2c-12 2c-12
2c-13
2c-13 2c-13 2c-13 2c-14
2c-14
2c-14
2c-14
2c-14 2c-14 2c-15
2c-15 2c-15
2c-15 2c-15 2c-15 2c-15 2c-15 2c-15 2c-16 2c-16 2c-16 2c-16 2c-16 2c-16 2c-16 2c-16 2c-16 2c-17 2c-17 2c-17 2c-17 2c-17 2c-17 2c-17 2c-17 2c-17 2c-18 2c-18 2c-18 2c-18 2c-18 2c-18
2c-18
Contents-l 2
HP 8360
User’s Handbook
Page 26
3.
INSTALLATION
Initial Inspection Equipment Supplied Options Available Preparation for Use
Power Requirements Line Voltage and Fuse Selection Power Cable Language Selection
How to View or Change a Language Selection from
the Front Panel .............
How to Select a Language on a Synthesizer without
a Front Panel ..............
HP-IB Address Selection
How to View or Change an HP-IB address from
the Front Panel .............
How to Prevent a Front Panel Change to an HP-IB
Address ................
How to Set the HP-IB Address on a Synthesizer
without a Front Panel ..........
Mating Connectors
10 MHz Frequency Reference Selection and Warmup
Time
Operating Environment
Chassis Kits
Rack Mount Slide Kit (Option 806)
Installation Procedure
Rack Flange Kit for Synthesizers with Handles
Removed (Option 908)
Installation Procedure
Rack Flange Kit for Synthesizers with Handles
Attached (Option 913)
Installation Procedure
Storage and Shipment
Environment Package the Synthesizer for Shipment
Converting HP 8340/41 Systems to HP 8360 Series
Systems
Manual Operation
Compatibility Front Panel Operation
Instrument Preset Conditions .......
System Connections
The HP 8510 Network Analyzer The HP The HP 83550 Series Millimeter-wave Source
Modules ................
The HP
Remote Operation
Language Compatibility Network Analyzer Language
................
..............
...............
...............
.............
........
.................
..............
...........
..............
..................
............
..................
.......
............
...........
............
...........
............
..............
.................
..................
..............
...............
...........
.............
.......
8757C/E
8970B
Scalar Network Analyzer
Noise Figure Meter
..............
...........
.........
......
. .
......
3-l
3-2 3-2 3-3 3-3 3-3 3-4 3-6
3-6
3-6 3-7
3-8
3-8
3-8 3-8
3-8
3-9
3-10 3-10 3-10
3-13 3-14
3-15 3-16 3-17 3-17 3-18
3-19
3-20 3-20 3-20 3-20 3-21 3-21 3-22
3-22 3-22 3-23 3-23 3-23
HP 8360
User’s Handbook
Contents-13
Page 27
Test and Measurement System Language . . .
Control Interface Intermediate Language . . .
Converting from Network Analyzer Language to
SCPI
. . . . . . . . . . . . . . . . . .
Numeric Suffixes . . . . . . . . . . . . . .
Status Bytes . . . . . . . . . . . . . . . .
4. OPERATOR’S CHECK and ROUTINE MAINTENANCE
Operator’s Checks
Service Information
Local Operator’s Check
Description Preliminary Check Main Check
Routine Maintenance
How to Replace the Line Fuse How to Clean the Fan Filter How to Clean the Cabinet How to Clean the Display Filter
5. Instrument History
How to Use Instrument History Change B Modulation
Pulse
...................
..................
....................
AM and Scan
Change A
...................
...............
..............
.............
.................
..............
.................
..............
.........
..........
...........
........
.........
................
3-23 3-23
3-23 3-24 3-24
4-l 4-l
4-2 4-2 . 4-2 4-3 4-4 4-4 4-5 4-6 4-6
5-l
5-3 5-5 5-5 5-6 5-7
Index
Contents-14
HP 8360
User’s Handbook
Page 28
Figures
O-l. Typical Serial Number Label . . . . . . . . . .
l-l. The HP 83620A Synthesized Sweeper . . . . . .
1-2. Display . . . . . . . . . . . . . . . . . . .
l-3.
Entry Area . . . . . . . . . . . . . . . . .
l-4.
CW Operation and Start/Stop Frequency Sweep .
l-5.
Center Frequency and Span Operation . . . . .
l-6.
Power Level and Sweep Time Operation . . . . .
l-7.
Continuous, Single, and Manual Sweep Operation
1-8. Marker Operation . . . . . . . . . . . . . .
l-9.
Saving and Recalling an Instrument State . . . .
l-10.
Power Sweep and Power Slope Operation . . . .
l-11.
ALC Circuit Externally Leveled . . . . . . . .
l-12.
Typical Diode Detector Response at
1-13. Leveling with a Power Meter . . . . . . . . . .
1-14. MM-wave Source Module Leveling . . . . . . .
1-15. MM-wave Source Module Leveling Using a Microwave
Amplifier . . . . . . . . . . . . . . . . .
1-16. Reverse Power Effects, Coupled Operation with
-8dBm
1-17. Reverse Power Effects, Uncoupled Operation
-8dBm
1-18. Creating a User Flatness Array Automatically
1-19. Creating a User Flatness Array . . . . . . .
l-20.
Creating Arbitrarily Spaced Frequency-Correctic
Pairs in a Swept mm-wave Environment
1-21. Scalar System Configuration . . . . . . . .
l-22.
Automatically Characterizing and Compensating for
a Detector . . . . . . . . . . . . . . . .
l-23.
Decision Tree for ALC Bandwidth Selection . . .
l-24.
SCPI Command Types . . . . . . . . . . . .
l-25.
A Simplified Command Tree . . . . . . . . . .
l-26.
Proper Use of the Colon and Semicolon . . . . .
l-27.
Simplified
l-28.
Voltage Controlled Oscillator Test . . . . . . .
l-29.
Simplified Program Message Syntax . . . . . . .
l-30.
Simplified Subsystem Command Syntax . . . . .
1-31. Simplified Common Command Syntax . . . . .
l-32.
Simplified Response Message Syntax . . . . . .
l-33.
Generalized Status Register Model . . . . . . .
l-34.
Typical Status Register Bit Changes . . . . . .
l-35.
Generalized Trigger Model . . . . . . . . . . .
1-36. Inside the Idle State . . . . . . . . . . . . .
Output . . . . . . . . . . . .
Output . . . . . . . . . . . .
SWEep
Command Tree . . . . . . .
25°C
. . . .
. .
th
wi
. . . . . .
on
. .
.
. .
vii
l-3 l-4 l-5 l-7 l-9
l-11
1-13 1-15 1-17
l-20 l-23 l-25 l-27 l-28
l-29
1-31
1-31
l-34 l-37
l-40 l-43
l-47 l-50 l-67 l-68 l-70
1-71
l-76
l-79 l-80
1-81
1-81
l-101 l-103 l-105 l-106
HP 8360
User’s Handbook
Contents-15
Page 29
l-37.
Inside the Initiate State ............
l-38.
Inside an Event Detection State ........
l-39.
Inside the Sequence Operation State ......
l-40.
The
INIT
Trigger Configuration ........
1-41. The TRIG Trigger Configuration ........
l-42.
HP 8360 Simplified Trigger Model .......
A-l. ALC System Simplified Block Diagram .....
A-2. Typical External Leveling Hookup .......
C-l. Auxiliary Interface Connector .........
C-2. HP-IB Connector and Cable ..........
C-3. Interface Signals of the Source Module Connector .
F-l. Basic User Flatness Configuration Using an HP
Power Meter ...............
F-2. User Flatness Correction Table as Displayed by the
Synthesizer ................
F-3. The Sources of ALC Calibration Correction Data . F-4. Array Configuration when the Correction Data
Frequency Span is a Subset of the Synthesizer
Frequency Span ..............
M-l. ALC Block Diagram .............
M-2.
Power Accuracy Over the AM Dynamic Range . .
M-3.
FM Deviation and Rate Limits .........
M-4.
ALC Block Diagram .............
M-5.
Pulse Modulation System ...........
M-6.
Video Feedthrough ..............
P-l. How (PRIOR) Works
S-l. Connections Required for a Two-Tone Scalar
3-l.
AC Power Cables Available ..........
3-2. Rear Panel HP-IB Switch ...........
3-3. Removing the Side Straps and Feet .......
3-4. Chassis Slide Kit ...............
3-5. Rack Mount Flanges for Synthesizers with Handles
3-6. Rack Mount Flanges for Synthesizers with Handles
4-l.
Replacing the Line Fuse ............
4-2. Removing the Fan Filter ...........
...............
Network Analyzer Measurement System ...
Removed .................
Attached
.................
437B
l-106 l-108 l-109 l-110 l-111 l-111
A-5 A-8
c-7 c-9
c-11
F-6
F-7 F-8
F-9 M-13 M-16 M-18 M-20 M-20 M-22
P-11
S-65
3-5
3-7 3-11 3-12
3-14
3-16
4-4
4-5
Contents-16
HP 8360
User’s Handbook
Page 30
Tables
l-l. Keys Under Discussion in This Section . . . . .
l-2. SWEep l-3.
SCPI Data Types . . . . . . . . . . . . . .
l-4.
Sample Synthesizer Commands . . . . . . . . .
C-l. Pin Description of the Auxiliary Interface . . . .
D-l. Mnemonics used to Indicate Status . . . . . . .
S-l. HP 8360 SCPI COMMAND SUMMARY . . . .
3-l.
Language HP-IB Addresses . . . . . . . . . .
3-2. Factory-Set HP-IB Addresses . . . . . . . . .
3-3. Rack Mount Slide Kit Contents . . . . . . . .
3-4. Rack Flange Kit for Synthesizers with Handles
3-5. Rack Flange Kit for Synthesizers with Handles
3-6. Instrument Preset Conditions for the HP
3-7. Numeric Suffixes . . . . . . . . . . . . . . .
3-8. Programming Language Comparison . . . . . .
4-l.
Fuse Part Numbers . . . . . . . . . . . . . .
Command Table . . . . . . . . . . .
Removed Contents . . . . . . . . . . . . .
Attached Contents . . . . . . . . . . . . .
8360/8340/8341 . . . . . . . . . . . . . .
1-21 1-71
l-82 l-87
C-8 D-7
s-19
3-6 3-7
3-10
3-13
3-15
3-20 3-24 3-25
4-4
HP 8360 User’s Handbook
Contents-17
Page 31

GETTING STARTED

1

What Is In This Chapter

Note
This chapter contains information on how to use the HP 8360 Series Synthesized Sweeper. The information is separated into three sections.
Basic
Advanced
Programming
If you are unpacking a new synthesizer, refer to the installation suggestions provided in the “INSTALLATION” chapter of this manual.
For the novice user unfamiliar with the HP 8360 Series Synthesized Sweepers. This section describes the basic features of the synthesizer.
For the user familiar with synthesizers, but not necessarily familiar with how to use the special
features of the HP 8360 series.
For the user wishing to program an HP 8360 Series Synthesized Sweeper. This section contains an introduction to Standard Commands
for Programmable Instruments language
(SCPI), Hewlett-Packard’s implementation of
IEEE-488.2-1987, and an introduction to the
Analyzer programming language.
Getting Started Introduction
l-1
Page 32
How To Use This
Chapter
To use this chapter effectively, refer to the tabbed section “Menu Maps”. Menu maps can be folded out to be viewed at the same time as the Getting Started information, as illustrated.
I ’ 1
Equipment Used In
Examples
The following table lists the equipment used in the operation examples shown in this chapter. You can substitute equipment, but be aware that you may get different results than those shown.

