• To prevent injury, do not transport the scale while the platform is loaded.
• For accurate weighing the scale must be placed on a flat, stable surface.
• For accurate weighing, verify proper operation according to the procedure described in this
manual before each use.
• Do not use in the presence of flammable materials.
• Operating at other voltages and frequencies than specified could damage the equipment.
• If the “LO BAT” indicator activates, for accurate weighing, replace the batteries or connect
the scale to an AC power source as soon as possible.
SPECIFICATIONS
GENERAL
Health o meter’s High Capacity Stand-On Scale Model 1101KL uses highly sophisticated
microprocessor technology. Each precision instrument is designed to provide accurate,
reliable and repeatable weight measurements and features that make the weighing process
simple, fast and convenient.
The actual weighing is performed by motion-sensing weighing technology, to determine the
weight of an active human.
The weight can be display in pounds or in kilograms. The scale includes two wheels and a
pillar, used to easily mobilize the scale from place to place.
The unit can be operated using its AC power adapter or by 6-C cell batteries.
SCALE SPECIFICATIONS
Capacity and Resolution
Power Requirements
Environmental
Physical Dimensions Platform
700 Lb x 0.2 Lb (310 Kg x 0.1 Kg)
Adapter 120V AC - 60Hz 6 x C size batteries
Operating temperatures: 50°F to 95°F (10°C to 35°C)
Storage temperatures: 30°F to 125°F (0°C to 50°C)
Humidity: 85%
Parts List:
(1) Platform Assembly
(1) Handrail Assembly
(4) Rubber Feet
(1) AC Adapter Holder
This 1101KL scale has been calibrated at the factory. Remove each assembly from its carton
and unwrap packing material carefully to prevent scratching the unit’s parts.
Figure 1. Assembly After Shipment
1. Place the platform assembly upside down on a level floor.
2. Screw the four feet onto the platform assembly.
3. Prepare the platform assembly for installing to the handrail assembly as follows (use
7/16” or 11mm wrench):
- Unscrew 2 screws from the both sides of the platform assembly.
- Unscrew 4 nuts from the both sides and from the back of the platform assembly.
- Keep these screws and nuts for reinstalling in the same locations.
4. Hold the handrail assembly perpendicular to the platform assembly, as shown in
Figure 1.
5. Attach the handrail assembly to the platform assembly, install 2 screws at both sides of
the platform and then 4 nuts at both sides and at the back of the platform.
6. Plug the handrail cable connector into the platform assembly.
7. Install the AC adapter holder on the interior right side of the handrail using two screws
attached to the holder assembly.
8. Adjust the rubber feet for equal stabilization of the scale.
The scale may be operated by an AC power source adapter or by 6-C cell batteries.
The scale will automatically switch to battery operation when an AC power source is absent.
3
SET UP
1. Plug the scale’s AC adapter into the handrail assembly and into the power source or
place batteries in the battery holder (see “Replacing Batteries” below).
2. Press the ON/OFF button to turn the scale on. The display will show “start” and
then ”0.0”.
3. When the self-test function is complete:
• The digital display should read
• “LB” should be indicated on the right side of the display. (To change the default
units of the scale to kilograms (Kg), please contact our Technical Support Center
at 1.800.638.3722 or 1.708.598.9100).
4. Place a weight [not to exceed 700 Lb (310 Kg)] on the scale. The digital display should
blink “
5. Press the LBS/KG button to select the weight mode, either in kilograms or pounds.
6. Press the REWEIGH button; the digital display should read again the weight of the item
on the scale.
7. Remove the weight from the scale. The digital display should continue to read the
weight of the item that was on the scale.
8. Press the ZERO button; the digital display should return to zero.
9. Disconnect the power source connector from the scale, then turn the scale on using
ON/OFF button. The scale should automatically switch to battery power.
NOTE
not corrected, refer to qualified service personnel.
------
“ until the weight calculation is complete.
------
: If the set up procedure failed, go to the troubleshooting instructions. If the problem is
”0.0”.
Replacing Batteries
1. Disconnect the scale from the AC power source.
2. Loosen two plastic knobs found underneath the display until the battery holder will be
released from the handrail assembly and will rest on the knobs. Carefully pull out the
cable to provide access to the connector. Disconnect battery holder cable connector
from the handrail cable connector by pressing the plastic button.
