• Repairs must only be carried out by personnel authorised to do so by the
manufacturer.
Interventions and modifications at centrifuges, which have been conducted
by persons not authorized by the HETTICH company, are at their own risk
and entail the loss off all guarantee and liability claims. In such an event any
guarantee claim or liability claim against the HETTICH company expire.
• Only original spare parts and original accessories licensed by the Hettich company
are allowed to be utilised.
If no original spare parts or no original accessories are used, any guarantee
claim or liability claim against the HETTICH company ceases to exist.
• Information about the operation of the centrifuge please see operating instructions.
• We reserve all rights for these technical documents.
2 Description of the centrifuge
These microprocessor controlled centrifuges mainly consist of the following electrical
components:
• Control panel, microprocessor controlled
− N-control panel, (without control knob)
− S-control panel, (with control knob)
• Supply board, microprocessor controlled
− for centrifuges with cooling
− for centrifuges without cooling
• Frequency converter (motor control), microprocessor controlled
• Motor with speed sensor (speedometer) and imbalance switch
• Braking chopper with brake resistor
• Lid lock
2.1 Control panel (A4, A5)
The control panel is the brain or the master of the centrifuge. This MASTER controls its
SLAVES via a serial data bus system. These SLAVES are the following components:
− supply board
− frequency converter
The control panel carries out the following tasks:
• Managing operator entries and controlling the LCD display
• Saving run programs:
− N-control panel: 3 programs
− S- control panel: 89 programs
• Controlling the components:
− supply board via the serial interface
− frequency converter via the release line and the serial interface
• Evaluating the speed sensor (speedometer)
• Evaluating the imbalance switch
6/92
• Evaluating the frequency converter error message line
• Evaluating the message line lid lock open/closed
• Triggering the relay for the voltage supply of the lid lock when the rotor has standstill
• Temperature measurement and evaluation of the temperature sensor at the top of
the centrifuge chamber
• Routine for input, saving and transmitting temperature offset values (offset values
are transmitted to the supply board each time the mains supply is switched on)
• Type of serial interface:
− N-control panel:
5 Volt interface with three wires (16-pole flat ribbon cable, pole 6, 8 and 11)
− S-control panel:
RS 485 interface via two wires (26-pole flat ribbon cable, pole 19 and 20)
• The voltage supply for the control panel is provided by the supply board via the flat
ribbon cable:
− N-control panel: + 15...20 Volt: pole 1, 2 and GND: pole 15, 16
− S-control panel: + 15...20 Volt: pole 1, 2, 25, 26 and GND: pole 15, 16, 17, 18
2.2 Supply board (A1)
The supply board carries out the following tasks:
• Current supply 18 V DC and 5 V DC for the supply board
• Current supply 18 V DC for the control panel
• Temperature measurement and evaluation of the temperature sensor at the bottom
of the centrifuge chamber
• Evaluating the overtemperature switch in the centrifuge chamber
• Temperature measurement and evaluation of the temperature sensor on the
condenser
• Evaluating the overpressure switch (not in ROTANTA 460 series)
• Evaluating the message line lid lock open/closed
• Evaluating the overtemperature switch on the brake resistor (when the overtempera-
ture switch opens, the frequency converter will be switched off)
• Voltage supply for the speed sensor (speedometer)
• Slot for speed sensor cable and transmission of the speed pulses to the control
panel
• Evaluating the speedometer pulses with regard to standstill
• Triggering the relay for the voltage supply of the lid lock when the rotor has standstill
• Slot for imbalance switch and direct transmission of the imbalance signal to the
control panel
• Processing the individual sensor temperatures:
− calculating the ACTUAL temperature
− controlling the cooling units
− controlling the heating
− controlling the solenoid valve
− controlling the condenser ventilator
7/92
• SLAVE behaviour (the requests and commands of the control panel are transmitted
via the serial interface):
− N-control panel:
5 Volt interface with three wires
− S-control panel:
RS 485 interface with two wires
• The 5 Volt interface with three wires is transferred to a RS 485 interface with two
wires.
Interface for the frequency converter: RS 485 interface via two wires
• Transmitting the main enable signal control panel ⇒ frequency converter
• Transmitting the error line frequency converter ⇒ control panel
• Evaluating the switching statuses of the key-operated switch (e.g. LOCK 1) and
transmitting the information to the control panel via the serial interface (only with
centrifuge with key-operated switch)
• Evaluating the jumper assignment and transmitting the information to the control
panel via the serial interface
− Machine version
− Cooling version
− Service mode
If the processor of the supply board detects an error, it switches off automatically. The
control panel continuously checks the status of the supply board via the serial interface.