Equipment Used In Examples

Equipment Recommended
Model Numbers
Power Meter Power Sensor Power Splitter Oscilloscope mm-Wave Source Module HP Power Amplifier Coupler Detector
HP
436A/437B
HP
8485A
HP
11667B
HP
1740A
83556A
HP
8349B
HP
11691D
HP
8474D
l-2
Getting Started Introduction
Page 33
Getting Started Basic
Introducing the
HP 8360 Series
Synthesized
Sweepers
PACKARD
lMENU SELEU
The HP 8360 Series Synthesized Sweepers are high performance, broadband frequency synthesizers.
PRESET
Figure l-l. The HP
(PRESET)
through a brief self-test. In the following examples, unless stated otherwise, begin by pressing
83620A
initializes the front panel settings and runs the synthesizer
Synthesized Sweeper
(PRESET).
Getting Started Basic
l-3
Page 34

Display Area

ACTIVE ENTRY AND DATA DISPLAY AREA
I
-MESSAGE LINE
SOFTKEY
LABEL AREA
I
\
SOFTKEYS
Figure
Active Entry and Data Display Area: This area typically displays
the frequency and power information of the current instrument state. When data entry is expected, the synthesizer uses all or part of this area to record the entries. The active entry arrow indicates the active entry function and its current value.
Message Line: This line is used to display:
ALC level status. Unlock information.
Timebase
RF output status.
Softkey Label Area: This area displays the name of the softkey
directly below it.
l-2.
status.
Display
(-->)
l-4
Getting Started Basic
Softkeys: These keys activate the functions indicated by the labels
directly above them.
Page 35

Entry Area

All function values are changed via the rotary knob and/or keys of the entry area.
ENTRY ENTRY ON
ON/OFF
LED
\,
ENTRY
\
ARROW KEY’S
ROTARY KNOB
/
TERMINATOR
KMS
NUMERIC
ENTRY KEYS
NEGATlVE SIGN/
BACKSPACE
Figure
l-3.
Entry Area
The following are active only when the synthesizer expects an input.
(ENTRY ON/OFF): This key lets you turn off or on the active entry
area. Turning off the entry area after a value is entered prevents accidental changes.
ENTRY ON LED: This LED lights when the entry area is active. Arrow Keys: The up/down arrow keys let you increase or decrease
a numeric value. The left/right arrow keys choose a significant digit indicated by an underline.
Rotary Knob: The rotary knob increases or decreases a numeric
value. The rotary knob can be used in combination with the
left/right arrow keys to change the increment size.
Terminator Keys: After the numeric entry keys are used to enter a
value, these keys define the units.
Negative Sign/Backspace Key: If a data entry is in progress, this
key backspaces over the last digit entered, otherwise a negative sign is entered.
Numeric Entry Keys: These keys enter specific numbers in the
active entry area and must be followed by one of the terminator keys before the function value changes.
Getting Started Basic
l-5
Page 36
CW Operation and

Start/Stop Frequency Sweep

CW Operation

Start/Stop Frequency
Sweep
CW operation is one of the major functions of the synthesizer, and is
easy to do using front panel keys. In CW operation, the synthesizer
produces a single, low-noise, synthesized frequency. Try this example:
Press(CW)(iJ@(J@@@@(7J@IGHz).
Check the active entry area. It indicates:
-->
cw: 12345.678000 MHz
The data display area indicates CW operation and the frequency
that you entered. The ENTRY ON LED is lit and the green SWEEP
LED is off.
Try other frequencies. Experiment with the rotary knob and the arrow keys as alternate methods of data entry.
The synthesizer can sweep a frequency span as wide as the frequency range of the instrument, or as narrow as 0 Hz (swept CW). In start/stop sweep operation, the synthesizer produces a sweep from the selected start frequency to the selected stop frequency. For example:
Press [START) @ 0 @ @ Press
ISTOP) (7J (J (7J @ (GHz.
IGHz).
The data display area indicates the start frequency and the stop frequency. The green SWEEP LED is on (periodically off when sweep is retracing). Because this is the active function, the active entry area indicates:
-->
STOP FREQUENCY: 7890.000000 MHz
Any subsequent entries change the stop frequency. To change the start frequency, press (START), which remains the active function until you press a different function key.
1-6 Getting Started Basic
Page 37
dLETT
.ARO
K
-
,
,-
ENlRRy
INSWMENT
SWEEP LED
CW Operation
1.
Press
Icw).
2. Enter
value.
3. Press terminator key.
STATE
cw
START
STOP
SOURCE MODULE INTERFACE
Figure 1-4. CW Operation and Start/Stop Frequency Sweep
start/stop Frequency Sweep
1.
Press
@TiF).
2.
Enter value.
3. Press terminator key.
4. Press
5.
6. Press terminator key.
(FiSj.
Enter value.
Getting Started Basic 1-7
Page 38

Center Frequency/Span Operation

Center frequency/span is another way of establishing swept
operation. This is just a different way of defining sweep limits. As an
example of center frequency/span operation:
Press
m(7J
Press
ISPAN) (iJ (GHz).
The synthesizer is now sweeping from 3.5 to 4.5 GHz (to view these
figures, press either
area indicates the center frequency, as well as, the span. Notice that
the green SWEEP LED is on.
While span is the active function, try the rotary knob and arrow
keys. This symmetrical increase or decrease of the frequency span
about the center frequency is one reason that center frequency/span
swept operation is used instead of start/stop frequency sweep.
Another example illustrates the subtleties of center frequency/span.
Press
(?ZFiK) @ LGHz)
Press
1SPAjj) @ (GHz)
Notice that the center frequency changed. This is because the center frequency could not accommodate a span of 8 GHz without exceeding
the lower frequency limit of the synthesizer’s specified frequency
range. If the low or high frequency range limits are exceeded, the
inactive (center or span) function is reset. Experiment with the
rotary knob and the arrow keys as alternate methods of data entry.
IGHz).
(START)
or
(STOP),
then
m).
The data display
1-6 Getting Started Basic
Page 39
SWEEP LED
CENTER
SPAN
Figure 1-5. Center Frequency and Span Operation
Center Frequent y
1.
Press
(jCENTEji).
2. Enter value.
3. Press terminator key.
Span Operation
1.
Press
m.
2. Enter value.
3. Press terminator key.
Getting Started Basic
l-9
Page 40

Power Level and Sweep Time Operation

Power Level Operation

Sweep Time Operation

The synthesizer can produce leveled power for CW, swept frequency, or power sweep operation. The selected power level can range from
-20
dBm
(-110
For practice: Press (POWER LEVEL) I-] @ @ area shows:
-->
POWER LEVEL: -20.00
If the selected power level is beyond the range of the synthesizer, the closest possible power is shown in both the data display area and the active entry area. If the selected power level exceeds the maximum leveled power the synthesizer is able to produce, the unleveled message UNLVLED appears on the message line. Experiment with the rotary knob and the arrow keys as alternate methods of data entry.
In typical applications the sweep time can vary tremendously, from milliseconds in a network analyzer system, to more than a minute in thermistor-based power meter systems. For this example, refer to the
“MENU MAP” section.
Press
@%Ki) @ [GHz).
Press
ISTOP) @ (GHz).
Press
lsWEEP
dBm
for option 001 synthesizers) to
dBm
@ 0
(ZJ (,,,).
(dB(mL).
+25 dBm.
The active entry
Watch the green SWEEP LED, it blinks every 2.5 seconds. The LED blinks at each retrace.
For the fastest sweep speed for which all specifications are guaranteed, the synthesizer must be in automatic sweep time selection.
Refer to menu map 8. Press SWEEP
Select more
Select
Notice that the active entry area indicates:
When the synthesizer is in automatic sweep time selection, the active entry area displays AUTO along with the current sweep time. Faster sweep speeds than this are possible, turn the rotary knob counter-clockwise until the display no longer changes. Notice that AUTO is no longer displayed
SwpTime
-->
SWEEP TIME:
[MENU).
if3
.
Aut
0.
100.0
mSec
AUTO
l-10
Getting Started Basic
Page 41
L”pI
.
.,WLETT
PACKARO
/
SWEEP TIME
Power Level Operation
1. Press
2. Enter value.
3. Press
CPOWER
IdBo).
SWEEP LED
LEVEL).
POWER LEVEL
Figure 1-6. Power Level and Sweep Time Operation
Sweep Time Operation
1. Press &WEEP
2. Enter value.
3. Press terminator key.
TtME].
Getting Started Basic
l-l 1
Page 42
Continuous,
and Manual
Operation
Single, Sweep
Continuous sweep is the operation mode set when the synthesizer is preset. It simply means that when the synthesizer is performing a swept operation, the sweeps will continuously retrace until a different sweep mode is selected. To choose this sweep mode, press
To change from continuous sweep to single sweep operation, press
($?@.
and switch to the single sweep mode. This initial keystroke cause’s the synthesizer to switch sweep modes, but it does not initiate a single sweep. A second keystroke (press sweep. When the synthesizer is in single sweep operation, the amber LED above the key lights. When the synthesizer is actually performing a sweep in single sweep mode, the green SWEEP LED lights.
The manual sweep mode lets you use the rotary knob to either sweep
from the start frequency to the stop frequency or to sweep power.
Refer to menu map 8, SYSTEM. Press
(CONT).
This causes the synthesizer to abort the sweep in progress
(PRESET).
sweep-retrace-sweep-
(%jGFJ
initiates a single
Press SWEEP Select Manual
The active entry area displays:
-->
SWEPT MANUAL: XXXXXXXXX MHz
Use the rotary knob to sweep from the start to the stop frequency. The green SWEEP LED is off in manual sweep mode because the sweeps are synthesized.
(j).
Beep.
1-12 Getting Started Basic
Page 43
SWEEP
LED
SINGLE CONT
Single Sweep
1. Press (SINGLE).
SWEEP MENU
Figure 1-7. Continuous, Single, and Manual Sweep Operation
Continuous Sweep
1. Press
c-1.
Manual Sweep
1. Press SWEEP (MENU).
2. Press Manual Sweep
3. Use the rotary knob to adjust frequency.
Getting Started Basic 1-13
Page 44