3. Raise the battery holder and simultaneously rotate it counterclockwise 1/4 turn.
Carefully remove the holder from the handrail assembly.
4. Replace the batteries with new ones.
5. Place the battery holder on the plastic knobs by rotating it 1/4 turn clockwise.
6. Plug the battery holder cable connector to the handrail cable connector.
7. Hold the battery holder close to the handrail assembly and tighten the two plastic knobs
to secure in place. Be careful not to damage the cable.
Figure 2. Replacing Batteries
4
SET UP
Figure 3. Control Panel
ITEM DESCRIPTION FUNCTION
1 ON/OFF button Turns scale ON and OFF.
2 REWEIGH button Allows repeated reweighing of person.
3 ZERO button Zeros Digital Display prior to weighing. Enables the
cancellation (tare) of weight if some object was placed
on the scale.
4 LBS/KG button Selects kilograms or pounds.
5 BAT indicator When ON, indicates scale batteries are low. The
batteries should be replaced.
6 ZERO indicator When ON, indicates the scale is stable and ready for
use.
7 Digital Display 6 Digit LCD Displays human weight in kilograms or pounds as
shown by the “LB” or “KG” indicator on the right side of
the display. When REWEIGH button is depressed,
display will blink “
determined.
8 LB indicator When ON, indicates displaying weight in pounds.
9 KG indicator When ON, indicates displaying weight in kilograms.
------
------
until patient’s weight is
”
5
OPERATION INSTRUCTIONS
1. Press the ON/OFF button to turn the scale on.
2. Wait until ”0.0” appears on the digital display.
3. Ask the patient to step on the scale. The display will blink “
person is determined.
4. If you wish to reweigh without asking the patient to step off and to step on the scale
again, press the REWEIGH button.
5. Ask the patient to step off the scale. The digital display should continue to read the
person’s weight until REWEIGH or ZERO button is pressed.
NOTE:
After 2 minutes of inactivity, the scale will turn itself off to conserve power. To restart
the scale press the ON/OFF button.
------
until the weight of the
------
”
MAINTENANCE
GENERAL
This section provides instructions for maintenance, cleaning, troubleshooting and operator
replaceable parts for the High Capacity Stand-On Scale Model 1101KL. Maintenance
operations other than those described in this section should be performed by qualified service
personnel.
MAINTENANCE
Before first use and after periods of non-use, check the scale for proper operation and
function. If the scale does not operate correctly, refer to qualified service personnel.
1. Check overall appearance of the total scale for any obvious damage, wear and tear.
2. Inspect AC adapter for cord cracking or fraying or for broken or bent prongs.
CLEANING
Proper care and cleaning is essential to ensure a long life of accurate and effective operation.
Disconnect the scale from the AC power source.
1. Clean all external surfaces with a clean damp cloth or tissue. Mild soap and water
solution may be used. Dry with a clean soft cloth.
2. Do not immerse the scale platter or pillar assembly into cleaning or other liquid solution.
3. Do not use Isopropyl Alcohol or other solutions to clean the display surface.
6
TROUBLESHOOTING
Refer to the following instructions to check and correct any failure before contacting service
personnel.
SYMPTOM POSSIBLE CAUSE CORRECTIVE ACTION
Scale does not turn on 1. Dead Battery
2. Faulty electrical outlet
3. Bad power supply
Questionable weight or
the scale does not zero
2. The display did not show
3. Scale is not placed on a
4. Scale is out of calibration 4. Check weight with
The display shows “OL” The load on the scale
The display shows “UL” The scale is in under-load
The display shows “LO BAT” The battery is low Replace batteries according
1. External object
interfering with the scale
”0.0” before weighing
level floor
exceeds the capacity
condition
1. Replace batteries
2. Use a different outlet
3. Replace adapter
1. Remove interfering
object from the scale
2. Ask the patient to get off
the scale, zero the scale
and begin weighing
process again
3. Place the scale on a
level floor and begin
weighing process again
known weight value
Remove the excess weight
and use the scale according
to its limits
Make sure the weighing tray
is placed on the base
to instructions
7
CALIBRATION PATH
The modes are accessible by holding certain keys pressed after power-up.
The keys must be pressed and held for at least three seconds.