The control panel defines the SET temperature and OFFSET temperatures via the
serial interface. The control panel inquires the ACTUAL temperature.
2.3 Frequency converter (A2)
The frequency converter carries out the following tasks:
• Generating the motor current supply
(three-phase current with variable frequency and voltage)
Functional description: The supply voltage is rectified, smoothened and
chopped into a pulse width pattern in three bridge
elements with a microprocessor.
• Monitoring the motor current
• Evaluating the overtemperature switch in the motor
• SLAVE behaviour (the requests and commands of the control panel are transmitted
via the serial interface):
RS 485 Volt interface with two wires
(5-pole flat ribbon cable, pole 2 and 3)
• Evaluating the main enable signal for the frequency converter
(5-pole flat ribbon cable, pole 1)
• Evaluating possible errors and handling the error line
(5-pole flat ribbon cable, pole 4)
• The following are defined by the control panel via the serial interface:
− Speed
− Starting and brake levels
− Control commands START, BRAKE, STOP
8/92
• LED status displays:
In the standby status the red LED illuminates
In the run status the green LED illuminates
In the error status the red LED flashes
If the frequency converter processor detects an error, it switches off automatically and
sets the error line. Then the control panel inquires the type of error via the serial
interface.
2.4 Special features
• The voltage supply of the motor-driven lock is switched via a relay on the supply
board.
This relay can pick up only if the control panel and the supply board have detected a
standstill.
• The N-control panel and the S-control panel are interchangeable.
• Multiprocessor concept:
If a processor stops working, the other two processors still continue monitoring their
area.
If there is a control panel failure, the frequency converter automatically switches off
the drive if no inquiries are made via the interface for longer than 60 seconds.
• Interface concept:
The information transmission is monitored with an additional check sum.
• Hardware concept:
All safety related switches are break contacts. This also ensures that loose contacts
and cable rupture can be detected.
2.5 Braking chopper (A3) / brake resistor (R1)
• The braking chopper transfers the electrical energy produced during braking, from a
voltage of approx. 390 Volt, to the brake resistor in a controlled manner.
• An overtemperature switch protects the brake resistor against fire. If the braking
chopper has a short circuit, the brake resistor becomes hot due to the high current,
the overtemperature switch opens the circuit and two relays on the supply board
separate the frequency converter from the mains supply.
2.6 Motor / tacho system
• The motor is a three-phase asynchronous motor with four pairs of poles.
• A speed sensor (speedometer) which is screwed onto the motor receives
− the rotor code information and
− the speed information (6 pulses per revolution)
from the magnets of the tacho ring attached to the rotor.
• The control panel monitors and regulates the ACTUAL speed.
− Double safety:
In addition, the frequency converter has been programmed in
such a way that it is not possible to set a speed value which is
higher than the permissible rotor speed.
• The control panel and the supply board monitor the rotor standstill.
− Double safety:
The lid can be opened only when the control panel and
the
supply board have detected a standstill.
9/92
2.7 Motor-driven lid lock
• The lid can be opened only if the motor of the lid lock is triggered. To do so, press
the key when the rotor is not moving and mains voltage exists. The jumper for the
lid lock must be plugged to the position "Mot" on the supply board.
• The motor locks the lid only if both lid brackets have actuated both switches in the lid
lock.
• You can start the centrifuge only if the lid is closed. A microswitch on the lid lock
detects the lid lock position.
2.8 Imbalance switch
• A switch (break contact) detects imbalance.
• Imbalance is detected only in running mode (starting, centrifuging and braking).
• If imbalance is detected, the drive is brought to a standstill.
2.9 Key-operated switch (A7, optionally)
• With the key-operated switch different program locks can be select, see chapter
"Option: key-operated switch (only for S-control panel)".
The key position will be displayed.
• If the key is removed, the selected program lock is locked and can be changed no
more.
2.10 Machine concept
• The machine version is set with jumper assignment A.4, A.5 and B.7 on the coding
strip of the supply board, see chapter "Setting the machine and cooling version".
• This machine version must comply with the control panel:
− N-control panel: correct EPROM
− S-control panel: correct initialization
• This machine version must comply with the installed and initialized frequency
converter.