Marker Operation

The synthesizer has five frequency markers that can be used as fixed
frequency “landmarks,” or as variable frequency pointers on a CRT
display. To view the marker features of the synthesizer on a CRT, connect the synthesizer as shown in Figure 1-8.
Refer to menu map 2, FREQUENCY. Press
[PRESET). Press (START) @ Press (STOP) 0 Press
[MARKER).
Select Marker Ml and enter @
@&).
[GHz).
(GHz.
The synthesizer is sweeping from 3 to 7 GHz, with a 100 ms sweep speed. A frequency marker is set at 4 GHz, which causes an intensified dot to appear on the CRT. To obtain an amplitude spike at that frequency, select
Amp1
Markers . Notice that you can set the amplitude of the spike with the rotary knob or entry keys. To return to the intensified dot representation, select
Amp1
Markers (asterisk
off).
Caution
Amplitude markers increase the output power at the marker frequency. Provide protection to devices that could be damaged.
For a second marker, select Marker M2 and enter @ 0 @
(GHz).
This process can be continued for all five markers. Note that the marker displayed in the active entry area is “active” and can be controlled by the rotary knob, arrow keys, and numeric entry keys.
Once the Ml and M2 markers are established, the marker sweep function, start/stop frequencies to those of markers Ml and
Ml--M2
softkey MI--M2
Sweep.
Notice that the synthesizer now is sweeping from
Sweep, temporarily changes the original
M2.
Select
4 to 5.5 GHz. Use this function to focus in on a selected portion of the frequency sweep. Select
Ml--M2
Sweep again. This turns the function off and returns the synthesizer to its original sweep parameters. To change the start/stop frequencies for the synthesizer,
not just temporarily, use the
softkey Start-Ml
Stop'M2
.
As an example of the delta marker function:
Select Marker M3 and enter @ 0 0
IGHz).
Select Delta Marker.
1-14 Getting Started Basic
The frequency difference between marker 3 and marker 1 is displayed, and the CRT trace is intensified between the two markers. The active entry area displays:
-->
DELTA MKR
(3-l)
: 2700.000000 MHz
Page 45
Marker 1 was chosen because it is selected as the delta marker reference. To change reference markers, select Delta Mkr Ref .
Select M2 as the reference. Watch the display change to indicate:
-->
DELTA MKR (3-2) : 1200.000000 MHz
You can choose any of the five markers as a reference, but when delta marker is on, if the reference marker has a frequency value higher than the last active marker, the difference between the frequencies is negative and is displayed as such by the synthesizer. The CRT display continues to intensify the difference between the two markers.
When delta marker is showing in the active entry area, the ENTRY area is active. Rotate the rotary knob and watch the frequency difference change. The last active marker (in this case, marker 3) changes frequency value, not the reference marker.
OSCILLOSCoPE
Marker Operation
1. Press
2. Select a marker key (
@i?GGiZ).
3. Enter value.
4. Press terminator key.
Ni
. . MS
Figure 1-6. Marker Operation
Delta Marker Operation
1. Press
).
2. Select a marker key ( #I . . NE
3. Enter value.
4. Press terminator key.
5. Select a different marker key (Ml . . . MS).
6. Enter value.
7. Press terminator key.
8. Select Delta
9. Select one of the previously chosen markers.
10. Press
11. Select Delta Marker
@ZiGiF].
(jZi?%J
l4kr
).
Ref .
Getting Started Basic 1-15
Page 46
Saving and Recalling an
Instrument State
The save/recall registers store and access a previously set instrument state. For example, set the synthesizer to sweep from 3 to 15 GHz at a -10 dB power level, with markers 1 and 2 set at 4.5 and 11.2 GHz.
Press [START) Press (STOP) Press (POWER
(7J (GHz).
(7J (?J (GHz.
LEVEL]
I-]
(iJ (TJ 0).
Press (MARKER). Select Marker Ml @ 0 @
IGHz).
Select Marker M2 0 0 0 @
(GHzl.
To save this instrument state in register 1, press (SAVE) (iJ. To verify that the synthesizer has saved this state:
Press
(PRESET). Press (RECALL) Press
[MARKER).
(iJ.
The active entry area displays:
-->
RECALL REGISTER: 1 RECALLED
Notice the sweep end points, power level, and the asterisks next to
the marker 1 and 2 key labels.
You can save instrument states in registers 1 through 8. Register 0 saves the last instrument state before power is turned off. When power is turned on, register 0 is automatically recalled.
1-16 Getting Started Basic
Page 47
RECALL
Figure 1-9. Saving and Recalling an Instrument State
Save
1. Setup synthesizer as desired.
2. Press
3. Press
[SAVE. a number
1
through
Recall
1. Press
2. Press
8.
@EGiIiJ.
a number
0 through
8.
Getting Started Basic
l-17
Page 48

Power Sweep and Power Slope Operation

Power Sweep Operation

The power sweep function allows the power output to be swept (positive or negative) when the synthesizer is in the CW frequency mode. The power output of the synthesizer determines the maximum leveled power sweep that can be accomplished. For this example refer to the “Menu Map” section.
Zero and calibrate the power meter. Connect the instruments as shown in Figure Press
@ @ [GHz).
Press (POWER Press (SWEEP @ Set the power meter to dB[REF] mode.
The synthesizer is ready to produce a 4 GHz CW signal at 0 power out, with a 2 second sweep rate whenever a single sweep is executed. The power meter is ready to measure the power level relative to a starting point of 0
Press POWER Select Power Sweep and enter @
Press
[$iKKJ.
LEVEL) (TJ (YiJ.
(,,,)
[SINGLE).
(MENU).
dBm.
(dB0)
l-10.
dBm
(asterisk on).
Watch the relative power indication on the power meter. At the end
of the sweep the power meter indicates +7 dB. The active entry area on the synthesizer indicates:
-->
POWER SWEEP:
Now enter @ @ function).
This time the power meter indicates less than the power sweep requested. Note that the synthesizer is unleveled, UNLVD. This
happens because the synthesizer’s output power at the start of the
sweep is 0 dB and the requested power sweep takes the synthesizer
beyond the range where it is able to produce leveled power. The range of the power sweep is dependent on the ALC range and can be offset if a step attenuator (Option 001) is present.
Select Power Sweep to turn this function off (no asterisk). Press
[POWER LEVEL) !_) Q (FJ
On the power meter, press dB[REF] to reset the reference level.
7.00 dB/SWP
(dB(m))
(power sweep is still the active entry
1-18 Getting Started Basic
Page 49
Select Power Sweep (asterisk on). Press
(SINGLE].

Power Slope Operation

The synthesizer performs a power sweep beginning at -20 ending at f5
dBm.
The power meter indicates
+25
dB.
dBm
and
This function allows for compensation of high frequency system or cable losses by linearly increasing the power output as the frequency increases. For this example refer to the “Menu Map” section.
Press Power Slope , the active entry area displays:
-->
RF SLOPE:
X. XX dB/GHz, where X is a numeric value.
Power slope is now active, notice that an asterisk is next to the key
label.
Use the entry keys, rotary knob, or arrow keys to enter a value for the linear slope.
Press Power Slope again to turn this feature off.
Getting Started Basic
l-19
Page 50
SYNTHESIZER
‘UT
POUER
IlETER
Power Sweep
1. Press POWER
2. Select
3. Enter a value.
4. Press terminator key.
Pouer Saeep
(jMENU).
Figure
.
l-10.
Power Sweep and Power Slope Operation
Power Slope
1. Press POWER
2. Select Power Slope
3. Enter a value.
4. Press terminator key.
(jj).
l-20
Getting Started Basic
Page 51
Advanced

