Haws 7501BL User Manual

No. 0510000497(2)

Model 7501BL/7501BL240

Heated Blanket for 7501 Portable Eyewash
INSTALLATION, OPERATION
&
MAINTENANCE INSTRUCTIONS

LIMITED WARRANTY

GIVEN BY HAWS.
NOTE TO INSTALLER: Please leave this information with the Maintenance Department.
SHOULD YOU EXPERIENCE DIFFICULTY WITH THE INSTALLATION OF THIS
1455 Kleppe Lane Sparks, NV 89431-6467 (775) 359-4712 Fax (775) 359-7424
E-mail: haws@hawsco.com website: www.hawsco.com
HAWS warrants that this specific product is guaranteed against defective material or poor workmanship for a period of one year from date of shipment. HAWS liability under this warranty shall be discharged by furnishing without charge F.O.B. HAWS Factory any goods, or part thereof, which shall appear to the Company upon inspection to be of defective material or not of first class workmanship, provided that claim is made in writing to Haws within a reasonable period after receipt of the product. Where claims for defects are made, the defective part or parts shall be delivered to the Company, prepaid, for inspection. HAWS will not be liable for the cost of repairs, alterations or replacements, or for any expense connected therewith made by the owner or his agents, except upon written authority from HAWS, Sparks, Nevada. HAWS will not be liable for any damages caused by defective materials or poor workmanship, except for replacements, as provided above. Buyer agrees that Haws has made no other warranties either expressed or implied in addition to those above stated, except that of title with respect to any of the products or equipment sold hereunder and that HAWS shall not be liable for general, special, or consequential damages claimed to arise under the contract of sale.
The emergency equipment manufactured by HAWS is warranted to function if installation and maintenance instructions provided are adhered to. The units also must be used for the purpose for which they were intended. This product is intended to supplement first-aid treatment. Due to widely varying conditions, Haws cannot guarantee that the use of this emergency equipment will prevent serious injury or the aggravation of existing or prior injuries
NO OTHER WARRANTIES EXPRESSED OR IMPLIED ARE AUTHORIZED, PROVIDED OR
MODEL PLEASE CALL:
TECHNICAL SUPPORT: 1-800-766-5612
.
FOR CUSTOMER SERVICE: 1-888-640-4297
09/16 7501BL/7501BL240 Page 1 of 5
TROUBLESHOOTING

PROBLEM

REPAIR CHECKLIST

breaker.
RECOMMENDED TOOLS: Assorted tools as needed for installing power cord (7501BL240 only).
LOCATION OF UNIT: The Model 7501BL/7501BL240 heated blanket should be installed over a 7501 tank in close proximity to potential accident areas. It should be clearly identified, free from obstruction, and easily accessible. This assembly should be wall mounted in accordance with the 7501 installation instructions. NOTE: Personal wash unit Models 7501BL/
7501BL240 shall not be exposed to ambient temperatures exceeding 38C (100F). ELECTRICAL SUPPLY: The Model 7501BL requires a 120 VAC, three-prong outlet. The
7501BL includes a heavy-duty 8 foot power cord with GFCI plug. The Model 7501BL240 requires a 240VAC outlet and is not supplied with a power cord. The Model 7501BL240 MUST BE PROTECTED BY a Class A G ROUND F AULT CIRCUIT INTERRUPTER (GFCI). No power cord is supplied with the 7501BL240 and must be installed by the user.
1. Heat trace does not turn on. 1. a. Not plugged in or no power to plug. Check GFCI plug (7501BL only) and circuit
b. If ambient temperature is high and doesn’t
allow the blanket temperature to drop below 68°F (20°C) the heater may not switch on. Heater settings are: Non-adjustable thermostat preset to switch on at 62°± 5°F (17°± 3°C) and switch off 77°± 5°F (25°± 3°C). Test the thermostat and heater function by applying ice in a circular motion to thermostat location. Apply ice to this location for approximately 3-5 minutes. The heat trace cable should become warm to the touch when operating properly. Thermostat is located on junction box.
09/16 7501BL/7501BL240 Page 2 of 5