• The machine version has a binary code:
Jumper /
Value
Version Machine A.4 A.5 B.7
1 2 16
19 ROTANTA 460 230 V / 115 V X X X
10/92
2.11 Cooling system
• The cooling version is set with jumper assignment A.0 to A.2 on the coding strip of
the supply board, see chapter "Setting the machine and cooling version".
• This cooling version must comply with the control panel:
− N-control panel: correct EPROM
− S-control panel: correct initialization
• The cooling version has a binary code:
Jumper /
Value
Version Cooling System A.0 A.1 A.2
1 2 4
0 No cooling system
1 1 compressor X
2 1 compressor, heating/cooling X
7 No cooling system, laboratory thermostat X X X
• Hardware control:
− If the lid is opened, the cooling or heating system switches off.
− If the overtemperature switch is opened, the heating system switches off.
• Software control:
− If the condenser temperature is higher than 58°C, the cooling system switches off.
− The processor on the supply board controls the cooling system.
− The control panel sets the nominal temperature and the temperature of the
temperature sensor T1 on the supply board via the serial interface.
− The control panel inquires the ACTUAL temperature via the serial interface.
− Temperature sensor:
T1 Temperature sensor at the top of the centrifuge chamber
This temperature is processed in the control panel.
T2 Temperature sensor at the bottom of the centrifuge chamber
This temperature is processed on the supply board.
The T2 temperature sensor housing also includes an overtemperature switch.
It switches off the drive in cooling centrifuges when the temperature is > 60°C and
in heating/cooling centrifuges if it is > 80°C.
Temperature determination:
During centrifugation the temperature is determined on the basis of the
relationship between T1 and T2. This occurs in different ways depending on the
speed range (see chapter "Diagram of the cooling version, sensor evaluation").
During standstill, the temperature is determined only with the T2 temperature
sensor.
The switching temperatures depend on the cooling version, see chapter "Diagram
of the cooling version“
T3 Temperature sensor on the condenser
This temperature is used for controlling the ventilator speed.
11/92
∆
g
2.11.1 Diagram of the cooling version
-4
centrifugationbrakin
set value -2,5°K
If the set temperature is >= 20° C, the cooling system
standstill
improvement of the displayed temperature
(can b e deac tivated by setting a j umper o n position B.3)
-3
OFF at
set value -1°K
set value -0,5°K
ON at
heating
switched off
remains switched off during standstill!
heating
switched off
heating
-1+1+2-2
OFF at
Set value
cooling
cooling
OFF at
set value -0,5°K
OFF at
set value +1°K
ON at
set value +0,5°K
ON at
set value +0,5°K
cooling
+3
ON at
set value +2°K
OFF at
set value +4°K
+4
+5
5/5 to 4/5 of n-max
4/5 to n-min
5/5 to n-min
5/5 to n-min
only sensor T2only sensor T2
coolingcooling
ON a)
at set value +5°K
+6+7
at set value +7°K
ON b)
Τ
sensor evaluation
temperature sensor temperature sensor
speed range 5/5 to 4/5 of n-max rotor
speed range 4/5 to 3/5 of n-max rotor
speed range 3/5 to 2/5 of n-max rotor
speed range 2/5 to 1/5 of n-max rotor
speed range 1/5 to n-min
T1 : temperature sensor at the top of the centrifuge chamber
T2 : temperature sensor at the bottom of the centrifuge chamber
T1 : T2
1 : 1
2 : 1
3 : 1
4 : 1
5 : 1
Explanation on:
The cooling will only be switched on 15 s after start
ON a) = cycle time 60 s ON / 6 min OFF
ON b) = cycle time 90 s ON / 6 min OFF
12/92
2.12 Ventilator control
• The ventilator control is a temperature-dependent phase control.
• If the cooling system switches on, the ventilator runs with a fixed start speed.
• When the condenser temperature rises, the ventilator runs faster.
• At 32 °C, or at the latest after 60 seconds, the ventilator speed regulation is
activated and the ventilator speed is regulated to achieve a constant temperature of
34°C at the condenser.
• If the cooling system is switched off and the condenser temperature falls, the
ventilator speed is reduced until minimum speed is achieved.
• If the temperature is lower than 28 °C, the phase control is not triggered, the
ventilator comes to a standstill.