Getting Started Advanced

This section of Chapter 1 describes the use of many of the unique
features of the HP 8360 Series Synthesized Sweepers. The format
used is similar to the one used on the previous pages. When referred to a menu map number, go to the Menu Map tab and unfold the menu map so that you can view it together with the text.
Some menus have more than one page of softkeys. Select the
more m/n softkey to view the next page of softkeys. more m/n is
not included in the keystrokes given in these procedures.
Table l-l. Keys
Paragraph Heading
Externally Leveling the Synthesizer
Working with Mixers/Reverse Power Effects
Working with Spectrum Analyzers/ Reverse Power Effects
3ptimizing
Synthesizer Performance
Under
Discussion in This Section
Keys
Leveling Point Coupling Factor
POWER LEVEL
Set
Leveling Point Pwr Mtr Range
Leveling Point Module Mdl Lev Menu
Uncoupl Atten
Leveling Mode
Leveling Mode Leveling Mode Search Fltness Menu Delete Menu Auto Fill Start
Auto Fill Stop
Auto Fill Mtr
FLTNESS ON/OFF
Enter Freq
Enter Corr
Freq Follow
List Menu
Copy List
Sweep Mode List
Atten
Meas
Menu
ExtDet
PwrMtr
Normal
ALCoff
Incr
Ext Det Cal
Getting Started Advanced
1-2 1
Page 52
Advanced
Table l-l.
Keys Under Discussion in This Section (continued)
Paragraph Heading
Optimizing Synthesizer Performance continued
Using Step Sweep
Creating and Using a Frequency List
Using the Security Features
Changing the Preset Parameters
Keys
Auto Track Peak RF Always Peak RF Once
Sap Span Cal Once Sap Span Cal Always
AM BW Cal Always AM BW Cal Once
FullUsr
AM On/Off
Cal
100%/V
AM On/Off IOdB/V Deep AM
USER DEFINED MENU ASSIGN
Step Sap Menu List Menu
Delete
Menu
Enter List Freq Enter List Offset Enter List Dwell Pt Trig Menu Zero Freq Save Lock Clear Memory Blank Display Save Usr Preset Preset Mode User
PRESET
1-22 Getting Started Advanced
For more
information,each
of these keys has
a separate entry in
“OPERATING and PROGRAMMING REFERENCE” chapter of
this handbook.
the
Page 53
Externally Leveling the Synthesizer
In externally leveled operations, the output power from the synthesizer is detected by an external sensor. The output of this detector is returned to the leveling circuitry, and the output power is automatically adjusted to keep power constant at the point of detection.
Leveling with
Detectors/Couplers
/Splitters
Figure l-11 illustrates a typical setup for external leveling. When externally leveled, the power level feedback is taken from the external
negative detector input rather than the internal detector. This feedback voltage controls the ALC system to set the desired RF output. Refer to Figure A-l in Chapter 2, for a block diagram of the synthesizer’s ALC circuitry.
SYNTHESIZER
NEGATIVE
Figure
l-1
1. ALC Circuit Externally Leveled
LEVELED OUTPUT
b
Getting Started Advanced 1-23
Page 54
To level externally:
1. Setup the equipment as shown. For this example, the detector/coupler setup is used.
2. Refer to menu map 1.
3. Press
4.
Select Leveling Point
5.
Set the coupling factor. Select Coupling Factor
(ALC).
(dB(m)).
ExtDet
.
c-) @ @
Note
Hint
Power splitters have a coupling factor of 0 dB.
Figure 1-12 shows the input power versus output voltage characteristics for typical HP diode detectors. From the chart, the leveled power at the diode detector input resulting from any
external level voltage setting may be determined. The range of power
adjustment is approximately -30
Automatically characterize and compensate for the detector used by performing a detector calibration. Refer to “Optimizing Synthesizer Performance, Using Detector Calibration,” later in this section.
dBm
to i-18
dBm.
l-24
Getting Started Advanced
Page 55
100
mV
iii
SQUARE LAW ASYMPTOTE
+20 d6V
+lO dBV
0
dBV
-10
-20
-30
dBV
dBV
dBV
10
1
.l
mV
mV
mV
DETECTOR INPUT POWER,
Figure 1-12. Typical Diode Detector Response at 25°C
dBm
-40
-50
-60
.-66
-70
-60
dBV
dBV
dBV
dBV
dBV
dBV
Getting Started Advanced
l-25
Page 56
External Leveling Used With the Optional Step Attenuator
Some external leveling applications require low output power from the synthesizer. The synthesizer automatically uncouples the attenuator from the ALC system for
Press (POWER LEVEL). Note the display. It shows:
all
external leveling points.
Hint
--> ATTEN
For example, leveling the output of a 30 dB gain amplifier to a level of -10
-40 range of the ALC modulator alone. If so, the LOW UNLVLED warning message is displayed. Inserting 40 dB of attenuation results in an ALC level of 0 20 GHz,30 dB attenuation is a better choice as it results in an ALC level of -10 that vary the power level.
For optimum display accuracy and minimum noise, the ALC level should be greater than -10 attenuation equal to the tens digit of output power. Example: desired output power = -43 dBm; use:
1. Press POWER
2.
To obtain flatness corrected power refer to “Optimizing Synthesizer Performance, Creating and Applying the User Flatness Correction
Array,” later in this section.
dBm
dBm
--> ATTEN:
Select Set
0 dB, POWER LEVEL: 0.00
requires the output of the synthesizer to be around
when leveled. At some frequencies this level is beyond the
dBm,
which is well within the range of the ALC. At
dBm.
This gives a margin for AM or other functions
dBm.
This is achieved by using
40 dB , ALC -3
(MENU).
Atten @ @ 0).
dBm
dBm
l-26
Getting Started Advanced
Page 57
Leveling with Power
Meters
Leveling with a power meter is similar to leveling with a diode
detector. Figure 1-13 shows the setup for power meter leveling.
SYNTHESIZER
POUER HETER
Figure 1-13. Leveling with a Power Meter
Hint
1. Set up the equipment as shown. Be sure to set the power meter to manual range mode and note the range.
2. Refer to menu map 1.
3. Press
4.
Select
5.
Select Pwr
a).
Leveling
Mtr
Point
Range.
PwrHts
.
Enter the range value set for the power
meter as noted in step 1.
6. Select Coupling Factor , press
@‘J (de(m)).
Unlike detector leveling, power meter leveling provides calibrated
power out of the leveled RF port.
To obtain flatness corrected power refer to “Optimizing Synthesizer Performance, Creating and Applying the User Flatness Correction Array,” later in this section.
Getting Started Advanced
l-27
Page 58
Leveling with MM-wave
Source Modules
Millimeter-wave source module leveling is similar to power meter leveling. The following figures illustrate the setups for leveling with a mm-wave source module.
SYNTHESIZER
Figure 1-14. MM-wave Source Module Leveling
High power model synthesizers can externally, level mm-wave source modules to maximum specified power without a microwave amplifier.
1-28 Getting Started Advanced
Page 59
RF OUT
AORPTER
(IF REQUIRED)
6
RF
IN
nICROUAVE
AWPLIFIER
0
-0
RF
OUT
I’ll-LINE
SOURCE
4
Figure 1-15. MM-wave Source Module Leveling Using a Microwave Amplifier
1. Set up the equipment as shown.
2. Refer to menu map 1.
3. Select Leveling Point Module.
4.
Select Mdl Lev Menu.
5.
Select Module Leveling Pt Auto or Front or Rear, depending
on where the interface connection is made.
All of the ALC data necessary to communicate properly with the synthesizer is exchanged via the SOURCE MODULE INTERFACE.
NODULE
Hint
To obtain flatness corrected power refer to “Optimizing Synthesizer Performance, Creating and Applying the User Flatness Correction Array,” later in this section.
Getting Started Advanced 1-29
Page 60
Working with
Mixers/Reverse
Power Effects
Note
Uncoupled operation applies to Option 001 synthesizers only.
Uncoupled operation is useful when working with mixers. Figure 1-16 shows a hypothetical setup where the synthesizer is providing a small signal to a mixer. The synthesizer output is -8 which in Leveling Node ALC Level = -8
dBm.
Normal
results in
ATTEN =
The mixer is driven with an LO of
dBm,
0
dB,
+lO dBm,
and has LO to RF isolation of 15 dB. The resulting LO feedthrough of -5
dBm
enters the synthesizer’s OUTPUT port, goes through the attenuator with no loss, and arrives at the internal detector. Depending on frequency, it is possible for most of this energy to enter
the detector. Since the detector responds to its total input power regardless of frequency, this excess energy causes the leveling circuit to reduce its output. In this example the reverse power is actually larger than the ALC level, which may result in the synthesizer output being shut off.
Figure 1-17 shows the same setup, with uncoupled operation used to produce the same -8 ALC Level = +2
dBm
output. In this case,
dBm.
The ALC level is 10 dB higher, and the
ATTEN
= -10
dB,
attenuator reduces the LO feedthrough by 10 dB. Thus the detector sees a
-l-2 dBm
desired signal versus a possible -15
dBm
undesired signal. This 17 dB difference results in a maximum 0.1 shift in the synthesizer output level. To set the synthesizer to the attenuator uncoupled mode as discussed in this example, do the following:
.
I-30
Getting Started Advanced
1. Press POWER
2. Select Set
Atten
(MENU).
and press
(iJ @J (dB(m)).
This step does two things, it uncouples the attenuator from the rest of the ALC system, and it lets you set an attenuator value, in this case, 10 dB.
3. Press [POWER
LEVEL)
(?J (j-l).
This sets the ALC level to
+2 dBm.
For more information on the ALC or setting power level, refer to
IALC)
or
(POWER LEVEL)
in Chapter 2.
Page 61
swrNEsl2ER WlTN OPflON Do1
MEASURES -8
DETECTOR
MC
LEVEL
Figure
ig-y-
dBm
l-16.
Reverse Power Effects, Coupled Operation with
sYNTNEsl2ER WITH OPTlON
DETECTOR
MUISURES REVERSEPOWER
001
-5
dBm
RF OUTPUT
I
-5dBm
-6dBm
MIXER
IF
Output
Q
LO
Ll%EL
= +lO
LO
dBm
RF LEVEL
CONTROL
MEASURES +2
DETECTOR
dBm
MC
LmEL
Figure 1-17. Reverse Power Effects, Uncoupled Operation with
MC LEVEL
-
+2
dBm
_
,, Q
iI 0
I
ATTENUATOR
10 dB
DETECTOR MEASURES -15 REVERSE POWER
d&n j
I
I
-5dBm
I I
I
-6dBm
Getting Started Advanced
IF
Output
= +lO
dBm
l-3
1
Page 62

Working with Spectrum Analyzers/Reverse Power Effects

Reverse power is a problem with spectrum analyzers that do not
have preselection capability. Some analyzers have as much as
+5 dBm
frequencies. The effects of reverse power are less in the heterodyne band (0.01 to 2.3 GHz)
broadband matching. Similarly, at frequencies above 2.3 GHz, reverse
power that is within 10 MHz of the synthesizer’s frequency may be
partially absorbed by the YIG filter. If the frequency difference is
small enough to be within the leveling system bandwidth (typically
10 kHz CW, 200 kHz sweep or AM), the effect of reverse power is amplitude modulation of the synthesizer’s output. The AM rate equals the difference in RF frequencies. Reverse power problems may be treated by using the unleveled mode. There are two unleveled modes, ALC off and search.
To set the synthesizer to the ALC off mode:
1. Refer to menu map 1.
2. Press
3.
Select Leveling Mode
In this mode, the synthesizer provides RF power with no ALC correction and therefore requires a power meter to set a particular power.
LO feedthrough coming out of their RF input, at some
where the power amplifier provides some
(ALC).
ALCoff
.
To set the synthesizer to the search mode:
1. Press
2.
Select Leveling Mode Search.
In this mode, the synthesizer is in the normal ALC mode until the
desired power level is reached, then the ALC is disconnected.
(ALC.
1-32 Getting Started Advanced
Page 63

Optimizing Synthesizer Performance

Creating and Applying
the User Flatness
Correction Array
The following examples demonstrate the user flatness correction
feature:
1. Using an HP
data for a swept 4 to 10 GHz measurement.
2. Manually entering correction data for a stepped (List Mode) measurement.
3. Making swept mm-wave measurements, automatically entering correction data for an arbitrary list of correction frequencies.
4. Making scalar analysis measurements with automatically-entered correction data that compensates for power variations at the output of a directional bridge.
Each example illustrates how to set up correction tables for a different measurement requirement. Modify the instrument setups shown to suit your particular needs. Completed correction tables may be easily edited if more correction data is required for your measurement. Additional correction frequencies may be added
by using the auto fill feature or by entering correction frequencies individually. The auto fill feature adds but does not delete correction frequencies.
There are two basic front-panel methods of creating a flatness
correction array. The first and quickest method is to use an HP power meter. Refer to Figure 1-18 for the setup. The second method is just as accurate, but requires a little more interaction between the operator and the instruments. Figure 1-19 shows the setup for the
second method.
437B
power meter to automatically enter correction
437B
Getting Started Advanced 1-33
Page 64
Creating a User Flatness Array Automatically, Example 1
In this example, a flatness array containing correction frequencies
from 4 to 10 GHz at 1 GHz intervals is created. An HP
438B
power
meter controlled by the synthesizer through the interface bus is used
to enter the correction data into the flatness array. For this example, refer to menu map 5, POWER.
1. The equipment setup shown in Figure 1-18 assumes that if the setup has an external leveling configuration, the steps necessary to correctly level have been followed. If you have questions about external leveling refer to earlier paragraphs titled, “Externally Leveling the Synthesizer.”
Setup Power Meter
2. Zero and calibrate the power meter/sensor.
3. Enter the appropriate power sensor calibration factors into the power meter.
4. Enable the power meter/sensor cal factor array. For operating
information on the HP
437B
power refer to its operating and
service manual.
5. Connect the power sensor to the point where corrected power is
desired.
HP
$778
TER
M
------B-B
I
COmxTEo
,
OUTFIJT
PORT
------
POUER
POYER NE
SENSOR
Figure 1-16. Creating a User Flatness Array Automatically
1-34 Getting Started Advanced
Page 65
Setup Synthesizer Parameters
On the synthesizer, press (PRESET).
6.
7.
FREQUENCY
8.
(POWER LEVEL) (TJ
Access User Flatness Correction Menu
ISTART) @ LGHz), LSTOP) 0 @ LGHz).
m.
Press POWER
9.
10.
Select Delete Menu Delete All . This step insures that the
flatness array is empty.
11.
Press
(6%).
soft key menu.
12
Enter the frequency points at which the correction information will be taken. Choose either the point-by-point entry method
Enter Freq or the automatic frequency point generation
Auto Fill Start.
[MENU).
Leave the delete menu and return to the previous
Select
For this example,select
Fitness
Menu.
Auto Fill Start
@IGHz).
13
Select Auto Fill Stop
Notice that a frequency list starting at 4 and ending at 10 GHz with an increment value of 1 GHz is created.
Enter Correction Data into Array
14.
Select
Mtr
Meas
Menu Measure
is now under synthesizer control and is performing the sequence of steps necessary to generate the correction information at each frequency point.
a@=,
Auto Fill
Cars
All.
Incr a[GHz).
The power meter
If an HP-IB error message is displayed verify that the interface
connections are correct. Check the HP-IB address of the power meter and ensure that it is the same address the synthesizer is using (address 13 is assumed). Refer to the menu map 8, System,
for the key sequence necessary to reach softkey Meter Adrs .
Enable User Flatness Correction
When the operation is complete, (a message is displayed) the
15.
flatness correction array is ready to be applied to your setup.
Disconnect the power meter/sensor and press [FLTNESS
(amber LED on). The power produced at the point where the
power meter/sensor was disconnected is now calibrated at the frequencies and power level specified above.
Getting Started Advanced
ON/OFF)
l-35
Page 66
Creating a User Flatness Array, Example 2
This example shows how to use the synthesizer and a power meter in manual entry mode. This example also introduces two features of the synthesizer. The
process and the frequencies.
The frequency follow feature automatically sets the source to a CW test frequency equivalent to the active correction frequency in the user flatness correction table. The front panel arrow keys are used to move around the correction table and enter frequency-correction pairs. Simultaneously, the synthesizer test frequency is updated to the selected correction frequency without exiting the correction table.
To further simplify the data entry process, the synthesizer allows you to enter correction data into the user flatness correction table by adjusting the front panel knob until the desired power level is displayed on the power meter. The user flatness correction algorithm automatically calculates the appropriate correction and enters it into the table. If you already have a table of correction data prepared, it can be entered directly into the correction table using the front-panel keypad of the synthesizer.
softkey
softkey
Freq Follow simplifies the data entry
List Mode sets up a list of arbitrary test
With the list mode feature, you may enter the test frequencies into a
table in any order and specify an offset (power) and/or a dwell time
for each frequency. When list mode is enabled, the synthesizer steps
through the list of frequencies in the order entered.
The user flatness correction feature has the capability of copying and entering the frequency list into the correction table. Since the offset in the list mode table is not active during the user flatness correction data entry process, the value of the correction data is determined as if no offset is entered. When user flatness correction and list mode
(with offsets) are enabled, the synthesizer adjusts the output power by an amount equivalent to the sum of the correction data and offset for each test frequency. You must make sure that the resulting power level is still within the ALC range of the synthesizer.
1-36 Getting Started Advanced
Page 67
Figure 1-19. Creating a User Flatness Array
For this example, refer to menu map 5, POWER.
1.
The equipment setup shown in Figure 1-19 assumes that if your setup has an external leveling configuration, the steps necessary to correctly level have been followed. If you have questions about external leveling refer to earlier paragraphs titled, “Externally Leveling the Synthesizer
.”
Setup Power Meter
2.
Zero and calibrate the power meter/sensor. Connect the power sensor to the point where flatness corrected
3.
power is desired.
Setup Synthesizer Parameters
4.
On the synthesizer, press (PRESET).
(POWER LEVEL)
5.
dBm (PO
Create A Frequency List
6.
On the synthesizer, press FREQUENCY
7.
Select List Menu Enter
@ 0).
max -
This sets the test port power to
Ppath loss).
(E).
List Freq @ (GHz).
This enters 5
GHz as the first frequency in the list array. Entering a frequency
automatically sets the offset to 0 dB and the dwell to 10 ms. Enter 18, 13, 11, and 20 GHz to complete this example array.
8.
Getting Started Advanced 1-37
+5
Page 68
Access User Flatness Correction Menu
9. Press POWER
10.
Select Delete Menu Delete All. This step insures that the
(z).
Select
Fitness
Menu.
flatness array is empty.
11. Press
(=I.
Leave the delete menu and return to the previous
soft key menu.
12. Select Copy List This step copies the frequency list into the correction table in sequential order.
13. Select Freq Follow. This sets the synthesizer to CW frequency mode to facilitate taking correction information. As you scroll through the correction cells, the synthesizer produces the corresponding CW frequency at 0
dBm.
14. Select Enter Corr . This allows correction value entry.
15. Press
(FLTNESS ON/OFF).
This step enables user flatness correction.
16. For 5 GHz, set the appropriate power sensor cal factor on the power meter.
17. Use the synthesizer rotary knob to adjust for a measurement of
0.00
dBm
on the power meter. Notice that a correction value is
entered at 5 GHz.
18. Use the up arrow key to increment to the next correction cell.
19. For 11 GHz, set the appropriate power sensor cal factor on the power meter.
20. Use the synthesizer rotary knob to adjust for a measurement of
0.00
dBm
on the power meter.
21. Repeat this sequence of steps until all the frequency points have a correction value entered.
Activate List Mode
22. Press SWEEP
(MENU].
Select Sweep Mode List .
23. The flatness correction array is ready to be applied to your setup. Disconnect the power meter/sensor. The power produced at the point where the power meter/sensor was disconnected is now calibrated at the frequencies and power level specified above.
l-38
Getting Started Advanced
Page 69
Swept mm-wave Measurement with Arbitrary Correction Frequencies, Example 3
The focus of this example is to use user flatness correction to obtain flat power at the output of the HP 83550 series mm-wave source modules. In this case we will use non-sequential correction
frequencies in a swept 26.5 to 40 GHz measurement with an
HP 83554 source module. The time it takes for a large quantity of power meter measurements
can be long, therefore, we selected non-sequential correction
frequencies to target specific points or sections of the measurement
range that we assume are more sensitive to power variations. This
greatly expedites setting up the user flatness correction table. The
amount of interpolated correction points between non- sequential correction frequencies varies. This example uses automatically enter correction data into the array.
theHP 437B
to
Note
Turn off the synthesizer before connecting to the source module
interface (SMI)
cable, or damage may result.
Getting Started Advanced 1-39
Page 70
SYNTHESIZER
SYNTHFSIZER
HICROURVE RNPLIFIER
HP 4378
POUER flETER
HP
‘l37B
POULR NFTFR
l-40
Getting Started Advanced
Creating
Figure
l-20.
Arbitrarily Spaced Frequency-Correction Pairs in a Swept mm-wave
Environment
For this example, refer to menu map 5, POWER.
1. The equipment setup shown in Figure
l-20
assumes that you have followed the steps necessary to correctly level the configuration. If you have questions about external leveling refer to earlier paragraphs titled, “Externally Leveling the Synthesizer.”
Setup Power Meter
2. Zero and calibrate the power meter/sensor.
3. Connect the power sensor to test port.
4. Enter and store in the power meter, the power sensor’s cal factors for correction frequencies to be used.
Page 71
Note
U, V, and W-band power sensors are not available from Hewlett-Packard. For these frequencies use the Anritsu ML83A Power Meter with the MP715-004 (40 to 60 (50 to 75
GHz),
or the MP81B (75 to 110 GHz) power sensors. Since
GHz),
the MP716A
the Anritsu model ML83A Power Meter is not capable of internally storing power sensor cal factors, you must manually correct the data entry. Refer to example 2 for information on manual entry of correction data.
Setup Synthesizer Parameters