Centrifuge type Start speed Maximum speed Minimum speed
ROTANTA 460 R / RS
230 V / 50 Hz
ROTANTA 460 R / RS
230 V / 60 Hz
• The ventilator runs in a temperature-controlled manner also if the lid is open.
• If there is condenser overtemperature (overpressure in the system) i.e. temperature
sensor T3 indicates a temperature > 58 °C (3.31 V), the cooling system and the
drive are switched off, but the condenser ventilator continues running.
2.13 Ventilator
approx. 2000 RPM approx. 2700 RPM approx. 900 RPM
approx. 1900 RPM approx. 2950 RPM approx. 820 RPM
• The ventilator cools down the refrigerant which flows through the condenser.
• The ventilator speed is controlled with the temperature sensor T3.
• This speed control keeps the pressure in the cooling system constant as far as
possible.
2.14 Offset compensation
• The offset compensation is carried out to balance the tolerances of the temperature
sensors and the electronic system.
After replacing: Offset compensation:
− a temperature sensor
− the supply board
− the control panel
− the control panel EPROM
Compensation of the concerned temperature
sensor.
Compensation of the temperature sensors T2
and T3.
Compensation of the temperature sensor T1
and read out the old offset values of T2 and
T3 and enter them in the new control panel.
Compensation is not required
13/92
2.15 Safety devices
Mains input
Mains switch
Radio interference
suppression filter
Frequency converter
Motor
Cooling system
PE connection
3 Troubleshooting procedures
• Fuses in installation in which centrifuge is installed are intact.
• Supply voltage present at (see circuit diagram):
− Connecting cable
− Appliance plug
− Overvoltage protection (F4)
− Mains switch (Q1)
− Supply board, plugs X21 and X29.
• Look for the displayed error code in the chapter "Error messages".
• Remedy the error according to the instructions.
• Carry out a functional check after every repair and whenever a component is
replaced, see chapter "Functional check after a repair".
overvoltage protection
⇒
with thermal overload protection
⇒
between supply board and frequency converter
⇒
electronically protected
⇒
overtemperature switch > 135 °C
⇒
overtemperature switch in the centrifuge chamber,
⇒
software monitoring of the temperature sensor T3 on the
condenser
additional earthing of the device (only with 115 V model)
⇒
14/92
4 Error messages
4.1 MAINS RESET
• Switch off the mains switch.
• Wait for 10 seconds and then switch on the mains switch again.
4.2 Brief description
Error designation No. Brief description Page
TACHO-ERROR 01 Speedometer pulses break down during rotation 18
TACHO-ERROR 02 No speedometer pulses after start command 18
IMBALANCE (03) Imbalance on motor axle 19
CONTROL-ERROR 04 Lid lock error, lid opened without recognizing that
motor had stopped
N > MAX 05 Excessive speed error, 250 RPM above n-max of
rotor
CONTROL-ERROR 06 Lid lock - error
Message:
from control panel → lid lock open
from supply board → lid lock closed
CONTROL-ERROR 07 Lid lock - error
Message:
from control panel → lid lock closed
from supply board → lid lock open
CONTROL-ERROR 08 Standstill error
Message:
from control panel → standstill
from supply board → no standstill
CONTROL-ERROR 09 Standstill error
Message:
from control panel → no standstill
from supply board → standstill
CONTROL-ERROR 99 S-control panel - error: hardware failure in error
management
INDEFINITE S-control panel - error: not specified error 36
BATT Battery is empty 36
35
17/92
4.3 Description and elimination of errors
TACHO - ERROR 01
Error During centrifugation the speedometer pulses are interrupted.
Consequences
of the error
Cause of error/
measurements
Delete error
code
TACHO - ERROR 02
Error After starting the centrifugation the control panel receives no
The rotor slows down until it stops.
After the rotor stops, there is a DC braking for 30 sec.
A MAINS RESET during slowing-down causes a DC braking for 3
minutes.
After the DC braking, the “open the lid” release takes place.
Further cooling until the SET temperature is achieved.
• Speed sensor (speedometer) is defective or loose contact in the
plug. Measure the speedometer pulses on the supply board (A1) at
plug X5, pin 4 - pin 2.
• Flat ribbon cable to the control panel (A4, A5) or flat ribbon cable to
the frequency converter (A2) is defective.
• Supply board (A1) or control panel (A4, A5) or frequency converter
(A2) is defective.