Turn on the synthesizer and press
5.
[PRESET).
The following occurs:
n The source module’s frequency span is displayed on the
synthesizer.
n The synthesizer’s leveling mode is automatically changed
internal to “module leveling”.
n The source module’s maximum specified power is set and
displayed.
6.
Press FREQUENCY (START) @ @ 0 @
IGHz).
The frequency sweep is set from 26.5 to 40 GHz.
7.
Press
(POWER LEVEL)
+7 dBm
Access User Flatness Correction Menu
Press POWER
8.
9.
Select Delete Menu Delete All. This step insures that the
for maximum power to the device under test.
CE).
0 @.
The source module power is set to
Select Fltness Menu.
(GHz, m@
flatness array is empty.
Press
10.
softkey
(PRIOR).
menu.
Leave the delete menu and return to the previous
from
@
11.
Select Enter Freq
(!?J @ 0 @ (GHz),
to enter 26.5 GHz as
the first correction frequency. Enter 31, 32.5, and 40 GHz to
complete the list. Notice that the frequencies are arbitrarily
spaced.
Enter Correction Data into Array
12.
Select Mtr
Meas
Menu Measure
Corr
All. The power meter
is now under synthesizer control and is performing the sequence of steps necessary to generate the correction information at each frequency point.
If an HP-IB error message is displayed verify that the interface
connections are correct. Check the HP-IB address of the power
meter and ensure that it is the same address the synthesizer is
Getting Started Advanced 1-41
Page 72
using (address 13 is assumed). Refer to the menu map 8, System, for the key sequence necessary to reach
Enable User Flatness Correction
softkey
Meter Adrs .
13. When the operation is complete, (a message is displayed) the flatness correction array is ready to be applied to your setup.
14. To save the synthesizer parameters including the correction table in an internal register, press
ISAVE) 0.
(n = number 1 through 8).
15. Disconnect the power meter/sensor and press
(FLTNESS ON/OFF)
[amber LED on). The power produced at the point where the power meter/sensor was disconnected is now calibrated at the
frequencies and power level specified above.
l-42
Getting Started Advanced
Page 73
Scalar Analysis Measurement with User Flatness Corrections, Example 4
The following example demonstrates how to setup a scalar analysis measurement (using an HP 8757 Scalar Network Analyzer) of a 2 to 20 GHz test device such as, an amplifier. User flatness correction
is used to compensate for power variations at the test port of a
directional bridge. Follow the instructions to set up the synthesizer,
then configure the system as shown in Figure 1-21.
Note
The synthesizer’s rear panel language and address switches must be set to 7 and 31 (all l’s), to change the language or address of the synthesizer from the front panel. The programming language must be
set to Analyzer. Refer to menu map 8, System, to find the location of softkey Programming Language Analyzer (asterisk on = active language).
SIONRL
SYNTHESIZER
NETIN-’
DETECTORDETECTOR
OIRECTIONRL
I”_
SCF
ILRR
RNRLYZEN
I
Figure
DETECTOR
l-2
1. Scalar System Configuration
Example Overview
In this example you use an HP
437B
power meter to automatically enter correction data into the array. It is necessary to turn off the HP 8757 System Interface (controlled from the front-panel of the analyzer) so that the synthesizer can temporarily control the power meter over HP-IB . When the correction data entry process is complete, enable user flatness correction and set the desired test port power level. Then store the correction table and synthesizer configuration in the same register that contains the analyzer configuration. Re-activate the HP 8757 System Interface and recall
Getting Started Advanced 1-43
Page 74
the stored register. Make sure that user flatness correction is still enabled before making the measurement.
Note
When an HP
437B
power meter is used to automatically enter the
correction data, the correction calibration routine automatically turns
off any active modulation, then re-activates the modulation upon
the completion of the data entry process. Therefore, the scalar pulse modulation that is automatically enabled in a scalar measurement system is disabled during an HP
437B
correction calibration.
The user flatness correction array cannot be stored to a disk. You must make sure that the array is stored in one of the eight internal registers. Recalling a file from an HP 8757 disk will not erase the current array; therefore you may recall an array from an internal register, then recall an associated file from a disk.
For this example, refer to menu map 5, POWER.
1. The equipment setup shown in Figure 1-21 assumes that you have followed the steps necessary to correctly level the
configuration. If you have questions about external leveling
refer to earlier paragraphs titled, “Externally Leveling the
Synthesizer
.”
2. On the analyzer, press [PRESET). Reset the analyzer and
synthesizer to a known state.
Setup System Parameters
3. On the synthesizer, press FREQUENCY
(STOP)@@m.S
et t e s n
h yth
esizer for a frequency sweep of
(START) @ (GHz,
2 to 20 GHz.
4. Press [POWER
LEVEL)
0 @.
Where n = maximum available
power.
5. On the analyzer, set up the appropriate measurement
(i.e. gain for an amplifier). Calibrate the measurement (thru and short/open calibration). Press
ISAVE) (iJ
to store the analyzer’s
configuration and synthesizer parameters in storage register 1.
6. Turn off the HP 8757 System Interface. Use the analyzer
SYSINTF
ON OFF
softkey
found under the SYSTEM menu to
deactivate the system interface.
Access User Flatness Correction Menu
7. On the synthesizer, press POWER
Fltnesa
8.
Select Delete Menu Delete All . This step insures that the
Menu.
Cm).
Select
flatness array is empty.
1-44 Getting Started Advanced
Page 75
9. Press (PRIOR). Leave the delete menu and return to the previous soft key menu.
10. Select Auto Fill Start @
m).
Set the first frequency in
correction table to 2 GHz.
11. Auto Fill Stop @ @J
(GHz).
Set the last frequency in
correction table to 20 GHz.
12. Auto Fill
Incr 0 @ @ INIHz).
Set the frequency increment to
every 100 MHz from 2 to 20 GHz.
Setup Power Meter
13. Zero and calibrate the power meter/sensor.
14. Connect the power sensor to test port.
15. Enter and store in the power meter, the power sensor’s cal factors for correction frequencies to be used.
Enter Correction Data into Array
16. Select
Mtr
Meas
Menu Measure
Corr
All . The power meter is now under synthesizer control and is performing the sequence of steps necessary to generate the correction information at each frequency point.
If an HP-IB error message is displayed verify that the interface connections are correct. Check the HP-IB address of the power meter and ensure that it is the same address the synthesizer is using (address 13 is assumed). Refer to the menu map 8, System,
for the key sequence necessary to reach
Enable User Flatness Correction
17.
When the operation is complete, (a message is displayed) the
softkey
Meter Adrs .
flatness correction array is ready to be applied to your setup.
18. Disconnect the power meter/sensor.
19. On the synthesizer, press [POWER
n
= PO max -
Ppath
for maximum leveled power at the test
loss
LEVEL)
0 (dBm).
Where
port.
20. To save the synthesizer parameters including the correction table in an internal register, press
ISAVE) (XJ.
(n = number 1 through 8).
Reactivate the HP 8757 System Interface
21. Set the analyzer to SYSINTF ON, the analyzer and synthesizer
preset.
22. Press (RECALL) 0, Recall the synthesizer parameters from storage
register 1
Getting Started Advanced
l-45
Page 76
23. On the synthesizer, press [FLTNESS
ON/OFF)
(amber LED on). The
power produced at the point where the power meter/sensor was disconnected is now calibrated at the frequencies and power level specified above.
1-46 Getting Started Advanced
Page 77
Using Detector
Calibration
Detector calibration is useful for characterizing and compensating for negative diode detectors used in external leveling. Detectors may be characterized by three operating regions as shown in Figure 1-12: the square law, the linear, and the transition region. The following steps use an HP
437B
to automatically characterize the operating regions and use this information to automatically compensate for the detector being used. The equipment setup shown in Figure assumes that
.the
steps necessary to correctly externally level have
l-22
been followed. Refer to menu map 9, USER CAL.
HP
$378
POUER NETER
Figure
l-22.
Automatically Characterizing and Compensating for a Detector
1. Connect the power meter as shown.
2. Zero and calibrate the power meter/sensor.
3. Enter the appropriate power sensor calibration factors into the
power meter.
4. Enable the power meter/sensor cal factor array. For operating information on the HP
437B
power meter refer to its operating and service manual.
5. Connect the power sensor to the output of the coupler (or splitter).
6. On the synthesizer, set the power level and start/stop frequency information as desired.
7.
Press
[USERCAL).
8.
Select Ext Det Gal . The power meter is now under synthesizer
control and is performing the sequence of steps necessary to generate the compensation information.
Getting Started Advanced 1-47
Page 78
If an HP-IB error message is displayed verify that the interface
connections are correct. Check the HP-IB address of the power meter and ensure that it is the same address the synthesizer is using (address 13 is assumed). Refer to the menu map 8, System,
for the key sequence necessary to reach softkey Meter
A&s
9. When the operation is complete, (a message is displayed) disconnect the power meter/sensor. The synthesizer has stored
the compensation information in its memory and is using it to
calibrate the detector’s output voltage relative to power.
.
1-48 Getting Started Advanced
Page 79
Using the Tracking
Feature
Peaking
Peaking is the function that aligns the output filter (YTM) so that
its
passband
mode. Use peaking to obtain the maximum available power and
spectral purity, and best pulse envelopes, at any given frequency above 2.35 GHz (or 2 GHz, when 2 GHz is the minimum frequency specified). The YTM is inactive for the low band frequencies (10 MHz to 2.35 GHz).
To peak at the present CW frequency:
Press (USER). Select Tracking Menu Peak RF Once.
This causes an instantaneous execution of the peaking function. This is a one-time implementation of the peaking, where the function is turned on and then turned off.
To peak at the present CW frequency, and continue to peak at new frequencies as they are entered:
Press (USER). Select Tracking Menu Peak RF Always.
is centered on the RF output, in CW or manual-sweep
Note
If “peak always” is on (denoted by an asterisk next to the key label) for an extended period of time, the peaking function will automatically repeak every seven minutes.
Tracking
Auto track is a more extensive version of peaking. It causes all of the YTM tracking calibration constants to be aligned and requires approximately 40 to 90 seconds to complete. Tracking is performed from 2.35 GHz (or 2.0 GHz) to the end of the specified frequency range.
If the synthesizer does not have a step attenuator, terminate the RF OUTPUT with a good attenuator or a power sensor to prevent mistracking.
To enhance the power output and spectral purity of swept modes, and to improve tracking performance (especially in harsh environments having wide temperature variations):
Press
(USERCAL).
Select Tracking Menu Auto Track
500
impedance match such as a 10 dB
I
Getting Started Advanced 1-49
Page 80

ALC Bandwidth Selection

The ALC bandwidth defaults at factory preset to the auto selection
ALC Bandwidth Select Auto which selects the appropriate
bandwidth (high bandwidth selection, the synthesizer determines which functions are activated and uses the decision tree shown in Figure
or low) for each application. To make the
NO
1
M( on?
-or-
SEh?
-or-
List
Low’ SW
FlVCp.lK#
-or-
SOP
swamp7
NO
T
Low Bw
I
l-23.
Figure
l-23.
Decision Tree for ALC Bandwidth Selection
l-50
Getting Started Advanced
Page 81

Using Step Sweep

1. Refer to menu map 2.
2. Press FREQUENCY
3.
Select
Step Swp Menu.
[e].
4.
Select
Step Size.
5.
Select Step Points.
6.
Determine the dwell time desired, select Step Dwell and enter
a value, or choose the dwell time determined by the ramp mode sweep time, select
7.
Determine the triggering scheme, select
Bus, or Ext
8. Press SWEEP (MENU).
9.
Select Sweep Mode Step, to activate the step frequency mode.
Enter the desired increment value.
Enter the number of points desired.
ISwe
Coupled .
Step Swp Pt Trig Auto ,
Getting Started Advanced
l-5 1
Page 82
Creating and Using
a Frequency List
1. Refer to menu map 2.
2. Press FREQUENCY
3.
Select List Menu.
To use the frequency points of a frequency list to create the frequency portion of the user flatness correction array:
1. Refer to menu map 5.
2. Press POWER
3.
Select
4.
Select Copy List .
Fltnesa
(hnENU).
(‘MENU).
Menu.
1-52 Getting Started Advanced
Page 83

Using the Security Features

To access the security menu:
1. Refer to menu map 8.
2. Press SYSTEM
3.
Select Security Menu.
@K).
Getting Started Advanced
l-53
Page 84

Changing the Preset Parameters

1. Setup the synthesizer in the desired operation state to be used as the preset state.
2. Refer to menu map 8.
3. Press SYSTEM
4.
Select Save User Preset.
5.
Select Preset Mode User.
(e).
Whenever the the operation state setup and saved in steps 1 and 4. The synthesizer displays:
*** USER DEFINED PRESET RECALLED *** and also gives you the option of selecting the factory preset state by creating a factory preset
(PRESET)
key is pressed, the synthesizer will return to
softkey.
1-54 Getting Started Advanced
Page 85
Programming

Getting Started Programming

HP-IB, the Hewlett-Packard Interface Bus, is the
instrument communication system between the synthesizer and up to
14 other instruments. Any instrument having HP-IB capability can be interfaced to the synthesizer, including non-HP instruments that have “GPIB,”
(these are common generic terms for HP-IB; all are electrically equivalent although IEC-625 uses a unique connector). This portion of the manual specifically describes interfacing the synthesizer to one
type of instrument: a computer.
The first part of this section provides general HP-IB information.
Later, the Standard Commands for Programmable Instruments language (SCPI)
For information on programming in the Control Interface Intermediate Language (CIIL), refer to a separate option 700 manual supplement.
“IEEE-488,” “
is introduced, and example programs are given.
ANSI
MC1.l,”
instrument-to-
or “IEC-625” capability
Getting Started Programming
l-55
Page 86
HP-IB General
Information

Interconnecting Cables

Instrument Addresses

HP-IB Instrument
Nomenclature
Figure
suitable cables, and describes the procedures and limitations for interconnecting instruments. Cable length restrictions, also described in Figure
Each instrument in an HP-IB network must have a unique address, ranging in value from 00-30(d
synthesizer is 19, but this can be changed using the My adrrs or rear panel switch as described in the reference chapter (Chapter
2) under the “8360 as the address for the synthesizer). Other instruments use a variety of procedures for setting the address, as described in their operating manuals, but typically either a rear panel switch or a front panel code is used.
An HP-IB instrument is categorized as a “listener,” “talker,” or
“controller,” depending on its current function in the network.
Listener
A listener is a device capable of receiving data or commands from other instruments. Any number of instruments in the HP-IB network can be listeners simultaneously.
C-2
shows the synthesizer rear-panel HP-IB connector and
C-2, must be observed to prevent transmission.
ecimal).
Adrs” entry (the examples in this section use 19
The default address for the
softkey
Programming the
Synthesizer
Talker
A talker is a device capable of transmitting data or commands to other instruments. To avoid confusion, an HP-IB system allows only one device at a time to be an active talker.
Controller
A controller is an instrument, typically a computer, capable of managing the various HP-IB activities. Only one device at a time can be an active controller.
The synthesizer can be controlled entirely by a computer (although the line POWER switch must be operated manually). Several functions are possible only by computer (remote) control. Computer programming procedures for the synthesizer involve selecting an HP-IB command statement, then adding the specific synthesizer (SCPI, Analyzer, or CIIL) programming codes to that statement to achieve the desired operating conditions. The programming codes can be categorized into two groups: Those that mimic front panel keystrokes; and those that are unique, and have no
front panel equivalent.
1-56 Getting Started Programming
Page 87
In the programming explanations that follow, specific examples are included that are written in a generic dialect of the BASIC language.
BASIC was selected because the majority of HP-IB computers have BASIC language capability. However, other languages can also be used.
HP-IB Command
Statements
Command statements form the nucleus of HP-IB programming; they are understood by all instruments in the network and, when combined with the programming language codes, they provide all management and data communication instructions for the system.
An explanation of the fundamental command statements follows. However, some computers use a slightly different terminology, or support an extended or enhanced version of these commands. Consider the following explanations as a starting point, but
for detailed information consult the BASIC language reference
manual, the I/O programming guide, and the HP-IB manual for the particular computer used.
Syntax drawings accompany each statement: All items enclosed by
a circle or oval are computer specific terms that must be entered exactly as described; items enclosed in a rectangular box are names of parameters used in the statement; and the arrows indicate a path that generates a valid combination of statement elements.
The eight fundamental command statements are as follows:
Abort
Abort abruptly terminates all listener/talker activity on the interface
bus, and prepares all instruments to receive a new command from the
controller. Typically, this is an initialization command used to place
the bus in a known starting condition. The syntax is:
interface
erlect
Cod.2
where the interface select code is the computer’s HP-IB I/O port, which is typically port 7. Some BASIC examples:
10 ABORT7 100
IF V>20 THEN ABORT 7
Related statements used by some computers:
ABORT10
(used by HP-80 series computers) HALT RESET
Getting Started Programming
1-57
Page 88
Remote
Remote causes an instrument to change from local control to remote control. In remote control, the front panel keys are disabled (except for the
(LOCAL]
key and the POWER switch), and the amber
REMOTE annunciator is lighted. The syntax is:
where the device selector is the address of the instrument appended to the HP-IB port number. Typically, the HP-IB port number is
7, and the default address for the synthesizer is 19, so the device
selector is 719. Some BASIC examples:
REMOTE 7
10
which prepares all HP-IB instruments for remote operation (although nothing appears to happen to the instruments until they are addressed to talk), or
10
REMOTE 719
which affects the HP-IB instrument located at address 19, or
10
REMOTE719,
721, 726, 715 which effects four instruments that have addresses 19, 21, 26, and 15. Related statements used by some computers: RESUME
Local Lockout
Local Lockout can be used in conjunction with REMOTE to disable the front panel [LOCAL) key. With the
(LOCAL]
key disabled, only the
controller (or a hard reset by the POWER switch) can restore local
control. The syntax is:
interface
select
code
w
A BASIC example:
10
20
1-56 Getting Started Programming
REMOTE 719 LOCAL LOCKOUT 7
Page 89
Local
Local is the complement to REMOTE, causing an instrument to return to local control with a fully enabled front panel. The syntax is:
Some BASIC examples:
10
LOCAL 7
which effects all instruments in the network, or
10 LOCAL 719 for an addressed instrument (address 19). Related statements used by some computers: RESUME
Clear
Clear causes all HP-IB instruments, or addressed instruments, to assume a “cleared” condition, with the definition of “cleared” being unique for each device. For the synthesizer:
1. All pending output-parameter operations are halted.
2. The parser (the software that interprets the programming codes)
is reset, and now expects to receive the first character of a
programming code.
The syntax is:
Some BASIC examples:
10 CLEAR7 to clear all HP-IB instruments, or
CLEAR 719
10
Getting Started Programming
l-59
Page 90
to clear an addressed instrument.
Related statements used by some computers: RESET CONTROL SEND The preceding statements are primarily management commands
that do not incorporate programming codes. The following two statements do incorporate programming codes, and are used for data communication.
output
Output is used to send function commands and data commands from the controller to the addressed instrument. The syntax is:
l-60
Getting Started Programming
where USING is a secondary command that formats the output in a particular way, such as a binary or ASCII representation of numbers.
The USING command is followed by “image items” that precisely define the format of the output; these image items can be a string of code characters, or a reference to a statement line in the computer program. Image items are explained in the programming codes where they are needed. Notice that this syntax is virtually identical to the
syntax for the ENTER statement that follows. A BASIC example:
100 OUTPUT 719; “programming codes” The many programming codes for the synthesizer are listed in the
“SCPI Command Summary” in Chapter 2.
Related statements used by some computers:
CONTROL
Page 91
CONVERT IMAGE IOBUFFER TRANSFER
Enter
Enter is the complement of OUTPUT, and is used to transfer data from the addressed instrument to the controller. The syntax is:
Note
ENTER is always used in conjunction with OUTPUT, such as:
100
OUTPUT 719; ” . . . programming codes . . .
110
ENTER 719; ‘I . . . response data.. .
ENTER statements are commonly formatted, which requires the secondary command USING and the appropriate image items. The most-used image items involve end-of-line (EOI) suppression, binary
inputs, and literal inputs. For example:
100
ENTER719USING "#,B";
suppresses the EOI sequence and C are to be filled with binary (B) data. As another example,
100
ENTER719
suppresses EOI, and indicates that string variable A$ is to be filled
with 123 bytes of literal data (123A).
Be careful when using byte-counting image specifiers. If the
requested number of bytes does not match the actual number available, data might be lost, or the program might enter an endless wait state.
USING
"#,
A, B, C
and indicates that variables A, B,
(#),
123A"; A$
‘I
‘I
The suppression of the EOI sequence is frequently necessary to prevent a premature termination of the data input. When not specified, the typical EOI termination occurs when an ASCII LF
Getting Started Programming 1-61
Page 92
(line feed) is received. However, the LF bit pattern could
coincidentally occur randomly in a long string of binary data, where it might cause a false termination. Also, the bit patterns for the ASCII CR (carriage return), comma, or semicolon might cause a false termination. Suppression of the EOI causes the computer to accept
all bit patterns as data, not commands, and relies on the HP-IB EOI
(end or identify) line for correct end-of-data termination.
Related statements used by some computers:
CONVERT IMAGE IOBUFFER
ON TIMEOUT
SET TIMEOUT
TRANSFER
This completes the HP-IB Command Statements subsection. The following material explains the SCPI programming codes, and shows how they are used with the OUTPUT and ENTER HP-IB command statements.
1-62 Getting Started Programming
Page 93
Getting Started with
SCPI
This section of Chapter 1 describes the use of the Standard Commands for Programmable Instruments language (SCPI). This section explains how to use SCPI commands in general. The
instrument command summary (at the end of this chapter) lists
the specific commands available in your instrument. This section presents only the basics of SCPI. If you want to explore the topic in greater depth, see the paragraph titled, “Related Documents.”