Open the lid. Turn the rotor by hand and carry out a MAINS RESET
while the rotor is rotating.
− In the S-control panel, CONTROL - ERROR 08 could appear in the
display after TACHO - ERROR 01 is deleted.
speedometer pulses.
Consequences
of the error
Cause of error/
measurements
Delete error
code
The rotor slows down until it stops.
After the rotor stops, there is a DC braking for 30 sec.
A MAINS RESET during slowing-down causes a DC braking for 3 min.
After the DC braking, the “open the lid” release takes place.
Further cooling until the SET temperature is achieved.
• Start implemented without rotor
• Motor is not connected
• Motor is defective
• Speed sensor (speedometer) is defective or loose contact in the
plug. Measure the speedometer pulses on the supply board (A1) at
plug X5, pin 4 - pin 2.
• Flat ribbon cable to the control panel (A4, A5) or flat ribbon cable to
the frequency converter (A2) is defective.
• No enable signal at the frequency converter (A2) (after a start, the
green LED on the frequency converter must illuminate), also refer to
FU / CCI – ERROR 60.
• Supply board (A1) or control panel (A4, A5) or frequency converter
(A2) is defective.
Open the lid. Turn the rotor by hand and carry out a MAINS RESET
while the rotor is rotating.
18/92
IMBALANCE (03)
Error Imbalance on the motor axle.
Consequences
of the error
Cause of error
Delete error
code
CONTROL - ERROR 04
Error Lid lock was opened during centrifugation.
Consequences
of the error
Cause of error
The centrifuge slows down until the “open the lid” release occurs.
Further cooling until the SET temperature is achieved.
• Weight difference in the rotor components
• Supporting lugs are not lubricated
• Wrong IMBALANCE MODE has been set (see chapter "Imbalance
Mode")
• Imbalance switch (S2) is not connected
• Imbalance switch (S2) is defective
• Loose contact in the cable or in the plug
• Flat ribbon cable to the control panel (A4, A5) is defective
• Control panel (A4, A5) or supply board (A1) is defective
Open the lid or carry out a MAINS RESET.
The centrifuge slows down until the “open the lid” release occurs.
Further cooling until the SET temperature is achieved.
• Lid lock (Y1) is defective and can be opened during centrifugation.
• Loose contact in the cable or in the plug
• Flat ribbon cable to the control panel (A4, A5) is defective
• Control panel (A4, A5) or supply board (A1) is defective
Delete error
code
N > MAX 05
Error Overspeed. The speed detected by the speed sensor is 250 RPM
Consequences
of the error
Cause of error
Delete error
code
Open the lid and close it again or carry out a MAINS RESET.
In the S-control panel, CONTROL – ERROR 06 or CONTROL – ERROR 08 could appear in the display after CONTROL – ERROR 04
is deleted.
higher than the maximum rotor speed (n-max rotor).
The centrifuge slows down until the “open the lid” release occurs.
Further cooling until the SET temperature is achieved.
• The insulation of the speed sensor cable is defective
• Loose contact in the speed sensor cable
• Speed sensor is defective
• Flat ribbon cable to the control panel (A4, A5) is defective
• Control panel (A4, A5) or frequency converter (A2) or supply board
(A1) is defective
Carry out a MAINS RESET.
19/92
CONTROL - ERROR 06
Error Lid lock message is faulty.
Control panel indicates: lid lock open
Supply board indicates: lid lock closed
Consequences
of the error
Cause of error
The centrifuge slows down until the “open the lid” release occurs.
Further cooling until the SET temperature is achieved.
• Flat ribbon cable to the control panel (A4, A5) is defective
• Control panel (A4, A5) or supply board (A1) is defective
Delete error
Carry out a MAINS RESET.
code
CONTROL - ERROR 07
Error Lid lock message is faulty.
Control panel indicates: lid lock closed
Supply board indicates: lid lock open
Consequences
of the error
Cause of error
The centrifuge slows down until the “open the lid” release occurs.
Cooling system is switched off.
• Flat ribbon cable to the control panel (A4, A5) is defective
• Supply board (A1) or control panel (A4, A5) is defective
Delete error
Carry out a MAINS RESET.
code
CONTROL - ERROR 08
Error Standstill message is faulty.
Control panel indicates: standstill
Supply board indicates: no standstill
Consequences
of the error
Cause of error
Braked slow-down.
Further cooling until the SET temperature is achieved.