Definitions of Terms

This section defines most terms when they are first used, you need a general understanding of the terms listed below before you continue.
controller
instrument
program
message
response
message
A controller is any computer used to communicate with a SCPI instrument. A controller can be a personal computer, a minicomputer, or a plug-in card in a card cage. Some intelligent instruments can also function as controllers.
An instrument is any device that implements SCPI. Most instruments are electronic measurement or stimulus devices, but this is not a requirement. Similarly, most instruments use an HP-IB interface for communication. The same concepts apply regardless of the instrument function or the type of interface used.
A program message is a combination of one or more properly formatted SCPI commands. Program messages always go from a controller to an instrument. Program messages tell the instrument how to make measurements and output signals.
A response message is a collection of data in specific SCPI formats. Response messages always go from an instrument to a controller or listening instrument. Response messages tell the controller about the
internal state of the instrument and about measured values.
command
query
A command is an instruction in SCPI. You
combine commands to form messages that control
instruments. In general, a command consists of mnemonics (keywords), parameters, and punctuation.
A query is a special type of command. Queries instruct the instrument to make response data available to the controller. Query mnemonics always end with a question mark.
Getting Started Programming 1-63
Page 94

Standard Notation

This section uses several forms of notation that have specific meaning.
Command Mnemonics
Many commands have both a long and a short form, and you must use either one or the other (SCPI does not accept a combination of the two). Consider the FREQuency command, for example. The short form is FREQ and the long form is FREQUENCY (this notation style is a shorthand to document both the long and short form of commands). SCPI is not case sensitive, so as FREQUENCY, but FREQ and FREQUENCY are the only valid forms of the FREQuency command.
Angle Brackets
Angle brackets indicate that the word or words enclosed represent something other than themselves. For example, <new line>
represents the ASCII character with the decimal value 10. Similarly,
C^END>means
in angle brackets have much more rigidly defined meaning than words used in ordinary text. For example, this section uses the word
“message” to talk about messages generally. But the bracketed words
<program
If you need them, you can find the exact definitions of words such as
<program message> in a syntax diagram.
that EOI is asserted on the HP-IB interface. Words
message> indicate a precisely defined element of SCPI.
fREquEnCy
is just as valid

How to Use Examples

It is important to understand that programming with SCPI actually requires knowledge of two languages. You must know the programming language of your controller (BASIC, C, Pascal) as well as the language of your instrument (SCPI). The semantic requirements of your controller’s language determine how the SCPI commands and responses are handled in your application.
Command Examples
Command examples look like this:
:FREQuency:CW?
This example tells you to put the string :FREQuency :CW? in the output statement appropriate to your application programming language. If you encounter problems, study the details of how the output statement handles message terminators such as If you are using simple OUTPUT statements in HP BASIC, this is taken care of for you. In HP BASIC, you type:
OUTPUT Source; ” : FREQuency : CW?”
Command examples do not show message terminators because
they are used at the end of every program message. “Details of
<new
line>.
1-64 Getting Started Programming
Page 95
Commands and Responses,”
discusses message terminators in more
detail.
Response Examples
Response examples look like this:
1.23
These are the characters you would read from an instrument after sending a query command. To actually pull them from the instrument into the controller, use the input statement appropriate to your application programming language. If you have problems, study the details of how the input statement operates. In particular, investigate how the input statement handles punctuation characters such as comma and semicolon, and how it handles
<new
line> and
EOI. To enter the previous response in HP BASIC, you type:
ENTER
Source;CW-frequency
Response examples do not show response message terminators because they are always <new line> C-END>. These terminators are typically automatically handled by the input statement. The paragraph titled “Details of Commands and Responses” discusses message terminators in more detail.
Getting Started Programming 1-65
Page 96

Essentials for Beginners

This subsection discusses elementary concepts critical to first-time users of SCPI. Read and understand this subsection before going on to another. This subsection includes the following topics:
Program and Response
Messages

Program and Response Messages

Subsystem Command Trees
Subsystem Command Tables
Reading Instrument Errors
Example Programs
To understand how your instrument and controller communicate
using SCPI, you must understand the concepts of program and response messages. Program messages are the formatted data sent from the controller to the instrument. Conversely, response messages are the formatted data sent from the instrument to the controller.
Program messages contain one or more commands, and response messages contain one or more responses.
These paragraphs introduce the basic types of messages sent between
instruments and controllers.
These paragraphs describe the
tree structure used in subsystem
commands.
These paragraphs present the
condensed tabular format used for
documenting subsystem commands. These paragraphs explain how to read
and print an instrument’s internal error messages.
These paragraphs contain two simple measurement programs that illustrate basic SCPI programming principles.
The controller may send commands at any time, but the instrument
sends responses only when specifically instructed to do so. The
special type of command used to instruct the instrument to send
a response message is the query. All query mnemonics end with a question mark. Queries return either measured values or internal instrument settings. Any internal setting that can be programmed with SCPI can also be queried.
Forgiving Listening and Precise Talking
SCPI uses the concept of forgiving listening and precise talking outlined in IEEE 488.2. Forgiving listening means that instruments are very flexible in accepting various command and parameter formats. For example, the synthesizer accepts either : POWer : STATe
ON or :POWer:STATe 1 to turn RF output on. Precise that the response format for a particular query is always the same. For example, if you query the power state when it is on (using : POWer : STATe?), the response is always 1, regardless of whether you previously sent : POWer : STATe 1 or : POWer : STATe ON.
1-66 Getting Started Programming
tallcing
means
Page 97
Page 98
root
level 1
c
BB
cc
DD
+I
level 2 EE FF GG
Figure
In the command tree shown in Figure the top is the root command, or simply the root. Notice that you must follow a particular path to reach lower level subcommands. For example, if you wish to access the GG command, you must follow the path AA to BB to GG.
Paths Through the Command Tree
To access commands in different paths in the command tree, you must understand how an instrument interprets commands. A special part of the instrument firmware, a purser, decodes each message sent to the instrument. The parser breaks up the message into component commands using a set of rules to determine the command tree path used. The parser keeps track of the current path, the level in the command tree where it expects to find the next command you send. This is important because the same keyword may appear in different paths. The particular path you use determines how the keyword is interpreted. The following rules are used by the parser:
8 Power On and Reset
l-25.
A Simplified Command Tree
l-25,
the command closest to
rtl
HH
JJ
l-68
Getting Started Programming
After power is cycled or after root.
n Message Terminators
A message terminator, such as a <new line> character, sets the current path to the root. Many programming languages have output statements that send message terminators automatically. The paragraph titled, discusses message terminators in more detail.
8
Colon
When it is between two command mnemonics, a colon moves the current path down one level in the command tree. For example, the colon in MEAS:VOLT specifies that VOLT is one level below When the colon is the first character of a command, it specifies that the next command mnemonic is a root level command. For example, the colon in command.
“Details of Commands and Responses,”
:
*RST,
INIT
specifies that
the current path is set to the
INIT
is a root level
MEAS.
Page 99
n Semicolon
A semicolon separates two commands in the same message without changing the current path.
w
Whitespace
White space characters, such as <tab> and <space>, are generally ignored. There are two important exceptions. White space inside a keyword, such as
:FREq uency,
is not allowed. You must use white
space to separate parameters from commands. For example, the
<space> between LEVel and 6.2 in the command
:POWer
: LEVel
6.2 is mandatory. White space does not affect the current path.
w
Commas
If a command requires more than one parameter, you must
separate adjacent parameters using a comma. Commas do not affect the current path.
w
Common Commands
Common commands, such as
*RST,
are not part of any subsystem. An instrument interprets them in the same way, regardless of the current path setting.
Figure
l-26
shows examples of how to use the colon and semicolon to
navigate efficiently through the command tree.
Getting Started Programming 1-69
Page 100
-
BB
cc
DD
l-h
EE
R
D
QQ
:AA:BB:EE;FF;GG
FF
4) vmv
:AA:BB:EE; :AA:DD:JJ
Figure
l-26.
In Figure typing.
GG
Proper Use of the Colon and Semicolon
l-26,
notice how proper use of the semicolon can save
rtl
HH
0 0 0
R Sets current path
to ROOT
N NO change to
current path
D Set current path
DOWN one level
JJ
l-70
Getting Started Programming
Sending this message:
:AA:BB:EE; FF; GG
Is the same as sending these three messages:
:AA:BB:EE
:AA:BB:FF
:AA:BB:GG
Loading...