• Flat ribbon cable to the control panel (A4, A5) is defective
• Control panel (A4, A5) or supply board (A1) is defective
Delete error
Carry out a MAINS RESET.
code
CONTROL - ERROR 09
Error Standstill message is faulty.
Control panel indicates: no standstill
Supply board indicates: standstill
Consequences
of the error
Cause of error
Braked slow-down.
Further cooling until the SET temperature is achieved.
• Flat ribbon cable to the control panel (A4, A5) is defective
• Control panel (A4, A5) or supply board (A1) is defective
Delete error
code
20/92
Carry out a MAINS RESET.
ROTORCODE 10
Error An invalid rotor code was read in during start-up.
Consequences
of the error
Cause of error
The centrifuge slows down until the “open the lid” release occurs.
Further cooling until the SET temperature is achieved.
• Magnetic code on the rotor is defective
• Speedometer system is defective
• Loose contact in speed sensor plug
• Motor is rotating in the wrong direction
Delete error
Open the lid or carry out a MAINS RESET.
code
MAINS INTERRUPT (11)
Error Mains interruption during centrifugation
Consequences
of the error
The centrifuge slows down until the “open the lid” release occurs.
− Switching on the mains during centrifugation causes slowing-down
until the “open the lid” release occurs.
− Switching on the mains when the rotor has stopped brings about
the “open the lid” release
Cause of error
• Power failure
• Loose contact in the electrical connections
• Flat ribbon cable to the control panel (A4, A5) is defective
Delete error
code
Open the lid and press the
− The error cannot be deleted with a MAINS RESET.
START
key
VERSION - ERROR 12
Error Differences in the initialization of supply board (jumper assignment
A.0 – A.7, B.7), control panel (EPROM) or frequency converter.
− In the S-control panel, VERSION – ERROR 12 also appears after
CONTROL – ERROR 92 is displayed
Consequences
Further operation is not possible.
of the error
Cause of error
• Incorrect jumper assignment (A.0 – A.7 and B.7) on the coding strip
of the supply board
Plug the jumper A.0 – A.7 and B.7 as described in chapter "Setting
the machine and cooling version".
• Carry out a MAINS RESET.
• Plug the service jumper at B.1 on the supply board, carry out a
MAINS RESET and re-initialize the machine
• A wrong EPROM is plugged in the control panel
• Supply board (A1) is defective
Delete error
Carry out a MAINS RESET.
code
21/92
N < MIN 13
Error Insufficient speed, motor slippage is too large.
The centrifuge control can readjust the speed by max. 5 % (control
limit).
This error is displayed if the rotor speed (ACTUAL speed) is lower
than the SET speed - 5%.
Consequences
of the error
Cause of error
Delete error
code
CONTROL - ERROR 21 - 25
Error Internal error in the N-control panel
Consequences
of the error
Cause of error
Delete error
code
CONTROL - ERROR 26
The centrifuge slows down until the “open the lid” release occurs.
Further cooling until the SET temperature is achieved.
• Motor is not running smoothly (bearing damage)
• Motor is defective
• Motor has a short-circuited coil (coil is defective)
• Loose contact in the electrical connections
• Frequency converter (A2) is defective
• Enable signal to the frequency converter (A2) was interrupted
during centrifugation (also see FU / CCI - ERROR 60).
Open the lid and carry out a MAINS RESET.
The centrifuge slows down until the “open the lid” release occurs.
• N-control panel (A4) is defective.
Carry out a MAINS RESET.
Error Incorrect feedback, control panel input does not recognize the
expected signal.
Consequences
of the error
Cause of error
Delete error
code
22/92
The centrifuge stops with error message.
• Control panel (A4, A5) is defective.
• Supply board (A1) is defective
• Flat ribbon cable between supply board (A1) and control panel (A4,
A5) is defective.
Carry out a MAINS RESET.
CONTROL - ERROR 27
Error Error in the EPROM of the N-control panel
Consequences
The centrifuge slows down until the “open the lid” release occurs.
of the error
Cause of error
Delete error
• EPROM is defective
Carry out a MAINS RESET.
code
N > ROTOR-MAX
Error Error in the entered program
Consequences
Further operation is not possible.
of the error
Cause of error SET speed or SET RCF is higher than the permissible rotor speed or
permissible rotor RCF.
Delete error
code
Open the lid.
Reduce the speed or RCF in the entered program to the permissible
rotor speed or permissible rotor RCF.
SER I/O - ERROR 30
Error The control panel has no connection to the components supply board
and frequency converter via serial interface.
Consequences
The centrifuge slows down until the “open the lid” release occurs.
of the error
Cause of error
• Flat ribbon cable to the control panel (A4, A5) is defective
• Supply board (A1) or control panel (A4, A5) is defective
Delete error
Carry out a MAINS RESET.
code
SER I/O - ERROR 31
Error The control panel has no connection to the component frequency
converter via the serial interface.
Consequences
of the error
Cause of error
The centrifuge slows down until the “open the lid” release occurs.
Further cooling until the SET temperature is achieved.
• Flat ribbon cable to the frequency converter (A2) is defective
• No voltage at the frequency converter (A2)
• Overtemperature switch (F2) on the brake resistor has opened or is
not connected
• Supply board (A1) or control panel (A4, A5) or frequency converter
(A2) is defective.
• Plug X5 on the frequency converter (A2) is not plugged or plugged
at the wrong position
Delete error
code
Carry out a MAINS RESET.
23/92
SER I/O - ERROR 32
Error The control panel has no connection to the component supply board via
the serial interface.
Consequences
The centrifuge slows down until the “open the lid” release occurs.
of the error
Cause of error
• The false machine or cooling version is set on the supply board (A1).
• Supply board (A1) or control panel (A4, A5) is defective.
Delete error
Carry out a MAINS RESET.
code
SER I/O - ERROR 33
Error The control panel does not receive correct data from the connected
components (supply board, frequency converter).
Consequences
The centrifuge slows down until the “open the lid” release occurs.
of the error
Cause of error
• Flat ribbon cable to the control panel (A4, A5) is defective
• Supply board (A1) or control panel (A4, A5) or frequency converter
(A2) is defective.
Delete error
Carry out a MAINS RESET.
code
SER I/O - ERROR 34
Error The control panel does not receive correct data from the frequency
converter.
Consequences
of the error
Cause of error
The centrifuge slows down until the “open the lid” release occurs.
Further cooling until the SET temperature is achieved.
• Flat ribbon cable to the control panel (A4, A5) is defective
• Flat ribbon cable to the frequency converter (A2) is defective
• Supply board (A1) or control panel (A4, A5) or frequency converter
(A2) is defective.
Delete error
Carry out a MAINS RESET.
code
SER I/O - ERROR 35
Error The control panel does not receive correct data from the supply board.
Consequences
The centrifuge slows down until the “open the lid” release occurs.
of the error
Cause of error
• Flat ribbon cable to the control panel (A4, A5) is defective
• Supply board (A1) or control panel (A4, A5) is defective
Delete error
Carry out a MAINS RESET.
code
24/92
SER I/O - ERROR 36
Error The frequency converter cannot acknowledge a command and sends
the NAK signal to the control panel.
NAK (no acknowledge).
Consequences
of the error
Cause of error
Delete error
code
SER I/O - ERROR 37
Error The supply board cannot acknowledge a command and sends the
Consequences
of the error
Cause of error
Delete error
code
°C / * - ERROR 50
The centrifuge slows down until the “open the lid” release occurs.
Further cooling until the SET temperature is achieved.
• Flat ribbon cable to the control panel (A4, A5) is defective
• Flat ribbon cable to the frequency converter (A2) is defective
• Frequency converter (A2) is defective
• Supply board (A1) or control panel (A4, A5) is defective
Carry out a MAINS RESET.
NAK signal to the control panel. NAK (no acknowledge)
The centrifuge slows down until the “open the lid” release occurs.
• Flat ribbon cable to the control panel (A4, A5) is defective
• Supply board (A1) is defective
• Control panel (A4, A5) or frequency converter (A2) is defective
Carry out a MAINS RESET.
Error Overpressure in the cooling system (only in centrifuges with an
overpressure switch on the plug X11)
No overpressure switch has been integrated in the ROTANTA 460
series. On the supply board, pin 2 and pin 3 on plug X11 are
connected with each other.
Consequences
of the error
Cause of error Plug X11 on the supply board (A1) is not plugged or has a loose
Delete error
code
The centrifuge slows down until the “open the lid” release occurs.
Cooling system is switched off.
contact.
Carry out a MAINS RESET.
25/92
°C / * - ERROR 51
Error Overtemperature in the condenser, temperature > 58 °C.
Consequences
of the error
Cause of error/
measurements
Delete error
code
°C / * - ERROR 52
Error Overtemperature in the centrifuge chamber
Consequences
of the error
Cause of error/
measurements
The centrifuge slows down until the “open the lid” release occurs.
Cooling system is switched off.
• Measurement on the supply board (A1) at plug X1, pin 1 - pin 2
(plug X1 plugged)
Voltage at 25 °C : 2.98 V (± 20 mV)
at 58 °C : 3.31 V (± 20 mV)
• Condenser is dirty
• Ventilator is defective
• Cable of the temperature sensor on the capacitor (T3) is defective
• Loose contact in the plug
• Supply board (A1) is defective
Carry out a MAINS RESET.
The centrifuge slows down until the “open the lid” release occurs.
Cooling system is switched off.
• Measurement on the supply board (A1) at plug X2, pin 1 - pin 2
(plug X2 plugged)
Voltage : 0 V (overtemperature switch closed)
Voltage : 18 V (overtemperature switch open)
• Cable of the overtemperature switch is defective
• Loose contact in the plug
• Supply board (A1) is defective
Delete error
code
26/92
Carry out a MAINS RESET.
°C / * - ERROR 53
Error The temperature sensor at the bottom of the centrifuge chamber has a
short circuit or has opened the circuit.
Voltage < 0.50 V = short circuit
Voltage > 4.50 V = circuit has been opened
Consequences
of the error
Cause of error/
measurements
Delete error
code
°C / * - ERROR 54
Error The temperature sensor at the top of the centrifuge chamber has a
Consequences
of the error
The centrifuge slows down until the “open the lid” release occurs.
Cooling system is switched off.
• Measurement on the supply board (A1) at plug X2, pin 4 - pin 5
(plug X2 plugged)
Voltage at 25 °C : 2.98 V (± 20 mV)
A temperature change of 1 °K causes a voltage change of 10 mV.
• Temperature sensor at the bottom of the centrifuge chamber (T2) is
defective
• Sensor cable is defective
• Loose contact in the plug
• Supply board (A1) is defective
Carry out a MAINS RESET.
short circuit or has opened the circuit.
Permissible range: 2.50 V = -23 °C to 3.77 V = +104 °C
The centrifuge slows down until the “open the lid” release occurs.
Cooling system is switched off.
Cause of error/
measurements
Delete error
code
• Measurement on the control panel (A4, A5) at plug X100 or X2, pin
1 - pin 2 (plug X100 or X2 plugged)
Voltage at 25 °C : 2.98 V (± 20 mV)
A temperature change of 1 °K causes a voltage change of 10 mV.
• Temperature sensor at the top of the centrifuge chamber (T1) is
defective
• Sensor cable is defective
• Loose contact in the plug
• Control panel (A4, A5) is defective
Carry out a MAINS RESET.
27/92
°C / * - ERROR 55
Error The temperature sensor on the condenser has a short circuit or has
opened the circuit.
Voltage < 0.50 V = short circuit
Voltage > 4.50 V = circuit has been opened
Consequences
of the error
Cause of error/
measurements
Delete error
code
LOCK - ERROR 57
Error The analogous value in the program locking plug X27 indicates that
Consequences
of the error
The centrifuge slows down until the “open the lid” release occurs.
Cooling system is switched off.
• Measurement on the supply board (A1) at plug X1, pin 1 - pin 2
(plug X1 plugged)
Voltage at 25 °C : 2.98 V (± 20 mV)
A temperature change of 1 °K causes a voltage change of 10 mV.
• Temperature sensor on the capacitor (T3) is defective
• Sensor cable is defective
• Loose contact in the plug
• Supply board (A1) is defective
Carry out a MAINS RESET.
the voltage is outside the voltage range.
Permissible voltage range: 0.9 V - 3.8 V.
Further operation is not possible.
Cause of error/
measurements
Delete error
code
• Measurement on the supply board at plug X27, pin 1 - pin 2 (plug
X27 plugged)
Key-operated switch position 1 (LOCK1) ≈ 1.4 V
2 (LOCK2) ≈ 2.1 V
3 ≈ 3.3 V
• Key-operated switch board is defective
• Cable to the key-operated switch board is defective
• Loose contact in the wire layout
• Supply board (A1) is defective
Carry out a MAINS RESET.
28/92
Loading...
+ 64 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.